Developing ICP-MS-based methods for regulatory purposes: One example from a contract laboratory

Size: px
Start display at page:

Download "Developing ICP-MS-based methods for regulatory purposes: One example from a contract laboratory"

Transcription

1 Developing ICP-MS-based methods for regulatory purposes: One example from a contract laboratory Douglas C. Baxter, Emma Engström, Dieke Sörlin, Ilia Rodushkin ALS Scandinavia AB, SE Luleå, Sweden Right Solutions Right Partner 1 Right Solutions Right Partner

2 U and Th isotopes in water Why (develop a method for U and Th isotopes in water)? Is there a market? Is there a standard method? Are there reference materials available? Is there legislation? What could possibly go wrong? 2 Right Solutions Right Partner

3 U and Th isotopes in water Why? Customer enquiries Is there a market? Yes, if legislation exists or customers ask for a service Is there a standard method? Yes, ISO :2016 Are there reference materials available? Yes, available from IRMM Is there legislation? Yes, 2013/51/EURATOM What could possibly go wrong? 3 Right Solutions Right Partner

4 Is there legislation? 2013/51/EURATOM 4 Right Solutions Right Partner

5 Is there legislation? 2013/51/EURATOM 240 µg/l 0,12 µg/l Detection limits not too challenging 1,6 µg/l 0,00086 µg/l 5 Right Solutions Right Partner

6 Are there reference materials available? U and Th reference materials Costs 520 EUR 1900 EUR 6 Right Solutions Right Partner

7 Is there a standard method? 7 Right Solutions Right Partner

8 Is there a standard method? Spectral interferences Isobaric (none tabulated for U or Th) Polyatomic (none tabulated for U or Th; hydrides mentioned) Note that complicated polyatomic interferences exist Taylor et al., Int. J. Mass Spectrom., 243 (2005) 71 8 Right Solutions Right Partner

9 Is there a standard method? Spectral interferences Suggested solutions (few details given) Sector field instruments (correct some polyatomic interf.) Hydride interferences cannot be resolved unless resolution is improved 5-fold 9 Right Solutions Right Partner

10 Is there a standard method? Spectral interferences Suggested solutions (few details given) Sector field instruments Mathematical corrections (none mentioned for U or Th) Does that mean that none are needed? 10 Right Solutions Right Partner

11 Is there a standard method? Spectral interferences Suggested solutions (few details given) Sector field instruments Mathematical corrections Collision and reaction cell technology No interferences? Don t need these! 11 Right Solutions Right Partner

12 Is there a standard method? Non-spectral interferences (ISO ) Suggested solutions to non-spectral interferences Matrix matching or sample dilution Dilution of samples will eliminate non-spectral interferences. Detection limits needed to fulfill requirements in legislation are easily met even after samples are diluted 10- or even 100-fold. 12 Right Solutions Right Partner

13 Is there a standard method? Non-spectral interferences (ISO ) Suggested solutions to non-spectral interferences Matrix matching or sample dilution Internal standards 18 Avoid internal standards that are present in the sample or that may cause spectral interferences. 204 Tl is interfered by 204 Pb and 204 Hg. 13 Right Solutions Right Partner

14 Is there a standard method? Non-spectral interferences (ISO ) Suggested solutions to non-spectral interferences Matrix matching or sample dilution Internal standards Inclusion of a suitable internal standard in the method will allow correction for non-spectral interferences and correction for instrumental drift. 14 Right Solutions Right Partner

15 Is there a standard method? Non-spectral interferences (ISO ) Suggested solutions to non-spectral interferences Matrix matching or sample dilution Internal standards Plasma condition optimization Optimize instrumental conditions during method validation stage. Study effects of e.g. nebulizer gas flow on U and Th isotopes as well as on the chosen internal standard in both the standards and in real samples. Vanhaecke et al., J. Anal. At. Spectrom., 8 (1993) Right Solutions Right Partner

16 Is there a standard method? Non-spectral interferences (ISO ) Suggested solutions to non-spectral interferences Matrix matching or sample dilution Internal standards Plasma condition optimization 16 Right Solutions Right Partner

17 Is there a standard method? Non-spectral interferences (ISO ) Suggested solutions to non-spectral interferences Matrix matching or sample dilution Internal standards Plasma condition optimization Standard addition technique Should never be needed if the method: (i) is based on correctly-optimized instrumental conditions; (ii) includes a suitable internal standard. Remember that standard additions will not correct for spectral interferences. 17 Right Solutions Right Partner

18 Is there a standard method? Non-spectral interferences (ISO ) Suggested solutions to non-spectral interferences Matrix matching or sample dilution Internal standards Plasma condition optimization Standard addition technique Isotope dilution (costs, purity issues) 18 Right Solutions Right Partner

19 What could possibly go wrong? 2013/51/EURATOM Only mentions 234 U and 238 U Customers may require 235 U and even 236 U, as well 19 Right Solutions Right Partner

20 What could possibly go wrong? The reference materials Price does not include delivery Many carriers will not handle radioactive material Register materials with national radiation safety authority Complete inventory change documents Log all uses of material Be prepared for radiation safety authority audit 20 Right Solutions Right Partner

21 What could possibly go wrong? Look into other resources E.g. ref. [15] in ISO What is abundance sensitivity? Pourcelot et al., J. Environ. Monitor., 13 (2011) Right Solutions Right Partner

22 What could possibly go wrong? Google abundance sensitivity icp ms 22 Right Solutions Right Partner

23 What could possibly go wrong? The standard ISO No mention of abundance sensitivity ( 234 U, 236 U) No info on how to correct for hydrides ( 235 U 1 H on 236 U) Mathematical correction for abundance sensitivity and hydrides is needed Abundance sensitivity correction requires measurement of tailing on low- and highmass sides of a single isotope. Hydride correction is element specific. 23 Right Solutions Right Partner

24 What could possibly go wrong? Everything Remember to be critical Be prepared for paperwork 24 Right Solutions Right Partner

NKS-B ICP User Seminar Program Overview

NKS-B ICP User Seminar Program Overview Program Overview Monday, 25th September Tuesday, 26th September Wednesday, 27th Sept. H. H. Koch, building 112 H. H. Koch, building 112 Buildings 202 and 204 8:00 Registration Radioanalytical method development

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Application Note Consumer products Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Effective removal of tungsten-based interferences on five Hg isotopes using MS/MS Authors

More information

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment : ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Analytical Approaches to Arsenic

Analytical Approaches to Arsenic Analytical Approaches to Arsenic (and Selenium) Analysis in Foods Brian Jackson Trace Element Analysis Core Dartmouth College Arsenic in food Arsenic levels in rice and rice products FDA data WHO data

More information

Sample Analysis Design PART III

Sample Analysis Design PART III Sample Analysis Design PART III Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode Lead isotope analysis: Removal of Hg isobaric interference from using ICP-QQQ in MS/MS mode Application note Authors Geochemistry and isotope analysis Glenn Woods Agilent Technologies, LDA UK Ltd., Stockport,

More information

Lecture 7: Atomic Spectroscopy

Lecture 7: Atomic Spectroscopy Lecture 7: Atomic Spectroscopy 1 Atomic spectroscopy The wavelengths of absorbance and emission from atoms in the gas phase are characteristic of atomic orbitals. 2 In the lowest energy transition, the

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,

More information

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Application note Petrochemical Author Lieve Balcaen, Frank Vanhaecke 1, Glenn Woods 2, Martín

More information

Atomic Absorption & Atomic Fluorescence Spectrometry

Atomic Absorption & Atomic Fluorescence Spectrometry Atomic Absorption & Atomic Fluorescence Spectrometry Sample Atomization Atomic Absorption (AA) Atomic Fluorescence (AF) - Both AA and AF require a light source - Like Molecular Absorption & Fluorescence,

More information

Very recently Roussis and Proulx clearly demonstrated

Very recently Roussis and Proulx clearly demonstrated APPLICATION NOTE Deconvolution of Isobaric Interferences in Mass Spectra Juris Meija and Joseph A. Caruso Department of Chemistry, University of Cincinnati, Cincinnati, Ohio, USA The concept of isobar

More information

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS FIELD APPLICATION REPORT ICP MASS SPECTROMETRY Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS Authors: Kenneth R. Neubauer, Ph.D. Perkin Elmer Life and Analytical Sciences 710

More information

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry Zhichao Lin, Kathryn Emanuele, Stephanie Healey, and Patrick Regan Analytical Branch Winchester Engineering and

More information

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution APEX-ACM Ca Ratios Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution Abstract High resolution ICP-MS is used

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass

More information

Sources of Errors in Trace Element and Speciation Analysis

Sources of Errors in Trace Element and Speciation Analysis Sources of Errors in Trace Element and Speciation Analysis Zoltan Mester, National Research Council of Canada, Institute for National Measurement Standards Outline Definitions Sources of errors in the

More information

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software Alain Desprez Agilent Technologies 1 The Total Solution Choice of Mass Spectrometers 7800 quadrupole ICP-MS cost-effective, 3

More information

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA

More information

Sample Analysis Design PART II

Sample Analysis Design PART II Sample Analysis Design PART II Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2.

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2. JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS Deliverable 1.2.32 (June 2015) Author : Guillaume Labarraque 1 Introduction High purity

More information

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction

More information

Sample Analysis Design Polyatomic Interferences

Sample Analysis Design Polyatomic Interferences Sample Analysis Design Polyatomic Interferences More serious than isobaric interferences Result from possible, short-lived combination of atomic species in the plasma or during ion transfer Common recombinants

More information

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Petrochemical Authors Ed McCurdy & Don Potter Agilent Technologies Ltd. Lakeside Cheadle Royal

More information

Sample Analysis Design Internal Standard

Sample Analysis Design Internal Standard Internal Standard Advantages: Fluctuations are monitored in each sample/ calibration / blank Disadvantages: Assume that behavior of IS is the same as the analyte Instrumental Drift Correct for instrument

More information

Fergus Keenan Shona McSheehy Julian Wills

Fergus Keenan Shona McSheehy Julian Wills Novel QCell Technology for Inference Removal in ICP-MS - Combining Low Mass Filtration with Kinetic Energy Discrimination Fergus Keenan Shona McSheehy Julian Wills 1 The world leader in serving science

More information

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS Accurate analysis of neptunium in a uranium matrix, using ICP-QQQ with MS/MS Application note Nuclear, environmental Authors Garry Duckworth, Springfields Fuels Ltd, K, Glenn Woods, Agilent Technologies,

More information

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos

More information

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS.

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS. ICP-MS plasma 17OO High Resolution Multi-Collector ICP-MS www.nu-ins.com Plasma 1700 & Sapphire 1700 Plasma 1700: Large Geometry Multi-Collector Inductively Coupled Plasma Mass Spectrometer For more information

More information

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES DoD Environmental Data Quality Workshop, St. Louis, MO IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES Jamie Fox Brooks Applied Labs www.brooksapplied.com

More information

Agenda. Recent Advances in Detection of Heavy Metals in Foods. NQAC Organization

Agenda. Recent Advances in Detection of Heavy Metals in Foods. NQAC Organization Agenda Recent Advances in Detection of Heavy Metals in Foods NQAC Nestlé Quality Assurance Centre Supporting factors for reliable results? Daniel Hammer Nestlé Quality Assurance Centre Singapore 2...by

More information

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS APPLICATION NOTE 434 Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS Authors Introduction Marcus Manecki, Daniel Kutscher, Shona McSheehy

More information

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences

More information

Trace Analyses in Metal Matrices Using the ELAN DRC II

Trace Analyses in Metal Matrices Using the ELAN DRC II www.perkinelmer.com Trace Analyses in Metal Matrices Using the ELAN DRC II Introduction Analyses of matrices containing high levels of metals present a challenge for ICP-MS. First, the concentrations of

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

Single particle analysis using the Agilent 7700x ICP-MS

Single particle analysis using the Agilent 7700x ICP-MS Single particle analysis using the Agilent 7700x ICP-MS Application note Authors Materials Sébastien Sannac 1, Soheyl Tadjiki 2, Evelin Moldenhauer 3 1 Agilent Technologies, France 2 Postnova Analytics

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

STRONTIUM ISOTOPES AS NATURAL TRACERS IN RESERVOIR OILFIELD AND IN GROUNDWATER SYSTEMS

STRONTIUM ISOTOPES AS NATURAL TRACERS IN RESERVOIR OILFIELD AND IN GROUNDWATER SYSTEMS 009 International Nuclear Atlantic Conference - INAC 009 Rio de Janeiro, RJ, Brazil, September7 to October, 009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-994-0-8 STRONTIUM ISOTOPES AS

More information

Fractionation of the stable silicon isotopes: Analytical method developments and selected applications in geochemistry

Fractionation of the stable silicon isotopes: Analytical method developments and selected applications in geochemistry 2007:39 LICENTIATE T H E S I S Fractionation of the stable silicon isotopes: Analytical method developments and selected applications in geochemistry Emma Engström Luleå University of Technology Department

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

Product Descriptions and Offerings

Product Descriptions and Offerings Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental analysis. Laboratories in industries ranging from environmental science, oil and

More information

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS ICP-MS Environmental Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS Application Note Steve Wilbur Analysis of arsenic and selenium in seawater at trace

More information

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples n-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples Application note Authors Giuseppe Centineo 1, Jose Ángel Rodríguez Castrillón 1, Esther Muñoz

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

Intermediates. THF Solutions. Expanded Tetrahydrofuran Offer for Pharma Customers in Europe

Intermediates. THF Solutions. Expanded Tetrahydrofuran Offer for Pharma Customers in Europe Intermediates THF Solutions Expanded Tetrahydrofuran Offer for Pharma Customers in Europe BASF We create chemistry About BASF At BASF, we create chemistry and have been doing so for 150 years. Our portfolio

More information

RISK ASSESSMENT, IMPLEMENTATION AND VALIDATION OF ELEMENTAL IMPURITIES FOR ICH Q3D, USP <232><233> AND EP 5.20

RISK ASSESSMENT, IMPLEMENTATION AND VALIDATION OF ELEMENTAL IMPURITIES FOR ICH Q3D, USP <232><233> AND EP 5.20 LIFE SCIENCE I TECHNICAL BULLETIN JUNE 2015 RISK ASSESSMENT, IMPLEMENTATION AND VALIDATION OF ELEMENTAL IMPURITIES FOR ICH Q3D, USP AND EP 5.20 AUTHOR: THOMAS DEVADDER, TEAM LEADER QC AAS/ICP/PARTICLES/TGA

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Agilent 8900 Triple Quadrupole ICP-MS

Agilent 8900 Triple Quadrupole ICP-MS Agilent 8900 Triple Quadrupole ICP-MS Technical Overview Introduction Agilent is the worldwide market leader in quadrupole ICP-MS, and the only supplier of triple quadrupole ICP-MS (ICP-QQQ). Introduced

More information

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Author Application Environmental Tetsuo Kubota Agilent Technologies, Japan Introduction Regulations concerning activities such

More information

Institute of Isotopes Co., Ltd.

Institute of Isotopes Co., Ltd. Carbon-14 Labelled PESTICIDES Institute of Isotopes Co., Ltd. Konkoly Thege M. u. 29-33., Budapest, 1121 HUNGARY Tel.: +36 1 392 2577, +36 1 395 9081 Fax: +36 1 395 9247, +36 1 392 2575 commerce@izotop.hu

More information

TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS)

TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS) NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS) EXECUTIVE SUMMARY Secondary Ion Mass Spectrometry (SIMS) is used for elemental

More information

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES Technical Note 43332 Key Words Overcoming Interferences, Element Finder plug-in Introduction Interferences are common in

More information

EVIDENCE FOR THE POSSIBLE EXISTENCE OF A LONG-LIVED SUPERHEAVY NUCLEUS WITH ATOMIC MASS NUMBER A = 292 AND ATOMIC NUMBER Z = 122 IN NATURAL Th

EVIDENCE FOR THE POSSIBLE EXISTENCE OF A LONG-LIVED SUPERHEAVY NUCLEUS WITH ATOMIC MASS NUMBER A = 292 AND ATOMIC NUMBER Z = 122 IN NATURAL Th International Journal of Modern Physics E Vol. 19, No. 1 (2010) 131 140 c World Scientific Publishing Company DOI: 10.1142/S0218301310014662 EVIDENCE FOR THE POSSIBLE EXISTENCE OF A LONG-LIVED SUPERHEAVY

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,

More information

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Use of ICP-MS in analysing radioisotopes Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Inductively Coupled Plasma Mass Spectrometry (ICP-MS) History ICP-AES

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis Application note Authors Environmental, Food, Chemical Wim Proper 1, Ed McCurdy 2 and Junichi Takahashi 3 1 Eurofins Analytico,

More information

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically

More information

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

Experiment: Nuclear Chemistry 1

Experiment: Nuclear Chemistry 1 Experiment: Nuclear Chemistry 1 Introduction Radiation is all around us. There are two main types of radiation: ionizing and non-ionizing. We will focus on radioactivity or ionizing radiation (though non-ionizing

More information

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 Tatiana Trejos, M.Sc Florida International University Department of Chemistry and Biochemistry International Forensic Research Institute Outline

More information

SmartNotes. What is meant by the term interference?

SmartNotes. What is meant by the term interference? Interference Removal on ICP-OES icap 7000 Plus Series ICP-OES SmartNotes What is meant by the term interference? In ICP-OES we speak of interference when a result is biased either by other components in

More information

Advantages of ICP-QQQ for nuclear waste characterisation

Advantages of ICP-QQQ for nuclear waste characterisation NKS-B ICPM Users seminar, 2017 Advantages of ICP-QQQ for nuclear waste characterisation Phil Warwick University of Southampton GAU-Radioanalytical Laboratories A major provider of radioanalytical expertise

More information

110 Lowell Rd Hudson, NH USA Tel:

110 Lowell Rd Hudson, NH USA Tel: 110 Lowell Rd Hudson, NH 03051 USA Tel: 603-886-8400 www.teledyneleemanlabs.com Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental

More information

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED Application Note: 40852 Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED - Interference-free analysis of 29 elements in environmental

More information

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document: SPECIFICATION SPECIFICATION Inductively Coupled Plasma Mass Spectrometry System 1. Scope This specification describes the requirements for an Inductively Coupled Plasma Mass Spectrometry System ( System

More information

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium, 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Application Note Environmental China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Fast, accurate quantitative analysis of 16 elements using an Agilent 7800 ICP-MS with

More information

Low level Os isotopic measurements using multiple ion counting

Low level Os isotopic measurements using multiple ion counting APPLICATION NOTE 30355 Low level Os isotopic measurements using multiple ion counting Authors Introduction Jean Louis Birck, Delphine Limmois, Institut de Physique du Globe de Paris, Sorbonne Paris Cité,

More information

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS application Note ICP Mass Spectrometry Authors: David Bass Daniel Jones PerkinElmer, Inc. Shelton, CT USA Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS Abstract This application

More information

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology Shona McSheehy Ducos The world leader in serving science Elemental Analysis What are you trying to accomplish?

More information

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Autenticity control, provenance testing and fraud detection using mass spectrometry

Autenticity control, provenance testing and fraud detection using mass spectrometry Autenticity control, provenance testing and fraud detection using mass spectrometry Abdal-Azim Al-Terkawi Ersan Özelci Melissa-Jane Monks Narendra Lagumaddepalli Venkatareddy Seyed Mohsen Jebreiil Khadem,

More information

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis Agilent ICP-MS Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis High Sample Matrix Tolerance Superior & Simple Interference Removal Ultra Wide Linear

More information

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards SeCtiOn 5 [STOCK AND CUSTOM] Assurance Multi-Element CLP Standards Your Science is Our Passion. Assurance CLPStandards SPEX CertiPrep offers acomplete series of multi-element standards and blanks designed

More information

Chapter 5: Writing Linear Equations Study Guide (REG)

Chapter 5: Writing Linear Equations Study Guide (REG) Chapter 5: Writing Linear Equations Study Guide (REG) 5.1: Write equations of lines given slope and y intercept or two points Write the equation of the line with the given information: Ex: Slope: 0, y

More information

Introduction to Weather Data Cleaning

Introduction to Weather Data Cleaning Introduction to Weather Data Cleaning Speedwell Weather Limited An Introduction Providing weather services since 1999 Largest private-sector database of world-wide historic weather data Major provider

More information

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS APPLICATION NOTE 43326 Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS Authors Introduction Tomoko Vincent1, Simon Lofthouse2,

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS Application note Environmental Author Sarah Hill and Rebeca Santamaria-Fernandez

More information

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals Validating performance of an Agilent 7800 ICP MS for USP / & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals Authors Lindsey Whitecotton, Ed McCurdy, Craig Jones and Amir Liba

More information

Accurate Determination of TiO 2 Nanoparticles in Complex Matrices using the Agilent 8900 ICP-QQQ

Accurate Determination of TiO 2 Nanoparticles in Complex Matrices using the Agilent 8900 ICP-QQQ Accurate Determination of TiO 2 Nanoparticles in Complex Matrices using the Agilent 8900 ICP-QQQ Authors Application note Environmental, food, cosmetics, materials Michiko Yamanaka Agilent Technologies,

More information

Analytical Method Determination of airborne metal concentrations by ICP-MS Sampling using Solu-Sert filter capsules

Analytical Method Determination of airborne metal concentrations by ICP-MS Sampling using Solu-Sert filter capsules Analytical Method Determination of airborne metal concentrations by ICP-MS Sampling using Solu-Sert filter capsules Method Technical Manager Mickaël Calosso, M. Sc., Chemist Contributor to this version

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

Chemistry Radioactive Decay Neatly answer all questions completely for credit. Show all work.

Chemistry Radioactive Decay Neatly answer all questions completely for credit. Show all work. Teacher Notes Time: 90 minutes (plus 30 minutes for teacher preparation) Overview: Radioactive isotopes are unstable. All radioactive matter decays, or breaks down, in a predictable pattern. Radioactive

More information

Oxygenation Mechanism of Ions in Dynamic Reaction Cell ICP-MS

Oxygenation Mechanism of Ions in Dynamic Reaction Cell ICP-MS ANALYTICAL SCIENCES JULY 2013, VOL. 29 747 2013 The Japan Society for Analytical Chemistry Oxygenation Mechanism of Ions in Dynamic Reaction Cell ICP-MS Tomohiro NARUKAWA and Koichi CHIBA National Metrology

More information

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application Metals Analysis Authors Ed McCurdy, Glenn Woods and Don Potter Agilent

More information

ANALYSIS CERTIFICATE OF

ANALYSIS CERTIFICATE OF CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 INORGANIC VENTURES is an

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Application note Semiconductor Author Katsuo Mizobuchi Masakazu

More information