1 Radio Frequency Interactions with Medical Devices Justin Peterson University of Western Ontario MBP 3970Z April 1, 2011

2 Introduction Magnetic Resonance Imaging (MRI) systems are an extremely important technique when considering the internal structure of an object, in this case a human. If a person experiences an accident and suffers a severely broken joint a device is used to stabilize the joint to ensure proper healing. If the accident is severe enough an MRI is needed to check for damage in other areas that could otherwise go unnoticed by a physician. Human tissue naturally undergoes heating near the surface of the skin during an MRI but an electrically conductive, elongated device could concentrate the Radio Frequencies (RF) and cause damaging effects [1]. To prevent damaging the tissue of the patient, tests must be performed on each device to ensure the degree of heating is acceptable. The factors affecting the rate of heating in the device include the strength of the magnet, the geometry and composition of the device, and the position in the bore. The testing should focus on the worst case scenario that the device could experience during an MRI scan. The first objective of this study was to determine the relationship between radial position in the bore and rate of temperature change of the device; the second objective was to determine the ideal worst case scenario for future testing. Theory The three main components of an MRI system are the main magnet, the gradient coils and the RF coils. The main magnet produces a uniform magnetic field along the length of the bore; this field aligns the hydrogen nuclei in what is to be imaged. The gradient coils produce a magnetic field that varies in strength along a plane (in any orientation) of the operator s choice, which make the strength of the magnetic field dependent on position in the bore. The RF coils produce a pulse, which causes the hydrogen nuclei to precess about an axis; the nuclei will precess at different speeds according to the position on the plane produced by the gradient coils. The nuclei release energy as they stop precessing; the RF coils detect this energy and send information to a computer, which builds the image [1]. The MRI was a 3T Trio model manufactured by Siemens. A phantom is a general representation of the body on the MRI bed; it is large enough for the device to be placed representative of where it would be in a real scenario. The phantom is made of an electrical insulator and is non-magnetic; it must be filled with a material that has a conductivity similar to that of human tissue. The gel that filled the phantom had a conductivity of 4.80 ms/cm. The device is comprised of two stainless steel pins connected by a carbon fiber core surrounded in a fiberglass sheath. The stainless steel pins have threads in one end, which are to anchor the device into bone. The carbon fiber/fiberglass rod connects the pins outside the body at some distance from the skin surface. 2

3 Methods Before the tests were performed the phantom gel was made using distilled water, Sodium Chloride (NaCl), and Poly-acrylic acid, partial sodium salt (PAA). Pour 8 L of distilled water into a container and measure the conductivity (should be 0 ms/cm). Next, 1.32 g/l NaCl was added to the distilled and mixed thoroughly and the conductivity was measured to be 2.40 ms/cm. Next, 10.0 g/l of PAA was mixed into the water/nacl solution. The entire solution was mixed for a total of 10 minutes (until the mixture was completely smooth ); this conductivity was measured to be 4.80 ms/cm. The part and batch numbers of the PAA were kg and MKAA3066, respectively. The part and batch numbers of the NaCl were and This process was repeated five times until a total of 40 L of gel was made, 30 L of the gel was poured into the phantom to a depth of 10 cm. The probes used to measure the change in temperature during the tests were optical sensors accurate to 0.2 C and a program named Neoptix collected the data. There were eight temperature sensors placed throughout the phantom gel and the device. Figure 1: Temperature Probe Placement 1: Tip of the furthest pin in the gel 2: Tip of the closest pin in the gel 3: Mid-way up the furthest pin out of the gel 4: Mid-way up the closest pin out of the gel 5: Along the left side of the phantom 6: Outside of the phantom (Ambient) 7: Along the right side of the phantom 8: Outside of the phantom (Ambient) Figure 2: The device was placed in five different radial positions: -240mm, -210mm, - 160mm, 0mm, and +230mm with 0 being the center of the MRI bore. In all five tests the device was anchored with a space of 550 mm from the bottom of the bar to the phantom gel. 3

4 Results All the numbers and figures in this study were calculated and plotted using Microsoft Excel Temperature ( C) Temperature Change -160 mm From Centre of Bore Sensor 1 Sensor 2 Sensor 3 Sensor 4 Sensor 5 Sensor 6 Sensor Sensor 8 Time (s) Figure 3: The temperature change at eight probes placed on the device, in the gel and in the surrounding environment. The start and end of RF exposure are at 60 and 240 seconds, respectively. Temperature( C) Temperature Change +230 mm from centre of Bore Time (s) Sensor 1 Sensor 2 Sensor 3 Sensor 4 Sensor 5 Sensor 6 Sensor 7 Sensor 8 Figure 4: The temperature change at eight probes placed on the device, in the gel and in the surrounding environment. The start and end of RF exposure are at 60 and 140 seconds, respectively. 4

5 Varying Radial Position of Device on Temperature Temperature ( C) mm mm mm 0 mm mm Time (s) Figure 5: Sensor 1 data from five tests at different radial positions along the bore of the MRI. Data was collected during the first 20 seconds of each test. dt/dt ( C/s) Rate of Temperature Change Dependency on Radial Position in MR Bore Radial Position in Bore (mm) Figure 6: Rate of temperature change at five different positions along the radius of the bore calculated from Sensor 1 data during the first 20 seconds of data collection. 5

6 In Figures 3 and 4, there was 60 seconds of steady temperature data before the RF exposure was turned on. In Figure 3 the device pin temperature increased at a rate of C/s while in Figure 4 the device pin increased at a rate of C/s. Almost instantly after the exposure stopped, the rate of temperature change began decreasing. Figure 5 depicts the temperature change with time for five radial positions along the MRI bore, using the first 20 seconds of this data the rate of temperature change for each position was calculated. Figure 6 shows spatially how the temperature rate of change varies with radial position along the bore. Discussion In this study, the effect of radial position on rate of temperature change due to radio frequency was tested. Between all eight of the optical temperature probes, Sensor 1 and 2 showed a rise in temperature large enough to look at (Figure 3). For simplicity, only Sensor 1 was analyzed because Sensor 2 had a smaller rate of temperature change. This is consistent with a study by the American Society for Testing and Materials (2009) [1] ; the ends of an elongated device will experience the greatest amount of heating. Figure 1 shows the position of Sensors 1 and 2 as the ends of the device, although Sensors 3 and 4 were also near the ends of the device but they did not undergo heating comparable with the first two sensors. This could be due to the induced electric field from the gradient coils, which cause an induced current around the loop of the device. This induced current travels around the device from one conductive pin to the bar to the other conductive pin but when the electrons attempt to migrate to the other pin. The gel acts as a resistor in the circuit and as a result the pin and the gel in the vicinity experiences a greater amount of heating. If the circuit were flowing in a clockwise direction around the device, it would explain why Sensor 1 was hotter than Sensor 2. Sensor 1 was chosen as a common probe among all tests because it experienced the highest rate of temperature change among all probes. As seen in Figure 6, the highest rates of temperature change that were experienced by the device occur along the edge of the MRI bore (furthest away from the center). As the device position moved from the edge of the bore to the center the rate of heating decreased from a maximum of C/s to a minimum of C/s. A possible explanation for this is the induced electric field associated with the gradient coils of the MRI. The electric field produced by the gradient coils is related to the strength and direction of the rate of change of the magnetic field [2]. This electric field is generated along the length of the bore, which is why the device experiences the highest rate of heating in the configuration tested, with the bar parallel with the field. The electric field is strongest near the coils that produce them or furthest from the center of the bore. 6

7 There is a strong trend that suggests the worst case scenario for device testing along the radial position would be along the edge of the MRI bore. This was not significant because a statistical analysis could not be performed with a small sample size of one for each group of testing. If the sample size were increased, there would be a parabolic trend between radial position and rate of temperature change much stronger than that seen in Figure 6. An interesting area of study would be testing the device in different positions along the length of the MRI bore. Since the device experienced a greater amount of heating on the furthest pin, a test of the device farthest away from the torso might result in a greater amount of heating on both pins. These results are accurate for a Siemens 3T Trio device only, other MRI devices with differing magnetic field strength and manufacturing specifications will produce different results [1]. Conclusion The two objectives of this study were to determine the relationship between radial position in an MRI bore and rate of temperature change of a passive medical implant and to determine the ideal worst case scenario for future testing. As radial distance increased from the center of the bore, the rate of temperature change increased as a parabolic trend. A larger sample size could have produced significant results and a better parabolic trend. The worst case scenario for radial position is along the edge of the MR bore (farthest from the center). An interesting area of research would be to test the device along the length of the bore to determine where exactly the ideal worst case scenario would be for testing of similar devices. These results can be used to refine and update the current standards when testing passive medical devices in the future. 7

8 References [1] Miller, Stephen. Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging. Tech. no. F a. Pennsylvania, Print. [2] Glover, P. M., and R. Bowtell. "Measurement of Electric Fields Induced in a Human Subject Due to Natural Movements in Static Magnetic Fields or Exposure to Alternating Magnetic Field Gradients." Physics in Medicine and Biology 53.2 (2008): Web. 1 Apr

By: Brett Empringham. Supervisor: Dr. Blaine A. Chronik. Medical Biophysics 3970Z: Six Week Project. University of Western Ontario

By: Brett Empringham. Supervisor: Dr. Blaine A. Chronik. Medical Biophysics 3970Z: Six Week Project. University of Western Ontario Relationship Between Insertion Depth and MRI Induced Heating of External Stabilizing Brace By: Brett Empringham Supervisor: Dr. Blaine A. Chronik Medical Biophysics 3970Z: Six Week Project University of

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging History Nuclear magnetic resonance was first described by Isidor Rabi in 1938 - Columbia University, New York City, (Nobel Prize Nobel Prize in Physics 1944) 1946 - Edward Mills

More information

Magnetic resonance imaging MRI

Magnetic resonance imaging MRI Magnetic resonance imaging MRI Introduction What is MRI MRI is an imaging technique used primarily in medical settings that uses a strong magnetic field and radio waves to produce very clear and detailed

More information

AQA Physics /7408

AQA Physics /7408 AQA Physics - 7407/7408 Module 10: Medical physics You should be able to demonstrate and show your understanding of: 10.1 Physics of the eye 10.1.1 Physics of vision The eye as an optical refracting system,

More information

John R. de Bruyn, Leesa Fleury, Corey Rae McRae, Will Handler, Blaine Chronik Department of Physics and Astronomy Western University

John R. de Bruyn, Leesa Fleury, Corey Rae McRae, Will Handler, Blaine Chronik Department of Physics and Astronomy Western University John R. de Bruyn, Leesa Fleury, Corey Rae McRae, Will Handler, Blaine Chronik Department of Physics and Astronomy Western University the MR scanner the MR image High dc magnetic field Rapidly varying magnetic

More information

RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES

RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES 1. What are potential consequences to patients and personnel should there be a release of gaseous

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 11 pages Written test, 9 December 2010 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: none. "Weighting": All problems weight equally.

More information

NUCLEAR MAGNETIC RESONANCE. The phenomenon of nuclear magnetic resonance will be used to study magnetic moments of nuclei.

NUCLEAR MAGNETIC RESONANCE. The phenomenon of nuclear magnetic resonance will be used to study magnetic moments of nuclei. 14 Sep 11 NMR.1 NUCLEAR MAGNETIC RESONANCE The phenomenon of nuclear magnetic resonance will be used to study magnetic moments of nuclei. Theory: In addition to its well-known properties of mass, charge,

More information

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle L 36 Modern Physics [2] X-rays & gamma rays How lasers work Medical applications of lasers Applications of high power lasers Medical imaging techniques CAT scans MRI s Modern physics ideas are strange!

More information

Fundamental MRI Principles Module Two

Fundamental MRI Principles Module Two Fundamental MRI Principles Module Two 1 Nuclear Magnetic Resonance There are three main subatomic particles: protons neutrons electrons positively charged no significant charge negatively charged Protons

More information

/14/$ IEEE 113. A. Absorption Material. B. External Fixation Device and ASTM phantom

/14/$ IEEE 113. A. Absorption Material. B. External Fixation Device and ASTM phantom MRI Heating Reduction for External Fixation Devices Using Asorption Material Xin Huang, Jianfeng Zheng, Ji Chen ECE department University of Houston Houston, USA {xhuang12, jchen18}@uh.edu Xin Wu, Mari

More information

MRI Homework. i. (0.5 pt each) Consider the following arrangements of bar magnets in a strong magnetic field.

MRI Homework. i. (0.5 pt each) Consider the following arrangements of bar magnets in a strong magnetic field. MRI Homework 1. While x-rays are used to image bones, magnetic resonance imaging (MRI) is used to examine tissues within the body by detecting where hydrogen atoms (H atoms) are and their environment (e.g.

More information

Magnetic Resonance Imaging (MRI)

Magnetic Resonance Imaging (MRI) Magnetic Resonance Imaging Introduction The Components The Technology (MRI) Physics behind MR Most slides taken from http:// www.slideworld.org/ viewslides.aspx/magnetic- Resonance-Imaging- %28MRI%29-MR-Imaging-

More information

Fundamental MRI Principles Module 2 N. Nuclear Magnetic Resonance. X-ray. MRI Hydrogen Protons. Page 1. Electrons

Fundamental MRI Principles Module 2 N. Nuclear Magnetic Resonance. X-ray. MRI Hydrogen Protons. Page 1. Electrons Fundamental MRI Principles Module 2 N S 1 Nuclear Magnetic Resonance There are three main subatomic particles: protons positively charged neutrons no significant charge electrons negatively charged Protons

More information

MRI in Practice. Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK. John Talbot MSc, DCRR

MRI in Practice. Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK. John Talbot MSc, DCRR MRI in Practice Third edition Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK and Carolyn Kaut RothRT(R) (MR) (CT) (M) (CV) Fellow SMRT (Section for Magnetic

More information

Proc. Intl. Soc. Mag. Reson. Med. 22 (2014) 4866.

Proc. Intl. Soc. Mag. Reson. Med. 22 (2014) 4866. 4038 WORST-CASE ANALYSIS OF RF-INDUCED HEATING DURING MRI SCANNING IN A GENERIC MULTI- COMPONENT ORTHOPEDIC MEDICAL IMPLANT APPLYING THE DESIGN OF EXPERIMENT METHOD (DOE),2, Gregor Schaefers, Juan D. Sánchez,

More information

(INCLUDING THIS FRONT PAGE)

(INCLUDING THIS FRONT PAGE) I'IFIITIIBIFI UNIVERSITY OF SCIEI'ICE RITD TECHNOLOGY FACULTY OF HEALTH AND APPLIED SCIENCES DEPARTMENT OF NATURAL AND APPLIED SCIENCES QUALIFICATION: BACHELOR OF SCIENCE (MAJOR AND MINOR) QUALIFICATION

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging http://www.qldxray.com.au/filelibrary/mri_cardiovascular_system_ca_0005.jpg Magnetic Resonance Imaging 1 Overview 1. The magnetic properties of nuclei, and how they behave in strong magnetic fields. 2.

More information

1. Write the relation for the force acting on a charge carrier q moving with velocity through a magnetic field in vector notation. Using this relation, deduce the conditions under which this force will

More information

Introduction to Medical Imaging. Medical Imaging

Introduction to Medical Imaging. Medical Imaging Introduction to Medical Imaging BME/EECS 516 Douglas C. Noll Medical Imaging Non-invasive visualization of internal organs, tissue, etc. I typically don t include endoscopy as an imaging modality Image

More information

Science 30 Unit C Review Outline GCCHS. Negatively charged Positively charged Coulomb Conductor Electric potential difference

Science 30 Unit C Review Outline GCCHS. Negatively charged Positively charged Coulomb Conductor Electric potential difference Science 30 Unit C Review Outline GCCHS Negatively charged Positively charged Coulomb Conductor Electric potential difference volt voltage Insulator Test body Gravitational field Field lines Solar wind

More information

Magnetic Torque Physics 2150 Experiment No. 11 University of Colorado

Magnetic Torque Physics 2150 Experiment No. 11 University of Colorado Experiment 11 1 Introduction Magnetic Torque Physics 2150 Experiment No. 11 University of Colorado In this experiment, you will study how a magnet moment interacts with a magnetic field. You will measure

More information

Elec Eng 3BA3: Structure of Biological Materials

Elec Eng 3BA3: Structure of Biological Materials Elec Eng 3BA3: Structure of Biological Materials Page 1 of 12 Day Class Instructor: Dr. I. C. BRUCE Duration of Examination: 3 Hours McMaster University Final Examination December 5, 2008 This examination

More information

Physics 272. Prof. Douglas C. Hamilton Dec. 13, 2011

Physics 272. Prof. Douglas C. Hamilton Dec. 13, 2011 Physics 272 Prof. Douglas C. Hamilton Dec. 13, 2011 Final Exam 1. Tuesday, Dec. 20 1:30-3:30 pm 2. In this room. 3. Closed book and notes. Formula sheet provided. 4. Covers material from entire semester

More information

Chapter 1 - Introduction to Human Anatomy and Physiology

Chapter 1 - Introduction to Human Anatomy and Physiology Chapter 1 - Introduction to Human Anatomy and Physiology 1.3 - LEVELS OF ORGANIZATION Early investigators, limited in their ability to observe small structures such as cells, focused their attention on

More information

A method for imaging and spectroscopy. using γ-rays and magnetic resonance

A method for imaging and spectroscopy. using γ-rays and magnetic resonance A method for imaging and spectroscopy using γ-rays and magnetic resonance Y. Zheng et al. Nature 537, 19775 (Sep 2016) Parker Brue PHY 492 Abstract Background: Magnetic resonance imaging and Nuclear imaging

More information

ASSESSMENT UNIT PH5: ELECTROMAGNETISM, NUCLEI & OPTIONS. Use black ink or black ball-point pen. Do not use gel pen or correction fluid.

ASSESSMENT UNIT PH5: ELECTROMAGNETISM, NUCLEI & OPTIONS. Use black ink or black ball-point pen. Do not use gel pen or correction fluid. Surname Other Names Centre Number 2 Candidate Number GCE A level 1325/01 PHYSICS ASSESSMENT UNIT PH5: ELECTROMAGNETISM, NUCLEI & OPTIONS A.M. MONDAY, 18 June 2012 1 3 4 hours 1325 010001 ADDITIONAL MATERIALS

More information

MRI Physics I: Spins, Excitation, Relaxation

MRI Physics I: Spins, Excitation, Relaxation MRI Physics I: Spins, Excitation, Relaxation Douglas C. Noll Biomedical Engineering University of Michigan Michigan Functional MRI Laboratory Outline Introduction to Nuclear Magnetic Resonance Imaging

More information

Construction of a Variable Electromagnet and Gauss Meter

Construction of a Variable Electromagnet and Gauss Meter Construction of a Variable Electromagnet and Gauss Meter Tanushree Bezbaruah, Chinmoy Bharadwas, Reinkosi Mam, Barnali Devi, Diganta Kumar Sarma # Department of Physics, B. Borooah College, Ulubari, Guwahati-781007,

More information

Let us start with a general overview of MRI...

Let us start with a general overview of MRI... MRI made easy 5 Let us start with a general overview of MRI... The single steps of an MRI examination can be described quite simply: 1. the patient is placed in a magnet, 2. a radio wave is sent in, 3.

More information

Electrical Engineering 3BA3: Structure of Biological Materials

Electrical Engineering 3BA3: Structure of Biological Materials Electrical Engineering 3BA3: Structure of Biological Materials Day Class Instructor: Dr. I. C. BRUCE Duration of Examination: 3 Hours McMaster University Final Examination December, 2004 This examination

More information

You MUST TAKE THE FINAL, even if you are a senior!!! If you are sick that day, you will have to make it up before you are allowed to graduate!

You MUST TAKE THE FINAL, even if you are a senior!!! If you are sick that day, you will have to make it up before you are allowed to graduate! LAHS Physics 2006-2007 End-of-Year Final Review Problems Your End-of-Year Physics Final covers the material in Physics during semester two. Juniors and Seniors will take Finals on the SAME Senior Final

More information

Chapter 5. Electromagnetic Induction

Chapter 5. Electromagnetic Induction Chapter 5 Electromagnetic Induction Overview In the last chapter, we studied how a current produces a magnetic field. Here we will study the reverse effect: A magnetic field can produce an electric field

More information

ELECTRON SPIN RESONANCE & MAGNETIC RESONANCE TOMOGRAPHY

ELECTRON SPIN RESONANCE & MAGNETIC RESONANCE TOMOGRAPHY ELECTRON SPIN RESONANCE & MAGNETIC RESONANCE TOMOGRAPHY 1. AIM OF THE EXPERIMENT This is a model experiment for electron spin resonance, for clear demonstration of interaction between the magnetic moment

More information

9. Which of the following is the correct relationship among power, current, and voltage?. a. P = I/V c. P = I x V b. V = P x I d.

9. Which of the following is the correct relationship among power, current, and voltage?. a. P = I/V c. P = I x V b. V = P x I d. Name: Electricity and Magnetism Test Multiple Choice Identify the choice that best completes the statement. 1. Resistance is measured in a unit called the. a. ohm c. ampere b. coulomb d. volt 2. The statement

More information

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics Magnetic Resonance Imaging Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics pal.e.goa@ntnu.no 1 Why MRI? X-ray/CT: Great for bone structures and high spatial resolution Not so great

More information

6.5. RESEARCH ON THE CONDUCTIVITY OF VARIOUS WATER SAMPLES

6.5. RESEARCH ON THE CONDUCTIVITY OF VARIOUS WATER SAMPLES 6.5. RESEARCH ON THE CONDUCTIVITY OF VARIOUS WATER SAMPLES Purpose of experiment Determine the conductivity of various water samples. Tasks of experiment: Determine the average conductivity of tap water,

More information

P3.1. Medical Applications Higher Tier. Question. Check. the. Command word. Link sentences. with a range of. Connectives. Make sure you include

P3.1. Medical Applications Higher Tier. Question. Check. the. Command word. Link sentences. with a range of. Connectives. Make sure you include P3.1 Medical Applications Higher Tier Route: Question-Draft 1 Question Check the Command word Link sentences with a range of Connectives Make sure you include the following Keywords Edit your first draft

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 10 pages Written test, 12 December 2012 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: None. Pocket calculator not allowed "Weighting":

More information

Student Sheet PA.1: What s Inside Electric Devices?

Student Sheet PA.1: What s Inside Electric Devices? Student Sheet PA.1: What s Inside Electric Devices? Directions: Use this student sheet to complete Investigation PA.1. Table A. What s Inside? Device What It Does Power Source Number of Components Components

More information

Chapter 22. Induction

Chapter 22. Induction Chapter 22 Induction Induced emf A current can be produced by a changing magnetic field First shown in an experiment by Michael Faraday A primary coil is connected to a battery A secondary coil is connected

More information

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS TSOKOS OPTION I-2 MEDICAL IMAGING Reading Activity Answers IB Assessment Statements Option I-2, Medical Imaging: X-Rays I.2.1. I.2.2. I.2.3. Define

More information

Physics 102 Spring 2006: Final Exam Multiple-Choice Questions

Physics 102 Spring 2006: Final Exam Multiple-Choice Questions Last Name: First Name: Physics 102 Spring 2006: Final Exam Multiple-Choice Questions For questions 1 and 2, refer to the graph below, depicting the potential on the x-axis as a function of x V x 60 40

More information

How Can We Determine the Actual Percentage of H 2 O 2 in a Drugstore Bottle of Hydrogen Peroxide?

How Can We Determine the Actual Percentage of H 2 O 2 in a Drugstore Bottle of Hydrogen Peroxide? How Can We Determine the Actual Percentage of H 2 O 2 in a Drugstore Bottle of Hydrogen Peroxide? Hydrogen peroxide, H 2 O 2, is easily oxidized. Dilute solutions are used as a disinfectant, in commercial

More information

NMR and MRI : an introduction

NMR and MRI : an introduction Intensive Programme 2011 Design, Synthesis and Validation of Imaging Probes NMR and MRI : an introduction Walter Dastrù Università di Torino walter.dastru@unito.it \ Introduction Magnetic Resonance Imaging

More information

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Ordinary Level

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Ordinary Level www.xtremepapers.com UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Ordinary Level *5533380653* PHYSICS 5054/02 Paper 2 Theory May/June 2007 1 hour 45 minutes Candidates

More information

Passive diffusion. Name: Leo Plese School: XV. gymansium Zagreb. Research work in biology 1 st grade of high school

Passive diffusion. Name: Leo Plese School: XV. gymansium Zagreb. Research work in biology 1 st grade of high school Research work in biology 1 st grade of high school 2013-2014 Passive diffusion Name: Leo Plese School: XV. gymansium Zagreb Leo Plese, e- mail: Plese.Leo@gmail.com 2 Passive diffusion Table of Contents

More information

A Brief Introduction to Medical Imaging. Outline

A Brief Introduction to Medical Imaging. Outline A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length

More information

Wednesday 21 June 2017 Morning

Wednesday 21 June 2017 Morning Oxford Cambridge and RSA Wednesday 21 June 2017 Morning A2 GCE PHYSICS A G485/01 Fields, Particles and Frontiers of Physics *6808554474* Candidates answer on the Question Paper. OCR supplied materials:

More information

Radioisotopes and PET

Radioisotopes and PET Radioisotopes and PET 1 Radioisotopes Elements are defined by their number of protons, but there is some variation in the number of neutrons. Atoms resulting from this variation are called isotopes. Consider

More information

ELG7173 Topics in signal Processing II Computational Techniques in Medical Imaging

ELG7173 Topics in signal Processing II Computational Techniques in Medical Imaging ELG7173 Topics in signal Processing II Computational Techniques in Medical Imaging Topic #1: Intro to medical imaging Medical Imaging Classifications n Measurement physics Send Energy into body Send stuff

More information

12.01 Determination of the isoelectric point of an amino acid (glycine)

12.01 Determination of the isoelectric point of an amino acid (glycine) Biochemistry LEB 12 Determination of the isoelectric point of an amino acid (glycine) What you can learn about Isoelectric point Acidic anions Basic cations Zwitterions Equivalence (inflection) points

More information

PRELIMINARY EXAMINATION 2018 H2 PHYSICS 9749/01. Paper 1 SEP 2018

PRELIMINARY EXAMINATION 2018 H2 PHYSICS 9749/01. Paper 1 SEP 2018 PRELIMINARY EXAMINATION 2018 H2 PHYSICS 9749/01 Paper 1 SEP 2018 Additional Materials: Multiple Choice Answer Sheet Duration: 1 hour DO NOT OPEN THIS BOOKLET UNTIL YOU ARE TOLD TO DO SO READ THESE INSTRUCTIONS

More information

Lecture 12 February 11, 2016

Lecture 12 February 11, 2016 MATH 262/CME 372: Applied Fourier Analysis and Winter 2016 Elements of Modern Signal Processing Lecture 12 February 11, 2016 Prof. Emmanuel Candes Scribe: Carlos A. Sing-Long, Edited by E. Bates 1 Outline

More information

London Examinations IGCSE

London Examinations IGCSE Centre No. Candidate No. Surname Signature Initial(s) Paper Reference(s) 4420/1F London Examinations IGCSE Physics Paper 1F Foundation Tier Monday 31 October 2005 Morning Time: 1 hour 30 minutes Examiner

More information

Summer AP Assignment 2016

Summer AP Assignment 2016 Summer AP Assignment 2016 Mr. Michael Wichart (Rm. 109) wichart.m@woodstown.org Summer Assignment Goals The main goal of this assignment is to review some of the material that was taught during Honors

More information

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline)

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline) maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline) Intended use: Storage: Safety: After 1:10 dilution of the PBS

More information

The NMR Inverse Imaging Problem

The NMR Inverse Imaging Problem The NMR Inverse Imaging Problem Nuclear Magnetic Resonance Protons and Neutrons have intrinsic angular momentum Atoms with an odd number of proton and/or odd number of neutrons have a net magnetic moment=>

More information

Chapter 25 & 28 Solutions

Chapter 25 & 28 Solutions Chapter 25 & 28 Solutions Q25.9. Reason: The original field is into the page within the loop and is changing strength. The induced, counterclockwise current produces a field out of the page within the

More information

The object experiencing the field is called the TEST OBJECT

The object experiencing the field is called the TEST OBJECT A field is an invisible area of influence around an object If you place another object in the field it will experience a force 1 object is said to be causing the field The object experiencing the field

More information

1. The diagram shows the electric field lines produced by an electrostatic focussing device.

1. The diagram shows the electric field lines produced by an electrostatic focussing device. 1. The diagram shows the electric field lines produced by an electrostatic focussing device. Which one of the following diagrams best shows the corresponding equipotential lines? The electric field lines

More information

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar. 1 Doppler echocardiography & Magnetic Resonance Imaging History: - Langevin developed sonar. - 1940s development of pulse-echo. - 1950s development of mode A and B. - 1957 development of continuous wave

More information

The Photon Concept. Modern Physics [2] How are x-rays produced? Gamma rays. X-ray and gamma ray photons. X-rays & gamma rays How lasers work

The Photon Concept. Modern Physics [2] How are x-rays produced? Gamma rays. X-ray and gamma ray photons. X-rays & gamma rays How lasers work Modern Physics [2] X-rays & gamma rays How lasers work Medical applications of lasers Applications of high power lasers Medical imaging techniques CAT scans MRI s The Photon Concept a beam of light waves

More information

Magnetism. a) Ferromagnetic materials are strongly attracted to magnets. b) Paramagnetic materials are weakly attracted to magnets

Magnetism. a) Ferromagnetic materials are strongly attracted to magnets. b) Paramagnetic materials are weakly attracted to magnets Magnetism Types of Magnetic Materials Magnetic substances can be classified into three basic groups, according to their response to a magnet. Note the strength and direction of the interaction. a) Ferromagnetic

More information

FALL 2004 Midterm Exam #2, Part A

FALL 2004 Midterm Exam #2, Part A Physics 152 FALL 2004 Midterm Exam #2, Part A Roster No.: Score: 17 points possible Exam time limit: 50 minutes. You may use a calculator and both sides of ONE sheet of notes, handwritten only. Closed

More information

2. A proton is traveling with velocity v, to the right, through a magnetic field pointing into the page as indicated in the figure below.

2. A proton is traveling with velocity v, to the right, through a magnetic field pointing into the page as indicated in the figure below. 1. An electron has a mass of 9.11 x 10-31 kg and its charge is -1.6 x 10-19 C. The electron is released from rest in a vacuum between two flat, parallel metal plates that are 10 cm apart. The potential

More information

NMR/MRI examination (8N080 / 3F240)

NMR/MRI examination (8N080 / 3F240) NMR/MRI examination (8N080 / 3F240) Remarks: 1. This test consists of 3 problems with at total of 26 sub-questions. 2. Questions are in English. You are allowed to answer them in English or Dutch. 3. Please

More information

8.5. Applications of Electric and Magnetic Fields. RFID Chips

8.5. Applications of Electric and Magnetic Fields. RFID Chips Applications of Electric and Magnetic Fields If you have a pet dog or cat, you probably have tags on it that give information on how to contact you if the pet ever gets lost. Unfortunately, these tags

More information

1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level)

1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level) 1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Capacitance (Chapter 18): Physics (A-level) Every capacitor has two leads, each connected to a metal plate, where in between there is an insulating

More information

Question 1. Question 2

Question 1. Question 2 Question 1 Figure 29-26 shows cross sections of two long straight wires; the left-hand wire carries current i 1 directly out of the page. The net magnetic field due to the two currents is to be zero at

More information

Rad Tech 4912 MRI Registry Review. Outline of the Registry Exam: Certification Fees

Rad Tech 4912 MRI Registry Review. Outline of the Registry Exam: Certification Fees Rad Tech 4912 MRI Registry Review Outline of the Registry Exam: Category: # of questions: A. Patient Care 30 B. Imaging Procedures 62 C. Data Acquisition and Processing 65 D. Physical Principles of Image

More information

EE 3324 Electromagnetics Laboratory

EE 3324 Electromagnetics Laboratory EE 3324 Electromagnetics Laboratory Experiment #3 Inductors and Inductance 1. Objective The objective of Experiment #3 is to investigate the concepts of inductors and inductance. Several inductor geometries

More information

Two point charges, A and B, lie along a line separated by a distance L. The point x is the midpoint of their separation.

Two point charges, A and B, lie along a line separated by a distance L. The point x is the midpoint of their separation. Use the following to answer question 1. Two point charges, A and B, lie along a line separated by a distance L. The point x is the midpoint of their separation. 1. Which combination of charges would yield

More information

Electromagnetism II. (a) enav Na enav Cl (b) enav Na + enav Cl (c) enav Na (d) enav Cl (e) zero

Electromagnetism II. (a) enav Na enav Cl (b) enav Na + enav Cl (c) enav Na (d) enav Cl (e) zero Electromagnetism II 1. Salt water contains n sodium ions (Na + ) per cubic meter and n chloride ions (Cl ) per cubic meter. A battery is connected to metal rods that dip into a narrow pipe full of salt

More information

Can a Magnetic Field Produce a Current?

Can a Magnetic Field Produce a Current? Can a Magnetic Field Produce a Current? In our study of magnetism we learned that an electric current through a wire, or moving electrically charged objects, produces a magnetic field. Could the reverse

More information

Medical Imaging Physics Spring Quarter Week 9-1

Medical Imaging Physics Spring Quarter Week 9-1 Medical Imaging Physics Spring Quarter Week 9-1 NMR and MRI Davor Balzar balzar@du.edu www.du.edu/~balzar Intro MRI Outline NMR & MRI Guest lecturer fmri Thursday, May 22 Visit to CUHSC It s not mandatory

More information

BASIC MRI PHYSICS SPIN GYMNASTICS Don Plewes PhD, Walter Kucharczyk MD

BASIC MRI PHYSICS SPIN GYMNASTICS Don Plewes PhD, Walter Kucharczyk MD BASIC MRI PHYSICS SPIN GYMNASTICS Don Plewes PhD, Walter Kucharczyk MD Introduction To understand MRI, it is first necessary to understand the physics of proton Nuclear Magnetic Resonance (NMR). The most

More information

Electromagnetics in Medical Physics

Electromagnetics in Medical Physics Electromagnetics in Medical Physics Part 4. Biomagnetism Tong In Oh Department of Biomedical Engineering Impedance Imaging Research Center (IIRC) Kyung Hee University Korea tioh@khu.ac.kr Dot Product (Scalar

More information

Physics 115. Magnetic forces, Coils, Induction. General Physics II. Session 29

Physics 115. Magnetic forces, Coils, Induction. General Physics II. Session 29 Physics 115 General Physics II Session 29 Magnetic forces, Coils, Induction R. J. Wilkes Email: phy115a@u.washington.edu Home page: http://courses.washington.edu/phy115a/ 5/22/14 1 Lecture Schedule Today

More information

Teacher Notes on: NMR Spectroscopy

Teacher Notes on: NMR Spectroscopy Teacher Notes on: NMR Spectroscopy What is it? Nuclear Magnetic Resonance (NMR) spectroscopy is (arguably) the most powerful tool available for determining the structure of organic compounds. This technique

More information

Light, Electricity and Liquids. By Vitaliy Zamsha and Vladimir Shevtsov

Light, Electricity and Liquids. By Vitaliy Zamsha and Vladimir Shevtsov Light, Electricity and Liquids updated By Vitaliy Zamsha and Vladimir Shevtsov Abstract The authors of this article revealed their long-awaited work, which represents some effects in liquids under the

More information

QUESTION 1. Answer: N

QUESTION 1. Answer: N MAGNETISM Review QUESTION 1 A proton speeding through a synchrotron at 3.0 10 7 m/s experiences a magnetic field of 3.0 T that is produced by the steering magnets in the synchrotron. What is the magnetic

More information

PHYSICS A 2825/02 Health Physics

PHYSICS A 2825/02 Health Physics THIS IS A LEGACY SPECIFICATION ADVANCED GCE PHYSICS A 2825/02 Health Physics *CUP/T60952* Candidates answer on the question paper OCR Supplied Materials: None Other Materials Required: Electronic calculator

More information

Waves Final Review. Name: Date: 1. On which one of the following graphs is the wavelength λ and the amplitude a of a wave correctly represented?

Waves Final Review. Name: Date: 1. On which one of the following graphs is the wavelength λ and the amplitude a of a wave correctly represented? Name: Date: Waves Final Review 1. On which one of the following graphs is the wavelength λ and the amplitude a of a wave correctly represented? A. Displacement λ a Distance along wave B. Displacement λ

More information

Physics 3204 Final Review (Based on June 2014 exam) Name:

Physics 3204 Final Review (Based on June 2014 exam) Name: Physics 3204 Final Review (Based on June 2014 exam) Name: Scan this QR code (or go to http://www.ed.gov.nl.ca/edu/k12/evaluation/physics3204/ june_2014_physics_3204_outcome_report.pdf) to see the Exam

More information

DO PHYSICS ONLINE MOTORS AND GENERATORS MAGNETIC FIELDS

DO PHYSICS ONLINE MOTORS AND GENERATORS MAGNETIC FIELDS DO PHYSICS ONLINE MOTORS AND GENERATORS MAGNETIC FIELDS Powerful magnets are essential components in motors and generators. Some electric motors and generators rely upon a combination of a permanent and

More information

Use quantitative and qualitative analyses to identify the powder in Mr. Orlow s car.

Use quantitative and qualitative analyses to identify the powder in Mr. Orlow s car. Case File 7 Drug Tests: Identifying an unknown chemical Use quantitative and qualitative analyses to identify the powder in Mr. Orlow s car. Police Report Patrol officers pulled over Mr. Yuri Orlow for

More information

Can a Magnetic Field Produce a Current?

Can a Magnetic Field Produce a Current? Can a Magnetic Field Produce a Current? In our study of magnetism we learned that an electric current through a wire, or moving electrically charged objects, produces a magnetic field. Could the reverse

More information

4 pt. (in J) 3.A

4 pt. (in J) 3.A Mark Reeves - Physics 22, Fall 2011 1 A point charge of mass 0.0699 kg and charge q = +6.87 µc is suspended by a thread between the vertical parallel plates of a parallel-plate capacitor, as shown in the

More information

Wednesday 23 January 2013 Afternoon

Wednesday 23 January 2013 Afternoon Wednesday 23 January 2013 Afternoon A2 GCE PHYSICS A G485/01 Fields, Particles and Frontiers of Physics *G411600113* Candidates answer on the Question Paper. OCR supplied materials: Data, Formulae and

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) The UV-VIS spectrometer is a portable ultraviolet light and visible light spectrophotometer, combining a spectrometer and a light source/cuvette

More information

Intermediate Physics PHYS102

Intermediate Physics PHYS102 Intermediate Physics PHYS102 Dr Richard H. Cyburt Assistant Professor of Physics My office: 402c in the Science Building My phone: (304) 384-6006 My email: rcyburt@concord.edu My webpage: www.concord.edu/rcyburt

More information

Tissue Characteristics Module Three

Tissue Characteristics Module Three Tissue Characteristics Module Three 1 Equilibrium State Equilibrium State At equilibrium, the hydrogen vector is oriented in a direction parallel to the main magnetic field. Hydrogen atoms within the vector

More information

ELEC ENG 3BA3: Structure of Biological Materials

ELEC ENG 3BA3: Structure of Biological Materials Name: Student Number: ELEC ENG 3BA3: Structure of Biological Materials Day Class Instructor: Dr. I. C. BRUCE Duration of Examination: 3 Hours McMaster University Final Examination December 2011 This examination

More information

Detection at a Distance for More Sensitive MRIDetection at a D...

Detection at a Distance for More Sensitive MRIDetection at a D... March 15, 2004 Detection at a Distance for More Sensitive MRI Contact: Paul Preuss, paul_preuss@lbl.gov Advanced Search Search Tips BERKELEY, CA! Alexander Pines and his colleagues have discovered a remarkable

More information

Chemical Engineering. Mike Jim. English 8. Mr. Rinkevich

Chemical Engineering. Mike Jim. English 8. Mr. Rinkevich Chemical Engineering Mike Jim English 8 Mr. Rinkevich May 28, 2012 Jim 1 Mike Jim Mr. Rinkevich English 8 28 May 2012 Chemical Engineering In the field of engineering, a chemical engineer is the profession

More information

19.1 Laws of Magnetism page The Earth can be regarded as a large bar magnet with its south pole in the northern hemisphere.

19.1 Laws of Magnetism page The Earth can be regarded as a large bar magnet with its south pole in the northern hemisphere. Chapter 19 Magnetism Learning Outcomes After completing this chapter, students should be able to: 1. state the properties of magnets 2. describe induced magnetism 3. describe electrical methods of magnetisation

More information

Don t Copy This. Michael Faraday 3/1/13. Chapter 25: EM Induction and EM Waves. Key Terms:

Don t Copy This. Michael Faraday 3/1/13. Chapter 25: EM Induction and EM Waves. Key Terms: 3/1/13 Chapter 25: EM Induction and EM Waves Key Terms: Induced Induced emf current Flux Lenz s Law waves Photons EM Don t Copy This Last chapter we learned that a current can create a magnetic field.

More information

BLUE-PRINT II XII Physics

BLUE-PRINT II XII Physics BLUE-PRINT II XII Physics S.No. UNIT VSA SA I SA II LA Total (1 Mark) (2 Marks) (3Marks) (5 Marks) 1. Electrostatics 1(1) 4(2) 3(1) - 8(4) 2. Current Electricity - 2(1) - 5(1) 7(2) 3. Magnetic Effect of

More information