Parameter estimation using macroscopic diffusion MRI signal models

Size: px
Start display at page:

Download "Parameter estimation using macroscopic diffusion MRI signal models"

Transcription

1 Parameter estimation using macroscopic diffusion MRI signal models Hang Tuan Nguyen 3, Denis Grebenkov 2, Dang Van Nguyen 1,3, Cyril Poupon 3, Denis Le Bihan 3, Jing-Rebecca Li 1,3 1 INRIA Saclay-Equipe DEFI,CMAP, Ecole Polytechnique Route de Saclay, 91128, Palaiseau Cedex, France 2 Laboratoire de Physique de la Matière Condensée, CNRS Ecole Polytechnique, F Palaiseau Cedex, France 3 NeuroSpin, CEA Saclay Center Gif-sur-Yvette Cedex France jingrebecca.li@inria.fr 1. Introduction Abstract. Macroscopic models of the diffusion MRI (dmri) signal can be helpful to understanding the relationship between the tissue microstructure and the dmri signal. We study the least squares problem associated with estimating tissue parameters such as the cellular volume fraction, the residence times, and the effective diffusion coefficients using a recently developed macroscopic model of the dmri signal called the Finite Pulse Kärger model that generalizes the original Kärger model to non-narrow gradient pulses. In order to analyze the quality of the estimation in a controlled way, we generated synthetic noisy dmri signals by including the effect of noise on the exact signal produced by the Finite Pulse Kärger model. The noisy signals were then fitted using the macroscopic model. Minimizing the least squares, we estimated the model parameters. The bias and standard deviations of the estimated model parameters as a function of the signal to noise ratio (SNR) were obtained. We discuss the choice of the b-values, the least square weights, the extension to experimentally obtained dmri data as well noise correction. In diffusion magnetic resonance imaging (dmri), water motion inside a biological tissue is monitored in order to characterize the tissue microstructure. Varying magnetic field gradients and diffusion times, one acquires the macroscopic signal (at millimeter spatial resolution) which aggregates statistically averaged information about the domain at the micrometer scale. Inferring the microstructure from the macroscopic signal is a formidable inverse problem (Grebenkov 2007). The complexity of brain tissue does not allow the extraction of the whole microstructure. Rather, one focuses on estimating some macroscopic parameters of the tissue whose changes can be related to physiological or pathological modifications. In the diffusion tensor model (Basser et al. 1994) water diffusion was modeled as a Gaussian process whereas the diffusion kurtosis model (Jensen et al. 2005, Chabert et al. 2005) included additional degrees of freedom for a better fit of the dmri signal. The dmri signal has been also modeled as the sum of two or multiple decaying exponentials, representing, for example, fast and slow Gaussian diffusion pools. Although the explicit form of this representation makes the bi-exponential model very attractive, naive identifications of the extracted volume fractions of the pools as volume

2 Parameters estimation using macroscopic diffusion MRI signal models 2 fractions of the physiological constituents of the tissue, i.e., the cells (slow diffusion) and the extracellular space (fast diffusion), have failed (Niendorf et al. 1996, Mulkern et al. 1999, Clark & Le Bihan 2000, Lee & Springer 2003, Schwarcz et al. 2004, Ababneh et al. 2005, Chin et al. 2002, Stanisz 2003). There also have been macroscopic models that explicitly use certain biological quantities as model parameters. For example, if the neurites are modeled by cylinders, while the extracellular space as a Gaussian pool, the total dmri signal can be represented as the sum of the two explicitly know signals from both compartments (Sen & Basser 2005, Assaf et al. 2008, Jespersen et al. 2007). The radius of neurites appears here as a parameter of the model. Note that this explicit representation relies on the Gaussian phase approximation (Grebenkov 2007) and assumes no water exchange between the compartments. The Kärger model (Karger et al. 1988) allows for water exchange between effective Gaussian compartments and incorporates macroscopic parameters such as the cellular volume fraction, the residence times, as well as the effective diffusion coefficients. When the gradient pulses are narrow (as compared to diffusion time), the time evolution of the signals in the different compartments can be described by a system of coupled ordinary differential equations (ODEs). The ODE system can be solved analytically, yielding an explicit dependence of the dmri signal on the model parameters. This explicit representation of the dmri signal can then be used to estimate the residence times and the volume fractions of water molecules in the effective Gaussian diffusion pools. Fieremans et al. proposed a coarse graining argument to justify the concept of co-existing pools of water molecules, and validated the Kärger model by Monte Carlo simulations (Fieremans et al. 2010) (see also (Meier et al. 2003, Nilsson et al. 2010) for Monte Carlo simulations). In particular, the slow exchange between compartments (i.e., low permeability) was shown to be necessary for compartment separation. In spite of these limitations, the Kärger model has been often used in biological tissue imaging to invert for model parameters (Waldeck et al. 1997, Stanisz et al. 1997, Pfeuffer et al. 1998, Lee & Springer 2003, Quirk et al. 2003, Meier et al. 2003, Roth et al. 2008, Aslund et al. 2009, Nilsson et al. 2009). Recently, a new macroscopic model for the dmri signal was formulated (Coatléven et al. 2014, Li et al. 2014) using homogenization techniques (Bensoussan et al. 1978). The compartments of this macroscopic model are the biological cells and the extracellular space that are separated from each other by cell membranes. By matching asymptotic expansions of the solution of the microscopic Bloch-Torrey partial differential equation (PDE) in different compartments, the dmri signal is given as the solution of a system of coupled ODEs. These ODEs are similar to that of the Kärger model, but they have time-dependent coefficients so that the solution is obtained numerically. As the main advantage, this new macroscopic model is not subject to the narrow pulse restriction, in contrast to the Kärger model. The gradient pulses can be long in duration and can have an arbitrary temporal profile. The work of (Coatléven et al. 2014) can be thought of as giving a sound mathematical basis to the Kärger model as well as generalizing it beyond narrow gradient pulses. In this paper, we consider the weighted least squares problem associated with estimating the macroscopic model parameters such as the cellular volume fraction, the residence times and the effective diffusion coefficients using the Finite Pulse Kärger (FPK) model. There have been many works on estimating the parameters of the diffusion tensor and the diffusion kurtosis models in the presence of noise and noise correction, see (Veraart et al. 2011, Veraart, Sijbers, Sunaert, Leemans & Jeurissen 2013, Veraart, Rajan, Peeters, Leemans, Sunaert & Sijbers 2013, Iima

3 Parameters estimation using macroscopic diffusion MRI signal models 3 et al. 2014, Collier et al. 2014), and references therein. The major difference between the diffusion tensor or kurtosis models and the FPK model that we consider in this paper is that the data cannot be easily represented as a linear function of the model parameters (as it was done, e.g., in the diffusion tensor model by taking the logarithm of the signal). In other words, one has to deal with nonlinear fitting so that the parameters estimation becomes much more involved. To proceed with our study, we first solve the Bloch-Torrey equation for the complex transverse magnetization on a geometry containing spherical cells of various sizes to show that the total complex transverse magnetization is accurately represented by the solution of the Finite Pulse Kärger (FPK) model over an appropriate range of b-values. In this way, FPK model is identified as an appropriate macroscopic fitting frame for this geometrical example. We then investigate in depth the robustness of parameters estimation within this model. Following the standard scheme, we first simulate the dmri signal within the FPK model with prescribed parameters, add noise, and then fit the noisy signal to the FPK model by minimizing the least squares in order to estimate the model parameters. The quality of the estimation procedure can be accessed by comparing the estimated parameters to their true values. The bias and standard deviations of the estimated model parameters as a function of the signal to noise ratio (SNR) were obtained. We discuss the choice of the b-values, the least square weights, the extension to experimentally obtained dmri data as well noise correction. 2. Theory To validate the Finite Pulse Kärger (FPK) model, we define a simplified geometrical model of a biological tissue in a computational box C = [ L 1 /2, L 1 /2] [ L 2 /2, L 2 /2] [ L 3 /2, L 3 /2] and assume that C is periodically repeated in all three dimensions to make up the imaging voxel. Inside C, there are sphere compartments, Ω j, j = 2, (where each Ω j is a sphere), while the remaining set Ω 1 = C\ j 2 Ωj is called the extracellular compartment Microscopic Bloch-Torrey equation A ground truth dmri signal can be computed by solving the multiple compartment Bloch-Torrey PDE in C (Torrey 1956, Grebenkov 2010) M l (r, t g) = if(t)(γg r) M l (r, t g) + (D 0 M l (r, t g)) (r Ω l ), (1) t where M l (r, t g) is the complex-valued transverse water proton magnetization at r = (r 1, r 2, r 3 ) in compartment Ω l, g = (g 1, g 2, g 3 ) is the diffusion-encoding gradient, f(t) is its effective temporal profile, i is the imaginary unit, γ = rad/s/tesla is the gyromagnetic ratio of the water proton, and D 0 is the intrinsic diffusion coefficient of water molecules. The PDEs in Eq. (1) are supplemented with two interface conditions on the interface Γ ln between any two adjacent compartments Ω l and Ω n. One interface condition is the continuity of magnetization flux: D 0 ( M l (r, t g) n l (r) ) = D 0 ( M n (r, t g) n n (r)), r Γ ln, (2) where n l (r) and n n (r) are the outward-pointing normals to Ω l and Ω n at r. second interface condition, The D 0 ( M l (r, t g) n l (r) ) = κ ( M n (r, t g) M l (r, t g) ), r Γ ln, (3)

4 Parameters estimation using macroscopic diffusion MRI signal models 4 incorporates a permeability κ across Γ ln and models the ease with which water crosses the interface. The larger the κ, the easier the passage of water. We add the initial condition: M l (r, 0 g) = M 0, r Ω l, l, (4) where the uniform excitation of the nuclei over the whole voxel is assumed, with M 0 being the initial magnetization. Since the computational box C = [ L 1 /2, L 1 /2] [ L 2 /2, L 2 /2] [ L 3 /2, L 3 /2] is periodically repeated, we add the boundary conditions on C following (Xu et al. 2007): M(r, t g) rk = L k /2 = M(r, t g) rk =L k /2 ei θk(t), k = 1, 2, 3, (5) M(r, t g) M(r, t g) = r k r k e i θk(t), k = 1, 2, 3, (6) rk =L k /2 rk = L k /2 for each of the faces perpendicular to the three coordinate axes, where θ k (t) = t γ g k L k f(s) ds. 0 Now we define the compartment magnetization as the integral of the magnetization in Ω l : M l P DE(b, t) := M l (r, t g) dr, r Ω l 0 t T E, (7) where T E is the echo time at which the dmri signal is acquired, and the b-value is defined as: T E ( u 2 b(g) = γ 2 g 2 du f(s)ds). (8) 0 0 The dmri signal measured in experiments (without the imaging gradients and T 2 effects) corresponds to P P S P DE (b) := M l P DE(b, T E) = M l (r, T E g) dr, r Ω l (9) l=0 where P is the number of compartments. Because the computational box C is periodically repeated, S P DE (b) provides the dmri signal in the whole voxel. In a dmri experiment, the T E and sequence f(t) are usually fixed while g is varied in amplitude and/or in direction to obtain the signal at different b-values. In the rest of the paper, we will set the initial magnetization to M 0 = 1 C, so that S P DE (b = 0) = 1 and M l P DE(b, t = 0) = Ωl C := v l, where Ω l is the volume of Ω l, C is the volume of the computational box, and v l is the volume fraction of Ω l. A standard pulsed-gradient spin-echo (PGSE) sequence is composed of two rectangular gradient pulses of duration δ separated by diffusion time : 1 t s < t < t s + δ, f(t) = 1 t s + < t < t s + + δ, (10) 0 otherwise, where t s is the start of the first pulse. Although both gradient pulses are of the same polarity in a usual spin-echo sequence, the second gradient pulse is effectively inverted (i.e., f(t) = 1) by the refocusing 180 rf pulse. The b-value from Eq. (8) becomes l=1 b = γ 2 g 2 δ 2 ( δ/3). (11)

5 Parameters estimation using macroscopic diffusion MRI signal models Macroscopic Kärger model The Kärger model (Karger et al. 1988) is formulated under the narrow pulse approximation, δ, for which the compartment magnetizations M l KAR(b, t), l = 1,, P satisfy dm m KAR(b, t) P = δ 2 γ 2 g T D m 1 g + dt τ lm M m KAR(b, t) l=1, l m (12) P 1 + τ ml M l KAR(b, t), m = 1,, P, l=1, l m where D m is the effective diffusion tensor in compartment m, and 1/τ lm is the transfer rate from m-th compartment to l-th compartment. Note that in the case of two compartments, τ 12 can be interpreted as the residence time in the first (extracellular) compartment. The volume fractions v l are normalized as P v l = 1. (13) l=1 These coupled ODEs are subject to the initial conditions: M l KAR(b, 0) = v l, l = 1,, P. (14) While the dmri signal is the sum of the M l KAR from all the compartments at t = : P M l KAR(b, ), l=1 we found that the Kärger model can be made more accurate when the signal is evaluated at t = δ/3, similar to the appearance of the factor δ/3 in the b-value in Eq. (11). Thus, we will use the expression P S KAR (b) = M l KAR(b, δ/3) l=1 for the dmri signal attenuation when comparing the Kärger model with the Finite Pulse Kärger model to be described next. The Kärger ODEs with constant coefficients are easy to solve analytically or numerically. For the two-compartment case (P = 2, i.e., one extracellular compartment Ω 1 and one intracellular compartment Ω 2 ), the Kärger dmri signal is given explicitly, where q δγ g, D ± 1 2 S KAR (b) = ρ exp( D + q 2 t) + (1 ρ) exp( D q 2 t), (D 1 g + D 2 g) + 1 q 2 ( 1 τ τ 21 ) ( ± (Dg 2 Dg) ( 1 q 2 1 )) 2 + τ 21 τ 12 4 q 4, τ 12 τ 21 ρ (v1 D 1 g + v 2 D 2 g) D D + D.

6 Parameters estimation using macroscopic diffusion MRI signal models Macroscopic Finite Pulse Kärger (FPK) model In (Coatléven et al. 2014), a macroscopic Finite Pulse Kärger (FPK) model was obtained from the microscopic multiple-compartment Bloch-Torrey PDE (1-6) using periodic homogenization theory. The narrow pulse assumption on the diffusion encoding sequence was not needed. The Kärger model can be retrieved from the FPK model in the narrow pulse limit: δ. The FPK model describes the time evolution of the compartment magnetizations as dm m F P K(b, t) P = c(t)γ 2 g T D m 1 g + dt τ lm M m F P K(b, t) l=1, l m (15) P 1 + τ ml M l F P K(b, t), m = 1,, P, l=1, l m where the function c(t) was derived in (Coatléven et al. 2014) to be ( t 2 c(t) := f(s)ds) (16) 0 for any profile f(t). For the standard PGSE sequence from Eq. (10), one finds t 2, t s t t s + δ, c(t) = δ 2, t s + δ < t t s +, (t δ) 2, t s + < t t s + + δ. In the narrow pulse regime, δ, Eq. (17) yields the coefficient δ 2 of the Kärger model. For connected Ω l C, the derivation of (Coatléven et al. 2014) shows that from which (17) 1 τ ml := κ Γml Ω l, (l m), (18) τ lm τ ml = Ωm Ω l = vm v l, (19) where Γ ml is the surface area of the interface between Ω m and Ω l. We note here that Eq. (18) has been used for the Kärger model as well. The initial condition of the FPK model is the same as for the Kärger model, namely, Eq. (14), and the dmri signal is P S F P K (b) = M l F P K(b, T E). (20) l= Effective diffusion tensors Now we define the effective diffusion tensors describing diffusion in the compartments in the long time limit. When the computational box C is small compared to the diffusion distance, and under the assumption that C is periodically repeated,

7 Parameters estimation using macroscopic diffusion MRI signal models 7 the effective diffusion tensor D m in compartment Ω m can be rigorously defined as (Coatléven et al. 2014) D m j,k = 1 v m D 0 W j (r) e k dr, j, k = 1,, 3, (21) Ω m where e k is the unit vector in the k th direction; one has to solve three Laplace equations over C for three unknown functions W j : (D 0 W j (r) ) = 0, r Ω m, j = 1,, 3, (22) subject to impermeable interface condition on Ω m : W j (r) n(r) = 0, r Ω m, (23) and the following boundary conditions on C: W j (r) rk = L k /2 = W j (r) rk =L k /2 δ j,kl k, k = 1,, 3, (24) W j (r) = W j (r), k = 1,, 3, (25) r k r k rk = L k /2 rk =L k /2 where δ j,k = 1 if k = j, and δ j,k = 0, otherwise. It is easy to show that if Ω m is the union of spheres then its effective diffusion tensor vanishes: D m = 0. (26) In other words, restricted diffusion in a compact isolated region becomes negligible in the long time limit. We note that the expressions (18, 26) have been used in previous works for the Kärger model (for example, in (Fieremans et al. 2010)) Limiting cases If there is no exchange between the compartments, the macroscopic dmri signal under the Gaussian phase approximation is P S NOEX (b) = v m exp ( Dg m b ), (27) m=1 where the effective diffusivity in compartment m in the gradient direction u g g/ g 2 is D m g u g D m u g. In the opposite limit of complete exchange, the macroscopic dmri signal is ( ( P ) ) S COMP EX (b) = exp v m Dg m b. (28) The Kärger model and the FPK model are formulated to investigate the intermediate exchange regime that falls between the two extreme cases in Eqs. (27, 28). One important property of the FPK model (and its narrow pulse limit, the Kärger model) is that the apparent diffusion coefficient (ADC) extrapolated at b = 0, i.e., log S(b) ADC 0 = lim, b 0 b m=1

8 Parameters estimation using macroscopic diffusion MRI signal models 8 is independent of time and of the permeability κ: ADC0 KAR = ADC0 F P K COMP EX = ADC0 = ADC NOEX 0 = P v m Dg m. Thus, two necessary conditions for the applicability of the above macroscopic models are: (i) the diffusion time should be long enough that the measured ADC 0 does not change (much) with time, and (ii) the cell membrane permeability κ should be low enough that it does not affect (much) the ADC Quality of macroscopic models We constructed the computational box C = [ 5µm, 5µm] 3 shown in Fig. 1(a) containing 76 spherical cells with a range of radii between 0.6µm and 2.55µm. Formally, one could consider 76 different compartments for the 76 spheres, each with its volume fraction and surface-to-volume ratio. However, we just combine the 76 spheres to form one compartment Ω 2. The remaining extracellular space forms another compartment Ω 1. The corresponding volume fractions are v 2 = 0.65 and v 1 = The surface-to-volume ratio is Γ 21 / Ω 2 = 1.85µm 1. We consider that Γ 21 / Ω 2 is a good approximation of the average surface-to-volume ratios of all the 76 spheres. Note that this approximation would be exact if all the spheres had the same radius. Setting the intrinsic diffusion coefficient D 0 = mm 2 /s and the permeability κ = 10 5 m/s, one finds the true model parameters to be { } U true = v 2 = 0.65, Dg 1 = mm 2 /s, D 2 = 0, τ 12 = 54ms, (29) where the true value of D 1 g is computed numerically as earlier explained in Sec The value D 1 g = mm 2 /s is rather high compared to the values found in the literature for brain tissue due to the geometry chosen (spherical cells). We first check the quality of the macroscopic models as an approximation of the ground truth dmri signal obtained from the Bloch-Torrey equation. Figure 1(b) shows that the two-compartment FPK model gives a good approximation to the signal from the Bloch-Torrey PDE. For the PGSE sequence with δ = = 40ms, the narrow pulse assumption is not satisfied, and the original Kärger model fails to approximate the dmri signal. In turn, the FPK model provides a very accurate approximation over the typical range of b-values up to 4000s/mm Materials and Methods The parameters estimation problem consists in fitting experimentally measured dmri signal to an appropriate macroscopic model in order to extract the model parameters. In order to investigate the sensitivity and robustness of the parameters estimation problem, we replace experimental data by synthetically generated data to which we added Gaussian noise in the real and imaginery parts. This allows us to check the influence of various factors on the stability and the quality of the parameters estimation. We show results for two examples. In the first example, the synthetic dmri data is generated by solving the FPK model for the 76-spheres geometry using the true model parameters in Eq. (29) (see Sec. 2.6). In the second example, the synthetic dmri data is obtained by evaluating the bi-exponetial fit of published in vivo data in the rat brain cortex (Pyatigorskaya et al. 2014). m=1

9 Parameters estimation using macroscopic diffusion MRI signal models 9 (a) Finite elements mesh of C (b) DMRI signals Figure 1. (a) The computational box C = [ 5µm, 5µm] 3 contains 76 spheres with radii between 0.6µm and 2.55µm. The volume fraction of the spheres and of the extracellular space are v 2 = 0.65 and v 1 = 0.35, respectively. We set κ = 10 5 m/s, D 0 = mm 2 /s. (b) The dmri signals: S P DE (b), S F P K (b), S KAR (b), S NOEX (b), S COMP EX (b). The parameters of the PGSE sequence are δ = = 40ms Synthetic dmri signal First example For the 76-spheres example (see Sec. 2.6) we generated the dmri signal by using the FPK model with the true parameters in Eq. (29) and simulating two PGSE sequences with {δ = = 25ms; δ = = 40ms}. (30) Second example For our second example, we refer to the in vivo experimental signals of the rat cortex obtained for PGSE at δ = 2.5 ms and several diffusion times (Pyatigorskaya et al. 2014). Here we focus on two diffusion times: = 10, 30 ms. It was shown that the experimental data over b-values in the range [0, 4000] s/mm 2 can be well fitted using a bi-exponential function: where S = f s exp( bd s ) + (1 f s ) exp( bd f ), δ, ms, ms f s D f, mm 2 /s D s, mm 2 /s We use the above bi-exponential description to generate dmri signals and we simulate the same PGSE sequences as in the in vivo experiment: {δ = 2.5ms, = 10ms; δ = 2.5ms, = 30ms}. (31) For both examples, the dmri signals were computed at a set of 42 b-values: { 50k s/mm 2, k = , b k = (32) 100(k 3) s/mm 2, k =

10 Parameters estimation using macroscopic diffusion MRI signal models 10 Note the first few low b-values increase in intervals of 50 s/mm 2 (e.g., b 1 = 50 s/mm 2, b 2 = 100 s/mm 2, etc.) and the later b-values increase in intervals of 100 s/mm 2 (e.g., b 6 = 300 s/mm 2, b 7 = 400 s/mm 2, etc.). The set of b-values is within the range typically used in experimental settings for brain imaging. We denote by S(b k ) the simulated dmri signal at b k Noisy signal In order to quantify the effect of experimental noise in the parameters estimation problem, we add Gaussian noise to the real and imaginary components of the simulated signal S(b k ) to produce noisy signals: ( ) 2 ( S σ,i (b k ) = S(b k ) + σni,k Re + σn Im i,k ) 2, (33) where k = 1,, 42 (42 b-values), i = 1,, N (N random samples of noisy signal), σ is the level of noise, while Ni,k Re Im and Ni,k are independent Gaussian variables with zero mean and unit variance. In practice, we save all generated noises Ni,k Re Im and Ni,k for N = 1000 samples of noise and use them repeatly when comparing different noise levels and parameters estimation strategies. We will generate results for 5 different noise levels σ {0.0125, 0.025, 0.05, 0.1, 0.2}. Note that the imaginary part of the simulated signal S(b k ) is zero due to the rephasing of the nuclei. The above scheme of adding Gaussian noises to the real and imaginary parts of the signal results in the so-called Rice distributed noisy magnitude signal (see, for example, (Gudbjartsson & Patz 1995)) which is considered as a benchmark for dmri measurements. In other words, the noisy magnitude signal S σ,i (b k ) from Eq. (33) is a random variable from the Rice distribution (Rice 1944). The expected value and the variance of this variable are where S σ,i (b k ) = σf (S(b k )/σ), (34) var{s σ,i (b k )} = σ 2( 2 + [S(b k )/σ] 2 [F (S(b k )/σ)] 2), (35) F (x) = [ ] π/2 e x2 /4 (1 + x 2 /2)I 0 (x 2 /4) + (x 2 /2)I 1 (x 2 /4), (36) and I n (z) are the modified Bessel functions of the first kind. When σ S(b k ) (i.e., x 1), one gets F (x) x + 1 2x + 1 8x 3 + O(x 5 ) (37) so that the mean value S σ,i (b k ) approaches the noiseless signal S(b k ) as expected. One can check that this approximate relation is accurate within 5% even at x = 1 (i.e., when S(b) = σ), while its relative error is 0.3% at x = 2. In the opposite limit σ S(b k ) (i.e., x 1), F (x) π/2(1 + O(x 2 )) so that S σ,i (b k ) π/2 σ. In other words, when the noise level is too large as compared to the noiseless signal, one measures the noise level. This situation is not relevant in practice.

11 Parameters estimation using macroscopic diffusion MRI signal models 11 According to Eq. (35), the variance of the noisy signal monotonously grows from σ 2 (2 π/2) 0.43σ 2 at x = 0 to σ 2 at x =. One can conclude that the variance does not change significantly, even for very noisy signals. It is also convenient to define the signal-to-noise ratio as where SNR k = F (x) = 1 x S σ,i (b k ) var {Sσ,i (b k )} = S(b k) σ F (S(b k )/σ), (38) ( ) 2 + x 2 1/2 [F (x)] 2 1. (39) Using Eq. (37) for large x, the correction factor F (x) can be written as F (x) x x 4 + O(x 6 ). (40) For instance, F (5) In other words, when x is large, SNR k is close to S(b k )/σ, as expected. For the first example (76-spheres), Fig. 2(a) shows the bias of the signal at σ = 0.2. One can see that the presence of noise tends to increase the signal: mean values of noisy signal (shown by circles and by solid line) lie above the noiseless signal (shown by dashed line). While this bias is negligible for small σ (not shown), it cannot be ignored at noises as large as σ = 0.2. Figure 2(b) presents the signal-to-noise ratio from Eq. (38) for all σ and δ = = 40ms. The decrease of the dmri signal S(b) with b-values results in decreasing SNR at large b-value. At the highest considered noise level σ = 0.2, SNR remains around 5 for almost all b-values. (a) Signal bias (b) SNR Figure 2. First example (76-spheres geometry) with δ = = 40ms: (a) Comparison between S(b) (noiseless signal) and S σ,i (b) (mean noisy signal); (b) SNR. Signal S(b) obtained by solving the FPK model using the true model parameters for the 76-spheres geometry. Symbols show the empirical mean values of S σ,i (b) over 1000 realizations. For the second example (bi-exponential fit of rat cortex data), Fig. 3(a) shows the bias of the signal at σ = 0.2, δ = 2.5ms, = 10ms. As earlier, the presence of noise tends to increase the signal. Figure 3(b) presents the signal-to-noise ratio (SNR) as a function of b-value for all σ, at δ = 2.5ms, = 10ms.

12 Parameters estimation using macroscopic diffusion MRI signal models 12 (a) Signal bias (b) SNR Figure 3. Second example (adapted from in vivo data) for δ = 2.5ms, = 10ms: (a) Comparison between S(b) (noiseless signal) and S σ,i (b) (mean noisy signal); (b) SNR. Signal S(b) obtained using bi-exponential fit from in vivo rat brain cortex experimental data. Symbols show the empirical mean values of S σ,i (b) over 1000 realizations. To perform the parameters estimation on noisy data using a subset, K {1,, 42}, of b-values, we define the SNR as SNR 1, i.e., the signal-to-noise ratio at the smallest (nonzero) b-value that we use: b 1 = 50s/mm The least squares problem In order to estimate the FPK model parameters, we fit the noisy signal to the FPK model in a weighted least squares (WLS) sense. For the first example, the 4-parameters set to be estimated is U = { s 0, v 2, D 1 g, τ 12}, (41) where the residence time τ 12 is related to the permeability κ according to Eq. (18), and D 1 g is the extracellular effective diffusion coefficient in direction g = (1, 0, 0), s 0 is the extrapolated value of the dmri signal at b = 0. The following constraints are imposed: τ 12 v2 = τ 21 v 1, v 1 + v 2 = 1, D 2 = 0, because Ω 2 is the union of spheres. For the second example, we note that the effective diffusion coefficients of two compartments have to be found. This is different than the spheres example where D 2 was set to zero. To shorten the computational time, we set s 0 to its true value (s 0 = 1) and the 4-parameters set to be estimated for this example is U = { v 2, D 1 g, D 2 g, τ 12}. (42) For a given set U of macroscopic model parameters, we denote the FPK signal by S F P K (b, U). For each random sample of noisy signal S σ,i (b), we solve the following least squares problem: min U S σ,i (b k ) s 0 S F P K (b k, U) 2 w k, (43) k K

13 Parameters estimation using macroscopic diffusion MRI signal models 13 where K {1,, 42} is a subset of the full list of simulated b-values, and w k are the associated weights. The solution of Eq. (43) which is denoted as UF i P K, is random because it depends on a random realization of noisy signal. Repeating this procedure for N = 1000 samples, we compute the mean values (MV) of the estimated parameters and their standard deviations (SD). In what follows, we focus on two cases: identical weights w k = 1 and w k = 1/var{S σ,i (b k )} from Eq. (35) Numerical resolution In order to assess the quality and robustness of the parameters estimation with the FPK model, we perform the following analysis. (i) We solve the least squares problem in Eq. (43) by an exhaustive search in the parameter space. For the first example, the 4-dimensional parameter space is U = { s0, v 2, D 1 g, τ 12}. The limits of the search space were chosen to be [0.9, 1.1] U 3, where the limits U 3 for {v 2, D 1 g, τ 12 } are U 3 = {[0.4, 0.9]} { [1.0, 3.0] 10 3 mm 2 /s } {[1, 150] ms}. We created and saved a search table for U 3 that has the values of S F P K (b k, U), k = 1,, 42, with a uniform discretization in 51 points in [0.4, 0.9], 41 in [1.0, 3.0] 10 3 mm 2 /s, and 150 in [1, 150] ms. Previously, the idea of using search tables for parameter estimation in dmri has appeared in (Iima et al. 2014). The exhaustive search for s 0 was conducted in [0.9, 1.1] using 9 uniform discretization points. The range for s 0 is justified because the synthetic signal is already normalized to S(b = 0) = 1. In summary, the precision of our search method is mm 2 /s in Dg, in v 2, 1 ms in τ 12 and in s 0. The ODE system (21) in the FPK model is solved numerically in Matlab by the function ode45 with absolute and relative tolerances of For the second example, we conduct an exhaustive search in 4-dimensional parameter space for U = (v 2, Dg, 1 Dg, 2 τ 12 ). The limits of the search space was chosen to be: U 4 = {[0.25, 0.55]} { [0.5, 3.0] 10 3 mm 2 /s } { [0, 0.5] 10 3 mm 2 /s } {[1, 400] ms}, We created and saved a search table for U 4 that has the values of S F P K (b k, U), k = 1,, 42, with a uniform discretization in 31 points in [0.25, 0.55], 51 in [0.5, 3.0] 10 3 mm 2 /s for Dg 1 and 11 in [0.0, 0.5] 10 3 mm 2 /s for Dg, 2 and 400 in [1, 400] ms. The precision of the search table is the same as for the first example. Because δ is much smaller than, it is sufficient to use the Kärger model for the second example. (ii) We characterize the quality of the fit by the root of the mean squared error (RMSE) of the weighted LS fitting, normalized by the weights, k K RMSE i = S σ,i(b k ) s 0 S FPK (b k, U) 2 w k k K w. (44) k This quantity characterizes the accuracy of the LS fitting for one random realization of the noisy signal. In what follows, we will present the empirical mean values and the

14 Parameters estimation using macroscopic diffusion MRI signal models 14 standard deviations of the estimated parameters and of RMSE i after averaging over 1000 realizations, as functions of the noise level (plotted against the SNR from Eq. (38)). In addition, for the first example, because the true values of the FPK model parameters are known, the mean values of the estimated parameters are compared to their true values. Note that while the true values of v 2 and τ 12 are determined by the microstructure and the permeability, the true value of D 1 g is computed from Eq. (21) after solving numerically Eqs. (22 24) by the same finite elements method as for the Bloch-Torrey equation (Nguyen et al. 2014). 4. Results and discussion 4.1. First example We compare four parameters estimation strategies: Choice 1: sequence δ = = 40ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1; Choice 2: sequence δ = = 40ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1/var{S σ,i (b k )} from Eq. (35); Choice 3: Sequence δ = = 25ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1; Choice 4: sequence δ = = 40ms, b-values are taken from the first 21 of the 42 values in Eq. (32) and repeated once (for a total of 42 b-values, 21 of them distinct), LS weighting w k = 1. The b-values range from 50s/mm 2 to 1700s/mm 2. In this way, we compare the reference case (Choice 1) with other cases in order to reveal the respective roles of weighting by the reciprocal of the variance (Choice 2), using shorter diffusion time (Choice 3), and using a smaller range of b-values (Choice 4). First we show in Fig. 4 the normalized RMSE defined in Eq. (44). One can see that all four strategies provide similar RMSE, both in the mean values and the standard deviations. The mean values of the RMSE increased from 1% at SNR = 100 to 18% at SNR = 4.9. The standard deviations increase from 0.1% to 3%.

15 Parameters estimation using macroscopic diffusion MRI signal models 15 (a) MV RMSE (b) SD RMSE Figure 4. Comparison between four strategies described in Sec. 4.1 for the 76-spheres example: empirical mean values (MV, as a percentage) in (a) and standard deviations (SD, as a percentage) in (b) of RMSE as functions of the SNR at the lowest b-value. The SNR is defined in Eq. (38). In Fig. 5 we show the mean values and standard deviations (as percentages) of the parameters, D 1 g, v 2, τ 12, of the four strategies (the parameters are normalized by their true values for convenience). We see that (i) As the SNR decreases from 77 to 4.9, for Choices 1 and 2, the mean values of D 1 g decrease from 101% to 96% of the true values. For Choice 3, the mean values decrease from 102% to 98% of the true value. For Choice 4, the mean value decreases from 102% to 93% of the true values. For Choices 1 and 2, the standard deviations of D 1 g increase from 23% to 26% of the true values. For Choice 3, the standard deviations increase from 20% to 35% of the true values. For Choice 4, the standard deviations increase from 30% to 40% of the true values. (ii) As the SNR decreases from 77 to 9.7, for Choices 1 and 2, the mean values of v 2 decrease from 98% to 92% of the true values. For Choice 3, the mean values decrease from 100% to 95% of the true values. For Choice 4, the mean values decrease from 96% to 93% of the true values. For Choices 1 and 2, the standard deviations of v 2 increase from 10% to 21% of the true values. For Choice 3, the standard deviations increase from 8% to 19% of the true values. For Choice 4, the standard deviations increase from 15% to 18% of the true values. (iii) As the SNR decreases from 77 to 4.9, for Choices 1,2, and 3, the mean values of τ 12 increase from 110% to 185% of the true values. For Choice 4, the mean values increase from 140% to 155% of the true values. For Choices 1,2, and 3, the standard deviations of τ 12 increase from 30% to 100% of the true values. For Choice 4, the mean values increase from 70% to 120% of the true values.

16 Parameters estimation using macroscopic diffusion MRI signal models 16 (a) MV D 1 g (D 1 g ) true (b) SD D 1 g (D 1 g ) true (c) MV v2 v 2 true (d) SD v2 v 2 true (e) MV τ 12 τ 12 true (f) SD τ 12 τ 12 true Figure 5. Comparison between four strategies described in Sec. 4.1 to estimate {s 0, D 1 g, v 2, τ 12 } for the 76-spheres example. Results from 1000 samples of noisy signals for 5 noise levels. The empirical mean values and standard deviations of the estimated parameters: D 1 g, v 2, τ 12, normalized by their true values from Eq. (29), are shown as functions of the SNR (at the lowest b-value). The SNR is defined in Eq. (38).

17 Parameters estimation using macroscopic diffusion MRI signal models In vivo brain imaging In Fig. 6 we compare the following three estimation strategies: Choice 1: sequence δ = 2.5ms, = 30ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1; Choice 2: sequence δ = 2.5ms, = 30ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1/var k {S σ,i (b k )} from Eq. (35); Choice 3: sequence δ = 2.5ms, = 10ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1. Choice 4: sequence δ = 2.5ms, = 30ms, b-values are taken from the first 21 of the 42 values in Eq. (32) and repeated once (for a total of 42 b-values, 21 of them distinct), LS weighting w k = 1. The b-values range from 50s/mm 2 to 1700s/mm 2. Results from 1000 samples of noisy signals for 5 noise levels are shown. In Fig. 6 we show the mean values and standard deviations of the estimated parameters, Dg, 1 v 2, τ 12, Dg, 2 of the strategies described above. The mean values of the estimated parameters are not normalized by their true values as in the previous example because the true values are not known. The standard deviations of the estimated parameters are normalized by their mean values and expressed as a percentage. We see that (i) As the SNR decreases from 77 to 4.9, for Choices 1,2,3,4, the mean values of D 1 g increase from 1.0mm 2 /s to 1.2mm 2 /s. For Choices 1,2,3, the standard deviations of D 1 g increase from 15% to 40% of the mean value. For Choice 4, the standard deviations increase from 15% to 55% of the mean values. (ii) For Choices 1 and 2, the mean values of v 2 vary between 0.36 to For Choice 3, the mean values vary between 0.35 to For Choice 4, the mean values vary between 0.38 to The trend of the mean values is not monotonic with SNR. For Choices 1,2,3, the standard deviations of v 2 vary between 30% to 38% of the mean values. For Choice 4, the standard deviations vary between 26% to 34% of the mean values. (iii) As the SNR decreases from 77 to 4.9, for Choices 1,2, the mean values of τ 12 increase from 70ms to 225ms. For Choice 3, the mean values increase from 30ms to 160ms. For Choice 4, the mean value increases from 70ms to 160ms. Between SNR=9.7 and SNR = 77, for Choices 1,2, and 4, the standard deviations of τ 12 are around 120% of the mean values. Between SNR=9.7 and SNR = 77, for Choice 3, the standard deviations of τ 12 are around 180% of the mean values. There is a decrease in the normalized standard deviations at SNR = 4.9 because the mean values are very large. (iv) As the SNR decreases from 77 to 4.9, for Choices 1,2,3, the mean value of D 2 g decreases from 0.2mm 2 /s to 0.05mm 2 /s. For Choice 4, the mean values of D 2 g are around 2.5mm 2 /s. For Choices 1,2,3, the standard deviations of D 2 g increase from 60% to 145% of the mean values. For Choice 4, the standard deviations increase from 55% to 90% of the mean values.

18 Parameters estimation using macroscopic diffusion MRI signal models 18 (a) MV D 1 g(mm 2 /s) (b) SD D 1 g (%) (c) MV v 2 (d) SD v 2 (%) (e) MV τ 12 (ms) (f) SD τ 12 (%) (g) MV D 2 g(mm 2 /s) (h) SD D 2 g (%) Figure 6. Comparison between four strategies described in Sec. 4.2 to estimate {v 2, D 1 g, D 2 g, τ 12 } for the in vivo rat brain cortex experimental data. Results from 1000 samples of noisy signals for 5 noise levels. The empirical mean values and standard deviations of the estimated parameters (v 2, D 1 g, D 2 g, τ 12 ), are shown as functions of the SNR at the lowest b-value. The SNR is defined in Eq. (38). The mean values are not normalized by the true values of the parameters as in the previous example since the true values are not known. The standard deviations of the estimated parameters are normalized by their mean values and expressed in percentages.

19 Parameters estimation using macroscopic diffusion MRI signal models Bias-corrected signal When the noise level σ is significant, the mean values of the noisy signal lie above the noiseless signal, as illustrated in Figs. 2 and 3. As a consequence, fitting noisy signals S σ,i (b) by the (noiseless) model signal S F P K (b) may yield a strong bias in estimated parameters. This bias can be corrected by replacing the model signal S F P K (b) by the expected value S σ,i (b) = σf (S F P K (b)/σ) from Eq. (34). When the noise level σ is not known a priori, it can be considered as an additional unknown parameter in the LS problem, i.e., added to macroscopic model parameters U. We note that this simple correction method relies on the assumption of the Rician noise (from which Eq. (34) was derived). As discussed in (Gudbjartsson & Patz 1995, Sijbers et al. 1998, Veraart et al. 2011), this assumption is satisfied in most dmri measurements. We compare three choices (for simplicity, s 0 will be set to 1 instead of being a searched parameter in the LS method also for the 76-spheres example, just like in the rat cortex example): No NC : no noise correction, in which case one solves the LS problem in Eq. (43) as in earlier sections; NC 1 : noise correction with known noise level σ, in which case one solves the LS problem min S σ,i (b k ) σf (S F P K (b k, U)/σ) 2 w k. (45) U k K NC 2 : noise correction with unknown noise level σ, in which case the LS problem becomes (S σ,i (b k ) βf (S F P K (b k, U)/β) 2 w k, (46) min U,β k K where β is an additional fitting parameter (the best fit noise level). The parameter β adds a new dimension to the search space. We used the following discretization of β in the global search: β [0.5, 0.6,, 1.5] σ, where σ is the true noise level. In Fig. 7 we show the results for the 76-spheres example. We used the sequence δ = = 40ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1. We see at SNR = 9.7, the small bias in D 1 g (96% of true value) is corrected by NC 1 (98% of true value) and NC 2 (98% of true value), with essentially no change in estimated v 2 and τ 12. At SNR = 4.9, the small bias in D 1 g (95% of true value) is corrected by NC 1 (98% of true value) and NC 2 (99% of true value), with an improvement in τ 12, going from 190% of the true value to 160% for NC 1 and 135% for NC 2, and a worse estimated v 2 : going from 96% of the true value to 91% for NC 1 and 90% for NC 2. There is very little change in the standard deviations.

20 Parameters estimation using macroscopic diffusion MRI signal models 20 (a) MV D 1 g (D 1 g ) true (b) SD D 1 g (D 1 g ) true (c) MV v2 v 2 true (d) SD v2 v 2 true (e) MV τ 12 τ 12 true (f) SD τ 12 τ 12 true Figure spheres example. Noise correction using Rice distribution. The sequence is δ = = 40ms, LS problem uses the full list of 42 b-values in Eq. (32) and LS weighting w k = 1. The strategies are described in Sec The mean values and standard deviations of the estimated parameters, Dg, 1 v 2, τ 12, normalized by their true values, are shown. True parameters are given by Eq. (29). The SNR of the estimation strategy is defined as the SNR at the lowest b-value. The SNR is defined in Eq. (38). In Fig. 8 we show the results for the rat cortex bi-exponential fit example. We used the sequence δ = 2.5ms, = 30ms, the full list of 42 b-values in Eq. (32), LS weighting w k = 1. We see at smallest SNR shown (=4.25), the positive bias for D 1 g

21 Parameters estimation using macroscopic diffusion MRI signal models 21 and τ 12 are corrected by NC 1 and NC 2. For v 2, the no correction case has a negative bias. NC 1 corrected the negative bias in v 2, whereas NC 2 produced a positive bias of a similar magnitude compared to the no correction case. For D 2 g, the no correction case has a negative bias, and both NC 1 and NC 2 resulted in a positive bias of a similar magnitude compared to the no bias case. (a) MV D 1 g(mm 2 /s) (b) MV v 2 (c) MV τ 12 (d) MV D 2 g(mm 2 /s) Figure 8. Noise correction using Rice distribution. The strategies are described in Sec The sequence is δ = 2.5ms, = 30ms, the LS problem uses the full list of 42 b-values in Eq. (32) and LS weighting w k = 1. The mean values of the estimated parameters, D 1 g, v 2, τ 12, D 2 g, are shown. The SNR of the estimation strategy is defined as the SNR at the lowest b-value. The SNR is defined in Eq. (38). Signal using bi-exponential fit from in vivo rat brain cortex experimental data Discussion Since the current dmri spatial resolution does not allow one to directly image the complete tissue microstructure, inferring the macroscopic parameters of biological tissues from voxel-level dmri signals is standard practice. Clearly, the inference quality depends strongly on the chosen macroscopic model. As one moves from simpler to more complicated macroscopic models, careful studies of the quality and sensitivity of the estimated macroscopic parameters under realistic imaging conditions need to be conducted. Such studies have been conducted for the diffusion tensor model (see, for example, (Veraart, Sijbers, Sunaert, Leemans & Jeurissen 2013, Collier et al. 2014)), as well as the diffusion kurtosis model (see (Veraart et al. 2011, Veraart, Rajan, Peeters,

22 Parameters estimation using macroscopic diffusion MRI signal models 22 Leemans, Sunaert & Sijbers 2013)). By taking the logarithm of dmri signal data, both the diffusion tensor and the diffusion kurtosis models become linear in the macroscopic parameters. Linear parameters estimation is inherently easier to solve than its nonlinear counterpart and there exist well-developed theory and numerical tools to solve the linear parameters estimation problem. By contrast, the parameters estimation problem associated with some other common macroscopic models, such as the multiple exponential model, the cylindrical neurites model, or the Kärger model, cannot be made linear by a simple transformation of the dmri signal data. The parameters estimation problem associated with these models remain non-linear. Nevertheless, it is hoped that the additional difficulties in parameters estimation will be justified by other advantages of these macroscopic models, such as better sensitivities to physiological or pathological variations in tissue. It is in this spirit that we undertake the present study of the FPK model which was derived from the microscopic Bloch-Torrey equation by homogenization techniques. In fact, it is a promising choice for revealing the relation between the macroscopic parameters and the microstructure. The FPK model generalizes the Kärger model (Karger et al. 1988) to non-narrow pulses and the Kärger model has already been used to invert for model parameters from dmri signals (Waldeck et al. 1997, Stanisz et al. 1997, Pfeuffer et al. 1998, Lee & Springer 2003, Quirk et al. 2003, Meier et al. 2003, Roth et al. 2008, Aslund et al. 2009, Nilsson et al. 2009). We used two sets of data. The first set came from the solution of the Bloch-Torrey equation on a microstructure containing 76 spheres of variable radii. We checked that this solution can be accurately approximated by the FPK model. For numerical convenience, the synthetic signal generated by the FPK model was used. The second set of data came from the in vivo rat cortex data (Pyatigorskaya et al. 2014). These data were shown to be accurately approximated by the bi-exponential model. Once again, we used the simpler representation by the bi-exponential model just for numerical purposes. We showed that these data can as well be accurately approximated by the FPK model. The synthetic dmri signals were then corrupted by adding Gaussian noises to their real and imaginary parts. This numerical procedure models the way how the experimental signal gets noisy, and yields the Rician noise, as expected. Fitting the noisy signal by weighted least squares method to the FPK signal allows one to estimate the macroscopic parameters of this model. Our goal was to quantify how the quality of the parameters estimation depends on the noise level, sequence timing, weighting coefficients, and the range of b-values. In what follows, we discuss the main conclusions of this study. (i) The mean values of the two estimated parameters Dg 1 and v 2 remain close to their true values and weakly depend on the noise level (or SNR), see Fig. 5a,c. The PGSE sequence with shorter pulses ( = δ = 25ms instead of = δ = 40ms) yields a slightly better estimation. In turn, the third parameter τ 12 exhibits a strong bias from its true value (Fig. 5e), ranging from 10% at the smallest noise (SNR 80) to 85% at the highest noise (SNR 5). This observation suggests that the FPK signal weakly depends on τ 12 so that even small noise can significantly affect the inference quality. (ii) There is almost no difference between the standard LS method (with identical weights w k = 1) and the LS method weighted by the inverse of variances of the signal (w k = 1/var{S σ,i (b k )}). In general, the correction by variances assigns smaller weights to more noisy points and thus improves the overall quality of the LS fit. In the case

23 Parameters estimation using macroscopic diffusion MRI signal models 23 of Rician noise, the variance of the noisy signal weakly depends on b-values (Sec. 3.2) and thus has no impact on the estimation quality. We conclude that the standard LS method can be employed. Fitting the signal over a shorter range of b-values also has a weak effect. (iii) The standard deviations of all estimated parameters (or, equivalently, the relative errors as they are plotted in Fig. 5b,d,f), decrease with the SNR, as expected. However, their values remain significant even at the smallest noise level. For instance, the relative errors vary between 20% and 40% for Dg, 1 8% and 20% for v 2, and 20% and 120% for τ 12. As expected, the worst estimation occurs for the residence time τ 12, re-affirming our suggestion about the weak dependence of the FPK signal on this parameter. The PGSE with shorter diffusion time yields smaller standard deviations of Dg 1 and v 2, and essentially the same standard deviation of τ 12. We also note that the variations between mean values of the estimated parameters among different estimation strategies (Fig. 5a,c,e)) are below standard deviations. For this reason, we conclude that all considered estimation strategies are of comparable quality, though the shorter diffusion time can yield a slight improvement. (iv) When the noise level is high, the mean noisy signal deviates significantly from the (noiseless) FPK signal S(b) that was used for the above parameters estimation (Fig. 2). In other words, fitting noisy signals by S(b) could bias the estimated parameters. In order to correct for this bias, we replaced the noiseless signal S(b) from the FPK model by the mean noisy signal S σ,i (b). Under the assumption of Rician noise, one can write S σ,i (b) = σf (S(b)/σ), with the explicit formula (34) for the function F (x). We compared three estimation strategies: use of noiseless signal S(b), use of biascorrected signal σf (S(b)/σ) with known noise level σ, and use of bias-corrected signal σf (S(b)/σ) with unknown noise level σ which is added to fitting parameters. We showed that the use of the bias-corrected signal improves the estimation of the worst parameter τ 12 at the highest noise level (Fig. 7c). Another advantage of this scheme is a possibility to estimate the noise level directly from the noisy signal. However, the standard deviations of all three parameters remain almost unchanged (Fig. 7b,d,f). We conclude that the use of bias-corrected signal may be advantageous but it does not allow to reduce standard deviations (i.e., uncertainty range of estimated parameters). (v) Similar analysis was performed on the second set of data coming from the in vivo rat cortex. Since the true values of the macroscopic parameters of the FPK model are not known in this case, it is more difficult to assess the quality of parameters estimation. Our goal here was just to illustrate a potiential application of the parameters estimation on experimental data (Fig. 6). The mean values of the estimated Dg 1 are similar for all 4 strategies, its standard deviations are higher when using smaller b-values (Fig. 6a,b) than the other three strategies. The quality of the estimated v 2 is similar among the 4 strategies (Fig. 6c,d). The mean values of the estimated τ 12 are different and the standard deviations are higher when using the shorter diffusion time compared to the other 3 strategies (Fig. 6e,f), leading us to conclude that the longer diffusion time should be used. The quality of the estimated Dg 2 is the best when using the smaller b-values (Fig. 6g,h). As earlier, the worst estimated parameter is the transfer time τ 12, for which the standard deviation is comparable to the mean value. The next worst is Dg. 2 The other parameters, Dg 1 and v 2, can be estimated more accurately. Although the quality of the parameters estimation remains modest, the FPK model can be considered as a promising tool for inferring the macroscopic parameters of a biological tissue. The worst estimated parameter is the transfer time between the

Outline of the talk How to describe restricted diffusion? How to monitor restricted diffusion? Laplacian eigenfunctions in NMR Other applications Loca

Outline of the talk How to describe restricted diffusion? How to monitor restricted diffusion? Laplacian eigenfunctions in NMR Other applications Loca Laplacian Eigenfunctions in NMR Denis S. Grebenkov Laboratoire de Physique de la Matière Condensée CNRS Ecole Polytechnique, Palaiseau, France IPAM Workshop «Laplacian Eigenvalues and Eigenfunctions» February

More information

Numerical simulation and macroscopic model formulation for diffusion magnetic resonance imaging in the brain

Numerical simulation and macroscopic model formulation for diffusion magnetic resonance imaging in the brain Numerical simulation and macroscopic model formulation for diffusion magnetic resonance imaging in the brain Jing-Rebecca Li Equipe DEFI, CMAP, Ecole Polytechnique Institut national de recherche en informatique

More information

Numerical simulation and macroscopic model formulation for diffusion magnetic resonance imaging in the brain

Numerical simulation and macroscopic model formulation for diffusion magnetic resonance imaging in the brain Numerical simulation and macroscopic model formulation for diffusion magnetic resonance imaging in the brain Jing-Rebecca Li Equipe DEFI, CMAP, Ecole Polytechnique Institut national de recherche en informatique

More information

Understanding brain micro-structure using diffusion magnetic resonance imaging (dmri)

Understanding brain micro-structure using diffusion magnetic resonance imaging (dmri) Understanding brain micro-structure using diffusion magnetic resonance imaging (dmri) Jing-Rebecca Li Equipe DEFI, CMAP, Ecole Polytechnique Institut national de recherche en informatique et en automatique

More information

Ordinary Least Squares and its applications

Ordinary Least Squares and its applications Ordinary Least Squares and its applications Dr. Mauro Zucchelli University Of Verona December 5, 2016 Dr. Mauro Zucchelli Ordinary Least Squares and its applications December 5, 2016 1 / 48 Contents 1

More information

Analytical and numerical study of the apparent diffusion coefficient in diffusion MRI at long diffusion times and low b- values

Analytical and numerical study of the apparent diffusion coefficient in diffusion MRI at long diffusion times and low b- values Analytical and numerical study of the apparent diffusion coefficient in diffusion MRI at long diffusion times and low b- values J.R. Li 1,2*, D. Le Bihan 2, T.Q. Nguyen 1, D. Grebenkov 3, C. Poupon 2,

More information

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics Magnetic Resonance Imaging Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics pal.e.goa@ntnu.no 1 Why MRI? X-ray/CT: Great for bone structures and high spatial resolution Not so great

More information

An Analytical Model of Water Diffusion and Exchange in White Matter from Diffusion MRI and Its Application in Measuring Axon Radii

An Analytical Model of Water Diffusion and Exchange in White Matter from Diffusion MRI and Its Application in Measuring Axon Radii An Analytical Model of Water Diffusion and Exchange in White Matter from Diffusion MRI and Its Application in Measuring Axon Radii Wenjin Zhou, Student Member, IEEE, and David H. Laidlaw, Senior Member,

More information

Downloaded 02/05/15 to Redistribution subject to SIAM license or copyright; see

Downloaded 02/05/15 to Redistribution subject to SIAM license or copyright; see SIAM J. APPL. MATH. Vol. 74, No., pp. 516 546 c 14 Society for Industrial and Applied Mathematics Downloaded /5/15 to 18.93.16.74. Redistribution subject to SIAM license or copyright; see http://www.siam.org/journals/ojsa.php

More information

Advanced Topics and Diffusion MRI

Advanced Topics and Diffusion MRI Advanced Topics and Diffusion MRI Slides originally by Karla Miller, FMRIB Centre Modified by Mark Chiew (mark.chiew@ndcn.ox.ac.uk) Slides available at: http://users.fmrib.ox.ac.uk/~mchiew/teaching/ MRI

More information

From the microstructure to diffusion MRI, and back

From the microstructure to diffusion MRI, and back From the microstructure to diffusion MRI, and back Denis S. Grebenkov May 5, 2015 Contents 1 Introduction 2 2 Mathematical background 3 2.1 Bloch-Torrey equation...................... 3 2.2 Boundary conditions.......................

More information

Principles of Nuclear Magnetic Resonance Microscopy

Principles of Nuclear Magnetic Resonance Microscopy Principles of Nuclear Magnetic Resonance Microscopy Paul T. Callaghan Department of Physics and Biophysics Massey University New Zealand CLARENDON PRESS OXFORD CONTENTS 1 PRINCIPLES OF IMAGING 1 1.1 Introduction

More information

Physics of MR Image Acquisition

Physics of MR Image Acquisition Physics of MR Image Acquisition HST-583, Fall 2002 Review: -MRI: Overview - MRI: Spatial Encoding MRI Contrast: Basic sequences - Gradient Echo - Spin Echo - Inversion Recovery : Functional Magnetic Resonance

More information

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution.

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution. Introduction to MRI Spin & Magnetic Moments Relaxation (T1, T2) Spin Echoes 2DFT Imaging Selective excitation, phase & frequency encoding K-space & Spatial Resolution Contrast (T1, T2) Acknowledgement:

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Thermodynamics of nuclei in thermal contact

Thermodynamics of nuclei in thermal contact Thermodynamics of nuclei in thermal contact Karl-Heinz Schmidt, Beatriz Jurado CENBG, CNRS/IN2P3, Chemin du Solarium B.P. 120, 33175 Gradignan, France Abstract: The behaviour of a di-nuclear system in

More information

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam (closed book, 1 sheets of notes double sided allowed, no calculator or other electronic devices allowed) 1. Ultrasound Physics (15 pt) A) (9

More information

Active Imaging with Dual Spin-Echo Diffusion MRI

Active Imaging with Dual Spin-Echo Diffusion MRI Active Imaging with Dual Spin-Echo Diffusion MRI Jonathan D. Clayden 1, Zoltan Nagy 2, Matt G. Hall 3,4, Chris A. Clark 1, and Daniel C. Alexander 3,4 1 Institute of Child Health 2 Wellcome Trust Centre

More information

DIFFUSION MAGNETIC RESONANCE IMAGING

DIFFUSION MAGNETIC RESONANCE IMAGING DIFFUSION MAGNETIC RESONANCE IMAGING from spectroscopy to imaging apparent diffusion coefficient ADC-Map anisotropy diffusion tensor (imaging) DIFFUSION NMR - FROM SPECTROSCOPY TO IMAGING Combining Diffusion

More information

NIH Public Access Author Manuscript Med Image Comput Comput Assist Interv. Author manuscript; available in PMC 2014 May 19.

NIH Public Access Author Manuscript Med Image Comput Comput Assist Interv. Author manuscript; available in PMC 2014 May 19. NIH Public Access Author Manuscript Published in final edited form as: Med Image Comput Comput Assist Interv. 2009 ; 12(0 1): 919 926. Bias of Least Squares Approaches for Diffusion Tensor Estimation from

More information

Magnetization Gradients, k-space and Molecular Diffusion. Magnetic field gradients, magnetization gratings and k-space

Magnetization Gradients, k-space and Molecular Diffusion. Magnetic field gradients, magnetization gratings and k-space 2256 Magnetization Gradients k-space and Molecular Diffusion Magnetic field gradients magnetization gratings and k-space In order to record an image of a sample (or obtain other spatial information) there

More information

Diffusion Magnetic Resonance Imaging Part 1: Theory & Methods

Diffusion Magnetic Resonance Imaging Part 1: Theory & Methods Diffusion Magnetic Resonance Imaging Part 1: Theory & Methods Benjamin M. Ellingson, Ph.D. Assistant Professor of Radiology, Biomedical Physics and Bioengineering Dept. of Radiological Sciences UCLA Neuro-Oncology

More information

Nuclear Magnetic Resonance Imaging

Nuclear Magnetic Resonance Imaging Nuclear Magnetic Resonance Imaging Simon Lacoste-Julien Electromagnetic Theory Project 198-562B Department of Physics McGill University April 21 2003 Abstract This paper gives an elementary introduction

More information

Diffusion-Weighted MRI may be used to measure the apparent diffusion coefficient of water in tissue.

Diffusion-Weighted MRI may be used to measure the apparent diffusion coefficient of water in tissue. Specialty Area: MR Physics for Physicists Speaker: Jennifer A. McNab, Ph.D. Assistant Professor, Radiology, Stanford University () Highlights The Bloch-Torrey equation is a generalization of the Bloch

More information

DWI acquisition schemes and Diffusion Tensor estimation

DWI acquisition schemes and Diffusion Tensor estimation DWI acquisition schemes and Diffusion Tensor estimation A simulation based study Santiago Aja-Fernández, Antonio Tristán-Vega, Pablo Casaseca-de-la-Higuera Laboratory of Image Processing L A B O R A T

More information

Suppression of Static Magnetic Field in Diffusion Measurements of Heterogeneous Materials

Suppression of Static Magnetic Field in Diffusion Measurements of Heterogeneous Materials PIERS ONLINE, VOL. 5, NO. 1, 2009 81 Suppression of Static Magnetic Field in Diffusion Measurements of Heterogeneous Materials Eva Gescheidtova 1 and Karel Bartusek 2 1 Faculty of Electrical Engineering

More information

EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 9 Solutions. Due April 29th, 2013

EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 9 Solutions. Due April 29th, 2013 EE5E/BIOE65 Spring 013 Principles of MRI Miki Lustig This is the last homework in class. Enjoy it. Assignment 9 Solutions Due April 9th, 013 1) In class when we presented the spin-echo saturation recovery

More information

Investigation of Multicomponent MRI Relaxation Data with Stochastic Contraction Fitting Algorithm

Investigation of Multicomponent MRI Relaxation Data with Stochastic Contraction Fitting Algorithm Investigation of Multicomponent MRI Relaxation Data with Stochastic Contraction Fitting Algorithm Thesis Presented in Partial Fulfillment of the Requirements for the Degree of Master in Mathematical Science

More information

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY Spatial encoding in Magnetic Resonance Imaging Jean-Marie BONNY What s Qu est an image ce qu une? image? «a reproduction of a material object by a camera or a related technique» Multi-dimensional signal

More information

Duration-Based Volatility Estimation

Duration-Based Volatility Estimation A Dual Approach to RV Torben G. Andersen, Northwestern University Dobrislav Dobrev, Federal Reserve Board of Governors Ernst Schaumburg, Northwestern Univeristy CHICAGO-ARGONNE INSTITUTE ON COMPUTATIONAL

More information

The NMR Inverse Imaging Problem

The NMR Inverse Imaging Problem The NMR Inverse Imaging Problem Nuclear Magnetic Resonance Protons and Neutrons have intrinsic angular momentum Atoms with an odd number of proton and/or odd number of neutrons have a net magnetic moment=>

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Spin Echo Imaging Hahn Spin Echo

More information

Characterizing Non-Gaussian Diffusion by Using Generalized Diffusion Tensors

Characterizing Non-Gaussian Diffusion by Using Generalized Diffusion Tensors Magnetic Resonance in Medicine 51:924 937 (2004) Characterizing Non-Gaussian Diffusion by Using Generalized Diffusion Tensors Chunlei Liu, 1,2 Roland Bammer, 1 Burak Acar, 3 and Michael E. Moseley 1 *

More information

Basic MRI physics and Functional MRI

Basic MRI physics and Functional MRI Basic MRI physics and Functional MRI Gregory R. Lee, Ph.D Assistant Professor, Department of Radiology June 24, 2013 Pediatric Neuroimaging Research Consortium Objectives Neuroimaging Overview MR Physics

More information

MR Fundamentals. 26 October Mitglied der Helmholtz-Gemeinschaft

MR Fundamentals. 26 October Mitglied der Helmholtz-Gemeinschaft MR Fundamentals 26 October 2010 Mitglied der Helmholtz-Gemeinschaft Mitglied der Helmholtz-Gemeinschaft Nuclear Spin Nuclear Spin Nuclear magnetic resonance is observed in atoms with odd number of protons

More information

Diffusion Tensor Imaging (DTI): An overview of key concepts

Diffusion Tensor Imaging (DTI): An overview of key concepts Diffusion Tensor Imaging (DTI): An overview of key concepts (Supplemental material for presentation) Prepared by: Nadia Barakat BMB 601 Chris Conklin Thursday, April 8 th 2010 Diffusion Concept [1,2]:

More information

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY Spatial encoding in Magnetic Resonance Imaging Jean-Marie BONNY What s Qu est an image ce qu une? image? «a reproduction of a material object by a camera or a related technique» Multi-dimensional signal

More information

NMR Imaging in porous media

NMR Imaging in porous media NMR Imaging in porous media What does NMR give us. Chemical structure. Molecular structure. Interactions between atoms and molecules. Incoherent dynamics (fluctuation, rotation, diffusion). Coherent flow

More information

The Basics of Magnetic Resonance Imaging

The Basics of Magnetic Resonance Imaging The Basics of Magnetic Resonance Imaging Nathalie JUST, PhD nathalie.just@epfl.ch CIBM-AIT, EPFL Course 2013-2014-Chemistry 1 Course 2013-2014-Chemistry 2 MRI: Many different contrasts Proton density T1

More information

D-eigenvalues of diffusion kurtosis tensors

D-eigenvalues of diffusion kurtosis tensors Journal of Computational and Applied Mathematics 221 (2008) 150 157 www.elsevier.com/locate/cam D-eigenvalues of diffusion kurtosis tensors Liqun Qi a,, Yiju Wang b, Ed X. Wu c a Department of Applied

More information

Electromagnetic fields and waves

Electromagnetic fields and waves Electromagnetic fields and waves Maxwell s rainbow Outline Maxwell s equations Plane waves Pulses and group velocity Polarization of light Transmission and reflection at an interface Macroscopic Maxwell

More information

PhD THESIS. prepared at INRIA Sophia Antipolis

PhD THESIS. prepared at INRIA Sophia Antipolis PhD THESIS prepared at INRIA Sophia Antipolis and presented at the University of Nice-Sophia Antipolis Graduate School of Information and Communication Sciences A dissertation submitted in partial satisfaction

More information

If we want to analyze experimental or simulated data we might encounter the following tasks:

If we want to analyze experimental or simulated data we might encounter the following tasks: Chapter 1 Introduction If we want to analyze experimental or simulated data we might encounter the following tasks: Characterization of the source of the signal and diagnosis Studying dependencies Prediction

More information

Fast and Accurate HARDI and its Application to Neurological Diagnosis

Fast and Accurate HARDI and its Application to Neurological Diagnosis Fast and Accurate HARDI and its Application to Neurological Diagnosis Dr. Oleg Michailovich Department of Electrical and Computer Engineering University of Waterloo June 21, 2011 Outline 1 Diffusion imaging

More information

Applications of Spin Echo and Gradient Echo: Diffusion and Susceptibility Contrast

Applications of Spin Echo and Gradient Echo: Diffusion and Susceptibility Contrast Applications of Spin Echo and Gradient Echo: Diffusion and Susceptibility Contrast Chunlei Liu, PhD Department of Electrical Engineering & Computer Sciences and Helen Wills Neuroscience Institute University

More information

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar. 1 Doppler echocardiography & Magnetic Resonance Imaging History: - Langevin developed sonar. - 1940s development of pulse-echo. - 1950s development of mode A and B. - 1957 development of continuous wave

More information

Generalized likelihood ratio tests for complex fmri data

Generalized likelihood ratio tests for complex fmri data Generalized likelihood ratio tests for complex fmri data J. Sijbers a and A. J. den Dekker b a Vision Lab, University of Antwerp, Groenenborgerlaan 171, B-2020 Antwerpen, Belgium; b Delft Center for Systems

More information

The Measure of Diffusion Skewness and Kurtosis in Magnetic Resonance Imaging

The Measure of Diffusion Skewness and Kurtosis in Magnetic Resonance Imaging The Measure of Diffusion Skewness and Kurtosis in Magnetic Resonance Imaging Xinzhen Zhang, Chen Ling, Liqun Qi, and Ed Xuekui Wu April 19, 008, Revised on September 4, 008 This paper is dedicated to the

More information

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging G16.4426 Medical Imaging Physics of Magnetic Resonance Imaging Riccardo Lattanzi, Ph.D. Assistant Professor Department of Radiology, NYU School of Medicine Department of Electrical and Computer Engineering,

More information

Contrast Mechanisms in MRI. Michael Jay Schillaci

Contrast Mechanisms in MRI. Michael Jay Schillaci Contrast Mechanisms in MRI Michael Jay Schillaci Overview Image Acquisition Basic Pulse Sequences Unwrapping K-Space Image Optimization Contrast Mechanisms Static and Motion Contrasts T1 & T2 Weighting,

More information

Complex Analysis MATH 6300 Fall 2013 Homework 4

Complex Analysis MATH 6300 Fall 2013 Homework 4 Complex Analysis MATH 6300 Fall 2013 Homework 4 Due Wednesday, December 11 at 5 PM Note that to get full credit on any problem in this class, you must solve the problems in an efficient and elegant manner,

More information

Modern Methods of Data Analysis - WS 07/08

Modern Methods of Data Analysis - WS 07/08 Modern Methods of Data Analysis Lecture VIc (19.11.07) Contents: Maximum Likelihood Fit Maximum Likelihood (I) Assume N measurements of a random variable Assume them to be independent and distributed according

More information

III, Diffusion, and Susceptibility. August 25, Departments of Mathematics and Applied Math and Computational Science University of Pennsylvania

III, Diffusion, and Susceptibility. August 25, Departments of Mathematics and Applied Math and Computational Science University of Pennsylvania III,, and Departments of Mathematics and Applied Math and Computational Science University of Pennsylvania August 25, 2010 Copyright Page All material in this lecture, except as noted within the text,

More information

A Riemannian Framework for Denoising Diffusion Tensor Images

A Riemannian Framework for Denoising Diffusion Tensor Images A Riemannian Framework for Denoising Diffusion Tensor Images Manasi Datar No Institute Given Abstract. Diffusion Tensor Imaging (DTI) is a relatively new imaging modality that has been extensively used

More information

Lightning Phenomenology Notes Note 23 8 Jan Lightning Responses on a Finite Cylindrical Enclosure

Lightning Phenomenology Notes Note 23 8 Jan Lightning Responses on a Finite Cylindrical Enclosure Lightning Phenomenology Notes Note 23 8 Jan 2014 Lightning Responses on a Finite Cylindrical Enclosure Kenneth C. Chen and Larry K. Warne Sandia National Laboratories, P. O. Box 5800, Albuquerque, NM 87185,

More information

arxiv: v1 [physics.chem-ph] 6 Oct 2011

arxiv: v1 [physics.chem-ph] 6 Oct 2011 Calculation of the Minimum Ignition Energy based on the ignition delay time arxiv:1110.1163v1 [physics.chem-ph] 6 Oct 2011 Jens Tarjei Jensen a, Nils Erland L. Haugen b, Natalia Babkovskaia c a Department

More information

Precise Inference and Characterization of Structural Organization (PICASO) of tissue from molecular diffusion

Precise Inference and Characterization of Structural Organization (PICASO) of tissue from molecular diffusion Precise Inference and Characterization of Structural Organization (PICASO) of tissue from molecular diffusion Lipeng Ning, Evren Özarslan, Carl-Fredrik Westin and Yogesh Rathi Journal Article Original

More information

On Signal to Noise Ratio Tradeoffs in fmri

On Signal to Noise Ratio Tradeoffs in fmri On Signal to Noise Ratio Tradeoffs in fmri G. H. Glover April 11, 1999 This monograph addresses the question of signal to noise ratio (SNR) in fmri scanning, when parameters are changed under conditions

More information

RAD229: Midterm Exam 2015/2016 October 19, Minutes. Please do not proceed to the next page until the exam begins.

RAD229: Midterm Exam 2015/2016 October 19, Minutes. Please do not proceed to the next page until the exam begins. RAD229: Midterm Exam 2015/2016 October 19, 2015 ---- 75 Minutes Name: Student ID: General Instructions: 1. Write your name legibly on this page. 2. You may use notes including lectures, homework, solutions

More information

Electromagnetic Waves

Electromagnetic Waves Physics 8 Electromagnetic Waves Overview. The most remarkable conclusion of Maxwell s work on electromagnetism in the 860 s was that waves could exist in the fields themselves, traveling with the speed

More information

The dynamics of small particles whose size is roughly 1 µmt or. smaller, in a fluid at room temperature, is extremely erratic, and is

The dynamics of small particles whose size is roughly 1 µmt or. smaller, in a fluid at room temperature, is extremely erratic, and is 1 I. BROWNIAN MOTION The dynamics of small particles whose size is roughly 1 µmt or smaller, in a fluid at room temperature, is extremely erratic, and is called Brownian motion. The velocity of such particles

More information

Basis of MRI Contrast

Basis of MRI Contrast Basis of MRI Contrast MARK A. HORSFIELD Department of Cardiovascular Sciences University of Leicester Leicester LE1 5WW UK Tel: +44-116-2585080 Fax: +44-870-7053111 e-mail: mah5@le.ac.uk 1 1.1 The Magnetic

More information

Laplacian transport towards irregular interfaces: the mathematics

Laplacian transport towards irregular interfaces: the mathematics Laplacian transport towards irregular interfaces: the mathematics Denis S GREBENKOV Laboratoire de Physique de la Matière Condensée CNRS Ecole Polytechnique, France Email: denisgrebenkov@polytechniqueedu

More information

Quantification of water compartmentation in cell suspensions by diffusion-weighted and T 2 -weighted MRI

Quantification of water compartmentation in cell suspensions by diffusion-weighted and T 2 -weighted MRI Available online at www.sciencedirect.com Magnetic Resonance Imaging 26 (2008) 88 102 Quantification of water compartmentation in cell suspensions by diffusion-weighted and T 2 -weighted MRI Yiftach Roth

More information

arxiv:hep-ex/ v1 2 Jun 2000

arxiv:hep-ex/ v1 2 Jun 2000 MPI H - V7-000 May 3, 000 Averaging Measurements with Hidden Correlations and Asymmetric Errors Michael Schmelling / MPI for Nuclear Physics Postfach 03980, D-6909 Heidelberg arxiv:hep-ex/0006004v Jun

More information

Waves in plasma. Denis Gialis

Waves in plasma. Denis Gialis Waves in plasma Denis Gialis This is a short introduction on waves in a non-relativistic plasma. We will consider a plasma of electrons and protons which is fully ionized, nonrelativistic and homogeneous.

More information

Chemistry 431. Lecture 23

Chemistry 431. Lecture 23 Chemistry 431 Lecture 23 Introduction The Larmor Frequency The Bloch Equations Measuring T 1 : Inversion Recovery Measuring T 2 : the Spin Echo NC State University NMR spectroscopy The Nuclear Magnetic

More information

Laplace Transforms Chapter 3

Laplace Transforms Chapter 3 Laplace Transforms Important analytical method for solving linear ordinary differential equations. - Application to nonlinear ODEs? Must linearize first. Laplace transforms play a key role in important

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu MRI advantages Superior soft-tissue contrast Depends on among

More information

On the estimation of the K parameter for the Rice fading distribution

On the estimation of the K parameter for the Rice fading distribution On the estimation of the K parameter for the Rice fading distribution Ali Abdi, Student Member, IEEE, Cihan Tepedelenlioglu, Student Member, IEEE, Mostafa Kaveh, Fellow, IEEE, and Georgios Giannakis, Fellow,

More information

Rich Tomography. Bill Lionheart, School of Mathematics, University of Manchester and DTU Compute. July 2014

Rich Tomography. Bill Lionheart, School of Mathematics, University of Manchester and DTU Compute. July 2014 Rich Tomography Bill Lionheart, School of Mathematics, University of Manchester and DTU Compute July 2014 What do we mean by Rich Tomography? Conventional tomography reconstructs one scalar image from

More information

Bayesian multi-tensor diffusion MRI and tractography

Bayesian multi-tensor diffusion MRI and tractography Bayesian multi-tensor diffusion MRI and tractography Diwei Zhou 1, Ian L. Dryden 1, Alexey Koloydenko 1, & Li Bai 2 1 School of Mathematical Sciences, Univ. of Nottingham 2 School of Computer Science and

More information

HT HT HOMOGENIZATION OF A CONDUCTIVE AND RADIATIVE HEAT TRANSFER PROBLEM, SIMULATION WITH CAST3M

HT HT HOMOGENIZATION OF A CONDUCTIVE AND RADIATIVE HEAT TRANSFER PROBLEM, SIMULATION WITH CAST3M Proceedings of HT2005 2005 ASME Summer Heat Transfer Conference July 17-22, 2005, San Francisco, California, USA HT2005-72193 HT2005-72193 HOMOGENIZATION OF A CONDUCTIVE AND RADIATIVE HEAT TRANSFER PROBLEM,

More information

Lecture #7 In Vivo Water

Lecture #7 In Vivo Water Lecture #7 In Vivo Water Topics Hydration layers Tissue relaxation times Magic angle effects Magnetization Transfer Contrast (MTC) CEST Handouts and Reading assignments Mathur-De Vre, R., The NMR studies

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 10 pages Written test, 12 December 2012 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: None. Pocket calculator not allowed "Weighting":

More information

1 Lyapunov theory of stability

1 Lyapunov theory of stability M.Kawski, APM 581 Diff Equns Intro to Lyapunov theory. November 15, 29 1 1 Lyapunov theory of stability Introduction. Lyapunov s second (or direct) method provides tools for studying (asymptotic) stability

More information

Spotlight on Laplace s Equation

Spotlight on Laplace s Equation 16 Spotlight on Laplace s Equation Reference: Sections 1.1,1.2, and 1.5. Laplace s equation is the undriven, linear, second-order PDE 2 u = (1) We defined diffusivity on page 587. where 2 is the Laplacian

More information

Towards Quantitative Measurements of Tissue Microstructure using Temporal Diffusion Spectroscopy. Hua Li

Towards Quantitative Measurements of Tissue Microstructure using Temporal Diffusion Spectroscopy. Hua Li Towards Quantitative Measurements of Tissue Microstructure using Temporal Diffusion Spectroscopy By Hua Li Dissertation Submitted to the Faculty of the Graduate School of Vanderbilt University in partial

More information

IEICE TRANS. INF. & SYST., VOL.E96 D, NO.6 JUNE

IEICE TRANS. INF. & SYST., VOL.E96 D, NO.6 JUNE 1387 PAPER An Explanation of Signal Changes in DW-fMRI: Monte Carlo Simulation Study of Restricted Diffusion of Water Molecules Using 3D and Two-Compartment Cortical Cell Models Shizue NAGAHARA, a), Student

More information

Report submitted to Prof. P. Shipman for Math 540, Fall 2009

Report submitted to Prof. P. Shipman for Math 540, Fall 2009 Dynamics at the Horsetooth Volume 1, 009. Three-Wave Interactions of Spin Waves Aaron Hagerstrom Department of Physics Colorado State University aaronhag@rams.colostate.edu Report submitted to Prof. P.

More information

Numerical Methods for ODEs. Lectures for PSU Summer Programs Xiantao Li

Numerical Methods for ODEs. Lectures for PSU Summer Programs Xiantao Li Numerical Methods for ODEs Lectures for PSU Summer Programs Xiantao Li Outline Introduction Some Challenges Numerical methods for ODEs Stiff ODEs Accuracy Constrained dynamics Stability Coarse-graining

More information

We have seen that the total magnetic moment or magnetization, M, of a sample of nuclear spins is the sum of the nuclear moments and is given by:

We have seen that the total magnetic moment or magnetization, M, of a sample of nuclear spins is the sum of the nuclear moments and is given by: Bloch Equations We have seen that the total magnetic moment or magnetization, M, of a sample of nuclear spins is the sum of the nuclear moments and is given by: M = [] µ i i In terms of the total spin

More information

1 Fundamentals. 1.1 Overview. 1.2 Units: Physics 704 Spring 2018

1 Fundamentals. 1.1 Overview. 1.2 Units: Physics 704 Spring 2018 Physics 704 Spring 2018 1 Fundamentals 1.1 Overview The objective of this course is: to determine and fields in various physical systems and the forces and/or torques resulting from them. The domain of

More information

Symmetry of the Dielectric Tensor

Symmetry of the Dielectric Tensor Symmetry of the Dielectric Tensor Curtis R. Menyuk June 11, 2010 In this note, I derive the symmetry of the dielectric tensor in two ways. The derivations are taken from Landau and Lifshitz s Statistical

More information

Slow symmetric exchange

Slow symmetric exchange Slow symmetric exchange ϕ A k k B t A B There are three things you should notice compared with the Figure on the previous slide: 1) The lines are broader, 2) the intensities are reduced and 3) the peaks

More information

NIH Public Access Author Manuscript J Magn Reson. Author manuscript; available in PMC 2012 June 1.

NIH Public Access Author Manuscript J Magn Reson. Author manuscript; available in PMC 2012 June 1. NIH Public Access Author Manuscript Published in final edited form as: J Magn Reson. 2011 June ; 210(2): 233 237. doi:10.1016/j.jmr.2011.03.012. Effect of gradient pulse duration on MRI estimation of the

More information

Lecture 3 Fiber Optical Communication Lecture 3, Slide 1

Lecture 3 Fiber Optical Communication Lecture 3, Slide 1 Lecture 3 Optical fibers as waveguides Maxwell s equations The wave equation Fiber modes Phase velocity, group velocity Dispersion Fiber Optical Communication Lecture 3, Slide 1 Maxwell s equations in

More information

Gauss s Law & Potential

Gauss s Law & Potential Gauss s Law & Potential Lecture 7: Electromagnetic Theory Professor D. K. Ghosh, Physics Department, I.I.T., Bombay Flux of an Electric Field : In this lecture we introduce Gauss s law which happens to

More information

The Spike Response Model: A Framework to Predict Neuronal Spike Trains

The Spike Response Model: A Framework to Predict Neuronal Spike Trains The Spike Response Model: A Framework to Predict Neuronal Spike Trains Renaud Jolivet, Timothy J. Lewis 2, and Wulfram Gerstner Laboratory of Computational Neuroscience, Swiss Federal Institute of Technology

More information

NMR, the vector model and the relaxation

NMR, the vector model and the relaxation NMR, the vector model and the relaxation Reading/Books: One and two dimensional NMR spectroscopy, VCH, Friebolin Spin Dynamics, Basics of NMR, Wiley, Levitt Molecular Quantum Mechanics, Oxford Univ. Press,

More information

On the Blurring of the Funk Radon Transform in Q Ball Imaging

On the Blurring of the Funk Radon Transform in Q Ball Imaging On the Blurring of the Funk Radon Transform in Q Ball Imaging Antonio Tristán-Vega 1, Santiago Aja-Fernández 1, and Carl-Fredrik Westin 2 1 Laboratory of Image Processing, University of Valladolid, Spain

More information

Supplementary Information: Dependence of nuclear spin singlet lifetimes on RF spin-locking power

Supplementary Information: Dependence of nuclear spin singlet lifetimes on RF spin-locking power Supplementary Information: Dependence of nuclear spin singlet lifetimes on RF spin-locking power Stephen J. DeVience a, Ronald L. Walsworth b,c, Matthew S. Rosen c,d,e a Department of Chemistry and Chemical

More information

K ex. Conformational equilibrium. equilibrium K B

K ex. Conformational equilibrium. equilibrium K B Effects of Chemical Exchange on NMR Spectra Chemical exchange refers to any yprocess in which a nucleus exchanges between two or more environments in which its NMR parameters (e.g. chemical shift, scalar

More information

Part III: Sequences and Contrast

Part III: Sequences and Contrast Part III: Sequences and Contrast Contents T1 and T2/T2* Relaxation Contrast of Imaging Sequences T1 weighting T2/T2* weighting Contrast Agents Saturation Inversion Recovery JUST WATER? (i.e., proton density

More information

COPYRIGHTED MATERIAL. Production of Net Magnetization. Chapter 1

COPYRIGHTED MATERIAL. Production of Net Magnetization. Chapter 1 Chapter 1 Production of Net Magnetization Magnetic resonance (MR) is a measurement technique used to examine atoms and molecules. It is based on the interaction between an applied magnetic field and a

More information

Problem Set Number 01, MIT (Winter-Spring 2018)

Problem Set Number 01, MIT (Winter-Spring 2018) Problem Set Number 01, 18.377 MIT (Winter-Spring 2018) Rodolfo R. Rosales (MIT, Math. Dept., room 2-337, Cambridge, MA 02139) February 28, 2018 Due Thursday, March 8, 2018. Turn it in (by 3PM) at the Math.

More information

Spectral Broadening Mechanisms

Spectral Broadening Mechanisms Spectral Broadening Mechanisms Lorentzian broadening (Homogeneous) Gaussian broadening (Inhomogeneous, Inertial) Doppler broadening (special case for gas phase) The Fourier Transform NC State University

More information

Expressions for the covariance matrix of covariance data

Expressions for the covariance matrix of covariance data Expressions for the covariance matrix of covariance data Torsten Söderström Division of Systems and Control, Department of Information Technology, Uppsala University, P O Box 337, SE-7505 Uppsala, Sweden

More information

Pore Length Scales and Pore Surface Relaxivity of Sandstone Determined by Internal Magnetic Fields Modulation at 2 MHz NMR

Pore Length Scales and Pore Surface Relaxivity of Sandstone Determined by Internal Magnetic Fields Modulation at 2 MHz NMR The OpenAccess Journal for the Basic Principles of Diffusion Theory, Experiment and Application Pore Length Scales and Pore Surface Relaxivity of Sandstone Determined by Internal Magnetic Fields Modulation

More information

A cylinder in a magnetic field (Jackson)

A cylinder in a magnetic field (Jackson) Problem 1. A cylinder in a magnetic field (Jackson) A very long hollow cylinder of inner radius a and outer radius b of permeability µ is placed in an initially uniform magnetic field B o at right angles

More information