Determination of Short-term Exposure with a Direct Reading Photoionization Detector

Size: px
Start display at page:

Download "Determination of Short-term Exposure with a Direct Reading Photoionization Detector"

Transcription

1 Ann. occup. Hyg., Vol. 48, No. 1, pp , British Occupational Hygiene Society Published by Oxford University Press DOI: /annhyg/meg079 Determination of Short-term Exposure with a Direct Reading Photoionization Detector P. POIROT 1 *, I. SUBRA 1, F. GÉRARDIN 1, V. BAUDIN 2, S. GROSSMANN 3 and M. HÉRY 4 1 Département Ingénierie des Procédés; 2 Service Epidémiologie en Entreprise; 3 Département Polluant et Santé; 4 Direction Scientifique, INRS, BP 27, F Vandoeuvre Cedex, France Received 21 October 2002; in final form 12 June 2003 The use of a direct reading photoionization detector (PID) to determine short-term solvent exposures is described in the present paper. To assess the relevance of such a total exposure evaluation it was necessary to compare it with the real concentration of pollutants. This comparison was made by measuring in parallel with the PID determination the concentration of each pollutant using a standard technique, i.e. sampling on charcoal tubes and subsequent analysis by gas chromatography. Laboratory tests showed that the linearity of the answer of the PID is good for many compounds and for a mixture of these compounds. Similar tests were carried out for painters in workplaces with the same good correlations (determination coefficient r 2 close to 1) between the PID response and the real concentration of the pollutants measured on the sampling tubes. The use of PID also allowed determination of the exposure profile of the workers and comparison of the short-term exposure to the corresponding limit values. Many cases of the short-term limit values being exceeded were revealed by use of the PID, although very few cases of the long-term limit values have been found by the usual sampling (charcoal tube) and analytical (gas chromatography) methods. Keywords: photoionization detector; short-term exposure; solvent; limit values INTRODUCTION To protect workers from diseases such as irritation, chronic or irreversible tissue damage and narcosis, organizations responsible for occupational safety and health have laid down specific short-term exposure limit values. In the case of the American Conference of Governmental Industrial Hygienists (ACGIH), whose values are widely referred to throughout the world, the corresponding values are the TLV-STEL (threshold limit value short-term exposure limit), which should be checked over a period of 15 min, and the TLV-C (ceiling), for which instantaneous monitoring is recommended whenever possible (ACGIH, 2002). These short-term exposure limit values are the counterpart of the well-known long-term TLV-TWA (time-weighted average). Comparing the exposure of workers to these shortterm limit values is often difficult and time *Author to whom correspondence should be addressed. Tel: ; fax: ; pascal.poirot@inrs.fr consuming work for industrial hygienists (Kumagai and Matsunuga, 1995, 1999): determining the most appropriate monitoring time may be difficult and the occupational hygienist is often obliged to increase the number of sampling operations; the detection limits of the analytical methods are sometimes poor and in some cases a 15 min sampling may be too short for the conventional analytical techniques. However, the use of direct reading photoionization air detectors (PIDs) permits, under certain conditions, the evaluation of occupational exposure to organic solvent mixtures, taking into account possible short duration peak concentrations. A PID has recently been used to measure the benzene exposure of a truck driver during the loading of gasoline (Drummond, 1997), the TWA concentration being determined in parallel by sampling on a charcoal tube. Assuming that the response of the PID is linear, the author can estimate the work site conditions with a correlation of 75

2 76 P. Poirot et al. exposure and activity. Another study of short-term exposure to solvent mixture was performed on construction painters (Coy et al., 2000). The principle of simultaneous sampling on charcoal tubes and PID was also adopted: the data were highly correlated. As part of an epidemiological study devoted to the neurological effects of solvents among apprentice painters (Poirot et al., unpublished results), our laboratory was asked to assess the exposure of house painters. This assessment involved evaluating the mean exposure during a workshift as well as the exposure profile and took into account the variations in the exposure level over the workshift. Special emphasis was put on the exposure peaks. At the same time, a study of occupational hazards among workers employed in the reclamation of an industrial site led us to suspect the occurrence of short but high exposure to a broad range of organic solvents. In both cases the mixture of pollutants was complex and the quantity of each product sampled over a short period was small, making the analysis difficult and time consuming. The use of a direct reading PID to determine these short-term exposures is described in the present paper. MATERIALS AND METHODS As the measurement principle of the PID is based on the determination of all photoionizable gases and vapors, it does not distinguish between the individual pollutants present in the atmosphere. Consequently, the concentration is measured in the form of a composite index. To assess the relevance of such a total exposure evaluation it was necessary to compare it to the real concentration of pollutants. This comparison was made by measuring the concentration of each pollutant using a standard technique, i.e. sampling on charcoal tubes and subsequent analysis by gas chromatography, in parallel with the PID determination. Photoionization detector A direct reading PID (2020 PE Photovac) was used. The energy of the UV lamp was 10.6 ev. The equipment was calibrated with a 100 p.p.m. calibration gas (isobutylene). In use, the concentration is measured every second, but a 1 min averaging interval was set because the data logger can only store 1000 entries. In these conditions the monitor can operate for 16 h before the data logger is filled. After sampling the data were transferred to a personal computer. Sampling and analysis on charcoal tubes Two-section coconut charcoal tubes (SKC ) were used at a sampling rate of 0.1 l/min (Gilian LFS 113 sampling pumps). The analysis was performed by gas chromatography. In both cases (house painters and site remediation), the charcoal was desorbed in 1 ml of carbon disulfide with a low benzene content (CS 2, Aldrich ). The pollutants were different in the two experiments: petroleum spirit, i.e. a petroleum cut from C 8 to C 12 for the painters, and a mixture of industrial solvents for the site reclamation (chlorinated hydrocarbons, ketones, naphtenic hydrocarbons, etc.), but the chromatographic conditions were the same. A wide bore capillary column (60 m long, 0.75 mm i.d. Supelco SPB1 column) was used with flame ionization detection (FID). Analysing petroleum spirits is always difficult on account of the large number of components contained in the mixture. When it is not possible to obtain a reference sample of the petroleum spirit used, a qualitative analysis is carried out on a tube sampled in the work atmosphere to determine which petroleum spirit currently available in the laboratory is analytically closest to the sample to be determined. This petroleum spirit is then taken as the reference and used to plot the calibration curve. For the quantitative analysis, about a dozen peaks evenly distributed over the chromatogram are chosen amongst the major ones. It was checked that the results obtained by this method do not differ significantly from the results that would have been obtained with calibration with the petroleum spirit itself considered. Tests in the laboratory The proportionality of the response of the PID according to the concentration of the product tested was checked in the laboratory. These tests were carried out successively for two different petroleum spirits, p-xylene, m-xylene, styrene and ethylbenzene, in controlled atmospheres. They consisted of sampling simultaneously on a charcoal tube and by a PAM sampling inlet, both placed in the cell of generation benches where the spatial and temporal homogeneity had been tested previously (Gradiski et al., 1978; Castel et al., 1998). The sampling time ranged from 5 to 120 min according to the concentration of the product generated in the bench cell. Tests in workplace atmospheres The same type of comparison (PID versus charcoal tube) was carried out during the house painting and industrial site reclamation operations. The sampling devices were worn by volunteers for 3 8 h during their workshift. These comparisons were in fact complementary to the principal aim of the study which was, as mentioned previously, to determine the exposure profile of the workers and to compare the short-term exposure to the corresponding limit values. As few products have been assigned short-term exposure limits, such as a TLV-STEL or a TLV-C, the comparison was made to the excursion limits defined by the ACGIH:

3 Direct reading photoionization detector 77 Table 1. Limit values of solvents encountered during the study (expressed in mg/m 3 ) France: VME a Assigned by analogy with 1-methoxy-2-propanol. UK: Long-term exposure limit ACGIH: TLV- TWA France: VLE UK: Short-term ACGIH: TLVexposure limit STEL n-hexane Cyclohexane Ethanol Isopropanol Isobutanol n-butanol Acetone Methyl ethyl ketone Ethyl acetate n-butyl acetate Methoxy propanol acetate 1-Ethoxy-2-propanol 425 a Methylglycol acetate Ethylglycol Ethylglycol acetate Butylglycol Butylglycol acetate Butyldiglycol Dichloromethane Dichloroethane Trichloromethane Trichloroethane Trichloroethylene Perchloroethylene Benzene Toluene C petroleum C6 C12 hydrocarbons Stoddard solvent 525 Xylene Ethylbenzene C9 C12 aromatic hydrocarbons Excursions in worker exposure levels may exceed 3 times the TLV-TWA for no more than a total of 30 minutes during a work-day, and under no circumstances should they exceed 5 times the TLV-TWA, provided that the TLV- TWA is not exceeded. (ACGIH, 2002) For exposure of the house painters, the TLV-TWA of Stoddard solvent was taken into account, i.e. 525 mg/m 3 (100 p.p.m.). To determine the exposure of the house painters to other mixtures of solvents and that of the workers employed in industrial site reclamation, where the atmosphere was polluted by various solvents, the long-term threshold limit of the mixture was considered exceeded when the formula recommended by the ACGIH (I TWA = C 1 /TLV 1 + C 2 /TLV C n /TLV n, with C n the atmospheric concentration of component n and TLV n its TLV- TWA) exceeded 1. In the same way an I 3TWA index was considered to take into account the above mentioned ACGIH recommendation for excursion limits. When this index (I 3TWA = C 1 /3TLV 1 + C 2 /3TLV C n /3TLV n ) exceeds 1, the short-term threshold limit of the mixture is considered exceeded. The comparison to the short-term limit values established for solvents corresponds to a particular case in that, as can be seen in Table 1, most shortterm limits established by the ACGIH or the French Department of Employment for these compounds have not been established at 3 times the corresponding TLV-TWAs, but rather at levels ranging from 1.25 to 2 times this long-term value. A second index, I STEL, was then calculated from the ACGIH

4 78 P. Poirot et al. formula with TLV = TLV-STEL (or French VLE, roughly equivalent to the American TLV-STEL) or TLV = 1.5 TLV-TWA in the absence of a TLV- STEL. If the ACGIH and French Department of Employment lists are considered, this value of 1.5 is actually the most common ratio between a TLV- STEL and the corresponding TLV-TWA, and it is an intermediate value in the range generally encountered. This second index I STEL was then considered exceeded when its value was >1. RESULTS Tests in the laboratory For some compounds, the response factors (RF) (real concentration/pid response) of the PID were provided by the apparatus manufacturer, but for complex mixtures such as Stoddard solvent (whose composition varies depending on the origin) it was necessary to determine this RF. Experiments were carried out in the laboratory to check (or determine) this factor and to verify whether the response of the PID to a mixture was the sum of the responses to the different constituents of the mixture. The linearity of the response of the PID was tested between 0 and 100 p.p.m. for Stoddard solvent and between 0 and 500 p.p.m. for toluene, methyl ethyl ketone, ethylbenzene and for a 1/1/1 (w/w/w) mixture of these last three compounds. The results are summarized in Table 2. The linearity is good for all these compounds and mixtures and the RFs are close to the values provided by the apparatus manufacturer. The RF obtained for the toluene/methyl ethyl ketone/ ethylbenzene mixture in the experimental chamber was compared to the theoretical RF calculated from the RFs obtained for each of the components, taking into account their relative weights in the composition of the vapor (which could be calculated from their relative molecular weights as the entire mixture was vaporized). The two results are similar. With y the real concentration (p.p.m.) and x the PID response (p.p.m.), the equations are: y 1 = 0.65x for the measurement of the three-component mixture in the experimental chamber; y 2 = 0.68x 0.05 for the theoretical value established from the RFs of the PID to each of the components tested separately in the experimental chamber. This corresponds to values of y 1 = 68 p.p.m. and y 2 = 69 p.p.m. for x = 100 p.p.m., y 1 = 136 p.p.m. and y 2 = p.p.m. for x = 200 p.p.m. and y 1 = 340 p.p.m. and y 2 = p.p.m. for x = 500 p.p.m. From these results it can be considered that if the relative composition of an atmospheric mixture remains constant, the PID can be used to watch the concentration development over time. Tests in workplace atmospheres Similar tests were carried out for painters in workplaces with the same good correlation (determination coefficient r 2 close to 1) between the PID response and the real concentration of the pollutants measured on the sampling tubes. These results are summarized in Table 3. The correlation was established between Table 2. Comparative response factors (RF) determined by the PID manufacturer and from measurements in an INRS laboratory Compound RF according to manufacturer RF calculated from measurements in the laboratory (determination coefficient r 2 ) Petroleum spirit (A) No data y = 0.61x (r 2 = 0.99) Toluene (B) y = 0.5x y = 0.53x (r 2 = 0.99) Methyl ethyl ketone (C) y = 0.8x y = 0.91x 3.90 (r 2 = 0.99) Ethylbenzene (D) y = 0.5x y = 0.51x (r 2 = 0.99) 1/1/1 (w/w/w) B/C/D mixture No data y = 0.65x (r 2 = 1) x = response of PID (p.p.m.); y = concentration of pollutant (p.p.m.). Table 3. Correlation between pollutant concentration measured in workplace atmospheres and response of the PID Compound No. of values Pollutant concentration = f (response of PID) Exposure index = f (response of PID) Petroleum spirit (E) 15 y = 0.58x Petroleum spirit (F) 4 y = 0.67x Polyurethane paint (1) 8 z = x Epoxy paint (2) 11 z = x Determination coefficient (r 2 ) x = response of PID (p.p.m.); y = pollutant concentration (p.p.m.); z = index of exposure (C 1 /LV 1 + C 2 /LV C n /LV n ); 1, the solvents in the polyurethane paint are methyl ethyl ketone, toluene, methoxy propanol acetate, xylene, ethylbenzene and C 9 aromatic hydrocarbons; 2, the solvents in the epoxy paint are isopropanol, methyl ethyl ketone, toluene, n-butyl acetate, xylene and ethyl benzene.

5 Direct reading photoionization detector 79 Fig. 1. Exposure of a painter during polyurethane floor painting with a roller. Fig. 2. Exposure of a painter during epoxy floor painting with a roller. the PID response and either the Stoddard solvent concentration or the exposure expressed in the form of an I TWA, as defined previously for mixtures (equivalent correlations would have been obtained with the exposure expressed in the form of an I 3TWA or an I STEL ). The corresponding components of the paints are defined in Table 3. Two examples of the exposure profile of the painters are given in Figures 1 and 2. The case described in Fig. 1 involves the painting of a floor with a polyurethane paint exposing the worker to methyl ethyl ketone, toluene, methoxy propanol acetate, xylene, ethylbenzene and C 9 aromatic hydrocarbons. Figure 2 describes the exposure of a worker using an epoxy paint on a floor (isopropanol, methyl ethyl ketone, toluene, n-butyl acetate, xylene and ethylbenzene). In both cases the PID values were recorded on a 1 min integration basis. Although the comparison with the TLV-TWA does not highlight any excess over the limit value expressed in the form of an I TWA, the use of the PID shows several intervals during which the I STEL is exceeded. On the other hand, if the TLV-STELs (or their French equivalents the VLEs) are given up on behalf of the I 3TWA level, no excursion over this reference short-term index over 1 is registered. As part of the epidemiological study mentioned previously (Poirot et al., unpublished results), a total of 125 personal samples on charcoal tubes were taken to assess the mean exposure of the painters during a workshift. The exposure profile was also recorded in parallel for 43 of these 125 samples in parallel by means of the PID, i.e. the probe of the PID and the inlet of the charcoal tube were situated in the imme-

6 80 P. Poirot et al. Table 4. Respective exceedings of exposure indices established from short-term and long-term limit values Activities exposing to Charcoal tube measurements No. of Comparison to I TWA samples No. of samples Maximum time >1 averaged value PID measurements No. of Comparison to I STEL samples No. of samples with instantaneous index value >1 Maximum instantaneous value Comparison to I 3TWA No. of Maximum samples with instantaneous instantaneous value index value >1 Petroleum spirit paints and varnishes Indoor emulsion paints Indoor primers Indoor polyurethane floor paints Indoor epoxy floor paints Neoprene glue Floor vitrification diate vicinity and the measurements performed over the same period. For only 7 of the 11 painting activities studied was at least one sample taken with a PID. The comparative occurrences of excesses over the limit value, depending on whether the long-term limit or the short-term limit is considered, are shown in Table 4. Only the 7 activities for which both charcoal tube and PID measurements were taken are considered in this table. Although exceeding the I TWA index is rare, 15 of the 43 samples taken with the PID reveal at least one instantaneous (in fact determined for a 1 min integration time) value of the I STEL > 1. This number decreases drastically if the limit is set at I 3TWA. It is worth noting in the case of the painters that the exposure is due to a defined product with a relatively constant composition (even if it is likely to vary slightly over time because of differences in the vapor pressures of the different components of the paint). For the polluted soil reclamation work the qualitative and quantitative variations in the atmospheric composition are higher because the pollution is not homogeneous over the entire area treated and the correlation is therefore poorer (r 2 = 0.87). To show this variability, the data from which this correlation was calculated are given in Table 5: they (the samples on charcoal tubes and the concentration recordings with the PID) were collected in the immediate vicinity of a sifting machine through which materials polluted by organic solvents were riddled. Although the composition of the atmosphere is likely to vary significantly from one sample to another for the same pollutant, reflecting the differences in the composition of the soil pollution, the number of products in the mixture is high enough to minimize the influence of this variability. In fact, the most significant contribution to the exposure index is due to trichloromethane, the limit value of which is low enough (25 mg/m 3 ) to make the atmospheric concentration of this compound a major influence. An example of the concentration profile recorded with the PID near the sifting machine is shown in Fig. 3. The variations in this particular profile are also representative of the five other recordings used to determine the correlation between the results of the charcoal samplings and the results of the PID (Table 5). Simultaneously, personal samples were taken on charcoal tubes on workers whose work was to avoid blocking of the sifting machine, which led them to intervene in the middle of the most polluted atmosphere, over the sifter. The exposure of these workers is summarized in Table 6. The pollutants detected are the same as in the area samples taken in parallel with the PID measurements, and the contribution of trichloromethane to the exposure index is also the most important. An example of such a profile is given in Fig. 3. DISCUSSION The choice of the reference limit value is especially important in the case of lack of short-term limit values. Depending on whether the equivalent of the TLV-STEL is fixed at 3 (in conformity with the official philosophy of the ACGIH) or 1.5 (according to its practice in the case of solvents) times the TLV- TWA, the number of excursions above the limit value will vary greatly. In the two cases of painters described in Figs 1 and 2, only methoxy propanol acetate and C 9 hydrocarbons were not assigned a short-term limit value (TLV-STEL or VLE). Their contribution to the pollution index being limited

7 Table 5. Atmospheric concentrations measured on the site remediation from which the correlation between PID response and charcoal tube sample was established Sample Concentration (p.p.m.) Methyl Trichloroethanethanethylenbutylketonethylenbenzentoluenbenzene Isobutanol Dichloro- Trichloro- Benzene Trichloro- Methyliso- Toluene Perchloro- Ethyl- Xylene Ethyl- Trimethyl- Exposure ethyl ketonemethane index <dl 0.2 <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl 0.8 <dl 0.1 <dl <dl Direct reading photoionization detector 81

8 Table 6. Exposure of workers in a polluted setting measured by sampling on a charcoal tube Sample Concentration (p.p.m.) Methyl ethyl ketone Trichloromethane Benzene Isobutanol Dichloroethane Trichloroethane Trichloroethylene Methylisobutylketone Toluene Xylene Perchloroethylene Ethylbenzene Ethyltoluene Trimethylbenzene <dl <dl <dl <dl <dl <dl <dl <dl <dl 0.1 <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl <dl 1.4 <dl 0.1 <dl Exposure index 82 P. Poirot et al. dl, detection limit.

9 Direct reading photoionization detector 83 Fig. 3. Concentration profile near the screen during remediation of the polluted site. (~10% of the total index for each), neglecting the possibility of using the I STEL and using only the I 3TWA comparison level would constitute, in our opinion, a very poor use of the limit value system. In these cases (methoxy propanol acetate and C 9 hydrocarbons), we preferred determining an arbitrary short-term value at 1.5 times the long-term value, which represents the most common ratio between the corresponding shortterm and long-term limit values for the compounds present in these workplace atmospheres. Whatever the answer to this debate of choice of reference limit value, the use of the PID can be helpful in the appreciation of exposure at the workplace. Considered in aggregate with exposure index values ranging from 0.05 to 0.83 (Table 5), the exposure of workers involved in the remediation of the polluted area examined in this paper could be considered as lower than the exposure limit values. However, examination of the concentration profiles recorded on charcoal tubes during work near the sifting machine revealed peaks 8 10 times higher than the mean value measured during the shift (Fig. 3). As these concentration profiles are representative of the exposure of workers intervening directly in the sifting machine, i.e. in the most polluted zone, it can be suspected that the short-term exposure of these workers may have exceeded the short-term limit values defined previously (I STEL or I 3TWA ). In reality, with personal exposure index values ranging from ~0.44 to 0.83 measured for the workers in charge of preventing blockage of the sifter, exposure peaks reaching levels 3.5 (0.44 8) to 8 ( ) times the I TWA are likely to have occurred. In terms of occupational health and safety, such information is important to ensure protection of the workers, either by installing a ventilation system on the sifter or by issuing the workers with a respirator when they have to intervene in this sifting machine. The limits of the method must nevertheless be kept in mind. For example, although trichloromethane cannot be detected by the PID as the energy of the lamp used (10.6 ev) is too low, its contribution to the exposure index is important. Insofar as it can be assumed that the concentration of trichloromethane varies in the same way as other detectable vapors, this non-detection is not a major drawback, provided that the relative concentrations of trichloromethane and detectable compounds are not too unbalanced in favor of trichloromethane. Another requirement for the relevance of establishing a correlation between index of exposure (in the case of a mixture of solvents) and PID response is that the relative concentrations of the different compounds of the mixture must not vary significantly during the measurement. From the results obtained in this study (in particular from the results of Table 3, where the petroleum spirits are themselves a mixture of various compounds with different vapor pressures), it looks likely that with determination coefficients close to 1, a good estimation of the actual exposure can be performed with the PID. The duration of the sampling (ranging from 3 to 8 h) probably minimized the influence of the differences in the vapor pressures of the various compounds, since the time was long enough to allow a stable atmosphere to settle and the activity of the workers did not change throughout the day. Whatever the case, there is an obvious contradiction in deducing an instantaneous value (PID measurement) from a time-weighted average one (charcoal tube measurement): given the measurement uncertainties (location of the sampling probe, analytical uncertainties, temperature differences from one day to the next, etc.), we consider that the information provided by the PID must mainly be considered as an indication which should be confirmed by other measurements (possibly done with the PID). The uncertainty on short-term measurements performed by other methods, such as charcoal tubes for example, is also great and they are not

10 84 P. Poirot et al. always so easy to use (determination of the pertinent moment to sample for example). CONCLUSION This paper shows the advantages and limitations of using a PID to assess the exposure of workers over short periods. Although the composition of the working atmosphere was constant enough (qualitatively and quantitatively) to ensure the relevance of the PID results for the two activities considered in this study, this property should be verified in all cases. However, because of the non-specificity of the response of the apparatus for mixtures, a number of measurements by a specific method (e.g. charcoal tubes with chromatographic analysis) should be carried out in parallel with the PID measurements to determine the relative concentration of the components of the mixture. This standardization should be completed at the beginning of the experiment and checked at regular intervals during the period when the PID is used. Under these conditions the PID is an excellent tool to show possible developments in the exposure of a worker. At the present stage of our research, it is not yet possible to determine the optimal conditions under which the technique developed in the present paper will show the best quality cost ratio. Given the rather good correlation coefficients obtained in various situations, it looks likely that four or five pairs of samples (PID + charcoal tubes) are enough provided that they are taken at concentrations ranging across the exposures encountered in the studied tasks and that the relative concentrations of the different compounds in the workplace atmosphere are relatively stable. Using the PID to compare short-term limit values is also a time consuming and difficult task, as the examples described in this paper show, but this difficulty is not directly related to the principle or the performance of the apparatus. Independently of the specificity problems mentioned above, this study has revealed contradictions between the TLV-STEL system and the excursion limit system, at least for pollutants such as solvents for which the TLV-STEL (when this exists) is generally established at 1.5 times the corresponding TLV-TWA. REFERENCES ACGIH. (2002) Threshold limit values for chemical substances and physical agents and biological exposure indices. Cincinnati, OH: American Conference of Governmental Industrial Hygienists. Castel B, Lefevre C, Lhuillier F, Delcourt, J. Sandino JP. (1998) Détermination du benzène par échantillonnage passif: essais interlaboratoires. Cah Notes Documentaires; 171: Coy JD, Bigelow PL, Buchan RM, Tessari JD, Parnell JO. (2000) Field evalaution of a portable photionization detector for assesing exposure to solvent mixtures. Am Ind Hyg Assoc J; 61: Drummond I. (1997) On-the-fly calibration of direct reading photoionization detectors. Am Ind Hyg Assoc J; 58: Gradiski D, Bonnet P, Raoult G, Magadur JL. (1978) Toxicité aiguë comparée par inhalation des principaux solvants aliphatiques chlorés. Arch Mal Prof; 39: Kumagai S, Matsunuga I. (1995) Changes in the distribution of short-term exposure concentration with different averaging times. Am Ind Hyg Assoc J; 56: Kumagai S, Matsunuga I. (1999) Within-shift variability of short-term exposure to organic solvent in indoor workplace. Am Ind Hyg Assoc J; 60:

Questions, Myths and Misconceptions about Using Photoionization Detectors

Questions, Myths and Misconceptions about Using Photoionization Detectors Questions, Myths and Misconceptions about Using Photoionization Detectors Solvent, fuel and other VOC vapours are pervasively common in many workplace environments. Increased awareness of the toxicity

More information

Short-Period Air Sampling by Direct-reading Instruments

Short-Period Air Sampling by Direct-reading Instruments Short-Period Air Sampling by Direct-reading Instruments OHAO Meeting, Toronto, October 13, 2011 Ian Drummond and Paul Bozek, Dalla Lana School of Public Health, University of Toronto Program for today's

More information

HYDROCARBONS, AROMATIC 1501

HYDROCARBONS, AROMATIC 1501 HYDROCARBONS, AROMATIC 1501 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1501, Issue 3 EVALUATION: Full Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table 2

More information

IV. DIRECT MEASUREMENT LABS. Introduction

IV. DIRECT MEASUREMENT LABS. Introduction IV. DIRECT MEASUREMENT LABS VI-1 I. Purpose Introduction An industrial hygienist frequently needs to evaluate airborne gas and vapor concentrations "on the spot". In other cases, he needs to track down

More information

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed

More information

Dose-Response Analysis in Animal

Dose-Response Analysis in Animal Environmental Health Perspectives Vol. 42, pp. 9-13, 1981 Dose-Response Analysis in Animal Studies: Prediction of Human Responses by Yves Alarie* An animal bioassay has been used to evaluate a series of

More information

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs Vapor Intrusion Sampling Options: Performance Data for s, Badges, and Sorbent Tubes for VOCs Linda S. Coyne SKC Inc., 863 Valley View Road, Eighty Four, PA 1533 George Havalias, Maria C. Echarte American

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Evaluation of Photoionization Detector Performance for the Measure of Toluene Vapors

Evaluation of Photoionization Detector Performance for the Measure of Toluene Vapors Chemistry Journal Vol. 1, No. 3, 2015, pp. 116-123 http://www.aiscience.org/journal/cj Evaluation of Photoionization Detector Performance for the Measure of Toluene Vapors Azimi Pirsaraei Seyed Reza 1,

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

GB Translated English of Chinese Standard: GB NATIONAL STANDARD Translated English of Chinese Standard: GB18581-2009 www.chinesestandard.net Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA ICS 87.040 G 51 GB 18581-2009 Replacing GB

More information

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No Validation of Tetrahydrofuran Using SKC Passive Sampler Cat. No. 575-002 Page 1 of 10 Publication 1841 Issue 1407 Research Report Validation of Tetrahydrofuran Using the SKC Cat. No. 575-002 Passive Sampler

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC Aysel T. ATIMTAY Middle East Technical University, Environmental Engineering Department, 06531 Ankara Turkey (aatimtay@metu.edu.tr)

More information

TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks

TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks Robert Truesdale, RTI International Heidi Hayes, Eurofins Air Toxics Ian Hers, Golder Associates 1 RTI International is a trade name

More information

Introduction to IH Analytical Chemistry

Introduction to IH Analytical Chemistry Introduction to IH Analytical Chemistry Sampling and Analysis Manning, Eide, Van Etten (Apr 2013) Foreword Thank you for purchasing the Introduction to IH Analytical Chemistry self-study workbook. Course

More information

Technical/Application Article 01 Version 1.2

Technical/Application Article 01 Version 1.2 Technical/Application Article 01 Version 1.2 th 25 January 2017 WRH What is a PID? Introduction PID is the abbreviation for Photo-Ionization Detector. A PID is a hand-held, personal, or fixed wallmounted

More information

Training Coarse ECOPROBE 5 CHEMISTRY RS DYNAMICS

Training Coarse ECOPROBE 5 CHEMISTRY RS DYNAMICS Training Coarse ECOPROBE 5 CHEMISTRY Contents Ecoprobe analytical principles Photoionization Infrared spectroscopy Oxygen measurement Pollutants Types and responses Head space New application of Ecoprobe

More information

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the

More information

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003 ALCOHOLS IV 1403 (1) HOCH 2 CH 2 OCH 3 MW: (1) 76.09 CAS: (1) 109-86-4 RTECS: (1) KL5775000 (2) HOCH 2 CH 2 OCH 2 CH 3 (2) 90.12 (2) 110-80-5 (2) KK8050000 (3) HOCH 2 CH 2 O(CH 2 ) 3 CH 3 (3) 118.17 (3)

More information

Equilibrated Vapor Concentrations for Bicomponent Organic Solvents

Equilibrated Vapor Concentrations for Bicomponent Organic Solvents J Occup Health 1998; 40: 13 136 Journal of Occupational Health Equilibrated Vapor Concentrations for Bicomponent Organic Solvents Hajime HORI 1 and Isamu TANAKA 1 Department of Environmental Management

More information

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method

More information

METHYLTIN CHLORIDES 5526

METHYLTIN CHLORIDES 5526 METHYLTIN CHLORIDES 5526 Formula: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 5526, Issue 1 EVALUATION: FULL Issue 1: 15 March 2003 OSHA : See Table 1 NIOSH: See Table 1 ACGIH: See Table 1

More information

Research Report. Validation of Methyl Methacrylate (MMA) using SKC Passive Sampler Abstract

Research Report. Validation of Methyl Methacrylate (MMA) using SKC Passive Sampler Abstract Research Report Validation of using SKC Passive Sampler 575-002 Abstract A sampling method for Methyl Methacrylate in air has been validated for concentration levels from 10 to 200 ppm and for exposure

More information

HAZARD COMMUNICATION SAFETY PROGRAM

HAZARD COMMUNICATION SAFETY PROGRAM Hazard Communication Program HAZARD COMMUNICATION SAFETY PROGRAM 1. Purpose. To provide an effective, written hazard communication program in compliance with company, State and Federal regulatory requirements.

More information

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR: SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD 21044 PREPARED FOR: HOWARD COUNTY PUBLIC SCHOOL SYSTEM 10910 ROUTE 108 ELLICOTT

More information

MEASURING NANOPARTICLE EXPOSURE

MEASURING NANOPARTICLE EXPOSURE MEASURING NANOPARTICLE EXPOSURE APPLICATION NOTE NSAM-001 The Nanoparticle Surface Area Monitor reports the surface area of inhaled particles deposited in the lung Nanoparticle Exposure There is increasing

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

Prima PRO Process Mass Spectrometer

Prima PRO Process Mass Spectrometer APPLICATION NOTE Prima PRO Process Mass Spectrometer No. xxxx Controlling emissions by multi-component analysis of Volatile Organic Compounds (VOC) Authors: Peter Traynor, Thermo Fisher Scientific, Franklin,

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Prelab Reading Assignment: Laboratory Techniques in Organic Chemistry, 4 th Ed. Chapter 19

Prelab Reading Assignment: Laboratory Techniques in Organic Chemistry, 4 th Ed. Chapter 19 CHEM 213 Technique Experiments Experiment 5: Column Chromatography Number of labs - one Reactions performed None Chemicals used: Fluorene-fluorenone mixture, hexanes, methylene chloride, silica gel Supplies

More information

Media blank result for a single-section sampler (e.g., membrane filter). Media blank result for back section of a sampler.

Media blank result for a single-section sampler (e.g., membrane filter). Media blank result for back section of a sampler. IV. GLOSSARY OF ABBREVIATIONS, DEFINITIONS, AND SYMBOLS AAS Acceptable range (biological) Accuracy ACGIH Ashing ASV B B b B f Bias Bioaerosol Biological monitoring Blank Atomic absorption spectrophotometry.

More information

Date: March 6, 2008 RWDI Reference #: W B Pages (Including Cover):

Date: March 6, 2008 RWDI Reference #: W B Pages (Including Cover): RWDI AIR Inc. 650 W oodlawn R oa d W est, Guelph, Ontari o, Ca nad a N1K 1B8 Tel: (51 9) 823-1 31 1 E-mail: info@rwdi.com Fax: (51 9) 823-1 31 6 W eb: www. r wdi.com Date: March 6, 2008 RWDI Reference

More information

Breakthrough Times for Vapors of Organic Solvents with Low Boiling Points in Steady-state and Pulsating Flows on Respirator Cartridges

Breakthrough Times for Vapors of Organic Solvents with Low Boiling Points in Steady-state and Pulsating Flows on Respirator Cartridges Industrial Health, 1996, 34, 125-131 125 Breakthrough Times for Vapors of Organic Solvents with Low Boiling Points in Steady-state and Pulsating Flows on Respirator Cartridges Shigeru TANAKA1, Miki HANEDA2~,

More information

Volatile Organic Compounds (VOC) Probes

Volatile Organic Compounds (VOC) Probes Volatile Organic Compounds (VOC) Probes New Plug and Play accessory probes Compatible with the new 9565 VelociCalc and 7575 Qtrak Also compatible with the 9555 VelociCalc and 7565 Qtrak Firmware must be

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

Solid Phase Micro Extraction of Flavor Compounds in Beer

Solid Phase Micro Extraction of Flavor Compounds in Beer Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental

More information

Compliance with occupational limit values by applying workplace measurements

Compliance with occupational limit values by applying workplace measurements Occupational Limit Values for Hazardous Substances - Healthy working conditions in a global economy Conference under the German Presidency of the European Council, Dortmund, Germany, 7-8 May 2007 Compliance

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

HAZARD COMMUNICATION PROGRAM

HAZARD COMMUNICATION PROGRAM HAZARD COMMUNICATION PROGRAM UNIVERSITY RISK MANAGEMENT Occupational Safety and Health Programs 19 Hagood Avenue, Suite 908 Charleston SC 29425 843-792-3604 Revised: January, 2015 TABLE OF CONTENTS Safety

More information

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory. VINYL CHLORIDE Method no.: 04 Matrix: Air Target concentration: 3 1 ppm (2.5 mg/m ) (OSHA PEL) Procedure: Collection on charcoal (two-tubes in-series), desorption with carbon disulfide, analysis by gas

More information

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority: Title: Chemical Hygiene Written Program Effective Date: November 2005 Revision Date: February 8, 2017 Issuing Authority: Responsible Officer: VP, Facilities and Construction Management Director Environmental

More information

Method for the determination of 1,3-butadiene

Method for the determination of 1,3-butadiene Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY 1. SCOPE AND APPLICATION: METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) Date 02/24/96 Page 1 1.1. This method is designed to measure the concentration

More information

Latest Amendments: Ministry of Health, Labour and Welfare Notification No. 191 of April 27, 2001 In Japanese

Latest Amendments: Ministry of Health, Labour and Welfare Notification No. 191 of April 27, 2001 In Japanese Working Environment Measurement Standards Ministry of Labour Notification No. 46 of April 22, 1976 Latest Amendments: Ministry of Health, Labour and Welfare Notification No. 191 of April 27, 2001 In Japanese

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

PRO LIGNO Vol. 11 N pp

PRO LIGNO Vol. 11 N pp INFLUENCE OF CLASSIC FINISHED SURFACES OF MASSIVE WOOD ON INDOOR ENVIRONMENT Daniela TESAŘOVÁ Mendel University in Brno, Faculty of Forestry and Wood Technology Str. Zemědělská nr. 1, 61300 Brno, Czech

More information

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES J Kouzaki 1*, S Sato 1, S Nakai 1, Y Shirasuna 2, K Hirano 2 1 Graduate School of Environmental and Information

More information

BEST PRACTICE GUIDELINES

BEST PRACTICE GUIDELINES In collaboration with : iw BEST PRACTICE GUIDELINES Safe use of gloves for the handling of solvents Safe use of gloves for the handling of solvents September 2018 Foreword The European Solvents Industry

More information

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website: R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J. 07852 Website: www.reactnj.org Email: info@reactnj.org RE: Air Test Results R.E.A.C.T. has facilitated an ambient air test

More information

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15

More information

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography Fast Analysis of Aromatic Solvent with.8 mm ID GC column Application Gas Chromatography Authors Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Ying Lun Road Waigaoqiao Free Trade Zone Shanghai 3 P.R.

More information

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction Purge and Trap (P&T) concentrators

More information

Methods for the determination of vinyl chloride

Methods for the determination of vinyl chloride Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

DETECTOR TUBES AND PUMPS

DETECTOR TUBES AND PUMPS DETECTOR TUBES AND PUMPS 08-00-02 FEATURES Quick and inexpensive to use. A reliable method of testing more than 170 hazardous gases and vapors. Kwik-Draw Deluxe and Gas Tester II pumps offer automatic

More information

Strategies For Selecting Air Sampling Methods AIHA-Florida Section Spring Conference

Strategies For Selecting Air Sampling Methods AIHA-Florida Section Spring Conference Strategies For Selecting Air Sampling Methods 2011 AIHA-Florida Section Spring Conference Course Outline What are we going to sample for Sampling media Active sampling, diffusive samplers, direct read

More information

Real-Time Detection: From Gisclard et al.: A Simple Device for Air Analysis. AIHA Quarterly, 14(1):23-25 (1953)

Real-Time Detection: From Gisclard et al.: A Simple Device for Air Analysis. AIHA Quarterly, 14(1):23-25 (1953) Real-Time Detection: 1953 From Gisclard et al.: A Simple Device for Air Analysis. AIHA Quarterly, 14(1):23-25 (1953) Sampling Gases and Vapors Gas: A state of matter characterized by very low density and

More information

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 8260 in conjunction

More information

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract. Application Note 8 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-5 Application Note Environmental, Canister, TO-5, Air Monitoring, Air Toxics Abstract This application

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

EPA TO-17 Volatile Organic Compounds

EPA TO-17 Volatile Organic Compounds EPA TO-17 Volatile Organic Compounds Method TO-17 is used to analyze samples for volatile organic compounds collected on multi-bed sorbent tubes, which are thermally desorbed and cryo-focused on the capillary

More information

IMPCA METHANOL REFERENCE SPECIFICATIONS

IMPCA METHANOL REFERENCE SPECIFICATIONS IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION IMPCA METHANOL REFERENCE SPECIFICATIONS a.i.s.b.l. International Methanol Producers & Consumers Association i.v.z.w. Avenue de Tervueren 270

More information

Gloves: choice and use

Gloves: choice and use Gloves: choice and use Adopted April 15 th, 2003 Comité sectoriel de sécurité du Département de chimie Université Laval Wearing gloves: why? Gloves are worn for protection. However, gloves do not always

More information

VOC TEST REPORT AgBB

VOC TEST REPORT AgBB Bostik SA 253 rue du Président Wilson 93211 La Plaine St Denis Cedex FRANCE Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT

More information

Hazard Communication Policy

Hazard Communication Policy Table of Contents I. Purpose... 2 II. Scope... 2 III. Policy... 2 III.A. Responsibilities... 2 III.A.1. Management... 2 III.A.2. Environmental Health and Safety Department (EH&S)... 3 III.A.3. Supervisors...

More information

Rocky Mountains Cooperative Ecosystem Studies Unit (RM-CESU) RM-CESU Cooperative Agreement Number: J

Rocky Mountains Cooperative Ecosystem Studies Unit (RM-CESU) RM-CESU Cooperative Agreement Number: J Rocky Mountains Cooperative Ecosystem Studies Unit (RM-CESU) RM-CESU Cooperative Agreement Number: J1580050167 Yellowstone Winter Use Personal Exposure Monitoring Prepared For: National Park Service Report

More information

HAZARD COMMUNICATION and GHS. Environmental Health and Safety

HAZARD COMMUNICATION and GHS. Environmental Health and Safety HAZARD COMMUNICATION and GHS Environmental Health and Safety RIGHT-TO-KNOW LAW OSHA created the Hazard Communication Standard in 1983. Applies to all places of employment where employees are exposed to

More information

VOC TEST REPORT CDPH

VOC TEST REPORT CDPH Euro Trade Flooring SL. Pol. Ind. Can Estella c/galileo no. 11 08635 Sant Esteve Sesrovires BARCELONA SPAIN Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing

More information

High Purity Chromasolv Solvents

High Purity Chromasolv Solvents Research Chemicals High Purity Chromasolv Solvents Part of Honeywell Research Chemical s Exclusive Riedel-de Haën Brand Portfolio The Chromasolv family of solvents are used in a variety of analytical applications

More information

"Employee Right To Know" Environmental, Health, and Safety Department

Employee Right To Know Environmental, Health, and Safety Department EASTERN ILLINOIS UNIVERSITY, Charleston, Illinois MODEL WRITTEN HAZARD COMMUNICATION PROGRAM "Employee Right To Know" Environmental, Health, and Safety Department SELF HELP SOURCES FOR MSDS INFORMATION:

More information

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline Fast : Cycles in

More information

ORGANOPHOSPHORUS PESTICIDES 5600

ORGANOPHOSPHORUS PESTICIDES 5600 ORGANOPHOSPHORUS PESTICIDES 5600 Formula: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 5600, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : Table 2 NIOSH: Table 2 ACGIH: Table 2 PROPERTIES:

More information

Method for the determination of dimethyl sulfate

Method for the determination of dimethyl sulfate German Social Accident Insurance Deutsche Gesetzliche Unfallversicherung Analytical Subcommittee of the Chemistry Board of Experts* Carcinogenic substances Order number: BGI 505-7-05 Established methods:

More information

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA 94804-3614 Ph. 510-236-2325; Fax 510-236-2335 E-mail info@berkeleyanalytical.com VOC Emissions from Building Products Customer & Building

More information

Industrial Hygiene Report

Industrial Hygiene Report Industrial Hygiene Report Control of Hydrochloric Acid Vapors in a Lab Setting September, 24 2012 Sentry Air Systems 6999 W. Little York Rd, Suite P1 Houston, TX 77040 BACKGROUND: Hydrochloric acid, an

More information

Target Compound Results Summary

Target Compound Results Summary Phone/ (86)88-4800 / (86)88-471 http://www.emsl.com to1lab@emsl.com Customer ID: EMSL0A K12312.D 28 cc 0 0/01/2017 K12327.D 21 cc 600 Target Compound Results Summary Propylene 11-07-1 42.08 ND.0 ND 8.6

More information

LATEST TECHNOLOGY IN Safe handling & Recovery OF Solvents in Pharma Industry

LATEST TECHNOLOGY IN Safe handling & Recovery OF Solvents in Pharma Industry LATEST TECHNOLOGY IN Safe handling & Recovery OF Solvents in Pharma Industry TYPICAL SOLVENT USE IN Pharma Industry Usage of solvents in an API process development is for: Diluent to carry out reaction

More information

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA 94804-3614 Ph. 510-236-2325; Fax 510-236-2335 E-mail info@berkeleyanalytical.com VOC Emissions from Building Products Customer & Building

More information

#03 Laboratory Ventilation and Risk Assessment Exercise John Cullen, Monroe Community College, Rochester, NY 14623

#03 Laboratory Ventilation and Risk Assessment Exercise John Cullen, Monroe Community College, Rochester, NY 14623 #0 Laboratory Ventilation and Risk Assessment Exercise John Cullen, Monroe Community College, Rochester, NY 1462 INTRODUCTION This exercise is designed to introduce students to the analysis of health risks

More information

Sampling for Gases and Vapors

Sampling for Gases and Vapors Sampling for Gases and Vapors EOH 466B Evaluating the Occupational Environment Spring 2008 Gases and Vapors Gases: don't exist in liquid state at STP e.g.: nitrous oxides, ozone, carbon monoxide Vapors:

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method

More information

Globally Harmonized System of Classification and Labeling of Chemicals (GHS) LC-1033p Rev. 06/16

Globally Harmonized System of Classification and Labeling of Chemicals (GHS) LC-1033p Rev. 06/16 Globally Harmonized System of Classification and Labeling of Chemicals (GHS) Rev. 06/16 Agenda Introduction to Globally Harmonized System of Classifying and Labeling Chemicals Hazard & Precautionary Statements

More information

Hazard Communication 29 CFR Right-To-Know 12 NYCRR Part 820. NYS PESH Training and Education

Hazard Communication 29 CFR Right-To-Know 12 NYCRR Part 820. NYS PESH Training and Education Hazard Communication 29 CFR 1910.1200 Right-To-Know 12 NYCRR Part 820 1 Purpose To protect employees from chemical hazards in the workplace 2 Hazardous Chemical Any chemical which poses a physical or health

More information

Hazard Communication Program

Hazard Communication Program SOUTHWESTERN COMMUNITY COLLEGE Hazard Communication Program Revised January 2005 TABLE OF CONTENTS 1. Purpose 1 2. Responsibilities 1 3. Access to the Written Program 1 4. Hazard Determination and Inventory

More information

Application Equipment means a device or equipment used to apply solvent, sealant, adhesive, coating, ink, or polyester resin materials.

Application Equipment means a device or equipment used to apply solvent, sealant, adhesive, coating, ink, or polyester resin materials. RULE 102. DEFINITIONS. (Adopted 10/18/1971, revised 1/12/1976, readopted 10/23/1978, revised 7/11/1989, 7/10/1990, 7/30/1991, 7/18/1996, 4/17/1997, 1/21/1999, 5/20/1999, 6/19/2003, 1/20/2005, 6/19/2008,

More information

Acetals, more than solvents. Expanding Chemistry 8-9 November 2017

Acetals, more than solvents. Expanding Chemistry 8-9 November 2017 Acetals, more than solvents Expanding Chemistry 8-9 November 2017 Acetals RANGE 2 Acetals Name Methylal Formula CH3-O-CH2-O-CH3 Ethylal CH3-CH2-O-CH2-O-CH2-CH3 Propylal CH3-CH2-CH2-O-CH2-O-CH2-CH2-CH3

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Fabriksvej 19 6000 Kolding DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 13 August 2015 Test Report- VOC emission regulations

More information

method for monitoring exposure to gasoline vapour in air revision 2002

method for monitoring exposure to gasoline vapour in air revision 2002 method for monitoring exposure to gasoline vapour in air revision 2002 Prepared by: P. E. Tindle Reviewed for CONCAWE by: A. Bianchi E. Loader C. Money G. Pizzella J. Urbanus (Technical Co-ordinator) Reproduction

More information

UV-PCO DEVICE FOR INDOOR VOCS REMOVAL: INVESTIGATION ON MULTIPLE COMPOUNDS EFFECT

UV-PCO DEVICE FOR INDOOR VOCS REMOVAL: INVESTIGATION ON MULTIPLE COMPOUNDS EFFECT UV-PCO DEVICE FOR INDOOR VOCS REMOVAL: INVESTIGATION ON MULTIPLE COMPOUNDS EFFECT WH Chen *, JS Zhang and ZB Zhang Dept. of Mechanical and Aerospace Engineering, Syracuse University, Syracuse, NY 3244,

More information

Available online at Procedia Engineering 7 (2010) Procedia Engineering 00 (2010)

Available online at   Procedia Engineering 7 (2010) Procedia Engineering 00 (2010) Available online at www.sciencedirect.com Procedia Engineering 7 (2010) 392 398 Procedia Engineering 00 (2010) 000 000 Procedia Engineering www.elsevier.com/locate/procedia www.elsevier.com/locate/procedia

More information

EXPERIMENT 7 - Distillation Separation of a Mixture

EXPERIMENT 7 - Distillation Separation of a Mixture EXPERIMENT 7 - Distillation Separation of a Mixture Purpose: a) To purify a compound by separating it from a non-volatile or less-volatile material. b) To separate a mixture of two miscible liquids (liquids

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Aps Fabriksvej 19 6000 Kolding Denmark Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 14 November 2014 Test Report- VOC emission

More information

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP Application Note Joseph M. Levy Michael Kraft Abstract Agilent Equipment 7890A

More information

Accepted for publication in Analytical Chemistry, August, 2011 SUPPORTING INFORMATION FOR. address:

Accepted for publication in Analytical Chemistry, August, 2011 SUPPORTING INFORMATION FOR.  address: Accepted for publication in Analytical Chemistry, August, 0 SUPPORTING INFORMATION FOR Microfabricated Gas Chromatograph for the Selective Determination of Trichloroethylene Vapor at Sub-Parts-Per-Billion

More information

Spontaneous Desorption of Organic Solvents from

Spontaneous Desorption of Organic Solvents from Industrial Health, 1987, 25, 73--81 Spontaneous Desorption of Organic Solvents from Carbon Cloth Miyuki KASAHARA and Masayuki IKEDA Depertment of Environmental Health, Tohoku University School of Medicine,

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe CONICA AG Industriestr. 26 CH-8207 Schauffhausen Switzerland Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 31 July 2014 Test Report-

More information

Summary of Results Work Order No

Summary of Results Work Order No Summary of Results Work Order No. 2502295 Sample Name: C07550 - Indoor The TVOCs concentration of this sample was found to be 440 ug/m 3. The top 3 Volatile Organic Compounds identified in this sample

More information