Trace Element Analysis: Time and Cost Savings with TXRF. Bruker AXS Microanalysis GmbH Berlin, Germany
|
|
- Rudolph Hood
- 6 years ago
- Views:
Transcription
1 Trace Element Analysis: Time and Cost Savings with TXRF Bruker AXS Microanalysis GmbH Berlin, Germany
2 Welcome Today s Topics Introduction to XRF and TXRF The S2 PICOFOX benchtop TXRF system Application studies - Liquid samples - urine - Suspensions - blood - Solid samples - NaCl (pharma) - Micro particles - nano particles Method comparison & summary Interactive Q & A Your Host Arkady Buman Business Development Mgr Madison, WI, USA Speakers Dr. Hagen Stosnach Applications Scientist TXRF Berlin, Germany Dr. Armin Gross Global Product Manager TXRF Berlin, Germany 2
3 Introduction to XRF and TXRF Armin Gross
4 Principles of X-ray fluorescence (XRF) spectroscopy An X-ray quantum hits an inner shell electron in a (sample) atom. The electron is removed leaving the atom in an excited state 2 2. The missing inner shell electron is replaced by an electron from an outer shell 3. The energy difference between the inner and outer shell is balanced by the emission of a photon (fluorescence radiation) 4
5 Principles of X-ray fluorescence spectroscopy The energy, and therefore the wavelength, of the X-ray fluorescence radiation is characteristic for the different chemical elements. QUALITATIVE ANALYSIS The intensity of the X-ray fluorescence radiation is, in first approximation, proportional to the element concentration. QUANTITATIVE ANALYSIS Low Z High Z 5
6 Principles of X-ray fluorescence spectroscopy Common XRF optics X-ray tube Sample Detector Beam angle: 45 o / 45 o 6
7 Principles of X-ray fluorescence spectroscopy Samples for common XRF spectrometry (ED and WD XRF): Solids (cut, polished and made into suitable shape) Powders (as pressed pellets, fused beads or loose powders in liquid cups) Liquids (in liquid cups) Necessary sample amount: from 1 g to 10 g!!! 7
8 Principles of total reflection X-ray fluorescence (TXRF) spectroscopy Total reflection X-ray fluorescence spectroscopy Detector X-ray tube Sample carrier Beam angle: 0 o / 90 o 8
9 Principles of total reflection X-ray fluorescence spectroscopy Samples for total reflection X-ray fluorescence spectroscopy: Powders: Liquids: Direct preparation or as suspension Direct preparation always as a thin film, micro fragment or suspension of a powder necessary sample amount: Low µg respectively µl range 9
10 Principles of total reflection X-ray fluorescence spectroscopy Quantification in total reflection X-ray fluorescence spectroscopy In TXRF the samples are prepared as thin films or layers Therefore matrix effects are negligible QUANTIFICATION IS POSSIBLE THROUGH: Sensitivity of the instrument for element lines (energy-dependent) Net intensity of element lines Known concentration of an internal standard element Negligible absorption of primary beam and fluorescence radiation 10
11 The instrument S2 PICOFOX Benchtop TXRF spectrometer S2 PICOFOX Metal-ceramic X-ray tube - Mo anode - air-cooled - optionally other tubes available Multilayer monochromator XFlash Silicon Drift Detector - electro-thermally cooled kcps Automatic version - 25 sample cassette 11
12 S2 PICOFOX Element sensitivity The S2 PICOFOX detects elements from Na(11) to U(92) The element sensitivities depend on the atomic number The sensitivity factors are calibrated ex works Element sensitivity K-lines L-lines Atomic number 12
13 Audience poll Please use your mouse to answer the question on your screen: How do you currently treat your samples before analysis? Check all that apply: No treatment Grinding Dissolving Dilution Extraction Digestion 13
14 Why Wait for Sample Prep! Ease of Use and Versatility of TXRF Analysis Hagen Stosnach
15 S2 PICOFOX Application overview Focus on medical and pharmaceutical applications Sample type Liquids Suspensions Solids, powders Particles Application urine analysis whole blood and blood serum testing raw materials in pharmaceutical production, e.g. NaCl nano particles 15
16 Sample preparation Liquid samples You ll need just a few steps for the preparation of liquid samples fill sample in micro tube add internal standard homogenize pipette on carrier 16
17 Sample preparation Final steps dry by heat / vacuum load the instrument start data aquisition 17
18 Liquid samples Urine Introduction The trace element content of urine is an indicator for the human health status Task Participation in round robin test for urine samples Sample preparation and measurement Direct application after addition of a Ga standard Measurement time 1000 s After treatment with HNO 3 directly on the carrier, samples were measured again 18
19 Liquid samples Urine Challenges of urine samples testing: Detection limits in the low ppb range required High amount of chlorine and calcium disturb TXRF measurements with sum peaks of matrix elements Solution Vaporize Cl with HNO 3 Perform second measurement of same sample TXRF spectrum of urine sample grey = before, blue = after HNO 3 treatment 19
20 Liquid samples Urine Results TXRF received certification for the elements As, Se, Zn Conclusion TXRF allows precise and accurate urine analysis after direct sample preparation Element TXRF (µg/l) Ref. (µg/l) LLD (µg/l) Sample A As Se Zn Sample B As Se Zn
21 Sample preparation Suspensions Suspensions can be analyzed just after dilution dilute sample with distilled water add internal standard homogenize pipette on carrier 21
22 Suspensions Blood analysis Introduction Analysis of nutrition-relevant elements (Cu, Fe, Zn, Se) Analysis of other e.g toxic elements and Pt (chemotherapy) Task Analysis of whole blood and blood serum standards Sample preparation Serum - 1:10 dilution with water (p.a. grade) - Addition of Ga for internal standardization Whole blood - 1:1 dilution with water (p.a. grade) - Addition of Ga for internal standardization 22
23 Suspensions Blood analysis Results I Serum reference standard (t = 600 s) good correlation for Cu, Zn & Se overestimation of Fe improved calibration curve required Note Confidence level of reference values is up to 35 % (!) Reference values (µg/l) TXRF values (µg/l) Se Cu Zn Fe 23
24 Suspensions Blood analysis Results II Whole blood reference standard (t = 600 s) application of optimized blood calibration excellent correlation for Fe, Cu, Zn, Se and others Reference values (µg/l) P S K Ca Rb Sr Pb Se Cu Zn Fe TXRF values (µg/l) 24
25 Suspensions Blood analysis Reproducibility Precise TXRF results although sample digestion was avoided Serum (ClinCheck L2) Whole blood (Seronorm L2) Element Unit TXRF Std. dev. Reference 1) Std. dev. TXRF Std. dev. Reference 2) Std. dev. Fe mg/l 440 7, ,9 0,09 1,964 0,20 Cu µg/l 66 2,2 62 2, Zn µg/l 501 4, , Se µg/l 12 0, , ) : Sector-Field Inductively-Coupled Plasma Mass Spectroscopy 2) : Atomic Absorption Spectroscopy 25
26 Sample preparation Solid and powder samples I Solids are ground to fine particle size and resuspended for direct analysis without digestion fill powder in mortar grind carefully weigh about mg transfer to tube 26
27 Sample preparation Solid and powder samples II suspend in detergent solution add standard homogenize pipette on carrier 27
28 Solid samples Pharmaceuticals purity control Example: Raw material NaCl Analytical task analysis of impurities with regard to the limit value of 1 mg/kg for As in pure NaCl Samples NaCl (p.a. grade), Carl Roth GmbH - As: < 0.4 mg/kg same sample, spiked with As Preparation Dissolution of some mg in 1 ml water Resuspension of about 50 mg in 1.5 ml detergent (Triton X-100) 28
29 Solid samples Pharmaceuticals purity control Recovery results Trace element concentrations of Ca, Ti, Cr, Fe, Ni, Cu, Zn and Br were determined simultaneously Measured values As (mg/kg) 5,0 4,0 3,0 2,0 1,0 y = 1,1157x + 0,0986 R 2 = 0,9972 0,0 0,0 1,0 2,0 3,0 4,0 5,0 Spiked As (mg/kg) 29
30 Solid samples Pharmaceuticals purity control Detection limits (600 s) Conc. LLD (mg/kg) Ca 2,4 0,57 Ti 1,8 0,25 Cr 0,85 0,17 Fe 11 0,13 Ni 0,29 0,09 Cu 0,34 0,08 Zn 0,33 0,07 As 1,2 0,07 Se (i.s.) 20 0,07 Br 19 0,08 30
31 Sample preparation Microparticles Microparticles are measured semi-quantitatively and non-destructively dab vacuum grease on carrier pick-up some particles with a (glass) rod drop particles on grease 31
32 Particles Characterization of nanoparticles Analytical question element ratios in CdSe nanoparticles coated with ZnS Analytical issues extremely small sample amount (R&D) non-destructive method preferred TXRF measurement transfer of nanoparticles to quartz carrier by cotton bud standardless quantification Results even smallest sample amounts allow the determination of element ratios in nanoparticles 32
33 Particles Characterization of nanoparticles Results Element ratios of nanoparticles even smallest sample amounts allow the determination of element ratios in nanoparticles S2 PICOFOX Standardless analysis applied Ratio (wt.-%) Sample 1 Sample 2 Sample 3 0 Zn/S Cd/Se Zn/Cd Measured ratios of 3 samples versus target value ( ) 33
34 Summary and Method Comparison Armin Gross
35 Summary biological matrices Wait Wait for Sample Prep! TXRF can be easily applied to all biological matrices Sample digestion is hardly required Biological matrix Typ. volume Blood - whole blood 500 µl Blood - serum, small volumes < 10 µl Urine ml Sample preparation for TXRF 1 : 1 dilution with H 2 O, addition of internal Ga standard 1 :2 dilution with H 2 O, pipetting on carrier addition of 1 µl Ga standard solution direct addition of internal standard, fume off Chlorine by HNO 3 Tissue homogenates µl 1 : 1 dilution with Y standard solution Seminal fluid µl direct addition of internal standard Cerebrospinal fluid µl direct addition of internal standard Mother s milk ml direct addition of internal standard Tear fluid µl to ml direct addition of internal standard 35
36 Summary biological matrices Low levels of detection Detection limits are in the low to middle ppb range 1000 Detection Limits in Biological Matrices urine whole blood blood serum 100 (µg/l) 10 1 Ca V Cr Mn Fe Co Ni Cu Zn As Se Br Rb Sr Hg Pb 36
37 Summary pharmaceuticals TXRF meets Pharmacopeia Metals with significant safety concern (EMEA): TXRF with Mo excitation fulfills required LLDs Method development for W excitation not yet finished Element Concentration (ppm) Excitation mode 3s LLD (ppm) Pt 1 Mo 0,06 Pd 1 W Ir 1* Mo 0,06 Rh 1* W Ru 1* W Os 1* (volatile, not analyzable) Mo 2,5 W Ni 2,5 Mo 0,08 Cr 2,5 Mo 0,19 V 2,5 Mo 0,24 *) subclass limit for total amount of Ir, Rh, Ru, Os 37
38 TXRF versus AAS & ICP Time-to-result Solid samples and suspensions Shorter time needed from sampling to the final quantitative result TXRF ICP AAS 38
39 TXRF versus AAS & ICP Productivity Factory calibrations one time Immediate results for unknown matrices standardless Shorter learning curve lab standardization ICP calibration about 30% of daily worktime AAS calibration about 15% of daily worktime TXRF automatic gain calibration about 2 % of daily worktime 39
40 TXRF versus AAS & ICP Cost of ownership Use of expensive accessories (AAS: graphite tubes) Need for consumables (noble gases, standards) Complicated sample preparation equipment and hazardous chemicals Expensive maintenance contracts TXRF ICP-MS ICP-OES AAS 40
41 Comparison TXRF versus AAS & ICP-OES The S2 PICOFOX can be an alternative to AAS and/or a complement for existing ICP-OES systems Major benefits of the S2 PICOFOX fast and simple sample preparation flexibility with regard to sample types easy multi-element analysis without external calibration low maintenance and operating costs 41
42 Any Questions? Please type in questions you may have for any of our speakers in the Question form and then click Submit.
43 Thank you for attending! Please take a moment to complete the brief survey on your screen. Your feedback is very important to us. Copies of this presentation and related TXRF resource materials will be ed to you.
44 Meet us at: IFPAC/PAT: International Forum for Process Analytical Chemistry & Technology Jan 26-28, Baltimore, MD SME & CME: Society for Mining, Metallurgy and Exploration & National Western Mining Conference Feb 23-25, Denver, CO PITTCON: Mar 8-13, Chicago, IL Frontiers in XRF Seminar Northwest: Your free 1-day seminar on the theory, instrumentation, preparation and application of ED and WDXRF Apr 25, Seattle, WA ISWM: International Symposium on Metallomics Jun 7-10, Cincinnati, OH Denver X-ray Conference: Jul 27-31, Colorado Springs, CO 44
Cost-effective Trace Element Analysis with TXRF
1 Cost-effective Trace Element Analysis with TXRF Welcome Today s Topics Introduction to XRF and TXRF The S2 PICOFOX benchtop TXRF system Why wait for sample prep? Application studies Environment Medical/Pharma
More informationShorten your Time-to-Result
Shorten your Time-to-Result Effective quality and process control requires the shortest time-to-result possible. This is the time needed from sampling to the final quantitative result. Any advantage results
More informationS2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF
S2 PICOFOX Spectrometry Solutions Innovation with Integrity TXRF S2 PICOFOX True Trace Analysis with XRF for the First Time! You need to know the concentration of trace elements in environmental samples?
More informationS2 PICOFOX. Innovation with Integrity. Benchtop TXRF Spectrometer TXRF
S2 PICOFOX Benchtop TXRF Spectrometer Innovation with Integrity TXRF S2 PICOFOX - Fast Trace Element Analysis with XRF The S2 PICOFOX is the world s first portable benchtop spectrometer for fast quantitative
More informationTRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 236 ABSTRACT TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Hagen Stosnach Röntec GmbH,
More informationTrace elemental analysis of challenging samples in the mining and resource industry. Bruker Nano GmbH, Berlin, Germany April 8, 2015
Trace elemental analysis of challenging samples in the mining and resource industry Bruker Nano GmbH, Berlin, Germany April 8, 2015 4/8/2015 Welcome Today s Topics Introduction to TXRF learn about X-ray
More informationDetermination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS
Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction
More informationICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer
Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full
More informationMS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF
2016 Fall Semester MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF Byungha Shin Dept. of MSE, KAIST 1 Course Information Syllabus 1. Overview of various characterization techniques (1 lecture)
More informationOES - Optical Emission Spectrometer 2000
OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the
More informationOverview of X-Ray Fluorescence Analysis
Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic
More informationApplication of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater
PRAMANA c Indian Academy of Sciences Vol. 76, No. 2 journal of February 2011 physics pp. 361 366 Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater
More informationElemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer
Elemental analysis by X-ray f luorescence Sequential benchtop WDXRF spectrometer Elemental analysis is one of the most important fundamental measurements made for industrial quality control and research
More informationMT Electron microscopy Scanning electron microscopy and electron probe microanalysis
MT-0.6026 Electron microscopy Scanning electron microscopy and electron probe microanalysis Eero Haimi Research Manager Outline 1. Introduction Basics of scanning electron microscopy (SEM) and electron
More informationThermo Scientific icap RQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry
More informationKey Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company
Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences
More informationEnhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS
Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,
More informationWalter Curlook Materials Characterization Laboratory Wallberg Building, 184 College Street, Room 147
Walter Curlook Materials Characterization Laboratory Wallberg Building, 184 College Street, Room 147 Walter Curlook Materials Processing Laboratory Wallberg Building, 184 College Street, Room 53 Introduction
More informationElemental Scientific. Spike Recovery. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast
More informationICP-OES Application Note Number 35
ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable
More informationToday s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy
Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy
More information7) Applications of Nuclear Radiation in Science and Technique (1) Analytical applications (Radiometric titration)
7) Applications of Nuclear Radiation in Science and Technique (1) (Radiometric titration) The radioactive material is indicator Precipitation reactions Complex formation reactions Principle of a precipitation
More informationUNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.
UNIVERSITI SAINS MALAYSIA Second Semester Examination Academic Session 2004/2005 March 2005 KAA 502 Atomic Spectroscopy Time: 3 hours Please make sure this paper consists of FIVE typed pages before answering
More informationNot Just for Paint Chips Anymore. High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions
Not Just for Paint Chips Anymore High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions Objective Develop a sample preparation procedure for analysis of a range of sample
More informationElemental Scientific. Spike Recovery. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast
More informationTest Method: CPSC-CH-E
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure
More informationElemental analysis by X-ray f luorescence. High performance EDXRF elemental analyzer
Elemental analysis by X-ray f luorescence High performance EDXRF elemental analyzer Simple elemental analysis 1. Prepare samples No digestion; minimal sample preparation. Direct non-destructive measurement
More informationNOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation.
Determination of Cadmium and Lead in powdered infant formula by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction NOTICE: This method: - has to be considered only
More informationS8 TIGER Series 2 for ASTM D 6443
Lab Report XRF 139 S8 TIGER Series 2 for ASTM D 6443 Standard Test Method for Determination of Ca, Cl, Cu, Mg, P, S and Zn in Unused Lubricating Oils and Additives Introduction Lubricating oils are generally
More informationUsing cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products
APPLICATION NOTE Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products AN43381 Dr. Jianfeng Cui, Thermo Fisher Scientific, Bremen, Germany Key Words al
More informationWorldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample
Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13 Determination of Major, Minor and Trace Elements in a Clay Sample IAEA Laboratories, Seibersdorf November 2017 CONTENTS
More informationPortable type TXRF analyzer: Ourstex 200TX
Excerpted from Adv. X-Ray. Chem. Anal., Japan: 42, pp. 115-123 (2011) H. Nagai, Y. Nakajima, S. Kunimura, J. Kawai Improvement in Sensitivity and Quantification by Using a Portable Total Reflection X-Ray
More informationStandardless Analysis by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited
by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited XRF Standardless Analysis In this talk What is meant by standardless analysis? Fundamental Parameters
More informationDetermination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors
Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,
More informationAnalysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy
Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Hiroshi Onodera Application & Research Center, JEOL Ltd. Introduction um, PBB and PBDE) are subject to usage restrictions in Europe.
More informationDirect Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry
Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.
More informationX-Ray Fluorescence and Natural History
X-Ray Fluorescence and Natural History How XRF Helps XRF can be used both quantitatively (homogenous samples) and quantitatively (heterogenous samples).! Trace elements in a fossil can help identify source,
More informationMulti-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS
Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction
More informationFebruary 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:
February 20, 2012 Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Subject: ICP-MS Report Job Number: S0CHG688 Dear Joe: Please find enclosed the procedure report for the analysis
More informationElectron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?
Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn? EPMA - what is it? Precise and accurate quantitative chemical analyses of micron-size
More informationDetermination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS
APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity
More informationPRINCIPLE OF ICP- AES
INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively
More informationDirect Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ
Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction
More informationQuantitative XRF Analysis. algorithms and their practical use
Joint ICTP-IAEA School on Novel Experimental Methodologies for Synchrotron Radiation Applications in Nano-science and Environmental Monitoring Quantitative XRF Analysis algorithms and their practical use
More informationInnovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer
Innovative 0.01 0.1 1 10 Intuitive Concentration (ppb) R = 0.99999 Versatile Hydra II C Mercury Analyzer At Teledyne Leeman Labs; atomic spectroscopy is our business our only business. We are industry
More informationElemental analysis of river sediment using the Agilent 4200 MP-AES
Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction
More informationatomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive
Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific
More informationFast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES
Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction
More informationU.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP
Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High
More informationUsing FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples
A P P L I C A T I O N N ot e Atomic Absorption Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples Introduction The Flow Injection Mercury System (FIMS) is a dedicated system
More information110 Lowell Rd Hudson, NH USA Tel:
110 Lowell Rd Hudson, NH 03051 USA Tel: 603-886-8400 www.teledyneleemanlabs.com Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental
More informationBATTERY INDUSTRY STANDARD ANALYTICAL METHOD
BATTERY INDUSTRY STANDARD ANALYTICAL METHOD For the Determination of Mercury, Cadmium and Lead in Alkaline Manganese Cells Using AAS, ICP-AES and "Cold Vapour" European Portable Battery Association (EPBA)
More informationhigh temp ( K) Chapter 20: Atomic Spectroscopy
high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment
More informationDirect Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS
Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide
More informationHigh Accuracy EUV Reflectometry and Scattering at the Advanced Light Source
High Accuracy EUV Reflectometry and Scattering at the Advanced Light Source Eric Gullikson Lawrence Berkeley National Laboratory 1 Reflectometry and Scattering Beamline (ALS 6.3.2) Commissioned Fall 1994
More informationProduct Descriptions and Offerings
Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental analysis. Laboratories in industries ranging from environmental science, oil and
More informationAEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD
Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 219 AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Takao Moriyama 1), Atsushi Morikawa 1), Makoto
More information1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:
1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for
More information2101 Atomic Spectroscopy
2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.
More informationFundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company
Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical
More information1 of 5 14/10/ :21
X-ray absorption s, characteristic X-ray lines... 4.2.1 Home About Table of Contents Advanced Search Copyright Feedback Privacy You are here: Chapter: 4 Atomic and nuclear physics Section: 4.2 Absorption
More informationATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light
ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light by these elements in order to measure their concentration.
More informationLAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong
LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong Introduction: X-ray fluorescence (XRF) spectrometer is an x-ray instrument used for routine, relatively
More informationThe Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction
application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and
More informationHands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment
: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it
More informationProf. Dr. Biljana Škrbić, Jelena Živančev
5 th CEFSER Training Course Analysis of chemical contaminants in food and the environment Faculty of Technology, University of Novi Sad, Novi Sad, Republic of Serbia 7-11 May 2012 Analysis of heavy elements
More informationData report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 18, 2010 SUMMARY
Data report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 8, 200 SUMMARY This report summarizes the quality assurance performed during elemental
More informationAltitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory
Altitude influence of elemental distribution in grass from Rila mountain Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory I. Introduction The application of modern instrumental
More informationAnalysis Repeatability of Trace and Major Elements in a Water Sample
Analysis Repeatability of Trace and Major Elements in a Water Sample Agnès COSNIER HORIBA Scientific Longjumeau, France Keywords: environment Elements: Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li,
More informationUltra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES
Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction
More informationED(P)XRF: SCREENING ANALYSIS AND QUANTITATIVE ANALYSIS with POLARIZED
384 ED(P)XRF: SCREENING ANALYSIS AND QUANTITATIVE ANALYSIS with POLARIZED X-RAYS R. Schramm, J. Heckel, K. Molt 2 Spectra Analytical Instruments, Kleve, Germany. 2Gerhard-Mercator- University Duisburg,
More informationMeeting the Challenges of Soil Analysis with the Avio 200 ICP-OES
APPLICATION NOTE ICP-Optical Emission Spectroscopy Author: Nick Spivey PerkinElmer, Inc. Shelton, CT Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES Introduction Micronutrients contained
More informationTrace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell
Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle
More informationAnalysis of domestic sludge using the Agilent 4200 MP-AES
Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal
More informationEXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy
EXPERIMENT 7 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium
More informationCHEMICAL COMPOUNDS MOLECULAR COMPOUNDS
48 CHEMICAL COMPOUNDS - Dalton's theory does not mention this, but there is more than one way for atoms to come together to make chemical compounds! - There are TWO common kinds of chemical compound, classified
More informationAgilent Technologies at TIAFT 2013
Agilent Technologies at TIAFT 2013 Analytical approaches for the measurement of trace metals in forensic samples Funchal, Madeira September 2-6 CRIMINALISTIC TOXICOLOGY DNA INVESTIGATION DOCUMENT EXAMINATION
More informationPOWER + FLEXIBILITY IN ELEMENTAL ANALYSIS
POWER + FLEXIBILITY IN ELEMENTAL ANALYSIS Energy Dispersive X-Ray Fluorescence Spectrometry (EDXRF) is an atomic spectroscopy that provides elemental composition analysis of materials. The elemental range
More informationAnalysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS
TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically
More informationMulti Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE
1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for
More informationINTRODUCTION TO MEASUREMENTS
INTRODUCTION TO MEASUREMENTS Introduction: In the first three experiments you will learn how to use equipment that to measure length, mass, and volume. You will apply the rules about significant figures
More informationPittcon External Seminar Template X-ray Analysis in Petrochemical Industry: Challenges and Solutions
Pittcon External Seminar Template X-ray Analysis in Petrochemical Industry: Challenges and Solutions Al Martin, Christopher Shaffer, Thermo Fisher Scientific, West Palm Beach, USA 1 The world leader in
More information3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement
3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing, but also as an energy carrier in the cooling or heating cycle. As vast amounts
More informationBasic Digestion Principles
Basic Digestion Principles 1 From Samples to Solutions Direct Analytical Method Solid Sample Problems: Mech. Sample Preparation (Grinding, Sieving, Weighing, Pressing, Polishing,...) Solid Sample Autosampler
More informationOptical Atomic Spectroscopy
Optical Atomic Spectroscopy Methods to measure conentrations of primarily metallic elements at < ppm levels with high selectivity! Two main optical methodologies- -Atomic Absorption--need ground state
More informationThermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials
Thermo Scientific ICP-MS solutions for the semiconductor industry Maximize wafer yields with ultralow elemental detection in chemicals and materials Trace impurity analysis in chemicals an used in the
More informationAtomic Emission Spectra. and. Flame Tests. Burlingame High School Chemistry
Atomic Structure Atomic Emission Spectra and Flame Tests Flame Tests Sodium potassium lithium When electrons are excited they bump up to a higher energy level. As they bounce back down they release energy
More informationDirect Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction
NOTICE: This method: Direct Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction - has to be considered only as guideline
More informationBORON IN GLASS DETERMINATION USING WDXRF
269 ABSTRACT BORON IN GLASS DETERMINATION USING WDXRF Alexander Seyfarth Bruker AXS Inc., 5465 East Cheryl Parkway, Madison, Wisconsin 53711-5373 The application of wavelength-dispersive XRF to the analysis
More informationApplications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry
Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS
More informationChemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy
Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental
More informationApplication field / Industry branch:
Application field / Industry branch: Chemistry / Polymer Industry Electronics Energy Nutrition / Agriculture Geology / Mining Semiconductor Technology Clinical Chemistry / Medicine / Sanitation / Health
More information1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:
1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for
More informationX-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination
Uppsala University Department of Physics and Astronomy Laboratory exercise X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination
More informationIntroduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960
Introduction to X-ray Photoelectron Spectroscopy (XPS) X-ray Photoelectron Spectroscopy (XPS), also known as Electron Spectroscopy for Chemical Analysis (ESCA) is a widely used technique to investigate
More informationCHEMICAL COMPOUNDS MOLECULAR COMPOUNDS
48 CHEMICAL COMPOUNDS - Dalton's theory does not mention this, but there is more than one way for atoms to come together to make chemical compounds! - There are TWO common kinds of chemical compound, classified
More information4.1 Atomic structure and the periodic table. GCSE Chemistry
4.1 Atomic structure and the periodic table GCSE Chemistry All substances are made of atoms this is cannot be chemically broken down it is the smallest part of an element. Elements are made of only one
More informationSingle Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#
1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for
More informationDeterminations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission
0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer
More informationAnalysis of Soil and Sewage Sludge in the Field with a Portable ED-XRF Spectrometer
AN APPLICATION BRIEF FROM SPECTRO ANALYTICAL INSTRUMENTS When results matter Analysis of Soil and Sewage Sludge in the Field with a Portable ED-XRF Spectrometer Introduction Arsenic, barium, cadmium, chromium,
More information