Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Size: px
Start display at page:

Download "Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases"

Transcription

1 Agilent 129 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract The Agilent 129 Infi nity Quaternary Pump supports fast and convenient method development due to its ability to run ternary and quaternary gradients. This helps to transfer a method containing non-ms compatible mobile phases using nonvolatile buffers to a method using MS compatible mobile phases. The stepwise change from one modifi er to the other permits keeping track of the elution change. This Technical Overview describes how a method, based on phosphate buffer, was changed to a MS compatible method through a stepwise replacement of buffers containing ammonium acetate or formic acid. The developed MS method was transferred to the Agilent 129 Infi nity Binary LC/MS system. 2 1 % phosphate buffer % FA % phosphate buffer % FA

2 Introduction The 129 Infinity Quaternary Pump supports fast and convenient method development due to its ability to run ternary and quaternary gradients. Ternary and quaternary gradients are often used to keep buffer and modifier concentration of a mobile phase mixture constant while using two other channels to run a water/organic gradient. This workflow is extremely helpful for method optimization without preparing several mobile phase sets. The optimum concentration can easily be found, by changing the percentage of the modifier channel from run to run. It is also possible to test different modifiers in one sequence by applying two ternary gradients in series using first, the modifier placed in channel C, and then the modifier placed in channel D. This Technical Overview demonstrates a workflow to transfer a method with mobile phases not compatible with MS into a method with MS-compatible mobile phases. If, for example, a method was developed using NaH 2 PO 4 as buffer, the direct transfer to a MS is not possible. A method compatible with MS needs buffers or modifiers which evaporate in the ion source, such as ammonium acetate or formic acid. Direct injection of the compound mixture onto a LC/MS system using the new modifier implies that the compounds are all known and identification is done through MS spectra. If unknown compounds are present in the compound mixture, the correct correlation between peaks of the LC chromatogram and the MS chromatogram might not be possible, because the elution of the peaks could have changed using the new modifier. Experimental Instrumentation Agilent 129 Infinity Quaternary LC System Agilent 129 Infinity Quaternary Pump Agilent 129 Infinity Autosampler Agilent 129 Infinity Thermostat Agilent 129 Infinity Thermostatted Column Compartment Agilent 129 Infinity Diode Array Detector Agilent 129 Infinity Binary LC/MS System Agilent 129 Infinity Binary Pump Agilent 129 Infinity Autosampler Agilent 129 Infinity Thermostat Agilent 129 Infinity Thermostatted Column Compartment Agilent 129 Infinity Diode Array Detector Agilent 614 Quadrupole Mass Spectrometer Acquisition and Evaluation Software OpenLAB CDS ChemStation version C.1.4 A solution is to stepwise increase the concentration of the MS compatible modifier while, at the same time, the phosphate buffer concentration is decreasing on the 129 Infinity Quaternary UV LC system. By applying a stepwise change, peak tracking is relieved using peak heights or peak areas and UV spectra for identification. Having identified the peak elution difference for the MS compatible modifier, the new method can be passed on to a LC/MS system. Model number G44A G4226A G13B G1316C G4212A Model number G42A G4226A G13B G1316C G4212A G614A 2

3 Chromatographic Conditions Agilent 129 Infinity Quaternary LC Parameter Setting Column Agilent ZORBAX RRHT Eclipse Plus C18, 4.6 mm, 1.8 µm (p/n ) Mobile phases A = water 4% B = methanol 4%, C = mm KH 2 PO 4, ph 7 (adjusted with NaOH) D = mm Ammonium acetate, ph 6.9, or D = 1% formic acid solution, ph 2.1 Gradient 1 4% B to 6% B in utes, at utes 8% B, C=%, D=% Gradient 2 4% B to 6% B in utes, at utes 8% B, C=8%, D=2% Gradient 3 4% B to 6% B in utes, at utes 8% B, C=6%, D=4% Gradient 4 4% B to 6% B in utes, at utes 8% B, C=4%, D=6% Gradient 4% B to 6% B in utes, at utes 8% B, C=2%, D=8% Gradient 6 4% B to 6% B in utes, at utes 8% B, C=%, D=% Flow rate 1.2 ml/ Injection volume µl Column temperature 4 C Detection 24/, Ref 4/8, Hz Transfer from phosphate buffer to ammonium acetate or formic acid solution. Increasing D in 2% steps and decreasing C in 2% steps simultaneously. Chromatographic Conditions Agilent 129 Infinity Binary LC/MS Parameter Setting Column Agilent ZORBAX RRHT Eclipse Plus C18, 2.1 mm, 1.8 µm (p/n ) Mobile phases A = water +.1% formic acid, B = methanol Gradient % B to 6% B in utes, at utes 8% B Flow rate.4 ml/ Injection volume. µl Column temperature C Detection MSD full scan with extracted ion mode Ionization mode API-ES Polarity positive Mass range to 6 Fragmentor 7 Peak width.2 utes 3

4 Analyzed Compounds Tricyclic antidepressants, used as example compounds, were purchased from Sigma Aldrich Chemie GmbH, Germany. The compounds were dissolved in distilled water. H Results and Discussion Due to the structure of these molecules, the separation strongly depends on the ph. The pka values of these compounds are > 9 except for trimipra (pka = 8). Choose a buffer eluent with 2 ph units above or below the pka values of the analytes. The original LC method was based on a phosphate buffer at ph = 7. Ammonium acetate was chosen as MS compatible solvent because the 6.9 ph was close to the ph of the phosphate buffer. Formic acid was used as MS compatible modifier in a second experiment, because it is frequently used for MS analysis. Amitriptyline Imiprae Nortriptyline H 3 C H 3 C H H HCl H Protriptyline Trimiprae, R and S racemate Table 1. Molecular weights and pka values for analyzed compounds. Compound pka value Molecular weight MSD extracted ion Nortriptyline HCl Protriptyline HCl Amitriptyline HCl Trimiprae maleate salt Imiprae HCl

5 Using the 129 Infinity Quaternary Pump, four eluents are deployable. Water was placed in channel A, and methanol was placed in channel B. The eluents containing the modifiers were placed in channels C and D. Channel D contained mm ammonium acetate solution at ph 6.9, or 1% formic acid solution at ph 2.1. The resulting concentration of ammonium acetate in the final experiment was mm, respectively.1% formic acid. To demonstrate the transfer from a non-ms compatible method to a MS compatible method, the following experiments were performed: The 129 Infinity Quaternary Pump was used to analyze tricyclic antidepressant with mm phosphate buffer (channel C) at ph 7. Stepwise addition of ammonium acetate solution (channel D) at ph 6.9. Several methods with decreasing C and increasing D were applied. Six methods had to be set up to go from % C to % D in 2% steps. Stepwise addition of 1% formic acid solution (channel D) at ph 2.1. Several methods with decreasing C and increasing D were applied. Six methods had to be set up to go from % C to % D in 2% steps. Transfer of developed MS method with formic acid to a 129 Infinity Binary LC/MS system Methods were obtained which were compatible with MS. For peak tracking, peak heights and UV-spectra were used Amitriptyline Nortriptyline Protriptyline Trimiprae Imiprae Figure 1. UV-spectra of analyzed compounds, used for identification in addition with peak height.

6 Figures 2 and 3 show an overlay of the six chromatograms with stepwise change of modifiers. Using the ammonium acetate modifier, the retention times shifted to lower values while the elution series remained unchanged. The resolution between the first peaks decreased compared to the chromatogram obtained using the phosphate buffer. Nevertheless, sufficient separation was obtained with ammonium acetate at ph 6.9. The same procedure was repeated with formic acid as the modifier. In this example, the elution series had changed, (Figure 3). All peaks shifted to lower retention times and resolution was significantly reduced. Amitriptylene was now the first peak and nortriptylene the last peak. Protriptylene and trimiprae coeluted and amitriptyline shifted in front of nortriptylene. For identification, the UV spectra were used and the different peak heights. At 6% formic acid, it became obvious that protriptylene and trimiprae will coelute. In a further step, the method was transferred to a column with smaller id and lower flow rates. Typically, MS analysis is done with 2.1-mm id columns and flow rates of about. ml/. 1 % phosphate buffer % NH 4 2% phosphate buffer 8% NH 4 4% phosphate buffer 6% NH 4 1. Protriptylene 2. Trimiprae 3. Nortriptylene 4. Imiprae. Amitriptylene 6% phosphate buffer 4% NH 4 8% phosphate buffer 2% NH 4 % phosphate buffer % NH Figure 2. Separation of antidepressants using ammonium acetate as MS compatible modifier % phosphate buffer % FA 2% phosphate buffer 8% FA 1. Protriptylene 2. Trimiprae 3. Nortriptylene 4. Imiprae. Amitriptylene 6 4% phosphate buffer 6% FA 4 1 6% phosphate buffer 4% FA % phosphate buffer 2% FA % phosphate buffer % FA Figure 3. Separation of antidepressants using formic acid (FA) as MS compatible modifiers. 6

7 The MS method based on formic acid was transferred to a 129 Infinity Binary LC/MS system, (Figure 4). A ZORBAX RRHT Eclipse Plus C18, 2.1 mm, 1.8 µm column was used. The flow was reduced to.4 ml/, the column temperature was raised to C and the injection volume was reduced to. µl. The developed MS method with formic acid provided complete elution of peaks within utes with clear identification TIC Imiprae Trimiprae Amitriptylene Protriptylene Nortriptylene Figure 4. Analys is of tricyclic antidepressants after transfer to the Agilent 129 Infinity Binary LC/MS system using extracted ion chromatograms. Conclusion The transfer of a non-ms compatible method to a method which is applicable to a LC/MS system can be done on an Agilent 129 Infinity Quaternary LC System using a dedicated workflow. The modifier, which is not compatible with MS, is stepwise replaced by a modifier which is usable on a MS. Peak tracking was done using peak height and UV spectra. The stepwise change to the different modifier facilitated the tracking of peaks. At the end, a method was created which was compatible with MS and the elution series was known and could be compared to the primary LC chromatogram. 7

8 This information is subject to change without notice. Agilent Technologies, Inc., 13 Published in the USA, June 1, EN

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 129 Infinity LC System with Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract During

More information

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts

More information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors

More information

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule

More information

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Application Note Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies

More information

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis Automated Switching Between D-LC and Comprehensive D-LC Analysis The Agilent 90 Infinity D-LC Solution Technical Overview Author Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Abstract This

More information

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note LC walk-up system using the Agilent 12 Series LC Method Development Solution and Agilent Easy Access software Test of reaction kinetics, column scouting, and impurity checks with one LC system and up to

More information

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates Using the Agilent 9 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates Technical Overview Author Florian Rieck Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical

More information

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography Technical Overview 2D-LC Conventional 1D-LC Abstract This Technical Overview presents

More information

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection Application Note Small Molecules The Agilent 26 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection Method Development Using Method Scouting Wizard Authors Stefan Bieber Analytical

More information

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent Infinity LC System with Diode Array Detector Application Note Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany [mau]

More information

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation Technical Overview The Agilent InfinityLab D-LC Solution with Active Solvent Modulation Achieving Improved Resolution and Sensitivity for Challenging Combinations of Separation Conditions Author Sonja

More information

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note Developing a fast, generic method for rapid resolution LC with quadrupole MS detection Combining the Agilent 6140 quadrupole MS and multimode source with the Agilent 1200 Series Rapid Resolution LC system

More information

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China Rapid Deteration of Eight Related Aromatic Acids in the p-phthalic Acid Mother Liquid Using an Agilent 126 Infinity LC System and an Agilent Poroshell 12 SB-C18 Column Application Note Author Xinlei Yang

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis

More information

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 2 PFP Columns Application ote Small Molecule Pharmaceuticals Authors Rongjie Fu Agilent Technologies (Shanghai) Co., Ltd Linhai Zhang

More information

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Analysis of Pesticides by Supercritical Fluid Chromatography/ Mass Spectrometry Optimizing the Agilent 126 Infi nity Analytical SFC System in Combination with the Agilent 646 Triple Quadrupole LC/MS Application

More information

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy Analytical Quantification of Deoxygenated Compounds in Catalytic Reaction Samples as an Evaluation Technique to Detere Reaction Conversion from a Bioderived Oil in Novel Biofuel Testing Application Note

More information

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 129 Infinity LC system Application Note Food and Beverages Author Srividya Kailasam

More information

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system Application Note Food Testing & Agriculture Authors Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Keiko

More information

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF Application Note Donghui Yi and Christine Miller Jon D. Williams, GlaxoSmithKline Introduction Mass spectrometry (MS) has become a core technology

More information

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine Comprehensive 2D-LC Analysis of Chinese Herbal Medicine The Agilent 1290 Infinity 2D-LC Solution Application Note Small Molecule Pharmaceuticals & Generics Author Sonja Krieger Agilent Technologies, Inc.

More information

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent

More information

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Food Author Michael Woodman Agilent Technologies, Inc. 285 Centerville Road Wilgton, DE 1988-161 USA Abstract Soft drinks,

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine pplication Note Small Molecule Pharmaceuticals, Generics ccurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine igh-resolution sampling D-LC uthors Rongjie Fu and Xinlei

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India QbD Based Method Development on an Agilent 129 Infinity UHPLC System Combined with a Seamless Method Transfer to HPLC Using Intelligent System Emulation Technology Application Note Pharmaceutical QA/QC

More information

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Application Note Forensic and Toxicology - Criminalistics Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Using the Agilent 126 Infinity II SFC and Detection by Coupled

More information

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 129 Infi nity LC, and 646 Triple Quadrupole LC/MS Application Note Clinical Research Authors Linda Côté and Christophe

More information

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract. Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28

More information

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Marco Karsten, Bas, Dolman, Giovanni Maio, Frank Steiner, Holger Franz, Frank Arnold and Remco Swart LC Packings,

More information

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Application. Authors. Abstract. Introduction. Hydrocarbon Processing Fast Analysis Method for Rubber Chemical Antidegradants Using Rapid Resolution Liquid Chromatography (RRLC) Systems with Rapid Resolution High Throughput (RRHT) Columns Application Hydrocarbon Processing

More information

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-exyl Columns to Separate Estrogens Application Note Pharmaceutical, Environmental Authors John W. enderson Jr. and William J. Long Agilent

More information

Application Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA

Application Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA Fast Method Development of Salicylic Acid Process Impurities using Agilent ZRBAX Rapid Resolution High Throughput Columns with the Agilent 12 Series Method Development Solution Controlled by AutoChrom

More information

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs Value of Eclipse Plus C18 and ph Selectivity to Analyze Ae-Containing Drugs Application Pharmaceuticals Authors John W. Henderson Jr., William J. Long, and Cliff Woodward Agilent Technologies 285 Centerville

More information

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Application Note Inorganic Ions, Water Testing, Minerals, Metals, Basic Chemicals Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Authors Anne Mack, Adam Bivens

More information

Application Note. Edgar Naegele. Abstract

Application Note. Edgar Naegele. Abstract Fast identification of main drug metabolites by quadrupole time-of-flight LC/MS Measuring accurate MS and MS/MS data with the Agilent 651 Q-TOF LC/MS and identification of main meta-bolites by comparison

More information

Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TOF LC/MS Combined with Advanced MassHunter Data Processing Software

Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TOF LC/MS Combined with Advanced MassHunter Data Processing Software Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TF LC/MS Combined with Advanced Massunter Data Processing Software Application Note Authors Siji Joseph Agilent Technologies LSCI Bangalore,

More information

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food Agilent Application Solution Analysis of acaricides in honey Application Note Food Authors mau 12 1 A.G. Huesgen Agilent Technologies, Inc. Waldbronn, Germany 8 6 4 2 2 4 6 8 1 12 14 16 18 Abstract Varroosis

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

Agilent LC & LC-MS Cannabis Testing

Agilent LC & LC-MS Cannabis Testing Agilent LC & LC-MS Cannabis Testing Potency Analysis for Cannabinoids Mike Adams 1, A. Roth 1, Joan Stevens 2, Karen Kaikaris 1, & Sue D Antonio 2 1 CWC LABS, 2 Agilent Technologies, UV detection with

More information

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 2 Columns Application Note Pharma, BioPharma, and Clinical Research Author William Long Agilent Technologies, Inc. Introduction Steroids

More information

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc. Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 126 Infinity Hybrid SFC/UHPLC System Improving peak shape and providing wider UV selectivity

More information

ACD/AutoChrom Assisted Method Development for Challenging Separations. Vera Leshchinskaya February 7, 2018

ACD/AutoChrom Assisted Method Development for Challenging Separations. Vera Leshchinskaya February 7, 2018 ACD/AutoChrom Assisted Method Development for Challenging Separations Vera Leshchinskaya February 7, 2018 Introduction Resolution of isomeric species is one of the most challenging areas of separation

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times The Secrets of Rapid HPLC Method Development Choosing Columns for Rapid Method Development and Short Analysis Times Rapid Analysis Is More Than Run Time It is developing a method to meet a goal and developing

More information

The new Agilent 1290 Infinity Quaternary LC

The new Agilent 1290 Infinity Quaternary LC The new Agilent 1290 Infinity Quaternary LC How binary pump-like performance can enhance your lab efficiency Sebastian Krahe Produktspezialist HPLC Agilent Technologies Page 1 The 1290 Infinity Quaternary

More information

Agilent All Ions MS/MS

Agilent All Ions MS/MS Agilent All Ions MS/MS Workflow Overview A Determine fragment ions for LC/MS Quant method B Develop final Quant method Develop LC/MS Qualitative Analysis method Process data with Find by Formula Build

More information

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Technical Note Agilent s SampliQ WAX provides Applications for strongly acidic, acidic and neutral compounds Excellent

More information

770-9P. Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase

770-9P. Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase 77-9P Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase J.T. Lee 1 Maria Esther Rodriguez Rosas 2 and Thomas E.

More information

Modernizing the USP Monograph for Acetaminophen

Modernizing the USP Monograph for Acetaminophen Modernizing the USP Monograph for cetaophen pplication Note Small Molecule Pharmaceuticals and Generics uthor Rongjie Fu gilent Technologies (Shanghai) Co. Ltd bstract new HPLC method was developed for

More information

Analysis of E.coli Tryptic Digest and Intact Protein Using the Agilent 1290 Infinity 2D-LC Solution with Diode Array Detection and Q-TOF LC/MS

Analysis of E.coli Tryptic Digest and Intact Protein Using the Agilent 1290 Infinity 2D-LC Solution with Diode Array Detection and Q-TOF LC/MS Analysis of E.coli Tryptic Digest and Intact Protein Using the Agilent Infinity D-LC Solution with Diode Array Detection and Q-TOF LC/MS Application Note Proteomics & Protein Sciences Authors Gerd Vanhoenacker,

More information

Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen

Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters Application Angelika Gratzfeld-Huesgen Abstract This application has been verified using an Agilent

More information

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution Detection of Trace Level Pharmaceuticals in Drinking Water by nline SPE Enrichment with the Agilent Infinity Series nline-spe Solution Application Note Environmental Author Edgar Naegele Agilent Technologies,

More information

Application Note. Authors. Abstract. Food Testing & Agriculture. Stephen E. Reichenbach University of Nebraska Lincoln, NE, USA

Application Note. Authors. Abstract. Food Testing & Agriculture. Stephen E. Reichenbach University of Nebraska Lincoln, NE, USA 2-Dimensional separation of polyphenols in beverages using the Agilent 129 Infinity 2D-LC Solution and software-assisted 2D-LC data analysis for comparison of ingredients Application Note Food Testing

More information

The Analysis of Organophosphate Pesticides by LC/MS Application

The Analysis of Organophosphate Pesticides by LC/MS Application The Analysis of Organophosphate Pesticides by LC/MS Application LC-MS Authors Paul Zavitsanos Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Paul Yang MOE Ontario Canada

More information

mau 200 ph min

mau 200 ph min mau ph - 6 8 TAKE A LOOK INSIDE TWIN Technology TWIN (Two-In-One ) Technology is what gives Gei its superior performance edge. During the final stage of silica manufacturing, a unique silica-organic layer

More information

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS Application Note Authors E. Michael Thurman and Imma Ferrer Center for Environmental

More information

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract Detection and identification of impurities in pharmaceutical drugs Computer-assisted extraction, profiling and analysis of Q-TF data for determination of impurities using Agilent Massunter software Application

More information

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction Application Note Consumer Products Authors Devon C. Zimmerman, Henry F. Rossi III, Jacqueline Rizzo, Daniel W. Keating,

More information

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract. Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS Application Food Analysis Authors Sheher Mohsin Agilent Technologies, Inc. Schaumburg, Il USA Michael Zumwalt

More information

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide What is the MassHunter METLIN Metabolite PCD/PCDL? 2 Where to find more information 2 Kit Content 3 Installation 5 Before you start 5 Install MassHunter

More information

Application Note. Edgar Naegele. Abstract

Application Note. Edgar Naegele. Abstract Agilent MassHunter Fast computer aided analysis of LC/ESI-TF data from complex natural product extracts Part 1: Analysis of Agilent 6210 TF data with the Molecular Feature Extractor in MassHunter Workstation

More information

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes Application Note Food Safety Author Sunil Kulkarni Agilent Technologies

More information

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster Welcome to our E-Sear: Choosing HPLC Columns for Faster Analysis Smaller and Faster High Throughput/Fast LC Requires. Short columns 0 mm or shorter Small particle sizes. µm Rapid Resolution or new.8 µm

More information

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software Application ote Drug discovery and development: Metabolite Identifi cation Authors Yuqin Dai,

More information

Analysis of Positional Isomers with Agilent Poroshell 120 PFP Columns

Analysis of Positional Isomers with Agilent Poroshell 120 PFP Columns Analysis of Positional Isomers with Agilent Poroshell PFP olumns Application ote Environmental Authors William Long and Jonathan Horton Agilent Technologies, Inc. Introduction The Agilent Poroshell PFP

More information

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539) Determination of ormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell P Column (EPA 539) Application Note Environmental Authors Rong-jie Fu and Chen-ao (Andy) Zhai Agilent Technologies

More information

penta-hilic UHPLC COLUMNS

penta-hilic UHPLC COLUMNS penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and

More information

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution

More information

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software Application Sergey Galushko Ralf Honsberg Abstract The development of an HPLC method is

More information

Polymer analysis by GPC-SEC. Technical Note. Introduction

Polymer analysis by GPC-SEC. Technical Note. Introduction Polymer analysis by GPC-SEC Technical Note Introduction Gel Permeation Chromatography (GPC), also referred to as Size Exclusion Chromatography (SEC) is a mode of liquid chromatography in which the components

More information

Computer-assisted analysis of complex natural product extracts

Computer-assisted analysis of complex natural product extracts Computer-assisted analysis of complex natural product extracts Detection of known and identification of unknown compounds from Q-TF mass spectrometry with the Agilent MassHunter Metabolite ID software

More information

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application. Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source Application Toxicology Authors Jürgen Wendt Agilent Technologies Sales and Support

More information

Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria

Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria Application Note Authors Yuqin Dai and Steven M. Fischer Agilent Technologies, Inc. Santa Clara, CA, USA Abstract

More information

Shodex TM ODP2 HP series columns

Shodex TM ODP2 HP series columns HPLC Columns Shodex TM ODP2 HP series columns Better retention of highly polar substances Technical notebook No. 6 Contents 1. Introduction 1-1. Specifications 1-2. Eluent Compatibility of ODP2 HP Series

More information

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120 LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120 Application Note Food Testing & Agriculture Author Irina Dioumaeva Agilent Technologies, Inc. Abstract

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

Acclaim Mixed-Mode WCX-1

Acclaim Mixed-Mode WCX-1 Acclaim Mixed-Mode WCX-1 Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28 PRODUCT MANUAL for the Acclaim Mixed-Mode WCX-1 Columns 4.6 x 150 mm, P/N (068353) 4.6 x 250 mm, P/N (068352)

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Accurate Mass Analysis of Hydraulic Fracturing Waters

Accurate Mass Analysis of Hydraulic Fracturing Waters Application Note Environmental Accurate Mass Analysis of Hydraulic Fracturing Waters Using the Kendrick mass defect with the Agilent LC/Q-TOF MS Authors E. Michael Thurman and Imma Ferrer Department of

More information

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors. Sensitive Femtogram Determination of Aflatoxins B 1, B 2, G 1 and G 2 in Food Matrices using Triple Quadrupole LC/MS Application Note Food Safety Authors Yang Chen and Jack Cappozzo National Center for

More information

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application Pharmaceuticals Author Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808

More information

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS Application Note Forensic Toxicology Authors Martin Josefsson, and Markus Roman National Board of Forensic Medicine Linköping,

More information

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection. Agilent Application Solution Analysis of PAHs in soil according to EPA 3 method with UV and fluorescence detection Application Note Environmental Authors Sonja Volk, Angelika Gratzfeld-Huesgen Agilent

More information

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column Application Note 95 Determination of Hydrochloride Purity Using the Acclaim PA Column INTRODUCTION -based medications are prescribed for several heart and blood pressure indications. The fastacting formulations

More information

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry Application Note Food safety Authors Ismael Flores and Carlos Sepulveda Agrolab México Km 7 Carretera Pachuca-Actopan

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Choosing Columns and Conditions for the Best Peak Shape

Choosing Columns and Conditions for the Best Peak Shape Choosing Columns and Conditions for the Best Peak Shape What is Good Peak Shape and Why is it Important? Good peak shape can be defined as a symmetrical or gaussian peak and poor peak shape can include

More information

Method Development in Comprehensive 2D-LC Finding the Most Orthogonal Separation Systems for RPLC RPLC Using Column and Solvent Screening

Method Development in Comprehensive 2D-LC Finding the Most Orthogonal Separation Systems for RPLC RPLC Using Column and Solvent Screening Method Development in Comprehensive 2D-LC Finding the Most Orthogonal Separation Systems for RPLC RPLC Using Column and Solvent Screening Technical Overview Author Sonja Krieger Agilent Technologies, Inc.

More information

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 643 LC/MS/MS Application Note Forensic Toxicology Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca Wagner,

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining Dave Weil, Ph.D. and Jim Lau, Ph.D. Typical Method Conditions: 1260 UHPLC

More information

The Agilent 1290 Infinity II Multi Method Solution

The Agilent 1290 Infinity II Multi Method Solution The Agilent 129 Infinity II Multi Method Solution Switch between seven applications on a single LC system by using automatic column and solvent selection Application ote Food Testing & Agriculture Author

More information

Fast Screening Methods for Analgesics and Non-Steroidal Anti-Inflammatory Drugs by HPLC

Fast Screening Methods for Analgesics and Non-Steroidal Anti-Inflammatory Drugs by HPLC Fast Screening Methods for Analgesics and on-steroidal Anti-Inflammatory Drugs by HPLC Application ote Small Molecule Pharmaceuticals and Generics Author William Long Agilent Technologies, Inc. Introduction

More information

for Acclaim Mixed-Mode WCX-1

for Acclaim Mixed-Mode WCX-1 for Acclaim Mixed-Mode WCX- Product Manual for the Acclaim Mixed-Mode WCX- Column Page of 8 Product Manual for Acclaim Mixed-Mode WCX- Columns Acclaim Mixed-Mode WCX-, 3µm, Analytical column, 3.0 x 50mm

More information