How to Make Photometric & Colorimetric Measurements of Light Sources using an Ocean Optics Spectrometer and SpectraSuite Software

Size: px
Start display at page:

Download "How to Make Photometric & Colorimetric Measurements of Light Sources using an Ocean Optics Spectrometer and SpectraSuite Software"

Transcription

1 How to Make Photometric & Colorimetric Measurements of Light Sources using an Ocean Optics Spectrometer and SpectraSuite Software This is a concise guide to setting up and calibrating your Ocean Optics spectrometer for making measurements of the illuminance and colour of a light source. Ocean Optics SpectraSuite software is a very powerful and versatile application, but light source measurements only require that you use a small proportion of its capabilities. The aim of this guide is to acquaint the operator with only that information necessary to perform light source measurements. For a more complete review of SpectraSuite, please refer to the operator manual that comes as a PDF file on the software installation CD. In terms of hardware, you will have an Ocean Optics spectrometer (e.g. USB2000+) with optical fibre and cosine diffuser (CC-3). You will also need a calibrated reference light source (e.g. LS-1-CAL). This guide assumes that you have previously installed SpectraSuite on your computer and copied the calibration files for the reference light source onto the computer (e.g. save them on the Desktop). A. Hardware Preparations 1. Connect the spectrometer to a USB port on your PC using the supplied USB cable. 2. Launch the SpectraSuite software application. 3. Remove the dust cap from the input to the spectrometer and attach one end of the optical fibre. 4. Attach the CC-3 cosine diffuser to the other end of the optical fibre. 5. Insert the CC-3 into the opening on the end of your reference light source. On the LS-1-CAL, you may first need to remove the SMA connector using the Allen key supplied (the SMA connector is left in place if the CC-3 diffuser is not fitted to the fibre, but we strongly recommend that you do not use a bare fibre for photometric or colorimetric measurements). 6. Insert the power jack into the rear of the calibrated light source and turn it on. (N.B. we recommend that you keep a log book to record how long you operate the calibrated light source for each time you use it. The calibration of the LS-1-CAL is valid for 50 hours of operation). Pro-Lite Technology LLP, 2009 Page 1 of 20

2 B. Performing a New Calibration The first time that you use the spectrometer (and at regular intervals thereafter), you need to create a new absolute irradiance calibration as follows. Please note that when you launch SpectraSuite, a spectrum graph opens automatically. This is called Graph (A). Ocean Optics refer to this chart as displaying spectral data in Scope Mode. Be very careful not to use Scope Mode data; scope mode is useful only for gauging the level of light relative to the spectrometer s dynamic range. Scope Mode simply charts the spectral power distribution in units of counts (Y axis) versus wavelength (x-axis). It is important to understand that the spectrum displayed in Scope Mode neither allows you to infer quantitative nor qualitative information about a light source. In other words, the shape of the spectrum will be displayed incorrectly in scope mode. This is because the raw data does not take account of the spectral sensitivity of the spectrometer and sampling optics. Even if you don t need to determine absolute light levels, and simply wish to see the spectral shape of a light source, you must perform a calibration against a reference light source. This is called a Relative Irradiance calibration, but this guide will explain how to perform an Absolute Irradiance calibration. An absolute irradiance calibration not only reports the absolute value of the light (spectral irradiance is reported in μw/cm2.nm), it also correctly displays the spectral power distribution (the spectral shape). Pro-Lite Technology LLP, 2009 Page 2 of 20

3 1. Select File, New, New Absolute Irradiance Measurement to open the Absolute Irradiance Calibration Wizard. 2. Select New Spectral Acquisition, then click on Next. Pro-Lite Technology LLP, 2009 Page 3 of 20

4 3. The spectral source (i.e. spectrometer) that you wish to calibrate will be highlighted as shown below. Click on Next. 4. Select New Calibration then click on Next. Pro-Lite Technology LLP, 2009 Page 4 of 20

5 5. Click on the Set Automatically button to set the optimum integration time. The software will choose a measurement integration time that corresponds to about 85% of full scale. For the USB2000+, the full scale is about 65,000 counts, so the optimum integration time will be when the Last Peak Value is about 55,000 counts. 6. Click on the Advanced Settings button and click in the check box next to Electric Dark Correction. 7. Set Scans to Average to 3, then click on Next. 8. Store the Reference Spectrum by clicking on the yellow light bulb icon then press Next. Remember to switch your reference light source off at this stage to help maintain the calibration life of your lamp. Pro-Lite Technology LLP, 2009 Page 5 of 20

6 9. Store a Dark Spectrum. First you must block off all light going into the spectrometer. The best way of doing this is to remove the fibre from the input to the spectrometer and replace the dust cap. Once you have blocked the light, click on the black light bulb and press Next. Pro-Lite Technology LLP, 2009 Page 6 of 20

7 10. Load the calibration data file for your reference lamp into SpectraSuite. To do this, click on Browse and navigate to the location that you saved the calibration files to (e.g. the Desktop). Select the file with the _CC suffix in the file name this is the correct file for the case of the CC-3 cosine diffuser fitted to the input of the optical fibre. An example of the file name is LSC2105_CC.LMP. Now press Next. 11. Select Fibre Diameter and enter the value 3900 (this is the diameter of the CC-3 diffuser). Pro-Lite Technology LLP, 2009 Page 7 of 20

8 12. Click on the Save to File button to save your new calibration. The default file location chosen by SpectraSuite is C:/Program Files/Ocean Optics/SpectraSuite/spectrasuite/bin. Use a naming convention that helps you to remember what the calibration relates to, for example you could use the file name CC-3 Irradiance Cal 23 Sept 09.IrradCal. 13. When you press Finish, the calibration wizard closes and a new chart will appear in SpectraSuite called Absolute Irradiance (B). Note that the other chart Graph (A) continues to display the raw counts versus wavelength data from the spectrometer. This is termed scope mode. Please note that in scope mode, the shape of the spectrum for any light source will be displayed erroneously scope mode should not be used for anything other than judging whether the light level is sufficient. Pro-Lite Technology LLP, 2009 Page 8 of 20

9 C. Performing a Photometric Measurement Once you have created and saved an absolute irradiance calibration, you can now start to make photometric measurements on one of your own light sources. With the spectrometer equipped with an optical fibre with cosine diffuser at its input, you are able to measure the illuminance (in lux) from a light source. Remember that illuminance varies with the distance between the receiver and the light source, so be sure to record your working distance when measuring illuminance. Point the optical fibre towards your light source and hold it in place. 1. Click on the I icon in the tool bar at the top right of the main SpectraSuite tool bar to perform a new absolute irradiance measurement. This is a similar process to the calibration that you have just performed, but now you will load the calibration you just created rather than creating a new calibration. 2. Select New Spectral Acquisition, then click on Next. Pro-Lite Technology LLP, 2009 Page 9 of 20

10 3. The spectral source (i.e. spectrometer) that you wish to use for the measurement will be highlighted as shown below. Click on Next. 4. Select the calibration that you just created from the list of saved calibrations. Press Next. Pro-Lite Technology LLP, 2009 Page 10 of 20

11 5. Select Fibre Diameter. The collection diameter of 3900 should already be showing. Press Next. 6. Click on the Set Automatically button to set the optimum integration time for your test light source. The software will choose a measurement integration time that corresponds to about 85% of full scale. For the USB2000+, the full scale is about 65,000 counts, so the optimum integration time will be when the Last Peak Value is about 55,000 counts. Please note that the USB2000+ has a minimum integration time of 1ms; if the light level is too high (i.e. above 55,000 counts) at 1ms, you will need to move the fibre further away from the light source. Press Next. Pro-Lite Technology LLP, 2009 Page 11 of 20

12 7. Click on the Advanced Settings button and click in the check box next to Electric Dark Correction. 8. Set Scans to Average to 3, then click on Next. 9. Store a Dark Spectrum. First you must block off all light going into the spectrometer. The best way of doing this is to remove the fibre from the input to the spectrometer and replace the dust cap. Once you have blocked the light, click on the black light bulb and press Finish. Pro-Lite Technology LLP, 2009 Page 12 of 20

13 10. Select Show Data in New Graph and then press Accept. 11. Click in the new chart area to activate the controls. Click on the candle icon at the far right in the tool bar directly above the chart area. Pro-Lite Technology LLP, 2009 Page 13 of 20

14 12. Click the Accept button when the Configure Photometry dialogue box opens. 13. The Photometry Report window will appear as shown below. Note that we have not defined the steradian field of view of the spectrometer so the values displayed for candelas will be erroneous. Similarly, you are not using an integrating sphere so the value for lumens will also be wrong. The illuminance (lux) value on the other hand is correct! Note the use of engineering format in the reported values E2 means 2 x 10 2 which equals 200, while E3 equals 2 x 10 3 or 2,000. Pro-Lite Technology LLP, 2009 Page 14 of 20

15 D. Performing a Colour Measurement Once you have created and saved an absolute irradiance calibration, you can also start to make colour measurements on your own light sources. The colour parameters which SpectraSuite computes include CIE chromaticity xy (2º observer), u v (10º observer), CCT (correlated colour temperature), dominant wavelength and colour rendering. Point the optical fibre towards your light source and hold it in place. 1. Click on the chromaticity diagram symbol (multicoloured triangle) in the tool bar at the top right of the main SpectraSuite tool bar. 2. In the New Colour Measurement dialogue box, select New Absolute Irradiance Processing and press Accept. Pro-Lite Technology LLP, 2009 Page 15 of 20

16 3. The spectral source (i.e. spectrometer) that you wish to use for the measurement will be highlighted as shown below. Click on Next. 4. Select Get Irradiance Calibration from file and click on Next. Pro-Lite Technology LLP, 2009 Page 16 of 20

17 5. Select the calibration that you just created from the list of saved calibrations. Press Next. 6. Select Fibre Diameter. The collection diameter of 3900 should already be showing. Press Next. Pro-Lite Technology LLP, 2009 Page 17 of 20

18 7. Click on the Set Automatically button to set the optimum integration time for your test light source. The software will choose a measurement integration time that corresponds to about 85% of full scale. For the USB2000+, the full scale is about 65,000 counts, so the optimum integration time will be when the Last Peak Value is about 55,000 counts. Please note that the USB2000+ has a minimum integration time of 1ms; if the light level is too high (i.e. above 55,000 counts) at 1ms, you will need to move the fibre further away from the light source. 8. Click on the Advanced Settings button and click in the check box next to Electric Dark Correction. 9. Set Scans to Average to 3, then click on Next. Pro-Lite Technology LLP, 2009 Page 18 of 20

19 10. Store a Dark Spectrum. First you must block off all light going into the spectrometer. The best way of doing this is to remove the fibre from the input to the spectrometer and replace the dust cap. Once you have blocked the light, click on the black light bulb and press Finish. 11. Select the Emissive Mode (this is always the selection when measuring light sources). Also select reference illuminant E for normalisation (this affects the calculation of dominant wavelength) then press Accept. Pro-Lite Technology LLP, 2009 Page 19 of 20

20 12. The Colour Report window and CIE Degree Observer Chromaticity Diagram will appear as shown below. Prepared by: Robert Yeo Pro-Lite Technology 24 h October 2009 Pro-Lite Technology LLP, 2009 Page 20 of 20

Measuring ph with Smart Cuvettes

Measuring ph with Smart Cuvettes 1 Measuring ph with Smart Cuvettes Hardware Setup for ph Measurements ph is measured in Absorbance mode. Configure your spectroscopy components for Absorbance measurements. Sample Absorbance setups for

More information

Ocean Optics Educational Spectrometers. Getting Started Manual

Ocean Optics Educational Spectrometers. Getting Started Manual Ocean Optics Educational Spectrometers Getting Started Manual We here at Ocean Optics want you to know that we appreciate your business. To speak to a Technical Support representative or an Ocean Optics

More information

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Phys316 Exploration 2: Verifying Stefan-Boltzmann Relationship Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Where A is the effective radiating area,

More information

Miniature Fiber Optic Spectrometry StellarNet Tutorial

Miniature Fiber Optic Spectrometry  StellarNet Tutorial StellarNet Tutorial Using SpectraWiz VBA-NIST-CRI/CQS for MS Excel Contents This tutorial describes the basic measurement procedure for using a StellarNet SpectroRadiometer system with the free SpectraWiz

More information

Experiment 4 Radiation in the Visible Spectrum

Experiment 4 Radiation in the Visible Spectrum Experiment 4 Radiation in the Visible Spectrum Emission spectra can be a unique fingerprint of an atom or molecule. The photon energies and wavelengths are directly related to the allowed quantum energy

More information

2. To measure the emission lines in the hydrogen, helium and possibly other elemental spectra, and compare these to know values.

2. To measure the emission lines in the hydrogen, helium and possibly other elemental spectra, and compare these to know values. 4.1. Purpose 1. To record several elemental emission spectra using arc lamps filled with each element using the Ocean Optics USB650 spectrometer. 2. To measure the emission lines in the hydrogen, helium

More information

Spectrometer User s Guide

Spectrometer User s Guide Spectrometer User s Guide (Order Codes: V-SPEC, SPRT-VIS, SP-VIS, SP-UV-VIS, ESRT-VIS) The spectrometer is a portable light spectrophotometer, combining a spectrometer and a light source/cuvette holder.

More information

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) The UV-VIS spectrometer is a portable ultraviolet light and visible light spectrophotometer, combining a spectrometer and a light source/cuvette

More information

FIELD SPECTROMETER QUICK-START GUIDE FOR FIELD DATA COLLECTION (LAST UPDATED 23MAR2011)

FIELD SPECTROMETER QUICK-START GUIDE FOR FIELD DATA COLLECTION (LAST UPDATED 23MAR2011) FIELD SPECTROMETER QUICK-START GUIDE FOR FIELD DATA COLLECTION (LAST UPDATED 23MAR2011) The ASD Inc FieldSpec Max spectrometer is a precision instrument designed for obtaining high spectral resolution

More information

Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer.

Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer. Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer. The screen shown in Fig. 1 may be displayed. If status line displays

More information

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

UV-Vis optical fiber assisted spectroscopy in thin films and solutions UV-Vis optical fiber assisted spectroscopy in thin films and solutions Description UV-Visible absorption and transmission spectra provide fundamental information for all experiments related to the attenuation

More information

DOAS measurements of Atmospheric Species

DOAS measurements of Atmospheric Species Practical Environmental Measurement Techniques: DOAS measurements of Atmospheric Species Last change of document: April 14, 2014 Supervisor: Dr. Andreas Richter, room U2090, tel 62103, and Dr. Folkard

More information

THERMAL RADIATION. The electromagnetic radiation emitted by a hot tungsten filament will be studied.

THERMAL RADIATION. The electromagnetic radiation emitted by a hot tungsten filament will be studied. THERMAL.1 THERMAL RADIATION The electromagnetic radiation emitted by a hot tungsten filament will be studied. Theory: The Stefan-Boltzmann Law relates the rate at which an object radiates thermal energy

More information

Jasco V-670 absorption spectrometer

Jasco V-670 absorption spectrometer Laser Spectroscopy Labs Jasco V-670 absorption spectrometer Operation instructions 1. Turn ON the power switch on the right side of the spectrophotometer. It takes about 5 minutes for the light source

More information

Measurement method for the proficiency testing program

Measurement method for the proficiency testing program APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law Vernier Spectrometer 1 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You

More information

Quick Start Guide New Mountain Visit our Website to Register Your Copy (weatherview32.com)

Quick Start Guide New Mountain Visit our Website to Register Your Copy (weatherview32.com) Quick Start Guide New Mountain Visit our Website to Register Your Copy (weatherview32.com) Page 1 For the best results follow all of the instructions on the following pages to quickly access real-time

More information

Experiment C-10 Titration of a Strong Acid and a Strong Base

Experiment C-10 Titration of a Strong Acid and a Strong Base 1 Experiment C-10 Titration of a Strong Acid and a Strong Base Objectives To study the titration process. To follow changes in the ph during the titration process while adding a strong base to a strong

More information

St art. rp m. Km /h 1: : : : : : : : : : : : :5 2.5.

St art. rp m. Km /h 1: : : : : : : : : : : : :5 2.5. modified 22/05/14 t 3:5 2.5 3:5 5.0 3:5 7.5 4:0 0.0 4:0 2.5 4:0 5.0 4:0 7.5 4:1 0.0 4:1 2.5 4:1 5.0 4:1 7.5 4:2 0.0 Km /h 0 25 50 75 100 125 150 175 200 225 rp m 0 250 0 500 0 750 0 100 00 125 00 1 2 3

More information

Magnetic Fields. Experiment 1. Magnetic Field of a Straight Current-Carrying Conductor

Magnetic Fields. Experiment 1. Magnetic Field of a Straight Current-Carrying Conductor General Physics Lab Department of PHYSICS YONSEI University Lab Manual (Lite) Magnetic Fields Ver.20181029 NOTICE This LITE version of manual includes only experimental procedures for easier reading on

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Physics Department. Experiment 03: Work and Energy

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Physics Department. Experiment 03: Work and Energy MASSACHUSETTS INSTITUTE OF TECHNOLOGY Physics Department Physics 8.01 Fall Term 2010 Experiment 03: Work and Energy Purpose of the Experiment: In this experiment you allow a cart to roll down an inclined

More information

DAY LABORATORY EXERCISE: SPECTROSCOPY

DAY LABORATORY EXERCISE: SPECTROSCOPY AS101 - Day Laboratory: Spectroscopy Page 1 DAY LABORATORY EXERCISE: SPECTROSCOPY Goals: To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are

More information

Computer simulation of radioactive decay

Computer simulation of radioactive decay Computer simulation of radioactive decay y now you should have worked your way through the introduction to Maple, as well as the introduction to data analysis using Excel Now we will explore radioactive

More information

Lab VI Light Emitting Diodes ECE 476

Lab VI Light Emitting Diodes ECE 476 Lab VI Light Emitting Diodes ECE 476 I. Purpose This experiment examines the properties of light emitting diodes. The use of the monochromator for studying the spectrum of light sources is also examined.

More information

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Rate Law Determination of Crystal Violet Hydroxylation Revised 5/22/12 RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Adapted from "Chemistry with Computers" Vernier Software, Portland OR, 1997

More information

Software BioScout-Calibrator June 2013

Software BioScout-Calibrator June 2013 SARAD GmbH BioScout -Calibrator 1 Manual Software BioScout-Calibrator June 2013 SARAD GmbH Tel.: ++49 (0)351 / 6580712 Wiesbadener Straße 10 FAX: ++49 (0)351 / 6580718 D-01159 Dresden email: support@sarad.de

More information

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Operation Manual SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC. 4.18 Foreword

More information

MEASUREMENT: PART II

MEASUREMENT: PART II 1 MEASUREMENT: PART II Copyright: Department of Chemistry, University of Idaho, Moscow, ID 83844-2343, 2013. INTRODUCTION Read and/or review Section 1.7 and Figure 7.5 in your textbook. The first part

More information

Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis

Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis Mike Spilde Spring IOM Seminar February 5, 2008 Qualitative Analysis Why use qualitative scans? Elemental ID (especially trace

More information

Our Drop Counter sensor now features housing for two electrode sensors, an anti-twist mechanism, an indicator LED and two cable guides.

Our Drop Counter sensor now features housing for two electrode sensors, an anti-twist mechanism, an indicator LED and two cable guides. The Drop Counter sensor is an optical sensor that accurately records the number of drops of titrant added during a titration. The Drop Counter sensor software can automatically convert the number of drops

More information

Kinetics of Crystal Violet Bleaching

Kinetics of Crystal Violet Bleaching Kinetics of Crystal Violet Bleaching Authors: V. C. Dew and J. M. McCormick* From Update March 12, 2013 with revisions Nov. 29, 2016 Introduction Chemists are always interested in whether a chemical reaction

More information

Diffraction Gratings and Spectral Lines

Diffraction Gratings and Spectral Lines Diffraction Gratings and Spectral Lines 1. Introduction Chemists long knew that when certain elements were burned they emitted characteristic colors. Sodium, for example, in a flame emits a particular

More information

Lab Activity H4 It s Snow Big Deal

Lab Activity H4 It s Snow Big Deal Lab Activity H4 It s Snow Big Deal OUTCOMES After completing this lab activity, the student should be able to use computer-based data acquisition techniques to measure temperatures. draw appropriate conclusions

More information

Alpha-Energies of different sources with Multi Channel Analyzer

Alpha-Energies of different sources with Multi Channel Analyzer Physical Structure of Matter Radioactivity Alpha-Energies of different sources with Multi Channel Analyzer What you can learn about Decay series Radioactive equilibrium Isotopic properties Decay energy

More information

M61 1 M61.1 PC COMPUTER ASSISTED DETERMINATION OF ANGULAR ACCELERATION USING TORQUE AND MOMENT OF INERTIA

M61 1 M61.1 PC COMPUTER ASSISTED DETERMINATION OF ANGULAR ACCELERATION USING TORQUE AND MOMENT OF INERTIA M61 1 M61.1 PC COMPUTER ASSISTED DETERMINATION OF ANGULAR ACCELERATION USING TORQUE AND MOMENT OF INERTIA PRELAB: Before coming to the lab, you must write the Object and Theory sections of your lab report

More information

Applying MapCalc Map Analysis Software

Applying MapCalc Map Analysis Software Applying MapCalc Map Analysis Software Generating Surface Maps from Point Data: A farmer wants to generate a set of maps from soil samples he has been collecting for several years. Previously, he would

More information

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P ) Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P2522015) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik Experiment:

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Dr. White 11 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To prepare a Beer s

More information

PolarSync Quick Start

PolarSync Quick Start PolarSync Quick Start Installation and Use In this Quick Start guide, we will cover installing the PolarSync program and using it as a teacher, student or guest. I. Installing PolarSync... 1 II. Teacher

More information

Circular Motion and Centripetal Force

Circular Motion and Centripetal Force [For International Campus Lab ONLY] Objective Measure the centripetal force with the radius, mass, and speed of a particle in uniform circular motion. Theory ----------------------------- Reference --------------------------

More information

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5.

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5. HMS-5000 Manual Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K - Manual version: ver 5.01- www.ecopia21.co.kr - Table of contents - 1. Hardware Installation

More information

How Do We Get Light from Matter: The Origin of Emission

How Do We Get Light from Matter: The Origin of Emission 1 How Do We Get Light from Matter: The Origin of Emission Lines ORGANIZATION Pre-Lab: Origins of Lines Mode: inquiry, groups of 2 Grading: lab notes and post-lab questions Safety: no special requirements

More information

Lab 5: Calculating an equilibrium constant

Lab 5: Calculating an equilibrium constant Chemistry 162 The following write-up is inaccurate for the particular chemicals we are using. Please have all sections up through and including the data tables ready before class on Wednesday, February

More information

Common Acids and Bases

Common Acids and Bases Common Acids and Bases Objective To determine the ph values of common household substances. Modules and Sensors Computer with NeuLog TM Software Data Cable, 10cm S98242-48ND USB Module - S98242-45ND ph

More information

Designing a Quilt with GIMP 2011

Designing a Quilt with GIMP 2011 Planning your quilt and want to see what it will look like in the fabric you just got from your LQS? You don t need to purchase a super expensive program. Try this and the best part it s FREE!!! *** Please

More information

Polar alignment in 5 steps based on the Sánchez Valente method

Polar alignment in 5 steps based on the Sánchez Valente method 1 Polar alignment in 5 steps based on the Sánchez Valente method Compared to the drift alignment method, this one, allows you to easily achieve a perfect polar alignment in just one step. By "perfect polar

More information

CLEA/VIREO PHOTOMETRY OF THE PLEIADES

CLEA/VIREO PHOTOMETRY OF THE PLEIADES CLEA/VIREO PHOTOMETRY OF THE PLEIADES Starting up the program The computer program you will use is a realistic simulation of a UBV photometer attached to a small (diameter=0.4 meters) research telescope.

More information

Remember that C is a constant and ë and n are variables. This equation now fits the template of a straight line:

Remember that C is a constant and ë and n are variables. This equation now fits the template of a straight line: CONVERTING NON-LINEAR GRAPHS INTO LINEAR GRAPHS Linear graphs have several important attributes. First, it is easy to recognize a graph that is linear. It is much more difficult to identify if a curved

More information

To Create a Simple Formula using the Point and Click Method:

To Create a Simple Formula using the Point and Click Method: To Create a Simple Formula that Adds Two Numbers: Click the cell where the formula will be defined (C5, for example). Type the equal sign (=) to let Excel know a formula is being defined. Type the first

More information

Protocol for Use of BCMP Jasco J-815 Circular Dichroism Spectropolarimeter

Protocol for Use of BCMP Jasco J-815 Circular Dichroism Spectropolarimeter Protocol for Use of BCMP Jasco J-815 Circular Dichroism Spectropolarimeter Getting Started 1. Turn on the nitrogen: Open the valve on the nitrogen tank (max 10 psi) so that the pressure gauge on the left

More information

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5.

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5. HMS-5000 Manual Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K - Manual version: ver 5.01- www.ecopia21.co.kr - Table of contents - 1. Hardware Installation

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law The primary objective of this experiment is to determine the concentration of an unknown cobalt (II) chloride solution. You will use a Vernier SpectroVis

More information

Operation of the Bruker 400 JB Stothers NMR Facility Department of Chemistry Western University

Operation of the Bruker 400 JB Stothers NMR Facility Department of Chemistry Western University Operation of the Bruker 400 JB Stothers NMR Facility Department of Chemistry Western University 1. INTRODUCTION...3 1.1. Overview of the Bruker 400 NMR Spectrometer...3 1.2. Overview of Software... 3 1.2.1.

More information

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base INTRODUCTION Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base Chemists frequently make use of the equivalent weight (eq. wt.) as the basis for volumetric calculations.

More information

PHYS133 Lab 7 The HR Diagram

PHYS133 Lab 7 The HR Diagram PHYS133 Lab 7 Goals: Measure brightness of various stars in the Pleiades star cluster in two different wavelength bands. Create an HR diagram based on the data taken. Use the distance modulus to determine

More information

FLS980 Series Reference Guide

FLS980 Series Reference Guide FLS980 Series Reference Guide Integrating Sphere for Measurements of Fluorescence Quantum Yields and Spectral Reflectance Revision 1 Copyrights Copyright 2016 Edinburgh Instruments Ltd. All rights reserved.

More information

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK) LM-79-08 Test Report For AOK LED LIGHT COMPANY LIMITED (Brand Name: AOK) Building 1, St George's Science and Technology Industrial Park, Shajin Street, Shenzhen, Guangdong Province, China Zip 518104 Outdoor

More information

ISSP User Guide CY3207ISSP. Revision C

ISSP User Guide CY3207ISSP. Revision C CY3207ISSP ISSP User Guide Revision C Cypress Semiconductor 198 Champion Court San Jose, CA 95134-1709 Phone (USA): 800.858.1810 Phone (Intnl): 408.943.2600 http://www.cypress.com Copyrights Copyrights

More information

User Manual RGB Lasersysteme GmbH Qstick Spectrometer Version: Date: November 04, 2013

User Manual RGB Lasersysteme GmbH Qstick Spectrometer Version: Date: November 04, 2013 User Manual User Manual RGB Lasersysteme GmbH Qstick Spectrometer Version: 1.0.3 Date: November 04, 2013 This document is protected by copyright. Do not copy or publish this document or parts of it without

More information

Safety: BE SURE TO KEEP YOUR SMART CART UPSIDE-DOWN WHEN YOU RE NOT ACTIVELY USING IT TO RECORD DATA.

Safety: BE SURE TO KEEP YOUR SMART CART UPSIDE-DOWN WHEN YOU RE NOT ACTIVELY USING IT TO RECORD DATA. Why do people always ignore Objective: 1. Determine how an object s mass affects the friction it experiences. 2. Compare the coefficient of static friction to the coefficient of kinetic friction for each

More information

In Association With. Introduces. TPT Contractor

In Association With. Introduces. TPT Contractor In Association With Introduces TPT Contractor Software for Contractors with TENSIONMETERS Table of Contents Capabilities 4 Setup and Use 7 The Dashboard 12 Tower Setup Form 18 Field Calibration Verification

More information

Experiment 14 It s Snow Big Deal

Experiment 14 It s Snow Big Deal Experiment 14 It s Snow Big Deal OUTCOMES After completing this experiment, the student should be able to: use computer-based data acquisition techniques to measure temperatures. draw appropriate conclusions

More information

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Application note Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Introduction Applications that use spectrometers to measure the light energy of radiant sources require an irradiance-calibration,

More information

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA.

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA. LM-79-08 Test Report For REMPHOS TECHNOLOGIES LLC (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA LED Lamp Model name(s): RPT-LEDBIAX-1500LM-4000K Representative (Tested) Model: RPT-LEDBIAX-1500LM-4000K

More information

O P E R A T I N G M A N U A L

O P E R A T I N G M A N U A L OPERATING MANUAL WeatherJack OPERATING MANUAL 1-800-645-1061 The baud rate is 2400 ( 8 bits, 1 stop bit, no parity. Flow control = none) To make sure the unit is on line, send an X. the machine will respond

More information

Acid-Base Titration. Sample

Acid-Base Titration. Sample Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to

More information

Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3)

Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3) Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3) 1.0 Purpose - This technical procedure shall be followed for the operation of the GRIM 3. 2.0 Scope - This procedure applies

More information

Obtain an optical "bench" setup (there should be three sliding mounts on the calibrated horizontal bar. The setup is shown in the diagram below.

Obtain an optical bench setup (there should be three sliding mounts on the calibrated horizontal bar. The setup is shown in the diagram below. Astronomy 100 Name(s): Exercise 4: Telescopes and spectroscopy Once the various focal issues are resolved, magnification of a small image is a significant consideration for a telescope. Though a planet

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Saddleback College Dr. White 1 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To

More information

NVLAP LAB CODE LM Test Report. For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A)

NVLAP LAB CODE LM Test Report. For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A) LM-79-08 Test Report For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A) 7 F, Jinchangda Industrial Park,zhangkengjing,GuanLan, Baoao,Shenzhen,Guangdong,China 2x2 Luminaires for Ambient Lighting of Interior

More information

Ideal Gas Law and Absolute Zero

Ideal Gas Law and Absolute Zero Experiment IX Ideal Gas Law and Absolute Zero I. Purpose The purpose of this lab is to examine the relationship between the pressure, volume and temperature of air in a closed chamber. To do this, you

More information

LAB: Photometry of the Pleiades Cluster

LAB: Photometry of the Pleiades Cluster LAB: Photometry of the Pleiades Cluster ASTR 203 - Instructors Olszewski & Rigby Due IN CLASS on Oct. 30 You may work with 1 partner. If you do, only turn in 1 assignment with both your names on it! You

More information

Life Cycle of Stars. Photometry of star clusters with SalsaJ. Authors: Daniel Duggan & Sarah Roberts

Life Cycle of Stars. Photometry of star clusters with SalsaJ. Authors: Daniel Duggan & Sarah Roberts Photometry of star clusters with SalsaJ Authors: Daniel Duggan & Sarah Roberts Photometry of star clusters with SalsaJ Introduction Photometry is the measurement of the intensity or brightness of an astronomical

More information

Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader

Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader 08 Jan 15 Page 1 of 18 Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader The BMG LABTECH CLARIOstar Microplate Readers were tested for compatibility with the Life Technologies Z -LYTE Assay

More information

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK) LM-79-08 Test Report For AOK LED LIGHT COMPANY LIMITED (Brand Name: AOK) Building 1, St George's Science and Technology Industrial Park, Shajin Street, Shenzhen, Guangdong Province, China Zip 518104 Outdoor

More information

Paramaters. Resolution 0.2 m (for measurements < 25 meas. / sec )

Paramaters. Resolution 0.2 m (for measurements < 25 meas. / sec ) Paramaters Method of altitude measurement: Measuring Range measurement of pressure changes 0-5500 m Resolution 0.2 m (for measurements < 25 meas. / sec ) Accuracy: Supply Voltage: Current consumption:

More information

Linear Motion with Constant Acceleration

Linear Motion with Constant Acceleration Linear Motion 1 Linear Motion with Constant Acceleration Overview: First you will attempt to walk backward with a constant acceleration, monitoring your motion with the ultrasonic motion detector. Then

More information

Zetasizer Nano-ZS User Instructions

Zetasizer Nano-ZS User Instructions Zetasizer Nano-ZS User Instructions 1. Activate the instrument computer by logging in to CORAL. If needed, log in to the local instrument computer Username: zetasizer. Password: zetasizer. 2. Instrument

More information

Possible Prelab Questions.

Possible Prelab Questions. Possible Prelab Questions. Read Lab 2. Study the Analysis section to make sure you have a firm grasp of what is required for this lab. 1) A car is travelling with constant acceleration along a straight

More information

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) LM-79-08 Test Report For Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) Rm 1902, Easey Comm Bldg,253-261 Hennessy Rd Wanchai, HONG KONG Four-Foot Linear Replacement Lamps Model name(s): HC-T8-4FT-22W-ID(X,Clear)

More information

Information Dependent Acquisition (IDA) 1

Information Dependent Acquisition (IDA) 1 Information Dependent Acquisition (IDA) Information Dependent Acquisition (IDA) enables on the fly acquisition of MS/MS spectra during a chromatographic run. Analyst Software IDA is optimized to generate

More information

Quantum States and Spectra of Gases

Quantum States and Spectra of Gases Page 1 of 12 RECOMMENDED READINGS Quantum States and Spectra of Gases ONE WEIGHT 1) R. Harris: Modern Physics, Ch.4, pp. 122-124; Ch.7 pp. 238-268; Ch.8.4-8.9. Addison Wesley, 2008. 2) PASCO Wireless Spectrometer

More information

NVLAP LAB CODE LM Test Report. For. LED Wav es,inc. (Brand Name: LED Wav es)

NVLAP LAB CODE LM Test Report. For. LED Wav es,inc. (Brand Name: LED Wav es) LM-79-08 Test Report For LED Wav es,inc (Brand Name: LED Wav es) 4100 1st Ave,The Esquire Building,Brooklyn, NY 11232 2x2 Luminaires for Ambient Lighting of Interior Commercial Spaces Model name(s): SL2-2X2-40W-11

More information

PHY221 Lab 2 - Experiencing Acceleration: Motion with constant acceleration; Logger Pro fits to displacement-time graphs

PHY221 Lab 2 - Experiencing Acceleration: Motion with constant acceleration; Logger Pro fits to displacement-time graphs Page 1 PHY221 Lab 2 - Experiencing Acceleration: Motion with constant acceleration; Logger Pro fits to displacement-time graphs Print Your Name Print Your Partners' Names You will return this handout to

More information

Creating Empirical Calibrations

Creating Empirical Calibrations 030.0023.01.0 Spreadsheet Manual Save Date: December 1, 2010 Table of Contents 1. Overview... 3 2. Enable S1 Calibration Macro... 4 3. Getting Ready... 4 4. Measuring the New Sample... 5 5. Adding New

More information

Sample Alignment (2D detector) Part

Sample Alignment (2D detector) Part Sample Alignment (2D detector) Part Contents Contents 1 How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 6 2 Sample alignment sequence...13

More information

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) LM-79-08 Test Report For Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) Rm 1902, Easey Comm Bldg,253-261 Hennessy Rd Wanchai, HONG KONG Four-Foot Linear Replacement Lamps Model name(s): HC-T8-4FT-13W-ID(X,

More information

41. Sim Reactions Example

41. Sim Reactions Example HSC Chemistry 7.0 41-1(6) 41. Sim Reactions Example Figure 1: Sim Reactions Example, Run mode view after calculations. General This example contains instruction how to create a simple model. The example

More information

NVLAP LAB CODE LM Test Report. For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503

NVLAP LAB CODE LM Test Report. For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503 Report No.: GZE150726-B LM-79-08 Test Report For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503 Four-Foot Linear Replacement Lamps Model name(s): IGTTU-1501850F-4BC

More information

Software. Tip. Software

Software. Tip. Software Our comprehensive library of data acquisition and display software helps you make the most of your spectroscopic measurements and analysis. No matter what operating system you prefer, chances are excellent

More information

LightLab I N T E R N A T I O N A L

LightLab I N T E R N A T I O N A L Report of Test LL15785 Performance Summary Luminous flux 540 lm Bare Lamp Power 7.76W SHR Nominal 0.75 SHR Maximum 0.79 PREPARED FOR : Indice Ecotech Pty. Ltd., Richmond, VIC 3121. Page 1 of 7 Performance

More information

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products White Paper Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products 1. Introduction The LED industry is growing rapidly and this naturally brings up an important need for reliable

More information

LED Entryway Light RoHS

LED Entryway Light RoHS 72508 LED Entryway Light RoHS QPL ID # PSMN8JUU Model: 72508 Input Voltage 120-277 VAC 60HZ Input Current 0.5A Max @120V Input Power 25W Power Factor PF 0.95 OVERALL LAMP PARAMETERS Luminance 3,121 LM

More information

Technical notes: using the optical/ellipsometric film thickness measurement setup.

Technical notes: using the optical/ellipsometric film thickness measurement setup. Technical notes: using the optical/ellipsometric film thickness measurement setup. Light source Optical probe Spectrometer Sample stage 1) Place a sample (for example, Si/SiO 2 wafer coated with parylene)

More information

Acid-Base Titration. Evaluation copy

Acid-Base Titration. Evaluation copy Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to

More information

Standard-Tech Co. Ltd Testing Center

Standard-Tech Co. Ltd Testing Center Test Datasheet Test Location: [ ] 1. B802, No.11 Caipin Road, Guangzhou Science City, Guangzhou, [ ] Guangdong, China 2. R108, 1 st Floor No.69 GuangPu West Road, Guangzhou Science City, Guangzhou, Guangdong,

More information

Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra

Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra OUTLINE OF THE PROCEDURE A) Measure the extinction spectra of each of your samples. B) Model the extinction spectrum of

More information

Simple circuits - 3 hr

Simple circuits - 3 hr Simple circuits - 3 hr Resistances in circuits Analogy of water flow and electric current An electrical circuit consists of a closed loop with a number of different elements through which electric current

More information

LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl350u1

LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl350u1 LM-79-08 Test Report For Simkar Corporation (Brand Name: ) LED Luminaire Model name(s):salho2xl350u1 Test & Report By: Garman Mo Engineer: Garman Mo Date: Feb.20,2017 Review By: Manager: Tommy Liang Note:

More information

C101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science

C101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science C101-E112 BioSpec-nano Shimadzu Spectrophotometer for Life Science Power of small. BioSpec-nano BioSpec-nano Shimadzu Spectrophotometer for Life Science Quick and Simple Nucleic Acid Quantitation Drop-and-Click

More information