1 12th International Conference & Workshop Medical Physics in the Baltic States, 2015 RAMAN SPECTROSCOPY OF POLYMERIZATION PROCESSES IN npag AND nmag DOSE GELS Neringa VAIČIŪNAITĖ 1, Rimas ŠEPERYS 2 1 Physics Department, Kaunas University of Technology 2 SME Šeši partneriai 1

2 INTRODUCTION Polymer gels have been proven to be a valuable tool for determination of beam dose characteristics and 3D radiation dose measurements for medical purposes [1-3]. A typical gel dosemeter consists of water, gelatin agent, monomers and a cross-linker comonomer [4,5]. Many chemical compositions have been proposed for dosimetric purposes, but only a few have been carefully tested [4,5]. To evaluate the interaction between monomer consumption and polymer production is very important [6]. 2

3 AIM OF THE STUDY The aim of this study was to identify the radiation interaction mechanisms on photon irradiated npag and nmag polymer gels in accordance with analysis of characteristic vibrational modes of dose gel components

4 POLYMER GELS PREPARATION AND IRRADIATION npag nmag 89% Highly purified water 86% Highly purified water 5% Gelatine 8% Gelatine 3% Acrylaimde 6% Methacrylic acid 3% N; N- methylenebisacrylamide 10 mmol/l Hydroxymethyl phosphonium chloride - 10 mmol/l Hydroxymethyl phosphonium chloride Fig.1. Preparation of the dose gel. npag and nmag polymer gels were irradiated by Co 60 photon source of teletherapy unit (ROKUS M) by 2 Gy dose. Dose rate of 0.03 Gy/ min, SSD 100 cm, field opening cm. 4

5 POLYMER GELS PREPARATION AND IRRADIATION Fig. 2. Micro Raman invia Renishaw c) d) Fig. 3. Non irradiated npag gel, Non irradiated nmag gel, c) npag irradiated by 2 Gy Co-60, d) nmag irradiated by 2 Gy Co-60. Micro Raman (invia Renishaw) vibrational spectroscopy of the irradiated copolymerized npag and nmag samples in cuvettes was performed on the 7th post-irradiation day.

6 RESULTS AND DISCUSSION npag Fig.4. Chemical structure of acrylamide, polyacrylamide. Fig.5. Micro -Raman spectra of npag polymer dose gel: non irradiated, irradiated by 2 Gy by Co 60 photon source. 6

7 npag RESULTS AND DISCUSSION FWHM = 310,9 FWHM = 1229,1 c) d) Fig.6. Polymer development observed in Raman spectra of normoxic polyacrylamide (npag) gel in the range of cm -1 and cm -1., c) Non irradiated npag gel;, d) irradiated npag gel. FWHM = 121,7 FWHM = 213,6

8 RESULTS AND DISCUSSION nmag Fig.7. Chemical structure of methacrylic acid, polymetchacrylic acid. Fig. 8. Raman spectra of nmag polymer dose gel: non irradiated, irradiated by 2 Gy by Co60 photon source. 8

9 nmag RESULTS AND DISCUSSION FWHM = 111,4 c) FWHM = 1198,8 d) Fig. 9. Polymer development observed in Raman spectra of normoxic methacrylic acid (nmag) gel at the range of cm-1 and cm-1., c) Non irradiated nmag gel;, d) irradiated nmag gel. FWHM = 102,8 FWHM = 290,7

10 CONCLUSIONS Photon irradiation of Co 60 was applied to normoxic npag and nmag polymer dose gels to perform a polymerization process. The Micro-Raman spectra and full-width half-maximum calculations, showed the incensement of polymer representing vibrational bands and sharp reduction of monomer fraction. The reduction of BIS proved the theory of cross-linker consumption and its importance in network structure formation. The results agreed with the studies of other researchers, however more tests are needed to evaluate the consumption rate of monomers and cross-linker. 10

11 Thank You for your attention 11

12 REFERENCES 1. Papoutsaki M., Maris T.G., Pappas E. Dosimetric characteristics of a new polymer gel and their dependence on post-preparation and post-irradiation time: Effect on X-ray beam profile measurements. Physica Medica 9, p McAuley K.B. Fundamentals of Polymer Gel Dosimeters. Journal of Physics: Conference Series 56, p , doi: / /56/1/ De Deene Y. Essential characteristics of polymer gel dosimeters. Third International Conference on Radiotherapy Gel Dosimetry, Journal of Physics: Conference Series 3, p , doi: / /3/1/ Baldock C., De Deene Y.,Doran S. Topical Review: Polymer gel dosimetry. Phys Med Biol, doi: / /55/5/r Bong J., Choi K., Yu S.C. Raman Spectroscopy of Irradiated Normoxic Polymethacrylic Acid Gel Dosimeter. Bull. Korean Chem. Soc., Vol. 32, No. 2, p doi /bkcs Maryanski M. J., Zastavker Y. Z., and Gore J. C. Radiation dose distributions in three dimensions fromtomographic optical density scanning of polymer gels: II. Optical properties of the BANG polymer gel. Phys. Med. Biol. 41, p

Raman Spectroscopy of Irradiated Normoxic Polymethacrylic Acid Gel Dosimeter

Raman Spectroscopy of Irradiated Normoxic Polymethacrylic Acid Gel Dosimeter Raman Spectroscopy of Irradiated Gel Dosimeter Bull. Korean Chem. Soc. 2011, Vol. 32, No. 2 625 DOI 10.5012/bkcs.2011.32.2.625 Raman Spectroscopy of Irradiated Normoxic Polymethacrylic Acid Gel Dosimeter

More information

The Effects of Crosslinker and Monomer to the Polymerization of Polymethacrylamide Gel Dosimeters by Direct and Indirect Methods

The Effects of Crosslinker and Monomer to the Polymerization of Polymethacrylamide Gel Dosimeters by Direct and Indirect Methods American Journal of Applied Sciences 2 (8): 1248-1255, 2005 ISSN 1546-9239 2005 Science Publications The Effects of Crosslinker and Monomer to the Polymerization of Polymethacrylamide Gel Dosimeters by

More information

INVESTIGATIONS OF THE FRICKE GEL (FXG) DOSIMETER DEVELOPED AT IPEN IRRADIATED WITH 60 Co GAMMA RAYS

INVESTIGATIONS OF THE FRICKE GEL (FXG) DOSIMETER DEVELOPED AT IPEN IRRADIATED WITH 60 Co GAMMA RAYS 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 INVESTIGATIONS OF

More information

On the relation between the spatial dose integrity and the temporal instability of polymer gel dosimeters

On the relation between the spatial dose integrity and the temporal instability of polymer gel dosimeters INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 9 () 57 5 PHYSICS IN MEDICINE AND BIOLOGY PII: S-955()877-5 On the relation between the spatial dose integrity and the temporal instability of polymer gel

More information

Chapter XII: Chemical dosimeters

Chapter XII: Chemical dosimeters Chapter XII: Chemical dosimeters 1 Chemical dosimeters Introduction Fricke dosimeter Polymer gel dosimeter Alanine dosimeter Radiochromic film 2 Principle of chemical dosimeters In a chemical dosimeter

More information

Mathematical Modeling of PAG and NIPAM-Based. Polymer Gel Dosimeters contaminated by oxygen and

Mathematical Modeling of PAG and NIPAM-Based. Polymer Gel Dosimeters contaminated by oxygen and Mathematical Modeling of PAG and NIPAM-Based Polymer Gel Dosimeters contaminated by oxygen and inhibitor. Valeria I Koeva 1, Shahab Daneshvar 1, Robert J Senden 1, A H M Imam 1, L John Schreiner 2,3 and

More information

Temperature increases associated with polymerization of irradiated PAG dosimeters

Temperature increases associated with polymerization of irradiated PAG dosimeters INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (02) 1435 1448 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)26033-9 Temperature increases associated with polymerization of irradiated PAG dosimeters

More information

Calibration of an alanine/agarose gel

Calibration of an alanine/agarose gel Institutionen för medicin och vård Avdelningen för radiofysik Hälsouniversitetet Calibration of an alanine/agarose gel Sara Olsson Department of Medicine and Care Radio Physics Faculty of Health Sciences

More information

Contrast mechanisms in magnetic resonance imaging

Contrast mechanisms in magnetic resonance imaging Journal of Physics: Conference Series Contrast mechanisms in magnetic resonance imaging To cite this article: M Lepage and J C Gore 2004 J. Phys.: Conf. Ser. 3 78 View the article online for updates and

More information

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW NPL REPORT IR 1 Gel dosimetry for direct measurement of dose distributions: Progress report J A D PEARCE and D C CROSSLEY NOT RESTRICTED OCTOBER 2007 National Physical Laboratory Hampton Road Teddington

More information

Dose enhancement effect of gold nanoparticles on MAGICA polymer gel in mega voltage radiation therapy

Dose enhancement effect of gold nanoparticles on MAGICA polymer gel in mega voltage radiation therapy Volume 11, No 1 International Journal of Radiation Research, Jan 2013 Dose enhancement effect of gold nanoparticles on MAGICA polymer gel in mega voltage radiation therapy M. Khadem-Abolfazli 1,2, M. Mahdavi

More information

Absorption spectra variations of EBT radiochromic film from radiation exposure

Absorption spectra variations of EBT radiochromic film from radiation exposure INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N35 N4 PHYSICS IN MEDICINE AND BIOLOGY doi:.88/3-955/5/3/n2 NOTE Absorption spectra variations of EBT radiochromic film from radiation exposure M

More information

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION DOI: 10.1038/NPHOTON.2013.97 Supplementary Information Far-field Imaging of Non-fluorescent Species with Sub-diffraction Resolution Pu Wang et al. 1. Theory of saturated transient absorption microscopy

More information

Use of a radioactive check device for redundancy check of ionization chambers

Use of a radioactive check device for redundancy check of ionization chambers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman

More information

Orientational behavior of liquid-crystalline polymers with amide groups

Orientational behavior of liquid-crystalline polymers with amide groups Advances in Materials 2014; 3(6): 89-93 Published online January 06, 2015 (http://www.sciencepublishinggroup.com/j/am) doi: 10.11648/j.am.20140306.14 ISSN: 2327-2503 (Print); ISSN: 2327-252X (Online) Orientational

More information

CHARACTERIZATION OF AAm/MBA HYDROGELS PREPARED BY RADIATION INDUCED POLYMERIZATION

CHARACTERIZATION OF AAm/MBA HYDROGELS PREPARED BY RADIATION INDUCED POLYMERIZATION Physica Macedonica 61, (2012) p. 73-78 ISSN 1409-7168 CHARACTERIZATION OF AAm/MBA HYDROGELS PREPARED BY RADIATION INDUCED POLYMERIZATION N. Mahmudi 1* and S. Rendevski 2 1 Faculty of Natural Science and

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

Chapter VIII: Photographic films

Chapter VIII: Photographic films Chapter VIII: Photographic films 1 Photographic films First (integrating) dosimeters to be developed based on photographic techniques dosifilms Dosifilms were more and more replaced by dosimeters using

More information

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range

More information

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers Small Field Dosimetric Measurements with TLD-1, Alanine, and Ionization Chambers S. Junell a, L. DeWerd a, M. Saiful Huq b, J. Novotny Jr. b, M. uader b, M.F. Desrosiers c, G. Bednarz b a Department of

More information

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.264-271 Variation of Dose Distribution by Detectors for Narrow Beam T. Fujisaki, H.

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

General characteristics of radiation dosimeters

General characteristics of radiation dosimeters General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

ph dependent thermoresponsive behavior of acrylamide-acrylonitrile UCSTtype copolymers in aqueous media

ph dependent thermoresponsive behavior of acrylamide-acrylonitrile UCSTtype copolymers in aqueous media Electronic Supplementary Material (ESI) for Polymer Chemistry. This journal is The Royal Society of Chemistry 2016 Supporting Information ph dependent thermoresponsive behavior of acrylamide-acrylonitrile

More information

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE Page no. 5.1 Introduction 132 5.2 Methods and measurements 132 5.3 Saturation thickness of elements and polymers

More information

City University of Hong Kong

City University of Hong Kong City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

6 Egypt. J. Rad. Sci. Applic., Vol. 29, No. 1-2, pp (2016)

6 Egypt. J. Rad. Sci. Applic., Vol. 29, No. 1-2, pp (2016) 6 Egypt. J. Rad. Sci. Applic., Vol. 29, No. 1-2, pp. 75-84 (2016) Physical Study Comparison of Neutral Red dye Prepared by Different Techniques for Low Dose Monitor Application E. Fahim and M. El-Kelany

More information

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements Hans-Georg Menzel International Commission on Radiation Units and Measurements (ICRU) The principal objective of ICRU

More information

DEVELOPMENT OF A TRANSPARENT AND DEFORMABLE TWO DIMENSIONAL RADIOCHROMIC GEL DOSIMETER. POURIA ATAEI, B.Eng

DEVELOPMENT OF A TRANSPARENT AND DEFORMABLE TWO DIMENSIONAL RADIOCHROMIC GEL DOSIMETER. POURIA ATAEI, B.Eng GEL DOSIMETRY i DEVELOPMENT OF A TRANSPARENT AND DEFORMABLE TWO DIMENSIONAL RADIOCHROMIC GEL DOSIMETER by POURIA ATAEI, B.Eng A Thesis Submitted to the School of Graduate Studies in Partial Fulfillment

More information

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 1 Medical Radiation Physics Group, University of Oldenburg and Pius

More information

J. López-Sánchez*,, A. Serrano,Ø, A. Del Campo Ø, M. Abuín,, O. Rodríguez de la Fuente,, N. Carmona, Macroscopic aspect of the samples

J. López-Sánchez*,, A. Serrano,Ø, A. Del Campo Ø, M. Abuín,, O. Rodríguez de la Fuente,, N. Carmona, Macroscopic aspect of the samples J. López-Sánchez*,, A. Serrano,Ø, A. Del Campo Ø, M. Abuín,, O. Rodríguez de la Fuente,, N. Carmona, Departamento de Física de Materiales, Universidad Complutense de Madrid, 28040 Madrid Unidad Asociada

More information

Vibrational properties of isotopically enriched materials: the case of calcite Ben Xu, Anna Hirsch, Leeor Kronik, and Kristin M.

Vibrational properties of isotopically enriched materials: the case of calcite Ben Xu, Anna Hirsch, Leeor Kronik, and Kristin M. Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 8 Electronic Supporting Information Vibrational properties of isotopically enriched materials: the

More information

A Study on Effective Source-Skin Distance using Phantom in Electron Beam Therapy

A Study on Effective Source-Skin Distance using Phantom in Electron Beam Therapy Journal of Magnetics 19(1), 15-19 (2014) ISSN (Print) 1226-1750 ISSN (Online) 2233-6656 http://dx.doi.org/10.4283/jmag.2014.19.1.015 A Study on Effective Source-Skin Distance using Phantom in Electron

More information

Supporting information. ph- and Thermal-Responsive Multishape Memory Hydrogel

Supporting information. ph- and Thermal-Responsive Multishape Memory Hydrogel Supporting information ph- and Thermal-Responsive Multishape Memory Hydrogel Xiao-Lei Gong,,, # Yao-Yu Xiao,, # Min Pan, Yang Kang, Bang-Jing Li,* and Sheng Zhang* State Key Laboratory of Polymer Materials

More information

Spectroscopic Quantification of Kinetic Rate Constants for Epoxy-Acrylate Hybrid Photopolymerization

Spectroscopic Quantification of Kinetic Rate Constants for Epoxy-Acrylate Hybrid Photopolymerization Spectroscopic Quantification of Kinetic Rate Constants for Epoxy-Acrylate Hybrid Photopolymerization Brian Dillman University of Iowa, Chemical and Biochemical Eng. Dept., Iowa City, USA Julie L. P. Jessop

More information

Using Visible Laser Based Raman Spectroscopy to Identify the Surface Polarity of Silicon Carbide

Using Visible Laser Based Raman Spectroscopy to Identify the Surface Polarity of Silicon Carbide Supporting Information for Using Visible Laser Based Raman Spectroscopy to Identify the Surface Polarity of Silicon Carbide Submitted by Yi-Chuan Tseng, a Yu-Chia Cheng, a Yang-Chun Lee, a Dai-Liang Ma,

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego Physics of Novel Radiation Modalities Particles and Isotopes Todd Pawlicki, Ph.D. UC San Diego Disclosure I have no conflicts of interest to disclose. Learning Objectives Understand the physics of proton

More information

Radiation exposure of personnel during IORT: radiation protection aspects.

Radiation exposure of personnel during IORT: radiation protection aspects. Radiation exposure of personnel during IORT: radiation protection aspects. L. Strigari 1, V. Bruzzaniti 1, V. Landoni 1, A. Soriani 1, S.Teodoli 1, M. Benassi 1 1 Lab. Fisica Medica e Sistemi Esperti,

More information

IRRADIATION EFFECTS ON POLY (VINYL CHLORIDE)

IRRADIATION EFFECTS ON POLY (VINYL CHLORIDE) IRRADIATION EFFECTS ON POLY (VINYL CHLORIDE) L.COSTA, V. BRUNELLA, P. BRACCO Dipartimento di Chimica IFM, Università di Torino, Italy E-mail: brunella@ch.unito.it We have studied the electron beam effects

More information

Plastic Scintillation Detectors: Principle and Application to Radiosurgery

Plastic Scintillation Detectors: Principle and Application to Radiosurgery Plastic Scintillation Detectors: Principle and Application to Radiosurgery Luc Beaulieu Professor, Department of Physics, Université Laval Medical Physicist and Head of Research, Department of Radiation

More information

Calibration of a pencil ionization chamber with and without preamplifier

Calibration of a pencil ionization chamber with and without preamplifier Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof.

More information

Radiation protection issues in proton therapy

Radiation protection issues in proton therapy Protons IMRT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Overview of presentation 1. Proton therapy: An overview 2. Radiation protection issues: Staff 3. Radiation

More information

Chapter 9: Calibration of Photon and Electron Beams

Chapter 9: Calibration of Photon and Electron Beams Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation

More information

Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields

Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields SLAC-PUB-7901 August 1998 Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields John J. Jummo and James C. Liu Presented at 12th International Conference

More information

NEUTRON SPECTROMETRY WITH BUBBLE DETECTORS

NEUTRON SPECTROMETRY WITH BUBBLE DETECTORS 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 NEUTRON SPECTROMETRY

More information

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator Determination of Ambient Dose quivalent at INFLPR 7 MeV Linear Accelerator F. Scarlat, A. Scarisoreanu, M. Oane,. Badita,. Mitru National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest-Magurele,

More information

The effect of dose calculation uncertainty on the evaluation of radiotherapy plans

The effect of dose calculation uncertainty on the evaluation of radiotherapy plans The effect of dose calculation uncertainty on the evaluation of radiotherapy plans P. J. Keall a) and J. V. Siebers Department of Radiation Oncology, Medical College of Virginia Hospitals, Virginia Commonwealth

More information

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento

More information

THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR*

THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR* SLAC-PUB-5122 Rev March 1991 w THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR* N. E. Ipe, R. J. Donahue, and D. D. Busick Stanford Linear Accelerator Center Stanford University,

More information

Investigation of dose rate effects in radiation chemistry with laser-driven pulsed sources at ELI Beamlines

Investigation of dose rate effects in radiation chemistry with laser-driven pulsed sources at ELI Beamlines Investigation of dose rate effects in radiation chemistry with laser-driven pulsed sources at ELI Beamlines Martin PŘEČEK, Ph.D. ELI Beamlines Research program 4 Applications Institute of Physics (FZÚ),

More information

Conclusion and Future Work

Conclusion and Future Work Chapter 7 7. Chapter 7 and Future Work Chapter 7 Abstract This chapter gives the details of correlations of the spectroscopic investigation results with those available from other studies and also summarizes

More information

Laser-Accelerated protons for radiation therapy

Laser-Accelerated protons for radiation therapy Laser-Accelerated protons for radiation therapy E Fourkal, I Velchev,, J Fan, J Li, T Lin, C Ma Fox Chase Cancer Center, Philadelphia, PA Motivation Proton beams provide better conformity to the treatment

More information

The volume effect of detectors in the dosimetry of small fields used in IMRT

The volume effect of detectors in the dosimetry of small fields used in IMRT The volume effect of detectors in the dosimetry of small fields used in IMRT Wolfram U. Laub a) Department of Medical Physics, University of Tübingen, Germany Tony Wong b) Department of Radiation Oncology,

More information

Introduction to microdosimetry! Manuel Bardiès, INSERM UMR 892, Nantes, France

Introduction to microdosimetry! Manuel Bardiès, INSERM UMR 892, Nantes, France Introduction to microdosimetry! Manuel Bardiès, INSERM UMR 892, Nantes, France manuel.bardies@inserm.fr Introduction to microdosimetry! In general, NM dosimetry: macrodosimetry, but:! macrodosimetry macroscopic

More information

The Radiation Effect of 60 Co Gamma Rays on Polycarbonate detector.

The Radiation Effect of 60 Co Gamma Rays on Polycarbonate detector. The Radiation Effect of 60 Co Gamma Rays on Polycarbonate detector. Ana Maria Sisti Galante *, Letícia Lucente Campos Radiation Metrology Centre Institute of Energetic and Nuclear Research, Av. Prof. Lineu

More information

DOWNLOAD OR READ : INFRARED AND RAMAN SPECTROSCOPY CONCEPTS AND APPLICATIONS PDF EBOOK EPUB MOBI

DOWNLOAD OR READ : INFRARED AND RAMAN SPECTROSCOPY CONCEPTS AND APPLICATIONS PDF EBOOK EPUB MOBI DOWNLOAD OR READ : INFRARED AND RAMAN SPECTROSCOPY CONCEPTS AND APPLICATIONS PDF EBOOK EPUB MOBI Page 1 Page 2 infrared and raman spectroscopy concepts and applications infrared and raman spectroscopy

More information

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin

More information

Electronic Supplemental Information. Metal nanoparticle catalyzed cyclobutane cleavage reaction

Electronic Supplemental Information. Metal nanoparticle catalyzed cyclobutane cleavage reaction Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2015 Electronic Supplemental Information Metal nanoparticle catalyzed cyclobutane cleavage reaction

More information

Terpolymerization of 2-ethoxy ethylmethacrylate, styrene and maleic anhydride: determination of the reactivity ratios

Terpolymerization of 2-ethoxy ethylmethacrylate, styrene and maleic anhydride: determination of the reactivity ratios Bull. Mater. Sci., Vol. 27, No. 3, June 2004, pp. 243 249. Indian Academy of Sciences. Terpolymerization of 2-ethoxy ethylmethacrylate, styrene and maleic anhydride: determination of the reactivity ratios

More information

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination

More information

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

The Radiological Physics Center s standard dataset for small field size output factors

The Radiological Physics Center s standard dataset for small field size output factors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 The Radiological Physics Center s standard dataset for small field size output factors David S. Followill, 1a Stephen F. Kry, 1 Lihong

More information

CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK

CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK 161 CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK 7.1 SUMMARY OF THE PRESENT WORK Nonlinear optical materials are required in a wide range of important applications, such as optical

More information

Atoms, Radiation, and Radiation Protection

Atoms, Radiation, and Radiation Protection James E. Turner Atoms, Radiation, and Radiation Protection Third, Completely Revised and Enlarged Edition BICENTENNIAL J 0 1 8 0 Q 71 z m z CAVILEY 2007 1 ;Z z ü ; m r B10ENTENNIAL WILEY-VCH Verlag GmbH

More information

Comparison of some human tissues and some commonly used thermoluminescent dosimeters for photon energy absorption

Comparison of some human tissues and some commonly used thermoluminescent dosimeters for photon energy absorption American Journal of Physics and Applications 2014; 2(6): 145-149 Published online December 16, 2014 (http://www.sciencepublishinggroup.com/j/ajpa) doi: 10.11648/j.ajpa.20140206.15 ISSN: 2330-4286 (Print);

More information

International Journal of Scientific & Engineering Research, Volume 7, Issue 2, February-2016 ISSN

International Journal of Scientific & Engineering Research, Volume 7, Issue 2, February-2016 ISSN 1465 International Journal of Scientific & Engineering Research, Volume 7, Issue, February-016 ISSN 9-5518 Investigation of Medical Neutron Radiation Doses Using Two types of bubble Detectors Hsin-Ell

More information

Synthesis and spectroscopic characterization of completely isotactic polyacrylonitrile. Jun-Ting Zou, Yu-Song Wang, Wen-Min Pang, Lei Shi,* and Fei Lu

Synthesis and spectroscopic characterization of completely isotactic polyacrylonitrile. Jun-Ting Zou, Yu-Song Wang, Wen-Min Pang, Lei Shi,* and Fei Lu Synthesis and spectroscopic characterization of completely isotactic polyacrylonitrile Jun-Ting Zou, Yu-Song Wang, Wen-Min Pang, Lei Shi,* and Fei Lu Hefei National Laboratory for Physical Sciences at

More information

Synthesis and Characterization of Grafted Polystyrene with Acrylic Acid Using Gamma-Irradiation

Synthesis and Characterization of Grafted Polystyrene with Acrylic Acid Using Gamma-Irradiation Australian Journal of Basic and Applied Sciences, 7(8): 746-750, 2013 ISSN 1991-8178 Synthesis and Characterization of Grafted Polystyrene with Acrylic Acid Using Gamma-Irradiation Moayad abd al-hassan

More information

Determining how uncertainties in optical properties affect light dose calculations

Determining how uncertainties in optical properties affect light dose calculations Determining how uncertainties in optical properties affect light dose calculations Julia Sandell 1, Jarod Finlay, Timothy Zhu 1 Department of Physics, University of Pennsylvania Radiation Oncology, Hospital

More information

Preparation and Characterization of Hydrogels

Preparation and Characterization of Hydrogels Chapter 2 Preparation and Characterization of Hydrogels 2.1 Materials and Methods 2.1.1 Materials 2-Acrylamido-2-methylpropane sulfonic acid (AMPS) was obtained from Vinati Organic Ltd. Acrylamide (AM),

More information

APPLIED RADIATION PHYSICS

APPLIED RADIATION PHYSICS A PRIMER IN APPLIED RADIATION PHYSICS F A SMITH Queen Mary & Westfield College, London fe World Scientific m Singapore * New Jersey London Hong Kong CONTENTS CHAPTER 1 : SOURCES of RADIATION 1.1 Introduction

More information

Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up

Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up G. Arico1,2, B. Hartmann2, J. Jakubek3, S. Pospisil3, O. Jäkel1,2,4, M. Martisikova1,2 1 University Clinic

More information

Damage to Molecular Solids Irradiated by X-ray Laser Beam

Damage to Molecular Solids Irradiated by X-ray Laser Beam WDS'11 Proceedings of Contributed Papers, Part II, 247 251, 2011. ISBN 978-80-7378-185-9 MATFYZPRESS Damage to Molecular Solids Irradiated by X-ray Laser Beam T. Burian, V. Hájková, J. Chalupský, L. Juha,

More information

Dosimetric Properties of the Field Sizes of 12MV Photon Beams: A Monte Carlo Study

Dosimetric Properties of the Field Sizes of 12MV Photon Beams: A Monte Carlo Study Journal of Nuclear and Particle Physics 2015, 5(3): 52-57 DOI: 10.5923/j.jnpp.20150503.02 Dosimetric Properties of the Field Sizes of 12MV Photon Beams: A Monte Carlo Study Maged Mohammed 1,2,*, E. Chakir

More information

Supplementary information

Supplementary information Electronic Supplementary Material (ESI) for Nanoscale. This journal is The Royal Society of Chemistry 2016 Supplementary information 1. Composition of the nanoparticles 2. Dynamic light scattering 3. Scanning

More information

Investigation of the relationship between linear attenuation coefficients and CT Hounsfield units using radionuclides for SPECT

Investigation of the relationship between linear attenuation coefficients and CT Hounsfield units using radionuclides for SPECT Applied Radiation and Isotopes 66 (2008) 1206 1212 www.elsevier.com/locate/apradiso Investigation of the relationship between linear attenuation coefficients and CT Hounsfield units using radionuclides

More information

Progress in Nuclear Science and Technology, Volume 6,

Progress in Nuclear Science and Technology, Volume 6, DOI: 1.15669/pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp. 1-16 ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi

More information

Polymer Chemistry Prof. Dibakar Dhara Department of Chemistry Indian Institute of Technology, Kharagpur

Polymer Chemistry Prof. Dibakar Dhara Department of Chemistry Indian Institute of Technology, Kharagpur Polymer Chemistry Prof. Dibakar Dhara Department of Chemistry Indian Institute of Technology, Kharagpur Lecture - 10 Radical Chain Polymerization (Contd.) (Refer Slide Time: 00:28) Welcome back, and we

More information

Investigation of EBT3 radiochromic film s response to humidity

Investigation of EBT3 radiochromic film s response to humidity Received: 19 October 2017 Revised: 10 January 2018 Accepted: 20 March 2018 DOI: 10.1002/acm2.12337 RADIATION ONCOLOGY PHYSICS Investigation of EBT3 radiochromic film s response to humidity Elsa Y. Leon-Marroquın

More information

Structure-Property Investigation of Functional Resins for UV-Curable Gaskets

Structure-Property Investigation of Functional Resins for UV-Curable Gaskets Structure-Property Investigation of Functional Resins for UV-Curable Gaskets Joel D. Schall and Eric Edo-Hernandez Henkel Corporation Rocky Hill, CT USA Introduction The main purpose of this work was to

More information

Nur Farizah Ayub, Shahrir Hashim and Nadia Adrus. Faculty of Chemical Engineering, Universiti Teknologi Malaysia, UTM Skudai, Johor, MALAYSIA

Nur Farizah Ayub, Shahrir Hashim and Nadia Adrus. Faculty of Chemical Engineering, Universiti Teknologi Malaysia, UTM Skudai, Johor, MALAYSIA Nur Farizah Ayub, Shahrir Hashim and Nadia Adrus Faculty of Chemical Engineering, Universiti Teknologi Malaysia, UTM Skudai, Johor, MALAYSIA INTRODUCTION Research Background: Hydrogels Three dimensional

More information

EBT2 Dosimetry of X-rays produced by the electron beam from PFMA-3, a Plasma Focus for medical applications

EBT2 Dosimetry of X-rays produced by the electron beam from PFMA-3, a Plasma Focus for medical applications EBT2 Dosimetry of X-rays produced by the electron beam from PFMA-3, a Plasma Focus for medical applications E Ceccolini 1, F Rocchi 1,2, D Mostacci 1, M Sumini 1 and A Tartari 3 1 Montecuccolino Nuclear

More information

Keywords: mechanical property, polymer optical fibre, radiation, transmission.

Keywords: mechanical property, polymer optical fibre, radiation, transmission. 5 th Australasian Congress on Applied Mechanics, ACAM 27 1-12 December 27, Brisbane, Australia Radiation damage to polymer optical fibres C. Yan 1, S.H. Law 2, N. Suchowerska 3,4 and S.H. Hong 5 1 School

More information

THERMOLUMINESCENCE PROPERTIES OF OLIVINE

THERMOLUMINESCENCE PROPERTIES OF OLIVINE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 THERMOLUMINESCENCE PROPERTIES

More information

The vibrational spectroscopy of polymers

The vibrational spectroscopy of polymers D. I. BOWER Reader in Polymer Spectroscopy Interdisciplinary Research Centre in Polymer Science & Technology Department of Physics, University of Leeds W.F. MADDAMS Senior Visiting Fellow Department of

More information

Spectroscopy of Polymers

Spectroscopy of Polymers Spectroscopy of Polymers Jack L. Koenig Case Western Reserve University WOMACS Professional Reference Book American Chemical Society, Washington, DC 1992 Contents Preface m xiii Theory of Polymer Characterization

More information

Dose Response and Optical Properties of Dyed Poly Vinyl Alcohol-Trichloroacetic Acid Polymeric Blends Irradiated with Gamma-Rays

Dose Response and Optical Properties of Dyed Poly Vinyl Alcohol-Trichloroacetic Acid Polymeric Blends Irradiated with Gamma-Rays American Journal of Applied Sciences 6 (12): 2071-2077, 2009 ISSN 1546-9239 2009 Science Publications Dose Response and Optical Properties of Dyed Poly Vinyl Alcohol-Trichloroacetic Acid Polymeric Blends

More information

Available online at Nuclear Instruments and Methods in Physics Research B 265 (2007)

Available online at   Nuclear Instruments and Methods in Physics Research B 265 (2007) Available online at www.sciencedirect.com Nuclear Instruments and Methods in Physics Research B 65 (007) 379 384 NIM B Beam Interactions with Materials & Atoms www.elsevier.com/locate/nimb Synthesis, characterization

More information

Towards the general application of water calorimetry as absorbed dose standard for radiotherapy dosimetry

Towards the general application of water calorimetry as absorbed dose standard for radiotherapy dosimetry Towards the general application of water calorimetry as absorbed dose standard for radiotherapy dosimetry A. Krauss, M. Bambynek, H.-J. Selbach Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116

More information

Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance

Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance Michael P. Barnes 1,2 a and Peter B. Greer 1,3

More information

CVD Diamond History Introduction to DDL Properties of Diamond DDL Proprietary Contact Technology Detector Applications BDD Sensors

CVD Diamond History Introduction to DDL Properties of Diamond DDL Proprietary Contact Technology Detector Applications BDD Sensors Diamond Detectors CVD Diamond History Introduction to DDL Properties of Diamond DDL Proprietary Contact Technology Detector Applications BDD Sensors Kevin Oliver CEO Alex Brown Sales & Marketing 20 May,

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 156 Copper Nanoparticles: Green Synthesis Characterization Y.Suresh*1, S.Annapurna*2, G.Bhikshamaiah*3, A.K.Singh#4 Abstract Present work describes the synthesis nanoparticles using papaya extract as a

More information

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators David S. Followill,*

More information