Laboratory Quality Control Report: Why is it Important?

Size: px
Start display at page:

Download "Laboratory Quality Control Report: Why is it Important?"

Transcription

1 Laboratory Quality Control Report: Why is it Important? Bradley J. Austin, Mike Daniels and Brian E. Haggard Arkansas Water Resources Center University of Arkansas System Division of Agriculture FS October

2 Laboratory Quality Control Report: Why is it Important? Bradley J. Austin, Mike Daniels, and Brian E. Haggard Arkansas Water Resources Center University of Arkansas System Division of Agriculture The Arkansas Water Resources Center (AWRC) maintains a fee-based water quality lab that is certified through the Arkansas Department of Environmental Quality (ADEQ). The AWRC Water Quality Lab analyzes water samples for a variety of constituents, using standard methods for the analysis of water samples (APHA 2012). Whether you have one or several water samples tested, the lab generates a report of values for each parameter that you have analyzed, which is provided to the client. Included with every water quality report is a Lab Quality Control (QC) report for each of the parameters analyzed within the package. The Lab QC report provides important information about the performance of the methods used to test your water sample(s). What is Quality Control? Quality control consists of the lab techniques that demonstrate the precision or repeatability and accuracy of each specific analytical method; and in doing so, provides confidence that the values reported in your water quality report are correct. These techniques include incorporating blanks, standards, sample duplicates, and laboratory sample spikes into every group of samples analyzed at the Water Quality Lab. Below in table 1 you will find a list of terms that will be helpful in understanding the rest of this fact sheet. Analyte Term Concentration Method Detection Limit Reagent Grade Water Reporting Limit Sample Blank Standard Sample Digestion Sample Duplicate Sample Matrix Sample Spike Standard Definition This is the parameter of concern. This is the measure of the amount of analyte contained per unit of volume in your sample. Minimum concentration measured with 99% confidence that the true value is greater than zero (see fact sheet FS ). Water that is free of all analytes that your sample is being tested for. The lowest quantified level within an analytical methods operational range (see fact sheet FS ). This is a sample of reagent grade water. Blanks are used to assess for possible contamination. This is a sample that has a known concentration of the analyte that your water is being tested for. Standards are used to test for the accuracy of the method. Process by which all nitrogen and phosphorus in a sample are converted to nitrate and soulble reactive phosphorus, respectively. This is a re-analysis of your sample, and is used to determine the precision of the method. This is a sample with a known concentration of analyte added to it prior to lab processing and analysis. This is a sample with a known concentration of analyte added to it prior to lab processing and analysis. This is a sample that has a known concentration of the analyte that your water is being tested for. Standards are used to test for the accuracy of the method. 2

3 So, What is the Difference Between Accuracy and Precision? Accuracy of an analysis describes how close the measured values are to the true values. If you think about accuracy in the form of target shooting, having high accuracy means hitting the target as close to the bullseye as possible. Precision of an analysis describes how similar measured values are to each other, regardless of how accurate or inaccurate the analysis may be. Staying with the target analogy, having high precision means having a tight grouping of all of your shots. High Accuracy Low Precision Low Accuracy High Precision High Accuracy High Precision In order to provide you with the most reliable data possible, the goal for water quality lab is to be both as accurate and precise as possible. The different tables in the Lab QC Report allow you to see the labs performance with respect to accuracy and precision while testing the different parameters in your water sample Blank Data Table Sample blanks are used to look for possible sources of contamination to your sample from the collection of the sample to the analysis of your sample in the lab. There are several blanks listed below which are defined by what stage of sample processing they are included. Digested Blank: A digested blank is a sample of reagent grade water that goes through all the steps of sample preparation in the lab before the samples are analyzed. This blank is used to assess for possible contamination introduced during sample preparation activities. The AWRC Water Quality Lab only includes this blank when analyzing samples for total nitrogen and total phosphorus. Method Blank: The method blank is a sample of reagent grade water that is analyzed with your water samples. This blank is included to assess for possible contamination associated with the analytical instrumentation. Other water quality labs may refer to this blank as the Instrument Blank. Almost all AWRC Water Quality Lab analyses include a method blank. There is variability in every laboratory method and while a value of zero is preferred for blanks, the blanks reported in your lab QC report should be less than the reporting limit (APHA 2012). The reporting limit is the lowest quantified level within an analytical method; this value is specific to each analyte and can be found in the analytical report next to the value determined for your sample. Sample blanks are used to look for possible sources of contamination to your sample. 3

4 Standards Data Table As indicated earlier, the goal of the AWRC Water Quality Lab is to provide you with accurate data, or values as close to the true value as possible. Standards provide a way for the lab to have samples with a known concentration. Then by comparing the known concentration with the measured concentration the Lab can assess the accuracy of their measurement. There are three types of standards which may or may not be digested depending on the method. Calibration Standard: Calibration standards are incorporated throughout the sample run to verify that the analytical equipment s performance has not changed significantly since the initial calibration. Quality Control Standard (QCS): The QCS is made from stock solution different from that of the calibration standard and is used to help verify the accuracy of the calibration standard; this also helps verify the accuracy of the value(s) reported for your sample(s). Digested Standards: In analyses where your sample goes through a digestion process, such as total nitrogen and total phosphorus, both the calibration standards and QCS standards are also digested following the same process as your sample. This process helps assess the efficiency of the digestion method. For a sample run to be approved, the measured value of each of the standards analyzed must be no more than +/- 10% of the expected value. From the standpoint of the table provided in the Lab QC Report, both the lower and higher limits are listed to either side of the value obtained as a quick reference to whether or not the obtained values are acceptable. The percent recovery column describes the accuracy of the analysis and should be between 90 and 110% of the expected concentration. What this means with respect to your sample(s) is that if the method and analytical equipment are accurately measuring the standards to within +/- 10% of the expected value, then the same is true for your sample(s). Duplicate Data Table Sample duplicates are used to assess the repeatability or precision of sample collection and analysis procedures. For duplicates the lab analyzes your sample twice and the closer the two measured values are, the more precise the analysis is. Lab Duplicate: A lab duplicate is used to assess the precision of the laboratory methods and analytical equipment. Which sample is analyzed for the duplicate is chosen randomly throughout the full sample run, so in your Lab QC Report you may not likely find a duplicate data table for every parameter analyzed. Within the duplicate data table you will find the value measured the first time in the Result column followed by the value measured the second time in the Duplicate column. The last column of the table is labeled % RPD which stands for Relative Percent Difference. The %RPD is calculated as the absolute difference in the result and duplicate, divided by the average of the two values then multiplied by 100. When examining this value, the closer to zero the better as this indicates a high level of precision. Overall, a %RPD value of 30% or less should be viewed as acceptable. Samples with reported values below the reporting limit can produce greater variability, which may result in %RPD values much greater than 30%; but this does not mean that your data is not meaningful. %RPD = Result - Duplicate (Result + Duplicate)/2 x 100 4

5 Recovery Data Table The recovery data table shows results for sample spikes. The sample spike is used to evaluate analyte recovery in a sample matrix and is similar to a calibration or QC standard. The calibration and QC standards are made with a pure stock solution, so no chemical component of the stock solution should interfere with the analysis of the analyte of concern. However, this is not necessarily the case for actual water samples. Chemical constituents that occur in your water sample may influence the ability of the analytical instruments to accurately detect and measure the concentrations of the analyte of concern; this is known as a matrix effect. A laboratory sample spike can be used to detect a matrix effect. For the sample spike, calibration standard is added to a duplicate sample to achieve a known spike level, or increase in analyte concentration. Following the analysis the Result (value obtained from the un-altered sample) is compared to the Result+Spike spike (value of the sample spike). The % Recovery is calculated as the Result+Spike spike minus the Result, divided by the spike level, and then multiplied by 100. As with the standards the % Recovery should be no more than 110% and no less than 90%. Values outside this range might suggest other analytes in your water sample are creating a matrix effect and the reported value(s) should be viewed with caution. What is Quality Assurance? %RPD = Spike Result Spike Level x 100 Quality control is just part of the overall Quality Assurance and Quality Control (QA/QC) protocol. Quality Assurance refers to laboratory protocols that specify the measures that are taken by the lab personnel to produce defensible data with known precision and accuracy (APHA 2012). Essentially QA is the steps that the AWRC Water Quality Lab takes to ensure the values they report in your water quality analytical report are as close to the actual values as possible. Following the analysis of your water sample both the analytical and Lab QC reports are reviewed by two individuals. The first individual verifies that the numbers in the report were correctly entered in from the original analysis, and looks for errors in the analysis. The QA/QC Reviewer provides a second set of eyes to look over your water quality report to ensure that the data provided is accurate. 5

6 Literature Cited APHA (American Public Health Association) Standard Methods for the Examination of Water and Wastewater (22nd edn.). American Public Health Association: Washington D.C pp. Austin, B.J., J.T. Scott, M. Daniels, B.E. Haggard Water Quality Reporting Limits, Method Detection Limits, and Censored Values: What Does it All Mean?. Arkansas Water Resources Center, Fayetteville, Arkansas, FS : 8 pp. How to Cite This Fact Sheet Austin, B.J., M. Daniels, and B.E. Haggard Laboratory Quality Control Report: Why is it Important?. Arkansas Water Resources Center, Fayetteville, AR, FS : 06 pp. Arkansas Water Resources Center awrc@uark.edu College of Engineering 203 Engineering Hall University of Arkansas Fayetteville, AR

Water Quality Reporting Limits, Method Detection Limits, and Censored Values: What Does It All Mean?

Water Quality Reporting Limits, Method Detection Limits, and Censored Values: What Does It All Mean? Water Quality Reporting Limits, Method Detection Limits, and Censored Values: What Does It All Mean? Bradley J. Austin, J. Thad Scott, Mike Daniels and Brian E. Haggard Arkansas Water Resources Center

More information

Water Quality MDLs, PQLs, and Censored Values What Does It All Mean?

Water Quality MDLs, PQLs, and Censored Values What Does It All Mean? Water Quality MDLs, PQLs, and Censored Values What Does It All Mean? Brad Austin, Thad Scott, M. Daniels, and Brian Haggard Arkansas Water Resources Center University of Arkansas System s Division of Agriculture

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015 Laboratory Techniques 100: Back To Basics Carol Injasoulian Lab Manager City of Bay City April 29,2015 QA/QC for Environmental Measurement What is quality assurance (QA)? Broad plan to maintain quality

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

Measurement Uncertainty: A practical guide to understanding what your results really mean.

Measurement Uncertainty: A practical guide to understanding what your results really mean. Measurement Uncertainty: A practical guide to understanding what your results really mean. Overview General Factors Influencing Data Variability Measurement Uncertainty as an Indicator of Data Variability

More information

And how to do them. Denise L Seman City of Youngstown

And how to do them. Denise L Seman City of Youngstown And how to do them Denise L Seman City of Youngstown Quality Control (QC) is defined as the process of detecting analytical errors to ensure both reliability and accuracy of the data generated QC can be

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. 1 REVISOR 4740.2100 4740.2100 QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. Subpart 1. Scope. This part applies to laboratories performing testing under the inorganic chemistry, metals,

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

Water & Wastewater. Laboratory Math

Water & Wastewater. Laboratory Math Water & Wastewater Laboratory Math VWEA Good Lab Practices Meeting Presented by: Wayne Staples Virginia DEQ July 24, 2017 Required to: Math in the Laboratory Determine a test result Alkalinity calculation

More information

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016 QA/QC in the Wastewater Laboratory Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016 What is Quality Assurance? Per Standard Methods for the Examination of Water and Wastewater, 21 st

More information

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA)

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) QA/QC Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) Is a description of the policies, procedures, techniques

More information

American Environmental Testing Laboratory Inc.

American Environmental Testing Laboratory Inc. American Environmental Testing Laboratory Inc. 2834 & 2908 North Naomi Street Burbank, CA 91504 DOHS NO: 1541, LACSD NO: 10181 Tel: (888) 288-AETL (818) 845-8200 Fax: (818) 845-8840 www.aetlab.com Ordered

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS Attention:Steve Allum 802 Drury Lane Burlington, ON L7R 4L3 Your Project #: LEAD Your C.O.C. #: na Report #: R4042481 Version: 1 - Final MAXXAM JOB #: B6C8371 Received: 2016/06/22, 08:00 Sample Matrix:

More information

Results of the EPA Method 1631 Validation Study

Results of the EPA Method 1631 Validation Study Results of the EPA Method 1631 Validation Study U.S. EPA Office of Water Office of Science and Technology Engineering and Analysis Division Washington, DC 20460 July 1996 Acknowledgments This report was

More information

CHAPTER VI ANALYTICAL METHODS & QUALITY CONTROL

CHAPTER VI ANALYTICAL METHODS & QUALITY CONTROL CHAPTER VI ANALYTICAL METHODS & QUALITY CONTROL Section A. Section B. Section C. Introduction Definitions and Terms Laboratory QA/QC 1. Sample Preservation and Holding Times 2. Sample Receiving 3. Sample

More information

This procedure describes the monitoring activities in a laboratory quality control (QC) program to ensure the quality of test results.

This procedure describes the monitoring activities in a laboratory quality control (QC) program to ensure the quality of test results. Page 1 of 8 Home Science & Research Field Science and Laboratories Laboratory Manual Science & Research Volume II - Assuring the Quality of Test Results ORA-LAB.5.9 ORA Laboratory Procedure Food and Drug

More information

December 30, Laboratory Report This Page is to be Stamped Introduction: This report package contains total of 5 pages divided into 3 sections:

December 30, Laboratory Report This Page is to be Stamped Introduction: This report package contains total of 5 pages divided into 3 sections: December 30, 2018 ENVIRONMENTAL AGRICULTURAL Meadowlark Ranches Mutual Water Co. Lab ID : SP 1816365 Attn: Mike Hadley P.O. Box 606 Santa Ynez, CA 93460-0606 Customer : 2-21676 Laboratory Report This Page

More information

Glossary of Common Laboratory Terms

Glossary of Common Laboratory Terms Accuracy A measure of how close a measured value is to the true value. Assessed by means of percent recovery of spikes and standards. Aerobic Atmospheric or dissolved oxygen is available. Aliquot A measured

More information

Understanding and Interpreting Soil and Plant Tissue Lab Reports

Understanding and Interpreting Soil and Plant Tissue Lab Reports Understanding and Interpreting Soil and Plant Tissue Lab Reports Dirk Holstege Director, UC Davis Analytical Laboratory dmholstege@ucdavis.edu 530-752-0148 224 Hoagland Hall UC Davis Anlab.ucdavis.edu

More information

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) 683 METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and

More information

Method Validation and Accreditation

Method Validation and Accreditation SELAMAT Mycotoxins Workshop China, 11-15th December 2006 Method Validation and Accreditation Dr Hamide Z Şenyuva Senior Research Scientist TÜBİTAK-ATAL, TURKEY hamide.senyuva@tubitak.gov.tr SELAMAT Mycotoxins

More information

ANALYTICAL METHODS & QUALITY CONTROL

ANALYTICAL METHODS & QUALITY CONTROL CHAPTER 6 ANALYTICAL METHODS & QUALITY CONTROL Section A. Section B. Section C. Introduction Definitions and Terms Laboratory QA/QC 1. Sample Preservation and Holding Times 2. Sample Receiving 3. Sample

More information

Appendix A. Glossary of Definitions, Acronyms, Abbreviations and Symbols. March 1997

Appendix A. Glossary of Definitions, Acronyms, Abbreviations and Symbols. March 1997 Appendix A Glossary of Definitions, Acronyms, Abbreviations and Symbols March 1997 Glossary Of Definitions, Acronyms, Abbreviations And Symbols 1.0 Standard Acronyms, Abbreviations, And Symbols 1.1 Units

More information

APPENDIX II QUALITY CONTROL SUMMARY TABLES

APPENDIX II QUALITY CONTROL SUMMARY TABLES APPENDIX II QUALITY CONTROL SUMMARY TABLES Table II-1: Completeness Checklist Quality Assurance/Quality Control Questions? Comments? 1. Was the report signed by the responsible applicant approved representative?

More information

Laboratory 101: A Guide to Understanding your Testing Laboratory

Laboratory 101: A Guide to Understanding your Testing Laboratory Laboratory 101: A Guide to Understanding your Testing Laboratory Jim Cronin Product Ecology Manager Environmental Monitoring and Technologies, Inc. (EMT) Morton Grove, Illinois Abstract: Laboratory 101:

More information

Analytical Measurement Uncertainty APHL Quality Management System (QMS) Competency Guidelines

Analytical Measurement Uncertainty APHL Quality Management System (QMS) Competency Guidelines QMS Quick Learning Activity Analytical Measurement Uncertainty APHL Quality Management System (QMS) Competency Guidelines This course will help staff recognize what measurement uncertainty is and its importance

More information

One Stop DIY Shop. PO Box 104 New Germany, MN Carbonyl Assay. Samples Received 09/09/2014. Analysis Report ( )

One Stop DIY Shop. PO Box 104 New Germany, MN Carbonyl Assay. Samples Received 09/09/2014. Analysis Report ( ) One Stop DIY Shop PO Box 104 New Germany, MN 55367 Carbonyl Assay Samples Received 09/09/2014 Analysis Report (0914-72) GC/MS Analysis Pentanedione (aka Acetyl propionyl) Diacetyl Enthalpy Analytical,

More information

Application of Detection and Quantification Concepts to Chlorine Residual Measurements

Application of Detection and Quantification Concepts to Chlorine Residual Measurements Application of Detection and Quantification Concepts to Chlorine Residual Measurements Charles D. Hertz, Ph.D. PADEP Proposed Disinfection Requirements Rule Stakeholder Meeting Harrisburg, PA March 9,

More information

Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University. and

Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University. and Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: Date: OEWRI Quality Assurance

More information

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Timberline Ammonia-001 Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Prepared By: Dr. Edward F. Askew Date: June 24, 2011 Table of Contents 1. SCOPE

More information

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Proposed Procedures for Determining the Method Detection Limit and Minimum Level Proposed Procedures for Determining the Method Detection Limit and Minimum Level Published by: ACIL Environmental Services Section Technical Committee Revision 3.0 3/8/006 PROCEDURES These procedures set

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) SOP-316 391 S. Vidal St., Sarnia, Ontario, N7T 7L1 Phone: (519) 344-4747 Fax: (519) 344-2350 E-Mail: info@lambtonscientific.com Method

More information

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350)

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350) Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350) Ann Strahl Technical Support Remediation Division TCEQ 512-239-2500 astrahl@tceq.state.tx.us 1 Data

More information

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager

More information

COLA Mass Spec Criteria

COLA Mass Spec Criteria Wednesday April 5, 2017 2pm COLA Mass Spec Criteria Kathy Nucifora, MPH, MT(ASCP) COLA, Accreditation Division Director Columbia, MD DESCRIPTION: Kathy Nucifora, COLA Accreditation Division Director, will

More information

Accuracy: An accurate measurement is a measurement.. It. Is the closeness between the result of a measurement and a value of the measured.

Accuracy: An accurate measurement is a measurement.. It. Is the closeness between the result of a measurement and a value of the measured. Chemical Analysis can be of two types: Chapter 11- Measurement and Data Processing: - : Substances are classified on the basis of their or properties, such as - : The amount of the sample determined in

More information

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES Adopted from ICH Guidelines ICH Q2A: Validation of Analytical Methods: Definitions and Terminology, 27 October 1994. ICH Q2B: Validation of Analytical

More information

Quick Nic Juice. 122 Indian Springs Drive #5 Sandwich, IL Carbonyl Assay Samples Received 3/16/2015. Analysis Report ( )

Quick Nic Juice. 122 Indian Springs Drive #5 Sandwich, IL Carbonyl Assay Samples Received 3/16/2015. Analysis Report ( ) Quick Nic Juice 122 Indian Springs Drive #5 Sandwich, IL 60548 Carbonyl Assay Samples Received 3/16/2015 Analysis Report (0315-538) HPLC/UV Analysis Acetoin (aka 3-Hydroxy-2-butanone) Diacetyl 2,3-Pentanedione

More information

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION Written by Pamela Doolittle, University of Wisconsin- Madison, pssemrad@wisc.edu 2014 In this experiment you will determine the phosphorus (P) concentration

More information

GOOD TEST PORTIONS. QA / QC: A Critical Need for Laboratory Sampling

GOOD TEST PORTIONS. QA / QC: A Critical Need for Laboratory Sampling + GOOD TEST PORTIONS QA / QC: A Critical Need for Laboratory Sampling + Quality Assurance Quality Control + GOOD Test Portion Working Group Members n Jo Marie Cook, FL Department of Ag & Consumer Services

More information

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP. NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta

More information

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

Quality Control Procedures for Graphite Furnace AAS using Avanta Software Application Note Quality Control Procedures for Graphite Furnace AAS using Avanta Software Introduction With the arrival of quality standards in the analytical laboratory, an ever increasing emphasis is

More information

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Table of Contents Section 1: Summary of Method Section 2: Scope and Application Section

More information

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,: sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES The National Association of Testing Authorities (NATA) requirements for accreditation have undergone

More information

EPA Office of Water Method Update Rule Final August 28, 2017 William Lipps October 2017

EPA Office of Water Method Update Rule Final August 28, 2017 William Lipps October 2017 EPA Office of Water Method Update Rule Final August 28, 2017 William Lipps October 2017 1 / 9 The update consists of minor changes, edits, revisions, and 7 new methods 80 70 60 50 40 30 20 10 0 Standard

More information

OBJECTIVE DATA PACKAGE DELIVERABLES

OBJECTIVE DATA PACKAGE DELIVERABLES PAGE: 1 of 10 OBJECTIVE This Standard Operating Procedure (SOP) describes the requirements for the analytical data packages that will be generated in association with the Stage 2 Delaware River Estuary

More information

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18 Draft PS 18 APPENDIX A Dynamic Spiking Procedure (4-4-2013) A1. Scope and Application 1.1 This appendix to Performance Specification 18 describes the procedure and performance requirements for dynamic

More information

~ ALG ANALYTICAL LABORATORY GROUP

~ ALG ANALYTICAL LABORATORY GROUP August 19, 2016 Doug Pigman Greater Albany Public Schools 3610 Grand Prairie Albany, OR 97322 TEL: (541) 967-4513 FAX RE: Waverly Retest Order No.: 1608769 Dear Doug Pigman: Analytical Laboratory Group

More information

KESTREL. April 26, Mr. Stephen Evans Maine Yankee 321 Old Ferry Road Wiscasset, Maine 04578

KESTREL. April 26, Mr. Stephen Evans Maine Yankee 321 Old Ferry Road Wiscasset, Maine 04578 KESTREL ENVIRONMENTAL TECHNOLOGIES, INC. Mr. Stephen Evans Maine Yankee 321 Old Ferry Road Wiscasset, Maine 04578 RE: Project No.: Maine Yankee Soil Samples Lab Name: Katahdin Analytical Services, Westbrook,

More information

Analytical Report

Analytical Report Analytical Report 386228 for Natural Soap Formulas Project Manager: Wanda Baligad Formula 206 Natural Soap Formulas 20-AUG-10 3231 NW 7th Avenue, Boca Raton, FL 33431 Ph:(561) 447-7373 Fax:(561) 447-6136

More information

Introduction to Measurement

Introduction to Measurement Units and Measurement Introduction to Measurement One of the most important steps in applying the scientific method is experiment: testing the prediction of a hypothesis. Typically we measure simple quantities

More information

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800)

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800) 2773 Downhill Drive Steamboat Springs, CO 80487 (800) 334-5493 Analytical Report November 19, 2004 Report to: Debbie Furlow 602 Lehmberg Road Columbus, MS 39702 Bill to: Accounts Payable PO Box 2487 Columbus,

More information

Instructions for the teacher

Instructions for the teacher Constant 19 different words commonly used in solving equations. Help your students quickly master these words via two self-correcting puzzles! Complete with a thorough terminology reference section, instructions,

More information

Percent Solids Determination

Percent Solids Determination Title: Percent Solids Page 1 of 10 Percent Solids Determination References: Method 2540G, Standard Methods For the Examination of Water and Wastewater, APHA 18 th edition, 1992. 1. Scope and Application

More information

Table of Contents I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:...

Table of Contents I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:... Section Table of Contents Page No I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:... 4 (A) Introduction... 4 (B) Data Package

More information

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94. Ethanol Project Chemical Analysis and Testing Task Laboratory Analytical Procedure Procedure Title: LAP-013 HPLC Analysis of Liquid Fractions of Process Samples for Monomeric Sugars and Cellobiose Author:

More information

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations

More information

SUGGESTED PROCEDURES FOR

SUGGESTED PROCEDURES FOR DRAFT The NELAC Institute Presents: SUGGESTED PROCEDURES FOR LOQ TNI plays well with Others These suggested procedures are compliant with NELAC and TNI standards, and can provide a dataset suitable for

More information

Hydra-C Application Note: 1076

Hydra-C Application Note: 1076 Hydra-C Application Note: 1076 The Determination of Mercury in Samples by U.S. EPA SOW 846 Method 7473 using the Hydra-C Mercury Analyzer Introduction The accurate determination of mercury in various samples

More information

MPCA Water Quality Database

MPCA Water Quality Database Appendix : MPCA Water Quality Database meta-data descriptions and requirements Project information 1 Project ID Project name Project purpose Start date Name of the monitoring project Reason why the monitoring

More information

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection Role of Validation Method Validation An overview Confirms the fitness for purpose of a particular analytical method. ISO definition: Conformation by examination and provision of objective evidence that

More information

Standard Operating Procedure for Performing a QA Review of an Aldehyde Data Batch

Standard Operating Procedure for Performing a QA Review of an Aldehyde Data Batch North Carolina Division of Air Quality Quality Assurance Project Plan 2.52 Aldehydes by TO-11a Section 4.1 Raleigh Central Office Responsibilities Aldehyde Data Review Section 2.52.4.1 Page 1 of 7 Standard

More information

A Laboratory Guide to Method Validation, (Eurachem).

A Laboratory Guide to Method Validation, (Eurachem). Page 1 of 19 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6.3 C. added NOTE:) 7. Definitions 8. Records 9. Supporting

More information

Understanding the Uncertainty Associated with Analytical Results: Sources, Control and Interpretation of Results. Marc Paquet, Chemist M.Sc.

Understanding the Uncertainty Associated with Analytical Results: Sources, Control and Interpretation of Results. Marc Paquet, Chemist M.Sc. Understanding the Uncertainty Associated with Analytical Results: Sources, Control and Interpretation of Results Marc Paquet, Chemist M.Sc. 1 Topics Misconceptions; Facts; FAQs; Sources of uncertainty;

More information

Update on ASTM and Standard Methods method development activities

Update on ASTM and Standard Methods method development activities Update on ASTM and Standard Methods method development activities William Lipps, Eurofins Eaton Analytical Monrovia South Bend www.eatonanalytical.com Full Service Laboratories Eurofins Lancaster Laboratories

More information

Source: Chapter 5: Errors in Chemical Analyses

Source: Chapter 5: Errors in Chemical Analyses Source: Chapter 5: Errors in Chemical Analyses Measurements invariably involve errors and uncertainties. it is impossible to perform a chemical analysis that is totally free of errors or uncertainties

More information

Enclosed is a copy of your laboratory report for test samples received by our laboratory on Thursday, February 25, 2010.

Enclosed is a copy of your laboratory report for test samples received by our laboratory on Thursday, February 25, 2010. Environmental Conservation Laboratories, Inc. 10775 Central Port Drive Orlando FL, 32824 407.826.5314 407.850.6945 Phone: FAX: www.encolabs.com Friday, March 5, 2010 Jones Edmunds & Associates (JO005)

More information

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant SWGDRUG GLOSSARY These definitions were developed and adopted by the SWGDRUG Core Committee from a variety of sources including The United Nations Glossary of Terms for Quality Assurance and Good Laboratory

More information

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4 page1 of 8 TNI Standard VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 January 2016 Description This TNI Standard has been taken through all of the voting stages and has

More information

Section A LABORATORY QUALITY ASSURANCE/QUALITY CONTROL. Her Majesty the Queen in Right of the Province of British Columbia 2007 All Rights Reserved.

Section A LABORATORY QUALITY ASSURANCE/QUALITY CONTROL. Her Majesty the Queen in Right of the Province of British Columbia 2007 All Rights Reserved. Section A LABORATORY QUALITY ASSURANCE/QUALITY CONTROL Her Majesty the Queen in Right of the Province of British Columbia 2007 All Rights Reserved. TABLE OF CONTENTS SECTION A Laboratory Quality Assurance/Quality

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

QAM-Q-101 Laboratory Quality Control

QAM-Q-101 Laboratory Quality Control 1.0 Applicability and Purpose 1.1. This procedure applies to operations of the analytical laboratory at the Texas Institute for Applied Environmental Research (TIAER), Tarleton State University, Stephenville,

More information

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer

More information

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS EL-VM-0 ENVIRONMENTAL LABORATORY SECTOR VOLUME MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS Module : Quality Systems for Radiochemical Testing Working Draft

More information

Appendix A - Test Methods Method 301--Field Validation of Pollutant Measurement Methods from Various Waste Media

Appendix A - Test Methods Method 301--Field Validation of Pollutant Measurement Methods from Various Waste Media Appendix A - Test Methods Method 301--Field Validation of Pollutant Measurement Methods from Various Waste Media 1. APPLICABILITY AND PRINCIPLE 1.1 Applicability. This method, as specified in the applicable

More information

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards SeCtiOn 5 [STOCK AND CUSTOM] Assurance Multi-Element CLP Standards Your Science is Our Passion. Assurance CLPStandards SPEX CertiPrep offers acomplete series of multi-element standards and blanks designed

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

California Environmental Protection Agency Air Resources Board

California Environmental Protection Agency Air Resources Board California Environmental Protection Agency Air Resources Board PROCEDURE FOR THE ANALYSIS OF PARTICULATE ANIONS AND CATIONS IN MOTOR VEHICLE EXHAUST BY ION CHROMATOGRAPHY Standard Operating Procedure No.

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES STANDARD OPERATING PROCEDURES for Spectrophotometric Measurement Chl a, Chl b, and Bulk Carotenoids using Methanol Solvent Prepared by EcoAnalysts, Inc. 1420 South Blaine Street, Suite 14 Moscow, ID 83843

More information

TNI V1M Standard Update Guidance on Detection and Quantitation

TNI V1M Standard Update Guidance on Detection and Quantitation TNI V1M4 2016 Standard Update Guidance on Detection and Quantitation GUID-3-109-Rev0 January 30, 2019 This material represents the opinion of its authors. It is intended solely as guidance and does not

More information

Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit

Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit Method(s): Laboratory ID Laboratory Name Address Phone Lab Manager/Supervisor QA Officer Analyst(s) Auditor Date(s) of Audit NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE

More information

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University. Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared

More information

Stream Water Quality Assessment of Long Point Gully, Graywood Gully, and Sand Point Gully: Conesus Lake Tributaries Spring 2012

Stream Water Quality Assessment of Long Point Gully, Graywood Gully, and Sand Point Gully: Conesus Lake Tributaries Spring 2012 The College at Brockport: State University of New York Digital Commons @Brockport Technical Reports Studies on Water Resources of New York State and the Great Lakes 6-2012 Stream Water Quality Assessment

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

LABORATORY REPORT. If you have any questions concerning this report, please do not hesitate to call us at (800) or (574)

LABORATORY REPORT. If you have any questions concerning this report, please do not hesitate to call us at (800) or (574) LABORATORY REPORT If you have any questions concerning this report, please do not hesitate to call us at (800) 332-4345 or (574) 233-4777. This report may not be reproduced, except in full, without written

More information

SECTION D.2 AMMONIA NITROGEN

SECTION D.2 AMMONIA NITROGEN SECTION D.2 AMMONIA NITROGEN CEDR Method Code: NH4F L01 a) Scope and Application i) This method describes the determination of low-level ammonia nitrogen concentrations in filtered samples taken from fresh

More information

LABORATORY REPORT. If you have any questions concerning this report, please do not hesitate to call us at (800) or (574)

LABORATORY REPORT. If you have any questions concerning this report, please do not hesitate to call us at (800) or (574) LABORATORY REPORT If you have any questions concerning this report, please do not hesitate to call us at (800) 332-4345 or (574) 233-4777. This report may not be reproduced, except in full, without written

More information

Spectrophotometric Determination of Ferrocyanide in Effluents

Spectrophotometric Determination of Ferrocyanide in Effluents Spectrophotometric Determination of Ferrocyanide in Effluents ECN-0025-1 INTRODUCTION This method is used to determine the concentration of ferrocyanide ion in photoprocessing solution effluents. The ion

More information

Method 8270C PAH. FD- Field Duplicate Samples

Method 8270C PAH. FD- Field Duplicate Samples Mr. John Rendall Maine Yankee 321 Old Ferry Road Wiscasset, Maine 04578 RE: Project No.: Maine Yankee Sediment Samples Lab Name: Katahdin Analytical Services, Westbrook, Maine Site Name: Maine Yankee Nuclear

More information

Applications of Ion-Selective Electrodes

Applications of Ion-Selective Electrodes Applications of Ion-Selective Electrodes Analyte Ammonia Carbon dioxide Chloride Chlorine residual Cyanide Fluoride Nitrate Nitrogen oxide/nitrite Oxygen, dissloved Sulfide Electrode type Gas sensing Gas

More information

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG Quantification of residual Protein A ligands in the presence of excess amounts of IgG Product number P0020456 P0020457 Product Name Gyrolab Protein A Kit for MabSelect SuRe Ligand Gyrolab Protein A Kit

More information

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE Use objectives: state purpose for which results will be used. Specifications: How good should the numbers be

More information

Matrix Effects and Uncertainty. A framework for further study. The two analytical problems

Matrix Effects and Uncertainty. A framework for further study. The two analytical problems Matrix Effects and Uncertainty Michael Thompson Birkbeck College (University of London) m.thompson@bbk.ac.uk A framework for further study A review of interference effects and their correction in chemical

More information

Biol/Chem 4900/4912. Forensic Internship Lecture 5

Biol/Chem 4900/4912. Forensic Internship Lecture 5 Biol/Chem 4900/4912 Forensic Internship Lecture 5 Quality Assurance/ Quality Control Quality Assurance A set of activities that ensures that development and/or maintenance processes are adequate in order

More information