The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results

Size: px
Start display at page:

Download "The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results"

Transcription

1 Online Supplementary Information for the paper: The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results Franco Pavese 1, Anna Szmyrka-Grzebyk 1, Peter P.M. Steur 2 A. Chemical impurities effect The chemical impurities, reported in the column of chemical corrections of Table 3 of the paper, show that: 1 the corrections for these impurities may be even more critical that the ones for the isotopic composition. The experience of the International Project on isotopic neon 13 2 has shown that the availability from KRISS of excellent assays also on the chemical impurities was a basic asset to obtain an outstanding top-accuracy overall correction of the results on the measured samples in that Project; The lack of data on some critical impurities, such as hydrogen in neon, can impair the validity of the correction for chemical impurities and the usefulness of the isotopic correction. [15 It is the authors opinion that the CCT should consider the need, for top accuracy, of the application of corrections, by using the SIE or similar methods S1 3. Supplementary references S1. Fellmuth B., Hill K.D., Pearce J.V., Peruzzi A., Steur P.P.M., Zhang J., Guide to the Realization of the ITS-90. Fixed Points: Influence of the Impurities. BIPM, CCT, update of 21 October 2015 (2015) S2. Pavese F., Metrologia 46, 47 (2009) S3. Pavese F., Metrologia 48, 268 (2011) B. Specific conditions for each laboratories concerning the K2.(x) comparisons BNM-INM (now LNE-LCM) The sample BCMH2O of neon used in K2 is not traceable to isotopic assays. However, from the STAR inter-comparison [12] it can be inferred that the bottle was the same as for INM sample 99/2, for which an isotopic assay is available [15]. 1 Effect on Ttp,Ne: for H2 ( 7 3) K/10 6 ; for N2 ( 8 3) K/ The reference number refers to the main paper References.

2 Later another measurement was available in the frame of the comparison K2.4 in , using cell Ne02/1 (see Table 2 in the paper also for the next references to samples), for which an isotopic assay exist: this result is also reported. Finally, the BNM-INM datum in K2 for set #2 was a clear outlier, though within its very large uncertainty. CNR-IMGC (now INRIM) The sample of neon used in K2 was sealed in cell 3Ne, later outsourced by the IMGC to NIM, and the IMGC reference was moved to cell 1Ne, using a different bottle of gas (see Table 3 in the paper). However, as explained in 4 and its Online Supplementary Information, the thermal results obtained with cells 1Ne and 3Ne are not consistent with the different measured compositions: the difference in the isotopic correction according to the assays amounts to (1Ne 3Ne) = 195(30) K. The difference (T 1Ne T 3Ne ) was 44(110) K in INRIM measurements 4 and 29(76) K in 12 the latter will be used in the following. On the other hand, the isotopic assays have some reasons for being less accurate than claimed because the sample available for the assays was quite small: the bottles only contained 2 mmol and 3 mmol, respectively, of gas, decades old. In addition, in the set of observed T L vs 22 x values, the T L (3Ne) looks as an outlier being too low by K, consistent with the above discrepancy. In Table 3 of the paper the values for sample 1Ne are also provided. INRIM performed a comparison of some key cells in [4]. This is referred in Table 3 as INRIM comparison, having INRIM cell Ec2Ne as the reference, now INRIM reference. KRISS This laboratory was not considered with respect to the CCT-K2 comparison until recently S4, when the isotopic composition of the sample of neon used for the (open cell) realisation during CCT -K2 became available. In addition, in S4 the realisation of the triple point of neon is also reported using samples from three different gas bottles whose isotopic composition is known from a calibrated mass spectrometer. NBS (now NIST) The sample of neon used for the (open cell) realisation during CCT-K2 was taken from a known bottle, whose isotopic assay was made available in For NIST, it is known that the same gas was used also for the data of CCT-K1 by direct realisations of the ITS-90 using an interpolating constantvolume gas thermometer. NPL The sample of neon used for the sealed cell realization during CCT-K2 for thermometer was sealed in the NPL cell Ne2 and Ne1: NPL assumed the two cells having been filled from the same bottle, for which isotopic assays became available in The data for thermometer came instead from calibration by comparison with the NPL reference thermometer.

3 NRC The sample of neon used for the sealed cell realization during CCT-K2 was sealed in the NRC cell F15, taken from a bottle whose isotopic assays became available only in The same bottle is said by NRC to have been used for sealing cell F17 [S5]. See hereinafter and the main text PTB The sample of neon used for the sealed cell realization during CCT-K2 was sealed in the PTB cell Ne-7, taken from a bottle for which isotopic assays became available in The PTB cell Ne-12, sealed later from the same bottle, is known to be basically identical to cell Ne-7, T = 8(66) K. Only thermometer was used for the K2 equivalence table. PTB was the co-ordinator of a subsequent comparison [12], reported in Table 3 as Star inter-comparison. INRIM cell Ec2Ne is used as the reference cell also for this exercise. PRMI (later VNIIFTRI) The sample of neon used for participation to the CCT-K2 is unknown. In all instances, the PRMI withdrew from the K2. Later, a bilateral comparison, K2.1, was performed with NRC. The pilot still used a thermometer calibrated on cell F15, so that no calculation was necessary to refer to the CCT-K2 KCRV. The VNIIFTRI thermometers were calibrated against an average realisation of the ITS-90, so that no data on the sample of neon used are available. No correction for isotopic composition can be performed. In , the PRMI participated in the first International Intercomparison of fixed points in sealed cells 34 with sealed cell MC-897, whose relationship with other still-existing sealed cells exist, and quite recently it was re-measured in the frame of the Star inter-comparison 12, where it was found (MC-897 3Ne) = 90(92) K. The isotopic composition of the gas used to fill this cell was provided by the VNIIFTRI with low resolution, corresponding to an uncertainty component of 80 K. NMI (then NMI-VSL, now VSL) NMI withdrew from its initial participation to the CCT-K2. Later a bilateral comparison, K2.3, was performed with NRC. VSL used a sealed neon cell produced by IMGC, 12Ne. NRC was unable to perform a calibration on the cell F15 used at the time of CCT-K2, and had to use a chain of calibrations to relate the last cell used, Cu-M-1, to cell F15, through cell F17 (53 K colder than F15). The resulting difference resulted to be (Cu-M-1 F15) = 165(200) K. Instead, the same difference obtained from the isotopic composition corrections is 342(95) K (see more below under INTiBS for NRC). On the Report of comparison K2.3, VSL performed the isotopic corrections for e-h2 and H 2 O using the official equations available in On the contrary for Neon, VSL used a home-made evaluation, 157(7) K; for the latter now the official correction for the neon sample sealed in cell 12Ne is available and is: 123(20) K. The difference with the VSL estimate is irrelevant considering the uncertainty of the overall temperature deviations.

4 INTiBS This Institute participated in a trilateral comparison, K2.4, with BNM and NRC (as the pilot). INTiBS used a sealed neon cell produced by the IMGC (E3Ne), with known isotopic composition. As in the case of K2.3, NRC was unable to perform a calibration based on the cell F15 used at the time of CCT-K2, and had to use a chain of calibrations to relate the last cell used, Cu-M-1, to cell F15, through cell F17. Assuming that cells F15 and F17 were effectively filled from the same neon bottle, the resulting difference resulted to be (Cu-M-1 F15) = 165(200) K. NMIJ-AIST This Institute participated in a trilateral comparison, K2.5, with INRIM and NRC, where the supplied temperature values are corrected for isotopic composition. Two thermometers were calibrated for this exercise, and the values are supplied for the neon isotopic composition corrected to the reference one [36], one sample being almost coincident with the latter. Only cell Ne-5 is the NMIJ-AIST reference, cell Ne-2 (see INRIM inter-comparison) being known to be hotter by 31(50) K. Supplementary references S4. Yang I., Gam K.S., Joung W., Kim Y.G., Int. J. Thermophysics 36, 2072 (2015) S5. K. Hill, Private Communication. The statement could not be verified. C. Results of other direct-cell comparisons In a direct cell comparison under European Project MULTICELLS [38], differences were found as follows, with respect to cell 7Ne, supplied by INRIM to PTB sealed in 1986 and of the batch 6Ne - 11Ne, the same of cell #20 in this paper): INRIM Eb1Ne 0.024(18) mk; 3 INRIM Ec2Ne 0.065(17); LNE-INM Ne99/2 = 0.31(20) mk. Under the same Project, at VSL the difference (LNE-INM Ne02-1 INRIM 3Ne) was found to be 0.07(45) mk. From previous inter-comparisons, the bilateral DoE of the Comparison Int84, as reported in [34], are the following with respect to IMGC-CNR reference cell where the difference between cells 1Ne and 3Ne was then set to 0.00 mk: ASMW (later PTB) 0.34 mk, INM 0.03 mk, NRC 0.04 mk, NRLM = 0.13 mk and PRMI (VNIIFTRI) 0.06 mk; U was estimated to be 0.3 mk. INM and VNIIFTRI cells are traceable to present data. 3 Here the uncertainty in parenthesis is the expanded one, U, which is the overall measurement uncertainty budget, so including the isotopic effect.

5 Good results were also found in the direct inter-comparison of sealed cells performed at INRIM [4], more restricted than the STAR inter-comparison of Fig. 4: (PTB Ne-7 INRIM Ec2Ne) = 0.057(52) mk; 3 (NMIJ Ne-2 INRIM Ec2Ne) = 0.006(57) whence (NMIJ Ne-5 INRIM Ec2Ne) = 0.043(57); (NPL Ne2 INRIM Ec2Ne) = 0.154(53) mk; (INTiBS E3Ne INRIM Ec2Ne) = 0.029(54) mk (from Fig. 2); (VSL 12Ne INRIM Ec2Ne) = 0.049(54) mk (from Fig. 2). Further, (15Ne Ec2Ne) = 0.054(58) mk; (1Ne Ec2Ne) = 0.131(66). Furthermore, using cell INRIM Ec2Ne as reference, via PTB Ne-12 by knowing that it is mk hotter than PTB cell Ne-7 as measured in [12] where the latter is the reference, the differences to PTB Ne-7 ar found to be: (3Ne Ne-7) = 0.066(45) mk and (12Ne Ne-7) = 0.062(48) mk, respectively. 3 In addition, (E1Ne Ne7) = 0.023(45) mk. Thus, being PTB Ne-12 hotter than INRIM Ec2Ne by 0.054(33) mk so (Ne-7 Ec2Ne) = mk one finds (E3Ne Ec2Ne) = mk and (12Ne Ec2Ne) = mk, respectively, and (E1Ne Ec2Ne) = mk.

THE ITS-90 AFTER DEFINITION OF NEON ISOTOPIC REFERENCE COMPOSITION: EXTENT OF THE ISOTOPIC EFFECT ON PREVIOUS INTER-COMPARISON RESULTS

THE ITS-90 AFTER DEFINITION OF NEON ISOTOPIC REFERENCE COMPOSITION: EXTENT OF THE ISOTOPIC EFFECT ON PREVIOUS INTER-COMPARISON RESULTS THE ITS-90 AFTER DEFINITION OF NEON ISOTOPIC REFERENCE COMPOSITION: EXTENT OF THE ISOTOPIC EFFECT ON PREVIOUS INTER-COMPARISON RESULTS Franco Pavese 1*, Anna Szmyrka-Grzebyk 1, Peter P.M. Steur 2 1 INTiBS,

More information

Investigation of low-temperature fixed points by an international star intercomparison of sealed triple-point cells

Investigation of low-temperature fixed points by an international star intercomparison of sealed triple-point cells IOP PUBLISHING Metrologia 49 (2012) 257 265 Investigation of low-temperature fixed points by an international star intercomparison of sealed triple-point cells METROLOGIA doi:10.1088/0026-1394/49/3/257

More information

CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to K

CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to K CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to 273.16 K K. D. Hill 1, T. Nakano 2, P. Steur 3 1 Institute for National Measurement Standards,

More information

INTERCOMPARISON OF WATER TRIPLE POINT CELLS FROM INTIBS AND INRIM 1. INTRODUCTION

INTERCOMPARISON OF WATER TRIPLE POINT CELLS FROM INTIBS AND INRIM 1. INTRODUCTION METROLOGY AND MEASUREMENT SYSTEMS VOL. XV, NUMBER 4 (2008) A. SZMYRKA-GRZEBYK 1, A. KOWAL 1, L. LIPIŃSKI 1, PETER P. M. STEUR 2, R. DEMATTEIS 2 1 Institute of Low Temperature and Structure Research PAN,

More information

European Association of National Metrology Institutes

European Association of National Metrology Institutes European Association of National Metrology Institutes EURAMET GUIDELINES ON TEMPERATURE: Extrapolation of SPRT calibrations below the triple point of argon, 83.8058 K, and traceability in baths of liquid

More information

2. CURRENT DEFINITION OF THE TEMPERATURE UNIT

2. CURRENT DEFINITION OF THE TEMPERATURE UNIT METROLOGY AND MEASUREMENT SYSTEMS VOL. XV, NUMBER 2 (2008) LESZEK LIPIŃSKI, ANNA SZMYRKA-GRZEBYK Institute of Low Temperature and Structure Research Polish Academy of Sciences, Wrocław e-mail: A. Szmyrka@int.pan.wroc.pl

More information

Measuring Systems for Thermometer Calibration in Low-Temperature Range

Measuring Systems for Thermometer Calibration in Low-Temperature Range Int J Thermophys (2011) 32:2466 2476 DOI 10.1007/s10765-011-1055-2 Measuring Systems for Thermometer Calibration in Low-Temperature Range A. Szmyrka-Grzebyk L. Lipiński H. Manuszkiewicz A. Kowal A. Grykałowska

More information

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin Franco Pavese INRIM, Thermodynamic Division, Torino, Italy f.pavese@inrim.it F.Pavese 1 Temperature scales

More information

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN Michael Stock and Stéphane Solve Bureau International des Poids et Mesures (BIPM) Pavillon de Breteuil,

More information

Report from the Mise en Pratique Task Group: the next international temperature scale and the mise en pratique for the definition of the kelvin

Report from the Mise en Pratique Task Group: the next international temperature scale and the mise en pratique for the definition of the kelvin Report from the Mise en Pratique Task Group: the next international temperature scale and the mise en pratique for the definition of the kelvin Dean Ripple 1, Bernd Fellmuth 2, Joachim Fischer 3, Graham

More information

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam Cryostat for Fixed-Point Calibration of Capsule-Type SPRTs I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam International Journal of Thermophysics Journal of Thermophysical Properties and Thermophysics and Its

More information

CCT Working Group for Humidity

CCT Working Group for Humidity CCT/17-41 CCT Working Group for Humidity Report to the CCT 2017 Stephanie Bell CCT Working Group for Humidity Terms of reference: to advise the CCT on matters relating to humidity; to pursue harmonization

More information

Comparison between the mass spectrometer and speed-of-sound measurements of M Ar for the k B determination

Comparison between the mass spectrometer and speed-of-sound measurements of M Ar for the k B determination Comparison between the mass spectrometer and speed-of-sound measurements of M Ar for the k B determination Inseok Yang Korea Research Institute of Standards and Science (KRISS) 04 Feb. 2015 Contribution

More information

LINKING SIM MASS COMPARISONS TO THE KCRV ON 1 kg. Luis O. Becerra CENAM Querétaro, Qro., Mexico,

LINKING SIM MASS COMPARISONS TO THE KCRV ON 1 kg. Luis O. Becerra CENAM Querétaro, Qro., Mexico, LINKING SIM MASS COMPARISONS TO THE KCRV ON 1 kg Luis O. Becerra CENAM Querétaro, Qro., Mexico, lbecerra@cenam.mx Abstract. Within the frame of SIM (Sistema Interamericano de Metrología), inter laboratory

More information

WORKING GROUP FOR CONTACT THERMOMETRY REPORT TO CCT June 2017

WORKING GROUP FOR CONTACT THERMOMETRY REPORT TO CCT June 2017 CCT/17-57 WORKING GROUP FOR CONTACT THERMOMETRY REPORT TO CCT June 2017 Members Joachim Fischer (PTB) chairman, Michael de Podesta (NPL), Vladimir Kytin (VNIIFTRI), Tohru Nakano (NMIJ/AIST), Laurent Pitre

More information

Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn. Aimo Pusa MIKES Finland

Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn. Aimo Pusa MIKES Finland Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn Aimo Pusa 09.02.2009 MIKES Finland 2 Content Page Content 3 Foreword 4 Chapter 1 1.1 General 6 1.2 Characteristics of the transducers 8 1.3

More information

Overview. TC Thermometry Graham Machin. Introduction to TC-T Annual meeting Selected comparisons TC-T training for future thermal metrologists

Overview. TC Thermometry Graham Machin. Introduction to TC-T Annual meeting Selected comparisons TC-T training for future thermal metrologists TC Thermometry Graham Machin Euramet GA 2 June 2015 Overview Introduction to TC-T Annual meeting Selected comparisons TC-T training for future thermal metrologists Workshops COOMET TC-T links EMPIR project

More information

CCT-K5: Comparison of local realizations of the ITS-90 between the silver point and 1700 C using vacuum tungsten strip lamps as transfer standards

CCT-K5: Comparison of local realizations of the ITS-90 between the silver point and 1700 C using vacuum tungsten strip lamps as transfer standards CCT-K5: Comparison of local realizations of the ITS-90 between the silver point and 1700 C using vacuum tungsten strip lamps as transfer standards Prepared by M.J. de Groot, E.W.M. van der Ham and R. Monshouwer

More information

APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range C to C

APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range C to C APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range -38.8344 C to 419.57 C Final report Prepared by. Joung (coordinator) and K. S. Gam Korea Research Institute of Standards and Science

More information

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon Tohru Nakano and Osamu Tamura National Metrology Institute of Japan (NMIJ), AIST Tsukuba, Japan

More information

How do you really know what the temperature is? Michael de Podesta

How do you really know what the temperature is? Michael de Podesta How do you really know what the temperature is? Michael de Podesta TECO: Madrid September 2016 GOLDEN RULE OF TALKS One talk: one thing This Talk: TWO THINGS! Thing 1: The definitions of the SI units of

More information

WORKING GROUP 4 REPORT TO CCT 7 June 2005

WORKING GROUP 4 REPORT TO CCT 7 June 2005 WORKING GROUP 4 REPORT TO CCT 7 June 2005 Members: R L Rusby (Chairman), M R Moldover, J Fischer, D R White, P P M Steur, assisted by R P Hudson and M Durieux (honorary), K D Hill (co-opted in place of

More information

Argon Triple-Point Device for Calibration of SPRTs

Argon Triple-Point Device for Calibration of SPRTs Int J Thermophys (2015) 36:229 239 DOI 10.1007/s10765-014-1768-0 Argon Triple-Point Device for Calibration of SPRTs B. Kołodziej H. Manuszkiewicz A. Szmyrka-Grzebyk L. Lipiński A. Kowal P. P. M. Steur

More information

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Jin Wan Chung 1, Shih Mean Lee, Sungjun Lee 1 1 Korea Research Institute of Standards

More information

APMP-K3: KEY COMPARISON OF REALIZATIONS OF THE ITS-90 OVER THE RANGE o C TO o C Final Report Prepared by M. K. Nguyen and M. J.

APMP-K3: KEY COMPARISON OF REALIZATIONS OF THE ITS-90 OVER THE RANGE o C TO o C Final Report Prepared by M. K. Nguyen and M. J. APMP-K3: KEY COMPARISON OF REALIZATIONS OF THE ITS-90 OVER THE RANGE 38.8344 o C TO 419.527 o C Final Report Prepared by M. K. Nguyen and M. J. Ballico National Measurement Institute of Australia Lindfield,

More information

The BIPM key comparison database, Aug /10

The BIPM key comparison database, Aug /10 BIPM.EM-K11.b, EUROMET.EM.BIPM-K11, APMP.EM.BIPM-K11.1, EUROMET.EM.BIPM-K11.5, EUROMET.EM.BIPM-K11.6, SIM.EM.BIPM-K11.b, COOMET.EM.BIPM-K11, APMP.EM.BIPM-K11.4 and APMP.EM.BIPM-K11.3 Key comparison BIPM.EM-K11.b

More information

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. A comparison of primary platinum-iridium kilogram mass

More information

CCM short note on the dissemination process after the proposed redefinition of the kilogram

CCM short note on the dissemination process after the proposed redefinition of the kilogram CCM short note on the dissemination process after the proposed redefinition of the kilogram Consultative Committee for Mass and Related Quantities 1. Introduction This note proposes how the mise en pratique

More information

Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68

Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68 Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68 8th APMP/TCQM Gas CRM Workshop June 10 th, 2010 Tukuba, Ibaraki, Japan Nobuyuki Aoki, Takuya Shimosaka, Nobuhiro Matsumoto, Kenji

More information

Preparation of nitrous oxide (N 2 O) in air standard being traceable to SI by gravimetric method

Preparation of nitrous oxide (N 2 O) in air standard being traceable to SI by gravimetric method Preparation of nitrous oxide ( O) in air standard being traceable to SI by gravimetric method 15th WMO Meeting of Experts on Carbon Dioxide, Other Greenhouse Gases, and Related Tracer Measurement Techniques

More information

Final Report on COOMET Key Comparison of Capacitance at 10 pf (COOMET.EM-K4)

Final Report on COOMET Key Comparison of Capacitance at 10 pf (COOMET.EM-K4) State Enterprise All-Ukrainian State Scientific and Production Center of Standardization, Metrology, Certification and Protection of Consumer (SE Ukrmetrteststandard ) Approved by the chairman of TC 1.3

More information

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom Force Key Comparison EUROMET.M.F-K2 (50 kn and 100 kn) EURAMET Project No 518 Final Report 14 July 2014 Pilot: NPL, United Kingdom Co-authors: Renato Reis Machado (INMETRO, Brazil), Petr Kašpar (CMI, Czech

More information

On the redefinition of the kilogram. Dr Philippe RICHARD

On the redefinition of the kilogram. Dr Philippe RICHARD On the redefinition of the kilogram Dr Philippe RICHARD Agenda 1. Recent history of IPK and new SI 2. The CCM conditions 3. The joint CCM and CCU roadmap 4. The mise en pratique of the definition of the

More information

DRAFT B, Final Report on CIPM key comparison of 1 kg standards in stainless steel (CCM.M-K1)

DRAFT B, Final Report on CIPM key comparison of 1 kg standards in stainless steel (CCM.M-K1) 1 DRAFT B, Final Report on CIPM key comparison of 1 kg standards in stainless steel (CCM.M-K1) C. Aupetit 1, L.O. Becerra, N. Bignell 3, W. Bich 4, G.D. Chapman 5, J.W. Chung 6, J. Coarasa 7, S. Davidson

More information

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2)

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2) 1 Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K) 1. Introduction L.O. Becerra 1, W. Bich, N. Bignell 3, G.D. Chapman 4, J.W. Chung 5 S. Davidson 6, M. Gläser

More information

EUROMET Project 702 EUROMET.M.D-K4

EUROMET Project 702 EUROMET.M.D-K4 EUROMET TC-M Meeting 2007 Thursday, 01 March 2007 EUROMET Project 702 EUROMET.M.D-K4 Comparison of the calibrations of high resolution hydrometers for liquid density determinations Salvatore Lorefice INRIM,

More information

On the possible future revision of the SI

On the possible future revision of the SI Bureau International des Poids et Mesures On the possible future revision of the SI Martin Milton (Director) and Michael Stock (Director of Electricity and Mass) BIPM 2 nd July 2013 Bureau International

More information

Report from TCTF/TCL JWG on Optical Frequency Metrology

Report from TCTF/TCL JWG on Optical Frequency Metrology Report from TCTF/TCL JWG on Optical Frequency Metrology Masami Yasuda 1 and Tetsuya Ido 2 1 Time Standards Group, National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial

More information

Final Report for the APMP.T-K4 (Draft B on October 27, 2011)

Final Report for the APMP.T-K4 (Draft B on October 27, 2011) Final Report for the APMP.T-K4 (Draft B on October 27, 2011) Comparison of Realizations of Aluminum Freezing-Point Temperatures Prepared by K. S. Gam (coordinator) and W. Joung Korea Research Institute

More information

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv EUROMET 545: Final report 25/9/28 page 1 of 48 Comparison of national air kerma standards for ISO 437 narrow spectrum series in the range 3 kv to 3 kv L. Büermann (a), M. O Brien (b), D. Butler (c), I.

More information

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002 Final Report CCEM Comparison of 10 pf Capacitance Standards Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002 Abstract Comparison of electrical standards must be carried out periodically

More information

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years Int J Thermophys (2017) 38:161 DOI 10.1007/s10765-017-2291-x TEMPMEKO 2016 Development of the High-Temperature Dew-Point Generator Over the Past 15 Years R. Bosma 1 J. Nielsen 2 A. Peruzzi 1 Received:

More information

WORKING GROUP 4 REPORT TO CCT 25 June 2008

WORKING GROUP 4 REPORT TO CCT 25 June 2008 WORKING GROUP 4 REPORT TO CCT 25 June 2008 CCT/08-13/rev Members: Joachim Fischer (PTB) chairman, Michael DePodesta (NPL), Ken Hill (NRC), Mike Moldover (NIST), Laurent Pitre (LNE-INM/CNAM), Peter Steur

More information

CCL-K5. CMM 1D: Step Gauge and Ball Bars. Final Report

CCL-K5. CMM 1D: Step Gauge and Ball Bars. Final Report Physikalisch-Technische Bundesanstalt CIPM Key Comparison CCL-K5 CMM 1D: Step Gauge and Ball Bars Final Report V3.2 Otto Jusko, PTB, Braunschweig, September 2009 CCL-K5_Final_Report_V3.2(corr).doc Page

More information

Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm nm

Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm nm NPL REPORT COM 1 NPL REPORT OP 12 Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm - 2500 nm TERESA GOODMAN 1, WILLIAM SERVANTES 1, EMMA WOOLLIAMS 1, PETER SPERFELD 2,

More information

Activity measurements of the radionuclide 111 In for the LNE-LNHB, France in the ongoing comparison BIPM.RI(II)-K1.In-111

Activity measurements of the radionuclide 111 In for the LNE-LNHB, France in the ongoing comparison BIPM.RI(II)-K1.In-111 Activity measurements of the radionuclide 111 In for the LNE-LNHB, France in the ongoing comparison BIPM.RI(II)-K1.In-111 C. Michotte*, G. Ratel*, S. Courte*, E. Verdeau, M.-N. Amiot *BIPM, LNE-LNHB, France

More information

EVALUATION OF THE UNCERTAINTY IN THE REALIZATION OF THE PLTS-2000 AT NMi

EVALUATION OF THE UNCERTAINTY IN THE REALIZATION OF THE PLTS-2000 AT NMi EVALUATION OF THE UNCERTAINTY IN THE REALIZATION OF THE PLTS-000 AT NMi A. Peruzzi and M.J. de Groot Nmi van Swinden Laboratorium, Delft, The Netherlands 1 Abstract The Provisional Low Temperature Scale

More information

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS Jan P.M. de Vreede Department of Electricity, Radiation and Length NMi Van Swinden Laboratorium Thijsseweg 11, 2629 JA Delft, the

More information

Boltzmann Project. Determination of k and redefinition 2008: imera+ Project coordinated by PTB

Boltzmann Project. Determination of k and redefinition 2008: imera+ Project coordinated by PTB Boltzmann Project CCT/17-28 2002: Study on k determination with DCGT u r (k) 2 ppm 2005: 1st Boltzmann workshop with 2004: international experts TEMPMEKO: at PTB Lecture with first idea for new definition

More information

Measurement & Uncertainty - Concept and its application

Measurement & Uncertainty - Concept and its application Measurement & Uncertainty - Concept and its application June. 20 2012 JongOh Choi (choijongoh@kriss.re.kr) I don t know what I don t know. If we don t express what we know in the form of numbers, we really

More information

Centro Nacional de Metrología. CIPM Key Comparison CCL KC-6

Centro Nacional de Metrología. CIPM Key Comparison CCL KC-6 CIPM CCL Key comparison: Calibration of CMM 2-D Artifacts 1/162 Centro Nacional de Metrología CIPM Key Comparison CCL KC-6 Calibration of Coordinate Measuring Machine (CMM) Two-dimensional (2-D) Artifacts

More information

Final Report. APMP.EM-K4.1 APMP Key Comparison of Capacitance at 10 pf

Final Report. APMP.EM-K4.1 APMP Key Comparison of Capacitance at 10 pf Final Report APMP.EM-K4.1 APMP Key Comparison of Capacitance at 10 pf H. L. Johnson National Measurement Institute, Australia February 09 1 Contents 1 Introduction...5 Participants and organisation of

More information

Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153

Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153 Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153 G. Ratel*, C. Michotte*, M. Reinhard, D. Alexiev, L. Mo *BIPM, ANSTO, Australia

More information

Temperature measurement

Temperature measurement Temperature measurement and control in length metrology Dr. Youichi Bitou Length standard section National Metrology Institute of Japan (NMIJ) 1 Outline Theoretical treatment of temperature measurement

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays D.T. Burns, W. Tiefenböck* and J. Witzani* Bureau International des Poids et Mesures, Pavillon

More information

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization

More information

C. Michotte 1, G. Ratel 1, S. Courte 1, Y. Nedjadi 2, C. Bailat 2, L. Johansson 3, Y. Hino 4. Abstract

C. Michotte 1, G. Ratel 1, S. Courte 1, Y. Nedjadi 2, C. Bailat 2, L. Johansson 3, Y. Hino 4. Abstract Update of the BIPM comparison BIPM.RI(II)-K1.Ho-166m activity measurements to include the IRA and the NPL and a re-evaluation of the degrees of equivalence for the APMP.RI(II)-K2.Ho-166m comparison C.

More information

Fundamental Temperature Measurement: Re-defining the Boltzmann Constant How do you really know what the temperature is? Michael de Podesta

Fundamental Temperature Measurement: Re-defining the Boltzmann Constant How do you really know what the temperature is? Michael de Podesta Fundamental Temperature Measurement: Re-defining the Boltzmann Constant How do you really know what the temperature is? Michael de Podesta NPL December 2014 Also my wonderful colleagues Gavin Sutton, Robin

More information

Modern Gas-Based Temperature and Pressure Measurements

Modern Gas-Based Temperature and Pressure Measurements Modern Gas-Based Temperature and Pressure Measurements THE INTERNATIONAL CRYOGENICS MONOGRAPH SERIES General Editors K. D. Timmerhaus, Chemical Engineering Department University of Colorado, Boulder, Colorado

More information

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW NPL REPORT ENG 17 Report on EURAMET key comparison of multiples and submultiples of the kilogram (EURAMET.M.M-K2.1) M Perkin NOT RESTRICTED July 2009 National Physical Laboratory Hampton Road Teddington

More information

Comité consultatif de thermométrie (CCT) Consultative Committee for Thermometry (CCT)

Comité consultatif de thermométrie (CCT) Consultative Committee for Thermometry (CCT) Comité consultatif de thermométrie (CCT) Consultative Committee for Thermometry (CCT) Rapport de la 19 e session (septembre 1996) Report of the 19th Meeting (September 1996) Bureau international des poids

More information

On the proposed re definition of the SI

On the proposed re definition of the SI On the proposed re definition of the SI Martin Milton Director, BIPM Metrology Summer School, Varenna Thursday 30th June 2016 The International System of Units (SI) Prefixes Derived units Base units The

More information

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, S. Vörös 2 and B. Hofstetter-Boillat 2 1

More information

CCM-Working Group on Gravimetry (CCM-WGG) Alessandro Germak INRIM, Torino, Italy

CCM-Working Group on Gravimetry (CCM-WGG) Alessandro Germak INRIM, Torino, Italy CCM-Working Group on Gravimetry (CCM-WGG) Alessandro Germak INRIM, Torino, Italy Terms of Reference To propose key comparisons to the CCM; To maintain contact with international organizations and stakeholders

More information

Final Report on APMP.T-K7

Final Report on APMP.T-K7 Final Report on APMP.T-K7 Key Comparison of water triple point cells Shu-Fei Tsai 1, Rod White 2, Jun Tamba 3, Kazuaki Yamazawa 3, Mong-Kim Ho 4, C M Tsui 5, Ghufron Zaid 6, Aditya Achmadi 6, Kee Sool

More information

Summary of the ICRM Life Sciences Working Group Meeting

Summary of the ICRM Life Sciences Working Group Meeting Summary of the ICRM Life Sciences Working Group Meeting 19-20 November 201 National Physical Laboratory, Teddington, UK Jeffery T. Cessna Nuclear Medicine Metrology Meeting, 21 November 2014, NPL, Teddinton,

More information

Determination of Ar molar mass for the Boltzmann constant. Inseok Yang, Jin Bok Lee, Jin Seog Kim Korea Research Institute of Standards and Science

Determination of Ar molar mass for the Boltzmann constant. Inseok Yang, Jin Bok Lee, Jin Seog Kim Korea Research Institute of Standards and Science Determination of molar mass for the Boltzmann constant Inseok Yang, Jin Bok Lee, Jin Seog Kim Korea Research Institute of Standards and Science IPQ, 13 October 2016 Acoustic determination of k and molar

More information

Final Report CCM.V-K2 Comparison

Final Report CCM.V-K2 Comparison Final Report CCM.V-K2 Comparison C. Patrick Maggi, David Trowbridge, M Thomas Zubler Cannon Abstract This report describes the second CCM key comparison in capillary viscometry at nine National Metrology

More information

Project 887 Force Uncertainty Guide. Andy Knott, NPL EUROMET TC-M Lensbury, Teddington, United Kingdom 1 March 2007

Project 887 Force Uncertainty Guide. Andy Knott, NPL EUROMET TC-M Lensbury, Teddington, United Kingdom 1 March 2007 Project 887 Force Uncertainty Guide Andy Knott, NPL EUROMET TC-M Lensbury, Teddington, United Kingdom 1 March 2007 Background EA-10/04 (EAL-G22) EUROMET Calibration Guide EM/cg/04.01/p Participants SMD

More information

Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 C to +20 C

Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 C to +20 C NPL REPORT ENG 57 Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 to +20 S Bell, M Stevens, H Abe, R Benyon, R Bosma,

More information

REPORT OF THE WORKING GROUP ON HARDNESS (WGH) ACTIVITIES (2005) Prepared by A. Germak (Chairman) and S. Low (Secretary)

REPORT OF THE WORKING GROUP ON HARDNESS (WGH) ACTIVITIES (2005) Prepared by A. Germak (Chairman) and S. Low (Secretary) REPORT OF THE WORKING GROUP ON HARDNESS (WGH) ACTIVITIES (2005) Prepared by A. Germak (Chairman) and S. Low (Secretary) The Working Group on Hardness held its sixth meeting at the National Institute of

More information

G. Ratel 1, C. Michotte 1, D. Kryeziu 1, M. Moune 2, A. Iwahara 3. Abstract

G. Ratel 1, C. Michotte 1, D. Kryeziu 1, M. Moune 2, A. Iwahara 3. Abstract Update of the BIPM comparison BIPM.RI(II)-K1.Cr-51 to include new activity measurements for the LNE-LNHB (France) and the pilot study result of the LNMRI (Brazil) G. Ratel 1, C. Michotte 1, D. Kryeziu

More information

EMRP JRP SIB-05. NewKILO. Developing a practical means of disseminating the redefined kilogram (June 2012 May 2015) Stuart Davidson, NPL, UK

EMRP JRP SIB-05. NewKILO. Developing a practical means of disseminating the redefined kilogram (June 2012 May 2015) Stuart Davidson, NPL, UK EMRP JRP SIB-05 NewKILO Developing a practical means of disseminating the redefined kilogram (June 2012 May 2015) Stuart Davidson, NPL, UK CCM M-e-P Workshop BIPM, November 2012 Background This Project

More information

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Points between PTB and LATU with PTB as Pilot Laboratory O. Robatto, Departamento de Metrología Científica, Laboratorio Tecnológico del Uruguay(LATU),

More information

Calibration of temperature sensors within Length Standard Section of NMIJ

Calibration of temperature sensors within Length Standard Section of NMIJ Calibration of temperature sensors within Length Standard Section of NMIJ Dr. Akiko Hirai Length Standard Section National Metrology Institute of Japan (NMIJ) 1 Contents Temperature measurement system

More information

A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards

A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards NPL REPORT CMAM 36(a) A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards This Report Supersedes NPL Report CMAM 36 I Severn SEPTEMBER 2000

More information

The steps to an uncertainty budget

The steps to an uncertainty budget The steps to an uncertainty budget Emma Woolliams 4 April 2017 Prepared by Paul Miller http://www.emceoc.org Uncertainty Where to start? What to do? How to be consistent? Make it easy. At the end of this

More information

Comparability and Traceability Point of view from a metrological institute

Comparability and Traceability Point of view from a metrological institute Comparability and Traceability Point of view from a metrological institute Ed W.B. de Leer NMi Van Swinden Laboratory Delft, The Netherlands 1 My Discussions in Metrologyland 2 The CCQM in 1996 3 Traceability

More information

Comparison between the CENAM (Mexico) 150 kn and the INRiM (Italy) 1 MN force standard machines

Comparison between the CENAM (Mexico) 150 kn and the INRiM (Italy) 1 MN force standard machines IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. Comparison between the CENAM (Mexico) 150 kn and the

More information

Maintaining and disseminating the kilogram following its redefinition

Maintaining and disseminating the kilogram following its redefinition Metrologia PAPER OPEN ACCESS Maintaining and disseminating the kilogram following its redefinition To cite this article: 2017 Metrologia 54 S99 View the article online for updates and enhancements. Related

More information

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report Michele M. Schantz, Reenie M. Parris, and Willie E. May National Institute of Standards and Technology (NIST)

More information

Lab i x i u i Lab i x i u i / (mg/kg) / (mg/kg) / (mg/kg) / (mg/kg)

Lab i x i u i Lab i x i u i / (mg/kg) / (mg/kg) / (mg/kg) / (mg/kg) MEASURAND : Mass fraction of total Arsenic in brown rice flour NOMINAL VALUE : 0.05 mg/kg to 0.5 mg/kg MEASURAND : Mass fraction of Cadmium in brown rice flour NOMINAL VALUE : 0.1 mg/kg to 0.5 mg/kg x

More information

BIPM.RI(II)-K1.Am-241, CCRI(II)-K2.Am-241, and COOMET.RI(II)-K2.Am.241

BIPM.RI(II)-K1.Am-241, CCRI(II)-K2.Am-241, and COOMET.RI(II)-K2.Am.241 MEASURAND : Key comparison BIPM.RI(II)-K1.Am-241 x i : u i : Equivalent activity of 241 Am result of measurement carried out in the SIR for the sample submitted by laboratory i combined standard uncertainty

More information

Measurement & Uncertainty - Basic concept and its application

Measurement & Uncertainty - Basic concept and its application Measurement & Uncertainty - Basic concept and its application Feb. 7. 2012 JongOh CHOI (choijongoh@kriss.re.kr) Knowledge and measurement When you can measure what you are speaking about, and express it

More information

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays D T Burns Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312

More information

VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz

VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz M. Zhao 1,3 and G. F. Strouse 2 1 Fluke Corporation, Hart Scientific Division, American Fork, Utah 84003, U.S.A. 2 National Institute

More information

G. Ratel*, C. Michotte*, Y. Hino** *BIPM and **NMIJ, Japan

G. Ratel*, C. Michotte*, Y. Hino** *BIPM and **NMIJ, Japan Update of the BIPM comparison BIPM.RI(II)-K1.Ce-139 of activity measurements of the radionuclide 139 Ce to include the 2004 NMIJ result and links for the 2004 regional comparison APMP.RI(II)-K2.Ce-139

More information

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017 Metrology Principles for Earth Observation: the NMI view Emma Woolliams 17 th October 2017 Interoperability Decadal Stability Radiometric Accuracy Identical worldwide Century-long stability Absolute accuracy

More information

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW NPL REPORT ENG 4 Report on EUROMET key comparison of 1 kg standards in stainless steel (EUROMET.M.M-K4) M Perkin NOT RESTRICTED February 008 National Physical Laboratory Hampton Road Teddington Middlesex

More information

Abstract. 1. Introduction

Abstract. 1. Introduction The importance of establishing a very-high temperature radiation thermometry measurement capability at the National Metrology Institute of South Africa (NMISA) Speaker: Dr Efrem K Ejigu National Metrology

More information

Guide to the Realization of the ITS-90

Guide to the Realization of the ITS-90 Bureau International des Poids et Mesures Guide to the Realization of the ITS-90 Consultative Committee for Thermometry under the auspices of the International Committee for Weights and Measures CONTENTS

More information

Improving the accuracy of NO 2 and Ozone monitoring

Improving the accuracy of NO 2 and Ozone monitoring Improving the accuracy of NO 2 and Ozone monitoring J. Viallon, E. Flores, F. Idrees, P. Moussay, and R.I. Wielgosz Standards and Measurements for Clean Air 14 October 2016, IPQ Caparica, Portugal Air

More information

REEVALUATION OF THE MEASUREMENT UNCERTAINTY OF UTC TIME TRANSFER

REEVALUATION OF THE MEASUREMENT UNCERTAINTY OF UTC TIME TRANSFER REEVALUATION OF THE MEASUREMENT UNCERTAINTY OF UTC TIME TRANSFER Z. Jiang, W. Lewandowski, G. Panfilo, and G. Petit Bureau International des Poids et Mesures (BIPM) Pavillon de Breteuil F-92312, SEVRES

More information

Euromet project 691 Calibration Inter-comparison of a 5 litre Volume Standard

Euromet project 691 Calibration Inter-comparison of a 5 litre Volume Standard Euromet project 691 Calibration Inter-comparison of a 5 litre Volume Standard Peter Lau SP Technical Research Institute of Sweden Measurement Technology SP Report 2007:10 Euromet project 691 Calibration

More information

Highlights in the work of the BIPM during 2017

Highlights in the work of the BIPM during 2017 Highlights in the work of the BIPM during 2017 Dr Martin Milton BIPM Director 18 October 2017 BIPM highlights of 2017 01 Introduction to the BIPM 02 Liaison work 03 Laboratory work Physical, Time, Ionising

More information

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM 83.8058 K TO 1234.93 K: METHODS AND UNCERTAINTIES Gregory F. Strouse National Institute of Standards and Technology, Gaithersburg, Maryland, USA ABSTRACT

More information

CCQM-K28: Tributyltin in sediment. Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK

CCQM-K28: Tributyltin in sediment. Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK CCQM-K28: Tributyltin in sediment Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK a Institute for National Measurement Standards, National Research Council

More information

BIPM.RI(II)-K1.F-18, BIPM.RI(II)-K4.F18, CCRI(II)-K3.F-18 and APMP.RI(II)-K3.F-18

BIPM.RI(II)-K1.F-18, BIPM.RI(II)-K4.F18, CCRI(II)-K3.F-18 and APMP.RI(II)-K3.F-18 BIPM.RI(II)-K1.F-18, BIPM.RI(II)-K4.F18, CCRI(II)-K3.F-18 and APMP.RI(II)-K3.F-18 MEASURAND : Equivalent activity of 18 F Key comparison BIPM.RI(II)-K1.F-18 result of measurement carried out in the SIR

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES BUREAU INTERNATIONAL DES POIDS ET MESURES Bilateral comparison of 1 Ω and 10 kω standards (ongoing BIPM key comparisons BIPM.EM-K13.a and 13.b) between the SMD (Belgium) and the BIPM October 2017 Final

More information

IAEA stable isotope reference materials: addressing the needs of atmospheric greenhouse gas monitoring.

IAEA stable isotope reference materials: addressing the needs of atmospheric greenhouse gas monitoring. IAEA stable isotope reference materials: addressing the needs of atmospheric greenhouse gas monitoring. S.Assonov, M.Gröning and A. Fajgelj IAEA, Vienna 18th GGMT meeting, La Jolla, Sept 15, 2015 CO2,

More information