1 Chapter 5 Errors in Chemical Analysis 熊同銘 tmhsiung@gmail.com Slide 1 of 19 Contents in Chapter Accuracy, Precision and Bias 5.2 Types of Errors in Experimental Data 5.3 Systematic Error Slide 2 of 19 1
2 5.1 Accuracy, Precision and Bias Precise Imprecise Unbiased Biased Accurate = Precise + Unbias Slide 3 of 19 Which one is accurate? (A) (B) μ μ (C) μ (D) μ μ: true value Slide 4 of 19 2
3 Bias: A measure of how far the analytical result is from the true value caused by systematic error(s). Precision: A measure of how similar the replicate measurements to each other caused by random error(s), NOT how close to the true value. Accuracy: A combination of precision and bias, which describe the degree of agreement between the analytical result and the true value. Slide 5 of The mean and Median Mean x N i 1 N x i Median: The central value in a set of replicate measurements. (for even number of data points, the median is the average of central pair) * The median is used advantageously when a set of data contain an outlier, a result that differs significantly from others in the set. Slide 6 of 19 3
4 5.1.2 Evaluation of Precision Standard deviation (s) s N i 1 ( x x) i N 1 2 Relative standard deviation (RSD) (also called coefficient of variation, CV): s RSD 100% x Variance (v) v = s 2 Slide 7 of Evaluation of Accuracy Absolute error E x i x t x t : true or accepted value Relative error E r x i x x t t 100% Slide 8 of 19 4
5 5.2 Types of Errors in Experimental Data Systematic (Determinate) error: Errors that have a known source, they affect accuracy, measurements in only one way(either positive or negative). It can be avoided, detected, and corrected. (Section 5.3) Random (Indeterminate or Statistic) error: Uncertainties resulting from fluctuation of uncontrolled variables and it can not be attributed to any cause. (Chapter 6) Gross error: An occasional error, neither random nor systematic, that results in occurrence of a questionable outlier result. Slide 9 of Systematic Error Sources of Systematic Errors (1) Instrumental errors: caused by nonideal instrument behavior, by faulty calibrations, or by use under inappropriate conditions. For example: A 100 ml volumetric flask, which exact volume is ml.. Slide 10 of 19 5
6 (2) Personal errors: result from the carelessness, inattention, or personal limitations of the experimenter. For example: The ability to see a change in the color of an indicator at the end point of a titration. Consistently overestimating or underestimating the value on an instrument s readout scale. Failing to calibrate glassware and instrumentation. Misinterpreting procedural directions. Slide 11 of 19 (3) Method errors: arise from nonideal chemical or physical behavior of analytical systems. Classical analytical method for example: More or less excess of titrant required for color change in titrimetry by indicator (titration error). Instrumental analytical method for example: S meas = kc A + S blank - Sensitivity (proportionality constant), k, may be incorrect because of the matrix interferences. - Reagents impurities and contamination may introduce the signal of blank (S blank ). Slide 12 of 19 6
7 5.3.2 The Effect of Systematic Errors on Analytical Results Constant (systematic) error: Those systematic errors they are independent of the size of the sample being analyzed. Proportional (systematic) error: Those systematic errors they decrease or increase in proportion to the size of the sample being analyzed. Slide 13 of 19 (1) Method errors on Instrumental analytical method for example: S meas = kc A + S blank - the blank signal, S blank, is an example of a constant determinate error. - The proportionality constant k may be affected by proportional errors. * Blank determination: The process of performing all steps of an analysis in the absence of analyte. Blank determinations reveal errors due to interfering contaminants from the reagents and vessels employed in the analysis. * Sample matrix: The whole constituents of the sample other than analyte. Difference in sample matrix may affect the proportionality constant. Slide 14 of 19 7
8 (2) A simulation of constant error and proportional error Theoretical estimation of constant error Theoretical estimation of proportional error Slide 15 of Detecting and Minimizing Systematic Errors (1) For Instrumental Errors: Periodic calibration of the equipment (2) For Personal Errors: check instrument readings, notebook entries, and calculations systematically. carefully choosing the analytical method or using an automated procedure. self discipline with the assistance of duplicate analysis and interlaboratory comparison. Slide 16 of 19 8
9 (3) For Method Errors: Analysis of standard reference materials (SRMs) Independent analysis with another reliable method Analysis with the variation in sample size Blank determination Spike analysis Slide 17 of 19 Homework (Due 2018/10/4) 1. Explain the difference between (a) precision and accuracy (b) random error and systematic error (c) constant error and proportional error (d) absolute error and relative error (e) mean and median 2. Name three types of systematic error. 3. Describe at least three examples of systematic errors that might occur while weighing a solid chemical on an analytical balance. 4. Describe at least three examples of systematic errors that might occur while using a pipet to transfer a known volume of liquid. 5. The color change of a chemical indicator requires an overtitration of 0.03 ml. What is the smallest volume of titrant such that the relative error is less than 0.1%? Slide 18 of 19 9
10 End of Chapter 5 Slide 19 of 19 10
Source: Chapter 5: Errors in Chemical Analyses
Source: Chapter 5: Errors in Chemical Analyses Measurements invariably involve errors and uncertainties. it is impossible to perform a chemical analysis that is totally free of errors or uncertainties
More informationHarris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR Chapter 3. Experimental Error -There is error associated with every measurement. -There is no way to measure the true
More informationHarris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR Chapter 3. Experimental Error -There is error associated with every measurement. -There is no way to measure the true
More informationStatistics: Error (Chpt. 5)
Statistics: Error (Chpt. 5) Always some amount of error in every analysis (How much can you tolerate?) We examine error in our measurements to know reliably that a given amount of analyte is in the sample
More information-However, this definition can be expanded to include: biology (biometrics), environmental science (environmetrics), economics (econometrics).
Chemometrics Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information. -However, this definition can be expanded to include: biology (biometrics), environmental
More informationChemistry Calibration of a Pipet and Acid Titration
Chemistry 3200 Today you are given a chance to brush up on some of the techniques that you will be using during the remainder of the semester. Lab grades will be based on obtaining the correct answer in
More informationEstimation of Parameters
CHAPTER Probability, Statistics, and Reliability for Engineers and Scientists FUNDAMENTALS OF STATISTICAL ANALYSIS Second Edition A. J. Clark School of Engineering Department of Civil and Environmental
More informationADVANCED ANALYTICAL LAB TECH (Lecture) CHM
ADVANCED ANALYTICAL LAB TECH (Lecture) CHM 4130-0001 Spring 2013 Professor Andres D. Campiglia Textbook: Principles of Instrumental Analysis Skoog, Holler and Crouch, 5 th Edition, 6 th Edition or newest
More informationData Analysis I. CU- Boulder CHEM-4181 Instrumental Analysis Laboratory. Prof. Jose-Luis Jimenez Spring 2007
Data Analysis I CU- Boulder CHEM-4181 Instrumental Analysis Laboratory Prof. Jose-Luis Jimenez Spring 2007 Presentation will be posted on course web page based on lab manual, Skoog, web links 5 Objective
More information03.1 Experimental Error
03.1 Experimental Error Problems: 15, 18, 20 Dr. Fred Omega Garces Chemistry 251 Miramar College 1 Making a measurement In general, the uncertainty of a measurement is determined by the precision of the
More informationWhy the fuss about measurements and precision?
Introduction In this tutorial you will learn the definitions, rules and techniques needed to record measurements in the laboratory to the proper precision (significant figures). You should also develop
More informationCh 3. EXPERIMENTAL ERROR
Ch 3. EXPERIMENTAL ERROR 3.1 Measurement data how accurate? TRUE VALUE? No way to obtain the only way is approaching toward the true value. (how reliable?) How ACCURATE How REPRODUCIBLE accuracy precision
More informationMAFF VALIDATED METHODS FOR THE ANALYSIS OF FOODSTUFFS MAFF VALIDATED METHOD V19 APRIL 1992 METHOD FOR ACIDITY IN HONEY
MAFF VALIDATED METHODS FOR THE ANALYSIS OF FOODSTUFFS MAFF VALIDATED METHOD V19 APRIL 1992 METHOD FOR ACIDITY IN HONEY Also published in the Journal of the Association of Public Analysts, 28, 171-175 Correspondence
More informationMETHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)
METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE) 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to ground water, drinking, surface, and saline waters, and domestic and industrial wastes.
More informationChem 321 Lecture 4 - Experimental Errors and Statistics 9/5/13
Chem 321 Lecture 4 - Experimental Errors and Statistics 9/5/13 Student Learning Objectives Experimental Errors and Statistics The tolerances noted for volumetric glassware represent the accuracy associated
More informationLecture 3: Chapter 3
Lecture 3: Chapter 3 C C Moxley UAB Mathematics 26 January 16 3.2 Measurements of Center Statistics involves describing data sets and inferring things about them. The first step in understanding a set
More informationLaboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015
Laboratory Techniques 100: Back To Basics Carol Injasoulian Lab Manager City of Bay City April 29,2015 QA/QC for Environmental Measurement What is quality assurance (QA)? Broad plan to maintain quality
More informationChapter 8: Sampling, Standardization, and Calibration
Chapter 8: Sampling, Standardization, and Calibration A chemical analysis uses only a small fraction of the available sample, the process of sampling is a very important operation. Knowing how much sample
More informationProposed Procedures for Determining the Method Detection Limit and Minimum Level
Proposed Procedures for Determining the Method Detection Limit and Minimum Level Published by: ACIL Environmental Services Section Technical Committee Revision 3.0 3/8/006 PROCEDURES These procedures set
More informationCompulsory textbook. Recommended textbooks
Compulsory textbook Fundamentals of Analytical Chemistry, D. A. Skoog, D. M. West, F. J. Holler and S. R. Crouch, Brooks/Cole, 2004, 8th Edition Recommended textbooks Principles of Instrumental Analysis,
More informationNaOH (aq) + HCl (aq) NaCl (aq) + H 2 O (l)
EXPERIMENT 21 Molarity of a Hydrochloric Acid Solution by Titration INTRODUCTION Volumetric analysis is a general term meaning any method in which a volume measurement is the critical operation; however,
More informationChapter 13. Titrations in Analytical Chemistry
Chapter 13 Titrations in Analytical Chemistry Titrations in Analytical Chemistry Titration methods are based on determining the quantity of a reagent of known concentration that is required to react completely
More informationALLOWAY METHOD OUTLINE
ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:
More informationGlossary of Common Laboratory Terms
Accuracy A measure of how close a measured value is to the true value. Assessed by means of percent recovery of spikes and standards. Aerobic Atmospheric or dissolved oxygen is available. Aliquot A measured
More informationMETHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)
METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) 1.0 SCOPE AND APPLICATION 1.1 Method 7196 is used to determine the concentration of dissolved hexavalent chromium [Cr(VI)] in EP/TCLP characteristic extracts
More informationSaleh Nasiri (Ph.D.) IBTO-Research Center
Quality and safety improvement of blood component products in blood transfusion by using statistical methods and control charts in screening test results Saleh Nasiri (Ph.D.) IBTO-Research Center What
More informationPropagation of Uncertainty
76 nalytical Chemistry 2.1 lthough we will not derive or further justify the rules presented in this section, you may consult this chapter s additional resources for references that discuss the propagation
More informationANALYTICAL CHEMISTRY 1 LECTURE NOTES
ANALYTICAL CHEMISTRY 1 LECTURE NOTES FUNDAMENTALS OF PRE ANALYSES TOPIC 1: Theory of Errors 1.0 Introduction Analytical chemistry is a specialised aspect of chemistry that deals with both qualitative analysis
More informationMeasurements Chapter 3
Measurements Chapter 3 Analytical Chemistry is the science of chemical measurement. Its object is the generation, treatment and evaluation of signals from which information is obtained on the composition
More informationQuality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE
Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE Use objectives: state purpose for which results will be used. Specifications: How good should the numbers be
More informationTopic 2 Measurement and Calculations in Chemistry
Topic Measurement and Calculations in Chemistry Nature of Measurement Quantitative observation consisting of two parts. number scale (unit) Examples 0 grams 6.63 10 34 joule seconds The Fundamental SI
More informationCHEM 3420 /7420G Instrumental Analysis
CHEM 3420 /7420G Instrumental Analysis Prof. Brian Gibney 2411 Ingersoll bgibney@brooklyn.cuny.edu Course Introduction Prof. Brian R. Gibney B.S. Chemistry (ACS Certified) Ph.D. Chemistry Brooklyn College,
More informationAnd how to do them. Denise L Seman City of Youngstown
And how to do them Denise L Seman City of Youngstown Quality Control (QC) is defined as the process of detecting analytical errors to ensure both reliability and accuracy of the data generated QC can be
More informationMercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution
Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table
More informationUnit 3 Chemistry - Volumetric Analysis
Unit 3 Chemistry Volumetric Analysis Volumetric analysis is a quantitative chemical analysis used to determine the unknown concentration of one reactant [the analyte] by measuring the volume of another
More informationRATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION
Rate Law Determination of Crystal Violet Hydroxylation Revised 5/22/12 RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Adapted from "Chemistry with Computers" Vernier Software, Portland OR, 1997
More informationChapter 3 Experimental Error
Chapter 3 Experimental Error Homework Due Friday January 27 Problems: 3-2, 3-5, 3-9, 3-10, 3-11, 3-12, 3-14, 3-19 Chapter 3 Experimental Error Uncertainties They are everywhere!! We need to learn to understand
More informationProtocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)
Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations
More informationComparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011
Comparing the MICROLAB 600 to tric Glassware and Air Displacement Pipettes June 2011 tric Glassware and Air Displacement Pipettes Table of Contents ``Abstract...3 ``Introduction...4 ``Method & Results...5
More informationCALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu
CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS A.Gnanavelu UNCERTAINTY (U) Every measurement has error which is unknown and unknowable. This unknown error is called measurement uncertainty. Types of Error
More informationWet Chemical Analysis: Volumetric. Wet Chemical Analysis: Volumetric. Wet Chemical Analysis: Volumetric. Volumetric analysis
hemical Measurement Wet hemical Analysis: Volumetric Quantitative Analysis Quantitative analysis seeks to quantify the chemical species in a sample hapter 1 Broadly speaking, an analysis can also aim to
More informationELISA QUALITY ASSURANCE Analytical Phase
LOGO ELISA QUALITY ASSURANCE Analytical Phase - DR. ALI MIRJALILI QUALITY ASSURANCE Dept. PISHTAZ TEB DIAGNOSTICS 01/02/92 Quality improvement congress Definitions: Quality Assurance Quality Control:-
More informationFerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV
Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),
More informationRATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION
Rate Law Determination of Crystal Violet Hydroxylation Revised 10/21/14 RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Adapted from "Chemistry with Computers" Vernier Software, Portland OR, 1997
More information11.1 Uncertainty and error in measurement (1 Hour) 11.2 Uncertainties in calculated results (0.5 Hour) 11.3 Graphical techniques (0.
Chapter 11 Measurement and Data Processing Page 1 Students are to read and complete any part that requires answers and will submit this assignment on the first day of class. You may use internet sources
More informationRevision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:
ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;
More informationMETHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)
METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) 1.0 SCOPE AND APPLICATION 1.1 This automated method is applicable to ground water, drinking and surface waters, and domestic and industrial wastes
More informationWhat is measurement uncertainty?
What is measurement uncertainty? What is measurement uncertainty? Introduction Whenever a measurement is made, the result obtained is only an estimate of the true value of the property being measured.
More informationCHAPTER 3a. INTRODUCTION TO NUMERICAL METHODS
CHAPTER 3a. INTRODUCTION TO NUMERICAL METHODS A. J. Clark School of Engineering Department of Civil and Environmental Engineering by Dr. Ibrahim A. Assakkaf Spring 1 ENCE 3 - Computation in Civil Engineering
More informationPowder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)
Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.
More information3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement
3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing, but also as an energy carrier in the cooling or heating cycle. As vast amounts
More information1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction
1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)
More informationChem 2115 Experiment #7. Volumetric Analysis & Consumer Chemistry Standardization of an unknown solution and the analysis of antacid tablets
Chem 2115 Experiment #7 Volumetric Analysis & Consumer Chemistry Standardization of an unknown solution and the analysis of antacid tablets OBJECTIVE: The goals of this experiment are to learn titration
More informationDetermination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS
Determination of Total Bromine and Iodine Page 2 of 5 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes an inductively coupled plasma optical emission spectrometry method
More information2.3 Estimating PDFs and PDF Parameters
.3 Estimating PDFs and PDF Parameters estimating means - discrete and continuous estimating variance using a known mean estimating variance with an estimated mean estimating a discrete pdf estimating a
More informationDetermination of Orthophosphate Ion
Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the
More informationExperiment 1 - Mass, Volume and Graphing
Experiment 1 - Mass, Volume and Graphing In chemistry, as in many other sciences, a major part of the laboratory experience involves taking measurements and then calculating quantities from the results
More informationComparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013
Comparing the Microlab 600 to tric Glassware and Air Displacement Pipettes January 2013 White Paper: Comparing the Microlab 600 to tric Table of Contents Abstract... 3 Introduction... 4 Methods & Results...
More informationPowder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)
Silica DOC316.53.01132 Heteropoly Blue Method 1 Method 8186 0.010 to 1.600 mg/l SiO 2 (LR, spectrophotometers) Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Scope and application: For boiler
More informationGETTING THE END POINT TO APPROXIMATE. Two hours
Chem 1312 Handout Experiment ONE Laboratory Time Required Special Equipment and Supplies Objective Safety First Aid GETTING THE END POINT TO APPROXIMATE THE EQUIVALENCE POINT Two hours Balance Potassium
More informationMethod Validation. Role of Validation. Two levels. Flow of method validation. Method selection
Role of Validation Method Validation An overview Confirms the fitness for purpose of a particular analytical method. ISO definition: Conformation by examination and provision of objective evidence that
More informationVAM Project Development and Harmonisation of Measurement Uncertainty Principles
VAM Project 3.2.1 Development and Harmonisation of Measurement Uncertainty Principles Part (d): Protocol for uncertainty evaluation from validation data V J Barwick and S L R Ellison January 2000 LGC/VAM/1998/088
More informationAssessment of Accuracy and Precision
2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter
More informationMeasurements, Sig Figs and Graphing
Measurements, Sig Figs and Graphing Chem 1A Laboratory #1 Chemists as Control Freaks Precision: How close together Accuracy: How close to the true value Accurate Measurements g Knowledge Knowledge g Power
More informationA Volumetric Analysis Lab Report Answers
We have made it easy for you to find a PDF Ebooks without any digging. And by having access to our ebooks online or by storing it on your computer, you have convenient answers with a volumetric analysis
More informationSchedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)
Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday
More informationDR/4000 PROCEDURE NITRATE. Using Powder Pillows
DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit
More informationEXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy
EXPERIMENT 7 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium
More informationCadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows
Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationPhenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator
Alkalinity DOC316.53.01166 Phenolphthalein and Total Alkalinity Method 8203 10 to 4000 mg/l as CaCO 3 Digital Titrator Scope and application: For water, wastewater and seawater. Test preparation Before
More informationChapter 13. Titrations in Analytical Chemistry
Chapter 13 Titrations in Analytical Chemistry Titrations are based on a reaction between the analyte ( 分析物 ) and a standard reagent known as titrant ( 滴定液 ). 1 Titration methods (titrimetric methods) are
More informationLecture 3. - all digits that are certain plus one which contains some uncertainty are said to be significant figures
Lecture 3 SIGNIFICANT FIGURES e.g. - all digits that are certain plus one which contains some uncertainty are said to be significant figures 10.07 ml 0.1007 L 4 significant figures 0.10070 L 5 significant
More informationTitration of HCl with Sodium Hydroxide
Titration of HCl with Sodium Hydroxide Lab Report for the Subject of Advanced Chemistry Anon Durongpisitkul, Karis Katekovit, Varun Saketharam,Thanon Thamvorapol, Chanon Anektanasup- January 28, 2017 1
More informationDirect Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry
Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.
More informationIdentification Of The Common Laboratory Glassware, Pipettes And Equipment. BCH 312 [Practical]
Identification Of The Common Laboratory Glassware, Pipettes And Equipment BCH 312 [Practical] (1) Identification of the common laboratory glassware : Conical flasks and beakers: Graduated cylinders Volumetric
More informationSpectrophotometric Determination of Ferrocyanide in Effluents
Spectrophotometric Determination of Ferrocyanide in Effluents ECN-0025-1 INTRODUCTION This method is used to determine the concentration of ferrocyanide ion in photoprocessing solution effluents. The ion
More informationBiol/Chem 4900/4912. Forensic Internship Lecture 5
Biol/Chem 4900/4912 Forensic Internship Lecture 5 Quality Assurance/ Quality Control Quality Assurance A set of activities that ensures that development and/or maintenance processes are adequate in order
More informationCHEM 334 Quantitative Analysis Laboratory
Calibration of Volumetric Glassware Introduction Volumetric glassware is a class of glass vessels that are calibrated to contain or deliver certain volumes of substances. Graduated cylinders, pipettes
More informationExperiment 5: Determining the Stoichiometry and Products of a Redox Reaction
Experiment 5: Determining the Stoichiometry and Products of a Redox Reaction Reading: Chapter sections 4.4-4.6 and 20.1-20.2 in your course text and this lab handout Ongoing Learning Goals: To use a scientific
More informationChemistry Lab: Introduction to Measurement
Name Hour Chemistry Lab: Introduction to Measurement (adapted from Flinn ChemTopic Labs) Introduction Much of what we know about the physical world has been obtained from measurements made in the laboratory.
More informationObjectives. Concepts in Quality Control Data Management. At the conclusion, the participant should be able to:
Concepts in Quality Control Data Management Bio-Rad Laboratories Quality System Specialist Thomasine Newby Objectives At the conclusion, the participant should be able to: Describe how laboratory quality
More informationDetermination of Orthophosphate Ion
Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the
More informationChem 2115 Experiment #7. Volumetric Analysis & Consumer Chemistry Standardization of an unknown solution, analysis of vinegar & antacid tablets
Chem 2115 Experiment #7 Volumetric Analysis & Consumer Chemistry Standardization of an unknown solution, analysis of vinegar & antacid tablets OBJECTIVE: The goals of this experiment are to learn titration
More informationDetermination of total hardness of water by complexonometric titration (EDTA)
Determination of total hardness of water by complexonometric titration (EDTA) SCOPE AND FIELD OF APPLICATION: The present method is used for the determination of total hardness of water in drinking water,
More information1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )
DR/4000 PROCEDURE Method 8023 1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to 0.700 mg/l Cr 6+ ) Scope and Application: For water and wastewater;usepa accepted for reporting for
More informationTetraphenylborate Method Method to 7.0 mg/l K Powder Pillows
Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information
More informationAccuracy and Precision of Laboratory Glassware: Determining the Density of Water
Accuracy and Precision of Laboratory Glassware: Determining the Density of Water During the semester in the general chemistry lab, you will come into contact with various pieces of laboratory glassware.
More informationCalibration of Volumetric Glassware. Prepared by Allan Fraser May 2016 APPLICATION Note 1
Calibration of Volumetric Glassware Prepared by Allan Fraser May 2016 APPLICATION Note 1 TABLE OF CONTENTS TABLE OF CONTENTS... i LIST OF TABLES... ii 1. 2. 3. 4. 5. 6. 7. 8. Purpose... 1 Scope... 1 Principle...
More informationMETHYL ETHYL KETONE PEROXIDE 3508
METHYL ETHYL KETONE PEROXIDE 3508 C 8 H 16 O 4 MW: 176.21 CAS: 1338-23-4 RTECS: EL9450000; 21837 METHOD: 3508, Issue 2 EVALUATION: PARTIAL Issue 1: 15 August 1987 Issue 2: 15 August 1994 OSHA : NIOSH:
More informationCalibration (The Good Curve) Greg Hudson EnviroCompliance Labs, Inc.
Calibration (The Good Curve) Greg Hudson EnviroCompliance Labs, Inc. greghudson@envirocompliance.com www.envirocompliance.com Abstract It should come as no surprise that the correlation coefficient is
More informationAnethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin
OIVMABS15 Anethole. Gas chromatograhpy determination of transanethole in Type II method SCOPE This method is suitable for the determination of transanethole in aniseedflavoured spirit drinks using capillary
More informationAnalytical Chemistry. The Analytical Process Sample Preparation. Dr. A. Jesorka, 6112,
Analytical Chemistry The Analytical Process Sample Preparation Dr. A. Jesorka, 6112, aldo@chalmers.se Lecture Summary 1. The analytical process. More than measurement! Several steps from problem to report.
More informationMETHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)
METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,
More informationDR/4000 PROCEDURE NITRATE. 2. The display will show:
Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number
More informationBasic Digestion Principles
Basic Digestion Principles 1 From Samples to Solutions Direct Analytical Method Solid Sample Problems: Mech. Sample Preparation (Grinding, Sieving, Weighing, Pressing, Polishing,...) Solid Sample Autosampler
More informationFor this lab, you will determine the purity of the aspirin by titration and by spectrophotometric analysis.
Introduction: ommercially prepared aspirin tablets are not considered 100% pure acetylsalicylic acid. Most aspirin tablets contain a small amount of binder which helps prevent the tablets from crumbling.
More informationApplication Determination of isocyanate (NCO-) content
Determination of isocyanate (NCO-) content date: 20.06.2013 page 1 from 10 Use This method is applicable for material containing reactive isocyanate groups.the isocyanate is reacted with di-n-butyl amine
More informationChlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell
Hardness DOC316.53.01044 Chlorphosphonazo Colorimetric Method Method 8374 8 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow Scope and application: For boiler and ultrapure water. Test preparation
More informationMETHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.
METHOD 9200 NITRATE 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to the analysis of ground water, drinking, surface, and saline waters, and domestic and industrial wastes. Modification can be
More informationSample Analysis Design Internal Standard
Internal Standard Advantages: Fluctuations are monitored in each sample/ calibration / blank Disadvantages: Assume that behavior of IS is the same as the analyte Instrumental Drift Correct for instrument
More information