A REVIEW ON RAPID RESOLUTION LIQUID CHROMATOGRAPHY OF AGILENT1200 RRLC SERIES
|
|
- Christal Grant
- 5 years ago
- Views:
Transcription
1 WORLD JOURNAL OF PHARMACY AND PHARMACEUTICAL SCIENCES Haripriya et al. SJIF Impact Factor Volume 4, Issue 07, Review Article ISSN A REVIEW ON RAPID RESOLUTION LIQUID CHROMATOGRAPHY OF AGILENT1200 RRLC SERIES P.Haripriya*, Sharadha Srikanth, A.Ajitha and V.Umamaheswararao Department of Pharmaceutical Analysis and Quality Assurance, CMR College of Pharmacy, Kandlakoya (v), Medchal Road, Hyderabad , T.S, India. Article Received on 16 May 2015, Revised on 07 June 2015, Accepted on 29 June 2015 *Correspondence for Author P.Haripriya Department of Pharmaceutical Analysis and Quality Assurance, CMR College of Pharmacy, Kandlakoya (v), Medchal Road, Hyderabad , T.S, India. ABSTRACT In this Application Note an Agilent 1200 Series Rapid Resolution LC system was provide a liquid chromatography offering ultra fast and highest analysis speed high improved sensitivity. High temperature up to 100 C on certain columns allows more selectivity flexibility and reduces solvent viscosity to allow even faster separation. This is achieved by using sub-2 micron (STM) column particle which means that for high flow rates or long columns additional pressure is required to drive the mobile phase through the column. Detectors with high data rates preserve the resolution of Very fast peaks eluting from the RRLC The needed analytical instrumentation requires optimal cost-ofownership instruments with high reliability, high flexibility, and ease of use. column particle, Ultra fast. KEYWORDS: Rapid Resolution Liquid Chromatography, Small INTRODUCTION TO LIQUID CHROMATOGRAPHY Chromatography is a powerful analytical method that is widely used for the separation, identification and determination of the chemical components in complex mixtures. It was first demonstrated by the Russian botanist Milkhail Tswett in 1906 when he separated different plant pigments by passing an extract Of the leaves through a glass column packed with calcium carbonate the term chromatography came from Greek words, chroma meaning "color and graphein meaning to write". Tswett s invention went virtually unnoticed for several decades. Chromatography may be defined as the separation of a mixture of components based on the distribution coefficients for each component between two phases; a Vol 4, Issue 07,
2 stationary phase and a mobile phase. The mobile phase is liquid consisting of a solvent or mixture of solvents and the stationary phase is a solid support. [7] CLASSIFICATION OF CHROMATOGRAPHIC SEPARATIONS (1)Adsorption chromatography; this is the traditional separation Mode in which the stationary phase is solid and the mobile phase is liquid. The mixture components adsorb on the surface of the solid phase. The more strongly a solute is adsorbed, the slower it travels the column. (2) Partition chromatography; the stationary phase is liquid bonded to a solid Support and the mobile phase is a liquid. The separation of the mixture components is based on their equilibration between the liquid on the stationary phase surface and the mobile phase. In the liquid-liquid partition chromatography, if a polar stationary phase with a non-polar mobile is used, polar compounds are retained and separated. This is called normal- phase chromatography. If non-polar used, non-polar compounds are retained and separated. This is called reversed-phase chromatography. (3)In ion exchange chromatograph, the stationary phase consists of either cationic exchange or anionic exchange groups attached to polymeric or silica materials and The mobile phase consists of an electrolyte. (4) In size exclusion chromatography, the separation of a mixture of components is based on the size of the molecules. Larger molecules are excluded from the pores of the stationary phase particles and eluted first, where as small molecules are retained in the pores for a while be for elution. [7] TYPES OF HPLC CHROMATOGRAPHY TECHNIQUES HPLC is compared with the classical techniques are characterized by Rapid Resolution Liquid chromatography (RRLC) Ultra Performance Liquid chromatography (UPLC) Ultra Fast Liquid chromatography (UFLC) Nano Liquid chromatography (NANO LC) FEATURES OF RAPID RESOLUTION LIQUID CHROMATOGRAPHY (RRLC) The design concept of the 1200 Rapid Resolution LC (RRLC) system was to provide a liquid chromatography offering ultra fast and highest analysis speed and high improved sensitivity Vol 4, Issue 07,
3 and pressure at a minimum yet which retain full functionality for standard HPLC applications. It is the fastest and most efficient and flexible LC system in the world. It has become an increasingly useful approach to achieve higher throughput, improved sensitivity and reduce costs. The RRLC system enable faster analysis ( theoretically up to 20x) than with conventional HPLC. This is achieved by using sub-2 micron (STM) column particle which means that for high flow rates or long columns additional pressure is required to drive the mobile phase through the column. The RRLC flow path is optimized to produce minimal backpressure and ZORBAX RRHT columns have an engineered particle size distribution that produces significantly less back pressure those other STM columns. High temperature up to 100 C on certain columns allows more selectivity flexibility and reduces solvent viscosity to allow even faster separation. High flow rates up to 5ml/min can be used for ultra -fast separations. The adjustable delay volume filly supports 2.1 to 4.6 mm i.d. columns. A low dispersion tubing kit and low Volume flow cells minimize peak dispersion for narrow bore Columns. Detectors with high data rates preserve the resolution of Very fast peaks eluting from the RRLC. [10] RRLC has been applied for the analysis of nutritional components (catechins) in green tea using the same approach. In this study, the following goals were achieved. 1) Demonstration of ease of conversion of a traditional HPLC method to a RRLC method and 2) To investigate the potential of RRLC and RRLC/MS on 1.8 µ m porous particles packed into short columns operating at a high flow rate compared to the performance of 5 µ m porous particles packed conventional columns. [10] Fig 1: Rapid resolution liquid chromatography (RRLP) Vol 4, Issue 07,
4 Fig 2: Schematic diagram of RRLC 2. PRINCIPLE The underlying principles of this evolution are governed by the van Demeter equation, which is an empirical formula that describes The relationship between linear velocity (flow rate) and plate height (HETP or 1/column efficiency) H=A+B/u + Cu A-Eddy s diffusion B- Longitudinal diffusion C-Concentration u- Linear Velocity Column efficiency: N=16RT²/W² N-number of theoretical plates RT-retention time W-path widt Resolution is proportional to the square root of N. But since N is inversely proportional to particle size ( Resolution (Rs) equation. Rs= 4 (α-1/α) (k/k+1) Equation.2 Resolution is proportional to the square root of N. But since N is inversely proportional to particle size (dp). N 1/dp Equation.3. Vol 4, Issue 07,
5 Efficiency (N) is increased by smaller Particle size (1.7µm). N is also inversely proportional to the square of the peak width(w).so as the particle size decreases to increase N and subsequently Rs, an increase in sensitivity is obtained. Also, peak height(h) is inversely proportional to the peak width(w). H 1/W Equation.4 An increase in sensitivity is obtained, since narrower peaks are taller peaks. Narrower peaks also mean more peak capacity per unit time in gradient separations, desirable for many applications. eg. peptidemapping. N L/dp Equation.5 Efficiency(N) is proportional to column length(l) and inversely proportional to the particle size.(14). Fig 3: van Demeter plot 3. INSTRUMENTATION RRLC analyses were performed on an Agilent 1200 SL HPLC equipped with a binary pump and a micro-vacuum degasser, a multi-wavelength (MW) detector, an auto-sampler, a thermostatted column compartment, and a Luna C 18-HST column (2.5_m,100mm 3.0 mm) from phenomenex (Torrance,CA,US). [11] DESCRIPTON To achieve lowest baseline noise, the standard delay volume configuration is Recommended for the 1200 Series RRLC pump module. Vol 4, Issue 07,
6 The injection volume and the sample dissolution solvent are important. Care must be taken that the compounds are focused at the top of the column, to avoid peak dispersion due to the injection, which would cause a reduced peak height. In order to achieve this, the sample should be dissolved in a solvent composition of lower elution strength than the mobile phase. The column temperature should not be too low, to avoid long retention of the peaks on the column. This also creates peak dispersion and a lower peak Height. Selection of the optimum detector cell depends on the id of the column being used. Typically, the longer the path length, the better the signal-to-noise ratio. The data rate of the UV detector should be selected according to the actual peak width. Higher than necessary data rates should be avoided because of higher noise levels. Available UV detectors are the Agilent 1200 Series Diode Array Detector SL Plus and the new Agilent 1200 Series Variable Wavelength Detector SL, with data rates of 160 Hz and significantly lower noise and drift levels. For highest sensitivity in the UV range, the new Agilent 1200 Series VWD SL Plus is the optimum choice [11] Fig: 4 Instrumentation of RRLC DEGASSER µ-degasser: Maximum baseline stability and minimum noise by continues vacuum source Vacuum Degasser: Flow rate: up to 10ml/min, Internal volume: 12 ml. Micro Degasser: Flow rate: up to 5ml/min, Internal volume: 1ml. [1] Vol 4, Issue 07,
7 Fig 5: Vacuum degasser Micro degasser WORKING OF AUTOSAMPLER IN RRLC During the injection routine of the autosampler, the sample loop, the inside of the needle, the seat capillary and the main channel of the injection valve are in the flow path, and remain there throughout the duration of run. This means. These parts are flushed continuously with mobile phase during the complete analysis. It is only during aspiration of the sample that the injection valve is switched out of the flow path. In this position, the pump effluent is led directly to the column. Prior to injection, the outside surfaces of the needle are washed with fresh solvent. This is achieved using the flush port of the autosampler, and prevents contamination of the needle seat. The flush port of the autosampler is refilled with fresh solvent by a peristaltic pump that is installed in the autosampler housing. The flush port has a volume of about 680 μl, and the pump delivers 6 ml/min. Setting the wash time to 10 seconds means that the flush port volume is refilled more than once with fresh solvent, which is sufficient in most cases to clean the outside of the needle. Flushing and leaning of the autosampler is thus achieved so as to have no (or) zero carry over. [11] Fig6:Autosampler SL injection system Vol 4, Issue 07,
8 PUMPS A steady pump pressure (usually about psi) is needed to ensure reproducibility and accuracy Proper maintenance of the pump system will minimize down time. Commonly used pumps in RRLC are binary pumps and quaternary pumps. BINARY PUMP SL The pump has the facility for mixing two solvents for binary gradients and of course may be used in the isocratic mode. By choosing a system from the fast quaternary/binary LC system through to the ultra-fast rapid resolution system you can achieve new levels of productivity, processing more sample in less time, while maintaining and even improving data quality. Short columns with sub-2 µm particles offer a unique opportunity to dramatically reduce analysis time by increasing the flow rate without losing separation performance. In order to take full advantage of their power, you need a system that can do more than handle high back pressure. The answer is a fully integrated solution, optimized for system volumes, cycle times, detection rates and temperature control. Configurable delay volume (120µl and µl) for uncompromised support of narrow and std bore columns ( mm ID). Standard delay volume ( µl) Low delay volume (120 µl) ml/min flow rate range for highest speed on narrow and standard bore RRHT columns (full P range). 600 bar pressure for higher speed and resolution in RRLC. Electronic damping control for lowest baseline noise Perfect choice for fast and precise gradients using LC/MS, as well as UV-only systems Fully exploits the speed and separation potential of ZORBAX Rapid Resolution HT Columns. Fig 7: Binary Pump SL Vol 4, Issue 07,
9 Binary pump Flow range: ml/min, for fast gradient analysis (column id: ) BENEFITS OF COLUMNS PACKED WITH SUB-2MICROPARTICLES FASTER CHROMATOGRAPHY There are several advantages of having short run times. High Throughput labs now have higher capacity and can analyze more samples in less time, also means lower costs. [1] HIGHER RESOLUTION Long columns packed with smaller particles result in higher efficiency and higher resolution. This is important for analysis of complex samples from metabolomics or proteomics. Also, applications such as impurity profiling can benefit from higher separation powder. Even the LC/MS analysis of drug in biological fluids can benefit from the higher peak capacity, because of the reduced interference from ion suppression. In general, higher separation power provides more confidence in the analytical results. [1] Fig 8: Peak capities of more than 700 can be achieved using a ZORBAX RRHT SB-C 18column (2.1 x 150 mm, 1.8 µm) to analyze tryptic digest of Bs. FRICTIONAL HEATING Forcing mobile phase through the column at higher pressure and higher flow rates generates heat. The resulting temperature gradients have an impact on the column efficiency. POWER= F.p F=Flow rate P = pressure Powerful column Thermostatting (for example, using a water bath) generates a strong radial temperature gradient, which leads to significant loss in column efficiency. Still-air-column Vol 4, Issue 07,
10 thermostatting reduces the radial temperature gradient and therefore reduces the efficiency losses, but a higher column outlet temperature has to be accepted. At lower back-pressure, performance losses due to frictional heat are minimized so that 4.6/3mm inner sub-2-micron columns still deliver superior efficiencies compared with the respective 2.1 mm inner diameter columns [1] COLUMN 1) THERMOSTATTED COLUMN COMPARTMENT SL The heart of the system is the column where separation occurs 100 C for higher speed and resolution in RRLC. New low volume heat exchanger for minimized peak broadening on 2.1mm ID columns. 600 bar 2Ps/10Pt micro valve for minimum cycle times by ACR (alternating column regeneration, which saves wash & regeneration time) PCC (post-column cooling) and improved peltier temperature control (< +/ C) for minimized UV noise under most demanding conditions (< 50µAU at high flow rates and temperatures) Fig 9: Thermostatted Column Compartment Sl 2) COLUMN DIAMETERS Column optimized for 2.1mm, 3 and 4.6 mm I.D are available in RRLC for improving its efficiency. Vol 4, Issue 07,
11 Fig 10: Low delay configuration for 2.1 mm inner diameter columns Fig 11: medium delay volume configuration for 2.1mm Id column with highest Uv sensitivity OPTIMIZED CONFIGURATION FOR 3 AND 4.6 MM I.D. COLUMNS: Fig 12: Standard delay volume configuration for 3 and 4.0mm ID columns with highest UV sensitivity. DETECTORS The detector is used to sense the presence of a compound passing through and to provide an electronic signal to a data-acquisition device. Vol 4, Issue 07,
12 DAD/MWD SL s: (DIODE-ARRAY DETECTOR/MULTIPLE WAVE LENGTH DETECTOR) Up to 100% resolution gain in UFLC by 80Hz data rate Highest sensitivity in UFLC with < 50uAU noise by Low noise cell design and electronics, ETC (Electronic Temperature Control),PCC (Post column cooling using TCC SL) Support of 2.1, 3.0 and 4.6mm ID columns by 3 flow cells Standard: 13 ul / 10 mm, Semi- Micro: 5 ul / 6 mm, Micro: 2 ul / 3 m The Agilent 1200 Series diode-array and multiple-wavelength detectors: Dual lamp design for highest sensitivity from 190 to 950 nm; Programmable slit for easy optimization of sensitivity, linearity and spectral resolution; Low-noise electronics and electronic temperature control for lowest detection limits, even under unstable ambient conditions A range of nine flow cells for highest application flexibility; and Easy upgrade to the 80 Hz sampling rate for high speed separations. Multiple wavelength detector SL for ultra-fast multi-electrical analysis. It has 8signals, 80 HZ data rate. Diode Array Detector SL for multi-electrical analysis. It have 5 signals, 20 HZ data rate [5] VWD SL (VARIABLE WAVE LENGTH DETECTOR SL) VWD SL for ultra-fast, programmable single ë analysis, 1 signal, 13 Hz data rate Maximum resolution in UFLC by 55Hz sampling rate. Three flow cells for support of narrow and std bore columns ( mm ID), Std: 14ul / 10mm, Semi-micro: 5ul / 6mm, Micro: 1ul / 5mm [5] FIG 13: Influence of the detector rate on resolution DAD SL and MWD SL Vol 4, Issue 07,
13 TYPES OF DETECTORS AVAILABLE IN AGILENT SERIES Different UV detector flow cells for use with 2.1, 3.0 and 4.6 mm inner diameter columns are available. Fast UV and MS detectors with data rates up to 160 Hz (1200 Series VWD SL Plus), 80HZ (1200 series DAD SL, MWD SL) and up to 40 HZ for MS application are available. The 1200 Series fluorescence detector (FLD) with 37 Hz data acquisition, and the 1200 series refractive index detector (RID) are also compatible with the Agilent 1200 series RRLC system. A stepwise upgrade from 1100 series to 1200 series RRLC is possible. [9] FIG NO: 14.Diagrammatic representation of differences/changes between HPLC and RRLC Advantages over HPLC The major high - throughput RRLC are the increase in throughput. The reduction in the analysis cost. The shortening in analysis time is due to the use of a shorter column length high resolution separation capability ultra fast and highest analysis speed It is most improve sensitivity High temperature up to 100 C on certain columns allows more selectivity flexibility It is offers true fast LC analysis [11] Vol 4, Issue 07,
14 6. Applications of RRLC RRLC-tandem mass spectrometry method for the determination of endocrine disrupting chemicals (EDCs) examples: Bentazone, salicyliacid, silica gel pharmaceuticals and personal care products (PPCPs) in waste water irrigated soils. Analysis for quality control of Rhodiolarosea roots and commercial standardized products. It is applicable for Herbal produces examples: Panax and Epimedium species It is Applicable for not only pharmaceuticals compound examples: Methanol, vitamins B6, B9, B12, Tri ethylamine but also for chemical compounds example: Acetanilide, Acetophenone, octanophenone, Tubufenozide Scalability as a Function of Column Dimensions Using ZORBAX Rapid Resolution HT Columns for the Analysis of the Pharmaceutical Triamcinolone Impurity Profiling with the Agilent 1200 Series LC System Part 1: Structure Elucidation of Impurities with LC/MS Polycyclic Aromatic Hydrocarbon (PAH) Separations Using ZORBAX Eclipse PAH The High-Resolution Reversed-Phase HPLC Separation of Licorice Root Extracts using Long Rapid resolution HT 1.8-um Columns The Analysis of Benzodiazepines in Hair Using RRHT LC/MS/MS Determination of Benzodiazepines in Oral Fluid Using LC/MS/MS More speed, better resolution and lower LOD using liquid chromatography and fluorescence detection - Comparing the 1100 Series LC to the 1200 Series Rapid Resolution system [9] 7. CONCLUSION In this study, RRLC offers improved run times and increased sensitivity over conventional HPLC based methods. We applied the fast RRLC technique to the analysis of general chemical compounds. RRLC separations could be easily achieved from many traditional HPLC methods. Most applications of these small particle columns have been in the area of pharmaceutical analysis. Using a 1200 RRLC system equipped with higher acquisition rate of detector, low dead volume system configuration and combined with a high-pressure HPLC system with 600 bar pressure capability, the ultrafast RRLC analysis could be done with satisfactory analytical precision. In an isocratic example, one cycle analysis time by traditional HPLC/DAD method that required 25 min could be shortened to a RRLC/MS method to 0.8 min, a factor of 31. The same method transfer approach was carried out in a gradient separation. RRLC technique is useful for not only pharmaceuticals but also for chemical compounds currently analyzed by conventional HPLC on 250mm 4.6 mm i.d. Vol 4, Issue 07,
15 columns. In the future, the fast RRLC technique will be used more widely in other HPLC fields. 8. REFERENCES 1. Angelika Gratzfeld-Huesgen, Michael Frank, Christian Gotenfels. Agilent Series RRLC Optimization Guide., 1200; Chen F, Ying GG, Yang JF, Zhao JL, Wang L, J Environ Sci Health B., 2010 Oct; 45(7): Frank.M, ìachieving fastest analyses with Agilent 1200 Se-rise RRLC System and Application Note Publication Number 2006; EN, March 4. Gerber. F, Krummen. M, Potgeter.H, Roth A. Siffrin. C and Spoendli c.j chromatograph., 2004; A1036: Greibrokk T. and Andersen T., J. Chromatograph, 2003; A1000: Ian Acworth, Darwin Asa, Eddie Goodall1, John Christensen, and Ryan McCarthy Available from: URL: 9. Pickering Laboratories, Application Note, Publ. No B-CA 5, Tatsunari Yoshida, Ronald E. Majors and Hiroki Kumagai, Chromatography, 2007; Vol.28 (No.2). 11. Takakura Cho, Hachioji Shi Tokyo Agilent Technologies, Inc Japan. 12. Agilent Technologies, Inc 2850 Centerville Road Wilmington, Delaware USA. 13. URL: http//schimadzu.co.uk. 14. Van Demeter JJ, Zuiderweg EJ, Klinkenberg A.Longitudinal diffusion resistance to mass transfer as causes of non ideality in chromatography. Chem. Eng. Sci., 1956; 5: Yuan-Chun Maa,b,, Xiao-Qiang Wanga,b, FeiFei Houa, Jie Ma, Mai Luoa, Alice (Chena, Peter Jin, Shane Lua, Iris Xuaa (RRLC) analysis and studies on Journal of Pharmaceutical and Biomedical Analysis., 2011; 54 : Vol 4, Issue 07,
LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note
LC walk-up system using the Agilent 12 Series LC Method Development Solution and Agilent Easy Access software Test of reaction kinetics, column scouting, and impurity checks with one LC system and up to
More informationAgilent 1200 Series Rapid Resolution LC and Rapid Resolution LC/MS Optimization Guide
Agilent 1200 Series Rapid Resolution LC and Rapid Resolution LC/MS Optimization Guide Authors: Angelika Gratzfeld-Huesgen, Michael Frank, Christian Gotenfels Table of Contents Introduction 5 1. Theory
More informationPerformance characteristics of the Agilent 1290 Infinity Quaternary Pump
Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof
More informationHigh Performance Liquid Chromatography
High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).
More informationHigh Resolution Fast LC
High Resolution Fast LC Easier Than You Think Rita Steed LC Columns Application Engineer May 9, 2013 What is High Resolution Fast LC? Maintain Resolution with Faster Run Time Increased Resolution with
More informationApplication Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India
Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis
More informationApplication Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India
Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts
More informationWelcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster
Welcome to our E-Sear: Choosing HPLC Columns for Faster Analysis Smaller and Faster High Throughput/Fast LC Requires. Short columns 0 mm or shorter Small particle sizes. µm Rapid Resolution or new.8 µm
More informationPractical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies
Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies Title Objectives Demonstrate Agilent Method Translation
More informationPerformance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography
Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography Technical Overview 2D-LC Conventional 1D-LC Abstract This Technical Overview presents
More informationSeparation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.
Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28
More informationLuna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency
Luna 2.5 µm C18(2)-HST Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Table of Contents Part 1 Theory 1.1 Abstract...3 1.2 Introduction...3 Part 2 Set Up 2.1
More informationRapid Screening and Analysis of Components in Nonalcoholic Drinks Application
Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Food Author Michael Woodman Agilent Technologies, Inc. 285 Centerville Road Wilgton, DE 1988-161 USA Abstract Soft drinks,
More informationApplication. Authors. Abstract. Introduction. Hydrocarbon Processing
Fast Analysis Method for Rubber Chemical Antidegradants Using Rapid Resolution Liquid Chromatography (RRLC) Systems with Rapid Resolution High Throughput (RRHT) Columns Application Hydrocarbon Processing
More informationUsing the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates
Using the Agilent 9 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates Technical Overview Author Florian Rieck Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical
More informationApplication Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.
Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 126 Infinity Hybrid SFC/UHPLC System Improving peak shape and providing wider UV selectivity
More informationTransferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system
Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system Technical Note Introduction The HP 1090 Series HPLC system, introduced in 1983, is regarded to be an outstanding
More informationPolymer analysis by GPC-SEC. Technical Note. Introduction
Polymer analysis by GPC-SEC Technical Note Introduction Gel Permeation Chromatography (GPC), also referred to as Size Exclusion Chromatography (SEC) is a mode of liquid chromatography in which the components
More informationApplication Note. Edgar Naegele. Abstract
Agilent MassHunter Fast computer aided analysis of LC/ESI-TF data from complex natural product extracts Part 1: Analysis of Agilent 6210 TF data with the Molecular Feature Extractor in MassHunter Workstation
More informationInstrumental Chemical Analysis
L2 Page1 Instrumental Chemical Analysis Chromatography (General aspects of chromatography) Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 2 nd semester,
More informationLC and LC/MS Column Selection Flow Chart
LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the
More informationHPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25
High Performance Liquid Chromatography (HPLC) Harris Chapter 25 12/1/2005 Chem 253 - Chapter 25 1 HPLC Separation of nonvolatile or thermally unstable compounds. If the analyte/sample can be found to be
More informationMethod Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions
Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions Today, ultra-high-performance liquid chromatography (UHPLC) has taken a firm foothold in the analytical laboratory.
More informationIntroduction to Chromatography
Introduction to Chromatography Dr. Sana Mustafa Assistant Professor Department of Chemistry, Federal Urdu University of Arts, Science & Technology, Karachi. What is Chromatography? Derived from the Greek
More informationLC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography
LC III: HPLC What is HPLC? Originally referred to as High-Pressure Liquid Chromatography Now more commonly called High Performance Liquid Chromatography In general: The instrument controlled version of
More informationMaximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System
Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and
More informationLC Technical Information
LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC
More informationAutomated Switching Between 1D-LC and Comprehensive 2D-LC Analysis
Automated Switching Between D-LC and Comprehensive D-LC Analysis The Agilent 90 Infinity D-LC Solution Technical Overview Author Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Abstract This
More informationAnalysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector
Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent Infinity LC System with Diode Array Detector Application Note Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany [mau]
More informationHigh Pressure/Performance Liquid Chromatography (HPLC)
High Pressure/Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) is a form of column chromatography that pumps a sample mixture or analyte in a solvent (known as the
More informationOpen Column Chromatography, GC, TLC, and HPLC
Open Column Chromatography, GC, TLC, and HPLC Murphy, B. (2017). Introduction to Chromatography: Lecture 1. Lecture presented at PHAR 423 Lecture in UIC College of Pharmacy, Chicago. USES OF CHROMATOGRAPHY
More informationIntroduction to Chromatographic Separations
Introduction to Chromatographic Separations Analysis of complex samples usually involves previous separation prior to compound determination. Two main separation methods instrumentation are available:
More informationImprove Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application
Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application Pharmaceuticals Authors John W. Henderson Jr., Nona Martone, and Cliff Woodward
More informationLiquid Chromatography
Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors
More informationChemistry Instrumental Analysis Lecture 31. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 31 High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) Solvent Delivery
More informationMODERN HPLC FOR PRACTICING SCIENTISTS
MODERN HPLC FOR PRACTICING SCIENTISTS Michael W. Dong Synomics Pharmaceutical Services, LLC Wareham, Massachusetts WILEY- INTERSCIENCE A JOHN WILEY & SONS, INC., PUBLICATION Preface xv 1 Introduction 1
More informationExceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens
Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-exyl Columns to Separate Estrogens Application Note Pharmaceutical, Environmental Authors John W. enderson Jr. and William J. Long Agilent
More informationDetermination of Caffeine by HPLC
Determination of Caffeine by HPLC Introduction It was a long history before real high performance liquid chromatography (HPLC) had evolved. The very first indication of a chromatographic separation was
More informationPeptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen
Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters Application Angelika Gratzfeld-Huesgen Abstract This application has been verified using an Agilent
More informationLuminescence transitions. Fluorescence spectroscopy
Luminescence transitions Fluorescence spectroscopy Advantages: High sensitivity (single molecule detection!) Measuring increment in signal against a dark (zero) background Emission is proportional to excitation
More informationHPLC Workshop 16 June 2009 What does this do? Chromatography Theory Review Several chromatographic techniques Even though each method utilizes different techniques to separate compounds, the principles
More informationDeveloping a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note
Developing a fast, generic method for rapid resolution LC with quadrupole MS detection Combining the Agilent 6140 quadrupole MS and multimode source with the Agilent 1200 Series Rapid Resolution LC system
More informationION CHROMATOGRAPHY SYSTEM S 150
ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved
More informationLECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools
LECTURE 2 Advanced Separation Science Techniques Present and Future Separation Tools Jack Henion, Ph.D. Emeritus Professor, Analytical Toxicology Cornell University Ithaca, NY 14850 Lecture 2, Page 1 Contents
More informationThe Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation
Technical Overview The Agilent InfinityLab D-LC Solution with Active Solvent Modulation Achieving Improved Resolution and Sensitivity for Challenging Combinations of Separation Conditions Author Sonja
More informationAnalysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns
Analysis of Saponins in Spine Date Seed using Agilent ZRBAX Solvent Saver T (.8 µm) Columns Application Note Food and Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies, Inc. 4 Ying Lun
More informationPacked Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents
No. 089 SEPARATION REPORT Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS Table of Contents 1. Introduction 1 2. Column Specification 1 3. Features of Packing Materials
More informationHPLC Background Chem 250 F 2008 Page 1 of 24
HPLC Background Chem 250 F 2008 Page 1 of 24 Outline: General and descriptive aspects of chromatographic retention and separation: phenomenological k, efficiency, selectivity. Quantitative description
More informationInstrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A.
Instrumental Analysis II Course Code: CH3109 Chromatographic &Thermal Methods of Analysis Part 1: General Introduction Prof. Tarek A. Fayed What is chemical analysis? Qualitative analysis (1) Chemical
More informationChromatography. Gas Chromatography
Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture
More informationThe Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times
The Secrets of Rapid HPLC Method Development Choosing Columns for Rapid Method Development and Short Analysis Times Rapid Analysis Is More Than Run Time It is developing a method to meet a goal and developing
More informationHPLC Praktikum Skript
HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,
More informationBiochemistry. Biochemical Techniques HPLC
Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know
More informationEasy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator
Technical Note 75 Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator Introduction The goal of every chromatographic optimization is a method that sufficiently resolves all
More informationAgilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity
Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule
More informationDetermination of Polymer Modifier in Asphalt
Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.
More informationFast methods for the determination of ibuprofen in drug products
APPLICATION NOTE 779 Fast s for the determination of ibuprofen in drug products Authors Sylvia Grosse, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Germering, Germany Keywords Pharmaceutical,
More informationImpact of Instrument Dispersion on Performance of HPLC Capillary Columns
Impact of Instrument Dispersion on Performance of HPLC Capillary Columns Wendy Roe, Richard A. Henry, and Hillel Brandes Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA T413131 sigma-aldrich.com/analytical
More informationChromatographic Separation
What is? is the ability to separate molecules using partitioning characteristics of molecule to remain in a stationary phase versus a mobile phase. Once a molecule is separated from the mixture, it can
More informationChapter content. Reference
Chapter 7 HPLC Instrumental Analysis Rezaul Karim Environmental Science and Technology Jessore University of Science and Technology Chapter content Liquid Chromatography (LC); Scope; Principles Instrumentation;
More informationChromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?
Chromatography- Separation of mixtures CHEM 212 What is solvent extraction and what is it commonly used for? How does solvent extraction work? Write the partitioning coefficient for the following reaction:
More informationHigh Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension
High Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension Minnesota Chromatography Forum Spring Symposium Dwight Stoll and Peter W. Carr Department of Chemistry University
More informationLIQUID CHROMATOGRAPHY
LIQUID CHROMATOGRAPHY RECENT TECHNIQUES HPLC High Performance Liquid Chromatography RRLC Rapid Resolution Liquid Chromatography UPLC Ultra Performance Liquid Chromatography UHPLC Ultra High Pressure Liquid
More informationAnalytical Chemistry
Analytical Chemistry Chromatographic Separations KAM021 2016 Dr. A. Jesorka, 6112, aldo@chalmers.se Introduction to Chromatographic Separations Theory of Separations -Chromatography Terms Summary: Chromatography
More informationComprehensive 2D-LC Analysis of Chinese Herbal Medicine
Comprehensive 2D-LC Analysis of Chinese Herbal Medicine The Agilent 1290 Infinity 2D-LC Solution Application Note Small Molecule Pharmaceuticals & Generics Author Sonja Krieger Agilent Technologies, Inc.
More informationUltrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system
Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 129 Infinity LC system Application Note Food and Beverages Author Srividya Kailasam
More informationAssay Transfer from HPLC to UPLC for Higher Analysis Throughput
MAY 2005 SEPARATION SCIENCE REDEFINED 31 Assay Transfer from HPLC to UPLC for Higher Analysis Throughput A typical HPLC assay was transferred and optimized for a Waters ACQUITY UPLC system to achieve both
More informationChem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS
Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable
More informationAnalysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system
Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system Application Note Food Testing & Agriculture Authors Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Keiko
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationCHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.
CHROMATOGRAPHY The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments. THEORY OF CHROMATOGRAPHY: Separation of two sample components in
More informationApplication Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India
QbD Based Method Development on an Agilent 129 Infinity UHPLC System Combined with a Seamless Method Transfer to HPLC Using Intelligent System Emulation Technology Application Note Pharmaceutical QA/QC
More informationApplication Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China
Rapid Deteration of Eight Related Aromatic Acids in the p-phthalic Acid Mother Liquid Using an Agilent 126 Infinity LC System and an Agilent Poroshell 12 SB-C18 Column Application Note Author Xinlei Yang
More informationHigh Performance Liquid Chromatography
STANDARDBASE techniques: High Performance Liquid Chromatography Drenthe College, The Netherlands 1. Introduction HPLC. High Performance Liquid Chromatography High Performance Liquid Chromatography (HPLC)
More informationModernizing the USP Monograph for Acetaminophen
Modernizing the USP Monograph for cetaophen pplication Note Small Molecule Pharmaceuticals and Generics uthor Rongjie Fu gilent Technologies (Shanghai) Co. Ltd bstract new HPLC method was developed for
More informationLC Column Troubleshooting. Do you have an equivalent column for my LC column?
LC Column Troubleshooting Do you have an equivalent column for my LC column? LC columns from different suppliers may have very different retention properties even if the bonding is the same. Although there
More informationChromTech HPLC System
ChromTech HPLC System High-prssure Gradient System (A) High-pressure pump: 2 pcs UV/VIS detector: 1 piece Gradient control module: 1 unit High-pressure gradient mixer: 1 piece Single channel chromatography
More informationQuantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection
Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,
More informationProduct Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns
Product Bulletin HPLC Columns ACE LC/MS and Rapid Analysis HPLC Columns 0 mm, 30 mm, 35 mm and 50 mm column lengths.0,., 3.0, 4.0 and 4.6 mm column diameters Configured for High Sample Throughput Specially
More informationAgilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases
Agilent 129 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract The Agilent
More informationPackings for HPLC. Packings for HPLC
Summary of packings for HPLC In analytical HPLC, packings with particle sizes of 3 to 10 µm are preferred. For preparative separation tasks, also particles with diameters larger than 10 µm are applied.
More informationPRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC
PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC CHROMATOGRAPHY Laboratory technique for the Separation of mixtures Chroma -"color" and graphein
More informationRapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS
Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent
More informationChemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets
Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets Liquid chromatography was developed by Tswett in early 1900 s and was shown to be a powerful separation
More informationCourse goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography
Emqal module: M0925 - Quality parameters and optimization in is a separation technique used for quantification of mixtures of analytes Svein.mjos@kj.uib.no Exercises and lectures can be found at www.chrombox.org/emq
More informationAgilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX
Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Technical Note Agilent s SampliQ WAX provides Applications for strongly acidic, acidic and neutral compounds Excellent
More informationPrinciples of Gas- Chromatography (GC)
Principles of Gas- Chromatography (GC) Mohammed N. Sabir January 2017 10-Jan-17 1 GC is a chromatographic technique utilizes gas as the mobile phase which is usually an inert gas (Hydrogen, Helium, Nitrogen
More informationLEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.
LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT Thomas Wenzel, Bates College In-class Problem Set Extraction Problem #1 1. Devise a scheme to be able to isolate organic acids, bases
More informationChapter 23 Introduction to Analytical Separations
Chapter 23 Introduction to Analytical Separations Homework Due Monday April 24 Problems 23-1, 23-2, 23-7, 23-15, 23-27, 23-29, 23-32 Analytical Separations: Universal approach to analyzing complex mixtures
More informationApplication Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA
Fast Method Development of Salicylic Acid Process Impurities using Agilent ZRBAX Rapid Resolution High Throughput Columns with the Agilent 12 Series Method Development Solution Controlled by AutoChrom
More informationMassHunter METLIN Metabolite PCD/PCDL Quick Start Guide
MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide What is the MassHunter METLIN Metabolite PCD/PCDL? 2 Where to find more information 2 Kit Content 3 Installation 5 Before you start 5 Install MassHunter
More informationLC800. Smart HPLC. Until now UHPLC From now Smart HPLC
LC800 Smart HPLC Until now UHPLC From now Smart HPLC Smart HPLC Leads to Ultimate Performance Patent Pending S LC800 is a completely new and unique concept, designed for maximum performance in high resolution
More informationBasic Principles for Purification Using Supercritical Fluid Chromatography
Basic Principles for Purification Using Supercritical Fluid Chromatography Jo-Ann M. Jablonski, Christopher J. Hudalla, Kenneth J. Fountain, Steven M. Collier, and Damian Morrison Waters Corporation, Milford,
More informationApplication. Introduction. Authors. Abstract. Hydrocarbon Processing
Fast Analysis of Phenolic Antioxidants and Erucamide Slip Additives in Polypropylene Homopolymer Formulations Using Rapid Resolution Liquid hromatography (RRL) with Rapid Resolution High Throughput (RRHT)
More informationAnalysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS
Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Application Note Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies
More informationHow proteins separate on reverse-phase HPLC
1 Reverse Phase How proteins separate on reverse-phase HPLC RP chromatography separates proteins through the interaction of the hydrophobic foot of the protein with a nonpolar surface of the particle RP
More informationSuccessfully Scaling and Transferring HPLC and UPLC Methods
Successfully Scaling and Transferring HPLC and UPLC Methods Esa Lehtorinne Esa_Lehtorinne@waters.com Tel: +358-9-5659 6288 Fax: +358-9-5659 6282 Waters Finland Kutomotie 16 00380 Helsinki 2013 Waters Corporation
More informationFigure 1. Structures for Vitamin B2 and Vitamin B1.
CH 461 & CH 461H F 18 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment
More informationpenta-hilic UHPLC COLUMNS
penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and
More information