Compton Camera. Compton Camera

Size: px
Start display at page:

Download "Compton Camera. Compton Camera"

Transcription

1 Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract a source image without the use of collimators or masks. Compton cameras have been built and tested for diverse applications including Compton telescopes for gamma-ray astronomy, as replacements for lead-collimated medical imagers, for industrial imaging, and for radioactive waste management. The Compton camera normally consists of two detectors. The first detector is designed to maximize the likelihood of Compton scattering and the second is designed to absorb the energy remaining after the Compton scatter. The useful range extends from kev to MeV. In contrast, the conventional camera is limited to at most a few hundred kev. Furthermore, it could obtain three dimensional images from a fixed position on one side of the source without the need for tomography. Compton Effect Compton Camera The Compton camera principle As illustrated in fig., a gamma photon emitted from point A with a known energy E, Compton scattering within the first detector at point B and undergoing some further interaction in the second detector at point C. If we assume that the coordinates of points B and C and the energy deposited in the first and second interaction are E and E. Then we can get E = E +E, and the angle θ between the incident direction and the direction of the scattered gamma ray is given by: cos( θ ) = + mc [], E E

2 where m c is the rest mass of electron. The direction of the incident gamma ray is confined to the cone shown in Fig. with its axis along the direction of the scattered gamma ray and its apex at the position of the interaction with the front plane detector. A large number of scattering events from a point source of gamma rays will define multiple cones which intersect at the location of the source. Generally the incident gamma ray energy E, is known or can be determined from peaks in the singles energy spectrum from each detector or in the coincident summed spectrum. A window is then set on the sum peak E + E = E. From events which fall within the window, the image is reconstructed on a plane or on the surface of a sphere centered on the front detector. The intersection of the cone with the image plane, as shown in fig., produces a curve which traces a conic section, typically an ellipse but parabolic and hyperbolic sections are possible for large angle scattering. A large number of events will produce multiple traces which intersect at the source. The simplest reconstruction is just a superposition of the traces which will build up a peak at the source position. The angular resolution of the reconstruction is limited by both the energy resolution and the position resolution of the front and back plane detectors and by the separation between the two detector planes. The Compton camera design

3 The first imaging gamma-ray detector was the Anger camera consisting of a pin-hole collimator in front of a position sensitive scintillator. The first Compton telescope was essentially a collimator rather than an imager, with an angular resolution of. It consisted of a single scintillator cell for the front plane and an array of nine cells in the back plane. The more recent UCR gamma-ray telescope consists of two arrays of 6 one meter long scintillator rods. The rods are viewed at each end by separate photomultiplier (PM) tubes for position sensitivity, achieving an angular resolution of. The basic concept of Compton camera comprises a detector system made of two parts, a scatter and an absorber. The direction of incident gamma-rays, Compton-scattering from the scatterer to the absorber, is reconstructed by analyzing the interaction positions and deposited energies in both detectors, i.e. by extracting the scattering angle, and applying the Compton formula. Hence the most relevant design criteria are:. Maximize the probability that the incident gamma-rays scatter by just a single scattering interaction from the scatterer to the absorber and that they are absorbed in the later. This determines the total detection efficiency and the sensitivity of the Compton camera.. Minimize the uncertainty with which the positions and energies of the gamma-ray interactions are determined and maximize the distance between scatterer and absorber. This maximizes the accuracy of determining the scattering angle and hence optimizes the resolution of the Compton camera. The scatter detector should have a high Compton scattering cross section. There are a number of options for absorbing the energy of the Compton scattered photon, from high density detectors such as sodium iodide, NaI(Tl), or bismuth germinate scintillators to multiple layers of silicon or germanium strip detectors. The desired events consist of a Compton scatter in the first detector followed by a full energy absorption in the absorption detector. Time coincidences are used to associate events in one detector with the other. In addition, since the energy of the photons emitted by the radionuclide is known, the use of germanium detectors means that the excellent energy resolution can also be exploited. Assuming that E is known, the uncertainty in the angle θ due to the error in the measured energy E is given below: Δθ mc = sin( θ )( E E ) ΔE The sin(θ) term in the denominator results in a large uncertainty for small angle scattering. The E terms in the denominator indicate that the uncertainty decreases rapidly with gamma ray energy. The finite width and depth of the detectors in each plane contribute to the geometrical uncertainty Δθ g, which represents an uncertainty in the vector connecting the positions of interactions in the two detector planes. The errors add in quadrature to give a total contribution to the angular uncertainty as follows: tan ( Δθ ) = tan ( Δθ ) + tan ( Δ θ g )

4 The choice of detector material is affected by the ratio of the photoelectric absorption to the Compton scattering cross section, which increases rapidly with atomic number Z. The former dominates below about 4 kev for Ge while the Compton Effect remains dominant down to about 6 kev for Si. For high Z semiconductors such as CdTe and HgI, the cross-over occurs at 3-4 kev. From Eq. [], the energy deposited in the first plane, E = E E = E E + α( cos( θ )) where α = E / m o c is the ratio of the incident photon energy to the electron rest mass. When this ratio is small, the energy deposited is also small and can fall below the noise threshold of the detector. All of these factors combine to make Ge marginal for 99m Te gamma rays at 4 kev where a 5 kev energy threshold eliminates all events with less than 3. The optimum geometry of the detector system is influenced by the above factors which suggest an arrangement that favors scattering angles of 3 or greater. The Klein-Nishina equations for Compton scattering show that the scattering probability per unit solid angle is largest for small angle scattering. However, the available solid angle increases with the scattering angle so that the probability per unit scattering angle is peaked around 3 for 66 kev gamma rays as shown in figure below. While only a single detector can be placed at, there is room for several detectors placed in a ring at 3. This also impacts on the question of open versus closepacked arrays of detectors. Close-packed arrays tend to favor small angle scattering whereas open arrays can be arranged at larger scattering angles. In addition, there is the increased probability of multiple scatterings into adjacent detectors in a closepacked array, which limits the useful detector thickness. Further, the ellipses generated by back-projections from adjacent detectors will tend to be similar and thus give poorer point spread functions than for detectors with larger separations. For these reasons, systems using open arrays generally have the advantage. Detectors Type

5 The Monte Carlo simulations suggest that both Si and Ge are feasible scattering detectors. The excellent efficiency that results from using Ge as the first detector has been demonstrated by Singh et al. and is the detector of choice for their ECGC system. Moreover, it is necessary to stack individual silicon detectors to achieve competitive efficiency. However, the need for cryogenic cooling of Ge, the expense of Ge detectors and the interesting advances recently made in silicon detector manufacture have led us to start investigations using silicon detectors. A crude prototype consisting of a single 5 X 5 cm microstrip detector operated in coincidence with a Pho gamma camera is being developed in collaboration with members of the high-energy physics group at Imperial College. Having established a basic working system, additional planes of Si will be added thus increasing the sensitivity. An interesting possibility that will also be examined is the use of two-dimensionally position-sensitive silicon drift detectors as the first detector. The potentially excellent energy resolution afforded by the low capacitance of these devices and their outstanding position resolution make them attractive. Conclusion Compton Cameras have advantages over mechanically collimated systems of having a wider field of view and of using a much larger fraction of the gamma rays emitted by the source in order to produce an image. This gives them increased sensitivity allowing improved signal-to-noise ratio, shorter counting time, or when used for medical imaging, it could reduce the radiation dosage to the patient. Compare to conventional detector, it can image high energy photons ( kev~ MeV), and it has the ability to obtain 3D image from a fixed position on one side of the source without the need for tomography. With the improvement of reconstruction algorithms and hardware, the Compton camera could apply to modern SPECT and PET scanners. Reference [] G.W. Philips, Gamma-ray imaging with Compton cameras, Nuclear Instruments & Methods in Physics Research B 99 (995) [] W. Gast et al., Gamma-ray imaging with segmented tracking detectors. [3] G. J. Royle et al., Design of a Compton camera for imaging 66 kev radionuclide distributions, Nuclear Instruments & Methods in Physics Research A 348 (994) [4] C. Solomon et al., Gamma Ray Imaging with Silicon Detectors A Compton Camera for Radionuclide Imaging in Medicine, Nuclear Instruments & Methods in Physics Research A 73 (988) [5] Singh M, Brechner RR. Experimental test-object study of electronically collimated SPECT. J Nucl Med 99;3:78-86.

Compton Camera with PositionSensitive Silicon Detectors

Compton Camera with PositionSensitive Silicon Detectors University of Ljubljana Faculty of mathematics and physics Andrej Studen Compton Camera with PositionSensitive Silicon Detectors Doctoral thesis Supervisor: Professor Marko Mikuž Outline: Motivation Basic

More information

Compton suppression spectrometry

Compton suppression spectrometry Compton suppression spectrometry In gamma ray spectrometry performed with High-purity Germanium detectors (HpGe), the detection of low intensity gamma ray lines is complicated by the presence of Compton

More information

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector ISSN: 1791-4469 Copyright 2012 Hellenic Naval Academy A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector K. Zachariadou a,c, K. Karafasoulis b,c, S. Seferlis c, I. Papadakis

More information

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 2, APRIL 2012 469 Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions Weiyi Wang, Member,

More information

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: NUCLEAR MEDICINE Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Introduction to Medical Imaging: Physics, Engineering and Clinical Applications, by Nadine Barrie Smith

More information

A Concept of Cylindrical Compton Camera for SPECT

A Concept of Cylindrical Compton Camera for SPECT A Concept of Cylindrical Compton Camera for SPECT A. Bolozdynya, C.E. Ordonez, and W. Chang Rush-Presbyterian-St. Luke's Medical Center, Chicago, IL 60612 Abstract A novel concept of a Compton camera for

More information

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na Ellen Simmons 1 Contents Introduction Review of the Types of Radiation Charged Particle Radiation Detection Review of Semiconductor

More information

Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine

Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine Journal of Physics: Conference Series PAPER OPEN ACCESS Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine To cite this article: L Yu Dubov et al 2017 J. Phys.:

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 26 Radiation Detection & Measurement II Spiritual Thought 2 I would not hold the position in the Church I hold today had I not followed

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore

More information

A novel design of the MeV gamma-ray imaging detector with Micro-TPC

A novel design of the MeV gamma-ray imaging detector with Micro-TPC Elsevier Science 1 Journal logo A novel design of the MeV gamma-ray imaging detector with Micro-TPC R.Orito *,H.Kubo,K.Miuchi,T.Nagayoshi,A.Takada,T.Tanimori,M.Ueno Department of Physics,Graduate School

More information

Lecture 5: Tomographic nuclear systems: SPECT

Lecture 5: Tomographic nuclear systems: SPECT Lecture 5: Tomographic nuclear systems: SPECT Field trip this saturday at 11 AM at UWMC meet in main hospital lobby at 11 AM if you miss the 'boat', page me at 540-4950 should take ~1 to 1.5 hours, depending

More information

III. Energy Deposition in the Detector and Spectrum Formation

III. Energy Deposition in the Detector and Spectrum Formation 1 III. Energy Deposition in the Detector and Spectrum Formation a) charged particles Bethe-Bloch formula de 4πq 4 z2 e 2m v = NZ ( ) dx m v ln ln 1 0 2 β β I 0 2 2 2 z, v: atomic number and velocity of

More information

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,

More information

Measurement of the transverse diffusion coefficient of charge in liquid xenon

Measurement of the transverse diffusion coefficient of charge in liquid xenon Measurement of the transverse diffusion coefficient of charge in liquid xenon W.-T. Chen a, H. Carduner b, J.-P. Cussonneau c, J. Donnard d, S. Duval e, A.-F. Mohamad-Hadi f, J. Lamblin g, O. Lemaire h,

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Gamma ray coincidence and angular correlation

Gamma ray coincidence and angular correlation University of Cape Town Department of Physics Course III laboratory Gamma ray coincidence and angular correlation Introduction Medical imaging based on positron emission tomography (PET) continues to have

More information

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -

More information

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Position Sensitive Germanium Detectors for the Advanced Compton Telescope Position Sensitive Germanium Detectors for the Advanced Compton Telescope R.A. Kroeger 1, W.N. Johnson 1, J.D. Kurfess 1, B.F. Phlips, P.N. Luke 3, M. Momayezi 4, W.K. Warburton 4 1 Naval Research Laboratory,

More information

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts M.L. McConnell, D.J. Forrest, J. Macri, M. McClish, M. Osgood, J.M. Ryan, W.T. Vestrand and C. Zanes Space Science Center University

More information

Radionuclide Imaging MII Positron Emission Tomography (PET)

Radionuclide Imaging MII Positron Emission Tomography (PET) Radionuclide Imaging MII 3073 Positron Emission Tomography (PET) Positron (β + ) emission Positron is an electron with positive charge. Positron-emitting radionuclides are most commonly produced in cyclotron

More information

Study of the feasibility of a compact gamma camera for real-time cancer assessment

Study of the feasibility of a compact gamma camera for real-time cancer assessment Study of the feasibility of a compact gamma camera for real-time cancer assessment L. Caballero Instituto de Física Corpuscular - CSIC - University of Valencia; C/Catedrático José Beltrán, 2; E-46980;

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Structure of Biological Materials

Structure of Biological Materials ELEC ENG 3BA3: Structure of Biological Materials Notes for Lecture #19 Monday, November 22, 2010 6.5 Nuclear medicine imaging Nuclear imaging produces images of the distribution of radiopharmaceuticals

More information

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation S. Moriuchi, M.Tsutsumi2 and K. Saito2 Nuclear safety technology Center,

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography Presenter: Difei Wang June,2018 Universität Bonn Contents 2 / 24 1 2 3 4 Positron emission Detected events Detectors and configuration Data acquisition Positron emission Positron

More information

The Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002)

The Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002) The Compton Effect Martha Buckley MIT Department of Physics, Cambridge, MA 02139 marthab@mit.edu (Dated: November 26, 2002) We measured the angular dependence of the energies of 661.6 kev photons scattered

More information

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy Taka Tanaka (SLAC/KIPAC) 9/19/2007 SLAC Advanced Instrumentation Seminar Outline Introduction CdTe Diode

More information

Detection and measurement of gamma-radiation by gammaspectroscopy

Detection and measurement of gamma-radiation by gammaspectroscopy Detection and measurement of gamma-radiation by gammaspectroscopy Gamma-radiation is electromagnetic radiation having speed equal to the light in vacuum. As reaching a matter it interact with the different

More information

Applied Radiation and Isotopes

Applied Radiation and Isotopes Applied Radiation and Isotopes 69 () 6 67 Contents lists available at ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Evaluation of dual g-ray imager with

More information

PHYS 3650L - Modern Physics Laboratory

PHYS 3650L - Modern Physics Laboratory PHYS 3650L - Modern Physics Laboratory Laboratory Advanced Sheet Photon Attenuation 1. Objectives. The objectives of this laboratory exercise are: a. To measure the mass attenuation coefficient at a gamma

More information

Compton Imaging with Coincidence Measurements for Treaty Verification Purposes

Compton Imaging with Coincidence Measurements for Treaty Verification Purposes 1 Compton Imaging with Coincidence Measurements for Treaty Verification Purposes Scott Wandel, The Pennsylvania State University Abstract Nuclear arms control is a cooperative effort among many government

More information

Modular Survey Spectrometer and Compton Imager

Modular Survey Spectrometer and Compton Imager 1 Modular Survey Spectrometer and Compton Imager Audrey MacLeod Ionizing Radiation Standards Measurement Science and Standards National Research Council Canada On behalf of the SCoTSS collaboration (Silicon

More information

IPNDV Working Group 3: Technical Challenges and Solutions Nuclear Material (3) Technology Data Sheet

IPNDV Working Group 3: Technical Challenges and Solutions Nuclear Material (3) Technology Data Sheet Nuclear Material (NM) Technology Name: Gamma-Ray Imaging Physical Principle/Methodology of Technology: Gamma-ray imaging provides the location and shape information of gamma-ray emitting radionuclides.

More information

Sample Spectroscopy System Hardware

Sample Spectroscopy System Hardware Semiconductor Detectors vs. Scintillator+PMT Detectors Semiconductors are emerging technology - Scint.PMT systems relatively unchanged in 50 years. NaI(Tl) excellent for single-photon, new scintillation

More information

1420 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE /$ IEEE

1420 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE /$ IEEE 1420 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE 2010 Multitracing Capability of Double-Scattering Compton Imager With NaI(Tl) Scintillator Absorber Hee Seo, Student Member, IEEE, Chan Hyeong

More information

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National

More information

arxiv: v1 [physics.ins-det] 29 Jun 2011

arxiv: v1 [physics.ins-det] 29 Jun 2011 Performance simulation of a MRPC-based PET Imaging System arxiv:1106.5877v1 [physics.ins-det] 29 Jun 2011 A. Banerjee, S. Chattopadhyay April 16, 2018 Abstract The low cost and high resolution gas-based

More information

THE COMPTON EFFECT Last Revised: January 5, 2007

THE COMPTON EFFECT Last Revised: January 5, 2007 B2-1 THE COMPTON EFFECT Last Revised: January 5, 2007 QUESTION TO BE INVESTIGATED: How does the energy of a scattered photon change after an interaction with an electron? INTRODUCTION: When a photon is

More information

Effect of the time characteristics of the Compton camera on its performance

Effect of the time characteristics of the Compton camera on its performance Effect of the time characteristics of the Compton camera on its performance Chibueze Zimuzo Uche Howell Round Michael Cree School of Engineering University of Waikato Introduction Anger camera Compton

More information

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra 22.101 Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra References: W. E. Meyerhof, Elements of Nuclear Physics (McGraw-Hill, New York, 1967),

More information

A. I, II, and III B. I C. I and II D. II and III E. I and III

A. I, II, and III B. I C. I and II D. II and III E. I and III BioE 1330 - Review Chapters 7, 8, and 9 (Nuclear Medicine) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in

More information

Radionuclide Imaging MII Detection of Nuclear Emission

Radionuclide Imaging MII Detection of Nuclear Emission Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor

More information

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Marcus Ressner, PhD, Medical Radiation Physicist, Linköping University Hospital Content What is Nuclear medicine? Basic principles of Functional

More information

A Brief Introduction to Medical Imaging. Outline

A Brief Introduction to Medical Imaging. Outline A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length

More information

Efficiency and Attenuation in CdTe Detectors

Efficiency and Attenuation in CdTe Detectors Efficiency and Attenuation in CdTe Detectors Amptek Inc. Bob Redus, May 5, 00 Amptek s XR-00T-CdTe is a high performance x-ray and gamma ray detector system. Like Amptek s other XR00 products, a detector

More information

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums Nuclear Science and Techniques 4 (13) 63 Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums QIAN Nan 1 WANG Dezhong 1,* WANG Chuan ZHU Yuelong MAUERHOFER Eric

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science

More information

Nuclear Physics and imaging in Medical, Security and Environmental applications

Nuclear Physics and imaging in Medical, Security and Environmental applications Nuclear Physics and imaging in Medical, Security and Environmental applications Andy Boston ajboston@liv.ac.uk Overview of presentation We are utilising segmented HPGe & CZT for Nuclear Structure Physics

More information

NEW DETECTION TECHNIQUES FOR GAMMA RADIATIONS AND POSSIBLE APPLICATIONS

NEW DETECTION TECHNIQUES FOR GAMMA RADIATIONS AND POSSIBLE APPLICATIONS National Scientific Session of the Academy of Romanian Scientists ISSN 2067-2160 Spring 2009 147 NEW DETECTION TECHNIQUES FOR GAMMA RADIATIONS AND POSSIBLE APPLICATIONS Dorel BUCURESCU 1 Abstract. One

More information

Progress in using prompt gammas for ion range monitoring during hadrontherapy

Progress in using prompt gammas for ion range monitoring during hadrontherapy Progress in using prompt gammas for ion range monitoring during hadrontherapy J. Krimmer 1, M. Chevallier 1, J. Constanzo 1, M. Dahoumane 1, D. Dauvergne 1, M. De Rydt 1,2, G. Dedes 1, N. Freud 3, J.M.

More information

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide Nuclear Medicine Imaging Systems: The Scintillation Camera List of Nuclear Medicine Radionuclides Tc99m 140.5 kev 6.03 hours I-131 364, 637 kev 8.06 days I-123 159 kev 13.0 hours I-125 35 kev 60.2 days

More information

www.aask24.com www.aask24.com www.aask24.com P=Positron E= Emission T=Tomography Positron emission or beta plus decay (+ ) is a particular type of radioactive decay, in which a proton inside a radionuclide

More information

Performance of the Gamma-ray Imaging Detector with Micro-TPC

Performance of the Gamma-ray Imaging Detector with Micro-TPC Performance of the Gamma-ray Imaging Detector with Micro-TPC Cosmic-Ray Group Dept. of Physics, Graduate School of Science, Kyoto University R.Orito Orito,H.Kubo,K.,H.Kubo,K.Miuchi,T.Nagayoshi, A.Takada,A.Takeda,T.Tanimori

More information

Drift plane. substrate (20ÉIm polyimide) 200ÉIm. Back strip (180ÉIm width) Base (Ceramic) Anode strip (10ÉIm width) Cathode strip (100ÉIm width)

Drift plane. substrate (20ÉIm polyimide) 200ÉIm. Back strip (180ÉIm width) Base (Ceramic) Anode strip (10ÉIm width) Cathode strip (100ÉIm width) Proceedings of the Second International Workshop on EGS, 8.-1. August, Tsukuba, Japan KEK Proceedings -, pp.11-17 Development of Gamma-Ray Direction Detector Based on MSGC T. Nagayoshi 1, H. Kubo 1, A.

More information

X- & γ-ray Instrumentation

X- & γ-ray Instrumentation X- & γ-ray Instrumentation Used nuclear physics detectors Proportional Counters Scintillators The Dark Ages Simple collimators HEAO A1 & A2: 2 x 8 degree field of view Confusion limit is about 200 sources

More information

A Germanium Detector with Optimized Compton Veto for High Sensitivity at Low Energy

A Germanium Detector with Optimized Compton Veto for High Sensitivity at Low Energy LLNL-TR-518852 A Germanium Detector with Optimized Compton Veto for High Sensitivity at Low Energy S. Friedrich December 6, 2011 Disclaimer This document was prepared as an account of work sponsored by

More information

arxiv:astro-ph/ v1 26 Sep 2003

arxiv:astro-ph/ v1 26 Sep 2003 Applications of Gas Imaging Micro-Well Detectors to an Advanced Compton Telescope arxiv:astro-ph/0309731v1 26 Sep 2003 P. F. Bloser a,, S. D. Hunter a, J. M. Ryan b, M. L. McConnell b, R. S. Miller b,,

More information

Initial Studies in Proton Computed Tomography

Initial Studies in Proton Computed Tomography SCIPP Initial Studies in Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz,

More information

Application of Nuclear Physics

Application of Nuclear Physics Application of Nuclear Physics Frontier of gamma-ray spectroscopy 0.1 IR visible light UV soft X-ray X-ray hard X-ray gamma-ray 1 10 100 1e3 1e4 1e5 1e6 energy [ev] Photoelectric effect e - Compton scattering

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

Interaction of particles with matter - 2. Silvia Masciocchi, GSI and University of Heidelberg SS2017, Heidelberg May 3, 2017

Interaction of particles with matter - 2. Silvia Masciocchi, GSI and University of Heidelberg SS2017, Heidelberg May 3, 2017 Interaction of particles with matter - 2 Silvia Masciocchi, GSI and University of Heidelberg SS2017, Heidelberg May 3, 2017 Energy loss by ionization (by heavy particles) Interaction of electrons with

More information

Researchers at the University of Missouri-Columbia have designed a triple crystal

Researchers at the University of Missouri-Columbia have designed a triple crystal Childress, N. L. and W. H. Miller, MCNP Analysis and Optimization of a Triple Crystal Phoswich Detector, Nuclear Instruments and Methods, Section A, 490(1-2), 263-270 (Sept 1, 2002). Abstract Researchers

More information

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber Wesley Ketchum and Abe Reddy EWI Group, UW REU 2006 Outline Neutrino Physics Background Double Beta Decay and the Majorana

More information

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy ISOCS / LabSOCS Calibration software for Gamma Spectroscopy Counts Setup Hardware Peak Shaping Parameters Rise Time Flat Top Pole-zero Number of Channels Signal Gain Setting up a detector for measurement

More information

Development of a new MeV gamma-ray camera

Development of a new MeV gamma-ray camera Development of a new MeV gamma-ray camera ICEPP Symposium February 16, 2004 Hakuba, Nagano, Japan Kyoto University Atsushi Takeda H. Kubo, K. Miuchi, T. Nagayoshi, Y. Okada, R. Orito, A. Takada, T. Tanimori,

More information

Chem 481 Lecture Material 3/20/09

Chem 481 Lecture Material 3/20/09 Chem 481 Lecture Material 3/20/09 Radiation Detection and Measurement Semiconductor Detectors The electrons in a sample of silicon are each bound to specific silicon atoms (occupy the valence band). If

More information

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16530 Quantitative Assessment of Scattering

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

Advanced Pulse-Shape Analysis and Implementation of Gamma-Ray Tracking in a Position-Sensitive Coaxial HPGe Detector

Advanced Pulse-Shape Analysis and Implementation of Gamma-Ray Tracking in a Position-Sensitive Coaxial HPGe Detector Advanced Pulse-Shape Analysis and Implementation of Gamma-Ray Tracking in a Position-Sensitive Coaxial HPGe Detector by Austin Lee Kuhn B.S. (United States Merchant Marine Academy) 1997 M.S. (University

More information

Interaction of charged particles and photons with matter

Interaction of charged particles and photons with matter Interaction of charged particles and photons with matter Robert Miyaoka, Ph.D. Old Fisheries Center, Room 200 rmiyaoka@u.washington.edu Passage of radiation through matter depends on Type of radiation

More information

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or Let's remind... What is nuclear medicine? Nuclear medicine can be broadly divided into two branches "in vitro" and "in vivo" procedures. There are numerous radioisotopic "in vitro" procedures for genotyping

More information

Imaging Hard X-Ray Compton Polarimeter SOI Sensor Prototype Specification

Imaging Hard X-Ray Compton Polarimeter SOI Sensor Prototype Specification Imaging Hard X-Ray Compton Polarimeter SOI Sensor Prototype Specification Hiro Tajima 1 and Gary S. Varner 2 1 Contact: htajima@slac.stanford.edu Stanford Linear Accelerator Center Stanford University,

More information

Detector R&D at KIPAC. Hiro Tajima Kavli InStitute of Particle Astrophysics and Cosmology

Detector R&D at KIPAC. Hiro Tajima Kavli InStitute of Particle Astrophysics and Cosmology Detector R&D at KIPAC Hiro Tajima Kavli InStitute of Particle Astrophysics and Cosmology Detector R&D Overview Si detector ASIC Integration GLAST GeV Gamma-ray Observatory ASIC DAQ Next generation X-ray

More information

Soft Gamma-ray Detector (SGD) for the NeXT mission and beyond Astro-E2 Tadayuki Takahashi

Soft Gamma-ray Detector (SGD) for the NeXT mission and beyond Astro-E2 Tadayuki Takahashi T.Takahashi, Astronomy with Radioactivies IV,2003 Soft Gamma-ray Detector (SGD) for the NeXT mission and beyond Astro-E2 Tadayuki Takahashi Institute of Space and Astronautical Science (ISAS, Japan) 2005

More information

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification

More information

AIR FORCE INSTITUTE OF TECHNOLOGY

AIR FORCE INSTITUTE OF TECHNOLOGY A PORTABLE COMPTON GAMMA-RAY CAMERA DESIGN THESIS Brian L. Evans, Captain, USAF AFIT/GAP/ENP/95D-05 *" BBmammon STATEMENT % % tjte aambuaea Uxüasuted DEPARTMENT OF THE AIR FORCE AIR UNIVERSITY AIR FORCE

More information

NGN PhD Studentship Proposal

NGN PhD Studentship Proposal NGN PhD Studentship Proposal Note that proposals will be assessed against both the quality of the scientific content and of the proposed training experience. Proposed supervisors (lead first) Dr Laura

More information

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE Ján Haščík, Branislav Vrban, Jakub Lüley, Štefan Čerba, Filip Osuský, Vladimír Nečas Slovak University of Technology

More information

ORTEC AN34 Experiment 10 Compton Scattering

ORTEC AN34 Experiment 10 Compton Scattering EQUIPMENT NEEDED FROM ORTEC 113 Preamplifier (2 ea.) TRUMP-PCI-2K MCA System including suitable PC operating Windows 98/2000/XP (other ORTEC MCAs may be used) 266 Photomultiplier Tube Base (2 ea.) 4001A/4002D

More information

Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation.

Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation. Problem Solving picture θ f = 10 m s =1 cm equation rearrange numbers with units θ factors to change units s θ = = f sinθ fθ = s / cm 10 m f 1 m 100 cm check dimensions 1 3 π 180 radians = 10 60 arcmin

More information

There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine.

There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine. Measurement of radioactivity. Radioactive decay is a random process and therefore fluctuations are expected in the radioactivity measurement. That is why measurement of radioactivity must be treated by

More information

TRADITIONALLY, Compton imaging has been performed

TRADITIONALLY, Compton imaging has been performed 1618 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 4, AUGUST 2004 4 Compton Imaging Using a 3-D Position-Sensitive CdZnTe Detector Via Weighted List-Mode Maximum Likelihood Carolyn E. Lehner, Student

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Towards Proton Computed Tomography

Towards Proton Computed Tomography SCIPP Towards Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz, CA 95064 V.

More information

6: Positron Emission Tomography

6: Positron Emission Tomography 6: Positron Emission Tomography. What is the principle of PET imaging? Positron annihilation Electronic collimation coincidence detection. What is really measured by the PET camera? True, scatter and random

More information

E. Caroli(1), R. M. Curado da Silva(2), J.B. Stephen(1), F. Frontera(1,3), A. Pisa (3), S. Del Sordo (4)

E. Caroli(1), R. M. Curado da Silva(2), J.B. Stephen(1), F. Frontera(1,3), A. Pisa (3), S. Del Sordo (4) E. Caroli(1), R. M. Curado da Silva(2), J.B. Stephen(1), F. Frontera(1,3), A. Pisa (3), S. Del Sordo (4) 1. INAF/IASF-Bologna, Italia 2. Departamento de Fisica, Univerisidade de Combra, Portugal 3. Dipartimento

More information

Figure 1. Decay Scheme for 60Co

Figure 1. Decay Scheme for 60Co Department of Physics The University of Hong Kong PHYS3851 Atomic and Nuclear Physics PHYS3851- Laboratory Manual A. AIMS 1. To learn the coincidence technique to study the gamma decay of 60 Co by using

More information

Results of a Si/CdTe Compton Telescope

Results of a Si/CdTe Compton Telescope Results of a Si/CdTe Compton Telescope Kousuke Oonuki a,b, Takaaki Tanaka a,b, Shin Watanabe a, Shin ichiro Takeda a,b, Kazuhiro Nakazawa a, Takefumi Mitani a,b, Tadayuki Takahashi a,b, Hiroyasu Tajima

More information

Queen s University PHYS 352

Queen s University PHYS 352 Page 1 of 5 Queen s University Faculty of Applied Science; Faculty of Arts and Science Department of Physics, Engineering Physics and Astronomy PHYS 352 Measurement, Instrumentation and Experiment Design

More information

K 40 activity and Detector Efficiency

K 40 activity and Detector Efficiency K 40 activity and Detector Efficiency Your goal in this experiment is to determine the activity of a salt substitute purchased in a local store. The salt subsitute is pure KCl. Most of the potassium found

More information

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH Semiconductor X-Ray Detectors Tobias Eggert Ketek GmbH Semiconductor X-Ray Detectors Part A Principles of Semiconductor Detectors 1. Basic Principles 2. Typical Applications 3. Planar Technology 4. Read-out

More information

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle L 36 Modern Physics [2] X-rays & gamma rays How lasers work Medical applications of lasers Applications of high power lasers Medical imaging techniques CAT scans MRI s Modern physics ideas are strange!

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

PhD Qualifying Exam Nuclear Engineering Program. Part 1 Core Courses

PhD Qualifying Exam Nuclear Engineering Program. Part 1 Core Courses PhD Qualifying Exam Nuclear Engineering Program Part 1 Core Courses 9:00 am 12:00 noon, November 19, 2016 (1) Nuclear Reactor Analysis During the startup of a one-region, homogeneous slab reactor of size

More information

Radiation (Particle) Detection and Measurement

Radiation (Particle) Detection and Measurement Radiation (Particle) Detection and Measurement Radiation detection implies that the radiation interacts (e.g. leaves at least part of its energy) in the material. A specific material is chosen, because

More information

Simulation of 4π HPGe Compton-Suppression spectrometer

Simulation of 4π HPGe Compton-Suppression spectrometer Vol. 9(2), pp. 13-19, 30 January, 2014 DOI: 10.5897/IJPS2013.4075 ISSN 1992-1950 2014 Academic Journals http://www.academicjournals.org/ijps International Journal of Physical Sciences Full Length Research

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

Radioisotopes and PET

Radioisotopes and PET Radioisotopes and PET 1 Radioisotopes Elements are defined by their number of protons, but there is some variation in the number of neutrons. Atoms resulting from this variation are called isotopes. Consider

More information