1 TECHNICAL DATASHEET OVIVO ANALYZER UREA 1601 Ovivo Switzerland AG Hauptstrasse 192 CH 4147 Aesch Switzerland Phone Fax

2 Analyzer UREA 1601 Vers Page 2 of 12

3 Table of Contents OVERVIEW... 4 TYPICAL SCREEN VIEWS... 7 IMPLEMENTATION, LAYOUT... 9 ORDERING INFORMATION WHERE TO FIND US Page 3 of 12 Analyzer UREA 1601 Vers

4 OVERVIEW Technical description of OVIVO urea analyzer The working principle of the system is based on a chromatographic separation by combination (coupling) of three specific properties of urea. 1. The first property is related to the non-ionic nature of urea. This is quite unique, as N-compounds are generally ionic, either anionic (e.g. nitrate) or cationic (e.g. amines). Thus, urea will have shorter retention time, if separated on a cation exchange liquid chromatographic column in respect to cationic compounds, or longer retention times in respect to an anionic exchange chromatographic column. Therefore, it is possible to isolate urea from other N-compounds by liquid chromatographic methods. 2. The oxidation behavior of urea: After exposition to short wave UV-light, urea will first undergo a disproportionation reaction, the breakdown to carbon dioxide and ammonium. If UV-wavelength and intensity is sufficient then ammonium will rapidly be oxidized to nitrite and then nitrate. Other organic compounds will be converted to carbon dioxide, organic N-compounds will be converted to carbon dioxide and nitrate. Heterocyclic compounds will also produce elemental nitrogen if nitrogen is present in a vicinal fashion. 3. The high molar extinction coefficient of nitrate at wavelengths < 230 nm. Due to this, nitrate can be detected and quantified at very low concentrations by spectrometric methods. Other salts found in natural or technical waters do not have this property; nitrate is almost (except e.g. bromine and iodine) the only species that is detected at short wave UV. Flow diagram of the OVIVO urea analyzer: The water sample passes through a multiplexer into the actual analyzer. The mobile phase (a salt buffer) is pre-cleaned by a special designed oxidation reactor (DOCOX). Then, a defined volume of aqueous sample, picked up by a loop, is inserted into the continuous stream of the mobile phase. It is thereafter fed to a HPLC pump and then to a chromatographic column in which urea is separated from other N-species, like nitrate and ammonium. After leaving the column the solution is fed to a short-wave UVoxidation lamp (DONOX) in which urea is converted to nitrate, while nitrate remains unaffected and ammonium is converted to nitrate. Thereafter the liquid is fed to a UVdetector (220 nm) and all nitrate is quantified. All species elute according to their specific retention times and retention time is used for identification while peak area units are used for quantification. Analyzer UREA 1601 Vers Page 4 of 12

5 ANALYZER FEATURES Automatic sampling after analysis is finished Designed for continuous operation (24/7) Low consumption of chemicals Minimal maintenance needs Wide range of urea concentration SOFTWARE FEATURES Automatic operation after initial start Automatic integration and interpretation Diagnostic options Options for manual injection of a sample DETECTORS Multi wavelength UV-Detector set to 220 nm using a Deuterium lamp Page 5 of 12 Analyzer UREA 1601 Vers

6 PERFORMANCE Quantification Range: 2 ppb.150 ppb as Urea Accuracy 5-10 % Repeatability 5% Number of sampling channels Max. 4 Volume per sample Average Sample time Operation mode OPERATING ENVIRONMENT Humidity Temperature Protection Class FACILITY REQUIREMENTS Phosphate Buffer: Sample pressure Drain KH 2 PO 4 (CAS ) Na 2 HPO 4 (CAS ) < 100 ml 60 min Automatic/selectable sequence or manual single measurements Online or grab sample analysis. 0-90% relative humidity, non-condensing 5-40 C ( F) IP 42 (NEMA 2) 0.65 ml/min 0.75 g/l 0.45 g/l Max. 10 bar (145 psi) Gravity Electrical Supply Power Consumption Inlet cable section 1.5 mm 2 (AWG 14) 115/230VAC 50/60Hz (Single Phase L, N, PE) 0.5 kw (0.75 hp) COMMUNICATION INTERFACE Signal Outputs Signal Inputs/Outputs Serial Interface Up to 4 Analog Outputs (active powered 24VDC current loop 0/4...20mA) Digital Interface for channel activation/deactivation, maintenance information, system alarms Not given (LAN for internal use only) HARDWARE Size (W x D x H) 656 x 653 x 1800 mm (25.8 x 25.7 x 70.9 ) Weight 155 kg (310 lbs) Analyzer UREA 1601 Vers Page 6 of 12

7 TYPICAL SCREEN VIEWS Screen 1 Main screen - overview of results from previous samples Page 7 of 12 Analyzer UREA 1601 Vers

8 TYPICAL SCREEN VIEWS Screen 2 actual measurement overview Screen 3 diagnostics overview Analyzer UREA 1601 Vers Page 8 of 12 Screen 4: diagnostics overview

9 IMPLEMENTATION, LAYOUT Front view Side View Page 9 of 12 Analyzer UREA 1601 Vers

10 IMPLEMENTATION, LAYOUT Required Space Required Bore Holes Analyzer UREA 1601 Vers Page 10 of 12

11 ORDERING INFORMATION Parts List OVIVO UREA 1601-A P S Measuring system ppb UREA based on Chromatographic Separation principle special version acc. to customer 's specification Sampling / Channels 1 1 Channel with direct on line process connection limited to 10 bar feed pressure 2 2 Channels with direct on line process connection limited to 10 bar feed pressure 3 3 Channels with direct on line process connection limited to 10 bar feed pressure 4 4 Channels with direct on line process connection limited to 10 bar feed pressure 6 4 Channels 6/4mm connection limited to < 0,5 bar feed pressure 7 4 Channels 1/4 inch connection limited to < 0,5 bar feed pressure S special version acc. to customer 's specification Power supply VAC / 50Hz VAC / 60Hz VAC / 50Hz VAC / 60Hz Communication 1 0/4...20mA and Modbus TCP 2 0/4...20mA and Profibus DP S special version acc. to customer 's specification Additional N No Performance Certificate C Performance Certificate S special version acc. to customer 's specification Maintenance OVIVO UREA 1601-M Set of UV Lamp DOCOX & DONOX OVIVO UREA 1601-M D2-Lamp for detector OVIVO UREA 1601-M Column TSK HW40S OVIVO UREA 1601-M UV Detector UVD OVIVO UREA 1601-M HPLC pump OVIVO UREA 1601-M Degasser Unit OVIVO UREA 1601-M Magnetic Valve Set (2 pieces) OVIVO UREA 1601-M Transformers Tools OVIVO UREA 1601-M Analyzer Operation and Maintenance Tool Set Calibration / Consumables OVIVO UREA 1601-M Calibration Kit for solution preparation OVIVO UREA 1601-M Pack of 12 C-EXR cartridges OVIVO UREA 1601-M Pack of 12 A-EXR cartridges Note: technical data are subject to changes without notice. Page 11 of 12 Analyzer UREA 1601 Vers

12 WHERE TO FIND US EUROPE Ovivo France S.A.S. Chemin des Fontaines Bernin France Phone +33 (0) info.fr@ovivowater.com Ovivo Germany GmbH Holderäckerstr Stuttgart Germany Phone +49-(0) info.de@ovivowater.com Ovivo Germany GmbH Regionalbüro Dresden Heidestraße 34B Dresden Germany Phone +49 (0) info.de@ovivowater.com Ovivo Holland BV Energieweg 1 Zoeterwoude 2382 Netherlands Phone + 31 (0) info.nl@ovivowater.com Ovivo Switzerland AG Hauptstrasse Aesch Switzerland Phone Info.CH@ovivowater.com ASIA Ovivo Shanghai Co., Ltd SOHO Bld. 8, #115. Lane 572 Bibo Road Shanghai, Pudong China Phone info.cn@ovivowater.com Ovivo Singapore Pte Ltd 10 Ang Mo Kio Street 65, #05-14 TechPoint Ang Mo Kio, Singapore Phone info.sg@ovivowater.com Ovivo Taiwan Co., Ltd No. 81, Schueli Road Hsin Chu City 300 Taiwan Phone Info.TW@ovivowater.com NORTH AMERICA Ovivo USA, LLC 1700 Alma, Suite 250 Plano, TX United States Phone info@ovivowater.com Ovivo USA, LLC th Ave, Suite D, Vancouver, Washington United States Phone info@ovivowater.com Ovivo USA, LLC Saratoga Technology + Energy Park 20 Tech Trail, Suite 101 Malta, NY United States Phone info@ovivowater.com Visit our website! Analyzer UREA 1601 Vers Page 12 of 12

TECHNICAL DATASHEET OVIVO ANALYZER THM1501. Ovivo Switzerland AG Hauptstrasse 192 CH 4147 Aesch Switzerland

TECHNICAL DATASHEET OVIVO ANALYZER THM1501. Ovivo Switzerland AG Hauptstrasse 192 CH 4147 Aesch Switzerland TECHNICAL DATASHEET OVIVO ANALYZER THM1501 Ovivo Switzerland AG Hauptstrasse 192 CH 4147 Aesch Switzerland Phone +41 61 755 85 11 Fax +41 61 755 85 56 www.ovivowater.com Analyzer THM1501 Vers. 201510 Page

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Marco Karsten, Bas, Dolman, Giovanni Maio, Frank Steiner, Holger Franz, Frank Arnold and Remco Swart LC Packings,

More information

How do you explain HPLC? 2 AZURA Educational System for tomorrow s HPLC professionals

How do you explain HPLC? 2 AZURA Educational System for tomorrow s HPLC professionals How do you explain HPLC? 2 AZURA Educational System for tomorrow s HPLC professionals AZURA Compact HPLC Analytical HPLC for the training of future HPLC professionals We believe that a lively and entertaining

More information

NEW!! Reliable Chlorine Monitoring. Online Residual Chlorine Monitor. Detailed Specifications. Clean Water through Innovative Technology

NEW!! Reliable Chlorine Monitoring. Online Residual Chlorine Monitor. Detailed Specifications. Clean Water through Innovative Technology Clean Water through Innovative Technology NEW!! Instrument Overview Reliable Chlorine Monitoring HF scientific developed the OnLine Residual Chlorine Monitor specifically for drinking water, wastewater

More information

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column Application Note 95 Determination of Hydrochloride Purity Using the Acclaim PA Column INTRODUCTION -based medications are prescribed for several heart and blood pressure indications. The fastacting formulations

More information

473-SHX Dew Point Mirror

473-SHX Dew Point Mirror Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable

More information

ABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5000 Temperature programmed gas chromatograph

ABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5000 Temperature programmed gas chromatograph ABB MEASUREMENT & ANALYTICS DATA SHEET PGC5000 Temperature programmed gas chromatograph 2 P G C 5 0 0 0 T E M P E R AT U R E P R O G R A M M E D D S/ P G C 5 0 0 0 T P - E N Measurement & Analytics The

More information

ION CHROMATOGRAPHY SYSTEM S 150

ION CHROMATOGRAPHY SYSTEM S 150 ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved

More information

ISEmax CAM40/CAS40. Technical Information

ISEmax CAM40/CAS40. Technical Information Technical Information Online measurement of nutrient parameters Ion-selective electrode system for the continuous measurement of ammonium and nitrate Application The ion-selective electrode system works

More information

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System G. Maio 1, F. Steiner 1, H. Franz 1, M. Tracy 2, X. Liu 2, F. Arnold 1 1 Dionex Softron GmbH, D-21 Germering,

More information

ADI 2040 Process Analyzer

ADI 2040 Process Analyzer ADI 2040 Process Analyzer Multi-purpose wet chemical analysis Multi-purpose Wet Chemical Analysis ADI 2040 The Metrohm Applikon ADI 2040 Process Analyzer is a multipurpose wet chemical analyzer, that has

More information

Information Technology Solutions Neptune Datasheet

Information Technology Solutions Neptune Datasheet Information Technology Solutions Neptune Datasheet Suspended Solids Analyzer Neptune is the first purpose built in-situ analyzer for measuring the suspended solids of slurries. Specifically developed for

More information

Agilent G2350A Atomic Emission Detector (AED)

Agilent G2350A Atomic Emission Detector (AED) Agilent G0A Atomic Emission Detector (AED) Specifications Elements Specifications Wavelength Minimum Detectable Selectivity Dynamic Measurement Group Element(s) (nm) Level (pg/sec) Over Carbon Range Compound

More information

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER Technical Description Version 2 BGRIMM AUTOMATION, South 4th Ring Road West, Beijing, China TEL: +86(10)59069762 Fax: +86(10)68360101 Contents BGRIMM AUTOMATION

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).

More information

Cation Exchange HPLC Columns

Cation Exchange HPLC Columns Cation Exchange HPLC Columns Hamilton offers seven polymeric packing materials for cation exchange separations. Type Recommended Application(s) PRP-X00 PRP-X00 PRP-X800 HC-0 HC-7 Ca + HC-7 H + HC-7 Pb

More information

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System Application Update 55 Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free System Introduction Hydrogen peroxide is an essential chemical in the fabrication

More information

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Trace Cations in Power Plant Waters Containing Morpholine Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion

More information

Nexera UC Unified Chromatography

Nexera UC Unified Chromatography Nexera UC Unified Chromatography The latest addition to the chromatography toolbox Dr. Gesa J. Schad Shimadzu Europa GmbH A brief history of SFC ϒ Late 1800 s: it was found that heavy, non-volatile organic

More information

FLASH CHROMATOGRAPHY

FLASH CHROMATOGRAPHY Turnkey Laboratories Solutions FLASH CHROMATOGRAPHY FC 3065 EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES Analytical Technologies Limited An ISO 9001 Certified Company

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 18 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

columns IonPac SCS 1 Silica Cation Separator

columns IonPac SCS 1 Silica Cation Separator columns IonPac SCS Silica Cation Separator The SCS Silica Cation Separator is designed for use with nonsuppressed conductivity detection, or single-column ion chromatography (SCIC). This column is particularly

More information

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS Jason Lai, Jia Wang, Brad Hart, Pavel Aronov, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Andy Jacobs, Haibo

More information

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Application Note Inorganic Ions, Water Testing, Minerals, Metals, Basic Chemicals Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Authors Anne Mack, Adam Bivens

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

UV3000. Accurate, precise, and portable ambient gas point analyzer

UV3000. Accurate, precise, and portable ambient gas point analyzer UV3000 Accurate, precise, and portable ambient gas point analyzer The Cerex UV3000 is a multifunction analyzer designed to detect part per billion (ppb) to percent level concentrations of multiple gases

More information

Dionex IonPac AS28-Fast-4µm column

Dionex IonPac AS28-Fast-4µm column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column

More information

ProPac WCX-10 Columns

ProPac WCX-10 Columns ProPac WCX-10 Columns Guidance for column use Tips to maximize column lifetime ProPac WCX-10 Column Tips and Tricks This guide provides essential information and invaluable guidelines for mobile phases,

More information

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis Application Note AN89.V.1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Page 1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Prior to LC-MS/MS Analysis Sample Preparation

More information

COMPREHENSIVE LC X LC ANALYSIS OF FLAVORING INGREDIENTS IN DISTILLED SPIRITS

COMPREHENSIVE LC X LC ANALYSIS OF FLAVORING INGREDIENTS IN DISTILLED SPIRITS PO-CON1574E COMPREHENSIVE LC X LC ANALYSIS OF FLAVORING INGREDIENTS IN DISTILLED SPIRITS HPLC 2015 PSB-MULTI-01 Gesa J. Dr. Schad; Brigitte Dr. Bollig; Robert Ludwig; Björn T. Dr. Erxleben Shimadzu Europa

More information

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation chromatography/size-exclusion chromatography (GPC/SEC)

More information

Monitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2

Monitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2 Application Summary Analytes: (nitrogen trichloride) Detector: OMA-300 Process Analyzer Process Stream: headspace gas above saturated CCl 4 solvent Typical Measurement Range: 0-20,000 ppm Introduction

More information

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,

More information

Chromatographic Analysis

Chromatographic Analysis Chromatographic Analysis Distribution of Analytes between Phases An analyte is in equilibrium between the two phases [S 1 ] [S 2 ] (in phase 1) (in phase 2) AS [S2 ] K 2 A S [S1 ] 1 AS, A 1 S Activity

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

CCS140 and CCS141. Technical Information. Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly

CCS140 and CCS141. Technical Information. Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly Technical Information CCS140 and CCS141 Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly Application Oxidising agents such as chlorine or anorganic

More information

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20544 Key Words Hydrophilic, HILIC,

More information

columns IonPac AS17-C Anion Exchange Column

columns IonPac AS17-C Anion Exchange Column columns IonPac AS-C Anion Exchange Column The IonPac AS-C is a hydroxide-selective anion exchange column designed for fast gradient separation of inorganic anions. The key application for the AS-C column

More information

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 15 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof

More information

Zirconia: the Ideal Substrate for Ion-Exchange LC and LC-MS

Zirconia: the Ideal Substrate for Ion-Exchange LC and LC-MS Zirconia: the Ideal Substrate for Ion-Exchange LC and LC-MS EAS 2005 Bingwen Yan 1, Clayton V. McNeff 1, Richard A. Henry 2, and David S. Bell 2 1 ZirChrom Separations, Inc., 617 Pierce Street, Anoka,

More information

Fast Separation of Vastly Different Compounds by Isocratic HPLC

Fast Separation of Vastly Different Compounds by Isocratic HPLC Fast Separation of Vastly Different Compounds by Isocratic HPLC Aimee N. Heyrman and Yury Zelechonok Resolution Systems, Inc. 590 E. 32nd St. Holland. MI, 49423 SIELC Technologies, 15 E. Palatine Rd, Suite

More information

UPLC for Synthetic Peptides: A User`s Perspective

UPLC for Synthetic Peptides: A User`s Perspective Tides 2007 Conference Las Vegas, May 23, 2007 UPLC for Synthetic Peptides: A User`s Perspective Holger Hermann, Jens Donath* Lonza AG, Switzerland Welcome to Lonza! One of the leading CMO Production facilities

More information

Mercury Stack Monitor SM-4

Mercury Stack Monitor SM-4 Member of the envea Group Mercury Stack Monitor SM-4 Automatic continuous emission monitor (CEM) for mercury Continuous operation Sample dilution directly at stack - works with virtually any sample matrix

More information

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic

More information

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale,

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Mark Tracy, Xiaodong Liu, and Ian Acworth, Thermo Fisher Scientific, Sunnyvale, CA,

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

ONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL

ONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL ONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL APPLICATIONS Accurate, repeatable, and reliable temperature

More information

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0. System Description The VUV PIONA + Analyzer is a VGA-100 Gas Chromatography (GC) detector configured for PIONA compound analysis by the VUV PIONA + method. The VUV PIONA + method uses relatively simple

More information

HiTrap HIC Selection Kit

HiTrap HIC Selection Kit GE Healthcare Data File 18-1143-21 AC Hydrophobic interaction chromatography HiTrap HIC Selection Kit Introduction The HiTrap HIC Selection Kit consists of seven hydrophobic interaction chromatography

More information

DEVELOPMENT OF HPLC METHOD FOR ANALYSIS OF NITRITE AND NITRATE IN VEGETABLE

DEVELOPMENT OF HPLC METHOD FOR ANALYSIS OF NITRITE AND NITRATE IN VEGETABLE Journal of Agricultural, Food and Environmental Sciences UDC 635.546.173/.175]:543.544.5.068.7 Original scientific paper DEVELOPMENT OF HPLC METHOD FOR ANALYSIS OF NITRITE AND NITRATE IN VEGETABLE A. Najdenkoska*

More information

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward Bruker Daltonics EASY-nLC Tailored HPLC for nano-lc-ms Proteomics think forward Nano-HPLC World-Class Performance with a Small Footprint Bruker Daltonics presents a nano-lc system, perfectly integrated

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience. Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience. Pure Consumables More focus on your application The Pure consumable portfolio offers an unrivaled range of products

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

JAPAN QUALITY. Automated Clinical Analyzer

JAPAN QUALITY. Automated Clinical Analyzer JAPAN QUALITY Automated Clinical Analyzer Higher test performance and operability with new functions The clinical analyzer in the next stage 1 The best to the requirements for functionality and usability

More information

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules columns Acclaim Mixed-Mode WCX- for Separating Basic Molecules The Acclaim Mixed-Mode WCX- is a novel, high-efficiency, silica-based column specially designed for separating various basic analytes. This

More information

High Pressure/Performance Liquid Chromatography (HPLC)

High Pressure/Performance Liquid Chromatography (HPLC) High Pressure/Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) is a form of column chromatography that pumps a sample mixture or analyte in a solvent (known as the

More information

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks bjective: The purpose of this experiment is to quantify the caffeine content of a diet soda sample using Ultra-High Performance

More information

Lufft MARWIS-UMB Mobile Advanced Road Weather Information Sensor

Lufft MARWIS-UMB Mobile Advanced Road Weather Information Sensor Lufft MARWIS-UMB Mobile Advanced Road Weather Information Sensor a passion for precision passion pour la précision pasión por la precisión passione per la precisione a www.lufft.com/wondermadeingermany

More information

ActiPix SDI300 Surface Dissolution Imaging System

ActiPix SDI300 Surface Dissolution Imaging System Product Note: PN002B Figure 1. SDI300 Surface Dissolution Imaging System Introduction The ActiPix SDI300 is a powerful UV area imaging system which enables quantitative imaging of surface phenomena for

More information

Varian Galaxie Chromatography Data System for Preparative HPLC

Varian Galaxie Chromatography Data System for Preparative HPLC Varian Galaxie Chromatography Data System for Preparative HPLC By Gary Burce Varian, Inc. 2700 Mitchell Drive, Walnut Creek, CA 95498 USA Abstract Galaxie is an ideal chromatography data system for the

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Reprint. Project Summary. Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado

Reprint. Project Summary. Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado ChemScan Process Analyzer Reprint Project Summary Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado Published by: Applied Spectrometry Associates,

More information

The first address for Titration. TITRONIC 300 and TitroLine 5000

The first address for Titration. TITRONIC 300 and TitroLine 5000 The first address for Titration TITRONIC 300 and TitroLine 5000 Selection table titration piston burettes TITRONIC and automatic titrators TitroLine The most important features of titrators TitroLine and

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Christopher Borton AB SCIEX Golden, Colorado verview Described here is the analysis of

More information

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Agilent 129 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract The Agilent

More information

Fine Particles: Why We Care

Fine Particles: Why We Care Fine Particles: Why We Care Visibility/Radiative Forcing Health Effects A function of chemical composition PM2.5 Mostly 1) Sulfate 2) Carbonaceous - Organic - Elemental (Soot) 3) Metals, minerals, Metals,

More information

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539) Determination of ormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell P Column (EPA 539) Application Note Environmental Authors Rong-jie Fu and Chen-ao (Andy) Zhai Agilent Technologies

More information

N-NITROSODIETHANOLAMINE (NDELA) IN COSMETICS

N-NITROSODIETHANOLAMINE (NDELA) IN COSMETICS As of: February, 2011 N-NITROSODIETHANOLAMINE (NDELA) IN COSMETICS Cosmetics, particularly mascara and skin care products, partly show very high concentration of the carciogenic compound N-Nitrosodiethanolamine

More information

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. 1 Scope This document describes a chromatographic method which allows the identification as well as the determination of

More information

Product Description and Specification

Product Description and Specification Product Description and Specification Catalog No 7 Synthra [ 11 C]Choline Description Synthra [ 11 C]Choline is a completely automated synthesis system for routine production of [ 11 C]choline. Using in-target

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

ChromTech HPLC System

ChromTech HPLC System ChromTech HPLC System High-prssure Gradient System (A) High-pressure pump: 2 pcs UV/VIS detector: 1 piece Gradient control module: 1 unit High-pressure gradient mixer: 1 piece Single channel chromatography

More information

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure. High performance liquid chromatography (HPLC) is a much more sensitive and useful technique than paper and thin layer chromatography. The instrument used for HPLC is called a high performance liquid chromatograph.

More information

Multi-element process analyzer

Multi-element process analyzer Multi-element process analyzer Elemental analysis by X-ray fluorescence Compact multi-element process analyzer for liquid streams or f ixed position web applications Featuring advanced third generation

More information

Process Instrumentation analyzers

Process Instrumentation analyzers PCA300 Family Chlorine, ph, ORP and Temperature Analyzers Backlit LCD display Nema 4X protection.10 DPD chlorine measurement method Colorimeter diagnostics Reagent reminder Amplified ph/temperature probe

More information

Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level

Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level Overview Function s are for measuring the liquid levels in wells, tanks, channels and dams. Sensor Diaphragm p U const. U +2 (brown) I EM I - (blue) 0, U B s are submersible sensors for hydrostatic level

More information

PRODUCT SPECIFICATIONS

PRODUCT SPECIFICATIONS FEATURES AND BENEFITS* Up to 14 year DC life 56V DC working voltage Resistive cell balancing Overvoltage outputs High power density 4U, half-rack package UL registered TYPICAL APPLICATIONS UPS systems

More information

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry O. Scheibner, 1 P. van Baar, 2 F. Wode, 2 U. Dünnbier, 2 K. Akervik, 3 J. Humphries, 3 M. Bromirski 1

More information

AM23 - Texas A&M University Health Science Purchase Order

AM23 - Texas A&M University Health Science Purchase Order Page 1 of 7 AM23 - Texas A&M University Health Science Purchase Order P.O. Date: 9/22/2016 Purchase Order Number AM23-17-P033061 SHOW THIS NUMBER ON ALL PACKAGES, INVOICES AND SHIPPING PAPERS. V E N D

More information

Suppression in anion chromatography. More sensitive analysis of anions and organic acids

Suppression in anion chromatography. More sensitive analysis of anions and organic acids Suppression in anion chromatography More sensitive analysis of anions and organic acids What is suppression? 02 Suppression plays a key role in the analysis of anions and organic acids using ion-exchange

More information

CCS140 and CCS141. Technical Information. Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly

CCS140 and CCS141. Technical Information. Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly Technical Information CCS140 and CCS141 Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly Application Oxidising agents such as chlorine or anorganic

More information

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry Olaf Scheibner, 1 Patrizia van Baar, 2 Florian Wode, 2 Uwe Dünnbier, 2 Kristi Akervik, 3 Jamie Humphrie,

More information

Acclaim Mixed-Mode WCX-1

Acclaim Mixed-Mode WCX-1 Acclaim Mixed-Mode WCX-1 Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28 PRODUCT MANUAL for the Acclaim Mixed-Mode WCX-1 Columns 4.6 x 150 mm, P/N (068353) 4.6 x 250 mm, P/N (068352)

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Determination of Caffeine by HPLC

Determination of Caffeine by HPLC Determination of Caffeine by HPLC Introduction It was a long history before real high performance liquid chromatography (HPLC) had evolved. The very first indication of a chromatographic separation was

More information

Automated Clinical Analyzer

Automated Clinical Analyzer Automated Clinical Analyzer Improved ease of operation and better test efficiency, and provides the optimal solution for routine operations, STAT operation and a broad range of clinical analysis requirements.

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

The Plus for your lab!

The Plus for your lab! BioPhotometer plus NEW! analysis system The Plus for your lab! The easy, rapid and reliable BioPhotometer plus analysis system See it clearly! UVette Precision in plastics The patented* 1 UVette is the

More information

CTC 155 / 350 / 660 Dry-block calibrator

CTC 155 / 350 / 660 Dry-block calibrator CTC 155 / 350 / 660 Dry-block calibrator Wide temperature range: CTC-155: -25 to 155 C CTC-350: 28 to 350 C CTC-660: 28 to 660 C Fast calibration External reference sensor External sensor control Easy

More information

xlogic SuperRelay is a compact and expandable CPU replacing mini PLCs, multiple timers, relays and counters.

xlogic SuperRelay is a compact and expandable CPU replacing mini PLCs, multiple timers, relays and counters. SuperRelay S U P E R R E L A Y, T H E P E R F E C T A L T E R N A T I V E T O L O W C O S T P L C s A N D B A S I C R E L A Y S I N T H I S B R O C H U R E : ELC 18 Standard Models ELC 18 Economy Models

More information

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Xiaodong Liu and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA Overview Review mixed-mode column technology

More information