MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA USING CdZnTe DETECTOR

Size: px
Start display at page:

Download "MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA USING CdZnTe DETECTOR"

Transcription

1 MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA USING CdZnTe DETECTOR M.Matsumoto 1,S.Miyajima 1,A.Yamamoto 1,T.Yamazaki 1 and Kanamori 2 1 Sch. of Allied Health Sciences, Osaka Univ., 1-7 Yamadaoka, Suita, Osaka Japan 2 Dep. of Electric Engineering,Fukui Univ. Tech.,3-6-1 Gakuen,Fukui Japan INTRODUCTION We directly measured diagnostic X-ray photon spectra produced by a single-phase 2-pulse diagnostic X-ray unit driven at 50 to 100 kv and a few ma(1), a mammographic X-ray unit at 25 to 32 kv and 3 times of 80 mas(2) and an X-ray computed tomography (CT) scanner at 120 kv and 3 ma(3) have been analyzed. Since detected spectra were distorted by the response of CdZnTe(CZT) detector and did not present the true photon spectra, the correction by the stripping procedurec(4) was applied. The response function of detector used in this procedure has been determined by an evaluation of interactions (K-escape, coherent scattering, Compton scattering) and incomplete charge collection calculated using the Monte Carlo method(4-7). Castro et al.(7) have attempted to perform the direct measurement of diagnostic X-ray spectra using a low-efficiency CdTe detector. They analyzed photon spectra produced by a portable diagnostic X-ray spectra unit from 45 to 100kV. They have used only an average value of K α (25 kev) radiation on account of the little difference between Cd(23 kev) and Te(27 kev) absorption coefficients and have not considered K β radiations of Cd and Te. They have used the constant value for the incomplete charge collection and have not considered Compton scattering. Therefore their correction of spectra have been bad accuracy. But we have used K α1,k α2,k β1,k β2 radiations of Cd,Zn and Te respectively and have used the weight function for the incomplete charge collection and have considered Compton scattering. The Monte Carlo simulations were continued with changing the important factors[mean path length of hole λ h, dead layer of CZT crystal (3 3 2 mm 3 ) and weight factor W q ] of incomplete charge collection until the fittest response function was found out. Therefore our correction of spectra have improved the accuracy. Corrected photon spectra were compared with the diagnostic X-ray spectral data published by Bureau of Radiological Health (BRH) measured using a Ge detector. Exposures were calculated from the corrected photon spectra and compared with exposures measured using an ionization chamber. METHOD 1. DIRECT MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA We have directly measured diagnostic X-ray spectra using the Ge (PRINCETONGAMMA-TECH NIGP ) and CZT (TOYO MEDIC RAMTEC 413) detectors. Photon spectra transmitted through an object(1 mmal to 6 mmal) produced by a single-phase 2-pulse diagnostic X-ray unit (SHIMADZU UD150L-5 and CIRCLEX 1/2U13CN-25) driven at 50 to 100 kv and a few ma have been analyzed(1). Each exposure attenuation curve for 50 to 100 kv is calculated and is compared with the curve measured by an ionization chamber (VICTOREEN RADCON MODEL 500 and TYPE 0.1MA PROBE). 2. DIRECT MEASUREMENT OF MAMMOGRAPHIC X-RAY SPECTRA We have directly measured mammographic X-ray spectra using a 35μm-collimator and the CZT detector. Photon spectra transmitted through 0.03 mm Mo or mm Rh filter and object (0.1 mm Al to 1.2 mm Al) produced by a mammographic X-ray unit (GE SENOGRAPHE DMR) at 25 to 32 kv and 3 times of 80 mas have been analyzed(2). Each exposure attenuation curve for 25 to 32 kv is calculated and is compared with the curve measured by an ionization chamber (PTW-FREIBURG TYPE 23344). 3. DIRECT MEASUREMENT COMPUTED TOMOGRAPHY(CT) X-RAY SPECTRA We have directly measured computed tomography (CT) X-ray spectra using the CZT detector. Photon spectra transmitted through object (1 mm Al to 8 mm Al) produced by a CT scanner (HITACHI LSCT) driven at 120 kv and 3 ma have been analyzed(3). The exposure attenuation curve for 120 kv and 3 ma is calculated and is compared with the curve measured by an ionization chamber (VICTOREEN RADCON MODEL 500 and TYPE 0.1MA PROBE). 4. CORRECTION OF MEASURED X-RAY SPECTRA 4.1 CALCULATION OF INCOMPLETE CHARGE COLLECTION The response functions of CZT detector are obtained by the simulation using Monte Carlo method(4-7) to correct measured spectra obtained with the detector. Monte Carlo method are a well-established method

2 for simulating X-ray photon interactions (K-escape, coherent scattering and Compton scattering processes) in the detector using the random number generator. For an incident X-ray photon, the depth of the initial interaction is determined from the total detector crystal cross section. One of the three processes is then selected on the basis of their relative cross sections by random-number sampling. The simulation is continued until the total energy of the X-ray photon will be absorbed as the charge carriers(pairs of electrons and holes) in the detector. If the charge carriers drifting in the detector are trapped and can not reach the collecting electrode, the output pulse height decreases. In this simulation the charge carrier trapping in the CZT detector had to be considered as follows. For a planar type detector, the collected charge Q(t) on the electrode is represented by Q(t) = (q 0 /d) [drift distance of electron + drift distance of hole], (a) where d is the thickness of the detector, and q 0 is the charge carriers created by photon interaction. In Fig. 1, FIG. 1. Simulation model considering the carrier trapping in the CdZnTe semiconductor detector. are created at the interaction position and the distance from this position to the cathode is indicated by x. The the incident X-ray transmitted through the dead layer which is an inactive layer for X-rays, and the charge carriers average collected charge, which is dependent on the position of interaction x, is described as Q(x). The value of Q(x) has twoequations according to the length between the mean free path λ h of hole and x. The mean free path λ h of hole is an average length of holes passed without collisions to electrons. For x λ h, For x > λ h, Q(x) = q 0 Q(x)=q 0 [1- W q (x - λ h )/d ] (b) (c) Where W q is weight factor. The energy depositions in the CZT detector calculated by Monte Carlo method are multiplied by the weight function, which is defined in the square parentheses of equation (c). Therefore, except for the condition [1], the energy multiplied by the weight function is less than the energy when the charge carrier trapping is not considered. In this way, response functions can be simulated considering the charge carrier trapping in the CZT detector as stated in FITTEST MONOENERGETIC RESPONSE FUNCTION Monoenergetic response function and full energy peak efficiency for the CZT detector have been determined by an evaluation of interactions and incomplete charge collection using Monte Carlo method elaborated on a NEC personal computer PC-9821 Ra20. In this simulation, important factors are mean path length of hole λ h, dead layer of the CZT crystal and weight factor W q for incomplete charge collection. This simulation are continued with changing the important factors(λ h =0.005 cm to 0.15 cm at every cm interval, dead layer=0.01 cm to 0.10 cm at every 0.01 cm interval and W q =0.1 to 1.0 at every 0.1 interval) until the fittest response function is found out. The values of factors are decided from the following reasons; (1) the calculated full energy peak efficiency is coincided with measured full energy peak efficiency using isotopes( 241 Am, 109 Cd and 57 Co) as shown in Fig.2, (2) the corrected spectra for the CZT detector are coincide

3 FIG. 2. Comparison of calculated full energy peak efficiencies for any factors and measured full energy peak efficiencies using isotopes( 241 Am, 109 Cd and 57 Co). FIG. 3. Comparison of (a)measured and (b)corrected γ-ray spectra of an isotope 241 Am for Ge(solid line) and CZT(dotted line) detectors. FIG. 4. Comparison of exposure attenuation curves obtained from the CZT spectra and the ionization chamber at (a)70 kv and (b)100 kv. with those for a Ge detector as shown in Fig.3, and (3) exposure attenuation curves calculated from corrected spectra for the CZT detector are coincided with those measured using the ionization chamber as shown in Fig.4. For example the monoenergetic response functions for the CZT detector with λ h =0.025 cm, dead layer=0.06 cm and W q =0.3 obtained at 50, 100 and 150 kev are shown in Fig.5 respectively. The full energy peak efficiencies obtained for the Ge and CZT detectors are shown in Fig. 2. The efficiency for the CZT detector

4 rapidly decreases at higher energy ranges than 20 kev. Therefore the photon spectra in high intensity X-ray beams can be measured by the CZT detector. 4.3 SRRIPPING PROCEDURE Castro et al.(7) have used only an average value of K α (25 kev) radiation on account of the little difference between Cd(23 kev) and Te(27 kev) absorption coefficients and have not considered K β radiations of Cd and Te. They have used the constant value for the incomplete charge collection and have not considered Compton scattering. Therefore their correction of spectra has been bad accuracy. But we use FIG. 5. Monoenergetic response functions obtained at 50, 100 and 150 kev for the CZT detector with λ h = cm, dead layer = 0.06 cm and W q = 0.3. K α1,k α2,k β1,k β2 radiations of Cd,Zn and Te,respectively and the weight function for the incomplete charge collection described in 4.1. And we consider Compton scattering. Therefore our correction of spectra improves the accuracy. We modify the stripping formula presented by Castro et al. into the following formula; Emax N t (E 0 ) = {N d (E 0 ) - R(E 0, E) N t (E)} / ε(e 0 ) (4) E=E 0 Where N t (E 0 ): true number of photons of energy E 0, N d (E 0 ): number of photons detected of energy E 0, E max : maximum photon energy of the spectrum, R(E 0, E): monoenergetic response function of events of energy E due to interactions including incomplete charge collection, ε(e 0 ): full energy peak efficiency. The stripping procedure is applied step by step starting from the maximum photon energy E max. From the counts N d (E 0 ) of every energy E 0 is subtracted the monoenergetic response function R(E 0, E) in Fig. 5 of the true contents of each energy channel greater than E 0. The energy channel contents N t (E 0 ) so obtained are divided by the full energy peak efficiency ε(e 0 ) in Fig. 2. Consequently an example of the results is shown in Fig. 3. Fig. 3 shows the comparison of (a)measured and (b)corrected γ-ray spectra of an isotope 241 Am for Ge(solid line) and CZT(dotted broken line) detectors. The corrected spectrum for the CZT detector coincides with that for the Ge detector. RESULTS AND DISCUSSION The comparison of corrected diagnostic X-ray spectra at (a)70 kv and (b)100 kv for Ge and CZT detectors are shown in Fig. 6. The corrected spectra at (a)70 kv and (b)100 kv for the CZT detector coincide with those for the Ge detector. Each exposure attenuation curve for (a)70 kv and (b)100 kv is calculated and is compared with the curve measured by the ionization chamber in Fig. 7. The obtained results with the CZT detector agree with those obtained by the ionization chamber.

5 FIG. 6. Comparison of corrected diagnostic X-ray spectra at (a)70 kv and (b)100 kv for Ge and CZT detectors. FIG. 7. Comparison of exposure attenuation curves obtained from the CZT spectra and the ionization chamber at (a)70 kv and (b)100 kv. FIG. 8. Comparison of corrected mammographic(mo-mo) X-ray spectra at (a)25 kv and (b) 30kV for the CZT and Ge detectors. The comparison of corrected mammographic(mo-mo) X-ray spectra at (a)25 kv and (b)30 kv for the Ge and the CZT detectors are shown in Fig. 8. The corrected spectrum(a) at (a)25 kv and (b)30 kv for the CZT detector coincides with the spectral data(8) for the Ge detector published by Bureau of Radiological Health(BRH). Each exposure attenuation curve for (a)25 kv(mo-mo) and (b)28 kv(rh-rh) is calculated and is compared with the curve measured by the ionization chamber in Fig. 9. The obtained curves with the CZT detector agree with those obtained by the ionization chamber. The comparison of directly measured (at 120 kv, 3 ma) and corrected primary X-ray spectra of a CT scanner is shown in Fig. 10(a). In Fig. 10(a) the dotted line is the direct measured spectrum and the solid line is

6 the corrected spectrum. The corrected spectrum obtained with the CZT detector roughly agrees with CT X-ray spectral data(9) obtained using the Ge detector published by BRH shown in Fig. 10(b). The exposure attenuation curve at 120 kv and 3 ma is calculated and is compared with the curve measured by the ionization chamber in Fig. 11. The curve obtained with the CZT detector agrees with the curve obtained by the ionization chamber. FIG. 9. Comparison of exposure attenuation curves obtained from the CZT spectra and the ionization chamber at (a)25 kv(mo-mo) and (b)28 kv(rh-rh). FIG. 10. Comparison of (a)photon spectra measured and corrected with (b)ct X-ray spectral data(9) of BRH. FIG. 11. Comparison of exposure attenuation curves obtained from the CZT spectra and the ionization chamber at 120 kv and 3 ma. CONCLUSION Our purpose is to measure directly some diagnostic X-ray spectra using a low-efficiency CdZnTe(CZT)

7 detector developed recently and to find out the fittest response function of CZT detector to correct the measured spectra. Photon spectra produced by a single-phase 2-pulse diagnostic X-ray unit driven at 50 to 100 kv and a few ma, a mammographic X-ray unit at 25 to 32 kv and 3 times of 80 mas and an X-ray computed tomography (CT) scanner at 120 kv and 3 ma have been analyzed. Since detected spectra were distorted by the response of CZT detector and did not present the true photon spectra, the correction by the stripping procedure was applied. The response function of detector used in this procedure has been determined by an evaluation of interactions (Kescape, coherent scattering, Compton scattering) and incomplete charge collection calculated using the Monte Carlo method. The Monte Carlo simulations were continued with changing the important factors(mean path length of hole λ h, dead layer of CZT crystal and weight factor W q ) of incomplete charge collection until the fittest response function was found out. Corrected photon spectra were compared with the diagnostic X-ray spectral data published by Bureau of Radiological Health (BRH) measured using a Ge detector. Exposures were calculated from the corrected photon spectra and compared with exposures measured using an ionization chamber. These results obtained using CZT detector agreed with the diagnostic X-ray spectral data of BRH and exposures obtained by the ionization chamber. Therefore we think that the CdZnTe detector is able to use for quality control and quality assurance of some diagnostic X-ray units. REFERENCES 1. M.Matsumoto, H.Kanamori, T.Toragaito and A.Taniguchi, KEK Proceedings 96-4, Radiation Detectors and Their Uses, (1996) 2. M.Matsumoto,H.Kanamori, A.Taniguchi and M.Yoshida, Medical&Biological Engineering&Computing 35, Supplement Part 2 (1997) M.Matsumoto, H.Kubota, H.Kanamori and M.Yoshida, International Congress on Imaging Science (Antwerp,1998) Vol. 2 of Proceedings, W.W.Seelentag and W.Panzer, Phys. Med. Biol. 24, (1979) 5. C.S.Chen, K.Doi, C.Vyborny, H-P. Chan and G.Holje, Medical Physics 7(6), (1980). 6. C-T.Chen, H-P. Chan and K.Doi, Energy responses of germanium planar detectors used measurement of x-ray spectra in the energy range from 12 to 300keV:Monte Carlo simulation studies, Research Report UCHI-DR/84-01(Univ. Chicago 1984). 7. E.Di.Castro,R.Pani,R.Pellegrini and C.Bacci, Phys. Med. Biol. 29, (1984) 8. R.Fewell and R.E.Shuping, Handbook of mammographic X-ray spectra, Bureau of Radiological Health, Rockville, Maryland, U.S.A. (1978). 9. T.R.Fewell, R.E.Shuping and K.R.Hawkins,Jr., Handbook of computed tomography X-ray spectra, Bureau of Radiological Health, Rockville, Maryland, U.S.A. (1981).

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector A. H. D. Rasolonjatovo 1, T. Shiomi 1, T. Nakamura 1 Y. Tsudaka 2, H. Fujiwara 2, H. Araki 2, K. Matsuo 2, H. Nishizawa 2 1 Cyclotron

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

In order to get the energy response of the CZT detector, the measurements were done using the mono-energetic X-ray beam sources from 10 to 40 kev and

In order to get the energy response of the CZT detector, the measurements were done using the mono-energetic X-ray beam sources from 10 to 40 kev and Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.144-151 Development of Gamma Ray Monitor Using CdZnTe Semiconductor Detector A. H.

More information

Compton Camera. Compton Camera

Compton Camera. Compton Camera Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract

More information

Efficiency and Attenuation in CdTe Detectors

Efficiency and Attenuation in CdTe Detectors Efficiency and Attenuation in CdTe Detectors Amptek Inc. Bob Redus, May 5, 00 Amptek s XR-00T-CdTe is a high performance x-ray and gamma ray detector system. Like Amptek s other XR00 products, a detector

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 26 Radiation Detection & Measurement II Spiritual Thought 2 I would not hold the position in the Church I hold today had I not followed

More information

Development and characterization of 3D semiconductor X-rays detectors for medical imaging

Development and characterization of 3D semiconductor X-rays detectors for medical imaging Development and characterization of 3D semiconductor X-rays detectors for medical imaging Marie-Laure Avenel, Eric Gros d Aillon CEA-LETI, DETectors Laboratory marie-laure.avenel@cea.fr Outlines Problematic

More information

Photon and primary electron arithmetics in photoconductors for digital mammography: Monte Carlo simulation studies

Photon and primary electron arithmetics in photoconductors for digital mammography: Monte Carlo simulation studies Journal of Instrumentation OPEN ACCESS Photon and primary electron arithmetics in photoconductors for digital mammography: Monte Carlo simulation studies To cite this article: T Sakellaris et al View the

More information

Multi-energy CT: Future Directions. Acknowledgements. Overview 7/23/2014. Taly Gilat Schmidt, PhD. Kevin Zimmerman Franco Rupcich Steven Haworth

Multi-energy CT: Future Directions. Acknowledgements. Overview 7/23/2014. Taly Gilat Schmidt, PhD. Kevin Zimmerman Franco Rupcich Steven Haworth Multi-energy CT: Future Directions Taly Gilat Schmidt, PhD Department of Biomedical Engineering Marquette University Acknowledgements Kevin Zimmerman Franco Rupcich Steven Haworth Results in this talk:

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Compton suppression spectrometry

Compton suppression spectrometry Compton suppression spectrometry In gamma ray spectrometry performed with High-purity Germanium detectors (HpGe), the detection of low intensity gamma ray lines is complicated by the presence of Compton

More information

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use Final report on DOE project number DE-FG07-99ID13772 High Pressure Xenon Gamma-Ray Spectrometers for Field Use Principle Investigator: Glenn K. Knoll Co-investigator: David K. Wehe, Zhong He, University

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation S. Moriuchi, M.Tsutsumi2 and K. Saito2 Nuclear safety technology Center,

More information

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 3, JUNE 2002 1547 A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate Wanno Lee, Gyuseong Cho, and Ho

More information

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na Ellen Simmons 1 Contents Introduction Review of the Types of Radiation Charged Particle Radiation Detection Review of Semiconductor

More information

Computational Toolset for X-ray Spectral Analysis

Computational Toolset for X-ray Spectral Analysis Key Engineering Materials Vols. 321-323 (26) pp. 16-165 online at http://www.scientific.net (26) Trans Tech Publications, Switzerland Computational Toolset for -ray Spectral Analysis Cheol-Soon Shon 1,

More information

A Brief Introduction to Medical Imaging. Outline

A Brief Introduction to Medical Imaging. Outline A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length

More information

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY Applications of Nuclear Techniques (CRETE3) International Journal of Modern Physics: Conference Series Vol. 27 (24) 4644 (8 pages) The Authors DOI:.42/S294546446 PERFORMANCE IMPROVEMENT OF CZT DETECTORS

More information

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples)

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples) Detecting high energy photons Interactions of photons with matter Properties of detectors (with examples) Interactions of high energy photons with matter Cross section/attenution length/optical depth Photoelectric

More information

Determination of the activity of radionuclides

Determination of the activity of radionuclides BUREAU NATIONAL DE MÉTROLOGIE COMMISSARIAT À L'ÉNERGIE ATOMIQUE LABORATOIRE NATIONAL HENRI BECQUEREL Note technique LNHB/04-33 Determination of the activity of radionuclides contained in volume samples

More information

Detection and measurement of gamma-radiation by gammaspectroscopy

Detection and measurement of gamma-radiation by gammaspectroscopy Detection and measurement of gamma-radiation by gammaspectroscopy Gamma-radiation is electromagnetic radiation having speed equal to the light in vacuum. As reaching a matter it interact with the different

More information

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices Week 7: Ch. 0 Spec. w/ Scintillation Ctrs. multiplier Devices Spectroscopy with Scint. Counters -- gamma-ray interactions, reprise -- observed spectra --- spectral components, backscatter --- summing --

More information

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA The Underwater Spectrometric System Based on CZT Detector for Survey of the Bottom of MR Reactor Pool 13461 Victor Potapov, Alexey Safronov, Oleg Ivanov, Sergey Smirnov, Vyacheslav Stepanov National Research

More information

Stability and Characteristics of Large CZT Coplanar Electrode Detectors

Stability and Characteristics of Large CZT Coplanar Electrode Detectors 272 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 48, NO. 3, JUNE 2001 Stability and Characteristics of Large CZT Coplanar Electrode Detectors José M. Pérez, Zhong He, Senior Member, IEEE, and David K. Wehe,

More information

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Position Sensitive Germanium Detectors for the Advanced Compton Telescope Position Sensitive Germanium Detectors for the Advanced Compton Telescope R.A. Kroeger 1, W.N. Johnson 1, J.D. Kurfess 1, B.F. Phlips, P.N. Luke 3, M. Momayezi 4, W.K. Warburton 4 1 Naval Research Laboratory,

More information

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra 22.101 Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra References: W. E. Meyerhof, Elements of Nuclear Physics (McGraw-Hill, New York, 1967),

More information

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia 7 th Workshop on European Collaboration for Higher Education and Research in Nuclear Engineering & Radiological Protection Bruxelles, Belgique 30 May - 1 June 2011 Activities at the Laboratory of the Nuclear

More information

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 2001-22, p.37-42 EVALUATION OF ABSORPTION EFFICIENCY FOR NIS TUNNEL JUNCTION DETECTOR WITH SEGMENTED ABSORBERS R. Nouchi, I. Yamada,

More information

Analysis of γ spectrum

Analysis of γ spectrum IFM The Department of Physics, Chemistry and Biology LAB 26 Analysis of γ spectrum NAME PERSONAL NUMBER DATE APPROVED I. OBJECTIVES - To understand features of gamma spectrum and recall basic knowledge

More information

PHYS 3650L - Modern Physics Laboratory

PHYS 3650L - Modern Physics Laboratory PHYS 3650L - Modern Physics Laboratory Laboratory Advanced Sheet Photon Attenuation 1. Objectives. The objectives of this laboratory exercise are: a. To measure the mass attenuation coefficient at a gamma

More information

Chapter V: Cavity theories

Chapter V: Cavity theories Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly

More information

THE spectroscopic performance of large volume CdZnTe

THE spectroscopic performance of large volume CdZnTe 3098 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 6, DECEMBER 2004 Analysis of Detector Response Using 3-D Position-Sensitive CZT Gamma-Ray Spectrometers Feng Zhang, Student Member, IEEE, Zhong He,

More information

Drift plane. substrate (20ÉIm polyimide) 200ÉIm. Back strip (180ÉIm width) Base (Ceramic) Anode strip (10ÉIm width) Cathode strip (100ÉIm width)

Drift plane. substrate (20ÉIm polyimide) 200ÉIm. Back strip (180ÉIm width) Base (Ceramic) Anode strip (10ÉIm width) Cathode strip (100ÉIm width) Proceedings of the Second International Workshop on EGS, 8.-1. August, Tsukuba, Japan KEK Proceedings -, pp.11-17 Development of Gamma-Ray Direction Detector Based on MSGC T. Nagayoshi 1, H. Kubo 1, A.

More information

Calibration of a pencil ionization chamber with and without preamplifier

Calibration of a pencil ionization chamber with and without preamplifier Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof.

More information

RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4. James C. Liu, W. R. Nelson and R. Seefred

RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4. James C. Liu, W. R. Nelson and R. Seefred SLAC-PUB-8606 (September 5, 2000) RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4 James C. Liu, W. R. Nelson and R. Seefred Stanford Linear Accelerator Center, MS 48, P. O. Box 4349, Stanford,

More information

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE Page no. 5.1 Introduction 132 5.2 Methods and measurements 132 5.3 Saturation thickness of elements and polymers

More information

anti-compton BGO detector

anti-compton BGO detector 1 2 3 Q β - measurements with a total absorption detector composed of through-hole HPGe detector and anti-compton BGO detector 4 5 Hiroaki Hayashi a,1, Michihiro Shibata b, Osamu Suematsu a, Yasuaki Kojima

More information

Chapter Seven (Nuclear Detectors)

Chapter Seven (Nuclear Detectors) Al-Mustansiriyah University College of Science Physics Department Fourth Grade Nuclear Physics Dr. Ali A. Ridha Chapter Seven (Nuclear Detectors) Ionizing radiation is rarely detected directly. Instead,

More information

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors Jean-Marc Dinten, Jean-Louis Amans, Loïck Verger, Olivier Peyret CEA-LETI, MINATEC, Recherche

More information

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Poster No.: C-0403 Congress: ECR 2014 Type: Scientific Exhibit Authors:

More information

III. Energy Deposition in the Detector and Spectrum Formation

III. Energy Deposition in the Detector and Spectrum Formation 1 III. Energy Deposition in the Detector and Spectrum Formation a) charged particles Bethe-Bloch formula de 4πq 4 z2 e 2m v = NZ ( ) dx m v ln ln 1 0 2 β β I 0 2 2 2 z, v: atomic number and velocity of

More information

New trends in CdTe detectors for X and γ-ray applications

New trends in CdTe detectors for X and γ-ray applications New trends in CdTe detectors for X and γ-ray applications Olivier Limousin CEA Saclay / DSM / DAPNIA Service d Astrophysique France New developments in photodetection, Beaune 2002 / Solid state detectors

More information

Radiation Dose, Biology & Risk

Radiation Dose, Biology & Risk ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.

More information

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Radiation Detection. 15 th Annual OSC Readiness Training Program. Radiation Detection 15 th Annual OSC Readiness Training Program www.oscreadiness.org GM Detectors 15 th Annual OSC Readiness Training Program www.oscreadiness.org 1 A closer look 15 th Annual OSC Readiness

More information

Researchers at the University of Missouri-Columbia have designed a triple crystal

Researchers at the University of Missouri-Columbia have designed a triple crystal Childress, N. L. and W. H. Miller, MCNP Analysis and Optimization of a Triple Crystal Phoswich Detector, Nuclear Instruments and Methods, Section A, 490(1-2), 263-270 (Sept 1, 2002). Abstract Researchers

More information

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry

More information

Rad T 290 Worksheet 2

Rad T 290 Worksheet 2 Class: Date: Rad T 290 Worksheet 2 1. Projectile electrons travel from a. anode to cathode. c. target to patient. b. cathode to anode. d. inner shell to outer shell. 2. At the target, the projectile electrons

More information

INTERACTIONS OF RADIATION WITH MATTER

INTERACTIONS OF RADIATION WITH MATTER INTERACTIONS OF RADIATION WITH MATTER Renée Dickinson, MS, DABR Medical Physicist University of Washington Medical Center Department of Radiology Diagnostic Physics Section Outline Describe the various

More information

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University Notes to accompany Lab demonstrations by Barry Diacon, Technician, Department of

More information

Gamma-Spectrum Generator

Gamma-Spectrum Generator 1st Advanced Training Course ITCM with NUCLEONICA, Karlsruhe, Germany, 22-24 April, 2009 1 Gamma-Spectrum Generator A.N. Berlizov ITU - Institute for Transuranium Elements Karlsruhe - Germany http://itu.jrc.ec.europa.eu/

More information

THE ACTIVITY CALIBRATOR

THE ACTIVITY CALIBRATOR A.O.U. OSPEDALI RIUNITI di TRIESTE S.C. di FISICA SANITARIA THE ACTIVITY CALIBRATOR Dr. Maria Rosa Fornasier 1 INDEX GENERAL FEATURES DETECTOR DESIGN CALIBRATION PROCEDURE - EFFECTS OF AN EXTERNAL SHIELD

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore

More information

For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric

For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric effect. 1 The objectives of today s lecture are to identify

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

MEASUREMENT AND DETECTION OF RADIATION

MEASUREMENT AND DETECTION OF RADIATION MEASUREMENT AND DETECTION OF RADIATION Second Edition Nicholas Tsoulfanidis University of Missouri-Rolla Ж Taylor &Francis * Publishers since I79H CONTENTS Preface to the First Edition Preface to the Second

More information

arxiv:astro-ph/ v1 6 Mar 2001

arxiv:astro-ph/ v1 6 Mar 2001 Investigation of charge sharing among electrode strips for a CdZnTe detector E. Kalemci a,1 J. L. Matteson a arxiv:astro-ph/0103097v1 6 Mar 2001 Abstract a Center for Astrophysics and Space Sciences, University

More information

Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam

Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam Radhakrishnan B Nair 1*, Ramakrishnan G 2, Chandralingam S 3 and Kurup PGG 1 1 Apollo Speciality Hospital, Chennai, India

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 11 pages Written test, 9 December 2010 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: none. "Weighting": All problems weight equally.

More information

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16530 Quantitative Assessment of Scattering

More information

Physics of Radiotherapy. Lecture II: Interaction of Ionizing Radiation With Matter

Physics of Radiotherapy. Lecture II: Interaction of Ionizing Radiation With Matter Physics of Radiotherapy Lecture II: Interaction of Ionizing Radiation With Matter Charge Particle Interaction Energetic charged particles interact with matter by electrical forces and lose kinetic energy

More information

Spectral Filtering for Improving Quality of Material Discrimination Using Dual Energy X-rays

Spectral Filtering for Improving Quality of Material Discrimination Using Dual Energy X-rays Spectral Filtering for Improving Quality of Material Discrimination Using Dual X-rays Y. M. Gil, Y. S. Lee, M. H. Cho, and W. Namgung POSTECH, PAL POSTECH Abstract The well-known dual energy method of

More information

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS TSOKOS OPTION I-2 MEDICAL IMAGING Reading Activity Answers IB Assessment Statements Option I-2, Medical Imaging: X-Rays I.2.1. I.2.2. I.2.3. Define

More information

Estimate of Large CZT Detector Absolute Efficiency

Estimate of Large CZT Detector Absolute Efficiency 2010 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 4, AUGUST 2002 Estimate of Large CZT Detector Absolute Efficiency J. M. Pérez, Member, IEEE,Z.He, Senior Member, IEEE, D.K.Wehe, Senior Member, IEEE,

More information

X-ray ionization yields and energy spectra in liquid argon

X-ray ionization yields and energy spectra in liquid argon E-print arxiv:1505.02296 X-ray ionization yields and energy spectra in liquid argon A. Bondar, a,b A. Buzulutskov, a,b,* A. Dolgov, b L. Shekhtman, a,b A. Sokolov a,b a Budker Institute of Nuclear Physics

More information

Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure

Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure ILA PILLALAMARRI Earth Atmospheric & Planetary Sciences Neutron Activation Analysis

More information

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of

More information

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 2, APRIL 2012 469 Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions Weiyi Wang, Member,

More information

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers Nuclear Instruments and Methods in Physics Research A 441 (2000) 459}467 Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers Z. He *, W.Li, G.F. Knoll, D.K. Wehe,

More information

Lecture 3 - Compton Scattering

Lecture 3 - Compton Scattering Lecture 3 - Compton Scattering E. Daw March 0, 01 1 Review of Lecture Last time we recalled that in special relativity, as in pre-relativistic dynamics, the total energy in an interaction or collision

More information

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical ORIGINAL ARTICLE Development of MGDA Software for Calculation of Patient Specific Mean Glandular Dose during Mammography 1Meherun Nahar, 2 M Sazzad, 3 Abdus Sattar Mollah and 4 Mir Mohammad Akramuzzaman

More information

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN Problem P7 Stéphanie Ménard Dosimetry Department 92262 Fontenay-aux aux-roses FRANCE What are the applications of Gamma-Ray Spectrometry in Radiological Protection and in Safety? In the environment: after

More information

Chapter 16 Basic Precautions

Chapter 16 Basic Precautions Chapter 16 Basic Precautions 16.1 Basic Principles of Radiation Protection The four basic methods used to control radiation exposure are time, distance, shielding, and contamination control. The first

More information

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification

More information

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields Journal of NUCLEAR SCIENCE and TECHNOLOGY, Vol. 44, No. 2, p. 109 113 (2007) ARTICLE A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard

More information

Slides by: Prof. Abeer Alharbi

Slides by: Prof. Abeer Alharbi Slides by: Prof. Abeer Alharbi electromagnetic radiation of high energy. They are produced by sub-atomic particle interactions, such as electron-positron annihilation, neutral pion decay, radioactive decay,

More information

Attenuation of Radiation in Matter. Attenuation of gamma particles

Attenuation of Radiation in Matter. Attenuation of gamma particles Attenuation of Radiation in Matter In this experiment we will examine how radiation decreases in intensity as it passes through a substance. Since radiation interacts with matter, its intensity will decrease

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

arxiv:physics/ v1 3 Aug 2006

arxiv:physics/ v1 3 Aug 2006 Gamma Ray Spectroscopy with Scintillation Light in Liquid Xenon arxiv:physics/6834 v1 3 Aug 26 K. Ni, E. Aprile, K.L. Giboni, P. Majewski, M. Yamashita Physics Department and Columbia Astrophysics Laboratory

More information

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National

More information

Chapter 8 Hyper-Pure Germanium Detector

Chapter 8 Hyper-Pure Germanium Detector Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 8-1 8.1. Introduction Chapter 8 Hyper-Pure Germanium Detector Silicon semiconductor detectors described in the previous chapter have depletion

More information

Results of a Si/CdTe Compton Telescope

Results of a Si/CdTe Compton Telescope Results of a Si/CdTe Compton Telescope Kousuke Oonuki a,b, Takaaki Tanaka a,b, Shin Watanabe a, Shin ichiro Takeda a,b, Kazuhiro Nakazawa a, Takefumi Mitani a,b, Tadayuki Takahashi a,b, Hiroyasu Tajima

More information

Radionuclide Imaging MII Detection of Nuclear Emission

Radionuclide Imaging MII Detection of Nuclear Emission Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor

More information

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: 1 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: XRF spectroscopy 1 exam question. Notes on: www.nuigalway.ie/nanoscale/3rdspectroscopy.html XRF books: Analytical Chemistry,

More information

Performance of semi-insulating. insulating GaAs-based radiation detectors: Role of key physical parameters of base material

Performance of semi-insulating. insulating GaAs-based radiation detectors: Role of key physical parameters of base material Institute of Electrical Engineering Slovak Academy of Sciences, Bratislava Slovak republic Performance of semi-insulating insulating GaAs-based radiation detectors: Role of key physical parameters of base

More information

LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH

LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH Lesson Objectives At the end of the lesson, student should able to: Define attenuation Explain interactions between x-rays and matter in

More information

APPLIED RADIATION PHYSICS

APPLIED RADIATION PHYSICS A PRIMER IN APPLIED RADIATION PHYSICS F A SMITH Queen Mary & Westfield College, London fe World Scientific m Singapore * New Jersey London Hong Kong CONTENTS CHAPTER 1 : SOURCES of RADIATION 1.1 Introduction

More information

GEANT4 Simulations of. the LAXPC Performance. H. M. Antia. Tata Institute of Fundamental Research

GEANT4 Simulations of. the LAXPC Performance. H. M. Antia. Tata Institute of Fundamental Research GEANT4 Simulations of the LAXPC Performance H. M. Antia Tata Institute of Fundamental Research GEANT4 simulations of LAXPC detector were used to estimate Efficiency of background rejection Efficiency and

More information

Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S****

Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S**** Photon Interaction Cross Section of Materials in Heterogeneous Energy Spectrum of Medical Diagnostic X-Ray Beam Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S**** *Senior Medical Physicist,

More information

Phonon-Mediated Distributed Transition- Edge-Sensor X-ray Detectors

Phonon-Mediated Distributed Transition- Edge-Sensor X-ray Detectors Phonon-Mediated Distributed Transition- Edge-Sensor X-ray Detectors Steven W. Leman Department of Physics, Stanford University TES 3 at Gainesville 18Aug06 Goals, applications and detector basics Energy

More information

ON THE OMNIPRESENT BACKGROUND GAMMA RADIATION OF THE CONTINUOUS SPECTRUM

ON THE OMNIPRESENT BACKGROUND GAMMA RADIATION OF THE CONTINUOUS SPECTRUM ON THE OMNIPRESENT BACKGROUND GAMMA RADIATION OF THE CONTINUOUS SPECTRUM R, Banjanac, D. Maletić, D. Joković, N. Veselinović, A. Dragić, V. Udovičić, I. Aničin Institute of Physics, University of Belgrade,

More information

Spatial resolved efficiency determination of CdZnTe semiconductor detectors with a collimated gamma-ray source for the C0BRA experiment

Spatial resolved efficiency determination of CdZnTe semiconductor detectors with a collimated gamma-ray source for the C0BRA experiment Institute of Nuclear and Particle Physics, Prof. K. Zuber Spatial resolved efficiency determination of CdZnTe semiconductor detectors with a collimated gamma-ray source for the C0BRA experiment Arnd Sörensen

More information

Chapiter VII: Ionization chamber

Chapiter VII: Ionization chamber Chapiter VII: Ionization chamber 1 Types of ionization chambers Sensitive volume: gas (most often air direct measurement of exposure) ionization chamber Sensitive volume: semiconductor (silicon, germanium,

More information

Effect of the time characteristics of the Compton camera on its performance

Effect of the time characteristics of the Compton camera on its performance Effect of the time characteristics of the Compton camera on its performance Chibueze Zimuzo Uche Howell Round Michael Cree School of Engineering University of Waikato Introduction Anger camera Compton

More information

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector ISSN: 1791-4469 Copyright 2012 Hellenic Naval Academy A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector K. Zachariadou a,c, K. Karafasoulis b,c, S. Seferlis c, I. Papadakis

More information

Progress in Nuclear Science and Technology, Volume 6,

Progress in Nuclear Science and Technology, Volume 6, DOI: 1.15669/pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp. 1-16 ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi

More information

Recent advances and future perspectives of gamma imagers for scintimammography

Recent advances and future perspectives of gamma imagers for scintimammography 3rd International Conference on Imaging Technologies in Biomedical Sciences: ITBS2005 Innovation in Nuclear and Radiological Imaging: From Basic Research to Clinical Application Milos Conference Center,

More information

HEAVY ion therapy is an advantageous modality over the

HEAVY ion therapy is an advantageous modality over the Measurement of electron density and effective atomic number using dual-energy x-ray CT T. Tsunoo, M. Torikoshi, Y. Ohno, M. Endo, M. Natsuhori, T. Kakizaki, N. Yamada, N. Ito, N. Yagi, and K. Uesugi Abstract

More information