Argon Triple-Point Device for Calibration of SPRTs

Similar documents
I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

Measuring Systems for Thermometer Calibration in Low-Temperature Range

CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to K

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

INTERCOMPARISON OF WATER TRIPLE POINT CELLS FROM INTIBS AND INRIM 1. INTRODUCTION

Investigation of low-temperature fixed points by an international star intercomparison of sealed triple-point cells

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

European Association of National Metrology Institutes

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

2. CURRENT DEFINITION OF THE TEMPERATURE UNIT

THE ITS-90 AFTER DEFINITION OF NEON ISOTOPIC REFERENCE COMPOSITION: EXTENT OF THE ISOTOPIC EFFECT ON PREVIOUS INTER-COMPARISON RESULTS

Calibration of temperature sensors within Length Standard Section of NMIJ

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results

CRYOGENIC CONDUCTION COOLING TEST OF REMOVABLE PANEL MOCK-UP FOR ITER CRYOSTAT THERMAL SHIELD

Realisation and study of a CO 2 triple point thermometric cell. Réalisation et étude d une cellule thermométrique au point triple du CO 2

Report from the Mise en Pratique Task Group: the next international temperature scale and the mise en pratique for the definition of the kelvin

Received: 17 December 2009 / Accepted: 11 March 2010 / Published online: 10 April 2010 Springer Science+Business Media, LLC 2010

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

A calibration facility for automatic weather stations

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Introduction to Blackbody Sources

Available online at ScienceDirect. Physics Procedia 67 (2015 ) Calibration of a HTS based LOX 400mm level sensor

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS

Low temperature thermometry in high magnetic fields. VII. Cernox sensors to 32 T

Measurements of temperature on LHC thermal models

Available online at ScienceDirect. Physics Procedia 67 (2015 ) Superfluid helium heat pipe. P.

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

He II Heat transfer through a Corrugated Tube - Test Report

High-Pressure Volumetric Analyzer

PRECISION MEASUREMENT OF SELF-HEATING EFFECT OF RESISTANCE THERMOMETERS INSTALLED ON MATERIAL ARTIFACTS

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

Development of high-efficiency Stirling cryocoolers for high temperature superconducting motors

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

Automated volume measurement for weihts using acoustic volumeter

HIMARC Simulations Divergent Thinking, Convergent Engineering

Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

Contribution to the study of neon-nitrogen mixtures at low temperatures

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

Put sufficient ice cubes into water (1 M) and wait for equilibrium (both exist) (1 M)

Boltzmann Project. Determination of k and redefinition 2008: imera+ Project coordinated by PTB

1. Thermal energy is transferred through the glass windows of a house mainly by. D. radiation and convection. (1)

Application of the Multi-current Transient Hot-Wire Technique for Absolute Measurements of the Thermal Conductivity of Glycols

(ii) the total kinetic energy of the gas molecules (1 mark) (iii) the total potential energy of the gas molecules (1 mark)

PAPER 2 THEORY QUESTIONS

Part I. Temperature Measurements in the Range from 0.1 K to 300 K

PROGRESS IN THE DETERMINATION OF THE BOLTZMANN CONSTANT WITH DIELECTRIC-CONSTANT GAS THERMOMETRY. C. Gaiser, T.Zandt, B. Fellmuth

Temperature Measurement

SIMPLE LIQUID NITROGEN COMPARATOR MODEL 461 User Maintenance Manual/Handbook

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz

A simple calorimetric method to avoid artifacts in a controversial field: the ice calorimeter

Name Partner. Thermal Physics. Part I: Heat of Vaporization of Nitrogen. Introduction:

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge

EURAMET Project 858: Hydrostatic weighing exchange of experiences. Project outline

Critical Current Properties of HTS Twisted Stacked-Tape Cable in Subcooled- and Pressurized-Liquid Nitrogen

DEVELOPMENT OF A FLOW-THROUGH CELL FOR ACCURATE MEASUREMENTS OF LOW ELECTROLYTIC CONDUCTIVITY

Reproduction of Fleischmann and Pons experiments

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Russell B. Schweickart and Gary Mills

Feasibility of HTS DC Cables on Board a Ship

The Speed of Sound in Air

How do you really know what the temperature is? Michael de Podesta

XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil

The Vacuum Case for KATRIN

Calorimetric Principles and TAM III

International Temperature Scale of 1990 From Wikipedia, the free encyclopedia

Measurements of ultralow temperatures

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

CHAPTER 4 THERMAL CONDUCTIVITY AND VISCOSITY MEASUREMENTS

ECNG3032 Control and Instrumentation I

APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range C to C

Copyright 2008, University of Chicago, Department of Physics HEAT CAPACITY OF METALS

DEVICE FOR MEASUREMENT OF THERMAL EMISSIVITY AT CRYOGENIC TEMPERATURES

2012 Thermodynamics Division C

Approximation of ITS-90 with High Temperature Thermometers up to 1085 C

Flow-meter calibrations for the ID of the ATLAS Experiment

Critical parameters of

Numerical Simulation on Flow and Heat Transfer in Oscillating Heat Pipes

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

19. Thermometry in Magnetic Fields

Apparatus to measure high-temperature thermal conductivity and thermoelectric power of small specimens

Superfluid helium and cryogenic noble gases as stopping media for ion catchers Purushothaman, Sivaji

Department of Mechanical Engineering ME 96. Free and Forced Convection Experiment. Revised: 25 April Introduction

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

Performance analysis of the active magnetic regenerative refrigerator for 20 K

Introduction. Measurement of temperature is generally considered to be one of the simplest and most accurate measurements performed in engineering.

Cambridge International Examinations Cambridge Ordinary Level

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale

Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter

PHY3901 PHY3905. Hall effect and semiconductors Laboratory protocol

Comparison tests of cellular glass insulation for the development of cryogenic insulation standards

40P (2 x 60 x 60) = 2.5 x 10 6 (4200)(5) P = 1.82 x 10 5 W

Transcription:

Int J Thermophys (2015) 36:229 239 DOI 10.1007/s10765-014-1768-0 Argon Triple-Point Device for Calibration of SPRTs B. Kołodziej H. Manuszkiewicz A. Szmyrka-Grzebyk L. Lipiński A. Kowal P. P. M. Steur F. Pavese Received: 9 August 2013 / Accepted: 15 October 2014 / Published online: 30 October 2014 The Author(s) 2014. This article is published with open access at Springerlink.com Abstract This paper presents an apparatus for the calibration of long-stem platinum resistance thermometers at the argon triple point (T 90 = 83.8058 K), designed at the Institute of Low Temperature and Structural Research, Poland (INTiBS). A hermetically sealed cell filled at the Istituto Nazionale di Ricerca Metrologica, Italy with high purity gas (6N) is the main element of this apparatus. The cell is placed in a cryostat fully immersed in liquid nitrogen. A temperature-controlled shield ensures the quasi-adiabatic condition needed for proper realization of the phase transition. A system for correcting the temperature distribution along the thermometer well is also implemented. The cell cooling and argon solidification is carried out by filling the thermometer well with liquid nitrogen. A LabVIEW computer program written at INTiBS automatically controls the triple-point realization process. The duration of a melting plateau in the apparatus lasts for about 24 h. The melting width for F between 20 % and 80 % was <0.3 mk. The reproducibility of the plateau temperature is better than ±20 µk. Keywords Argon triple point Platinum thermometer Temperature scale 1 Introduction The triple point of argon is one of the low-temperature fixed points of the International Temperature Scale of 1990, ITS 90 (T 90 = 83.8058 K) needed for the calibration of B. Kołodziej (B) H. Manuszkiewicz A. Szmyrka-Grzebyk L. Lipiński A. Kowal F. Pavese Instytut Niskich Temperatur i Badań Strukturalnych PAN, Okólna 2, 50-422 Wrocław, Poland e-mail: b.kolodziej@int.pan.wroc.pl P. P. M. Steur F. Pavese Istituto Nazionale di Ricerca Metrologica, Strada delle Cacce 73-91, 10-135 Turin, Italy

230 Int J Thermophys (2015) 36:229 239 standard platinum resistance thermometers, with both long stem (SPRT) and capsule type (CSPRT). Several models of cryostat and systems for the calibration of SPRTs at the triple point of argon were built by different institutions [1 6] since more than thirty years. CSPRTs are calibrated in small sealed cells under good adiabatic conditions [7 10]. Because of the size of SPRTs, it is difficult to build a cryostat for its calibration in a similar way. At the Institute of Low Temperature and Structure Research (INTiBS), Wroclaw, a new cryostat model for calibration of long-stem SPRTs at the argon triple point was built. Based on a long-term experience in work with sealed cells used for calibration of CSPRTs in the low-temperature range, the thermometric cell for the calibration of long-stem SPRTs was designed. The cell was made in cooperation with the Istituto Nazionale di Ricerca Metrologica (INRIM), Turin. The cell is placed in a vacuum chamber immersed in a liquid nitrogen bath allowing realization of the argon triple point and calibration of both types of standard platinum thermometers long stem and capsule type in a quasi-adiabatic condition. The characteristics of the system and the testing results are reported, showing that reproducible plateaux with the realized temperature spanning <300 µk can go on for several dozens of hours. 2 Construction of the System 2.1 Argon Cell A sketch of the argon cell is shown in Fig. 1. The thermometric cell wall is made of stainless steel. The lower part of the cell outer diameter is 22 mm and has a length of 60 mm, properly sized to the dimensions of the platinum sensor of the SPRTs. Inside that part of the cell, a heat exchanger made of a copper block is installed. At the bottom of the cell, a Teflon bracket supports the copper block in a stable position. The upper part of the thermometric cell is enlarged to a diameter of 60 mm in order to increase the cell volume for reducing gas (argon) pressure at room temperature below 6 MPa. The cell contains an accurately measured amount of about 0.35 mol of high-purity argon (6N, Messer GmbH, Germany). The impurities of the gas declared by the producer are as follows: oxygen, nitrogen, water vapor, hydrocarbons, and CO/CO 2, all below 0.5 vppm. The mechanical part of the cell was made at INTIBS. It was filled at INRIM with the argon mentioned before and permanently sealed using an INRIM pinch-weld technique [11]. Through the central part of the cell, a stainless tube (thermometer well diameter of 8 mm) passes up to a flange. In the lower part of the cell, a heater H1 is used for performing the melting with the heat pulse technique. The temperature of the cell is also monitored by a small thermometer PRT1. 2.2 Cryostat The cryostat design is presented in Fig. 2. The argon cell surrounded by superinsulation and a copper thermal shield is located in the vacuum chamber (see Fig. 3).

Int J Thermophys (2015) 36:229 239 231 Fig. 1 Argon cell design The chamber is sealed by an indium gasket and immersed in a liquid nitrogen Dewar (50 L). A turbo-molecular vacuum system is used to maintain vacuum insulation at a pressure inside the chamber below 10 4 Pa. For obtaining stable thermal conditions in the cryostat during the argon triple-point realization, the level of liquid nitrogen above the vacuum chamber is maintained constant (near 10 cm). For this purpose, a device is used consisting of a plastic foam cup, a suspension cord, and a counterweight. The cryostat is immersed at a constant depth in liquid nitrogen, and its weight is balanced with buoyancy and with the weight of a counterweight. Measurements can be performed without refilling nitrogen for up to 12 h. To ensure sufficient thermal contact between the thermal shield and the liquid nitrogen outside the vacuum chamber, bronze springs, as shown in Fig. 4, are attached to the bottom of the shield. The springs ensure an initial cool-down of the shield in a reasonably short time. The temperature of the shield is then controlled with a PID (proportional-integral-derivative) controller. For this purpose, a heater H2 and a thermometer PRT2 are mounted on the shield. The cell cool-down and argon solidification are carried out by filling the thermometer well with liquid nitrogen. The SPRT thermometer is outside the well at this moment and the temperature of the cell is monitored by thermometer PRT1. The heater H1 mounted at the lower part of the cell allows increase in the cell temperature step-wise up to the argon triple point. The upper part of the thermometric well is immersed in liquid nitrogen and has a lower temperature than the cell during the phase transition. This will probably cause a

232 Int J Thermophys (2015) 36:229 239 Fig. 2 Cryostat design heat exchange by convection along the gap between the thermometer and the thermometric well wall, resulting in a reduction of the observed temperature of the triple point. A system for correcting the temperature distribution along the thermometer well was implemented. To reduce the disturbing effect of convection, the upper part of the well is heated by the heater H3 to the temperature T 2 monitored by thermometer PRT3 (see Fig. 2). The influence of the temperature of the upper part of the thermometer well (T 2 ) on the observed phase transition temperature is shown in Fig. 5. The measurement was performed using a glass-sheath SPRT thermometer. A curve a was obtained when an aperture between SPRT and the thermometric well was large (approximately 1 mm in diameter). When the aperture width is decreased by filling it with the thin Teflon tape in the space close to the heater H3 along the thermometer sheath, the influence of temperature T 2 on the temperature of the cell is much lower (curve b ). The thermal resistance R s between the cell and the thermal shield was determined in an experiment consisting in the step change of the shield temperature and simultaneous measurement of changes in cell temperature drift. The cell temperature was below the argon triple-point temperature, and argon was totally solidified. R s was calculated with

Int J Thermophys (2015) 36:229 239 233 Fig. 3 Argon cell: (a) with super-insulation, (b) with thermal shield, and (c) vacuum housing Fig. 4 Bronze springs for obtaining thermal contact between the shield and vacuum chamber T E R s = C (dt/dt), (1) where T E is the temperature difference of the thermal shield before and after a step change, C is the cell thermal capacity, and (dt/dt) means the change of the cell temperature drift after thermal shield temperature modification. R s was determined to be equal to 1170 K W 1 ensuring sufficient adiabatic conditions during the realization of the phase transition and thermometer calibration. The entrance to the thermometer well is located in the gaseous nitrogen evaporated from the Dewar (Fig. 2). This prevents air from condensing in the well. The entrance to the well should always be placed below the top-end level of the Dewar.

234 Int J Thermophys (2015) 36:229 239 Fig. 5 Influence of the temperature of the upper part of the thermometric well (T 2 ) on the temperature observed at the triple point of argon Fig. 6 Measuring system design: M1, M2 thermometer resistance bridges, Z current supply, Vm voltmeter, Rt PID temperature controller, T long stem thermometer, Mp vacuum meter 2.3 Measuring System The measuring system is schematically shown in Fig. 6. The SPRT resistance is measured with a thermometric resistance bridge, M1, ensuring a measurement accuracy of 0.1 ppm. For monitoring the cell temperature, thermometer PRT1 is used. Its resistance is measured with another resistance bridge, M2 (accuracy of 1 ppm). The heater, H1, mounted on the cell is supplied by a current source (Advantech C/A 1723). A multimeter Vm measures the voltage on the heater (four-wire technique is used). The thermal shield is controlled with a PID controller with a nominal resolution of 1 mk.

Int J Thermophys (2015) 36:229 239 235 Fig. 7 Argon melting plateaux measured with the continuous heat method. P cell heater power The thermal gradient on the shield was measured with two calibrated PRTs attached to the bottom and to the top part of the shield. The maximum temperature difference between the temperature on different parts of the screen was 50 mk. The process of the argon triple-point realization is controlled by a LabVIEW computer program. The program lets the operator observe and control the thermal conditions in the cryostat and measure the temperatures of the cell (PRT1), the thermal shield (PRT2), and the thermometer well (PRT3). The SPRT resistance is measured automatically as well. The reading noise level of the SPRT resistance measurements was equivalent to 32 µk when the thermometer current was 1 ma. The observed self-heating of the SPRT was 1 mk: all SPRT resistance values were corrected to zero current. 3 Results The triple point of argon in this system was realized using both the continuous and pulse heating methods. In Fig. 7, the shape of the melting plateaux achieved during continuous heating of the cell is presented for two power values 49 mw and 25 mw. The determined heat of transition Q was 380 J. The discrepancy between the measured heat of transition and the expected one based on the quantity of enclosed substance was about 35 J. Also, the thermal resistance R CS for F 50 % (where F is the percentage of the molten argon) between the solid argon and a wall of the cell was estimated using the heat pulse method [12]. The temperature inside the cell before and after the heat pulse was measured. The thermal resistance value R CS was calculated with R CS = T CS P PAS, (2) where T CS is the cell temperature difference before and after the heat pulse and P PAS is the power of the heat pulse applied to the cell with cell heater H1. R CS is about 0.02 K W 1. Three argon melting curves determined using the heat pulse method are presented in Fig. 8a, b. These melting curves were obtained still without temperature control of the thermometer well. The transition width for F between 20 % and 80 % is

236 Int J Thermophys (2015) 36:229 239 Fig. 8 (a) Argon melting curves measured with the heat pulse method as a function of F and (b) argon melting curves measured with the heat pulse method as a function of 1/F <0.3 mk, and the reproducibility of the argon triple-point temperature for F 50 % is better than 40 µk. Figure 8b shows the melting curves as a function R = f (1/F). For the purpose of SPRT calibrations, the value at F = 50 % is taken as T tp,ar (argon triple-point temperature): the difference with respect to the true liquidus point T L (a temperature at 1/F = 1, extrapolated from the range 2 1/F 8) is less than 50 µk. The expanded uncertainty of the SPRT calibration is better then 0.3 mk for k = 2. The uncertainty budget of the SPRT calibration is presented in Table 1. The most significant uncertainty budget components described below are determination of the plateau value, propagation of the TPW uncertainty, and perturbing heat exchange effect. These three components determine over 90 % of the total standard uncertainty of the SPRT calibration in the argon cell. 3.1 Determination of the Plateau Value The value is estimated for F = 50 %. The uncertainty is evaluated from the melting range in the interval F = 20 % to 80 % assuming a rectangular distribution. The value of this uncertainty component was estimated from the results shown in Fig. 8a

Int J Thermophys (2015) 36:229 239 237 Table 1 Uncertainty budget for the calibration of long-stem SPRTs at the triple point of argon Uncertainty component Standard uncertainty (mk) Chemical impurity From the producers certificate using the first cryoscopic constant [13] 0.025 Determination of the plateau value The value is estimated for F = 50 %. The uncertainty is evaluated from the 0.087 melting range in the interval F = 20 % to 80 % assuming a rectangular distribution. Value of this uncertainty component was estimated from the results shown in Fig. 8a( T = 270 µk) assuming a symmetric, rectangular probability distribution Plateau repeatability The value is estimated for F = 50 %. The uncertainty evaluation is from the difference between the maximum and minimum value assuming a normal distribution Perturbing heat exchange The uncertainty evaluation is from the difference between value of temperature at the bottom of the thermometer well and 5 cm above the bottom, for F = 50 % assuming a symmetric, rectangular probability distribution Hydrostatic head correction The value of this component is an estimation of the uncertainty of the distance between the argon level in the cell and the middle point of the sensor 0.023 0.043 0.010 Propagation from TPW uncertainty 0.061 Self-heating 0.022 Bridge measurement 0.009 Bridge uncertainty for SPRT 25 Standard resistor 0.005 Resistor uncertainty and temperature coefficient Standard uncertainty 0.12 Expanded uncertainty 0.24 ( T = 270 µk) assuming a symmetric, rectangular probability distribution using u 1 (T tpar ) = T 2 3 (3) 3.2 Propagation of the TPW Uncertainty The uncertainty of the long-stem SPRT calibration in the TPW u(t TPW ) is <280 µk. The ratio R(T tpar )/R(TPW) is 0.22. The propagation of the TPW uncertainty component value u 2 (T tpar ) is determined from u 2 (T tpar ) = u(t TPW ) R(T tpar) R(TPW) (4)

238 Int J Thermophys (2015) 36:229 239 3.3 Perturbing Heat Exchange The heat exchange between the cell and the external environment during the realization of the triple point can affect the measurement. The influence of this effect can be estimated by measuring the immersion profile in the thermometer well in the cell. The difference T +5 between the value of the temperature at the bottom of the thermometer well and 5 cm above the bottom for F = 50 % is <150 µk. The uncertainty component value, assuming a symmetric rectangular probability distribution is determined using u 3 (T tpar ) = T +5 2 3 (5) Small as for the long-stem thermometer, the uncertainty value of the perturbing heat exchange is achieved by the use of heater H3 (Fig. 2). The heater cooperating with the PRT3 thermometer and temperature controller Rt is a compensating heat flux along the thermometer well. 4 Conclusion A new cryostat for the calibration of long-stem SPRTs at the argon triple point in quasi-adiabatic conditions was tested using continuous and pulse heating methods. The results obtained show that the reproducibility of the phase transition is about 40 µk. The expanded uncertainty of the SPRT calibration is better then 0.3 mk. The discrepancy between the measured heat of transition and the expected one based on the quantity of enclosed substance is still under investigation. Acknowledgment The presented work was done under the project of the National Center of Research and Development (Poland) No. PBS1/A9/6/2012 (ID:179275). Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. References 1. J. Ancsin, J.M. Phillips, Rev. Sci. Instrum. 55, 1321 (1984) 2. G. Furukawa, in Temperature: Its Measurement and Control, in Science and Industry, vol. 6, ed. by J.F. Schooley (AIP, New York, 1992), pp. 265 269 3. I. Yang, C.H. Song, K.H. Kang, Y.G. Kim, K.S. Gam, Int. J. Thermophys. 29, 1740 (2008) 4. R. Ding, M. Zhao, E. Nerdrum, D. Meier, Int. J. Thermophys. 32, 2252 (2011) 5. M. Dobre, I. Didialaoui, Y. Hermier, Int. J. Thermophys. 32, 1573 (2011) 6. I. Didi-Alaoui, D. Fiorillo, A. Vergé, F. Sparasci, D. Jouin, Y. Hermier, in Proceedings of Ninth International Temperature Symposium (Los Angeles), Temperature: Its Measurement and Control, in Science and Industry, vol. 8, ed. by C.W. Meyer, A.I.P. Conference Proceedings 1552 (AIP, Melville, NY, 2013), pp. 474 479 7. F. Pavese, J. Ancsin, D.N. Astrov, J. Bonhoure, G. Bonnier, G.T. Furukawa, R.C. Kemp, H. Maas, R.L. Rusby, H. Sakurai, L. Shan-Kang, Metrologia 20, 127 (1984) 8. A.G. Steele, B. Fellmuth, D.I. Head, Y. Hermier, K.H. Kang, P.P.M. Steur, W.L. Tew, Metrologia 39, 551 (2002)

Int J Thermophys (2015) 36:229 239 239 9. F. Pavese, D. Ferri, I. Peroni, A. Pugliese, P.P.M. Steur, B. Fellmuth, D. Head, L. Lipiński, A. Peruzzi, A. Szmyrka-Grzebyk, L. Wolber, in Temperature: Its Measurement and Control, in Science and Industry, vol. 7, ed. by D.C. Ripple (AIP, New York, 2003), pp. 173 178 10. B. Fellmuth, L. Wolber, D.I. Head, Y. Hermier, K.D. Hill, T. Nakano, F. Pavese, A. Peruzzi, R.L. Rusby, V. Shkraba, A.G. Steele, P.P.M. Steur, A. Szmyrka-Grzebyk, W.L. Tew, L. Wang, D.R. White, Metrologia 49, 257 (2012) 11. F. Pavese, G. Molinar, Modern Gas-Based Temperature and Pressure Measurements, 2nd edn. (Springer, New York, 2013), p. 110 12. F. Pavese, G. Molinar, Modern Gas-Based Temperature and Pressure Measurements, 2nd edn. (Springer, New York, 2013), pp. 80 81 13. F. Pavese, Metrologia 48, 268 (2011)