MEASURING H 2 S IN CRUDE OIL FOR QUALITY CONTROL & TRANSPORTATION SAFETY
|
|
- Muriel Marshall
- 6 years ago
- Views:
Transcription
1 MEASURING IN CRUDE OIL FOR QUALITY CONTROL & TRANSPORTATION SAFETY COQA Meeting 2015 Marriott West Loop, Houston Wesley Kimbell Analytical Systems Keco
2 Why measure in crude oil? Safety Product Quality
3 Why measure in crude oil? Safety Personnel protection Transportation Product Quality
4 Why measure in crude oil? Safety Product Quality Personnel protection Transportation Off-spec products Poisoned catalysis Environmental concern
5 Why measure in crude oil? Safety Personnel protection 1. gas is toxic and deadly 2. Even dissolved in crude is deadly
6 CHARACTERISTICS Gas with rotten egg smell Specific gravity: vs air LEL 4.3%, UEL 46% Highly corrosive Toxic
7 Why measure in crude oil? Safety Transportation 1. Prevent car derailments 2. Corrosion in vessels Pipelines Railcars Holding tanks Truck tanks
8 Why measure in crude oil? Safety Product Quality Transportation Personnel protection Off-spec products Poisoned catalysts Environmental concern
9 Why measure in crude oil? Product Quality Control 1. Off-spec products
10 Why measure in crude oil? Product Quality Control 1. Off-spec products 2. Poisoned catalysts
11 Why measure in crude oil? Product Quality Control 1. Off-spec products 2. Poisoned catalysts 3. Environmental concern
12 How do we measure H2S in crude oil? Direct optical measurement difficult: Crude oil is opaque in nature Complex compositional characteristics High viscosity crude oils difficult to transport in flow lines/tubing
13 How do we measure H2S in crude oil? Question: What is the best method to measure H2S in liquid samples like crude oil?
14 How do we measure H2S in crude oil? Answer: Transfer from liquid sample to headspace Upon saturation, the headspace concentration and the liquid sample concentration are proportional (Henry s Law)
15 How do we measure H2S in crude oil? Henry s Law: "At a constant temperature, the amount of a given gas that dissolves in a given type and volume of liquid is directly proportional to the partial pressure of that gas in equilibrium with that liquid."
16 Method of sampling Headspace Headspace concentration function of: 1. Saturation time/volume 2. Temperature 3. Vapor pressure of analyte 4. Concentration of analyte
17 Method of sampling Headspace
18 Method of sampling Headspace Glass Container Headspace Proportional concentration Crude Oil Zoomed to molecular level
19 Method of sampling Headspace Goal: Extract the headspace vapor of the liquid sample for analysis in the gas phase Solution: Headspace sampling Sample Transfer Stripper with ASI Membrane Technology (proprietary)
20 Method of sampling Headspace
21 Method of sampling Headspace
22 Method of sampling Headspace (dynamic) Purge Air In Headspace vapor to STS / Detector Dip tube REMEMBER: Headspace concentration is a function of: 1. Saturation time/volume 2. Temperature 3. Vapor pressure of analyte 4. Concentration of analyte
23 Method of sampling Headspace Headspace vapor can be dirty Mists will coat the detector and detector flow lines Results in in-gassing and out-gassing Leads to inaccurate readings over a short period of time
24 Method of sampling Headspace Clean the headspace vapor before analysis Significantly increase accuracy Significantly decrease maintenance
25 Method of sampling Membrane technology (proprietary) Sample Transfer Stripper with ASI Membrane Technology 5
26 Method of sampling Membrane technology (proprietary)
27 Method of sampling Membrane technology (proprietary) Membrane physically blocks headspace vapor mists Prevents fouling of detector Prevents liquid carry-over 5 Radically reduced maintenance Maintains accuracy of analyzer Maintains composition of headspace vapor
28 Method of sampling Membrane technology (proprietary) Headspace ASI Membrane Technology Dirty Crude oil vapor Zoomed to molecular level
29 Method of Analysis Difficulty of quantifying H2S in crude oil
30 Method of Analysis Difficulty of quantifying H2S in crude oil Crude oil is compositionally complex Aromatics make quantification difficult Common problematic components: SO2, ethyl-mercaptan, methylmercaptan and other sulfur compounds Crude compositional changes not uncommon
31 Method of Analysis Comparison Measurement Technology Colorimetric Ratiometric Tape Method - ASTM approved ASTM methods YES D , D , & D related Specific to H2S only YES Directly measures H2S without interference. No false positives. UV Absorbance Spectrophotometry X NO X NO Interference with mercaptans (naturally occurring), SO2 and other sulfurs. May give false positives. Requires complicated chemometric modeling but not full-proof Accuracy ± 2% of full scale typical ±10% at 0-10ppm FS ± 3-5% of FS Electrochemical Cell X NO X NO Interference with mercaptans, hydrogen, CO, ammonia, etc. giving false positives. Sensor Zero Drift Calibration Stability NO Does not suffer from zero drift YES Maintains accuracy despite severe process changes X YES 5% in 24 hours for low ppm ranges X NO Even slight process changes requires re-calibration. X YES X NO Requires monthly or bi-monthly calibrations Range PPB, PPM and up to 100% ranges with high accuracy X Low ppm ranges not reliable X Low ppm ranges not reliable Calibration Stability Long-term even in demanding environments and process changes X Unstable if process changes even slightly. Requires calibrations Maintenance Consumables LOW Tape change only 4 times per year LOW tape & solution change every 3 or 4 months only LOW Optics require frequent cleaning once a month or more X Light source and scrubber materials require replacement X HIGH Minimal NOTE: Data based on typical manufacturers published data
32 Method of Analysis Colorimetric Ratiometric Tape Detector
33 Method of Analysis Colorimetric Ratiometric Tape Detector The chemical formulation for this reaction is as follows: H 2 O Pb(OAc) 2 + PbS + 2HOAc White colored This reaction has three characteristics which allow it to be applied with unique analytical precision: 1. The reactants are colorless (white). 2. The product (PbS) is colored. 3. In more than a century of use, hydrogen sulfide is the only practical reactant found that produces a colored product on lead acetate (Pb(OAc) 2 tape.
34 Method of Analysis Colorimetric Ratiometric Tape Detector The rate (r) of this reaction (forming PbS) is as follows: r = k Pb(OAc) 2 x Where k, the proportionality factor, is called the rate constant. NOTE: The very large (more than 1000 times) concentration of Pb(OAc) 2 on the sensing tape simplifies the reaction as follows: [Pb(OAc) 2 ] 0 = [Pb(OAc) 2 ] 1 PbS 1 = K 2 (<.1%) Therefore, r n = kk 2 ( - PbS n ) For simplicity: kk 2 = K r n = K( = PbS n )
35 Method of Analysis Colorimetric Ratiometric Tape Detector By maintaining a constant concentration, with continuous flow of the sample into the sample chamber, the rate of reaction equation is further reduced to: r = K Therefore, by maintaining the control conditions of: 1. A large quantity of Pb(OAc) 2 2. Constant concentration. 3. Sufficient moisture. The rate of reaction (i.e. the rate of darkening) is linearly proportional concentration. Measuring the rate of darkening is therefore directly equivalent to measuring the concentration.
36 Method of Analysis Colorimetric Ratiometric Tape Detector The rate of reaction (i.e. the rate of darkening) is linearly proportional concentration.
37 Method of Analysis Colorimetric Ratiometric Tape Detector In actual practice, the rate of darkening is determined by measuring the change in darkness from second to second.
38 Method of Analysis Colorimetric Ratiometric Tape Detector 80 PPM tape darkness Y 40 PPM 0 b 0 X time The general equation for the lines of the graph above is: y = ax + b b is the zero drift or offset term in conventional equipment that is often adjusted to zero (zero drift)
39 Method of Analysis Colorimetric Ratiometric Tape Detector 80 PPM tape darkness Y 40 PPM 0 b 0 X time
40 Colorimetric Ratiometric Tape Detector Advantages Significant advantages of Tape Method Detector Only method inherently specific to No vapor processor or scrubbers required to block interfering components No complicated chemometric calibration modeling needed (spectrophotometry) Measures only Hydrogen Sulfide no matter the compositional background Sensor cannot be damaged by samples
41 Colorimetric Ratiometric Tape Detector Improvements to design Significant technological advancements to Tape Method: Improved tape drive system Reliable and rugged One tape reel > 150 analyses Fewer moving parts Simplest detection method available
S is referred to as sour. The sulfide loading lends an unpleasant odor and makes the water corrosive. Additionally, the presence of H 2
Application Summary Analytes: S (hydrogen sulfide), (ammonia) Detection Technology: OMA-300 Process Analyzer Process Stream: stripped sour water Introduction In the context of petrochemical refining and
More informationApplication Note S/SL
Analysis of Ultra low Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and Chemiluminescence according to ASTM D Ultra low detection limits Excellent Sensitivity, Repeatability &
More informationPros and Cons of Water Analysis Methods
Water Lens, LLC 4265 San Felipe, Suite 1100 Houston, Texas 77027 Office: (844) 987-5367 www.waterlensusa.com Pros and Cons of Water Analysis Methods Prepared by: Adam Garland, CTO Water Lens, LLC ICP-MS/OES
More information[RNH]SO3CH3 [RNH]SO 4 CH [RNH]I
Product bulletin Karl Fischer Tritration The basics of Karl Fischer Tritration Karl Fischer titration is an analytical technique to measure the amount of water contained in various samples, which may be
More informationUnderstanding ph Troubleshooting and Diagnostic information
Understanding ph Troubleshooting and Diagnostic information The desire is to achieve an accurate, reliable measurement with a reasonable electrode life expectancy while minimizing frequency or complexity
More informationCyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations
Cyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations Introduction Fluid catalytic cracking (FCC) is a major unit operation in refineries around the world. FCC is used to convert
More informationUV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability
UV Hound Series Portable UVDOAS Multi-gas Analyzers Multi-Gas Capability Dependable High Quality Air Monitoring Individual VOCs PPB Sensitivity Non-Contact Optical Measurement Automated Reporting Configurable
More information3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement
3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing, but also as an energy carrier in the cooling or heating cycle. As vast amounts
More information2Fe 2 O 3 +3H 2 S FeS+FeS x +S+3H 2 O
Elemental analysis of hydrocarbon streams using Dry colorimetry analyzers, a catalyst saviour Bushra Dawood, Application Coordinator C.I. Analytics www.cianalytics.com The Petrochemical industry has refined
More informationMonitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2
Application Summary Analytes: (nitrogen trichloride) Detector: OMA-300 Process Analyzer Process Stream: headspace gas above saturated CCl 4 solvent Typical Measurement Range: 0-20,000 ppm Introduction
More informationSpectrophotometric Determination of Iron
Spectrophotometric Determination of Iron INTRODUCTION Many investigations of chemical species involve the interaction between light and matter. One class of these investigations, called absorbance spectrophotometry,
More informationCatalytic bead sensors are used primarily to detect
Chapter 3 Catalytic Combustible Gas Sensors Catalytic bead sensors are used primarily to detect combustible gases. They have been in use for more than 50 years. Initially, these sensors were used for monitoring
More informationSOUR WATER ANALYSIS USING A DUAL- COLUMN HEADSPACE SAMPLING SYSTEM
SOUR WATER ANALYSIS USING A DUAL- COLUMN HEADSPACE SAMPLING SYSTEM Andy Burgess, David Greaves, Daniel Murphy and Sam Rolley Applied Analytics, Inc. 29 Domino Drive, Concord, MA 01742 KEYWORDS Sour Water,
More informationBe prepared to discuss the quantitative comparison method in the oral exam.
Subject: Ring Experiment III 8 Shell and Tube Heat Exchanger Control The shell and Tube Heat Exchanger has two control valves: one on the process fluid flowing to the tubes and one on the cooling water
More informationChapter 3 Electrochemical methods of Analysis-Potentiometry
Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator
More informationPermeation Measurement Testing Techniques. Michelle Stevens MOCON, Inc.
Permeation Measurement Testing Techniques Michelle Stevens MOCON, Inc. Permeation Permeation Permeation is simply the flux of molecules through a material normalized to the pressure drop across the film.
More informationSULFUR RECOVERY ANALYZERS; UV Diode-Array Fiber- Optics Technology
SULFUR RECOVERY ANALYZERS; UV Diode-Array Fiber- Optics Technology Yoav Barshad Applied Analytics, Inc. 29 Domino Drive Concord, Massachusetts, 01742 KEYWORDS Sulfur recovery, H2S, SO2, tail gas, UV, Spectroscopy,
More informationInstrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013
0510212 Instrumental Chemical Analysis Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 Chapter 1 Electroanalytical Methods Electroanalytical Chemistry Electroanalytical
More informationQuestions, Myths and Misconceptions about Using Photoionization Detectors
Questions, Myths and Misconceptions about Using Photoionization Detectors Solvent, fuel and other VOC vapours are pervasively common in many workplace environments. Increased awareness of the toxicity
More informationISA QATAR : 90 th Seminar
ANALYTICAL INSTRUMENTATION & MAINTENANCE SYSTEMS (AIMS) ISA QATAR : 90 th Seminar UTILIZATION OF LASERS IN THE FIELD OF GAS ANALYZERS Presenter: ZAHEER JUDDY Electromagnetic Spectrum Electromagnetic Spectrum
More information3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement
3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing, but also as an energy carrier in the cooling or heating cycle. As vast amounts
More informationThe Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)
The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of
More informationAlternative Monitoring Technologies: A Brief Overview
Alternative Monitoring Technologies: A Brief Overview a technical solution to meet every need Cemtek Environmental Inc. 3041 S. Orange Ave. Santa Ana, CA 92707 800-400-0200 www.cemteks.com 1 What s New?
More informationThe design and operational theory of ph electrodes is a very complex subject, explored only briefly here. What is important to understand is that thes
ph measurement A very important measurement in many liquid chemical processes (industrial, pharmaceutical, manufacturing, food production, etc.) is that of ph: the measurement of hydrogen ion concentration
More informationTheory of Headspace Sampling
Theory of Headspace Sampling Contents 1 Basics 2 2 Static headspace sampling 2 2.1 Preconcentration time and volume.......................... 3 2.2 Sample temperature..................................
More informationComparison of HCD5000 DewPoint Duo to Condumax II
EXECUTIVE SUMMARY This document compares the technology and products of ZEGAZ Instruments (US) to that of Michell Instruments (UK) in the field of hydrocarbon and water dewpoint measurement. INTRODUCTION
More informationThermo Scientific Electrochemistry Products
Comparing Performance of DO Methods The world leader in serving science in Water and Seawaters Thermo Scientific Electrochemistry Products Dissolved Oxygen Measurement Oxygen is essential for most of the
More informationUV3000. Accurate, precise, and portable ambient gas point analyzer
UV3000 Accurate, precise, and portable ambient gas point analyzer The Cerex UV3000 is a multifunction analyzer designed to detect part per billion (ppb) to percent level concentrations of multiple gases
More informationA New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores
Application Note 37890312 A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Keywords CNSolution 9310 On-line Cyanide Analyzer Cyanidation
More informationTHE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS
International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp
More informationThe Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer
The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS Agilent Technologies/Wasson ECE Monomer Analyzer Application Authors Fred Feyerherm 119 Forest Cove Dr. Kingwood, TX
More informationCollision Theory. Collision theory: 1. atoms, ions, and molecules must collide in order to react. Only a small number of collisions produce reactions
UNIT 16: Chemical Equilibrium collision theory activation energy activated complex reaction rate reversible reaction chemical equilibrium law of chemical equilibrium equilibrium constant homogeneous equilibrium
More informationSTATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016
APPLICATIONS Moisture monitoring of air or gas (SF6) Suitable for indoor or outdoor Industrial, medical or aerospace fields HIGHLIGHTS High voltage circuit breakers commonly used for distribution and transmission
More informationREACTION RATES AND EQUILIBRIUM
Name Date Class 18 REACTION RATES AND EQUILIBRIUM SECTION 18.1 RATES OF REACTION (pages 541 547) This section explains what is meant by the rate of a chemical reaction. It also uses collision theory to
More informationSERVOTOUGH SpectraScan
SERVOTOUGH SpectraScan Continuous, real-time analysis of light hydrocarbons Jan Hordijk - Servomex Hydrocarbon stream analysis 5 major technologies used in HydroCarbon land for gas: IR (Infra Red) Laser
More informationMaking Improvements to Today s Natural Gas Analysis Systems
Making Improvements to Today s Natural Gas Analysis Systems Hiroyuki Aikawa, Ryosuke Kamae, Yuki Hashi, PhD, Shimadzu (China) Co., Ltd. 1. Introduction In recent years, driven mainly by rising energy demand,
More informationHOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY
HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY Accurate temperature measurement is key for controlling the stability and repeatability of many temperature-critical processes.
More informationaa + bb cc + dd Equation 1
Experiment: The Determination of K eq for FeSCN 2+ Introduction For any reversible chemical reaction at equilibrium, the concentrations of all reactants and products are constant or stable. There is no
More informationMonitoring Flammable Vapors and Gases in Industrial Processes
Flammability Hazards Industrial fires and explosions happen more frequently than most people think. They cause downtime, property damage, injury and sometimes death. These fires and explosions result from
More informationMercury Vapour Analysers: Finding the Right Fit for Your Needs
Mercury Vapour Analysers: Finding the Right Fit for Your Needs The toxic effects of mercury vapour on human health have been well documented throughout the years. mercury detection with few interferences
More information1. Base your answer to the following question on information below and on your knowledge of chemistry.
1. Base your answer to the following question on information below and on your knowledge of A sample of nitric acid contains both ions and ions. This sample has a ph value of 1. Write a name of the positive
More informationOn-line XRF analysis of multiple elements in liquid process streams C-QUAND
An ideal tool for multiple element analysis using on-line XRF technology, from low ppm to percentage levels C-QUAND On-line XRF analysis of multiple elements in liquid process streams Introduction The
More informationAn Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples
An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples Abstract Color-Tec is a field-based analytical method developed by Ecology and
More informationExperiment 1: Thin Layer Chromatography
Experiment 1: Thin Layer Chromatography Part A: understanding R f values Part B: R f values & solvent polarity Part C: R f values & compound functionality Part D: identification of commercial food dye
More informationModule: 7. Lecture: 36
Module: 7 Lecture: 36 DIMETHYL FORMAMIDE INTRODUCTION Dimethylformamide is an organic compound and denotes as DMF. The name is derived from the fact that it is a derivative of formamide, the amide of formic
More informationMonomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Monomer Analysis Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Polymer Grade Monomer Analysis Monomer Analysis Impurities in feedstocks can adversely
More informationFast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.
Application Note 36720111 Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary PLOT Column Keywords Pulsed Flame Photometric
More informationNew advances in folded pathlength technology for Process Tunable Diode Laser Absorption Spectrometers (TDLAS)
New advances in folded pathlength technology for Process Tunable Diode Laser Absorption Spectrometers (TDLAS) Jean-Nicolas Adami, PhD Head of Strategic Product Group Gas Analytics Mettler-Toledo GmbH,
More informationFlow Solution FS 3700 Automated Chemistry Analyzer
Method Abstract Total Cyanide by ISO 14403 Document #42600116 Flow Solution FS 3700 Automated Chemistry Analyzer Total Cyanide by In-Line Ultraviolet Digestion, Gas Diffusion, and Photometric Detection,
More informationZirconium in Organic Applications
Zirconium in Organic Applications INTRODUCTION Laboratory studies and case histories have established zirconium as the most corrosion resistant material of construction in many production methods involving
More informationRS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers
RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed
More informationModule: 7. Lecture: 36
Module: 7 Lecture: 36 DIMETHYL FORMAMIDE INTRODUCTION Dimethylformamide is an organic compound and denotes as DMF. The name is derived from the fact that it is a derivative of formamide, the amide of formic
More informationBECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple
BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover
More informationTesting the Ammonia Content of Class F Fly Ash
2011 World of Coal Ash (WOCA) Conference May 9-12, 2011 in Denver, CO, USA http://www.flyash.info/ Testing the Ammonia Content of Class F Fly Ash Steven Roy Butler 1 1 Lafarge North America, Lockport,
More informationThe RAID-M had the highly sensitive TDI library in Library A of the instrument as shown in Table 1. Table 1: Highly sensitive TDI library in ppb.
Frank Thibodeau, Bruker Detection, Billerica, MA USA Franziska Lange, Bruker Detection, Leipzig, Germany INTRODUCTION Ion Mobility Spectrometry (IMS) is the scientific technique of measuring the drift
More informationTechnical/Application Article 01 Version 1.2
Technical/Application Article 01 Version 1.2 th 25 January 2017 WRH What is a PID? Introduction PID is the abbreviation for Photo-Ionization Detector. A PID is a hand-held, personal, or fixed wallmounted
More informationFlammable/Combustible Liquid Storage
8.12.1 RESPONSIBILITY University personnel are responsible for using approved and safe methods for storing flammable/combustible liquids. Personnel must store flammable and combustible liquids in accordance
More informationMonthly Report: Chevron Richmond Community Air Monitoring Program. Report Number: RCAMP_MO_11 Date: February RCAMP_MO_11 Page 1 of 26
Monthly Report: Chevron Richmond Community Air Monitoring Program Report Number: RCAMP_MO_11 Date: February 2014 RCAMP_MO_11 Page 1 of 26 Table of Contents Table of Contents... 2 Executive Summary... 4
More informationReprint. Project Summary. Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado
ChemScan Process Analyzer Reprint Project Summary Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado Published by: Applied Spectrometry Associates,
More informationShifting Equilibrium. Section 2. Equilibrium shifts to relieve stress on the system. > Virginia standards. Main Idea. Changes in Pressure
Section 2 Main Ideas Equilibrium shifts to relieve stress on the system. Some ionic reactions seem to go to completion. Common ions often produce precipitates. > Virginia standards CH.3.f The student will
More informationApril 23, 2015 Lab 1 - Gravimetric Analysis of a Metal Carbonate
Lab 1 - Gravimetric Analysis of a Metal Carbonate Purpose - to determine the identity of the cation in X 2 CO 3 by gravimetric analysis Lab Highlights Include: SLOW gravity filtration Primary Experimental
More informationQuestion 1: Solution 1:
Book Name: Selina Concise Question 1: Comment, sulphuric acid is referred to as: (a) King of chemicals (b) Oil of vitriol Solution 1: (a) Sulphuric acid is called King of Chemicals because there is no
More informationNitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)
Methodology Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) (Cartridge Part #A002603) 1.0 Scope and Application 1.1 This method is used to determine the concentration of nitrate
More informationA general statement governing all systems in a state of dynamic equilibrium follows:
Chapter 20 Experiment: LeChâtelier s Principle: Buffers OBJECTIVES: Study the effects of concentration and temperature changes on the position of equilibrium in a chemical system. Study the effect of strong
More informationCP Chemistry Study Guide Test 1 (Ch 1 and 2)
Key Concepts: CP Chemistry Study Guide Test 1 (Ch 1 and 2) What is alchemy and how did it influence chemistry? What are the steps to the scientific method? Why are communication and collaboration important
More informationDetection limit: grain, feed 500 ppb; milk 50 ppb; cream, cheese 5 ppb
Product information Background Deoxynivalenol (DON) Deoxynivalenol, called vomitoxin, is a toxic metabolite mainly produced by Fusarium graminearum. It is mainly found in wheat, barley, corn and feed.
More information3 H 2 (g) + N 2 (g) 2 NH 3 (g)
TPIC 7: EQUILIBRIUM Dynamic equilibrium chapter 16.1 Chemical equilibrium chapters 16.2 16.3, 16.5 pen system: 2 Na (s) + Cl 2 (g) 2 NaCl goes virtually to completion. 3 2 (g) + N 2 (g) 2 N 3 (g) 2 (l)
More informationIon Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015
Ion Selective Electrodes for the Laboratory Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Outline Review ISE measurement technology How to properly calibrate
More informationCORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG
Introduction CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG A mass flowmeter is a system that provides a measurement of
More informationChapter 6 Notes. Section 4-5 Nonmetals
Chapter 6 Notes Section 4-5 Nonmetals Section 4-5 Objectives 8) Contrast the reactivity and abundance of elements in the Group 3A through Group 8A. 9) Name an important element in each group of p-block
More informationPIOX / Analytical Measuring & Sales Methods. FLEXIM GmbH
PIOX / Analytical Measuring & Sales Methods PIOX - Process Analyzers PIOX S Non-intrusive Ultrasonic Measurement Concentration Mass Flow Density HCL Nitric Acid Sulfuric Acid PIOX R - Process Refract meter
More informationStandard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1
Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,
More informationReaction Rates and Equilibrium
CHAPTER 7 14 SECTION Chemical Reactions Reaction Rates and Equilibrium KEY IDEAS As you read this section, keep these questions in mind: How can you increase the rate of a reaction? What does a catalyst
More informationFlow Solution FS 3700 Automated Chemistry Analyzer
Method Abstract Total Cyanide by ASTM D7511 Document #43671116 Flow Solution FS 3700 Automated Chemistry Analyzer Total Cyanide by In-Line Ultraviolet Digestion and Amperometric Detection ASTM D7511, Segmented
More informationClassification of Matter
Matter in Chemistry Classification of Matter Elements: substances that are made from one type of atom only. Compounds: substances that are made from more than one type of atom chemically bonded together.
More informationFlow Monitoring Technologies in Oil and Gas. Don Ford, Technical Specialist Spartan Controls Ltd.
Flow Monitoring Technologies in Oil and Gas Don Ford, Technical Specialist Spartan Controls Ltd. Agenda Flow technologies at a glance Oil and Gas at a glance Select Flow Technologies in Depth Differential
More informationStudies of a Precipitation Reaction
Studies of a Precipitation Reaction Prelab Assignment Read the entire lab. Write an objective and any hazards associated with this lab in your laboratory notebook. Answer the following 6 questions in your
More informationExperiment 7: Adsorption Spectroscopy I, Determination of Iron with 1,10 Phenanthroline
Procedure Experiment 7: Adsorption Spectroscopy I, Determination of Iron with 1,10 Phenanthroline John Smith CHE226.001 4/10/95 In this experiment, we aim to determine the Na concentration of an unknown
More informationlems. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement task.
TOC the Chemical Industry TOC Measurement Application in Handbook Chemical Industry 3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing,
More informationICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer
Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full
More informationC (s) + O 2 (g) CO 2 (g) S (s) + O 2 (g) SO 2 (g)
Combustion The rapid combination of oxygen with a substance. A major type of chemical reaction. When elemental carbon or carbon-containing compounds burn in air, oxygen combines with the carbon to form
More informationChapter 15 Equilibrium
Chapter 15. Chemical Equilibrium Common Student Misconceptions Many students need to see how the numerical problems in this chapter are solved. Students confuse the arrows used for resonance ( )and equilibrium
More informationUnderstanding catalytic LEL combustible gas sensor performance
: Understanding catalytic LEL combustible gas sensor performance These four conditions are frequently diagrammed as the "Fire Tetrahedron". If any side of the tetrahedron is missing, incomplete or insubstantial;
More informationIdentification of an Unknown Compound through Mass Correlations
EXPERIMENT Identification of an Unknown Compound through Mass Correlations PURPOSE To carry out a series of decomposition reactions for five different unknown, and use stoichiometry in order to identify
More informationCHEM Principles of Chemistry II. Chapter 17 - Electrochemistry
CHEM 1212 - Principles of Chemistry II Chapter 17 - Electrochemistry electrochemistry is best defined as the study of the interchange of chemical and electrical energy 17.1 Galvanic Cells an oxidation-reduction
More informationCHEMISORPTION OF ARSINE ON ACTIVATED, IMPREGNATED CARBON ADSORBENTS
CHEMISORPTION OF ARSINE ON ACTIVATED, IMPREGNATED CARBON ADSORBENTS Tara Sewell, Alex Balboa, Gregory W. Peterson and Christopher J. Karwacki US Army Edgewood Chemical and Biological Center, Aberdeen Proving
More informationModule 20: Corrosion Control and Sequestering Answer Key (edited April 2017)
Module 20: Corrosion Control and Sequestering Answer Key (edited April 2017) Bureau of Safe Drinking Water, Department of Environmental Protection 1 UNIT 1 EXERCISE Exercise for Unit 1 Background and Properties
More informationIntroduction to electrochemistry
Introduction to electrochemistry Oxidation reduction reactions involve energy changes. Because these reactions involve electronic transfer, the net release or net absorption of energy can occur in the
More informationQualitative Analysis I - Cations
1 Qualitative Analysis I - Cations Purpose: To separate and identify several metal cations from a mixture of cations, and to analyze an unknown sample of cations Introduction Qualitative analysis is the
More informationCombustible Gas Catalytic Bead (0-100 %LEL) Part No FM Performance Certified 1,4 FM Performance Certified 1
Sensor Data Sheet Document No. 365-2211-31 (Rev D) Combustible Gas Catalytic Bead (0-100 %LEL) Part No. 823-0211-31 FM Performance Certified 1,4 FM Performance Certified 1 Minimum Indicated Concentration...
More informationDEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING SRM NAGAR, KATTANKULATHUR-603203 EI 2302 ANALYTICAL INSTRUMENTS QUESTION BANK UNIT I COLORIMETRY AND SPECTROPHOTOMETRY Part A 1. State Lambert
More informationCone differential pressure flow meter. Operating manual WIDE PLUS PRECISION INSTRUMENTS CO., LTD.
Cone differential pressure flow meter Operating manual WIDE PLUS PRECISION INSTRUMENTS CO., LTD. INDEX Introduction 2 Principle Application The characteristics of flow meter Specification Technical features
More informationChemistry. ANSWERS and MARKING SCHEME. Final Examination Preliminary Course General Instructions. Total Marks 64
ANSWERS and MARKING SCHEME Chemistry Final Examination Preliminary Course 2003 General Instructions Reading time 5 minutes Working time 120 minutes Write using black or blue pen Draw diagrams using pencil
More informationCHEMICAL EQUILIBRIUM. Chapter 15
Chapter 15 P a g e 1 CHEMICAL EQUILIBRIUM Examples of Dynamic Equilibrium Vapor above a liquid is in equilibrium with the liquid phase. rate of evaporation = rate of condensation Saturated solutions rate
More informationATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light
ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light by these elements in order to measure their concentration.
More informationi i ne. (1) i The potential difference, which is always defined to be the potential of the electrode minus the potential of the electrolyte, is ln( a
We re going to calculate the open circuit voltage of two types of electrochemical system: polymer electrolyte membrane (PEM) fuel cells and lead-acid batteries. To do this, we re going to make use of two
More informationAnswers to spectroscopy questions. 1. Consider the spectrum below. Questions a f refer to this spectrum.
Answers to spectroscopy questions. 1. Consider the spectrum below. Questions a f refer to this spectrum. a. Is the spectrum above a band spectrum or a line spectrum? This is a band spectra, there are what
More informationWHAT YOU NEED. Visible Spectrophotometer. Micropipettes. Cuvettes. Parafilm. Timer µL
WHAT YOU NEED Visible Spectrophotometer Micropipettes 10-100µL 100-1000µL Cuvettes Parafilm Timer WHAT TESTS CAN BE DONE Acetic Acid Amino Acid Nitrogen Ammonia Citric Acid Glucose & Fructose Lactic Acid
More informationH 2 O WHAT PROPERTIES OF WATER MAKE IT ESSENTIAL TO LIFE OF EARTH? Good solvent High Surface tension Low vapor pressure High boiling point
Unit 9: Solutions H 2 O WHAT PROPERTIES OF WATER MAKE IT ESSENTIAL TO LIFE OF EARTH? Good solvent High Surface tension Low vapor pressure High boiling point Water is a polar molecule. It experiences hydrogen
More informationHYDROCARBONS: STRUCTURE & PROPERTIES
YDROARBONS: STRUTURE & PROPERTIES PURPOSE: To discover the physical and chemical properties of alkanes, alkenes, and aromatic hydrocarbons. To identify an unknown hydrocarbon by comparing it to known samples.
More information