SOUR WATER ANALYSIS USING A DUAL- COLUMN HEADSPACE SAMPLING SYSTEM

Size: px
Start display at page:

Download "SOUR WATER ANALYSIS USING A DUAL- COLUMN HEADSPACE SAMPLING SYSTEM"

Transcription

1 SOUR WATER ANALYSIS USING A DUAL- COLUMN HEADSPACE SAMPLING SYSTEM Andy Burgess, David Greaves, Daniel Murphy and Sam Rolley Applied Analytics, Inc. 29 Domino Drive, Concord, MA KEYWORDS Sour Water, Headspace, H 2 S, NH 3, Spectroscopy ABSTRACT In the petroleum refining process one of the main byproducts is wastewater rich in hydrogen sulfide and ammonia, commonly referred to as sour water. These chemicals must be removed to prevent contamination and corrosion. A process called Sour Water Stripping accomplishes this removal. Accurate measurement of hydrogen sulfide and ammonia concentrations at the outlet of the stripper provides critical process validation information and allows for corrective action in the event of stripper failure. To carry out this measurement in real time, a dual-column headspace sampling system is utilized in conjunction with a UV-diode array based spectrometer. This system allows for the simultaneous measurement of hydrogen sulfide and ammonia with a single device. A detailed discussion of sour water measurement benefits, the unique features of headspace sampling, and experimental results are presented. INTRODUCTION Sour water is defined as any refinery process water that contains hydrogen sulfide (H 2 S). With such a broad definition, sources of sour water vary significantly. Atmospheric crude columns and vacuum crude towers produce sour water from condensed stripping steam removed by overhead condensing systems. Steam cracking and fluid catalytic cracking units produce sour water as condensates from steam used in injection stripping and aeration [1]. Refineries create sour water as a byproduct in many of their processes. Unfortunately, sour water contains ammonium salts such as ammonium bisulfide and ammonium Session 7.3: Page 1

2 carbonate. These ammonium salts can cause several problems. In an extensive study of corrosion rates in sour water, a number of variables were considered including ammonium bisulfide concentration, velocity, H 2 S partial pressure, temperature, chloride concentration, hydrocarbon content, and chemical treatments. It was confirmed that the concentration of ammonium bisulfide is the most influential variable contributing to corrosion [5]. Ammonium salts can also settle on orifices causing plugging and increase maintenance concerns. As a result, refineries need to remove the hydrogen sulfide and ammonia (NH 3 ) from the water to avoid the formation of these ammonium salts when the water is reused for other plant processes. To remove the H 2 S and NH 3, a sour water stripper is often utilized. The plant operators need to know how efficient the stripping process is and if the water is suitable for reintroduction. TECHNOLOGY To address the need brought to light in the introduction, Applied Analytics developed a system utilizing the OMA-300 for the measurement of H 2 S and NH 3 in sour water. It utilizes an ultraviolet-visible full-spectrum spectrometer and a dual column headspace sampling approach. The combined system offers many benefits. It allows a large dynamic range in the measurement of the analytes due to flexibility in wavelength selection. Cross interference from moisture is avoided, because the measurement is being done in the UV- VIS region where water doesn t absorb light. Finally, consumables and/or reagents are not necessary for operation. The technology is broken down into two sections for ease of explanation: an analytical section and a sample conditioning section. ANALYTICAL The analytical configuration consists of the following equipment: a xenon flash lamp, an ultraviolet-visible full-spectrum spectrometer, collimating lenses, fiber optic cables, and a flow through cell. This layout can be seen in Figure 1. Absorbance is defined using the following equation: A = log 10 (I 0 /I) (1) Where: A = Absorbance I 0 = Light In I = Light Out Session 7.3: Page 2

3 FIGURE 1. ANALYTICAL HARDWARE The measurement algorithm used for analysis is a classical least squares regression of Beer s Law [2]: A l = e l bc (2) Where: A = Absorbance l = Wavelength e = Molar Absorptivity b = Path Length c = Concentration The spectrometer measures I 0 during its zero calibration and I when the analyte is in the flow through cell of path length b. e l is found by introducing a known concentration of the analyte into the flow-through cell and is saved for use in subsequent readings to determine unknown concentrations of the analyte. This would be adequate if the absorbance at a given wavelength could be attributed to only one component. However, this isn t the case with H 2 S and NH 3 in the ultraviolet region. At each wavelength in the UV where NH 3 absorbs, so does H 2 S. This is illustrated in the following spectra (Figures 2 & 3). Session 7.3: Page 3

4 H2S (gas) Absorbance Curve ) U (A e c n a rb o s b A Wavelength (nm) FIGURE 2. H 2 S ABSORBANCE CURVE NH3 (gas) Absorbance Curve ) U 0.7 (A e c 0.6 n a rb 0.5 o s b 0.4 A Wavelength (nm) FIGURE 3. NH 3 ABSORBANCE CURVE Session 7.3: Page 4

5 Therefore, in this case the equation must be extended to the following: A lx = e 1lx * b * c 1 + e 2lx * b * c 2 (3) A ly = e 1ly * b * c 1 + e 2ly * b * c 2 (4) Where lx could be 217nm and ly could be 220nm, for example (the over determination is omitted for simplicity). The final concentration readings are determined by solving for c 1 and c 2 using the previous equations. Accordingly, before the gas can be analyzed it must be coerced into the vapor phase. This is accomplished by the headspace sample conditioning system. SAMPLE CONDITIONING The heart of the headspace system is the stripping column (Figure 4). Its dimensions are 24 in height by 2 in diameter. The column is sealed with gaskets that are held together by sanitary clamps. The column is filled with Pall rings for efficient mass transfer between the liquid sample and the carrier gas. The column is heated with three band heaters and controlled with a thermocouple and a temperature controller. The flow rates of the carrier gas and the sample liquid are controlled by variable area flow meters with integrated flow control valves. A backpressure regulator regulates the pressure in the stripping column and the flow through cell. During operation, the liquid sample enters through a fitting on the top of the column and exits through a fitting on the bottom of the column. A carrier gas is introduced through another fitting on the bottom of the column and exits through a fourth fitting on the top of the column. This establishes a countercurrent flow between the liquid sample and the carrier gas. The Pall rings allow for efficient mass transfer between the liquid sample and the carrier gas by greatly increasing the total internal surface area of the column. After the carrier gas leaves the column it flows to the flow through cell for measurement. The headspace system is based on the principle of Henry s Law: C = k*p gas (5) Where: P = Partial Pressure k = Henry s Law Constant c = Concentration Henry s Law states that the amount of a gas dissolved in a solution at a given temperature is directly proportional to the partial pressure of the gas above the solution [3]. Session 7.3: Page 5

6 FIGURE 4. HEADSPACE SAMPLE CONDITIONING SYSTEM Therefore, given constant conditions such as temperature, pressure, carrier gas flow rate and liquid sample flow rate the headspace gas being measured can be correlated to the concentration of the liquid sample. Although the headspace sampling method can be used in a variety of applications, it is most advantageous in measurements of components with high volatility in opaque samples that have interfering components with low volatility. Sour water is a good example of such a stream, since it can often contain phenols that were formed from reactions between steam and cyclic hydrocarbons. Phenols have a very strong absorbance in the low UV region ( nm), which would make measurement of the H 2 S or NH 3 in the liquid phase very difficult. Fortunately, phenols also have a much higher boiling point (see Table I) than H 2 S or NH 3. As a result, only very low phenol concentrations can exist in the headspace gas, thus making the measurement easier by minimizing the interfering component. Session 7.3: Page 6

7 TABLE I. BOILING POINT COMPARISON Phenol H 2 S NH 3 Boiling Point EXPERIMENT The objective of the experiment is to quantitatively demonstrate the typical accuracy and stability that the headspace sampling system offers when correlated with a liquid sample of H 2 S and NH 3 in water. PARAMETERS A number of parameters need be determined before starting the experiment. The proper liquid sample flow rate shall be constant. The liquid sample flow rate was experimentally determined by filling the column with water and draining it with just gravity as a motive force. The column was packed with Pall rings, as it would be in normal operation, during this test. The time in which it takes the water to fully evacuate the column was recorded. The volume of water as it leaves the column was captured in a graduated beaker, thus allowing the determination of the volume of the column with Pall rings installed. That information was then used to find the volumetric flow rate (V/t). Since Henry s Law is dependent upon temperature, it was controlled and kept constant at a temperature that promotes low solubility of the gas in the water. The dependence of the solubility on temperature is described by the following equations [4]: k = k t * exp [ (-enthalpy/r) * ((1/T) (1/T t )) ] (6) (-d*ln(k))/(d*(1/t)) = enthalpy/r (7) Where k = Henry s constant of a specific gas (M/atm) k t = Henry s constant of a specific gas at a temperature of K (M/atm) R = universal gas constant (8.314 (J/mol*k)) T = System temperature (K) T t = K (-d*ln(k))/(d*(1/t)) = Temperature Dependence (K) TABLE II. HENRY S LAW CONSTANT & TEMPERATURE DEPENDENCE H 2 S NH 3 Henry's Law Constant (M/atm) Temperature Dependence (K) Session 7.3: Page 7

8 Table II gives a value for temperature dependence, which according to Equation 7 is equal to enthalpy/r. That value was substituted into Equation 6 to find the Henry s Law Constant at different temperatures. Then Equation 5 was used to predict the concentration of the gas in the water. TABLE III. HENRY S LAW CONSTANT FOR H 2 S AT DIFFERENT TEMPERATURES k (M/atm) k t (M/atm) E/R (K) T (K) T t (K) TABLE IV. HENRY S LAW CONSTANT FOR NH 3 AT DIFFERENT TEMPERATURES k (M/atm) k t (M/atm) E/R (K) T (K) T t (K) Tables III and IV show that as temperature increases, the Henry s Law constant and the concentration of the dissolved gas in the water decreases. Therefore, it is beneficial to choose a high temperature for the column set point because it forces more of the dissolved gas from the liquid phase into the vapor phase. A larger concentration in the vapor phase results in more absorbance in the path length of the flow cell. Furthermore, the accuracy of the analyzer increases as absorbance increases. Keeping this in mind, along with practical limitations of the setup, 50 degrees C was chosen for the set point of the column. The carrier gas flow rate was also kept constant. The ideal carrier flow rate was set based on the highest flow rate that was achieved without decreasing the concentration of the analyte in the stripping carrier gas. This was determined experimentally by analyzing the carrier gas at several carrier gas flow rates and recording the results. Session 7.3: Page 8

9 The analyzer (UV absorbance) was calibrated on gas standards of H 2 S in N 2 and NH 3 in N 2. These spectral standards were used to establish an extinction coefficient for each gas at each wavelength in the range of interest. This calibration was done prior to water testing. It is also important to note that ph plays an important role in sour water stripping. At low ph levels near 5.5, NH 3 tends to be held in solution and H 2 S tends to be more volatile. However, at higher ph values near 9.0, NH 3 is more volatile and H 2 S tends to stay in solution [1]. For the purposes of these experiments a constant ph was assumed. PROCEDURE A sample of water without H 2 S or NH 3 was introduced into the column. The temperature of the column, the carrier gas flow rate and the liquid sample flow rate were held constant at their optimal values. This sample was used as a zero reference by the analyzer. A sample of water with a known concentration of H 2 S was prepared in a reservoir large enough to hold the volume required to run a 1-hour stability test. The temperature of the column, the carrier gas flow rate and the liquid sample flow rate were held constant at their optimal values. The concentration log function on the analyzer was activated and the system was left to run for 1 hour. The start time and the end time were recorded. At the end of the 1-hour run, a reservoir of clean water was connected to wash the system. Second, third, and fourth samples of water with different known concentrations of H 2 S were prepared and introduced to the system in the same manner as the first. A 1-hour run was not required as these samples were for the purpose of testing accuracy, not stability. After the system was washed with clean water again, a sample of water with a known concentration of NH 3 was prepared in a reservoir large enough to hold the volume required to run a 1-hour stability test. The concentration log function on the analyzer was activated and the system was left to run for 1 hour. The start time and the end time were recorded. At the end of the 1-hour run, a reservoir of clean water was connected to wash the system. Second, third, and fourth samples of water with different known concentrations of NH 3 were prepared and introduced into the system in the same manner as the first. RESULTS The data retrieved from the concentration log file was used to show the stability and accuracy of the headspace system using H 2 S and NH 3 concentrations in water. Session 7.3: Page 9

10 TABLE V. H 2 S STABILITY TABLE Test Duration (hours) 1 Reading Stored (1/sec) 60 Average Value (PPM) Maximum Value (PPM) Minimum Value (PPM) Standard Deviation (PPM) 6.54 Slope (PPM/reading).36 Slope * 60 (PPM/hour) 2.35 H 2 S Stability Concentration (ppm) Time (seconds) FIGURE 5. H 2 S STABILITY GRAPH The first set of data is the stability of H 2 S in water over 1 hour. The average value over the 1-hour run was determined to be , with the maximum reading at and the minimum reading at The standard deviation of the data set is 6.54 and it has a slope of.36. TABLE VI. NH 3 STABILITY Test Duration (hours) 1 Reading Stored (1/sec) 60 Average Value (PPM) Maximum Value (PPM) Minimum Value (PPM) Standard Deviation (PPM) 4.09 Slope (PPM/reading) 0.02 Slope * 60 (PPM/hour) 1.10 Session 7.3: Page 10

11 NH 3 Stability ) m p (p n tio tra n e c n o C Time (seconds) FIGURE 6. NH 3 STABILITY GRAPH The second set of data is the stability of NH 3 in water over 1 hour. The average value over the 1-hour run was determined to be , with the maximum reading at and the minimum reading at The standard deviation of the data set is 4.09 and it has a slope of TABLE VII. H 2 S ACCURACY TABLE Liquid Concentration Sample 5 (% of Cal Sample) 43.5 Liquid Concentration Sample 6 (% of Cal Sample) Liquid Concentration Sample 7 (% of Cal Sample) Liquid Concentration Sample 8 (% of Cal Sample) 100 Gas Concentration Sample 5 (PPM) 1300 Gas Concentration Sample 6 (PPM) 2600 Gas Concentration Sample 7 (PPM) 3800 Gas Concentration Sample 8 (PPM) 4200 Linearity R 2 Value.9996 The third set of data is the accuracy of H 2 S in water. The gas concentration reading for Sample 1 was 1300ppm. The gas concentration reading for Sample 2 was 2600ppm. The gas concentration reading for Sample 3 was 3800ppm. The gas concentration reading for Sample 4 was 4200ppm. The linearity (R 2 value) of the gas concentration readings was Session 7.3: Page 11

12 Gas Concentration (PPM) H2S Accuracy R² = Liquid Concentration (% of Cal Sample) FIGURE 7. H 2 S ACCURACY GRAPH TABLE VIII. NH 3 ACCURACY TABLE Liquid Concentration Sample 5 (% of Cal Sample) 25 Liquid Concentration Sample 6 (% of Cal Sample) 50 Liquid Concentration Sample 7 (% of Cal Sample) 75 Liquid Concentration Sample 8 (% of Cal Sample) 100 Gas Concentration Sample 5 (PPM) Gas Concentration Sample 6 (PPM) Gas Concentration Sample 7 (PPM) Gas Concentration Sample 8 (PPM) Linearity R 2 Value The fourth set of data is the accuracy of NH 3 in water. The gas concentration reading for Sample 5 was ppm. The gas concentration reading for Sample 6 was ppm. The gas concentration reading for Sample 7 was ppm. The gas concentration reading for Sample 8 was The linearity (R 2 value) of the gas concentration readings was Session 7.3: Page 12

13 Gas Concentration (PPM) NH3 Accuracy R² = Liquid Concentration (% of Cal Sample) FIGURE 8. NH 3 ACCURACY GRAPH CONCLUSIONS Refineries create sour water as a byproduct in many of their processes. Sour water has been proven to cause corrosion and plugging due to the presence of ammonium salts, such ammonium bisulfide. Therefore, it is beneficial for refineries to remove the hydrogen sulfide and ammonia so the water can be reused or discarded. This is often done through a stripping process. Measurement of the H 2 S and NH 3 in sour water occurs afterwards to ensure the effectiveness of the stripper before the water is sent for reuse. The measurement is not particularly easy without advanced sampling techniques. Therefore, a sampling technique was developed that used some of the same principles of stripping that the refineries themselves use to remove H 2 S and NH 3 from the sour water stream. After the concentrations of each of these chemicals are measured in the gas phase, they are correlated back to a liquid concentration. In conclusion, the results from the stability test demonstrate a drift of less than 0.5% full scale. Also, the headspace system yields a linearity of or better when comparing carrier gas to sample stream analyte concentrations. Coupled with the benefits mentioned in the introduction, these analytical results make the UV-VIS full-spectrum spectrometer Session 7.3: Page 13

14 with the headspace sample conditioning system an ideal solution for the measurement of H 2 S and NH 3 in sour water. REFERENCES 1. Armstrong, Tim, Scott, Bruce, Taylor, Kin, and Gardner, Art, Sour Water Stripping, Today s Refinery, June, Mark, Howard, Workman, Jerome, Classical Least Squares, Part 1: Mathematical Theory, Spectroscopy, May, Zumdahl, Steven S., Zumdahl, Susan A., Chemistry 6 th Edition Houghton Mifflin, Boston, Massachusetts, Sander, Rolf, Compilation of Henry s Law Constants for Inorganic and Organic Species of Potential Importance in Environmental Chemistry, Max Planck Institute of Chemistry, April Horvath, Richard J., Lagand, Vishal V., Srinivasan, Sridhar, Kane, Russel D., Prediction and Assessment of Ammonium Bisulfide Corrosion Under Refinery Sour Water Service Conditions Part 2., NACE International, Session 7.3: Page 14

S is referred to as sour. The sulfide loading lends an unpleasant odor and makes the water corrosive. Additionally, the presence of H 2

S is referred to as sour. The sulfide loading lends an unpleasant odor and makes the water corrosive. Additionally, the presence of H 2 Application Summary Analytes: S (hydrogen sulfide), (ammonia) Detection Technology: OMA-300 Process Analyzer Process Stream: stripped sour water Introduction In the context of petrochemical refining and

More information

Reprinted from February Hydrocarbon

Reprinted from February Hydrocarbon February2012 When speed matters Loek van Eijck, Yokogawa, The Netherlands, questions whether rapid analysis of gases and liquids can be better achieved through use of a gas chromatograph or near infrared

More information

Physicochemical Processes

Physicochemical Processes Lecture 3 Physicochemical Processes Physicochemical Processes Air stripping Carbon adsorption Steam stripping Chemical oxidation Supercritical fluids Membrane processes 1 1. Air Stripping A mass transfer

More information

SULFUR RECOVERY ANALYZERS; UV Diode-Array Fiber- Optics Technology

SULFUR RECOVERY ANALYZERS; UV Diode-Array Fiber- Optics Technology SULFUR RECOVERY ANALYZERS; UV Diode-Array Fiber- Optics Technology Yoav Barshad Applied Analytics, Inc. 29 Domino Drive Concord, Massachusetts, 01742 KEYWORDS Sulfur recovery, H2S, SO2, tail gas, UV, Spectroscopy,

More information

Cyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations

Cyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations Cyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations Introduction Fluid catalytic cracking (FCC) is a major unit operation in refineries around the world. FCC is used to convert

More information

Monitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2

Monitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2 Application Summary Analytes: (nitrogen trichloride) Detector: OMA-300 Process Analyzer Process Stream: headspace gas above saturated CCl 4 solvent Typical Measurement Range: 0-20,000 ppm Introduction

More information

MEASURING H 2 S IN CRUDE OIL FOR QUALITY CONTROL & TRANSPORTATION SAFETY

MEASURING H 2 S IN CRUDE OIL FOR QUALITY CONTROL & TRANSPORTATION SAFETY MEASURING IN CRUDE OIL FOR QUALITY CONTROL & TRANSPORTATION SAFETY COQA Meeting 2015 Marriott West Loop, Houston Wesley Kimbell Analytical Systems Keco www.h2sanalyzer.com Why measure in crude oil? Safety

More information

A New Hyperspectral Spherical-Cavity Absorption Meter

A New Hyperspectral Spherical-Cavity Absorption Meter A New Hyperspectral Spherical-Cavity Absorption Meter David R. Dana and Robert A. Maffione HOBI Labs, Inc. Adapted from the poster presentation at Ocean Sciences 2006, Honolulu, Hawaii INTRODUCTION How

More information

ISA QATAR : 90 th Seminar

ISA QATAR : 90 th Seminar ANALYTICAL INSTRUMENTATION & MAINTENANCE SYSTEMS (AIMS) ISA QATAR : 90 th Seminar UTILIZATION OF LASERS IN THE FIELD OF GAS ANALYZERS Presenter: ZAHEER JUDDY Electromagnetic Spectrum Electromagnetic Spectrum

More information

Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application

Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application FFeature Article Article Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application Yoko NAKAI HORIBA s CS-Series chemical concentration monitors that use ultraviolet

More information

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES Exercise 2-4 Titration of a Buffer Solution EXERCISE OBJECTIVES To define the terms buffer solution and buffer capacity; To titrate a buffer solution with a weak acid solution; To plot a graph using the

More information

Method for Nitrite determination on Low-Range Samples

Method for Nitrite determination on Low-Range Samples Method for Nitrite determination on Low-Range Samples 1. PRINCIPLE This method is designed for the determination of nitrite in soil, water and other forms of samples. The sample is prepared offline if

More information

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II) : Absorption Spectroscopy of Cobalt(II) OBJECTIVES In successfully completing this lab you will: prepare a stock solution using a volumetric flask; use a UV/Visible spectrometer to measure an absorption

More information

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law Vernier Spectrometer 1 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You

More information

SERVOTOUGH SpectraScan

SERVOTOUGH SpectraScan SERVOTOUGH SpectraScan Continuous, real-time analysis of light hydrocarbons Jan Hordijk - Servomex Hydrocarbon stream analysis 5 major technologies used in HydroCarbon land for gas: IR (Infra Red) Laser

More information

Mass Transfer Operations I Prof. Bishnupada Mandal Department of Chemical Engineering Indian Institute of Technology, Guwahati

Mass Transfer Operations I Prof. Bishnupada Mandal Department of Chemical Engineering Indian Institute of Technology, Guwahati Mass Transfer Operations I Prof. Bishnupada Mandal Department of Chemical Engineering Indian Institute of Technology, Guwahati Module - 4 Absorption Lecture - 4 Packed Tower Design Part - 3 Welcome to

More information

Final Exam. Physical Constants and Conversion Factors. Equations

Final Exam. Physical Constants and Conversion Factors. Equations Final Exam Instructions: This exam is worth 100 points. Some questions allow a choice as to which parts are answered. Do not answer more parts than are requested. velocity of light in a vacuum: 3.0x10

More information

Operating Instructions for Ammonium ISE Specifications

Operating Instructions for Ammonium ISE Specifications Operating Instructions for Ammonium ISE Specifications Range: The Ammonium Electrode responds to uncomplexed ion activity over the range 1 x 10-1M to 1 x 10-6M. Linear detection limit is about 10-5M Interference's:

More information

MASS TRANSFER AND GAS ABSORPTION EQUIPMENT

MASS TRANSFER AND GAS ABSORPTION EQUIPMENT MASS TRANSFER AND GAS ABSORPTION EQUIPMENT Mark J McCready University of Notre Dame Indiana, USA TOPICS Review of heat transfer and heat exchangers Some fundamental aspects of mass transfer Analysis of

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

2B Technologies, Inc. An InDevR Company

2B Technologies, Inc. An InDevR Company 2B Technologies, Inc. An InDevR Company Technical Note No. 40 UV-Absorbing Interferences in Ozone Monitors Date: 22 April 2015 Author: John Birks Background Ozone measurements by absorbance of the 253.7-nm

More information

Photochemistry of Nitro-Polycyclic Aromatic Hydrocarbons. August 9, 2011

Photochemistry of Nitro-Polycyclic Aromatic Hydrocarbons. August 9, 2011 Photochemistry of Nitro-Polycyclic Aromatic Hydrocarbons August 9, 2011 Photochemistry focuses on the absorption of light by atoms and molecules leading to chemical reactions. With the absorption of light,

More information

Chromatography & instrumentation in Organic Chemistry

Chromatography & instrumentation in Organic Chemistry Chromatography & instrumentation in Organic Chemistry What is Chromatography? Chromatography is a technique for separating mixtures into their components in order to analyze, identify, purify, and/or quantify

More information

Lab 5: Calculating an equilibrium constant

Lab 5: Calculating an equilibrium constant Chemistry 162 The following write-up is inaccurate for the particular chemicals we are using. Please have all sections up through and including the data tables ready before class on Wednesday, February

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law The primary objective of this experiment is to determine the concentration of an unknown cobalt (II) chloride solution. You will use a Vernier SpectroVis

More information

Carbon dioxide removal processes by alkanolamines in aqueous organic solvents Hamborg, Espen Steinseth

Carbon dioxide removal processes by alkanolamines in aqueous organic solvents Hamborg, Espen Steinseth University of Groningen Carbon dioxide removal processes by alkanolamines in aqueous organic solvents Hamborg, Espen Steinseth IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's

More information

Physical Separations and Chromatography

Physical Separations and Chromatography Lab #5A & B: Physical Separations and Chromatography Individual Objectives: At the end of these experiments you should be able to: Ø Distinguish between Rf and tr; chromatograph and chromatogram; adsorption

More information

WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY

WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY University of Puget Sound Department of Chemistry Chem 230 WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY INTRODUCTION Nitrogen is an essential nutrient

More information

Chemistry 141 Laboratory Spectrometric Determination of Iron Concentration Lab Lecture Notes 8/29/2011 Dr. Abrash

Chemistry 141 Laboratory Spectrometric Determination of Iron Concentration Lab Lecture Notes 8/29/2011 Dr. Abrash Chemistry 141 Laboratory Spectrometric Determination of Iron Concentration Lab Lecture Notes 8/29/2011 Dr. Abrash What is the purpose of this experiment? We re going to learn a way to quantify the amount

More information

EXPERIMENT 6 INTRODUCTION TO SPECTROSCOPY

EXPERIMENT 6 INTRODUCTION TO SPECTROSCOPY EXPERIMENT 6 INTRODUCTION TO SPECTROSCOPY INTRODUCTION Much of what we know about the structures of atoms and molecules has been learned through experiments in which photons (electromagnetic radiation

More information

Distillation of Liquids: Separation of 2-Propanol from Water by Fractional Distillation

Distillation of Liquids: Separation of 2-Propanol from Water by Fractional Distillation Distillation of Liquids: Separation of 2-Propanol from Water by Fractional Distillation Introduction: Distillation is the process of vaporizing a liquid, condensing the vapor, and collecting the condensate

More information

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes M. W. Teague*, Tonya Felix*, M. K. Hudson, and R. Shanks *Department of Chemistry, Hendrix College, Conway, AR

More information

ph Probe ReallyEasyData com

ph Probe ReallyEasyData com ph Probe 9200001 Uses Whether you re monitoring ph for chemistry, physical science, life science, or earth science activity, this ph meter offers accurate readings in a convenient format. Use the RED ph

More information

UNIT 4 Part 2 Measurements of Environmental Air Pollution Parameters j

UNIT 4 Part 2 Measurements of Environmental Air Pollution Parameters j UNIT 4 Part 2 Measurements of Environmental Air Pollution Parameters j 1 Environmental Air Pollution The addition of any harmful material (having harmful effects on our lives) to our atmosphere is called

More information

Sampling. Information is helpful in implementing control measures for reducing pollutant concentration to acceptable levels

Sampling. Information is helpful in implementing control measures for reducing pollutant concentration to acceptable levels Types of pollutant sampling and measurement: Air quality monitoring: Sampling and measurement of air pollutants generally known, as air quality monitoring. It is an integral component of any air pollution

More information

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,

More information

Rule 2. Rule 1. Rule 4. Rule 3. Rule 5. Rule 6. Rule 7. Rule 8

Rule 2. Rule 1. Rule 4. Rule 3. Rule 5. Rule 6. Rule 7. Rule 8 Rule 1 Follow the directions in your course reader, of your teaching assistant and of your instructor. They are usually much more experienced doing chemistry. Rule 3 When in doubt, ask. This will make

More information

Physics 197 Lab 11: Spectrometer

Physics 197 Lab 11: Spectrometer Physics 197 Lab 11: Spectrometer Equipment: Item Part # Qty per Team # of Teams Red Tide Spectrometer Vernier V-Spec 1 7 7 Computer with Logger Pro 1 7 7 Optical Fiber Assembly For Red Tide 1 7 7 Ring

More information

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Ammonium Ion Electrode is used to quickly, simply, accurately, and economically

More information

8 Enthalpy of Reaction

8 Enthalpy of Reaction E x p e r i m e n t Enthalpy of Reaction Lecture and Lab Skills Emphasized Calculating the heat and enthalpy of reactions. Writing net ionic equations. Using Hess s law to determine the enthalpy of a reaction.

More information

(Refer Slide Time: 00:00:43 min) Welcome back in the last few lectures we discussed compression refrigeration systems.

(Refer Slide Time: 00:00:43 min) Welcome back in the last few lectures we discussed compression refrigeration systems. Refrigeration and Air Conditioning Prof. M. Ramgopal Department of Mechanical Engineering Indian Institute of Technology, Kharagpur Lecture No. # 14 Vapour Absorption Refrigeration Systems (Refer Slide

More information

Experiment 12: SPECTROSCOPY: EMISSION & ABSORPTION

Experiment 12: SPECTROSCOPY: EMISSION & ABSORPTION Sample Experiment 12: SPECTROSCOPY: EMISSION & ABSORPTION Purpose: Emission and absorption spectroscopy is to be explored from different perspectives in a multipart experiment. Part I: Certain elements

More information

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal. L A B O R A T O R Y C O M P E T E N C E Compact Knowledge: Absorbance Spectrophotometry Flexible. Reliable. Personal. The interaction of light with molecules is an essential and well accepted technique

More information

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at 2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Advanced Pharmaceutical Analysis

Advanced Pharmaceutical Analysis Lecture 2 Advanced Pharmaceutical Analysis IR spectroscopy Dr. Baraa Ramzi Infrared Spectroscopy It is a powerful tool for identifying pure organic and inorganic compounds. Every molecular compound has

More information

Measurement Technique and its Application to Trace Components of Atmospheric Gas

Measurement Technique and its Application to Trace Components of Atmospheric Gas F e a t u r e A r t i c l e Feature Article Measurement Technique and its Application to Trace Components of Atmospheric Gas Junji Kato From early on, HORIBA has been developing and adopting various measurement

More information

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur 603 203 DEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING QUESTION BANK V SEMESTER EI6501 Analytical Instruments Regulation 2013 Academic

More information

Brooklyn College Department of Chemistry

Brooklyn College Department of Chemistry Brooklyn College Department of Chemistry Instrumental Analysis (Chem 42/790) Atomic Absorption Spectroscopy An atomic absorption spectrometer is used in this experiment to analyze a copper-base alloy for

More information

Corresponding Author: Jason Wight,

Corresponding Author: Jason Wight, COMPARISON OF COMBUSTION, CHEMICAL, AND NEAR-INFRARED SPECTROSCOPIC METHODS TO DETERMINE SOIL ORGANIC CARBON Jason P Wight, Fred L. Allen, M. Z., Zanetti, and T. G. Rials Department of Plant Sciences,

More information

Determination of an Equilibrium Constant

Determination of an Equilibrium Constant Last updated 1/29/2014 - GES Learning Objectives Students will be able to: Determine the numerical value of an equilibrium constant from measured concentrations of all reaction species. Use an absorption

More information

STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION. Price: $5.00. Revised March 31, 1997

STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION. Price: $5.00. Revised March 31, 1997 STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION Revised March 31, 1997 Price: $5.00 a. an arrangement where the water pipe to be injected with chlorine enters

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability UV Hound Series Portable UVDOAS Multi-gas Analyzers Multi-Gas Capability Dependable High Quality Air Monitoring Individual VOCs PPB Sensitivity Non-Contact Optical Measurement Automated Reporting Configurable

More information

Experiment 7A ANALYSIS OF BRASS

Experiment 7A ANALYSIS OF BRASS Experiment 7A ANALYSIS OF BRASS FV 10/21/10 MATERIALS: Spectronic 20 spectrophotometers, 2 cuvettes, brass sample, 7 M HNO 3, 0.100 M CuSO 4, 2 M NH 3, two 50 ml beakers, 100 ml beaker, two 25 ml volumetric

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

A Study of Beer s Law Prelab

A Study of Beer s Law Prelab 1. What is the purpose of this experiment? A Study of Beer s Law Prelab 2. Using the absorbance versus wavelength curve given in Figure I, determine the approximate value of max of the dye used to construct

More information

Validating Slope Spectroscopy Methods: A Formula for Robust Measurements

Validating Slope Spectroscopy Methods: A Formula for Robust Measurements CWP0802261A White Paper Validating Slope Spectroscopy Methods: A Formula for Robust Measurements February 26, 2008 I-Tsung Shih, PhD Mark Salerno 1.0 Abstract Demands on measurement systems are ever increasing.

More information

Distillation is a method of separating mixtures based

Distillation is a method of separating mixtures based Distillation Distillation is a method of separating mixtures based on differences in their volatilities in a boiling liquid mixture. Distillation is a unit operation, or a physical separation process,

More information

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Standardizing a Solution of Sodium Hydroxide. Evaluation copy Standardizing a Solution of Sodium Hydroxide Computer 6 It is often necessary to test a solution of unknown concentration with a solution of a known, precise concentration. The process of determining the

More information

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013 Absorbance Experiment 13 Dilutions and Data Handling in a Spreadsheet rev 1/2013 GOAL: This lab experiment will provide practice in making dilutions using pipets and introduce basic spreadsheet skills

More information

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15 Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15 Abstract: This lab was performed to synthesize acetyl salicylic acid or aspirin from a carboxylic acid and an alcohol. We had learned in class

More information

UV Quality Assurance. Dark Current Correction

UV Quality Assurance. Dark Current Correction UV Quality Assurance This document presents the Quality Assurance protocol used for the measurement of broad spectrum UV spectra according to the guidelines given by the United States Environmental Protection

More information

Optical In-line Control of Web Coating Processes with UV-VIS-NIR Spectroscopy. AIMCAL Europe Web Coating Conference Peter Lamparter

Optical In-line Control of Web Coating Processes with UV-VIS-NIR Spectroscopy. AIMCAL Europe Web Coating Conference Peter Lamparter Optical In-line Control of Web Coating Processes with UV-VIS-NIR Spectroscopy AIMCAL Europe Web Coating Conference 2012 13.6.2012 Peter Lamparter Optical In-line Control of Web Coating Processes with UV-VIS-NIR

More information

SEPARATION TECHNIQUES

SEPARATION TECHNIQUES SEPARATION TECHNIQUES If a substance does not dissolve in a solvent, we say that it is insoluble. For example, sand does not dissolve in water it is insoluble. Filtration is a method for separating an

More information

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of

More information

FEATURE ARTICLE. The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process. Takaaki Yada.

FEATURE ARTICLE. The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process. Takaaki Yada. FEATURE ARTICLE The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process Takaaki Yada Cleaning Cleaning Cleaning Cleaning Wet process Wet Etching process SPM Organic

More information

FLOW-THROUGH CAPACITOR ASSEMBLY FOR THE TREATMENT OF A. Field of application

FLOW-THROUGH CAPACITOR ASSEMBLY FOR THE TREATMENT OF A. Field of application FLOW-THROUGH CAPACITOR ASSEMBLY FOR THE TREATMENT OF A FLUID Field of application The present invention regards a flow-through capacitor assembly for the treatment of a fluid, according to the preamble

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES Exercise 2-2 Titration of a Strong Acid EXERCISE OBJECTIVES To describe the effect of a ph variation on a chemical indicator; To titrate water containing a strong base solution with a strong acid solution;

More information

Figure 4-1: Pretreatment schematic

Figure 4-1: Pretreatment schematic GAS TREATMENT The pretreatment process consists of four main stages. First, CO 2 and H 2 S removal stage which is constructed to assure that CO 2 would not exceed 50 ppm in the natural gas feed. If the

More information

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing Automated Preparation of Simulated Distillation Standards and Samples for ASTM Methods D2887, D7213, D7398 and D6352 using the 7693A System with Easy SamplePrep Software Author Roger L. Firor Agilent Technologies,

More information

Benefits of Heat Stable Salts in Tail Gas Treaters

Benefits of Heat Stable Salts in Tail Gas Treaters Benefits of Heat Stable Salts in Tail Gas Treaters Clayton E. Jones 1, R. Scott Alvis 2, Ralph H. Weiland 2 Optimized Gas Treating, Inc. 1 Technical Development Centre, 112 Cimarron Park Loop, Suite A,

More information

Aspirin Synthesis H 3 PO 4

Aspirin Synthesis H 3 PO 4 Aspirin Synthesis Experiment 10 Aspirin is the common name for the compound acetylsalicylic acid, widely used as a fever reducer and as a pain killer. Salicylic acid, whose name comes from Salix, the willow

More information

Design of A chemical Absorption System for The Separation of Propane/Propylene Mixture

Design of A chemical Absorption System for The Separation of Propane/Propylene Mixture Design of A chemical Absorption System for The Separation of Propane/Propylene Mixture Reda Zein, Ahmed F. Nassar, Tarek M. Mostafa Chemical Engineering Department Cairo University Giza Egypt reda.zein@eng1.cu.edu.eg

More information

Physics P202, Lab #12. Rydberg s Constant

Physics P202, Lab #12. Rydberg s Constant Physics P0, Lab #1 Rydberg s Constant The light you see when you plug in a hydrogen gas discharge tube is a shade of lavender, with some pinkish tint at a higher current. If you observe the light through

More information

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25 High Performance Liquid Chromatography (HPLC) Harris Chapter 25 12/1/2005 Chem 253 - Chapter 25 1 HPLC Separation of nonvolatile or thermally unstable compounds. If the analyte/sample can be found to be

More information

S reacts with the catalyst s metallic surface to substitute sulfur atoms for oxygen atoms.

S reacts with the catalyst s metallic surface to substitute sulfur atoms for oxygen atoms. Application Summary Analytes: S (hydrogen sulfide), sulfiding agent Detector: OMA-300 S Analyzer Process Stream: reactor effluent gas Introduction Catalyst presulfiding is a practice which reduces the

More information

Lecture (9) Reactor Sizing. Figure (1). Information needed to predict what a reactor can do.

Lecture (9) Reactor Sizing. Figure (1). Information needed to predict what a reactor can do. Lecture (9) Reactor Sizing 1.Introduction Chemical kinetics is the study of chemical reaction rates and reaction mechanisms. The study of chemical reaction engineering (CRE) combines the study of chemical

More information

TRACERCO Diagnostics FCCU Study

TRACERCO Diagnostics FCCU Study TRACERCO Diagnostics FCCU Study 2 TRACERCO Diagnostics TRACERCO Diagnostics FCCU Study Introduction The Fluidized Catalytic Cracking Unit (FCCU) is the economic heart of a modern refinery. Even small increases

More information

MEASUREMENT: PART II

MEASUREMENT: PART II 1 MEASUREMENT: PART II Copyright: Department of Chemistry, University of Idaho, Moscow, ID 83844-2343, 2013. INTRODUCTION Read and/or review Section 1.7 and Figure 7.5 in your textbook. The first part

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Systems Engineering Spring Group Project #1: Process Flowsheeting Calculations for Acetic Anhydride Plant. Date: 2/25/00 Due: 3/3/00

Systems Engineering Spring Group Project #1: Process Flowsheeting Calculations for Acetic Anhydride Plant. Date: 2/25/00 Due: 3/3/00 10.551 Systems Engineering Spring 2000 Group Project #1: Process Flowsheeting Calculations for Acetic Anhydride Plant Date: 2/25/00 Due: 3/3/00 c Paul I. Barton, 14th February 2000 At our Nowhere City

More information

THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION

THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION JPACSM 5 Process Analytical Chemistry THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION Peter Traynor and Steven Trimuar Thermo Electron Corporation Houston, TX Robert G. Wright Thermo

More information

1901 Application of Spectrophotometry

1901 Application of Spectrophotometry 1901 Application of Spectrophotometry Chemical Analysis Problem: 1 Application of Spectroscopy Organic Compounds Organic compounds with single bonds absorb in the UV region because electrons from single

More information

UV3000. Accurate, precise, and portable ambient gas point analyzer

UV3000. Accurate, precise, and portable ambient gas point analyzer UV3000 Accurate, precise, and portable ambient gas point analyzer The Cerex UV3000 is a multifunction analyzer designed to detect part per billion (ppb) to percent level concentrations of multiple gases

More information

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C LEADX.01-1 LEAD (Colorimetric) PRINCIPLE SCOPE Organic matter in the sample is destroyed by ignition in the presence of sulfuric acid. The residue is dissolved in dilute acid, and the lead is complexed

More information

CHAPTER 3 MODELLING AND ANALYSIS OF THE PACKED COLUMN

CHAPTER 3 MODELLING AND ANALYSIS OF THE PACKED COLUMN 37 CHAPTER 3 MODELLING AND ANALYSIS OF THE PACKED COLUMN Absorption in a chemical process refers to a mass transfer between gas and liquid which transfers one or more components from the gas phase to the

More information

Experiment 9. Emission Spectra. measure the emission spectrum of a source of light using the digital spectrometer.

Experiment 9. Emission Spectra. measure the emission spectrum of a source of light using the digital spectrometer. Experiment 9 Emission Spectra 9.1 Objectives By the end of this experiment, you will be able to: measure the emission spectrum of a source of light using the digital spectrometer. find the wavelength of

More information

The Synthesis and Analysis of Aspirin

The Synthesis and Analysis of Aspirin The Synthesis and Analysis of Aspirin Computer 22 Aspirin, the ubiquitous pain reliever, goes by the chemical name acetylsalicylic acid. One of the compounds used in the synthesis of aspirin is salicylic

More information

SYNTHESIS OF 1-BROMOBUTANE Experimental procedure at macroscale (adapted from Williamson, Minard & Masters 1 )

SYNTHESIS OF 1-BROMOBUTANE Experimental procedure at macroscale (adapted from Williamson, Minard & Masters 1 ) SYNTHESIS OF 1-BROMOBUTANE Experimental procedure at macroscale (adapted from Williamson, Minard & Masters 1 ) Introduction 1-bromobutane is a primary alkyl halide (primary alkyl) and therefore it is produced

More information

2 Copyright Pearson Education, Inc., or its affiliates. All Rights Reserved.

2 Copyright Pearson Education, Inc., or its affiliates. All Rights Reserved. CHEMISTRY & YOU Chapter 17 Thermochemistry 17.1 The Flow of Energy 17. Measuring and Expressing Enthalpy Changes 17.3 Heat in Changes of State 17.4 Calculating Heats of Reaction Why does sweating help

More information

EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW

EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW INTRODUCTION Chemical changes are generally accompanied by energy changes; energy is absorbed or evolved, usually as heat. Breaking chemical bonds in reactants

More information

OPTICAL GAIN AND LASERS

OPTICAL GAIN AND LASERS OPTICAL GAIN AND LASERS 01-02-1 BY DAVID ROCKWELL DIRECTOR, RESEARCH & DEVELOPMENT fsona COMMUNICATIONS MARCH 6, 2001 OUTLINE 01-02-2 I. DEFINITIONS, BASIC CONCEPTS II. III. IV. OPTICAL GAIN AND ABSORPTION

More information

Advancements in Tracer Flow Testing; Development of Real-Time Technology for Flow and Enthalpy Measurements

Advancements in Tracer Flow Testing; Development of Real-Time Technology for Flow and Enthalpy Measurements Proceedings World Geothermal Congress 2010 Bali, Indonesia, 25-29 April 2010 Advancements in Tracer Flow Testing; Development of Real- Technology for Flow and Enthalpy Measurements Matthew Broaddus, J.

More information

Lab Investigation 4 - How could you make more of this dye?

Lab Investigation 4 - How could you make more of this dye? Lab Investigation 4 - How could you make more of this dye? USING SPECTROSCOPY TO DETERMINE SOLUTION CON- CENTRATION Guiding Question How could you make more of this dye? INTRODUCTION A solution is a homogeneous

More information

25 Instruments for Optical Spectrometry

25 Instruments for Optical Spectrometry 25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.

More information

UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water

UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water C. H. Tan a, Y. C. Moo a, M. Z. Matjafri a and H. S. Lim a a School of Physics, Universiti Sains Malaysia, 118 Pulau Pinang, Malaysia.

More information