1 1 Supporting Information Quantitative 1 H NMR with External Standards: Use in Preparation of Calibration Solutions for Algal Toxins and other Natural Products Ian W. Burton, Michael A. Quilliam and John A. Walter* 8 9 Institute for Marine Biosciences 10 National Research Council of Canada xford Street, Halifax N.S. Canada B3H 3Z1 12

2 12 Contents Supplementary discussion: Note: Numbering of references is the same as for the main manuscript. 15 SD1 Proportionality of NMR signal intensities to concentrations SD2. RF pulses and relaxation effects. SD3. Temperature effect on spin level populations. SD4. Probe Q-damping and effect on pulse width. SD5. Data processing and integration of spectra. 20 SD6 Summary of Conclusions Tables: Table S1. Response (integrated intensity I x p 360 ) of caffeine methyl resonances to single 90 o pulses, probe T, for varying transmitter frequencies covering the full 15 ppm spectral width. Table S2. Caffeine solutions (4 mm) in H 2 at varying NaCl concentration, Probe T: Effects of probe Q-damping on integrated resonance intensities I(H 2 ), I(caff) and p Table S3. Comparison of product of solvent resonance integral per mole and p 90, for 28 H 2 in dilute caffeine solutions, neat CHCl 3 and neat CH 3 H, probe T. 29 Table S4. Effects of mistuning and mismatching of probe T on product of integrated peak intensity I(H2) and p360, for the H 2 resonance of a caffeine solution containing trace GdCl Table S5. Concentrations of dilute solutes in H 2 measured by NMR against the integral of the H 2 solvent signal, compared with determinations using gravimetry or NMR using an external standard. 2

3 Table S6. Caffeine, theophylline, arginine and saxitoxin (STX) solutions in H 2 : product of integral I(H 2 ) of solvent resonance (single 90 o pulse) and p 90 for probe X, over a 3 week period. 38 Table S7. Linearity of response: Caffeine average resonance intensity (3 samples, 39 methyl resonances) vs gravimetric concentration (mm), probe X. 40 Table S8. Sucrose Concentrations determined by NMR with probe X against five external caffeine standards prepared gravimetrically, compared with gravimetric concentrations of same sucrose solutions. Figures: Fig. S1. Structures of compounds used for concentration determinations by NMR, including the shellfish toxins domoic acid, gymnodimine, okadaic acid and some of the Paralytic Shellfish Poisoning (PSP) toxins. Fig. S2. Effects of increased probe damping on signal intensity and p 90 for probe T. Fig. S3. Linearity of response and comparison of integration of same datasets by two different operators. Fig. S4. Concentrations of sucrose solutions determined by interpolation of 1 H NMR integrals against a calibration curve of caffeine integrals vs gravimetric concentration, probe X. Fig. S5. Concentration of theophylline solutions determined by NMR with a 90 o pulse against 4 mm standard caffeine solutions, vs gravimetric concentration of theophylline, probe X. Fig. S6. Concentration of theophylline solutions determined by NMR with 47 o pulse against 4 mm standard caffeine solutions, vs gravimetric concentration of theophylline, probe X. 3

4 59 60 SD1 Proportionality of NMR signal intensities to concentrations. We will restrict discussion to pulse Fourier-transformed spectra integrated in the frequency domain. The strength I A of the 1 H NMR signal (per proton in a given chemical environment) produced by an analyte A, compared to that (I S ) of a standard S recorded under identical conditions, is proportional to the ratio N A /N S, where N A and N S are the respective numbers of 1 H nuclei in A and S, assumed to have the same geometrical distribution within the sensitive area of the probe. Where protons are in more than one chemically-distinct environment, independent measurements of I may be obtainable from one spectrum. As N A and N S are directly proportional to the concentrations C A and C S respectively, I A = k A C A and I S = k S C S and the molar ratio C A /C S may be determined directly from the intensity ratio if k A = k S. The experimental tests and discussion cover the steps necessary to ensure k A = k S when standards and analytes are in separate tubes SD2. RF pulses and relaxation effects Rapid repetition of a one-pulse acquisition sequence produces partial saturation of resonances. This can be avoided by use of a flip angle << 90 o, 18 but a 90 o pulse with increased delay between pulses lessens the dependence of signal strength on precise setting of the pulse length. As the objective is to avoid saturation effects, the ultimate S/N per unit time will be similar no matter which alternative is adopted. The total time between pulses is normally the sum of a very short delay of a few µsec to allow recovery from the pulse, a fixed acquisition time ca 2-4 s, and a relaxation delay of 0-20 s. With 90 o pulses virtually complete relaxation (to within 0.7% of the maximum) occurs after a time 5T 1, where T 1 is the spin-lattice relaxation time for a given nucleus. T 1 can also be 4

5 measured if necessary for assurance of complete relaxation. The 90 o pulses should be short enough to provide uniform excitation over the full spectral width SW (i.e., reciprocal of pulse length >> SW). T 1 s are normally sufficiently long that relaxation during the recovery delay is negligible SD3. Temperature effect on spin level populations When using external standards for NMR quantitation, temperatures must be the same for both analyte and standard samples as each is assumed to have the same equilibrium Boltzmann distribution of spin level populations. For two spin (I = ½) levels in a magnetic field B o at high absolute temperature T A, such that γhb o /4π << kt A, the equilibrium spin population difference is closely approximated by: p 1 p 2 = p 1 γhb/2πkt A = p A where γ is the nuclear gyromagnetic ratio, and k is Boltzmann s constant. For a small change in temperature from T A to T B, (where T << T A ) the proportional change in population difference is therefore: ( p A p B )/ p A = T/T B Thus at room temperature (20 o C = 293 o K), a 1 o K increase of temperature would cause a 1/293 (0.34 %) decrease in spin level population difference, and a corresponding decrease in the detected magnetization SD4. Probe Q-damping and effect on pulse width The effects of sample RF susceptibility and conductivity on the probe response ( Q factor of the tuned circuit) may be profound if the sample tube diameter is large 5

6 compared to the separation of the probe coils. The effects can be visualized with the probe tuning display (RF bridge and swept frequency source) usually provided with current commercial NMR spectrometers, or by an external sweep generator and bridge connected to the probe. For all quantitative work with a separate external standard, the probe must be tuned and matched to the preamplifier (usually 50 Ω) for each sample, at the observing frequency. As probe tuning usually varies during sample spinning, samples should be static. A highly-conducting sample such as a salt solution will broaden the matched probe response corresponding to a lowered (damped) Q. As the intensity of the NMR signal is proportional to Q, 27 damping is a major factor preventing the condition k A = k S from being satisfied. However, increased damping also increases the 90 o (and 360 o ) pulse lengths (p 90 and p 360 respectively), and these readily measurable quantities can be used as a check and a correction for damping when the same coil provides both RF irradiation (at constant power) and detection of the nuclear magnetization, as in the probes used for our experiments. With fully protonated solvents, the probe damping may also be measured and compensated by integrating the solvent resonances of analyte and standard following a single 90 o pulse without presaturation, as the 1 H concentration for each solvent is calculable for dilute solutions from the known solvent densities. When RF irradiation and NMR detection are performed with the same coil, and all amplifiers are tuned and matched at the resonant frequency so that their characteristics are reproducible between samples, an inverse relation between p 90 and signal intensity is consistent with the principle of reciprocity. This states 28 that the emf ξ induced in an arbitrary conductor (coil) by time-dependent motion of a magnetic moment vector m at 6

7 some point in its vicinity, is related to the magnetic field vector b that would be produced at that point by unit current in the conductor: ξ = -( / t)(b.m). The total emf is the sum of contributions ξ j from nuclear moments across the whole sample. 28 During RF pulse irradiation, at an arbitrary position j, p 90 1/(B 1j ) XY, where (B 1j ) XY is the component at point j of the alternating RF field B 1 in the XY plane perpendicular to the static field B 0 defined along the Z axis. If B 1j makes an angle θ to Z (which is also the direction of the equilibrium magnetization): p 90j = k 1 /[ B 1j Sin θ] The measurement of p 90 (or p 360, which may be determined more precisely), quantifies the effect of the medium (damping) on the magnitude of the RF magnetic field produced by the coil at a particular point j. Media of higher conductivity will result in longer p 360. The signal (emf) in the same coil, produced by precessing nuclei at the same position j, may be measured in response to a pulse p X producing arbitrary (but known) flip angle, usually but not necessarily 90 o. Following the pulse p X, the nuclear magnetization component in the XY plane, (m j ) XY, precessing at constant frequency ω about the Z direction, induces an emf ξ j given by: ξ j = - ( / t)(b j.m j ) = k 2 ( / t)(b 1j.m j ) as b j is in the same direction as B 1j. The vector b j is not time dependent. The time derivative of m j is: i(ωt + φ) ( / t)m j = iω (m j ) XY e where φ is an arbitrary phase angle. Therefore ξ j = iω m jxy e i(ωt + φ) k 2 B 1j Sinθ 7

8 The integrated signal intensity I j recorded after amplification, detection and Fourier transformation is proportional to the amplitude of ξ j : I j = k 3 ω m jxy k 2 B 1j Sinθ Therefore the product p 90j I j = { k 1 /[ B 1j Sin θ]}{k 3 ω m jxy k 2 B 1j Sinθ} = k 1 k 2 k 3 ω m jxy i.e. p 90j I j has a constant proportionality to m jxy at an arbitrary point j in the sample, provided that the flip angle p X is constant and the same coil is used for excitation and detection. RF susceptibility and conductivity effects which influence the coupling of the precessing nuclear magnetization to the coil will have the same effect on the radiofrequency field (produced by the same coil at the same frequency) that causes the excitation of the same parcel of nuclei. When a measurement of p 90 = p 360 /4 is made, the null at the 360 o pulse is a sample average, as the criterion is zero integral of the dispersive residual signal. Some nuclear magnetization elements are rotated by more, and some by 163 less, than 360 o due to B 1 inhomogeneities. When comparing different samples the distribution of B 1 inhomogeneities will be the same as the samples extend well beyond the sensitive region of the coil. Magnetic susceptibility-induced differences in B 1 distribution should be negligible in these slightly paramagnetic or diamagnetic solutions, so the p 360 measurement should cover the same geometrical distribution of nuclear magnetization in different samples. The fact that any change in Ι due to dielectic and conductivity effects in the sample is exactly the same as the proportional change in 1/p 90, provides a direct means for compensation of probe damping effects when comparing different samples, and removes one of the major obstacles to quantitative use of NMR with separate external standards. 8

9 Probe damping may be minimized by use of a detection coil that is large in comparison to the sample tube diameter, or by reducing the concentration of conductive samples. An alternative approach (the ERETIC method 22,23 ) is to provide a small exponentiallydamped RF signal, at a frequency within the detected range, to an untuned (e.g. 13 C) coil in the NMR probe, and calibrate the area of the resulting peak to the resonance areas for a standard sample. It is claimed that this peak area will be influenced by probe damping to the same extent as 1 H resonances; however, for quantitative comparisons it would still be necessary to calibrate changes in flip angle SD5. Data processing and integration of spectra Factors affecting the accuracy and precision in data processing of quantitative NMR spectra have been discussed in general 3,7,9,26 and in the context of natural products. 18 We will restrict discussion to Fourier-transformed spectra integrated in the frequency domain. Free induction decays (FID s) must be processed with apodization functions that leave the first point of the FID unaltered and do not cause excessive line broadening, e.g. a decreasing exponential corresponding to line broadening less than a typical linewidth, or a half-gaussian function having value 1.0 at zero time. Alteration of the first data point by apodization will change the integrated areas of the tranformed spectrum rendering quantitation impossible. Following Fourier transformation, baseline correction may be necessary, for which it is preferable to use reproducible means such as subtraction of linear or simple polynomial 194 segments over small regions of the spectrum. 18 Integration should be performed 195 numerically taking care to include all the area in the wings of individual resonances. 9

10 Baselines should be adjusted so that integral displays are horizontal at the start and end of the integrated range, and should be performed at a constant scaling factor for all samples. When lines are Lorentzian, the integral range ± R must be large compared to the linewidth ν at ½ height, to incorporate all the area that is significant above noise: e.g. to incorporate 99% or more of the area requires R/ ν > In practice, this restriction contributes little to inaccuracy in integration of spectra from natural products as linewidths are usually comparable across the spectrum and are only approximately Lorentzian in shape owing to residual field inhomogeneities. The inclusion of 1 H 13 C satellite resonances within particular integration ranges has to be considered on a caseby-case basis, 31 as their overlap with 1 H 12 C resonances varies with field as well as from one compound to another. Published studies of integration reproducibility with 1 H spectra of a drug 31 and a pure natural product 18 showed integration errors of 0.5% to 2%, repeatability within one laboratory being better than reproducibility between different laboratories SD6 Summary of Conclusions Quantitative NMR using precision tubes and external standards enables the direct measurement of analyte concentrations without contaminating the sample with internal standards or compromising subsequent use of the solutions with other techniques, such as LC - MS. The method is also advantageous for use with hazardous materials as samples may be kept in sealed tubes throughout. Expensive deuterated solvents can be avoided, overlap of standard with analyte resonances is not a concern, impurities in the standard are physically separated from the analyte, and standards may be prepared with greater 10

11 accuracy. The 1 H spectrum provides in addition a measure of the analyte purity and integrity. Analytes and standards in different solvents or in salt solutions may be compared by quantitative corrections of Q-damping effects via ratios of analyte/standard p 360 values, or by measurements of solvent resonance integrals following a single 90 o pulse. The factors affecting the accuracy and precision of quantitative NMR using external standards have been explored and their relative influence is summarized in Table 2. Provided that all samples are accurately tuned and matched at the observing frequency, and that sample temperatures are constant, and that adequate relaxation delays are used, the major contributions to error arise from the accuracy and purity of the gravimetrically prepared external standards, the purity of the sample (particularly freedom from protoncontaining impurities producing resonances that overlap with the analyte, polymerization of the analyte, and macromolecules), subjectivity in the integration of spectra, and intermolecular NEs if presaturation is used to suppress resonances of non-deuterated solvents. Integration subjectivity is highly sample-dependent. With pure analytes having sharp well-separated resonances (e.g. caffeine) different experienced operators processing and integrating the same dataset can agree to better than 0.3%. Results with natural products that have more complex spectra and which may contain residual protonated impurities are influenced by the choice of peaks to integrate and baseline correction, and the only option may be to average the results obtained by two or more operators on the same dataset. Spectral overlaps with proton-containing impurities may sometimes be detected by 2D TCSY spectra. Impurities that adsorb the analyte will lead to broadening or 11

12 indetectability of resonances and must be avoided. These problems also occur to the same extent when internal standards are used and may be further compounded by overlap of analyte and standard resonances. Intermolecular NE s are not accurately predictable, although as illustrated by the agreement between NMR concentration measurements and other methods (Table 1, Fig. 2) they are likely to be negligible within error for with most 1 H resonances in natural products. A practical strategy is to disregard peaks for quantitation if their intensity is significantly enhanced relative to others, and to employ structural knowledge of the molecule to decide which H atoms are likely candidates for this effect. The potential problems peculiar to NMR with external standards contribute little to the overall accuracy and precision. With single-coil RF irradiation and detection, differences in damping between analyte and standard may be corrected from the ratio of 360 o pulse lengths. A series of measurements with two probes and a variety of solvents and solutes showed that the product of the molar response of the solvent resonance and the 360 o (or 90 o ) pulse was constant, for a given probe, to better than 1%, the standard deviation being typically less than 0.5%. This error also includes any variation in sample tube diameter, and is consistent with the tolerances claimed by the manufacturer. ther possible sources of variation included within this figure are temperature effects on level populations, drift of total amplifier gain, digitizer response and integration. Constancy of the total response over a period of months has been demonstrated so it is possible to use the technique with confidence using periodic calibrations against standards. 12

13 Although we have worked with samples in non-deuterated solvents owing to the ultimate uses of the reference materials, the methods are more readily utilized with deuterated solvents as presaturation is unnecessary and field-frequency lock can be used. We have also shown that dilute solutions (mm range) of analytes in otherwise pure non-deuterated solvents may be quantified with a spectrometer having highly linear amplification stages and accurately calibrated receiver attenuation, by recording the 269 solvent-suppressed analyte signal with multiple 90 o pulses at high gain, against the solvent signal from the same solution recorded from a single 90 o pulse at low gain. This method is faster as it does not require external standards, exact tuning, precisely matched sample tubes, damping corrections, or multiple samples, and it could be used with probes having separate irradiation and detection coils. Necessary precautions are to ensure dryness of solvents, and the same signal path in the instrument for measurements of solvent and analyte signals

14 Table S1. Response (integrated intensity I x p 360 ) of caffeine methyl resonances to single 90 o pulses, probe T, for varying transmitter frequencies covering the full 15 ppm spectral width. Sample was mm caffeine in D 2 with added GdCl 3 for rapid relaxation (linewidth of caffeine resonances ca 4 Hz, delay between pulses > 60s). Effect of varying transmitter offset on integrated peak intensity I x p360 Sample: Caffeine mm in D2 plus trace GdCl3 Me resonances (pk 2, pk 3, pk 4) Single 90 deg pulses, no presat, p360 = 23.3 us, delay between pulses 60s, temp 20C Transmitter I x p360 of pk 2 I x p360 of pk 3 I x p360 of pk 4 offset (Hz) Transmitter on residual H peak Transmitter on pk Distance from edge of SW: pk 4 at 430Hz pk 2 at 264Hz, pk 3 at 32 Hz Av: SD: Av excl : SD: Numbers in italics not used in averages. Response uniform to 1.0% SD across 15 ppm (7508 Hz) SW to within 264 Hz of edge. This includes both filtering and pulse power contributions to non-uniformity of response. 14

15 Table S2. Caffeine solutions (4 mm) in H 2 at varying NaCl concentration, Probe T: Effects of probe Q-damping on integrated resonance intensities I(H 2 ), I(caff) and p 90. NaCl conc p(90) I(H2) I(H2) x p(90) Av I(caff)/H Av I(caff)/H Av I(H2) xp90 mm (µs) x p(90) per mole H Av: sd: Note: concentration of H 2 in a 100 mm NaCl solution at 20 o C is that of pure H 2 (CRC Handbook of Chemistry and Physics, 60 th edition, 1979, p D-261). 15

16 Table S3. Comparison of product of solvent resonance integral per mole and p 90, for H 2 in dilute caffeine solutions, neat CHCl 3 and neat CH 3 H, probe T. These tests were run four months later than those in Table S1 and used a different set of RF filters, so absolute intensities are not strictly comparable with those in Table S Neat solvent I(solv)/H Av solv x p90 Sample concentration single scan per Mole (M) p(90) Caff 1mM, NaCl 1mM T&M T&M T&M Caff 1mM, NaCl 100mM T&M CHCl3 Neat T&M MeH Neat T&M NS=1 T&M T&M Av: sd: Caff 1mM, NaCl 1mM Mismatch Caff 1mM, NaCl 100mM Mismatch T&M = probe tuned and matched 16

17 Table S4. Effects of mistuning and mismatching of probe T on product of integrated peak intensity I(H2) and p360, for the H 2 resonance of a caffeine solution containing trace GdCl 3. Effect of mistuning and mismatching on integrated peak intensity I x p360 Sample: Caffeine mm in H2 plus trace GdCl3 Probe T Single 90 deg pulses, no presat, delay between pulses 60s, temp 20C I (H2) is integrated intensity of H2 resonance plus caffeine peaks (I x p360) / p360 (us) I (H2) I x p360 (I x p360) tuned & matched Tuned and matched: Tuned, not matched: (match capacitor at one extreme of range) Tuned, not matched: Match capacitor at other extreme of range) Matched, not tuned: (gross mistuning) Matched, not tuned: (transmitter frequency at 1/3 of tuning display depth) 17

18 Table S5. Caffeine, theophylline, arginine and saxitoxin (STX) solutions in H 2 : product of integral I(H 2 ) of solvent resonance (single 90 o pulse) and p 90 for probe X, over a 3 week period. Day when Compound Conc NaCl conc data I(H2) p360 p90 p90*i(h2) recorded Caffeine 4 mm Caffeine 4 mm 1 mm Caffeine 4 mm 5 mm Caffeine 4 mm 10 mm Caffeine 4 mm 20 mm Caffeine 4 mm 50 mm Caffeine 4 mm 100 mm Theophylline 2 mm Theophylline 2 mm Theophylline 2 mm Theophylline 4 mm Theophylline 4 mm Theophylline 4 mm Theophylline 1mM Theophylline 1mM Theophylline 1mM USP Caffeine 4 mm USP Caffeine 4 mm USP Caffeine 4 mm sd Av all Caffeine, Theophylline: Av of solutions without NaCl: STX STX STX Arginine 4 mm Arginine 4 mm Arginine 4 mm STX STX STX Av all STX, Arginine: Av all solutions:

19 Table S6. Linearity of response: Caffeine average resonance intensity (3 samples, methyl resonances) vs gravimetric concentration (mm), probe X. Caffeine: Signal intensity vs Gravimetric concentration, Probe X Grav conc Av integral verall Av SD (mm) ea. Sample

20 Table S7. Sucrose Concentrations determined by NMR with probe X against five external caffeine standards prepared gravimetrically, compared with gravimetric concentrations of same sucrose solutions. Peaks showing anomalous intensity attributable to presaturation (caffeine CH, two sucrose ranges) have not been used. Probe damping did not alter between solutions, so corrections for this were not needed. Caffeine std conc (mm) Suc conc Gravimetric by caffeine sucrose conc Sucrose conc (mm) by NMR against each caffeine std Av suc conc calib. Curve (mm) Caff conc x Av I(suc) / Av I(caff) by NMR (mm) sd (mm)

21 Table S8. Concentrations of dilute solutes in H 2 measured by NMR against the integral of the H 2 solvent signal, compared with determinations using gravimetry or NMR using an external standard: S8(a) 40 mm caffeine; S8(b) domoic acid; S8(c) decarbomyl neosaxitoxin (dcne). For S8 (a) and S8(b) the integrals of the analyte and H 2 signals were taken from the same spectrum measured with probe T. The calculations assume negligible change in H 2 concentration ( M at 20 o C) due to presence of analyte. Comparisons of domoic acid with caffeine standards were corrected for differences in p 360. Table S8(a) and S8(b): Sample Gravimetric Conc from SD Conc from SD conc (mm) H2 signal ext caffeine (mm) (mm) (a) Caffeine Av: (b) Domoic acid Av: Error incl. rel. error of caffeine stds: For S8 (c), the integrals for the analyte were measured with a presaturation sequence at a high receiver gain setting (RG = 32) while the H 2 integral was measured with a single pulse sequence at low gain (RG = 1), using Probe X. The gain ratio was calibrated by comparing the integrals of a residual HD signal in D 2 (sealed sample) seven times at each RG setting, using a single pulse sequence with a 60 s delay. 21

22 339 Table S8(c). Decarbamoyl neosaxitoxin, probe X. Concentration (mm) of two decarbamoyl neosaxitosin (dcne) solutions determined against caffeine, with p90 correction, Probe X Range # Norm I*p90/scan sd Conc sd of means dcne A Av dcne B Av Av of dcne A and dcne B Calibration of gain ratio for attenuator settings RG 32 / RG 1 The following attenuator ratios obtained with same sample of HD in D2, p360=84.0us, p90=21.0us Attenuator ratios: RG32/RG1 RG = 1 RG = 32 RG32/RG1 Dataset # Integ tox0323y 1, , tox0326y 1, , , , , Av: sd 0.37 Concentration (mm) of the same dcne solutions determined against solvent H2 No p90 correction needed as H2 and dcne in same sample Probe X Range # Norm I/scan sd Conc sd dcne A Conc of dcne A: Av sd of means dcne B Conc of dcne B: Av sd of means Av of dcne A and dcne B:

23 343 H H 3 C N CH 3 N N N H H N N N N CH 3 H H H H H H H CH 3 CH 3 Caffeine Theophylline Sucrose H H CH 3 H 3 C CH CH H H N CH N Domoic Acid H Gymnodimine H H 3 C H kadaic Acid CH 3 H CH 3 H H H H 3 C H CH 3 H R 1 R 2 R 3 R R 1 R 4 H NH N NH 2 + STX H H H NE H H H GTX2 H H S - 3 NH 2 GTX3 H S - + H 2 N N NH 3 H H GTX1 H H S3- H GTX4 H S - 3 H R 2 R 3 GTX5 H H H dcstx H H H dcgtx2 H H S - 3 H dcgtx3 H S - 3 H NHS 3 Fig. S1. Structures of compounds used for concentration determinations by NMR, including the shellfish toxins domoic acid, gymnodimine, okadaic acid and some of the Paralytic Shellfish Poisoning (PSP) toxins

24 348 H2 Integral/H TBI Av Caff Int/H (2-4) TBI H2 Integral y = x R 2 = Integral y = x R 2 = [NaCl] mm [NaCl] mm a. y = -( ± )x ± 0.35 b. y = -( ± )x ± R 2 = R 2 = deg (us) TBI 90 deg pulse length (us) y = x R 2 = [NaCl] mm c. y = ( ± )x ± R 2 = Fig. S2. Effects of increased probe damping on signal intensity and p 90 for probe T: Concentration dependence of: a. I(H 2 ); b. I(caff); c. p 90 for 4 mm caffeine solutions containing increasing concentrations of NaCl. 24

25 Integral Integral y = (430.4 ± 1.53)x 25.8 ± 36.9, R 2 = Caffeine integral vs gravimetric concentration: perator B y = 430.4x R 2 = [Caffeine] mm Caffeine integral vs gravimetric concentration: perator A y = 429.5x R 2 = [Caffeine] mm y = (429.5 ± 1.28)x 32.7 ± 31.1, R 2 = Fig. S3. Linearity of response and comparison of integration of same datasets by two different operators: Integral of caffeine methyl resonances (in H 2 solution) vs gravimetric concentration (av. of 3 solutions at each concentration, arbitrary integral units, integrated on same scale, probe X). The preamplifier used in these experiments was subsequently replaced. 25

26 Caffeine integral a. y = (429.5 ± 1.28)x 32.7 ± 31.1, R 2 = NMR conc. of sucrose (mm) Caffeine integral vs gravimetric concentration: calibration curve 0 y = 429.5x R 2 = Caffeine gravimetric conc (mm) Sucrose quantitation by caffeine calibration curve y = x R 2 = Gravimetric conc. of sucrose (mm) b. y = ( ± )x ± , R 2 = Fig. S4. Concentrations of sucrose solutions determined by interpolation of 1 H NMR integrals against a calibration curve of caffeine integrals vs gravimetric concentration, probe X: a. Caffeine calibration curve. b. Sucrose concentrations by NMR from interpolation of caffeine calibration curve, vs gravimetric sucrose concentration. See Table S7. 26

27 Theo NMR conc against caff (mm) a. y = ( ± )x, R 2 = Theo NMR conc against caff (mm) Theophylline NMR conc vs gravimetric conc Theophylline NMR conc vs gravimetric conc y = x R 2 = Gravimetric Conc (mm) y = x R 2 = Gravimetric Conc (mm) b. y = ( ± )x ± , R 2 = Fig. S5. Concentration of theophylline solutions determined by NMR with a 90 o pulse against 4 mm standard caffeine solutions, vs gravimetric concentration of theophylline, probe X: a. Linear regression through origin and data points. b. Linear regression through data points only. 27

28 Theo NMR conc against caff (mm) a. y = ( ± )x, R 2 = Theo NMR conc against caff (mm) Theophylline NMR conc vs gravimetric conc determined against 4 mm caffeine, 47 deg pulse y = x R 2 = Gravimetric Conc'n (mm) Theophylline NMR conc vs gravimetric conc determined against 4 mm caffeine, 47 deg pulse y = x R 2 = Gravimetric Conc'n (mm) b. y = ( ± )x ± , R 2 = Fig. S6. Concentration of theophylline solutions determined by NMR with 47 o pulse against 4 mm standard caffeine solutions, vs gravimetric concentration of theophylline, probe X: a. Linear regression through origin and data points. b. Linear regression through data points only. 28

Measuring Spin-Lattice Relaxation Time

Measuring Spin-Lattice Relaxation Time WJP, PHY381 (2009) Wabash Journal of Physics v4.0, p.1 Measuring Spin-Lattice Relaxation Time L.W. Lupinski, R. Paudel, and M.J. Madsen Department of Physics, Wabash College, Crawfordsville, IN 47933 (Dated:

More information

NMR PRAKTIKUM. Data processing Data acquisition... 17

NMR PRAKTIKUM. Data processing Data acquisition... 17 NMR PRAKTIKUM 1. INTRODUCTION... 2 1.1. Description of a Spectrometer... 2 1.2. Principle of a NMR Experiment... 4 1.2.1. 1D NMR experiment... 4 1.2.2. 2D NMR experiment... 5 2. PRACTICAL PART... 8 2.1.

More information

Chapter 7. Nuclear Magnetic Resonance Spectroscopy

Chapter 7. Nuclear Magnetic Resonance Spectroscopy Chapter 7 Nuclear Magnetic Resonance Spectroscopy I. Introduction 1924, W. Pauli proposed that certain atomic nuclei have spin and magnetic moment and exposure to magnetic field would lead to energy level

More information

Chemistry 431. Lecture 23

Chemistry 431. Lecture 23 Chemistry 431 Lecture 23 Introduction The Larmor Frequency The Bloch Equations Measuring T 1 : Inversion Recovery Measuring T 2 : the Spin Echo NC State University NMR spectroscopy The Nuclear Magnetic

More information

NMR, the vector model and the relaxation

NMR, the vector model and the relaxation NMR, the vector model and the relaxation Reading/Books: One and two dimensional NMR spectroscopy, VCH, Friebolin Spin Dynamics, Basics of NMR, Wiley, Levitt Molecular Quantum Mechanics, Oxford Univ. Press,

More information

Spectral Broadening Mechanisms

Spectral Broadening Mechanisms Spectral Broadening Mechanisms Lorentzian broadening (Homogeneous) Gaussian broadening (Inhomogeneous, Inertial) Doppler broadening (special case for gas phase) The Fourier Transform NC State University

More information

Polarised Nucleon Targets for Europe, 2nd meeting, Bochum 2005

Polarised Nucleon Targets for Europe, 2nd meeting, Bochum 2005 Polarised Nucleon Targets for Europe, nd meeting, Bochum Temperature dependence of nuclear spin-lattice relaxations in liquid ethanol with dissolved TEMPO radicals H. Štěpánková, J. Englich, J. Kohout,

More information

PROTEIN NMR SPECTROSCOPY

PROTEIN NMR SPECTROSCOPY List of Figures List of Tables xvii xxvi 1. NMR SPECTROSCOPY 1 1.1 Introduction to NMR Spectroscopy 2 1.2 One Dimensional NMR Spectroscopy 3 1.2.1 Classical Description of NMR Spectroscopy 3 1.2.2 Nuclear

More information

Ferdowsi University of Mashhad

Ferdowsi University of Mashhad Spectroscopy in Inorganic Chemistry Nuclear Magnetic Resonance Spectroscopy spin deuterium 2 helium 3 The neutron has 2 quarks with a -e/3 charge and one quark with a +2e/3 charge resulting in a total

More information

Relaxation, Multi pulse Experiments and 2D NMR

Relaxation, Multi pulse Experiments and 2D NMR Relaxation, Multi pulse Experiments and 2D NMR To Do s Read Chapter 6 Complete the end of chapter problems; 6 1, 6 2, 6 3, 6 5, 6 9 and 6 10. Read Chapter 15 and do as many problems as you can. Relaxation

More information

NMR Spectroscopy Laboratory Experiment Introduction. 2. Theory

NMR Spectroscopy Laboratory Experiment Introduction. 2. Theory 1. Introduction 64-311 Laboratory Experiment 11 NMR Spectroscopy Nuclear Magnetic Resonance (NMR) spectroscopy is a powerful and theoretically complex analytical tool. This experiment will introduce to

More information

τ 1 > 1/J - if this lifetime is significantly shortened, the coupling (splitting of the signal) will not be observed

τ 1 > 1/J - if this lifetime is significantly shortened, the coupling (splitting of the signal) will not be observed It is often advantageous to reverse or remove the splitting caused by spin-spin coupling This is called spin decoupling Spin decoupling (or just decoupling) can be used for several reasons - to simplify

More information

Physical Background Of Nuclear Magnetic Resonance Spectroscopy

Physical Background Of Nuclear Magnetic Resonance Spectroscopy Physical Background Of Nuclear Magnetic Resonance Spectroscopy Michael McClellan Spring 2009 Department of Physics and Physical Oceanography University of North Carolina Wilmington What is Spectroscopy?

More information

Biochemistry 530 NMR Theory and Practice

Biochemistry 530 NMR Theory and Practice Biochemistry 530 NMR Theory and Practice Gabriele Varani Department of Biochemistry and Department of Chemistry University of Washington Lecturer: Gabriele Varani Biochemistry and Chemistry Room J479 and

More information

Spectral Broadening Mechanisms. Broadening mechanisms. Lineshape functions. Spectral lifetime broadening

Spectral Broadening Mechanisms. Broadening mechanisms. Lineshape functions. Spectral lifetime broadening Spectral Broadening echanisms Lorentzian broadening (Homogeneous) Gaussian broadening (Inhomogeneous, Inertial) Doppler broadening (special case for gas phase) The Fourier Transform NC State University

More information

Overhauser Magnetometers For Measurement of the Earth s Magnetic Field

Overhauser Magnetometers For Measurement of the Earth s Magnetic Field Overhauser Magnetometers For Measurement of the Earth s Magnetic Field By: Dr. Ivan Hrvoic GEM Systems Inc. (Magnetic field Workshop on Magnetic Observatory Instrumentation Espoo, Finland. 1989) TABLE

More information

Chem 325 NMR Intro. The Electromagnetic Spectrum. Physical properties, chemical properties, formulas Shedding real light on molecular structure:

Chem 325 NMR Intro. The Electromagnetic Spectrum. Physical properties, chemical properties, formulas Shedding real light on molecular structure: Physical properties, chemical properties, formulas Shedding real light on molecular structure: Wavelength Frequency ν Wavelength λ Frequency ν Velocity c = 2.998 10 8 m s -1 The Electromagnetic Spectrum

More information

NMR Spectroscopy: A Quantum Phenomena

NMR Spectroscopy: A Quantum Phenomena NMR Spectroscopy: A Quantum Phenomena Pascale Legault Département de Biochimie Université de Montréal Outline 1) Energy Diagrams and Vector Diagrams 2) Simple 1D Spectra 3) Beyond Simple 1D Spectra 4)

More information

NMR course at the FMP: NMR of organic compounds and small biomolecules - II -

NMR course at the FMP: NMR of organic compounds and small biomolecules - II - NMR course at the FMP: NMR of organic compounds and small biomolecules - II - 16.03.2009 The program 2/76 CW vs. FT NMR What is a pulse? Vectormodel Water-flip-back 3/76 CW vs. FT CW vs. FT 4/76 Two methods

More information

Biophysical Chemistry: NMR Spectroscopy

Biophysical Chemistry: NMR Spectroscopy Spin Dynamics & Vrije Universiteit Brussel 25th November 2011 Outline 1 Pulse/Fourier Transform NMR Thermal Equilibrium Effect of RF Pulses The Fourier Transform 2 Symmetric Exchange Between Two Sites

More information

1H 1D-NOE Difference Spectra and Spin-Saturation Transfer Experiments on the GN500

1H 1D-NOE Difference Spectra and Spin-Saturation Transfer Experiments on the GN500 UGN526 VVM-21JUN88CD VVM-31OCT91UD 1H 1D-NOE Difference Spectra and Spin-Saturation Transfer Experiments on the GN500 Double-resonance experiments are techniques which use a second irradiating field (B

More information

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution.

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution. Introduction to MRI Spin & Magnetic Moments Relaxation (T1, T2) Spin Echoes 2DFT Imaging Selective excitation, phase & frequency encoding K-space & Spatial Resolution Contrast (T1, T2) Acknowledgement:

More information

The NMR Inverse Imaging Problem

The NMR Inverse Imaging Problem The NMR Inverse Imaging Problem Nuclear Magnetic Resonance Protons and Neutrons have intrinsic angular momentum Atoms with an odd number of proton and/or odd number of neutrons have a net magnetic moment=>

More information

Carbon 13 NMR NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY

Carbon 13 NMR NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY PRINCIPLE AND APPLICATION IN STRUCTURE ELUCIDATION Carbon 13 NMR Professor S. SANKARARAMAN Department of Chemistry Indian Institute of Technology Madras Chennai

More information

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics Magnetic Resonance Imaging Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics pal.e.goa@ntnu.no 1 Why MRI? X-ray/CT: Great for bone structures and high spatial resolution Not so great

More information

Ammonium Sulfate as a Standard for 33S-NMR Spectra

Ammonium Sulfate as a Standard for 33S-NMR Spectra 612 J. Jpn. Oil Chem. Soc. (YUKAGAKU) ORIGINAL Ammonium Sulfate as a Standard for 33S-NMR Spectra Yoshio KOSUGI Department of Applied Chemistry, School of Engineering, Nagoya University (Chikusa ku, Nagoya-shi,

More information

- Basic understandings: - Mapping interactions:

- Basic understandings: - Mapping interactions: NMR-lecture April 6th, 2009, FMP Berlin Outline: Christian Freund - Basic understandings: Relaxation Chemical exchange - Mapping interactions: -Chemical shift mapping (fast exchange) Linewidth analysis

More information

Nuclear Magnetic Resonance

Nuclear Magnetic Resonance Nuclear Magnetic Resonance PRINCIPLES OF NMR SPECTROSCOPY Contents Principles of nuclear magnetic resonance The nmr spectrometer Basic principles in nmr application NMR tools used to obtain information

More information

The Basics of Magnetic Resonance Imaging

The Basics of Magnetic Resonance Imaging The Basics of Magnetic Resonance Imaging Nathalie JUST, PhD nathalie.just@epfl.ch CIBM-AIT, EPFL Course 2013-2014-Chemistry 1 Course 2013-2014-Chemistry 2 MRI: Many different contrasts Proton density T1

More information

Physikalische Chemie IV (Magnetische Resonanz) HS Solution Set 2. Hand out: Hand in:

Physikalische Chemie IV (Magnetische Resonanz) HS Solution Set 2. Hand out: Hand in: Solution Set Hand out:.. Hand in:.. Repetition. The magnetization moves adiabatically during the application of an r.f. pulse if it is always aligned along the effective field axis. This behaviour is observed

More information

C NMR Spectroscopy

C NMR Spectroscopy 13.14 13 C NMR Spectroscopy 1 H and 13 C NMR compared: both give us information about the number of chemically nonequivalent nuclei (nonequivalent hydrogens or nonequivalent carbons) both give us information

More information

Spin Dynamics Basics of Nuclear Magnetic Resonance. Malcolm H. Levitt

Spin Dynamics Basics of Nuclear Magnetic Resonance. Malcolm H. Levitt Spin Dynamics Basics of Nuclear Magnetic Resonance Second edition Malcolm H. Levitt The University of Southampton, UK John Wiley &. Sons, Ltd Preface xxi Preface to the First Edition xxiii Introduction

More information

Correcting Lineshapes in NMR Spectra

Correcting Lineshapes in NMR Spectra Correcting Lineshapes in NMR Spectra Colin Vitols, Pascal Mercier June 2006 In this note we present a method for removing lineshape distortions from nuclear magnetic resonance (NMR) spectra prior to more

More information

10.4 Continuous Wave NMR Instrumentation

10.4 Continuous Wave NMR Instrumentation 10.4 Continuous Wave NMR Instrumentation coherent detection bulk magnetization the rotating frame, and effective magnetic field generating a rotating frame, and precession in the laboratory frame spin-lattice

More information

Nuclear Magnetic Resonance Imaging

Nuclear Magnetic Resonance Imaging Nuclear Magnetic Resonance Imaging Simon Lacoste-Julien Electromagnetic Theory Project 198-562B Department of Physics McGill University April 21 2003 Abstract This paper gives an elementary introduction

More information

Magnetic Resonance Spectroscopy EPR and NMR

Magnetic Resonance Spectroscopy EPR and NMR Magnetic Resonance Spectroscopy EPR and NMR A brief review of the relevant bits of quantum mechanics 1. Electrons have spin, - rotation of the charge about its axis generates a magnetic field at each electron.

More information

16.1 Introduction to NMR Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy 4/11/2013

16.1 Introduction to NMR Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy 4/11/2013 What is spectroscopy? NUCLEAR MAGNETIC RESONANCE (NMR) spectroscopy may be the most powerful method of gaining structural information about organic compounds. NMR involves an interaction between electromagnetic

More information

Double-Resonance Experiments

Double-Resonance Experiments Double-Resonance Eperiments The aim - to simplify complicated spectra by eliminating J-couplings. omonuclear Decoupling A double resonance eperiment is carried out using a second rf source B 2 in addition

More information

Nuclear magnetic resonance spectroscopy

Nuclear magnetic resonance spectroscopy nuclear spin transitions O Nuclear magnetic resonance spectroscopy 1 H, 13 C, 2-dimensional which transitions? wavelength and intensity; ppm what happens if we change the environment of the nucleus? substituent

More information

NMR = Nuclear Magnetic Resonance

NMR = Nuclear Magnetic Resonance NMR = Nuclear Magnetic Resonance NMR spectroscopy is the most powerful technique available to organic chemists for determining molecular structures. Looks at nuclei with odd mass numbers or odd number

More information

With that first concept in mind, it is seen that a spinning nucleus creates a magnetic field, like a bar magnet

With that first concept in mind, it is seen that a spinning nucleus creates a magnetic field, like a bar magnet NMR SPECTROSCOPY This section will discuss the basics of NMR (nuclear magnetic resonance) spectroscopy. Most of the section will discuss mainly 1H or proton spectroscopy but the most popular nuclei in

More information

It is possible to choose the temperature for each experiment by setting a temperature under the Temp pane (under the Standard panel).

It is possible to choose the temperature for each experiment by setting a temperature under the Temp pane (under the Standard panel). 1 2 The study queue gives a lot of flexibility for lining up experiments: they can be run at different temperatures or at different times. You must respect the instrument limits: do not submit experiments

More information

Shimming of a Magnet for Calibration of NMR Probes UW PHYSICS REU 2013

Shimming of a Magnet for Calibration of NMR Probes UW PHYSICS REU 2013 Shimming of a Magnet for Calibration of NMR Probes RACHEL BIELAJEW UW PHYSICS REU 2013 Outline Background The muon anomaly The g-2 Experiment NMR Design Helmholtz coils producing a gradient Results Future

More information

General NMR basics. Solid State NMR workshop 2011: An introduction to Solid State NMR spectroscopy. # nuclei

General NMR basics. Solid State NMR workshop 2011: An introduction to Solid State NMR spectroscopy. # nuclei : An introduction to Solid State NMR spectroscopy Dr. Susanne Causemann (Solid State NMR specialist/ researcher) Interaction between nuclear spins and applied magnetic fields B 0 application of a static

More information

BMB/Bi/Ch 173 Winter 2018

BMB/Bi/Ch 173 Winter 2018 BMB/Bi/Ch 173 Winter 2018 Homework Set 8.1 (100 Points) Assigned 2-27-18, due 3-6-18 by 10:30 a.m. TA: Rachael Kuintzle. Office hours: SFL 220, Friday 3/2 4:00-5:00pm and SFL 229, Monday 3/5 4:00-5:30pm.

More information

Measuring Protein Concentrations by NMR Spectroscopy

Measuring Protein Concentrations by NMR Spectroscopy Published on Web 02/04/2006 Measuring Protein Concentrations by NMR Spectroscopy Gerhard Wider* and Lars Dreier Contribution from the Institut für Molekularbiologie und Biophysik ETH Zürich, CH-8093 Zürich,

More information

INTRODUCTION TO NMR and NMR QIP

INTRODUCTION TO NMR and NMR QIP Books (NMR): Spin dynamics: basics of nuclear magnetic resonance, M. H. Levitt, Wiley, 2001. The principles of nuclear magnetism, A. Abragam, Oxford, 1961. Principles of magnetic resonance, C. P. Slichter,

More information

NMR Spectroscopy of Polymers

NMR Spectroscopy of Polymers UNESCO/IUPAC Course 2005/2006 Jiri Brus NMR Spectroscopy of Polymers Brus J 1. part At the very beginning the phenomenon of nuclear spin resonance was studied predominantly by physicists and the application

More information

Introduction to 1D and 2D NMR Spectroscopy (4) Vector Model and Relaxations

Introduction to 1D and 2D NMR Spectroscopy (4) Vector Model and Relaxations Introduction to 1D and 2D NMR Spectroscopy (4) Vector Model and Relaxations Lecturer: Weiguo Hu 7-1428 weiguoh@polysci.umass.edu October 2009 1 Approximate Description 1: Energy level model Magnetic field

More information

Nuclear Magnetic Resonance Spectroscopy

Nuclear Magnetic Resonance Spectroscopy Nuclear Magnetic Resonance Spectroscopy Structural Elucidation Nuclear magnetic resonance spectroscopy is the name given to the technique which exploits the magnetic properties of nuclei and measures their

More information

Nuclear Magnetic Resonance Spectroscopy Chem 4010/5326: Organic Spectroscopic Analysis Andrew Harned

Nuclear Magnetic Resonance Spectroscopy Chem 4010/5326: Organic Spectroscopic Analysis Andrew Harned Nuclear Magnetic Resonance Spectroscopy Chem 4010/5326: Organic Spectroscopic Analysis 2015 Andrew Harned NMR Spectroscopy NMR Spectroscopy All nuclei have a nuclear spin quantum number (I) I = 0, 1/2,

More information

Topics. The concept of spin Precession of magnetic spin Relaxation Bloch Equation. Bioengineering 280A Principles of Biomedical Imaging

Topics. The concept of spin Precession of magnetic spin Relaxation Bloch Equation. Bioengineering 280A Principles of Biomedical Imaging Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2006 MRI Lecture 1 Topics The concept of spin Precession of magnetic spin Relaxation Bloch Equation 1 Spin Intrinsic angular momentum of

More information

Asian Journal of Chemistry; Vol. 25, No. 4 (2013),

Asian Journal of Chemistry; Vol. 25, No. 4 (2013), Asian Journal of Chemistry; Vol. 25, No. 4 (213), 214-218 http://dx.doi.org/1.14233/ajchem.213.13346 Observation of Triplet Traces Obtained with Inversion Recovery Method in Both Residual Water- and H

More information

Chapter 13 Nuclear Magnetic Resonance Spectroscopy

Chapter 13 Nuclear Magnetic Resonance Spectroscopy Organic Chemistry, 6 th Edition L. G. Wade, Jr. Chapter 13 Nuclear Magnetic Resonance Spectroscopy Jo Blackburn Richland College, Dallas, TX Dallas County Community College District 2006, Prentice Hall

More information

NMR of CeCoIn5. AJ LaPanta 8/15/2016

NMR of CeCoIn5. AJ LaPanta 8/15/2016 NMR of CeCoIn5 AJ LaPanta 8/15/2016 In Co-NMR measurements on CeCoIn5, we see an increasing peak width below 50K. We interpret this as the growth of antiferromagnetic regions surrounding Cadmium dopants

More information

NMR SATELLITES AS A PROBE FOR CHEMICAL

NMR SATELLITES AS A PROBE FOR CHEMICAL NMR SATELLITES AS A PROBE FOR CHEMICAL INVESTIGATIONS SHIzuo FUJIWARA, Yogi ARATA, HIR0sHI OZAWA and MASAYUKI KuruGI Department of Chemistry, The University of Tokyo, Japan ABSTRACT Satellite lines in

More information

e 2m p c I, (22.1) = g N β p I(I +1), (22.2) = erg/gauss. (22.3)

e 2m p c I, (22.1) = g N β p I(I +1), (22.2) = erg/gauss. (22.3) Chemistry 26 Molecular Spectra & Molecular Structure Week # 7 Nuclear Magnetic Resonance Spectroscopy Along with infrared spectroscopy, nuclear magnetic resonance (NMR) is the most important method available

More information

NMR Spectroscopy. for 1 st B.Tech INTRODUCTION Lecture -1 Indian Institute of Technology, Dhanbad

NMR Spectroscopy. for 1 st B.Tech INTRODUCTION Lecture -1 Indian Institute of Technology, Dhanbad NMR Spectroscopy for 1 st B.Tech Lecture -1 Indian Institute of Technology, Dhanbad by Dr. R P John & Dr. C. Halder INTRODUCTION Nucleus of any atom has protons and neutrons Both Proton and Neutron has

More information

Nuclear Magnetic Resonance (NMR)

Nuclear Magnetic Resonance (NMR) Nuclear Magnetic Resonance (NMR) E E increases with increasing magnetic field strength Boltzmann distribution at thermal equilibrium: N (m=-1/2) /N (m=+1/2) = e ( E/kT) with E = γ(h/2π)b o NMR Physical

More information

Biophysical Chemistry: NMR Spectroscopy

Biophysical Chemistry: NMR Spectroscopy Nuclear Magnetism Vrije Universiteit Brussel 21st October 2011 Outline 1 Overview and Context 2 3 Outline 1 Overview and Context 2 3 Context Proteins (and other biological macromolecules) Functional characterisation

More information

NMR Spectroscopy. Guangjin Hou

NMR Spectroscopy. Guangjin Hou NMR Spectroscopy Guangjin Hou 22-04-2009 NMR History 1 H NMR spectra of water H NMR spectra of water (First NMR Spectra on Water, 1946) 1 H NMR spectra ethanol (First bservation of the Chemical Shift,

More information

Concentration measurements by PULCON using X-filtered or 2D NMR spectra

Concentration measurements by PULCON using X-filtered or 2D NMR spectra MAGNETIC RESONANCE IN CHEMISTRY Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/mrc.1838 Concentration measurements by PULCON using X-filtered or 2D NMR spectra Lars Dreier

More information

Schematic for resistivity measurement

Schematic for resistivity measurement Module 9 : Experimental probes of Superconductivity Lecture 1 : Experimental probes of Superconductivity - I Among the various experimental methods used to probe the properties of superconductors, there

More information

Natural abundance solid-state 95 Mo MAS NMR of MoS 2 reveals precise 95 Mo anisotropic parameters from its central and satellite transitions

Natural abundance solid-state 95 Mo MAS NMR of MoS 2 reveals precise 95 Mo anisotropic parameters from its central and satellite transitions Electronic Supplementary Information for: Natural abundance solid-state 95 Mo MAS NMR of MoS 2 reveals precise 95 Mo anisotropic parameters from its central and satellite transitions Hans J. Jakobsen,*

More information

Center for Sustainable Environmental Technologies, Iowa State University

Center for Sustainable Environmental Technologies, Iowa State University NMR Characterization of Biochars By Catherine Brewer Center for Sustainable Environmental Technologies, Iowa State University Introduction Nuclear magnetic resonance spectroscopy (NMR) uses a very strong

More information

Polarized solid deuteron targets EU-SpinMap Dubrovnik

Polarized solid deuteron targets EU-SpinMap Dubrovnik Experimentalphysik I Arbeitsgruppe Physik der Hadronen und Kerne Prof. Dr. W. Meyer G. Reicherz, Chr. Heß, A. Berlin, J. Herick Polarized solid deuteron targets EU-SpinMap 11.10.2010 Dubrovnik Polarized

More information

Biophysical Chemistry: NMR Spectroscopy

Biophysical Chemistry: NMR Spectroscopy Relaxation & Multidimensional Spectrocopy Vrije Universiteit Brussel 9th December 2011 Outline 1 Relaxation 2 Principles 3 Outline 1 Relaxation 2 Principles 3 Establishment of Thermal Equilibrium As previously

More information

Room Temperature Quantum Coherence and Rabi Oscillations in Vanadyl Phthalocyanine: Toward Multifunctional Molecular Spin Qubits

Room Temperature Quantum Coherence and Rabi Oscillations in Vanadyl Phthalocyanine: Toward Multifunctional Molecular Spin Qubits Room Temperature Quantum Coherence and Rabi Oscillations in Vanadyl Phthalocyanine: Toward Multifunctional Molecular Spin Qubits Matteo Atzori, Lorenzo Tesi, Elena Morra, Mario Chiesa, Lorenzo Sorace,

More information

COPYRIGHTED MATERIAL. Basic Principles. 1.1 Introduction

COPYRIGHTED MATERIAL. Basic Principles. 1.1 Introduction . Introduction Basic Principles The field of spectroscopy is in general concerned with the interaction between matter and electromagnetic radiation. Atoms and molecules have a range of discrete energy

More information

Laboratory and Rotating frames

Laboratory and Rotating frames Laborator and Rotating frames The coordinate sstem that we used for the previous eample (laborator frame) is reall pathetic. The whole sstem is spinning at ω o, which makes an kind of analsis impossible.

More information

Nuclear spin maser with a novel masing mechanism and its application to the search for an atomic EDM in 129 Xe

Nuclear spin maser with a novel masing mechanism and its application to the search for an atomic EDM in 129 Xe Nuclear spin maser with a novel masing mechanism and its application to the search for an atomic EDM in 129 Xe A. Yoshimi RIKEN K. Asahi, S. Emori, M. Tsukui, RIKEN, Tokyo Institute of Technology Nuclear

More information

Structure Determination: Nuclear Magnetic Resonance Spectroscopy

Structure Determination: Nuclear Magnetic Resonance Spectroscopy Structure Determination: Nuclear Magnetic Resonance Spectroscopy Why This Chapter? NMR is the most valuable spectroscopic technique used for structure determination More advanced NMR techniques are used

More information

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam (closed book, 1 sheets of notes double sided allowed, no calculator or other electronic devices allowed) 1. Ultrasound Physics (15 pt) A) (9

More information

Magnetic resonance imaging MRI

Magnetic resonance imaging MRI Magnetic resonance imaging MRI Introduction What is MRI MRI is an imaging technique used primarily in medical settings that uses a strong magnetic field and radio waves to produce very clear and detailed

More information

Carbon and Heteronuclear NMR on the Bruker

Carbon and Heteronuclear NMR on the Bruker Carbon and Heteronuclear NMR on the Bruker There are several different types of carbon spectra such as a normal qualitative spectrum, DEPT, coupled, and those with and without NOE. This handout discusses

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu MRI advantages Superior soft-tissue contrast Depends on among

More information

The NMR Spectrum - 13 C. NMR Spectroscopy. Spin-Spin Coupling 13 C NMR. A comparison of two 13 C NMR Spectra. H Coupled (undecoupled) H Decoupled

The NMR Spectrum - 13 C. NMR Spectroscopy. Spin-Spin Coupling 13 C NMR. A comparison of two 13 C NMR Spectra. H Coupled (undecoupled) H Decoupled Spin-Spin oupling 13 NMR A comparison of two 13 NMR Spectra 1 oupled (undecoupled) 1 Decoupled 1 Proton Decoupled 13 NMR 6. To simplify the 13 spectrum, and to increase the intensity of the observed signals,

More information

A Hands on Introduction to NMR Lecture #1 Nuclear Spin and Magnetic Resonance

A Hands on Introduction to NMR Lecture #1 Nuclear Spin and Magnetic Resonance A Hands on Introduction to NMR 22.920 Lecture #1 Nuclear Spin and Magnetic Resonance Introduction - The aim of this short course is to present a physical picture of the basic principles of Nuclear Magnetic

More information

NMR BMB 173 Lecture 16, February

NMR BMB 173 Lecture 16, February NMR The Structural Biology Continuum Today s lecture: NMR Lots of slides adapted from Levitt, Spin Dynamics; Creighton, Proteins; And Andy Rawlinson There are three types of particles in the universe Quarks

More information

NMR and MRI : an introduction

NMR and MRI : an introduction Intensive Programme 2011 Design, Synthesis and Validation of Imaging Probes NMR and MRI : an introduction Walter Dastrù Università di Torino walter.dastru@unito.it \ Introduction Magnetic Resonance Imaging

More information

Nuclear Magnetic Resonance Imaging

Nuclear Magnetic Resonance Imaging Nuclear Magnetic Resonance Imaging Jeffrey A. Fessler EECS Department The University of Michigan NSS-MIC: Fundamentals of Medical Imaging Oct. 20, 2003 NMR-0 Background Basic physics 4 magnetic fields

More information

Nuclear Magnetic Resonance Spectroscopy: Tools for Structure Determination

Nuclear Magnetic Resonance Spectroscopy: Tools for Structure Determination Nuclear Magnetic Resonance Spectroscopy: Tools for Structure Determination Chung-Ming Sun Department of Applied Chemistry National Chiao Tung University Hualien 300, Taiwan Introduction NMR (Nuclear Magnetic

More information

Certificate of analysis Certified reference material code: CRM-00-NEO

Certificate of analysis Certified reference material code: CRM-00-NEO Nº 2/PMR3 Certificate of analysis Certified reference material code: CRM--NEO Name: Neosaxitoxin dihydrochloride in.3 M HCl aq. Batch: 14-1 Description: This Certified Reference Material (CRM) is a solution

More information

NUCLEAR MAGNETIC RESONANCE. The phenomenon of nuclear magnetic resonance will be used to study magnetic moments of nuclei.

NUCLEAR MAGNETIC RESONANCE. The phenomenon of nuclear magnetic resonance will be used to study magnetic moments of nuclei. 14 Sep 11 NMR.1 NUCLEAR MAGNETIC RESONANCE The phenomenon of nuclear magnetic resonance will be used to study magnetic moments of nuclei. Theory: In addition to its well-known properties of mass, charge,

More information

NMR course at the FMP: Practical aspects

NMR course at the FMP: Practical aspects NMR course at the FMP: Practical aspects 04.05.2009 The program 2/68 The dynamic range Solvent supression Quadrature detection Folding/Aliasing Phasing in the indirect dimensions 3/68 The dynamic range

More information

Chapter 13: Nuclear Magnetic Resonance (NMR) Spectroscopy direct observation of the H s and C s of a molecules

Chapter 13: Nuclear Magnetic Resonance (NMR) Spectroscopy direct observation of the H s and C s of a molecules hapter 13: Nuclear Magnetic Resonance (NMR) Spectroscopy direct observation of the s and s of a molecules Nuclei are positively charged and spin on an axis; they create a tiny magnetic field + + Not all

More information

Supporting Information Elucidating Lithium-Ion and Proton Dynamics in Anti- Perovskite Solid Electrolytes

Supporting Information Elucidating Lithium-Ion and Proton Dynamics in Anti- Perovskite Solid Electrolytes Electronic Supplementary Material (ESI) for Energy & Environmental Science. This journal is The Royal Society of Chemistry 2018 Supporting Information Elucidating Lithium-Ion and Proton Dynamics in Anti-

More information

High-Resolutio n NMR Techniques i n Organic Chemistry TIMOTHY D W CLARIDGE

High-Resolutio n NMR Techniques i n Organic Chemistry TIMOTHY D W CLARIDGE High-Resolutio n NMR Techniques i n Organic Chemistry TIMOTHY D W CLARIDGE Foreword Preface Acknowledgements V VI I X Chapter 1. Introduction 1.1. The development of high-resolution NMR 1 1.2. Modern

More information

I690/B680 Structural Bioinformatics Spring Protein Structure Determination by NMR Spectroscopy

I690/B680 Structural Bioinformatics Spring Protein Structure Determination by NMR Spectroscopy I690/B680 Structural Bioinformatics Spring 2006 Protein Structure Determination by NMR Spectroscopy Suggested Reading (1) Van Holde, Johnson, Ho. Principles of Physical Biochemistry, 2 nd Ed., Prentice

More information

Introduction to Relaxation Theory James Keeler

Introduction to Relaxation Theory James Keeler EUROMAR Zürich, 24 Introduction to Relaxation Theory James Keeler University of Cambridge Department of Chemistry What is relaxation? Why might it be interesting? relaxation is the process which drives

More information

COPYRIGHTED MATERIAL. Production of Net Magnetization. Chapter 1

COPYRIGHTED MATERIAL. Production of Net Magnetization. Chapter 1 Chapter 1 Production of Net Magnetization Magnetic resonance (MR) is a measurement technique used to examine atoms and molecules. It is based on the interaction between an applied magnetic field and a

More information

Agilent s new solution for obtaining routinely quantitative results from NMR measurements. Magnetic Resonance Systems

Agilent s new solution for obtaining routinely quantitative results from NMR measurements. Magnetic Resonance Systems Agilent s new solution for obtaining routinely quantitative results from NMR measurements. 1 Magnetic Resonance Systems The Scope of Analytical Chemistry Analytical Chemistry is the study of the separation,

More information

Magnetic Resonance Imaging in a Nutshell

Magnetic Resonance Imaging in a Nutshell Magnetic Resonance Imaging in a Nutshell Oliver Bieri, PhD Department of Radiology, Division of Radiological Physics, University Hospital Basel Department of Biomedical Engineering, University of Basel,

More information

Nuclear Magnetic Resonance Spectroscopy

Nuclear Magnetic Resonance Spectroscopy Chapter 5 Nuclear Magnetic Resonance Spectroscopy http://www.yteach.co.uk/page.php/resources/view_all?id=nuclear_magnetic _resonance_nmr_spectroscopy_spin_spectrometer_spectrum_proton_t_pag e_5&from=search

More information

Supporting information S1. Procedures Materials and methods Preparation of the reagents Preparation of the calibration standard solutions

Supporting information S1. Procedures Materials and methods Preparation of the reagents Preparation of the calibration standard solutions Electronic Supplementary Material (ESI) for Analytical Methods. This journal is The Royal Society of Chemistry 2015 Supporting information S1. Procedures Materials and methods Six oil samples were purchased

More information

Principles of Magnetic Resonance Imaging

Principles of Magnetic Resonance Imaging Principles of Magnetic Resonance Imaging Hi Klaus Scheffler, PhD Radiological Physics University of 1 Biomedical Magnetic Resonance: 1 Introduction Magnetic Resonance Imaging Contents: Hi 1 Introduction

More information

Nuclear Magnetic Resonance (NMR)

Nuclear Magnetic Resonance (NMR) Nuclear Magnetic Resonance (NMR) Nuclear Magnetic Resonance (NMR) The Nuclear Magnetic Resonance Spectroscopy (NMR) is one of the most important spectroscopic methods to explore the structure and dynamic

More information

An Introduction to NMR Spectroscopy. The types of information accessible via high resolution NMR include:

An Introduction to NMR Spectroscopy.  The types of information accessible via high resolution NMR include: 1 of 40 An Introduction to NMR Spectroscopy 1 NMR 13C NMR The types of information accessible via high resolution NMR include: 1. Functional group analysis (chemical shifts) 2. Bonding connectivity and

More information

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging G16.4426 Medical Imaging Physics of Magnetic Resonance Imaging Riccardo Lattanzi, Ph.D. Assistant Professor Department of Radiology, NYU School of Medicine Department of Electrical and Computer Engineering,

More information

Supplemental Material to the Manuscript Radio frequency magnetometry using a single electron spin

Supplemental Material to the Manuscript Radio frequency magnetometry using a single electron spin Supplemental Material to the Manuscript Radio frequency magnetometry using a single electron spin M. Loretz, T. Rosskopf, C. L. Degen Department of Physics, ETH Zurich, Schafmattstrasse 6, 8093 Zurich,

More information