MEMBRANE COVERED POLAROGRAPHIC OXYGEN SENSOR MANUFACTURING THEORETICAL CONSIDERATIONS
|
|
- Molly Brown
- 5 years ago
- Views:
Transcription
1 ANALELE ŞTIINŢIFICE ALE UNIVERSITĂŢII AL. I. CUZA IAŞI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 006 MEMBRANE COVERED POLAROGRAPHIC OXYGEN SENSOR MANUFACTURING THEORETICAL CONSIDERATIONS Alina Manole 1, I. Neacsu, M-O. Apostu 3, V. Melnig 1 KEYWORDS: Clark electrode, polarography, diffusion oxygen models The operating principle of the Clark type polarographic oxygen electrode is based on the use of an electrolyte solution contained within the electrode assembly to transport oxygen from an oxygen-permeable membrane to the metal cathode. In this paper are reconsidered theoretical assumption of one layer and two layer diffusion models and the evaluation of optimal calibration and error measurements. 1. INTRODUCTION Electrochemical oxygen analysers are based on electrochemical reduction of O at a negatively polarized electrode. This principle lies in the bases of the Clark-type oxygen-sensitive electrode which was proposed in 1956 by Leland Clark [1, ]. Clark studied the electrochemistry of oxygen gas reduction at platinum (Pt) metal electrodes, pioneering the use of Pt electrodes as an oxygen sensor. (Pt electrodes used to detect oxygen electrochemically are often referred as Clark electrodes. ) An electrode is an electrical conductor used to make contact with a metallic part of a circuit (e.g. a semiconductor, an electrolyte or ionised gases). The word was coined by the scientist Michael Faraday from the Greek words elektron (meaning amber, from which the word electricity is derived) and hodos, a way. An electrode in an electrochemical cell is referred to as either an anode or a cathode, words that were also coined by Faraday. The anode is defined as the electrode at which electrons leave the cell and oxidation occurs, and the cathode as the electrode at which electrons enter the cell and reduction occurs. Each electrode may become either the anode or the cathode depending on the voltage applied to the cell. A bipolar electrode is an electrode that functions as the anode of one cell and the cathode of another cell. 1 Faculty of Physics, Al. I. Cuza University, Bd. Carol I, Nr. 11A, , Iasi Faculty of Biology, Al. I. Cuza University, Bd. Carol I, Nr. 0A, , Iasi 3 Faculty of Chemistry, Al. I. Cuza University, Bd. Carol I, Nr. 11A, , Iasi
2 ALINA MANOLE, I. NEACSU, M-O. APOSTU, V. MELNIG 6. THEORETICAL CONSIDERATIONS If an electrode of noble metal such as platinum or gold is made 0.6 to 0.8 V negative with respect to a suitable reference electrode such as Ag/AgCl or an calomel electrode in a neutral KCl solution (Fig. 1), the oxygen dissolved in the liquid is reduced at the surface of the noble metal following the reactions [3]: Cathodic reaction: O + H O + e - H O + OH H O + e - OH Anodic reaction: Ag + Cl - AgCl + e - Overall reaction: 4 Ag + O + H O + 4 Cl - 4 AgCl + 4 OH - Fig. 1 Setup for polarography. The oxygen reduction at the noble metal surface can follow two pathways: 4 electron pathway - the oxygen in the bulk diffuses to the surface of the cathode and is converted to OH - via H O (path a in Fig. ) electron pathway - the intermediate H O diffuses directly out of the cathode surface into the bulk liquid (path b in Fig. ). The oxygen reduction path may change depending on surface condition of the noble metal. This is probably the cause for time-dependent current drift of polarographic probes. Since the hydroxyl ions are constantly being substituted for chloride ions as the reaction starts, KCl or NaCl has to be used as the electrolyte. When the electrolyte is depleted of Cl -, it has to be replenished. Fig. Alternative pathways of oxygen reduction at cathode surface.
3 7 THEORETICAL AND PRACTICAL If we study the current-voltage diagram of the electrode (called a polarogram) we can observe that the current output of the electrode increases, in the beginning, with the increase of the negative voltage applied to the noble metal electrode (the cathode) and because the reaction of oxygen at the cathode is very fast and the rate of reaction is limited by the diffusion of oxygen to the cathode surface, the current becomes saturated (Fig. 3a). When the negative voltage is further increased, the current output of the electrode increases rapidly due to other reactions, mainly, the reduction of water to hydrogen. If a fixed voltage in the plateau region (for example, - 0.6V) is applied to the cathode, the current output of the electrode can be linearly calibrated to the dissolved oxygen (Fig. 3b). It has to be noted that the current is proportional not to the actual concentration but to the activity or equivalent partial pressure of dissolved oxygen, which is often referred to as oxygen tension. Originally developed for measuring oxygen gas, it is only a matter of polarity, whether the electrode senses hydrogen or oxygen gas. For hydrogen measurements +600 mv (vs. Ag/AgCl) are supplied. The electrode is calibrated using two gas mixtures of known oxygen (or hydrogen) concentration. (a) (b) Fig. 3 (a) Current-to-voltage diagram at different oxygen tensions; (b) Calibration obtained at a fixed polarization voltage of -0.6V. Sensor Output and Design Variables. One Layer Model A first-order diffusion model of the Clark electrode is illustrated in Fig. 4 []. The membrane electrolyte electrode system is considered to act as a one-dimensional diffusion system with the partial pressure at the membrane surface equal to the equilibrium partial pressure p 0 and that at the cathode equal to zero.
4 ALINA MANOLE, I. NEACSU, M-O. APOSTU, V. MELNIG 8 The following assumptions are made [4]: 1. The cathode is well polished and the membrane is tightly fit over the cathode surface such that the thickness of the electrolyte layer between the membrane and the cathode is negligible.. The liquid around the sensor is well agitated that the partial pressure of oxygen at the membrane surface is the same as that of the bulk liquid. 3. Oxygen diffusion occurs only in one direction, perpendicular to the cathode surface. Fick s nd law describes the unsteady-state diffusion in the membrane: p p = D m (1) t x D m - the oxygen diffusivity in the membrane; x - the distance from the cathode surface. The initial and boundary conditions are: p = 0 at t = 0 () p = 0 at x = 0 (3) p = p o at x = d m (4) d m - the membrane thickness; p o - the partial pressure of oxygen in the bulk liquid The solution of eq. (1) with the boundary conditions is: p x = + m m (5) p 0 d m n= 1 nπ d m The current output of the electrode is proportional to the oxygen flux at the cathode surface: I = NFADm ( C/ x) x= 0 (6) = NFAP ( p/ x) n nπx ( 1) sin exp( n π D t d ) m N - number of electrons per mole of oxygen reduced; F - Faraday constant (= 96,500 C/mol); A - surface area of the cathode; P m - oxygen permeability of the membrane The permeability P m is related to diffusivity by: P m = D msm (7) S m - the oxygen solubility of the membrane From eqs. (5) and (6), the current output I t of the electrode is: P = + m n I t NFA p 0 1 ( 1) exp( n π D mt d m ) (8) d m n=1 The pressure profile within the membrane and the current output under steady-state conditions can be obtained from eqs. (5) and (8), respectively: p d = (9) p and 0 d m p x= 0 m I s = NFA p 0 (10) d m
5 9 THEORETICAL AND PRACTICAL At steady state, the pressure profile is linear and the current output is proportional to the oxygen partial pressure in the bulk liquid. Eq. (10) forms the basis for oxygen diffusion measurement by the sensor. Eq. (8) shows that the rapidness of the sensor response is characterized by the response time parameter: d m τ = (11) D m For a rapid response (small τ) the membrane must be thin and/or diffusion coefficient, D m, high. Adjusting d m (rather than D m ) is more effective in adjusting τ (because τ depends on the square of d m ). Eqs. (10) and (11) indicate that the design variables for a diffusion oxygen sensor (DO sensor) are P m, d m, D m and A. Fig. 4 One layer electrode model. Fig. 5 Two layer electrodemodel Two Layer Model In reality, the assumption made earlier is not entirely satisfactory, because when a DO sensor is used to measure DO concentration in two different liquids at the same partial pressure of oxygen, the readings are not the same. This indicates that the sensor output depends, to a certain extent, on the properties of the liquid. The effect of liquid layer on sensor current output can be estimated using the two layer model witch considers the membrane and the liquid film as shown in Fig. 5. At steady state, the oxygen flux J through each layer in Fig. 5 should be the same: J = Kp 0 = k L ( p 0 p m ) (1) = k mp m K - the overall mass transfer coefficient; k L - mass transfer coefficient for liquid film; k m - mass transfer coefficient for membrane The inverse of the mass transfer coefficient can be termed as the mass transfer resistance:
6 ALINA MANOLE, I. NEACSU, M-O. APOSTU, V. MELNIG 30 1 K 1 1 = + (13) k L k m Equation (13) says that the overall mass transfer resistance, 1/K, is the sum of the liquid phase mass transfer resistance, 1/k L, and the membrane phase mass transfer resistance, 1/k m. The derivation is based on Ohm s law analogy; J is considered the current and p the voltage. The individual resistances can be replaced by: 1 d L d m = + (14) K PL Pm d L- liquid film thickness; P L - the oxygen permeability of the liquid film When the individual mass transfer resistances are considered the steady state sensor output becomes: p m I s = NFA p 0 (15) d where d is defined by: Alternatively, I s can be written as: d + I d L = d m Pm (16) PL p 0 s = NFA (17) d m Pm + d L PL The time constant τ of eq. (11) can be modified to: d τ = (18) D m where d is defined by: D m d = d m + d L (19) D L When placed in a stagnant liquid, the DO sensor produces a diffusion gradient extending outside the membrane and into the liquid. When the liquid is stirred, the diffusion gradient can no longer be extended beyond the liquid film around the membrane. Since the diffusion gradient becomes steeper with decreasing liquid film thickness, the current output of the sensor increases with increase in liquid velocity as shown in Fig. 6. Also, the response time of the sensor increases as the liquid velocity decreases. This so-called "flow sensitivity" is greater for a sensor with a larger cathode because the size of the stagnant diffusion field is proportionally with the size of the cathode. For proper operation of the sensor, the liquid has to be stirred beyond a certain level to maintain membrane control of oxygen diffusion. The critical velocity, V c, of the liquid is the velocity where the probe output reaches 95% (95-99% depending on the definition) of the steady state value. For a given liquid, V c is smaller for smaller cathodes. For example, with a 5 μm Teflon membrane, a cathode of 5 mm diameter requires V c of 70 cm/s in water, whereas only 5 cm/s is required for 5 μm
7 31 THEORETICAL AND PRACTICAL cathode. When the cathode diameter is less than 1 μm, the sensor becomes insensitive to liquid flow even without the membrane because, the diffusion field of the cathode is so small that it is always contained inside the minimum liquid boundary layer around the cathode. One obvious effect of the cathode size is the area effect. Eq. (17) shows that the current output is directly proportional to the sensor area A. When the current produced by the sensor is too small (when the cathode area is small), the sensor signal tends to be more susceptible to noise. The current output should be greater than 10-6 to 10-7 range for ease of signal amplification. From eq. (17), the condition for a membrane-controlled diffusion becomes: d m /Pm >> d L/PL (0) To achieve this condition, a relatively thick membrane with low oxygen permeability has to be used (this contradicts the requirement for a fast sensor response). In this case the oxygen sensor output depends only on membrane properties as given by eq. (10) and the sensor calibrated in one liquid can be used in other liquids without recalibration. In reality, however, there is always a liquid film (however thin it may be) and this causes variations in calibration in different liquids. Fig. 6 Effect of cathode diameter on flow dependency of DO sensor. Running experiments it has been observed that there is 1 to 5% increase in sensor output current per Celsius degree increase in temperature (Fig. 7a). From eq. (17) we can see that the temperature effect comes from P m and P L because they are functions of temperature. In the case of membrane diffusion control, the temperature dependency of I s should come entirely from P m and when the liquid film resistance is not negligible, both P m and P L contribute to the temperature dependency. Generally, P m is expressed as: Pm = Pm *exp(-e/rt) (1) E - the activation energy for the permeation (8.8 kcal/g-mol for polyethylene and kcal/g-mol for polypropylene membranes) [5].
8 ALINA MANOLE, I. NEACSU, M-O. APOSTU, V. MELNIG 3 (a) (b) Fig. 7 Temperature dependency of DO sensor. From Fig. 7b we can observe that as the temperature increases, the diffusivity increases and this makes the sensor respond faster. Calibration and k L a Measurement For liquid phase calibration an air saturated and also a nitrogen saturated water is prepared by passing air bubbles into a small volume (100 ml) of water, then the voltage output for both water solutions is measured. To obtain proper calibration the liquids have to be stirred at high speed. This is so-called a two point calibration. To perform gas phase calibration the sensor is exposed to air and also to nitrogen, as the gas phase, and then the two calibrations are compared [6]. For measuring the response time a step change in oxygen partial pressure in the measurement medium has to be done and then the sensor response is determined. The sensor can be approximated as a first order system: c c p = τ p (dc p / dt) () c - the oxygen concentration in the measurement sample; c p - the oxygen concentration measured by the sensor; τ p - the sensor time constant When a step change is made in c (by transferring the sensor from air into nitrogen saturated, stirred water), the sensor output decreases roughly exponentially (not exactly exponentially because the sensor may not be a true first order system). The time constant τ p is the time when the sensor response reaches 63.7% of the ultimate response (Fig. 8a). The solution to eq. () with the following boundary condition is an exponential function: c = 1 at t = 0 (3) c c p = 1 exp( t / τ p ) (4) A normalized concentration is used: c of 1 means 100% air saturation and 0 means nitrogen saturation. Eq. (4) indicates that when t = τ p, c/c p will be The time constant τ p can also be determined conveniently by using an integral method - the area above the response curve is equal to τ p (see Fig. 8b). This method is especially useful when there is a lot of noise in the measured signal.
9 33 THEORETICAL AND PRACTICAL Fig. 8 (a) Sensor response time measurement; (b) integral method for measuring the sensor time constant. Measurement of k L a k L a, the liquid phase overall volumetric mass transfer coefficient, represents the oxygen absorption capability of a bioreactor [4]. A DO sensor is used frequently to measure k L a. The reactor is first sparged with nitrogen and at time zero, the nitrogen is switched to air. The oxygen mass balance in the reactor yields: * dc dt = k La( c c) (5) c - the oxygen concentration in the reactor; c* - the oxygen concentration at the gasliquid interface. This equation can be rearranged to: * c c = τ k ( dc dt) (6) where τ k =1 k L a (7) Eqs. () and (6) can be solved simultaneously to obtain an expression for k L a and τ k can be obtained graphically as shown in Fig. 9 when τ p is known. Fig. 9 Measurement of k L a by integral method. It is very important to know that the magnitude of τ p depends on the liquid velocity in the vicinity of the sensor. Therefore, if a τ p measured at one agitation rate is used for measuring k L a for different agitation rates, the results will be in error. A safe
10 ALINA MANOLE, I. NEACSU, M-O. APOSTU, V. MELNIG 34 way is to used the same agitation rate for both τ p and τ k measurements. However, if τ k is much greater than τ p, such a precaution is not necessary. 3.CONCLUSIONS As the models reveals, there is the possibility of constructing an oxygen sensor, based on oxygen diffusion limited process (the Clark electrode), that consists of an anode and cathode in contact with an electrolyte solution (usually potassium chloride, 1 M). It is covered at the tip by a semi-permeable polymer membrane, which is permeable to gases but not contaminants and reducible ions of the sample. The cathode is in a glass envelope in the body of the electrode. The anode has a larger surface that provides stability and guards against drift due to concentration of the po electrolyte. This silver/silver chloride (Ag/AgCl) anode provides electrons for the cathode reaction. When most of the mass transfer resistance is confined in the membrane, the electrode system can measure oxygen tension in various liquids. This is the basic operating principle of the membrane covered polarographic DO probe (DO sensor) (Fig. 10) [7]. Fig. 10 A Clark-type oxygen-sensitive electrode. REFERENCES 1. T.C. Tan, and C.C. Liu, 1991, Principles and fabrication materials of electrochemical sensors. In: Chemical Sensor Technology. Kodansha Ltd., Vol. 3.. L.C. Clark, Monitor and control of blood and tissue oxygen tension. Trans. Am.Soc. Artif. Internal Org., Magda Aflori, I. Neacşu, V. Melnig, 006. Analysis methods of oxygen consumption in biological systems, Scientific Annals of University of Agricultural Sciences and Veterinary Medicine Ion Ionescu de la Brad IA I, Tom XLIX. 4. Young H. Lee, Biosensors, Engineering Biotechnology Gateway Project, Drexel University, 5. Park, J.H., N.O. Kaplan, E.P. Kennedy, 1966.Current Aspects of Biochemical Energetics, Academic Press, New York, David R. Caprette, Experimental Biosciences, Introductory Laboratory Bios 11, Rice University. 7. Trinity College Dublin, Biochemistry Laboratory: Manual for Senior Freshman Science.
METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES
METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES ELECTROCHEMISTRY Basic principle Electrodes are used to selectively measure particular ions Instruments utilizing electrodes measure the potential
More informationChapter 18 Electrochemistry. Electrochemical Cells
Chapter 18 Electrochemistry Chapter 18 1 Electrochemical Cells Electrochemical Cells are of two basic types: Galvanic Cells a spontaneous chemical reaction generates an electric current Electrolytic Cells
More informationELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY
ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY ELECTROCHEMISTY ELECTROCHEMISTRY IS THE STUDY OF CHEMICAL REACTIONS THAT RESULT IN THE FLOW OF ELECTRONS (CURRENT) OR THE DEVELOPMENT
More informationElectro Analytical Methods
CH 2252 Instrumental Methods of Analysis Unit II Electro Analytical Methods Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam
More informationUnit 2 B Voltammetry and Polarography
Unit 2 B Voltammetry and Polarography Voltammetric methods of Analysis What is Voltammetry? A time-dependent potential is applied to an electrochemical cell, and the current flowing through the cell is
More informationThermo Scientific Electrochemistry Products
Comparing Performance of DO Methods The world leader in serving science in Water and Seawaters Thermo Scientific Electrochemistry Products Dissolved Oxygen Measurement Oxygen is essential for most of the
More informationELECTROCHEMISTRY Chapter 14
ELECTROCHEMISTRY Chapter 14 Basic Concepts: Overview of Electrochemical Process at Constant T, P (14-1) ΔG = ΔG o + RT ln Q = w elec (maximum) = qe = ItE (exp) (E intensive parameter, q extensive) = nfe
More informationKentaro INOUE. Introduction. Measurement principle (membrane polarographic method)
FFeature Article Article Performance of the Dissolved Oxygen Monitor Used in the Semiconductor Wet Process; Low Concentration Monitoring, High Temperature, Small Amount of Sampling Volume, Chemical Resistance
More informationMembrane Electrodes. Several types
Membrane Electrodes Electrical connection Several types - Glass membrane electrode - Liquid membrane electrode - Solid State membrane electrode - Permeable membrane electrode seal 0.1 M HCl Filling solution
More informationIn all electrochemical methods, the rate of oxidation & reduction depend on: 1) rate & means by which soluble species reach electrode surface (mass
Voltammetry Methods based on an electrolytic cell Apply potential or current to electrochemical cell & concentrations change at electrode surface due to oxidation & reduction reactions Can have 2 or 3
More information3/24/11. Introduction! Electrogenic cell
March 2011 Introduction Electrogenic cell Electrode/electrolyte interface Electrical double layer Half-cell potential Polarization Electrode equivalent circuits Biopotential electrodes Body surface electrodes
More informationChapter 25. Voltammetry
Chapter 5. Voltammetry Excitation Signal in Voltammetry Voltammetric Instrumentation Hydrodynamic Voltammetry Cyclic Voltammetry Pulse Voltammetry High-Frequency and High-Speed Voltammetry Application
More informationInstrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013
0510212 Instrumental Chemical Analysis Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 Chapter 1 Electroanalytical Methods Electroanalytical Chemistry Electroanalytical
More informationANALYSIS METHODS OF OXYGEN CONSUMPTION IN BIOLOGICAL SYSTEMS
ANALELE ŞTIINŢIFICE ALE UNIVERSITĂŢII AL. I. CUZA IAŞI Tomul III, s. Biofizică, Fizică medicală şi Fizica mediului 2007 ANALYSIS METHODS OF OXYGEN CONSUMPTION IN BIOLOGICAL SYSTEMS Magda Aflori 1, I. Neacşu
More informationn. log a ox a red
Amperometry &Voltammetry Non-equilibrium electrochemistry! Based on Electrolytic Cells---apply external voltage to pair of electrodes to force rxn to occur--get current flow---current α [conc] o E elect
More informationChapter 3 Electrochemical methods of Analysis-Potentiometry
Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator
More informationElectrochemistry. Michael Faraday s law of electromagnetic induction says that whenever a conductor is
Surname 1 Name Course Instructor Date Electrochemistry 1. Faraday s Law Michael Faraday s law of electromagnetic induction says that whenever a conductor is positioned in a changeable magnetic field emf
More informationChem 321 Lecture 16 - Potentiometry 10/22/13
Student Learning Objectives Chem 321 Lecture 16 - Potentiometry 10/22/13 In lab you will use an ion-selective electrode to determine the amount of fluoride in an unknown solution. In this approach, as
More informationAGITATION AND AERATION
AGITATION AND AERATION Although in many aerobic cultures, gas sparging provides the method for both mixing and aeration - it is important that these two aspects of fermenter design be considered separately.
More informationChapter 14: Electrodes and Potentiometry
Yonsei University Chapter 14: Electrodes and Potentiometry The use of electrodes to measure voltages that provide chemical information is called potentiometry (ion-selective electrode, ion-sensing field
More informationUnit 2 Electrochemical methods of Analysis
Unit 2 Electrochemical methods of Analysis Recall from Freshman Chemistry: Oxidation: Loss of electrons or increase in the oxidation number Fe 2 e - Fe 3 Reduction: Gain of electrons or decreases in the
More informationBasic Concepts in Electrochemistry
Basic Concepts in Electrochemistry 1 Electrochemical Cell Electrons Current + - Voltage Source ANODE Current CATHODE 2 Fuel Cell Electrons (2 e) Current - + Electrical Load ANODE Current CATHODE H 2 2H
More informationChem 321 Lecture 17 - Potentiometry 10/24/13
Student Learning Objectives Chem 321 Lecture 17 - Potentiometry 10/24/13 Electrodes The cell described in the potentiometric chloride titration (see 10/22/13 posting) consists of a Ag/AgCl reference electrode
More informationELECTROCHEMISTRY Chapter 19, 4.9
ELECTROCHEMISTRY Chapter 19, 4.9 Overview of an Electrochemical Process at Constant T and P ΔG = ΔG o + RT ln Q = welec (maximum) Note: I below stands for current measured in amperes = qecell = ItEcell
More informationMass transfer by migration & diffusion (Ch. 4)
Mass transfer by migration & diffusion (Ch. 4) Mass transfer equation Migration Mixed migration & diffusion near an electrode Mass transfer during electrolysis Effect of excess electrolyte Diffusion Microscopic
More informationChemical Quantities in Blood. 11. Biochemistry-1 Electrochemical Measurements of Biochemical Quantities. Hematocrit
11. Biochemistry-1 Electrochemical Measurements of Biochemical Quantities Chemical Quantities in Blood Living cells are chemical factories where the input is metabolic food and the output is waste products.
More informationCyclic Voltammetry. Fundamentals of cyclic voltammetry
Cyclic Voltammetry Cyclic voltammetry is often the first experiment performed in an electrochemical study of a compound, biological material, or an electrode surface. The effectiveness of cv results from
More informationCHM 213 (INORGANIC CHEMISTRY): Applications of Standard Reduction Potentials. Compiled by. Dr. A.O. Oladebeye
CHM 213 (INORGANIC CHEMISTRY): Applications of Standard Reduction Potentials Compiled by Dr. A.O. Oladebeye Department of Chemistry University of Medical Sciences, Ondo, Nigeria Electrochemical Cell Electrochemical
More informationCh. 13 Fundamentals of Electrochemistry
Ch. 13 Fundamentals of Electrochemistry 13.1 13-1. Basic Concepts of electrochemistry redox reaction : reactions with electron transfer oxidized : loses electrons reduced : gains electrons Fe 3+ + V 2+
More informationCHEM Pharmacy Week 9: Nernst Equation. Dr. Siegbert Schmid School of Chemistry, Rm 223 Phone:
CHEM1612 - Pharmacy Week 9: Nernst Equation Dr. Siegbert Schmid School of Chemistry, Rm 223 Phone: 9351 4196 E-mail: siegbert.schmid@sydney.edu.au Unless otherwise stated, all images in this file have
More informationCombination ph, ORP, T Electrode with Transmitter
Combination ph, ORP, T Electrode with Transmitter Technical Information and Manual 2014 Osorno Enterprises Inc. The content of this manual is protected by copyright. All rights reserved. OCS/1406 Combination
More informationChemistry Instrumental Analysis Lecture 22. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 22 Measures potential under very low currents. The cell is 2 half cells. Consist of a reference electrode, indicator electrode, and potential measuring device.
More informationReference electrode. Calomel electrode Hg in contact with Hg(I) chloride Ag/AgCl 15-2
Potentiometry Potential measurements of electrochemical cells Ion selective methods Reference electrode Indicator electrode Potential measuring device Reference electrode Indicator electrodes Ion specific
More informationModeling as a tool for understanding the MEA. Henrik Ekström Utö Summer School, June 22 nd 2010
Modeling as a tool for understanding the MEA Henrik Ekström Utö Summer School, June 22 nd 2010 COMSOL Multiphysics and Electrochemistry Modeling The software is based on the finite element method A number
More informationELECTROCHEMISTRY. these are systems involving oxidation or reduction there are several types METALS IN CONTACT WITH SOLUTIONS OF THEIR IONS
Electrochemistry 1 ELECTROCHEMISTRY REDOX Reduction gain of electrons Cu 2+ (aq) + 2e > Cu(s) Oxidation removal of electrons Zn(s) > Zn 2+ (aq) + 2e HALF CELLS these are systems involving oxidation or
More information470 Lecture #7 of 18
Lecture #7 of 18 470 471 Q: What s in this set of lectures? A: Introduction, Review, and B&F Chapter 1, 15 & 4 main concepts: Section 1.1: Redox reactions Chapter 15: Electrochemical instrumentation Section
More informationMASS TRANSFER COEFFICIENTS DURING AERATION BY A SELF-ASPIRATING IMPELLER
th European Conference on Mixing Warszawa, - September MASS TRANSFER COEFFICIENTS DURING AERATION BY A SELF-ASPIRATING IMPELLER Czesław Kuncewicz, Jacek Stelmach Lodz University of Technology, Department
More informationElectrolysis and Faraday's laws of Electrolysis
Electrolysis and Faraday's laws of Electrolysis Electrolysis is defined as the passage of an electric current through an electrolyte with subsequent migration of positively and negatively charged ions
More informationAnalysis of cations and anions by Ion- Selective Electrodes (ISEs)
Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Purpose: The purpose of this assignment is to introduce potentiometric measurements of ionic species by ion selective electrodes (ISEs)
More informationChemistry 1011 TOPIC TEXT REFERENCE. Electrochemistry. Masterton and Hurley Chapter 18. Chemistry 1011 Slot 5 1
Chemistry 1011 TOPIC Electrochemistry TEXT REFERENCE Masterton and Hurley Chapter 18 Chemistry 1011 Slot 5 1 18.5 Electrolytic Cells YOU ARE EXPECTED TO BE ABLE TO: Construct a labelled diagram to show
More informationElectro Chemistry Part-II 1. Faraday s laws of electrolysis are related to the 1) Molar mass of the electrolyte 2) Equivalent weight of the cation or anion 3) Molecular mass of the electrolyte 4) Atomic
More informationTopic: APPLIED ELECTROCHEMISTRY. Q.1 What is polarization? Explain the various type of polarization.
Topic: APPLIED ELECTROCHEMISTRY T.Y.B.Sc Q.1 What is polarization? Explain the various type of polarization. Ans. The phenomenon of reverse e.m.f. brought about by the presence of product of electrolysis
More informationKJ 2051 Coulometric titration of hydrochloric acid
KJ 2051 Page 1 of 4 F.G. Banica KJ 2051 Coulometric titration of hydrochloric acid By F.G. Banica, 22 September 2006 This document includes a short theory of this experiment. See Ref. [1] for experimental
More informationSpecific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes
Turk J Chem 24 (2000), 95 99 c TÜBİTAK Specific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes Nil ERTAŞ Ege University, Faculty of Science, Department
More informationIntroduction of Electrode Processes (Ch. 1)
Introduction of Electrode Processes (Ch. 1) Introduction Electrochemical cells & reactions: thermodynamics and potentials Nonfaradaic processes & electrode-solution interface Capacitance and charge of
More informationProf. Mario L. Ferrari
Sustainable Energy Mod.1: Fuel Cells & Distributed Generation Systems Dr. Ing. Mario L. Ferrari Thermochemical Power Group (TPG) - DiMSET University of Genoa, Italy Lesson II Lesson II: fuel cells (electrochemistry)
More informationElectrochemistry. Redox reactions. Half Reactions. Nernst Equation Ion selective electrodes
Electrochemistry Nernst Equation Ion selective electrodes Redox reactions oxidation - loss of electrons M n+ M n+1 + e - M is oxidized - reducing agent reduction - gain of electrons N n+ + e - N n-1 N
More informationCurrent based methods
Current based methods Amperometric and voltammetric sensors More significant influence on analytical parameters (sensitivity, selectivity, interferences elimination) kind of method, potential range, electrode
More informationChapter 2. Materials and Methods
Chapter 2 Materials and Methods 2. Materials and Methods This chapter describes the chemicals, reagents and instruments used for carrying out this study. A brief discussion of the methods used for the
More informationMEASUREMENT OF ph: INTRODUCTION TO BUFFER AND BASICS OF ph Meter
Theory module: 04 MEASUREMENT OF ph: INTRODUCTION TO BUFFER AND BASICS OF ph Meter Introduction ph is a measure of the relative amount of hydrogen and hydroxide ions in an aqueous solution. In any collection
More informationUniversity of Technology Corrosion Engineering Lecturer: Basheer Ahmed Chemical Engineering Dept. 4 th Class
Example 1 Determine the corrosion rate of carbon steel in salt solution from the following laboratory data. Consider the corrosion rate by a)mpy, b)mdd. C) Calculate the current density in μa/cm 2 Δ W
More information17.1 Redox Chemistry Revisited
Chapter Outline 17.1 Redox Chemistry Revisited 17.2 Electrochemical Cells 17.3 Standard Potentials 17.4 Chemical Energy and Electrical Work 17.5 A Reference Point: The Standard Hydrogen Electrode 17.6
More informationElectrochemistry. Goal: Understand basic electrochemical reactions. Half Cell Reactions Nernst Equation Pourbaix Diagrams.
Electrochemistry Goal: Understand basic electrochemical reactions Concepts: Electrochemical Cell Half Cell Reactions Nernst Equation Pourbaix Diagrams Homework: Applications Battery potential calculation
More informationBatteries (Electrochemical Power Sources)
Batteries (Electrochemical Power Sources) 1. Primary (single-discharge) batteries. => finite quantity of the reactants 2. Secondary or rechargeable batteries => regeneration of the original reactants by
More information13 AN ELECTROCHEMICAL METHOD FOR ACCELERATED TESTING OF CHLORIDE DIFFUSIVITY IN CONCRETE
13 AN ELECTROCHEMICAL METHOD FOR ACCELERATED TESTING OF CHLORIDE DIFFUSIVITY IN CONCRETE TIEWEI. ZHANG and ODD E. GJ@RV Division of Building Materials, The Norwegian Institute of Technology - NTH N - 7034
More informationFundamental molecular electrochemistry - potential sweep voltammetry
Fundamental molecular electrochemistry - potential sweep voltammetry Potential (aka voltammetric) sweep methods are the most common electrochemical methods in use by chemists today They provide an efficient
More informationName AP CHEM / / Collected Essays Chapter 17
Name AP CHEM / / Collected Essays Chapter 17 1980 - #2 M(s) + Cu 2+ (aq) M 2+ (aq) + Cu(s) For the reaction above, E = 0.740 volt at 25 C. (a) Determine the standard electrode potential for the reaction
More informationANALYSIS OF LEAD IN SEAWATER
ANALYSIS OF LEAD IN SEAWATER BY DIFFERENTIAL PULSE POLAROGRAPHY Introduction Electrochemical methods of analysis can be used for the quantitative analysis of any electroactive species any species that
More informationPOTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background
POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA Background In this experiment, students will familiarize themselves with potentiometric titration, practice using the first derivative to find the equivalence
More information#13 Electrochemical Cells
#13 Electrochemical Cells If a copper strip is placed in a solution of copper ions, one of the following reactions may occur: Cu 2+ + 2e - Cu Cu Cu 2+ + 2e - The electrical potential that would be developed
More informationChapter 7 Electrochemistry
Chapter 7 Electrochemistry Outside class reading Levine: pp. 417 14.4 Galvanic cells: pp. 423 14.5 types of reversible electrodes 7.6.1 Basic concepts of electrochemical apparatus (1) Electrochemical apparatus
More informationChapter 24. Electrogravimetry and Coulometry
Chapter 24 Electrogravimetry and Coulometry Dynamic Electrochemical Methods of analysis Electrolysis Electrogravimetric and Coulometric Methods For a cell to do any useful work or for an electrolysis to
More informationGas transfer rates If either phase concentration can not be predicted by Henry's law then there will be a transfer of mass across the interface until
Gas transfer rates If either phase concentration can not be predicted by Henry's law then there will be a transfer of mass across the interface until equilibrium is reached. The mechanisms and rate expressions
More informationEVALUATION OF Ag/AgCl SENSORS FOR IN-SITU MONITORING OF FREE CHLORIDE CONCENTRATION IN REINFORCED CONCRETE STRUCTURES
EVALUATION OF Ag/AgCl SENSORS FOR IN-SITU MONITORING OF FREE CHLORIDE CONCENTRATION IN REINFORCED CONCRETE STRUCTURES Farhad Pargar, Dessi Koleva, Oguzhan Copuroglu, Eduard Koenders, Klaas vab Breugel
More informatione - Galvanic Cell 1. Voltage Sources 1.1 Polymer Electrolyte Membrane (PEM) Fuel Cell
Galvanic cells convert different forms of energy (chemical fuel, sunlight, mechanical pressure, etc.) into electrical energy and heat. In this lecture, we are interested in some examples of galvanic cells.
More informationREADING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.
CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE READING Skoog, Holler and Crouch: Chapter 23 and Appendix 3. A. INTRODUCTION Potentiometry is a static electroanalytical method in which the potential
More informationMeasurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste
Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major
More informationAP Questions: Electrochemistry
AP Questions: Electrochemistry I 2 + 2 S 2O 2-3 2 I - + S 4O 2-6 How many moles of I 2 was produced during the electrolysis? The hydrogen gas produced at the cathode during the electrolysis was collected
More informationChapter 22. Bulk Electrolysis: Electrogravimetry and Coulometry. Definition. Features of Bulk Electrolysis Cells
Chapter 22 Bulk Electrolysis: Electrogravimetry and Coulometry Definition Bulk Electrolysis deals with methods that involve electrolysis producing a quantitative change in oxidation state Example: In a
More informationOxidation state. Electrochemical Techniques OCN Nov. 25, Redox chemistry refresher. Intro to electrochemistry. Electrochemical techniques
Electrochemical Techniques OCN 633 - Nov. 25, 2013 Brian Glazer glazer@hawaii.edu Redox chemistry refresher Life on Earth is comprised of e - transfer reactions Intro to electrochemistry voltaic cells,
More informationELECTROCHEMISTRY. these are systems involving oxidation or reduction there are several types METALS IN CONTACT WITH SOLUTIONS OF THEIR IONS
Electrochemistry 1 ELECTROCHEMISTRY REDOX Reduction gain of electrons Cu 2+ (aq) + 2e > Cu(s) Oxidation removal of electrons Zn(s) > Zn 2+ (aq) + 2e HALF CELLS these are systems involving oxidation or
More informationThe design and operational theory of ph electrodes is a very complex subject, explored only briefly here. What is important to understand is that thes
ph measurement A very important measurement in many liquid chemical processes (industrial, pharmaceutical, manufacturing, food production, etc.) is that of ph: the measurement of hydrogen ion concentration
More information9.1 Introduction to Oxidation and Reduction
9.1 Introduction to Oxidation and Reduction 9.1.1 - Define oxidation and reduction in terms of electron loss and gain Oxidation The loss of electrons from a substance. This may happen through the gain
More informationElectrochemistry SYBSc 2017
Electrochemistry SYBSc 2017 Definition It is a branch in chemistry which deals with the qualitative and quantitative studies of chemical changes brought about by the passage of electricity. It is also
More information(i) Voltameter consist of a vessel, two electrodes and electrolytic solution.
Electrochemistry is the branch of physical chemistry which deals with the relationship between electrical energy and chemical changes taking place in redox reactions i.e., how chemical energy produced
More informationReview of temperature distribution in cathode of PEMFC
Project Report 2008 MVK 160 Heat and Mass Transport May 08, 2008, Lund, Sweden Review of temperature distribution in cathode of PEMFC Munir Ahmed Khan Department of Energy Sciences, Lund Institute of Technology,
More informationGENERAL INSTRUCTIONS GENERAL PREPARATION
GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required
More informationEquilibrium electrochemistry
Equilibrium electrochemistry The principles of thermodynamics can be applied to solutions of electrolytes. For that we need to take into account activity coefficients: they differ significantly from 1
More informationEXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION
EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION I. INTRODUCTION This experiment introduces the direct-measurement approach to potentiometry. Principal purposes of the study are a) to understand quantitative
More informationWhere does Physical Chemistry fit into your course in Dentistry?
Where does Physical Chemistry fit into your course in Dentistry? Acidogenic bacteria in dental plaque can rapidly metabolise certain carbohydrates to acid endproducts. In the mouth, the resultant change
More informationSUPPLEMENTARY INFORMATION
doi:10.1038/nature17653 Supplementary Methods Electronic transport mechanism in H-SNO In pristine RNO, pronounced electron-phonon interaction results in polaron formation that dominates the electronic
More informationModeling of Liquid Water Distribution at Cathode Gas Flow Channels in Proton Exchange Membrane Fuel Cell - PEMFC
Modeling of Liquid Water Distribution at Cathode Gas Flow Channels in Proton Exchange Membrane Fuel Cell - PEMFC Sandro Skoda 1*, Eric Robalinho 2, André L. R. Paulino 1, Edgar F. Cunha 1, Marcelo Linardi
More informationContents. 2. Fluids. 1. Introduction
Contents 1. Introduction 2. Fluids 3. Physics of Microfluidic Systems 4. Microfabrication Technologies 5. Flow Control 6. Micropumps 7. Sensors 8. Ink-Jet Technology 9. Liquid Handling 10.Microarrays 11.Microreactors
More information4.07 Ions in Aqueous Solution (5 points)
2nd/3rd year Physical Chemistry Practical Course, Oxford University 4.07 Ions in Aqueous Solution (5 points) Outline In this experiment, you will (i) determine molar conductivities of electrolyte solutions
More informationLecture 4. Conductance sensors. ChemFET. Electrochemical Impedance Spectroscopy. py Practical consideration for electrochemical biosensors.
Lecture 4 Conductance sensors. ChemFET. Electrochemical Impedance Spectroscopy. py Practical consideration for electrochemical biosensors. Conductivity I V = I R=, L - conductance L= κa/, l Λ= κ /[ C]
More informationCHAPTER 6 Modern Theory Principles LECTURER SAHEB M. MAHDI
CHAPTER 6 Modern Theory Principles LECTURER SAHEB M. MAHDI Modern Theory principles in Corrosion and their applications :- Corrosion studies can be carried-out by two methods 1 Thermodynamics. or 2 By
More informationlect 26:Electrolytic Cells
lect 26:Electrolytic Cells Voltaic cells are driven by a spontaneous chemical reaction that produces an electric current through an outside circuit. These cells are important because they are the basis
More informationLab #14: Electrochemical Cells
Lab #14: Electrochemical Cells Objectives: 1. To understand the nature of electrochemical cells. 2. To construct a table listing the reduction potentials of a series of metal ions, in order of ease of
More informationSynthesis and Characterization of Proton-Conducting Oxides as Hydrogen Transport Membranes
Synthesis and Characterization of Proton-Conducting Oxides as ydrogen Transport Membranes Lin Li and Enrique Iglesia Department of Chemical Engineering, University of California at Berkeley, Division of
More informationElectroanalytical Chemistry techniques covered to date
Electroanalytical Chemistry techniques covered to date Potentiometry based on galvanic cell Controlled Potential Electrolysis electrolytic Chronoamperometry electrolytic cell Chronopotentiometry electrolytic
More informationELECTROCHEMISTRY (SUMMER VACATION WORKSHEET)
ELECTROCHEMISTRY (SUMMER VACATION WORKSHEET) QUESTION CARRYING 1 MARK 1. What is the effect of temperature on molar conductivity? 2. Why is it not possible to measure single electrode potential? 3. Name
More informationComponents of output signal in Chronoamperometry
Chronoamperometry Stationary electrode Unstirred = mass transport by diffusion Constant potential Measure current vs time Theory assume Ox + n e - Red - both Ox and Red are soluble - reversible reaction
More informationChapter 9: Polyprotic Acids and Bases
Amino acids Isoionic point Isoelectronic point 9-6 Isoelectric & Isoionic ph 9-1 Diprotic A&B Leucine: acidic & basic forms Leucine: intermediate form Monoprotic Polyprotic 9-5 Fractional compositions
More informationVoltammetry. Voltammetry and Polarograph. Chapter 23. Polarographic curves -- Voltammograms
Chapter 23 Voltammetry Voltammetry and Polarograph Electrochemistry techniques based on current (i) measurement as function of voltage (E appl ) Voltammetry Usually when the working electrode is solid,
More informationElectro Chemical Comparative Studies Of Ortho, Meta and Para Nitro Phenols
International journal of Emerging Trends in Science and Technology Electro Chemical Comparative Studies Of Ortho, Meta and Para Nitro Phenols Authors M. S. Selvakumar 1*, Dr. N. Xavier 2, Dr. V. Jeyabal
More informationUnit - 3 ELECTROCHEMISTRY VSA QUESTIONS (1 - MARK QUESTIONS) 3. Mention the purpose of salt-bridge placed between two half-cells of a galvanic cell?
Unit - 3 ELECTROCHEMISTRY 1. What is a galvanic cell? VSA QUESTIONS (1 - MARK QUESTIONS) 2. Give the cell representation for Daniell Cell. 3. Mention the purpose of salt-bridge placed between two half-cells
More informationVI. EIS STUDIES LEAD NANOPOWDER
VI. EIS STUDIES LEAD NANOPOWDER 74 26. EIS Studies of Pb nanospheres Impedance (valid for both DC and AC), a complex resistance occurs when current flows through a circuit (composed of various resistors,
More informationSecondary Ag/AgCl Pseudo-Reference Electrode on Silicon Substrate
Secondary Ag/AgCl Pseudo-Reference Electrode on Silicon Substrate F. L. Almeida a, M. B. A. Fontes b, C. Jimenez c and I. Burdallo c a Laboratório de Sistema Integráveis LSI, University of São Paulo, São
More information8. ELECTROCHEMICAL CELLS. n Electrode Reactions and Electrode Potentials a. H 2 2H + + 2e. Cl 2 + 2e 2Cl. H 2 + Cl 2 2H + + 2Cl ; z = 2
8. ELECTROCHEMICAL CELLS n Electrode Reactions and Electrode Potentials 8.1. a. H H + + e Cl + e Cl H + Cl H + + Cl ; z = E = E RT F ln ( a H +a Cl ) b. Hg(l)+ Cl Hg Cl + e H + + e H Hg + H + + Cl Hg Cl
More informationBAE 820 Physical Principles of Environmental Systems
BAE 820 Physical Principles of Environmental Systems Inter phase mass transfer Dr. Zifei Liu Mass transfer between two phases For a two phase system not at equilibrium (concentrations in the two phases
More information