TRACE ELEMENTS IN BLOOD MEASURED BY PIXE

Size: px
Start display at page:

Download "TRACE ELEMENTS IN BLOOD MEASURED BY PIXE"

Transcription

1 2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: TRACE ELEMENTS IN BLOOD MEASURED BY PIXE Suene Bernardes, Manfredo H. Tabacniks, Marcel D.L. Barbosa and Márcia A. Rizzutto Instituto de Física - Universidade de São Paulo Rua do Matão, travessa R 187, , São Paulo, SP suene@if.usp.br ABSTRACT The abstract is a very brief summary highlighting main accomplishments, what is new, and how it relates to the state-of-the-art. The quantitative measurement of trace elements in blood (and plasma) and its comparison with data from literature is shown. To optimize the analytical conditions and detection limits of blood analysis by PIXE (Particle Induced X-ray Emission), a study on several sample preparation methods was made. PIXE analysis was done using a 2.4 MeV proton beam and two Si(Li) detectors with respectively 50µm Be and 130 µm thick Mylar filters. Elementary concentrations were calculated relative to an internal gallium standard. A total of 13 trace elements were detected with concentrations ranging from 10 ng/g to 1 mg/g. 1. INTRODUCTION Blood is a suspension of cells (erythrocytes, leucocytes and platelets) in a complex fluid, the plasma, which is mainly constituted of water, minerals, vitamins, proteins, carbohydrates, and lipids. Serum differs from the plasma by the removal of coagulation factors. The minerals act as catalytic factors for many biochemical reactions and play an important role in the hormone production. In the plasma, the minerals can be chemically free or aggregated to proteins and other organic molecule [1]. The continuous development of increasingly more sensitive trace element methods, is shifting the research of trace elements in biological materials from simple toxicological contaminants to the better understanding of their function in the organisms. It is known that many trace elements are essential for life, but are also toxic, if in excess. The measurement of very small amounts of trace elements in diminutive organic samples is a research theme by itself, since sample preparation methods, data quantification and detection limits need to be tested and improved. The objective of this work was to measure the trace elements in sanguineous plasma by PIXE (Particle Induced X-ray Emission) and to compare the results with data from literature attempting to guide the answers of two basic questions: a) Which are the elements of biological interest? b) What is the necessary detection limit? PIXE is an elementary spectrometric method that relies on the detection of characteristics X-rays emitted by bombarding a sample with an energetic (~MeV) ion beam (H +, He +,..). PIXE is sensitive to all the elements in the periodic table above the aluminum, with a primary detection limit for solid samples of the order of 1 µg/g which can be extended to the ng/g level for liquid samples after a pre-concentration procedure [2].

2 2. EARLIER MEASUREMENTS The development of many element sensitive techniques and the systematic measurement of trace elements in materials and more specifically in biological materials can be tracked down to early 1950 [3, 4]. Initially, and probably aiming toxicological studies, many investigations dealt with the detection of poisonous contaminants at relatively high concentrations (like Pb, Hg, Cd etc.) [5]. In the search to understand the role of trace and ultra-trace elements in the environment and in the living bodies, the search for trace elements required decreasingly lower detection limits, while also reducing the amount of sample material needed. The multielementary and almost non-destructive physical spectrometric methods are specially suited for this kind of analysis. Starting around the 50ties, the X-ray Fluorescence analysis and the Neutron Activation Analysis, extended and complemented the data of the widespread monoelementary Atomic Absorption (and Emission) Spectrometry, a common equipment available in many chemical laboratories almost since the beginning of the 20th century. With the development and worldwide dissemination of the PIXE analysis in the 70ties [6], the search for trace elements in the environment and biology got a remarkable push. Thanks to its low sensitivity (in the ppm range), speediness and allowing yet smaller samples, when compared to the XRF, PIXE analysis grew in less than 10 years to a mature and powerful micro-analytical method [2]. Yet, there was still the need to lower the detection limits down to sub ng/g (ppb) level. The Inductive Coupled Plasma Atomic Emission (ICP-AES) and its successor, the ICP - Mass Spectrometer, developed in the 80ties [7, 8], achieved the ultimate trace element detection limits at a reasonable cost. The question nowadays changed from how many elements were detected to what is the necessary detection limit or which elements need to be detected. Figure 1. Compilation of published data on trace element concentration (ng/g) in whole blood. Squares ( ) refer to medians while ( ) indicate data from a unique source. Straight lines indicate the maximum-minim range of published data.

3 A survey in 10 of the most accessible journals and publications supplied data for elementary concentration data for 68 elements in whole blood. The data were compiled in simple scatterplots of their median values for plasma and serum in Figure 1, ordered according to their atomic number. The maximum-minimum ranges, when available, were also plotted as straight lines. Concentration data range over 11 orders of magnitude and include samples from different healthy populations, most measured by PIXE and ICP [9-21]. 3. MATERIAL AND METHODS To verify the actual PIXE detection limits and the agreement with published data, three plasma samples were analyzed by PIXE. The samples were micro-pipetted on Nuclepore (10 µm thick Makrofol ) filters stretched on 25mm diameter PVC rings. The filter surface was prepared by pipetting 20µl of a plasma solution spiked with gallium solution, as an internal standard resulting in 50 µg/ml of Ga in the final solution. The picture of a sample ready for PIXE analysis is shown in Figure 2. PIXE analysis was done using a 2.04 MeV proton beam collimated to a diameter of 5mm. X-rays were measured with two Si(Li) detectors, respectively with a 50µm Be filter and a 150µm Mylar filter, tailored to record respectively the low energy and the high energy X-ray spectra. The samples were irradiated during 10 minutes with a beam current of 20 na. The complete PIXE setup at LAMFI [22] is described elsewhere [23]. The x-ray spectra were analyzed using the well-known non-linear multiple gaussian fitting program AXIL [24]. Quantitative calibration of the PIXE setup was made using a set of mono-elementary evaporated thin film standards (mostly from Micromatter Co.) and a semi-empirical fit. Figure 2. Pipetted sample on polymer film ready for analysis. 4. RESULTS Fourteen elements were positively identified in the PIXE spectra. Data were converted to mass concentration using the internal Ga standard concentration in Equation 1 I R C Z = CGa (1) SR

4 where I R represents the relative-to-gallium X-ray intensity of element with atomic number Z, C Z is its measured concentration, (µg ml -1 ), C Ga represents the Gallium internal standard concentration (50 µg/ml ) and S R is the relative-to-gallium yield of the PIXE system. The measured average concentrations are reproduced in Table 1. The quoted uncertainties are the standard deviation of the averaged values. Table 1. Elementary mass concentrations in sanguineous plasma measured by PIXE Element Average concentration (µg/ml) Si 2.2 ± 2.2 P 49.1 ± 2.9 S (0.63 ± 0.16)10 3 Cl (6.0 ± 1.3)10 3 K 372 ± 35 Ca 77 ± 15 Ti 0.16 ± 0.14 Fe 1.4 ± 0.9 Co ± Ni ± Cu 1.11 ± 0.37 Zn 1.55 ± 0.63 Se 0.19 ± DISCUSSION Several materials were tested as a substrate for the pipetted blood samples: thin polymer films like Mylar, Nuclepore, Kimfol and Kapton, graphite discs and analytical ashless paper filters. Blood was noticed not to adhere on Mylar or Kimfol and regular graphite presented too much contamination. Kapton and Nuclepore thin films, which showed good adhesion to blood, were used in the present study. Paper filters presented good adhesion and high sample load but showed low uniformity of the sample mass across the filter. Measured elementary concentrations were compared to whole blood and serum values from literature in Figure 3. The results are in good agreement with the literature, except for phosphorous whose value is below expectation and nickel with results above expectation. It may be the consequence of sample preparation or may just be a regional effect. It must be stressed out that the measured trace elements in plasma are being compared to literature data of whole blood and serum measurements. The results are expected to compare to serum values, since serum differs from plasma by the absence of coagulation factors. The difference

5 between whole blood and serum (or plasma) is evident in the iron concentrations, which is partially removed with the erythrocytes, leucocytes and platelets. The need for better detection limits is quite visible in Figure 3 where no data are lower than 1 ppm. Actually, inspecting Figure 1, it is found that around 18 more elements were expected to be measured, if the PIXE detection limit for liquid samples could be set to its nominal detection limit at about 1 ppb. Figure 1. Water-filled borehole using conventional discrete ordinates formulation for the transport equation. Elementary concentrations in serum (S) and whole blood (WB) compared to the PIXE data measured (M) in this work. The error bars indicate one standard deviation. 6. CONCLUSIONS PIXE analysis of plasma samples deposited on paper filter seems to be a promising technique for measuring trace elements in blood above the 10 ppb wgt concentration. Besides the need to succeed reaching the nominal PIXE detection limits, a further 10-fold reduction of the detection limits is needed to cover most of the possible important trace elements in blood serum, which would make PIXE analysis a very competitive technique in bio-trace-elements research. This can be achieved by pre-concentrating the liquid samples and by reducing the film thickness of the substrates. ACKNOWLEDGMENTS The authors like to acknowledge M.R. Antonio and M.V.S. Lima for help running the accelerator, N. Added for helpful discussions and CNPq for financial support.

6 REFERENCES 1. T. Verrastro, T.F. Lorenzi, S. Wendel Neto, Hematologia e Hemoterapia. Ed. Atheneu. São Paulo, Brazil, (1996). 2. S.A.E. Johansson and J.L. Campbell, PIXE, A Novel Technique for Elemental Analysis. John Wiley and Sons, (1998). 3. W. Maenhaut, in B. Ottar and J.M. Pacyna eds. Control and Fate of Atmospheric Heavy Metals, NATO ARW Series, Kluver Ac. Pub. Amsterdam., (1988). 4. R. Jenkins, R.W. Gould and D. Gedke, Quantitative X-ray Spectrometry, Marcel Dekker, New York, USA. (1981). 5. W. Maenhaut, L. De Reu., H.A. Van Rinsvelt, G.S. Roessler, J.W. Swanson, and M.D. Williams, IEEE Trans. on Nucl. Science. NS-28-2, pp (1981). 6. T.B. Johansson, K.R. Akselsson and S.A.E. Johansson, Nucl. Instr. Meth., 84, pp. 141 (1970). 7. P.J. Potts, A Handbook of Silicate Rock Analysis. Chapman & Hall, pp (1987). 8. M.F. Giné-Rosias, ed. Espectrometria de Massas com Fonte de Plasma (ICP-MS). CGP/CENA. Piracicaba, Brasil, (1999). 9. H. Afarideh, A. Amirabadi, S.M. Hadji-Saeid, K. Kaviani and E. Zibafar, Nuclear Instruments and Methods in Physics Research, B109/110, pp , (1996). 10. E. Bárány, I.A. Bergdahl, L.E. Bratteby, T. Lundh, G. Samuelson, A. Schütz, S. Skerfving and A. Oskarsson, Toxicology Letters, 134, pp (2002). 11. C.P. Case, Clin. Chem., 47:2, pp (2001). 12. J. Dombovári, Zs. Varga, J.S. Becker, J. Mátyus and L. Papp, Atomic Spectroscopy, Jul/Aug., 22(4), (2001). 13. K. Inagaki and H. Haraguchi, Analyst, 125, pp (2000). 14. S. Mousavi-Yeganeh, F. Ebrahimy-Fakhar and F. Enayati, Nuclear Instruments and Methods in Physics Research, B3, (1984). 15. C.S. Muñiz, J.M.M. Gayón, J.I.G. Alonso and A. Sanz-Medel, J. Anal. At. Spectrom., 14, pp (1999). 16. I. Rodushkin, F. Ödman and S. Branth, Fresenius J. Anal. Chem., 364, pp (1999). 17. M. St antná, I. Nemcová and J. Zúka, Analytical Letters, 32(13), pp (1999). 18. H. Vanhoe, C. Vandecasteele, J. Versieck and R. Dams, Anaytica Chimica Acta, 244, pp (1991). 19. H. Vanhoe, R. Dams, C. Vandecasteele and J. Versieck, Anaytica Chimica Acta, 281, pp (1993). 20. H. Vanhoe, F.V. Allemeersch, J. Versieck and R. Dams, Anayst, 118, pp (1993). 21. H. Vanhoe, R. Dams and J. Versieck, J. Analytical Atomic Spectrometry, 9, pp (1994) M.H. Tabacniks, in Nuclear Physics, Eds. S.R. Souza, O.D. Gonçalves, C.L. Lima, L. Tomio, V.R. Vanin, World Scientific, Singapore, (1998). 24. P. Van Espen, K. Janssens, & I. Swenters, AXIL X-Ray Analysis software. Camberra Packard, Benelux.

FLUORINE IN WATER AND TEETH MEASURED WITH PIGE

FLUORINE IN WATER AND TEETH MEASURED WITH PIGE 2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 FLUORINE IN WATER AND TEETH

More information

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS ANTOANETA ENE 1, I. V. POPESCU 2, T. BÃDICÃ 3, C. BEªLIU 4 1 Department of Physics, Faculty

More information

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater PRAMANA c Indian Academy of Sciences Vol. 76, No. 2 journal of February 2011 physics pp. 361 366 Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

More information

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DEVELOPMENT

More information

Within the vast field of atomic physics, collisions of heavy ions with atoms define

Within the vast field of atomic physics, collisions of heavy ions with atoms define Chapter 1 Introduction Within the vast field of atomic physics, collisions of heavy ions with atoms define one of the most active areas of research. In the last decades, the design and construction of

More information

Overview of X-Ray Fluorescence Analysis

Overview of X-Ray Fluorescence Analysis Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic

More information

BLOOD BIOCHEMISTRY EVALUATION IN HORSES HYPERIMMUNE SERA PRODUCTION

BLOOD BIOCHEMISTRY EVALUATION IN HORSES HYPERIMMUNE SERA PRODUCTION 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 BLOOD BIOCHEMISTRY

More information

Quantitative XRF Analysis. algorithms and their practical use

Quantitative XRF Analysis. algorithms and their practical use Joint ICTP-IAEA School on Novel Experimental Methodologies for Synchrotron Radiation Applications in Nano-science and Environmental Monitoring Quantitative XRF Analysis algorithms and their practical use

More information

Ion-beam techniques. Ion beam. Electrostatic Accelerators. Van de Graaff accelerator Pelletron Tandem Van de Graaff

Ion-beam techniques. Ion beam. Electrostatic Accelerators. Van de Graaff accelerator Pelletron Tandem Van de Graaff Ion-beam techniques RBS Target nucleus Ion beam STIM RBS: Rutherford backscattering ERD: Elastic recoil detection PIXE: Particle induced x-ray emission PIGE: Particle induced gamma emission NRA: Nuclear

More information

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 EFFICIENCY SIMULATION OF

More information

Applications of proton-induced X-ray emission technique in materials and environmental science

Applications of proton-induced X-ray emission technique in materials and environmental science Ovidius University Annals of Chemistry Volume 20, Number 1, pp. 35-39, 2009 Applications of proton-induced X-ray emission technique in materials and environmental science Antoaneta ENE* a, Ion V. POPESCU

More information

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory Altitude influence of elemental distribution in grass from Rila mountain Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory I. Introduction The application of modern instrumental

More information

I. 17. PIXE Analysis of Trace Heavy-Metals in River Water Using an Ion-Exchange Cellulose Filter Paper

I. 17. PIXE Analysis of Trace Heavy-Metals in River Water Using an Ion-Exchange Cellulose Filter Paper CYRIC Annual Report 2001 I. 17. PIXE Analysis of Trace Heavy-Metals in River Water Using an Ion-Exchange Cellulose Filter Paper Amartaivan Ts., Yamazaki H., Ishii K., Matsuyama S., Takahashi Y., Satoh

More information

Particle-Induced X-Ray Emission Spectrometry (PIXE)

Particle-Induced X-Ray Emission Spectrometry (PIXE) Particle-Induced X-Ray Emission Spectrometry (PIXE) Edited by SVEN A. E. JOHANSSON Department of Nuclear Physics Lund Institute of Technology Lund, Sweden JOHN L. CAMPBELL Department of Physics University

More information

THERMOLUMINESCENCE PROPERTIES OF OLIVINE

THERMOLUMINESCENCE PROPERTIES OF OLIVINE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 THERMOLUMINESCENCE PROPERTIES

More information

Summary of test results for Daya Bay rock samples. by Patrick Dobson Celia Tiemi Onishi Seiji Nakagawa

Summary of test results for Daya Bay rock samples. by Patrick Dobson Celia Tiemi Onishi Seiji Nakagawa Summary of test results for Daya Bay rock samples by Patrick Dobson Celia Tiemi Onishi Seiji Nakagawa October 2004 Summary A series of analytical tests were conducted on a suite of granitic rock samples

More information

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS 45 ABSTRACT FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS W. T. Elam, Robert B. Shen, Bruce Scruggs, and Joseph A. Nicolosi EDAX, Inc. Mahwah, NJ 70430 European Community Directive 2002/95/EC

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information

TECHNIQUE OF MEASURING THE FIRST IONIZATION COEFFICIENT IN GASES

TECHNIQUE OF MEASURING THE FIRST IONIZATION COEFFICIENT IN GASES 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 TECHNIQUE OF MEASURING

More information

S.K. Sahoo 1 *, Z.S. Zunic 2, R. Kritsananuwat 1, H. Arae 1 and S. Mishra 1

S.K. Sahoo 1 *, Z.S. Zunic 2, R. Kritsananuwat 1, H. Arae 1 and S. Mishra 1 S.K. Sahoo 1 *, Z.S. Zunic 2, R. Kritsananuwat 1, H. Arae 1 and S. Mishra 1 1, Japan 2 Institute of Nuclear Sciences Vinca, Beograd, Serbia E-mail:sahoo@nirs.go.jp 3rd TRE-ICEP Veszprém-Hungary, May 16-18,

More information

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences

More information

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical

More information

PIXE Protocol Setup for Cluster Analysis of Archeological Samples

PIXE Protocol Setup for Cluster Analysis of Archeological Samples PIXE Protocol Setup for Cluster Analysis of Archeological Samples Mohamad Roumié Accelerator Laboratory, Lebanese Atomic Energy Commission, National Council for Scientific Research, Airport Road, P.O.Box:

More information

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017 MICROMATTER TECHNOLOGIES INC. #1-8333 130 th Street, Surrey, B.C. V3W 7X4, Canada Phone: +1.604.594.9720 or 1.866.966.2286 Fax: +1.604.594.9726 info@micromatter.com www.micromatter.com Technical Note 2015-01A

More information

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF S2 PICOFOX Spectrometry Solutions Innovation with Integrity TXRF S2 PICOFOX True Trace Analysis with XRF for the First Time! You need to know the concentration of trace elements in environmental samples?

More information

Portable type TXRF analyzer: Ourstex 200TX

Portable type TXRF analyzer: Ourstex 200TX Excerpted from Adv. X-Ray. Chem. Anal., Japan: 42, pp. 115-123 (2011) H. Nagai, Y. Nakajima, S. Kunimura, J. Kawai Improvement in Sensitivity and Quantification by Using a Portable Total Reflection X-Ray

More information

SYSTEM FOR PIXE TRACE ELEMENT ANALYSIS AT MTF. M. Beňo, J. Dobrovodský, D. Vaňa, S. Minárik

SYSTEM FOR PIXE TRACE ELEMENT ANALYSIS AT MTF. M. Beňo, J. Dobrovodský, D. Vaňa, S. Minárik SYSTEM FOR PIXE TRACE ELEMENT ANALYSIS AT MTF M. Beňo, J. Dobrovodský, D. Vaňa, S. Minárik Slovak University of Technology in Bratislava, Faculty of Materials Science and Technology in Trnava (MTF), Advanced

More information

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Hiroshi Onodera Application & Research Center, JEOL Ltd. Introduction um, PBB and PBDE) are subject to usage restrictions in Europe.

More information

Treatment of effluents from petroleum production by electron beam irradiation

Treatment of effluents from petroleum production by electron beam irradiation Radiation Physics and Chemistry 71 (2004) 443 447 Treatment of effluents from petroleum production by electron beam irradiation Celina Lopes Duarte a, *, Lucia Limoeiro Geraldo b, Oswaldo de Aquino P.

More information

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 511 FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON

More information

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR A. C. Neiva 1, J. N. Dron 1, L. B. Lopes 1 1 Escola Politécnica da Universidade de São Paulo

More information

GUAMA RIVER AND GUAJARA BAY, BELEM - AMAZON REGION

GUAMA RIVER AND GUAJARA BAY, BELEM - AMAZON REGION 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 210 Pb GEOCHRONOLOGY

More information

7) Applications of Nuclear Radiation in Science and Technique (1) Analytical applications (Radiometric titration)

7) Applications of Nuclear Radiation in Science and Technique (1) Analytical applications (Radiometric titration) 7) Applications of Nuclear Radiation in Science and Technique (1) (Radiometric titration) The radioactive material is indicator Precipitation reactions Complex formation reactions Principle of a precipitation

More information

DIFFERENCES IN TRACE ELEMENT CONCENTRATIONS IN WHOLE BLOOD OF SMALL SIZED ANIMALS USING ANAA

DIFFERENCES IN TRACE ELEMENT CONCENTRATIONS IN WHOLE BLOOD OF SMALL SIZED ANIMALS USING ANAA 2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DIFFERENCES IN TRACE ELEMENT

More information

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF 2016 Fall Semester MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF Byungha Shin Dept. of MSE, KAIST 1 Course Information Syllabus 1. Overview of various characterization techniques (1 lecture)

More information

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 236 ABSTRACT TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Hagen Stosnach Röntec GmbH,

More information

MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES

MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES THE PUBLISHING HOUSE PROCEEDINGS OF THE ROMANIAN ACADEMY, Series A, OF THE ROMANIAN ACADEMY Volume 18, Number 1/2017, pp. 42 49 MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES Ion GRUIA University

More information

Supporting Information. Facile design of phase separation for microfluidic. droplet-based liquid phase microextraction as a front end to

Supporting Information. Facile design of phase separation for microfluidic. droplet-based liquid phase microextraction as a front end to Supporting Information Facile design of phase separation for microfluidic droplet-based liquid phase microextraction as a front end to electrothermal vaporization-icpms for the analysis of trace metals

More information

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS 255 Yoshiyuki Kataoka 1, Naoki Kawahara 1, Shinya Hara 1, Yasujiro Yamada 1, Takashi Matsuo 1, Michael Mantler 2 1 Rigaku

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water

UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water C. H. Tan a, Y. C. Moo a, M. Z. Matjafri a and H. S. Lim a a School of Physics, Universiti Sains Malaysia, 118 Pulau Pinang, Malaysia.

More information

PERFLUORINATED POLYMER GRAFTING: INFLUENCE OF PRE- IRRADIATION CONDITIONS

PERFLUORINATED POLYMER GRAFTING: INFLUENCE OF PRE- IRRADIATION CONDITIONS 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 PERFLUORINATED

More information

Physics 100 PIXE F06

Physics 100 PIXE F06 Introduction: Ion Target Interaction Elastic Atomic Collisions Very low energies, typically below a few kev Surface composition and structure Ion Scattering spectrometry (ISS) Inelastic Atomic Collisions

More information

ECE Semiconductor Device and Material Characterization

ECE Semiconductor Device and Material Characterization ECE 4813 Semiconductor Device and Material Characterization Dr. Alan Doolittle School of Electrical and Computer Engineering Georgia Institute of Technology As with all of these lecture slides, I am indebted

More information

A new detector for neutron beam monitoring

A new detector for neutron beam monitoring A new detector for neutron beam monitoring European Organization for Nuclear Research (CERN), Geneva, Switzerland in collaboration with Commissariat à l Energie Atomique (CEA), Saclay, France, Instituto

More information

Novel Plastic Microchannel-Based Direct Fast Neutron Detection

Novel Plastic Microchannel-Based Direct Fast Neutron Detection Novel Plastic Microchannel-Based Direct Fast Neutron Detection D. Beaulieu, P. de Rouffignac, D. Gorelikov, H. Klotzsch, J. Legere*, J. Ryan*, K. Saadatmand, K. Stenton, N. Sullivan, A. Tremsin Arradiance

More information

2.3 Particle Induced X-Ray Emission PIXE

2.3 Particle Induced X-Ray Emission PIXE 2.3 Particle Induced X-Ray Emission PIXE The previous section concentrated on X-ray fluorescence. This section discusses a different X-ray production technique that can lead to the development of 2-D/3-D

More information

Applications of Nuclear Analytical Techniques in Geoscience

Applications of Nuclear Analytical Techniques in Geoscience Applications of Nuclear Analytical Techniques in Geoscience J. Aspiazu a,1, J. López a, J. Ramírez a, M. E. Montero b, P. Villaseñor a a Intituto Nacional de Investigaciones Nucleares (ININ), Carretera

More information

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 219 AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Takao Moriyama 1), Atsushi Morikawa 1), Makoto

More information

X-Ray Fluorescence and Natural History

X-Ray Fluorescence and Natural History X-Ray Fluorescence and Natural History How XRF Helps XRF can be used both quantitatively (homogenous samples) and quantitatively (heterogenous samples).! Trace elements in a fossil can help identify source,

More information

Particle Analysis of Environmental Swipe Samples

Particle Analysis of Environmental Swipe Samples IAEA-SM-367/10/07 Particle Analysis of Environmental Swipe Samples D. DONOHUE, S. VOGT, A. CIURAPINSKI, F. RUEDENAUER, M. HEDBERG Safeguards Analytical Laboratory International Atomic Energy Agency Vienna,

More information

Out of vacuum characterisation of surfaces: a possible approach?

Out of vacuum characterisation of surfaces: a possible approach? Out of vacuum characterisation of surfaces: a possible approach? Lorenzo Giuntini Dipartimento di Fisica dell Università and Sezione INFN Firenze, Italy XVIII Congresso Nazionale sulla Scienza e Tecnologia

More information

XUV 773: X-Ray Fluorescence Analysis of Gemstones

XUV 773: X-Ray Fluorescence Analysis of Gemstones Fischer Application report vr118 HELM UT FISCHER GMBH + CO. KG Institut für Elektronik und Messtechnik Industriestrasse 21-7169 Sindelfingen, Germany Tel.: (+49) 731 33- - Fax: (+49) 731 33-79 E-Mail:

More information

Measurement of thin oil layers by PIXE-α

Measurement of thin oil layers by PIXE-α Measurement of thin oil layers by PIXE-α P. Brisset, D. Chambellan and J. de Sanoit CEA-LIST Département des technologies du Capteur et du Signal 91191 Gif-sur-Yvette Cedex, France International conference

More information

Data report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 18, 2010 SUMMARY

Data report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 18, 2010 SUMMARY Data report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 8, 200 SUMMARY This report summarizes the quality assurance performed during elemental

More information

Selected Topics in Physics a lecture course for 1st year students by W.B. von Schlippe Spring Semester 2007

Selected Topics in Physics a lecture course for 1st year students by W.B. von Schlippe Spring Semester 2007 Selected Topics in Physics a lecture course for 1st year students by W.B. von Schlippe Spring Semester 2007 Lecture 11 1.) Determination of parameters of the SEMF 2.) α decay 3.) Nuclear energy levels

More information

Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis

Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis Bull. Mater. Sci., Vol. 25, No. 5, October 2002, pp. 435 441. Indian Academy of Sciences. Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis M K TIWARI*,

More information

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy. UNIVERSITI SAINS MALAYSIA Second Semester Examination Academic Session 2004/2005 March 2005 KAA 502 Atomic Spectroscopy Time: 3 hours Please make sure this paper consists of FIVE typed pages before answering

More information

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE Copyright JCPDS-International Centre for Diffraction Data 2012 ISSN 1097-0002 281 IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE Susumu Imashuku 1, Deh Ping Tee 1, Yasukazu

More information

DETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS

DETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS 7 th INTERNATIONAL MULTIDISCIPLINARY CONFERENCE Baia Mare, Romania, May 17-18, 2007 ISSN-1224-3264 DETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS Mariana Dobra, Vasile Viman,

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science

More information

BIOMEDICAL APPLICATIONS OF SYNCHROTRON X-RAY FLUORESCENCE

BIOMEDICAL APPLICATIONS OF SYNCHROTRON X-RAY FLUORESCENCE Vol. 86 (1994) ACTA PHYSICA POLONICA A No. 5 Proceedings of the ISSSRNS,94, Jaszowiec 1994 BIOMEDICAL APPLICATIONS OF SYNCHROTRON X-RAY FLUORESCENCE W.M. KWIATEK H. Niewodniczański Institute of Nuclear

More information

Spectroscopy on Mars!

Spectroscopy on Mars! Spectroscopy on Mars! Pathfinder Spirit and Opportunity Real World Friday H2A The Mars Pathfinder: Geological Elemental Analysis On December 4th, 1996, the Mars Pathfinder was launched from earth to begin

More information

IN THE NAME OF ALLAH, THE MOST MERCIFUL AND COMPASSIONATE

IN THE NAME OF ALLAH, THE MOST MERCIFUL AND COMPASSIONATE IN THE NAME OF ALLAH, THE MOST MERCIFUL AND COMPASSIONATE Ion Beam Analysis of Diamond Thin Films Sobia Allah Rakha Experimental Physics Labs 04-03-2010 Outline Diamond Nanostructures Deposition of Diamond

More information

Determination of Copper in Uniformly-Doped Silicon Thin Films by Isotope-Dilution Inductively Coupled Plasma Mass Spectrometry

Determination of Copper in Uniformly-Doped Silicon Thin Films by Isotope-Dilution Inductively Coupled Plasma Mass Spectrometry Determination of Copper in Uniformly-Doped Silicon Thin Films Bull. Korean Chem. Soc. 2001, Vol. 22, No. 2 205 Determination of Copper in Uniformly-Doped Silicon Thin Films by Isotope-Dilution Inductively

More information

The scanning microbeam PIXE analysis facility at NIRS

The scanning microbeam PIXE analysis facility at NIRS Nuclear Instruments and Methods in Physics Research B 210 (2003) 42 47 www.elsevier.com/locate/nimb The scanning microbeam PIXE analysis facility at NIRS Hitoshi Imaseki a, *, Masae Yukawa a, Frank Watt

More information

PHITS calculation of the radiation field in HIMAC BIO

PHITS calculation of the radiation field in HIMAC BIO PHITS calculation of the radiation field in HIMAC BIO Ondřej Ploc, Yukio Uchihori, Hisashi Kitamura, Lembit Sihver National Institute of Radiological Sciences, Chiba, Japan Nuclear Physics Institute, Prague,

More information

Surface analysis techniques

Surface analysis techniques Experimental methods in physics Surface analysis techniques 3. Ion probes Elemental and molecular analysis Jean-Marc Bonard Academic year 10-11 3. Elemental and molecular analysis 3.1.!Secondary ion mass

More information

EBSS: Hands-on Activity #8. Operation of an electromagnetic isotope separator (EMIS) Dan Stracener Jon Batchelder

EBSS: Hands-on Activity #8. Operation of an electromagnetic isotope separator (EMIS) Dan Stracener Jon Batchelder EBSS: Hands-on Activity #8 Operation of an electromagnetic isotope separator (EMIS) Dan Stracener Jon Batchelder Schematic of RIB Production at the HRIBF ISOL: Isotope Separation On-Line 2 Managed by UT-Battelle

More information

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes Nuclear Instruments and Methods in Physics Research B 213 (24) 267 271 www.elsevier.com/locate/nimb X-ray spectrometry with Peltier-cooled large area avalanche photodiodes L.M.P. Fernandes, J.A.M. Lopes,

More information

AND LIMITATIONS K. BALOGH

AND LIMITATIONS K. BALOGH SENSITIVITY OF THE 40 Ar 39 Ar METHOD: NEW POSSIBILITIES AND LIMITATIONS K. BALOGH Institute of Nuclear Research, Hungarian Academy of Sciences, 4026 Debrecen, Bem t. 18/c, Hungary; balogh@moon.atomki.hu

More information

Past searches for kev neutrinos in beta-ray spectra

Past searches for kev neutrinos in beta-ray spectra Past searches for kev neutrinos in beta-ray spectra Otokar Dragoun Nuclear Physics Institute of the ASCR Rez near Prague dragoun@ujf.cas.cz supported by GAČR, P203/12/1896 The ν-dark 2015 Workshop TUM

More information

NUCLEAR EMISSIONS FROM TITANIUM HYDRIDE/DEUTERIDE INDUCED BY POWERFUL PICOSECOND LASER BEAM

NUCLEAR EMISSIONS FROM TITANIUM HYDRIDE/DEUTERIDE INDUCED BY POWERFUL PICOSECOND LASER BEAM NUCLEAR EMISSIONS FROM TITANIUM HYDRIDE/DEUTERIDE INDUCED BY POWERFUL PICOSECOND LASER BEAM A. S. ROUSSETSKI P.N. Lebedev Physical Institute Russian Academy of Sciences, 3 Leninsky prospect, 119991 Moscow,

More information

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials 296 Time-Resolved μ-xrf and Elemental Mapping of Biological Materials K. Tsuji 1,2), K. Tsutsumimoto 1), K. Nakano 1,2), K. Tanaka 1), A. Okhrimovskyy 1), Y. Konishi 1), and X. Ding 3) 1) Department of

More information

Analysis Repeatability of Trace and Major Elements in a Water Sample

Analysis Repeatability of Trace and Major Elements in a Water Sample Analysis Repeatability of Trace and Major Elements in a Water Sample Agnès COSNIER HORIBA Scientific Longjumeau, France Keywords: environment Elements: Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li,

More information

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis MT-0.6026 Electron microscopy Scanning electron microscopy and electron probe microanalysis Eero Haimi Research Manager Outline 1. Introduction Basics of scanning electron microscopy (SEM) and electron

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

Stable isotope. Relative atomic mass

Stable isotope. Relative atomic mass Stable isotope Relative atomic mass Mole fraction 106 Cd 105.906 460 0.012 45 108 Cd 107.904 183 0.008 88 110 Cd 109.903 007 0.124 70 111 Cd 110.904 183 0.127 95 112 Cd 111.902 763 0.241 09 113 Cd 112.904

More information

1 of 5 14/10/ :21

1 of 5 14/10/ :21 X-ray absorption s, characteristic X-ray lines... 4.2.1 Home About Table of Contents Advanced Search Copyright Feedback Privacy You are here: Chapter: 4 Atomic and nuclear physics Section: 4.2 Absorption

More information

Determination of the activity of radionuclides

Determination of the activity of radionuclides BUREAU NATIONAL DE MÉTROLOGIE COMMISSARIAT À L'ÉNERGIE ATOMIQUE LABORATOIRE NATIONAL HENRI BECQUEREL Note technique LNHB/04-33 Determination of the activity of radionuclides contained in volume samples

More information

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-ray Photoelectron Spectroscopy Introduction Qualitative analysis Quantitative analysis Charging compensation Small area analysis and XPS imaging

More information

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES 249 PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES F. I. G. M. Borges, S. J. C. do Carmo, T. H. V. T. Dias, F.

More information

Available online at ScienceDirect. Energy Procedia 71 (2015 ) Yokohama, Kanagawa, Japan

Available online at   ScienceDirect. Energy Procedia 71 (2015 ) Yokohama, Kanagawa, Japan Available online at www.sciencedirect.com ScienceDirect Energy Procedia 71 (21 ) 22 26 The Fourth International Symposium on Innovative Nuclear Energy Systems, INES-4 Design of a proton-induced quasimonochromatic

More information

Materials Analysis Using Fast Ions

Materials Analysis Using Fast Ions A. Denker, W. Bohne, J. Rauschenberg,J. Röhrich, E. Strub Ionenstrahllabor Hahn-Meitner-Institut Berlin Materials Analysis Using Fast Ions Introduction: Energy Loss PIXE Proton Induced X-ray Emission RBS

More information

Ionization Techniques Part IV

Ionization Techniques Part IV Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector

More information

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS 302 ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS M. Song 1, K. Mitsuishi 1, M. Takeguchi 1, K. Furuya 1, R. C. Birtcher 2 1 High Voltage Electron Microscopy Station, National

More information

PRELIMINARY CONCEPT OF A ZERO POWER NUCLEAR REACTOR CORE

PRELIMINARY CONCEPT OF A ZERO POWER NUCLEAR REACTOR CORE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 PRELIMINARY CONCEPT OF

More information

Sample Analysis Design. Solution Mode

Sample Analysis Design. Solution Mode Sample Analysis Design Solution Mode Step I Sample preparation The quality of your data will only be as good as the quality of your sample i.e. did you adequately prepare your sample in the clean lab?

More information

Secondary ion mass spectrometry (SIMS)

Secondary ion mass spectrometry (SIMS) Secondary ion mass spectrometry (SIMS) Lasse Vines 1 Secondary ion mass spectrometry O Zn 10000 O 2 Counts/sec 1000 100 Li Na K Cr ZnO 10 ZnO 2 1 0 20 40 60 80 100 Mass (AMU) 10 21 10 20 Si 07 Ge 0.3 Atomic

More information

MODULE 4.3 Atmospheric analysis of particulates

MODULE 4.3 Atmospheric analysis of particulates MODULE 4.3 Atmospheric analysis of particulates Measurement And Characterisation Of The Particulate Content 1 Total particulate concentration 1 Composition of the particulate 1 Determination of particle

More information

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K. 783 SCOPE AND LIMITATIONS XRF ANALYSIS FOR SEMI-QUANTITATIVE Introduction Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K. Historically x-ray fluorescence spectrometry has

More information

MEASUREMENT CAPABILITIES OF X-RAY FLUORESCENCE FOR BPSG FILMS

MEASUREMENT CAPABILITIES OF X-RAY FLUORESCENCE FOR BPSG FILMS , MEASUREMENT CAPABILITIES OF X-RAY FLUORESCENCE FOR BPSG FILMS K.O. Goyal, J.W. Westphal Semiconductor Equipment Group Watkins-Johnson Company Scotts Valley, California 95066 Abstract Deposition of borophosphosilicate

More information

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009] 2 Vol. 36(1) [Jan. Feb. 2009] BRIEFING h233i Elemental Impurities Procedures. This proposed new general test chapter is the second of two being developed to replace the general test chapter Heavy Metals

More information

TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS

TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS W.L. Griffin 1, 3, Sonal Rege 1, Rondi M. Davies 1, 2, Simon Jackson 1, Suzanne Y. O Reilly 1 1.ARC National Key

More information

Simulations of Ion Beam Analysis with laser-driven proton sources at Politecnico di Milano. Francesco Mirani Frascati, February 21ˢ, 2018

Simulations of Ion Beam Analysis with laser-driven proton sources at Politecnico di Milano. Francesco Mirani Frascati, February 21ˢ, 2018 Simulations of Ion Beam Analysis with laser-driven proton sources at Politecnico di Milano Francesco Mirani Frascati, February 21ˢ, 2018 The ENSURE team at Politecnico di Milano Matteo Passoni Associate

More information

Trace elemental analysis of challenging samples in the mining and resource industry. Bruker Nano GmbH, Berlin, Germany April 8, 2015

Trace elemental analysis of challenging samples in the mining and resource industry. Bruker Nano GmbH, Berlin, Germany April 8, 2015 Trace elemental analysis of challenging samples in the mining and resource industry Bruker Nano GmbH, Berlin, Germany April 8, 2015 4/8/2015 Welcome Today s Topics Introduction to TXRF learn about X-ray

More information

Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry

Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry 2005 The Japan Society for Analytical Chemistry 143 Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry M. K. TIWARI, A. K. SINGH, and K. J. S. SAWHNEY Synchrotron

More information

Trace Element Analysis: Time and Cost Savings with TXRF. Bruker AXS Microanalysis GmbH Berlin, Germany

Trace Element Analysis: Time and Cost Savings with TXRF. Bruker AXS Microanalysis GmbH Berlin, Germany Trace Element Analysis: Time and Cost Savings with TXRF Bruker AXS Microanalysis GmbH Berlin, Germany Welcome Today s Topics Introduction to XRF and TXRF The S2 PICOFOX benchtop TXRF system Application

More information