ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011

Size: px
Start display at page:

Download "ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011"

Transcription

1 ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011 1,2-DIBROMOETHANE (EDB) AND 1,2-DIBROMO-3-CHLOROPROPANE (DBCP) IN WATER BY MICROEXTRACTION AND GAS CHROMATOGRAPHY SOP #: Revision 1.1, 1995 & SW Originator: Section Supervisor: QA Manager Date: Date: Date: Page 1 of 23

2 Section 1.0 Scope and Application 1.1 This is a gas chromatographic (GC) method applicable to the determination of the following compounds in ground water and finished drinking water. ANALYTE 1,2-Dibromoethane 1,2-Dibromo-3-chloropropane 1.2 When this method is used to analyze unfamiliar samples for any or all of the analytes above, analyte identifications must be confirmed by at least one additional qualitative technique. Section 2.0 Summary of Method mL of sample are salted with approximately 6 g of baked sodium chloride and are extracted with 2-mL hexane. Two L of the extract are then injected into a gas chromatograph equipped with a linearized electron capture detector for separation and analysis. Aqueous calibration standards are extracted and analyzed in an identical manner to the samples in order to compensate for possible extraction losses. Section 3.0 Definitions 3.1 Acceptance Criteria specified limits placed on characteristics of an item, process, or service defined in required documents. 3.2 Analytical Batch composed of prepared environmental samples which are analyzed together as a group. It can include prepared sample originating from various environmental matrices and cannot exceed 20 samples. 3.3 Calibration to determine, by measurement or comparison with a standard, the correct value of each scale reading on a meter, instrument, or other device. The levels of the applied calibration standard should bracket the range of planned or expected sample measurements. 3.4 Calibration Curve the graphical relationship between the known values, such as concentrations, of a series of calibration standards and their instrument response. Page 2 of 23

3 3.5 Calibration Standard (CAL) -- A solution prepared from the primary dilution standard solution. The CAL is used to calibrate an instrument. 3.6 Field Blank or Trip Blank blank prepared in the field or laboratory by filling a clean container with pure de-ionized water and appropriate preservative, if any, for the specific sampling activity being undertaken. 3. High Quality Pure Reagent Water water in which no target analytes or interferences are detected as required by the analytical method. 3.8 Holding Times (Maximum Allowable Holding Times) the maximum times that samples may be held prior to analysis and still be considered valid or not compromised. 3.9 Internal Standard a known amount of standard added to a test portion of a sample as a reference for evaluating and controlling the precision and bias of the applied analytical method Laboratory Reagent Blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware, equipment, solvents, reagents, internal standards, and surrogates that are used with other samples. The LRB is used to determine if method analytes or other interferences are present in the laboratory environment, the reagents, or the apparatus Laboratory Fortified Blank (LFB) -- An aliquot of reagent water to which a known quantity of the method analyte is added in the laboratory. The LFB is analyzed exactly like a sample, and its purpose is to determine whether the method is in control, and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit Laboratory Performance Check Solution (LPC) -- A solution of method analytes, surrogate compounds, and internal standards used to evaluate the performance of the instrument system with respect to a defined set of method criteria Matrix the component or substrate that contains the analyte of interest Matrix Spike (MS) or Laboratory Fortified Matrix (LFM) a sample prepared by adding a known mass of target analyte to a specified amount of matrix sample for which an independent estimate of target analyte concentration is available. Used to determine the effect of the matrix on a method s recovery efficiency. The background concentration of the analyte in the sample matrix must be determined in a separate aliquot and the measured value in the LFM corrected for background concentration. Page 3 of 23

4 3.15 Matrix Spike Duplicate (MSD) or Laboratory Fortified Matrix Duplicate (LFMD) a second replicate of spike prepared in the laboratory and analyzed to obtain a measure of the precision of the recovery for each analyte Primary Dilution Standard Solution (PDS) - A solution of several analytes purchased from a commercial source and diluted as needed to prepare calibration solutions and other needed analyte solutions. 3.1 Procedural Standard Calibration - A calibration method where aqueous calibration standards are prepared and processed (e.g., purged, extracted, and/or derivatized) in exactly the same manner as a sample. All steps in the process from addition of sampling preservatives through instrumental analyses compensates for any inefficiencies in the processing procedure Quality Control the overall system of technical activities whose purpose is to measure and control the quality of a product or service so that it meets the needs of users Raw Data any original factual information from a measurement activity or study recorded in a laboratory notebook, worksheets, records, memoranda, notes, or exact copies thereof that are necessary for the reconstruction and evaluation of the report activity Reference Method a method of known and documented accuracy and precision issued by an organization recognized as competent to do so Standard Operating Procedure (SOP) a written document which details the method of an operation, analysis or action whose techniques and procedures are thoroughly prescribed and which is accepted as the method for performing certain routine or repetitive tasks Stock Standard Solution -- A concentrated solution containing a single certified standard that is a method analyte, or a concentrated solution of a single analytes prepared in the laboratory with an assayed reference compound. Stock standard solutions are used to prepare primary dilution standards Surrogate a substance with properties that mimic the analyte of interest. It is unlikely to be found in environment samples and is added to them for quality control purposes. Calculated as percent recovery. Page 4 of 23

5 Section 4.0 Interferences 4.1 Method interferences may be caused by contaminants in solvents, reagents, glassware, and other sample processing apparatus that lead to anomalous peaks or elevated baselines in liquid chromatograms. The LRB is used as a tool to monitor for the presence of interferences. 4.2 Glassware used for this procedure includes pre-cleaned volatile vials (VOA vials) and pre-cleaned 2 ml auto-sampler vials. 4.3 Interfering contamination may occur when a sample containing low concentrations of compounds is analyzed immediately after a sample containing relatively high concentrations of compounds. Syringes and injection port liners must be cleaned carefully or replaced as needed. After analysis of a sample containing high concentrations of compounds, a solvent blank or an LRB is analyzed to ensure that accurate values are obtained for the next sample. If suspicion of carryover exists, reanalysis of the affected samples is performed. 4.4 Samples and standards must be contained in the same solvent to ensure chromatographic comparability. 4.5 Current column technology suffers from the fact that EDB at low concentrations may be masked by very high levels of dibromochloromethane (DBCM), a common disinfection by-product of chlorinated drinking waters. Section 5.0 Safety 5.1 The toxicity or carcinogenicity of each chemical and reagent used in this method has not been precisely defined. However, each one must be treated as a potential health hazard, and exposure to these chemicals should be minimized. Each analyst is responsible for adherence to the procedures outlined in the Alloway Chemical Hygiene Plan. 5.2 Some method analytes have been tentatively classified as known or suspected human or mammalian carcinogens. Pure standard materials and stock standard solutions of these compounds should be handled with suitable protection to skin, eyes, etc. 5.3 Safety glasses, gloves, and lab coats must be worn when handling samples and solvents. Safety glasses must be worn when handling glassware and Page 5 of 23

6 chromatographic columns. Safety glasses must be worn when handling equipment within the laboratory. 5.4 EDB and DBCP have been tentatively classified as known or suspected human or mammalian carcinogens. Pure standard materials and stock standard solutions of these compounds should be handled accordingly. Section 6.0 Equipment and Supplies 6.1 Sample Bottle - 40 ml VOA vials (pre-cleaned) fitted with Teflon lined screw cap. 6.2 Autosampler vials - Autosampler vials (pre-cleaned) suitable for use on the appropriate autosampler. 6.3 Micro Syringes 10, 25 and 100 l gas-tight micro-syringes. 6.4 Pipetters and 8.0 ml transfer pipetters. 6.5 Gas Chromatograph - Analytical system complete with temperature programmable GC suitable for use with capillary columns and all required accessories including syringes, analytical columns, gasses, detector and data collection Column 1 RTX-CLPesticidesII 30-m x 0.53-mm I.D. x 0.42-m Film (Restek Catalog Number 11340), or equivalent Conditions Oven - 50C, 4 degrees C/min to 60 degrees C/min hold 6 min, then 10 degrees C/min to 145 degrees C then 20 degrees C/min to 200 degrees C and hold 2.25 minutes. Inlet Temperature 0C, 5 degrees C/min to 110 degrees C, then 10 degrees C/min to 180 degrees C, then 15 degrees C/min to 225 degrees C and hold for 4 minutes. Detector C Injector 2.5 l Flow electronic pressure control set at 3.0 psi. Page 6 of 23

7 6.5.2 Column 2 RTX CLPesticides 30m x 0.53 mm I.D. x 0.50 m Film (Restek Catalog Number 11140), or equivalent Conditions Oven - 50C, 4 degrees C/min to 60 degrees C hold 6 min, then 10 degrees C/min to 145 degrees C then 20 degrees C/min to 200 degrees C and hold 2.25 minutes. Inlet Temperature 0C, 5 degrees C/min to 110 degrees C, then 10 degrees C/min to 180 degrees C, then 15 degrees C/min to 225 degrees C and hold for 4 minutes. Detector C Injector 2.5 l Flow electronic pressure control set at Detector - Electron capture. This detector has been proven effective in the analysis of fortified reagent and artificial ground waters. Section.0 Reagents and Calibration Standards.1 Hexane - Pesticides quality or better..2 Methanol - Purge and Trap Grade, or equivalent.3 Sodium Chloride - Baked for 0.5 hours at 400C.4 Stock Standard Solution (200 ug/ml) - Ultra Scientific, DWN-504, or equivalent.5 Calibration Spike Standard 50-ug/L. See Alloway s Reagent traceability software for the preparation of this spiking standard..6 Second Source Standard (200 ug/ml) Restek catalog number or equivalent.. Laboratory Fortified Blank (LFB) and Laboratory Fortified Sample Matrix (LFM) solution - 50-ug/L. See Alloway s reagent traceability for the preparation of the 50-µg/L spike solution. Page of 23

8 .8 High Quality Reagent Water Laboratory DI water meets method blank requirements. Section 8.0 Sample Collection, Preservation, Shipment, and Storage 8.1 Sample Collection Replicate field reagent blanks (FRB) must be handled along with each sample set, which is composed of the samples collected from the same general sampling site at approximately the same time. At the laboratory, fill a minimum of two sample bottles with reagent water, seal, and ship to the sampling site along with the sample bottles. Wherever a set of samples is shipped and stored, it must be accompanied by the FRB Collect all samples in 40 ml VOA bottles into which 3 mg of sodium thiosulfate crystals have been added to the empty bottles just prior to shipping to the sampling site. Alternatively, 5 L of freshly prepared sodium thiosulfate solution (0.04 g/ml) may be added to empty 40 ml VOA bottles just prior to sample collection When sampling from a water tap, open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 minutes). Adjust the flow to about 500 ml/min and collect samples from the flowing stream When sampling from a well, fill a wide-mouth bottle or beaker with sample, and carefully fill 40 ml VOA bottles. 8.2 Sample Preservation and Storage The sample must be iced or refrigerated at 6C or less away from light from the time of collection until extraction Sodium thiosulfate must be added to avoid the possibility of reactions which may occur between residual chlorine and indeterminate contaminants present in some solvents, yielding compounds which may subsequently interfere with the analysis. The presence of sodium thiosulfate will arrest the formation of DBCM (see Section 4.5). Method 8011 samples should be acidified to a ph < 2.0 to avoid the possibility of microbial degradation which may periodically affect these analytes contained in other ground water matrices. Page 8 of 23

9 8.2.3 Holding Times - Preservation study results indicate that the analytes present in the samples are stable for 14 days when stored under these conditions. All samples must be extracted and analyzed within 14 days of collection and when necessary, extracts may be stored in tightly capped vials at 6 degrees C for up to 24 hours. Section 9.0 Calibration and Standardization 9.1 Establish GC operating parameters equivalent to those indicated in Section 6.5. The GC system is calibrated using the external standard technique. 9.2 External Standard Calibration Procedure Prepare seven standards by adding 5, 10, 20, 30, 40, 50, 60 ul of the stock solution (0.05-µg/L) to 35 mls of high quality reagent water. Curve concentrations range from , , , , 0.051, 0.014, and ug/l Each day, extract and analyze each calibration standard according to the procedure (Section 11). Tabulate response (peak area) against concentration for each compound. Calculate the response factor (RF) for each analyte and surrogate. Use a linear or quadratic regression without forcing the curve through zero. Regression curve fits >0.995 should be appropriate. Average Response Factor must be less than 20% RSD for EPA Method Samples (<10% for EPA Method 8011). 9.3 Continuing Calibration Verification (CCV) A calibration standard at mid concentration containing all of the compounds must be analyzed every 10 samples during analysis. For each compound the calculated value must be within 30% of the working value. If not corrective action must be taken and the check standard re-analyzed. If the reanalyzed CCV meets the requirements of 0-130% recovery then analysis may proceed. If the reanalyzed CCV is high (>130%) and the results of samples are below the reporting levels, the results of the samples can be reported without qualification. If the reanalyzed CCV is high (>130 %) indicating a high bias, and the results of samples are detected at or above the reporting limit the samples need to be reanalyzed following the acceptance of a new calibration curve or a completed corrective action. If the reanalyzed CCV is low (<0%) indicating a low bias, results for samples exceeding a regulatory limit may be reported and Page 9 of 23

10 qualified using the statement: The CCV was below the lowest limit indicating a low bias. This result exceeds the regulatory limit and is considered a reportable value under the TNI standards. If the reanalyzed CCV is biased low (<0%) and the results of the samples do not exceed a regulatory limit then the samples must be reanalyzed after acceptance of a new calibration curve The CCV must be from a different source than that of the standard used in the initial calibration curve. 9.4 Instrument Performance - Check the performance of the entire analytical system daily using data gathered from analyses of reagent water blanks, standards, and the QC check standard Significant peak tailing must be corrected. Tailing problems are generally traceable to active sites on the GC column, improper column installation, or the operation of the detector Check the precision between replicate analyses. A properly operating system should perform with an average relative standard deviation of less than 10%. Poor precision is generally traceable to pneumatic leaks, especially at the injection port. Section 10.0 Quality Control 10.1 Minimum quality control (QC) requirements are initial demonstration of capability and detection limit studies and an ongoing analysis of laboratory reagent blanks (LRB), laboratory performance check solutions (LPC), laboratory fortified sample matrix (LFM), laboratory fortified blanks (LFB), and quality control samples (QCS) to evaluate and document data quality. Ongoing data quality checks are compared with established performance criteria to determine if the results of analyses meet the performance characteristics of the method Initial Demonstration of Capability Prepare four samples at a concentration of g/L for each analyte. Analyze each sample according to procedures beginning in Section Calculate the average concentration (X) in g/l, and the standard deviation of the concentration(s) in g/l, for each analyte. %RSD of the four samples must be <20%. Page 10 of 23

11 For each analyte the recovery value for all four of these samples must fall in the range of 30% of true value. For those compounds that meet the acceptable criteria, performance is considered acceptable and sample analysis may begin. For those compounds that fail these criteria, the instrument should be optimized and this procedure must be repeated using fresh samples until satisfactory performance has been demonstrated The initial demonstration of capability is used primarily to preclude a laboratory or analyst from analyzing unknown samples via a new, unfamiliar method prior to obtaining some experience with it Method Detection Limit For each analyte, determine the method detection limit (MDL). Prepare a minimum of seven LFB s at a low concentration. Extract and analyze each replicate according to Section The LFB s should be prepared over a period of several days so that day to day variations are reflected in precision measurements. The standard deviation of the average of seven results for each analyte is then multiplied by to obtain the calculated MDL. The reporting limit check standard when analyzed daily may be used to calculate the MDL Laboratory Reagent Blanks (LRB) - With each batch, the analyst must demonstrate that all glassware and reagent interferences are under control. The LRB consists of reagent water extracted as with the samples. Each time a set of samples is extracted or reagents are changed, a LRB must be analyzed. If a target analyte is detected but falls below the reporting limit, then samples may be reported. If an analyte of interest is detected in the LRB at or above the reporting limit, samples that have results below the reporting limit can be reported with qualification. Sample results that are 10 times higher than the LRB can be reported with the following qualifying statement: Analyte detected in the Method Blank is 1/10 of the amount detected in this sample. Evaluate data accordingly. If the sample result is not 10 times higher than the result in the method blank then the samples should be reanalyzed or if reported the results must be qualified with the following statement: Analyte detected in the Method Blank is 1/10 of the amount detected in this sample. Evaluate data accordingly. If within the retention time window of any analyte of interest the LRB produces a peak that would prevent the determination of that analyte, determine the source of contamination and eliminate the interference before processing samples. Page 11 of 23

12 10.5 Assessing Laboratory Performance - Laboratory Fortified Blank (LFB) The laboratory must analyze at least one laboratory fortified blank (LFB) sample with every 10 samples or one per sample set (all samples extracted within a 24 hour period) whichever is greater. Calculate the accuracy as percent recovery (X). The recovery (X) should be between 0% and 130% of the expected value. If the LFB meets limits the results can be reported. If the initial LFB fails the 0-130% recovery criteria then it can be analyzed a second time. \ If the reanalyzed LFB meets requirements then analysis may proceed. If the reanalyzed LFB is exceeded high (>130%) and the results of the samples are below the reporting level, then the results can be reported using the following qualification statement: The LFB was above the upper limit indicating a high bias. This non-detect result is not affected by the high bias of the LFB. If the reanalyzed LFB is exceeded high (>130%) and the results of the sample are detected at or above the reporting level then the samples need to be re-analyzed following acceptance of a new calibration curve or following corrective action. If the reanalyzed LFB is exceeded low (<0%), indicating a low bias, then the results for samples that exceed a regulatory limit may be reported using the following qualifying statement: The LFB was below the lower limit indicating a low bias. This sample result exceeds the regulatory limit and is considered reportable under the TNI standards. If the reanalyzed LFB is exceeded low, indicating a low bias then the results for samples that do not exceed a regulatory limit need to be reanalyzed following acceptance of a new calibration curve or following corrective action Assessing Analyte Recovery - Laboratory Fortified Sample Matrix (LFM) The laboratory must add a known concentration to a minimum of 10% of the routine samples or one sample concentration per set, whichever is greater. The concentration of g/l will be used for the LFM. Results of these analyses should have a recovery of 65%-135% of the expected value. If the recovery fall outside of the acceptable limits, report the sample result with the following qualification statement: The matrix spike fell outside method limits (biased high) or (biased low). The result for this sample should be evaluated accordingly. Page 12 of 23

13 10. Assessing Analyte Precision Laboratory Fortified Sample Matrix Duplicate (LFSMD) / Sample Duplicate (SD) If the sample is not expected to contain target analytes then a LFSMD is performed on at least 10% of the samples. If the sample is expected to contain target analytes then a sample duplicate is used If the LFSMD option is used then the corrected recovery must be the same as in Section The %RPD on a Sample duplicate and the LFSMD should be < 10%.If the relative percent difference (RPD) is results can be reported If the %RPD is not report only the result of the sample used for the duplicate with the following statement: The duplicate fell outside method limits and the result should be evaluated accordingly Reporting limit check (RLC) The RLC is extracted following same procedures as samples. One RLC is extracted every extraction batch to make sure that the equipment is capable of seeing analytes below the reporting limits. The check is analyzed using a µg/l concentration. The recovery of the RLC should be 60%-140% of the expected value. If recovery is outside of these limits, corrective action must be taken and the problem resolved prior to the analysis of any samples It is required that a laboratory periodically establish its ability to distinguish Dibromochloromethane (DBCM) from EDB once per year. To do this start with reagent water that is fortified at 50 g/l DBCM (Stock Standard can be obtained from VOC analyst), and during sample extraction fortify this sample at the Reporting Limit for EDB, and analyse. There should be no interferences between DBCM and EDB and the recovery should be % of the expected value. Section 11.0 Analytical Procedures 11.1 Micro-extraction and Analysis Remove samples and standards from storage and allow them to reach room temperature Remove container cap if in place. Remove approximately 6.2 ml from the 40 ml VOA vials and weigh, record as wet weight. Page 13 of 23

14 From the 6.2 mls removed from the 40 ml vial, check the ph and residual chlorine levels. Record both values on the extraction benchsheet. Chlorine level should be non-detect. (for method 8011, ph should be <2.0.) Add approximately 6 grams of baked sodium chloride Fortify Calibration standards with the appropriate amount of the 50-µg/L calibration stock standard solution (5, 10, 20,30,40,50, and 60-µL) Fortify CCVs, LFB, LFM, LFMD with 40-µL of the stock standard solution Add 2.0 ml of hexane with 2 ml pipetter. Recap and shake vigorously by hand for 1 minute. Allow the water and hexane layers to separate. (If stored at this stage, keep the container upside down) Remove cap and carefully transfer hexane into the autosampler vial using a disposable glass pipette Transfer the autosampler vials to the autosampler and inject 2.5 L into the gas chromatograph Determination of Sample Volume Weigh empty 40 ml VOA vial and record weight Subtract the weight of empty vial from wet weight to get sample weight/volume. The volume correction factor = 35/sample volume Setting up the Sequence On computer, pull down Sequence Scroll and choose New Sequence Pull down Sequence Scroll and choose Save Sequence As. Save the sequence with the date that the run is started Pull down Sequence Scroll and choose Sequence Parameters. Change the subdirectory to the date that is being used for the sequence One the computer, pull down Sequence Scroll and choose Sequence Table From the prep sheets, correctly type in each standard, sample and other information into the sequence table to meet quality control requirements. Page 14 of 23

15 Pull down the sequence scroll and choose Print Sequence. Click in the boxes in front of the sequence parameters, method, and injection port. Choose print and make two copies Load all appropriate vials into their correct positions on the autosampler tray using the printed sequence log Prepare instrument for analysis When all samples are loaded and the instrument is ready, choose Start Sequence NOTE: The vials must be recapped after being pierced if the vial will be re-injected for subsequent confirmation analysis Gas Chromatography Section 6.5 summarizes the recommended operating conditions for the GC Calibrate the system daily as described in Section Inject 2.5 L of the sample extract. Record the resulting peak size in area units If the response for the peak exceeds the working range of the system, dilute the extract and reanalyze Reviewing Data Load Method If a curve was run, choose the correct curve for the method. Work the curve according to Section 9 to make sure that it meets all criteria. If all criteria is met, save the new curve Examine all check standards and quality control samples before any samples to ensure that the data is reportable. Use manual integrations if necessary (printing a before and after copy and including the reason code, date, and initials) Load Data File Page 15 of 23

16 After Data File is loaded, all samples will be listed on the left-hand side of the box. Double click on each sample to pull it up on the screen. If a sample is clean, print only a copy from the column that was used for the check standards. If a target compound is detected, copies from both columns must be printed. Page 16 of 23

17 11.6 Identification of Analytes Identify a sample component by comparison of its retention time to the retention time of a reference chromatogram. If the retention time of an unknown corresponds, within limits, to the retention time of a standard compound on both columns, then identification is considered positive The width of the retention time window used to make identifications should be based upon measurements of actual retention time variations of standards over the course of a day. Three times the standard deviation can be used to calculate a suggested window size for a compound. However, the experience of the analyst should weigh heavily in the interpretation of chromatograms. Section 12.0 Calculations and Quantitative Analysis 12.1 Provided the GC data system has been properly calibrated, the results as obtained from the data system should be used to calculate the concentration of the sample. Use the following formula to calculate the concentration: C V F C DF 1 V I Where: Response Factor (RF) C 1 = Concentration from instrument (g/l) V F = Final Volume of Extract DF = Dilution factor C = Concentration in µg/l VI = Volume of sample in ml RF A s Cis A C is s Where: A s = Response for parameter in standard A is = Response for internal standard in standard C is = Concentration (µg/l) of internal standard C s = Concentration (µg/l) of parameter Page 1 of 23

18 Mean Response Factor (RF) RF RFi n Where: RF i = individual Response factor (i n) Soil Concentration (µg/kg) example C g kg A x Is Vt A RFV W D is i s Where: A x = Area of compound of interest RF = Response factor V i = Volume of extract added for purging D = Dilution factor W s = Weight of sample in grams V t = Volume of total extract A is = Area of internal standard I s = Internal standard conc Section 13.0 Data Assessment 13.1 The primary analyst bears the responsibility of producing accurate data. A documented review is performed that includes a check of the following parameters: Calibration verification QC criteria Calculation checks Data entry using the LIMS worksheet Review of all data to ensure proper peak integration The initial data review is documented on a checklist that is initiated and dated by the primary analyst. A secondary review must be performed by an individual trained to the standard operating procedure. The specific procedures for data review can be found in Alloway s data review and manual integration SOPs Data qualifiers are used whenever deviations occur while analyzing the samples. The qualifiers are included on the Certificate of Analysis that is presented to the client. Qualifiers include: samples that were not properly preserved, failed surrogate recovery, failed LFM recovery, qualified due to LRB contamination, estimated or elevated reporting limit due to sample matrix interference, calibration failure (high bias) but sample was <RL for the compound. Page 18 of 23

19 13.3 If acceptance limits are not met corrective actions are taken as described in Alloway s Corrective Action SOP. Section 14.0 Report Generation 14.1 A report is generated after each analysis run is complete. The report includes the US EPA Method number, sample identification, date acquired, acquisition method, date processed, processing method, analyst s name, chromatogram and quantitative results. If any manipulation is performed, the result is saved and another report is printed and added to the original. Section 15.0 Pollution Prevention and Waste Management 15.1 (For information about pollution prevention that may be applicable to laboratory practices, consult Less is Better: Laboratory Chemical Management for Waste Reduction available from the American Chemical Society s Department of Government Relations and Science Policy, th Street N.W., Washington, D.C ) 15.2 ( The Waste Management Manual for Laboratory Personnel, also available from the American Chemical Society at the address above.) 15.3 It is the laboratory s responsibility to comply with the requirements as stated in the Alloway Chemical Hygiene Plan for waste disposal. Section 16.0 References 16.1 EPA Method 504.1, Revision 1.1, Supplement III, 1995 Version, 1,2- Dibromoethane (EDB) and 1,2-Dibromo-3-Chloropropane (DBCP) In Water By Microextraction And Gas Chromatography, Environmental Monitoring Systems Laboratory, Office of Research and Development, USEPA, Cincinnati, Ohio 45268, plus subsequent updates SW-846 Method 8011 plus subsequent updates. Section 1.0 Revision History 1.1 Revision 3 issued to update calibration standard concentrations, add glassware preparation procedures, modify format, remove ATEL/Aqua Tech references and add revision history section. Page 19 of 23

20 1.2 Revision 4 issued to include Sections 6,, 9.2 (differentiated the %RSD criteria between (<20%) and 8011 (<10%), , Removed single point calibration option, reworked Section 11, and added Section 18 QC acceptance criteria, added 10.9 demonstrated the ability to distinguish DBCP from EDB. 1.3 Revision 5.0 issued to update section 9.0 Calibration and Standardization. Section CCV was moved to section 9.3. Added section 9.4 Instrument Performance. Modified and expanded Section 10.0 Quality Control. Added ph and residual chlorine checks to section Revision 6 issued to add section to recap vials after being pierced if additional injections are deemed necessary, to add last sentence to section Revision issued 3/9/15. Revisions include updates to and preservation check limits. Updated instrument conditions in sections and Section 10.8 rewritten to reflect section 9.6 of EPA Method. Updated format. Page 20 of 23

21 Section 18.0 Data Assessment and Acceptance Criteria for Quality Control Measures SOP Reference Requirement Specification and Frequency Acceptance Criteria Section Section Initial Calibration Continuing Calibration Verification (CCV) Use external standard calibration technique to generate an average RRF, first or second order calibration curve only. Verify initial calibration by analyzing a calibration verification standard (second source) at the beginning of an analytical run, after every 10 samples, and at the end of the analytical run. When average response is used, the %RSD must be less than 20% for Method and < 10% for When first or second order regression is used, the correlation coefficient must be greater than A minimum of 5 standards are required for a linear regression. A minimum of 6 standards is required for second order regression. If criteria are not met corrective action is required such as injection port maintenance, new column, and/or recalibration. If Average response is used for the calibration then the response of the analyte should be ± 30% of the average response in the curve. If first or second order regression is used, then the recovery should be ±30% of the true value. 1. If the recovery is below criteria, then the results for the analysis must be considered invalid. 2. If the recovery is above criteria, and there are NO detections in the sample then the results can be reported with a qualifier, otherwise the results should be considered invalid. Page 21 of 23

22 Section 10.8 Reporting Limit Standard (RLC). Daily, prior to the analysis of field samples. The recoveries of the analytes should meet the criteria %, if the recovery is high and no target analytes are detected in the sample then the results can be reported. If the recovery is low then the system does not meet the required sensitivity and analysis cannot proceed until corrective action is taken to restore system sensitivity. Section 10.4 Section 10.5 Laboratory Reagent Blank / Method Blank (LRB) Laboratory Fortified Blank (LFB) Daily, or with each extraction batch of up to 10 samples (1 in 10), whichever is more frequent. Daily, or with each extraction batch of 10 samples or at a rate of 1 in 10 samples Demonstrate that all target analytes are below the PQL, and confirm that possible interferences do not prevent quantification of method Analytes. If target analytes are detected in the method blank then extraction must be considered invalid if the analytes are detected in the samples. Results of LFB analyses at medium and high fortifications should be 0-130% of the true value for each analyte. 1. If the recovery is below method criteria. Then the results for that extraction batch should be considered invalid. 2. If the recovery is above criteria, and there are NO detections in the sample then the results can be reported with a qualifier, otherwise the results should be considered invalid. Page 22 of 23

23 Section 10.6 Section 10. Section Section Laboratory Fortified Sample Matrix (LFSM) Laboratory Fortified Sample Matrix Duplicate (LFSMD) or Field Duplicates (FD) Sample Holding Time Extract Holding Time Analyze one LFSM per analysis batch (10 samples or less) fortified with method analytes at a concentration close to but greater than the native concentration. Calculate LFSM recoveries. Extract and analyze at least one FD or LFSMD with each extraction batch (10% samples or less). A LFSMD may be substituted for a FD when the frequency of detects are low. Calculate RSDs. 14 days with appropriate preservation and storage 24 Hours with appropriate preservation and storage Recoveries that are not within % of the fortified amount may indicate a matrix effect and must be qualified. Target analyte RSDs for the LFMD or FD should be <10%. If the criteria are greater than 10% then the greater of the two results are reported and may indicate matrix effect and must be qualified. Sample results are valid only if samples are extracted within sample hold time. Sample results are valid only if extracts are analyzed within extract hold time. Page 23 of 23

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS PAGE: 1 of 18 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

METHOD 8000A GAS CHROMATOGRAPHY

METHOD 8000A GAS CHROMATOGRAPHY METHOD 8000A GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Gas chromatography is a quantitative technique useful for the analysis of organic compounds capable of being volatilized without being decomposed

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Table of Contents Section 1: Summary of Method Section 2: Scope and Application Section

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82 ALLOWAY Method Outline Standard Laboratory Method: SM4500-F - B and SM4500-F-C Parameter: Total Fluoride via Distillation Method: Distillation then Specific Ion Date Issued: 01/04/82 Date Revised: 04/09/02,

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

Quantitative determination of cannabinoids in Hemp and Hemp products by High Performance Liquid Chromatography using a UV Detector

Quantitative determination of cannabinoids in Hemp and Hemp products by High Performance Liquid Chromatography using a UV Detector SOP # 04/30/6 04/30/6 of 4 Mike Klasner Mike Klasner 04/30/6 Quantitative determination of cannabinoids in Hemp and Hemp products by High Performance Liquid Chromatography using a UV Detector.0 PURPOSE.

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY 1. SCOPE AND APPLICATION: METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) Date 02/24/96 Page 1 1.1. This method is designed to measure the concentration

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

Method 548. July 1990 EPA EMSL-Ci. J.W. Hodgeson

Method 548. July 1990 EPA EMSL-Ci. J.W. Hodgeson Determination of Endothall in Drinking Water by Aqueous Derivatization, Liquid-Solid Extraction, and Gas Chromatography with Electron-Capture Detection July 1990 EPA EMSL-Ci J.W. Hodgeson Determination

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2. Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES 701.0 Version 2.3 July 2018 1. SCOPE AND APPLICATION This method is used for the determination

More information

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method Rating Guide Appendix for the CCME Reference Method for the Canada-Wide Standard for Petroleum (PHC) in Soil - Tier 1 Method (Note: Checklist incorporates requirements from Dec 2000 version + Addendum

More information

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. 1 REVISOR 4740.2100 4740.2100 QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. Subpart 1. Scope. This part applies to laboratories performing testing under the inorganic chemistry, metals,

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

METHOD 547 DETERMINATION OF GLYPHOSATE IN DRINKING WATER BY DIRECT-AQUEOUS- INJECTION HPLC, POST-COLUMN DERIVATIZATION, AND FLUORESCENCE DETECTION

METHOD 547 DETERMINATION OF GLYPHOSATE IN DRINKING WATER BY DIRECT-AQUEOUS- INJECTION HPLC, POST-COLUMN DERIVATIZATION, AND FLUORESCENCE DETECTION METHOD 547 DETERMINATION OF GLYPHOSATE IN DRINKING WATER BY DIRECT-AQUEOUS- INJECTION HPLC, POST-COLUMN DERIVATIZATION, AND FLUORESCENCE DETECTION July 1990 T.W. Winfield W.J. Bashe (Technology Applications

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State

More information

DETERMINATION OF PHENOLS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) Revision 1.

DETERMINATION OF PHENOLS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) Revision 1. METHOD 528 DETERMINATION OF PHENOLS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) Revision 1.0 J.W. Munch - April 2000 NATIONAL EXPOSURE

More information

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay... Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials (HPLC grade)... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation...

More information

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University. Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared

More information

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7 Troy, NY 12180 Date 01/25/18 Page 1 of 7 USGS Water Science Center Laboratory, Troy, NY Ion Chromatography Standard Operating Procedure 1. Scope and Application 1.1 Analytes Chloride, Nitrate, and Sulfate

More information

Date 10/22/97 Page 1 Revision 4.1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO)

Date 10/22/97 Page 1 Revision 4.1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO) Page 1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY 1. SCOPE AND APPLICATION: METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO) 1.1. This method is designed to measure the concentration of diesel

More information

LOW-RANGE TOTAL CHLORINE

LOW-RANGE TOTAL CHLORINE LOW-RANGE TOTAL CHLORINE Thermo Orion Method AC2072 Revision 4 3/14/2002 Accepted by U.S. EPA on March 20, 2002 for wastewater compliance monitoring Accepted by U.S. EPA on September 31, 2002 for drinking

More information

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS 1.0 SCOPE AND APPLICATION 1.1 Method 8100 is used to determine the concentration of certain polynuclear aromatic hydrocarbons (PAH). Table 1 indicates compounds

More information

Version 1.0 October 2007

Version 1.0 October 2007 EPA Document #: 815-B-07-002 METHOD 536 DETERMINATION OF TRIAZINE PESTICIDES AND THEIR DEGRADATES IN DRINKING WATER BY LIQUID CHROMATOGRAPHY ELECTROSPRAY IONIZATION TANDEM MASS SPECTROMETRY (LC/ESI-MS/MS)

More information

SECTION D.2 AMMONIA NITROGEN

SECTION D.2 AMMONIA NITROGEN SECTION D.2 AMMONIA NITROGEN CEDR Method Code: NH4F L01 a) Scope and Application i) This method describes the determination of low-level ammonia nitrogen concentrations in filtered samples taken from fresh

More information

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Timberline Ammonia-001 Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Prepared By: Dr. Edward F. Askew Date: June 24, 2011 Table of Contents 1. SCOPE

More information

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division METHOD 375.2 DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING SYSTEMS

More information

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP. NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta

More information

Introduction. The following definitions may help you better understand the components of the data report.

Introduction. The following definitions may help you better understand the components of the data report. Introduction This document is designed to help our clients understand the quality control requirements and limitations of data reporting. There are three sections to this document. The first section will

More information

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS 1.0 SCOPE AND APPLICATION 1.1 Method 5000 provides general guidance on the selection of sample preparation methods (purge-and-trap, extraction,

More information

METHOD DETERMINATION OF INORGANIC ANIONS IN DRINKING WATER BY ION CHROMATOGRAPHY. Revision 1.0

METHOD DETERMINATION OF INORGANIC ANIONS IN DRINKING WATER BY ION CHROMATOGRAPHY. Revision 1.0 METHOD 300.1 DETERMINATION OF INORGANIC ANIONS IN DRINKING WATER BY ION CHROMATOGRAPHY Revision 1.0 John D. Pfaff (USEPA, ORD, NERL) - Method 300.0, (1993) Daniel P. Hautman (USEPA, Office of Water) and

More information

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices Environmental Express 2345A Charleston Regional Parkway Charleston, SC 29492 800-343-5319 Table of Contents 1. Scope and Application...

More information

METHOD DETERMINATION OF CARBONYL COMPOUNDS IN DRINKING WATER BY FAST GAS CHROMATOGRAPHY

METHOD DETERMINATION OF CARBONYL COMPOUNDS IN DRINKING WATER BY FAST GAS CHROMATOGRAPHY METHOD 556.1 DETERMINATION OF CARBONYL COMPOUNDS IN DRINKING WATER BY FAST GAS CHROMATOGRAPHY Revision 1.0 September 1999 J.W. Munch, US EPA Office of Research and Development, NERL D.J. Munch, US EPA,

More information

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager

More information

Revision 1.0. April 2000

Revision 1.0. April 2000 METHOD 515.4 DETERMINATION OF CHLORINATED ACIDS IN DRINKING WATER BY LIQUID-LIQUID MICROEXTRACTION, DERIVATIZATION, AND FAST GAS CHROMATOGRAPHY WITH ELECTRON CAPTURE DETECTION Revision 1.0 April 2000 S.C.

More information

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin OIVMABS15 Anethole. Gas chromatograhpy determination of transanethole in Type II method SCOPE This method is suitable for the determination of transanethole in aniseedflavoured spirit drinks using capillary

More information

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94. Ethanol Project Chemical Analysis and Testing Task Laboratory Analytical Procedure Procedure Title: LAP-013 HPLC Analysis of Liquid Fractions of Process Samples for Monomeric Sugars and Cellobiose Author:

More information

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7 Troy, NY 12180 Date: 03/14/2012 Page 1 of 7 USGS New York Water Science Center Soil and Low Ionic Strength Water Quality Laboratory Electrometric Measurements and Titration Standard Operating Procedure

More information

APPENDIX A TO PART 136 METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND INDUSTRIAL WASTEWATER METHOD 609 NITROAROMATICS AND ISOPHORONE

APPENDIX A TO PART 136 METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND INDUSTRIAL WASTEWATER METHOD 609 NITROAROMATICS AND ISOPHORONE APPENDIX A TO PART 136 METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND INDUSTRIAL WASTEWATER METHOD 609 NITROAROMATICS AND ISOPHORONE 1. Scope and Application 1.1 This method covers the determination

More information

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY METHOD 335.4 DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 1.0 August 1993 ENVIRONMENTAL MONITORING

More information

Volatile Hydrocarbons (VH) in Waters by GC/FID

Volatile Hydrocarbons (VH) in Waters by GC/FID PBM/Hydrocarbon Revision Date: March 9, 2017 Volatile Hydrocarbons (VH) in Waters by GC/FID Parameters Volatile Hydrocarbons (nc6-nc10) in waters Analyte Symbols and EMS Codes Analyte Symbol Approx MDL

More information

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2 Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY METHOD 350.1 DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

STANDARD OPERATING PROCEDURES SOP: 1824 PAGE: 1 of 22 REV: 0.0 DATE: 04/21/95 ANALYSIS OF METHYL PARATHION IN SOIL SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1824 PAGE: 1 of 22 REV: 0.0 DATE: 04/21/95 ANALYSIS OF METHYL PARATHION IN SOIL SAMPLES BY GC/MS PAGE: 1 of 22 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Sample Storage 3.2 Holding Times 4.0 INTERFERENCES AND POTENTIAL PROBLEMS

More information

Uncontrolled Copy. SOP Ethylene Glycol Quantitation by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Uncontrolled Copy. SOP Ethylene Glycol Quantitation by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay... Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials (HPLC grade)... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation

More information

Analysis of Total PCBs and PCB Congeners and Trans-nonachlor in Fish by Gas Chromatography/ Negative Chemical Ionization Single Ion Mass Spectrometry

Analysis of Total PCBs and PCB Congeners and Trans-nonachlor in Fish by Gas Chromatography/ Negative Chemical Ionization Single Ion Mass Spectrometry Analysis of Total PCBs and PCB Congeners and Trans-nonachlor in Fish by Gas Chromatography/ Negative Chemical Ionization Single Ion Mass Spectrometry Standard Operating Procedure SOP No. HC 519.D (Replaces:

More information

Chloride by Flow Injection Analysis (FIA)

Chloride by Flow Injection Analysis (FIA) Methodology Chloride by Flow Injection Analysis (FIA) (Cartridge Part #A001568) 1.0 Scope and Application 1.1 This method is used for the determination of chloride in drinking water, surface water, saline

More information

Hexavalent Chromium by Flow Injection Analysis (FIA)

Hexavalent Chromium by Flow Injection Analysis (FIA) Methodology Hexavalent Chromium by Flow Injection Analysis (FIA) (Cartridge Part #A002831) 1.0 Scope and Application 1.1 This method is used for the determination of hexavalent chromium in water, including

More information

TO-15 Checklist Determination of VOCs in Air by GC-MS

TO-15 Checklist Determination of VOCs in Air by GC-MS LAB ID: DATE: LAB NAME: ASSESSOR NAME: Method Number: TO-15 Checklist Determination of VOCs in Air by GC-MS SOP Number: Revision Number: SOP Date: Personnel records observed: Data records observed: Revision

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.

More information

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Sample Throughput In Purge And Trap Analysis Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in

More information

Revision 1.0 May 2005

Revision 1.0 May 2005 METHOD 314.1 DETERMINATION OF PERCHLORATE IN DRINKING WATER USING INLINE COLUMN CONCENTRATION/MATRIX ELIMINATION ION CHROMATOGRAPHY WITH SUPPRESSED CONDUCTIVITY DETECTION Revision 1.0 May 2005 Herbert

More information

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Alkalinity, USEPA by Flow Injection Analysis (FIA) Methodology Alkalinity, USEPA by Flow Injection Analysis (FIA) (Cartridge Part #A002745) 1.0 Scope and Application 1.1 This method is used for the determination of alkalinity in drinking water, surface

More information

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,: sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied

More information

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations OUTLINE OF METHOD This gas chromatography/mass spectrometry (GC/MS) method provides

More information

METHOD 3600C CLEANUP

METHOD 3600C CLEANUP METHOD 3600C CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible

More information

METHOD 551. July J.W. Hodgeson A.L. Cohen

METHOD 551. July J.W. Hodgeson A.L. Cohen METHOD 551 DETERMINATION OF CHLORINATION DISINFECTION BYPRODUCTS AND CHLORINATED SOLVENTS IN DRINKING WATER BY LIQUID-LIQUID EXTRACTION AND GAS CHROMATOGRAPHY WITH ELECTRON-CAPTURE DETECTION July 1990

More information

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800)

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800) 2773 Downhill Drive Steamboat Springs, CO 80487 (800) 334-5493 Analytical Report November 19, 2004 Report to: Debbie Furlow 602 Lehmberg Road Columbus, MS 39702 Bill to: Accounts Payable PO Box 2487 Columbus,

More information

Shaker Table Extraction

Shaker Table Extraction Title: Shaker Table Extraction Page 1 of 11 Shaker Table Extraction References: This standard operating procedure (SOP) is a performance-based method. This SOP describes the procedure as developed by Alpha

More information

Percent Solids Determination

Percent Solids Determination Title: Percent Solids Page 1 of 10 Percent Solids Determination References: Method 2540G, Standard Methods For the Examination of Water and Wastewater, APHA 18 th edition, 1992. 1. Scope and Application

More information

United States EPA Method 415.3

United States EPA Method 415.3 United States EPA Method 415.3 Application Note Abstract In order to ensure drinking water is safe for human consumption, water treatment plants often add disinfectants to drinking water. The disinfectants,

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

Total Petroleum and Saturated Hydrocarbons by Gas Chromatography/Flame Ionization Detector

Total Petroleum and Saturated Hydrocarbons by Gas Chromatography/Flame Ionization Detector Title: TPH-SHC Page 1 of 18 Total Petroleum and Saturated Hydrocarbons by Gas Chromatography/Flame Ionization Detector References: USEPA, Method 8015D Nonhalogenated Organics Using GC/FID, in Test Methods

More information

Catecholamines Urine LC-MS/MS Analysis Kit User Manual

Catecholamines Urine LC-MS/MS Analysis Kit User Manual Page 1 / 13 Catecholamines Urine LC-MS/MS Analysis Kit User Manual ZV-3031-0200-20 200 Page 2 / 13 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3 4. WARNING

More information

Seawater Extraction Procedure for Trace Metals

Seawater Extraction Procedure for Trace Metals Title: Seawater Extraction Page 1 of 10 Seawater Extraction Procedure for Trace Metals References: Danielson, L., B. Magnusson, and S. Westerlund. 1978. An improved metal extraction procedure for the determination

More information

PCB Congener Analysis of XAD-2 Resins and GFF Filters Using GC/ECD

PCB Congener Analysis of XAD-2 Resins and GFF Filters Using GC/ECD ` PCB Congener Analysis of XAD-2 Resins and GFF Filters Using GC/ECD Eric Crecelius and Lisa Lefkovitz Pacific Northwest National Laboratory Battelle Marine Sciences Laboratory 1529 West Sequim Bay Road

More information

Appendix A. Glossary of Definitions, Acronyms, Abbreviations and Symbols. March 1997

Appendix A. Glossary of Definitions, Acronyms, Abbreviations and Symbols. March 1997 Appendix A Glossary of Definitions, Acronyms, Abbreviations and Symbols March 1997 Glossary Of Definitions, Acronyms, Abbreviations And Symbols 1.0 Standard Acronyms, Abbreviations, And Symbols 1.1 Units

More information

METHOD 509 DETERMINATION OF ETHYLENE THIOUREA (ETU) IN WATER USING GAS CHROMATOGRAPHY WITH A NITROGEN-PHOSPHORUS DETECTOR. Revision 1.0.

METHOD 509 DETERMINATION OF ETHYLENE THIOUREA (ETU) IN WATER USING GAS CHROMATOGRAPHY WITH A NITROGEN-PHOSPHORUS DETECTOR. Revision 1.0. METHOD 509 DETERMINATION OF ETHYLENE THIOUREA (ETU) IN WATER USING GAS CHROMATOGRAPHY WITH A NITROGEN-PHOSPHORUS DETECTOR Revision 1.0 December 1992 D.J. Munch and R.L. Graves T.M. Engel and S.T. Champagne

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 19 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Sample Storage 3.2 Holding Times 4.0 INTERFERENCES AND POTENTIAL PROBLEMS

More information

California Environmental Protection Agency Air Resources Board

California Environmental Protection Agency Air Resources Board California Environmental Protection Agency Air Resources Board PROCEDURE FOR THE ANALYSIS OF PARTICULATE ANIONS AND CATIONS IN MOTOR VEHICLE EXHAUST BY ION CHROMATOGRAPHY Standard Operating Procedure No.

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 55 ONTENTS 1.0 OBJETIVE 2.0 APPLIABILITY 3.0 DESRIPTION 3.1 Sample ontainers, Preservative, Storage and Holding Times 3.1.1 Objective 3.1.2 Requirements 3.1.3 Evaluation Procedure 3.1.4 Action

More information

METHOD 3600B CLEANUP

METHOD 3600B CLEANUP METHOD 3600B CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

METHOD July J.W. Hodgeson. W.J. Bashe (Technology Applications Inc.) T.V. Baker (Technology Applications Inc.)

METHOD July J.W. Hodgeson. W.J. Bashe (Technology Applications Inc.) T.V. Baker (Technology Applications Inc.) METHOD 550.1 DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS IN DRINKING WATER BY LIQUID-SOLID EXTRACTION AND HPLC WITH COUPLED ULTRAVIOLET AND FLUORESCENCE DETECTION July 1990 J.W. Hodgeson W.J. Bashe

More information

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of phenols in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicable to the quantitative determination of

More information

Revision 1.0. September, J.W. Munch

Revision 1.0. September, J.W. Munch EPA Document #: EPA/600/R-05/052 METHOD 529 DETERMINATION OF EXPLOSIVES AND RELATED COMPOUNDS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS)

More information

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis Technical Procedure for Solid Phase Extraction of THC and THC-COOH 1.0 Purpose - This procedure specifies the required elements for the extraction of THC and THC-COOH using United Technologies Styre Screen

More information

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Dehydroepiandrosterone 3-sulfate, Primary Standard D-065 Page 1 of 6 Catalog Number: D-065 Lot: H 3 C Expiration: Description: February 2019 Dehydroepiandrosterone

More information

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) Methodology Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) (Cartridge Part #A002603) 1.0 Scope and Application 1.1 This method is used to determine the concentration of nitrate

More information

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke British American Tobacco Group Research & Development Method - Determination of ammonia in mainstream smoke 1 SCOPE OF APPLICATION The method is applicable to quantitative determination of the yields of

More information

DETERMINATION OF PHENYLUREA COMPOUNDS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND HIGH PERFORMANCE LIQUID CHROMATOGRAPHY WITH UV DETECTION

DETERMINATION OF PHENYLUREA COMPOUNDS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND HIGH PERFORMANCE LIQUID CHROMATOGRAPHY WITH UV DETECTION METHOD 532. DETERMINATION OF PHENYLUREA COMPOUNDS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND HIGH PERFORMANCE LIQUID CHROMATOGRAPHY WITH UV DETECTION Revision 1.0 June 2000 M. V. Bassett, S.C. Wendelken,

More information

Application. Gas Chromatography February Introduction

Application. Gas Chromatography February Introduction Ambient Headspace Analysis with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography February 1998 Authors Matthew S. Klee and Chin Kai Meng Agilent Technologies, Inc. 2850 Centerville

More information

Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments

Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments Title: AVS-SEM Page 1 of 8 Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments Reference: Determination of acid-volatile sulfide and simultaneously-extracted metals in sediments using

More information

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 525.2

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 525.2 ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD DETERMINATION OF ORGANIC COMPOUNDS IN DRINKING WATER BY LIQUID-SOLID EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY Marion, OHIO Originator:

More information

USEPA CONTRACT LABORATORY PROGRAM NATIONAL FUNCTIONAL GUIDELINES FOR ORGANIC DATA REVIEW

USEPA CONTRACT LABORATORY PROGRAM NATIONAL FUNCTIONAL GUIDELINES FOR ORGANIC DATA REVIEW OSWER 9240.1-05A-P PB99-963506 EPA540/R-99/008 October 1999 USEPA CONTRACT LABORATORY PROGRAM NATIONAL FUNCTIONAL GUIDELINES FOR ORGANIC DATA REVIEW Office of Emergency and Remedial Response U.S. Environmental

More information