Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Size: px
Start display at page:

Download "Reference Dosimetry for Megavoltage Therapy Beams: Electrons"

Transcription

1 Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX

2 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams. Med. Phys. 26, p (1999) TG-51 Based on an Absorbed Dose to Water (in Water) Standard ( 60 Co): Conversion to Absorbed dose at other energies and modalities (electrons) is by correction factors, k ecal, k R50, and P gr (electrons). Based on Bragg Gray cavity theory. All of the chamber correction factors introduced in TG-21 are incorporated into K ecal, and k R50. Many chamber specific corrections are averaged into classes of chambers, so only a few k R50 values are needed.

3 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams. Med. Phys. 26, p (1999) TG-51 Requires Calibration in Water ( at least annually): Plastics may be used for monthly output checks, but must be referenced to the water calibration by a correction factor. Electrons Energy characterized by depth of 50% of max dose, d 50 Reference measurement depth, d ref, is related to d 50.

4 Ideal: Absorbed Dose Calibration Factor D = M N Q D,w Q N D,w Where: is the chamber Absorbed Dose Calibration Factor specific for the energy and modality of the beam being measured. (N K for Kilovolt X-rays) Pragmatic: Too expensive to be practical We understand the behavior of chambers in the megavoltage range. Cobalt 60 is still available, and very reproducible. Cobalt 60 is a near perfect beam to use as the reference energy for chamber calibration. D = M 60 N Co D,w other energy/modality specific corrections.

5 Chamber Calibration Factor Obtain from ADCL Chamber waterproofing material: 1 mm Acrylic (PMMA) wall Provided by ADCL Waterproof chambers

6 Calibrate P-P P Chamber ADCL s: TG-39, in water 60 Co Evidence of problems with k ecal User: (Recommended) TG-39 in water/plastic Cylindrical Chamber is NIST traceable TG-51 In water high energy electron R 50 near 7.5 cm compare with cylindrical chamber

7 Electron Equation k Q D Q w = M P Q gr k ' R50 k ecal N 60 Co D,w [Gy] P Q gr k ' R 50 = k R 50 (Eq 6) P Q gr = Gradient Correction k ' R 50 = non - gradient component of energy correction factor k ecal is conversion of N D,w from 60 Co to electrons

8 For Electrons k = P Q Q gr k R 50 (Eq 4) Q P gr = k = k ' R R50 50 ecal correction for gradient at the point of calibration (d ref ) k (Eq 5) P Q gr = ( ) ref + 0.5rref M ( d ) M d raw ref (Eq 21)

9 TG-51 Measurements Electrons: Look up k ecal for your chamber. Search for Imax and I50 (use 0.5 r cav shift) Determine R 50 Determine d ref and k R50 Move chamber to physical d ref (no shift) Measure P pol and P ion Move chamber to d ref r cav Calculate the gradient correction, P gr

10 k Electrons ecal k ecal dummy factor carries into Q N D,w N Co D, w Q D,w ( ecal = dose calib factor at reference electron energy d 50 = 7.5cm) Simplifies the calculations Allows for a specific electron calibration factor to be introduced later. Includes L/ρ and P repl for R 50 = 7.5 cm and L/ρ, P repl, P wall for Cobalt 60. N ecal

11

12 TG-51 Measurements Electrons: Look up k ecal for your chamber. Search for Imax and I50 (use 0.5 r cav shift) Determine R 50 Determine d ref and k R50 Move chamber to physical d ref (no shift) Measure P pol and P ion Move chamber to d ref r cav Calculate the gradient correction, P gr

13 Electrons: Reference Conditions Field Size: assure full lateral scatter for R cm x 10 cm for R 50 < 8.5 cm 20 cm x 20 cm for R 50 > 8.5 cm Reference Depth, d ref d ref = 0.6R [cm] (Eq 18) Nominal SSD

14 Electrons Clinical % dd Measure % ionization Shift to effective point of measurement Convert ionization to dose -- using TG-25 (revised L/ρ) Use %dd to shift from d ref to d max, not %dion Parallel Plate and diode need no shift

15 Effective point of measurement Electron source x r cav Center of chamber Effective pt of measurement d eff = d meas -f (r cav ) f = 0.5 r cav for electrons f = 0 for parallel plate inner surface of front window

16 Effective Point of Measurement Water surface Effective depth Physical depth X X Cylindrical Parallel plate

17

18 TG-51 Measurements Electrons: Look up k ecal for your chamber. Search for Imax and I50 (use 0.5 r cav shift) Determine R 50 Determine d ref and k R50 Move chamber to physical d ref (no shift) Measure P pol and P ion Move chamber to d ref r cav Calculate the gradient correction, P gr

19 Beam Quality Specification for Electron Beams Specified by R 50 R 50 = depth (cm) at which dose falls to 50% of max for a 10 cm x 10 cm field. (20cm x 20cm for R 50 > 8.5cm) R 50 R 50 = I (cm) [I 50 < 10cm] = I (cm) [I 50 >10cm] (Eq 16) (Eq 17) {I 50 = depth of 50% ionization}

20 TG-51 Measurements Electrons: Look up k ecal for your chamber. Search for Imax and I50 (use 0.5 r cav shift) Determine R 50 Determine d ref and k R50 Move chamber to physical d ref (no shift) Measure P pol and P ion Move chamber to d ref r cav Calculate the gradient correction, P gr

21 Reference depth: dref d ref = 0.6R [cm]

22 Energy Dependent Factor - chamber specific (Includes ratio of L/ρ P rep P cel for arbitrary electron energy to that for R 50 = 7.5 cm) ' well behaved (observed ) k R50 ' k R50 ' k R50 (cyl) = e -( R50/3.67) (Eq 19) (cyl cham for 2cm < R 50 < 9cm), 0.2% error for Farmer chambers. Fig 5 & 7 (cyl) Fig 6 & 8 (pp)

23 For Electrons For Electrons For Electrons 60 Cobalt evaluated at cel wall repl water air 7.5 energy R evaluated at cel repl water air ecal P P P L P P L k 50 ρ ρ = = energy R at evaluated cel repl water air 50 energy R the at evaluated cel repl water air 50 R P P L P P L k' = ρ ρ =

24

25 TG-51 Measurements Electrons: Look up k ecal for your chamber. Search for Imax and I50 (use 0.5 r cav shift) Determine R 50 Determine d ref and k R50 Move chamber to physical d ref (no shift) Measure P pol and P ion Move chamber to d ref r cav Calculate the gradient correction, P gr

26 Phantoms Water only Annual calibration Plastics: Weekly/Monthly Compare with water at annual calibration.

27 Chamber Protection Waterproof Chambers no protective sleeve needed Other Chambers 1 mm thick Acrylic protective sleeve

28 Chambers: parallel plate vs cylindrical Electrons: Parallel Plate Chambers: RECOMMENDED Cylindrical Chambers: ACCEPTABLE P-P Chamber required for R cm

29 Chamber Position Cylindrical Chamber (e - ) Clinical depth dose data: effective pt of measurement Beam quality specification: effective pt of measurement Calibration: Physical center of chamber gradient correction included in P gr

30 M = P ion P TP P elec P pol M raw [C or rdg] (Eq 8) P ion = Collection efficiency correction P TP = Temp-Press correction P elec = Electrometer factor P pol = Polarity Correction M raw = uncorrected charge reading

31 TG-51 Measurements Electrons: Look up k ecal for your chamber. Search for Imax and I50 (use 0.5 r cav shift) Determine R 50 Determine d ref and k R50 Move chamber to physical d ref (no shift) Measure P pol and P ion Move chamber to d ref r cav Calculate the gradient correction, P gr

32 P Pulsed ion Beam V H = normal operating potential of chamber V L = reduced potential on chamber - H M raw L M raw P ion ( V ) H = VH 1.00 VL H M raw V L M raw V = raw reading with full potential = raw reading with reduced potential H L V L < V H /2 (Eq 11)

33 1.005 Response (Norm to 1st full bias) d Monitor Δ NEL, NEL' X PTW, PTW' Cap-G, Cap-C 6 dm Sequence #

34 P ion using the equations in TG-51 For V H /V L = 2 continuous radiation pulsed radiation P io n M H /M L

35 Polarity Correction P pol M + = raw raw (Eq 9) 2M - M raw M + (M - ) is the charge collected with positive (negative) polarity on the collector M raw = charge collected with normal polarity

36 TG-51 Measurements Electrons: Look up k ecal for your chamber. Search for Imax and I50 (use 0.5 r cav shift) Determine R 50 Determine d ref and k R50 Move chamber to physical d ref (no shift) Measure P pol and P ion Move chamber to d ref r cav Calculate the gradient correction, P gr

37 For Electrons Q P gr = gradient correction P Q gr ( ref + 0.5rcav ) M ( d ) M d = (Eq 21) raw ref Q P gr = M (dref + 0.5rcav) M raw(dref)

38 Whole equation with %dd% D Q Q ' Co = Mraw PTP Pelec Pion Ppol Pgr kr50 k ecal N 60 w D,w (1/pdd) [Gy]

39 Implementation

40 Current Implementation Status Current Implementation Status TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) Only 8% to go Nov-99 May-00 Nov-00 May-01 Nov-01 May-02 Nov-02 May-03 Nov-03 May-04 Nov-04 May-05 Nov-05 May-06 Nov-06 May-07 Nov-07 May-08 Nov-08 May-09

41 What s s the Holdup? New Air Kerma standard TG-51 TG-21 depending on the chamber Time and effort required (everyone is very busy) New equipment requirements (chambers and phantoms)

42 Equipment Needs Properly sized liquid water phantom (30x30x30 cm 3 ) Don t use the scanning tank Adequate scatter conditions Easy reproducible setup

43 Chamber Holder and Positioner Holder Versatile to hold different chambers Rigid (sensitive volume perpendicular to water surface) No lateral displacement with depth Accurate sub-millimeter placement at any depth Verify accuracy prior to initial use Remote electronic control is nice

44 Ion Chambers TG-51 ion chambers vs NEW ion chambers Most are similar in design but now waterproof 1. Wall material 2. Radius of air cavity 3. Presence of Al electrode 4. Wall thickness AAPM working group to determine the k R50, k ecal for new chambers

45 Ion Chambers - Electrons Parallel-plate or cylindrical chambers okay Cylindrical for energies > 6 MeV per protocol (R cm) Cylindrical = parallel plate if care in placement Always use a parallel plate chamber for 4 MeV beams Caution as to where the inside surface of the front window is located

46 Ion Chambers - Electrons All chambers must have an ADCL calibration coefficient EXCEPT PARALLEL PLATE CHAMBERS AAPM recommendation is to cross calibrate parallel plate chamber with cylindrical chamber in a high energy electron beam (worksheet C a la TG-39) ADCL N D,w good TG-51 k ecal bad Use of (N D,w k ecal ) results in an error of 1-2% ONE EXCEPTION Exradin P11 seems to be okay AAPM working group determining new k ecal values

47 Measurement Techniques Accurate placement of cylindrical ion chamber at depth Whether manual or electronic motor driven there must be a starting reference point 1. Surface method Two techniques Air Correct Position Water

48 Measurement Techniques Cut ruler down 2. Cowboy method to minimize surface area U-shape Water plastic surface attached flush with end of ruler Accuracy depends on cutting ruler Used for reference starting point Periodic check of depth Ion chamber Cut ruler by the chamber radius and wall thickness weights

49 Measurement Techniques Parallel plate ion chambers 1. Flat surface makes it easy to measure depth 2. Accurate ruler needed 3. Must know where the inside surface of the front window is located

50 Beam Quality Conversion Factors Electrons k R 50 Only small figures, no tables Good figures at: apers/tg51_figures.pdf

51 Beam Quality Conversion Factors Electrons 4 MeV beams (R 50 < 2.0 cm) Only use parallel plate chamber Need to extrapolate curve Equation good down to 1 cm 1

52 Charge Measurements M = P ion P TP P elec P pol M raw P TP correction factor Mercury thermometers and barometers most accurate (but they are no longer kosher) Hg barometers T&G corrections needed Quality aneroid or digital can be used Check annually against a standard Digital purchased with a calibration does not mean accurate but rather what it read at certain pressures or temperatures

53 Charge Measurements P elec correction factor ADCL calibration for each scale needed P pol correction factor Change polarity requires irradiation (600 to 800 cgy) to re-equilibrate chamber Use of eq 9 in TG-51 requires that you preserve the sign of the reading or P pol = M + raw 2 raw raw P pol should be near unity for cylindrical chambers and slightly larger correction for parallel plate chambers + M M

54 Charge Measurements A-12' P ion correction Δ NEL'' factor Response norm to 1st rdg (-300V) for a beam High dose rate capabilities result in higher P ion Change in bias requires irradiation ( cgy) to reequilibrate chamber. PTW'' Monitor X PTW'' 6 d10 A-12 NEL'' 18 x 6 d10 Cl 2100CD (210 5) data (75 mu) dated 19-Feb-00 Monitor's drift due to Ktp & machine fluctuation (All other chamber data are norm to monitor) d5 Sequence # 6 dm 20 dm

55 Charge Measurements P ion correction factor Use eqs. 11 and 12 to calculate P ion As a check if using V H /V L = 2 (within 0.1%) Pulsed beam : P ion = M H /M L if M H /M L < 1.02 Continuous beam : P ion = {(M H /M L 1)/3}+1 P ion depends on chamber, beam energy, linac and beam modality Tends to increase with energy

56 Charge Measurements Electron beam gradient (P gr ) correction factor Only for cylindrical ion chambers Ratio of readings at two depths ( d 0.5r ) ( d ) The reading at d ref +0.5r cav should have the same precision as the reading at d ref since: Dose = M(d ref ) (many factors) M(d ref +0.5r cav ) P gr M ref + = M raw ref cav M(d ref )

57 Charge Measurements Electron beam gradient (P gr ) correction factor E < 12 MeV; P gr >1.000 E 12 MeV; P gr Why? Because for low electron energies d ref = d max and this places the eff. pt. of measurement in the buildup region thus a ratio of readings greater than At higher electron energies d ref is greater than d max and as such the eff. Pt. of measurement is on the descending portion of the depth dose curve thus a ratio of readings less than

58 Charge Measurements Physical depth Percent Dose x x 6 MeV Dose Depth (cm) Effective depth ( d 0.5r ) M ref + M raw ( d ) ref cav

59 Charge Measurements Physical depth 80 Percent Dose Effective depth Depth (cm) x d ref x d ref+0.5rcav ( d 0.5r ) M ref + M raw ( d ) ref cav

60 Clinical Depth Dose For electrons depth dose correction for 16 MeV is significant (~98.5% - 16 MeV and ~95.5% - 20 MeV) Caution!!! Super big problem if you use % depth ionization data (3-5% error for high energy electron beams)

61 Summary Implementation is straightforward 1. Must read the protocol and follow the prescriptive steps 2. Many suggestions to clarify confusion have been made 3. RPC will assist you and answer questions Differences between TG-51 and other protocols such as TG-21 and TRS 398 are minimal.

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin NCCAAPM meeting April 17, 2015 Larry DeWerd has partial interest in Standard Imaging Inc. Determination of your uncertainty

More information

Updating reference dosimetry a decade after TG-51

Updating reference dosimetry a decade after TG-51 Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual

More information

Dosimetry: Electron Beams

Dosimetry: Electron Beams ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann

More information

R. C. Tailor and W. F. Hanson. Department of Radiation Physics. The University of Texas, M.D. Anderson Cancer Center. Houston, Texas 77030

R. C. Tailor and W. F. Hanson. Department of Radiation Physics. The University of Texas, M.D. Anderson Cancer Center. Houston, Texas 77030 Page 1 of 231 Calculated absorbed-dose ratios, TG51/TG21, for most idely used cylindrical and parallel-plate ion chambers over a range of photon and electron energies R. C. Tailor and W. F. Hanson Department

More information

ABSTRACT. Keywords: Megavoltage, dosimetry, TG51 protocol, TG21 protocol, parallel-plate chambers, crosscomparison. INTRODUCTION

ABSTRACT. Keywords: Megavoltage, dosimetry, TG51 protocol, TG21 protocol, parallel-plate chambers, crosscomparison. INTRODUCTION CALCULATED TG51 TO TG21 ABSORBED-DOSE RATIOS FOR MOST WIDELY USED CYLINDRICAL AND PARALLEL-PLATE ION CHAMBERS OVER A RANGE OF PHOTON AND ELECTRON ENERGIES R.C. Tailor, and W.F. Hanson ABSTRACT The change

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:

More information

Chapter 9: Calibration of Photon and Electron Beams

Chapter 9: Calibration of Photon and Electron Beams Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation

More information

) for Varian TrueBeam high-dose-rate therapy beams

) for Varian TrueBeam high-dose-rate therapy beams JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Ion recombination correction factors (P ion ) for Varian TrueBeam high-dose-rate therapy beams Stephen F. Kry, 1a Richard Popple,

More information

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy M. Saiful Huq, PhD, FAAPM, FInstP Dept. of Radiation Oncology, University of

More information

Electron beam water calorimetry measurements to obtain beam quality conversion factors

Electron beam water calorimetry measurements to obtain beam quality conversion factors Electron beam water calorimetry measurements to obtain beam quality conversion factors Bryan R. Muir, a) Claudiu D. Cojocaru, Malcolm R. McEwen, and Carl K. Ross Measurement Science and Standards, National

More information

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA

More information

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Progress in calculations of k Q for TG-51

Progress in calculations of k Q for TG-51 1 Progress in calculations of k Q for TG-51 D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University Ottawa http://www.physics.carleton.ca/~drogers AAPM Charlotte,NC,

More information

Optimization of microionization chambers for small-field reference dosimetry

Optimization of microionization chambers for small-field reference dosimetry Optimization of microionization chambers for small-field reference dosimetry Jessica R. Snow University of Wisconsin-Madison Department of Medical Physics Medical Radiation Research Center NCCAAPM Fall

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams

NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams E. Chin 1, J. Seuntjens 1, H. Palmans 2, A. DuSautoy 2, D. Shipley 2, M. Bailey

More information

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams NEDERLANDSE COMMISSIE VOOR STRALINGSDOSIMETRIE Report 18 of the Netherlands Commission on Radiation Dosimetry

More information

Monitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics

Monitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics Monitor Unit Calculations for Photon and Electrons AAMD Meeting Raleigh, NC October 3, 2014 John P. Gibbons Chief of Clinical Physics Outline I. TG71 Formation and Charge II. Photon Calculations III. Electron

More information

Ionizing Radiation Dosimetry and Medical Physics

Ionizing Radiation Dosimetry and Medical Physics Ionizing Radiation Dosimetry and Medical Physics D.W.O. Rogers Ionizing Radiation Standards, Institute for National Measurement Standards NRC, Ottawa, K1A OR6 Present e-mail: drogers at physics.carleton.ca

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers Small Field Dosimetric Measurements with TLD-1, Alanine, and Ionization Chambers S. Junell a, L. DeWerd a, M. Saiful Huq b, J. Novotny Jr. b, M. uader b, M.F. Desrosiers c, G. Bednarz b a Department of

More information

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin AAPM July 2012 Larry DeWerd has a partial interest in Standard Imaging Traceability to NIST via calibration points

More information

Composite field dosimetry

Composite field dosimetry Composite field dosimetry Hugo Bouchard, PhD, MCCPM Senior Research Scientist Radiation dosimetry group National Physical Laboratory May 2014 Overview 1. Introduction Dosimetry protocols IAEA formalism

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation Signature: Pol J Radiol, 05; 80: 55-5 DOI: 0.659/PJR.89450 ORIGINAL ARTICLE Received: 05.03.7 Accepted: 05.06.9 Published: 05..5 Authors Contribution: A Study Design B Data Collection C Statistical Analysis

More information

ERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1

ERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1 ERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1 1. page 034 Table 3 (cont) Chambers SNC 100730 and 100740 are replaced, respectively, by model numbers SNC 100700-0 and 100700-1. Their radii are changed

More information

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 1 Medical Radiation Physics Group, University of Oldenburg and Pius

More information

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento

More information

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology

More information

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel

More information

A comparison of methods for monitoring photon beam energy constancy

A comparison of methods for monitoring photon beam energy constancy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon

More information

Use of a radioactive check device for redundancy check of ionization chambers

Use of a radioactive check device for redundancy check of ionization chambers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman

More information

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.264-271 Variation of Dose Distribution by Detectors for Narrow Beam T. Fujisaki, H.

More information

Chapter 10 Acceptance Tests and Commissioning Measurements

Chapter 10 Acceptance Tests and Commissioning Measurements Chapter 10 Acceptance Tests and Commissioning Measurements This set of 189 slides is based on Chapter 10 authored by J. L. Horton of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics:

More information

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,

More information

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS Malcolm McEwen Definitions: Standard instrument/measurement/artifact intended to define, realize, conserve

More information

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma 4.1b - Cavity Theory Lecture 2 Peter R Almond 2011 Overview of Lecture Exposure (W/e) air Exposure and Air Kerma Exposure Exposure is symbolized as X and defined by the ICRU as the quotient of dq by dm,

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator Determination of Ambient Dose quivalent at INFLPR 7 MeV Linear Accelerator F. Scarlat, A. Scarisoreanu, M. Oane,. Badita,. Mitru National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest-Magurele,

More information

Investigation of the standard temperature- pressure correction factor at low x-ray energies

Investigation of the standard temperature- pressure correction factor at low x-ray energies Investigation of the standard temperaturepressure correction factor at low x-ray energies D. J. La Russa, M. R. McEwen and D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

Outline. Indrin J. Chetty, AAPM 2006 Monte Carlo CE course. Indrin J. Chetty Henry Ford Hospital. David W. O. Rogers Carleton University

Outline. Indrin J. Chetty, AAPM 2006 Monte Carlo CE course. Indrin J. Chetty Henry Ford Hospital. David W. O. Rogers Carleton University AAPM Task Group Report No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning Indrin J. Chetty Henry Ford Hospital David W. O.

More information

Ionization Chamber. Pocket dosimeter

Ionization Chamber. Pocket dosimeter Ionization Chamber Pocket dosimeter 1 Ionization Chamber Pocket dosimeter 2 Radiation Quantities and Units Radiation measurements require specification of the radiation field at various points At the source

More information

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV

More information

The Radiological Physics Center s standard dataset for small field size output factors

The Radiological Physics Center s standard dataset for small field size output factors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 The Radiological Physics Center s standard dataset for small field size output factors David S. Followill, 1a Stephen F. Kry, 1 Lihong

More information

General characteristics of radiation dosimeters

General characteristics of radiation dosimeters General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Monte Carlo Simulation concerning Particle Therapy

Monte Carlo Simulation concerning Particle Therapy Monte Carlo Simulation concerning Particle Therapy Masaaki Takashina Graduate School of Medicine, Osaka University, Osaka 565-0871, Japan INTRODUCTION It is well known that the particle therapy has some

More information

NPL s progress towards absorbed dose standards for proton beams

NPL s progress towards absorbed dose standards for proton beams NPL s rogress towards absorbed dose standards for roton beams H. Palmans 1 R. Thomas 1 D. Shiley 1 A. Kacerek 2 1 National Physical Laboratory Teddington United Kingdom 2 Clatterbridge Centre of Oncology

More information

Quality-Assurance Check of Collimator and Phantom- Scatter Factors

Quality-Assurance Check of Collimator and Phantom- Scatter Factors Quality-Assurance Check of Collimator and Phantom- Scatter Factors Ramesh C. Tailor, David S. Followill, Nadia Hernandez, Timothy S. Zhu, and Geoffrey S. Ibbott. UT MD Anderson Cancer Center, Houston TX.

More information

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen Principles of applied dosimetry - illustrated by ionometry Lesson FYSKJM4710 Eirik Malinen Ionometry Ionometry: the art of measuring ionizations Number of ionizations proportional to dose Air filled ionization

More information

A Thesis. entitled. Micro Ionization Chamber. Vidheesha Arora. Master of Science Degree in. Biomedical Sciences: Medical Physics

A Thesis. entitled. Micro Ionization Chamber. Vidheesha Arora. Master of Science Degree in. Biomedical Sciences: Medical Physics A Thesis entitled An Investigation of the Polarity Effects in Small Field Based on the Orientation of the Micro Ionization Chamber by Vidheesha Arora Submitted to the Graduate Faculty as partial fulfillment

More information

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA GROUP I SPRING 2014 KEVIN JORDAN GRADUATE STUDENT, RADIOLOGICAL PHYSICS KARMANOS CANCER CENTER WAYNE STATE UNIVERSITY

More information

Equilibration of air temperature inside the thimble of a Farmer-type ion chamber

Equilibration of air temperature inside the thimble of a Farmer-type ion chamber Equilibration of air temperature inside the thimble of a Farmer-type ion chamber R. C. Tailor a) Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030

More information

Dosimetry. Introduction. General. CHAPTER 1

Dosimetry. Introduction. General. CHAPTER 1 CHAPTER 1 Dosimetry Introduction. a. The detection and repeated measurement of radioactive fallout fields produced by nuclear explosions will give important information affecting the operation of field

More information

Calibration of a pencil ionization chamber with and without preamplifier

Calibration of a pencil ionization chamber with and without preamplifier Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof.

More information

An introduction to IAEA TRS-483

An introduction to IAEA TRS-483 An introduction to IAEA TRS-483 P Andreo, Professor of Medical Radiation Physics Karolinska University Hospital and Karolinska Institute, Stockholm, Sweden Journées Scientifiques de la SFPM Toulouse 2018

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER*

COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER* Original Article COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER* Roseli T. Bulla 1, Linda V.E. Caldas 2 * Study developed at Instituto de Pesquisas Energéticas

More information

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors 6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator

More information

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT) 1- Ionizing radiation metrology for radiation protection 2- Metrological requirements for ionizing radiation measurement in radiotherapy and radiodiagnostics R. F. Laitano Part 2 Metrological traceability

More information

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation

More information

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration Pablo

More information

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range

More information

Air kerma rate measurements from a miniature x-ray source using free-air ionization chambers

Air kerma rate measurements from a miniature x-ray source using free-air ionization chambers Air kerma rate measurements from a miniature x-ray source using free- ionization chambers Stephen D. Davis 1, John A. Micka 1, Larry A. DeWerd 1, and Thomas W. Rusch 2 1 University of Wisconsin, Madison,

More information

IAEA TRS May 2001 (V.10A) PUBLISHED BY THE IAEA ON BEHALF OF IAEA, WHO, PAHO, AND ESTRO INTERNATIONAL ATOMIC ENERGY AGENCY IAEA

IAEA TRS May 2001 (V.10A) PUBLISHED BY THE IAEA ON BEHALF OF IAEA, WHO, PAHO, AND ESTRO INTERNATIONAL ATOMIC ENERGY AGENCY IAEA IAEA TRS-398 Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water Pedro Andreo, Dosimetry and Medical

More information

FOREWORD In 1987 the IAEA published a Code of Practice entitled Absorbed Dose Determination in Photon and Electron Beams: An International Code of

FOREWORD In 1987 the IAEA published a Code of Practice entitled Absorbed Dose Determination in Photon and Electron Beams: An International Code of IAEA-TECDOC-1173 Review of data and methods recommended in the international code of practice for dosimetry IAEA Technical Reports Series No. 381, The Use of Plane Parallel Ionization Chambers in High

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September

More information

Electron therapy Class 2: Review questions

Electron therapy Class 2: Review questions Electron therapy Class 2: Review questions 1 Raphex Question: T63, 2002 In what situation is electron backscatter likely to be a problem? A. Using 1cm of tissue equivalent bolus on the skin. B. Using a

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator

Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator Christoph Trauernicht Groote Schuur Hospital & University of Cape Town Method is based on: Background A method of

More information

Determination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source

Determination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source Polish Journal of Medical Physics and Engineering 2016;22(3):55-59 September 2016 doi: 10.1515/pjmpe-2016-0010 ISSN 1898-0309 Scientific Paper Determination of contributions of scatter and distance error

More information

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1) Rapport BIPM92/1 Determination of absorbed dose to water for 60Co by the scaling theorem M. Boutillon and A.M. Perroche (1) Bureau International des Poids et Mesures, F92312 Sevres Cedex Abstract A simple

More information

Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter

Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter Radiotherapy Standards Users Group meeting NPL, 5 th June 2007 Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter Mark Bailey, G A Bass, N D Lee, G A Mapp Contents 1. History

More information

Objectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process

Objectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process Tolerance Limits and Action Levels for IMRT QA Jatinder R Palta, Ph.D Siyong Kim, Ph.D. Department of Radiation Oncology University of Florida Gainesville, Florida Objectives Describe the uncertainties

More information

General Overview of Gas Filled Detectors

General Overview of Gas Filled Detectors GAS-FILLED DETECTOR General Overview of Gas Filled Detectors Gas-Filled Detectors Ion chamber Proportional counter G-M (Geiger-Miller) counter Diagram of a Generic Gas-Filled Detector A Anode High-voltage

More information

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer SLAC-PUB-8099 July 1999 Air Kerma Calibration Factors and k ch Values for PTW Soft X-ray, NACP and Roos Ionization Chambers at Very Low X-ray Energies (0.035 mm - 1.0 mm Al HVL) N. E. Ipe*, K. E. Rosser,

More information

B E A M Q U A L I T Y S P E C I F I C A T I O N I N K I L O-V O L T A G E R A D I O T H E R A P Y

B E A M Q U A L I T Y S P E C I F I C A T I O N I N K I L O-V O L T A G E R A D I O T H E R A P Y B E A M Q U A L I T Y S P E C I F I C A T I O N I N K I L O-V O L T A G E R A D I O T H E R A P Y Sibusiso Jozela A research report submitted to the Faculty of Science, University of the Witwatersrand,

More information

PHITS calculation of the radiation field in HIMAC BIO

PHITS calculation of the radiation field in HIMAC BIO PHITS calculation of the radiation field in HIMAC BIO Ondřej Ploc, Yukio Uchihori, Hisashi Kitamura, Lembit Sihver National Institute of Radiological Sciences, Chiba, Japan Nuclear Physics Institute, Prague,

More information

Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams

Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams Ramamoorthy Ravichandran,

More information

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National

More information

Recombination factor dependancy of high and low dose pulsed accelerators for electron beam energies

Recombination factor dependancy of high and low dose pulsed accelerators for electron beam energies Recombination factor dependancy of high and low dose plsed accelerators for electron beam energies Giseppe Felici Sordina IORT Technologies S.p.A. Talk Overview Dose per plse what is the so called high

More information

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Norio Saito, Tadahiro Kurosawa, Masahiro Kato, Yuichiro Morishita, Takahiro Tanaka, and Morihito

More information

Chapiter VII: Ionization chamber

Chapiter VII: Ionization chamber Chapiter VII: Ionization chamber 1 Types of ionization chambers Sensitive volume: gas (most often air direct measurement of exposure) ionization chamber Sensitive volume: semiconductor (silicon, germanium,

More information

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego Physics of Novel Radiation Modalities Particles and Isotopes Todd Pawlicki, Ph.D. UC San Diego Disclosure I have no conflicts of interest to disclose. Learning Objectives Understand the physics of proton

More information

Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling.

Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling. Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling. 5 Sergei Zavgorodni 1,2, Eyad Alhakeem 2,1 and Reid Townson 2,1 1 Department of

More information

Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers

Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers Klaus Derikum Physikalisch-Technische undesanstalt undesallee 100, 38116 raunschweig, Germany

More information

Instrumentation for Verification of Dose

Instrumentation for Verification of Dose Instrumentation for Verification of Dose MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 52 Essen, Germany, May 3 5, 2013 Consistent and harmonized dosimetry guidelines

More information

Visual Analysis of the Daily QA Results of Photon and Electron Beams of a Trilogy Linac over a Five-Year Period

Visual Analysis of the Daily QA Results of Photon and Electron Beams of a Trilogy Linac over a Five-Year Period International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2015, 4, 290-299 Published Online November 2015 in SciRes. http://www.scirp.org/journal/ijmpcero http://dx.doi.org/10.4236/ijmpcero.2015.44035

More information

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin

More information

The development of new devices for accurate radiation dose measurement

The development of new devices for accurate radiation dose measurement The development of new devices for accurate radiation dose measurement A guarded liquid ionization chamber and an electron sealed water calorimeter Kristin J. Stewart Department of Physics McGill University,

More information

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators David S. Followill,*

More information

Monte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET)

Monte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET) Monte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET) Christina Jarlskog Department of Radiation Physics Lund University, Malmö University Hospital (UMAS)

More information

The electron accelerator of the ISOF-CNR Institute: its characteristics and use

The electron accelerator of the ISOF-CNR Institute: its characteristics and use The electron accelerator of the ISOF-CNR Institute: its characteristics and use P. Fuochi, U. Corda, and M. Lavalle ISOF-CNR Institute, Via P. Gobetti 101, I-40129 Bologna, Italy LINAC Laboratory: ISOF-CNR,

More information

Manipulation on charged particle beam for RT benefit.

Manipulation on charged particle beam for RT benefit. High Electron Beam Dose Modification using Transverse Magnetic Fields Ion Chamber Response Modification under Strong Magnetic Field Conditions Sion Koren, Radiation Oncology Preface Manipulation on charged

More information

Radiation Quantities and Units

Radiation Quantities and Units Radiation Quantities and Units George Starkschall, Ph.D. Lecture Objectives Define and identify units for the following: Exposure Kerma Absorbed dose Dose equivalent Relative biological effectiveness Activity

More information

Small field dosimetry

Small field dosimetry Small ield dosimetry Hugo Palmans edaustron Wiener Neustadt Austria and National Physical Laboratory Teddington UK 1 Acknowledgements IAEA/AAP small and composite ield working group (Rodolo Alonso Pedro

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de

More information

Electron beam energy QA a note on measurement tolerances

Electron beam energy QA a note on measurement tolerances JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Electron beam energy QA a note on measurement tolerances Juergen Meyer, a Matthew J. Nyflot, Wade P. Smith, Landon S. Wootton, Lori

More information