EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

Size: px
Start display at page:

Download "EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD"

Transcription

1 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD Danilo C. Vasconcelos 1, Claubia Pereira 1, Sergio Gallardo 2, Zildete Rocha 3, Talita O. Santos 2 1 Departamento de Engenharia Nuclear Escola de Engenharia Universidade Federal de Minas Gerais Departamento de Engenharia Nuclear - PCA 1 - Anexo Engenharia, Av. Antônio Carlos, 6627 Campus UFMG CEP , Belo Horizonte, MG, Brazil 2 Departamento de Ingeniería Química y Nuclear Universitat Politècnica de València, Camino de Vera, s/n, Valencia, Spain 3 Centro de Desenvolvimento da Tecnologia Nuclear / Comissão Nacional de Energia Nuclear (CDTN/CNEN), Laboratório de Trítio, Caixa Postal 941,CEP , Belo Horizonte, Minas Gerais, Brazil ABSTRACT A gamma spectrometer including an HPGe detector is commonly used for environmental radioactivity measurements. In this work, a systematic study of the efficiency simulation of an HPGe coaxial detector has been performed in the energy range between 50 and 1500 kev using the Monte Carlo code MCNP. With this aim, a detailed model of the HPGe detector and a Marinelli beaker containing a certified gamma source has been done. Once the Pulse Height Distribution (PHD) registered in the detector is simulated, the efficiency curve is obtained by calculating the net peak areas and taking into account the certified activity of the source. In order to avoid the uncertainty due to the net area calculation, the simulated PHD is treated using the GAMMAVISION program, which is also used in the experimental spectra processing. The simulated and experimental efficiency curves are compared observing a satisfactory agreement. 1. INTRODUCTION A gamma spectrometer including an HPGe (High Purity Germanium) detector is commonly used for environmental radioactivity measurements. The Environmental Radioactivity Laboratory at Tritium - CDTN (Centro de Desenvolvimento da Tecnologia Nuclear) in Belo Horizonte, Brazil undertakes experimental work using this detector for laboratory measurements. To obtain reliable measurements of the environmental radionuclide activity, the knowledge of the absolute peak efficiency of the detector is required. It can be obtained by employing known radioactive standard source homogenously distributed in a Marinelli beaker of same dimension and similar composition. In the present work, the Monte Carlo method was applied complementary for detector efficiency calibration using version 5 of the MCNP code [1]. Monte Carlo N-particles Transport Code (MCNP) is an extremely advanced Monte Carlo program, which contains all the necessary cross-section data for neutron, photon, and electron transport calculations. The use of Monte Carlo simulation of a detector s response to incident photons is becoming increasingly important. Some authors used the MCNP code for Full Energy Peak Efficiency (FEPE) determinations [2].

2 The version used is suitable for modelling the detector response, since it contains a tally (F8), which is specific for detector Pulse Height Distribution (PHD) determination. Thus, the detector geometry is modelled with MCNP code, which simulates the detection process to obtain spectrum peaks [2]. Marinelli beaker geometries have been considered as well as water containing the standards for the simulation. Results obtained from Monte Carlo simulation have been compared with experimental measurements in order to validate the model. In this kind of semiconductor detectors (HPGe), germanium dead layer extremely affects the efficiency of low-energy photons. For this purpose, the Monte Carlo simulations are very useful to know the effect of the dead layer thickness on the efficiency, due to the fact that this value is normally not well known in experimental measurements Experimental measurements 2. METHODOLOGY A High Purity Germanium coaxial detector system has been used for experimental measurements. The detector model is a Canberra GC1519, with a closed end coaxial geometry. The main performance specifications of the detector are as follows: relative efficiency at 1.33 MeV Co-60 is 18.7%, FWHM resolution of 1.78 kev, FWTM resolution of 3.29 kev both at 1.33 MeV Co-60, and the peak-to-compton ratio for Co-60 is 44.7:1. The crystal diameter is 61mm and the length is 25.5 mm. The core hole diameter is 8 mm, and the core hole length is 15.5 mm. The end cap to crystal distance is 5 mm. The cryostat window material is aluminium with 1.5 mm thickness. The crystal has also a layer composed by Mylar and Kapton in front of the cryostat window. The effective thickness of the dead layer is not well known a cause of the existence of a transition zone between the inactive layer and the active germanium in the crystal whose thickness is very difficult to be accurately estimated, but for this kind of detectors it has been assumed a dead layer of 1.5 mm thick (including the outer and inner electrode thickness). The source used for measurements was a calibration gamma cocktail solution, covering the energy range between 80 and 1400 kev. The radionuclides contained in the source solution are listed in Table 1 together with the main peak energy, the branching ratio and the certified activity. The uncertainty of the experimental efficiency was estimated using the ISO standard [3] from certified activities and counting results. All of the measured gamma spectra were analysed using the GAMMAVISION program [3]. The mixed gamma-ray standard contains the following radionuclides: Ba-133, Cs-134, Cs-137 and Co-60. From the measurement of calibration sources, the experimental efficiencies in the energy range between 80 and 1400 kev were calculated. The experimental efficiency at energy E γ for a given measuring condition is: (1) where N Ei is the net area under the full-energy peak corresponding to Eγ energy photons emitted by a radionuclide with a known activity, A, f is the emission probability, m is the sample mass and t is the counting time [4].

3 Table 1. Gamma standard properties. Energy Branching (kev) ratio Activity (Bq) Gamma/s Ba Ba Ba Ba Ba Co Co Cs Cs Cs The Monte Carlo model A MCNP model has been developed for the system defined by the HP Ge detector and the Marinelli beaker. The Marinelli beaker was filled with water containing the calibration gamma solution sit directly on top of the detector providing relatively high-efficiency geometry, as it can be seen in Figure 1. This figure has been obtained using the Sabrina program [5]. Marinelli beaker Air Gamma solution in water Ge detector Aluminum holder Figure 1. Representation of the Marinelli beaker and the Ge detector.

4 The F8 (Pulse Height Distribution) tally has been used for photons and electrons. A detailed physics treatment, including photoelectric effect with fluorescence production, incoherent scattering with form factors and pair production, has been considered in the energy range between and 2.0 MeV. Simulating electron tracks requires large computation time, but it has been included in the model due to its influence on the spectrum. To decrease computing time, an energy cut off has been used: 30 kev for electron transport and 1 kev for photons. The number of histories has been set to 100 million in order to obtain a relative error less than 1% at every peak centroid. The use of the GEB (Gaussian Energy Broadening) card option provides a spectrum that can be compared with the experimental one in terms of resolution (FWHM and FWTM). The GEB card parameters have been chosen to reproduce the actual resolution calibration. Determination of peak efficiency from MCNP results has been performed applying the same method used by the GAMMAVISION program in experimental measurements. This program obtains the net area of each peak subtracting the background to the gross area of the peak. 3. RESULTS AND DISCUSSION The PHD obtained with MCNP corresponds to one-photon emitted by the source. In order to compare both experimental and simulated PHDs it is necessary to take into account the total number of gamma/s emitted by the source. For this purpose, the calculated PHD has been multiplied by the acquisition live time (86400 s) and by the total number of gamma/s emitted by the calibration gamma cocktail. The comparison of both PHDs is shown in Figure 2. As it can be seen in this figure, it has been achieved a satisfactory agreement between experimental and simulated measurements, even in the peak resolution (FWHM). Figure 2. Comparison of experimental and simulated Pulse Height Distribution.

5 The Compton edges are well simulated as well as the contribution of each radionuclide. Regarding to the photo peaks, it is important to remark that the simulation slightly overestimates the height of all of them, and consequently the efficiency. Experimental and simulated PHDs have been analyzed with GAMMAVISON in order to obtain the efficiency curve for each case. In Table 2 and Table 3 it can be seen the results of analysis (background counts, net area counts, intensity counts, 1-sigma uncertainty, resolution (FWHM) and efficiency) for experimental and simulated PHDs respectively. In Table 4, the ratio between simulated and experimental efficiency is shown. As it can be seen in this table, the values obtained with MCNP-GAMMAVISION slightly overestimate the efficiency for almost all the energies considered. For the comparison between measured and calculated efficiencies, a method of uncertainty propagation was applied considering a normal distribution with K = 1, providing a confidence level equal to In the MCNP calculations, the number of particles simulated has been chosen to achieve a relative error lower than 1% for each energy bin. Table 2. Experimental PHD analysis with GAMMAVISON. Energy (kev) g/s Background Counts Net Area Counts Intensity Counts/s Uncert 1 Sigma (%) FWHM (kev) Efficiency Ba Ba Ba Ba Ba Cs Cs Cs Co Co Table 3. MCNP PHD analysis with GAMMAVISON. Net Energy Background Intensity g/s Area (kev) Counts Counts/s Counts Uncert 1 Sigma (%) FWHM (kev) Efficiency Ba Ba Ba Ba Ba Cs Cs Cs Co Co

6 Table 4. Experimental and simulated efficiency obtained with GAMMAVISION. Energy Experimental MCNP Ratio (kev) efficiency efficiency MCNP/Exp Ba Ba Ba Ba Ba Cs Cs Cs Co Co Figure 3 shows the efficiency curve as a function of energy, for the ten experimental and simulated peaks studied. The efficiency of the detector varies with the incident photon energy and their interactions with the active volume. The maximum ratio MCNP/experiment is 1.12 corresponding to Cs-134 and Co-60. Figure 3. Experimental and simulated efficiency. Experimental errors have been characterized considering the uncertainty of standard cocktail provided by manufacturer and errors due to the experimental measurements. MCNP error is less than 1% for every efficiency value, so error bars cannot be seen in figure.

7 For the energy range considered, one of the most important parameter on the detection efficiency is the dead layer thickness. In order to analyze its effect, different MCNP models have been developed varying the dead layer (0.5, 1.0, 1.25 and 1.50 mm). When the dead layer increases, the low-energy efficiency decreases due to the fact that the low-energy photons do not reach the active detection volume. As it can be seen in the Figure 4 and Table 5, the effect is more evident for Ba-133, as expected. Normally, the precise dead layer value is not well known, and this kind of analysis can be useful for understanding its effect. When the dead layer thickness is increased from 0.5 to 1.5 mm, the ratio MCNP/Experiment is reduced from 2.11 to 1.0 in the case of Ba-133 (80.99 kev). This effect is also important in the higher energy range. In this case, the ratio MCNP/Experiment varies from 1.44 to 1.12 for Co MeV. Figure 4. Dead layer thickness effect on detection efficiency. However, in the case of Co-60, there is also another effect to be considered: the true coincidence effect. The effect of coincidence summing on efficiency calibration has been taken into account dividing experimental efficiencies by the correction factors (0.93 and for each Co-60 gamma line). It is important to remark that these factors are only valid for aqueous solutions [6]. True coincidence is not taken into account in MCNP.

8 Table 5. Dead layer thickness effect on detection efficiency. Ratio MCNP/Experiment. Energy Ratio model 1 Ratio model 2 Ratio model 3 Ratio model 4 (kev) /Experimental /Experimental /Experimental /Experimental Ba Ba Ba Ba Ba Cs Cs Cs Co Co Model 1: 1.5 mm; Model 2: 1.25 mm; Model 3: 1.0 mm and Model 4: 0.5 mm. The active volume of the detector also determines de efficiency of detection. If the geometrical dimensions of the crystal are fixed and dead layer is supposed to be known, then the only parameter affecting to the active volume is the core hole of the crystal. In order to test the importance of this hole, some MCNP models have been performed varying the diameter of the core hole (3, 4 and 5 mm). In Figure 5, it is represented the detector efficiency when the core hole diameter varies. It can be seen, that the effect is almost negligible for this energy range. It can be stated that the most important parameter affecting the detector efficiency is the dead layer thickness. Figure 5. Core hole diameter effect on detection efficiency.

9 4. CONCLUSIONS The MCNP code, based on the Monte Carlo method, is a useful tool to calibrate in efficiency an HP Ge detector as those commonly used in Environmental Radioactivity laboratories. The combined method MCNP-GAMMAVISION permits to compare the efficiency curve obtained by simulation with the experimental curve, avoiding errors and deviations due to the calculation method of the net peak area. The agreement between simulated and experimental efficiencies is satisfactory and allows studying relevant design parameters of this kind of detectors. Dead layer thickness is one of the most important parameter affecting the efficiency of the detection system. The effect of this parameter is more important at lowenergy range (below 400 kev), but it is not negligible for the high-energy range (up to Co- 60). It is stated that, for these conditions, the variation of the equivalent volume of the inner core hole is not relevant for the detection efficiency. ACKNOWLEDGMENTS The authors thank to the Centro de Desenvolvimento da Tecnologia Nuclear CDTN/CNEN, where this study was carried out. The authors also are grateful to FAPEMIG, CNPq and CAPES for financial assistance. REFERENCES 1. Briesmeister, J.F., MCNP A general Monte Carlo code for neutron and photon transport, version 5, Los Alamos National Laboratory, Report LA M, Ródenas, J.; Gallardo, S.; Ballester, S.; Primault, V.; Ortiz, J. Application of the Monte Carlo method to the analysis of measurement geometries for the calibration of a HP Ge detector in an environmental radioactivity laboratory. Nuclear Instruments and Methods in Physics Research B 263, , GammaVision, Gamma-ray Spectrum Analysis and MCA Emulation for MS Windows, Software User s Manual, version 32, V Debertin, K., Helmer, R.G. Gamma and X-Ray Spectrometry with Semiconductor Detectors. North-Holland, Amsterdam, K. A. Van Riper, Sabrina User s Guide Windows version, ed. White Rock Science, White Rock, NM, USA D. Arnold, O. Sima, Extension of the efficiency calibration of germanium detectors using the GESPECOR software, Appl. Radiat. Isotopes 61 (2004) 117.

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods N. Stefanakis 1 1 GMA Gamma measurements and analyses e.k. PO Box 1611, 72706 Reutlingen,

More information

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information.

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information. Experimental and Monte Carlo determination of Gamma Spectrometry Efficiency Pavel Dryak, Petr Kovář Czech Metrology Institute February 2009 Leading motive of our testing The producer send the drawing after

More information

Basic hands-on gamma calibration for low activity environmental levels

Basic hands-on gamma calibration for low activity environmental levels Basic hands-on gamma calibration for low activity environmental levels Iolanda Osvath Presented by Mats Eriksson Environment Laboratories Marine Environment Laboratories, Monaco Radiometrics Laboratory

More information

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN Problem P7 Stéphanie Ménard Dosimetry Department 92262 Fontenay-aux aux-roses FRANCE What are the applications of Gamma-Ray Spectrometry in Radiological Protection and in Safety? In the environment: after

More information

Indian J.Sci.Res.2(3) :25-32, 2011

Indian J.Sci.Res.2(3) :25-32, 2011 Indian J.Sci.Res.2(3) :25-32, 211 The International Atomic Energy Agency (IAEA) via technical cooperation and coordinated research projects (CRP), expert services, and fellowship awards developed a nondestructive,

More information

Retrieval Of GammaCell 220 Irradiator Isodose Curves With MCNP Simulations And Experimental Measurements

Retrieval Of GammaCell 220 Irradiator Isodose Curves With MCNP Simulations And Experimental Measurements 120 R.R. Rodrigues et al. Retrieval Of GammaCell 220 Irradiator Isodose Curves With MCNP Simulations And Experimental Measurements R.R. Rodrigues, S.E. Grynberg, A.V. Ferreira, L.C.M. Belo, and P.L. Squair

More information

Document downloaded from: This paper must be cited as:

Document downloaded from:  This paper must be cited as: Document downloaded from: http://hdl.handle.net/10251/47496 This paper must be cited as: Querol Vives, A.; Gallardo Bermell, S.; Ródenas Diago, J.; Verdú Martín, GJ. (2011). Parametric study of the X-ray

More information

Brazilian Journal of Physics ISSN: Sociedade Brasileira de Física Brasil

Brazilian Journal of Physics ISSN: Sociedade Brasileira de Física Brasil Brazilian Journal of Physics ISSN: 0103-9733 luizno.bjp@gmail.com Sociedade Brasileira de Física Brasil Araújo, Arione; Pereira, Claubia; Fortini Veloso, Maria Auxiliadora; Lombardi Costa, Antonella; Moura

More information

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods Asian J. Exp. Sci., Vol. 28, No. 2, 2014; 23-31 Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods Rahim Khabaz, Farhad Yaghobi

More information

Determination of the activity of radionuclides

Determination of the activity of radionuclides BUREAU NATIONAL DE MÉTROLOGIE COMMISSARIAT À L'ÉNERGIE ATOMIQUE LABORATOIRE NATIONAL HENRI BECQUEREL Note technique LNHB/04-33 Determination of the activity of radionuclides contained in volume samples

More information

High Purity Germanium Detector Calibration at ISOLDE

High Purity Germanium Detector Calibration at ISOLDE High Purity Germanium Detector Calibration at ISOLDE Guðmundur Kári Stefánsson Summer Student of Maria Borge September 5, 2013 Abstract: This Summer Student Project involved the test and calibration of

More information

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy ISOCS / LabSOCS Calibration software for Gamma Spectroscopy Counts Setup Hardware Peak Shaping Parameters Rise Time Flat Top Pole-zero Number of Channels Signal Gain Setting up a detector for measurement

More information

THERMAL HYDRAULIC MODELING OF THE LS-VHTR

THERMAL HYDRAULIC MODELING OF THE LS-VHTR 2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 THERMAL HYDRAULIC MODELING OF

More information

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE Ján Haščík, Branislav Vrban, Jakub Lüley, Štefan Čerba, Filip Osuský, Vladimír Nečas Slovak University of Technology

More information

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48 Introduction to Environmental Measurement Techniques 2016 Radioactivity Dana Pittauer (dpittauer@marum.de) 1of 48 Introduction Radioisotopes are of interest in environmental physics for several reasons:

More information

EXPERIMENTAL DETERMINATION OF NEUTRONIC PARAMETERS IN THE IPR-R1 TRIGA REACTOR CORE

EXPERIMENTAL DETERMINATION OF NEUTRONIC PARAMETERS IN THE IPR-R1 TRIGA REACTOR CORE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 EXPERIMENTAL DETERMINATION

More information

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data International Journal of Advanced Research in Physical Science (IJARPS) Volume 5, Issue 10, 2018, PP 24-28 ISSN No. (Online) 2349-7882 www.arcjournals.org Comparison of the Photo-peak Efficiencies between

More information

Calibration of A Small Anode Germanium Well Detector

Calibration of A Small Anode Germanium Well Detector Calibration of A Small Anode Germanium Well Detector DTU Nutech, Roskilde, Denmark Joonas Tikkanen Introduction Small anode germanium (SAGe) well detector manufactured by Canberra/Mirion Measurements possible

More information

Preparation of Standard Source as a Petri Dish for Plant by Using

Preparation of Standard Source as a Petri Dish for Plant by Using Preparation of Standard Source as a Petri Dish for Plant by Using 152 Eu Element Khalid H. AL-Ubaidi 1 Salam K. AL-Nasri 2 Auday T. AL-Bayati 1* 1.Department of Physics, College of Education / Ibn Al-

More information

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia 7 th Workshop on European Collaboration for Higher Education and Research in Nuclear Engineering & Radiological Protection Bruxelles, Belgique 30 May - 1 June 2011 Activities at the Laboratory of the Nuclear

More information

Characterization and Monte Carlo simulations for a CLYC detector

Characterization and Monte Carlo simulations for a CLYC detector Characterization and Monte Carlo simulations for a CLYC detector A. Borella 1, E. Boogers 1, R.Rossa 1, P. Schillebeeckx 1 aborella@sckcen.be 1 SCK CEN, Belgian Nuclear Research Centre JRC-Geel, Joint

More information

Maria Ângela de B. C. Menezes. Radojko Jaćimović. Cláubia Pereira

Maria Ângela de B. C. Menezes. Radojko Jaćimović. Cláubia Pereira SPATIAL DISTRIBUTION OF NEUTRON FLUX IN GEOLOGICAL LARGER SAMPLE ANALYSIS AT CDTN/CNEN, BRAZIL 1 Maria Ângela de B. C. Menezes 2 Radojko Jaćimović 3 Cláubia Pereira 1 Nuclear Technology Development Center/Brazilian

More information

Validation of the Monte Carlo Model Developed to Estimate the Neutron Activation of Stainless Steel in a Nuclear Reactor

Validation of the Monte Carlo Model Developed to Estimate the Neutron Activation of Stainless Steel in a Nuclear Reactor Joint International Conference on Supercomputing in Nuclear Applications and Monte Carlo 2010 (SNA + MC2010) Hitotsubashi Memorial Hall, Tokyo, Japan, October 17-21, 2010 Validation of the Monte Carlo

More information

A TEMPERATURE DEPENDENT ENDF/B-VI.8 ACE LIBRARY FOR UO2, THO2, ZIRC4, SS AISI-348, H2O, B4C AND AG-IN-CD

A TEMPERATURE DEPENDENT ENDF/B-VI.8 ACE LIBRARY FOR UO2, THO2, ZIRC4, SS AISI-348, H2O, B4C AND AG-IN-CD 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 A TEMPERATURE

More information

Monte Carlo Simulations for a Preliminary Design of TRIGA IPR-R1 PGAA Facility

Monte Carlo Simulations for a Preliminary Design of TRIGA IPR-R1 PGAA Facility J. Chem. Chem. Eng. 10 (2016) 256-270 doi: 10.17265/1934-7375/2016.06.002 Monte Carlo Simulations for a Preliminary Design of TRIGA IPR-R1 PGAA Facility D DAVID PUBLISHING Bruno Teixeira Guerra 1, 2, Alexandre

More information

Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor. Abstract

Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor. Abstract Organized and hosted by the Canadian Nuclear Society. Vancouver, BC, Canada. 2006 September 10-14 Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor

More information

Simulation of 4π HPGe Compton-Suppression spectrometer

Simulation of 4π HPGe Compton-Suppression spectrometer Vol. 9(2), pp. 13-19, 30 January, 2014 DOI: 10.5897/IJPS2013.4075 ISSN 1992-1950 2014 Academic Journals http://www.academicjournals.org/ijps International Journal of Physical Sciences Full Length Research

More information

EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY

EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY R.M. Keyser New Product Development, ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831 1 INTRODUCTION

More information

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DEVELOPMENT

More information

First tests of the big volume ultra low background gamma spectrometer

First tests of the big volume ultra low background gamma spectrometer First tests of the big volume ultra low background gamma spectrometer N. Todorović,, D. MrđaM rđa,, I. Bikit, M. Vesković,, S. Forkapić,, J. Slivka Departman za fiziku, PMF, Novi Sad 1. Introduction Even

More information

Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code

Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code Journal of Materials and Environmental Sciences ISSN : 2028-2508 Copyright 2017, University of Mohammed Premier Oujda Morocco J. Mater. Environ. Sci., 2017 Volume 8, Issue 12, Page 4560-4565 http://www.jmaterenvironsci.com

More information

arxiv:nucl-ex/ v2 21 Jul 2005

arxiv:nucl-ex/ v2 21 Jul 2005 Gamma-spectrometric uranium age-dating using intrinsic efficiency calibration arxiv:nucl-ex/0506029v2 21 Jul 2005 Cong Tam Nguyen and József Zsigrai Institute of Isotopes of the Hungarian Academy of Sciences

More information

Researchers at the University of Missouri-Columbia have designed a triple crystal

Researchers at the University of Missouri-Columbia have designed a triple crystal Childress, N. L. and W. H. Miller, MCNP Analysis and Optimization of a Triple Crystal Phoswich Detector, Nuclear Instruments and Methods, Section A, 490(1-2), 263-270 (Sept 1, 2002). Abstract Researchers

More information

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays Alison V. Goodsell, William S. Charlton alisong@tamu.edu, charlton@ne.tamu.edu Nuclear Security Science & Policy Institute Texas A&M University,

More information

Characterization of semiconductor detectors for γ-ray and x-ray spectrometry

Characterization of semiconductor detectors for γ-ray and x-ray spectrometry Author: Facultat de Física, Universitat de Barcelona, Diagonal 645, 08028 Barcelona, Spain. Advisor: José M. Fernández-Varea Semiconductor detectors are routinely used in γ-ray and x-ray spectrometry.

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore

More information

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA The Underwater Spectrometric System Based on CZT Detector for Survey of the Bottom of MR Reactor Pool 13461 Victor Potapov, Alexey Safronov, Oleg Ivanov, Sergey Smirnov, Vyacheslav Stepanov National Research

More information

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL A. Specification Gamma-spectrometry method is used to identify and determine the activity concentration

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science

More information

SOME ASPECTS OF MONTE CARLO SIMULATION FOR EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS

SOME ASPECTS OF MONTE CARLO SIMULATION FOR EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS ICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 SOME ASPECTS OF MONTE CARLO SIMULATION S.Hurtado Universidad de Sevilla SPAIN 1 MONTE CARLO SIMULATION DISADVANTAGES Self-absorption correction

More information

PROFILE S and C Series

PROFILE S and C Series P-type Semi-planar and Coaxial HPGe Detectors Exceptional Resolution and Stable, Low-Energy Efficiency Comprehensive Solutions to Demanding Applications Exceptional resolution and stable, low energy efficiency

More information

ARTICLE IN PRESS. Applied Radiation and Isotopes

ARTICLE IN PRESS. Applied Radiation and Isotopes Applied Radiation and Isotopes 68 (21) 1349 1353 Contents lists available at ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Standardization and measurement

More information

Performance of the True Coincidence Correction Method in GammaVision. Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L.

Performance of the True Coincidence Correction Method in GammaVision. Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L. Performance of the True Coincidence Correction Method in GammaVision Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L. ORTEC, PerkinElmer Instruments, Inc. 801 South Illinois Avenue, Oak Ridge,

More information

Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods

Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods Journal of Physics: Conference Series PAPER OPEN ACCESS Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods To cite this article: A Ruzzarin et al 2016 J. Phys.: Conf.

More information

Digital simulation of neutron and gamma measurement devices

Digital simulation of neutron and gamma measurement devices 3Security of radioactive materials and transport 3 2 Digital simulation of neutron and gamma measurement devices A.-L. WEBER (IRSN) 1 - Maximum activity that a radioactive element can present without being

More information

Inaccuracy Determination in Mathematical Model of Labsocs Efficiency Calibration Program

Inaccuracy Determination in Mathematical Model of Labsocs Efficiency Calibration Program Inaccuracy Determination in Mathematical Model of Labsocs Efficiency Calibration Program M Kuznetsov, T Nikishkin, S Chursin National Research Tomsk Polytechnic University, Tomsk, Russia E-mail: chursinss@tpu.ru

More information

The Best Gamma-Ray Detector

The Best Gamma-Ray Detector /sec The Best Gamma-Ray Detector What is the best detector for gamma-ray spectroscopy? Amptek, Inc. provides signal processing electronics which can be used with a variety of detectors (some available

More information

A coincidence method of thorium measurement

A coincidence method of thorium measurement A coincidence method of thorium measurement Nevenka Antovic a*, Perko Vukotic a and Nikola Svrkota b a Faculty of Natural Sciences and Mathematics, University of Montenegro, Cetinjski put b.b., 81000 Podgorica,

More information

Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure

Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure ILA PILLALAMARRI Earth Atmospheric & Planetary Sciences Neutron Activation Analysis

More information

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 2001-22, p.30-36 APPLICATION OF Ge SEMI-CONDUCTOR DETECTOR TO WHOLE-BODY COUNTER S. Kinase 1 2, H. Noguchi 1 and T. Nakamura 2 1 Japan

More information

Slides by: Prof. Abeer Alharbi

Slides by: Prof. Abeer Alharbi Slides by: Prof. Abeer Alharbi electromagnetic radiation of high energy. They are produced by sub-atomic particle interactions, such as electron-positron annihilation, neutral pion decay, radioactive decay,

More information

PRESSURE DROP OF FLOW THROUGH PERFORATED PLATES

PRESSURE DROP OF FLOW THROUGH PERFORATED PLATES 27 International Nuclear Atlantic Conference - INAC 27 Santos, SP, Brazil, September 3 to October 5, 27 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-2-1 PRESSURE DROP OF FLOW THROUGH

More information

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools 10th Nuclear Science Training Course with NUCLEONICA, Cesme, Turkey, 8-10 October, 2008 1 Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools A.N. Berlizov ITU - Institute for Transuranium

More information

THE COMPTON EFFECT Last Revised: January 5, 2007

THE COMPTON EFFECT Last Revised: January 5, 2007 B2-1 THE COMPTON EFFECT Last Revised: January 5, 2007 QUESTION TO BE INVESTIGATED: How does the energy of a scattered photon change after an interaction with an electron? INTRODUCTION: When a photon is

More information

Chapter 8 Hyper-Pure Germanium Detector

Chapter 8 Hyper-Pure Germanium Detector Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-1 Chapter 8 Hyper-Pure Germanium Detector 8.1. Introduction Silicon semiconductor detectors described in the previous chapter have depletion

More information

EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS

EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS NUCLEAR PHYSICS EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS MAGDALENA TOMA 1, OCTAVIAN SIMA 2, CARMEN CRISTACHE 1, FELICIA DRAGOLICI 1, LAURENÞIU

More information

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC A. Specification Whole body counting method is used to detect the gamma rays emitted by radio nuclides,

More information

NUCL 3000/5030 Laboratory 2 Fall 2013

NUCL 3000/5030 Laboratory 2 Fall 2013 Lab #2: Passive Gamma Spec Measurements in Decoding Natural Radioactivity in SLC Area Objectives a. Learn basics of gamma spectroscopy b. Learn the equipment in Counting stations #4, #5 and #8 c. Apply

More information

Compton suppression spectrometry

Compton suppression spectrometry Compton suppression spectrometry In gamma ray spectrometry performed with High-purity Germanium detectors (HpGe), the detection of low intensity gamma ray lines is complicated by the presence of Compton

More information

Research Article Sensitivity Analysis of the TRIGA IPR-R1 Reactor Models Using the MCNP Code

Research Article Sensitivity Analysis of the TRIGA IPR-R1 Reactor Models Using the MCNP Code International Nuclear Energy, Article ID 7, pages http://dx.doi.org/.1155/01/7 Research Article Sensitivity Analysis of the TRIGA IPR-R1 Reactor Models Using the MCNP Code C. A. M. Silva, 1 J. A. D. Salomé,

More information

Quality Assurance. Purity control. Polycrystalline Ingots

Quality Assurance. Purity control. Polycrystalline Ingots Quality Assurance Purity control Polycrystalline Ingots 1 Gamma Spectrometry Nuclide Identification Detection of Impurity Traces 1.1 Nuclides Notation: Atomic Mass Atomic Number Element Neutron Atomic

More information

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations WU Yongpeng ( ) 1,2, TANG Bin ( ) 2 1 Department of Engineering Physics,

More information

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT A Monte Carlo simulation has been developed using the Geant4 software

More information

FLUORINE IN WATER AND TEETH MEASURED WITH PIGE

FLUORINE IN WATER AND TEETH MEASURED WITH PIGE 2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 FLUORINE IN WATER AND TEETH

More information

Proposed model for PGAA at TRIGA IPR- R1 reactor of CDTN

Proposed model for PGAA at TRIGA IPR- R1 reactor of CDTN Proposed model for PGAA at TRIGA IPR- R1 reactor of CDTN AS LEAL, BT GUERRA, MABC Menezes, C Pereira Centre for Development of Nuclear Technology (CDTN), Brazilian Nuclear Energy Commission (CNEN), Av.

More information

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors 6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator

More information

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS 227 IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS Fusheng Li, Weijun Guo, and Robin P. Gardner Center for Engineering Applications of

More information

Identification of Naturally Occurring Radioactive Material in Sand

Identification of Naturally Occurring Radioactive Material in Sand Identification of Naturally Occurring Radioactive Material in Sand Michael Pope 2012 NSF/REU Program Physics Department, University of Notre Dame Advisors: Dr. Ed Stech, Dr. Michael Wiescher Abstract Radionuclides

More information

Gamma-Spectrum Generator

Gamma-Spectrum Generator 1st Advanced Training Course ITCM with NUCLEONICA, Karlsruhe, Germany, 22-24 April, 2009 1 Gamma-Spectrum Generator A.N. Berlizov ITU - Institute for Transuranium Elements Karlsruhe - Germany http://itu.jrc.ec.europa.eu/

More information

GAMMA RAY SPECTROSCOPY

GAMMA RAY SPECTROSCOPY GAMMA RAY SPECTROSCOPY Gamma Ray Spectroscopy 1 In this experiment you will use a sodium iodide (NaI) detector along with a multichannel analyzer (MCA) to measure gamma ray energies from energy level transitions

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS

ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS NUCLEAR PHYSICS ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS A. DUMITRESCU 1, GH. CǍTA-DANIL 1, 2 1 Physics Department, University Politehnica of Bucharest, Splaiul Independenţei

More information

anti-compton BGO detector

anti-compton BGO detector 1 2 3 Q β - measurements with a total absorption detector composed of through-hole HPGe detector and anti-compton BGO detector 4 5 Hiroaki Hayashi a,1, Michihiro Shibata b, Osamu Suematsu a, Yasuaki Kojima

More information

Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations

Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations 271 I. ORION 1,2 1 AND L. WIELOPOLSKI 2, 1 St. Luke=s/Roosevelt Hospital, Columbia University, New York, NY 10025,USA 2

More information

MCNP analysis and optimization of a double crystal phoswich detector

MCNP analysis and optimization of a double crystal phoswich detector EURONS I3 506065 JRA9 RHIB Report made during stay IEM-CSIC Madrid January 2008 MINISTERIO DE ASUNTOS EXTERIORES Y DE COOPERACIÓN AECI VICESECRETARÍA GENERAL MCNP analysis and optimization of a double

More information

Chapter 8 Hyper-Pure Germanium Detector

Chapter 8 Hyper-Pure Germanium Detector Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 8-1 8.1. Introduction Chapter 8 Hyper-Pure Germanium Detector Silicon semiconductor detectors described in the previous chapter have depletion

More information

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums Nuclear Science and Techniques 4 (13) 63 Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums QIAN Nan 1 WANG Dezhong 1,* WANG Chuan ZHU Yuelong MAUERHOFER Eric

More information

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method A Ruzzarin 1, P A L da Cruz 2, A L Ferreira Filho 2, A Iwahara 2 1 Laboratório de Instrumentação Nuclear/Programa

More information

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440 Jazan University College of Science Physics Department جاهعة جازان كلية العل وم قسن الفيزياء Lab Manual Nuclear Physics (2) 462 Phys 8 th Level Academic Year: 1439/1440 1 Contents No. Name of the Experiment

More information

Semiconductor Detector

Semiconductor Detector Semiconductor Detector General 1. Introduction A semiconductor detector is used for a. checking the radionuclidic purity of radiopharmaceuticals and calibration sources; b. the quantitative analysis of

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

The IAEA-CU world wide open proficiency test on the determination of radionuclides in soil, spinach and water

The IAEA-CU world wide open proficiency test on the determination of radionuclides in soil, spinach and water The IAEA-CU-2007-03 world wide open proficiency test on the determination of radionuclides in soil, spinach and water Laboratory s Final Report Laboratory Code: 146 (CuNo: 13949) Total Pages (with cover):

More information

Applied Radiation and Isotopes

Applied Radiation and Isotopes Applied Radiation and Isotopes 70 (2012) 2091 2096 Contents lists available at SciVerse ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Disintegration rate

More information

PROFILE S and C Series P-type Semi-planar and Coaxial HPGe Detectors

PROFILE S and C Series P-type Semi-planar and Coaxial HPGe Detectors P-type Semi-planar and Coaxial HPGe Detectors Exceptional Resolution and Stable, Low-energy Efficiency Comprehensive Solutions to Demanding Applications Exceptional resolution and stable, low energy efficiency

More information

Application of Optical Fiber in Radiation Detection Systems. Francisco Antônio Brandão Junior, Antonella Lombardi Costa* and Arno Heeren de Oliveira

Application of Optical Fiber in Radiation Detection Systems. Francisco Antônio Brandão Junior, Antonella Lombardi Costa* and Arno Heeren de Oliveira J. Chem. Chem. Eng. 10 (2016) 120-127 doi: 10.17265/1934-7375/2016.03.002 D DAVID PUBLISHING Application of Optical Fiber in Radiation Detection Systems Francisco Antônio Brandão Junior, Antonella Lombardi

More information

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University Notes to accompany Lab demonstrations by Barry Diacon, Technician, Department of

More information

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values. EXPERIMENT 2.1 GAMMA ENERGY CALIBRATION 1- Turn the power supply on to 900 V. Turn the NIM crate on to power the amplifiers. Turn the Oscilloscope on to check the gamma pulses. The main amplifier should

More information

TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY

TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY NUCLE A R FORENSIC S INTERN ATION A L TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY This document was designed and printed at Lawrence Livermore

More information

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry

More information

Airo International Research Journal October, 2015 Volume VI, ISSN:

Airo International Research Journal October, 2015 Volume VI, ISSN: 1 INTERACTION BETWEEN CHARGED PARTICLE AND MATTER Kamaljeet Singh NET Qualified Declaration of Author: I hereby declare that the content of this research paper has been truly made by me including the title

More information

EXPERIMENTAL DETERMINATION OF THE URANIUM ENRICHMENT RATIO

EXPERIMENTAL DETERMINATION OF THE URANIUM ENRICHMENT RATIO NUCLER PHYSICS EXPERIMENTL DETERMINTION OF THE URNIUM ENRICHMENT RTIO. LUC Horia Hulubei National Institute for Physics and Nuclear Engineering (IFIN-HH), 407 tomistilor Street, PO Box MG-6, Magurele,

More information

VERIFICATION OF MONTE CARLO CALCULATIONS OF THE NEUTRON FLUX IN THE CAROUSEL CHANNELS OF THE TRIGA MARK II REACTOR, LJUBLJANA

VERIFICATION OF MONTE CARLO CALCULATIONS OF THE NEUTRON FLUX IN THE CAROUSEL CHANNELS OF THE TRIGA MARK II REACTOR, LJUBLJANA International Conference Nuclear Energy for New Europe 2002 Kranjska Gora, Slovenia, September 9-12, 2002 www.drustvo-js.si/gora2002 VERIFATION OF MONTE CARLO CALCULATIONS OF THE NEUTRON FLUX IN THE CAROUSEL

More information

Alpha-Energies of different sources with Multi Channel Analyzer

Alpha-Energies of different sources with Multi Channel Analyzer Physical Structure of Matter Radioactivity Alpha-Energies of different sources with Multi Channel Analyzer What you can learn about Decay series Radioactive equilibrium Isotopic properties Decay energy

More information

Coincidence gamma-ray spectrometry

Coincidence gamma-ray spectrometry Downloaded from orbit.dtu.dk on: Dec 28, 2018 Coincidence gamma-ray spectrometry Markovic, Nikola; Roos, Per; Nielsen, Sven Poul Published in: Proceedings of the 11th Symposium of the Croatian

More information

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system

More information

arxiv: v2 [nucl-ex] 10 Jun 2009

arxiv: v2 [nucl-ex] 10 Jun 2009 Discrimination of gamma rays due to inelastic neutron scattering in AGATA A. Ataç,a, A. Kaşkaş a, S. Akkoyun a,, M. Şenyiğit a, T. Hüyük a, S. O. Kara a, J. Nyberg b a Department of Physics, Faculty of

More information

PHYS 3650L - Modern Physics Laboratory

PHYS 3650L - Modern Physics Laboratory PHYS 3650L - Modern Physics Laboratory Laboratory Advanced Sheet Photon Attenuation 1. Objectives. The objectives of this laboratory exercise are: a. To measure the mass attenuation coefficient at a gamma

More information

Compton Camera. Compton Camera

Compton Camera. Compton Camera Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract

More information

Validation of the MCNP computational model for neutron flux distribution with the neutron activation analysis measurement

Validation of the MCNP computational model for neutron flux distribution with the neutron activation analysis measurement Journal of Physics: Conference Series PAPER OPEN ACCESS Validation of the MCNP computational model for neutron flux distribution with the neutron activation analysis measurement To cite this article: K

More information

NGN PhD Studentship Proposal

NGN PhD Studentship Proposal NGN PhD Studentship Proposal Note that proposals will be assessed against both the quality of the scientific content and of the proposed training experience. Proposed supervisors (lead first) Dr Laura

More information