Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis
|
|
- Augustine Miles
- 6 years ago
- Views:
Transcription
1 Agilent ICP-MS Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis High Sample Matrix Tolerance Superior & Simple Interference Removal Ultra Wide Linear Range for Trace and Percent Level Elemental Concentration Determinations
2 Agilent products and solutions are intended to be used for cannabis quality control and safety testing in laboratories where such use is permitted under state/country law.
3 Benefits of ICP-MS Multi-elemental technique High sensitivity, ppb/ppt (even ppq!) for most elements Short analysis time (~ 3-4 min) Extremely linear with wide dynamic range Minimum number of interferences High Productivity Page 3
4 Agilent s History of Innovation in ICP-MS 1994 to Series introduced - World's first benchtop system. Hyperbolic profile quad, motorized torch XYZ, cool plasma 1998 First real time ICP-MS chromatographic software PlasmaChrom. T- mode reaction interface introduced Series 100, 200 & 300 introduced: 1st applications-specific ICP-MS Agilent 7500 Series introduced a, 7500i and 7500s - the next generation in ICP-MS instrumentation. 9 orders detector range 2001 Agilent 7500c launched 1 st generation ORS for high matrix samples New digital generators and LAN control introduced. First commercial GC- ICP-MS interface Agilent 7500cs launched 2 nd generation ORS for high purity semicon samples Agilent 7500ce launched 2 nd generation ORS for high matrix samples Low flow cell gas MFC s for Xe NH 3, O 2, etc added to 7500ce/cs Agilent acquires 100% of Agilent/Yokogawa joint venture 2007 Agilent 7500cx introduced: He only mode ICP-MS 2008 High Matrix Interface developed enables 2% TDS samples to be run by ICP-MS 2009 Agilent 7700 Series introduced replaces 7500 Series. MassHunter Software introduced - common platform with other Agilent MS. ISIS-DS Discrete sampling system, for ultra high throughput analysis 2012 January Agilent 8800 Introduced The world s first triple quad ICP-MS 2013 December Over ICP-MS units sold worldwide 2014 January Introduction of the 7900 ICP-MS Agilent 8800 ICP-QQQ Agilent 7500 Series Agilent 4500 Series Agilent 7700 Series Agilent 7900 Series Page 4
5 Current Agilent ICP-MS Portfolio Agilent 7900 Agilent 7800 Agilent 8900 (Industries 1 st Triple Quad ICP-MS!) Page 5
6 Why is Metals Analysis Important for Cannabis? Assure no toxic metals are present Product safety for Medical & Recreational Big Four Metals As, Cd, Pb, & Hg Determination of full elemental suite Additional elements necessary for horticulture aspects soil and potential contaminants Not just about the plant Irrigation Water Fertilizers/Added Nutrients Soil Ingredients used for Edibles Oil Extracts/Concentrates Delivery Devices Vaporizers
7 ICPMS with Enhanced ORS 4 Technology 2. The smallest droplets pass through the spray chamber and into the ion source - the plasma 3. The sample is desolvated and ionized in the plasma 6. The quadrupole mass spectrometer separates ions based on their mass to charge ratio. The selected ions continue on to the detector 1. The liquid sample is mixed with argon gas by the nebulizer to form an aerosol.. 4. Ions are extracted from the plasma by Sampler /Skimmer cones and extraction lenses in the interface region 5. Ion beam is deflected and then focused in Octopole. Gases can be used to remove interferences 7. Ions are measured using a discrete dynode detector providing 10(7800)-11(7900) orders of linear dynamic range. Page 7
8 3 Key Benefits/Components of the Agilent ICP-MS 1) Matrix Tolerance 2) Interference Removal 3) Dynamic Range High Matrix Introduction System (HMI) Handles tough sample matrices better than any other ICP-MS. Reduces sample prep time and error Better long-term stability Octopole Reaction System He Mode He Mode effectively removes common polyatomic interferences in the samples while maintaining sensitivity. Easy to use - He collision mode only Effective for a wide range of sample types Accurate measurements Wider Dynamic Range 10(7800) 11(7900) orders dynamic range: 0.1ppt(DL) to 1,000ppm Simplifies method development Easy sample prep. Improves productivity Page 8 July 11, 2017
9 3 Key Benefits of Agilent ICP-MS 1) Matrix Tolerance 2) Interference Removal 3) Dynamic Range High Matrix Introduction System (HMI) Handles tough sample matrices better than any other ICP-MS. Reduces sample prep time and error Better long-term stability Octopole Reaction System He Mode He Mode effectively removes common polyatomic interferences in the samples Easy to use - He collision mode only Effective for a wide range of sample types Accurate measurements Wider Dynamic Range 10 orders dynamic range: 0.1ppt(DL) to 1,000ppm Simplifies method development Easy sample prep. Improves productivity Page 9 July 11, 2017
10 Introducing the Ground-Breaking Agilent UHMI(7900) UHMI gas port UHMI (ultra High Matrix Interface) much more than just a simple T-piece UHMI uses optimized gas mixing geometry and sophisticated plasma/gasflow tuning algorithm to set reproducible conditions for predictable aerosol dilution rate Page 10 July 11, 2017
11 The Big Four Spiked at Different Salt Content NaCl Amount 75 As [ 25 ppb ] 114 Cd [ 50 ppb ] 208 Pb [ 50 ppb ] 201 Hg [ 1 ppb ] 0% % % % % % % % Average % RSD 5% 4% 1% 6% % Recovered 100% 97% 100% 93% 0.5g / 100ml = 0.5% 1g / 100ml = 0.5% 1.5g / 100ml = 0.5% 2g / 100ml = 0.5% 5g / 100ml = 0.5% 10g / 100ml = 0.5% 25g / 100ml = 25% 11
12 High TDS Will Result in Ionization Suppression Ionization Suppression in 0.3% NaCl Sc-45 Cr-52 Fe-56 Zn-66 Mo-95 In-115 1% HNO3 0.3%NaCl Confidentiality Label July 11,
13 Cerium Oxide Formation in the Plasma The tendency to form CeO in the plasma is expressed as a ratio of the cps at 156 to the cps at 140 (CeO/Ce ratio) Ce 16 O 140 Ce The CeO complex will break apart under hottler plasma conditions. A hotter plasma will have a lower CeO/Ce ratio. Page 13
14 The Importance of Oxides (CeO/Ce) CeO/Ce ratio is a performance criterion that all vendors cite, however the actual relevance of this is often overlooked. - Lower CeO/Ce ratio is the result of a hotter plasma. - A hotter plasma is the result of more electrons (joule heating) in the plasma Important factors affecting 140 Ce 16 O / 140 Ce : -Vapor loading in plasma (more water, more available Oxygen). -Plasma temperature (Higher temperature plasma more readily breaks down CeO complex). - A higher temperature plasma is the result of more electrons in the plasma. Lower CeO/Ce Ratio More CeO/Ce is expressed as the ratio of counts at mass 156 (CeO) divided by the counts at mass 140 (Ce). electrons available for ionization!!
15 Ionization Suppression as a Function of CeO With With 1.0%, 1.7%, suppression only at 10-25% 20-50%
16 Sample Introduction HMI Sample Introduction is: Low-flow (typically 0.15mL/min) Temperature stabilized (Peltier cooled spray chamber) Now features HMI(7800 shown) UHMI(7900), allowing auto setup of plasma conditions and much higher matrix tolerance
17 3 Key Benefits of Agilent ICP-MS 1) Matrix Tolerance 2) Interference Removal 3) Dynamic Range High Matrix Introduction System (HMI) Handles tough sample matrices better than any other ICP-MS. Reduces sample prep time and error Better long-term stability Octopole Reaction System He Mode He Mode effectively removes common polyatomic interferences in the samples Easy to use - He collision mode only Effective for a wide range of sample types Accurate measurements Wider Dynamic Range 10 orders dynamic range: 0.1ppt(DL) to 1,000ppm Simplifies method development Easy sample prep. Improves productivity Page 17 July 11, 2017
18 Polyatomic Interference Formation Ar 35 Cl 75 As Page 18
19 Troublesome Region of the Periodic Table: Polyatomics Interfernces from Ar, O, Cl, C, Na, Mg, Ca. Page 19
20 Principle of Helium Collision Mode and Kinetic Energy Discrimination (KED) Polyatomic ions Analyte ions Energy distribution of analyte and interfering polyatomic ions with the same mass Polyatomic ions Bias voltage rejects low energy (polyatomic) ions Analyte ions Energy Energy At cell entrance, analyte and polyatomic ion energies overlap. Energy spread of both groups of ions is narrow, due to ShieldTorch System Cell Entrance Energy loss from each collision with a He atom is the same for analyte and polyatomic ion, but polyatomics are bigger and so collide more often Cell Exit By cell exit, ion energies no longer overlap; polyatomics are rejected using a bias voltage step. Analyte ions have enough residual energy to get over step; polyatomics don t (energy discrimination) Page 20
21 Acid Matrices and IPA in NoGas Mode (HNO 3 + HCl + H 2 SO 4 + IPA) 2E5 cps CO 2 SO, SOH ClO ArC ClO ArO, CaO SO 2, S 2, ArN 2 H, SO 2 H Unspiked 5% HNO 3 + 5% HCl + 1% H 2 SO 4 + 1% IPA Matrix Ar 2, Ca 2, ArCa, S 2 O, SO 3 Unspiked Matrix ALL peaks are due to polyatomic interferences ArN ArCl CO 2 H ArC S 2, SO 2 Cl 2 ArOH, CaOH ArCO, ArCN ArS, Cl 2 Ar 2 ArCl Br, Ar 2 H Br, Ar 2 H SN ClO, NaS CaO, NaCl CaO ClN 2, CaOH, ArNa CaO, NaClH NaCl S 2, SO Mass No Gas Mode Cl 2 H ArS, Cl 2 ArS Ar 2 Page 21 July 11, 2017
22 Acid Matrices and IPA in He Mode (HNO 3 + HCl + H 2 SO 4 + IPA) 2E5 cps Unspiked 5% HNO 3 + 5% HCl + 1% H 2 SO 4 + 1% IPA Matrix ALL polyatomic interferences are removed in He Mode (same cell conditions) He Mode Page 22
23 Acid Matrices and IPA in He Mode + 10ppb spike Consistent sensitivity and perfect template match for all elements 2E5 cps 10ppb Spike in 5% HNO 3 + 5% HCl + 1% H 2 SO 4 + 1% IPA Matrix Consistent high sensitivity for all isotopes of all elements in He Mode Good signal for all spike elements in 10ppb spike. Perfect template fit for all elements ALL isotopes available for quantification / confirmation No residual interferences and no loss of analyte signal by reaction Mass He Mode Page 23 July 11, 2017
24 Elimination of ArO Interference on Fe-56 Pag
25 Cal 0, 1, 10 ppb Arsenic in 1%HNO3/0.5% HCl (ArCl interference on m/z 75) Page 25
26 3 Key Benefits of Agilent ICP-MS 1) Matrix Tolerance 2) Interference Removal 3) Dynamic Range High Matrix Introduction System (HMI) Handles tough sample matrices better than any other ICP-MS. Reduces sample prep time and error Better long-term stability Octopole Reaction System He Mode He Mode effectively removes common polyatomic interferences in the samples Easy to use - He collision mode only Effective for a wide range of sample types Accurate measurements Wider Dynamic Range 10 orders dynamic range: 0.1ppt(DL) to 1,000ppm Simplifies method development Easy sample prep. Improves productivity Page 26 July 11, 2017
27 High Concentration Elements and Trace Elements: Linear Dynamic Range Page 27
28 Largest Analytical Range of any ICP-MS Calibration ranges Hg (10 200ppt) NoGas Mode As ( ppt) He Mode Se ( ppt) He Mode Na ( ppm) He Mode These 4 plots were obtained under the same analytical conditions on the 7800 only the gas mode (NoGas for Hg) changed Hg As As Overall calibration range 10ppt (Hg, As, Se) to 1000 ppm (Na) in a single method - without attenuating ion transmission to increase working range 10 ppt Mercury 10 ppt Arsenic Na Typically, ICP-MS cannot measure above 200ppm Na without changing quad resolution or ion lens settings Hg Hg LOD on 7800 is about 2ppt 7800 can QUANTITATE at 10ppt! Agilent ICP-MS can do both of the above in the same run! Se Se 10 ppt Selenium Na 1000 ppm Sodium Good fit at 0.2ppm Page 28
29 7800 Nutritional Supplement Samples #1 -Major constituents can be determined at any dilution. -Trace toxics measureable at even the highest dilution. *A Single Helium mode is used across the entire mass range. Nutritional Supplements Very Similar in Composition to Edible Cannabis Products Page 29
30 7800 Nutritional Supplement Samples #2 -Concentrations across all dilutions for cations the same. -Agilent actually measures Fe at 56. -All elements using main isotope or on mass isotope. -Excellent sensitivity across all trace/toxic element dilutions (As, Cd, Pb). Page 30
31 Unique Performance of the 7800 & 7900 Better matrix tolerance than any other ICP-MS Higher plasma temperature (lower CeO/Ce ratio) under standard conditions than any other system 7800 with HMI: ~2-3% TDS & 7900 with UHMI: ~25% TDS Best performance with Helium cell gas eliminates need for reaction gases in all common applications 7800/7900 ORS 4 removes polyatomic interferences in complex matrices while maintaining excellent sensitivity. Wider dynamic range than any other quadrupole ICP-MS 7800: 10 orders of Magnitude (~1000s ppm) (3 ms Integration) 7900: 11 orders of Magnitude (~Percent) (0.1 ms Integration spnp) Page 31
32 Analysis of Cannabis - Sample Preparation 1. Sample is crushed and homogenized 2. Aliquot of sample is weighed (~ mg) 3. Addition of HNO 3 & HCl (to stabilize elements such as Hg) 4. Samples are digested using microwave digestion procedure 5. After digestion and cooling, digested material is brought to volume with Type 1 H 2 O for ICP-MS analysis. 6. Using the Agilent ICP-MS with HMI technology, samples are directly analyzed with no additional dilution required (minimizes potential for contamination from pipette tips and additional reagents)
33 Analysis of Cannabis Digestates by Agilent ICP-MS Calibration Curves (1%HNO 3 /0.5%HCl)
34 ICP-MS Results for Cannabis Digestates (ppb) Including Continuing Calibration QC and Spike Recoveries QC True Values 10.0 ug/l for As, Cd, & Pb, 1.0 ug/l for Hg Spike True Values 10.0 mg/l for As, Cd, & Pb, 0.1 mg/l for Hg
35 Agilent Mass Hunter Software Automatically Collects Quick Scan Data with Each Individual Sample Analysis. Provides elemental information for additional elements, even though they have not been included in the initial calibration standards.
36 I. Matrix Tolerance I. High Matrix Interface (HMI(7800) & UHMI(7900)) I. Aerosol dilution increases plasma temperature II. More efficient matrix breakdown and ionization efficiency III. Reduction of Polyatomics Begins in the Plasma on the Agilent II. Interference Removal (ORS 4 ) I. Low flow of non reactive gas (He), Small Volume Octopole provides high energy collisions resulting in polyatomic interference removal for multiple matrices during all analyses. II. ICP-MS Advantages Review and Summary Linear Dynamic Range (10 (7800) 11 (7900) Orders LDR) I. ICP-MS easily provides sub ppb down to ppt detection levels II. Large linear dynamic range allows for the quantitation of high concentration analytes, no dilution, single analysis, simple.
37 Potential for Future Applications New Industry, prepare for the future! HPLC ICP-MS Speciation Analysis As, Se, Cr.. GC ICP-MS Metals Determination in Organic Comp ppb/ppt Determination of P, S in pesticides
38 Page 38
Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio
Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio NEMC August 2018 Craig Jones Applications Specialist ICP-MS, ICP-MS/MS The Agilent Atomic Spectroscopy Lineup An Instrument for Any Application!
More informationApplications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry
Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS
More informationThe New Agilent 7700 Series ICP-MS
The New Agilent 7700 Series ICP-MS the best ICP-MS just got better! Presenter: Uwe Noetzel and Jerome Darrouzes ICP-MS Product Specialist Life Sciences and Chemical Analysis Global Footprint Liquid Separation
More informationEnhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS
Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,
More informationThe Agilent 7700x ICP-MS Advantage for Drinking Water Analysis
The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction
More informationMulti-Element Analysis of Cannabis using the Agilent 7800 ICP-MS
Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains
More informationSimple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS
Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,
More informationAnalysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ
Application Note Semiconductor Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Japan Introduction Hydrochloric
More informationComparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS
Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Technical Overview Abstract Inductively coupled plasma mass spectrometry
More informationDirect Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ
Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction
More informationFergus Keenan Shona McSheehy Julian Wills
Novel QCell Technology for Inference Removal in ICP-MS - Combining Low Mass Filtration with Kinetic Energy Discrimination Fergus Keenan Shona McSheehy Julian Wills 1 The world leader in serving science
More informationUnmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application
Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application Metals Analysis Authors Ed McCurdy, Glenn Woods and Don Potter Agilent
More informationDetermination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS
Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction
More informationAgilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology
Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass
More informationAgilent 7700 Series. The Smallest, Most Powerful ICP MS Ever Made
The All New Agilent 7700 Series The Smallest, Most Powerful ICP MS Ever Made Inductively Coupled Plasma Mass Spectrometry ICP MS is a fast, multi element, high sensitivity trace metals analysis technique
More informationDetermination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS
APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity
More informationTrace Analyses in Metal Matrices Using the ELAN DRC II
www.perkinelmer.com Trace Analyses in Metal Matrices Using the ELAN DRC II Introduction Analyses of matrices containing high levels of metals present a challenge for ICP-MS. First, the concentrations of
More informationRapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.
Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA
More informationDetermination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS
APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords
More informationTrace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell
Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle
More informationFundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company
Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical
More informationHandling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology
Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology Shona McSheehy Ducos The world leader in serving science Elemental Analysis What are you trying to accomplish?
More informationDetermination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode
Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki
More informationTechniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)
Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Petrochemical Authors Ed McCurdy & Don Potter Agilent Technologies Ltd. Lakeside Cheadle Royal
More informationThe ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode
The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro
More informationApplication note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS
Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS Application note Semiconductor Author Naoki Sugiyama Agilent Technologies, Tokyo,
More informationAgilent 8900 Triple Quadrupole ICP-MS
Agilent 8900 Triple Quadrupole ICP-MS Technical Overview Introduction Agilent is the worldwide market leader in quadrupole ICP-MS, and the only supplier of triple quadrupole ICP-MS (ICP-QQQ). Introduced
More informationAnalysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS
TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically
More informationKey Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company
Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences
More informationDirect Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS
Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide
More informationSimple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3
Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Author Application Environmental Tetsuo Kubota Agilent Technologies, Japan Introduction Regulations concerning activities such
More informationTotal Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS
Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos
More informationMulti-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS
Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction
More informationAgilent Technologies at TIAFT 2013
Agilent Technologies at TIAFT 2013 Analytical approaches for the measurement of trace metals in forensic samples Funchal, Madeira September 2-6 CRIMINALISTIC TOXICOLOGY DNA INVESTIGATION DOCUMENT EXAMINATION
More informationBasic and Application Inductively coupled plasma mass spectrometry (ICPMS)
Basic and Application Inductively coupled plasma mass spectrometry (ICPMS) 2 What is ICP MS Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry which is capable of detecting
More informationSemiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode
Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,
More informationAtomic Spectroscopy AA/ICP/ICPMS:
Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics
More informationThermo Scientific icap RQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry
More informationHigh Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast
High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,
More informationThermo Scientific icap TQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively
More informationDetermination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS
Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction
More informationAccurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS
Accurate analysis of neptunium in a uranium matrix, using ICP-QQQ with MS/MS Application note Nuclear, environmental Authors Garry Duckworth, Springfields Fuels Ltd, K, Glenn Woods, Agilent Technologies,
More informationAgilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent
More informationAnalysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS
ICP-MS Environmental Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS Application Note Steve Wilbur Analysis of arsenic and selenium in seawater at trace
More informationIMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES
DoD Environmental Data Quality Workshop, St. Louis, MO IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES Jamie Fox Brooks Applied Labs www.brooksapplied.com
More informationUltra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode
Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Application note Semiconductor Author Katsuo Mizobuchi Masakazu
More informationPerformance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis
Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis Application note Authors Environmental, Food, Chemical Wim Proper 1, Ed McCurdy 2 and Junichi Takahashi 3 1 Eurofins Analytico,
More informationSources of Errors in Trace Element and Speciation Analysis
Sources of Errors in Trace Element and Speciation Analysis Zoltan Mester, National Research Council of Canada, Institute for National Measurement Standards Outline Definitions Sources of errors in the
More informationChina Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS
Application Note Environmental China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Fast, accurate quantitative analysis of 16 elements using an Agilent 7800 ICP-MS with
More informationHands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment
: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it
More informationDirect Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System
Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Application Note Foods Author Glenn Woods ICP-MS Specialist Agilent Technologies UK Ltd. Lakeside Business
More informationDirect Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application
Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,
More informationRapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED
Application Note: 40852 Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED - Interference-free analysis of 29 elements in environmental
More informationLow Level Detection of Priority Inorganic Pollutants in Water. Anthony Neumann, MSc Darren Crichton, BSc
Low Level Detection of Priority Inorganic Pollutants in Water Anthony Neumann, MSc Darren Crichton, BSc Priority List of Pollutants Consists of 16 metals and/or transitional elements. Enter the human food
More informationDetermination of Trace Metals in Human Urine Using the NexION 300 ICP-MS
application Note ICP Mass Spectrometry Authors: David Bass Daniel Jones PerkinElmer, Inc. Shelton, CT USA Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS Abstract This application
More informationLead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode
Lead isotope analysis: Removal of Hg isobaric interference from using ICP-QQQ in MS/MS mode Application note Authors Geochemistry and isotope analysis Glenn Woods Agilent Technologies, LDA UK Ltd., Stockport,
More informationToday s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy
Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy
More informationSample Analysis Design PART II
Sample Analysis Design PART II Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis
More informationHigh-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)
High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,
More informationIntroduction to the New Agilent 7900
Introduction to the New Agilent 7900 Redefining ICP-MS Performance Spectroscopy Webinar February 2014 Presenter: Ed McCurdy ICP-MS Product Marketing, Agilent Technologies Agilent 7900 ICP-MS Introduction
More informationPartial Energy Level Diagrams
Partial Energy Level Diagrams 460 nm 323 nm 610 nm 330 nm 819 nm 404 nm 694 nm 671 nm 589 / 590 nm 767 / 769 nm Lithium Sodium Potassium Gas Mixtures Maximum Temperatures, C Air-Coal Gas 1825 Air-Propane
More informationInductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur
Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur Introduction What is ICP-MS? Inductively Coupled Plasma Mass Spectrometry Mass spectrometry
More informationAccurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS
APPLICATION NOTE 434 Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS Authors Introduction Marcus Manecki, Daniel Kutscher, Shona McSheehy
More informationAnalytical Approaches to Arsenic
Analytical Approaches to Arsenic (and Selenium) Analysis in Foods Brian Jackson Trace Element Analysis Core Dartmouth College Arsenic in food Arsenic levels in rice and rice products FDA data WHO data
More informationATOMIC SPECTROSCOPY APPLICATIONS IN THE CONTRACT ENVIRONMENTAL LABORATORY. Primer July 2015 Edition
ATOMIC SPECTROSCOPY APPLICATIONS IN THE CONTRACT ENVIRONMENTAL LABORATORY Primer July 2015 Edition > Search entire document Table of contents Agilent for the analysis of drinking water, waste water, soils
More informationThe 30-Minute Guide to ICP-MS
TECHNICAL NOTE The 30-Minute Guide to ICP-MS ICP - Mass Spectrometry A Worthy Member of the Atomic Spectroscopy Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been
More informationTotal elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS
APPLICATION NOTE 43326 Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS Authors Introduction Tomoko Vincent1, Simon Lofthouse2,
More informationSimplified workflows for the Thermo Scientific icap TQ ICP-MS
TECHNICAL NOTE 43401 Simplified workflows for the Thermo Scientific icap TQ ICP-MS Authors Marcus Manecki, Dhinesh Asogan Thermo Fisher Scientific Keywords Ease of use, Qtegra ISDS Software, Reaction Finder,
More informationAgilent ICP-MS Journal September 2009 Issue 39
Agilent ICP-MS Journal September 2009 Issue 39 7700 Series ICP-MS Special Issue Inside this Issue 2 Clearly Better ICP-MS: Introducing the New 7700 Series, Agilent 7700 Series Stars on YouTube 3 New ORS
More informationUltra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES
Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction
More informationSample Analysis Design PART III
Sample Analysis Design PART III Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis
More informationSpeciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry
APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos
More informationAccurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ
Application Note Consumer products Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Effective removal of tungsten-based interferences on five Hg isotopes using MS/MS Authors
More informationFebruary 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:
February 20, 2012 Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Subject: ICP-MS Report Job Number: S0CHG688 Dear Joe: Please find enclosed the procedure report for the analysis
More informationApplication. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.
Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid Application Semiconductor Authors Abe G. Gutiérrez Elemental Scientific 2440 Cuming St Omaha, NE 68131 USA abe@icpms.com
More informationFlorida State University Libraries
Florida State University Libraries Electronic Theses, Treatises and Dissertations The Graduate School 2013 Accurate and Precise Determination of Low Concentration Iron, Arsenic, Selenium, Cadmium, and
More informationQuantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS
Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS Application note Metals Analysis & Production Authors Naoki Sugiyama and Mineko Omori Agilent Technologies,
More informationReal World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.
Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology Part 2 of a 3 part series on Environmental Analysis Application Environmental Authors
More informationApplication of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs
Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs Application note Authors Food safety Kazuhiro Sakai 1, Junichi
More informationINDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS
More informationDefining quality standards for the analysis of solid samples
Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples
More informationLow level Os isotopic measurements using multiple ion counting
APPLICATION NOTE 30355 Low level Os isotopic measurements using multiple ion counting Authors Introduction Jean Louis Birck, Delphine Limmois, Institut de Physique du Globe de Paris, Sorbonne Paris Cité,
More informationICP-MS. Inductively Coupled Plasma Mass Spectrometry. A Primer
ICP-MS Inductively Coupled Plasma Mass Spectrometry A Primer Table of Contents ICP-MS Primer Section 1 Introduction to ICP-MS 1 History and Development of ICP-MS 2 Agilent Technologies - History in ICP-MS
More informationPROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]
2 Vol. 36(1) [Jan. Feb. 2009] BRIEFING h233i Elemental Impurities Procedures. This proposed new general test chapter is the second of two being developed to replace the general test chapter Heavy Metals
More informationFollowing documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:
SPECIFICATION SPECIFICATION Inductively Coupled Plasma Mass Spectrometry System 1. Scope This specification describes the requirements for an Inductively Coupled Plasma Mass Spectrometry System ( System
More informationDetermination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors
Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,
More informationAgilent 8800 Triple Quadrupole ICP-MS: Understanding oxygen reaction mode in ICP-MS/MS
Agilent 8800 Triple Quadrupole ICP-MS: Understanding oxygen reaction mode in ICP-MS/MS Agilent 8800 ICP-QQQ Technical Overview Introduction The use of oxygen as a reaction gas for the removal of interferences
More informationAnalysis of domestic sludge using the Agilent 4200 MP-AES
Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal
More informationIntroducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies
Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading
More informationPRINCIPLE OF ICP- AES
INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively
More informationSamples compliant with the WATER
ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions
More informationDISCLAIMER: This method:
Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered
More informationHigh-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application
High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application Environmental Author Neal Julien Midwest Research Institute
More informationRapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry
Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,
More informationElemental analysis of river sediment using the Agilent 4200 MP-AES
Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction
More informationChapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques
Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are
More informationLecture 7: Atomic Spectroscopy
Lecture 7: Atomic Spectroscopy 1 Atomic spectroscopy The wavelengths of absorbance and emission from atoms in the gas phase are characteristic of atomic orbitals. 2 In the lowest energy transition, the
More informationELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4
ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 Tatiana Trejos, M.Sc Florida International University Department of Chemistry and Biochemistry International Forensic Research Institute Outline
More informationOn-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples
n-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples Application note Authors Giuseppe Centineo 1, Jose Ángel Rodríguez Castrillón 1, Esther Muñoz
More informationRapid determination of five arsenic species in polished rice using HPLC-ICP-MS
Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS Application note Food safety Authors Bing Yue China National Center For Food Safety Risk Assessment, Beijing, China Juane
More information