EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION
|
|
- Darlene Shepherd
- 6 years ago
- Views:
Transcription
1 EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION I. INTRODUCTION This experiment introduces the direct-measurement approach to potentiometry. Principal purposes of the study are a) to understand quantitative relationships between electrode potential and concentration and b) to use that information to understand how absolute and relative concentration errors depend on errors in measured potentials. A secondary purpose is to use potential measurements to quantify the hydrogen peroxide concentration in an unknown sample and to determine the confidence interval for the results. II. OVERVIEW A. Chemical reactions The half reactions involved in this experiment are as follows: + O + 2 H + 2 e 2 H O (1a) H2 2 2 I I + 2 e (1b) Hydrogen peroxide is reduced to water and iodide is oxidized to triiodide. The net reaction is the sum of these two half reactions as follows: + - O + I + 2 H I 2 H O (1c) H This reaction is slow in the absence of a catalyst but quite fast in the presence of catalysts such as molybdate (Mo(VI)) and an excess of iodide that forces the reaction to completion by the formation of triiodide, I -. B. Rationale In the presence of a high iodide concentration, changes in triiodide concentration can be monitored potentiometrically. Accordingly, hydrogen peroxide concentration can be quantified potentiometrically by measuring the potential of an inert electrode immersed in the solution and using a calibration relationship to compute triiodide concentration, which is then converted to hydrogen peroxide concentration. C. Standardization In practical situations, the hydrogen peroxide concentration should be standardized. However, the main thrust of this experiment focuses on characteristics of potentiometric methods. Accordingly, to save time and to permit you to focus on the main thrust of the experiment, you will use the nominal concentration of the hydrogen peroxide in a % solution. D. Potentiometric detection It will be helpful to review some features of potentiometric detection before proceeding with the experimental discussion. 1. General. Potentiometry exploits electrochemical properties of chemical systems. In potentiometry, the aim is to measure the potential of a system without drawing any current from it, e.g. to measure the potential without causing electrochemical reactions that would change the concentration at the electrode surface. Typical examples of potentiometric sensors include the glass electrode used for ph measurements and ion-selective electrodes used for a variety of ionic analytes such as sodium, potassium, calcium, sulfide, etc. Relationships between electrode potential and concentration are described by the Nernst equation (see text). Potentiometry is frequently used in one of two modes, namely a direct-measurement mode where electrode potential is measured and used to compute concentration and a titrimetric mode where electrode potential is used to detect the end point in a titration. This experiment involves the direct-measurement mode. 1
2 2. Electrode potential. Potentiometric detection of triiodide at an inert (e.g. platinum) electrode is based on the fact that reaction 1b is reversible. The Nernst equation for this reaction is: E I/I = E 0 I/I + (RT/nF)ln(C I /C I ) = E 0 I/I - (RT/nF)lnC I + (RT/nF)lnC I (2) where E is the electrode potential, E 0 I/I is the standard reduction potential (See text) for reaction 1b, R is the ideal gas constant, T is temperature ( 0 K), n is the number of electrons involved in the reaction (n = 2 in this case), F is Faraday s constant, and C I and C I are the molar concentrations of triiodide and iodide, respectively. Because the iodide concentration is present in large excess and does not change significantly, it can be incorporated into the formal electrode potential, E 0. Substituting E 0 I/I = E 0 I/I - (RT/nF)lnC I, Eq. 2 can be simplified to: E I/I = EI/I 0 + (RT/nF)ln C I (). Cell voltage. The potential described above is the potential expected at an inert electrode immersed in a solution containing iodide and triiodide. Unfortunately, this potential can not be measured independently but must be measured relative to a reference electrode. The reference electrode used in this study is the saturated calomel (SCE) electrode and has a potential of V vs. the normal hydrogen electrode(nhe) at 25 0 C. The cell voltage you will measure is the difference between the indicator and reference electrodes. E Cell = E I/I - E SCE = E I/I C (4a) Substituting from Eq. we have: E Cell = E 0 I/I - E SCE + (RT/nF)ln C I (4b) Because both E 0 I/I and E SCE are constants, Equation 4b can be simplified to: E Cell = α + β ln C I (4c) where α = E 0 I/I - β ln C I - - E SCE and β = (RT/nF). In other words, the cell potential is expected to vary linearly with the logarithm of the molar concentration of triiodide ion. 4. Numerical values of constants. Using R = 8.14 J K -1 and F = 96,485 coulombs it follows that β = RT/nF = 8.62 x 10-5 T/n. Assuming n = 2 (See reaction 1b) and T = 298, it follows that β = x Given E 0 I/I = 0.56 V, E SCE = V, and β ln C I - = x 10-2 ln (0.2) = , it follows that α = 0.56 V V - ( V) = Accordingly, at 25 0 C, the cell potential is expected to vary with triiodide concentration as follows: E Cell (V) = 0.54 V x 10-2 ln C I (4d) This expression gives potentials in Volts, V. The ph meter you will use reads potential in millivolts, mv. The above expression is expressed in millivolts as follows: E Cell (mv) = 54 mv ln C I (4e) with C I expressed in mol/l. In practice, the cell voltage is expected to vary linearly with the logarithm of concentration (Eq. 4c) but the constants are expected to be somewhat different than those in Eqns. 4d and 4e. 5. Concentration vs. potential. Equation 4c is easily rearranged to be explicit in concentration. The relationship is E α C = exp( ) Cell (5) β In other words, the relationship between concentration and potential is an exponential function. Theoretical values of the constants for T = 298 and n = 2 are given above, with magnitudes depending on whether voltages are expressed in volts or millivolts. III. INSTRUMENTATION AND SOLUTIONS A. Extra glassware In addition to supplies in your drawers, each group will need twelve 100-mL volumetric flasks, twelve 100-mL beakers and a thermometer good to at least 0.1 degrees Celcius. 2
3 B. Instrumentation A ph meter equipped with a platinum indicator electrode and a saturated-calomel reference electrode pair will be used to measure potentials. Potentials will be measured and entered into laboratory notebooks manually. C. Solutions All solutions are to be prepared in deionized water and glassware should be rinsed in deionized water. 1. Solution provided Buffer/catalyst solution. For each liter of solution, add 0.1 g of (NH 4 ) 6 Mo 7 O 24 4H 2 O, 0 g of KH 2 PO 4, and 14 g of K 2 HPO 4 H 2 O to 1 L of water and stir until the salts are completely dissolved (600 ml/group). 2. Solutions you must prepare a. Composite reagent (1 L required). Prepare this solution fresh by dissolving 67 g of potassium iodide in about 500 ml of the buffer/catalyst solution and dilute to about 1L with DI water. Because the dissolution of KI is endothermic and light enhances air oxidation of iodide, place the solution in your desk to warm to room temperature in the dark while you prepare hydrogen peroxide solutions described below. b. Stock solutions of hydrogen peroxide. Three stock solutions of hydrogen peroxide will be used to prepare diluted solutions containing nominal hydrogen peroxide concentrations over a wide concentration range. The nominal molar concentration of hydrogen peroxide in a % solution is C H2O2 = (1000 ml/l) x 1g/mL x 0.0 (g/g) /4 g/mol = mol/l. i. Stock 1 ( M). Pipet 10 ml of % hydrogen peroxide into a 100-mL volumetric flask, dilute to volume with water and mix thoroughly. ii. Stock 2 ( M). Pipet ml of the Stock 1 solution into a 250-mL volumetric flask, dilute to volume with water and mix thoroughly. iii. Stock ( M). Pipet ml of Stock 2 hydrogen peroxide solution into a 250-mL volumetric flask, dilute to volume with water and mix thoroughly. IV. PROCEDURES Turn on the ph meter and set the mode switch to mv at least 20 min before you will begin to measure potentials. Briefly, the procedure involves reaction of hydrogen peroxide with iodide to produce triiodide and detection of the triiodide potentiometrically using a platinum indicator electrode and a saturated calomel reference electrode. Three sets of data will be obtained for each solution. A. Calibration Standards/Unknown. Stock solutions described above are used to prepare two sets of calibration standards covering a 150-fold concentration range. Do NOT dilute the following solutions to volume at this point. 1. Calibration standards (Set 1). Label five 100-mL volumetric flasks A-E. Using a buret, add 2.00, 5.00, 10.00, 20.00, and 0.00 ml of Stock solution to flasks labeled A-E, respectively. Set these flasks aside and proceed with the next step. 2. Calibration standards (Set 2). Label four 100-mL volumetric flasks F-I. Using a second buret, add 5.00, 10.00, and 0.00 ml of Stock 2 solution to flasks labeled F-I, respectively. Set these flasks aside and proceed with the next step.. Unknown. See instructions for Unknown for Experiment 8 on page 8.
4 B. Reaction step Pipet ml of the composite reagent into each of the 100-mL volumetric flasks containing calibration standards and unknowns. Dilute each to volume using deionized water and mix thoroughly. Wait approximately 5 min after preparing these solutions before making measurements. Note and record the appearances of the solutions. C. Measurement step Label several clean dry 100-mL beakers A-I and U. Transfer approximately 50 ml of each of the calibration standards and unknown solutions into each of the 100-mL beakers. Rinse the platinum/saturated calomel electrode (SCE) pair with deionized water and pat dry with lint-free tissue. 1. Calibration solutions Wipe the electrodes gently with a lint-free tissue (do not wash with water) and Immerse them in the most dilute of the calibration standards. Swirl briefly let stand until the potential stabilizes, and record the potential. (The potential should approach a steady-state value very quickly in this and the other standard solutions). Repeat this process for each of the calibration standards, working from the more dilute to the more concentrated solutions to minimize effects of carry-over among the solutions. Measure and record the temperature of at least three of the solutions. 2. Unknown solution After measuring the potential of the calibration standards, rinse the electrodes thoroughly with deionized water, pat them dry, and measure and record the potential of the unknown solution.. Replicate measurements Repeat the measurement process twice for each standard and unknown. You should have at least three complete sets of data for potential vs. concentration before you quit. D. Least significant digit Note and record the least significant digit (LSD( )) on the ph meter used to measure voltages. The least significant digit is the smallest voltage resolved by the readout scale on the ph meter. The precision of replicate measurements with a digital device can be no better than one half of the LSD (i.e. SD 0.5 LSD). V. DATA PROCESSING/ANALYSIS Principal emphasis here is on relationships between potential and concentration and error analysis. In all of the discussion below, it is assumed that signals are expressed in Volts (not mv) and that concentrations are expressed in mol/l (M). Note: The ph meter you will use reads cell voltage in millivolts. Although it is possible to work with millivolts (See Eq. 4e), there will be less confusion in some calculations if you change voltages to volts by multiplying by 1 x 10 - V/mV before you proceed with other calculations (See Eq. 4d). 4
5 A. Mean values and imprecision Calculate the mean values and standard deviations of cell voltages for the replicate measurements of the calibration standards. Use the individual standard deviations to calculate the pooled standard deviation (s E,p, See text) for all the data. Compare the pooled standard deviation to the least significant digit (Part IV-D). If the pooled standard deviation is greater than one half the least significant digit (s E,p 0.5 LSD), use the pooled standard deviation as the voltage error, s E, in calculations below. If the pooled standard deviation is less than 0.5 LSD, use 0.5 x LSD as the voltage error, s E, in calculations below. B. Potential/Concentration relationships Potential/concentration data will be plotted in three formats to help you visualize more completely the nature of the relationship. 1. Potential vs. concentration Potential (V) vs. concentration (M) will be plotted in two formats, on linear coordinates and as a semilogarithmic plot. a. Linear coordinates. Plot mean values of Potential (V) vs. Concentration (M) on linear coordinates. Using suitable software (e.g. Origin), fit an equation in the form of Eq. 4c to the data. A suitable syntax for the equation in Origin is: P 1+ P2* LN( X ) (6) where P1 is α and P2 is β (Eq. 4c). Use numbers in Eq. 4d (assuming data are in Volts) as initial estimates of P1 and P2. Include this graph and least-squares statistics in your report as Figure 1. Compare values of P1 = α and P2 = β to expected values in Eq. 4d. b. Semilogarithmic plot. Calculate values of the natural logarithms of concentrations (ln C I ) of the standards and prepare a plot of potential vs. ln C I. Fit a linear model to these data. Include this graph and least-squares statistics in your report as Figure Concentration vs. potential Plot Concentration (M) of standards as ordinate vs. Potential (V) as abscissa and fit a model similar to Eq. 5 to the data. The model in Origin syntax is: EXP(( X P1) / P2) (7) Symbols have the same significance (and approximate numerical values) as in Eq. 6. Include this graph and least-squares statistics in your report as Figure. C. Error analysis A primary purpose of this experiment is to understand how measurement errors, s E, are translated into absolute and relative concentration errors (s C and s c /C). We shall approach this both graphically and mathematically. Recall that the absolute concentration error is the measurement error divided by the sensitivity. The measurement error was quantified above (Part V-A). Accordingly, the next step in our error analysis is to determine the sensitivity. Error equations are summarized at the end of this experiment. 5
6 1. Sensitivity (Eq. EEP-5) Recall that the sensitivity is the slope of a plot of the dependent variable vs. the independent variable. In this case, it is the slope of a plot of cell voltage vs. concentration. (See Eq. EEP-5 in attached equations). Use the numerical value of β calculated above (Part V-B-1-a) to calculate values of sensitivity as a function of concentration. Prepare a plot of sensitivity vs. concentration, do a least-squares fit of EEP-5 (P1/X in Origin syntax) to the data and include this plot in your report as Figure 4. Rationalize the shape of this plot using the linear-coordinate plot of potential vs. concentration (Part V-B-1-a). 2. Concentration error (Eq. EEP-6) Calculate the concentration error (Eq. EEP-6) for each standard concentration. Prepare a plot of concentration error (mol/l) vs. concentration, fit Eq. EEP-6 to the data and include the plot in your report as Figure 5.. Relative concentration error (EEP-7) Use equation EEP-7 to describe how relative concentration error is expected to vary with concentration in direct-measurement potentiometry. D. Unknown concentration Use the relationship in Eq. 7 (Part V-B-2) to calculate the concentration (mol/l) of your unknown. Use this value to calculate the hydrogen peroxide concentration in the 250 ml volumetric flask after the first dilution. Report this concentration, the standard deviation, and the 95 % confidence interval in the usual way. Target values for unknown concentrations are 0.0 to 0.09 mol/l. VI. REPORT In addition to the usual information, your report should include a tabulation of all calculations, figures, and least-squares results requested above, and a discussion of each. Also, compare relationships between absolute and relative concentration errors vs. concentration in direct-measurement potentiometry and direct-measurement spectrophotometry with fixed transmittance error (Experiment 4). 6
7 ERROR EQUATIONS POTENTIOMETRY This paper summarizes an approach to the development of error equations for direct-measurement potentiometry. Sensitivity The symbol, Φ E,C, represents the sensitivity of potential, E, to changes in concentration, C. The sensitivity is quantified either as the slope of a plot of potential vs. concentration or the first derivative of a mathematical relationship between potential and concentration. Indicator reaction I + 2 e I (EEP-1) Equilibrium electrode potential [I 0 RT ] E = E + ln E - Ref nf [I ] (EEP-2) Constants R = 8.14 Coul/ 0 K mol, F = 96,485 Coul/Eq, n = 2, T = K, E Ref = V, [I - ] = 0.20 mol/l Simplified equation E = 0.54V ln[ I ] (EEP-) E = α + β ln C (Note: α = 0.54 V, β = V) (EEP-4) Sensitivity de d(α + βlnc) β d( k lnu) 1 du Φ E, C = = = (Note: = k ) (EEP-5) dc dc C dx u dx Concentration error se se C sc = = = se,c (EEP-6) ΦE,C β / C β (Note: Absolute concentration error increases linearly with concentration). Relative concentration error (RCE) sc 1 C 1 = s E,C = se,c (EEP-7) C C β β (Note: Relative concentration error is independent of concentration). 7
8 Instructions for Unknown for Experiment 8 Peroxide is somewhat unstable in air and must be diluted immediately prior to analysis. Therefore, you will need to do two dilutions of your unknown to get it into the proper concentration range for analysis by this procedure. Dilute unknown to volume in the 250 ml flask with distilled water and mix thoroughly. Pipet 25 ml of this into a clean 250 ml volumetric flask. Dilute to volume with composite reagent and mix thoroughly. Pipet 10 ml of this dilution into a clean 100 ml volumetric flask and proceed with the reaction step. 8
IODINE CLOCK REACTION KINETICS
Name: Section Chemistry 104 Laboratory University of Massachusetts Boston IODINE CLOCK REACTION KINETICS PRELAB ASSIGNMENT Calculate the initial concentration of H 2 O 2 that exists immediately after mixing
More informationMeasurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste
Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major
More informationPRACTICAL 3 ph AND BUFFERS
PRACTICAL 3 ph AND BUFFERS ph and Buffers Structure 3.1 Introduction 3.2 ph and Buffers: Basic Concept 3.2.1 ph 3.2.2 Buffers and Buffer Solutions 3.3 Methods for Determining ph Experiment 1: Measurement
More informationLowell High School AP Chemistry Spring 2009 REACTION KINETICS EXPERIMENT
Lowell High School AP Chemistry Spring 2009 REACTION KINETICS EXPERIMENT Complete the following for Pre-Lab on a clean sheet of paper: (1) In your own words, explain the following: a. why the I 2 concentration
More informationExperiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University
Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University Adapted from "Experiment 3: Analysis of Phosphate in Water," Laboratory Experiments
More informationChemistry Determination of Mixed Acids
Chemistry 3200 Acid-base titration is one of the most common operations in analytical chemistry. A solution containing an unknown amount of ionizable hydrogen can be titrated with a solution of standard
More informationREADING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.
CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE READING Skoog, Holler and Crouch: Chapter 23 and Appendix 3. A. INTRODUCTION Potentiometry is a static electroanalytical method in which the potential
More informationExperimental Procedure Lab 402
Experimental Procedure Lab 402 Overview Measured volume of several solutions having known concentrations of reactants are mixed in a series of trials. The time required for a visible color change to appear
More informationDetermination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)
Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with
More information[Ca 2+ ] = s (3) [IO - 3 ] = 2s (4)
E10 Chemical Equilibria: K sp of Calcium Iodate Objective! Understand the relation between the molar solubility and the solubility product constant of a sparingly soluble salt.! Measure the molar solubility
More informationDetermination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)
Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with
More informationSafety Note: Safety glasses and laboratory coats are required when performing this experiment
The Determination of Hypochlorite in Bleach Reading assignment: Burdge, Chemistry 4 th edition, section 4.6. We will study an example of a redox titration in order to determine the concentration of sodium
More informationExperiment 7 Buffer Capacity & Buffer Preparation
Chem 1B Dr. White 57 Experiment 7 Buffer Capacity & Buffer Preparation Objectives To learn how to choose a suitable conjugate acid- base pair for making a buffer of a given ph To gain experience in using
More informationSodium Chloride - Analytical Standard
Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes
More informationThe Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Chemistry 102 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
More informationAnalysis of cations and anions by Ion- Selective Electrodes (ISEs)
Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Purpose: The purpose of this assignment is to introduce potentiometric measurements of ionic species by ion selective electrodes (ISEs)
More informationTable of Contents. Purpose... 2 Background... 2 Prelab Questions... 3 Procedure:... 3 Calculations:... 4
Table of Contents Purpose... 2 Background... 2 Prelab Questions... 3 Procedure:... 3 Calculations:... 4 CHM 212 Experiment 4 Determination of the Ka of Potassium Hydrogen Phthalate (KHP) Using a Gran Plot
More informationGETTING THE END POINT TO APPROXIMATE. Two hours
Chem 1312 Handout Experiment ONE Laboratory Time Required Special Equipment and Supplies Objective Safety First Aid GETTING THE END POINT TO APPROXIMATE THE EQUIVALENCE POINT Two hours Balance Potassium
More informationRate Properties of an Iodide Oxidation Reaction
Rate Properties of an Iodide Oxidation Reaction GOAL AND OVERVIEW The rate law for the reduction reaction of peroxodisulfate (PODS) by iodide: S 2 O8 2 (aq) + 2 I (aq) I 2 (aq) + 2 SO4 2 (aq) will be determined.
More informationExperiment 2: The Rate of an Iodine Clock Reaction
Experiment 2: The Rate of an Iodine Clock Reaction Introduction: Some reactions, including most of the ones that you have seen before, occur so rapidly that they are over as soon as the reactants are mixed.
More informationChemistry 1B Experiment 17 89
Chemistry 1B Experiment 17 89 17 Thermodynamics of Borax Solubility Introduction In this experiment, you will determine the values of H and S for the reaction which occurs when borax (sodium tetraborate
More informationGENERAL INSTRUCTIONS GENERAL PREPARATION
GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required
More informationGENERAL INSTRUCTIONS GENERAL PREPARATION
GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.
More informationDetermination of the Equivalent Weight and Ionization Constant of a Weak Acid
Determination of the Equivalent Weight and Ionization Constant of a Weak Acid Introduction: The object of this experiment will be to determine the ionization constant, K a, and the equivalent weight of
More informationPractical 1P3 Electrode Potentials
Practical 1P3 Electrode Potentials What you should learn from this practical Science This experiment will familiarise you with the thermodynamics of solutions and show how easily thermodynamic quantities
More informationThe Decomposition of Hydrogen Peroxide. Evaluation copy
The Decomposition of Hydrogen Peroxide Computer 12 The decomposition of hydrogen peroxide in aqueous solution proceeds very slowly. A bottle of 3% hydrogen peroxide sitting on a grocery store shelf is
More informationCHEM 334 Quantitative Analysis Laboratory
Calibration of Volumetric Glassware Introduction Volumetric glassware is a class of glass vessels that are calibrated to contain or deliver certain volumes of substances. Graduated cylinders, pipettes
More informationThe Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.
Cole-Parmer Ammonium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium
More informationEXPERIMENT 1 REACTION RATE, RATE LAW, AND ACTIVATION ENERGY THE IODINE CLOCK REACTION
PURPOSE: To determine the Rate Law and the Activation Energy for a reaction from experimental data. PRINCIPLES: The Rate Law is a mathematical expression that predicts the rate of a reaction from the concentration
More information6 Acid Base Titration
E x p e r i m e n t Acid Base Titration Experiment : http://genchemlab.wordpress.com/-titration/ objectives To understand the concept of titration. To explain the difference between the analyte and standard
More informationPOTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA
POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA Introduction In this experiment, students will familiarize themselves with potentiometric titration, practice using the first derivative to find the equivalence
More informationCOLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com
More informationShown below is a sample titration curve for a diprotic acid. Note the two equivalence points.
EXPERIMENT 9 Titration Curve for a Polyprotic Acid INTRODUCTION Other than by strength and concentration, another way of classifying acids involves the number of H + ions an acid can donate. A monoprotic
More informationDetermining the Concentration of a Solution: Beer s Law
Determining the Concentration of a Solution: Beer s Law Vernier Spectrometer 1 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You
More informationDensity of Aqueous Sodium Chloride Solutions
Experiment 3 Density of Aqueous Sodium Chloride Solutions Prepared by Ross S. Nord and Stephen E. Schullery, Eastern Michigan University PURPOSE Determine the concentration of an unknown sodium chloride
More informationPrince George s Community College PL 2: CHARACTERIZATION OF A MONOPROTIC WEAK ACID BY POTENTIOMETRIC TITRATION
Prince George s Community College Name Section Partner(s) Date PL 2: CHARACTERIZATION OF A MONOPROTIC WEAK ACID BY POTENTIOMETRIC TITRATION PRE-LAB QUERIES 1. Complete the neutralization reactions given
More informationChapter 12 Tex-617, Determining Chloride in Concrete
Chapter 12 Tex-617, Determining Chloride in Contents: Section 1 Overview... 12-2 Section 2 Apparatus... 12-3 Section 3 Preparing Solutions... 12-4 Section 4 Procedures... 12-5 Section 5 Calculations...
More information1) A ph/mv meter or an ion meter, either line operated or portable.
COLE-PARMER Sodium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS The Cole-Parmer Sodium Ion Electrodes are used to quickly, simply, accurately, and economically measure sodium ion concentrations
More information[Agilent Technologies] [Electrochemistry Products] Quick Reference
Overview Using I9121 Ammonia Gas Sensing Probe on 3200I/M This document demonstrates how to use I 9121 Ammonia Gas Sensing Probe on 3200I or 3200 M to measure pn or the concentration of [NH 4 + ]. Applied
More informationExamining the Effect of Temperature on Reaction Rate
1 Purpose: To measure reaction rate at different temperatures for the reaction between persulfate ions, S2O8-2, and iodide ions, I -, and thereby determine the activation energy and frequency factor for
More informationLast Semester: Titration of a weak acid (vinegar) with a strong base. NaOH. Equivalence (End) Point. mols NaOH = mols Acid.
// VIRTUAL LAB & IDENTIFICATION OF AN UNKNOWN WEAK ACID RFS / Purpose: To conduct a titration, to use it to determine the Molar Mass and pk a of an unknown acid & from these two quantities, identify of
More information#13 Electrochemical Cells
#13 Electrochemical Cells If a copper strip is placed in a solution of copper ions, one of the following reactions may occur: Cu 2+ + 2e - Cu Cu Cu 2+ + 2e - The electrical potential that would be developed
More information(b) Write the chemical equation for the dissolution of NaOH in water.
Making a Solution and Measuring ph Prelab Assignment Read the entire lab. Write an objective and any hazards associated with this lab in your laboratory notebook. Answer the following 6 questions in your
More informationMETHOD 9040B. ph ELECTROMETRIC MEASUREMENT
METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least
More informationElectrochemistry LEC Potentiometric ph titration (phosphoric acid in a soft drink) What you need: What you can learn about
Electrochemistry LEC 06 What you can learn about Galvanic cell Types of electrodes Nernst equation Potentiometry Principle and tasks The cell voltage and the Galvani voltage of the electrodes of an galvanic
More informationPartner: Judy 29 March Analysis of a Commercial Bleach
Partner: Judy 29 March 2012 Analysis of a Commercial Bleach Purpose: The purpose of this lab is to determine the amount of sodium hypochlorite (NaClO) in commercial bleach. This can be done by forming
More informationSolubility Product Constant (K sp ) and the Common-Ion Effect for Calcium Iodate, a Salt of Limited Solubility
Solubility Product Constant (K sp ) and the Common-Ion Effect for Calcium Iodate, a Salt of Limited Solubility Purpose Determine the solubility product constant (K sp ) for a sparingly soluble salt. Study
More informationCOLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com
More informationExperimental Procedure. Lab 406
Experimental Procedure Lab 406 Overview This experiment is to be complete in cooperation with other chemists/chemist groups in the laboratory. In PART A, a standardized solution of hydrochloric acid is
More informationaa + bb cc + dd Equation 1
Experiment: The Determination of K eq for FeSCN 2+ Introduction For any reversible chemical reaction at equilibrium, the concentrations of all reactants and products are constant or stable. There is no
More informationAcid-Base ph Titration Introduction
Electronic Supplementary Material (ESI) for Chemistry Education Research and Practice. This journal is The Royal Society of Chemistry 2016 Appendix B: Example of Traditional Investigation Acid-Base ph
More informationTHE TEMPERATURE DEPENDENCE OF THE EQUILIBRIUM CONSTANT
Experiment 7B THE TEMPERATURE DEPENDENCE OF THE EQUILIBRIUM CONSTANT Prepared by Ross S. Nord, Chemistry Department, Eastern Michigan University PURPOSE To investigate the relationship between the equilibrium
More informationExperiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1
Experiment 13H 08/03/2017 AHRM THE REACTION OF RED FOOD COLOR WITH BLEACH 1 PROBLEM: Determine the rate law for the chemical reaction between FD&C Red Dye #3 and sodium hypochlorite. LEARNING OBJECTIVES:
More informationExperiment 7A ANALYSIS OF BRASS
Experiment 7A ANALYSIS OF BRASS FV 10/21/10 MATERIALS: Spectronic 20 spectrophotometers, 2 cuvettes, brass sample, 7 M HNO 3, 0.100 M CuSO 4, 2 M NH 3, two 50 ml beakers, 100 ml beaker, two 25 ml volumetric
More informationDensity of Aqueous Sodium Chloride Solutions
Experiment 3 Density of Aqueous Sodium Chloride Solutions Prepared by Ross S. Nord and Stephen E. Schullery, Eastern Michigan University PURPOSE Determine the concentration of an unknown sodium chloride
More informationExperiment 20: Analysis of Vinegar. Materials:
Experiment 20: Analysis of Vinegar Materials: graduated cylinder 6 M NaOH: Dilute Sodium Hydroxide 1000 ml Florence Flask & stopper KHC 8 H 4 O 4 : Potassium Hydrogen Phthalate (KHP) 125 ml Erlenmeyer
More informationPOTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background
POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA Background In this experiment, students will familiarize themselves with potentiometric titration, practice using the first derivative to find the equivalence
More informationEXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion
EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical
More informationDr. White Chem 1B Saddleback College 1. Experiment 15 Thermodynamics of the Solution Process
Dr. White Chem 1B Saddleback College 1 Experiment 15 Thermodynamics of the Solution Process Objectives To learn about the relationship between K and ΔG. To learn how the van't Hoff equation can be used
More informationTo see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide:
Weak Acid Titration v010516 You are encouraged to carefully read the following sections in Tro (3 rd ed.) to prepare for this experiment: Sec 4.8, pp 168-174 (Acid/Base Titrations), Sec 16.4, pp 769-783
More informationDetermining the Concentration of a Solution: Beer s Law
Determining the Concentration of a Solution: Beer s Law LabQuest 11 The primary objective of this experiment is to determine the concentration of an unknown nickel (II) sulfate solution. You will be using
More informationElectrochemistry LEC 06. The Nernst equation. What you need: What you can learn about. Principle and tasks
Electrochemistry LEC 06 What you can learn about Electrode potentials and their concentration dependence Redox electrodes Electrochemical cells What you need: Principle and tasks expresses how the electrical
More informationExperiment 7: ACID-BASE TITRATION: STANDARDIZATION OF A SOLUTION
Experiment 7: ACID-BASE TITRATION: STANDARDIZATION OF A SOLUTION Purpose: Determine molarity of a solution of unknown concentration by performing acid-base titrations Performance Goals: Apply the concepts
More informationEXPERIMENT NINE Part I - The Standardization of Thiosulfate Solutions
EXPERIMENT NINE Part I - The Standardization of Thiosulfate Solutions In general, thiosulfate solutions are standardized by indirect methods, Primary-standard oxidizing agents such as KIO 3, As 2 O 3,
More informationLABORATORY 2. ENZYME CATALYSIS
LABORATORY 2 STUDENT GUIDE LABORATORY 2. ENZYME CATALYSIS Objectives In this laboratory, you will observe the role of an enzyme (catalase) in conversion of hydrogen peroxide (H 2 O 2 ) to water and oxygen
More informationElectrochemical Cells
Electrochemical Cells PURPOSE To see how changes in concentration and ph affect the potential in an electrochemical cell, and confirm the Nernst equation. GOALS To examine how standard reduction potentials
More informationMETHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking
More informationEnzyme Catalysis Lab
AP Biology Name: Enzyme Catalysis Lab Objectives In this laboratory, you will observe the role of an enzyme (catalase) in conversion of hydrogen peroxide (H 2 O 2 ) to water and oxygen determine the rate
More informationKinetics; A Clock Reaction
Kinetics; A Clock Reaction Background This experiment involves the study of the rate properties, or chemical kinetics, of the following reaction between iodide ion (I - ) and bromate ion (BrO 3 - ) under
More informationThe Determination of an Equilibrium Constant
LabQuest 10 The equilibrium state of a chemical reaction can be characterized by quantitatively defining its equilibrium constant, Keq. In this experiment, you will determine the value of Keq for the reaction
More informationExperimental Procedure Lab 402
Experimental Procedure Lab 402 Overview One set of solutions having known molar concentrations of FeNCS 2+ is prepared for a calibration curve, a plot of absorbance versus concentration. A second set of
More informationSolubility of CaSO 4 Experiment 8
Solubility of CaSO 4 Experiment 8 Major Concepts and Learning Goals Application of the solubility product constant (K sp ) Saturated solutions Le Chatlier s Principle/Common ion effect Activities and activity
More informationThermodynamics of Borax Dissolution
Thermodynamics of Borax Dissolution Introduction In this experiment, you will determine the values of H, G and S for the reaction which occurs when borax (sodium tetraborate octahydrate) dissolves in water.
More informationTitration of a strong acid with a strong base with Cobra4
Titration of a strong acid with a strong base with Cobra4 TEC Related topics Strong and weak acids and bases, ph value, titration curves, equivalence point, potentiometry. Principle Hydrochloric acid is
More informationExperiment 13I THE REACTION OF RED FOOD COLOR WITH BLEACH 1
Experiment 13I FV 1/11/16 THE REACTION OF RED FOOD COLOR WITH BLEACH 1 PROBLEM: Determine the rate law for the chemical reaction between FD&C Red Dye #3 and sodium hypochlorite. LEARNING OBJECTIVES: By
More informationKa Acid Dissociation Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16
Ka Acid Dissociation Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16 I. Introduction Monoprotic acetic acid, CH 3 COOH is sometimes written as HCH 3 COO, HC
More informationDetermination of Orthophosphate Ion
Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the
More informationC = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)
GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Silver/Sulfide Ion Electrode is used to measure silver or sulfide ions in aqueous solutions. The two ions are virtually never present in solution
More informationTHE EFFECT OF TEMPERATURE AND CONCENTRATION ON REACTION RATE
THE EFFECT OF TEMPERATURE AND CONCENTRATION ON REACTION RATE INTRODUCTION FACTORS INFLUENCING REACTION RATE: The study of chemical reactions is not complete without a consideration of the rates at which
More informationDetermining the Concentration of a Solution: Beer s Law. Evaluation copy. Figure 1
Determining the Concentration of a Solution: Beer s Law Computer 17 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You will use a
More informationDetermination of Orthophosphate Ion
Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the
More informationStudies of a Precipitation Reaction
Studies of a Precipitation Reaction Prelab Assignment Read the entire lab. Write an objective and any hazards associated with this lab in your laboratory notebook. Answer the following 6 questions in your
More informationThe Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Calculator 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
More informationThis Exam Includes Three Exercises. It Is Inscribed On Three Pages Numbered From 1 to 3. The Use of a Non-programmable Calculator Is Allowed.
دورة سنة 2005 العادیة امتحانات شھادة الثانویة العامة وزارة التربیة و التعلیم العالي فرع العلوم العامة المدیریة العامة للتربیة داي رة الامتحانات الاسم: مسابقة في الكیمیاء الرقم: المدة: ساعتان This Exam
More informationChemistry Potentiometric Titration of a Chloride-Iodide Mixture
Chemistry 3200 Silver iodide, AgI, is much less soluble than AgCl. The solubility products of the two salts are 9.8 x 10 17 and 1.78 x 10 10, respectively. Therefore, if a mixture of I and Cl is titrated
More informationElectrochemistry. To use principles of electrochemistry to understand the properties of electrochemical cells and electrolysis.
Electrochemistry Objectives: To use principles of electrochemistry to understand the properties of electrochemical cells and electrolysis. Background: Part I: Galvanic Cells: A Galvanic cell is a device
More informationThe thermodynamics of the solubility of borax
Chemistry 1 6 3 The thermodynamics of the solubility of borax Purpose: To determine the thermodynamic quantities ΔH and ΔS for the solvation reaction of borax in water by measuring the solubility product
More informationThe Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.
COLE-PARMER NITRATE ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate
More informationExperiment 8 and 9 Weak Acids and Bases: Exploring the Nature of Buffers
Experiment 8 and 9 Weak Acids and Bases: Exploring the Nature of Buffers Pre-Laboratory Assignments Reading: Textbook Chapter 16 Chapter 17:1-3 This Laboratory Handout Pre-Laboratory Assignments: Complete
More information1iI1E. The Determination of 0 an Equilibrium Constant [LU. Computer
Computer The Determination of 0 an Equilibrium Constant Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant,
More informationEXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE
EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE INTRODUCTION Electrolytes are compounds that are present in solution as ions. They are more likely to be soluble in water than in most other liquids
More informationH + [ ] [ ] H + NH 3 NH 4. = poh + log HB +
Titration Lab: Determination of a pk a for an Acid and for a Base Theory A Brønsted-Lowry acid is a substance that ionizes in solution (usually aqueous, but it doesn t have to be, ammonia is often used
More informationPreLAD: b. KHP is a monoprotic acid, what are the number of moles of ionizable H + in the approximately 0.25 g of KHP?
LAD G.1 (pg! 1 of 6! ) What % of vinegar is really acetic acid? Name Per What part of the solution is really HC2H3O2 aka CH3COOH? What is the Ka of acetic acid? Introduction: The acid in vinegar is acetic
More informationDetermination of the K a of a Weak Acid and the K b of a Weak Base from ph Measurements
Experiment 6 Determination of the K a of a Weak Acid and the K b of a Weak Base from ph Measurements Pre-Lab Assignment Before coming to lab: Read the lab thoroughly. Answer the pre-lab questions that
More informationThe Thermodynamics of the Solubility of Borax
Experiment 10 Pre-Lab Assignment Before coming to lab: Read the lab thoroughly. Answer the pre-lab questions that appear at the end of this lab exercise. The questions should be answered on a separate
More informationEXPERIMENT 23 Lab Report Guidelines
EXPERIMENT 23 Listed below are some guidelines for completing the lab report for Experiment 23: For each part, follow the procedure outlined in the lab manual. Observe all safety rules, including wearing
More informationDetermining the Rate Law for a Chemical Reaction
Determining the Rate Law for a Chemical Reaction Purpose: To determine the reaction orders, rate law, and rate constant for the reaction between persulfate ions, SO8 -, and iodide ions, I - Introduction
More informationDETERMINATION OF K c FOR AN EQUILIBRIUM SYSTEM
DETERMINATION OF K c FOR AN EQUILIBRIUM SYSTEM 1 Purpose: To determine the equilibrium constant K c for an equilibrium system using spectrophotometry to measure the concentration of a colored complex ion.
More informationBuffers, Electrochemistry. Jan Pláteník & Tomáš Navrátil 2010/2011
Buffers, Electrochemistry Practical Lesson on Medical Chemistry and Biochemistry General Medicine Jan Pláteník & Tomáš Navrátil 2010/2011 1 BUFFERS AND BUFFER CAPACITY 1.1 Principle of buffering: A buffer
More informationC q T q C T. Heat is absorbed by the system H > 0 endothermic Heat is released by the system H < 0 exothermic
PLEASE REORD ALL DATA DIRETLY INTO YOUR LAB NOTEBOOKS Introduction Heating a substance is one of the simplest processes carried out in the chemical laboratory, and is usually accompanied by a rise in the
More information