1 Performance Characterization and Field Testing of a Portable Tandem Mass Spectrometer John Grossenbacher Mitch Wells, Garth Patterson, Brent Knecht, Dennis Barket, Jr., Brent Rardin, Mark Gregory, Jason Springston, Rob Donoho Griffin Analytical Technologies Inc. West Lafayette, IN
2 Griffin Analytical Technologies Technology transfer from Purdue s top-tier Analytical Chemistry Department Exclusive license in all fields of use Developing miniaturized, portable mass spectrometer-based instrumentation that provides bench-top performance in a portable package Chemistry Happens Outside the Lab So Should Analysis.
3 Field MS Applications Homeland Security/Defense Detection surrounding potential terrorist targets Transportation security Forward-deployed protection of troops On-site Environmental Monitoring Water quality Continuous air monitor R&D Tool for Chemical Warfare Testing
4 Advantages of the Cylindrical Ion Trap Operation under relatively high vacuum ~10-4 Torr MS/MS capabilities Selectivity Ease of machining/manufacturing Reduced voltage requirements
5 Ion Trap Comparison Standard Ion Trap Cylindrical Ion Trap electrodes mass resolution MS n capability Hyperbolic radius = 10 mm req d. voltage ~8000 V (at 1.1 MHz) mass range 1000 amu (typical) unit (typical) MS 4 (typical) Flat radius = 2.5 mm ~600 V (at 1.1 MHz) 350 amu (typical) unit (demonstrated) MS 3 (demonstrated) Result of Difference Simplified machining, lower operating voltage Smaller electronics, lower power consumption Range for target applications Performance comparable Performance comparable
6 Minotaur Instrument Minotaur Specifications: Mass Analyzer: Cylindrical Ion Trap Sample Inlet: Internal Membrane Ionization: Internal Electron Ionization (EI) MS3: Demonstrated Sensitivity: LODs of ppb to pptr, compound specific Mass Range: 350 Da/charge Resolution: ~unit Bath Gas: ambient air (~10-4 Torr) Detection: positive and negative ions Size: 60 cm x 45 cm x 20 cm (~1.9 cu. Ft.) Weight: under 50 lbs. (with case and pumps) Power Consumption: ~100 W
7 Griffin Analytical Technologies Instrument Control Software Scan Editor - development software Advanced Waveform Editing Data Collection
8 Griffin Analytical Technologies Instrument Control Software Data Analysis
9 Griffin Analytical Technologies Inlets and Sources Currently In Use MIMS Silastic PDMS tubular membrane (0.64 mm i.d., 1.19 mm o.d.), ~2 cm long Sampling flow rate 0.5 L/min Onboard sampling pump Internal EI In Development SPME inlet GC interface Other interfaces are under development including external ionization sources
10 Gas-phase Standards Dynamic VOC Standards Standard compound Dilution air Mass flow controller 4.10 dump Mass flow controller 41.0 Mixing coil to MMS tee
11 Gas-phase Standards Bag Standards Mass flow controller 10.0 Syringe containing standard solution 3 mil PFA Teflon bag Heated Glass tee 150 L Dilution Air
12 Sampling Protocol Continuous sampling with MIMS (toluene) 2. Pervaporation 1. Sample Membrane Trap and Release MIMS (Methyl Salicylate, Triethyl Phosphate (TEP), Dimethyl Methylphosphonate (DMMP) 1. Sample Membrane 2. Trap 3. Release Sorbent Trapping with Tenax (Methyl Salicylate) 1. Sample 2. Heat/Desorb 3. Trap 4. Release Tenax tube Membrane
13 Minotaur Performance: Real-time Monitoring of VOCs Toluene LOD = 15 ppb 91/92 peak area y x y R= x R 2 = Concentration (ppm) Abundance (arb. units) m/z
14 Minotaur Performance: Trap-and-Release of Semi-Volatiles Methyl Salicylate LOD = 300 pptr m/z 92, 120, 152 peak areas 2.5E5 1.5E5 5E4 Abundance (arb. units) m/z 8 ppb standard 152 y = 221.8x R 2 = Concentration (ppb) 1200
15 Minotaur Performance: Field Testing Methyl Salicylate with Sorption Tube Sampling m/z 92, 120, 152 peak areas 2.0E5 1.2E5 4E4 y = 65.1x R 2 = LOD = 77 ppb Concentration (ppb)
16 Minotaur Performance: Trap-and-Release of triethyl phosphate (TEP) 283 ppb standard LOD 150 ppb TIC (arbitrary units) 5E4 3E4 1E4 signal (arbitrary units) 1.6E5 1.2E5 8E4 4E4 y = 166.9x R 2 = time (minutes) concentration (ppb)
17 Under Development: External Ionization Atmospheric Sampling Glow Discharge Ionization (ASGDI) explosives detection CW agents high voltage sample in ions to CIT Abundance (arb. units) PDCH C 3 F - 7 C 4 F ~0.5 Torr (M-C 2 F 5 ) - (M-CF 3 ) m/z
18 Challenges with Current Technology Analyzer Vacuum Requirements Traps Quads Sectors E TOF D B ICR pressure (Torr)
19 Challenges with Current Technology Vacuum Pumping power consumer, delicate
20 Challenges with Current Technology Miniaturization of High Voltage Electronics Size limited by HV line separation and component spacing Relatively high quiescent power consumption Broadcast/Electromagnetic Interference (EMI)
21 Future Directions Develop innovative sample introduction/ionization techniques Tailored to specific applications Increased selectivity Funding secured Refinement of instrument package Volume: ~1.5 cu. ft. Weight: ~40 lbs Further Ruggedization for field use
22 Acknowledgements U.S. EPA, NSF, and DoD for funding Prof. R.G. Cooks The Griffin Team
Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications. Mitch Wells, Ph.D. Philip Tackett, Ph.D.
Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications Mitch Wells, Ph.D. Philip Tackett, Ph.D. HEMS Workshop 2017 FLIR Systems, Inc. Industrial Commercial Government
More informationIntroduction to GC/MS
Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules
More informationHandheld Miniature Ion Trap Mass Spectrometers
Anal. Chem. 2009, 81, 2421 2425 Handheld Miniature Ion Trap Mass Spectrometers Zheng Ouyang, Robert J. Noll, and R. Graham Cooks Downloaded via 148.251.232.83 on September 23, 2018 at 19:01:28 (UTC). See
More informationSeries 2. Continuous monitoring of trace chemical vapours
Series 2 TT24-7 Continuous monitoring of trace chemical vapours Series 2 TT24-7 TT24-7 Series 2 Continuous monitoring of trace chemical vapours The TT24-7 Series 2 is an advanced thermal desorption (TD)
More informationHand-Portable GC-TMS Instrument for Measurement of Chemical Agents and Hazardous Compounds in Harsh Environments
Hand-Portable GC-TMS Instrument for Measurement of Chemical Agents and Hazardous Compounds in Harsh Environments The 6 th Workshop on Harsh Environment Mass Spectrometry Cocoa Beach, Florida September
More informationMass Spectrometry in MCAL
Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring
More informationTHE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS
International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp
More informationLC-MS Based Metabolomics
LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile
More informationAnalysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software
Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software NEMC, Washington D.C. August 6 th 2012 Nicola Watson Environmental Specialist Markes International
More informationDetection of Chemical Warfare Agents by Transportable GC/MS
Detection of Chemical Warfare Agents by Transportable GC/MS Application Note Gas Chromatography/Mass Spectrometry Author Robert V. Mustacich Agilent Technologies, Inc. 5301 Stevens Creek Boulevard Santa
More informationMASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps
ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space
More informationOwlstone Nanotech Inc
Owlstone Nanotech Inc PIDplus Technical Paper USA POC: Mark Brennan mark.brennan@owlstonenanotech.com 400 Rella Boulevard Suite 160 Suffern, NY 10901 Tel: +1 845-533-4225 Fax: +1 845-533-4232 www.owlstonenanotech.com
More informationMM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE
MM 2 The new Generation of Mobile Mass Spectrometers Innovation with Integrity Defence CBRNE Picture by: MOWAG GmbH Mobile CWA Detection and Identification Low weight and high performance detector Introducing
More informationIonization Techniques Part IV
Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector
More informationAuthor. Abstract. Introduction
Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent
More informationMiniaturized Sensor for the Detection of Environmental Pollutants Alexander Graf 1, Ronald Stübner 1, Christian Kunath 1, Sebastian Meyer 1, Harald Schenk 1 1 Fraunhofer Institute for Photonic Microsystems
More informationMass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University
Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors
More informationFundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics
Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass
More informationFast Results Anytime, Anywhere
P R O D U C T N O T E Gas Chromatography/ Mass Spectrometry Key Features Portable, lightweight GC/MS with laboratory equivalent chromatographic resolution Operates under harsh conditions, providing routine
More informationMiniature Ambient Ionization Mass Spectrometry System For Analysis of Microorganisms
Miniature Ambient Ionization Mass Spectrometry System For Analysis of Microorganisms R. Graham Cooks 1 and Zheng Ouyang 2 1 Department of Chemistry and 2 Weldon School of Biomedical Engineering Purdue
More informationChip-Scale Mass Spectrometers for Portable Gas Analyzers Luis Fernando Velásquez-García. A. I. Akinwande, K. Cheung, and L.-Y Chen.
Chip-Scale Mass Spectrometers for Portable Gas Analyzers Luis Fernando Velásquez-García. A. I. Akinwande, K. Cheung, and L.-Y Chen. Microsystems Technology Laboratories (MTL) lfvelasq@mit.edu November
More informationCEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.
More informationCEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.
More informationTANDEM DIFFERENTIAL MOBILITY SPECTROMETER:
TANDEM DIFFERENTIAL MOBILITY SPECTROMETER: AN IONIZATION DETECTOR FOR GC WITH HIGH SPEED, SELECTIVITY, SMALL SIZE, AND LOW COST G.A. Eiceman & M. Menlyadiev Department of Chemistry & Biochemistry New Mexico
More informationThe Application of Method TO-15 to Naphthalene Measurements in Indoor Air
The Application of Method TO-15 to Naphthalene Measurements in Indoor Air Extended Abstract #13 Heidi C. Hayes and Diane J. Benton Air Toxics Ltd. 180 Blue Ravine Rd. Ste. B Folsom, CA 95630 INTRODUCTION
More informationApplication Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek
Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek
More informationExplosive Screening by Vapour Detection: Operational Aspects
Explosive Screening by Vapour Detection: Operational Aspects G. Fernandez de la Mora SEDET, Boecillo, Spain, gfdelamora@sedet.com November 14 th, 2013 1 ACES OVERVIEW Product: Air Cargo Explosives Screener
More informationMass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez
Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation
More informationDefining quality standards for the analysis of solid samples
Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples
More informationTypes of Analyzers: Quadrupole: mass filter -part1
16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron
More informationMass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers
More informationDetection of Chemical Agents in Water by Membrane- Introduction Mass Spectrometry
DETECTION OF CHEMICAL AGENTS IN WATER BY MIMS Detection of Chemical Agents in Water by Membrane- Introduction Mass Spectrometry John S. Morgan, Wayne A. Bryden, Rebecca F. Vertes, and Scott Bauer Membrane-Introduction
More informationThe development of portable chemical analyzers
Focus: HARSH ENVIRONMENT MASS SPECTROMETRY Development of a Palm Portable Mass Spectrometer Mo Yang, Tae-Young Kim, Hyun-Chul Hwang, Seok-Kyung Yi, and Do-Hoon Kim Analytical Instrumentation Research Institute,
More informationThermo Finnigan LTQ. Specifications
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications
More informationFinnigan LCQ Advantage MAX
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer
More informationDetection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS
Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?
More informationTalk Line-up. MassTech, Inc Columbia Gateway Drive Columbia, Maryland (443) Goals for MT Explorer 50 Development
A Field Deployable Ion Trap Mass Spectrometer with Atmospheric Pressure Interface Arnold Lee *, Alexander Misharin, Konstantin Novoselov, Victor Laiko, Vladimir M. Doroshenko Talk Line-up Goals for MT
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?
More informationVOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations
VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations Environmental Chemistry group, water Science Forum and the Separation Science Group
More informationAppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples
AppNote 2/2 Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples Pat Sandra Department of Organic Chemistry, University of Gent, Krijgslaan 281 S4, B-9 Gent, Belgium Erik Baltussen
More informationOptimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS
Application Note Optimization of Method Parameters for the Evaluation of USEPA Method 524.2 Using a Purge and Trap with GC/MS Objective By: Glynda Smith In this paper, an evaluation of different methods
More informationDesign and Use of Portable and Compact Sampling Systems for Mass Spectrometers. Kenneth Wright INFICON Inc, East Syracuse NY
Design and Use of Portable and Compact Sampling Systems for Mass Spectrometers. Kenneth Wright INFICON Inc, East Syracuse NY Ken.wright@inficon.com Outline Quick Review of PPA(RGA) Operation Instrumentation
More informationDEVELOPMENT OF HIGH PRESSURE MASS SPECTROMETRY FOR HANDHELD INSTRUMENTS. Kenion H. Blakeman. Chapel Hill 2015
DEVELOPMENT OF HIGH PRESSURE MASS SPECTROMETRY FOR HANDHELD INSTRUMENTS Kenion H. Blakeman A dissertation submitted to the faculty at the University of North Carolina at Chapel Hill in partial fulfillment
More informationMobile Mass Spectrometer in daily use Experiences made by non-scientific operators
Mobile Mass Spectrometer in daily use Experiences made by non-scientific operators Franziska Lange, Thomas Ludwig, Bruker Daltonik GmbH HEMS Workshop, Florida, USA September, 211 E 2 M Enhanced Environmental
More informationAgilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology
Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass
More informationCopyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.
Optimizing Sampling and Analytical Parameters for Soil Vapour Samples using Automated Thermal Desorption / Gas Chromatography / Mass Spectrometry (ATD/GC/MS) Stephen Varisco, Technical Manager, CARO Analytical
More informationMass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.
Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted
More informationHPLC-MS of the Initial Perfluorinated Surfactants and Shorter Chains:
Details of the Analytical Procedures and Reactor Configurations. HPLC-MS of the Initial Perfluorinated Surfactants and Shorter Chains: Analysis for initial PF surfactant and possible shorter-chain products
More informationMass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS
Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from
More informationTANDEM MASS SPECTROSCOPY
TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER
More informationEvaluation of a New Analytical Trap for Gasoline Range Organics Analysis
Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible
More informationAnalytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS
Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1
More informationThermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples
Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples
More informationFourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers
Fourier Transform Ion Cyclotron Resonance Mass Spectrometry MAGLAB Summer School 2015 Ryan P. Rodgers Electrospray Ionization Source Drying Gas Capillary Evaporation Chamber High Voltage Supply Charged
More informationMass Spectrometry (MS)
Mass Spectrometry (MS) Alternative names: Mass spectrometric (selective) detector (MSD) Spectrometry - methods based on interaction of matter and radiation Mass spectrometry - method based on formation
More informationAnalysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF
Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,
More informationCerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs
Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,
More informationRotating Field Mass Spectrometer (RFMS)
Rotating Field Mass Spectrometer (RFMS) Rotating Field Mass Spectrometer (RFMS) The RFMS is placed inside a Portable Sensor packaged engineered by JPL. We have fabricated a brass board and measured both
More informationExtrel Application Note
Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical
More informationValidation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System
Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method
More informationQuadrupole Miniaturization: Revisited
Quadrupole Miniaturization: Revisited Justin Jacobsky University of Pittsburgh Randall E. Pedder, Ph.D. Michael Klaric Christopher Taormina, Ph.D. Ardara Technologies L.P. Flashback to 2007 Eight years
More informationKore Overview 2013 Dr Barrie Griffiths Kore Technology Ltd.
Kore Overview 2013 Dr Barrie Griffiths Kore Technology Ltd. Kore Technology Ltd: Brief History 1982: Cambridge Mass Spectrometry: a Cambridge Consultants Spin Off formed to sell a TOF-MS-based Laser Ionisation
More informationMS Goals and Applications. MS Goals and Applications
MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis
More informationAgilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent
More informationAgilent G3212 GC-APCI Source
Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data
More informationMass Spectrometry (MS)
Kevin Burgess, February 20, 2017 1 Mass Spectrometry (MS) from chapter(s) in the recommended text A. Introduction Kevin Burgess, February 20, 2017 2 B. Components f Mass Spectrometers mass-to-charge. molecular
More informationTO-15 Checklist Determination of VOCs in Air by GC-MS
LAB ID: DATE: LAB NAME: ASSESSOR NAME: Method Number: TO-15 Checklist Determination of VOCs in Air by GC-MS SOP Number: Revision Number: SOP Date: Personnel records observed: Data records observed: Revision
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The
More informationChapter 3: Source Measurement Techniques
Chapter 3 Source Measurement Techniques Measurement Methods Method 18, Measurement of Gaseous Organic Compound Emissions by Gas Chromatography Method 25, Determination of Total Gaseous Non-Methane Organic
More informationPrima PRO Process Mass Spectrometer
APPLICATION NOTE Prima PRO Process Mass Spectrometer No. xxxx Controlling emissions by multi-component analysis of Volatile Organic Compounds (VOC) Authors: Peter Traynor, Thermo Fisher Scientific, Franklin,
More informationMAX300-BIO Bioreactor and Fermentation Gas Analyzer. Providing Solutions for Gas Analysis
+ MAX300-BIO Bioreactor and Fermentation Gas Analyzer Providing Solutions for Gas Analysis + Quadrupole Mass Spectrometry (QMS) For 50 years Extrel has supplied Quadrupole Mass Spectrometers systems and
More informationAutoresonant Ion Trap Mass Spectrometer The RGA Alternative
Autoresonant Ion Trap Mass Spectrometer The GA Alternative 1 What is a Mass Spectrometer? Ionizer Mass Spectrometer Mass Separator Magnetic Sector, Quadrupole or Ion Trap Detector Partial Pressure Measurement
More informationOver the past 15 years, efforts toward implementing
Focus: HARSH ENVIRONMENT MASS SPECTROMETRY Implementation of DART and DESI Ionization on a Fieldable Mass Spectrometer J. Mitchell Wells, Michael J. Roth, Adam D. Keil, John W. Grossenbacher, Dina R. [ustes,
More information- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma
THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which
More informationUS EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS
Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 8260 in conjunction
More informationApplication Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS
Application Note #1820230 Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS GC-MS Introduction USEPA Method 8270 [1] for semivolatiles is is used by by laboratories to measure
More informationAccurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.
Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID Application Note Volatile Organic Compounds (VOCs) Author Andrew Jones Activated
More informationMass Spectrometry. What is Mass Spectrometry?
Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance
More informationfor the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York
Mass Spectrometry for the Novice John Greaves and John Roboz (^>, yc**' CRC Press J Taylor & Francis Group Boca Raton London New York CRC Press is an imprint of the Taylor & Francis Croup, an informa business
More informationELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW
ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces
More informationDevelopment and Deployment of In-Situ Mass Spectrometers
Development and Deployment of In-Situ Mass Spectrometers R. Timothy Short Center for Ocean Technology College of Marine Science University of South Florida St. Petersburg, FL 33701-5016 phone (727) 553-3990
More informationChem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:
Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: I. BASIC ELEMENTS OF A MASS SPECTROMETER Inlet System or Chromatograph Ion Source Mass Analyzer Detector Computer II. ION SOURCES A. Electron
More informationOverview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC
AQualEER Chemistry faculty (Belgrade, Serbia, December 2017) Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC Emeritus Prof. Dr Jean-Claude Tabet 1,2,3 1 Institut Parisien de Chimie Moléculaire
More information(Refer Slide Time 00:09) (Refer Slide Time 00:13)
(Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer
More informationJianfei Peng et al. Correspondence to: Jianfei Peng Min Hu and Renyi Zhang
Supplement of Atmos. Chem. Phys., 17, 10333 10348, 2017 https://doi.org/10.5194/acp-17-10333-2017-supplement Author(s) 2017. This work is distributed under the Creative Commons Attribution 3.0 License.
More informationHEMS Low Power Carbon Nanotube Field Emission Electron Source for Chemical Ionization Mass Spectrometry
Low Power Carbon Nanotube Field Emission Electron Source for Chemical Ionization Mass Spectrometry HEMS 2015 Charles B. Parker, Erich J. Radauscher, Adam D. Keil, Mitch Wells, Jason J. Amsden, Jeffrey
More informationHeadspace Technology for GC and GC/MS: Features, benefits & applications
Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over
More informationTOC INCOMING LETTER OF TRANSMITTAL
E = Electronic H = Hardcopy *No Electronic Media Will Come in Password Protected, See SOW/PO for Acceptable Formats Vendor/Company Representative: JJ Furlong 03/27/17 Print First and Last Name Signature
More informationDETECTING AND MEASURING CHEMICAL WARFARE AGENTS IN REAL TIME USING THE TRACE ATMOSPHERIC GAS ANALYZER (TAGA)
On-Scene Coordinator Readiness Conference Phoenix, AZ November 15-19, 2004 DETECTING AND MEASURING CHEMICAL WARFARE AGENTS IN REAL TIME USING THE TRACE ATMOSPHERIC GAS ANALYZER (TAGA) David B. Mickunas
More informationULTRA-SENSITIVE SOLUTIONS FOR REAL-TIME TRACE GAS ANALYSIS
ULTRA-SENSITIVE SOLUTIONS FOR REAL-TIME TRACE GAS ANALYSIS PRODUCT PORTFOLIO PROTON TRANSFER REACTION - MASS SPECTROMETRY SERIES PTR-QiToF LoD < 1 pptv Resolution > 5,500 m/ m (FWHM) The IONICON PTR-QiTOF
More informationSTANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS
PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationApplication Note S/SL
Analysis of Ultra low Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and Chemiluminescence according to ASTM D Ultra low detection limits Excellent Sensitivity, Repeatability &
More informationEPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS
EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Christopher Borton AB SCIEX Golden, Colorado verview Described here is the analysis of
More informationElectrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
More informationIntroduction to LC-MS
Wednesday April 5, 2017 10am Introduction to LC-MS Amy Patton, MS Laboratory Manager, Pinpoint Testing, LLC Little Rock, AR DESCRIPTION: Amy Patton, laboratory manager for Pinpoint Testing, will begin
More informationVapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs
Vapor Intrusion Sampling Options: Performance Data for s, Badges, and Sorbent Tubes for VOCs Linda S. Coyne SKC Inc., 863 Valley View Road, Eighty Four, PA 1533 George Havalias, Maria C. Echarte American
More informationMS Goals and Applications. MS Goals and Applications
MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis
More informationChemistry Instrumental Analysis Lecture 34. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry
More informationChoosing the metabolomics platform
GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation
More information