ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

Size: px
Start display at page:

Download "ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER"

Transcription

1 ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2), D. Dreossi(3) (1) Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, C R Casaccia, cp 2400 Roma, Italy. (2) Società Elettra, impianto sincrotrone, Trieste (3) Università di Trieste, Trieste 1

2 INTRODUCTION Monochromatic synchrotron light provides better resolution and sharpness of the radiological images and the advantages of using such beams for mammography have been widely assessed since the past years. One of the research programmes of the Synchrotron ELETTRA operating in Trieste, Italy, is to dedicate a beam line of the accelerator for high-definition mammography in the framework of the SYRMA project (SYnchrotron Radiation for MAmmography). Accurate air-kerma measurements in this beam are then necessary in order to optimize the patient absorbed dose with respect to the definition of the radiographic image. A collaboration between the INMRI- and Synchrotron ELETTRA was started aiming at characterizing monochromatic synchrotron light beams in terms of air kerma. 2

3 OVERVIEW SYRMA beam line Radiation detectors Method Results: absolute air kerma measurements SYRMA monitors chambers calibration secondary standard free-air chamber calibration Conclusions 3

4 THE SYRMA BEAM LINE SCHEMATIC DIAGRAM OF THE LIGHT BEAM bending magnet of ELETTRA polychromatic laminar beam Light source Safety and control elements vacuum slits Si(111) double crystal monochromator Vacuum slits OPTICS HUTCH beam preparation (energy, flux, geometry) Monochromator Beam stopper air slits monochromatic laminar beam fast shutter Al filters Air Slits SETUP SYRMEP Fast shutters Slits, dose monitors EXPERIMENTAL HUTCH beam monitoring (dose monitors, exposure times, safety system) ENERGY SELECTION FLUX AND GEOMETRY SELECTION CCD beam monitors detector reference point low energy free-air chamber PATIENT ROOM Patient exposure X-RAY BEAM Schematic layout of the ELETTRA medical beam line for high definition mammography in the framework of the SYRMA project (SYnchrotron Radiation for MAmmography) at the Synchrotron ELETTRA in Trieste, Italy. 4

5 photons flux / mm -2 s -1 Absolute air-kerma measurement in a synchrotron light beam by ionization free-air chamber THE SYRMA BEAM LINE BEAM PHOTON FLUX AND CHARACTERISTICS GeV ; 300 ma 2.4 GeV ; 180 ma E / kev Photon flux values - as function of the photons energy - at the reference point for patient exposure, when synchrotron Elettra is operated at 2 GeV and 2.4 GeV with stored the maximum electron beam currents of 300 ma and 180 ma, respectively. Light source CHARACTERISTIC Relevant characteristics of the light source and x-ray beams for high definition mammography in the framework of the SYRMA project at the Synchrotron ELETTRA in Trieste, Italy. 5 VALUE Type Bending magnet Critical energy (at 2.0 GeV) 3.21 kev Critical energy (at 2.4 GeV) 5.59 kev Source size 100 µm x 1100 µm Horizontal angular acceptance (beam divergence) 7 mrad X-ray beam at reference point Energy range 8 kev - 35 kev Energy range for patient exposure 16 kev 25 kev Energy Resolution ΔE/E = 2*10-3 Typical photons fluxes at 15 kev 2 x 10 8 / mm -2 s -1 (at 2 GeV, 300 ma) 7 x 10 8 phot./mm -2 s -1 (at 2.4 GeV, 180 ma) Beam Size 120 mm x 4 mm Source-to-sample distance about 30 m (26.5 m in vacuum)

6 THE SYRMA BEAM LINE REFERENCE MEASUREMENT POINT (a) Speed controlled movable support (a) for patient exposure to obtain high definition mammography by scanning the patient through the beam itself at the Synchrotron ELETTRA. Detail in figure 4b shows the position of the mild compression system used for stretching and equalizing the breast tissues thickness. (b) The standard chambers are placed in this position for air kerma measurements, after removing the mild compression system. 6

7 METHOD Because of the specific irradiation conditions in the synchrotron light beam (energy, intensity and shape), it was decided to perform absolute air kerma measurements directly in the SYRMA beam by the standard lowenergy-free-air chamber (-FAC) of INMRI-. To this end, the diaphragm of the standard chamber -FAC (4 mm in radius) was replaced by a smaller one with a radius of 1.5 mm. A re-determination of relevant correction factors was made Two SYRMA monitor chambers - in fixed positions on the beam were then calibrated by absolute air-kerma rate measurements in the energy range from 8 kev to 30 kev. To assure the periodical calibration of these beam monitors the synchrotron users were equipped by a non-standard free-air chamber (-SEC) made at INMRI- and acting as a secondary standard. The chamber -SEC was calibrated in terms of air kerma in the SYRMA beam against the standard chamber -FAC (It will be periodically calibrated at the INMRI- in the low-energy filtered x- ray beams to check for stability). 7

8 BEAM STABILITY AND UNIFORMITY Before and after each series of calibration measurements: the fine angular alignments of the monochromator crystals was controlled as the energy of the monochromatic beam depend on monochromator position. the spatial uniformity and stability of the beam was accurately verified by measurements made by the standard chamber and by a digital detector based on a silicon charge coupled detector (CCD) that presents to the incident photons a grid of 2048x2024 pixels (each with typical dimensions of 14 µm x 14 µm). the homogeneous irradiation of the chamber diaphragm was verified by the CCD that provided beam imaging after the chambers. 3 mm 3.7 mm 191 mm 8

9 STANDARD IONIZATION CHAMBERS The standard free-air chamber for low energy x-rays (-FAC) and the secondary standard free-air chamber (-SEC) at the reference measurement point (inside the red circle) under the patient support, at the Synchrotron ELETTRA in Trieste, Italy. The two chambers use the same diaphragm. 9

10 BEAM MONITOR IONIZATION CHAMBERS Designed by ELETTRA for the laminar beam geometry with an entrance window larger than the largest beam size. The electrodes and windows are perpendicular to the beam and are realized in Mylar foils - 50 mm thick - covered on each side by a 50 nm layer of Al. HV ELECTRODES HV LIGHT BEAM The sensitive volume is constituted by two regions delimited by the two HV electrodes and separated by the central collecting electrode. The chambers are inside an Al box, shielded against RF. 10

11 EFFECTIVENESS OF THE MONITORING Photons fluence in the synchrotron light beam was never constant because it varies according to the electron beam currents stored in the ring of the synchrotron. Possibility to compare measurements carried out in different period is based on the confidence in the beam monitoring system. I IOC1 /A I IOC2 /A I -SEC /A IOC1/-SEC IOC2/-SEC MEDIA E E E+14 STD E E E+11 STD% 1.27% 1.28% 1.29% 0.03% 0.02% MaxVar% 5.02% 5.04% 5.06% 0.18% 0.17% 11

12 CORRECTION FACTORS K SC, K FL AND K E Determined by the Monte Carlo code PENELOPE for the standard chamber - FAC - with the 1.5 mm in radius chamber aperture - in monochromatic photon beams at the energies in the range from 8 kev to 30 kev. Relative standard uncertainty estimated for the k sc k fl k e product: within 0.1%. No significant difference in the k sc k fl k e values obtained for the smaller 1.5 mm diaphragm and the usual 4.0 mm diaphragm.

13 MASS AIR ATTENUATION COEFFICIENTS, μ/ρ μ/ρ values measured in synchrotron monochromatic beams (triangles) and ISO 4037 filtered x-radiation (circles), in the energy range from 8 kev to 30 kev. Standard uncertainties were estimated to be within 0.5 %. The difference from the experimental values obtained for filtered x radiation having a mean energy of the spectra E and those obtained for monochromatic photons having the same energy E, resulted in the range from 0.7 % to 1 % being values referring to filtered x radiation lower. The values given by Hubbell 95 (squares) are also reported for comparison and differ of about 0.9 % from those measured, being those measured lower. Such deviations would influence the corresponding correction factors by no more than 0.1 %.

14 ION RECOMBINATION CORRECTION I(1600 V)/I(400 V Ksat determination -FAC ksat < E+0 2.0E E E E E E E E E E-10 0 I(1600 V) /A I(2000 V)/I(250 V Ksat determination -SSC ksat < E-12 1E E-11 2E E-11 3E E-11 4E E-11 I(2000 V) /A I(400 V )/ I(100 V Ksat determination SYRMA-IOC ksat < E+00 1.E+05 2.E+05 3.E+05 4.E+05 5.E+05 6.E+05 7.E+05 I(400 V) /A The effect of ion recombination in the chamber exposed to the synchrotron light beam (with a kerma rate of up to 7 mgy s -1, was corrected for by the method described by Boutillon Standard uncertainty within 0.05%.

15 CALIBRATION RESULTS FOR SYRMA MONITOR Calibration coefficients of the SYRMA monitor chambers IOC 1(blue square) and IOC2 (red triangle) at reference ambient conditions (temperature of K, pressure of Pa and 50% of relative humidity). The associated standard uncertainty is 0.75%. Due to the SYRMA beam energy resolution ΔE/E and the chamber energy dependance, the possible variation of the calibration coefficient is 0.6% at each energy, in the energy range useful for patient exposure (from 16 kev to 25 kev).

16 CALIBRATION COEFFICIENTS FOR THE - SEC CALIBRATED FREE-AIR CHAMBER The associated standard uncertainty is within 0.80%. The difference between the determinations by monitor IOC1 (blue square) or IOC2 (red triangle). is within 0.15%. Due to the SYRMA beam energy resolution DE/E and the chamber energy dependance, the possible variation of the calibration coefficient is 0.03% at each energy, in the energy range useful for patient exposure (from 16 kev to 25 kev).

17 CONCLUSION Absolute air-kerma measurements by the standard free-air chamber for low energy x-rays (-FAC) were performed in the synchrotron light beam used for high definition mammography in the framework of the SYRMA project (SYnchrotron Radiation for MAmmography) at the Synchrotron ELETTRA in Trieste. These measurements were necessary to perform directly in the synchrotron light beam the calibration of the two beam monitors. The -SEC chamber was calibrated during the present measurements directly against the -FAC as a secondary standard to assure the periodical calibration of the synchrotron beam monitors. A procedure to verify the -SEC stability by measurements in lowenergy filtered x-ray beams, was established.

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de

More information

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Pierino De Felice and Marco Capogni - ENEA C.R. Casaccia Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA-INMRI) Roma ITALY Yoshio

More information

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron P. Berkvens and D. Häusermann European Synchrotron Radiation Facility BP 0, Grenoble Cedex 0, France Australian Synchrotron

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

Chapiter VII: Ionization chamber

Chapiter VII: Ionization chamber Chapiter VII: Ionization chamber 1 Types of ionization chambers Sensitive volume: gas (most often air direct measurement of exposure) ionization chamber Sensitive volume: semiconductor (silicon, germanium,

More information

Beamline practice at BL01B1 (XAFS) In-situ XAFS measurement of catalyst samples

Beamline practice at BL01B1 (XAFS) In-situ XAFS measurement of catalyst samples Beamline practice at BL01B1 (XAFS) In-situ XAFS measurement of catalyst samples ver. 2015/09/18 T. Ina, K. Kato, T. Uruga (JASRI), P. Fons (AIST/JASRI) 1. Introduction The bending magnet beamline, BL01B1,

More information

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays C Kessler 1, D T Burns 1, P Roger 1, V Sochor 2 1 Bureau International des Poids et

More information

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields Journal of NUCLEAR SCIENCE and TECHNOLOGY, Vol. 44, No. 2, p. 109 113 (2007) ARTICLE A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard

More information

Secondary Particles Produced by Hadron Therapy

Secondary Particles Produced by Hadron Therapy Iranian Journal of Medical Physics Vol. 12, No. 2, Spring 2015, 1-8 Received: March 10, 2015; Accepted: July 07, 2015 Original Article Secondary Particles Produced by Hadron Therapy Abdolkazem Ansarinejad

More information

Radiological safety studies for the TeraFERMI beamline at

Radiological safety studies for the TeraFERMI beamline at Radiological safety studies for the TeraFERMI beamline at FERMI@elettra K.Casarin 1, L. Fröhlich 2, G.Tromba 1, A.Vascotto 1 1 Elettra - Sincrotrone Trieste S.C.p.A., Trieste, Italy 2 Deutsches Elektronen-Synchrotron

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM Rapport BIPM-09/02 PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM D T Burns, C Kessler, P Roger BIPM A R El-Sersy NIS (Egypt) July 2009 BUREAU INTERNATIONAL DES POIDS ET

More information

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma 4.1b - Cavity Theory Lecture 2 Peter R Almond 2011 Overview of Lecture Exposure (W/e) air Exposure and Air Kerma Exposure Exposure is symbolized as X and defined by the ICRU as the quotient of dq by dm,

More information

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator Determination of Ambient Dose quivalent at INFLPR 7 MeV Linear Accelerator F. Scarlat, A. Scarisoreanu, M. Oane,. Badita,. Mitru National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest-Magurele,

More information

Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source

Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source P. K. Sahani 1,5, A. K. Das 2, Haridas G. 3, A. K. Sinha 4,5, B. N. Rajasekhar 2,5, T. A. Puntambekar 1 and N K Sahoo

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL J-B. Pruvost 1, P. Berkvens 2, F. Justine 1, C. Mage 1 1: Synchrotron SOLEIL BP.48, 91192

More information

Introduction to XAFS. Grant Bunker Associate Professor, Physics Illinois Institute of Technology. Revised 4/11/97

Introduction to XAFS. Grant Bunker Associate Professor, Physics Illinois Institute of Technology. Revised 4/11/97 Introduction to XAFS Grant Bunker Associate Professor, Physics Illinois Institute of Technology Revised 4/11/97 2 tutorial.nb Outline Overview of Tutorial 1: Overview of XAFS 2: Basic Experimental design

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation

More information

New irradiation zones at the CERN-PS

New irradiation zones at the CERN-PS Nuclear Instruments and Methods in Physics Research A 426 (1999) 72 77 New irradiation zones at the CERN-PS M. Glaser, L. Durieu, F. Lemeilleur *, M. Tavlet, C. Leroy, P. Roy ROSE/RD48 Collaboration CERN,

More information

Chapter V: Cavity theories

Chapter V: Cavity theories Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly

More information

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays D.T. Burns, W. Tiefenböck* and J. Witzani* Bureau International des Poids et Mesures, Pavillon

More information

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento

More information

Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY

Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY Yvonne Bergmann, Klaus Ott Helmholtz- Zentrum Berlin BESSY II Radiation Protection Department yvonne.bergmann@helmholtz-berlin.de

More information

A STUDY OF MULTIPLE REFRACTIVE SCATTERING OF MONOENERGETIC X-RAYS FROM ENSEMBLES OF MONODISPERSE SPHERES. Masterarbeit

A STUDY OF MULTIPLE REFRACTIVE SCATTERING OF MONOENERGETIC X-RAYS FROM ENSEMBLES OF MONODISPERSE SPHERES. Masterarbeit A STUDY OF MULTIPLE REFRACTIVE SCATTERING OF MONOENERGETIC X-RAYS FROM ENSEMBLES OF MONODISPERSE SPHERES Masterarbeit zur Erlangung des akademischen Grades Master of Science in Physics (M.Sc.) dem Fachbereich

More information

Comparison of Polychromatic and Monochromatic X-rays for Imaging

Comparison of Polychromatic and Monochromatic X-rays for Imaging Comparison of Polychromatic and Monochromatic X-rays for Imaging M. Hoheisel 1, P. Bernhardt 1, R. Lawaczeck 2, and H. Pietsch 2 1 Siemens AG Medical Solutions, Forchheim, Germany 2 Schering AG Imaging

More information

Shielding calculations for the design of new Beamlines at ALBA Synchrotron

Shielding calculations for the design of new Beamlines at ALBA Synchrotron Shielding calculations for the design of new Beamlines at ALBA Synchrotron A. Devienne 1, M.J. García-Fusté 1 1 Health & Safety Department, ALBA Synchrotron, Carrer de la Llum -6, 0890 Cerdanyola del Vallès,

More information

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 511 FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON

More information

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland Available on CMS information server CMS CR -2018/225 The Compact Muon Solenoid Experiment Conference Report Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland 27 September 2018 (v2, 19 November

More information

How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source

How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source What Makes Up The Canadian Light Source? 4. Storage Ring 5. Synchrotron Light 6. Beamline 1. Electron Gun 2. Linear Accelerator

More information

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3) PROPOSAL FOR EYE-LENS DOSEMETER CALIBRATION AND TYPE TESTING ORAMED WP2 J.-M. Bordy (1), J. Daures (1), M. Denozière (1), G. Gualdrini (2), M. Guijaume (3), E. Carinou (4),F. Vanhavere (5) (1) Laboratoire

More information

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005 Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp. 233 237 Advance Access publication 6 October 2005 doi:10.1093/rpd/nci353 DOSE BUILD UP CORRECTION FOR RADIATION MONITORS IN HIGH-ENERGY BREMSSTRAHLUNG

More information

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays Final Report 2010-01-12 Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays C. Kessler 1, P.J. Allisy-Roberts 1, D.T. Burns 1, A.S. Guerra 2,

More information

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors 6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator

More information

Quantitative determination of the effect of the harmonic component in. monochromatised synchrotron X-ray beam experiments

Quantitative determination of the effect of the harmonic component in. monochromatised synchrotron X-ray beam experiments Frascati Physics Series Vol. XXX (1997), pp. 000-000 Conference Title - Conference Town, Oct 3rd, 1996 Quantitative determination of the effect of the harmonic component in monochromatised synchrotron

More information

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical ORIGINAL ARTICLE Development of MGDA Software for Calculation of Patient Specific Mean Glandular Dose during Mammography 1Meherun Nahar, 2 M Sazzad, 3 Abdus Sattar Mollah and 4 Mir Mohammad Akramuzzaman

More information

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE Copyright(C)JCPDS-International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Vol.46 74 ISSN 1097-0002 LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE K. Chouffani 1, D. Wells

More information

Two Dimensional SR-Interferometer at PETRA III. Artem Novokshonov DESY, Tomsk Polytechnic University

Two Dimensional SR-Interferometer at PETRA III. Artem Novokshonov DESY, Tomsk Polytechnic University Two Dimensional SR-Interferometer at PETRA III Artem Novokshonov DESY, Tomsk Polytechnic University First Part: Two Dimensional Interferometer at Petra III Interferometer principal Schematic of the Interferometer

More information

EUV Reflectivity measurements on Acktar Sample Magic Black

EUV Reflectivity measurements on Acktar Sample Magic Black Report EUV Reflectivity measurements on Acktar Sample Magic Black S. Döring, Dr. K. Mann Laser-Laboratorium Göttingen e.v. October 28, 2011 Contents 1 Introduction 3 2 Setup 3 3 Measurements 4 4 Conclusion

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September

More information

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT) 1- Ionizing radiation metrology for radiation protection 2- Metrological requirements for ionizing radiation measurement in radiotherapy and radiodiagnostics R. F. Laitano Part 2 Metrological traceability

More information

Diffractometer. Geometry Optics Detectors

Diffractometer. Geometry Optics Detectors Diffractometer Geometry Optics Detectors Diffractometers Debye Scherrer Camera V.K. Pecharsky and P.Y. Zavalij Fundamentals of Powder Diffraction and Structural Characterization of Materials. Diffractometers

More information

PHITS calculation of the radiation field in HIMAC BIO

PHITS calculation of the radiation field in HIMAC BIO PHITS calculation of the radiation field in HIMAC BIO Ondřej Ploc, Yukio Uchihori, Hisashi Kitamura, Lembit Sihver National Institute of Radiological Sciences, Chiba, Japan Nuclear Physics Institute, Prague,

More information

David Martin High Precision Beamline Alignment at the ESRF IWAA, Grenoble 3-7 October 2016

David Martin High Precision Beamline Alignment at the ESRF IWAA, Grenoble 3-7 October 2016 David Martin High Precision Beamline Alignment at the ESRF IWAA, Grenoble 3-7 October 2016 OVERVIEW The ESRF has just completed the Phase I Upgrade programme. The Phase I Upgrade programme was centered

More information

Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS

Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS SLAC RADIATION PHYSICS NOTE RP-08-11 September 23, 2008 Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS J. Vollaire, A. Prinz Radiation Protection Department,

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

Introduction. X-Ray Production and Quality. Fluorescence Yield. Fluorescence X-Rays. Initiating event. Initiating event 3/18/2011

Introduction. X-Ray Production and Quality. Fluorescence Yield. Fluorescence X-Rays. Initiating event. Initiating event 3/18/2011 X-Ray Production and Quality Chapter 9 F.A. Attix, Introduction to Radiological Physics and Radiation Dosimetry Introduction Physics of x-ray generation Fluorescence x-rays Bremsstrahlung x-rays Beam quality

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays D T Burns 1, C Kessler 1, E Mainegra-Hing 2, H Shen 2 and M R McEwen 2 1 Bureau International

More information

DICOM Correction Item

DICOM Correction Item DICOM Correction Item Correction Number CP-687 Log Summary: of Modification Addition Name of Standard DICOM PS 3.16 2007 Rationale for Correction: When the Diagnostic X-Ray Radiation Dose Reporting SOP

More information

General characteristics of radiation dosimeters

General characteristics of radiation dosimeters General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Development of 2-Dimentional Imaging XAFS System at BL-4

Development of 2-Dimentional Imaging XAFS System at BL-4 Development of 2-Dimentional Imaging XAFS System at BL-4 Koichi Sumiwaka 1, Misaki Katayama 2, Yasuhiro Inada 2 1) Department of Applied Chemistry, College of Science and Engineering, Ritsumeikan, University,

More information

Initial Certification

Initial Certification Initial Certification Medical Physics Part 1 Content Guide Part 1 Content Guides and Sample Questions PLEASE NOTE: List of Constants and Physical Values for Use on the Part 1 Physics Exam The ABR provides

More information

Synchrotron Radiation a Tool for Precise Beam Energy Measurements at the ILC

Synchrotron Radiation a Tool for Precise Beam Energy Measurements at the ILC Synchrotron Radiation a Tool for Precise Beam Energy Measurements at the ILC K.Hiller, R.Makarov, H.J.Schreiber, E.Syresin and B.Zalikhanov a BPM based magnetic spectrometer example E b see LC-DET-2004-031

More information

X-Ray Diagnostics Commissioning at the LCLS

X-Ray Diagnostics Commissioning at the LCLS X-Ray Diagnostics Commissioning at the LCLS - Selected Studies - J. Welch, SLAC National Accelerator Laboratory Aug. 3-27, 2010 Commissioning Studies Microbunching Instability Laser Heater tune-up Gas

More information

ICTP-IAEA Joint Workshop on Nuclear Data for Science and Technology: Medical Applications. 30 September - 4 October, 2013

ICTP-IAEA Joint Workshop on Nuclear Data for Science and Technology: Medical Applications. 30 September - 4 October, 2013 2484-11 ICTP-IAEA Joint Workshop on Nuclear Data for Science and Technology: Medical Applications 30 September - 4 October, 2013 Experimental techniques (Nuclear reaction data, estimation of uncertainties)

More information

Electron Impact Ionization in EGSnrc

Electron Impact Ionization in EGSnrc Electron Impact Ionization in EGSnrc by Leila Lukhumaidze A thesis submitted to the Faculty of Graduate and Postdoctoral Affairs in partial fulfillment of the requirements for the degree of Master of Science

More information

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities.

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities. SLAC-PUB-7906 November, 1998 Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities. N. E. Ipe, A. Fassò, K.

More information

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range

More information

Ultrafast X-Ray-Matter Interaction and Damage of Inorganic Solids October 10, 2008

Ultrafast X-Ray-Matter Interaction and Damage of Inorganic Solids October 10, 2008 Ultrafast X-Ray-Matter Interaction and Damage of Inorganic Solids October 10, 2008 Richard London rlondon@llnl.gov Workshop on Interaction of Free Electron Laser Radiation with Matter Hamburg This work

More information

II) Experimental Design

II) Experimental Design SLAC Experimental Advisory Committee --- September 12 th, 1997 II) Experimental Design Theory and simulations Great promise of significant scientific and technological achievements! How to realize this

More information

Beam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI)

Beam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI) Beam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI), 2003, A dedicated proton accelerator for 1p-physics at the future GSI Demands facilities for

More information

Abstract. 1. Introduction

Abstract. 1. Introduction The New Upgrade of SESAME D.Einfeld1, R.H.Sarraf2, M.Attal3, H.Hashemi4, A.Elsisi5, A.Amro6, H.Hassanzadegan4, K.Tavakoli3, B.Kalantari7, S. Varnasery8, E. Al-Dmour8, D. Foudeh6, H.Tarawneh9, A.Aladwan7

More information

SYNCHROTRON RADIATION: NEW TREND IN X-RAY MAMMOGRAPHY

SYNCHROTRON RADIATION: NEW TREND IN X-RAY MAMMOGRAPHY Vol. 91 (1997) ACTA PHYSICA POLONICA A No. 4 Proceedings of the 3rd ISSSRNS Jaszowiec '96 SYNCHROTRON RADIATION: NEW TREND IN X-RAY MAMMOGRAPHY E. BURATTINI National Laboratory of Nuclear Physics, INFN

More information

Determination of Absolute Neutron Fluence to sub-0.1% uncertainty (and better)

Determination of Absolute Neutron Fluence to sub-0.1% uncertainty (and better) Determination of Absolute Neutron Fluence to sub-0.1% uncertainty (and better) Andrew Yue University of Maryland / NIST for the Alpha-Gamma Collaboration NIST-ILL-Sussex neutron lifetime experiments Neutron

More information

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN Problem P7 Stéphanie Ménard Dosimetry Department 92262 Fontenay-aux aux-roses FRANCE What are the applications of Gamma-Ray Spectrometry in Radiological Protection and in Safety? In the environment: after

More information

EXPERIMENTS CHARACTERIZING THE X-RAY EMISSION FROM A SOLID-STATE CATHODE USING A HIGH-CURRENT GLOW DISCHARGE

EXPERIMENTS CHARACTERIZING THE X-RAY EMISSION FROM A SOLID-STATE CATHODE USING A HIGH-CURRENT GLOW DISCHARGE EXPERIMENTS CHARACTERIZING THE X-RAY EMISSION FROM A SOLID-STATE CATHODE USING A HIGH-CURRENT GLOW DISCHARGE A.B. KARABUT AND S.A. KOLOMEYCHENKO FSUE SIA LUCH 24 Zheleznodorozhnaja Street, Podolsk, Moscow

More information

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:

More information

An experimental study of dual-energy CT imaging using synchrotron radiation

An experimental study of dual-energy CT imaging using synchrotron radiation Nuclear Science and Techniques 24 (2013) 020102 An experimental study of dual-energy CT imaging using synchrotron radiation HAO Jia 1,2 ZHANG Li 1,2 XING Yuxiang 1,2 KANG Kejun 1,2 1 Department of Engineering

More information

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc Stephen D. Davis and Larry A. DeWerd University of Wisconsin, Madison, WI 13 th UK Monte Carlo User Group meeting March

More information

JSPS Asien Science Seminar Synchrotron Radiation Science

JSPS Asien Science Seminar Synchrotron Radiation Science SESAME Synchrotron Light for Experimental Science and Applications in the Middle East JSPS Asien Science Seminar Synchrotron Radiation Science Dieter Einfeld Amman, October 2002 Introduction Based on the

More information

ATHENA / AD-1. First production and detection of cold antihydrogen atoms. ATHENA Collaboration. Rolf Landua CERN

ATHENA / AD-1. First production and detection of cold antihydrogen atoms. ATHENA Collaboration. Rolf Landua CERN ATHENA / AD-1 First production and detection of cold antihydrogen atoms ATHENA Collaboration Rolf Landua CERN 1 LONG TERM PHYSICS GOALS Antihydrogen = Hydrogen? CPT Gravity But... 2 FIRST GOAL PRODUCTION

More information

X-ray Absorption Spectroscopy Eric Peterson 9/2/2010

X-ray Absorption Spectroscopy Eric Peterson 9/2/2010 X-ray Absorption Spectroscopy Eric Peterson 9/2/2010 Outline Generation/Absorption of X-rays History Synchrotron Light Sources Data reduction/analysis Examples Crystallite size from Coordination Number

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

Radiation Protection At Synchrotron Radiation Facilities

Radiation Protection At Synchrotron Radiation Facilities 3 rd ILSF Advanced School on Synchrotron Radiation and Its Applications September 14-16, 2013 Radiation Protection At Synchrotron Radiation Facilities Ehsan Salimi Shielding and Radiation Safety Group

More information

ThomX Machine Advisory Committee. (LAL Orsay, March ) Ring Beam Dynamics

ThomX Machine Advisory Committee. (LAL Orsay, March ) Ring Beam Dynamics ThomX Machine Advisory Committee (LAL Orsay, March 20-21 2017) Ring Beam Dynamics A. Loulergue, M. Biagini, C. Bruni, I. Chaikovska I. Debrot, N. Delerue, A. Gamelin, H. Guler, J. Zang Programme Investissements

More information

Measurement of Beam Profile

Measurement of Beam Profile Measurement of Beam Profile The beam width can be changed by focusing via quadruples. Transverse matching between ascending accelerators is done by focusing. Profiles have to be controlled at many locations.

More information

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection R. Kreuger, C. W. E. van Eijk, Member, IEEE, F. A. F. Fraga, M. M. Fraga, S. T. G. Fetal, R. W. Hollander, Member, IEEE, L. M.

More information

Calculations of Neutron Yield and Gamma Rays Intensity by GEANT4

Calculations of Neutron Yield and Gamma Rays Intensity by GEANT4 Armenian Journal of Physics, 2016, vol. 9, issue 4, pp. 315-323 Calculations of Neutron Yield and Gamma Rays Intensity by GEANT4 R. Avagyan, R. Avetisyan, V. Ivanyan*, I. Kerobyan A.I. Alikhanyan National

More information

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays D T Burns Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312

More information

Calibration of the IXPE Instrument

Calibration of the IXPE Instrument Calibration of the IXPE Instrument Fabio Muleri (INAF-IAPS) On behalf of the IXPE Italian Team 13th IACHEC Meeting 2018 Avigliano Umbro (Italy), 9-12 April 2018 IXPE MISSION IXPE will (re-)open the polarimetric

More information

Research with Synchrotron Radiation. Part I

Research with Synchrotron Radiation. Part I Research with Synchrotron Radiation Part I Ralf Röhlsberger Generation and properties of synchrotron radiation Radiation sources at DESY Synchrotron Radiation Sources at DESY DORIS III 38 beamlines XFEL

More information

MEASUREMENT OF TEMPORAL RESOLUTION AND DETECTION EFFICIENCY OF X-RAY STREAK CAMERA BY SINGLE PHOTON IMAGES

MEASUREMENT OF TEMPORAL RESOLUTION AND DETECTION EFFICIENCY OF X-RAY STREAK CAMERA BY SINGLE PHOTON IMAGES Proceedings of IBIC212, Tsukuba, Japan MEASUREMENT OF TEMPORAL RESOLUTION AND DETECTION EFFICIENCY OF X-RAY STREAK CAMERA BY SINGLE PHOTON IMAGES A. Mochihashi, M. Masaki, S. Takano, K. Tamura, H. Ohkuma,

More information

Survey Meter OD-01 Address: Phone: Fax: URL:

Survey Meter OD-01 Address: Phone: Fax:   URL: Survey Meter OD-01 Dose meter and dose rate meter for the measurement of the ambient dose and dose rate equivalent H*(10), dh*(10)/dt and the directional dose and dose rate equivalent H'(0.07), dh'(0.07)/dt

More information

Mahmoud Abdellatief, PhD Materials Science BL Scientist SESAME Synchrotron Jordan

Mahmoud Abdellatief, PhD Materials Science BL Scientist SESAME Synchrotron Jordan Mahmoud Abdellatief, PhD Materials Science BL Scientist SESAME Synchrotron Jordan Outlines Introduction MS Layout Ray Tracing Source Front end Optics Experimental SESAME Synchrotron Synchrotron light for

More information

Experiments on deflection of charged particles using silicon crystals at. Proton Synchrotron (KEK(

Experiments on deflection of charged particles using silicon crystals at. Proton Synchrotron (KEK( Experiments on deflection of charged particles using silicon crystals at REFER ring (Hiroshima( University) and Proton Synchrotron (KEK( KEK) S. Sawada KEK High Energy Accelerator Research Organization

More information

RADIATION SOURCES AT SIBERIA-2 STORAGE RING

RADIATION SOURCES AT SIBERIA-2 STORAGE RING RADIATION SOURCES AT SIBERIA-2 STORAGE RING V.N. Korchuganov, N.Yu. Svechnikov, N.V. Smolyakov, S.I. Tomin RRC «Kurchatov Institute», Moscow, Russia Kurchatov Center Synchrotron Radiation undulator undulator

More information

X-ray Fluorescence Imaging Following Synchrotron Beam Excitation

X-ray Fluorescence Imaging Following Synchrotron Beam Excitation Conference on Applied Digital Imaging Techniques for Understanding the Palimpsest X-ray Fluorescence Imaging Following Synchrotron Beam Excitation Uwe Bergmann Stanford Synchrotron Radiation Laboratory

More information

4.2 Photon Beam Line Systems

4.2 Photon Beam Line Systems shielding walls will be ~1-m thick, and the transverse walls will be ~1.5-m thick. Polyethylene will be used to shield neutrons, and additional lead and/or steel will be used for localized shielding as

More information

AIRFLY: Measurement of the Air Fluorescence induced by electrons

AIRFLY: Measurement of the Air Fluorescence induced by electrons AIRFLY: Measurement of the Air Fluorescence induced by electrons Valerio Verzi INFN Sezione di Roma II For the Airfly collaboration 9 th Topical Seminar on Innovative Particle and Radiation Detectors 23-26

More information

Radiation Safety Assessment of the CLS Beamlines Using FLUKA Monte-Carlo Code

Radiation Safety Assessment of the CLS Beamlines Using FLUKA Monte-Carlo Code Radiation Safety Assessment of the CLS Beamlines Using FLUKA Monte-Carlo Code Mo Benmerrouche Fluka Advanced Workshop - Oct 07, 2010 Ericeira, Portugal M. Benmerrouche, HSE Manager http://www.lightsource.ca

More information

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS Leonardo de Holanda Mencarini 1,2, Claudio A. Federico 1,2 and Linda V. E. Caldas 1 1 Instituto de Pesquisas Energéticas e Nucleares IPEN,

More information

Basics of Synchrotron Radiation Beamlines and Detectors. Basics of synchrotron radiation X-ray optics as they apply to EXAFS experiments Detectors

Basics of Synchrotron Radiation Beamlines and Detectors. Basics of synchrotron radiation X-ray optics as they apply to EXAFS experiments Detectors Basics of Synchrotron Radiation Beamlines and Detectors Basics of synchrotron radiation X-ray optics as they apply to EXAFS experiments Detectors Important properties of Synchrotron Radiation Tunability

More information

Shielding of Ionising Radiation with the Dosimetry & Shielding Module

Shielding of Ionising Radiation with the Dosimetry & Shielding Module Shielding of Ionising Radiation with the Dosimetry & Shielding Module J. Magill Overview Biological Effects of Ionising Radiation - Absorber dose, Quality or Weighting Factor, Equivalent Dose Attenuation

More information

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.162 pissn 2508-4445, eissn 2508-4453 Secondary Neutron Dose Measurement for Proton Line Scanning

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Radiation Protection Issues at SOLARIS - Safety Assumptions and Project Status.

Radiation Protection Issues at SOLARIS - Safety Assumptions and Project Status. Radiation Protection Issues at SOLARIS - Safety Assumptions and Project Status. J. Wikłacz 1 1 National Synchrotron Radiation Centre SOLARIS, Jagiellonian University, Krakow, Poland Abstract SOLARIS, the

More information