READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.

Size: px
Start display at page:

Download "READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3."

Transcription

1 CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE READING Skoog, Holler and Crouch: Chapter 23 and Appendix 3. A. INTRODUCTION Potentiometry is a static electroanalytical method in which the potential of an indicating electrode in contact with a solution at equilibrium is measured in the absence of current flow across the electrochemical cell. The potential measured depends on the activity of ions according to Nernst equation. The platinum electrode used in CHE322L last semester, is an example of an electrode that is commonly used for potentiometric measurements. The potential of the Pt electrode is dependent on the concentration of the oxidized and reduced species of a redox couple according to the Nernst Equation. The Pt electrode is a non-selective electrode because it responds to the potential developed by many redox couples. Electrodes that respond selectively to the activity of a specific ion are commonly referred to as ion-selective electrodes. The first electrode of this type was the glass electrode, which responds selectively to the hydronium ion. The success of the glass ph electrode was followed by the development of glass electrodes of slightly different composition responding selectively to Na + and to K +. Since then, several ion-selective electrodes have been developed that respond selectively to a wide variety of ions, including F -, Cl -, I -, S 2-, ClO 4 -, NO 3 -, SO 4 2-, PO 4 3 -, Ca 2+, Pb 2+, Mg 2+, and Fe 2+. These electrodes all consist of an internal reference electrode separated from the test solution by a membrane which acts reversibly to a specific ion or group of ions. Three types of membranes are used (see next page from C&E News for schematic representations): glass, solid-state and liquid membrane electrode. In a glass electrode, a partially hydrated glass membrane separates the test solution from an inner solution. The internal reference electrode solution contains, in addition to the ions which determine the potential of the reference electrode, a constant activity of the ion of interest. In solid-state electrodes, the internal reference solution is separated from the test solution by a single crystal, or polycrystalline, insoluble salt, e.g. a silver halide. The salt acts as an ionic conductor for specific ions, and the potential developed across it depends on the activity of the specific ion in the sample solution according to the Nernst Equation. In a liquid membrane electrode, the membrane consists of a polymer for structural support, impregnated with a solvent immiscible in water. An ion carrier dissolved in the immiscible solvent gives the electrode its specificity. The K + selective electrode examined in this experiment is composed of polyvinyl-chloride (PVC), with bis(2-ethylhexyl)adipate as solvent, and valinomycin as the ion carrier. 1

2 CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE 2

3 CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE One of the limiting features of ion selective electrodes is that they respond to a number of ions, termed interferents, in addition to the ion of interest. For instance, the K + selective glass electrode responds more strongly to H + and Ag + than to K +, and is only moderately more selective for K + over Na +. The major advantage of the liquid membrane K + electrode, based on polyvinylchloride, bis(2-ethylhexyl)adipate, and valinomycin, is the high selectivity for K + over Na +, Ag +, and H +. Ion-selective electrodes measure the activity, and not the concentration, of the specific ion. The activity of a specific ion, ai, is related to its concentration, Ci, by ai = γ ici where γ i is the activity coefficient of the specific ion. The activity coefficient depends on the ionic strength, µ, of the solution according to the Debye-Hückel equation: logγ i = 0.5z 2 i µ where z i is the charge on the specific ion. This equation is reasonably accurate up to ionic strengths of about 0.1 M for 1:1 electrolytes. The ionic strength depends on the concentrations and charges of all the ions in the solution according to: µ 1 2 = zi 2 Ci At low ionic strengths, activities approach concentrations. At high ionic strengths, activities can be substantially less than concentrations. In such cases, concentrations can only be determined from ion-selective electrode measurements by accounting for the ionic strength effect. This can be done by: 1) preparing a calibration curve of electrode potential vs ion concentration in a medium of ionic strength comparable to that of the test solution; or 2) calculating the concentration from the measured activity using the above equations if the concentrations of other ionic species are known. The ion-selective electrodes respond to the activity of the specific ion according to the Nernst equation, which for the K + selective electrode is: E 2.3RT F = Ec + log a K (1) where E is the measured potential of the K + electrode relative to a standard reference electrode and a K is the potassium ion activity in the sample solution. E C is the standard cell potential; that portion of the K + electrode potential due to the internal reference electrode, internal solutions, and the specific membrane. "2.3 RT/F" is the Nernst factor 3

4 CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE (59.16 mv at 25 C for a ±1 ion). R and F are the gas and Faraday constants respectively, and T is the temperature in Kelvin. Because E C is different for each electrode membrane, a calibration curve of electrode potential relative to the potential of a reference electrode vs ion concentration or activity must be prepared. The response of ion selective electrodes is linear in log a i over only a finite range. It is necessary to prepare a calibration curve to determine this range, the value of E C, and the slope of the linear response curve. The response is termed "Nernstian" if the slope is equal to 2.3 RT/nF, or mv at 25 C for a ±1 ion. However, in many cases the slope is less than that predicted by the Nernst equation by 5 to 10 mv. If the response of a K + electrode is linear in log [K + ] at the concentration of a sample solution, it is possible to use Gran s method of multiple standard additions to determine the unknown concentration (1). This technique can be used to minimize matrix effects such as varying ionic strength, or the presence of interfering ions. To apply Gran s method, the observed parameter must be proportional to concentration. Rearranging equation 1 and taking the antilog gives equation 2, 10 E S E S + C = γ 10 [ K ] (2) added where S is the actual slope determined for the electrode in calibration solutions and γ is E S the activity coefficient in the unknown solution. A plot of 10 versus added [K + ] should give a straight line with a concentration axis intercept of -[K + ] unknown. B. EXPERIMENT SUMMARY In this experiment 1) potassium will be determined in diluted orange juice and tap water. 2) the effect of ionic strength on potentiometric measurements will be studied 4

5 POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE C. EXPERIMENTAL PROCEDURE Solutions 1) 1.00 M NaCl (1L) 2) M KCl (1L) 3) M KCl for storing electrode 4) Orange juice (20 ml per group) C.1 Electrode care The electrode used in this experiment an Orion Model K + selective electrode and a Ag/AgCl reference electrode. The potential developed between the electrodes is measured with an Orion Model 370 ph/mv meter. When not in use, the electrodes are stored in 0.01 M KCl. Prior to making any measurements, uncap the filling hole of the reference electrode and add the filling solution. Note that the filling solution for the Ag/AgCl electrode is 0.1 M NaCl. The K + selective electrode should not be exposed to organic solvents or to aqueous solutions containing organic solvents. C.2 Making measurements With the meter in STANDBY mode, raise the electrodes and rinse them with distilled water. Wipe down the sides and gently blot the ends of the electrodes with Kimwipes. Be careful. Vigorous blotting on the end can ruin the K + electrode. Place a beaker containing the solution to be measured along with a stirring bar on the magnetic stirrer. Set it to stir at a slow speed. Immerse the electrodes in the solution. Only the ends need to be immersed; avoid colliding with the stirring bar. Switch the meter to mv mode and record the reading once it has stabilized. (This should take less than one minute for [K + ] 10-5 M. At lower concentrations it may take longer) 5

6 POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE After the measurement has been made, switch the meter to STANDBY and rinse the electrodes. Immerse the electrodes in the next test solution or in the 0.01 M KCl storage solution. Do not leave the electrodes in air for longer than 2-3 minutes. C.3 Calibration of the K + Ion Selective Electrode The potassium selective electrode will be calibrated over its useful range by successive 10-fold dilutions of 0.1 M KCl. To maintain a constant ionic strength, 0.1 M NaCl is used to dilute. Begin by preparing 1 L of 0.1 M NaCl by a suitable dilution of the stock NaCl solution provided. Following the procedure outlined above, measure the potential of the electrode in a 20- to 30-mL portion of the 0.1 M KCl solution provided. While waiting for the reading to stabilize, prepare a 0.01 M KCl/0.09 M NaCl solution by pipetting 10 ml of 0.1 M KCI into a 100-mL volumetric flask, and making up to the mark with 0.1 M NaCl. Once the 0.01 M KCl solution is prepared, record the reading for the 0.1 M KCl solution in the table below. Rinse and transfer the electrodes to the 0.01 M KCl solution, and prepare a 10-3 M KCl solution by pipetting 10 ml of the 10-2 M KCl/0.09 M NaCl solution into a 100-mL volumetric flask. Dilute to volume with 0.1 M NaCl. When the 10-3 M K + solution is prepared, record the reading for the 10-2 M KC1 solution. Continue in this fashion until [KC1] = 1 x 10-7 M. Calibration Data for K + ISE [K + ], M E(mV) 1 x x x x x x x

7 POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE C.4 Determination of Potassium in Orange Juice Prepare a 1:10 dilution of orange juice by diluting 10-mL of orange juice with 0.1 M NaCl. With a pipet measure a 25-mL aliquot of the 1:10 dilution and measure its potential. Record the value in the table below. If the electrode does not stabilize, record several values and use the average. C.5 Standard Addition Procedure While monitoring the electrode potential in the same 25-mL aliquot, use a pipet to deliver 2 ml of 0.1 M KCl into the solution. Record the new value of the potential. Repeat this procedure until the total volume is 31 ml. (Again record several values of E after each addition if stability is a problem.) Standard Addition Calibration for K + in Diluted Orange Juice Solution Diluted Orange Juice Std. Add. 1 Std. Add. 2 Std. Add. 3 E (mv) C.6 Determination of K + in Tap Water Deliver a 50 ml sample tap water into a beaker. Insert the rinsed electrodes, and record the stabilized potential. Deliver 1 ml of 0.1 M NaCl by pipet to increase the ionic strength and re-record the potential. Then deliver 10 ml of 0.1 M NaCl, and record the potential when stabilized. Effect of Ionic Strength on Potentiometric Response for K + Solution Tap Water 1 ml NaCl 10 ml NaCl E (mv) 7

8 POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE D. DATA ANALYSIS AND QUESTIONS 1. For the data acquired in Part C.3.1, make a plot of electrode potential versus log [K + ]. Over what region of this plot is the response linear? Report your value for the slope, S. 2. Calculate the ionic strength for the 1 x 10-7 M KCl solution. Using Table a2-1 on page 996 in your text, estimate the activity coefficient of K +. What is the activity of K + for this solution? 3. Discuss why a plot of E vs log[k + ] conforms to equation 1, and a plot of E vs log a K is not necessary. 4. Using the calibration curve from question 1 and the potential measured for the diluted orange juice, determine [K + ] in undiluted orange juice. 5. Make a Gran s plot, as discussed in the Introduction, for the standard additions of KCl to orange juice. Evaluate the concentration in undiluted orange juice using this method. Discuss and compare the calibration curve and standard addition methods for evaluating ion concentrations. 6. Based on your data, and using the calibration curve method, make an estimate of [K + ] in tap water. Is your value meaningful? Why or why not? How did the apparent value of [K + ] change with the addition of NaCl? Is this what you would expect to happen? Explain. F. REFERENCES 1. G.D. Christian, J.E. O Reilly, Instrumental Analysis, 2nd Ed., Chapter K. Srinivasan, G.A. Rechnitz, Selective Studies on Liquid Membrane Ion- Selective Electrodes, Anal. Chem. 41, 1203 (1969). 3. G.J. Moody, J.D.R. Thomas, Selective Ion Sensitive Electrodes, Chapter 2, Merrow, London, D.A. Skoog, F.J. Holler and S.R. Crouch, Principle of Instrumental Analysis, 6 th Ed., Chapter 23 (2007) 8

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.

More information

Unit 2 Electrochemical methods of Analysis

Unit 2 Electrochemical methods of Analysis Unit 2 Electrochemical methods of Analysis Recall from Freshman Chemistry: Oxidation: Loss of electrons or increase in the oxidation number Fe 2 e - Fe 3 Reduction: Gain of electrons or decreases in the

More information

Fig. Electrochemical Cell/ Potentiometric Titration

Fig. Electrochemical Cell/ Potentiometric Titration Fig. Electrochemical Cell/ Potentiometric Titration The accurate, precise and effective potentiometric measurements can be made with the help of the following two types of electrodes namely : REFERENCE

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

Electrochemistry LEC 06. The Nernst equation. What you need: What you can learn about. Principle and tasks

Electrochemistry LEC 06. The Nernst equation. What you need: What you can learn about. Principle and tasks Electrochemistry LEC 06 What you can learn about Electrode potentials and their concentration dependence Redox electrodes Electrochemical cells What you need: Principle and tasks expresses how the electrical

More information

Operating Instructions for Ammonium ISE Specifications

Operating Instructions for Ammonium ISE Specifications Operating Instructions for Ammonium ISE Specifications Range: The Ammonium Electrode responds to uncomplexed ion activity over the range 1 x 10-1M to 1 x 10-6M. Linear detection limit is about 10-5M Interference's:

More information

Analysis of cations and anions by Ion- Selective Electrodes (ISEs)

Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Purpose: The purpose of this assignment is to introduce potentiometric measurements of ionic species by ion selective electrodes (ISEs)

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

Chem 321 Lecture 17 - Potentiometry 10/24/13

Chem 321 Lecture 17 - Potentiometry 10/24/13 Student Learning Objectives Chem 321 Lecture 17 - Potentiometry 10/24/13 Electrodes The cell described in the potentiometric chloride titration (see 10/22/13 posting) consists of a Ag/AgCl reference electrode

More information

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions. Cole-Parmer Ammonium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium

More information

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0) 150mV 300mV Mono/Reference Combination S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) 1376 340 456 Fax: +44 (0) 1376 340 453 Email:

More information

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

Chem 321 Lecture 16 - Potentiometry 10/22/13

Chem 321 Lecture 16 - Potentiometry 10/22/13 Student Learning Objectives Chem 321 Lecture 16 - Potentiometry 10/22/13 In lab you will use an ion-selective electrode to determine the amount of fluoride in an unknown solution. In this approach, as

More information

Chemistry Determination of Mixed Acids

Chemistry Determination of Mixed Acids Chemistry 3200 Acid-base titration is one of the most common operations in analytical chemistry. A solution containing an unknown amount of ionizable hydrogen can be titrated with a solution of standard

More information

Chemistry 213. Electrochemistry

Chemistry 213. Electrochemistry Chemistry 213 Electrochemistry Part A: Electrochemical Cells Objective Oxidation/reduction reactions find their most important use in the construction of voltaic cells (chemical batteries). In this experiment,

More information

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml) GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Silver/Sulfide Ion Electrode is used to measure silver or sulfide ions in aqueous solutions. The two ions are virtually never present in solution

More information

Physical Chemistry Laboratory I Experiment 3 Effect of Ionic Strength on the Solubility of CaSO 4 (Revised, 01/13/03)

Physical Chemistry Laboratory I Experiment 3 Effect of Ionic Strength on the Solubility of CaSO 4 (Revised, 01/13/03) Physical Chemistry Laboratory I Experiment 3 Effect of Ionic Strength on the Solubility of CaSO 4 (Revised, 01/13/03) It is generally assumed that solutions of strong electrolytes are completely dissociated

More information

1) A ph/mv meter or an ion meter, either line operated or portable.

1) A ph/mv meter or an ion meter, either line operated or portable. COLE-PARMER Sodium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS The Cole-Parmer Sodium Ion Electrodes are used to quickly, simply, accurately, and economically measure sodium ion concentrations

More information

EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION

EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION I. INTRODUCTION This experiment introduces the direct-measurement approach to potentiometry. Principal purposes of the study are a) to understand quantitative

More information

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA Background In this experiment, students will familiarize themselves with potentiometric titration, practice using the first derivative to find the equivalence

More information

Ion Selective Electrode Probe

Ion Selective Electrode Probe Rev. 1 Ion Selective Electrode Probe KDS-1064 Calcium KDS-1065 Ammonium KDS-1066 Nitrate KDS-1067 Chloride Type : Ion selective electrode Range : 1 M to 5x10-7 M(40,000 ~ 0.01 ppm) Sensor description The

More information

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic. GENERAL INSTRUCTIONS Introduction The Van London Co. Cyanide Ion Selective Electrode is used to measure dissolved cyanide ion in aqueous solutions. *IMPORTANT NOTICE* Acidic cyanide solutions produce hydrogen

More information

Table of Contents. Purpose... 2 Background... 2 Prelab Questions... 3 Procedure:... 3 Calculations:... 4

Table of Contents. Purpose... 2 Background... 2 Prelab Questions... 3 Procedure:... 3 Calculations:... 4 Table of Contents Purpose... 2 Background... 2 Prelab Questions... 3 Procedure:... 3 Calculations:... 4 CHM 212 Experiment 4 Determination of the Ka of Potassium Hydrogen Phthalate (KHP) Using a Gran Plot

More information

Complex equilibria and calculations of formation constants from potentiometric data

Complex equilibria and calculations of formation constants from potentiometric data Experiment 8: /Ag 2 Based Ion elective Electrode Determination of Lead/Hydroxide Equilibria While lead hydroxide does not give a good quantitative measure of lead, particularly in the presence of competing

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London Co. Nitrogen Oxide Gas-Sensing Electrode is used to measure nitrite in aqueous solutions. The measurement is not affected by sample color or turbidity.

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY

ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY ELECTROCHEMISTY ELECTROCHEMISTRY IS THE STUDY OF CHEMICAL REACTIONS THAT RESULT IN THE FLOW OF ELECTRONS (CURRENT) OR THE DEVELOPMENT

More information

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions. COLE-PARMER NITRATE ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate

More information

Electrochemistry LEC Potentiometric ph titration (phosphoric acid in a soft drink) What you need: What you can learn about

Electrochemistry LEC Potentiometric ph titration (phosphoric acid in a soft drink) What you need: What you can learn about Electrochemistry LEC 06 What you can learn about Galvanic cell Types of electrodes Nernst equation Potentiometry Principle and tasks The cell voltage and the Galvani voltage of the electrodes of an galvanic

More information

Electrochemistry. To use principles of electrochemistry to understand the properties of electrochemical cells and electrolysis.

Electrochemistry. To use principles of electrochemistry to understand the properties of electrochemical cells and electrolysis. Electrochemistry Objectives: To use principles of electrochemistry to understand the properties of electrochemical cells and electrolysis. Background: Part I: Galvanic Cells: A Galvanic cell is a device

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

2. Which of the following statements best describes the movement of electrons in an electrochemical cell?

2. Which of the following statements best describes the movement of electrons in an electrochemical cell? Exam 2 Chem 311 Evans Fall 2009 112: 2 pts each 1. Consider the following unbalanced redox equation: Pb (s) + PbO 2 (s) + 2 HSO 4 (aq) 2 PbSO 4 (s) Which species is being oxidized? A. HSO 4 B. Pb(s) C.

More information

CHAPTER 14: ELECTRODES AND POTENTIOMETRY

CHAPTER 14: ELECTRODES AND POTENTIOMETRY CHAPTER 14: ELECTRODES AND POTENTIOMETRY Chapter 14 Electrodes and Potentiometry Potentiometry : The use of electrodes to measure voltages that provide chemical information. (The cell voltage tells us

More information

Chemistry Instrumental Analysis Lecture 22. Chem 4631

Chemistry Instrumental Analysis Lecture 22. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 22 Measures potential under very low currents. The cell is 2 half cells. Consist of a reference electrode, indicator electrode, and potential measuring device.

More information

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA Introduction In this experiment, students will familiarize themselves with potentiometric titration, practice using the first derivative to find the equivalence

More information

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Potentiometry. Analytical Laboratory II

Potentiometry. Analytical Laboratory II Potentiometry Analytical Laboratory II Potentiometry is based on the measurement of the potential of an electrode system. Potentiometric measurement system consists of two electrodes called reference and

More information

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 0510212 Instrumental Chemical Analysis Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 Chapter 1 Electroanalytical Methods Electroanalytical Chemistry Electroanalytical

More information

Chapter 3 Electrochemical methods of Analysis-Potentiometry

Chapter 3 Electrochemical methods of Analysis-Potentiometry Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator

More information

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable.

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable. phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Silver/Sulfide Ion Electrodes are used to quickly, simply, accurately,

More information

Potentiometry: Ion-Selective Electrodes

Potentiometry: Ion-Selective Electrodes Potentiometry: Ion-Selective Electrodes Determination of the sodium monofluorophosphate content of a toothpaste sample using a fluoride ion-selective electrode. 1. Aims The aims of the experiment are to:

More information

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Calcium Ion Electrodes are used to quickly, simply, accurately, and economically

More information

Chapter 21. Potentiometry

Chapter 21. Potentiometry Chapter 21 Potentiometry 1 Potentiometric methods Potentiometric methods of analysis are based on measuring the potential of electrochemical cells without drawing appreciable currents. Applications: Determine

More information

Chemistry Potentiometric Titration of a Chloride-Iodide Mixture

Chemistry Potentiometric Titration of a Chloride-Iodide Mixture Chemistry 3200 Silver iodide, AgI, is much less soluble than AgCl. The solubility products of the two salts are 9.8 x 10 17 and 1.78 x 10 10, respectively. Therefore, if a mixture of I and Cl is titrated

More information

Electro Analytical Methods

Electro Analytical Methods CH 2252 Instrumental Methods of Analysis Unit II Electro Analytical Methods Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam

More information

F- Cl- Br- I- CN- S2- NO3-

F- Cl- Br- I- CN- S2- NO3- Middle of the 60 s First ISE for fluoride Na+ K+ Ag+ Ca2+ NH + 4 Cu2+ Pb2+ F- Cl- Br- I- CN- S2- NO3- Environment Food industry and agriculture Medecine, pharmaceutical, cosmetics Power plants Fluoride

More information

#13 Electrochemical Cells

#13 Electrochemical Cells #13 Electrochemical Cells If a copper strip is placed in a solution of copper ions, one of the following reactions may occur: Cu 2+ + 2e - Cu Cu Cu 2+ + 2e - The electrical potential that would be developed

More information

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY

WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY University of Puget Sound Department of Chemistry Chem 230 WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY INTRODUCTION Nitrogen is an essential nutrient

More information

Titration of a strong acid with a strong base with Cobra4

Titration of a strong acid with a strong base with Cobra4 Titration of a strong acid with a strong base with Cobra4 TEC Related topics Strong and weak acids and bases, ph value, titration curves, equivalence point, potentiometry. Principle Hydrochloric acid is

More information

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0) 150mV 300mV Mono/Reference Combination S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) 1376 340 456 Fax: +44 (0) 1376 340 453 Email:

More information

Chapter 24. Electrogravimetry and Coulometry

Chapter 24. Electrogravimetry and Coulometry Chapter 24 Electrogravimetry and Coulometry Dynamic Electrochemical Methods of analysis Electrolysis Electrogravimetric and Coulometric Methods For a cell to do any useful work or for an electrolysis to

More information

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge Standard Operating Procedure for GLNPO Specific Conductance: Marvin Palmer United States Environmental Protection Agency Great Lakes National Program Office Region 5 77 West Jackson 9th Floor Chicago,

More information

Buffer Preparation. Learning Objectives:

Buffer Preparation. Learning Objectives: Proteomics Buffer Preparation Buffer Preparation Maintaining the optimum ph during the biological sample processing is to maintain the proper functional and structural aspects of the sample. It is important

More information

Chapter 13 POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY

Chapter 13 POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY Chapter 13 POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY Ch.13 In this chapter Understand the concept of the various types of electrodes that can be used for measuring

More information

METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES

METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES ELECTROCHEMISTRY Basic principle Electrodes are used to selectively measure particular ions Instruments utilizing electrodes measure the potential

More information

Equilibrium and Ionic Strength Effects

Equilibrium and Ionic Strength Effects Equilibrium and Ionic Strength Effects Objectives You will determine the thermodynamic equilibrium constant for the reaction between iron(iii) ion and thiocyanate ion to form iron(iii)-thiocyanate. Fe

More information

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Ion Selective Electrodes for the Laboratory Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Outline Review ISE measurement technology How to properly calibrate

More information

LANTHANUM (III) SELECTIVE ELECTRODE BASE 1,10-DIAZA-4,7,13,16- TETRAOXACYCLOOCTADECANE-N,N -DIACETIC ACID AS IONOPHORE.

LANTHANUM (III) SELECTIVE ELECTRODE BASE 1,10-DIAZA-4,7,13,16- TETRAOXACYCLOOCTADECANE-N,N -DIACETIC ACID AS IONOPHORE. 1 LANTHANUM (III) SELECTIVE ELECTRODE BASE 1,10-DIAZA-4,7,13,16- TETRAOXACYCLOOCTADECANE-N,N -DIACETIC ACID AS IONOPHORE Suyanta a, *, a) Departement of Chemistry, Yogyakarta State University, Yogyakarta

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

EXPERIMENT 2 DETERMINATION OF K a USING THE CONDUCTANCE METHOD

EXPERIMENT 2 DETERMINATION OF K a USING THE CONDUCTANCE METHOD EXPERIMENT 2 DETERMINATION OF K a USING THE CONDUCTANCE METHOD Introduction Equilibrium Processes When a pure sample of liquid-state acetic acid (i.e., CH 3 COOH (l) /HAc (l) ) is added to a beaker of

More information

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014 All about ph! Is it as simple as it seems? The ph Workshop Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014 Common Questions: Measuring ph What is ph? Potential Hydrogen

More information

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions. OAKTON Cadmium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous

More information

Na+ K+ Ag+ Ca2+ NH + 4. F- Cl- Br- I- CN- S2- NO3-

Na+ K+ Ag+ Ca2+ NH + 4. F- Cl- Br- I- CN- S2- NO3- 1 1 HISTORY 2 Middle of the 60 s First ISE for fluoride Na+ K+ Ag+ Ca2+ NH + 4 Cu2+ Pb2+ F- Cl- Br- I- CN- S2- NO3- Until the 60 s the only ISE is in fact the ph electrode with a glass sensitive to protons.

More information

Ch. 14. ELECTRODES AND POTENTIOMETRY

Ch. 14. ELECTRODES AND POTENTIOMETRY Ch. 14. ELECTRODES AND POTENTIOMETRY 14.1 Analytical chemists design electrodes (voltage sensitive to conc. change) galvanic cells ion-selective electrodes ion-sensing field effect transistors potentiometry

More information

Electrochemistry. Conductivity of strong and weak electrolytes LEC 06. What you need: What you can learn about. Principle and tasks

Electrochemistry. Conductivity of strong and weak electrolytes LEC 06. What you need: What you can learn about. Principle and tasks LEC 06 Electrochemistry What you can learn about Kohlrausch s law Equivalent conductivity Temperature-dependence of conductivity Ostwald s dilution law Principle and tasks It is possible to differentiate

More information

Reference electrode. Calomel electrode Hg in contact with Hg(I) chloride Ag/AgCl 15-2

Reference electrode. Calomel electrode Hg in contact with Hg(I) chloride Ag/AgCl 15-2 Potentiometry Potential measurements of electrochemical cells Ion selective methods Reference electrode Indicator electrode Potential measuring device Reference electrode Indicator electrodes Ion specific

More information

Chemistry 325 Instrumental Methods of Analysis March 13, Final Exam. Name

Chemistry 325 Instrumental Methods of Analysis March 13, Final Exam. Name Final Exam Name Instructions: This exam is worth 100 points. Some questions allow a choice as to which parts are answered. Only answer the number of parts requested. 1. (32 points) Circle the best answer

More information

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Ion-selective electrodes: challenges and opportunities for on-site/on-line measurements

Ion-selective electrodes: challenges and opportunities for on-site/on-line measurements EURACEM Workshop on Quality Assurance of Measurements from Field to Laboratory, MIKES, Espoo Finland, 20-21 May 2013 Ion-selective electrodes: challenges and opportunities for on-site/on-line measurements

More information

we might also expect the reaction rate to be influenced by ph. In fact, the rate has been reported to follow the rate law:

we might also expect the reaction rate to be influenced by ph. In fact, the rate has been reported to follow the rate law: KINETICS Objective: The objective of this lab is to measure the rate of iron oxidation, to determine the order of the reaction, and thereby to gain familiarity with rate laws in both the differential and

More information

Experiment 5E BOTTLES WITHOUT LABELS: STUDIES OF CHEMICAL REACTIONS

Experiment 5E BOTTLES WITHOUT LABELS: STUDIES OF CHEMICAL REACTIONS Experiment 5E BOTTLES WITHOUT LABELS: STUDIES OF CHEMICAL REACTIONS FV 1-21-16 MATERIALS: Eight 50 ml beakers, distilled water bottle, two 250 ml beakers, conductivity meter, ph paper (A/B/N), stirring

More information

KJ 2051 Coulometric titration of hydrochloric acid

KJ 2051 Coulometric titration of hydrochloric acid KJ 2051 Page 1 of 4 F.G. Banica KJ 2051 Coulometric titration of hydrochloric acid By F.G. Banica, 22 September 2006 This document includes a short theory of this experiment. See Ref. [1] for experimental

More information

GETTING THE END POINT TO APPROXIMATE. Two hours

GETTING THE END POINT TO APPROXIMATE. Two hours Chem 1312 Handout Experiment ONE Laboratory Time Required Special Equipment and Supplies Objective Safety First Aid GETTING THE END POINT TO APPROXIMATE THE EQUIVALENCE POINT Two hours Balance Potassium

More information

Electrochemistry. Redox reactions. Half Reactions. Nernst Equation Ion selective electrodes

Electrochemistry. Redox reactions. Half Reactions. Nernst Equation Ion selective electrodes Electrochemistry Nernst Equation Ion selective electrodes Redox reactions oxidation - loss of electrons M n+ M n+1 + e - M is oxidized - reducing agent reduction - gain of electrons N n+ + e - N n-1 N

More information

POTASSIUM ION SELECTIVE ELECTRODE

POTASSIUM ION SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6733 012-06614A 9/97 POTASSIUM ION SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville,

More information

PRACTICAL 3 ph AND BUFFERS

PRACTICAL 3 ph AND BUFFERS PRACTICAL 3 ph AND BUFFERS ph and Buffers Structure 3.1 Introduction 3.2 ph and Buffers: Basic Concept 3.2.1 ph 3.2.2 Buffers and Buffer Solutions 3.3 Methods for Determining ph Experiment 1: Measurement

More information

General terms relevant to ion-selective electrodes

General terms relevant to ion-selective electrodes 8.3.2.1 General terms relevant to ion-selective electrodes Activity, Activity coefficient and Concentration See sections 1.3.8, 1.3.10 and 3.3. Calibration plot This is a plot of the cell emf, E (i.e.

More information

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions.

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions. OAKTON Lead Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in

More information

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 imagine explore learn Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 Overview Sodium is highly abundant in the lithosphere (Earth s outer shell) where it occurs as rock salt deposits as

More information

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 3 3. Electrode and Measurement Setup 4 Electrode

More information

EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical

More information

Determination of an Equilibrium Constant

Determination of an Equilibrium Constant Last updated 1/29/2014 - GES Learning Objectives Students will be able to: Determine the numerical value of an equilibrium constant from measured concentrations of all reaction species. Use an absorption

More information

ELECTROCHEMISTRY. these are systems involving oxidation or reduction there are several types METALS IN CONTACT WITH SOLUTIONS OF THEIR IONS

ELECTROCHEMISTRY. these are systems involving oxidation or reduction there are several types METALS IN CONTACT WITH SOLUTIONS OF THEIR IONS Electrochemistry 1 ELECTROCHEMISTRY REDOX Reduction gain of electrons Cu 2+ (aq) + 2e > Cu(s) Oxidation removal of electrons Zn(s) > Zn 2+ (aq) + 2e HALF CELLS these are systems involving oxidation or

More information

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Chapter 18 Electrochemistry. Electrochemical Cells

Chapter 18 Electrochemistry. Electrochemical Cells Chapter 18 Electrochemistry Chapter 18 1 Electrochemical Cells Electrochemical Cells are of two basic types: Galvanic Cells a spontaneous chemical reaction generates an electric current Electrolytic Cells

More information

Electrodes for ph measurements. - when you need to be sure...

Electrodes for ph measurements. - when you need to be sure... Electrodes for ph measurements - when you need to be sure... Different types Combined Reference Glass Electrode working principle Glass electrode Reference electrode Electrode body Internal reference element

More information

Eye on Ions: Electrical Conductivity of Aqueous Solutions

Eye on Ions: Electrical Conductivity of Aqueous Solutions Eye on Ions: Electrical Conductivity of Aqueous Solutions Pre-lab Assignment: Reading: 1. Chapter sections 4.1, 4.3, 4.5 and 4.6 in your course text. 2. This lab handout. Questions: 1. Using table 1 in

More information

Determination of Orthophosphate Ion

Determination of Orthophosphate Ion Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the

More information

AP Questions: Electrochemistry

AP Questions: Electrochemistry AP Questions: Electrochemistry I 2 + 2 S 2O 2-3 2 I - + S 4O 2-6 How many moles of I 2 was produced during the electrolysis? The hydrogen gas produced at the cathode during the electrolysis was collected

More information

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable.

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable. phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Carbon Dioxide Gas-Sensing Electrode is used to quickly,

More information

Determination of Orthophosphate Ion

Determination of Orthophosphate Ion Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the

More information

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out.

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out. Sign In Forgot Password Register username username password password Sign In If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out. ChemWiki

More information

EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY

EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY University of Puget Sound Department of Chemistry Chem 230 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY INTRODUCTION Nitrogen is an essential nutrient

More information