A Case for Periodic Calibration or Verification of RTDs

Size: px
Start display at page:

Download "A Case for Periodic Calibration or Verification of RTDs"

Transcription

1 A Case for Periodic Calibration or Verification of RTDs If you are experiencing audio problems please call the teleconference number below Phone Access code

2 Your Host and Presenter 2 Presenter Bill Bergquist, Principal Applications Engineer and RTDologistTM 35 years experience in temperature measurement with RTDs and thermocouples in the aerospace, industrial, and laboratory markets. Host Jeff Wigen, National Sales Manager 30 years in sales and marketing of custom designed made-to-order products for the industrial and biotech markets.

3 What We ll Cover Terminology Refresher What is Calibration? Verification? Why you should do it When you should do it How exactly do you go about doing it? 3

4 Terminology Refresher R0 = resistance at 0 C IR = insulation resistance Dewar = insulated container (Thermos bottle) SPRT = standard platinum resistance thermometer ITS-90 = International Temperature Scale of 1990 IPTS-68 = International Practical Temperature Scale 1968 TPW = triple point of water 0.01 C or Kelvin K = Kelvin temperature scale (used for ITS-90) mk or millik =.001 Kelvin NIST = National Institute of Standards and Technology NVLAP = National Voluntary Laboratory Accreditation Program A2LA = American Association for Laboratory Accreditation 4

5 A Case for Periodic Calibration/Verification No regular maintenance schedule for the measurement point Sensor began giving erratic readings Wasted energy Questionable product quality How long had this been a problem? Nobody knows for sure. 5

6 What is Calibration? Verification? 6 Calibration: Calibration is performed to verify sensor/instrument performance and characterize the R vs. T relationship Usually involves adjustment of the instrument however there typically is no adjustment to an RTD (can adjust the instrument to which it is connected) Maintain traceability of parameters with reference to national/international standards Verification: A check of the RTD to see if it meets the ASTM E1137 or IEC standards for R0, insulation resistance (RIR), and interchangeability tolerances Can be performed in the lab or field by comparison to a temperature standard, or measured with DMM

7 Why? 7 Insure accuracy of measurement Consistent product quality Meet Quality System requirements Meet 3rd party requirements Safety Efficient use of energy Poor accuracy = wasted $ Establish CONFIDENCE in the temperature measurement

8 Energy Cost and RTD Accuracy Process Fluid: Flow Rate: Control Temperature: Energy Cost: Water 25 GPM 100 F 6 / KW-hour Annual Cost of Energy Per F Error $1800 8

9 Why 9 Initial Calibration/Verification New plant or equipment commissioning Verify manufacturer data shipping and installation damage Insure accuracy of measurements Establish and record initial performance necessary to show trends Ongoing Calibration/Verification Minimize and control random and systematic errors Identify sensor and instrument errors Compare and complement the quality and reliability of measurements by comparison to standards Provide traceability to national standards, (e.g. NIST) Conformance to RTD standards ASTM 1137 and IEC Meet Regulatory Requirements (FDA, USDA, NRC)

10 When 10 Frequency of calibration Before and/or after a critical measurement Risk mitigation After an event 3rd party requirement Defined calendar schedule Manufacturer recommendation Your best guess

11 When 11 Frequency Process also dictates the calibration cycle Probe drift due to: Vibration Shock Temperature high temp = more drift Temperature cycling Product value Corrosion, erosion, product build-up

12 When 12 How time and temperature affect drift rate

13 When 13 How temperature cycling affects drift rate

14 Frequency 14 Set to meet your confidence level Guidelines Graph acceptable error vs. time std. deviations based on accumulated data Group together sensor data from similar locations Manufacturer recommendation Process characteristics Third party requirement Safety After an event Before and/or after critical measurement or batch Complete 5 cycles w/o shift then double the interval

15 How 15 Temperature scales Equations Options Methods Calibration Equipment & Software Rt= R0[1 + A * t + B * t2 + (t 100)C * t3]

16 How 16 Two types of RTD checks Verification or tolerance check Compare resistance to defined R vs. T such as IEC or ASTM Usually performed at the ice point. This may also take the form of a field check by using a standard inserted into the process or a nearby test well Characterization Calibrate at several temperatures and use equations to define R vs. T

17 How 17 Insulation resistance First and most important calibration/verification check Low IR can cause a low temperature measurement due to shunting between the sensing element wires Most IR failures are due to moisture and/or contaminants that may have entered the probe

18 Insulation Resistance Test method Lower resistance = lower measured temperature Test at 50 VDC IR should be >100 megohms at 25 C 18

19 Insulation Resistance 19 Low insulation resistance (RIR or IR) IR acts as a shunt resistor to the measurement circuit the lower the IR the higher the effect on the accuracy of the probe. The equation for calculating theoretical effect of IR on the measurement is basically the equation for calculating the overall resistance of resistor in parallel, where one resistor is the PRT (RPRT) and the other is the insulation resistance ( RIR) RMeasured Where: [ RPRT xrir ] [ RPRT RIR ] RMeasured = Resultant measured resistance RPRT = Resistance of PRT element RIR = Insulation resistance value So for example: a probe that reads 100Ω at 0ºC that then degrades to IR of 0.1 MΩ the measured resistance will be which equates to approximately -0.26ºC.

20 How 20 Ice Bath Check Resistance at 0 C is the most important and easiest to check Standard interchangeability tolerances established by either ASTM 1137, IEC 60751, or manufacturer

21 Ice Point Check 21 Using an Ice Bath and resistance readout, check resistance at 0 C Crushed ice, purified water, and an insulated container

22 Interchangeability 22 4 IEC Class B ASTM Grade B 3 2 IEC Class A ASTM Grade A Tolerance (± C) ASTM Grade A -1 IEC Class A -2 ASTM Grade B -3 IEC Class B -4 Temperature ( C) 800

23 Interchangeability 23 Standard Tolerance Defining Equation¹ ASTM E1137 ASTM E1137 IEC IEC IEC IEC Grade A Grade B Class AA Class A Class B Class C ± [ t ] ± [ t ] ± [ t ] ± [ t ] ± [ t ] ± [ t ] Note 1: t = absolute value of temperature of interest in C

24 How - Temperature Scales Evolution of standard temperature scales IPTS-27 IPTS-48 IPTS ITS-90 ITS-90 (International Temperature Scale) Released in 1990 The official international scale In better agreement with thermodynamic values than the IPTS-68 ITS-90 vs. IPTS-68 ITS-90 Uses resistance at TPW Most accurate Complex equations IPTS-68 Uses resistance at ice point Simpler equations Less accurate Callendar-Van Dusen equation

25 How - Equations 25 Callendar-Van Dusen Equation For the range between 0 C to 661 C the equation is Rt = R0(1 + At + Bt2) For the range between -200 C to 0 C the equation is Rt= R0[1 + At + Bt2 + (t 100)C * t3] t = temperature ( C) R = resistance at temperature t R0 = resistance at the ice point A, B, and C from the RTD standards: A = x 10-3 B = x 10-7 C = x 10-12

26 How - Methods 26 Methods of calibration Fixed point Comparison Laboratory Field

27 Comparison Calibration Most common method Comparison of unknown to known sensors Multiple sensors can be calibrated at the same time Equipment Meter, Standard PRT, Recorder, etc. (system) All add to uncertainty level The standard PRT should have an accuracy at least four times greater than the unit under test 27

28 Comparison Calibration 28 More practical and less expensive than fixed point temperature calibration Laboratory Typical uncertainty: C to 0.01 C Very high accuracy reference resistance bridge, standard PRT, calibration baths, etc. Uses some fixed point temperatures Field Typical uncertainty: 0.05 to 0.5 C Accurate reference meters, secondary PRTs, baths or dry-wells Instruments are field compatible

29 Typical Comparison System Setup 29

30 Equipment 30

31 Standard PRTs 31 Specifications Very fragile Used in laboratory environments Highest accuracy, high repeatability, low drift -328 to 1983 F (-200 to 1084 C), accurate to ±.0018 F (±.001 C)

32 Fluid Calibration Baths Range: -80 C to 500 C Fluids: Alcohol, Water, Silicone Oil, Salt Stability: < ±0.001 C to ±0.05 C Working depth: 12 to 18 Working diameter: 4 or Larger 32

33 Metal (hot) Block Baths Range: -30 C to 700 C Stability: ±0.02 C to ±0.05 C Working depth: 6 Portable 33

34 Thermometer Readout A readout device or data acquisition system is needed to display or record temperature data when performing a calibration 34

35 How - Calibration Options Factory Calibration Options Matched Calibration Match RTD to a transmitter Matched to a Temperature Readout Multiple Point Calibration -196, -38, 0, 100, 200, 300, 420, and 500 C 35

36 How Calibration Options Transmitters Matched to RTD Improved system accuracy 36

37 Outside Calibration Service Accredited Lab such as A2LA, NVLAP and others Defined QA program Training Documentation Complete solution RTDs Transmitters System Retention of records Scheduling Experience Support Guarantee 37

38 Summary 38 Calibrate or Verify to insure measurement confidence Verification IR and ice point check Calibration multiple points to characterize R vs. T Frequency Process dictates the calibration cycle Probe drift Vibration Shock Temperature range and cycling Product value Safety

39 Thank you for attending! Questions? Use the chat window to send us a question now Contact us later at bbergquist@burnsengineering.com or visit

40 BE educated Watch for upcoming RTDology events View previous recordings on YouTube. Search for RTDology Let us know if there is a topic you would like to see covered Find discussions from industry experts on LinkedIn: Temperature Measurement Group

41 How - Equations 41 Callendar-Van Dusen Equation Callendar-Van Dusen equations are interpolation equations which describe the temperature versus resistance relationship of industrial PRTs. Equation for the temperature range of 32 F (0 C) to 1562 F (850 C) is: For the temperature range of -392 F ( -200 C) to 32 F (0 C): Where: R(t) = the resistance of the PRT at the temperature t = temperature in deg. C. R0 = the nominal resistance of the PRT at 0 C A,B,C = calibration coefficients The nominal calibration coefficients for standard curves are: R0 A B C x x x 10-12

42 How - Equations 42 To determine the temperature from a measured resistance, a different set of equations and calibration coefficients are required. For temperatures greater than 0 C: t C = ((Rt-R0)/(Alpha*R0)) + Delta((t/100)-1)(t/100) For temperatures less than 0 C: t C = ((Rt-R0)/(Alpha*R0)) + Delta((t/100)-1)(t/100) + Beta((t/100)-1)(t/100)3 Where: t = temperature to be calculated Rt = measured resistance at unknown temperature R0 = resistance of the sensor at 0 C Alpha = , Delta = , Beta = To correctly determine the temperature from a given resistance, you must iterate the equations a minimum of five times. After each calculation, the new value of temperature (t) is plugged back into the equations. The calculated temperature value will converge on its true value. After five iterations, the calculated temperature should be within.001 C of the true value.

43 How - Equations 43 One more tidbit of information: A = Alpha x { 1+ (Delta/100)} C-1 B = -Alpha x Delta x 10-4 C-2 C = -Alpha X Beta x 10-8 C-4

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

ECNG3032 Control and Instrumentation I

ECNG3032 Control and Instrumentation I sensor ECNG3032 Control and Instrumentation I Lecture 1 Temperature Sensors Sensors The sensor is the first element in the measurement system. Measurand Transducer Principle Excitation Signal Interface

More information

Certificate of Accreditation to ISO/IEC 17025:2005

Certificate of Accreditation to ISO/IEC 17025:2005 United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018 New Thermometer Guidance Document from AASHTO Accreditation Program Maria Knake Asphalt Binder ETG Meeting May 10, 2018 Background Mercury reduction initiatives have hastened the need to find suitable

More information

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density High-precision calibration for temperature and density g/cm³ F K C DAkkS Calibration Laboratory for Temperature and Density Accredited DAkkS laboratory (D-K-15223-01-00) for temperature and density measurements

More information

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators Michael Coleman Fluke Corporation, Temperature Calibration Division, American Fork, Utah, USA mike.coleman@fluke.com

More information

STATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016

STATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016 APPLICATIONS Moisture monitoring of air or gas (SF6) Suitable for indoor or outdoor Industrial, medical or aerospace fields HIGHLIGHTS High voltage circuit breakers commonly used for distribution and transmission

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Page 1 of 5 CERTIFICATE OF CALIBRATION 1K Thermistor CERTIFICATE NO.: MANUFACTURER: Measurement Specialties DATE OF CALIBRATION: 26-Apr-17 TC15249 CUSTOMER: AMBIENT TEMPERATURE: (21 ± 3) C JOB NUMBER:

More information

Calibration of temperature sensors within Length Standard Section of NMIJ

Calibration of temperature sensors within Length Standard Section of NMIJ Calibration of temperature sensors within Length Standard Section of NMIJ Dr. Akiko Hirai Length Standard Section National Metrology Institute of Japan (NMIJ) 1 Contents Temperature measurement system

More information

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

ME 365 EXPERIMENT 5 FIRST ORDER SYSTEM IDENTIFICATION APPLIED TO TEMPERATURE MEASUREMENT SYSTEMS

ME 365 EXPERIMENT 5 FIRST ORDER SYSTEM IDENTIFICATION APPLIED TO TEMPERATURE MEASUREMENT SYSTEMS ME 365 EXPERIMENT 5 FIRST ORDER SYSTEM IDENTIFICATION APPLIED TO TEMPERATURE MEASUREMENT SYSTEMS Objectives: In this two week experiment, we will gain familiarity with first order systems by using two

More information

Temperature. 3

Temperature. 3 Temperature In 1848, Sir William Thomson (Lord Kelvin) stated the zero principle of dynamics. This principle enabled him to define thermodynamic temperature and to establish an objective method of measuring

More information

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating

More information

Designed to go where you work

Designed to go where you work Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers! Accurate and repeatable to ± 0.05 C over its full range, the 55A/55A Stik Thermometer is the new gold

More information

Exercise 1: Thermistor Characteristics

Exercise 1: Thermistor Characteristics Exercise 1: Thermistor Characteristics EXERCISE OBJECTIVE When you have completed this exercise, you will be able to describe and demonstrate the characteristics of thermistors. DISCUSSION A thermistor

More information

ph Probes Lab vs. Process Care, Calibrations and Comparison Presented by: Michael Blackburn Technical Sales Specialist at Hach Company

ph Probes Lab vs. Process Care, Calibrations and Comparison Presented by: Michael Blackburn Technical Sales Specialist at Hach Company ph Probes Lab vs. Process Care, Calibrations and Comparison Presented by: Michael Blackburn Technical Sales Specialist at Hach Company What Will We Cover Today? Calibrations vs. Validation Detection Methods

More information

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration 1551A Ex/155A Ex Intrinsically Safe Stik Thermometer The new gold standard of industrial temperature calibration Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers!

More information

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014 Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH Hannelore Bloemink, Jan Bijma Royal Netherlands Meteorological Institute (KNMI) Instrumentation Division P.O. Box 201, 3730 AE De Bilt, The Netherlands

More information

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE VALUATION OF TH ACCURACY OF A THRMOMTRIC RITANC BRIDG TO DIFFRNT VALU OF ATMOPHRIC PRUR Diego Almeida cuadorian tandards Institute, dalmeida@inen.gob.ec Abstract: The accuracy of a commercial thermometric

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

Temperature Sensors & Measurement

Temperature Sensors & Measurement Temperature Sensors & Measurement E80 Spring 2014 Contents Why measure temperature? Characteristics of interest Types of temperature sensors 1. Thermistor 2. RTD Sensor 3. Thermocouple 4. Integrated Silicon

More information

Experiment 3. Electrical Energy. Calculate the electrical power dissipated in a resistor.

Experiment 3. Electrical Energy. Calculate the electrical power dissipated in a resistor. Experiment 3 Electrical Energy 3.1 Objectives Calculate the electrical power dissipated in a resistor. Determine the heat added to the water by an immersed heater. Determine if the energy dissipated by

More information

CALIBRATION AND ACCURACY OF SECONDARY STANDARD PRT'S TO THE ITS-90

CALIBRATION AND ACCURACY OF SECONDARY STANDARD PRT'S TO THE ITS-90 CALIBRATION AND ACCURACY OF SECONDARY STANDARD PRT'S TO THE ITS-9 Presenter: Thomas J. Wiandt Southern California Edison 73 Fenwick Lane Westminster, CA 92683 Telephone: (714) 895-443 Fax: (714) 895-686

More information

Of all the physical properties, it is temperature, which is being measured most often.

Of all the physical properties, it is temperature, which is being measured most often. 1. Temperature 1.1 Introduction Of all the physical properties, it is temperature, which is being measured most often. Even the measurement of other physical properties,e.g. pressure, flow, level,, often

More information

M2 SERIES THERMOSTATS 0 F to 240 F, Narrow Differential, Hermetically Sealed ½

M2 SERIES THERMOSTATS 0 F to 240 F, Narrow Differential, Hermetically Sealed ½ M2 SERIES THERMOSTATS 0 F to 240 F, Narrow Differential, Hermetically Sealed ½ Introduction The Klixon M2 thermostat is constructed with a snap acting bimetal disc that serves as the actuating element.

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 PROTEMP MECHANICAL, INC. 3350 Scott Blvd, Bldg. #3 Santa Clara, CA 95054 Martin L. Reich Phone: 408 244 9821 CALIBRATION Valid To: August

More information

Exercise 1: Thermocouple Characteristics

Exercise 1: Thermocouple Characteristics The Thermocouple Transducer Fundamentals Exercise 1: Thermocouple Characteristics EXERCISE OBJECTIVE When you have completed this exercise, you will be able to describe and demonstrate the characteristics

More information

High Accuracy High Speed in Your Process

High Accuracy High Speed in Your Process DX-Series Handheld Precision IR Thermometers EXERGEN C O R P O R A T I O N The Only Certified Accurate NIST Traceable Infrared Temperature Instruments in the World* High Accuracy High Speed in Your Process

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ALABAMA SCALE & INSTRUMENT MARSHALL, INC. d.b.a. ASI Calibration Labs - Texas 1006 East End Blvd North Marshall, TX 75672 Lawson Hill

More information

Designing a Thermostat Worksheet

Designing a Thermostat Worksheet Designing a Thermostat Worksheet Most of us have a thermostat in our homes to control heating and cooling systems of our home. These important devices help us save energy by automatically turning off energy

More information

Stage 4 G-03: CONDUCTIVITY OF SOLUTIONS CP: USP. May 2015 INTRODUCTION

Stage 4 G-03: CONDUCTIVITY OF SOLUTIONS CP: USP. May 2015 INTRODUCTION 002-1601PDG.pdf 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Stage 4 G-03: CONDUCTIVITY OF SOLUTIONS CP: USP May 2015 INTRODUCTION This chapter provides information on how to apply electrical

More information

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors 1. Scope... 2 2. Introduction... 2 3. Selection of thermocouples

More information

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT JO Favreau 1, E Georgin 1, B Savanier 1, A. Merlone 2 1 CETIAT, 96100 Villeurbanne, France 2 INRIM, Torino, Italy Abstract. CETIAT is a calibration laboratory

More information

SEN TRONIC AG 1 A 6 6 / "

SEN TRONIC AG 1 A 6 6 / 1A 66/" 0!"#$%&'() %"*+", - %"*.", - /01234%( 34.+*!54%& 0*%/# "6#,7857.'.0" - 6#)9.:. &%&;! 0 &????'.&% )&" 8" @&& (++ '() %('.('/(#$!!! ' %! %!& ;!;8 ;!;8 0 &&'&&;! C;!C&(D"@@ &;! 0&&+%&;! C&=;!C&(D"@@

More information

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

Reference Only Spec. No. JEFL243B 0003F-01 P 1/ 6

Reference Only Spec. No. JEFL243B 0003F-01 P 1/ 6 Spec. No. JEFL243B 0003F-01 P 1/ 6 DC Common Mode Choke Coil PLT09HN2003R0P1B Reference Specification 1. Scope This reference specification applies to DC Common Mode Choke Coil (PLT09H Series). 2. Part

More information

Applications. Capacitance Tolerance G = ±2% J = ±5%

Applications. Capacitance Tolerance G = ±2% J = ±5% Surface Mount Metallized PPS Film Capacitor LDB, Unencapsulated Stacked Chip, Size 1206 1812, 16 and 50 VDC Overview Polyphenylene sulphide (PPS) film capacitor for surface mounting. Applications Typical

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V (Hereinafter called the Organization)

More information

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

STATUS OF THE WIGOS DEMONSTRATION PROJECTS STATUS OF THE WIGOS DEMONSTRATION PROJECTS Demonstration Project Morocco Strengthening Moroccan RIC Capacities (Submitted by Rabia Merrouchi, National Meteorological Service of Morocco (DMN)) February

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 AMERICAN INSTRUMENT CORPORATION Crossroads Centre 702 Rose Drive Hartland, WI 53029 Jeff Quinn Phone: 262 367 4409 CALIBRATION Valid

More information

CALIBRATION REPORT FOR THERMOHYDROMETER

CALIBRATION REPORT FOR THERMOHYDROMETER CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited

More information

Chip tantalum capacitors (Fail-safe open structure type)

Chip tantalum capacitors (Fail-safe open structure type) TCFG0J106M8R Chip tantalum capacitors (Fail-safe open structure type) TCFG Series Case Features 1) Safety design by open function built - in. 2) Wide capacitance range 3) Screening by thermal shock. Dimensions

More information

Part 2. Sensor and Transducer Instrument Selection Criteria (3 Hour)

Part 2. Sensor and Transducer Instrument Selection Criteria (3 Hour) Part 2 Sensor and Transducer Instrument Selection Criteria (3 Hour) At the end of this chapter, you should be able to: Describe the definition of sensor and transducer Determine the specification of control

More information

An Introduction to Insulation Resistance Testing

An Introduction to Insulation Resistance Testing An Introduction to Insulation Resistance Testing In a perfect world, electrical insulation would allow no current to flow through it. Unfortunately, a number of factors can over time result in the deterioration

More information

ME 105 Mechanical Engineering Laboratory Spring Quarter Experiment #2: Temperature Measurements and Transient Conduction and Convection

ME 105 Mechanical Engineering Laboratory Spring Quarter Experiment #2: Temperature Measurements and Transient Conduction and Convection ME 105 Mechanical Engineering Lab Page 1 ME 105 Mechanical Engineering Laboratory Spring Quarter 2010 Experiment #2: Temperature Measurements and Transient Conduction and Convection Objectives a) To calibrate

More information

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY Proceedings of the 6th International Conference on Mechanics and Materials in Design, Editors: J.F. Silva Gomes & S.A. Meguid, P.Delgada/Azores, 26-30 July 2015 PAPER REF: 5376 STATISTICAL PROCESS CONTROL

More information

METALLIZED POLYPROPYLENE AC FILTERING CAPACITORS

METALLIZED POLYPROPYLENE AC FILTERING CAPACITORS METALLIZED POLYPROPYLENE AC FILTERING CAPACITORS Technology looking ahead With the new LNF series a new level of safety for the AC capacitors has been reached: Active safety: When the capacitor is stressed

More information

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable

More information

Part Number Range Accuracy Printing WP * Probe Special User LDC with Page Logging included Feature Cal Backlight

Part Number Range Accuracy Printing WP * Probe Special User LDC with Page Logging included Feature Cal Backlight Table of Contents Comparison Chart Introduction Temperature Indicators & Controllers Portable Infrared Printing/Logging Page O2 O3 O10 O14 O30 O33 O1 Part Number Range Accuracy Printing WP * Probe Special

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Hards Laboratories cc, t/a Technology Solutions

More information

Humidity Calibration Solutions

Humidity Calibration Solutions Instruments Humidity Solutions www.michell.com Instruments Humidity Solutions The Importance of Regular The reliable operation of a hygrometer and indeed any measuring instrument, can only be verified

More information

Temperature Measurements

Temperature Measurements Engineering 80 Spring 2015 Temperature Measurements SOURCE: http://www.eng.hmc.edu/newe80/pdfs/vishaythermdatasheet.pdf SOURCE: http://elcodis.com/photos/19/51/195143/to-92-3_standardbody to-226_straightlead.jpg

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES BUREAU INTERNATIONAL DES POIDS ET MESURES Bilateral comparison of 1 Ω and 10 kω standards (ongoing BIPM key comparisons BIPM.EM-K13.a and 13.b) between the NIMT (Thailand) and the BIPM December 016 Final

More information

Activity 1: Investigating Temperature

Activity 1: Investigating Temperature Contents Activity Overview... 5 Quick Start Guide... 5 Software Installation... 5 Hardware Setup... 6 mytemp Getting Started Program... 10 General Tips and Tricks... 11 Activity 1: Investigating Temperature...

More information

- Copyright Dewesoft d.o.o., all rights reserved. Temperature measurement

- Copyright Dewesoft d.o.o., all rights reserved. Temperature measurement www.dewesoft.com - Copyright 2000-2019 Dewesoft d.o.o., all rights reserved. Temperature measurement Table of Contents Temperature and temperature scales 2 Types of temperature sensors 4 1. Thermocouples

More information

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS Gianalberto Grego, Fabio F. Muciaccia W.E.S.T. Srl, Milan, Italy west-hydro@libero.it ABSTRACT The

More information

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard 92A AC/DC Transfer Standard Technical Data Support for your most demanding ac measurement requirements ppm total uncertainty Traceable to national standards range: 2 mv to 00 V Frequency range: Hz to 1

More information

SIMPLE LIQUID NITROGEN COMPARATOR MODEL 461 User Maintenance Manual/Handbook

SIMPLE LIQUID NITROGEN COMPARATOR MODEL 461 User Maintenance Manual/Handbook SIMPLE LIQUID NITROGEN COMPARATOR MODEL 461 User Maintenance Manual/Handbook Isothermal Technology Limited, Pine Grove, Southport, PR9 9AG, England Tel: +44 (0)1704 543830 Fax: +44 (0)1704 544799 Internet:

More information

Chip tantalum capacitors (Fail-safe open structure type)

Chip tantalum capacitors (Fail-safe open structure type) TCFG1C6M8R Chip tantalum capacitors (Fail-safe open structure type) TCFG Series Case Features 1) Safety design by open function built - in. 2) Wide capacitance range 3) Screening by thermal shock. Dimensions

More information

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature

More information

LM34 Precision Fahrenheit Temperature Sensors

LM34 Precision Fahrenheit Temperature Sensors 1 Precision Fahrenheit Temperature Sensors 1 Features 3 Description 1 Calibrated Directly in Degrees Fahrenheit The series devices are precision integratedcircuit temperature sensors, whose output voltage

More information

For the National Voluntary Laboratory Accreditation Program

For the National Voluntary Laboratory Accreditation Program SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 Additel Corporation 2900 Saturn Street Brea, CA 92821 Eric Chavier Phone: 714-998-6899 Fax: 714-998-6999 E-mail: eric.chavier@additel.com URL: http://additel.com

More information

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance / maintenance / calibration Holger Dörschel, Dr Tilman Holfelder WMO International Conference on Automatic Weather Stations

More information

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,

More information

473-SHX Dew Point Mirror

473-SHX Dew Point Mirror Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES BUREAU INTERNATIONAL DES POIDS ET MESURES Bilateral comparison of 1 Ω and 10 kω standards (ongoing BIPM key comparisons BIPM.EM-K13.a and 13.b) between the SMD (Belgium) and the BIPM October 2017 Final

More information

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

White Paper. Perform Conductivity Measurements In Compliance with USP <645> Perform Conductivity Measurements In Compliance with USP Water is the most widely used substance, raw material, or ingredient in the production, processing and formulation of compendial articles.

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

Temperature measurement

Temperature measurement Temperature measurement 19 2 Example of connections of particular elements Temperature display units BGK85 (page 31) Option Computer, control, itp. Temperature measurement TFC TFL (page 23) (page 23) Temperature

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

CALIBRATION CERTIFICATE # 503

CALIBRATION CERTIFICATE # 503 CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion

More information

Usage limitations and accuracies of platinum resistance thermometers (DIN EN 60751:2009) in industrial applications

Usage limitations and accuracies of platinum resistance thermometers (DIN EN 60751:2009) in industrial applications Technical information Usage limitations and accuracies of platinum resistance thermometers (DIN EN 60751:2009) in industrial applications WIKA data sheet IN 00.17 General information Temperature is a measurement

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CONTROL COMPANY 12554 Galveston Road Suite B230 Webster, TX 77598 Aaron Judice Phone: 281 482 1714 CALIBRATION Valid To: March 31, 2018

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Welcome to the Isotech Journal of Thermometry Fundamentals of Thermometry Part 1 Temperature Scales Practical Calibration of Thermometers

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 RS CALIBRATION SERVICES, INC. 1047 Serpentine Lane Pleasanton, CA 94566 Mr. Ralph Sabiel Phone: 925 462 4217 CALIBRATION Valid To: April

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 EQUIPMENT SERVICES GROUP, LLC 3269 Hilton Rd. Ferndale, MI 48220 Michael Macauley Phone: 248 227 5310 CALIBRATION Valid To: December

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares

More information

Fine Tuning Coriolis Flow Meter Calibrations Utilizing Piece- Wise Linearization

Fine Tuning Coriolis Flow Meter Calibrations Utilizing Piece- Wise Linearization Fine Tuning Coriolis Flow Meter Calibrations Utilizing Piece- Wise Linearization Tonya Wyatt Global Chemical Industry Manager Emerson Process Management Micro Motion, Inc. AGA Operations Conference and

More information

Lab 1f Boiling Heat Transfer Paradox

Lab 1f Boiling Heat Transfer Paradox Lab 1f Boiling Heat Transfer Paradox OBJECTIVES Warning: though the experiment has educational objectives (to learn about boiling heat transfer, etc.), these should not be included in your report. - Obtain

More information

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook Reference Manual 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book and Flow Handbook www.rosemount.com Reference Manual 405 and 1595 405 Compact Orifice Series and 1595 Conditioning

More information

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard 92A AC/DC Transfer Standard Technical Data Support for your most demanding ac measurement requirements ppm total uncertainty Traceable to national standards range: 2 mv to 00 V Frequency range: Hz to 1

More information

Welcome. Functionality and application of RTD Temperature Probes and Thermocouples. Dipl.-Ing. Manfred Schleicher

Welcome. Functionality and application of RTD Temperature Probes and Thermocouples. Dipl.-Ing. Manfred Schleicher Welcome Functionality and application of RTD Temperature Probes and Thermocouples Dipl.-Ing. Manfred Schleicher This presentation informs about the basics of RTD Temperature Probes and thermocouples Functionality

More information

Standard Test Methods for Measurement of Thermal Expansion of Rock Using a Dilatometer 1

Standard Test Methods for Measurement of Thermal Expansion of Rock Using a Dilatometer 1 Designation: D 4535 85 (Reapproved 2000) Standard Test Methods for Measurement of Thermal Expansion of Rock Using a Dilatometer 1 This standard is issued under the fixed designation D 4535; the number

More information

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration User Manual Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio with NIST-Traceable Calibration Model 20250-09 THE STANDARD IN PRECISION MEASUREMENT Introduction The Digi-Sense

More information

Temperature Measurements

Temperature Measurements ME 22.302 Mechanical Lab I Temperature Measurements Dr. Peter Avitabile University of Massachusetts Lowell Temperature - 122601-1 Copyright 2001 A transducer is a device that converts some mechanical quantity

More information

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book and Flow Handbook www.rosemount.com 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book NOTICE Read

More information

Approximation of ITS-90 with High Temperature Thermometers up to 1085 C

Approximation of ITS-90 with High Temperature Thermometers up to 1085 C Approximation of ITS-9 with High Temperature Thermometers up to 185 C N.P. Moiseeva, A.I. Pokhodun D. I. Mendeleyev Institute of Metrology, St. Petersburg, Russia Abstract. The upper limit of the ITS-9

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

DATA SHEET. Thermocouple module with digital I²C-Interface - THMOD-I²C. Characteristic features. Areas of application. Features.

DATA SHEET. Thermocouple module with digital I²C-Interface - THMOD-I²C. Characteristic features. Areas of application. Features. Description Characteristic features Industrial temperature measuring method Wide measuring range, -270...+1360 C Digital I²C-interface Simple integration to micro-controller Scope of supply with thermocouple,

More information

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES XVII IMEKO World Congress Metrology in the 3rd Millennium June 7, 003, Dubrovnik, Croatia MODEL OR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION O THERMOCOUPLES Georges Bonnier, Eliane Renaot, DavorZvizdic,

More information

Chilled Mirror Hygrometry

Chilled Mirror Hygrometry Chilled Mirror Hygrometry Precision instruments for critical measurement and control applications Chilled Mirror Hygrometry Wherever precise, fundamental measurements are required in your laboratory or

More information

DUBLIN INSTITUTE OF TECHNOLOGY Kevin Street, Dublin 8.

DUBLIN INSTITUTE OF TECHNOLOGY Kevin Street, Dublin 8. Question Sheet Page 1 of 5 Instructions for the student: Question 1 is compulsory [40 marks] Attempt any two other questions [30 marks per question] The following must be made available during the examination:

More information

Edgetech Instruments Inc.

Edgetech Instruments Inc. 2015_07_ETI OVERVIEW Rev 3 MADE IN THE USA Edgetech Instruments (ETI) is a USA company based in Massachusetts. We have over 50 years of expertise in the design and manufacture of a comprehensive range

More information

DM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability

DM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability DM70 Hand-held Dewpoint Meter Truly Fast Response - Superior Long-term Stability DM70 - When Dry Air Is Important The Vaisala DM70 Hand-held Dewpoint Meter offers stable and fast measurement for industrial

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 JEM PRECISION 9615 56 Avenue NW Edmonton, Alberta T6E 0B2 Canada Robert Korthuis Phone: 780-440 1933 CALIBRATION

More information