1 Frank Thibodeau, Bruker Detection, Billerica, MA USA Franziska Lange, Bruker Detection, Leipzig, Germany INTRODUCTION Ion Mobility Spectrometry (IMS) is the scientific technique of measuring the drift time of ions across a uniform electric field. It is based on the principle that when different charged molecules (ions) are accelerated through an electric field (with a "drift flow" of gas slightly opposing them) they will travel at different speeds and reach the detector at different times. The advantages of IMS instruments include, relatively small footprint, fairly inexpensive, rugged, extremely fast and extremely sensitive. The purpose of this investigation was to evaluate the effectiveness of IMS technology for detection of TDI in foam plants. IMS technology has been assessed in numerous foam plants under field conditions. TDI was monitored in foam plants which produced automotive, molded and slab products. A handheld IMS instrument surveyed in and around production lines, hot and cold foam, storage tanks, transfer lines, and transfer operations. BACKGROUND All site visits were conducted with the Bruker handheld RAID-M IMS instrument. The IMS instrument was programmed with a TDI library fixed to detect and alarm at or below 1 ppb. Spectra were also collected at various locations at all foam plants for more detailed post analysis. Eight foam plants were visited in the US and Canada and two chemical terminal transfer sites which utilized TDI. Additionally, one plant with installed units for more than 20 months provided feedback. The observations, data and post spectral analyses are included the findings. The RAID-M had the highly sensitive TDI library in Library A of the instrument as shown in Table 1. Page 1 Table 1: Highly sensitive TDI library in ppb. The library above shows TDI readings in ppb, the current instrument version displays on the instrument interface in ppb readings, and records data in ppb.

2 BASICS OF ION MOBILITY IMS is based on the principle that when charged molecules (ions) are accelerated through an electric field (with a "drift flow" of gas slightly opposing them) they will travel at different speeds (due to different size and shape) and reach a detector at different times. All molecules are drawn through the drift region of the detector by the electric field, but larger ions with bigger cross sectional areas will interact with the drift gas longer thus arriving at the detector later than smaller ions. The analysis is carried out under atmospheric pressure which eliminates the need for large, heavy, power hunger pumps to generate a vacuum as is required for a mass spectrometer. As a result, handheld, portable ion mobility instruments are in use in many places, where they are used to detect chemical weapons, conventional weapons, drugs and industrial gases. IMS is an ionization-based time-of-flight technique performed at atmospheric pressure. The heart of IMS is the time-of-flight tube. Time-of-flight tubes come in different sizes and measure ions in single or dual polarity. Generally speaking, the longer the time-of-flight tube, the better the resolution or ability to differentiate between compounds. Other key components that comprise an IMS instrument are described in the following pages. MEMBRANE A continuous ambient air sample is drawn over a semi-permeable membrane. The membrane serves to protect the interior of the cell from particles and moisture, provide a degree of contaminant selectivity, and allow various levels of sensitivity based on ambient contamination conditions. It also enhances the analytical integrity of the instrument by controlling contamination and helping to maintain humidity at desired levels. REACTION CHAMBER The molecules of interest permeate through the membrane, and are transferred to the reaction region. There, the sample is ionized by low-level beta energy emitted by a sealed nickel-63 radiation source or other method such as high energy photoionization. Figure 1 describes the reaction region of an IMS. The sealed nickel-63 source is most commonly used in IMS because it is inexpensive and reliable. The radioactivity is extremely low and is often below threshold values in many countries for licensing and monitoring. Bruker IMS instruments are exempt by the Nuclear Regulatory Commission for licensing purposes and therefore do not require record keeping, controlled radiation areas or wipe testing. Page 2

3 Figure 1. IMS Reaction Region The ions are moved by an electric field to the exit of the reaction chamber. At this point, an electric voltage on a grid prevents the ions from entering into the drift tube. Figure 2. Time of Flight & the Drift Tube By reversing this voltage momentarily, ions are injected into the drift tube for a brief moment and attracted to the faraday plate detector with opposite polarity. The resultant detection parameters are than evaluated as the below figure demonstrates. Figure 2 illustrates the movement of ions in the electric field. Page 3

4 Figure 3. Ion Mobility Spectra with Monomer & Dimer Peak The drift times (in milliseconds) of the ions are corrected to compensate for varying temperature and pressure conditions during analysis. The resultant reduced ion mobility constant K o (in cm² V -1 s -1 ) for each ion along with peak intensity (in Pico amps) and peak area correspond to particular compounds in an IMS library. Figure 3 is a representation of the data generated from the detector in Bruker s proprietary analysis software. Some molecules form a monomer during ionization while others form dimers or even adducts. The formation of monomer or dimer peak systems is compound and concentration dependent. Presence of a monomer with dimer peak series in the positive mode provides confirmation and more confident identification of a chemical compound than the single monomer peak for that compound. With ion mobility, size and shape of the ionized compounds are important in characterization of species for not only single, but multiple compound detections. In some cases, chemical modifiers called dopants are intentionally added to the ionization region to react with the sample. Molecules from the sample source, any water vapor present, and dopant molecules may combine or react when compounds are introduced into the ionization region. The IMS technology uses different reactants ions in each polarity. These form what is called a reactant ion peak which is usually water vapor in the negative mode and a dopant like ammonia in the positive mode. The interaction of the reactant ions with the sampled air form new compounds (product-ions) measured in the time-of-flight tube. All information is vital to correct identification of the intended target species. Library development of course is the key to detecting any substance required. Known samples of analyte are introduced to an IMS instrument and the response observed. If the reactant ion peak diminishes in direct correlation to the introduced analyte peak, then the chemical of interest is detected with confidence. The chemical is then introduced at small concentrations intervals to determine the lowest limit of detection while observing the spectra. From this data, the library is then calibrated for this lone agent. With the upper and lower limits determined temperature and humidity differences are introduced with corresponding spectral reaction observed again. Common or suspected interferent Page 4

5 compounds could then be introduced to minimize masking effects. The resultant individual compound library is used by itself or added to another library. FINDINGS & OBSERVATIONS The beginning of any site survey started with a proof of concept usually in a lab testing the handheld IMS instrument (RAID-M manufactured by Bruker) with a TDI sample. The response was immediate with a reading from 4-16 ppb. Having established confidence that the RAID-M could detect TDI at very low levels, the site survey then commenced on the floor of the factory. Storage Storage tanks, transfer lines, valves, and joints were always surveyed with the handheld to observe any response. The instrument was held for seconds or laid underneath for minutes with no response for TDI in any plant or facility. This shows sound safety practices and monitoring by PFA member plants, but does not diminish the need for constant monitoring wherever TDI is stored or transferred as valves, hoses and fittings can wear out over time. Figure 4 is an example of the storage tank monitoring setup for Bruker s RAID-M. Figure 4. TDI Storage Tank Monitoring with IMS Page 5

6 Smaller amounts of TDI were also stored in drums with either pure TDI or mixed with MDI. Capped drums were surveyed on several occasions with only one response noted with an empty drum. The RAID-M alarmed after five minutes. This only occurred on one occasion. Sniffing with the cap removed on full drums of MDI/TDI mixture with 70%/30% ratio resulted in alarms in the range of 1-8 ppb. Half full or less resulted in no alarms after 1-2 minutes of exposure. Production Lines-Slab Slab foam production lines generally have barriers or side panels and may or may not be vented from the top. The reactor at the head of course has hot TDI and or MDI mixed with other chemicals as part the foam mixture. The hot TDI gases rendered a good signal or instrument response at the beginning of the line and decreased toward the end of the line as the foam formed and cooled. This produced a linear response as high as 12 ppb down to 1 ppb of TDI. Cutting a slab, depending on the temperature of the foam at the time, generally elicited a response of 1-4 ppb. An example of the instrument datalog results as shown below from a slab line foam sample are shown in Table 2. Date: ; Instrument RM_02708 No. ; Time Agent; Conc Bars; RI+; RI-; PBars; lib;state 375; 10:17:22 ; T;TDI; 6;ppb ; 4 ; 28%; 23%; 2; 2 ; A ;WORK 376; 10:17:25 ; T;TDI; 6;ppb ; 4 ; 29%; 22%; 2; 2 ; A ;WORK 377; 10:17:28 ; T;TDI; 6;ppb ; 4 ; 30%; 22%; 2; 2 ; A ;WORK 378; 10:17:32 ; T;TDI; 5;ppb ; 4 ; 26%; 23%; 3; 2 ; A ;WORK 379; 10:17:35 ; T;TDI; 5;ppb ; 4 ; 30%; 27%; 2; 2 ; A ;WORK 380; 10:17:38 ; T;TDI; 4;ppb ; 4 ; 28%; 26%; 2; 2 ; A ;WORK 381; 10:17:41 ; T;TDI; 4;ppb ; 4 ; 34%; 28%; 2; 2 ; A ;WORK 382; 10:17:44 ; T;TDI; 4;ppb ; 3 ; 33%; 29%; 2; 2 ; A ;WORK 383; 10:17:47 ; T;TDI; 4;ppb ; 3 ; 32%; 29%; 2; 2 ; A ;WORK 384; 10:17:50 ; T;TDI; 4;ppb ; 3 ; 32%; 30%; 2; 1 ; A ;WORK 385; 10:17:53 ; T;TDI; 3;ppb ; 3 ; 33%; 31%; 2; 1 ; A ;WORK Page 6

7 386; 10:17:56 ; T;TDI; 3;ppb ; 3 ; 33%; 31%; 2; 1 ; A ;WORK 387; 10:18:00 ; T;TDI; 3;ppb ; 3 ; 34%; 31%; 2; 1 ; A ;WORK 388; 10:18:03 ; T;TDI; 3;ppb ; 3 ; 34%; 31%; 2; 1 ; A ;WORK 389; 10:18:06 ; T;TDI; 3;ppb ; 3 ; 31%; 31%; 2; 1 ; A ;WORK 390; 10:18:09 ; T;TDI; 3;ppb ; 3 ; 32%; 29%; 2; 2 ; A ;WORK 391; 10:18:12 ; T;TDI; 3;ppb ; 3 ; 32%; 32%; 2; 1 ; A ;WORK 392; 10:18:15 ; T;TDI; 3;ppb ; 3 ; 32%; 33%; 2; 1 ; A ;WORK 393; 10:18:18 ; T;TDI; 3;ppb ; 3 ; 32%; 32%; 2; 1 ; A ;WORK 394; 10:18:21 ; T;TDI; 3;ppb ; 3 ; 32%; 32%; 2; 1 ; A ;WORK 395; 10:18:25 ; T;TDI; 2;ppb ; 3 ; 32%; 33%; 2; 1 ; A ;WORK 396; 10:18:28 ; T;TDI; 2;ppb ; 3 ; 33%; 32%; 2; 1 ; A ;WORK 397; 10:18:31 ; T;TDI; 2;ppb ; 3 ; 33%; 34%; 2; 1 ; A ;WORK 398; 10:18:34 ; T;TDI; 2;ppb ; 3 ; 33%; 34%; 2; 1 ; A ;WORK 399; 10:18:37 ; T;TDI; 2;ppb ; 3 ; 33%; 34%; 2; 1 ; A ;WORK 400; 10:18:40 ; T;TDI; 2;ppb ; 3 ; 33%; 35%; 2; 1 ; A ;WORK Table 2: Data log file of Slab Production Line The initial site visits at facilities used a library with HDI included. The were no actual sites visited that used HDI and false alarms did occur with HDI due to similar chemical vapors in the background. Because of this, the TDI only library was used in all subsequent visits. Of course, background spectral readings were taken and analyzed at all sites. The false positives here were a result of taking risks to provide an extremely sensitive library and detection capability. Adjustments in many cases have to be made and libraries customized to meet specific site requirements. In another example, while surveying the entire plant even outside TDI generating areas, the usual response from the handheld IMS unit was no response to any TDI present. In one exception, TDI at was at 1-4 ppb was observed in a general area where none was generated. This was a consistent signal throughout the day. The production of slab foam with TDI occurred behind a wall in another section of the plant. That section was vented from the ceiling when production was underway with readings up to 12 ppb in that area. Therefore the readings in the fabrication area of 1-4 ppb were either false readings or somehow after the production terminated and the fans were turned off, the TDI sank and then Page 7

8 migrated to the non -production area through outer doors. analysis is the only method to determine the ground truth. An air sample and subsequent GC/MS Spectral analysis of the fabrication area reveals many peaks surrounding the location where TDI would be detected. Figure 5 shows an example of one spectrum from the fabrication area. Figure 5: Spectra analysis of Fabrication Area The left spectral capture shows a normal TDI response with well-defined peak of TDI at The right spectral capture shows many close peaks surrounding the 1.46 location. This would cause interference and therefore a likely false reading. The library used was optimized for TDI sensitivity. A higher threshold level would not have caused the response in the detector. Molded Foam Molded foam production plants have some similarities to slab or rigid foam production in the methodology of using TDI. The storage and transfer operations are much the same however, the production lines tend to be much more self-contained. Generally the TDI is contained within the production line itself. TDI at 1-4 ppb range can be detected in vented breaks in the metal housing around the line. In one case, TDI was even observed at the top of the production enclosure. Along the production line itself, sampling with the IMS handheld responded with 1-8 ppb inside the moving line during production. Almost nothing was observed during non-production times. The hot Page 8

9 foam exposed when the mold opened did emit TDI. The handheld IMS was positioned from inch to two feet from the foam within the production enclosure. Sometimes TDI could be detected outside the enclosure, but seldom. It usually had to be within 2 feet of the hot foam. As the foam cooled over 5-10 minutes the signal would diminish to zero. Foam at ambient temperature produced no signal or response from the IMS instrument. Hot foam produced a TDI alarm at every site but cured, cooled foam produced no response. While collecting background spectra, it was observed that there was a very strong signal in the positive mode of a chemical vapor not in the library. Figure 6 is a spectrum containing the unknown chemical in positive mode. It reacted extremely well with the positive reactant ion peak which lends itself to excellent detection capability. There are theories as to what the chemical could be, but without an air sample analyzed by GC/MS, there is not definitive way of knowing. Once this occurs and a sample provided to the lab, a library can be calibrated to include it in any IMS instrument for the customer application. Page 9 Figure 6: Spectral Capture Positive Mode Unknown This chemical does permeate the plant and is common to several flexible foam plants, but very prominent in one. It may or may not be harmful. It does however completely dominate detection in the positive mode without affecting negative mode detection of TDI. Wall Mounted vs Handheld IMS Instruments Handheld units are certainly needed for conducting plant surveys, leak detection, monitoring transfer operations, defining limits of any contaminated zone, and monitoring the thoroughness of

10 decontamination with its speed, sensitivity and reliability. Wall mounted IMS would be used at critical points to continuously monitor for any fugitive TDI emissions. How can they be used in tandem? In one installed base, the wall mounted IMS positioned for 24/7 monitoring near an outside door detected TDI while transfer operations from a delivery truck were ongoing. The plant personnel responded with a handheld IMS for verification and detected TDI leaking from a transfer hose. It was discovered that the delivery company was using low quality hoses. The company was directed to use the proper hoses in the future thus preventing a more serious situation. Experimental During the last site visit, we decided to experiment with a new software upgrade to the RAID-M handheld. Samples were collected with a RAID-M with upgraded software and with a standard RAID-M then the results were compared in real time to the industry standard colorimetric instrument. A flexible foam production line with large buns off a long conveyor provided ample TDI emissions at levels below 20 ppb maximum. All three instruments were detecting TDI in tandem. Figure 7 is a picture of all three instruments alarming with roughly the same detection levels. Figure 7: RAID-M standard read bar readout; RAID-M w/new prototype ppb readout; Colorimetric reading The first RAID-M reads 3 bars or 2.1 to 4.1 ppb TDI. The second RAID-M reads directly 1-8 ppb, the colorimetric instrument reads 2.5 ppb. The instruments are certainly in agreement given standard margin of error. Page 10

11 The two RAID-M IMS units were almost perfectly in agreement, took only seconds to respond and alarm, and display in either bars or ppb readings. The colorimetric instrument took 2-3 minutes to alarm and in many cases did not detect any TDI while both IMS instruments alarmed, cleared down, and then reengaged simultaneously. The speed of the IMS allowed operators to ascertain the real location of TDI emissions. CONCLUSIONS 1. IMS instruments provide fast, reliable, accurate, and very sensitive TDI detection in foam plants. 2. TDI can be detected with IMS in hot foam until cooled some 5-10 minutes after forming. 3. TDI can be detected by IMS some 2 ft from a hot source depending on concentration. 4. Slab production lines cause dispersals of TDI within all open areas around the conveyor line at levels observed below 8 ppb. 5. The speed and accuracy of IMS instruments provides actionable information to protect the health and safety of plant workers with minimal resources which positively affects the bottom line. RECOMMENDATIONS 1. Consider IMS technology for real time, actionable monitoring for TDI emissions 2. Continue development of customized IMS libraries to meet industry needs. 3. Provide readouts in ppb for all IMS instruments to include in the datalogger. 4. Develop external display for wall mounted IMS units. Page 11

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

Miniaturized Sensor for the Detection of Environmental Pollutants Alexander Graf 1, Ronald Stübner 1, Christian Kunath 1, Sebastian Meyer 1, Harald Schenk 1 1 Fraunhofer Institute for Photonic Microsystems

More information

MM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE

MM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE MM 2 The new Generation of Mobile Mass Spectrometers Innovation with Integrity Defence CBRNE Picture by: MOWAG GmbH Mobile CWA Detection and Identification Low weight and high performance detector Introducing

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

Chemical Health and Safety General Program

Chemical Health and Safety General Program Chemical Health and Safety General Program I. Objective To establish minimum requirements for storage, handling and use of chemicals. II. Scope This process applies to employees and operations involved

More information

UALR Radiation Safety Office

UALR Radiation Safety Office UALR Radiation Safety Office ETAS-329 501-569 8210 Graduate Institute of Technology University of Arkansas at Little Rock Regulatory Authority Nuclear Regulatory Commission (NRC) EPA, DoE, DoT, OSHA Agreement

More information

New Test Results For Physical Separation Of Tritium From Noble Gases And It s Implications For Sensitivity And Accuracy In Air And Stack Monitoring

New Test Results For Physical Separation Of Tritium From Noble Gases And It s Implications For Sensitivity And Accuracy In Air And Stack Monitoring New Test Results For Physical Separation Of Tritium From Noble Gases And It s Implications For Sensitivity And Accuracy In Air And Stack Monitoring -Robert Goldstein, Ivan Mitev, Dell Williamson, Overhoff

More information

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Ensuring ultimate cleanliness for maximum adhesion Application Note Author John Seelenbinder Agilent Technologies,

More information

TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL

TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL NUCLE A R FORENSIC S INTERN ATION A L TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL This document was designed

More information

Secondary Ion Mass Spectroscopy (SIMS)

Secondary Ion Mass Spectroscopy (SIMS) Secondary Ion Mass Spectroscopy (SIMS) Analyzing Inorganic Solids * = under special conditions ** = semiconductors only + = limited number of elements or groups Analyzing Organic Solids * = under special

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Storing, using and disposing of unsealed radioactive substances in a Type C Laboratory: Extract of regulatory requirements

Storing, using and disposing of unsealed radioactive substances in a Type C Laboratory: Extract of regulatory requirements Storing, using disposing of unsealed radioactive substances in a Type C Laboratory: Extract of regulatory requirements Radiation Protection Control (Ionising Radiation) Regulations 2000 Requirements for

More information

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ Design and Construction of a High Energy X-Ray R&D Facility, and the Development and Optimization of Real Time Radioisotopic Characterization of Remote Handled Waste at MeV Energies. S. Halliwell, V.J.Technologies

More information

1. What would be the dose rate of two curies of 60Co with combined energies of 2500 kev given off 100% of the time?

1. What would be the dose rate of two curies of 60Co with combined energies of 2500 kev given off 100% of the time? 1.11 WORKSHEET #1 1. What would be the dose rate of two curies of 60Co with combined energies of 500 kev given off 100% of the time?. What would be the dose rate of 450 mci of 137Cs (gamma yield is 90%)?

More information

Prima PRO Process Mass Spectrometer

Prima PRO Process Mass Spectrometer APPLICATION NOTE Prima PRO Process Mass Spectrometer No. xxxx Controlling emissions by multi-component analysis of Volatile Organic Compounds (VOC) Authors: Peter Traynor, Thermo Fisher Scientific, Franklin,

More information

Characterization of Large Structures & Components

Characterization of Large Structures & Components Structures & Components KEY BENEFITS Key Drivers: Lack of good knowledge about the position, the identification and the radiological specification of contamination on or inside large components. Significant

More information

Questions, Myths and Misconceptions about Using Photoionization Detectors

Questions, Myths and Misconceptions about Using Photoionization Detectors Questions, Myths and Misconceptions about Using Photoionization Detectors Solvent, fuel and other VOC vapours are pervasively common in many workplace environments. Increased awareness of the toxicity

More information

UV3000. Accurate, precise, and portable ambient gas point analyzer

UV3000. Accurate, precise, and portable ambient gas point analyzer UV3000 Accurate, precise, and portable ambient gas point analyzer The Cerex UV3000 is a multifunction analyzer designed to detect part per billion (ppb) to percent level concentrations of multiple gases

More information

A. Identify the highly penetrating radioactive emission that exposed the photographic plates.

A. Identify the highly penetrating radioactive emission that exposed the photographic plates. Name Unit 3: Nuclear Chemistry Date Part 2 Questions 1. In 1896, Antoine H. Becquerel discovered that a uranium compound could expose a photographic plate wrapped in heavy paper in the absence of light.

More information

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading

More information

EAS 535 Laboratory Exercise Weather Station Setup and Verification

EAS 535 Laboratory Exercise Weather Station Setup and Verification EAS 535 Laboratory Exercise Weather Station Setup and Verification Lab Objectives: In this lab exercise, you are going to examine and describe the error characteristics of several instruments, all purportedly

More information

nmos IC Design Report Module: EEE 112

nmos IC Design Report Module: EEE 112 nmos IC Design Report Author: 1302509 Zhao Ruimin Module: EEE 112 Lecturer: Date: Dr.Zhao Ce Zhou June/5/2015 Abstract This lab intended to train the experimental skills of the layout designing of the

More information

Applied Nuclear Science Educational, Training & Simulation Systems

Applied Nuclear Science Educational, Training & Simulation Systems WWW.NATS-USA.COM Applied Nuclear Science Educational, Training & Simulation Systems North American Technical Services Bridging Technology with the Latest in Radiation Detection Systems The Center For Innovative

More information

Hydrogen & Chlorine monitoring for the Chlor-Alkali Industry

Hydrogen & Chlorine monitoring for the Chlor-Alkali Industry technical datasheet Application Note - AN9041 Hydrogen & Chlorine monitoring for the Chlor-Alkali Industry Measurement Technology Limited (MTL) part of Eaton's Crouse-Hinds business, design and manufacture

More information

Vacuum. Kai Schwarzwälder, Institut für Physik Universität Basel October 6 th 2006

Vacuum. Kai Schwarzwälder, Institut für Physik Universität Basel October 6 th 2006 Physics,, Technology and Techniques of the Vacuum Kai Schwarzwälder, Institut für Physik Universität Basel October 6 th 2006 Outline Introduction and basics Defintion of Vacuum Vacuum A vacuum is a volume

More information

Noble Gas Control Room Accident Filtration System for Severe Accident Conditions (N-CRAFT)

Noble Gas Control Room Accident Filtration System for Severe Accident Conditions (N-CRAFT) E-Journal of Advanced Maintenance Vol.7-1 (2015) 34-42 Japan Society of Maintenology Noble Gas Control Room Accident Filtration System for Severe Accident Conditions (N-CRAFT) Axel HILL *1, Dr. Cristoph

More information

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation chromatography/size-exclusion chromatography (GPC/SEC)

More information

Defining quality standards for the analysis of solid samples

Defining quality standards for the analysis of solid samples Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples

More information

M2 SERIES THERMOSTATS 0 F to 240 F, Narrow Differential, Hermetically Sealed ½

M2 SERIES THERMOSTATS 0 F to 240 F, Narrow Differential, Hermetically Sealed ½ M2 SERIES THERMOSTATS 0 F to 240 F, Narrow Differential, Hermetically Sealed ½ Introduction The Klixon M2 thermostat is constructed with a snap acting bimetal disc that serves as the actuating element.

More information

S8 TIGER Series 2 for ASTM D 6443

S8 TIGER Series 2 for ASTM D 6443 Lab Report XRF 139 S8 TIGER Series 2 for ASTM D 6443 Standard Test Method for Determination of Ca, Cl, Cu, Mg, P, S and Zn in Unused Lubricating Oils and Additives Introduction Lubricating oils are generally

More information

Laird Thermal Systems Application Note. Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment

Laird Thermal Systems Application Note. Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment Laird Thermal Systems Application Note Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment Contents Overview...3 Thermal Management Techniques...3 Application Examples...3

More information

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed

More information

Chlorine and Hydrogen Chloride Monitoring Utilizing Ion Mobility Spectroscopy (IMS)

Chlorine and Hydrogen Chloride Monitoring Utilizing Ion Mobility Spectroscopy (IMS) Chlorine and Hydrogen Chloride Monitoring Utilizing Ion Mobility Spectroscopy (IMS) Tad Bacon Jim Taylor Kurt Webber VP Engineering Sr. Chemical Engineer VP Sales Molecular Analytics Molecular Analytics

More information

12 Moderator And Moderator System

12 Moderator And Moderator System 12 Moderator And Moderator System 12.1 Introduction Nuclear fuel produces heat by fission. In the fission process, fissile atoms split after absorbing slow neutrons. This releases fast neutrons and generates

More information

Peltier Application Note

Peltier Application Note Peltier Application Note Early 19th century scientists, Thomas Seebeck and Jean Peltier, first discovered the phenomena that are the basis for today s thermoelectric industry. Seebeck found that if you

More information

Accelerator Facility Accident Report

Accelerator Facility Accident Report Accelerator Facility Accident Report 31 May 2013 Incorporated Administrative Agency - Japan Atomic Energy Agency Inter-University Research Institute - High Energy Accelerator Research Organization Subject:

More information

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed

More information

The detector and counter are used in an experiment to show that a radioactive source gives out alpha and beta radiation only.

The detector and counter are used in an experiment to show that a radioactive source gives out alpha and beta radiation only. ATOMS AND NUCLEAR RADIATION PART II Q1. The detector and counter are used in an experiment to show that a radioactive source gives out alpha and beta radiation only. Two different types of absorber are

More information

Fig 1. Power Tower during Operation

Fig 1. Power Tower during Operation Accurate Flux Calculations Using Thermographic IR cameras in Concentrated Solar Power Fields A. Eitan*, G. Naor*, R. Hayut*, I. Segev*, J. Golbert**, S. Pekarsky*, A. Zisken*, G. Medan*, A. Feigelstock*,

More information

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples

More information

FP release behavior at Unit-2 estimated from CAMS readings on March 14 th and 15 th

FP release behavior at Unit-2 estimated from CAMS readings on March 14 th and 15 th Attachment 2-11 FP release behavior at Unit-2 estimated from CAMS readings on March 14 th and 15 th 1. Outline of the incident and the issue to be examined At Unit-2 of the Fukushima Daiichi NPS, the reactor

More information

An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples

An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples Abstract Color-Tec is a field-based analytical method developed by Ecology and

More information

BETA-RAY SPECTROMETER

BETA-RAY SPECTROMETER 14 Sep 07 β-ray.1 BETA-RAY SPECTROMETER In this experiment, a 180, constant-radius magnetic spectrometer consisting of an electromagnet with a Geiger-Muller detector, will be used to detect and analyze

More information

Making Decisions with Insulation

Making Decisions with Insulation More on Heat Transfer from Cheresources.com: FREE Resources Making Decisions with Insulation Article: Basics of Vaporization Questions and Answers: Heat Transfer Experienced-Based Rules for Heat Exchangers

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

RADIATION-METER TM-91/TM-92

RADIATION-METER TM-91/TM-92 RADIATION-METER TM-91/TM-92 User s Manual EN 1 / 16 CONTENTS 1. introduction... 3 2. Safety Precaution... 4 3. Specification... 6 4. Identifying parts... 7 5. Operation Procedure... 8 6. Battery Replacement...

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature Control without Compromise: rapid changes, precise temperatures, powerful complex functionality.

More information

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which

More information

The PEAC-WMD Gamma Radiation Dose Calculator

The PEAC-WMD Gamma Radiation Dose Calculator The PEAC-WMD Gamma Radiation Dose Calculator During the last couple of months newsletters I ve discussed some of the new computational tools included in the PEAC-WMD 2007 (v5.5) application. This month

More information

Quantitative and Qualitative Determination of Organics in Water Using Electrospray Ionization Coupled with Ion Mobility Spectrometry

Quantitative and Qualitative Determination of Organics in Water Using Electrospray Ionization Coupled with Ion Mobility Spectrometry Quantitative and Qualitative Determination of Organics in Water Using Electrospray Ionization Coupled with Ion Mobility Spectrometry Final Progress Report 15 Mar 98 14 Sep 02 Agreement No. DAAG55-98-1-0107

More information

GUIDE TO LABORATORY SURVEYS. Introduction

GUIDE TO LABORATORY SURVEYS. Introduction APPENDIX - V GUIDE TO LABORATORY SURVEYS Introduction Routine laboratory surveys are an important part of the overall radiation safety program in a laboratory. Surveys provide a direct measure of the presence

More information

Pros and Cons of Water Analysis Methods

Pros and Cons of Water Analysis Methods Water Lens, LLC 4265 San Felipe, Suite 1100 Houston, Texas 77027 Office: (844) 987-5367 www.waterlensusa.com Pros and Cons of Water Analysis Methods Prepared by: Adam Garland, CTO Water Lens, LLC ICP-MS/OES

More information

TECHNICAL DESCRIPTION SPECTRAFLOW ON LINE ANALYZER for BELT CONVEYOR APPLICATION

TECHNICAL DESCRIPTION SPECTRAFLOW ON LINE ANALYZER for BELT CONVEYOR APPLICATION TECHNICAL DESCRIPTION SPECTRAFLOW ON LINE ANALYZER for BELT CONVEYOR APPLICATION TECHNICAL SPECIFICATION SPECTRAFLOW ON LINE ANALYZER FOR BELT CONVEYOR CONTENTS 1. SpectraFlow Technical Description...

More information

Utilization of Egyptian Research Reactor and modes of collaboration

Utilization of Egyptian Research Reactor and modes of collaboration Utilization of Egyptian Research Reactor and modes of collaboration Mohamed A. GAHEEN Director of ETRR-2, Egypt TM on Research Reactor Application for Materials Under High Neutron Fluence, Vienna, 17-21

More information

Extrel is widely respected for the quality of mass spectrometer systems that are

Extrel is widely respected for the quality of mass spectrometer systems that are Extrel is widely respected for the quality of mass spectrometer systems that are available to the world's top research scientists. In response to increasing requests for complete turn-key systems built

More information

Calibration of the Modular Neutron Array (MoNA)

Calibration of the Modular Neutron Array (MoNA) Calibration of the Modular Neutron Array (MoNA) Robert Pepin Gonzaga University ~1~ Calibration of the Modular Neutron Array (MoNA): Figure 1 - A rendering of the Modular Neutron Array In recent years

More information

Safety in the Chemistry Laboratory

Safety in the Chemistry Laboratory Safety in the Chemistry Laboratory CHAPTER1 Safety must be everyone s primary concern in the chemistry lab. Understanding and following all safety rules in the organic chemistry lab is critical to your

More information

Microscopy Cryostat System

Microscopy Cryostat System OF AMERICA, INC. Microscopy Cryostat System RC102-CFM Microscopy Cryostat offers fast cooldown, high efficiency, lowest thermal drift, excellent temperature stability and ultra low vibration Optical cryostat

More information

10.1 RADIOACTIVE DECAY

10.1 RADIOACTIVE DECAY 10.1 RADIOACTIVE DECAY When Henri Becquerel placed uranium salts on a photographic plate and then developed the plate, he found a foggy image. The image was caused by rays that had not been observed before.

More information

New Developments in Tritium Monitors

New Developments in Tritium Monitors Overhoff Technology Corporation New Developments in Tritium Monitors 1 Overhoff Technology Design and manufacture of Tritium in Air monitors with ionization chamber detectors Portable Fixed 2 Overhoff

More information

IEC Tritium Standard

IEC Tritium Standard detect and identify IEC 62303 Tritium Standard Dr. Alfred Klett Berthold Technologies, Bad Wildbad, Germany 21 st Annual Air Monitoring Users Group (AMUG) Meeting Palace Station Hotel, Las Vegas, Nevada,

More information

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS Agilent Technologies/Wasson ECE Monomer Analyzer Application Authors Fred Feyerherm 119 Forest Cove Dr. Kingwood, TX

More information

The Franck-Hertz Experiment Physics 2150 Experiment No. 9 University of Colorado

The Franck-Hertz Experiment Physics 2150 Experiment No. 9 University of Colorado Experiment 9 1 Introduction The Franck-Hertz Experiment Physics 2150 Experiment No. 9 University of Colorado During the late nineteenth century, a great deal of evidence accumulated indicating that radiation

More information

costech instruments ECS 4010 Elemental Combustion System CHNS-O

costech instruments ECS 4010 Elemental Combustion System CHNS-O instruments ECS 4010 Elemental Combustion System CHNS-O costech Costech Analytical Tecnologies Inc. 26074 Avenue Hall, Suite 14 Valencia, CA 91355 - USA Phone: (661) 297-2395 Fax: (661) 297-5492 e-mail:

More information

Accelerator Mass Spectroscopy

Accelerator Mass Spectroscopy Accelerator Mass Spectroscopy Accelerator Mass Spectroscopy (AMS) is a highly sensitive technique that is useful in isotopic analysis of specific elements in small samples (1mg or less of sample containing

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Characterization of Odor & Taste Components Using Olfactory Detection and Solid Phase Microextraction GC-MS

Characterization of Odor & Taste Components Using Olfactory Detection and Solid Phase Microextraction GC-MS Characterization of Odor & Taste Components Using Olfactory Detection and Solid Phase Microextraction GC-MS By Amy Porter, Ph.D., Operations Manager, Impact Analytical and Karen Griffin, Senior Analytical

More information

ISEmax CAM40/CAS40. Technical Information

ISEmax CAM40/CAS40. Technical Information Technical Information Online measurement of nutrient parameters Ion-selective electrode system for the continuous measurement of ammonium and nitrate Application The ion-selective electrode system works

More information

Sensing When the Cycle is Over. Using In-Mold Impedance Sensors in Thermoset Molding. Tom Trexler. Signature Control Systems

Sensing When the Cycle is Over. Using In-Mold Impedance Sensors in Thermoset Molding. Tom Trexler. Signature Control Systems Sensing When the Cycle is Over Using In-Mold Impedance Sensors in Thermoset Molding Tom Trexler International AVK-TV Conference Introduction Impedance sensing technology provides an important new method

More information

ATTACHMENT Mitigating Systems

ATTACHMENT Mitigating Systems ATTACHMENT 71111.01 INSPECTABLE AREA: Adverse Weather Protection CORNERSTONE: Initiating Events Mitigating Systems EFFECTIVE DATE: January 1, 2010 INSPECTION BASES: Weather conditions leading to loss of

More information

What happens during nuclear decay? During nuclear decay, atoms of one element can change into atoms of a different element altogether.

What happens during nuclear decay? During nuclear decay, atoms of one element can change into atoms of a different element altogether. When Henri Becquerel placed uranium salts on a photographic plate and then developed the plate, he found a foggy image. The image was caused by rays that had not been observed before. For his discovery

More information

Measurement of Tritium in Helium

Measurement of Tritium in Helium detect and identify Measurement of Tritium in Helium Dr. Alfred Klett Berthold Technologies, Bad Wildbad, Germany 22 nd Annual Air Monitoring Users Group (AMUG) Meeting Palace Station Hotel, Las Vegas,

More information

Clearance Monitoring. Chris Goddard.

Clearance Monitoring. Chris Goddard. Clearance Monitoring Chris Goddard Outline What is Clearance? Clearance Limits around Europe Measurement techniques Plastic scintillators Long Range Alpha Detection Example systems Thermo SAM12 VF FRM-2

More information

Exercise 4-3. Titration of Weak Acids EXERCISE OBJECTIVE DISCUSSION OUTLINE. The 5% rule DISCUSSION

Exercise 4-3. Titration of Weak Acids EXERCISE OBJECTIVE DISCUSSION OUTLINE. The 5% rule DISCUSSION Exercise 4-3 Titration of Weak Acids EXERCISE OBJECTIVE Titrate both a weak acid solution and a weak polyprotic acid solution with a strong base solution. Plot a graph using the titration data, analyze

More information

Harris: Quantitative Chemical Analysis, Eight Edition

Harris: Quantitative Chemical Analysis, Eight Edition Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of

More information

Assessing The Thermal Performance of Building Enclosure Materials Using A Medium-Size Hot Box Chamber Summer Meeting Torkan Fazli

Assessing The Thermal Performance of Building Enclosure Materials Using A Medium-Size Hot Box Chamber Summer Meeting Torkan Fazli Assessing The Thermal Performance of Building Enclosure Materials Using A Medium-Size Hot Box Chamber 2014 Summer Meeting Torkan Fazli Introduction The building construction industry consumes significant

More information

Nelson Environmental Technologies, Inc. ( Chlorine and Chlorine Monitoring 2011 November 8

Nelson Environmental Technologies, Inc. (  Chlorine and Chlorine Monitoring 2011 November 8 This Power Point presentation explains: 1. the forms of chlorine found in municipal water supplies; 2. how to monitor them; 3. and, methods of treating them in hemodialysis water treatment systems. 1 Water

More information

How many protons are there in the nucleus of the atom?... What is the mass number of the atom?... (Total 2 marks)

How many protons are there in the nucleus of the atom?... What is the mass number of the atom?... (Total 2 marks) Q1. The diagram shows an atom. How many protons are there in the nucleus of the atom?... What is the mass number of the atom?... (Total 2 marks) Page 1 of 53 Q2. The picture shows a man at work in a factory

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

PRESENTER'S GUIDE "HAZMAT LABELING" Training for the OSHA HAZARDOUS WASTE OPERATIONS and EMERGENCY RESPONSE (HAZWOPER) REGULATION

PRESENTER'S GUIDE HAZMAT LABELING Training for the OSHA HAZARDOUS WASTE OPERATIONS and EMERGENCY RESPONSE (HAZWOPER) REGULATION PRESENTER'S GUIDE "HAZMAT LABELING" Training for the OSHA HAZARDOUS WASTE OPERATIONS and EMERGENCY RESPONSE (HAZWOPER) REGULATION Quality Safety and Health Products, for Today... and Tomorrow OUTLINE OF

More information

Technical Note 103. Response Enhancement for the Loral CCDs. using Ultra-Violet Flooding. A. P. Oates (RGO)

Technical Note 103. Response Enhancement for the Loral CCDs. using Ultra-Violet Flooding. A. P. Oates (RGO) echnical Note 103 Response Enhancement for the oral CCDs using Ultra-Violet Flooding. A. P. Oates (RGO) July 1996 1 Response Enhancement for the oral CCDs using Ultra-Violet Flooding. 1. Introduction.

More information

Introduction. Principle of Operation

Introduction. Principle of Operation Introduction Ionizing radiation that is associated with radioactivity cannot be directly detected by our senses. Ionization is the process whereby the radiation has sufficient energy to strip electrons

More information

CCS140 and CCS141. Technical Information. Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly

CCS140 and CCS141. Technical Information. Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly Technical Information CCS140 and CCS141 Sensors for free chlorine Amperometric, membrane-covered sensors for installation in the CCA250 assembly Application Oxidising agents such as chlorine or anorganic

More information

Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications. Mitch Wells, Ph.D. Philip Tackett, Ph.D.

Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications. Mitch Wells, Ph.D. Philip Tackett, Ph.D. Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications Mitch Wells, Ph.D. Philip Tackett, Ph.D. HEMS Workshop 2017 FLIR Systems, Inc. Industrial Commercial Government

More information

Non Member State Summer Student Program at ISOLDE

Non Member State Summer Student Program at ISOLDE Program at ISOLDE LabVIEW readout of a gauge pressure meter for the VITO beam line. Atomic local studies in Graphene. Diana Sofía Valverde Mendez Supervisors: Torben Esmann Mølholt

More information

RECORD AND ANALYSE THE PRESSURE-ENTHALPY DIAGRAM FOR A COMPRESSION HEAT PUMP

RECORD AND ANALYSE THE PRESSURE-ENTHALPY DIAGRAM FOR A COMPRESSION HEAT PUMP Thermodynamics Heat cycles Heat Pump RECORD AND ANALYSE THE PRESSURE-ENTHALPY DIAGRAM FOR A COMPRESSION HEAT PUMP Demonstrate how an electric compression heat pump works Quantitatively investigate of the

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

INTRODUCTION Strained Silicon Monochromator Magnesium Housing Windows for Monochromator Shutter and Collimator Fission Detector HOPG Monochromator

INTRODUCTION Strained Silicon Monochromator Magnesium Housing Windows for Monochromator Shutter and Collimator Fission Detector HOPG Monochromator Design for a Four-Blade Neutron Interferometer INTRODUCTION Strained Silicon Monochromator The neutron beam used for this interferometer is separated from the NIST reactor's main beam using a strained

More information

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY PURPOSE This procedure provides instructions for developing, maintaining, and documenting, radiation safety procedures conducted at the Cyclotron Radiochemistry

More information

Multilayer Ceramic Chip Capacitors

Multilayer Ceramic Chip Capacitors HIGH VOLTAGE SERIES JARO high voltage series Multilayer Ceramic Capacitors are constructed by depositing alternative layers of ceramic dielectric materials and internal metallic electrodes, by using advanced

More information

L-35 Modern Physics-3 Nuclear Physics 29:006 FINAL EXAM. Structure of the nucleus. The atom and the nucleus. Nuclear Terminology

L-35 Modern Physics-3 Nuclear Physics 29:006 FINAL EXAM. Structure of the nucleus. The atom and the nucleus. Nuclear Terminology 9:006 FINAL EXAM L-5 Modern Physics- Nuclear Physics The final exam is on Monday MAY 7:0 AM - 9:0 AM in W90 CB The FE is not cumulative, and will cover lectures through 6. (50 questions) The last regular

More information

RADIATION SAFETY GUIDELINES FOR NON-USERS

RADIATION SAFETY GUIDELINES FOR NON-USERS RADIATION SAFETY GUIDELINES FOR NON-USERS This is a Read and Sign Awareness Training document. You should read and sign this document if you: 1. DO NOT work directly with radioactive materials, but 2.

More information

3 Types of Heat Transfer

3 Types of Heat Transfer 3 Types of Heat Transfer The movement of heat from a warmer object to a cooler object. Heat Transfer- 1. Conduction Heat transfer by direct contact of molecules. In other words, when one molecule runs

More information

Owlstone Nanotech Inc

Owlstone Nanotech Inc Owlstone Nanotech Inc PIDplus Technical Paper USA POC: Mark Brennan mark.brennan@owlstonenanotech.com 400 Rella Boulevard Suite 160 Suffern, NY 10901 Tel: +1 845-533-4225 Fax: +1 845-533-4232 www.owlstonenanotech.com

More information

Vacuum techniques (down to 1 K)

Vacuum techniques (down to 1 K) Vacuum techniques (down to 1 K) For isolation (deep Knudsen regime) liquid helium dewar / inner vacuum jacket Leak testing at level 10-11 Pa m3/s (10-10 mbar l/s) liquid helium dewar & transfer syphon

More information

PRINCIPLE OF ICP- AES

PRINCIPLE OF ICP- AES INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information