1 So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

2 GC Columns Wall Coated Open Tubulars Liquid phase coated capillaries Internal Diameter mm Length 5m 100m Porous Layer Open Tubulars Adsorbent coated capillaries Internal Diameter mm Length 5 60m Packed Columns Tubing packed with coated packings or porous particles Internal Diameter 2 4mm Length m

3 WCOT Column Types Agilent J&W has over 50 different stationary phase offerings

4 FactorFour TM Phases VF-1ms, VF-5ms, VF-5ht, VF-5ht UltiMetal VF-17ms, VF-17ms for PAH, VF-35ms, VF-200ms,VF-Xms, VF-23ms, VF-624ms, VF-DA, VF-1301ms, VF-Pesticides, VF-1701ms, VF-WAXms

5 20+ Different Specialty Phases Specialty phases are columns that are optimized to perform a specialized GC analysis. Column Typical Application DB-624 EPA and USP volatiles DB-VRX volatiles analysis HP-VOC volatiles analysis DB EPA Method DB EPA semi-volatiles analysis DB-608 EPA Method 608 DB-1701P EPA pesticides analysis DB-MTBE total petroleum hydrocarbon (TPH) HP-PONA petroleum hydrocarbon analysis DB-HT SimDis hi-temp simulated distillation DB-ALC1 & ALC2 blood alcohol analysis HP-88 fatty acid methyl ester (FAME)

6 Select TM Column Examples Environmental applications CP-Sil 88 for dioxins, Select mineral oil, CP-Select 624 CB Chiral applications CP-Chirasil Val, CP-Chirasil-DEX CB Chemical applications CP-Volamine, CP-Select CB for MTBE, CP-PONA C8, CP-Propox, Select Silanes, CP-SimDist UltiMetal TM, CP-Lowox TM Food and Beverage applications CB-Carbowax 400, Select FAME, CP-Sil 88 for FAME, CP-FFAP

7 PLOT Column Types PLOT columns are primarily, but not exclusively, used for the analysis of gases and low boiling point solutes (i.e., boiling point of solute is at or below room temperature). Agilent J&W PLOT columns begin with the designation of GS (Gas Solid) or HP-PLOT followed by a specific name 10 stationary phases GS-OxyPLOT GS-Alumina HP-PLOT Al2O3 M HP-PLOT Al2O3 S HP-PLOT Al2O3 KCl GS-OxyPLOT: oxygenates HP-PLOT Molesieve: O2, N2, CO, Methane HP-PLOT Alumina and GS-Alumina: complex hydrocarbon gas matrices, ethylene and propylene purity, 1,4-butadiene HP-PLOT Q: freons, sulfides HP-PLOT U: C1 to C7 hydrocarbons, CO2, Polar Hydrocarbons GS-GasPro: freons, sulfurs, inorganic gases GS-CarbonPLOT: inorganic and organic gases HP-PLOT MoleSieve GS-CarbonPLOT HP-PLOT Q HP-PLOT U GS-GasPro

8 PLOT GC Columns from Varian Line Porous Polymers CP-PoraBOND Q CP-PoraBOND U CP-PoraPLOT Q CP-PoraPLOT U CP-PoraPLOT S CP-PoraPLOT Q-HT CP-PoraPLOT amines Zeolites Molsieve 5A Alumina KCL Na 2 SO4 MAPD Multi Layer CP-Lowox Porous Silica SilicaPLOT Graphatised Carbon CP-CarboPLOT P7 CP-CarboBOND Select TM permanent gases Molsieve 13x

9 Typical Capillary Column Polyimide Coating Fused Silica Stationary Phase Expanded view of capillary tubing

10 Distribution Constant (K C ) Mobile Phase Stationary Phase K C = conc. of solute in stationary phase conc. of solute in mobile phase K C formerly written as K D

11 K C and Retention Fused Silica Tubing Stationary Phase Gas Flow Stationary Phase K c => Large retention K c => Small retention

12 K C and Peak Width Time of Elution Fused Silica Tubing Stationary Phase Gas Flow Stationary Phase K c => Large K c => Small

13 Resolution Getting One Peak Away From Another

14 Separation vs. Resolution Separation: time between peaks Resolution: time between the peaks while considering peak widths

15 Column Selection Principles 1. Choose a stationary phase 2. Column Diameter 3. Film Thickness 4. Column Length

16 Resolution N k Rs = 4 k 1 a 1 a Efficiency Retention Selectivity N = (gas, L, r c ) k = (T, d f, r c ) a = (T, phase) L = Length r c = column radius d f = film thickness T = temperature

17 Selectivity Relative spacing of the chromatographic peaks The result of all non-polar, polarizable and polar interactions that cause a stationary phase to be more or less retentive to one analyte than another

18 Separation Factor (a) a = k 2 k1 co-elution: a = 1 k 2 = retention factor of 2nd peak k 1 = retention factor of 1st peak

19 Optimizing Selectivity Match analyte polarity to stationary phase polarity -like dissolves like(oil and water don t mix) Take advantage of unique interactions between analyte and stationary phase functional groups

20 Polarity Nonpolar Molecules - generally composed of only carbon and hydrogen and exhibit no dipole moment (Straight-chained hydrocarbons (nalkanes)) Polar Molecules - primarily composed of carbon and hydrogen but also contain atoms of nitrogen, oxygen, phosphorus, sulfur, or a halogen (Alcohols, amines, thiols, ketones, nitriles, organo-halides, etc. Includes dipole-dipole interactions and H-bonding) Polarizable Molecules - primarily composed of carbon and hydrogen, but also contain unsaturated bonds (Alkenes, alkynes and aromatic compounds)

21 Selectivity 100% PEG Strong Dispersion Strong Dipole Moderate H Bonding 2 Compounds Polar Aromatic Hydrogen Bonding Dipole Toluene no yes no induced Hexanol yes no yes yes Phenol yes yes yes yes Decane no no no no Naphthalene no yes no induced Dodecane no no no no % Methyl Strong Dispersion No Dipole No H Bonding 1. Toluene 2. Hexanol 3. Phenol 4. Decane (C10) 5. Naphthalene 6. Dodecane (C12)

22 Selectivity is important but not everything Temperature limits will play a role as well. Bleed and Inertness can be critical factors in column selection.

23 Polarity vs Stability/Temperature Range Polarity Stability Temperature Range

24 Benefits of Low Bleed Phases PAH Sensitivity Using DB-35MS Commercially Available 35% phenyl column DB-35MS Benzo[ghi]perylene S/N = 15 Benzo[ghi]perylene S/N = Naphthalene 2. Acenaphthylene 3. Acenaphthene 4. Fluorene 5. Phenanthrene 6. Anthracene 7. Fluoranthene 8. Pyrene 9. Benz[a]anthracene 10. Chrysene 11. Benzo[b]fluoranthene 12. Benzo[k]fluoranthene 13. Benzo[a]pyrene 14. Indeno[1,2,3,- c,d]anthracene 15. Dibenz[a,h]anthracene 16. Benzo[g,h,i]perylene min. Columns: 30 m x 0.32 mm x 0.35 um. Carrier: H2, constant flow, 5 psi at 100 o C. Injector: 275 o C, splitless, 1 ul, 0.5-5ppm. Oven: 100 o C to 250 o C (5 min.) at 15 o C/min.,; then to 320 o C (10 min.) at 7.5 o C/min. Detector: FID, 320 o C.

25 Poor Inertness Can Cause Poor Peak Shape Sample: 0.5 ng on column loading of Short Mix Components with ISTD Column: Brand R 20m 0.18mm 0.18µm Carrier: Helium 37cm/sec, Ramped flow; 0.7ml/min (0.1min) to 1.3ml/min (15ml/min 2 ) Oven: 35 o C (2.5 min) to 80 o C (40 o C/min), 15 o C/min to 200 o C, 8 o C/min to 275 o C (2 min) Injection: 0.5µl, Splitless, 280 o C, purge flow 30ml/min at 0.75 min MSD: Transfer Line 290 o C, Source 300 o C, Quad 180 o C n-nitrosodimethylamine 2. Aniline 3. 1,4-Dichlorobenzene-d4 4. Benzoic Acid 5. Naphthalene-d8 6. Acenaphthene-d ,4-Dinitrophenol 8. 4-Nitrophenol 9. 2-Me-4,6-dinitrophenol Aminobiphenyl 11. Pentachlorophenol 12. Phenanthrene-d Benzidine 14. Chrysene-d ,3 -Dichlorobenzidine 16. Benzo[b]fluoranthene 17. Benzo[k]fluoranthene 18. Perylene-d12 Abundance Time (min) Agilent Restricted Month ##, 200X

26 Agilent J&W HP-5ms Ultra Inert Delivers Excellent Peak Shapes Sample: 0.5 ng on column loading of Short Mix components with ISTD Column: Agilent J &W HP-5ms Ultra Inert 20m 0.18mm 0.18µm Carrier: Helium 37cm/sec, Ramped flow; 0.7ml/min (0.1min) to 1.3ml/min (15ml/min 2 ) Oven: 35 o C (2.5 min) to 80 o C (40 o C/min), 15 o C/min to 200 o C, 8 o C/min to 275 o C (2 min) Injection: 0.5µl, Splitless, 280 o C, purge flow 30ml/min at 0.75 min MSD: Transfer Line 290 o C, Source 300 o C, Quad 180 o C n-nitrosodimethylamine 2. Aniline 3. 1,4-Dichlorobenzene-d4 4. Benzoic Acid 5. Naphthalene-d8 6. Acenaphthene-d ,4-Dinitrophenol 8. 4-Nitrophenol 9. 2-Me-4,6-dinitrophenol Aminobiphenyl 11. Pentachlorophenol 12. Phenanthrene-d Benzidine 14. Chrysene-d ,3 -Dichlorobenzidine 16. Benzo[b]fluoranthene 17. Benzo[k]fluoranthene 18. Perylene-d12 13 Abundance Time (min) Agilent Restricted Month ##, 200X

27 Compare the Difference Abundance Time (min) Abundance Time (min)

28 Agilent J&W Ultra Inert GC columns Highly Inert Less absorption Improved quantification low levels Agilent s stringent QC testing Improved linearity Better peak shape Column reproducibility Better accuracy Improved separation More assurance Less re-runs DB selectivity Easy method change Less revalidation High Efficiency format Fast GC methods Higher productivity

29 Stationary Phase Selection Existing information Critical separations Selectivity/Polarity Inertness and Temperature limits Application designed Examples: DB-VRX, DB-MTBE, DB-TPH, DB-ALC1, DB-ALC2, DB-HTSimDis, DB-Dioxin, HP-VOC, and now Select TM phases Choose the column phase that gives the best separation but not at the cost of robustness or ruggedness.

30 Resolution N k Rs = 4 k 1 a 1 a Efficiency Retention Selectivity N = (gas, L, r c ) k = (T, d f, r c ) a = (T, phase) L = Length r c = column radius d f = film thickness T = temperature

31 Column Diameter - Theoretical Efficiency 5 m Total Plates I.D. (mm) N ~ 112, n/m 23, m N ~ 112, , , m N ~ 112, m N ~ 112, k =

32 Column Diameter and Capacity I.D. (mm) Capacity (ng) Like Polarity Phase/Solute 0.25 µm film thickness

33 Column Diameter and Carrier Gas Flow Lower flow rates: Smaller diameter columns Higher flow rates: Larger diameter columns Low flow rates : GC/MS High flow rates: Headspace, purge & trap

34 Diameter Summary To increase Efficiency Resolution Pressure Capacity Flow rate Diameter Smaller Smaller Smaller Larger Larger

35 Resolution N k Rs = 4 k 1 a 1 a Efficiency Retention Selectivity N = (gas, L, r c ) k = (T, d f, r c ) a = (T, phase) L = Length r c = column radius d f = film thickness T = temperature

36 Film Thickness and Retention: Isothermal Thickness (µm) Retention Change Constant Diameter Normalized to 0.25 µm

37 Film Thickness and Resolution When solute k < 5 d f R (early eluters) or T When solute k > 5 (later eluters) d f or T R

38 Film Thickness and Bleed More stationary phase = More degradation products

39 Other Retention - Adsorption Analysis of Noble & Fixed Gases Using HP PLOT MoleSieve Column: Carrier: Oven: Sample: HP-PLOT/MoleSieve 30 m x 0.53 mm x 50 m HP part no P-MS0 Helium, 4 ml/min 35 C (3 min) to 120 C (5 min) at 25 C/min 250 l, split (ratio 50:1) Neon 2. Argon 3. Oxygen 4. Nitrogen 5. Krypton 6. Xenon Time (min)

40 Film Thickness Summary To Increase Retention Resolution (k<5) Resolution (k>5) Capacity Bleed Inertness Efficiency Make Film Thicker Thicker Thinner Thicker Thicker Thicker Thinner

41 Phase Ratio ( ) Column Dimensions Phase Ratio β 30 m x.53 mm x 3.0 m m x.32 mm x 1.8 m 44 β = r 2d f Kc = kβ (same phase ratio gives same retention)

42 Different Column I. D. Equal Phase Ratios Column: DB m, 0.53 mm, 3 m Carrier: Oven: Helium, 40(cm/sec) 65 C Injection: Split Detector: FID Column : DB m, 0.32 mm, 1.8 m Time (min)

43 Resolution N k Rs = 4 k 1 a 1 a Efficiency Retention Selectivity N = (gas, L, r c ) k = (T, d f, r c ) a = (T, phase) L = Length r c = column radius d f = film thickness T = temperature

44 Column Length and Efficiency (Theoretical Plates) Length (m) N 15 69, , , mm ID n/m = 4630 (for k = 5)

45 Column Length and Resolution R a N a L Length X 4 = Resolution X 2 t a L Upside = Cut a bunch off during routine inlet maintenance and not lose a lot of Resolution

46 Column Length VS Resolution and Retention Time Isothermal R= min R= min R= min 15 m 30 m 60 m Double the plates, double the time but not double the the resolution

47 Column Length and Cost 15m 30m 60m $ $ $ $ $ $ $

48 Length Summary To Increase Efficiency Resolution Analysis Time Pressure Cost Length Longer Longer Longer Longer Longer

49 Resolution Impact of Efficiency, Selectivity and Retention on Resolution Typical Method Development Parameters Increase N Increase Alpha Increase k' R s = N ½ /4 (a-1)/a k /(k +1) Plates: Alpha: k :

50 Conclusions Choose the stationary phase for the best separation in the least amount of time. (Use application specific columns when available) Temperature limits and inertness can play a role. Smaller diameter allows for shorter lengths but have less capacity. Choose film thickness for retention reasons. PLOT columns have strong retention, use where appropriate. Minimize length but have enough necessary for resolution.

Choosing the Correct GC Column Dimensions and Stationary Phase

Choosing the Correct GC Column Dimensions and Stationary Phase Choosing the Correct GC Column Dimensions and Stationary Phase Daron Decker Chromatography Technical Specialist Page 1 Nothing is useless it can always serve as a bad example Custom Column: 150 m x 250

More information

Selection of a Capillary

Selection of a Capillary Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector

More information

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension Simon Jones Application Engineer Things to Consider Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid?

More information

How To Select the Correct GC Column. Simon Jones Application Engineer

How To Select the Correct GC Column. Simon Jones Application Engineer How To Select the Correct GC Column Simon Jones Application Engineer Things to Consider Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected? What

More information

Selection of a Capillary GC Column

Selection of a Capillary GC Column Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer

More information

GC Resolution Do You See What I See?

GC Resolution Do You See What I See? GC Resolution Do You See What I See? Column Dimensions and Carrier Gas Optimization Deans Switch Page 1 Variables for Maximizing Resolution Optimized Stationary Phase Longer Column Length Decrease Internal

More information

Agilent J&W GC Column Selection Guide SPEED YOUR SELECTION WITH THIS ONE-STOP RESOURCE

Agilent J&W GC Column Selection Guide SPEED YOUR SELECTION WITH THIS ONE-STOP RESOURCE Agilent J&W GC Column Selection Guide SPEED YOUR SELECTION WITH THIS ONE-STOP RESOURCE Table of Contents The story behind Agilent J&W Advanced GC Columns... 1 Column Selection Principles... 3 Selecting

More information

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010 Secrets of GC Column Dimensions GC Columns and Consumables Mark Sinnott Application Engineer Folsom California March 12, 2010 Page 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution

More information

Secrets of GC Column Dimensions

Secrets of GC Column Dimensions Secrets of GC Column Dimensions GC Columns and Consumables Simon Jones Application Engineer May 20, 2008 Slide 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution Equation Changes

More information

Introduction and Principles of Gas Chromatography

Introduction and Principles of Gas Chromatography Introduction and Principles of Gas Chromatography Jaap de Zeeuw Restek, Middelburg, The Netherlands Jaap.dezeeuw@restek.com Definition and Uses of Gas Chromatography GC Components and Types of Columns

More information

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011 ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation Optimizing GC Parameters for Faster Separations with Conventional Instrumentation Anila I. Khan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 243 Key Words TraceGOLD fast GC analysis

More information

Introduction to Capillary GC

Introduction to Capillary GC ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

The Suite for Environmental GC Analysis

The Suite for Environmental GC Analysis The Suite for Environmental GC Analysis SGE Environmental GC Columns Performance Selectivity Delivery Promise www.sge.com SGE Environmental GC Columns The Suite for Environmental GC Analysis SGE GC Columns

More information

Agilent J&W GC Column Selection Guide

Agilent J&W GC Column Selection Guide Agilent J&W GC Column Selection Guide Rely on unsurpassed reproducibility, efficiency, and inertness. Speed your selection with this one-stop resource. Agilent J&W GC Column Selection Guide Table of Contents

More information

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms Application Note Author Laura Provoost Agilent Technologies, Inc. Introduction In multi-residue analysis different groups

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Introduction to Gas Chromatography

Introduction to Gas Chromatography Introduction to Gas Chromatography 31-1 Objectives To know what is chromatography To understand the mechanism of compound separation To know the basic of gas chromatography system 31-2 Chromatography Definition

More information

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes Application Note Environmental Authors Ken Lynam and Doris Smith Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

Speakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine

Speakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine Webcast Notes Type your questions in the Submit Question box, located below the slide window You can enlarge the slide window at any time by clicking on the Enlarge Slides button, located below the presentation

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software High Efficiency GC Columns Page 1 Variables for Shortening Run Times Stationary Phase Shorten Column Length

More information

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings Technical Overview Introduction Modern GC and GC/MS instruments are important analytical tools for accurate and

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask

More information

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column Application Note Environmental Author Doris Smith and Kenneth Lynam Agilent Technologies, Inc. 285 Centerville

More information

The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane

The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane Jarl Snider, D. J. Hotnisky, Kristi Sellers, Dinesh V. Patwardhan Ph. D. Restek Corporation www.restekcorp.com

More information

Introduction to Capillary GC

Introduction to Capillary GC Introduction to Capillary GC LC Columns and Consumables Simon Jones Chromatography Applications Engineer February 20, 2008 Page 1 Introduction to Capillary GC t r K c?? Kβ k = - tr t m? t m R s Page 2

More information

Chapter 27: Gas Chromatography

Chapter 27: Gas Chromatography Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid

More information

Volatile organic compounds (VOCs):

Volatile organic compounds (VOCs): Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined

More information

Low Bleed Stationary Phases for Gas Chromatography

Low Bleed Stationary Phases for Gas Chromatography Low Bleed Stationary Phases for Gas Chromatography Slide 4 Advantages of Low Bleed Phase Lower bleed compared to standard phases Better signal to noise (sensitivity) Higher upper temperature - shorter

More information

Rely on Rxi -PAH Columns

Rely on Rxi -PAH Columns Rely on Rxi -PAH Columns to Ensure Successful Polycyclic Aromatic Hydrocarbon (PAH) Analysis Optimized Efficiency, Selectivity, and Robustness Let You: Report accurate results with speed and confidence.

More information

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014 Developing Large Volume Injection (LVI) in Split / Splitless Inlets Philip J. Koerner Phenomenex Inc. NEMC 2014 Goal To produce a method that provides the largest response of analytes that will produce

More information

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications Chapter 27: Gas Chromatography Principles Instrumentation Detectors Columns and Stationary Phases Applications GC-MS Schematic Interface less critical for capillary columns Several types of Mass Specs

More information

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices Food Quality and Environmental Author Ken Lynam Agilent Technologies, Inc. 280 Centerville Road Wilmington, DE 19808 Abstract

More information

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns Brochure Brief Porous Layer Open Tubular (PLOT) columns have been replacing traditional packed columns

More information

Get Selective. By Jaap de Zeeuw

Get Selective. By Jaap de Zeeuw 34 Get Selective Modern narrow bore columns have made chromatographers lazy when it comes to stationary phase selection. Here s how getting back to basics in gas chromatography by using selectivity can

More information

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at

More information

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

CH 2252 Instrumental Methods of Analysis Unit V  Gas Chromatography.  M. Subramanian CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography M. Subramanian Assistant Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam 603

More information

2401 Gas (liquid) Chromatography

2401 Gas (liquid) Chromatography 2401 Gas (liquid) Chromatography Chromatography Scheme Gas chromatography - specifically gas-liquid chromatography - involves a sample being vaporized and injected onto the head of the chromatographic

More information

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS Application Note # CA 284780 Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS Abstract The analysis of semi-volatile organic compounds

More information

Trajan SGE GC Columns

Trajan SGE GC Columns Trajan Scientific and Medical Trajan SGE GC Columns Trajan Scientific and Medical Our focus is on developing and commercializing technologies that enable analytical systems to be more selective, sensitive

More information

Ultra-Inert chemistry for Trace Level Analysis

Ultra-Inert chemistry for Trace Level Analysis Ultra-Inert chemistry for Trace Level Analysis Cikui Liang, Ph.D. Challenges and Needs of Today s Laboratories Challenges Qualification/quantification of trace samples Keep instrument up and running Needs

More information

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs) environmental #59120 Applications note Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs) Polychlorinated biphenyls (PCBs) are

More information

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS 1.0 SCOPE AND APPLICATION 1.1 Method 8100 is used to determine the concentration of certain polynuclear aromatic hydrocarbons (PAH). Table 1 indicates compounds

More information

Evaluation of Capillary Columns for General Performance Parameters

Evaluation of Capillary Columns for General Performance Parameters Evaluation of Capillary s for General Performance Parameters Application Authors Mitch Hastings, Eberhardt R. Kuhn and Allen K. Vickers Agilent Technologies, Inc 91 Blue Ravine Road Folsom, CA 95630 USA

More information

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5 Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible

More information

Column Selection. there is more to life than a boiling point column. Jaap de Zeeuw Restek Corporation, The Netherlands. Copyrights: Restek Corporation

Column Selection. there is more to life than a boiling point column. Jaap de Zeeuw Restek Corporation, The Netherlands. Copyrights: Restek Corporation Column Selection there is more to life than a boiling point column Jaap de Zeeuw Restek Corporation, The Netherlands Stationary Phase Selectivity Boiling Point versus Solubility Retention of a compound

More information

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas GAS CHROMATOGRAPHY Mobile phase is a gas! Stationary phase could be anything but a gas Gas Chromatography (GC) GC is currently one of the most popular methods for separating and analyzing compounds. This

More information

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit Analysis times for semivolatile compounds can limit sample throughput and decrease overall lab productivity. You can significantly

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

The Role of the Carbon Cartridge in the One-Pass System

The Role of the Carbon Cartridge in the One-Pass System The Role of the Carbon Cartridge in the One-Pass System William Jones and Zoe Grosser, Horizon Technology, Inc. Key Words Solid phase extraction, spe, one-pass system, carbon cartridges Introduction Horizon

More information

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable

More information

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas Agilent Porous Polymer PLOT Columns: New Products, Expanded Uses, Prices Cut in Half! Product Brief Need improved resolution of small volatile compounds? Didn't try a PLOT column due to high price, short

More information

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid Gas Chromatography Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid Principles Instrumentation Applications 18-1 Retention Volumes Volumes rather than times

More information

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column Competitive Comparison Authors Ngoc-A Dang and Allen K. Vickers Agilent

More information

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column The World's First Non-metal 5% Phenyl Phase GC Column Rated to 430 C * Specially processed for thermal stability up to 430 C A true

More information

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Chemistry Instrumental Analysis Lecture 28. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 28 Two types in general use: -packed (stationary phase) -open tubular or capillary determine selectivity and efficiency of the sample. Column Materials Column

More information

Gas Chromatography. Introduction

Gas Chromatography. Introduction Gas Chromatography Introduction 1.) Gas Chromatography Mobile phase (carrier gas) is a gas - Usually N 2, He, Ar and maybe H 2 - Mobile phase in liquid chromatography is a liquid Requires analyte to be

More information

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Environmental Analysis Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution Application Note Abstract Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution Page 1 Polycyclic Aromatic Hydrocarbons (PAHs) are a large group of organic compounds

More information

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes Priority Analysis LC GC Core-Shell Technology GC Columns Kinetex PAH Expanded resolution with chemical selectivity specifically for PAHs Increased throughput and sensitivity with core-shell technology

More information

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS Application Note #1820230 Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS GC-MS Introduction USEPA Method 8270 [1] for semivolatiles is is used by by laboratories to measure

More information

GC Columns PLOT Columns

GC Columns PLOT Columns GC Columns PLOT Columns PLOT Column Selection...117 119 Fused Silica Silica BOND Columns...120 Fused Silica Alumina BOND Columns...121 124 Fused Silica Molecular Sieve 5A Columns...125 Fused Silica Porous

More information

BP1 PONA BPX1. GC Capillary Columns BP1 PONA and BPX1

BP1 PONA BPX1. GC Capillary Columns BP1 PONA and BPX1 GC Capillary Columns 100% Dimethyl Polysiloxane GC Columns and Applications ID (mm) Film Thickness (µm) Length (m) Temperature Limits ( C) Part No. 0.32 0.5 60-60 to 320/340 054069 0.32 1 60-60 to 320/340

More information

GC Instruments. GC Instruments - Sample Introduction

GC Instruments. GC Instruments - Sample Introduction GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes

More information

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9 Gas Chromatograph Application Note Author Matthew Giardina, Ph.D., Applications Chemist, Gas Phase Separations Division, Agilent Technologies,

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

Virtually Particle-Free Rt -Silica BOND Columns

Virtually Particle-Free Rt -Silica BOND Columns Petrochemical Applications Virtually Particle-Free Rt -Silica BOND s Provide Reliable PLOT Performance With Less Time Lost for Maintenance By Amanda Rigdon, Bill Bromps, Tom Vezza, and Jaap de Zeeuw Optimized

More information

Theory and Instrumentation of GC. GC Columns

Theory and Instrumentation of GC. GC Columns Theory and Instrumentation of GC GC Columns i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual.

More information

Activity in the FID Detection Port: A Big Problem if Underestimated

Activity in the FID Detection Port: A Big Problem if Underestimated Activity in the FID Detection Port: A Big Problem if Underestimated Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands It is commonly known in gas chromatography, that many problems can be

More information

Gas chromatography. Advantages of GC. Disadvantages of GC

Gas chromatography. Advantages of GC. Disadvantages of GC Advantages of GC Gas chromatography Fast analysis, typically minutes Effi cient, providing high resolution Sensitive, easily detecting ppm and often ppb Nondestructive, making possible on - line coupling;

More information

CHAPTER 6 GAS CHROMATOGRAPHY

CHAPTER 6 GAS CHROMATOGRAPHY CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1

More information

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions Michael D. Buchanan September 11, 2014 sigma-aldrich.com/analytical 2012 Sigma-Aldrich Co. All

More information

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC Gas Chromatography Gas Chromatography Presented By Mr. Venkateswarlu Mpharm KTPC What is Gas Chromatography? It is also known as Gas-Liquid Chromatography (GLC) GAS CHROMATOGRAPHY Separation of gaseous

More information

Application Note. Abstract. Authors. Environmental

Application Note. Abstract. Authors. Environmental GC/MS Analysis of European Union (EU) Priority Polycyclic Aromatic Hydrocarbons (PAHs) using an Agilent J&W DB-EUPAH GC Column with a Column Performance Comparison Application Note Environmental Authors

More information

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography Emqal module: M0925 - Quality parameters and optimization in is a separation technique used for quantification of mixtures of analytes Svein.mjos@kj.uib.no Exercises and lectures can be found at www.chrombox.org/emq

More information

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Chapter 23. Gas Chromatography What did they eat in the year 1,000? GC of Cholesterol and other lipids extracted from

More information

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2 Print version Gas Chromatography Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 1 Contents The primary components to a GC system 1. Carrier Gas System (including

More information

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples AppNote 2/2 Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples Pat Sandra Department of Organic Chemistry, University of Gent, Krijgslaan 281 S4, B-9 Gent, Belgium Erik Baltussen

More information

Theory and Instrumentation of GC. Chromatographic Parameters

Theory and Instrumentation of GC. Chromatographic Parameters Theory and Instrumentation of GC Chromatographic Parameters i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this

More information

Ch24. Gas Chromatography (GC)

Ch24. Gas Chromatography (GC) Ch24. Gas Chromatography (GC) 24.1 What did they eat in the year 1000? From 13 C content of cholesterol in ancient bone 13 C : 1.1%, 12 C: 98.9% 13 C/ 12 C ratio types of plants Bones of 50 people in Barton-on-Humber

More information

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA New Analytical Tools for the Determination of Persistent Organic Pollutants (POPs) in Fatty Food and Beverage Matrices Using QuEChERS Extraction/Cleanup and Gas Chromatography (GC) Analysis Katherine K.

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography) Gas Chromatography 1. Introduction. Stationary phases 3. Retention in Gas-Liquid Chromatography 4. Capillary gas-chromatography 5. Sample preparation and injection 6. Detectors (Chapter and 3 in The essence

More information

Experiment 6 Simple and Fractional Distillation

Experiment 6 Simple and Fractional Distillation Experiment 6 Simple and Fractional Distillation Vapor Pressure vs Temperature of Water Vapor Pressure vs Temperature of Water 25 Vapor Pressure vs Temperature of Water 25 Vapor Pressure (kpa) (kpa) 2 2

More information

New Gs BP GC Columns

New Gs BP GC Columns New Gs BP GC Columns GS-Tek What really makes GsBP GC columns different from other brand name columns? GsBP columns are developed and made in USA with proprietary manufacturing processes developed by column

More information

GC-CI-MS analysis of TMS derivatives

GC-CI-MS analysis of TMS derivatives GC-CI-MS analysis of TMS derivatives This method describes analysis of TMS derivatives with methane chemical ionisation (CI) rather than the more normal electron impact (EI) ionisation. Methane CI is a

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Analytical Trap Comparison for USEPA Method 8260C

Analytical Trap Comparison for USEPA Method 8260C Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component

More information

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Gas Chromatography (GC) In gas chromatography, the sample is vaporized and injected onto the head of a chromatographic

More information

GC Instruments. GC Instruments - Columns

GC Instruments. GC Instruments - Columns GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes

More information

NEW Agilent J&W Ultra Inert Capillary GC Columns. Proven performance that raises the bar on. consistent. column inertness.

NEW Agilent J&W Ultra Inert Capillary GC Columns. Proven performance that raises the bar on. consistent. column inertness. NEW Agilent J&W Ultra Inert Capillary GC Columns Proven performance that raises the bar on consistent column inertness. Finally a column technology that gives you better sensitivity and peak shape for

More information

REFINED RESTEK. innovative petrochemical solutions. Visit us at Innovative Chromatography Solutions

REFINED RESTEK. innovative petrochemical solutions. Visit us at   Innovative Chromatography Solutions RESTEK REFINED innovative petrochemical solutions Innovative bonding process minimizes particle release. More consistent flow means stable retention times. Outstanding peak symmetry improves impurity analysis.

More information

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid. Chromatography Various techniques for the separation of complex mixtures that rely on the differential affinities of substances for a gas or liquid mobile medium and for a stationary adsorbing medium through

More information

This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission

This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission cannot be held responsible for any use which may be made

More information

Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas

Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas Chromatography Joan Stevens, Ph.D. Sample Preparation Applications Scientist Chemistries and Supplies Division 1 October 27, 2011

More information

Water Injections in GC - Does Water Cause Bleed?

Water Injections in GC - Does Water Cause Bleed? Water Injections in GC - Does Water Cause Bleed? Eberhardt Kuhn Applications Chemist April 4, 2001 Practical Advice and Useful Tips for the Analysis of Semivolatile Organics by GC and GC/MS 11:00 a.m.

More information

CHEM340 Tutorial 4: Chromatography

CHEM340 Tutorial 4: Chromatography CHEM340 Tutorial 4: Chromatography 1. The data in the table below was obtained from a chromatogram obtained with a 10 cm liquid chromatography column. Under the conditions used, the compound uracil is

More information