Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements

Size: px
Start display at page:

Download "Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements"

Transcription

1 Practicum Spectroscopy Fall 2010 LIM Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements Jorge Ferreiro, study degree Chemistry, 5 th semester, fjorge@student.ethz.ch Alex Lauber, study degree Chemistry, 5 th semester, allauber@student.ethz.ch Assistant: Mohamed Tarik Abstract: LA-ICP-MS is a widely used method for trace and ultratrace element detection in solid samples. To get intense signals and low oxidation rates, daily optimization for ICP was necessary. Adjusting parameters were nebulizer gas flow rate, auxiliary gas flow rate, plasma gas flow, RF-power and ion lens voltage; first one being mainly important for optimal signals and low oxide formation. For MS, different dwell times can be set up, showing higher sensitivity for lower dwell time. After daily optimization the LOD and sensitivity for 30 elements of the reference material NIST 610 were calculated, showing higher LODs for unsensitive elements. A qualitative analysis of a 5 centime Swiss coin showed presence of Al, Cu, Na, Ni and Zn. Using a reference material and software Lamtrace a quantitative analysis of a steel sample was performed. In total 17 elements were identified and the results were accurate except for Cu and Cr. Zurich, the J. Ferreiro A. Lauber

2 1 Introduction Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry (LA-ICP-MS) is a widely used method for elemental analysis in solids. The sample is ablated with a pulsed laser beam and introduced as aerosol into the ICP-MS, where it s ionized (ICP) and detected (MS). Figure 1: Standard set-up for LA-ICP-MS equipement. [1] The sample is placed into an airtight, closed ablation-cell which is flushed with a constant He-flow (1 Lmin 1 ). Laser pulses with certain amount of energy are focussed on a solid sample surface. The pulse ablates material from the surface creating a microplasma of ions and neutral particles, which are carried by He/Ar to the plasma of the ICP. Two working modes for ablation are used: scanning mode and drilling mode. The inductively coupled plasma (ICP) atomizes and ionizes the aerosol produced by the laser. Plasma is a state of matter made of ions, electrons and atoms reaching a temperature of about 8000 K. A radio frequency (RF) coil surrounds a part of the torch. The plasma is ignited with a high voltage spark, which ionizes a first Aratom. The generated electron is accelerated and ionizes further Ar-atoms leading to a cascade reaction. The RF-coil supplies the plasma torch with energy to keep it stable. The plasma torch consists of three concentric tubes, which are usually made from quartz. These are shown as the outer tube, middle tube, and sample injector. The gas (usually Ar) used to form the plasma is passed between the outer and middle tubes at a flow rate of Lmin 1. A second gas flow, the auxiliary gas, passes between the middle tube and the sample injector normally at 1 Lmin 1 and is used to change the position of the base of the plasma relative to the tube 2

3 and the injector. A third gas flow, the nebulizer gas, also flowing normally at 1 Lmin 1 carries the sample, in the form of a fine-droplet aerosol, from the sample introduction system. In ICP-MS the ion beam comes through the interface into the mass analyzer. The role of the interface is to transport the ions efficiently from the plasma, which is at atmospheric pressure (760 Torr) to the mass spectrometer analyzer region, which is at approximately 10 6 Torr. The interface consists of two metallic cones (first known as sampler cone and second one known as skimmer cone) with very small orifices, which intermediary space is maintained at a pressure of 2 Torr with a mechanical roughing pump and are normally water-cooled. To focus the ions for the detector an ion lens system (mostly hexapole rods) are used. For the separation by means of a detection of the different elements of the sample, one uses a mass spectrometer (MS). There are many different mass spectrometers with different characteristics and functionalities. For ICP-MS Quadrupole-MS (Q- MS) is widely used because cheap and the sensitivity is sufficient for most analysis. The collimated ion beam enters the mass separator which usually consists of four hyperbolic rods. Each opposite pair of rods are connected with a DC and an AC voltage. According to Mathieu s equations for a given DC and AC amplitude, only ions of a given m ratio will have stable oscillating trajectories and reach the z detector. Widely used detectors are electron- and photomultiplier. For Q-MS a number of instrumental parameter can also be adjusted. The peakhopping mode of operation is generally selected for quantitative analysis. In such a case the dwell time is the period of time that is spend collecting data at a particular mass. Starting with the lowest mass to be analyzed, the instrument collects data (in form of counts-per-second (cps)) for a duration of the selected dwell time and then jumps to the next mass chosen and so on. Normally higher dwell times provoke a smaller error but also fewer mass sweeps in total acquistion period. To get a quantitative analysis of an unknown sample natural internal standards are widely used. (C A ) S = (C A) R (I is ) R (C is ) R (I A ) R (C is) S (I A ) S (I is ) S (1) C indicates concentrations (in ppm), I inentistiy (in cps), A analyte, S sample, is internal standard and R reference material. For all kind of analytical measurements there are always two very important parameters which are given: limit of detection LOD and sensitivity S. The LOD describes the lowest amount of substance which can be distinguished from the background: LOD = 3 σnet S (2) 3

4 where σ net is the background standard deviation and S is the sensitivity given by S = I c (3) where I is the intensity an c the concetration. 4

5 2 Experimental 2.1 General A quadrupole ICP-MS instrument Perkin Elmer SCIEX ELAN 6100 DRC II, controlled by the software called ELAN was used. For laser ablation a solid state Nd:YAG laser LSX-213, CETAC Technologies (λ = 213 nm) and a solid state Nd: YAG laser LSX-500, CETAC Technologies (λ = 266 nm) were used. A reference material silicate glass NIST 610 was used for the daily optimization. For the data evaluation Lamtrace and the program R were used. 2.2 Daily optimization In order to produce low oxidation rates in the ICP ThO+ should be lower than Th + 0.5%. To monitor complete atomization and ionization U+ 1. U + and Th + are Th + used because both ions have similar ionization energies, masses and concentrations (in NIST 610) and their major isotope has an abundance of > 99%. Adjusting parameters were nebulizer gas flow rate, auxiliary gas flow rate, plasma gas flow rate and ion lens voltage. The nebulizer gas flow rate ( Lmin 1 ) and auxiliary gas flow rate ( Lmin 1 ) were changed in a bigger range than the plasma gas flow rate ( Lmin 1 ). The RF-power was kept at 1350 W. 2.3 Acquisition of transient signals and mass spectra The dwell time influences the homogeneity of transient signals. The effect of three different dwell times (4ms, 10ms, 50 ms) was investigated on 30 elements of NIST 610. For six major elements (Ag, Cu, Li, Ni, Pb, Si) the transient signal was measured. 2.4 Mass spectra acquistion To identify the elements within NIST 610 two mass spectra were acquiered (one blank and one NIST 610). To acquire good mass spectra the multi channel analyzer (MCA) was set to measure 20 points per mass. For the elements Ag, Li, Ni, Pb 5

6 and Si the resolution R (in NIST 610 and blank) was calculated. R = m m (4) 2.5 Analysis of unknown sample The mass spectrum of a 5 centime Swiss coin was acquiered in hole drilling ablation mode to identify all present elements in the sample. The mass spectrum of the background was measured. For the detected elements a transient signal was acquired. 2.6 Analysis of steel sample After daily optimization of the ICP-MS system a steel sample was analyzed in hole drilling mode. In total 10 transient signals were measured: 2 times standard, 6 times sample and finally 2 times standard to assure that the measurement sequence didn t alter the conditions in the plasma. Ni was used as internal standard to perform the quantification of the sample [2]. 6

7 3 Results 3.1 Daily optimization Table 2 in the appendix summarizes the optimal parameter values for each lab day. I cps Nebulizer Gas Flow (L/min) Figure 2: t s (Top) Nebulizer gas flow rate for daily optimization: The best working conditions (low oxide formation and intense signal) are achieved with a flow rate of 0.8 Lmin 1. (Bottom) Auxiliary gas flow rate for daily optimization: With a flow rate of 0.8 Lmin 1 best signals are achieved. Both paramters were changed in a wide range. Colorcodes: Al(red), Na (black), Si (green), Th (blue), ThO (brown), U (yellow). I cps Auxiliary Gas Flow (L/min) t s

8 I cps Plasma Gas Flow (L/min) Figure 3: (Top) Plasma gas flow rate for daily optimization: Good working conditions are achieved with a flow rate of 17.5 Lmin 1. Although the effects of larger increase or decrease of the plasma flow rate wasn t sudied. (Bottom) Ion lens voltage calibration: The calibration should yield a good linear fit for the meausred points. The unit for y-axis is voltage-level. Colorcodes: Al(red), Na (black), Si (green), Th (blue), ThO (brown), U (yellow). t s 8

9 3.2 Acquistion of transient signals and mass spectra I cps t dwell = 4ms Figure 4: t s (Top) Transient signal acquistion for dwell time = 4ms in hole drilling mode for six elements: The transient signals are very homogeneous. (Bottom) Transient signal acquistion for dwell time = 10 ms in hole drilling mode for six elements: The signals begin to tickle and lose homogeneity. All signals descrease constanteley due to the hole drilling mode. Colorcodes: Ag (blue), Cu(dark green), Li (red), Ni (light green), Pb (dark blue), Si (dark red). I cps t dwell = 10ms t s 9

10 I cps t dwell = 50ms t s Figure 5: Transient signal acquistion for dwell time = 50 ms in hole drilling mode for six elements: Ag (blue), Cu(dark green), Li (red), Ni (light green), Pb (dark blue), Si (dark red). The strong tickeling of transient signals increases more comparing to 4ms and 10 ms. So it s clearly better to settle a lower dwell time. All signals descrease constanteley due to the hole drilling mode. 10

11 3.3 Mass spectra acquistion 40 Ar Na I cps 10 3 Ar ArN ArCl Figure 6: (Top) Mass spectra of gas blank: Very high intensities are shown for masses 40 and 80, which correspond to the plasma gas 40 Ar + respectively the Ar-dimer 80 Ar + 2. The signals in the low mass range show up due to non-metal elements of the matrix. (Bottom) Mass spectra of NIST 610: All desired NIST 610 elements are detected. m z 40 Ar I cps Mn 59 Co Ce Ag 159 Tb 175 Lu 95 Mo Sn Sb 197 Au Th Pb U Bi m z 11

12 3.4 Analysis of unknown sample 65 Cu I cps Na 27 Al 60 Ni 68 Zn Figure 7: (Top) Mass spectra of coin: In order to indentify the elements possibly present in the sample the whole mass spectrum was acquiered skipping m z = 40 (40 Ar + ) and m z = 80 (80 Ar + 2 ) to protect the detector from high concentration of Ar-species. (Bottom) Transient signals for identified elements: After identfication transient signals were meausured showing very homogenous distributioon despite of Na. m z I cps t s 12

13 3.5 Analysis of steel sample Element Corrected bg./cps bg./cps Calc. Concentration/ppm LOD/ppm B 35(13) 108(10) 41(12) 30(2) Ti 6800(1500) 333(15) 1500(300) 13.9(7) V 31000(1300) 664(33) 363(2) 1.04(7) Cr 24700(850) 410(13) 309(3) 0.88(7) Co (10000) 131(10) 3240(15) 0.51(5) Cu 98000(3000) 153(9) 4900(85) 2.0(2) Zr (82000) 120(10) 1200(210) 0.19(2) Nb (150000) 130(7) 4300(960) 0.30(3) Mo 18200(800) 152(13) 641(8) 1.5(1) Ag 725(48) 121(9) 12.5(5) 0.69(6) Sn 11200(331) 113(9) 184(6) 0.60(6) Sb 8100(260) 114(4) 108(3) 0.48(6) Ce 14000(17000) 112(12) 58(59) 0.17(1) Ta (96000) 116(7) 2800(700) 0.30(3) W 76000(4000) 106(6) 2012(35) 0.9(1) Table 1: Calculated concentrations, background(bg.), background correction and LOD for steel sample. Ni was used as internal standard. Ti, Co, Cu, Nb, Ta and W have high concentrations in steel. One can t figure out a trend for LOD because normally the values increase with heavier elements, but in general most elements have a very low LOD so they are easily detected. 4 Discussion 4.1 Daily optimization The nebulizer gas flow rate showed the largest influence on the signal and formation of oxides. Auxiliary gas flow rate and plasma gas flow didn t have such a large influence. Notice this results are for the intervalls in which the paramters were changed. Obviously the size of the particles in the aerosol will determine the qualitity of the signal. Normally the plasma gas flow will also have a strong influence on the sample mainly on the oxide formation but in this case the rate wasn t nearly changed. 13

14 4.2 Acquisition of transient signals and mass spectra A higher dwell time leads to a more homogenous background and slightly better intensity. In general higher dwell times lead to a longer sweep which can result in a loss of signal. So in this case the suggestion is to take a dwell time between 0.1 ms up to 10 ms because the mass scan is performed hopping in steps of 1 m z although the scan range is big. 4.3 Mass spectra acquistion With one point per mass (MCA = 1) one can t calculate a resolution for two adjacent signals since there is only a maximal value but it isn t possible to draw a line shape. Although the scan speed is higher than for multiple points per peak. One observes for heavier elements better mass resolution for two adjacent peaks. Because of isobaric and polyatomic interferences normally one has a loss of sensitivity. 4.4 Analysis of unknown samples Al, Cu, Na, Ni and Zn were identified through the mass spectrum of the coin. For all signals the transient signal was very homogenous despite for Na, which is one of the major contamination sources for analysis. Na is present in salt which is present on human skin. So the coin was always in contact with Na. The signal appaered to be inhomogenous because the analysis was performed in hole drilling mode and the Na is probably only present on the surface. 4.5 Analysis of steel sample Despite of Co and Cr for all detected elements the concentration was very accurate compared to the certified sample. 14

15 5 Literature [1] Thomas R., A beginner s guide to ICP-MS, Spectroscopy Tutorial [2] Longerich H.P., Jackson S.E., Günther D., J. of Analytical Spectrometry, 1996,11, [3] Uriano G.A.;Certifacte of Analysis - Standard Reference Material 1261a - AISI 4340 Steel, National Bureau of Standards, 1981, 56. [4] Uriano G.A.;Certifacte of Analysis - Standard Reference Material 1262a - AISI 4340 Steel, National Bureau of Standards, 1981, Appendix 6.1 Tables Parameter 1 st day 2 nd day 3 rd day Nebulizer gas flow rate [l/min] Auxiliary gas flow rate [l/min] Plasma gas flow [l/min] Table 2: Optimal parameter values summarized for each experimental day. The parameters remained quite constant for each working day. The nebulizer gas flow rate and auxiliary gas flow rate were changed in bigger ranges than the nebulizer gas flow. Although the last parameter will have a strong influence over the measurements. For the first day the oxide formation rate and complete ionization/atomization were calculated. ThO+ Th + = 0.2% and U+ Th + = 91% 15

16 Element bg./cps bg. corr./cps LOD/ppm S/cps ppm 1 Li (4) 28(7) Na (3) 100(21) Al (27) 28(8) Si (2665) 12(3) Ca (571) 100(26) Cr (14) 173(51) Mn (1) 206(54) Fe (70) 205(81) Co (5) 182(46) Ni (3) 150(39) Cu (3) 138(35) Zn (114) 51(25) Mo (2) 400(97) Pd Ag (8) 260(56) Cd (7) 147(48) Sn (5) 535(128) Sb (5) 291(66) Te Ba (2) 582(150) Ce (2) 555(135) Tb (2) 566(145) Lu (2) 501(134) Pt Au (8) 128(39) Pb (6) 289(63) Bi (4) 241(54) Th (3) 334(85) U (3) 321(76) Table 3: Background (bg.), corrected background, LOD and sensitivity for 30 analytes of NIST 610: Elements with high LODs have a very low sensitivity. 16

17 Isotope Resolution 7 Li 7 29 Si Ni Pb W 204 Table 4: Calculated Resolutions for 5 elements of NIST 610: For heavier elements the resolution is higher. Resolution was calculated at FWMH. Element Calc. concentration/ppm Sample 1262 a/ppm Ref a/ppm Rel. Error /% 11 B 41(12) Ti 1500(300) V 363(2) Cr 309(3) Co 32.40(15) Cu 4900(85) Zr 1200(210) Nb 4300(960) Mo 641(8) Ag 12.5(5) Sn 184(6) Sb 108(3) Ce 59(58) Ta 2800(700) W 2012(35) Table 5: Calculated concentrations, Sample 1262 a and reference (ref.) material 1261a and relative errors. The relative errors are high since the reference material was only used for calibration. Although the calculated concentrations are very accurate for some elements. [2],[3] 17

18 6.2 Program codes 6.3 Lamtrace files 6.4 Labjournal 18

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

Effect of Sampling Depth on the Analyte Response in Laser Ablation Inductively Coupled Plasma Mass Spectrometry

Effect of Sampling Depth on the Analyte Response in Laser Ablation Inductively Coupled Plasma Mass Spectrometry JLMN-Journal of Laser Micro/Nanoengineering Vol., No., Effect of Sampling Depth on the Analyte Response in Laser Ablation Inductively Coupled Plasma Mass Spectrometry Zhongke WANG *, Bodo Hattendorf and

More information

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe: February 20, 2012 Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Subject: ICP-MS Report Job Number: S0CHG688 Dear Joe: Please find enclosed the procedure report for the analysis

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which

More information

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) By Arianne Bazilio & Jacob Weinrich December 2012 Contents Introduction... 2 Sample Introduction... 3 Torch... 4 Interface... 6

More information

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Thermo Scientific icap TQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS Application note Metals Analysis & Production Authors Naoki Sugiyama and Mineko Omori Agilent Technologies,

More information

Sample Analysis Design PART II

Sample Analysis Design PART II Sample Analysis Design PART II Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

ICP-Mass Spectrometer

ICP-Mass Spectrometer ICP-Mass Spectrometer New Mass Spectrometers The main issue: sequential vs. simultaneous Scanning, peak hopping are sequential Like viewing a photo through a peephole One pixel at a time Other parts of

More information

Trace Analyses in Metal Matrices Using the ELAN DRC II

Trace Analyses in Metal Matrices Using the ELAN DRC II www.perkinelmer.com Trace Analyses in Metal Matrices Using the ELAN DRC II Introduction Analyses of matrices containing high levels of metals present a challenge for ICP-MS. First, the concentrations of

More information

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords

More information

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Application Note Semiconductor Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Japan Introduction Hydrochloric

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment : ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction

More information

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Use of ICP-MS in analysing radioisotopes Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Inductively Coupled Plasma Mass Spectrometry (ICP-MS) History ICP-AES

More information

Laser Spectroscopy on Bunched Radioactive Ion Beams

Laser Spectroscopy on Bunched Radioactive Ion Beams Laser Spectroscopy on Bunched Radioactive Ion Beams Jon Billowes University of Manchester Balkan School on Nuclear Physics, Bodrum 2004 Lecture 1. 1.1 Nuclear moments 1.2 Hyperfine interaction in free

More information

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information

Sample Analysis Design PART III

Sample Analysis Design PART III Sample Analysis Design PART III Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS

More information

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry Zhichao Lin, Kathryn Emanuele, Stephanie Healey, and Patrick Regan Analytical Branch Winchester Engineering and

More information

ICP-MS 基本原理. 陳麗真 Jennifer Chen Inorganic Products Specialist

ICP-MS 基本原理. 陳麗真 Jennifer Chen Inorganic Products Specialist ICP-MS 基本原理 陳麗真 Jennifer Chen Inorganic Products Specialist Email:Jennifer.Chen@perkinelmer.com What is the ICP-MS Inductively Coupled Plasma Mass Spectrometry can analyze almost all elements in the periodic

More information

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA

More information

Spectrometric Methods of Analysis. OCN 633 Fall 2013

Spectrometric Methods of Analysis. OCN 633 Fall 2013 Spectrometric Methods of Analysis OCN 633 Fall 2013 Plasma Emission and Plasma Mass Spectroscopy Two fields of elemental analysis undergoing the most study Myriad analytical applications Three categories

More information

Sample Analysis Design Polyatomic Interferences

Sample Analysis Design Polyatomic Interferences Sample Analysis Design Polyatomic Interferences More serious than isobaric interferences Result from possible, short-lived combination of atomic species in the plasma or during ion transfer Common recombinants

More information

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques ENVG 60500 FALL 2013 ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques HISTORY In the 1940s, arc and high-voltage spark spectrometry became widely utilized for metal analysis

More information

The 30-Minute Guide to ICP-MS

The 30-Minute Guide to ICP-MS TECHNICAL NOTE The 30-Minute Guide to ICP-MS ICP - Mass Spectrometry A Worthy Member of the Atomic Spectroscopy Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been

More information

a. An emission line as close as possible to the analyte resonance line

a. An emission line as close as possible to the analyte resonance line Practice Problem Set 5 Atomic Emission Spectroscopy 10-1 What is an internal standard and why is it used? An internal standard is a substance added to samples, blank, and standards. The ratio of the signal

More information

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS ANTOANETA ENE 1, I. V. POPESCU 2, T. BÃDICÃ 3, C. BEªLIU 4 1 Department of Physics, Faculty

More information

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically

More information

Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER

Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL 001.847.913.0777 for Certified, Refurbished Lab Equipment Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER TECHNICAL

More information

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+ INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ Applied'Analy+cal'and'Inorganic'Chemistry'Program' Department'of'Chemistry,'Faculty'of'Science' Mahidol'University' What+is+Plasma?+ Ar Ar + e - Plasma is an ionized

More information

ORBITAL DIAGRAM - A graphical representation of the quantum number "map" of electrons around an atom.

ORBITAL DIAGRAM - A graphical representation of the quantum number map of electrons around an atom. 160 ORBITAL DIAGRAM - A graphical representation of the quantum number "map" of electrons around an atom. 4p 3d 4s 3p 3s 2p 2s 1s Each blank represents an ORBITAL, and can hold two electrons. The 4s subshell

More information

Guide to the Extended Step-Pyramid Periodic Table

Guide to the Extended Step-Pyramid Periodic Table Guide to the Extended Step-Pyramid Periodic Table William B. Jensen Department of Chemistry University of Cincinnati Cincinnati, OH 452201-0172 The extended step-pyramid table recognizes that elements

More information

Lecture 7: Atomic Spectroscopy

Lecture 7: Atomic Spectroscopy Lecture 7: Atomic Spectroscopy 1 Atomic spectroscopy The wavelengths of absorbance and emission from atoms in the gas phase are characteristic of atomic orbitals. 2 In the lowest energy transition, the

More information

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are

More information

Partial Energy Level Diagrams

Partial Energy Level Diagrams Partial Energy Level Diagrams 460 nm 323 nm 610 nm 330 nm 819 nm 404 nm 694 nm 671 nm 589 / 590 nm 767 / 769 nm Lithium Sodium Potassium Gas Mixtures Maximum Temperatures, C Air-Coal Gas 1825 Air-Propane

More information

ORBITAL DIAGRAM - A graphical representation of the quantum number "map" of electrons around an atom.

ORBITAL DIAGRAM - A graphical representation of the quantum number map of electrons around an atom. 178 (MAGNETIC) SPIN QUANTUM NUMBER: "spin down" or "spin up" - An ORBITAL (region with fixed "n", "l" and "ml" values) can hold TWO electrons. ORBITAL DIAGRAM - A graphical representation of the quantum

More information

Fergus Keenan Shona McSheehy Julian Wills

Fergus Keenan Shona McSheehy Julian Wills Novel QCell Technology for Inference Removal in ICP-MS - Combining Low Mass Filtration with Kinetic Energy Discrimination Fergus Keenan Shona McSheehy Julian Wills 1 The world leader in serving science

More information

THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important ones are...

THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important ones are... Atomic Structure THE STRUCTURE OF ATOMS ATOMS Atoms consist of a number of fundamental particles, the most important ones are... Mass / kg Charge / C Relative mass Relative Charge PROTON NEUTRON ELECTRON

More information

Solutions and Ions. Pure Substances

Solutions and Ions. Pure Substances Class #4 Solutions and Ions CHEM 107 L.S. Brown Texas A&M University Pure Substances Pure substance: described completely by a single chemical formula Fixed composition 1 Mixtures Combination of 2 or more

More information

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Application Note - AN1508 Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Introduction Page 1 The combination of hydride generation with

More information

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid Application Semiconductor Authors Abe G. Gutiérrez Elemental Scientific 2440 Cuming St Omaha, NE 68131 USA abe@icpms.com

More information

Nucleus. Electron Cloud

Nucleus. Electron Cloud Atomic Structure I. Picture of an Atom Nucleus Electron Cloud II. Subatomic particles Particle Symbol Charge Relative Mass (amu) protons p + +1 1.0073 neutrons n 0 1.0087 electrons e - -1 0.00054858 Compare

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,

More information

TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS

TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS W.L. Griffin 1, 3, Sonal Rege 1, Rondi M. Davies 1, 2, Simon Jackson 1, Suzanne Y. O Reilly 1 1.ARC National Key

More information

-"l" also contributes ENERGY. Higher values for "l" mean the electron has higher energy.

-l also contributes ENERGY. Higher values for l mean the electron has higher energy. 170 - Giving the four parameters will uniquely identify an electron around an atom. No two electrons in the same atom can share all four. These parameters are called QUANTUM NUMBERS. PRINCIPAL QUANTUM

More information

Introduction to LIBS COMMUNITY USER WORKSHOP ON PLANETARY LIBS (CHEMCAM) DATA. Sam Clegg and the ChemCam team

Introduction to LIBS COMMUNITY USER WORKSHOP ON PLANETARY LIBS (CHEMCAM) DATA. Sam Clegg and the ChemCam team Lunar and Planetary Science Conference, March 18 th, 2015 NASA/JPL-Caltech/MSSS COMMUNITY USER WORKSHOP ON PLANETARY LIBS (CHEMCAM) DATA Introduction to LIBS Sam Clegg and the ChemCam team Creating LIBS

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

Summary of test results for Daya Bay rock samples. by Patrick Dobson Celia Tiemi Onishi Seiji Nakagawa

Summary of test results for Daya Bay rock samples. by Patrick Dobson Celia Tiemi Onishi Seiji Nakagawa Summary of test results for Daya Bay rock samples by Patrick Dobson Celia Tiemi Onishi Seiji Nakagawa October 2004 Summary A series of analytical tests were conducted on a suite of granitic rock samples

More information

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains

More information

Analysis Repeatability of Trace and Major Elements in a Water Sample

Analysis Repeatability of Trace and Major Elements in a Water Sample Analysis Repeatability of Trace and Major Elements in a Water Sample Agnès COSNIER HORIBA Scientific Longjumeau, France Keywords: environment Elements: Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li,

More information

1 Electrons and Chemical Bonding

1 Electrons and Chemical Bonding CHAPTER 13 1 Electrons and Chemical Bonding SECTION Chemical Bonding BEFORE YOU READ After you read this section, you should be able to answer these questions: What is chemical bonding? What are valence

More information

Optical Atomic Spectroscopy

Optical Atomic Spectroscopy Optical Atomic Spectroscopy Methods to measure conentrations of primarily metallic elements at < ppm levels with high selectivity! Two main optical methodologies- -Atomic Absorption--need ground state

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-ray Photoelectron Spectroscopy Introduction Qualitative analysis Quantitative analysis Charging compensation Small area analysis and XPS imaging

More information

Detection of iron species using inductively coupled plasma mass spectrometry under cold plasma temperature conditions

Detection of iron species using inductively coupled plasma mass spectrometry under cold plasma temperature conditions Spectrochimica Acta Part B 2001 123128 Analytical note Detection of iron species using inductively coupled plasma mass spectrometry under cold plasma temperature conditions Li-Shing Huang a,b, King-Chuen

More information

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction

More information

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis Agilent ICP-MS Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis High Sample Matrix Tolerance Superior & Simple Interference Removal Ultra Wide Linear

More information

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document: SPECIFICATION SPECIFICATION Inductively Coupled Plasma Mass Spectrometry System 1. Scope This specification describes the requirements for an Inductively Coupled Plasma Mass Spectrometry System ( System

More information

The Periodic Table. Periodic Properties. Can you explain this graph? Valence Electrons. Valence Electrons. Paramagnetism

The Periodic Table. Periodic Properties. Can you explain this graph? Valence Electrons. Valence Electrons. Paramagnetism Periodic Properties Atomic & Ionic Radius Energy Electron Affinity We want to understand the variations in these properties in terms of electron configurations. The Periodic Table Elements in a column

More information

Agilent Technologies at TIAFT 2013

Agilent Technologies at TIAFT 2013 Agilent Technologies at TIAFT 2013 Analytical approaches for the measurement of trace metals in forensic samples Funchal, Madeira September 2-6 CRIMINALISTIC TOXICOLOGY DNA INVESTIGATION DOCUMENT EXAMINATION

More information

Chapter 3: Elements and Compounds. 3.1 Elements

Chapter 3: Elements and Compounds. 3.1 Elements Chapter 3: Elements and Compounds 3.1 Elements An element is a fundamental substance that cannot be broken down by chemical or physical methods to simpler substances. The 118 known elements are nature

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur Introduction What is ICP-MS? Inductively Coupled Plasma Mass Spectrometry Mass spectrometry

More information

All chemical bonding is based on the following relationships of electrostatics: 2. Each period on the periodic table

All chemical bonding is based on the following relationships of electrostatics: 2. Each period on the periodic table UNIT VIII ATOMS AND THE PERIODIC TABLE 25 E. Chemical Bonding 1. An ELECTROSTATIC FORCE is All chemical bonding is based on the following relationships of electrostatics: The greater the distance between

More information

Radiometric Dating (tap anywhere)

Radiometric Dating (tap anywhere) Radiometric Dating (tap anywhere) Protons Neutrons Electrons Elements on the periodic table are STABLE Elements can have radioactive versions of itself called ISOTOPES!! Page 1 in your ESRT has your list!

More information

- Atomic line spectra are UNIQUE to each element. They're like atomic "fingerprints".

- Atomic line spectra are UNIQUE to each element. They're like atomic fingerprints. - Atomic line spectra are UNIQUE to each element. They're like atomic "fingerprints". - Problem was that the current model of the atom completely failed to explain why atoms emitted these lines. An orbit

More information

Topic 3: Periodicity OBJECTIVES FOR TODAY: Fall in love with the Periodic Table, Interpret trends in atomic radii, ionic radii, ionization energies &

Topic 3: Periodicity OBJECTIVES FOR TODAY: Fall in love with the Periodic Table, Interpret trends in atomic radii, ionic radii, ionization energies & Topic 3: Periodicity OBJECTIVES FOR TODAY: Fall in love with the Periodic Table, Interpret trends in atomic radii, ionic radii, ionization energies & electronegativity The Periodic Table What is the periodic

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Essential Chemistry for Biology

Essential Chemistry for Biology 1 Chapter 2 Essential Chemistry for Biology Biology and Society: More Precious than Gold A drought is a period of abnormally dry weather that changes the environment and one of the most devastating disasters.

More information

Ch. 8 Introduction to Optical Atomic Spectroscopy

Ch. 8 Introduction to Optical Atomic Spectroscopy Ch. 8 Introduction to Optical Atomic Spectroscopy 8.1 3 major types of Spectrometry elemental Optical Spectrometry Ch 9, 10 Mass Spectrometry Ch 11 X-ray Spectrometry Ch 12 In this chapter theories on

More information

Overview of X-Ray Fluorescence Analysis

Overview of X-Ray Fluorescence Analysis Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass

More information

4.1 Atomic structure and the periodic table. GCSE Chemistry

4.1 Atomic structure and the periodic table. GCSE Chemistry 4.1 Atomic structure and the periodic table GCSE Chemistry All substances are made of atoms this is cannot be chemically broken down it is the smallest part of an element. Elements are made of only one

More information

This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under contract

This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under contract This work was performed under the auspices of the U.S. Department of Energy by under contract DE-AC52-7NA27344. Lawrence Livermore National Security, LLC The ITER tokamak Tungsten (W) is attractive as

More information

PRINCIPLE OF ICP- AES

PRINCIPLE OF ICP- AES INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively

More information

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,

More information

-"l" also contributes ENERGY. Higher values for "l" mean the electron has higher energy.

-l also contributes ENERGY. Higher values for l mean the electron has higher energy. 175 - Giving the four parameters will uniquely identify an electron around an atom. No two electrons in the same atom can share all four. These parameters are called QUANTUM NUMBERS. PRINCIPAL QUANTUM

More information

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Application Note Foods Author Glenn Woods ICP-MS Specialist Agilent Technologies UK Ltd. Lakeside Business

More information

ICPMS Doherty Lecture 1

ICPMS Doherty Lecture 1 ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms

More information

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS A P P L I C AT I O N N O T E ICP - Mass Spectrometry Authors: Lee Davidowski Chady Stephan PerkinElmer, Inc. Shelton, CT The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical

More information

Element Cube Project (x2)

Element Cube Project (x2) Element Cube Project (x2) Background: As a class, we will construct a three dimensional periodic table by each student selecting two elements in which you will need to create an element cube. Helpful Links

More information