FTIR CEM Performance Specification Modification Considerations

Size: px
Start display at page:

Download "FTIR CEM Performance Specification Modification Considerations"

Transcription

1 FTIR CEM Performance Specification Modification Considerations EPA/ICAC Emissions Measurements Roundtable Meeting September 17, 2013 Peter G. Zemek MKS Instruments Marty Spartz Prism Analytical MultiGas FT-IR Technologies Automated HCl CEM Systems MKS Instruments On-Line Product Group 2 Tech Drive, Suite 201 Andover, MA Tel:

2 Performance Specifications for FTIR CEMS PERFORMANCE SPECIFICATIONS: PERFORMANCE SPECIFICATION 15 - FOR EXTRACTIVE FTIR CONTINUOUS EMISSIONS MONITOR SYSTEMS IN STATIONARY SOURCES RM RATA RUN COMPARISONS FOLLOW PERFORMANCE SPECIFICATION 2 SPECIFICATIONS AND TEST PROCEDURES FOR SO2 AND NOX CONTINUOUS EMISSION MONITORING SYSTEMS IN STATIONARY SOURCES DRAFT PERFORMANCE SPECIFICATION 18 AND TEST PROCEDURES FOR HCL CONTINUOUS EMISSION MONITORING SYSTEMS IN STATIONARY SOURCES

3 FTIR Reference Methods for CEMS REFERENCE METHODS: TO VALIDATE UPON INSTALLATION OR RATA METHOD 301 FIELD VALIDATION OF POLLUTANT MEASUREMENT METHODS FROM VARIOUS WASTE MEDIA METHOD 320* - MEASUREMENT OF VAPOR PHASE ORGANIC AND INORGANIC EMISSIONS BY EXTRACTIVE FOURIER TRANSFORM INFRARED (FTIR) SPECTROSCOPY (INCLUDES FTIR PROTOCOL) METHOD 321** - MEASUREMENT OF GASEOUS HYDROGEN CHLORIDE EMISSIONS AT PORTLAND CEMENT KILNS BY FOURIER TRANSFORM INFRARED (FTIR) SPECTROSCOPY METHOD EXTRACTIVE FTIR METHOD FOR THE MEASUREMENT OF EMISSIONS FROM THE MINERAL WOOL AND WOOL FIBERGLASS INDUSTRIES ASTM D STANDARD TEST METHOD FOR DETERMINATION OF GASEOUS COMPOUNDS BY EXTRACTIVE DIRECT INTERFACE FOURIER TRANSFORM INFRARED (FTIR) SPECTROSCOPY. *1 of 2 RM for EGU **Only acceptable RM for Portland Cement MACT

4 Confusion and Errors Between Methods/Specifications? TEST METHOD MEASUREMENT OF VAPOR PHASE ORGANIC AND INORGANIC EMISSIONS BY EXTRACTIVE FOURIER TRANSFORM INFRARED (FTIR) SPECTROSCOPY - (This is already a self-validating Compliance (Reference) Method with its own QA/QC) PS-15 states- For System Validation, Since FTIR CEMS have multicomponent capability, it is possible to perform more than one RM simultaneously, one for each target analyte. Interpretation, If use EPA M320, 321, 301, etc, ASTM D6348 you can RATA w/ftir

5 Comparison PS/RM for FTIR Parameter PS-15 PS-18 Draft EPA M320 ASTM D ,12 M301 Dates Promulgated Feb 2000 Feb 2014? Feb 2000, Protocol 1995 Feb 2003, '10,'12 June 1991, '04, '11 Cal Gas Direct Accuracy 5%, 7% w/bias NA 2% or 5% 5% NA Dynamic Spike Recovery +-30% +-15% +-30% +-30% +-30% to 50% # of Spike Runs 24 (M301), 3 (M320) 6? Interference test No (just mentions them) <3% span combined No bias check No Method Relative Accuracy (DQO) 20% (RM in denom) 15% (Emiss Limit in denom) SO2 (10% Emiss Limit) PS2 < 15% 2.5% 2% Acc 10% Prec 5% t-test comparison "Bias" Correction Yes? ">30% is not valid method", but apply a corr factor May Yes, above 10% Calibration Error Test 10% 3 pt, <5%, intercept <15% of Span No One Knows No depends on RM Dry Cal check YES (All Compounds) Daily? Yes, but select set..no HCl Yes, but select set..no HCl depends on RM 7-Day Drift No 3% of span/day No No No Detector linearity 3 point, 2%/5% yes? 3 point, 2%/5% 3 point, 2%/5% NA Detection Limitations depends on RM LOS, MDC#3, LOD MAU, OFU MDC#1, MDC#2*, MDC#3 Cell Volumes per Data Pt. 10 spiking, 5 sample 5? 5 5 NA PLQ*, LOD Conf Intervals depends on RM 99%? 95% 95% 99% LOD, 95% t-test * Reviewed/Revised Manual Data Verification NA (Yes RATA RM) NA (Yes RATA RM) YES YES NO Calibration Transfer Standard (CTS) 5%? 5% 5% NA *MDC #2 = PLQ (no longer used)

6 FTIR is a Special Case Most of the Methods/PS were written with traditional CEMs in mind Traditional CEMS/GC are destruction techniques and as such change daily due many factors, i.e. ionization plates, fuel, gas conditioning FTIR is an optical technique where sample is not changed in any way Therefore, it DOES not change unless it is not operating properly If one compound does not perform 1 day, all of the compounds will be biased, but if one compound does perform, they will all perform EPA has stated, If EPA library reference spectra are not available, use calibration standards to prepare reference spectra according to Section 6 of the FTIR Protocol EPA is stating that, if the spectra are good, the numbers will be good Important point! Initial certification pass = very little ongoing QA/QC needed

7 FTIR Do not Drift or Change Response MKS Instruments 7

8 FTIR has come far (err mid) FTIR Requirements for CEM has been burdened to the point that much of the daily QA/QC and initial certification is unwarranted and will prevent its use in industry as newer technology QA/QC becomes more efficient (FTIR Rules Outdated)-Discriminatory 40 CFR Parts 72 and 75 [FRL ] RIN 2060 AG46 Acid Rain Program; Continuous Emission Monitoring Rule Revisions 75.33(d) would make it a violation of the primary measurement requirement 75.10(a) to allow the annual monitor data availability to drop below 80.0 percent for SO2, NOX, or CO2. (Docket A 97 35, Item II B 16). Consistent with current practices, the Agency would continue to expect CEMS to achieve high data availability.., monitor downtime in excess of 5.0 percent may warrant appropriate investigation and follow-up activities. Uptime = Credits = $

9 Current PS-15 Requirements not necessary today PS-15 Required Reason Whats Changed Detector Linearity Manufacturer performed New Boards and Algorithms and Performance Based Cell leak test X-axis accuracy Resolution Check CTS will determine if leaks present Redundant Will fail relative accuracy Only 1 manufacturer has a problem and dynamic spike test Will fail relative accuracy Redundant and dynamic spike test Audit Spike Dynamic spike with Audit is Good if conc correct 100%T Line 5% Req Current algorithms correct for this Used in place of Dynamic Spike Faster computers/more stabile Temperatures Audit Spectra Silly Not possible unless all spectral functions are provided Submit Spectra for Analysis to EPA Really? Signal to Noise Calculation Meaningless Need multiple soft wares and training. No one else is required to do this Will be determined by LOD/LOS Daily Dry Zero, Mid, Span or ALL TARGET COMPOUNDS Does not represent actual target compound response (Need H2O), Contaminates System Need to measure HCl, NH3, CH2O, HF, HCN and other polar species at very low levels If dry cal gas spans/etc are required, then certain cmpds will take hours

10 Cement Plant CEM Data with no Filter Change/Back-Purge for 3 Months- End of Period Test A steady-state is reached until the moisture changes Once dry, HCl is liberated in very high concentrations

11 Recommended On-Going PS-15 On-Going QA/QC Performance Specifications (After Initial Installation Certification): QA/QC Checks that should be required on a daily basis or other basis are as follows: Background Times are at the Discretion of Manufacturer Daily CEM System Zero (2% of 1 st Span) Daily - CTS indirect gas check, preferably using Ethylene or dilution of CO2/H2O (System Check) Daily - Dynamic spike check based on a minimum of at least 1 target compound. (Preferably the component that is expected to be the hardest to recover during the dynamic spiking test due to its chemical or reactive nature, adherence to the wetted components, or has the highest spectral interferences) (20% Recovery at 10% 50% expected or 5x MDL * Special Cases Options should exist for HCl/NH3 streams) Semi-Annual - RATA test with an acceptable reference method (Beware traditional CEMs such as Chemiluminesense or CO/CO2

12 Calibration Transfer Standard (CTS) The FTIR CEM, once validated is accurate and precise if functioning properly Indirect CTS gas checks the instrumentation and sampling system functionality Used to determine: Leaks within the sampling system T95 Rise and Fall times Instrument performance results for Resolution Signal to noise ratio (SNR) Linearity Frequency accuracy Sample cell path-length Instrument Line Shape (ILS) Noise Equivalent Area (NEA) Detection Limit (DL) Spectral limits Ensures that the correct Analytical Method is loaded

13 CTS Ethylene Wet or Dry

14 Detection Limits - Which One? Needs Modification and Consistent Terminology 1 set of data MKS Instruments 14

15 PS-15/PS-18 Detection Limitations Modifications Updating (~LOS) M301 The PLQ is a limit determined by the standard deviation of an estimate of a concentration; if the standard deviation of the estimate exceeds a threshold, then that estimate is unacceptable. The LOD is a limit determined by the estimate of the concentration itself (therefore PLQ = MDC#2) The LOD is clearly a threshold. An estimate that cannot be distinguished from one resulting from a blank sample is unlikely to provide meaningful results. 1. Prepare 7 samples standard LOD1 2. Std Dev = S1 3. LOD0 = 3S1 4. LOD1 2LOD0, then S0 = S1 5. LOD1 > 2LOD0, then: 6. 2 additional standards <LOD samples std dev each = S2, S3 8. Plot S1, S2, S3 as f(conc) 9. Draw Regression Line 10.Extrapolate to zero HCl Std Dev as f(conc) y = x LOD = 0.02 ppm Std as f(c) Linear (Std as f(c))

16 Span? Must be Defined for FTIR Each Cal Point is a Span Multi-Span in 1 Method Must Define Span PS15

17 % Recovery For Your Consideration? Typical dynamic spike involves diluting the CEM sample flow by 10% to get 50% addition 5 ppm native, use 25 ppm HCl cylinder diluted 10x = ~2.5 ppm added to 5 ppm native Theoretical CEM response = ~ 7.5 ppm HCl High NH3 slip Recovery vs. Spike Level (5 ppm native HCl) (10 percent dilution) ppm added to 5 ppm native HCl Greater than 50% Rule (162%) Matrix Effect - is it wrong? Spike Level (ppm) ASTM allows up to150% of Native Conc. Fail because not 10% DF? Why? Still reports 5 ppm at stack correct 50% of Native Rule or 10% Flow Rule Wrong?

18 Recommendations Rewrite/Modify/Update PS15 Allow substitution of ASTM D for M320 or Correct/Modify/Update M320 Standardize on Terminology Establish a sub-committee of outside interested parties/stakeholders made up of experienced field technicians/scientists/engineers in FTIR to provide recommendations to EPA on changes to PS15 and M320 Make it understandable to the layman Try to keep the Politics Out of the Science MKS Instruments 18

Pilot-Plant Performance Testing to Support Development of a Dilution Extractive HCl Monitoring System

Pilot-Plant Performance Testing to Support Development of a Dilution Extractive HCl Monitoring System Pilot-Plant Performance Testing to Support Development of a Dilution Extractive HCl Monitoring System Ty Smith Cemtek Group Inc. and Ken Greaves Cemtek Instruments Inc. Background Cemtek is researching

More information

VALIDATION STUDY OF FTIR-BASED EMISSIONS MEASUREMENTS AT A MUNICIPAL WASTE COMBUSTOR

VALIDATION STUDY OF FTIR-BASED EMISSIONS MEASUREMENTS AT A MUNICIPAL WASTE COMBUSTOR AP-157 VALIDATION STUDY OF FTIR-BASED EMISSIONS MEASUREMENTS AT A MUNICIPAL WASTE COMBUSTOR Grant M. Plummer, Ph.D. Peter Zemek, Ph.D. Special Projects Manager Applications Manager Enthalpy Analytical,

More information

VOTING DRAFT STANDARD

VOTING DRAFT STANDARD page 1 of 7 VOTING DRAFT STANDARD VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 Description This Voting Draft Standard is a proposed revision of the 2009 standard (EL-

More information

Validation Study of FTIR-Based Emissions Measurements at a Municipal Waste Combustor

Validation Study of FTIR-Based Emissions Measurements at a Municipal Waste Combustor 13th North American Waste to Energy Conference May 23-25, 2005, Orlando, Florida USA NAWTEC13-3158 Validation Study of FTIR-Based Emissions Measurements at a Municipal Waste Combustor G. J. Aldina Jeffrey

More information

Gaseous CEMS technology

Gaseous CEMS technology GAS CEMS EXTRACTIVE IN-SITU COLD DRY HOT WET DRY DILUTION PROBE/ POINT IN-STACK CROSS STACK OUT OF STACK Gaseous CEMS technology Probe (at stack) Heated filter Heated sample line Condenser Pump Analyzers

More information

Developing an Improved Impinger-Based Method for Measuring Gaseous Hydrogen Chloride in Cement and Lime Kiln Emissions

Developing an Improved Impinger-Based Method for Measuring Gaseous Hydrogen Chloride in Cement and Lime Kiln Emissions Developing an Improved Impinger-Based Method for Measuring Gaseous Hydrogen Chloride in Cement and Lime Kiln Emissions Laura L. Kinner and James W. Peeler Emission Monitoring Inc. Raleigh, NC 27612 Daniel

More information

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4 page1 of 8 TNI Standard VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 January 2016 Description This TNI Standard has been taken through all of the voting stages and has

More information

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Ask The Expert Webinar Series EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Richard Burrows, Ph.D. Corporate Technical Director A Revision to the Method Detection Limit EPA

More information

White Paper. Overview: NDIR Definition:

White Paper. Overview: NDIR Definition: Title: NDIR Technology Overview, Compliance, and Comparison to Other Generally Available Gas Measurement Technologies TSN Number: 06 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\06

More information

How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert

How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert The LOD Procedure has changed Federal Register /Vol. 82, No. 165 /Monday, August 28, 2017 ACTION: Final rule. DATES: This regulation

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply EPAs New MDL Procedure What it Means, Why it Works, and How to Comply Richard Burrows TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure

More information

Quality Assurance of Automated Measuring Systems: practical implementation of pren 14181:2001. Results of field experience at two Italian plants.

Quality Assurance of Automated Measuring Systems: practical implementation of pren 14181:2001. Results of field experience at two Italian plants. Quality Assurance of Automated Measuring Systems: practical implementation of pren 14181:2001. Results of field experience at two Italian plants. Cipriano D. (1), Ferrari A. (2), Negri A. (1) (1) CESI

More information

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. 1 REVISOR 4740.2100 4740.2100 QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. Subpart 1. Scope. This part applies to laboratories performing testing under the inorganic chemistry, metals,

More information

: PM CEM Optical Technology and PS-11

: PM CEM Optical Technology and PS-11 : PM CEM Optical Technology and PS-11 Outline : Optical PM CEM Technology Overview : Opacity : Scatter Light : Review of PS-11 Initial Correlation Audit (ICA) : Requirements : Field Experience : QA/QC

More information

Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System

Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System Presented at the BevTech Conference, Albuquerque, NM 2018 by Charles M. Phillips Ph.D., Max Analytical Technologies Mark Taylor, Vice

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18 Draft PS 18 APPENDIX A Dynamic Spiking Procedure (4-4-2013) A1. Scope and Application 1.1 This appendix to Performance Specification 18 describes the procedure and performance requirements for dynamic

More information

Mercury Stack Monitor SM-4

Mercury Stack Monitor SM-4 Member of the envea Group Mercury Stack Monitor SM-4 Automatic continuous emission monitor (CEM) for mercury Continuous operation Sample dilution directly at stack - works with virtually any sample matrix

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

Appendix A - Test Methods Method 301--Field Validation of Pollutant Measurement Methods from Various Waste Media

Appendix A - Test Methods Method 301--Field Validation of Pollutant Measurement Methods from Various Waste Media Appendix A - Test Methods Method 301--Field Validation of Pollutant Measurement Methods from Various Waste Media 1. APPLICABILITY AND PRINCIPLE 1.1 Applicability. This method, as specified in the applicable

More information

Cyanide, Total or Weak Acid Dissociable, by Manual Distillation - PBM

Cyanide, Total or Weak Acid Dissociable, by Manual Distillation - PBM Inorganics Revision Date: March 8, 2017 (Draft) Cyanide, Total or Weak Acid Dissociable, by Manual Distillation - PBM Parameter Analytical Method Introduction Method Summary MDL(s) and EMS Analyte Codes

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

Laboratory 101: A Guide to Understanding your Testing Laboratory

Laboratory 101: A Guide to Understanding your Testing Laboratory Laboratory 101: A Guide to Understanding your Testing Laboratory Jim Cronin Product Ecology Manager Environmental Monitoring and Technologies, Inc. (EMT) Morton Grove, Illinois Abstract: Laboratory 101:

More information

Chlorine and Hydrogen Chloride Monitoring Utilizing Ion Mobility Spectroscopy (IMS)

Chlorine and Hydrogen Chloride Monitoring Utilizing Ion Mobility Spectroscopy (IMS) Chlorine and Hydrogen Chloride Monitoring Utilizing Ion Mobility Spectroscopy (IMS) Tad Bacon Jim Taylor Kurt Webber VP Engineering Sr. Chemical Engineer VP Sales Molecular Analytics Molecular Analytics

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

ABB Measurement & Analytics Applying Cavity Enhanced Absorption Spectroscopy (CEAS) to trace HCl measurement of coal-fired power plant emissions

ABB Measurement & Analytics Applying Cavity Enhanced Absorption Spectroscopy (CEAS) to trace HCl measurement of coal-fired power plant emissions White paper WP/ANALYTICAL/1 EN ABB Measurement & Analytics Applying Cavity Enhanced Absorption Spectroscopy (CEAS) to trace HCl measurement of coal-fired power plant emissions Cavity-enhanced laser absorption

More information

Robustness of the QAL2 calibration (EN 14181) Uncertainty on the results given by a calibrated AMS

Robustness of the QAL2 calibration (EN 14181) Uncertainty on the results given by a calibrated AMS Robustness of the 2 calibration (EN 14181) Uncertainty on the results given by a calibrated Jean.Poulleau*¹, Cécile.Raventos 1, Philippe.Fayolle², Emmanuel.Fiani 3 ¹ INERIS, Parc technologique ALATA BP

More information

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE) METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE) 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to ground water, drinking, surface, and saline waters, and domestic and industrial wastes.

More information

IMPROVE DETECTION AND QUANTITATION. Richard Burrows

IMPROVE DETECTION AND QUANTITATION. Richard Burrows WORKING WITH THE EPA TO IMPROVE DETECTION AND QUANTITATION Richard Burrows FEDERAL ADVISORY COMMITTEE ON DETECTION AND QUANTITATION ML First, a little history Lead in Albacore: Guide to Lead Pollution

More information

SECTION D.2 AMMONIA NITROGEN

SECTION D.2 AMMONIA NITROGEN SECTION D.2 AMMONIA NITROGEN CEDR Method Code: NH4F L01 a) Scope and Application i) This method describes the determination of low-level ammonia nitrogen concentrations in filtered samples taken from fresh

More information

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) SOP-316 391 S. Vidal St., Sarnia, Ontario, N7T 7L1 Phone: (519) 344-4747 Fax: (519) 344-2350 E-Mail: info@lambtonscientific.com Method

More information

Really, A Revised MDL Procedure

Really, A Revised MDL Procedure NEMC 2016 Really, A Revised MDL Procedure Richard Burrows, Ph.D. TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure in the Federal

More information

METHOD 8000A GAS CHROMATOGRAPHY

METHOD 8000A GAS CHROMATOGRAPHY METHOD 8000A GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Gas chromatography is a quantitative technique useful for the analysis of organic compounds capable of being volatilized without being decomposed

More information

Agilent Oil Analyzer: customizing analysis methods

Agilent Oil Analyzer: customizing analysis methods Agilent Oil Analyzer: customizing analysis methods Application Note Author Alexander Rzhevskii & Mustafa Kansiz Agilent Technologies, Inc. Introduction Traditionally, the analysis of used oils has been

More information

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Proposed Procedures for Determining the Method Detection Limit and Minimum Level Proposed Procedures for Determining the Method Detection Limit and Minimum Level Published by: ACIL Environmental Services Section Technical Committee Revision 3.0 3/8/006 PROCEDURES These procedures set

More information

Measurement Uncertainty: A practical guide to understanding what your results really mean.

Measurement Uncertainty: A practical guide to understanding what your results really mean. Measurement Uncertainty: A practical guide to understanding what your results really mean. Overview General Factors Influencing Data Variability Measurement Uncertainty as an Indicator of Data Variability

More information

TNI V1M Standard Update Guidance on Detection and Quantitation

TNI V1M Standard Update Guidance on Detection and Quantitation TNI V1M4 2016 Standard Update Guidance on Detection and Quantitation GUID-3-109-Rev0 January 30, 2019 This material represents the opinion of its authors. It is intended solely as guidance and does not

More information

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability UV Hound Series Portable UVDOAS Multi-gas Analyzers Multi-Gas Capability Dependable High Quality Air Monitoring Individual VOCs PPB Sensitivity Non-Contact Optical Measurement Automated Reporting Configurable

More information

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

Comparison of Particulate Monitoring Methods at Fort Air Partnership Monitoring Stations

Comparison of Particulate Monitoring Methods at Fort Air Partnership Monitoring Stations Comparison of Particulate Monitoring Methods at Fort Air Partnership Monitoring Stations Melanie Larsen Harry Benders RS Environmental (Tom Dann) March 13, 2014 Executive Summary Historically FAP has acquired

More information

FTIR GAS ANALYSIS WHITE PAPER INTRODUCTION TYPICAL APPLICATIONS

FTIR GAS ANALYSIS WHITE PAPER INTRODUCTION TYPICAL APPLICATIONS WHITE PAPER W FTIR GAS ANALYSIS INTRODUCTION FTIR (Fourier Transform InfraRed) spectroscopy is the most popular analytical technology for industrial applications requiring the continuous measurement of

More information

Update on ASTM and Standard Methods method development activities

Update on ASTM and Standard Methods method development activities Update on ASTM and Standard Methods method development activities William Lipps, Eurofins Eaton Analytical Monrovia South Bend www.eatonanalytical.com Full Service Laboratories Eurofins Lancaster Laboratories

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

Alternative Monitoring Technologies: A Brief Overview

Alternative Monitoring Technologies: A Brief Overview Alternative Monitoring Technologies: A Brief Overview a technical solution to meet every need Cemtek Environmental Inc. 3041 S. Orange Ave. Santa Ana, CA 92707 800-400-0200 www.cemteks.com 1 What s New?

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

Laboratory Analytical Data

Laboratory Analytical Data Laboratory Analytical Data NEWMOA Data Collection & Interpretation: State of Practice & Lessons Learned Jim Occhialini jocchialini@alphalab.com "Interface" verb meaning to blend, ally, coalesce, combine,

More information

Utilizing Data Trend Evaluations in Method Audits and Regulations Development

Utilizing Data Trend Evaluations in Method Audits and Regulations Development Utilizing Data Trend Evaluations in Method Audits and Regulations Development National Environmental Monitoring and The NELAC Institute (TNI) Conference 8/09/2018 Presented by Cornelius (Andy) Valkenburg

More information

AMBIENT AIR MONITORING MONTHLY DATA REPORT

AMBIENT AIR MONITORING MONTHLY DATA REPORT MAXXAM ANALYTICS #1 2080 39 Ave. NE, Calgary, AB T2E 6P7 maxxam.ca Toll Free 800-386-7247 Fax 403-219-3673 AMBIENT AIR MONITORING MONTHLY DATA REPORT PEACE RIVER AREA MONITORING PROGRAM COMMITTEE RENO

More information

Low Level Phosphorus: A Method Comparison Study

Low Level Phosphorus: A Method Comparison Study Low Level Phosphorus: A Method Comparison Study Katie Adams* Stephanie Le, Theresa McBride, Tom Pearson EPA Region 10 Laboratory August 2012 8/29/2012 1 U.S. Environmental Protection Agency Goal: 1 µg/l

More information

FLAME PHOTOMETRY AIM INTRODUCTION

FLAME PHOTOMETRY AIM INTRODUCTION FLAME PHOTOMETRY AIM INTRODUCTION Atomic spectroscopy is based on the absorption, emission or fluorescence process of light by atoms or elementary ions. Information for atomic scale is obtained in two

More information

DRAFT. The NELAC Institute Presents VERIFICATION OPTIONS FOR

DRAFT. The NELAC Institute Presents VERIFICATION OPTIONS FOR DRAFT The NELAC Institute Presents VERIFICATION OPTIONS FOR TNI Limit of Detection A laboratory's estimate of the minimum amount of an analyte in a given matrix that an analytical process can reliably

More information

THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION

THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION JPACSM 5 Process Analytical Chemistry THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION Peter Traynor and Steven Trimuar Thermo Electron Corporation Houston, TX Robert G. Wright Thermo

More information

Assessing Unknown Odors and Chemicals

Assessing Unknown Odors and Chemicals Treat yourself to better rental service. Thank You for Attending Today s Webinar: Assessing Unknown Odors and Chemicals Your Host Matt DeLacluyse Operations Mgr RAECO Rents mattd@raecorents.com Featured

More information

TO-15 Checklist Determination of VOCs in Air by GC-MS

TO-15 Checklist Determination of VOCs in Air by GC-MS LAB ID: DATE: LAB NAME: ASSESSOR NAME: Method Number: TO-15 Checklist Determination of VOCs in Air by GC-MS SOP Number: Revision Number: SOP Date: Personnel records observed: Data records observed: Revision

More information

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350)

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350) Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350) Ann Strahl Technical Support Remediation Division TCEQ 512-239-2500 astrahl@tceq.state.tx.us 1 Data

More information

PERFORMANCE VERIFICATION IONSCAN LS

PERFORMANCE VERIFICATION IONSCAN LS Page of 3 PERFORMANCE VERIFICATION IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ 07059 908--900 Copyright Smiths Detection 004 All rights reserved. Reproduction, adaptation or translation without

More information

Table of Contents I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:...

Table of Contents I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:... Section Table of Contents Page No I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:... 4 (A) Introduction... 4 (B) Data Package

More information

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High

More information

Results of the EPA Method 1631 Validation Study

Results of the EPA Method 1631 Validation Study Results of the EPA Method 1631 Validation Study U.S. EPA Office of Water Office of Science and Technology Engineering and Analysis Division Washington, DC 20460 July 1996 Acknowledgments This report was

More information

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800)

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800) 2773 Downhill Drive Steamboat Springs, CO 80487 (800) 334-5493 Analytical Report November 19, 2004 Report to: Debbie Furlow 602 Lehmberg Road Columbus, MS 39702 Bill to: Accounts Payable PO Box 2487 Columbus,

More information

PA DEP SOUTHWEST REGIONAL OFFICE

PA DEP SOUTHWEST REGIONAL OFFICE PA DEP SOUTHWEST REGIONAL OFFICE MEMO TO FROM Air Quality Permit File: OP-04-00043 / Thomas J. Joseph, P.E. Facilities Engineering Manager Air Quality Program THROUGH Mark R. Gorog, P.E. Environmental

More information

METHOD MIDGET IMPINGER HCl/Cl EMISSION SAMPLING TRAIN

METHOD MIDGET IMPINGER HCl/Cl EMISSION SAMPLING TRAIN METHOD 0051 MIDGET IMPINGER HCl/Cl EMISSION SAMPLING TRAIN 1.0 SCOPE AND APPLICATION This method describes the collection of hydrogen chloride (HCl, CAS Registry Number 7647010) and chlorine (Cl, CAS Registry

More information

Sample RM Analytical Report ( B) HPLC/UV Analysis Acetaldehyde, Acrolein, Crotonaldehyde, Formaldehyde

Sample RM Analytical Report ( B) HPLC/UV Analysis Acetaldehyde, Acrolein, Crotonaldehyde, Formaldehyde For the benefit of VaporFi, Inc. Sample RM080514 Analytical Report (0814-52B) HPLC/UV Analysis Acetaldehyde, Acrolein, Crotonaldehyde, Formaldehyde GC/MS/MS Analysis 1-Aminonaphthalene, 2-Aminonaphthalene,

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016 QA/QC in the Wastewater Laboratory Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016 What is Quality Assurance? Per Standard Methods for the Examination of Water and Wastewater, 21 st

More information

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA)

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) QA/QC Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) Is a description of the policies, procedures, techniques

More information

Hydra-C Application Note: 1076

Hydra-C Application Note: 1076 Hydra-C Application Note: 1076 The Determination of Mercury in Samples by U.S. EPA SOW 846 Method 7473 using the Hydra-C Mercury Analyzer Introduction The accurate determination of mercury in various samples

More information

Method 8270C PAH. FD- Field Duplicate Samples

Method 8270C PAH. FD- Field Duplicate Samples Mr. John Rendall Maine Yankee 321 Old Ferry Road Wiscasset, Maine 04578 RE: Project No.: Maine Yankee Sediment Samples Lab Name: Katahdin Analytical Services, Westbrook, Maine Site Name: Maine Yankee Nuclear

More information

Introduction to IH Analytical Chemistry

Introduction to IH Analytical Chemistry Introduction to IH Analytical Chemistry Sampling and Analysis Manning, Eide, Van Etten (Apr 2013) Foreword Thank you for purchasing the Introduction to IH Analytical Chemistry self-study workbook. Course

More information

SYNOPSIS OF CHANGES TO BC LABORATORY MANUAL, MARCH 2017

SYNOPSIS OF CHANGES TO BC LABORATORY MANUAL, MARCH 2017 SYNOPSIS OF CHANGES TO BC LABORATORY MANUAL, MARCH 2017 A. INTRODUCTION The British Columbia Ministry of Environment (MOE) has updated the Laboratory Manual with new analytical methods for use under the

More information

AOAC Official Method 2016.xx. Determination of Total Sulfur in Fertilizers by High Temperature Combustion

AOAC Official Method 2016.xx. Determination of Total Sulfur in Fertilizers by High Temperature Combustion AOAC Official Method 2016.xx Determination of Total Sulfur in Fertilizers by High Temperature Combustion Proposed First Action 2015 (Applicable for measuring total sulfur concentration in solid and liquid

More information

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote 1 April 2015 European Union s CODEX COMMITTEE ON PESTICIDE RESIDUES 47 th Session Beijing, China, 13 18 April 2015 AGENDA ITEM 8 PROPOSED DRAFT GUIDELINES ON PERFORMANCE CRITERIA SPECIFIC FOR METHODS OF

More information

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015 Laboratory Techniques 100: Back To Basics Carol Injasoulian Lab Manager City of Bay City April 29,2015 QA/QC for Environmental Measurement What is quality assurance (QA)? Broad plan to maintain quality

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD)

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD) 874 METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and supplies)

More information

Chemistry Instrumental Analysis Lecture 15. Chem 4631

Chemistry Instrumental Analysis Lecture 15. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 15 IR Instruments Types of Instrumentation Dispersive Spectrophotometers (gratings) Fourier transform spectrometers (interferometer) Single beam Double beam

More information

Analytical Performance & Method. Validation

Analytical Performance & Method. Validation Analytical Performance & Method Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

Process Analyzer and Sampling Systems

Process Analyzer and Sampling Systems Process Analyzer and Sampling Systems Course Price 3050 Course Description Analytical instruments used for online chemical analysis of process streams or plant environments are generally called Process

More information

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement 3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing, but also as an energy carrier in the cooling or heating cycle. As vast amounts

More information

! ## " # " &"& &") &"* )"- /!.IN-SITU! )"2 )", /101.3 KPa+/K "EN 14181!

! ##  #  && &) &* )- /!.IN-SITU! )2 ), /101.3 KPa+/K EN 14181! 34 EN 14181 %analyzer$ ' Stationary source emissions, Quality assurance of automated (BS EN 14181 )(measuring system Stationary source emissions, Automated measuring systems (BS EN 13284 2. *(Determination

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

Laboratory Evaluation of EPA Methods 26 and 26A for Analysis of Halogens and Halides in Stack Gas

Laboratory Evaluation of EPA Methods 26 and 26A for Analysis of Halogens and Halides in Stack Gas Laboratory Evaluation of EPA Methods 26 and 26A for Analysis of Halogens and Halides in Stack Gas 3002003347 Laboratory Evaluation of EPA Methods 26 and 26A for Analysis of Halogens and Halides in Stack

More information

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis LGC Quality ISO Guide 34:2009 DAkkS D-RM-14176-01-00 ISO/IEC 17025:2005 DAkkS D-PL-14176-01-00 ISO 9001:2008 DQS 102448 QM08 Primary Reference Standard Granisetron Hydrochloride

More information

Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications

Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications Application Note 35140710 Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications Keywords 1030S Solids Module Aurora 1030 TOC Analyzer High Temperature Catalytic

More information

Introduction to Gas Phase FTIR Spectroscopy

Introduction to Gas Phase FTIR Spectroscopy Introduction to Gas Phase FTIR Spectroscopy Introduction to FTIR spectroscopy FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared (IR) spectroscopy, radiation

More information

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS Agilent Technologies/Wasson ECE Monomer Analyzer Application Authors Fred Feyerherm 119 Forest Cove Dr. Kingwood, TX

More information

2 SPECTROSCOPIC ANALYSIS

2 SPECTROSCOPIC ANALYSIS 2 SPECTROSCOPIC ANALYSIS 2.1 Introduction Chemical analysis falls into two basic categories: qualitative what is present quantitative how much is present Spectroscopy is capable of both types of analysis,

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS FOR ERRORS AND UPDATES, PLEASE VISIT WWW.TSFX.COM.AU/VCE-UPDATES QUESTION 45 QUESTION 46 Answer is A QUESTION 47 The number of protons in the element.

More information

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air The Application of Method TO-15 to Naphthalene Measurements in Indoor Air Extended Abstract #13 Heidi C. Hayes and Diane J. Benton Air Toxics Ltd. 180 Blue Ravine Rd. Ste. B Folsom, CA 95630 INTRODUCTION

More information

Technical Contact. Randy Moore. (910) Highway 421 North Wilmington, NC 28401

Technical Contact. Randy Moore. (910) Highway 421 North Wilmington, NC 28401 NORTH CAROLINA DIVISION OF AIR QUALITY Air Permit Review Permit Issue Date: Facility Data Applicant (Facility s Name): Wilbara, LLC Facility Address: Wilbara, LLC SIC: 2819 / Industrial Inorganic Chemicals

More information

New good practices for emission measurements. and relevance of the SRM with respect to these requirements/practices CEM 2014

New good practices for emission measurements. and relevance of the SRM with respect to these requirements/practices CEM 2014 New good practices for emission measurements and relevance of the SRM with respect to these requirements/practices CEM 2014 Jean Poulleau Cécile Raventos Evolution of French rules for emission measurements

More information

The New MDL Procedure How To s. Presented by: Marcy Bolek - Alloway

The New MDL Procedure How To s. Presented by: Marcy Bolek - Alloway The New MDL Procedure How To s Presented by: Marcy Bolek - Alloway Proposed MDL Revision 2015 MUR How to obtain a copy of the proposed MDL revision? https://www.gpo.gov/fdsys/pkg/fr-2015-02-19/pdf/2015-02841.pdf

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

UV Quality Assurance. Dark Current Correction

UV Quality Assurance. Dark Current Correction UV Quality Assurance This document presents the Quality Assurance protocol used for the measurement of broad spectrum UV spectra according to the guidelines given by the United States Environmental Protection

More information