Istituto Nazionale di Ricerca Metrologica
|
|
- Claire Brooks
- 6 years ago
- Views:
Transcription
1 Istituto Nazionale di Ricerca Metrologica G. D Agostino, S. Desogus, A. Germak, C. Origlia and D. Quagliotti ABSOLUTE MEASUREMENTS OF THE FREE-FALL ACCELERATION g IN PANTELLERIA (ITALY) T.R. 76 November 2007 TECHNICAL REPORT I.N.RI.M. 1
2 ABSTRACT The work hereafter described was carried out on June 23-36, 2007 by the Istituto Nazionale di Ricerca Metrologica (INRIM) for the Istituto Nazionale di Geofisica e Vulcanologia (INGV) - Centro Nazionale Terremoti and Osservatorio Vesuviano. The experimental results of absolute measurements of the free-fall acceleration g carried out in Pantelleria are reported. Gravity measurements were performed with the new transportable absolute gravimeter IMGC-02. 2
3 CAPTIONS INDEX ABSTRACT... 2 CAPTIONS INDEX... 3 TABLES and FIGURE INDEX INTRODUCTION IMGC ABSOLUTE GRAVIMETER DESCRIPTION Measurement method Technical description MEASUREMENT UNCERTAINTY The instrumental uncertainty of the IMGC-02 absolute gravimeter EXPERIMENTAL RESULTS Khamma Porto REFERENCES
4 TABLES and FIGURE INDEX Figure 1. Picture of the new absolute gravimeter IMGC Figure 2.1. Schematic layout of the IMGC-02 Absolute Gravimeter... 7 Figure 2.2. GravisoftM manager front panel... 8 Figure 2.3. GravisoftPP post-processing front panel... 8 Table 3.1. Instrumental uncertainty of the IMGC-02 absolute gravimeter Figure Pictures of the observation station in Khamma Figure Plane of the building in Khamma Table Experimental results in Khamma Table Apparatus setup in Khamma Figure Time series in Khamma (rejected-red, accepted-white) Figure Data sets (average of 20 launches) collected in Khamma Figure Trajectory residuals (one launch-red, average-white) in Khamma Figure Density frequency graph in Khamma Figure Local barometric pressure acquired in Khamma Figure Launch chamber pressure in Khamma Figure Tide correction in Khamma Table Measurement uncertainty in Khamma Figure Pictures of the observation station in Porto Figure Plane of the building in Porto Table Experimental results in Porto Table Apparatus setup in Porto Figure Time series in Porto (rejected-red, accepted-white) Figure Data sets (average of 20 launches) collected in Porto Figure Trajectory residuals (one launch-red, average-white) in Porto Figure Density frequency graph in Porto Figure Ambient temperature acquired in Porto Figure Local barometric pressure acquired in Porto Figure Launch chamber pressure in Porto Figure Tide correction in Porto Table Measurement uncertainty in Porto
5 1 INTRODUCTION The measurement of the free-fall acceleration, g, has been performed with the new gravimeter IMGC-02. The apparatus (fig.1) is developed by INRIM /1/, and derives from that one previously realized in collaboration with the Bureau International des Poids et Mesures in Sèvres (BIPM) /2/. Several improvements characterize the IMGC-02, among them there is the automation of the instrument which allows to perform the measurement during the night, when the disturbance due to the environmental noise is minimum. All the measurement sessions have been recorded and stored in data files for postprocessing. If necessary, these files are delivered for future revision or checking. The software used is the GravisoftM 1.1 and GravisoftPP 1.1, developed and tested by INRIM. Figure 1. Picture of the new absolute gravimeter IMGC-02 5
6 2 IMGC ABSOLUTE GRAVIMETER DESCRIPTION 2.1 MEASUREMENT METHOD The free-fall acceleration g is measured by to tracking the vertical trajectory of a testbody subjected to the gravitational acceleration. The IMGC-02 absoulte gravimeter adopts the symmetric rise and falling method, where both the rising and falling trajectories of the test-mass are recorded. The raw datum consists in an array where each element represents the time correspondent to the passage of the test-body through equally spaced levels (or stations). A model function derived from the equation of motion is fitted to the raw datum in a least-squares adjustment. One of the parameter estimates is the acceleration experienced by the test-body during its flight. A measurement session consists of about 2000 launches. To assure the evaluated measurement uncertainty, the g value is obtained by averaging those launches which fulfill accepting criteria. 2.2 TECHNICAL DESCRIPTION A schematic layout of the apparatus is showed in fig The basic parts of the instrument are a Mach-Zehnder interferometer /3/ and a long-period (about 20 s) seismometer. The wavelength of a iodine stabilised He-Ne laser is used as the length standard. The inertial mass of a seismometer supports a cube-corner reflector, which is the reference mirror of the interferometer. The moving mirror of the interferometer is also a cube-corner retro-reflector and is directly subjected to the free falling motion. It is thrown upwards vertically by means of a launch pad in a vacuum chamber (about Pa). Interference fringes emerging from the interferometer are detected by a photo-multiplier. The output signal is sampled by a high-speed waveform digitizer synchronized to a Rb oscillator, used as the time standard. Equidistant stations are selected by counting a constant integer number of interference fringes (at present 1024); in particular consecutive stations are separated by a distance d = 1024 λ/2, being λ the wavelength of the laser radiation. The so called local fit method is used to time the interference signal /4/. In particular the time is computed by fitting the equation model of the interference of monochromatic waves to the interference fringe correspondent to the selected station. The space-time coordinates are processed in a least-squares algorithm, where a suitable model function is fitted to the trajectory datum. Each throw gives an estimate of the g value. A personal computer manage the instrument. The pad launch is triggered only if the system is found to be ready. In particular the software checks the pad launch state (loaded or unloaded) and the laser state (locked or unlocked). Environmental parameters such as the local barometric pressure and the temperature are acquired and stored for each launch. 6
7 seismometer interferometer photo-detector frame laser launch chamber computer electronics Figure 2.1. Schematic layout of the IMGC-02 Absolute Gravimeter The software used includes (i) the manager GravisoftM 1.1 (fig. 2.2) for driving the instrument and storing the measurement data and (ii) the post-processing GravisoftPP 1.1 (fig. 2.3) for elaborating the data-files. These programs were developed and tested on the LabVIEW8.2 platform. Geophysical corrections are applied: (i) the Earth tides and Ocean loading are computed with the ETGTAB, version 3.0, which uses the tidal potential development of Cartwright-Tayler-Edden (1973), (ii) the polar motion correction is computed starting from the daily pole coordinates x and y (rad) obtained from the International Earth Rotation Service (IERS). The measured gravity is normalized to a nominal pressure, taking into account a barometric factor f B = m s -2 mbar -1, as recommended by the IAG 1983 resolution n.9. Instrumental corrections are also applied: (i) the diffraction correction and the (ii) laser beam verticality. The g value associated to every measurement session is calculated as the average of n measurements and it is referred to a specific height from the floor surface. The measurement expanded uncertainty is evaluated according to the method of combination of uncertainties as suggested by the ISO GUM guide /5/. 7
8 Figure 2.2. GravisoftM manager front panel Figure 2.3. GravisoftPP post-processing front panel 8
9 3 MEASUREMENT UNCERTAINTY The uncertainty associated to the g measurement is evaluated by combining the contributions of uncertainty of the IMGC-02 absolute gravimeter, called the instrumental uncertainties to the contribution of uncertainty depending on the observation site. 3.1 THE INSTRUMENTAL UNCERTAINTY OF THE IMGC-02 ABSOLUTE GRAVIMETER Influence factors which are characteristic of the instrument are: vacuum level, nonuniform magnetic field, temperature gradient, electrostatic attraction, mass distribution, laser beam verticality, air gap modulation, length standard, time standard, retro-reflector balancing, radiation pressure and reference height. A detailed description of these phenomena concerning the present IMGC-02 absolute gravimeter can be found in /1/. Tab. 3.1 reports the quantitative assessment of the effect of every disturbing factor. The expanded uncertainty at the 95% confidence level (coverage factor k = 2.10 and 19 degrees of freedom) is estimated to be U = m s -2. The measurement uncertainty results from the combination of the instrument uncertainty with influence factors that are dependent from the observation site: Coriolis force, floor recoil and geophysical effects such as local barometric pressure, gravity tides, ocean loading and polar motion. Uncertainty tables, related to each observation site, are attached to the experimental results below described. 9
10 ctiinfluence parameters, x i Value Unit u i or a i Type A, s i Type B, a i Table 3.1. Instrumental uncertainty of the IMGC-02 absolute gravimeter Coreg on Type of distribution Equivalent variance Sensitivity coefficients Contribution to the variance Degrees of freedom, ν i Equivalent standard uncertainty Drag effect negligible Outgassing effect negligible Non-uniform magnetic field effect negligible Temperature gradient effect m s -2 ±1.5E E-09 U 1.1E E E E-09 Effect for Electrostatic negligible Mass distribution effect m s -2 ±5.0E E-09 rectangular 8.3E E E E-09 Laser beam verticality correction 6.6E-09 m s -2 ±2.1E E E-09 rectangular 1.5E E E E-09 Air gap modulation effect negligible Laser effect m s E E E E E E-09 Index of refraction effect negligible Beam divergence correction 1.52E-07 m s E E E E E E E-08 Beam share effect unknown unknown Clock effect m s E E-09 rectangular 3.6E E E E-09 Finges timing effect negligible Finite value of speed of light effect negligible Retroreflector balancing 0.0E+00 m ±1.0E E-04 rectangular 3.3E E E E-08 Radiation Pressure effect negligible Reference height 5.0E-01 m ±5.0E E-04 rectangular 8.3E E E E-10 Corr. 1.59E-07 m s -2 Variance 1.6E-15-4 m 2 s Combined standard uncertainty, u 4.0E-08 m s -2 Degrees of freedom, ν eff (Welch-Satterthwaite formula) Confidence level, p Coverage factor, k (calculated with t-student) Expanded uncertainty, U = ku 21 95% E-08 m s -2 Relative expanded uncertainty, U rel = U/g 8.4E-09 10
11 4 EXPERIMENTAL RESULTS 4.1. KHAMMA The observation station of Khamma is located on a room of the elementary school, fig Figure Pictures of the observation station in Khamma 11
12 The position of the measurement point, referred to the room is showed on the plan of the building, fig The orientation of the instrument is showed by the red triangle where black square represents the laser body. Figure Plane of the building in Khamma The instrument processed and stored 999 trajectories after executing 1338 throws. The number of trajectories useful for post-processing was 599. The measured data are filtered by applying rejecting criteria. The most critical factor is the visibility variation of the interference signal during the trajectory, which highlights an horizontal motion of the test-body. The effect due to the Coriolis force and the beam share are minimized by rejecting those launches with a decrease of visibility bigger that 10%. Outliers are found by applying the Chauvenet criterion to the estimating parameters such as the vertical gradient, the friction of residual air and to the estimated g value. The final g value is obtained by averaging 160 trajectories. Tab reports the most important experimental results. Other information concerning the apparatus setup is reported in tab
13 Table Experimental results in Khamma Observation Station: Khamma Date 2007, June Start/Stop time (UTC) 21:36:07 / 03:42:18 Geodetic longitude λ = Geodetic latitude ϕ = Topographic elevation H T = 170 m Nominal pressure at the observation site P N = mbar Pole coordinates in IERS system x = , y = Measurement parameters Total measurement time T m = 9.92 h Measurement rate m r = 140 h -1 Measurement drift m d = m s -2 h Total executed throws n et = 1388 Total processed and stored throws n ps = 999 Number of throws useful for post-processing n pp = 599 Temperature range T = (- -) C Local barometric pressure (mean) P = mbar χ 2 test (80% confidence level) χ 2 max = 19.8; χ 2 min = 7.0; χ 2 exp = 6.7 Corrections Laser beam verticality correction g bv = m s -2 Laser beam divergence correction g bd = m s -2 Polar motion correction g pm = m s -2 Tide and ocean loading correction (mean) g tol = m s -2 Local barometric pressure correction (mean) g bp = m s -2 Results corrected mean g value g mv = m s -2 Reference height h ref = m Number of throws accepted for the average n = 160 Experimental standard deviation s g = m s -2 Experimental standard deviation of the mean value s gm = m s -2 Measurement combined uncertainty u gm = m s -2 Measurement expanded uncertainty (p = 95%, ν = 38, k = 2.03) U gm = m s -2 Vertical gradient (INGV data) at (80 120) cm γ = (???.? ±?.?) 10-8 s -2 Table Apparatus setup in Khamma Instrument orientation Laser body to? direction (see figure) Fitting Model Laser modulation Fringe visibility threshold f vt = 10% Measurements each set n ma = 20 Waveform digitizer sampling frequency S f = 50 MHz Laser wavelength λ l = m Clock frequency f c = Hz Vertical gradient input γ = s -2 Rise station number n rs = 350 Leaved upper stations n sl = 20 Laser modulation frequency f lm = Hz Instrumental height h inst = m 13
14 The time series concerning the post-processed trajectories is reported in fig Data sets, each correspondent to the average of 20 launches, are reported in fig The apparatus experienced an oscillation of about m s -2. The averaged trajectory residuals after the measurement session are within ± m, fig Figure Time series in Khamma (rejected-red, accepted-white) Figure Data sets (average of 20 launches) collected in Khamma Figure Trajectory residuals (one launch-red, average-white) in Khamma 14
15 The histogram reported in fig represents the density frequency graph of the measurement session. The χ 2 test rejects the null hypothesis, i.e. the normal distribution, with a 20% risk error. Figure Density frequency graph in Khamma The sensor used to record the ambient temperature didn t work properly. The graph reporting the ambient temperature is missing. Fig reports the local barometric pressure acquired each launch. Figure Local barometric pressure acquired in Khamma Fig and fig report respectively the launch chamber pressure and the tide correction applied during the measurement station. Figure Launch chamber pressure in Khamma 15
16 Figure Tide correction in Khamma The measurement uncertainty in Khamma is summarized in tab It includes the instrumental uncertainty reported in tab REMARKS After the measurement session the interferometer was found aligned as well as at the beginning. There is not reason for suspecting a thermal drift of the apparatus. The absolute measurement of the free-fall acceleration is considered to be correct. 16
17 CoonTable Measurement uncertainty in Khamma Influence parameters, x i Value Unit u i or a i Type A, s i Type B, a i re g ctitype of distribution Equivalent variance Sensitivity coefficients Contribution to the variance Degrees of freedom, ν i Equivalent standard uncertainty Instrument uncertainty m s E E E E E E-08 Coriolis effect m s -2 ±2.9E E-08 rectangular 2.8E E E E-08 Floor recoil effect negligible Barometric pressure correction -3.1E-08 m s -2 ±1.0E E E-08 rectangular 3.3E E E E-09 Tide correction -2.9E-07 m s E E E E E E E-09 Ocean loading correction m s E E E E E E E-09 Polar motion correction -6.0E-09 m s -3 negligible -6.0E-09 Standard deviation of the mean value m s E E E E E E-08 Corr. -3.3E-07 m s -2 Variance 2.2E-15-4 m 2 s Combined standard uncertainty, u Degrees of freedom, ν eff (Welch-Satterthwaite formula) Confidence level, p Coverage factor, k (calculated with t-student) Expanded uncertainty, U = ku Relative expanded uncertainty, U rel = U/g 4.7E-08 m s % E-08 m s E-09 17
18 4.2 PORTO The observation station of Porto is located on a room of the elementary school, fig Figure Pictures of the observation station in Porto 18
19 The position of the measurement point, referred to the room is showed on the plan of the building, fig The orientation of the instrument is showed by the red triangle where black square represents the laser body. Figure Plane of the building in Porto The instrument processed and stored 856 trajectories after executing 987 throws. The number of trajectories useful for post-processing was 549. The measured data are filtered by applying rejecting criteria. The most critical factor is the visibility variation of the interference signal during the trajectory, which highlights an horizontal motion of the test-body. The effect due to the Coriolis force and the beam share are minimized by rejecting those launches with a decrease of visibility bigger that 10%. Outliers are found by applying the Chauvenet criterion to the estimating parameters such as the vertical gradient, the friction of residual air and the estimated g value. The final g value is obtained by averaging 102 trajectories. Tab reports the most important experimental results. Other information concerning the apparatus setup is reported in table
20 Table Experimental results in Porto Observation Station: Porto Date 2007, June 24 Start/Stop time (UTC) 08:20:28 / 13:12:15 Geodetic longitude λ = Geodetic latitude ϕ = Topographic elevation H T = 20 m Nominal pressure at the observation site P N = mbar Pole coordinates in IERS system x = , y = Measurement parameters Total measurement time T m = 4.88 h Measurement rate m r = 140 h -1 Measurement drift m d = m s -2 h Total executed throws n et = 683 Total processed and stored throws n ps = 549 Number of throws useful for post-processing n pp = 549 Temperature range T = ( ) C Local barometric pressure (mean) P = mbar χ 2 test (80% confidence level) χ 2 max = 17.2; χ 2 min = 5.6; χ 2 exp = 17.5 Corrections Laser beam verticality correction g bv = m s -2 Laser beam divergence correction g bd = m s -2 Polar motion correction g pm = m s -2 Tide and ocean loading correction (mean) g tol = m s -2 Local barometric pressure correction (mean) g bp = m s -2 Results corrected mean g value g mv = m s -2 Reference height h ref = m Number of throws accepted for the average n = 102 Experimental standard deviation s g = m s -2 Experimental standard deviation of the mean value s gm = m s -2 Measurement combined uncertainty u gm = m s -2 Measurement expanded uncertainty (p = 95%, ν = 34, k = 2.03) U gm = m s -2 Vertical gradient (INGV data 2005) at (80 120) cm γ = (???.? ±?.?) 10-8 s -2 Table Apparatus setup in Porto Instrument orientation Laser body to? direction (see figure) Fitting Model Laser modulation Fringe visibility threshold f vt = 10% Measurements each set n ma = 20 Waveform digitizer sampling frequency S f = 50 MHz Laser wavelength λ l = m Clock frequency f c = Hz Vertical gradient input γ = s -2 Rise station number n rs = 350 Leaved upper stations n sl = 20 Laser modulation frequency f lm = Hz Instrumental height h inst = m 20
21 The time series concerning the post-processed trajectories is reported in fig Data sets, each correspondent to the average of 20 launches, are reported in fig The apparatus experienced an oscillation of about m s -2. The averaged trajectory residuals after the measurement session are within ± m, fig Figure Time series in Porto (rejected-red, accepted-white) Figure Data sets (average of 20 launches) collected in Porto Figure Trajectory residuals (one launch-red, average-white) in Porto 21
22 The histogram reported in fig represents the density frequency graph of the measurement session. The χ 2 test rejects the null hypothesis, i.e. the normal distribution, with a 20% risk error. Figure Density frequency graph in Porto Fig and fig report respectively the ambient temperature and the local barometric pressure acquired each launch. Figure Ambient temperature acquired in Porto Figure Local barometric pressure acquired in Porto 22
23 Fig and fig reports respectively the launch chamber pressure and the tide correction applied during the measurement station. Figure Launch chamber pressure in Porto Figure Tide correction in Porto The measurement uncertainty in Palestrina is summarized in tab It includes the instrumental uncertainty reported in tab REMARKS The planned observation point, located at the middle of the room, showed a strong recoil effect. In particular the trajectory residuals highlighted a remarkable excitation of the resonance frequency of the seismometer inner spring. The instrument performance was significantly degraded. The relevant results are not reported because they are meaningless but collected data are available if necessary. The reported results concern a new observation point, located near the door (see fig ). During the observation session the alignment of the interferometer was affected by the horizontal displacement of the launch chamber. The measuring and fixed beams were not coaxial anymore. This was probably due to the slope of the floor. Luckily the displacement was linear and very slow, therefore it was possible to process the beginning of the session when the interferometer alignment was still acceptable. For this reason the absolute measurement of the local gravity field is considered to be correct. 23
24 CoonTable Measurement uncertainty in Porto Influence parameters, x i Value Unit u i or a i Type A, s i Type B, a i re g ctitype of distribution Equivalent variance Sensitivity coefficients Contribution to the variance Degrees of freedom, ν i Equivalent standard uncertainty Instrument uncertainty m s E E E E E E-08 Coriolis effect m s -2 ±2.9E E-08 rectangular 2.8E E E E-08 Floor recoil effect negligible Barometric pressure correction -3.1E-08 m s -2 ±1.0E E E-08 rectangular 3.3E E E E-09 Tide correction -2.9E-07 m s E E E E E E E-09 Ocean loading correction m s E E E E E E E-09 Polar motion correction -6.0E-09 m s -3 negligible -6.0E-09 Standard deviation of the mean value m s E E E E E E-08 Corr. -3.3E-07 m s -2 Variance 2.1E-15-4 m 2 s Combined standard uncertainty, u Degrees of freedom, ν eff (Welch-Satterthwaite formula) Confidence level, p Coverage factor, k (calculated with t-student) Expanded uncertainty, U = ku Relative expanded uncertainty, U rel = U/g 4.6E-08 m s % E-08 m s E-09 24
25 REFERENCES [1] D Agostino,G., Development and Metrological Characterization of a New Transportable Absolute Gravimeter, PhD Thesis, [2] Cerutti,G., Cannizzo,L., Sakuma,A., Hostache, J., A transportable apparatus for absolute gravity measurements, In: VDI-Berichte n. 212, 1974: p. 49. [3] Germak,A., Desogus,S., Origlia,C., Interferometer for the IMGC rise-and-fall absolute gravimeter", In: Metrologia, Special issue on gravimetry, Bureau Int Poids Mesures, BIPM, Pavillon De Breteuil, F-92312, Sèvres Cedex, France, 2002, Vol. 39, Nr. 5, pp [4] D Agostino,G., Germak,A., Desogus,S., Barbato,G. A Method to Estimate the Time-Position Coordinates of a Free-Falling Test-Mass in Absolute Grvimetry, In: Metrologia Vol. 42, No. 4, pp , August [5] Guide to the Expression of Uncertainty in Measurement, BIPM, IEC, ISCC, ISO, IUPAC, IUPAP, OIML, ISO,
Istituto Nazionale di Ricerca Metrologica
Istituto Nazionale di Ricerca Metrologica G. D Agostino (1), S. Desogus (1), A. Germak (1), C. Origlia (1), D. Quagliotti (1), G. Celli (2) and O. Francis (3) MEASUREMENTS OF THE ACCELERATION DUE TO GRAVITY
More informationIstituto Nazionale di Ricerca Metrologica
Istituto Nazionale di Ricerca Metrologica G. D Agostino, S. Desogus, A. Germak, C. Origlia and D. Quagliotti ABSOLUTE MEASUREMENTS OF THE FREE-FALL ACCELERATION g IN THE REPUBLIC OF PANAMA T.R. 111 March
More informationAbsolute and Relative Gravity Measurements at the Istituto Nazionale di Ricerca Metrologica (INRIM) of Turin (Italy) in June 2006
Absolute and Relative Gravity Measurements at the Istituto Nazionale di Ricerca Metrologica (INRIM) of Turin (Italy) in June 2006 Final Report September 2006 Prof. Dr. Olivier Francis University of Luxembourg
More informationAbsolute Gravity Observation at the National Measurement Institute
Record 2013/30 GeoCat 76733 Absolute Gravity Observation at the National Measurement Institute 18th June 2008 Nicholas Dando, Ray Tracey APPLYING GEOSCIENCE TO AUSTRALIA S MOST IMPORTANT CHALLENGES www.ga.gov.au
More informationTo the best of our knowledge, the FG5 gravimeter represents the current state-of-the-art in the measurement of absolute gravity.
FG5 gravity meter To the best of our knowledge, the FG5 gravimeter represents the current state-of-the-art in the measurement of absolute gravity. American Geophysical Union, EOS Journal Since its introduction
More informationABSOLUTE GRAVITY MEASUREMENTS
ABSOLUTE GRAVITY MEASUREMENTS AT THE CONRAD OBSERVATORIUM IN AUSTRIA IN JUNE 2012 Final Report August 2012 Prof. Dr. Olivier Francis University of Luxembourg Faculty of Sciences, Technology and Communication
More informationFinal Report. January Prof. Dr. Olivier Francis
ABSOLUTE AND RELATIVE GRAVITY MEASUREMENTS IN NUUK (GREENLAND) IN JULY 2005 Final Report January 2006 Prof. Dr. Olivier Francis University of Luxembourg Faculty of Sciences, Technology and Communication
More informationFG5 Absolute Gravimeter
FG5 Absolute Gravimeter Micro-g LaCoste Derek van Westrum, Ph.D. www.microglacoste.com derek@microglacoste.com FG5 Specifications Accuracy: 2 μgal (observed agreement between FG5 instruments) Precision:
More informationABSOLUTE GRAVITY MEASUREMENTS IN LOMONOSOV, SAINT-PETERSBURG, RUSSIA
ABSOLUTE GRAVITY MEASUREMENTS IN LOMONOSOV, SAINT-PETERSBURG, RUSSIA FINAL REPORT NOVEMBER 2004 Prof. Dr. Olivier Francis University of Luxembourg Faculty of Sciences, Technology and Communication Campus
More informationTest and Measurements conference 2016 DESIGN AND DEVELOPMENT OF FREE FALL ABSOLUTE GRAVITY MEASURING SYSTEM USING PNEUMATIC ACTUATORS
Test and Measurements conference 2016 DESIGN AND DEVELOPMENT OF FREE FALL ABSOLUTE GRAVITY MEASURING SYSTEM USING PNEUMATIC ACTUATORS Author: T Mokobodi Supervisor: Prof N Theron Co-authors: Oelof Kruger,
More informationIntermediary Report - August Prof. Dr. Olivier Francis and Prof. Dr. Tonie Van Dam
ABSOLUTE GRAVITY MEASUREMENTS IN YELLOWSTONE IN SEPTEMBER 2009 Intermediary Report - August 2010 Prof. Dr. Olivier Francis and Prof. Dr. Tonie Van Dam University of Luxembourg Faculty of Sciences, Technology
More informationResults of the Seventh International Comparison of Absolute Gravimeters ICAG-2005 at the Bureau International des Poids et Mesures, Sèvres
Chapter 7 the Seventh International Comparison of Absolute Gravimeters ICAG-2005 at the Bureau International des Poids et Mesures, Sèvres L. Vitushkin, Z. Jiang, L. Robertsson, M. Becker, O. Francis, A.
More informationRenewal of the gage-block interferometer at INRIM
Renewal of the gage-block interferometer at INRIM R. Bellotti, M. Franco, G. B. Picotto* and M. Pometto Istituto Nazionale di Ricerca Metrologica (INRIM), Strada delle Cacce 73, 10135 Torino, Italy * Corresponding
More informationRECOMMENDATION 1 (CI-2002): Revision of the practical realization of the definition of the metre
194 91st Meeting of the CIPM RECOMMENDATION 1 (CI-2002): Revision of the practical realization of the definition of the metre The International Committee for Weights and Measures, recalling that in 1983
More informationOptical Metrology Applications at TAS-I in support of Gravity and Fundamental Physics
Optical Metrology Applications at TAS-I in support of Gravity and Fundamental Physics Template reference : 100181670S-EN Stefano Cesare, Thales Alenia Space Italia, Torino Workshop GG/GGG: state of the
More informationby A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +
Bilateral Comparison of 1.018 V and 10 V Standards between the INTI (Argentina) and the BIPM, August to October 2009 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) by A.Tonina*, R.Iuzzolino*,
More informationComparison of 3 absolute gravimeters based on different methods for the e-mass project
1 Comparison of 3 absolute gravimeters based on different methods for the e-mass project Anne Louchet-Chauvet, Sébastien Merlet, Quentin Bodart, Arnaud Landragin and Franck Pereira Dos Santos, Henri Baumann,
More informationThe Stanford Gravitational Reference Sensor
The Stanford Gravitational Reference Sensor S. Buchman, B. Allard, G. Allen, R. Byer, W. Davis, D. DeBra, D. Gill, J. Hanson, G.M. Keiser, D. Lauben, I. Mukhar, N. A. Robertson, B. Shelef, K. Sun, S. Williams
More informationMachine Positioning Uncertainty with Laser Interferometer Feedback
Machine Positioning Uncertainty with Laser Interferometer Feedback The purpose of this discussion is to explain the major contributors to machine positioning uncertainty in systems with laser interferometer
More informationINTERNAL THALES ALENIA SPACE
Workshop Galileo Galilei (GG) and GGG lab prototype: state of the art and new possibilities Template reference : 100181670S-EN GG error budget from the end-to-end simulator developed at TAS-I Giuseppe
More informationAbsolute gravity measurements with a cold atom gravimeter
Absolute gravity measurements with a cold atom gravimeter Anne Louchet-Chauvet, Sébastien Merlet, Quentin Bodart, Tristan Farah, Arnaud Landragin, Franck Pereira Dos Santos LNE-SYRTE Observatoire de Paris
More informationTYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE
TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed
More informationConstruction of an absolute gravimeter using atom interferometry with cold 87. Rb atoms
Construction of an absolute gravimeter using atom interferometry with cold 87 Rb atoms Patrick Cheinet Julien Le Gouët Kasper Therkildsen Franck Pereira Dos Santos Arnaud Landragin David Holleville André
More informationSYRTE - IACI. AtoM Interferometry dual Gravi- GradiOmeter AMIGGO. from capability demonstrations in laboratory to space missions
SYRTE - IACI AtoM Interferometry dual Gravi- GradiOmeter AMIGGO from capability demonstrations in laboratory to space missions A. Trimeche, R. Caldani, M. Langlois, S. Merlet, C. Garrido Alzar and F. Pereira
More informationLast Revision: August,
A3-1 HALL EFFECT Last Revision: August, 21 2007 QUESTION TO BE INVESTIGATED How to individual charge carriers behave in an external magnetic field that is perpendicular to their motion? INTRODUCTION The
More informationSpeed of Light in Air
Speed of Light in Air Electromagnetic waves represent energy in the form of oscillating electric and magnetic fields which propagate through vacuum with a speed c = 2.9979246x10 8 m/s. Electromagnetic
More informationExperiment Design and Performance. G. Catastini TAS-I (BUOOS)
Experiment Design and Performance G. Catastini TAS-I (BUOOS) 10 EP and the GRT Einstein s General Relativity Theory Weak Equivalence Principle: all test particles at the same space-time point in a given
More informationThe CLIC active prealignment studies H. MAINAUD DURAND, T. TOUZE CERN
The CLIC active prealignment studies H. MAINAUD DURAND, T. TOUZE CERN Overview The alignment of CLIC Steps of alignment The active prealignment The situation of the studies on the active prealignment Studies
More informationFinal Report August 2010
Bilateral Comparison of 100 pf Capacitance Standards (ongoing BIPM key comparison BIPM.EM-K14.b) between the CMI, Czech Republic and the BIPM, January-July 2009 J. Streit** and N. Fletcher* *Bureau International
More informationA guide to expression of uncertainty of measurements QA4EO-QAEO-GEN-DQK-006
A guide to expression of uncertainty of measurements QA4EO-QAEO-GEN-DQK-006 A guide to expression of uncertainty of measurements Author: E-mail: Nigel Fox Nigel.Fox@npl.co.uk Editor: E-mail: Marie-Claire
More informationThe BIPM Watt Balance
CCM Workshop on the mise en pratique of the new definition of the kilogram 21 November 2012 The BIPM Watt Balance Hao FANG Mass Department Outline Motivations Main features Current status Future improvement
More informationLISA Pathfinder measuring pico-meters and femto-newtons in space
LISA Pathfinder measuring pico-meters and femto-newtons in space M Hewitson for the LPF team Barcelona, February 15th 2012 Observing from Space 2 Observing from Space 2 Observing from Space Push down to
More informationReviewers' comments: Reviewer #1 (Remarks to the Author):
Reviewers' comments: Reviewer #1 (Remarks to the Author): In their manuscript Dual Matter-Wave Inertial Sensors in Weightlessness Brynle Barrett et al. describe an dual species atom interferometer experiment
More informationTHE ACTIVE PREALIGNMENT OF THE CLIC COMPONENTS H. MAINAUD DURAND, T. TOUZE CERN
THE ACTIVE PREALIGNMENT OF THE CLIC COMPONENTS H. MAINAUD DURAND, T. TOUZE CERN Overview Introduction : the CLIC study The alignment of CLIC Steps of alignment The active prealignment The situation of
More informationCESIUM ATOMIC FOUNTAIN CLOCKS AT NMIJ
CESIUM ATOMIC FOUNTAIN CLOCKS AT NMIJ A. Takamizawa 1, S. Yanagimachi 1, Y. Shirakawa 2, K. Watabe 1, K. Hagimoto 1, and T. Ikegami 1 1 National Metrology Institute of Japan (NMIJ), AIST 2 Tokyo University
More informationLarge Momentum Beamsplitter using Bloch Oscillations
Large Momentum Beamsplitter using Bloch Oscillations Pierre Cladé, Saïda Guellati-Khélifa, François Nez, and François Biraben Laboratoire Kastler Brossel, UPMC, Ecole Normale Supérieure, CNRS, 4 place
More informationABSOLUTE AND RELATIVE GRAVITY MEASUREMENTS AT ETNA VOLCANO (ITALY)
ABSOLUTE AND RELATIVE GRAVITY MEASUREMENTS AT ETNA VOLCANO (ITALY) F. Greco 1, G. Currenti 1, G. D Agostino 2, C. Del Negro 1, A. Di Stefano 1, A. Germak 2, R. Napoli 1, C. Origlia 2, A. Pistorio 1, 3,
More informationBureau International des Poids et Mesures, Sèvres, France. *KEBS, Kenya Bureau of Standards P.O. Box , Nairobi, Kenya
Bilateral Comparison of 1.018 V and 10 V Standards between the KEBS (Kenya) and the BIPM January to March 2018 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) by S. Solve +, R. Chayramy +,
More informationAbsolute gravimetry. Gabriel Strykowski Geodynamics Department DTU-Space (formerly Danish National Space Center)
Absolute gravimetry by Gabriel Strykowski Geodynamics Department DTU-Space (formerly Danish National Space Center) gs@space.dtu.dk Contents 1. Instrumentation 2. Absolute gravity measurements: applications
More informationTraceable Mass Determination and Uncertainty Calculation
Traceable Mass Determination and Uncertainty Calculation Falko Hilbrunner, Thomas Fehling, Siego Mühlich, Mario Schreiber Sartorius Lab Instruments GmbH Weender Landstrasse 94-108 Goettingen, Germany ABSTRACT
More informationResults from a new interferometric ballistic differential gravity meter
BOLLETTINO DI GEOFISICA TEORICA ED APPLICATA VOL. 40, N. 3-4, pp. 487-496; SEP.-DEC. 1999 Results from a new interferometric ballistic differential gravity meter J. M. BROWN (1),T.M.NIEBAUER (2) and F.
More informationBRITISH PHYSICS OLYMPIAD BPhO Round 1 Section 2 18 th November 2016
BRITISH PHYSICS OLYMPIAD 2016-17 BPhO Round 1 Section 2 18 th November 2016 Instructions This question paper must not be taken out of the exam room. Time: 1 hour 20 minutes on this section. Questions:
More informationGravimètre Quantique Absolu : une utilisation opérationnelle des atomes froids pour la mesure de gravité. Dr. Jean Lautier-Gaud Journée CNFGG
Gravimètre Quantique Absolu : une utilisation opérationnelle des atomes froids pour la mesure de gravité Dr. Jean Lautier-Gaud Journée CNFGG Outline Brief overview of Muquans Absolute gravimetry with cold
More informationFORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).
FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO). Jorge C. Torres-Guzmán*, Amritlal Sawla**, Daniel Ramírez-Ahedo* * Centro Nacional de Metrología, México. **Physikalisch-Technische
More informationNIST ELECTROSTATIC FORCE BALANCE EXPERIMENT
NIST ELECTROSTATIC FORCE BALANCE EXPERIMENT John A. Kramar, David B. Newell, and Jon R. Pratt National Institute of Standards and Technology, Gaithersburg, MD, USA We have designed and built a prototype
More informationCharacterization of an Atom Interferometer Gravimeter with Classical Sensors for the Use in Geodesy and Geophysics
Characterization of an Atom Interferometer Gravimeter with Classical Sensors for the Use in Geodesy and Geophysics M. Schilling C. Freier V. Schkolnik M. Hauth H. Wziontek H.-G. Scherneck A. Peters and
More informationABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER
ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2),
More informationBloch oscillations of ultracold-atoms and Determination of the fine structure constant
Bloch oscillations of ultracold-atoms and Determination of the fine structure constant Pierre Cladé P. Cladé Bloch oscillations and atom interferometry Sept., 2013 1 / 28 Outlook Bloch oscillations of
More informationTypical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge
Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge 2090TN05 Revised 2004 March 26 Technical Note FPG8601 is a pressure standard designed to cover the range from
More informationA beam of coherent monochromatic light from a distant galaxy is used in an optics experiment on Earth.
Waves_P2 [152 marks] A beam of coherent monochromatic light from a distant galaxy is used in an optics experiment on Earth. The beam is incident normally on a double slit. The distance between the slits
More informationSCI404: Honors Physics
SCI404: Honors Physics This advanced course surveys all key areas: physical systems, measurement, kinematics, dynamics, momentum, energy, thermodynamics, waves, electricity, and magnetism, and introduces
More informationBureau International des Poids et Mesures
Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de
More informationDevelopement and operation of an electro-optical gravitational waves detector simulator as part of the space mission elisa/ngo
gravitational space mission PhD student : Pierre Gruning Thesis director : Hubert Halloin Contents gravitational Contents gravitational Contents gravitational Contents gravitational Plan gravitational
More informationIntercomparison of Relative and Absolute Gravimeters. Olivier FRANCIS
Intercomparison of Relative and Absolute Gravimeters Olivier FRANCIS The right gravimeter for the right job (signal to be detected) Absolute/relative Stability (accuracy)/precision (repetability) Spatial
More informationRevision of the Guide to the expression of uncertainty in measurement impact on national metrology institutes and industry
Revision of the Guide to the expression of uncertainty in measurement impact on national metrology institutes and industry Maurice Cox/Peter Harris National Physical Laboratory, Teddington, UK CCRI BIPM,
More informationStandard Practices for Air Speed Calibration Testing
Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument
More informationTowards an improved ILRS TRF contribution
Towards an improved ILRS TRF contribution Erricos C. Pavlis ILRS Analysis Coordinator JCET/ & NASA Goddard IERS Workshop on Conventions 2007 20-21 Sept. 2007, Sèvres, France Overview The ILRS Network Geometry
More informationState of the art cold atom gyroscope without dead times
State of the art cold atom gyroscope without dead times Remi Geiger SYRTE, Observatoire de Paris GDR IQFA Telecom Paris November 18 th, 2016 I. Dutta, D. Savoie, B. Fang, B. Venon, C. L. Garrido Alzar,
More informationLab 10 - Harmonic Motion and the Pendulum
Lab 10 Harmonic Motion and the Pendulum L10-1 Name Date Partners Lab 10 - Harmonic Motion and the Pendulum L (measured from the suspension point to the center of mass) Groove marking the center of mass
More informationPioneer anomaly: Implications for LISA?
Pioneer anomaly: Implications for LISA? Denis Defrère Astrophysics and Geophysics Institute of Liege (Belgium) Andreas Rathke EADS Astrium GmbH Friedrichshafen (Germany) ISSI Meeting - Bern November 10th
More informationPrecision neutron flux measurement with a neutron beam monitor
Journal of Physics: Conference Series OPEN ACCESS Precision neutron flux measurement with a neutron beam monitor To cite this article: T Ino et al 2014 J. Phys.: Conf. Ser. 528 012039 View the article
More informationReview of Anemometer Calibration Standards
Review of Anemometer Calibration Standards Rachael V. Coquilla rvcoquilla@otechwind.com Otech Engineering, Inc., Davis, CA Anemometer calibration defines a relationship between the measured signals from
More informationThe Kilogram Redefined:
The Kilogram Redefined: SI Units and Fundamental Constants Alan Steele, NRC Canada IPAC 18 Vancouver May 2, 2018 The Metre Convention: May 20, 1875 International Treaty that creates the BIPM and launches
More information2 Each satellite will have two test masses, each being the end mirror for an interferometer.
Ground Testing for LISA Test Masses with a Torsion Pendulum Matthew Schmidt Valdosta State University International REU: University of Trento, Italy Advisor: Dr. Bill Weber Abstract: One of the most important
More informationIndex. Cambridge University Press Data Analysis for Physical Scientists: Featuring Excel Les Kirkup Index More information
χ 2 distribution, 410 χ 2 test, 410, 412 degrees of freedom, 414 accuracy, 176 adjusted coefficient of multiple determination, 323 AIC, 324 Akaike s Information Criterion, 324 correction for small data
More informationDr. Jean Lautier-Gaud October, 14 th 2016
New generation of operational atomic clock: what perspectives for radio-astronomy & VLBI? Dr. Jean Lautier-Gaud October, 14 th 2016 Courtesy of Noel Dimarcq, SYRTE Content 1. Why is Muquans here? 2. What
More informationAtom Interferometry II. F. Pereira Dos Santos
Atom Interferometry II F. Pereira Dos Santos Organization of the lecture 1 : Applications of inertial sensors/gravimetry : A case study : the SYRTE atom gravimeter 3 : Towards more compact sensors 4 :
More informationUNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Advanced Subsidiary Level and Advanced Level
*1945541248* UNIVERSITY OF MRIGE INTERNTIONL EXMINTIONS General ertificate of Education dvanced Subsidiary Level and dvanced Level PHYSIS 972/1 Paper 1 Multiple hoice October/November 27 dditional Materials:
More informationTechnical Protocol of the CIPM Key Comparison CCAUV.V-K5
Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative
More informationZeeman Effect. Alex Povilus Physics 441- Fall 2003 December 20, 2003
Zeeman Effect Alex Povilus Physics 441- Fall 2003 December 20, 2003 Abstract The Zeeman Effect is observed by application of a strong magnetic field to a mercury vapor cell and exciting transitions by
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 16063-15 First edition 2006-08-01 Methods for the calibration of vibration and shock transducers Part 15: Primary angular vibration calibration by laser interferometry Méthodes
More informationCurriculum Map-- Kings School District Honors Physics
Curriculum Map-- Kings School District Honors Physics Big ideas Essential Questions Content Skills/Standards Assessment + Criteria Activities/Resources Motion of an object can be described by its position,
More informationAtom interferometry in microgravity: the ICE project
Atom interferometry in microgravity: the ICE project (4) G. Stern 1,2, R. Geiger 1, V. Ménoret 1,B. Battelier 1, R. Charrière 3, N. Zahzam 3, Y. Bidel 3, L. Mondin 4, F. Pereira 2, A. Bresson 3, A. Landragin
More informationSqueezed Light for Gravitational Wave Interferometers
Squeezed Light for Gravitational Wave Interferometers R. Schnabel, S. Chelkowski, H. Vahlbruch, B. Hage, A. Franzen, and K. Danzmann. Institut für Atom- und Molekülphysik, Universität Hannover Max-Planck-Institut
More informationGROUND NOISE STUDIES USING THE TAMA300 GRAVITATIONAL-WAVE DETECTOR AND RELATED HIGHLY-SENSITIVE INSTRUMENTS
GROUND NOISE STUDIES USING THE TAMA300 GRAVITATIONAL-WAVE DETECTOR AND RELATED HIGHLY-SENSITIVE INSTRUMENTS Akito ARAYA Earthquake Research Institute, University of Tokyo, 1-1-1, Yayoi, Bunkyo, Tokyo 113-0032,
More informationIntegrated Physics & Chemistry A
Integrated Physics & Chemistry A Credit by Exam This Credit by Exam can help you prepare for the exam by giving you an idea of what you need to study, review, and learn. To succeed, you should be thoroughly
More informationby S. Solve +, R. Chayramy +, M. Stock +, D. Vlad*
Bilateral Comparison of 1 V and 10 V Standards between the SMD (Belgium) and the BIPM, October to November 2014 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) by S. Solve +, R. Chayramy
More informationKey comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays
Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays C Kessler 1, D T Burns 1, P Roger 1, V Sochor 2 1 Bureau International des Poids et
More informationScience Curriculum Matrix
Science Curriculum Matrix Physics Version 1.0 beta June 2, 2008 This curriculum (core matrix) document will eventually become part of the Science Curriculum Matrix. We envision the Science Curriculum Matrix
More informationPA04/2 (JUN08PA0401) General CertiÞ cate of Education June 2008 Advanced Level Examination
Surname Other Names For Examiner s Use Centre Number Candidate Number Candidate Signature General CertiÞ cate of Education June 2008 Advanced Level Examination PHYSICS (SPECIFICATION A) Unit 4 Waves, Fields
More informationarxiv: v1 [physics.optics] 6 Jul 2009
v8 Measuring the Molecular Polarizability of Air M.J. Madsen, D.R. Brown, S.R. Krutz, and M.J. Milliman Department of Physics, Wabash College, Crawfordsville, IN 47933 (Dated: July 7, 2009) Abstract arxiv:0907.0782v1
More informationI. Yang, C. H. Song, Y.-G. Kim & K. S. Gam
Cryostat for Fixed-Point Calibration of Capsule-Type SPRTs I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam International Journal of Thermophysics Journal of Thermophysical Properties and Thermophysics and Its
More informationPractice Final C. 1. The diagram below shows a worker using a rope to pull a cart.
1. The diagram below shows a worker using a rope to pull a cart. 6. The graph below represents the relationship between gravitational force and mass for objects near the surface of Earth. The worker s
More informationLab 8: Gravity and Isostasy (35 points)
it's not the most important thing in your life right now. But what is important is gravity. Arnold Schwarzenegger as Colonel John Matrix, Commando (Check out this classic of American cinema!) Lab 8: Gravity
More informationMODERN INTERFEROMETRY
MODERN INTERFEROMETRY * No actual laser photons were harmed during the filming of this cover. Build Michelson, Sagnac, and Mach-Zehnder Interferometers Generate and Count Interference Fringes Manually
More informationCurriculum Map-- Kings School District- Physics
Curriculum Map-- Kings School District- Physics Big ideas Essential Questions Content Skills/Standards Assessment + Criteria Activities/Resources Motion of an object can be described by its position, speed,
More informationIn vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL
In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL J-B. Pruvost 1, P. Berkvens 2, F. Justine 1, C. Mage 1 1: Synchrotron SOLEIL BP.48, 91192
More information2. What are the 4 steps of the Scientific Method as described by Mr. Martin?
Ch.1 Study Guide Outline Study the Review that is posted on the website. Make a note card to use for the test. 1. What is science and physics? 2. What are the 4 steps of the Scientific Method as described
More informationProgress on Atom Interferometer (AI) in BUAA
Progress on Atom Interferometer (AI) in BUAA Group of Prof. FANG Jiancheng Beihang University ZHANG Yuchi, Hu Zhaohui, QI Lu, WANG Tongyu, WANG Tao 01.09.2011 7 th UK-China Workshop on Space Science and
More informationAutoresonant Ion Trap Mass Spectrometer The RGA Alternative
Autoresonant Ion Trap Mass Spectrometer The GA Alternative 1 What is a Mass Spectrometer? Ionizer Mass Spectrometer Mass Separator Magnetic Sector, Quadrupole or Ion Trap Detector Partial Pressure Measurement
More informationLAB 10: HARMONIC MOTION AND THE PENDULUM
163 Name Date Partners LAB 10: HARMONIC MOION AND HE PENDULUM Galileo reportedly began his study of the pendulum in 1581 while watching this chandelier swing in Pisa, Italy OVERVIEW A body is said to be
More informationHybrid Atom-Optical Interferometry for Gravitational Wave Detection and Geophysics
Hybrid Atom-Optical Interferometry for Gravitational Wave Detection and Geophysics Remi Geiger, SYRTE for the MIGA consortium EGAS 46, July 3rd 2014, Lille, France http://syrte.obspm.fr/tfc/capteurs_inertiels
More informationWhich one of the following correctly describes the velocities of the two bodies after the collision?
Q1.In which of the following do both quantities have the same unit? Electrical resistivity and electrical resistance. Work function Planck constant Pressure and the Young modulus. cceleration and rate
More informationSupplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688
Supplementary Comparison EURAMET.EMS19 EURAMET Project No. 688 Bilateral Comparison of Measurements of Current Transformers (CTs) Between and Final Report Hüseyin Çaycı (, Pilot Laboratory) January 211
More information1 General Definitions and Numerical Standards
This chapter provides general definitions for some topics and the values of numerical standards that are used in the document. Those are based on the most recent reports of the appropriate working groups
More informationPHY410 Optics Exam #3
PHY410 Optics Exam #3 NAME: 1 2 Multiple Choice Section - 5 pts each 1. A continuous He-Ne laser beam (632.8 nm) is chopped, using a spinning aperture, into 500 nanosecond pulses. Compute the resultant
More informationExperimental AMO eets meets M odel Model Building: Part I (Precision Atom Interferometry)
Experimental AMO meets Model Building: Part I (Precision Atom Interferometry) Interference of Rb atoms Chiow, et. al, PRL, 2011 Young s double slit with atoms Young s 2 slit with Helium atoms Interference
More informationCatapult tests for microgravity characterization of the MICROSCOPE accelerometers. Manuel Rodrigues On behalf ONERA & ZARM team
Catapult tests for microgravity characterization of the MICROSCOPE accelerometers Manuel Rodrigues mrodrig@onera.fr On behalf ONERA & ZARM team 1 Instrument Description SU sqm Sensor Unit (SU) = differential
More informationComparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2
Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,
More informationIntroduction to geodetic VLBI
Introduction to geodetic VLBI David Mayer Andreas Hellerschmied Johannes Böhm Harald Schuh and Johannes Böhm, Very Long Baseline Interferometry for Geodesy and Astrometry, in Guochang Xu (editor): Sciences
More information