There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine.

Size: px
Start display at page:

Download "There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine."

Transcription

1 Measurement of radioactivity. Radioactive decay is a random process and therefore fluctuations are expected in the radioactivity measurement. That is why measurement of radioactivity must be treated by statistical methods. In every measurement a deviation from the true value or error is likely to occur. There are two types of errors systematic and random. The accuracy of a measurement indicates how closely it agrees with the true value. The precision of a series of measurements describes the reproducibility and indicates the deviation from the average or mean value. The closer the measurement is to the average value, the higher the precision, whereas the closer the measurement is to the true value, the more accurate is the measurement. It is important to keep in mind, that a series of measurements may be quite precise but their value may be far from the true value. Precision can be improved by eliminating the random errors, better accuracy is obtained by removing both the random and systematic errors. The average or mean value is obtained by adding the values of all measurements divided by the number of measurements. The standard deviation indicates the deviation from the mean value and is a measure of precision. The standard deviation in radioactive measurements indicates the statistical fluctuations in radioactive disintegration. If the number of measurements is large, the distribution can be approximated by a Gaussian distribution even if the radioactive decay follows the Poisson distribution law. For practical reasons, only single measurement is obtained on radioactive sample instead of multiple repeat counts to determine the mean value. The precision of a count of a radioactive sample can be increased by accumulating a large number of counts in a single measurement because of decreasing the standard deviation. Interaction of radiation with matter. All radiations may interact with the atoms of the matter during their passage through it producing ionization and excitation of the atoms. These radiations are called ionizing radiation. The mechanism of interaction differ for particulate and electromagnetic type of radiation. The interaction of beta particles as a charged particles and gamma radiation as an electromagnetic radiation is the most important from the point of view of using them in nuclear medicine. Beta particles interact primarily with the electrons of the absorber atoms and rarely with the nucleus. Ionization occurs when beta particle transfers sufficient amount of its energy to the orbital electron and ejects it from the atom. As a result ion pair is formed. This process may rupture chemical bonds in the molecules. Ionization is used namely in the radiation therapy and also serves as a mean of the detection of charged particles in ion chambers. When energetic beta particles, namely electrons, pass close to the nucleus of the atom, they lose energy as a result of deceleration. This loss of energy appears as an x ray and is called bremsstrahlung. Bremsstrahlung production increases with the kinetic energy of the beta particles and the atomic number of the absorber. That is why high energy beta particles are stored in plastic rather than shielding by lead. Beta plus particles, positrons, combine with the orbital electrons and produce two 511 kev photons of gamma radiation, so called annihilation radiation, that are emitted in exactly opposite directions. This is the basis of positron emission tomography. Gamma rays interact with orbital electrons and if their energy is very high they may also interact with the nucleus of the absorber atoms. They travel a long path in the absorber before losing all energy and so they are called penetrating radiations. There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine. Photoelectric effect means transfer of all energy of gamma photon to an orbital electron, called photoelectron, and ejecting it from the atom. The photoelectron then loses its energy by ionization and excitation.

2 Compton scattering means transfer of only a part of energy of gamma photon to an electron and ejecting it. The gamma photon with less energy is deflected from its original direction. The scattered photon may then undergo further photoelectric or Compton interaction and the Compton electron may cause ionization or excitation. Depending on the photon energy and the density and thickness of the absorber, some photons may pass through the absorber without any interaction. Attenuation of gamma radiations by means of their interaction with absorber is an important factor in radiation protection. The term half value layer is defined as the thickness of the absorber that reduces the intensity of a photon beam by one half. It depends on the energy of the radiation and the atomic number of the absorber. Radiation detection. Interaction of ionizing radiations with the matter is also used for their detection and measurement. There are several principles of radiation detection in nuclear medicine. Some of them are used in radiation protection, others in measurement and imaging. The oldest principle is darkening of photographic emulsion. This principle is used in the personnel dosimetry. The film badge is most popular and cost effective for personnel monitoring and gives reasonably accurate readings of exposures from beta, gamma and x radiations. The film badge consists of a radiation sensitive film held in a plastic holder. Filters of copper and lead are attached to the holder to differentiate exposure from different types and energies of radiation. Another principle is thermoluminiscence. Several inorganic crystals (e.g. LiF) can accumulate radiation energy and hold it. If the crystal is heated from 300 to 400 degrees of Celsius, it emits light in amount proportional to the absorbed energy. Thermoluminiscent dosimeters, so called TLD, are mostly used as a finger dosimeters, so inorganic crystals are held in a plastic holders and plastic rings. It gives an accurate exposure reading and can be reused. Another principle is converting the energy of radiation to electric current. There are two basic principles based on ionization and excitation. First is ionization of gas molecules, the second is excitation and ionization of solid, liquid or plastic material, called scintilator, which emits photons of light after absorbing radiation. Light is then converted to the electric current by means of photomultiplier tube. Gas filled detectors collect the ion pairs as a current with the application of a voltage between two electrodes. The measured current is proportional to the aplied voltage and the amount of radiation. At a lower voltages from 50 to 300 V, only the primary ion pairs formed by the initial radiation are collected. Ionization chambers operate in this region. The detector is a cylindrical chamber with a central wire filled with air or different gases. These detectors are primarily used for measuring high intensity radiation. Dose calibrators and pocket dosimeters are the common ionization chambers used in nuclear medicine. The dose calibrator is one of the most essential instrument in nuclear medicine for measuring the activity of radionuclides and radiopharmaceuticals. It must be regularly checked for constancy, accuracy, linearity and geometry. At higher voltages from 1000 to 1200 V, the current becomes identical regardless of how many ion pairs are produced by the incident radiation. Geiger Müller counters operate in this region. Geiger Müller counters are used to monitor the radiation level in different work areas and they are called area monitors or survey meters. They are more sensitive than ionization chambers but they cannot discriminate between energies. They are almost 100% efficient for counting alpha and beta particles but have only 1 to 2% efficiency for counting gamma and x rays.

3 Scintillation detectors consist of scintilator emitting flashes of light after absorbing gamma or x radiation. The light photons produced are then converted to an electrical pulse by means of a photomultiplier tube. The pulse is amplified by a linear amplifier, sorted by a pulse height analyzer and then registred as a count. Different solid or liquid scintillators are used for different types of radiation. In nuclear medicine, sodium iodide solid crystals with a trace of thallium NaI(Tl) are used for gamma and x ray detection. The basic solid scintillation counter consists of na NaI(Tl) crystal or detector, a photomultiplier (PM) tube, a preamplifier, a linear amplifier, a pulse height analyzer (PHA) and a recording device. NaI(Tl) crystals are hermetically sealed in aluminium containers. They are fragile and must be handled with care. Room temperature should not be changed suddenly because of possibility of cracks in the crystal. In well counters and thyriod probes smaller and thicker crystals are used, whereas larger and thinner crystals are employed in imaging devices like gamma cameras. PM tube consists of a light sensitive photocathode facing the crystal, series of dynodes in the middle and an anode at the other end all enclosed in a vacuum glass tube. A high voltage about 1000 V is applied between the photocathode and the anode of the PM tube. The electron pulse reaching the anode is delivered to the preamplifier. The amplitude of the pulse is proportional to the number of light photons received by the photocathode and in turn to the energy of gamma ray photon absorbed in the crystal. The applied voltage must be very stable. A linear amplifier amplifies further the signal from the preamplifier and delivers it to the pulse height analyzer for analysis of its amplitude. A pulse height analyzer is a device that selects for counting only those pulses falling within preselect voltage intervals and rejects all others. Desired pulses are ultimately delivered to the recording devices such as scalers, computers, films and so on. In the output of scintillation counter a distribution of pulse heights will be obtained depicting a spectrum of gamma ray energies. In an ideal situation each gamma ray would be seen as a line on the gamma ray spectrum. In reality, the photopeak is broder, which is due to various statistical fluctuations in the process of forming the pulses. When gamma rays interact with the scintillation crystal by means of Compton scattering, the Compton electrons of variable energies result in a pulse height smaller than that of photopeak. Thus the gamma ray spectrum will show a continuum of pulses corresponding to Compton electron energies between zero and photopeak, so called Compton continuous spectrum. The relative hights of the photopeak and the Compton scattering depend on the photon energy as well as the size of the crystal. There are several basic characteristics of counters important from the point of view of using them in nuclear medicine procedures. Background of the detector means registered count rate without presence of any measured specimen. It is caused namely by cosmic radiation, natural radioactivity, radioactivity of building material and of the material the detecting system consists of. It can be minimized by shielding detector in lead cover, by using special material for its construction or by means of pulse height analyzer to exclude inappropriate radiation energy from detection. The energy resolution simply means the width or the sharpness of the photopeak or the ability to discriminate the gamma ray photons of similar energies. The detection efficiency is given by the observed count rate divided by the disintegration rate of a radioactive sample. It depends on the type and energy of the detected radiation, size and thickness of the detector crystal and geometric efficiency of the measuring. The dead time is the time period during which the counter is insensitive for the radiation detection. This time

4 enclosed time needed to process a radiation event starting from interaction in the crystal all the way up to forming and recording pulse. Dead time for Geiger Müller detectors is from 100 to 500 microseconds, for NaI(Tl) crystal from 0,5 to 5 microseconds. Dead time loss of counts is a serious problem at high count rates. Either count rates must be lowered or corrections must be made to the observed count rates. Scintillation detectors can be used as a part of both nonimaging and imaging devices. From the nonimaging devices, scintillation well counters and thyroid probes are used. The gamma well counter consists of a scintillation detector with a hole in the center, for a sample to be placed inside for increasing the geometric efficiency and hence the counting efficiency of the counter, and other associated electronics. Well counters are used namely for in vitro measurements of different samples. They are usually available with automatic sample changers and are mostly programmable with computers. Their major advantage is high detection efficiency which is from 50% to 70% for 140 kev gamma photons. The thyroid probe is a scintillation counter used for measuring radioacitivity above the thyroid gland to assess the uptake of 131I after its oral administration. In contrast to well counter the thyroid probe must be equipped with collimator, which limits the field of view. This is a cylindrical barrel made of lead and it covers all the detector including PM tube. It prevents the gamma radiations from other organs to reach the detector. Radionuclide imaging devices. Radionuclide imaging is based on the ability to detect electromagnetic radiation emitted from an injected radioactive tracer that has been taken up by the organ to be studied. The electromagnetic radiation used is in the form of gamma rays or x rays. The radiation absorbed by the detector is used to generate a digital image by the computer, which is then interpreted by the physician. This imaging device is called scintillation camera or gamma camera. It also employs sodium iodide scintillation detector and the associated electronics like nonimaging systems do. The most frequently used scintillation camera is of Anger type (Hal O. Anger invented it in the 1960), it means it has a large scintillation crystal which makes possible to detect radiation from the entire field of view simultaneously and therefore it is capable of recording dynamic as well as static images of the area of interest in the patient. Many sophisticated improvements have been made of the cameras over the years, but the basic principles of the operation are essentially the same. Like nonimaging probe also scintillation camera consists of a collimator, a scintillation crystal, PM tubes, a preamplifier, an amplifier, a pulse height analyser and recording or display devices. In addition it must have an X,Y positioning circuit to localize the point of interaction of gamma ray with the crystal. The operation of a camera is performed by a computer built in it and is very convenient for the staff. The detector head (collimator, scintillation crystal, PM tubes and amplifiers) is mounted on a stand called gantry, which moves the head to the appropriate position for patient imaging. Detectors have usually large (about 50 cm in diameter) circular or rectangular NaI(Tl) crystal with about 1 cm thickness. In front of the crystal, a collimator is attached to limit the field of view so that gamma radiations from outside the field of view cannot reach the crystal. Collimator is usually a plate of lead with many holes. Most frequently used are parallel hole colimators with holes parallel to each others and perpendicular to the detector face. They are of different types according to energy of radionuclides used for imaging and according to their spatial resolution. Thus we can distinguish high resolution, high sensitivity and all purpose (with compromise parameters) types or low, medium and high energy types. The spatial resolution of the parallel hole collimators decreases with the increasing distance of the object from the front of the collimator but the sensitivity is the same. That is why every data collection must be performed with the minimal space between collimator forhead and the patient body surface. The collimator of conical shape with one up to three holes on the top is called pinhole and is used for imaging of small and near to the surface lying organs, such as a thyroid gland, tight join or infant kidneys. It has very good resolution but very poor sensitivity. Nowadays also collimators with special converging holes called fan beam are made for small organs imaging, such as a brain. Also collimators with diverging holes can be used, namely in

5 cameras with small crystal to make possible imaging of large organs. Collimators designed for higher energy are thicker with thicker septa between holes to prevent penetration of photons through them. Gamma cameras have many photomultiplier tubes (up to 90) mounted to the back of the crystal with optical grease. They are used to be of hexagonal shape and the output from each is used to define the X and Y coordinate of the point of interaction of the gamma ray in the crystal by the use of an X,Y positioning circuit. The X and Y pulses are than projected on a cathode ray tube or oscilloscope to create image or can be stored in the computer in a square matrix for further processing. The larger the number of PM tubes, the better the spatial resolution on the image. The use of digital computers in nuclear medicine has considerably increased and today all nuclear medicine studies are being analyzed by the computers. Data from a gamma camera must be digitized by the analog todigital converter. The computer memory approximates the area of the detector as a square matrix from 32x32 up to 1024x1024 size. Each element of this matrix is called pixel and corresponds to a specific X and Y location in the detector. The number in each pixel corresponds to the number of pulses detected in this specific location of the crystal. In this manner of storing data in computer memory, called frame mode (the most common mode in nuclear medicine), we must preset the matrix size, the number of images (frames) in the study, and the time of collection of data per frame or total counts to be collected per frame. Computers are very important part of imaging devices, current cameras cannot operate without it, namely ECT is impossible without computers. The basic function of computers during image construction is to correct and maintain cameras performance parameters, such as high voltage in the PM tubes and photopeak setting in pulseheight analyzer. Another improvement of image quality is achieved by images smoothing and filtering, mathematical operation with images, background subtraction, creation of parametric and tomographic images, regions of interest (ROI) creation, dynamic curves creation and their mathematical processing with computing quantitative data of physiological processes, by using of interpolation to reduce digital raster effect on matrix image, smoothing images by means of temporal and spatial filters or by color coding according to number of counts in each pixel. Very important camera parameter, field of view uniformity, can be effectively improved by means of computers. Detector uniformity means a uniform response throughout the field of view. Even properly tuned and adjusted gamma cameras produce nonuniform images with count density variation of up to 10%. There are several possibilities of using computer for nonuniformity correction. These are: correction of number of counts in each pixel of image matrix to an average value, setting of the own photopeak for each pixel in image matrix, different gain of high voltage for each photomultiplier tube and spatial distortion correction. To ensure a high quality of images, quality control tests must be performed routinely on these devices. The most common tests are the positioning of the photopeak, field of view uniformity and background check, which must be performed daily. The spatial resolution of the camera should be checked weekly. This type of Anger gamma camera provide two dimensional images and that is why it is also called planar camera. According to the type of metabolic processes and the type of data recording we can distinguish two basal type of collecting data. During static acquisition ofimages we can evaluate distribution of radiopharmaceuticals in the organs, which does not change in time. To achieve good information density of images, we must preset proper acquisition time according to registered count rate or we must preset needed number of collected counts. Except acquiring images with the same size and shape as detector has we can also acquire a so called whole body images. In this type of data collection patient on the examining table passes under or over the detector and the image of radioactivity distribution in the whole body can be obtained. The velocity of the patient movement is again done by the registered count rate to ensure good image information density. Since radioactivity distribution does not change significantly, we can collect data for a longer time period and thus high resolution collimator should be used. The second type of collecting data, during which we can evaluate differently fast metabolic or functioning

6 processes in the body (blood flow), is called dynamic study. According to velocity of assessing processes we must preset number of images (frames) and duration of one image. Because distribution of radiopharmaceuticals changes rapidly and duration of one frame acquisition is limited (up to split second), high sensitivity collimators should be used to achieve sufficient image information density. A special kind of collecting data is gated or synchronized method. It is used for acquiring a periodical organ activity, such as heart mechanical function. In this example ECG signal is used for synchonization and by this way to collect data with very high temporal resolution is possible. In principle, computer divides each R to R interval into several intervals, for example twenty. Each interval has its own number, in this example from one to twenty, is presented as a frame in the computer memory and its duration is of tens miliseconds. After R interval detection by computer, data are stored into the frame number one, after the time of one frame duration is over data are stored into frame number two and so forth till the further R wave is detected. From this point all the process is repeated and by this way several hundreds heart cycles are summed into one, so image information density is sufficient for further processing. On planar images, the third dimension of displayed objects is obscured by superimposition of data. Therefore tomographic devices have been developed. Primarily, so called longitudinal tomography was used, which display sharply only one slice parallel to the face of detector. The images was not of good quality and thus nowadays only transversal tomography is used. In this technique multiple views are obtained at many agles around the long axis of the patient and images in the slices perpendicular to the detector face, transversal slices (each element of this slice is in the form of cube in the computer memory and is called voxel), are constructed by the computer. Since in nuclear medicine the source of radioactivity is in the patient body, and the gamma photons or x rays are emitted from it, we talk about emission computed tomography (ECT). This technique improves images quality and by this way physician interpretation by means of a better contrast between target organ and background, better topography of pathological sites and higher detection efficiency of pathological lesions, which is done by better spatial resolution predominantly. According to the type of radionuclide used we can distinguish two types of ECT. Single photon emission computed tomography (SPECT), which uses gamma emitting radionuclides (most of today used) and positron emission tomography (PET), which uses positron emitting radionuclides. The SPECT cameras have in principle the same detector as planar cameras have, but the gantry makes possible to rotate the detector around the patient. The minimal angle of rotation is 180 degrees at small angle incremets. The path around the patient can be circular or eliptical and current cameras have the possibility of so called body contouring pathway to minimize the distance between detector forhead and patient body surface. The data are stored in the computer for further processing and image reconstruction in the form of slices in three perpendicullar planes of transverse, coronal (frontal) and sagittal direction. There are different designs of SPECT cameras with one, two or three detectors. SPECT image is a type of static image, so sufficient image information density must be achieved. That is why about 20 to 30 minutes of time is needed to collect sufficient data. We must preset angle of rotation, angular step and time for one frame acquisition according to count rate detected. There are several mathematical algorithms to reconstruct images from acquired data. The most frequently used is so called filtered back projection. Filters are mathematical functions, which increase image quality and suppress artefacts. Various types of filters are commercially available in the form of software packages. Another methods for image reconstruction are Fourier transform or iterative methods. PET is based on the detection of coincidence of the two annihilation gamma photons that are emitted after interaction of positron with electron in the patient body. These two photons, which arise at the same time, have the same energy of 511 kev and are emitted in the angle of 180 degrees, are detected by the two detectors connected in coincidence. Data collected around the body axis are then used for image reconstruction. In current PET systems, many detectors (hundreds to thousands) are arranged in circular rings. The number of rings can be up to sixteen and are arranged in arrays. Each detector is connected to the opposite detector in the same ring by a coincidence circuit. The field of view is defined by the width of the array of detectors. For this device no collimator is needed. Data are collected in the computer in the frame mode. Image reconstruction is accomplished by the same method as in the SPECT does.

Radionuclide Imaging MII Detection of Nuclear Emission

Radionuclide Imaging MII Detection of Nuclear Emission Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor

More information

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: NUCLEAR MEDICINE Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Introduction to Medical Imaging: Physics, Engineering and Clinical Applications, by Nadine Barrie Smith

More information

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad

More information

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -

More information

Physics in Nuclear Medicine

Physics in Nuclear Medicine SIMON R. CHERRY, PH.D. Professor Department of Biomedical Engineering University of California-Davis Davis, California JAMES A. SORENSON, PH.D. Emeritus Professor of Medical Physics University of Wisconsin-Madison

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

A. I, II, and III B. I C. I and II D. II and III E. I and III

A. I, II, and III B. I C. I and II D. II and III E. I and III BioE 1330 - Review Chapters 7, 8, and 9 (Nuclear Medicine) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in

More information

Chapter Seven (Nuclear Detectors)

Chapter Seven (Nuclear Detectors) Al-Mustansiriyah University College of Science Physics Department Fourth Grade Nuclear Physics Dr. Ali A. Ridha Chapter Seven (Nuclear Detectors) Ionizing radiation is rarely detected directly. Instead,

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system

More information

Radionuclide Imaging MII Positron Emission Tomography (PET)

Radionuclide Imaging MII Positron Emission Tomography (PET) Radionuclide Imaging MII 3073 Positron Emission Tomography (PET) Positron (β + ) emission Positron is an electron with positive charge. Positron-emitting radionuclides are most commonly produced in cyclotron

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

Outline Chapter 14 Nuclear Medicine

Outline Chapter 14 Nuclear Medicine Outline Chapter 14 uclear Medicine Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther Introduction Detectors for nuclear

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 2 Radioisotopes Techniques Lecture - 3 GM Counting and

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links NM - introduction Relies on EMISSION of photons from body (versus transmission of photons through

More information

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Radiation Detection. 15 th Annual OSC Readiness Training Program. Radiation Detection 15 th Annual OSC Readiness Training Program www.oscreadiness.org GM Detectors 15 th Annual OSC Readiness Training Program www.oscreadiness.org 1 A closer look 15 th Annual OSC Readiness

More information

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or Let's remind... What is nuclear medicine? Nuclear medicine can be broadly divided into two branches "in vitro" and "in vivo" procedures. There are numerous radioisotopic "in vitro" procedures for genotyping

More information

Scintillation Detector

Scintillation Detector Scintillation Detector Introduction The detection of ionizing radiation by the scintillation light produced in certain materials is one of the oldest techniques on record. In Geiger and Marsden s famous

More information

Basic physics Questions

Basic physics Questions Chapter1 Basic physics Questions S. Ilyas 1. Which of the following statements regarding protons are correct? a. They have a negative charge b. They are equal to the number of electrons in a non-ionized

More information

Radioisotopes and PET

Radioisotopes and PET Radioisotopes and PET 1 Radioisotopes Elements are defined by their number of protons, but there is some variation in the number of neutrons. Atoms resulting from this variation are called isotopes. Consider

More information

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must

More information

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of

More information

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx 1 Candidates should be able to : HISTORY Describe the nature of X-rays. Describe in simple terms how X-rays are produced. X-rays were discovered by Wilhelm Röntgen in 1865, when he found that a fluorescent

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 26 Radiation Detection & Measurement II Spiritual Thought 2 I would not hold the position in the Church I hold today had I not followed

More information

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado Experiment 6 1 Introduction The Compton Effect Physics 2150 Experiment No. 6 University of Colorado In some situations, electromagnetic waves can act like particles, carrying energy and momentum, which

More information

Mitigation of External Radiation Exposures

Mitigation of External Radiation Exposures Mitigation of External Radiation Exposures The three (3) major principles to assist with maintaining doses ALARA are :- 1) Time Minimizing the time of exposure directly reduces radiation dose. 2) Distance

More information

Analysis of γ spectrum

Analysis of γ spectrum IFM The Department of Physics, Chemistry and Biology LAB 26 Analysis of γ spectrum NAME PERSONAL NUMBER DATE APPROVED I. OBJECTIVES - To understand features of gamma spectrum and recall basic knowledge

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -Monitoring

More information

Medical Physics. Nuclear Medicine Principles and Applications

Medical Physics. Nuclear Medicine Principles and Applications Medical Physics Nuclear Medicine Principles and Applications Dr Roger Fulton Department of PET & Nuclear Medicine Royal Prince Alfred Hospital Sydney Email: rfulton@mail.usyd.edu.au Lectures: http://www-personal.usyd.edu.au/~rfulton/medical_physics

More information

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung Contents Marcel MiGLiERiNi Nuclear Medicine, Radiology and Their Metrological Aspects. Radiation in Medicine. Dosimetry 4. Diagnostics & Therapy 5. Accelerators in Medicine 6. Therapy Planning 7. Nuclear

More information

Radiation Dose, Biology & Risk

Radiation Dose, Biology & Risk ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.

More information

electrons out of, or ionize, material in their paths as they pass. Such radiation is known as

electrons out of, or ionize, material in their paths as they pass. Such radiation is known as Detecting radiation It is always possible to detect charged particles moving through matter because they rip electrons out of, or ionize, material in their paths as they pass. Such radiation is known as

More information

Compton Camera. Compton Camera

Compton Camera. Compton Camera Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract

More information

Detection and measurement of gamma-radiation by gammaspectroscopy

Detection and measurement of gamma-radiation by gammaspectroscopy Detection and measurement of gamma-radiation by gammaspectroscopy Gamma-radiation is electromagnetic radiation having speed equal to the light in vacuum. As reaching a matter it interact with the different

More information

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide Nuclear Medicine Imaging Systems: The Scintillation Camera List of Nuclear Medicine Radionuclides Tc99m 140.5 kev 6.03 hours I-131 364, 637 kev 8.06 days I-123 159 kev 13.0 hours I-125 35 kev 60.2 days

More information

Radiation Protection & Radiation Therapy

Radiation Protection & Radiation Therapy Radiation Protection & Radiation Therapy For Medical Students Professor of Medical Physics Radiation Units Activity Number disintegrations per second (Curie, Becquerel) Exposure (Roentgen, C/kg) Absorbed

More information

Positron-Electron Annihilation

Positron-Electron Annihilation Positron-Electron Annihilation Carl Akerlof September 13, 008 1. Introduction This experiment attempts to explore several features of positron-electron annihilation. One of the attractive aspects of e

More information

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS TSOKOS OPTION I-2 MEDICAL IMAGING Reading Activity Answers IB Assessment Statements Option I-2, Medical Imaging: X-Rays I.2.1. I.2.2. I.2.3. Define

More information

www.aask24.com www.aask24.com www.aask24.com P=Positron E= Emission T=Tomography Positron emission or beta plus decay (+ ) is a particular type of radioactive decay, in which a proton inside a radionuclide

More information

Bases of radioisotope diagnostic methods

Bases of radioisotope diagnostic methods Medical, pharmaceutical applications of radioisotopes Bases of radioisotope diagnostic methods Dr. István Voszka Basis of application: radioisotopes have identical behavior in the organism to corresponding

More information

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics Nuclear Physics Laboratory Gamma spectroscopy with scintillation detectors M. Makek Faculty of Science Department of Physics Zagreb, 2015 1 1 Introduction The goal of this excercise is to familiarize with

More information

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics

More information

Radioactivity. The Nobel Prize in Physics 1903 for their work on radioactivity. Henri Becquerel Pierre Curie Marie Curie

Radioactivity. The Nobel Prize in Physics 1903 for their work on radioactivity. Henri Becquerel Pierre Curie Marie Curie Radioactivity Toward the end of the 19 th century, minerals were found that would darken a photographic plate even in the absence of light. This phenomenon is now called radioactivity. Marie and Pierre

More information

Waves & Radiation exam questions

Waves & Radiation exam questions National 5 Physics Waves & Radiation exam questions these questions have been collated from previous Standard Grade (Credit) and Intermediate 2 exams Thurso High School 1. A mountain climber carries a

More information

Radioactivity and Ionizing Radiation

Radioactivity and Ionizing Radiation Radioactivity and Ionizing Radiation QuarkNet summer workshop June 24-28, 2013 1 Recent History Most natural phenomena can be explained by a small number of simple rules. You can determine what these rules

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

Gamma Ray Spectroscopy

Gamma Ray Spectroscopy Gamma Ray Spectroscopy Uzair Latif, Imran Younus Department of Physics Lahore University of Management Sciences November 4, 2014 1 Objectives 1. To acquaint the students with some of the basic techniques

More information

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University Notes to accompany Lab demonstrations by Barry Diacon, Technician, Department of

More information

Absolute activity measurement

Absolute activity measurement Absolute activity measurement Gábor Veres, Sándor Lökös Eötvös University, Department of Atomic Physics January 12, 2016 Financed from the financial support ELTE won from the Higher Education Restructuring

More information

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na Ellen Simmons 1 Contents Introduction Review of the Types of Radiation Charged Particle Radiation Detection Review of Semiconductor

More information

Nuclear Physics and Astrophysics

Nuclear Physics and Astrophysics Nuclear Physics and Astrophysics PHY-30 Dr. E. Rizvi Lecture 4 - Detectors Binding Energy Nuclear mass MN less than sum of nucleon masses Shows nucleus is a bound (lower energy) state for this configuration

More information

INAYA MEDICAL COLLEGE (IMC) RAD LECTURE 1 RADIATION PHYSICS DR. MOHAMMED MOSTAFA EMAM

INAYA MEDICAL COLLEGE (IMC) RAD LECTURE 1 RADIATION PHYSICS DR. MOHAMMED MOSTAFA EMAM INAYA MEDICAL COLLEGE (IMC) RAD 232 - LECTURE 1 RADIATION PHYSICS DR. MOHAMMED MOSTAFA EMAM Radiation: It is defined as the process by which energy is emitted from a source and propagated through the surrounding

More information

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,

More information

Gamma ray coincidence and angular correlation

Gamma ray coincidence and angular correlation University of Cape Town Department of Physics Course III laboratory Gamma ray coincidence and angular correlation Introduction Medical imaging based on positron emission tomography (PET) continues to have

More information

INAYA MEDICAL COLLEGE (IMC) RAD LECTURE 1 RADIATION PHYSICS DR. MOHAMMED MOSTAFA EMAM

INAYA MEDICAL COLLEGE (IMC) RAD LECTURE 1 RADIATION PHYSICS DR. MOHAMMED MOSTAFA EMAM INAYA MEDICAL COLLEGE (IMC) RAD 232 - LECTURE 1 RADIATION PHYSICS DR. MOHAMMED MOSTAFA EMAM LECTURES & CLASS ACTIVITIES https://inayacollegedrmohammedemam.wordpress.com/ Password: drmohammedemam 16-02-2015

More information

08 - Miscellaneous and historical detectors

08 - Miscellaneous and historical detectors 08 - Miscellaneous and historical detectors Jaroslav Adam Czech Technical University in Prague Version 2 Jaroslav Adam (CTU, Prague) DPD_08, Miscellaneous and historical detectors Version 2 1 / 25 Streamer

More information

Radioactivity INTRODUCTION. Natural Radiation in the Background. Radioactive Decay

Radioactivity INTRODUCTION. Natural Radiation in the Background. Radioactive Decay Radioactivity INTRODUCTION The most common form of radiation is the electromagnetic wave. These waves include low energy radio waves, microwaves, visible light, x-rays, and high-energy gamma rays. Electromagnetic

More information

Physics 1000 Half Life Lab

Physics 1000 Half Life Lab Physics 1000 Half Life Lab Determination of Half-Life with a Geiger-Müller Counter Object: Apparatus: To understand the concept of half-life; to become familiar with the use of a Geiger-Müller counter;

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 10 pages Written test, 12 December 2012 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: None. Pocket calculator not allowed "Weighting":

More information

Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA ramsey

Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA   ramsey SPECIAL FEATURE: MEDICAL PHYSICS www.iop.org/journals/physed Nuclear medicine Ramsey D Badawi Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA E-mail: ramsey badawi@dfci.harvard.edu

More information

Quality Assurance. Purity control. Polycrystalline Ingots

Quality Assurance. Purity control. Polycrystalline Ingots Quality Assurance Purity control Polycrystalline Ingots 1 Gamma Spectrometry Nuclide Identification Detection of Impurity Traces 1.1 Nuclides Notation: Atomic Mass Atomic Number Element Neutron Atomic

More information

Gamma Spectroscopy. References: Objectives:

Gamma Spectroscopy. References: Objectives: Gamma Spectroscopy References: G.F. Knoll, Radiation Detection and Measurement (John Wiley & Sons, New York, 2000) W. R. Leo, Techniques for Nuclear and Particle Physics Experiments: A How-to Approach,

More information

Rad T 290 Worksheet 2

Rad T 290 Worksheet 2 Class: Date: Rad T 290 Worksheet 2 1. Projectile electrons travel from a. anode to cathode. c. target to patient. b. cathode to anode. d. inner shell to outer shell. 2. At the target, the projectile electrons

More information

Physics 23 Fall 1989 Lab 5 - The Interaction of Gamma Rays with Matter

Physics 23 Fall 1989 Lab 5 - The Interaction of Gamma Rays with Matter Physics 23 Fall 1989 Lab 5 - The Interaction of Gamma Rays with Matter Theory The nuclei of radioactive atoms spontaneously decay in three ways known as alpha, beta, and gamma decay. Alpha decay occurs

More information

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras?

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras? BioE 1330 - Exam 2 11/13/2018 Answer Sheet - Correct answer is A for all questions 1. Unlike CT, in nuclear medicine A. Bremsstrahlung is not used to produce high-energy photons. B. signal can be increased

More information

Chapter 16 Basic Precautions

Chapter 16 Basic Precautions Chapter 16 Basic Precautions 16.1 Basic Principles of Radiation Protection The four basic methods used to control radiation exposure are time, distance, shielding, and contamination control. The first

More information

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra 22.101 Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra References: W. E. Meyerhof, Elements of Nuclear Physics (McGraw-Hill, New York, 1967),

More information

Lab NUC. Determination of Half-Life with a Geiger-Müller Counter

Lab NUC. Determination of Half-Life with a Geiger-Müller Counter Lab NUC Determination of Half-Life with a Geiger-Müller Counter Object: Apparatus: To understand the concept of half-life; to become familiar with the use of a Geiger-Müller counter; to determine the half-lives

More information

ABSORPTION OF BETA AND GAMMA RADIATION

ABSORPTION OF BETA AND GAMMA RADIATION ABSORPTION OF BETA AND GAMMA RADIATION The purpose of this experiment is to understand the interaction of radiation and matter, and the application to radiation detection and shielding Apparatus: 137 Cs

More information

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia 7 th Workshop on European Collaboration for Higher Education and Research in Nuclear Engineering & Radiological Protection Bruxelles, Belgique 30 May - 1 June 2011 Activities at the Laboratory of the Nuclear

More information

CHAPTER 5. Department of Medical Physics, University of the Free State, Bloemfontein, South Africa

CHAPTER 5. Department of Medical Physics, University of the Free State, Bloemfontein, South Africa CHAPTE 5 STATISTICS FO ADIATIO MEASUEMET M.G. LÖTTE Department of Medical Physics, University of the Free State, Bloemfontein, South Africa 5.1. SOUCES OF EO I UCLEA MEDICIE MEASUEMET Measurement errors

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:

More information

SCINTILLATION DETECTORS AND PM TUBES

SCINTILLATION DETECTORS AND PM TUBES SCINTILLATION DETECTORS AND PM TUBES General Characteristics Introduction Luminescence Light emission without heat generation Scintillation Luminescence by radiation Scintillation detector Radiation detector

More information

Structure of Biological Materials

Structure of Biological Materials ELEC ENG 3BA3: Structure of Biological Materials Notes for Lecture #19 Monday, November 22, 2010 6.5 Nuclear medicine imaging Nuclear imaging produces images of the distribution of radiopharmaceuticals

More information

PHYS 3446 Lecture #12

PHYS 3446 Lecture #12 PHYS 3446 Lecture #12 Wednesday, Oct. 18, 2006 Dr. 1. Particle Detection Ionization Detectors MWPC Scintillation Counters Time of Flight 1 Announcements Next LPCC Workshop Preparation work Each group to

More information

2. Which of the following statements help(s) to explain why gas can fill the vessel containing it completely while liquid cannot?

2. Which of the following statements help(s) to explain why gas can fill the vessel containing it completely while liquid cannot? Name: Class: ( ) There are 30 questions. Time Allowed: 45 min 1. Kinetic theory explains the behaviour of a substance in terms of the behaviour of the molecules in it. Which of the following is/are the

More information

Year 12 Notes Radioactivity 1/5

Year 12 Notes Radioactivity 1/5 Year Notes Radioactivity /5 Radioactivity Stable and Unstable Nuclei Radioactivity is the spontaneous disintegration of certain nuclei, a random process in which particles and/or high-energy photons are

More information

Chapter 37 Early Quantum Theory and Models of the Atom. Copyright 2009 Pearson Education, Inc.

Chapter 37 Early Quantum Theory and Models of the Atom. Copyright 2009 Pearson Education, Inc. Chapter 37 Early Quantum Theory and Models of the Atom Planck s Quantum Hypothesis; Blackbody Radiation Photon Theory of Light and the Photoelectric Effect Energy, Mass, and Momentum of a Photon Compton

More information

The Nuclear Imaging Uncertainty Principle. Do Nuclear Cameras Really Work? Richard M. Fleming

The Nuclear Imaging Uncertainty Principle. Do Nuclear Cameras Really Work? Richard M. Fleming The Nuclear Imaging Uncertainty Principle. Do Nuclear Cameras Really Work? Richard M. Fleming Address correspondence to: Dr. R.M. Fleming 1697 Lone Oak Trail Reno, NV 89523 Rmfmd7@hotmail.com 12 October

More information

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440 Jazan University College of Science Physics Department جاهعة جازان كلية العل وم قسن الفيزياء Lab Manual Nuclear Physics (2) 462 Phys 8 th Level Academic Year: 1439/1440 1 Contents No. Name of the Experiment

More information

SiPM & Plastic Scintillator

SiPM & Plastic Scintillator SiPM & Plastic Scintillator Silicon photomultiplier coupled to plastic scintillator Lodovico Lappetito SiPM_PlasticScint_ENG - 28/04/2016 Pag. 1 Table of contents Introduction... 3 Plastic Scintillators...

More information

11 Gamma Ray Energy and Absorption

11 Gamma Ray Energy and Absorption 11 Gamma Ray Energy and Absorption Before starting this laboratory, we must review the physiological effects and the proper use of the radioactive samples you will be using during the experiment. Physiological

More information

Particle Energy Loss in Matter

Particle Energy Loss in Matter Particle Energy Loss in Matter Charged particles, except electrons, loose energy when passing through material via atomic excitation and ionization These are protons, pions, muons, The energy loss can

More information

This Week. 3/23/2017 Physics 214 Summer

This Week. 3/23/2017 Physics 214 Summer This Week Atoms and nuclei What are we made of? The periodic table Why does it stop? How were the elements made? Radioactive decay Useful but can be toxic Discovery of X Rays: Cathode Rays and TV sets

More information

Chapter NP-4. Nuclear Physics. Particle Behavior/ Gamma Interactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 IONIZATION

Chapter NP-4. Nuclear Physics. Particle Behavior/ Gamma Interactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 IONIZATION Chapter NP-4 Nuclear Physics Particle Behavior/ Gamma Interactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 IONIZATION 2.0 ALPHA PARTICLE INTERACTIONS 3.0 BETA INTERACTIONS 4.0 GAMMA INTERACTIONS

More information

THE COMPTON EFFECT Last Revised: January 5, 2007

THE COMPTON EFFECT Last Revised: January 5, 2007 B2-1 THE COMPTON EFFECT Last Revised: January 5, 2007 QUESTION TO BE INVESTIGATED: How does the energy of a scattered photon change after an interaction with an electron? INTRODUCTION: When a photon is

More information

General Physics (PHY 2140)

General Physics (PHY 2140) General Physics (PHY 2140) Lecture 19 Modern Physics Nuclear Physics Nuclear Reactions Medical Applications Radiation Detectors Chapter 29 http://www.physics.wayne.edu/~alan/2140website/main.htm 1 Lightning

More information

General Physics (PHY 2140)

General Physics (PHY 2140) General Physics (PHY 2140) Lightning Review Lecture 19 Modern Physics Nuclear Physics Nuclear Reactions Medical Applications Radiation Detectors Chapter 29 http://www.physics.wayne.edu/~alan/2140website/main.htm

More information

NATO HANDBOOK ON THE MEDICAL ASPECTS OF NBC DEFENSIVE OPERATIONS AMedP-6(B) PART I - NUCLEAR ANNEX A RADIATION DETECTION AND MEASUREMENT

NATO HANDBOOK ON THE MEDICAL ASPECTS OF NBC DEFENSIVE OPERATIONS AMedP-6(B) PART I - NUCLEAR ANNEX A RADIATION DETECTION AND MEASUREMENT NATO HANDBOOK ON THE MEDICAL ASPECTS OF NBC DEFENSIVE OPERATIONS AMedP-6(B) PART I - NUCLEAR RADIATION DETECTION AND MEASUREMENT 1 FEBRUARY 1996 NATO UNCLASSIFIED A ORIGINAL (Reverse Blank) TABLE OF CONTENTS

More information

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011 Photon Instrumentation First Mexican Particle Accelerator School Guanajuato Oct 6, 2011 Outline The Electromagnetic Spectrum Photon Detection Interaction of Photons with Matter Photoelectric Effect Compton

More information

Copyright 2008, University of Chicago, Department of Physics. Gamma Cross-sections. NaI crystal (~2" dia) mounted on photo-multiplier tube

Copyright 2008, University of Chicago, Department of Physics. Gamma Cross-sections. NaI crystal (~2 dia) mounted on photo-multiplier tube Gamma Cross-sections 1. Goal We wish to measure absorption cross-sections for γ-rays for a range of gamma energies and absorber atomic number. 2. Equipment Pulse height analyzer Oscilloscope NaI crystal

More information

Interaction of charged particles and photons with matter

Interaction of charged particles and photons with matter Interaction of charged particles and photons with matter Robert Miyaoka, Ph.D. Old Fisheries Center, Room 200 rmiyaoka@u.washington.edu Passage of radiation through matter depends on Type of radiation

More information

Chapter Four (Interaction of Radiation with Matter)

Chapter Four (Interaction of Radiation with Matter) Al-Mustansiriyah University College of Science Physics Department Fourth Grade Nuclear Physics Dr. Ali A. Ridha Chapter Four (Interaction of Radiation with Matter) Different types of radiation interact

More information

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI)

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI) Radioation Resolution and Sensitivity Nuclear Imaging PET + SPECT Radioactive Decay (EC,Ɣ), (β -,Ɣ), (I.T.,Ɣ) β + Projection imaging collimator needed one angular view Projection imaging coincidence imaging,

More information

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Marcus Ressner, PhD, Medical Radiation Physicist, Linköping University Hospital Content What is Nuclear medicine? Basic principles of Functional

More information

Theoretical questions for the final exam ED 2012.

Theoretical questions for the final exam ED 2012. Theoretical questions for the final exam ED 2012. 1. Radiation a) Properties and types of radiation b) Physical parameters of radiation 2. Law of attenuation of radiation a) Experimental interpretation

More information

Figure 1. Decay Scheme for 60Co

Figure 1. Decay Scheme for 60Co Department of Physics The University of Hong Kong PHYS3851 Atomic and Nuclear Physics PHYS3851- Laboratory Manual A. AIMS 1. To learn the coincidence technique to study the gamma decay of 60 Co by using

More information