Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Size: px
Start display at page:

Download "Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:"

Transcription

1 SPECIFICATION SPECIFICATION Inductively Coupled Plasma Mass Spectrometry System 1. Scope This specification describes the requirements for an Inductively Coupled Plasma Mass Spectrometry System ( System ) for the determination of trace, minor and major elements in environmental swipe samples and related applications. The System includes a Quadrupole Inductively Coupled Plasma Mass Spectrometer and associated equipment, such as sample introduction system (HF-resistant), sample changer, water-to-water cooling equipment for interface parts and RF generator (if applicable), electrical cables, control computer, printer, software. As essential part of the system, a collision or reaction cell is required for controlled interferences removal. The System shall be with full PC control of all instrument settings and compatible accessories. Environmental Sample Laboratory of the IAEA fulfils its service to the IAEA Department of Safeguards by performing bulk and particle analysis of environmental inspection samples. The analysed samples usually contain a few nanograms of uranium (10-9 g) and femtograms to picograms (10-15 g to g) of plutonium. Mass spectrometric techniques are implemented as key methods for isotopic analysis of uranium and plutonium, including minor isotopes 234 U and 236 U, as well as for age determination of traces of nuclear materials. With account to the requirements above, the ESL intends to purchase a quadrupole Inductively Coupled Plasma Mass Spectrometry System for screening of elemental composition and isotope ratios of U and Pu in swipe samples as well as in process solutions. For these purposes a mass spectrometer is required that possesses sufficiently high absolute sensitivity, high abundance sensitivity, and capability of simultaneous determination of actinide isotopes and concomitant stable elements that might produce interferences on actinide isotopes in ICP-MS analyses. The performance requirements for the System are based on the ESL procedures for inspection sample analysis. 2. Applicable Documents Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document: International System of Units (SI). International Bureau of Weights and Measures, 8th edition, 2006; and ISO/IEC Guide 99-12:2007, International Vocabulary of Metrology Basic and General Concepts and Associated Terms, VIM. The following documents from International Union of Pure and Applied Chemistry (IUPAC) shall be used for reference for the requirements in this document:

2 IUPAC Physical Chemistry Division. Quantities, Units and Symbols in Physical Chemistry. 3rd ed. (the Green Book ). By E. R. Cohen, T. Cvitaš, J. G. Frey, B. Holmström, K. Kuchitsu, R. Marquardt, I. Mills, F. Pavese, M. Quack, J. Stohner, H. L. Strauss, M. Takami, A. J. Thor. RSC Publications, Cambridge, UK (2007); IUPAC. Nomenclature of Inorganic Chemistry. by G. J. Leigh. Blackwell, Oxford (1990); IUPAC. Nomenclature of Inorganic Chemistry. IUPAC Recommendations by N. G. Connelly, T. Damhus, R. M. Hartshorn, A. T. Hutton. RSC Publishing, Cambridge, UK (2005); and IUPAC. Compendium of Chemical Terminology, 2nd ed. (the Gold Book ). By A. D. McNaught and A.Wilkinson. Blackwell Scientific Publications, Oxford (1997). In the event of conflict between the documents listed above and the content of this Specification, the content of this Specification shall take precedence to the extent of the conflict. 3. Definitions, Acronyms, and Abbreviations The following definitions, acronyms, and abbreviations shall apply throughout this Specification unless defined otherwise hereinafter: ICP-MS shall mean Inductively Coupled Plasma Mass Spectrometer ; Cps shall mean count per second ; m/z shall mean mass to charge ratio ; and Standard prefixes are used accordingly to the International System of Units. 4. Requirements 4.1. Functional and Performance Requirements The System shall meet the following functional and performance requirements: Detection limits calculated as three times the standard deviation of the blank and peak hopping at 1-point per mass, measured in diluted nitric acid, e.g., in 1% HNO3 shall be as defined in the table below: Element/Isotope Detection limits, ng L Fe 3 or smaller 115 In 0.1 or smaller 238 U 0.1 or smaller Sensitivity measured with a low-flow nebuliser at a sample uptake rate of 0.1 ml min -1 to 0.2 ml min -1 shall be as defined in the table below: Element/Isotope Sensitivity, cps (mg/l) -1

3 115 In greater than 4 x U greater than 4 x Ratio of oxides and doubly-charged species measured under identical operating conditions that are used to achieve sensitivity and detection-limit specifications shall be as defined the table below: Element Intensity Ratio, % 140 CeO+/ 140 Ce+ smaller than Ba++/ 137 Ba+ smaller than Background signal measured at m/z = 220 shall be smaller than 0.5 cps Short-term precision defined as the relative standard deviation (% RSD) and measured in 1 µg L -1 multi-element solution, for instance 24 Mg, 63 Cu, 114 Cd and 208 Pb, without internal standardization shall be smaller than 3% RSD over 20 minutes Long-term stability defined as the relative standard deviation (% RSD) and measured in 1 µg L -1 multi-element solution, for instance 24 Mg, 63 Cu, 114 Cd and 208 Pb, without internal standardization shall be smaller than 4% RSD over 2 hours Isotope-ratio precision defined, for example, for the isotope ratio of 107 Ag/ 109 Ag using a 25 µg L -1 solution shall be smaller than 0.1% RSD Abundance sensitivity defined as the intensity of a given isotope at spectral peak maximum, relative to the intensity of that isotope at 1 amu lower and at 1 amu higher than the mass position corresponding to peak maximum and measured at 238 U or 232 Th shall be better than 5.0 x 10-7 at low mass side of peak and better than 1.0 x 10-7 at high mass side of peak Mass range shall be amu or wider Mass calibration stability measured, e.g., using a 1 µg L -1 multi-element solution containing 7 Li, 24 Mg, 115 In, and 238 U shall be better than 0.05 amu over 8 hours of continuous operation Dynamic range shall be 9 orders of magnitude or larger Mass resolution M/ M measured with the 5% peak width definition according to IUPAC Compendium of Chemical Terminology using, e.g., 238 U will be larger than 300.

4 4.2. Technical Requirements The System shall meet the following technical requirements: The System shall be enabled for automated unattended plasma start up and shutdown; The System shall be enabled for true multitasking, allowing users to reprocess or review data from other worksheets while data is being collected for another application; The System shall be enabled for full PC control of plasma ignition, XYZ plasma position, nebuliser, plasma and auxiliary gases, gate valve, vacuum system, RF power, collision or reaction cell setup, detector setup, mass calibration, resolution, performance tests and peristaltic pump speed; The System shall be enabled full post run data editing with automatic redisplay of the results; All sample introduction equipment shall be HF-resistant; and The System shall be compatible with other sample introduction systems such as Aridus and APEX Q Essential Accessories The System shall include the following essential accessories: Collision or reaction cell that shall be installed for controlled interferences removal; Low-flow nebuliser and temperature-controlled (Peltier cooled) spray chamber that shall be installed; and Sample changer that shall be enabled for full PC control sample changer and that shall have racks for sample vials ranging from 2 ml up to 50 ml and that shall have a rinsing unit for the probe. The sample changer shall have a capacity for simultaneous handling of at least sixty 15 ml vials and at least twenty 50 ml vials. The sample changer shall have two rinse positions equipped with pumps. The ssample changer shall be acid

5 resistant and it shall not require any rinsing with water to remove diluted acid residue from the sample changer before the shutdown.

6 4.4. Spare parts and consumables The System shall include sufficient spare parts and consumables for two years operation such as, but not limited to, the following items and quantities: 5. Marking - Three (3) spare plasma torches; - Three (3) spare injectors ; - Three (3) skimmer cones; - Three (3) sampler cones; - Three (3) spare nebulizers; - One (1) spare temperature-controlled (Peltier cooled) spray chamber; and - appropriate tube sets and all necessary fittings for gas supply and sample introduction systems. The System shall have all safety markings in English language. 6. Packing The Supplier shall suitably pack the System for delivery into clean laboratory of ESL. 7. Quality Requirements The System shall be manufactured, shipped and installed in accordance with the ISO 9000:2000 quality assurance system or an equivalent quality assurance system. 8. Testing and Acceptance The System, prior to shipment, shall be tested for conformance of the System with manufacturer s performance specifications and the minimum requirements specified herein. The System, after installation, shall be tested by Contractor with the IAEA to demonstrate that the performance meets the manufacturer s performance specifications and the minimum requirements specified herein as determined by the IAEA. This on-site acceptance testing of the System shall include the following: (a) The Supplier shall demonstrate to the IAEA the performance for each of the requirements specified in Section 4 above;

7 (b) The Supplier shall demonstrate: Fully automatic operation overnight; Fully automatic and safe shut down of instrument if argon pressure becomes too low; and Fully automatic closing of analyser valve in case of power loss; (c) The Supplier shall demonstrate to the IAEA that for the software delivered with the System it shall be possible to export all of the information associated with an analysis in a single export operation including the following: a. The time of the measurement as well as the analysis time of the data reduction and export; b. All individual cycle data (selected peak intensities as well as ratios); and c. Overall summary data; The results of the on-site acceptance testing of the System shall be documented by the Supplier in an acceptance protocol, which will be signed by IAEA. 9. Installation and Training 9.1. The Supplier shall install the System in the Environmental Sample Laboratory in Seibersdorf, Austria The Supplier shall provide three-days applications training for up to three IAEA operators on the System s hardware and software and in addition one day of electronics diagnosis and repair training for up to two IAEA engineers. Both trainings shall be provided immediately after the installation of the System. 10. Deliverable Data Items The Supplier shall provide, in both hard copy and electronic format, manuals for operation, maintenance, and software use for the System. All documentation shall be in the English language. 11. Warranty Period The Supplier shall provide two-year warranty for the System.

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Thermo Scientific icap TQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos

More information

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment : ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information

Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER

Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL 001.847.913.0777 for Certified, Refurbished Lab Equipment Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER TECHNICAL

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Application Note - AN1508 Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Introduction Page 1 The combination of hydride generation with

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction

More information

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS

More information

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) By Arianne Bazilio & Jacob Weinrich December 2012 Contents Introduction... 2 Sample Introduction... 3 Torch... 4 Interface... 6

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Use of ICP-MS in analysing radioisotopes Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Inductively Coupled Plasma Mass Spectrometry (ICP-MS) History ICP-AES

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009] 2 Vol. 36(1) [Jan. Feb. 2009] BRIEFING h233i Elemental Impurities Procedures. This proposed new general test chapter is the second of two being developed to replace the general test chapter Heavy Metals

More information

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application Environmental Author Neal Julien Midwest Research Institute

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

Analysis Repeatability of Trace and Major Elements in a Water Sample

Analysis Repeatability of Trace and Major Elements in a Water Sample Analysis Repeatability of Trace and Major Elements in a Water Sample Agnès COSNIER HORIBA Scientific Longjumeau, France Keywords: environment Elements: Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li,

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords

More information

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Petrochemical Authors Ed McCurdy & Don Potter Agilent Technologies Ltd. Lakeside Cheadle Royal

More information

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction

More information

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis Agilent ICP-MS Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis High Sample Matrix Tolerance Superior & Simple Interference Removal Ultra Wide Linear

More information

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES V. MAIOROV, A. CIURAPINSKI, W. RAAB and V. JANSTA Safeguards Analytical Laboratory, International Atomic Energy

More information

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,

More information

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1 Vol. 36(1) [Jan. Feb. 2010] 1 Page 1 of 23 Time:10:03 Date:10/14/09 Instance: g:/pf/production/final PF36(1)/m5193.xml Template:s:/Pf/Template/PFRedesign/Pf353/PFR-pf-server-2009.3f BRIEFING h233i Elemental

More information

Environmental Sample Analysis Advances and Future Trends

Environmental Sample Analysis Advances and Future Trends Environmental Sample Analysis Advances and Future Trends D.Donohue Office of Safeguards Analytical Services Environmental Sample Laboratory International Atomic Energy Agency Contents Introduction Bulk

More information

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which

More information

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS Determination of Total Bromine and Iodine Page 2 of 5 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes an inductively coupled plasma optical emission spectrometry method

More information

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer Maintenance with OQ Operational Qualification OQ within the scope of a maintenance for the Atomic Absorption Spectrometer ZEEnit 700 Contents 1. GENERAL INFORMATION...4 2. DEFINITIONS...4 2.1 Validation...4

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur Introduction What is ICP-MS? Inductively Coupled Plasma Mass Spectrometry Mass spectrometry

More information

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry 1 IAEA-CN-184/168 Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry R. Williams Lawrence Livermore National Laboratory Livermore, California U.S.A. williams141@llnl.gov

More information

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode Lead isotope analysis: Removal of Hg isobaric interference from using ICP-QQQ in MS/MS mode Application note Authors Geochemistry and isotope analysis Glenn Woods Agilent Technologies, LDA UK Ltd., Stockport,

More information

The IUPAC Green Book - Units Dictators Source Book?

The IUPAC Green Book - Units Dictators Source Book? The IUPAC Green Book - Units Dictators Source Book? Prof. Dr. Jürgen Stohner FRSC ZHAW Zürich University of Applied Sciences, Institute of Chemistry and Biological Chemistry CH 8820 Wädenswil, Switzerland

More information

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover

More information

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High

More information

DISCLAIMER: This method:

DISCLAIMER: This method: Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered

More information

Atomic Spectroscopy AA/ICP/ICPMS:

Atomic Spectroscopy AA/ICP/ICPMS: Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics

More information

Trace Analyses in Metal Matrices Using the ELAN DRC II

Trace Analyses in Metal Matrices Using the ELAN DRC II www.perkinelmer.com Trace Analyses in Metal Matrices Using the ELAN DRC II Introduction Analyses of matrices containing high levels of metals present a challenge for ICP-MS. First, the concentrations of

More information

ICP-MS. High Resolution ICP-MS.

ICP-MS. High Resolution ICP-MS. ICP-MS attom ES High Resolution ICP-MS www.nu-ins.com Attom ES Enhanced Sensitivity Enhanced Speed Enhanced Selectivity Enhanced Software The Attom ES from Nu Instruments is a double focussing inductively

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter 1/5 Specification Hidex 300 SL-SLL 100-240V AC and 24 V DC. Features Model (425-020) Super Low Level (SLL) Automatic TDCR Liquid Scintillation Counter with Guard Detector and Low Level PMTs General Description

More information

á233ñ ELEMENTAL IMPURITIES PROCEDURES

á233ñ ELEMENTAL IMPURITIES PROCEDURES Second Supplement to USP 38 NF 33 Chemical Tests / á233ñ Elemental Impurities Procedures 1 á233ñ ELEMENTAL IMPURITIES PROCEDURES INTRODUCTION This chapter describes two analytical procedures (Procedures

More information

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces

More information

Automation of the radiochemical procedures for the sequential separation of radionuclides

Automation of the radiochemical procedures for the sequential separation of radionuclides LSC2017 - An International Conference on Advances in Liquid Scintillation Spectrometry, Copenhagen Denmark, 1 5 May 2017 Automation of the radiochemical procedures for the sequential separation of radionuclides

More information

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Application Note Semiconductor Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Japan Introduction Hydrochloric

More information

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software V5.2 Plutonium and Uranium Isotopic Analysis Software Advanced Isotopic Ratio Analysis Software for HPGe Gamma-Ray Spectra Analyzes Pu, and a wide variety of heterogeneous samples containing Pu, Am, U,

More information

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical

More information

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Automated Determination of PPQ Levels of Thorium in High Purity Copper PPQ Levels of Th in High Purity Cu TRUFAST ICPMS Automated Determination of PPQ Levels of Thorium in High Purity Copper Using the ESI TRUFAST System and ICPMS Detection By Nathan Saetveit, PhD and Dan

More information

Single particle analysis using the Agilent 7700x ICP-MS

Single particle analysis using the Agilent 7700x ICP-MS Single particle analysis using the Agilent 7700x ICP-MS Application note Authors Materials Sébastien Sannac 1, Soheyl Tadjiki 2, Evelin Moldenhauer 3 1 Agilent Technologies, France 2 Postnova Analytics

More information

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution APEX-ACM Ca Ratios Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution Abstract High resolution ICP-MS is used

More information

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS APPLICATION NOTE 43326 Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS Authors Introduction Tomoko Vincent1, Simon Lofthouse2,

More information

ICP-OES DETERMINATION OF IRON. Introduction:

ICP-OES DETERMINATION OF IRON. Introduction: ICP-OES DETERMINATION OF IRON. Introduction: Silver impregnated cloth has bene used in bandages and dressings to help kill bacteria associated with a wound. Cloth samples are to anaysed for silver and

More information

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials Thermo Scientific ICP-MS solutions for the semiconductor industry Maximize wafer yields with ultralow elemental detection in chemicals and materials Trace impurity analysis in chemicals an used in the

More information

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA

More information

The 30-Minute Guide to ICP-MS

The 30-Minute Guide to ICP-MS TECHNICAL NOTE The 30-Minute Guide to ICP-MS ICP - Mass Spectrometry A Worthy Member of the Atomic Spectroscopy Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been

More information

OptiMass 9500 ICP-oTOFMS. Orthogonal Time-Of-Flight The World s Fastest Benchtop ICP Mass Spectrometer. SCIENTIFIC EQUIPMENT

OptiMass 9500 ICP-oTOFMS. Orthogonal Time-Of-Flight The World s Fastest Benchtop ICP Mass Spectrometer. SCIENTIFIC EQUIPMENT OptiMass 9500 ICP-oTOFMS ogy for a nol Se e World Sensitive ch i ti v ns Te Orthogonal Time-Of-Flight The World s Fastest Benchtop ICP Mass Spectrometer SCIENTIFIC EQUIPMENT www.gbcsci.com ISO 9001 Quality

More information

Robust ICP-MS with ease of use and high productivity for routine laboratories

Robust ICP-MS with ease of use and high productivity for routine laboratories PRODUCTION SPECIFICATIONS Thermo Scientific icap RQ ICP-MS with XS Robust ICP-MS with ease of use and high productivity for routine laboratories Benefits User-friendly operation. Simple method development

More information

Spectrometric Methods of Analysis. OCN 633 Fall 2013

Spectrometric Methods of Analysis. OCN 633 Fall 2013 Spectrometric Methods of Analysis OCN 633 Fall 2013 Plasma Emission and Plasma Mass Spectroscopy Two fields of elemental analysis undergoing the most study Myriad analytical applications Three categories

More information

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains

More information

Fergus Keenan Shona McSheehy Julian Wills

Fergus Keenan Shona McSheehy Julian Wills Novel QCell Technology for Inference Removal in ICP-MS - Combining Low Mass Filtration with Kinetic Energy Discrimination Fergus Keenan Shona McSheehy Julian Wills 1 The world leader in serving science

More information

Alpha spectrometry systems. A users perspective. George Ham. 26 th May Date Month Year

Alpha spectrometry systems. A users perspective. George Ham. 26 th May Date Month Year Alpha spectrometry systems A users perspective George Ham Date Month Year 26 th May 2005 Centre for Radiation, Chemical and Environmental Hazards Radiation Protection Division formerly the National Radiological

More information

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2.

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2. JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS Deliverable 1.2.32 (June 2015) Author : Guillaume Labarraque 1 Introduction High purity

More information

Method development for analysis of single hot particles in Safeguards swipe samples

Method development for analysis of single hot particles in Safeguards swipe samples 1 IAEA-CN-184/177 Method development for analysis of single hot particles in Safeguards swipe samples Zs. Mácsik 1, N. Vajda 2, É. Széles 1, R. Katona 1 1 Institute of Isotopes, Hungarian Academy of Sciences,

More information

Low level Os isotopic measurements using multiple ion counting

Low level Os isotopic measurements using multiple ion counting APPLICATION NOTE 30355 Low level Os isotopic measurements using multiple ion counting Authors Introduction Jean Louis Birck, Delphine Limmois, Institut de Physique du Globe de Paris, Sorbonne Paris Cité,

More information

On the use of ICP-MS for measuring plutonium in urine

On the use of ICP-MS for measuring plutonium in urine Journal of Radioanalytical and Nuclear Chemistry, Vol. 243, No. 2 (2000) 397 401 On the use of ICP-MS for measuring plutonium in urine N. Baglan, 1 C. Cossonnet, 1 P. Pitet, 1 D. Cavadore, 2 L. Exmelin,

More information

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS) Basic and Application Inductively coupled plasma mass spectrometry (ICPMS) 2 What is ICP MS Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry which is capable of detecting

More information

IAEA-CN-184/159. Environmental Sample Analysis Advances and Future Trends. D. Donohue

IAEA-CN-184/159. Environmental Sample Analysis Advances and Future Trends. D. Donohue Environmental Sample Analysis Advances and Future Trends D. Donohue Department of Safeguards International Atomic Energy Agency Vienna, Austria Email D.Donohue@iaea.org IAEA-CN-184/159 Abstract. The environmental

More information

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES The National Association of Testing Authorities (NATA) requirements for accreditation have undergone

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application Analytical/Inorganic MM800 (a) Ion Exchange and ICP/MS of Uranium in Water 1.0 Scope and Application This procedure can be used to determine U concentration or isotopic-ratio composition in groundwater

More information

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid Application Semiconductor Authors Abe G. Gutiérrez Elemental Scientific 2440 Cuming St Omaha, NE 68131 USA abe@icpms.com

More information

PRINCIPLE OF ICP- AES

PRINCIPLE OF ICP- AES INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively

More information

NICKEL-63/59 IN WATER

NICKEL-63/59 IN WATER Analytical Procedure NICKEL-63/59 IN WATER 1. SCOPE 1.1. This is a method for the separation and measurement of nickel- 63/59 in water samples. 1.2. This method does not address all aspects of safety,

More information

STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION. Price: $5.00. Revised March 31, 1997

STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION. Price: $5.00. Revised March 31, 1997 STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION Revised March 31, 1997 Price: $5.00 a. an arrangement where the water pipe to be injected with chlorine enters

More information

CHAPTER 4: ANALYTICAL INSTRUMENTATION

CHAPTER 4: ANALYTICAL INSTRUMENTATION CHAPTER 4: ANALYTICAL INSTRUMENTATION 4.1 INTRODUCTION In this section, a review of the analytical instrumentation used during sample preparation and analysis is presented which includes an overview of

More information

The New Agilent 7700 Series ICP-MS

The New Agilent 7700 Series ICP-MS The New Agilent 7700 Series ICP-MS the best ICP-MS just got better! Presenter: Uwe Noetzel and Jerome Darrouzes ICP-MS Product Specialist Life Sciences and Chemical Analysis Global Footprint Liquid Separation

More information