METHOD ,2-DIBROMOETHANE AND 1,2-DIBROMO-3-CHLOROPROPANE BY MICROEXTRACTION AND GAS CHROMATOGRAPHY

Size: px
Start display at page:

Download "METHOD ,2-DIBROMOETHANE AND 1,2-DIBROMO-3-CHLOROPROPANE BY MICROEXTRACTION AND GAS CHROMATOGRAPHY"

Transcription

1 METHOD ,2-DIBROMOETHANE AND 1,2-DIBROMO-3-CHLOROPROPANE BY MICROEXTRACTION AND GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 This methd is applicable t the determinatin f the fllwing cmpunds in drinking water and grund water: Cmpund Name CAS N. a 1,2-Dibrmethane (EDB) ,2-Dibrm-3-chlrprpane (DBCP) a Chemical Abstract Services Registry Number. 1.2 Fr cmpunds and matrices ther than thse listed in Sectin 1.1, the labratry must demnstrate the usefulness f the methd by cllecting precisin and accuracy data n actual samples and prvide qualitative cnfirmatin f results by gas chrmatgraphy/mass spectrmetry (GC/MS). 1.3 The experimentally determined methd detectin limits (MDL) fr EDB and DBCP were calculated t be 0.01 µg/l. The methd has been shwn t be useful fr these analytes ver a cncentratin range f apprximately 0.03 t 200 µg/l. Actual detectin limits are highly dependent upn the characteristics f the gas chrmatgraphic system, sample matrix, and calibratin. 1.4 This methd is restricted t use by r under the Supervisin f analysts experienced in the use f gas chrmatgraphy and in the interpretatin f gas chrmatgrams. Each analyst must demnstrate the ability t generate acceptable results with this methd using the prcedure described in Sectin ,2-Dibrmethane and 1,2-Dibrm-3-chlrprpane have been tentatively classified as knwn r suspected human r mammalian carcingens. Pure standard materials and stck standard slutins f these cmpunds shuld be handled in a hd. A NIOSH/MESA apprved txic gas respiratr shuld be wrn when the analyst handles high cncentratins f these txic cmpunds. 2.0 SUMMARY OF METHOD 2.1 Thirty five ml f sample are extracted with 2 ml f hexane. Tw µl f the extract are then injected int a gas chrmatgraph equipped with a linearized electrn capture detectr fr separatin and analysis. Aqueus matrix spikes are extracted and analyzed in an identical manner as the samples in rder t cmpensate fr pssible extractin lsses. CD-ROM Revisin 0

2 2.2 The extractin and analysis time is 30 t 50 minutes per sample depending upn the analytical cnditins chsen. See Table 1 and Figure Cnfirmatry evidence is btained using a different clumn (Table 1). 3.0 INTERFERENCES 3.1 Impurities cntained in the extracting slvent (hexane) usually accunt fr the majrity f the analytical prblems. Reagent blanks shuld be analyzed fr each new bttle f hexane befre use. Indirect daily checks n the hexane are btained by mnitring the reagent blanks. Whenever an interference is nted in the methd r instrument blank, the labratry shuld reanalyze the hexane. Lw level interferences generally can be remved by distillatin r clumn chrmatgraphy, hwever, it is generally mre ecnmical t btain a new surce f hexane slvent. Interference-free hexane is defined as cntaining less than 0.01 µg/l f the analytes. Prtect interference-free hexane by string it in an area knwn t be free f rganchlrine slvents. 3.2 Several instances f accidental sample cntaminatin have been attributed t diffusin f vlatile rganics thrugh the septum seal int the sample bttle during shipment and strage. Trip blanks must be used t mnitr fr this prblem. 3.3 This liquid/liquid extractin technique extracts a wide biling range f nn-plar rganic cmpunds and, in additin, extracts sme plar rganic cmpunds. 3.4 EDB at lw cncentratins may be masked by very high cncentratins f dibrmchlrmethane (DBCM), a cmmn chlrinated drinking water cntaminant, when using the cnfirmatin clumn. 4.0 APPARATUS AND MATERIALS 4.1 Micrsyringe - 10, 25, and 100 µl with a 2 in. x in. needle (Hamiltn 702N r equivalent). 4.2 Gas Chrmatgraph The GC must be capable f temperature prgramming and shuld be equipped with a linearized electrn capture detectr and a capillary clumn splitless injectr Clumns Clumn A mm ID x 30 m fused silica capillary with dimethyl silicne mixed phase (Durawax-DX 3, 0.25 µm film, r equivalent) Clumn B (cnfirmatin clumn) mm ID x 30 m fused silica capillary with methyl plysilxane phase (DB-1, 0.25 µm film, r equivalent). CD-ROM Revisin 0

3 4.3 Vlumetric flasks, Class A - 10 ml. 4.4 Glass bttles - 15 ml, with Tefln lined screw caps r crimp tps. 4.5 Analytical balance g. 4.6 Graduated cylinder - 50 ml. 4.7 Transfer pipet. 5.0 REAGENTS 5.1 Reagent grade chemicals shall be used in all tests. Unless therwise indicated, it is intended that all reagents shall cnfrm t the specificatins f the Cmmittee n Analytical Reagents f the American Chemical Sciety, where such specificatins are available. Other grades may be used, prvided it is first ascertained that the reagent is f sufficiently high purity t permit its use withut lessening the accuracy f the determinatin. 5.2 Organic-free reagent water - All references t water in this methd refer t rganic-free reagent water, as defined in Chapter One. 5.3 Hexane, C H - UV grade (Burdick and Jacksn #216 r equivalent) Methyl alchl, CH OH - Demnstrated t be free f analytes Sdium chlride, NaCl - Pulverize a batch f NaCl and place it in a muffle furnace at rm temperature. Increase the temperature t 400 C fr 30 minutes. Stre in a capped bttle ,2-Dibrmethane (99%), C H Br, (Aldrich Chemical Cmpany, r equivalent) ,2-Dibrme-3-chlrprpane (99.4%), C H Br C1, (AMVAC Chemical Crpratin, Ls Angeles, Califrnia, r equivalent). 5.8 Stck standards - These slutins may be purchased as certified slutins r prepared frm pure standards using the fllwing prcedures: Place abut 9.8 ml f methanl int a 10 ml grund glass stppered vlumetric flask. Allw the flask t stand, unstppered, fr abut 10 minutes and weigh t the nearest g Use a 25 µl syringe and immediately add tw r mre drps (. 10 µl) f standard t the flask. Be sure that the standard falls directly int the alchl withut cntacting the neck f the flask Reweigh, dilute t vlume, stpper, and then mix by inverting the flask several times. Calculate the cncentratin in milligrams per liter (mg/l) frm the net gain in weight. When cmpund purity is assayed t be 96% r greater, the weight may be used withut crrectin t calculate the cncentratin f the stck standard. CD-ROM Revisin 0

4 5.8.4 Stre stck standards in 15 ml bttles equipped with Tefln lined screw-caps r crimp tps. Stck standards are stable fr at least fur weeks when stred at 4 C and away frm light. 5.9 Intermediate standard - Use stck standards t prepare an intermediate standard that cntains bth analytes in methanl. The intermediate standard shuld be prepared at a cncentratin that can be easily diluted t prepare aqueus calibratin standards that will bracket the wrking cncentratin range. Stre the intermediate standard with minimal headspace and check frequently fr signs f deteriratin r evapratin, especially just befre preparing calibratin standards. The strage time described fr stck standards als applies t the intermediate standard Quality cntrl (QC) reference sample - Prepare a QC reference sample cncentrate at 0.25 mg/l f bth analytes frm standards frm a different surce than the standards used fr the stck standard Check standard - Add an apprpriate vlume f the intermediate standard t an aliqut f rganic-free reagent water in a vlumetric flask. D nt add mre than 20 µl f an alchlic intermediate standard t the water r pr precisin will result. Use a 25 µl micrsyringe and rapidly inject the alchlic intermediate standard int the expanded area f the almst filled vlumetric flask. Remve the needle as quickly as pssible after injectin. Mix by inverting the flask several times. Discard the cntents cntained in the neck f the flask. Aqueus calibratin standards shuld be prepared every 8 hurs. 6.0 SAMPLE COLLECTION, PRESERVATION, AND STORAGE 6.1 See the intrductry material t this chapter, Organic Analytes, Sectin PROCEDURE 7.1 Recmmended Chrmatgraphic Cnditins Tw gas chrmatgraphy clumns are recmmended. Clumn A is a highly efficient clumn that prvides separatins fr EDB and DBCP withut interferences frm trihalmethanes. Clumn A shuld be used as the primary analytical clumn unless rutinely ccurring analytes are nt adequately reslved. Clumn B is recmmended fr use as a cnfirmatry clumn when GC/MS cnfirmatin is nt available. Retentin times fr EDB and DBCP n these clumns are presented in Table 1. Clumn A: Injectr temperature: 200 C. Detectr temperature: 290 C. Carrier gas (Helium)Linear velcity: 25 cm/sec. Temperature prgram: Initial temperature: Prgram: 40 C, hld fr 4 min. 40 C t 190 C at 8 C/min. CD-ROM Revisin 0

5 Final temperature: 190 C, hld fr 25 min., r until all expected analytes have eluted. See Figure 1 fr a sample chrmatgram and Table 1 fr retentin data. Clumn B: Injectr temperature: 200 C. Detectr temperature: 290 C. Carrier gas (Helium)Linear velcity: 25 cm/sec. Temperature prgram: Initial temperature: Prgram: Final temperature: See Table 1 fr retentin data. 7.2 Calibratin 40 C, hld fr 4 min. 40 C t 270 C at 10 C/min. 270 C, hld fr 10 min., r until all expected analytes have eluted Prepare at least five calibratin standards. One shuld cntain EDB and DBCP at a cncentratin near, but greater than, the methd detectin limit (Table 1) fr each cmpund. The thers shuld be at cncentratins that bracket the range expected in the samples. Fr example, if the MDL is 0.01 µg/l, and a sample expected t cntain apprximately 0.10 µg/l is t be analyzed, aqueus calibratin standards shuld be prepared at cncentratins f 0.03 µg/l, 0.05 µg/l, 0.10 µg/l, 0.15 µg/l, and 0.20 µg/l Analyze each calibratin standard and tabulate peak height r area respnse versus the cncentratin in the standard. Prepare a calibratin curve fr each cmpund. Alternatively, if the rati f respnse t cncentratin (calibratin factr) is a cnstant ver the wrking range (< 10% relative standard deviatin), linearity can be assumed and the average rati r calibratin factr can be used in place f a calibratin curve. 7.3 Sample preparatin Remve samples and standards frm strage and allw them t reach rm temperature Fr samples and field blanks cntained in 40 ml bttles, remve the cntainer cap. Discard a 5 ml vlume using a 5 ml transfer pipet. Replace the cntainer cap and weigh the cntainer with cntents t the nearest 0.1 g and recrd this weight fr subsequent sample vlume determinatin Fr calibratin standards, check standards, QC reference samples, and blanks, measure a 35 ml vlume using a 50 ml graduated cylinder and transfer it t a 40 ml sample cntainer. 7.4 Extractin CD-ROM Revisin 0

6 7.4.1 Remve the cntainer cap and add 7 g f NaCl t all samples Recap the sample cntainer and disslve the NaCl by shaking by hand fr abut 20 secnds Remve the cap and using a transfer pipet, add 2.0 ml f hexane. Recap and shake vigrusly by hand fr 1 minute. Allw the water and hexane phases t separate. If stred at this stage, keep the cntainer upside dwn Remve the cap and carefully transfer a sufficient amunt ( ml) f the hexane layer int a vial using a dispsable glass pipet Transfer the remaining hexane phase, being careful nt t include any f the water phase, int a secnd vial. Reserve this secnd vial at 4 C fr reanalysis if necessary. 7.5 Analysis Transfer the first sample vial t an autsampler set up t inject 2.0 µl prtins int the gas chrmatgraph fr analysis. Alternately, 2 µl prtins f samples, blanks and standards may be manually injected, using the slvent flush technique, althugh an aut sampler is strngly recmmended. 7.6 Determinatin f sample vlume Fr samples and field blanks, remve the cap frm the sample cntainer. Discard the remaining sample/hexane mixture. Shake ff the remaining few drps using shrt, brisk wrist mvements. Reweigh the empty cntainer with riginal cap and calculate the net weight f sample by difference t the nearest 0.1 g. This net weight is equivalent t the vlume f water extracted. 7.7 Calculatins Identify EDB and DBCP in the sample chrmatgram by cmparing the retentin time f the suspect peak t retentin times generated by the calibratin standards and the check standard Use the calibratin curve r calibratin factr t directly calculate the uncrrected cncentratin (C i) f each analyte in the sample (e.g. calibratin factr x respnse) Calculate the sample vlume (V ) as equal t the net sample s weight: V (ml) = grss weight (grams) - bttle tare (grams) s Calculate the crrected sample cncentratin as: Cncentratin (µg/l) = C x 35 i V s CD-ROM Revisin 0

7 7.7.5 Reprt the results fr the unknwn samples in µg/l. Rund the results t the nearest 0.01 µg/l r tw significant figures. 8.0 QUALITY CONTROL 8.1 Each labratry that uses this methd is required t perate a frmal quality cntrl prgram The labratry must make an initial determinatin f the methd detectin limits and demnstrate the ability t generate acceptable accuracy and precisin with this methd. This is established as described in Sectin In recgnitin f labratry advances that are ccurring in chrmatgraphy, the labratry is permitted certain ptins t imprve the separatins r lwer the cst f measurements. Each time such a mdificatin is made t the methd, the analyst is required t repeat the prcedure in Sectin 7.1 and The labratry must analyze a reagent and calibratin blank t demnstrate that interferences frm the analytical system are under cntrl every twenty samples r per analytical batch, whichever is mre frequent The labratry must, n an nging basis, demnstrate thrugh the analyses f QC reference samples and check standards that the peratin f the measurement system is in cntrl. The frequency f the check standard analyses is equivalent t 5% f all samples r every analytical batch, whichever is mre frequent. On a weekly basis, the QC reference sample must be run. 8.2 T establish the ability t achieve lw detectin limits and generate acceptable accuracy and precisin, the analyst must perfrm the fllwing peratins: Prepare seven samples each at a cncentratin f 0.03 µg/l Analyze the samples accrding t the methd beginning in Sectin 7.0. _ Calculate the average cncentratin (X) in µg/l and the standard deviatin f the cncentratins (s) in µg/l, fr each analyte using the seven results. Then calculate the MDL at 99% cnfidence level fr seven replicates as 3.143s. _ Fr each analyte in an aqueus matrix sample, X must be between 60% and 140% f the true value. Additinally, the MDL may nt exceed the 0.03 µg/l spiked cncentratin. If bth analytes meet the acceptance criteria, the system perfrmance is acceptable and analysis f actual samples can begin. If either analyte fails t meet a criterin, repeat the test. It is recmmended that the labratry repeat the MDL determinatin n a regular basis. CD-ROM Revisin 0

8 8.3 The labratry must demnstrate n a frequency equivalent t 5% f the sample lad r nce per analytical batch, whichever is mre frequent, that the measurement system is in cntrl by analyzing a check standard f bth analytes at 0.25 µg/l Prepare a check standard (0.25 µg/l) by diluting the intermediate standard with water t 0.25 µg/l Analyze the sample accrding t Sectin 7.0 and calculate the recvery fr each analyte. The recvery must be between 60% and 140% f the expected value fr aqueus matrices. Fr nn-aqueus matrices, the U.S. EPA will set criteria after mre interlabratry data are gathered If the recvery fr either analyte falls utside the designated range, the analyte fails the acceptance criteria. A secnd calibratin verificatin standard cntaining each analyte that failed must be analyzed. Repeated failure, hwever, will cnfirm a general prblem with the measurement system. If this ccurs, lcate and crrect the surce f the prblem and repeat the test. 8.4 On a weekly basis, the labratry must demnstrate the ability t analyze a QC reference sample Prepare a QC reference sample at 0.10 µg/l by diluting the QC reference sample cncentrate (Sectin 5.9) Fr each analyte in an aqueus matrix, the recvery must be between 60% and 140% f the expected value. When either analyte fails the test, the analyst must repeat the test nly fr that analyte which failed t meet the criteria. Repeated failure, hwever, will cnfirm a general prblem with the measurement system r faulty samples and/r standards. If this ccurs, lcate and crrect the surce f the prblem and repeat the test. Fr nn-aqueus matrices, the U.S. EPA will set criteria after mre interlabratry data are gathered. 8.5 Instrument perfrmance - Check the perfrmance f the entire analytical system daily using data gathered frm analyses f blanks, standards, and replicate samples Peak tailing significantly in excess f that shwn in the chrmatgram (Figure 1) must be crrected. Tailing prblems are generally traceable t active sites n the GC clumn r t the detectr peratin Check the precisin between replicate analyses. A prperly perating system shuld perfrm with an average relative standard deviatin f less than 10%. Pr precisin is generally traceable t pneumatic leaks, especially at the injectin prt. 9.0 METHOD PERFORMANCE 9.1 Methd detectin limits are presented in Table 1. Single labratry accuracy and precisin at several cncentratins in tap water are presented in Table 2. CD-ROM Revisin 0

9 9.2 In a preservatin study extending ver a 4 week perid, the average percent recveries and relative standard deviatins presented in Table 3 were bserved fr rganic-free reagent water (acidified), tap water and grund water. The results fr acidified and nn-acidified samples were nt significantly different REFERENCES 1. Optimizatin f Liquid-Liquid Extractin Methds fr Analysis f Organics in Water, EPA-600/S , Hendersn, J.E.; Peytn, G.R.; Glaze, W.H. Identificatin and Analysis f Organic Pllutants in Water; Keith, L.H., Ed; Ann Arbr Sci.: Ann Arbr, MI; Richard J.J.; Junk, G.A. Jurnal AWWA 1977, 69, Budde, W.L.; Eichelberger, J.W. Organic Analyses Using Gas Chrmatgraphy- Mass Spectrmetry; Ann Arbr Science: Ann Arbr, MI; Glaser, J.A.; et al. Envirnmental Science and Technlgy 1981, 15, Methds fr the Determinatin f Organic Cmpunds in Finished Drinking Water and Raw Surce Water; U.S. Envirnmental Prtectin Agency. Office f Research and Develpment. Envirnmental Mnitring and Supprt Labratry. ORD Publicatin Offices f Center fr Envirnmental Research Infrmatin: Cincinnati, OH CD-ROM Revisin 0

10 TABLE 1. CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS (MDL) FOR 1,2-DIBROMOETHANE (EDB) AND 1,2-DIBROMO-3-CHLOROPROPANE (DBCP) Retentin Time, Minutes Analyte Clumn A Clumn B MDL (µg/l) EDB DBCP Clumn A: Durawax-DX 3 Clumn B: DB-1 TABLE 2. SINGLE LABORATORY ACCURACY AND PRECISION FOR EDB AND DBCP IN TAP WATER Relative Number Spike Average Standard f Cncentratin Recvery Deviatin Analyte Samples (µg/l) (%) (%) EDB DBCP CD-ROM Revisin 0

11 TABLE 3. ACCURACY AND PRECISION AT 2.0 µg/l OVER A 4-WEEK STUDY PERIOD Average Relative Number Accuracy Std. Dev. Analyte 1 Matrix f Samples (% Recvery) (%) EDB RW-A GW GW-A TW TW-A DBCP RW-A GW GW-A TW TW-A RW-A = Organic-free reagent water at ph 2 GW = Grund water, ambient ph GW-A = Grund water at ph 2 TW = Tap water, ambient ph TW-A = Tap water at ph 2 CD-ROM Revisin 0

12 FIGURE 1. SAMPLE CHROMATOGRAM FOR EXTRACT OF WATER SPIKED AT µg/l WITH EDB AND DBCP CD-ROM Revisin 0

13 METHOD ,2-DIBROMOETHANE AND 1,2-DIBROMO-3-CHLOROPROPANE BY MICROEXTRACTION AND GAS CHROMATOGRAPHY CD-ROM Revisin 0

METHOD 5040A ANALYSIS OF SORBENT CARTRIDGES FROM VOLATILE ORGANIC SAMPLING TRAIN (VOST): GAS CHROMATOGRAPHY/MASS SPECTROMETRY TECHNIQUE

METHOD 5040A ANALYSIS OF SORBENT CARTRIDGES FROM VOLATILE ORGANIC SAMPLING TRAIN (VOST): GAS CHROMATOGRAPHY/MASS SPECTROMETRY TECHNIQUE METHOD 5040A ANALYSIS OF SORBENT CARTRIDGES FROM VOLATILE ORGANIC SAMPLING TRAIN (VOST): GAS CHROMATOGRAPHY/MASS SPECTROMETRY TECHNIQUE 1.0 SCOPE AND APPLICATION 1.1 Methd 5040 was frmerly Methd 3720 in

More information

METHOD 3650A ACID-BASE PARTITION CLEANUP. 1.1 Method 3650 was formerly Method 3530 in the second edition of this manual.

METHOD 3650A ACID-BASE PARTITION CLEANUP. 1.1 Method 3650 was formerly Method 3530 in the second edition of this manual. 1.0 SCOPE AND APPLICATION METHOD 3650A ACID-BASE PARTITION CLEANUP 1.1 Methd 3650 was frmerly Methd 3530 in the secnd editin f this manual. 1.2 Methd 3650 is a liquid-liquid partitining cleanup methd t

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information

NF Ethylcellulose Assay Procedure

NF Ethylcellulose Assay Procedure The Dw Chemical Cmpany September 2, 2011 NF Ethylcellulse Assay Prcedure Purpse This dcument is written t prvide guidance fr analysts cnducting the NF Ethylcellulse assay test methd. The prcedure utlined

More information

Method for Chloride Determination by Ferric Thiocyanate, Version 2.0

Method for Chloride Determination by Ferric Thiocyanate, Version 2.0 Methd fr Chlride Determinatin by Ferric Thicyanate, Versin 2.0 PRINCIPLE This methd is designed fr the determinatin f chlride in water, sil, and ther frms f samples. The sample is prepared ffline if necessary

More information

METHOD 8275 THERMAL CHROMATOGRAPHY/MASS SPECTROMETRY (TC/MS) FOR SCREENING SEMIVOLATILE ORGANIC COMPOUNDS

METHOD 8275 THERMAL CHROMATOGRAPHY/MASS SPECTROMETRY (TC/MS) FOR SCREENING SEMIVOLATILE ORGANIC COMPOUNDS METHOD 8275 THERMAL CHROMATOGRAPHY/MASS SPECTROMETRY (TC/MS) FOR SCREENING SEMIVOLATILE ORGANIC COMPOUNDS 1.0 SCOPE AND APPLICATION 1.1 Methd 8275 is a screening technique that may be used fr the qualitative

More information

Method for Ammonia Determination by Salicylate, Version 2.2

Method for Ammonia Determination by Salicylate, Version 2.2 Methd fr Ammnia Determinatin by Salicylate, Versin 2.2 PRINCIPLE This methd is designed fr the determinatin f ammnia in water, sil, and ther frms f samples. The sample is prepared ffline if necessary and

More information

METHOD 5041 PROTOCOL FOR ANALYSIS OF SORBENT CARTRIDGES FROM VOLATILE ORGANIC SAMPLING TRAIN (VOST): WIDE-BORE CAPILLARY COLUMN TECHNIQUE

METHOD 5041 PROTOCOL FOR ANALYSIS OF SORBENT CARTRIDGES FROM VOLATILE ORGANIC SAMPLING TRAIN (VOST): WIDE-BORE CAPILLARY COLUMN TECHNIQUE METHOD 5041 PROTOCOL FOR ANALYSIS OF SORBENT CARTRIDGES FROM VOLATILE ORGANIC SAMPLING TRAIN (VOST): WIDE-BORE CAPILLARY COLUMN TECHNIQUE 1.0 SCOPE AND APPLICATION 1.1 This methd describes the analysis

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

METHOD 8151 CHLORINATED HERBICIDES BY GC USING METHYLATION OR PENTAFLUOROBENZYLATION DERIVATIZATION: CAPILLARY COLUMN TECHNIQUE

METHOD 8151 CHLORINATED HERBICIDES BY GC USING METHYLATION OR PENTAFLUOROBENZYLATION DERIVATIZATION: CAPILLARY COLUMN TECHNIQUE METHOD 8151 CHLORINATED HERBICIDES BY GC USING METHYLATION OR PENTAFLUOROBENZYLATION DERIVATIZATION: CAPILLARY COLUMN TECHNIQUE 1.0 SCOPE AND APPLICATION 1.1 Methd 8151 is a capillary gas chrmatgraphic

More information

Method for Ammonia Determination by Salicylate in Water and TKN Samples, Version 2.0 PRINCIPLE SUMMARY COMPLIANCE

Method for Ammonia Determination by Salicylate in Water and TKN Samples, Version 2.0 PRINCIPLE SUMMARY COMPLIANCE Methd fr Ammnia Determinatin by Salicylate in Water and TKN Samples, Versin 2.0 PRINCIPLE This methd is designed fr the determinatin f ammnia in water, as well as samples digested by the Ttal Kjedahl Nitrgen

More information

Method for Nitrate Determination by Cadmium Reduction, Version 2.3

Method for Nitrate Determination by Cadmium Reduction, Version 2.3 Methd fr Nitrate Determinatin by Cadmium Reductin, Versin 2.3 PRINCIPLE This methd is designed fr the determinatin f nitrate in sil, water, and ther frms f samples. The sample is prepared ffline if necessary

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

UNIT 6 DETERMINATION OF FLASH AND FIRE POINT OF A LUBRICATING OIL BY OPEN CUP AND CLOSED CUP METHODS

UNIT 6 DETERMINATION OF FLASH AND FIRE POINT OF A LUBRICATING OIL BY OPEN CUP AND CLOSED CUP METHODS UNIT 6 DETERMINATION OF FLASH AND FIRE POINT OF A LUBRICATING OIL BY OPEN CUP AND CLOSED CUP METHODS Determinatin f Flash and Fire Pint f a Cup and Clsed Cup Structure 6. Intrductin Objectives 6. Experiment

More information

DEFENSE OCCUPATIONAL AND ENVIRONMENTAL HEALTH READINESS SYSTEM (DOEHRS) ENVIRONMENTAL HEALTH SAMPLING ELECTRONIC DATA DELIVERABLE (EDD) GUIDE

DEFENSE OCCUPATIONAL AND ENVIRONMENTAL HEALTH READINESS SYSTEM (DOEHRS) ENVIRONMENTAL HEALTH SAMPLING ELECTRONIC DATA DELIVERABLE (EDD) GUIDE DEFENSE OCCUPATIOL AND ENVIRONMENTAL HEALTH READINESS SYSTEM (DOEHRS) ENVIRONMENTAL HEALTH SAMPLING ELECTRONIC DATA DELIVERABLE (EDD) GUIDE 20 JUNE 2017 V1.0 i TABLE OF CONTENTS 1 INTRODUCTION... 1 2 CONCEPT

More information

Total Nitric Oxide Assay Kit

Total Nitric Oxide Assay Kit Ttal Nitric Oxide Assay Kit Catalg Number: EMSNOTOT Prduct descriptin A cmplete kit fr the quantitative determinatin f ttal nitric xide (NO) in bilgical fluids. The kit uses the enzyme nitrate reductase

More information

Molecule Labs 780 Clark Avenue Pittsburg, CA 94565

Molecule Labs 780 Clark Avenue Pittsburg, CA 94565 Mlecule Labs 780 Clark Avenue Pittsburg, CA 94565 Carbnyl Assay Vn Vape Nble 0.6% Analytical Reprt (0615-572C1) GC/MS Analysis (NT225-Draft) Diacetyl 2,3-Pentanedine (aka Acetyl prpinyl) nthalpy Analytical,

More information

USP Hypromellose Assay Procedure

USP Hypromellose Assay Procedure July 13, 2015 USP Hyprmellse Assay Prcedure Purpse This dcument is written t prvide guidance fr analysts cnducting the USP Hyprmellse assay test methd. The prcedure utlined in this dcument is based n the

More information

Competency Requirement Details for each badge (information added to badge metadata in

Competency Requirement Details for each badge (information added to badge metadata in Using digital badges fr develping high-schl chemistry labratry skills Nami Hennah and Michael K. Seery 1. Nrthamptn Schl fr Bys, Billing Rd, Nrthamptn NN1 5RT, UK. 2. EaStCHEM Schl f Chemistry, University

More information

Sampling and Analytical Methods for Wild Rice Waters. Environmental Analysis and Outcomes Division

Sampling and Analytical Methods for Wild Rice Waters. Environmental Analysis and Outcomes Division Sampling and Analytical Methds fr Wild Rice Waters July 2017 Envirnmental Analysis and Outcmes Divisin The analytical methds and sampling prcedures prvided in this dcument are incrprated by reference in

More information

Pipetting 101 Developed by BSU CityLab

Pipetting 101 Developed by BSU CityLab Discver the Micrbes Within: The Wlbachia Prject Pipetting 101 Develped by BSU CityLab Clr Cmparisns Pipetting Exercise #1 STUDENT OBJECTIVES Students will be able t: Chse the crrect size micrpipette fr

More information

METHOD OF TEST FOR THE DETERMINATION OF POTENTIAL ALKALI-CARBONATE REACTIVITY OF. CARBONATE ROCKS BY CHEMICAL COMPOSITION

METHOD OF TEST FOR THE DETERMINATION OF POTENTIAL ALKALI-CARBONATE REACTIVITY OF. CARBONATE ROCKS BY CHEMICAL COMPOSITION Labratry Testing Manual Date., 96 5 1 Page 1 f 5 METHOD OF TEST FOR THE DETERMINATION OF POTENTIAL ALKALI-CARBONATE REACTIVITY OF. CARBONATE ROCKS BY CHEMICAL COMPOSITION 1. SCOPE 1.1 This methd cvers

More information

Materials o o o o o o o o o

Materials o o o o o o o o o Experiment 3: Measurements, the Metric System & Density Objective The purpse f experiment is t becme familiar with the metric system by taking measurements using metric. Additinally, the purpse f this

More information

QUESTIONNAIRE. for RECOGNITION of GOOD LABORATORY PRACTICE. SRI LANKA ACCREDITATION BOARD for CONFORMITY ASSESSMENT. Instructions to the Applicant:

QUESTIONNAIRE. for RECOGNITION of GOOD LABORATORY PRACTICE. SRI LANKA ACCREDITATION BOARD for CONFORMITY ASSESSMENT. Instructions to the Applicant: SRI LANKA ACCREDITATION BOARD fr CONFORMITY ASSESSMENT QUESTIONNAIRE fr RECOGNITION f GOOD LABORATORY PRACTICE Instructins t the Applicant: 1. Please fill the questinnaire n yur wn judgment f activities

More information

Determining the Accuracy of Modal Parameter Estimation Methods

Determining the Accuracy of Modal Parameter Estimation Methods Determining the Accuracy f Mdal Parameter Estimatin Methds by Michael Lee Ph.D., P.E. & Mar Richardsn Ph.D. Structural Measurement Systems Milpitas, CA Abstract The mst cmmn type f mdal testing system

More information

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011 ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011 1,2-DIBROMOETHANE (EDB) AND 1,2-DIBROMO-3-CHLOROPROPANE (DBCP) IN WATER BY MICROEXTRACTION AND GAS CHROMATOGRAPHY SOP #: 504.1 Revision 1.1,

More information

THERMAL TEST LEVELS & DURATIONS

THERMAL TEST LEVELS & DURATIONS PREFERRED RELIABILITY PAGE 1 OF 7 PRACTICES PRACTICE NO. PT-TE-144 Practice: 1 Perfrm thermal dwell test n prtflight hardware ver the temperature range f +75 C/-2 C (applied at the thermal cntrl/munting

More information

BASD HIGH SCHOOL FORMAL LAB REPORT

BASD HIGH SCHOOL FORMAL LAB REPORT BASD HIGH SCHOOL FORMAL LAB REPORT *WARNING: After an explanatin f what t include in each sectin, there is an example f hw the sectin might lk using a sample experiment Keep in mind, the sample lab used

More information

RN52-STK2 Starter Kit

RN52-STK2 Starter Kit The RN52-STK2 Starter Kit has everything yu need t kick-start yur prject and is a great tl fr develpers: Prf f cncept prttypes as well as final cmmercial slutins. fr educatinal use: Natural sciences curses

More information

Physics 2010 Motion with Constant Acceleration Experiment 1

Physics 2010 Motion with Constant Acceleration Experiment 1 . Physics 00 Mtin with Cnstant Acceleratin Experiment In this lab, we will study the mtin f a glider as it accelerates dwnhill n a tilted air track. The glider is supprted ver the air track by a cushin

More information

Guidelines for laboratory safety and chemical use at the Abisko Scientific Research Station

Guidelines for laboratory safety and chemical use at the Abisko Scientific Research Station Guidelines fr labratry safety and chemical use at the Abisk Scientific Research Statin Welcme t the Abisk Scientific Research Statin labratry facilities. Labratry wrk can be hazardus. Many f the materials

More information

Unit 11 Solutions- Guided Notes. What are alloys? What is the difference between heterogeneous and homogeneous mixtures?

Unit 11 Solutions- Guided Notes. What are alloys? What is the difference between heterogeneous and homogeneous mixtures? Name: Perid: Unit 11 Slutins- Guided Ntes Mixtures: What is a mixture and give examples? What is a pure substance? What are allys? What is the difference between hetergeneus and hmgeneus mixtures? Slutins:

More information

Chapter 3 Forensic Laboratory Techniques Cengage Learning. All Rights Reserved

Chapter 3 Forensic Laboratory Techniques Cengage Learning. All Rights Reserved Chapter 3 Frensic Labratry Techniques Intrductin Objectives 1. Distinguish between physical and chemical prperties. 2. Describe presumptive and cnfirmatry tests. 3. Cmpare and cntrast different types f

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Absolute Precision in Particle Size Analysis

Absolute Precision in Particle Size Analysis Abslute Precisin in Particle Size Analysis Authr: Dr Graham Rideal Funder and Senir Scientist Whitehuse Scientific Ltd Summary Abslute precisin in particle size analysis can nly be achieved when the tw

More information

February 28, 2013 COMMENTS ON DIFFUSION, DIFFUSIVITY AND DERIVATION OF HYPERBOLIC EQUATIONS DESCRIBING THE DIFFUSION PHENOMENA

February 28, 2013 COMMENTS ON DIFFUSION, DIFFUSIVITY AND DERIVATION OF HYPERBOLIC EQUATIONS DESCRIBING THE DIFFUSION PHENOMENA February 28, 2013 COMMENTS ON DIFFUSION, DIFFUSIVITY AND DERIVATION OF HYPERBOLIC EQUATIONS DESCRIBING THE DIFFUSION PHENOMENA Mental Experiment regarding 1D randm walk Cnsider a cntainer f gas in thermal

More information

PHOTOSYNTHESIS THE PRACTICALS 16 APRIL 2014

PHOTOSYNTHESIS THE PRACTICALS 16 APRIL 2014 PHOTOSYNTHESIS THE PRACTICALS 16 APRIL 2014 Lessn Descriptin In this lessn, we will: Review the prcess f phtsynthesis Study the starch test in leaves Study the varius practicals testing phtsynthesis Lk

More information

ENSC Discrete Time Systems. Project Outline. Semester

ENSC Discrete Time Systems. Project Outline. Semester ENSC 49 - iscrete Time Systems Prject Outline Semester 006-1. Objectives The gal f the prject is t design a channel fading simulatr. Upn successful cmpletin f the prject, yu will reinfrce yur understanding

More information

2004 AP CHEMISTRY FREE-RESPONSE QUESTIONS

2004 AP CHEMISTRY FREE-RESPONSE QUESTIONS 2004 AP CHEMISTRY FREE-RESPONSE QUESTIONS 6. An electrchemical cell is cnstructed with an pen switch, as shwn in the diagram abve. A strip f Sn and a strip f an unknwn metal, X, are used as electrdes.

More information

A.P. CHEMISTRY. SOLUTIONS AND ACID BASE CHEMISTRY. p 1

A.P. CHEMISTRY. SOLUTIONS AND ACID BASE CHEMISTRY. p 1 A.P. CHEMISTRY. SOLUTIONS AND ACID BASE CHEMISTRY. p 1 (Nte: questins 1 t 14 are meant t be dne WITHOUT calculatrs!) 1.Which f the fllwing is prbably true fr a slid slute with a highly endthermic heat

More information

THERMAL-VACUUM VERSUS THERMAL- ATMOSPHERIC TESTS OF ELECTRONIC ASSEMBLIES

THERMAL-VACUUM VERSUS THERMAL- ATMOSPHERIC TESTS OF ELECTRONIC ASSEMBLIES PREFERRED RELIABILITY PAGE 1 OF 5 PRACTICES PRACTICE NO. PT-TE-1409 THERMAL-VACUUM VERSUS THERMAL- ATMOSPHERIC Practice: Perfrm all thermal envirnmental tests n electrnic spaceflight hardware in a flight-like

More information

How do scientists measure trees? What is DBH?

How do scientists measure trees? What is DBH? Hw d scientists measure trees? What is DBH? Purpse Students develp an understanding f tree size and hw scientists measure trees. Students bserve and measure tree ckies and explre the relatinship between

More information

Lab 4: Passive Transport & Graphing Data

Lab 4: Passive Transport & Graphing Data CWI Cncepts f Bilgy LAB Manual 42 Lab 4: Passive Transprt & Graphing Data The internal envirnment f all cells is a slutin. A slutin is a mixture f tw r mre substances that are evenly distributed thrughut.

More information

RN53 Radon Detector. RN53-STK Starter Kit

RN53 Radon Detector. RN53-STK Starter Kit The RN53-STK Starter Kit has everything yu need t kick-start yur prject and is a great tl fr develpers: Prf f cncept prttypes as well as final cmmercial slutins. fr educatinal use: Natural science curses

More information

Hypothesis Tests for One Population Mean

Hypothesis Tests for One Population Mean Hypthesis Tests fr One Ppulatin Mean Chapter 9 Ala Abdelbaki Objective Objective: T estimate the value f ne ppulatin mean Inferential statistics using statistics in rder t estimate parameters We will be

More information

Weathering. Title: Chemical and Mechanical Weathering. Grade Level: Subject/Content: Earth and Space Science

Weathering. Title: Chemical and Mechanical Weathering. Grade Level: Subject/Content: Earth and Space Science Weathering Title: Chemical and Mechanical Weathering Grade Level: 9-12 Subject/Cntent: Earth and Space Science Summary f Lessn: Students will test hw chemical and mechanical weathering can affect a rck

More information

CORESTA Guide N 10 March 2011

CORESTA Guide N 10 March 2011 CORESTA Guide N 10 March 2011 A User Guideline fr the Measurement f Diameter f Cigarettes and Filter Rds Sub-Grup Physical Test Methds 1 Review and Histry Issue Date Reasn 1.0 July 2009 Original issue.

More information

CHM112 Lab Graphing with Excel Grading Rubric

CHM112 Lab Graphing with Excel Grading Rubric Name CHM112 Lab Graphing with Excel Grading Rubric Criteria Pints pssible Pints earned Graphs crrectly pltted and adhere t all guidelines (including descriptive title, prperly frmatted axes, trendline

More information

Name: Block: Date: Science 10: The Great Geyser Experiment A controlled experiment

Name: Block: Date: Science 10: The Great Geyser Experiment A controlled experiment Science 10: The Great Geyser Experiment A cntrlled experiment Yu will prduce a GEYSER by drpping Ments int a bttle f diet pp Sme questins t think abut are: What are yu ging t test? What are yu ging t measure?

More information

/ / Chemistry. Chapter 1 Chemical Foundations

/ / Chemistry. Chapter 1 Chemical Foundations Name Chapter 1 Chemical Fundatins Advanced Chemistry / / Metric Cnversins All measurements in chemistry are made using the metric system. In using the metric system yu must be able t cnvert between ne

More information

Component Certification Specification for Manual and Automated Pool Cleaners (CCS-15965)

Component Certification Specification for Manual and Automated Pool Cleaners (CCS-15965) NSF Internatinal Recreatinal Water Prgram Cmpnent Certificatin Specificatin fr Manual and Autmated Pl Cleaners (CCS-15965) The Public Health and Safety Cmpany. TM Recreatinal Water Prgram Cmpnent Certificatin

More information

IR Emitter and Detector Product Data Sheet LTE-C9306 Spec No.: DS Effective Date: 04/20/2007 LITE-ON DCC RELEASE

IR Emitter and Detector Product Data Sheet LTE-C9306 Spec No.: DS Effective Date: 04/20/2007 LITE-ON DCC RELEASE IR Emitter and Detectr Prduct Data Sheet LTE-C936 Spec N.: DS5-27-1 Effective Date: 4/2/27 Revisin: - LITE-ON DCC RELEASE BNS-OD-FC1/A4 LITE-ON Technlgy Crp. / Optelectrnics N.9,Chien 1 Rad, Chung H, New

More information

Mathematics and Computer Sciences Department. o Work Experience, General. o Open Entry/Exit. Distance (Hybrid Online) for online supported courses

Mathematics and Computer Sciences Department. o Work Experience, General. o Open Entry/Exit. Distance (Hybrid Online) for online supported courses SECTION A - Curse Infrmatin 1. Curse ID: 2. Curse Title: 3. Divisin: 4. Department: 5. Subject: 6. Shrt Curse Title: 7. Effective Term:: MATH 70S Integrated Intermediate Algebra Natural Sciences Divisin

More information

Floating Point Method for Solving Transportation. Problems with Additional Constraints

Floating Point Method for Solving Transportation. Problems with Additional Constraints Internatinal Mathematical Frum, Vl. 6, 20, n. 40, 983-992 Flating Pint Methd fr Slving Transprtatin Prblems with Additinal Cnstraints P. Pandian and D. Anuradha Department f Mathematics, Schl f Advanced

More information

METHOD 551. July J.W. Hodgeson A.L. Cohen

METHOD 551. July J.W. Hodgeson A.L. Cohen METHOD 551 DETERMINATION OF CHLORINATION DISINFECTION BYPRODUCTS AND CHLORINATED SOLVENTS IN DRINKING WATER BY LIQUID-LIQUID EXTRACTION AND GAS CHROMATOGRAPHY WITH ELECTRON-CAPTURE DETECTION July 1990

More information

Semester 2 AP Chemistry Unit 12

Semester 2 AP Chemistry Unit 12 Cmmn In Effect and Buffers PwerPint The cmmn in effect The shift in equilibrium caused by the additin f a cmpund having an in in cmmn with the disslved substance The presence f the excess ins frm the disslved

More information

CHAPTER 3 INEQUALITIES. Copyright -The Institute of Chartered Accountants of India

CHAPTER 3 INEQUALITIES. Copyright -The Institute of Chartered Accountants of India CHAPTER 3 INEQUALITIES Cpyright -The Institute f Chartered Accuntants f India INEQUALITIES LEARNING OBJECTIVES One f the widely used decisin making prblems, nwadays, is t decide n the ptimal mix f scarce

More information

Bootstrap Method > # Purpose: understand how bootstrap method works > obs=c(11.96, 5.03, 67.40, 16.07, 31.50, 7.73, 11.10, 22.38) > n=length(obs) >

Bootstrap Method > # Purpose: understand how bootstrap method works > obs=c(11.96, 5.03, 67.40, 16.07, 31.50, 7.73, 11.10, 22.38) > n=length(obs) > Btstrap Methd > # Purpse: understand hw btstrap methd wrks > bs=c(11.96, 5.03, 67.40, 16.07, 31.50, 7.73, 11.10, 22.38) > n=length(bs) > mean(bs) [1] 21.64625 > # estimate f lambda > lambda = 1/mean(bs);

More information

Experiment #3. Graphing with Excel

Experiment #3. Graphing with Excel Experiment #3. Graphing with Excel Study the "Graphing with Excel" instructins that have been prvided. Additinal help with learning t use Excel can be fund n several web sites, including http://www.ncsu.edu/labwrite/res/gt/gt-

More information

Part One: Heat Changes and Thermochemistry. This aspect of Thermodynamics was dealt with in Chapter 6. (Review)

Part One: Heat Changes and Thermochemistry. This aspect of Thermodynamics was dealt with in Chapter 6. (Review) CHAPTER 18: THERMODYNAMICS AND EQUILIBRIUM Part One: Heat Changes and Thermchemistry This aspect f Thermdynamics was dealt with in Chapter 6. (Review) A. Statement f First Law. (Sectin 18.1) 1. U ttal

More information

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS 1.0 SCOPE AND APPLICATION 1.1 Method 5000 provides general guidance on the selection of sample preparation methods (purge-and-trap, extraction,

More information

AMERICAN PETROLEUM INSTITUTE API RP 581 RISK BASED INSPECTION BASE RESOURCE DOCUMENT BALLOT COVER PAGE

AMERICAN PETROLEUM INSTITUTE API RP 581 RISK BASED INSPECTION BASE RESOURCE DOCUMENT BALLOT COVER PAGE Ballt ID: Title: USING LIFE EXTENSION FACTOR (LEF) TO INCREASE BUNDLE INSPECTION INTERVAL Purpse: 1. Prvides a methd t increase a bundle s inspectin interval t accunt fr LEF. 2. Clarifies Table 8.6.5 Als

More information

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS PAGE: 1 of 18 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Wagon Markings Guidelines

Wagon Markings Guidelines Versin / Status: V 3.0 / apprved Wagn Markings Guidelines 1. Intrductin Article 4, para 4 f the Safety Directive (2004/49/EG amended by 2008/110/EC) stipulates the respnsibility f each manufacturer, maintenance

More information

Pattern Recognition 2014 Support Vector Machines

Pattern Recognition 2014 Support Vector Machines Pattern Recgnitin 2014 Supprt Vectr Machines Ad Feelders Universiteit Utrecht Ad Feelders ( Universiteit Utrecht ) Pattern Recgnitin 1 / 55 Overview 1 Separable Case 2 Kernel Functins 3 Allwing Errrs (Sft

More information

A New Evaluation Measure. J. Joiner and L. Werner. The problems of evaluation and the needed criteria of evaluation

A New Evaluation Measure. J. Joiner and L. Werner. The problems of evaluation and the needed criteria of evaluation III-l III. A New Evaluatin Measure J. Jiner and L. Werner Abstract The prblems f evaluatin and the needed criteria f evaluatin measures in the SMART system f infrmatin retrieval are reviewed and discussed.

More information

o o IMPORTANT REMINDERS Reports will be graded largely on their ability to clearly communicate results and important conclusions.

o o IMPORTANT REMINDERS Reports will be graded largely on their ability to clearly communicate results and important conclusions. BASD High Schl Frmal Lab Reprt GENERAL INFORMATION 12 pt Times New Rman fnt Duble-spaced, if required by yur teacher 1 inch margins n all sides (tp, bttm, left, and right) Always write in third persn (avid

More information

Unit 1 Study Guide Name Date Scientific Method Notes

Unit 1 Study Guide Name Date Scientific Method Notes Unit 1 Study Guide Name Date Scientific Methd Ntes 1) What is the difference between an bservatin and an inference? 2) What are the tw types f bservatins? Give examples f each type. 3) Define the fllwing:

More information

Standard Title: Frequency Response and Frequency Bias Setting. Andrew Dressel Holly Hawkins Maureen Long Scott Miller

Standard Title: Frequency Response and Frequency Bias Setting. Andrew Dressel Holly Hawkins Maureen Long Scott Miller Template fr Quality Review f NERC Reliability Standard BAL-003-1 Frequency Respnse and Frequency Bias Setting Basic Infrmatin: Prject number: 2007-12 Standard number: BAL-003-1 Prject title: Frequency

More information

Edexcel IGCSE Chemistry. Topic 1: Principles of chemistry. Chemical formulae, equations and calculations. Notes.

Edexcel IGCSE Chemistry. Topic 1: Principles of chemistry. Chemical formulae, equations and calculations. Notes. Edexcel IGCSE Chemistry Tpic 1: Principles f chemistry Chemical frmulae, equatins and calculatins Ntes 1.25 write wrd equatins and balanced chemical equatins (including state symbls): fr reactins studied

More information

YEAR 6 (PART A) Textbook 6A schema

YEAR 6 (PART A) Textbook 6A schema YEAR 6 (PART A) Textbk 6A schema Chapter 1 Numbers t 10 Millin Lessn 1 Reading and Writing Numbers t 10 Millin T create and identify numbers t 10 000 000; t write in numerals and wrds numbers t 10 000

More information

TUTORIAL B1.31 TB 538

TUTORIAL B1.31 TB 538 TUTORIAL B1.31 TB 538 RECOMMENDATIONS FOR TESTING OF SUPERCONDUCTING CABLES Cnvener: D Lindsay (US) Secretary: T Masuda (JP) Page 1 Cpyright DISCLAIMER Ownership f a CIGRE publicatin, whether in paper

More information

Name AP CHEM / / Chapter 1 Chemical Foundations

Name AP CHEM / / Chapter 1 Chemical Foundations Name AP CHEM / / Chapter 1 Chemical Fundatins Metric Cnversins All measurements in chemistry are made using the metric system. In using the metric system yu must be able t cnvert between ne value and anther.

More information

1. Transformer A transformer is used to obtain the approximate output voltage of the power supply. The output of the transformer is still AC.

1. Transformer A transformer is used to obtain the approximate output voltage of the power supply. The output of the transformer is still AC. PHYSIS 536 Experiment 4: D Pwer Supply I. Intrductin The prcess f changing A t D is investigated in this experiment. An integrated circuit regulatr makes it easy t cnstruct a high-perfrmance vltage surce

More information

Differentiation Applications 1: Related Rates

Differentiation Applications 1: Related Rates Differentiatin Applicatins 1: Related Rates 151 Differentiatin Applicatins 1: Related Rates Mdel 1: Sliding Ladder 10 ladder y 10 ladder 10 ladder A 10 ft ladder is leaning against a wall when the bttm

More information

, which yields. where z1. and z2

, which yields. where z1. and z2 The Gaussian r Nrmal PDF, Page 1 The Gaussian r Nrmal Prbability Density Functin Authr: Jhn M Cimbala, Penn State University Latest revisin: 11 September 13 The Gaussian r Nrmal Prbability Density Functin

More information

100KG/300KG/600KG/1000KG/2000KG MAGNETIC LIFTER

100KG/300KG/600KG/1000KG/2000KG MAGNETIC LIFTER Vlume 1 OPERATIONAL MANUAL MODEL: 100KG/300KG/600KG/1000KG/2000KG MAGNETIC LIFTER by BLUEROCK Tls 100KG/3 0 0 K G / 6 0 0 K G / 1 0 0 0 K G / 2 0 0 0 K G M A G N E T I C L I F T E R UNPACKING THE ITEM

More information

+ Charge attraction between a

+ Charge attraction between a 1 Types f Intermlecular Frces: Strength Interactin Picture Descriptin est Lndn Dispersin Frces (induced dipleinduced diple) + + + + + + Attractin between temprary induced diples in nn-plar mlecules. Diple-Diple

More information

Sediment Basin (SB) Description. Appropriate Uses. Design and Installation

Sediment Basin (SB) Description. Appropriate Uses. Design and Installation Descriptin A sediment basin is a temprary pnd built n a cnstructin site t capture erded r disturbed sil transprted in strm runff prir t discharge frm the site. Sediment basins are designed t capture site

More information

WRITING THE REPORT. Organizing the report. Title Page. Table of Contents

WRITING THE REPORT. Organizing the report. Title Page. Table of Contents WRITING THE REPORT Organizing the reprt Mst reprts shuld be rganized in the fllwing manner. Smetime there is a valid reasn t include extra chapters in within the bdy f the reprt. 1. Title page 2. Executive

More information

5.0 minutes. The temperature rose from

5.0 minutes. The temperature rose from QUESTION 1 The table belw describes features cmmn t a slutin calrimeter and a bmb calrimeter as well as differences. Features f Bth the Slutin and Bmb alrimeters W X Features f the Bmb alrimeter Only Y

More information

Rotator PTR-35. Operating instructions

Rotator PTR-35. Operating instructions Rtatr PTR-35 Versin 3.01 - January 2012 Page 2 Cntents 1 Safety...4 2 General Infrmatin...5 3 Getting Started...6 4 Operatin f PTR-35...8 5 Specificatins...11 6 Guarantee and service...13 Versin 3.01 -

More information

Department of Electrical Engineering, University of Waterloo. Introduction

Department of Electrical Engineering, University of Waterloo. Introduction Sectin 4: Sequential Circuits Majr Tpics Types f sequential circuits Flip-flps Analysis f clcked sequential circuits Mre and Mealy machines Design f clcked sequential circuits State transitin design methd

More information

Activity 2 Dimensional Analysis

Activity 2 Dimensional Analysis Activity 2 Dimensinal Analysis Gals! Develp cnversin factrs frm cmmn equalities.! Use cnversin factrs t cnvert between different units f measure.! Apply the cncept f dimensinal analysis t string tgether

More information

Temperature sensor / Dual Temp+Humidity

Temperature sensor / Dual Temp+Humidity www.akcp.cm Temperature sensr / Dual Temp+Humidity Intrductin Temperature sensrs are imprtant where ptimum temperature cntrl is paramunt. If there is an air cnditining malfunctin r abnrmal weather cnditins,

More information

Effect of Paramagnetic Ions in Aqueous Solution for Precision Measurement of Proton Gyromagnetic Ratio

Effect of Paramagnetic Ions in Aqueous Solution for Precision Measurement of Proton Gyromagnetic Ratio 240 Bulletin f Magnetic Resnance Effect f Paramagnetic Ins in Aqueus Slutin fr Precisin Measurement f Prtn Gyrmagnetic Rati Ae Ran Lim, Chang Suk Kim Krea Research Institute f Standards and Science, Taejn

More information

Temperature sensor / Dual Temp+Humidity

Temperature sensor / Dual Temp+Humidity www.akcp.cm Temperature sensr / Dual Temp+Humidity Intrductin Temperature sensrs are imprtant where ptimum temperature cntrl is paramunt. If there is an air cnditining malfunctin r abnrmal weather cnditins,

More information

The standards are taught in the following sequence.

The standards are taught in the following sequence. B L U E V A L L E Y D I S T R I C T C U R R I C U L U M MATHEMATICS Third Grade In grade 3, instructinal time shuld fcus n fur critical areas: (1) develping understanding f multiplicatin and divisin and

More information

For more information about how to cite these materials visit

For more information about how to cite these materials visit Authr(s): MELO 3D Prject Team, 2011 License: This wrk is licensed under the Creative Cmmns Attributin-ShareAlike 3.0 Unprted License. T view a cpy f this license, visit http://creativecmmns.rg/licenses/by-sa/3.0/.

More information

Tutorial 3: Building a spectral library in Skyline

Tutorial 3: Building a spectral library in Skyline SRM Curse 2013 Tutrial 3 Spectral Library Tutrial 3: Building a spectral library in Skyline Spectral libraries fr SRM methd design and fr data analysis can be either directly added t a Skyline dcument

More information

Acids and Bases Lesson 3

Acids and Bases Lesson 3 Acids and Bases Lessn 3 The ph f a slutin is defined as the negative lgarithm, t the base ten, f the hydrnium in cncentratin. In a neutral slutin at 25 C, the hydrnium in and the hydrxide in cncentratins

More information

Supporting information

Supporting information Electrnic Supplementary Material (ESI) fr Physical Chemistry Chemical Physics This jurnal is The wner Scieties 01 ydrgen perxide electrchemistry n platinum: twards understanding the xygen reductin reactin

More information

Group Color: Subgroup Number: How Science Works. Grade 5. Module 2. Class Question: Scientist (Your Name): Teacher s Name: SciTrek Volunteer s Name:

Group Color: Subgroup Number: How Science Works. Grade 5. Module 2. Class Question: Scientist (Your Name): Teacher s Name: SciTrek Volunteer s Name: Grup Clr: Subgrup Number: Hw Science Wrks Grade 5 Mdule 2 Class Questin: Scientist (Yur Name): Teacher s Name: SciTrek Vlunteer s Name: VOCABULARY Science: The study f the material wrld using human reasn.

More information

INFORMATION SHEET: LS8 LABORATORY SAFETY SAFE USE OF HIGH SPEED CENTRIFUGES

INFORMATION SHEET: LS8 LABORATORY SAFETY SAFE USE OF HIGH SPEED CENTRIFUGES INFORMATION SHEET: LS8 LABORATORY SAFETY SAFE USE OF HIGH SPEED CENTRIFUGES This dcument prvides guidance n hw t use high-speed centrifuges (which can, if nt perated crrectly, cause serius injury and damage)

More information

CONTENTS. VII. Functions VIII. Specifications IX. Included with tia NC-9900 X. Troubleshooting

CONTENTS. VII. Functions VIII. Specifications IX. Included with tia NC-9900 X. Troubleshooting CONTENTS I. Safety Precautins II. Intrductin III. Prir t Use IV. Prduct Descriptin (tia NC-9900) V. Using yur tia VI. Settings Unit f Measure Alarm Setup Bdy r Surface Mde Ttal Variatin Laser / Light Setup

More information

Study Group Report: Plate-fin Heat Exchangers: AEA Technology

Study Group Report: Plate-fin Heat Exchangers: AEA Technology Study Grup Reprt: Plate-fin Heat Exchangers: AEA Technlgy The prblem under study cncerned the apparent discrepancy between a series f experiments using a plate fin heat exchanger and the classical thery

More information

Forensic Science. Group: Background information

Forensic Science. Group: Background information Frensic Science Grup: Backgrund infrmatin One f the graduate students in the Department f Bilgy n the campus f the University f Flrida went missing. A week later, a crime scene was discvered n a remte

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

Application Note - Energy Metering at a Medium Voltage Connection Point using a Voltage Transformer

Application Note - Energy Metering at a Medium Voltage Connection Point using a Voltage Transformer Applicatin Nte - Energy Metering at a Medium Vltage Cnnectin Pint using a Vltage Transfrmer This dcument describes the cnnectin and cnfiguratin f a Janitza meter at a medium vltage (MV) cnnectin pint t

More information

Biocomputers. [edit]scientific Background

Biocomputers. [edit]scientific Background Bicmputers Frm Wikipedia, the free encyclpedia Bicmputers use systems f bilgically derived mlecules, such as DNA and prteins, t perfrm cmputatinal calculatins invlving string, retrieving, and prcessing

More information