1 Romanian Reports in Physics, Vol. 64, No. 4, P , 2012 NUCLEAR PHYSICS COUNT RATE EFFECTS ON PEAK SHAPE AND THROUGHPUT IN DIGITAL AND ANALOG GAMMA RAY SPECTROMETRY A. DUMITRESCU 1, GH. CǍTA-DANIL 1,2 1 Politehnica University of Bucharest, Physics Department, Splaiul Independenţei 313, , Bucharest, Romania, adrian_m_dumitrescu@yahoo.com 2 Horia Hulubei National Institute for Physics and Nuclear Engineering, Bucharest-Mǎgurele, Romania, cata-danil@physics.pub.ro Received February 7, 2012 Abstract. A comparative study of the count rate effects on peak shapes and throughput was performed using a digital and a conventional analogical gamma-ray spectrometer. Low (25%) and medium (60%) efficiency HPGe detectors were used. FWHM, FW1/5M, peak shape ratio and acquisition statistics were analyzed and compared. Key words: digital, gamma-ray, spectrometry, LN2 free cooling, HPGe, peak shape. 1. INTRODUCTION Developments of modern HPGe detectors using liquid nitrogen-free cooling methods increased the number of possible applications for high resolution gamma ray spectrometry. The use of mechanical cooling for these detectors degraded the spectroscopic performance at low energies in favor of application flexibility and customization [1]. In recent years, developments in digital spectroscopy signal processing made possible the customization of the acquisition environment through digital signal shaping and processing. This allows the user to optimize performance beyond the limitations of equivalent analogical solutions in specific applications [2, 3]. In order to determine the importance of acquisition parameter optimization and their effects on energy resolution and peak shape, in the present work, experiments were conducted using sets of purpose optimized parameters. Gamma-ray peak shape analysis is an important issue to be addressed in recent experimental spectroscopy [4]. In the present paper an analysis of count rate effects on energy resolution, peak shape and throughput is presented. The study compares peak shape and throughput characteristics for a digital desktop spectrometer and an analogical counterpart. Another goal of the presented experiments was determining the influence of detector efficiency in the comparison of the two spectrometers. Two high purity

2 958 A. Dumitrescu, Gh. Cǎta-Danil 2 germanium detectors were used for data acquisition, one with low efficiency (25%) and one with medium efficiency (60%). The detectors were cooled by using mechanical coolers. The measurements were conducted over a spectrum of count rates ranging from ~1 khz up to ~20 khz. The peaks to be analyzed were chosen in the 50 kev 1.5 MeV energy range. The recorded data showed notable differences between the two types of spectrometers both in terms of throughput and peak shape. A discussion of the obtained results is presented in Sections 4 and 5 of this paper. 2. ANALOG AND DIGITAL SPECTROMETERS Digital gamma-ray spectrometers have significantly improved since their introduction about two decades ago. They are functionally different from their analogical counterparts both in structure and in operation [3, 5]. The main two architectural differences between these spectrometric systems are the point where digitization occurs in the signal chain and the way the signal is filtered [5]. The difference in signal filtering translates into higher flexibility in acquisition customization for the digital system. This allows the user to optimize the acquisition parameters of digital spectrometers in order to better suit the application. In a digital system the signal processing starts by digitizing the analogical signal received from the detector s preamplifier, using an Analog to Digital Converter (ADC) [5, 6]. After the signal has been digitized, the stream from the ADCs is sent to a real time processing unit. This unit performs digital filtering for the incoming stream of data. The key difference from analogical signal filtering is in the type of filter used. Digital circuits are more suited for the implementation of finite impulse response filters, and in the case of gamma-ray spectroscopy, a trapezoidal filter [7, 8, 10]. The digital system s filter uses several timing parameters (L and G for each of the two filters), a signal threshold for the fast filter, and other standard parameters [7, 8]. The purpose of these parameters is to offer a large degree of flexibility in data acquisition. The different structure and functionality of analogical and digital spectrometers reflect finally in the shape of the energy spectra. In the present study are compared the two types of spectrometers, pointing out the differences in acquisition results through a range of input count rates of practical interest. 3. EXPERIMENT SETUP AND DATA COLLECTION METHODOLOGY The gamma-ray energies analyzed in this work were 59.5keV, 121.8keV, 356keV, 778keV, keV, keV and keV, obtained using a stack of four radioactive spectroscopic sources (Fig. 1) with activities ranging between 16 to 47 kbq.

3 3 Digital and analog gamma ray spectrometry 959 To analyze the effect of the detector s efficiency on results, experiments were carried out using two germanium detectors with different relative efficiencies: one with medium-high efficiency (60%) and the other one with a lower efficiency (25%) [9], both cooled by the mechanical cooler X-Cooler from ORTEC (see e.g. [9]). The measurements were organized in two sets. Acquisition parameters were experimentally determined in order to maximize the throughput in one set of experiments and to optimize the energy resolution in the other. Measurements were done increasing the count rate from 1 khz to 20 khz with about 1 khz increments step. As a result 80 different measurements were performed resulting in 160 spectra for analysis. All acquisitions were timed so that statistically relevant data could be collected at each of the given input count rates. Fig. 1 Layout of the experimental setup employed in the present study. Prior to data collection and acquisition parameter optimization, both systems were set up for the input signal. On the analog system the usual baseline restoration, gain setting and pole zero adjustments were done for the shaping amplifier for each of the two detectors. The parameters for the Spectroscopy Amplifier of analog system (gain, shaping time, and shaping method) were determined on the basis of several experimental trials [6]. Table 1 Parameters of the digital spectrometer optimizing throughput or resolution for each of the detectors used in the present study Parameter/Detector and Optimization 25GEM Throughput 25GEM Resolution 60GEM Throughput 60GEM Resolution Preamp Gain(mV/MeV) Decay Time (µs) Energy Filter Range Energy Filter Rise Time (µs) Energy Filter Flat Top (µs) Trigger Filter Threshold (kev) Trigger Filter Rise Time (µs) Trigger Filter Flat Top (µs)

4 960 A. Dumitrescu, Gh. Cǎta-Danil 4 On the digital system, the optimization of acquisition parameters was done through the software command interface of the Polaris DGF [7]. The acquisition parameters (preamp gain, decay time, rise time, flat top, trigger threshold presented in Table 1) were set up by monitoring the signal using the integrated software oscilloscope, and analyzing the resulting test spectra. Each detector and each optimization required a different set of acquisition parameters. The preamp gain was set up according to the dynamic energy range of the experiments (50 kev to 1.5 MeV). By choosing the dynamic range of the acquisition a software routine sets the preamp gain so that the chosen number of analysis channels fits the selected energy range. The preamplifier decay time is used to correct the energy of a pulse sitting on the falling slope of a previous pulse (pileup effects [7, 8]). A first estimate of the decay time was found using the dedicated software routine. A more accurate value was experimentally determined by acquiring test spectra at high count rates and examining the changes produced by this parameter in energy resolution. The rise time of the energy filter is primarily responsible for energy resolution [7]. Longer rise times result in better resolution but lower throughput. The values for the rise time were obtained by searching for the time values that lead to the lowest FWHM for the kev 60 Co peak. The flat top needs to be wide enough to accommodate the longest typical signal rise time from the detector [7]. The values for the flat top of this filter were obtained in the same way. The trigger threshold should be set as low as possible for best resolution [7, 10]. The threshold values for these sets of experiments were set just above the noise peaks. The parameters of the trigger filter were chosen experimentally to maximize pileup detection and rejection and also to minimize the noise level by discarding events associated with noise present in the input signal. 4. RESULTS AND DISCUSSION ON THROUGHPUT The throughput for digital systems is usually higher than in analogical systems because the process of event acquisition is different [3, 7]. Quantitatively, dead time is used to compare throughput between devices and experiments [6]. For each experiment dead time was computed as a sum of all the dead times associated with each event. The resulting dead times for each set of experiments were plotted in Fig. 2 as percentages of real time against input count rate. Two graphs were made in order to analyze and compare device performance for both detectors: one for energy resolution optimization (Fig. 2, panel 1) and one for event throughput optimization (Fig. 2, panel 2). The figures show that dead time rises with input count rate in an almost linear fashion. This is normal because the more incoming events are processed the bigger the load on the processing device and consequently, the larger dead times

5 5 Digital and analog gamma ray spectrometry 961 due to processing. The experimental set optimized for energy resolution produced about 20% 25% more dead time for the higher count rates compared to the one optimized for throughput. At low count rates there is 5 10% less dead time for the throughput optimized systems [10]. It can be concluded from Fig. 2 that throughput optimization for both spectrometry systems yields lower dead times compared to energy resolution optimizations. The analogical spectrometer produced slightly better throughput compared to the digital one in the case of energy resolution optimization for both detectors as shown in Fig. 2, panel 1. Generally, the advantage of a digital spectrometer is a better throughput compared to analogical counterparts. The energy resolution optimization comes at the cost of complex and time consuming calculations for pileup rejection, baseline restoration and signal filtering. Dead time in the case of the digital spectrometer acquiring data through a small efficiency detector (25GEM) is significantly lower (Fig. 2, panel 2) than all previous cases at all count rates. The rest of experimental sets using this optimization produced roughly the same throughput. It can be concluded that throughput optimization of the digital device coupled with the use of a smaller detector yielded significantly better throughput and about double the throughput compared to the energy resolution optimized experiments. This is the result both of trading throughput for energy resolution on the digital spectrometer and the use of a smaller efficiency detector which is less susceptible to noise factors. Fig. 2 Statistics for acquisitions; panel 1 Dead Time in resolution optimization; panel 2 Dead Time in resolution optimization; panel 3 Output count rate vs. input count rate.

6 962 A. Dumitrescu, Gh. Cǎta-Danil 6 A better view of the difference between resulting throughputs for the two sets of optimizations can be obtained by plotting output count rate (OCR) against input count rate (ICR) for the digital spectrometer (Fig. 2, panel 3). Optimization of the digital spectrometer for throughput results in significantly increased throughput performance, effectively doubling the number of processed events. It can be concluded that using a smaller efficiency detector results in better throughput. Both detectors lead to similar throughput while using the energy resolution optimization. 5. RESULTS AND DISCUSSION ON PEAK SHAPE Energy resolution and peak shape analysis was done for each peak at each of the considered count rates by analyzing the measured spectra. In the experimental spectra the background was subtracted then the peaks were fitted with Gaussian functions. For each peak FWHAM and FW1 /5M values were extracted and employed to evaluate the peak shape ratio defined in the present work as the ratio between FWHM and FW1/5M. The variation model of peak shape is presented in this section and plotted against the input count rate, for each of the analyzed peaks. The first experimental set recorded spectra using the analog spectrometer, the 25GEM detector and an energy resolution optimized setup for acquisition parameters. Here the FWHM is proportional to peak energy and also increases with input count rate. These results were expected as FWHM of the Gaussian fitting a gamma ray peak is proportional to the gamma-ray's total absorption peak s energy. With increasing input count rate, the peak broadens [5] and thus FWHM of the fitting Gaussian increases. The same observations can be made about FW1/5M. For the peaks located in the low energy range of the spectrum the peak shape ratio decreases more with input count rate (Fig. 3, panel 1). This means that the base of those peaks broadens more with input count rate increase compared to the high energy peaks in the same experimental set. Also it can be concluded for this set of experiments, that while FWHM increases with ICR at the same rate for all the peaks considered, FW1/5M increases at different rates depending on the energy of the gamma-ray peak (see resulting peak shapes of 59 kev, 121 kev and 356 kev peaks from Fig. 3). It is also noticeable that FW1/5M increases faster with count rate than FWHM leading to a descending trend for the peak shape ratio plotted against ICR. The best resolution was obtained by using the 25GEM detector while acquiring data through the analogical resolution optimized system. This was expected because analogical spectrometers generally offer better energy resolution than a digital desktop gamma spectroscopy system using low efficiency detectors [3]. The second experimental set recorded spectra using the analog spectrometer, the 25GEM detector and throughput optimized parameters. Results are shown in Fig. 3, panel 2. From this plot can be observed that FWHM and FW1/5M increase with ICR and are proportional to peak energy. Both peak measurements show much smaller increases with input count rate than in the case of resolution

7 7 Digital and analog gamma ray spectrometry 963 optimization. If the FWHM of all the peaks increased throughout the 20 khz range, for the energy resolution optimization experiment, with about 1keV, here the FWHM increases with approx. 0.5 kev. The same observation goes for FW1/5M which in the first set increased with about 1.5 kev, while in the second set increased with about 0.7 kev. Correlated with the small FWHM and FW1/5M values obtained in both cases this result shows that for the analogical spectrometer acquiring data from a small detector, throughput optimization yields better resolution for a large ICR range. The shape ratio is preserved much better with increasing ICRs than the first experimental set (Fig. 3, panel 1 and 2). The shape of the Gaussians does not present base broadening for the lower energy peaks (59 kev, 121 kev and 356 kev) like in the first set. Variations with count rate of the peak shape ratio are under 0.02 compared to the almost 0.1 variations in the peak shape ratio presented in Fig. 3, panel 1. This shows that the shape of the peaks in the spectra obtained here, are much more stable with ICR increase. Fig. 3 Peak shape ratio for the 25GEM detector experiments set; panel 1 analog, resolution optimized; panel 2 analog, throughput optimized; panel 3 digital, resolution optimized; panel 4 digital, throughput optimized (common legend presented in panel 2). The third experimental set recorded spectra by using the digital spectrometer, the 25GEM detector and energy resolution optimized parameters. Here, FWHM and FW1/5M don t increase with ICR at the same rate for all the considered peaks.

8 964 A. Dumitrescu, Gh. Cǎta-Danil 8 While FWHM of low energy peaks increases in value with about 2 kev the FWHM for the high energy peaks increases in value with only approx kev throughout the 20 khz ICR range. At low ICR, FWHM and FW1/5M values are comparable to the ones obtained in the first experimental set. The values show a significant increase at high count rates compared to the analog device ( kev difference at 20 khz). It can be concluded that energy resolution is better, especially at high count rates on the analog spectrometer. The lower energy peaks (59 kev, 121 kev and 356 kev) suffer significant base broadening in the middle of the ICR range with a maximum around 12 khz (Fig. 3, panel 3). This means that for ICRs in the 5 khz 10 khz interval, these peaks broaden and then, progressing to the 12 khz 17 khz the basis of the peaks narrows. It should be emphasized that this result could not be caused by a systematic measurement error as non sequential count rates were selected from the ICR range for each experiment. Further analysis on this issue is underway. The shape of high energy peaks is more stable with ICR compared to the case of the analog spectrometer. The fourth experimental set recorded spectra by using the digital spectrometer, the 25GEM detector and throughput optimized acquisition parameters. While the analogical spectrometer has slightly better performance with respect to energy resolution, the digital spectrometer shows less increase in FWHM and FW1/5M over the 20 khz ICR range. Overall, energy resolution in the case of throughput optimization on this detector is very similar between the two spectrometers. Peak shape ratio analysis shows similar results to the second experimental set (Fig. 3, panels 2 and 4). The peaks keep their shape ratio over the 20 khz ICR range. It can be concluded that using throughput optimization similar results are obtained on both spectrometers. The fifth experimental set collected spectra using the analog spectrometer, the 60GEM detector and energy resolution optimized acquisition parameters. Here the increase both in FWHM and FW1/5M with count rate is more significant compared to the measurements done with the 25GEM detector. The digital acquisition system offers slightly better resolution at lower energies. At higher energies the analogical system gains a slight advantage compared to the digital system with lower FWHM. Regarding peak shape ratio, results similar to the first experimental set were obtained (Fig. 4, panel 1). While FWHM increases with ICR at the same rate for all the considered peaks, FW1/5M increases at different rates depending on the energy of the gamma-ray peak (see peak at 121 kev). Here FW1/5M increases faster than FWHM leading to a clear descending trend for the peak shape ratio plotted against ICR. This is due to the fact that the larger volume of this detector favors creation of pileup events at the input of the analog spectrometer which lead to broadening of the peak s base. The base broadening trend over the higher ICRs is more prominent here than in all experimental sets (about a 0.15 difference in value at large count rates compared to small ones).

9 9 Digital and analog gamma ray spectrometry 965 Fig. 4 Peak shape ratio for the 60GEM detector experiments set; panel 1 analog, resolution optimized; panel 2 analog, throughput optimized; panel 3 digital, resolution optimized; panel 4 digital, throughput optimized (common legend presented in panel 2). The sixth experimental set collected spectra using the analog spectrometer, the 60GEM detector and throughput optimized parameters. FWHM and FW1/5M are similar to those obtained using the other detector in the second set. Both FWHM and FW1/5M show much smaller increases with ICR than in the case of resolution optimization in the previous experimental set. While the peak shape ratio obtained in the second experimental set presented a horizontal trend, here there is a clear descending trend with ICR increase (Fig. 4, panel 2). This is caused by the larger volume of this detector favoring creation of pileup events at the input of the analog spectrometer. This leads to broadening in the peak s base. The phenomenon is not as accentuated as in the previous experimental set (about a 0.06 difference in value at large count rates compared to small count rates). The seventh experimental set analyzed data using the digital spectrometer, the 60GEM detector and an energy resolution optimized setup for acquisition parameters. The results are shown in Fig. 4, panel 3. The most notable difference from the sixth experimental set is the fact that both FWHM and FW1/5M show much smaller increases with ICR. These results show better energy resolution over the entire ICR range compared with the sixth experimental set when the analog

10 966 A. Dumitrescu, Gh. Cǎta-Danil 10 spectrometer was optimized for resolution. This result shows that on a larger efficiency detector, the digital spectrometer performs better than its analog counterpart regarding energy resolution. This was an unlikely result as most spectrometers offer better resolution when processing a smaller number of events associated with smaller efficiency detectors [6]. Peak shape ratio was preserved to a degree similar to the results obtained with the analog spectrometer on the 25GEM detector (Fig. 3, panel 3). Peaks present negligible base broadening compared to the analog spectrometer acquiring data with the 60GEM detector. The eighth set of spectra was produced by using the digital spectrometer, the 60GEM detector and throughput optimized acquisition parameters. A notable difference from all other experimental sets occurs in the fact that both FWHM and FW1/5M show the smallest increases with ICR from all sets of data. This result confirms the fact that on a larger efficiency detector, the digital spectrometer performs better than its analog counterpart regarding peak shapes. Peak shape ratio (Fig. 4, panel 4) was preserved to a degree similar to the results obtained with the analog spectrometer on the 25GEM detector (Fig. 3, panel 4). The shape of the Gaussians presents negligible base broadening compared to the analog spectrometer acquiring data through the 60GEM detector. Compared to the equivalent 25GEM experimental set, the shape ratio is preserved over the 20 khz range in the same fashion. 6. CONCLUSIONS An experimental comparative study on ICR s effects on peak shape and throughput was done using an analog and a digital spectrometer, two detectors of different efficiencies and two optimization modes. Eight experimental sets were measured and the results showed that the best results for energy resolution were obtained using 25GEM detector, an energy resolution optimization and the analog spectrometer. Several experimental sets showed that acquisition parameters optimized for resolution can cause peak base broadening. An anomalous effect was observed for the third experimental set when the peak shape ratio value decreased and then increased with ICR, for low energy gamma peaks. This issue is currently subject for further investigation. Generally, throughput optimization yielded more stable peaks with low variations in the values of the FWHM, FW1/5M and peak shape ratio along the investigated ICR range. The digital spectrometer produced better results, compared to the analog one, when used with the 60GEM detector. The throughput optimized digital spectrometer preserved the quality of FWHM, FW1/5 and peak shape ratio throughout the 20 khz ICR range. In the future developments we expect to be able to improve the performances of the digital spectrometer due to its much higher flexibility in handling detector signals, comparatively with the classical, analog one.

11 11 Digital and analog gamma ray spectrometry 967 Acknowledgements. The authors acknowledge the financial support within the project POSDRU, PRODOC, financed through the contract POSDRU/88/1.5/S/ REFERENCES 1. R.M. Keyser, R. D. Blingham, T. R. Towmey, Improved performance in germanium detector gamma-spectrometers based on digital signal processing, Journal of Radioanalytical and Nuclear Chemistry, 276, 3, (2008). 2. R. GRZYWACZ, Applications of digital pulse processing in nuclear spectroscopy, Nuclear Instrum. Meth. Phys. Res. B 204, (2003). 3. D. T. Vo, P. A. Russo, T. E. Sampson, Comparison Between Digital Gamma-Ray Spectrometer (DSPec) and Standard Nuclear Instrumentation Methods (NIM) Systems, LA MS, UC-706 and UC-700, Los Alamos National Laboratory, A. H. Fatah, A. H. Ahmed, Analysis of Gamma-Ray Spectra Using Levenberg-Marquardt Method, World Academy of Science, Engineering and Technology, 73, G.F. Knoll, Radiation Detection and Measurement Third Edition, John Wiley & Son, Inc., New York, W. R. Leo, Techniques for Nuclear and Particle Physics Experiments, A How-to Approach, Springer- Verlag, Berlin, XIA LLC, Digital Gamma Finder (DGF) Polaris User s Manual, Version 3.40, Rev , Wuyun Xiao, Yixiang Wei, Xianyun Ai, Qi Ao, System simulation of digital pulse Spectrometer, Nuclear Instrum. Meth. Phys. Res. A 555, (2005). 9. Ortec, GEM Series HPGe Coaxial Detector System, Rev G, U.S.A, A. Dumitrescu, Comparison of a digital and an analogical gamma spectrometer at low count rates, U.P.B. Sci. Bull., Series A, 73, 4, (2011).

ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS

ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS NUCLEAR PHYSICS ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS A. DUMITRESCU 1, GH. CǍTA-DANIL 1, 2 1 Physics Department, University Politehnica of Bucharest, Splaiul Independenţei

More information

The Best Gamma-Ray Detector

The Best Gamma-Ray Detector /sec The Best Gamma-Ray Detector What is the best detector for gamma-ray spectroscopy? Amptek, Inc. provides signal processing electronics which can be used with a variety of detectors (some available

More information

Characterizing New Multi-Channel Peak Sensing ADC-Mesytec MADC-32

Characterizing New Multi-Channel Peak Sensing ADC-Mesytec MADC-32 Armenian Journal of Physics, 2016, vol. 9, issue 1, pp. 29-33 Characterizing New Multi-Channel Peak Sensing ADC-Mesytec MADC-32 A. Gyurjinyan 1,2, R. Avetisyan 1,2 1 Experimental Physics Division, Yerevan

More information

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software V5.2 Plutonium and Uranium Isotopic Analysis Software Advanced Isotopic Ratio Analysis Software for HPGe Gamma-Ray Spectra Analyzes Pu, and a wide variety of heterogeneous samples containing Pu, Am, U,

More information

EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY

EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY R.M. Keyser New Product Development, ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831 1 INTRODUCTION

More information

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER V.A. MOROZOV 1, N.V. MOROZOVA 1, T. BĂDICĂ 2, D. DELEANU 2,3, D. GHIŢĂ 2, S. PASCU 2,3 1 Joint

More information

High Purity Germanium Detector Calibration at ISOLDE

High Purity Germanium Detector Calibration at ISOLDE High Purity Germanium Detector Calibration at ISOLDE Guðmundur Kári Stefánsson Summer Student of Maria Borge September 5, 2013 Abstract: This Summer Student Project involved the test and calibration of

More information

The Shape of the Full-Energy Gamma Ray Peak in Germanium Detectors for Different Energies and Incident Geometries

The Shape of the Full-Energy Gamma Ray Peak in Germanium Detectors for Different Energies and Incident Geometries The Shape of the Full-Energy Gamma Ray Peak in Germanium Detectors for Different Energies and Incident Geometries Ronald M. Keyser ORTEC 801 South Illinois Avenue Oak Ridge, TN 37831 Abstract The accurate

More information

Efficient Algorithms for Pulse Parameter Estimation, Pulse Peak Localization And Pileup Reduction in Gamma Ray Spectroscopy M.W.Raad 1, L.

Efficient Algorithms for Pulse Parameter Estimation, Pulse Peak Localization And Pileup Reduction in Gamma Ray Spectroscopy M.W.Raad 1, L. Efficient Algorithms for Pulse Parameter Estimation, Pulse Peak Localization And Pileup Reduction in Gamma Ray Spectroscopy M.W.Raad 1, L. Cheded 2 1 Computer Engineering Department, 2 Systems Engineering

More information

AGATA preamplifier performance on large signals from a 241 Am+Be source. F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici

AGATA preamplifier performance on large signals from a 241 Am+Be source. F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici AGATA preamplifier performance on large signals from a 241 Am+Be source F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici AGATA Week - LNL (PD), Italy, 12-15 November 2007 Outline Recalls : Fast reset device

More information

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data International Journal of Advanced Research in Physical Science (IJARPS) Volume 5, Issue 10, 2018, PP 24-28 ISSN No. (Online) 2349-7882 www.arcjournals.org Comparison of the Photo-peak Efficiencies between

More information

A Digital Method for Dead Time Compensation in Nuclear Spectroscopy

A Digital Method for Dead Time Compensation in Nuclear Spectroscopy A Digital Method for Dead Time Compensation in Nuclear Spectroscopy Keyser, Ronald M., Gedcke, Dale A., Upp, Daniel L., Twomey, Timothy R., and Bingham, Russell D. ORTEC, PerkinElmer Instruments, Inc.

More information

Gamma Spectroscopy. References: Objectives:

Gamma Spectroscopy. References: Objectives: Gamma Spectroscopy References: G.F. Knoll, Radiation Detection and Measurement (John Wiley & Sons, New York, 2000) W. R. Leo, Techniques for Nuclear and Particle Physics Experiments: A How-to Approach,

More information

NUCL 3000/5030 Laboratory 2 Fall 2013

NUCL 3000/5030 Laboratory 2 Fall 2013 Lab #2: Passive Gamma Spec Measurements in Decoding Natural Radioactivity in SLC Area Objectives a. Learn basics of gamma spectroscopy b. Learn the equipment in Counting stations #4, #5 and #8 c. Apply

More information

PROFILE S and C Series

PROFILE S and C Series P-type Semi-planar and Coaxial HPGe Detectors Exceptional Resolution and Stable, Low-Energy Efficiency Comprehensive Solutions to Demanding Applications Exceptional resolution and stable, low energy efficiency

More information

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA Experimental Comparison between High Purity Germanium and Scintillator Detectors for Determining Burnup, Cooling Time and Decay Heat of Used Nuclear Fuel - 14488 Peter Jansson *, Sophie Grape *, Stephen

More information

Developing DRF for 1 X1 LaBr 3 Detector. Sangkyu Lee

Developing DRF for 1 X1 LaBr 3 Detector. Sangkyu Lee Developing DRF for 1 X1 LaBr 3 Detector Sangkyu Lee Detector Response Function Function: a function is a relation between a set of inputs and a set of permissible outputs with the property that each input

More information

An Improved U-Pu System for Field Pu Measurements

An Improved U-Pu System for Field Pu Measurements An Improved U-Pu System for Field Pu Measurements Markku J. Koskelo and Gary Gardner Canberra Industries, Inc., Meriden, CT, USA Abstract With high resolution gamma spectrometry using the 4 original Multi

More information

arxiv:nucl-ex/ v2 21 Jul 2005

arxiv:nucl-ex/ v2 21 Jul 2005 Gamma-spectrometric uranium age-dating using intrinsic efficiency calibration arxiv:nucl-ex/0506029v2 21 Jul 2005 Cong Tam Nguyen and József Zsigrai Institute of Isotopes of the Hungarian Academy of Sciences

More information

Mass of the electron m 0

Mass of the electron m 0 Mass of the electron m 0 1 Objective To determine the rest mass of the electron, m e, via γ-ray interactions (mainly Compton scattering and photoeffect) in a NaI scintillation detector. Based on the enclosed

More information

LANTHANUM BROMIDE SCINTILLATION DETECTOR FOR GAMMA SPECTROMETRY APPLIED IN INTERNAL RADIOACTIVE CONTAMINATION MEASUREMENTS

LANTHANUM BROMIDE SCINTILLATION DETECTOR FOR GAMMA SPECTROMETRY APPLIED IN INTERNAL RADIOACTIVE CONTAMINATION MEASUREMENTS U.P.B. Sci. Bull., Series A Vol. 73, Iss. 4, 211 ISSN 1223-727 LANTHANUM BROMIDE SCINTILLATION DETECTOR FOR GAMMA SPECTROMETRY APPLIED IN INTERNAL RADIOACTIVE CONTAMINATION MEASUREMENTS Mirela Angela SAIZU

More information

Figure 1. Decay Scheme for 60Co

Figure 1. Decay Scheme for 60Co Department of Physics The University of Hong Kong PHYS3851 Atomic and Nuclear Physics PHYS3851- Laboratory Manual A. AIMS 1. To learn the coincidence technique to study the gamma decay of 60 Co by using

More information

Performance of the True Coincidence Correction Method in GammaVision. Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L.

Performance of the True Coincidence Correction Method in GammaVision. Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L. Performance of the True Coincidence Correction Method in GammaVision Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L. ORTEC, PerkinElmer Instruments, Inc. 801 South Illinois Avenue, Oak Ridge,

More information

A gamma-gamma coincidence spectrometer for nuclear attribution and safeguard applications

A gamma-gamma coincidence spectrometer for nuclear attribution and safeguard applications IAEA-CN-184/174 A gamma-gamma coincidence spectrometer for nuclear attribution and safeguard applications Weihua Zhang, Jing Yi, Pawel Mekarski and Kurt Ungar Radiation Protection Bureau of Health Canada,

More information

PROFILE S and C Series P-type Semi-planar and Coaxial HPGe Detectors

PROFILE S and C Series P-type Semi-planar and Coaxial HPGe Detectors P-type Semi-planar and Coaxial HPGe Detectors Exceptional Resolution and Stable, Low-energy Efficiency Comprehensive Solutions to Demanding Applications Exceptional resolution and stable, low energy efficiency

More information

Semiconductor Detector

Semiconductor Detector Semiconductor Detector General 1. Introduction A semiconductor detector is used for a. checking the radionuclidic purity of radiopharmaceuticals and calibration sources; b. the quantitative analysis of

More information

Determination of the shielding power of different materials against gamma radiation

Determination of the shielding power of different materials against gamma radiation Determination of the shielding power of different materials against gamma radiation Chow Wing Yan, Yeung Chun Lap S.K.H Tsang Shiu Tim Secondary School Wong Ka Wing Baptist Lui Ming Choi Secondary School

More information

arxiv: v3 [nucl-ex] 12 Jan 2012

arxiv: v3 [nucl-ex] 12 Jan 2012 Fast-timing measurements in 95,96 Mo arxiv:2.539v3 [nucl-ex] 2 Jan 202 S Kisyov, S Lalkovski, N Mǎrginean 2, D Bucurescu 2, L Atanasova 3, D Balabanski 3, Gh Cata-Danil 2, I Cata-Danil 2, D Deleanu 2,

More information

arxiv: v1 [physics.ins-det] 14 Nov 2017

arxiv: v1 [physics.ins-det] 14 Nov 2017 Characteristics of GRIFFIN high-purity germanium clover detectors U. Rizwan a, A. B. Garnsworthy b, C. Andreoiu a, G. C. Ball b, A. Chester a, T. Domingo a, R. Dunlop c, G. Hackman b, E.T. Rand c, J.K.

More information

Researchers at the University of Missouri-Columbia have designed a triple crystal

Researchers at the University of Missouri-Columbia have designed a triple crystal Childress, N. L. and W. H. Miller, MCNP Analysis and Optimization of a Triple Crystal Phoswich Detector, Nuclear Instruments and Methods, Section A, 490(1-2), 263-270 (Sept 1, 2002). Abstract Researchers

More information

Measurements of photon scattering lengths in scintillator and a test of the linearity of light yield as a function of electron energy

Measurements of photon scattering lengths in scintillator and a test of the linearity of light yield as a function of electron energy Measurements of photon scattering lengths in scintillator and a test of the linearity of light yield as a function of electron energy Alexandra Huss August 31, 2013 Abstract The SNO+ experiment in Sudbury,

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies CHARACTERIZATION OF PHOSWICH WELL DETECTORS FOR RADIOXENON MONITORING Wolfgang Hennig 1, Hui Tan 1, William K. Warburton 1, Anthony Fallu-Labruyere 1, Konstantin Sabourov 1, Justin I. McIntyre 2, Matthew

More information

QuantumMCA QuantumNaI QuantumGe QuantumGold

QuantumMCA QuantumNaI QuantumGe QuantumGold QuantumMCA QuantumNaI QuantumGe QuantumGold Berkeley Nucleonics Corporation (San Rafael, CA) and Princeton Gamma Tech (Princeton, NJ) have partnered to offer gamma spectroscopy with either germanium or

More information

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 EFFICIENCY SIMULATION OF

More information

THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS. Carly Jean Smith

THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS. Carly Jean Smith THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS by Carly Jean Smith A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment

More information

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use Final report on DOE project number DE-FG07-99ID13772 High Pressure Xenon Gamma-Ray Spectrometers for Field Use Principle Investigator: Glenn K. Knoll Co-investigator: David K. Wehe, Zhong He, University

More information

ORTEC AN34 Experiment 10 Compton Scattering

ORTEC AN34 Experiment 10 Compton Scattering EQUIPMENT NEEDED FROM ORTEC 113 Preamplifier (2 ea.) TRUMP-PCI-2K MCA System including suitable PC operating Windows 98/2000/XP (other ORTEC MCAs may be used) 266 Photomultiplier Tube Base (2 ea.) 4001A/4002D

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement Fourth Edition Glenn F. Knoll Professor Emeritus of Nuclear Engineering and Radiological Sciences University of Michigan Ann Arbor, Michigan WILEY John Wiley & Sons,

More information

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Position Sensitive Germanium Detectors for the Advanced Compton Telescope Position Sensitive Germanium Detectors for the Advanced Compton Telescope R.A. Kroeger 1, W.N. Johnson 1, J.D. Kurfess 1, B.F. Phlips, P.N. Luke 3, M. Momayezi 4, W.K. Warburton 4 1 Naval Research Laboratory,

More information

GAMMA RAY SPECTROSCOPY FOR ARTIFICIAL CONTAMINATION EFFECTS EVALUATION IN LUMINISCENCE DATING OF ARTEFACTS FROM LOW DEPTH LAYERS IN SOUTHERN ROMANIA *

GAMMA RAY SPECTROSCOPY FOR ARTIFICIAL CONTAMINATION EFFECTS EVALUATION IN LUMINISCENCE DATING OF ARTEFACTS FROM LOW DEPTH LAYERS IN SOUTHERN ROMANIA * Romanian Reports in Physics, Vol. 64, No. 2, P. 381 386, 2012 NUCLEAR PHYSICS GAMMA RAY SPECTROSCOPY FOR ARTIFICIAL CONTAMINATION EFFECTS EVALUATION IN LUMINISCENCE DATING OF ARTEFACTS FROM LOW DEPTH LAYERS

More information

Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals

Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals Wolfgang Hennig, Hui Tan, William K Warburton, and Justin I McIntyre

More information

Alpha-Energies of different sources with Multi Channel Analyzer

Alpha-Energies of different sources with Multi Channel Analyzer Physical Structure of Matter Radioactivity Alpha-Energies of different sources with Multi Channel Analyzer What you can learn about Decay series Radioactive equilibrium Isotopic properties Decay energy

More information

Quality Assurance. Purity control. Polycrystalline Ingots

Quality Assurance. Purity control. Polycrystalline Ingots Quality Assurance Purity control Polycrystalline Ingots 1 Gamma Spectrometry Nuclide Identification Detection of Impurity Traces 1.1 Nuclides Notation: Atomic Mass Atomic Number Element Neutron Atomic

More information

Determination of e/m, mc 2, and the Relations Among Momentum, Velocity, and Energy of Relativistic Electrons

Determination of e/m, mc 2, and the Relations Among Momentum, Velocity, and Energy of Relativistic Electrons Determination of e/m, mc 2, and the Relations Among Momentum, Velocity, and Energy of Relativistic Electrons Edwin Ng MIT Department of Physics (Dated: November 16, 2011 Using a spherical electromagnet,

More information

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES Doru Stanga*, Pierino De Felice** *National Institute of R&D for Physics and Nuclear Engineering-Horia Hulubei(IFIN-HH)

More information

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce F.C.L. Crespi 1,2, F.Camera 1,2, N. Blasi 2, A.Bracco 1,2, S. Brambilla 2, B. Million 2, R. Nicolini 1,2, L.Pellegri 1, S. Riboldi

More information

1. Introduction. Etim I.P. 1, William E. S. 2, Ekwe S.O. 3. Licensed Under Creative Commons Attribution CC BY

1. Introduction. Etim I.P. 1, William E. S. 2, Ekwe S.O. 3. Licensed Under Creative Commons Attribution CC BY Alpha-Particle Spectroscopy and Ranges in Air Etim I.P. 1, William E. S. 2, Ekwe S.O. 3 1 Lecturer in the Department Of Physics, University of Calabar, Calabar, Nigeria 2 Concluded his Master program in

More information

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P ) Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P2522015) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik Experiment:

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement Third Edition Glenn F. Knoll Professor of Nuclear Engineering and Radiological Sciences University of Michigan Ann Arbor, Michigan John Wiley & Sons, Inc. New York/Chichester/Weinheim/Brisbane/Toronto/Singapore

More information

Stability and Characteristics of Large CZT Coplanar Electrode Detectors

Stability and Characteristics of Large CZT Coplanar Electrode Detectors 272 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 48, NO. 3, JUNE 2001 Stability and Characteristics of Large CZT Coplanar Electrode Detectors José M. Pérez, Zhong He, Senior Member, IEEE, and David K. Wehe,

More information

*Corresponding author,

*Corresponding author, Horia Hulubei National Institute of Physics and Nuclear Engineering g - IFIN HH Horia Hulubei National Institute of Physics and Nuclear Engineering, IFIN-HH ROMANIA *Corresponding author, e-mail bercea@nipne.ro

More information

Detection and measurement of gamma-radiation by gammaspectroscopy

Detection and measurement of gamma-radiation by gammaspectroscopy Detection and measurement of gamma-radiation by gammaspectroscopy Gamma-radiation is electromagnetic radiation having speed equal to the light in vacuum. As reaching a matter it interact with the different

More information

THE USE OF GE DETECTORS FOR (N,XN) CROSS-SECTION MEASUREMENTS AT INTENSE AND LOW FREQUENCY PULSED NEUTRON BEAMS. Abstract

THE USE OF GE DETECTORS FOR (N,XN) CROSS-SECTION MEASUREMENTS AT INTENSE AND LOW FREQUENCY PULSED NEUTRON BEAMS. Abstract THE USE OF GE DETECTORS FOR (N,XN) CROSS-SECTION MEASUREMENTS AT INTENSE AND LOW FREQUENCY PULSED NEUTRON BEAMS Strahinja Lukic, 1 Paule Baumann, 1 Catalin Borcea, 2 Erwin Jericha, 3 Stevan Jokic, 4 Aleksandra

More information

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48 Introduction to Environmental Measurement Techniques 2016 Radioactivity Dana Pittauer (dpittauer@marum.de) 1of 48 Introduction Radioisotopes are of interest in environmental physics for several reasons:

More information

Investigation of fast neutron spectroscopy capability of 7 Li and 6. Li enriched CLYC scintillator for nuclear physics experiments

Investigation of fast neutron spectroscopy capability of 7 Li and 6. Li enriched CLYC scintillator for nuclear physics experiments ANSRI Application of Novel Scintillators for Research and Industry 16 May, 1 th _ 6 th 16 Dublin, Ireland Investigation of fast neutron spectroscopy capability of 7 Li and 6 Li enriched CLYC scintillator

More information

DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON

DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON W. Hennig 1, H. Tan 1, W.K. Warburton 1, and J.I. McIntyre 2 XIA LLC 1, Pacific Northwest

More information

1 Introduction. 2 Method. Robert Metzger 1,*, Kenneth Van Riper 2, and George Lasche 3

1 Introduction. 2 Method. Robert Metzger 1,*, Kenneth Van Riper 2, and George Lasche 3 Uranium, radium and thorium in soils with high-resolution gamma spectroscopy, MCNP-generated efficiencies, and VRF non-linear full-spectrum nuclide shape fitting Robert Metzger 1,*, Kenneth Van Riper 2,

More information

IdentiFINDER Digital Hand Held Spectrometer & Dose Rate Meter for Portable Applications

IdentiFINDER Digital Hand Held Spectrometer & Dose Rate Meter for Portable Applications fire IdentiFINDER Digital Hand Held Spectrometer & The world s smallest spectrometer and dose rate meter designed for portable applications. safety security identifinder - CH (yellow) CZT and neutron detector

More information

SYSTEM OF MONITORING THE ATMOSPHERICAL RADON WITH AN IONIZATION CHAMBER DETECTOR TYPE IN PULSE MODE

SYSTEM OF MONITORING THE ATMOSPHERICAL RADON WITH AN IONIZATION CHAMBER DETECTOR TYPE IN PULSE MODE SYSTEM OF MONITORING THE ATMOSPHERICAL RADON WITH AN IONIZATION CHAMBER DETECTOR TYPE IN PULSE MODE Marian Romeo Călin, Adrian Cantemir Călin Horia Hulubei National Institute of Physics and Nuclear Engineering

More information

BEGe Detector studies update

BEGe Detector studies update BEGe Detector studies update Performance and analysis Dušan Budjáš Stefan Schönert Mikael Hult* MPI für Kernphysik Heidelberg * IRMM Geel MAX-PLANCK-INSTITUTNSTITUT FÜR KERNPHYSIK Outline 1. BEGe publication

More information

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy ISOCS / LabSOCS Calibration software for Gamma Spectroscopy Counts Setup Hardware Peak Shaping Parameters Rise Time Flat Top Pole-zero Number of Channels Signal Gain Setting up a detector for measurement

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator 88 The 1 st NPRU Academic Conference SCI-O11 Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator P. Limkitjaroenporn and W.Chewpraditkul Radiation Physics Laboratory,

More information

ELECTRONIC TECHNIQUES IN TIMING MEASUREMENTS FOR NUCLEAR STRUCTURE

ELECTRONIC TECHNIQUES IN TIMING MEASUREMENTS FOR NUCLEAR STRUCTURE U.P.B. Sci. Bull., Series A, Vol. 70, Iss. 4, 2008 ISSN 1223-7027 ELECTRONIC TECHNIQUES IN TIMING MEASUREMENTS FOR NUCLEAR STRUCTURE Dan Gabriel GHIŢĂ 1 Prezenta lucrare descrie în detaliu două metode

More information

Preparation of Standard Source as a Petri Dish for Plant by Using

Preparation of Standard Source as a Petri Dish for Plant by Using Preparation of Standard Source as a Petri Dish for Plant by Using 152 Eu Element Khalid H. AL-Ubaidi 1 Salam K. AL-Nasri 2 Auday T. AL-Bayati 1* 1.Department of Physics, College of Education / Ibn Al-

More information

Alpha-particle Stopping Powers in Air and Argon

Alpha-particle Stopping Powers in Air and Argon Alpha-particle Stopping Powers in Air and Argon Mohammad QH 1* and Maghdid HA 2 1 Department of Physics, College of Science, Raparin University, Sulaimanyah, Iraqi Kurdistan, Iraq 2 Department of Physics,

More information

Estimate of Large CZT Detector Absolute Efficiency

Estimate of Large CZT Detector Absolute Efficiency 2010 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 4, AUGUST 2002 Estimate of Large CZT Detector Absolute Efficiency J. M. Pérez, Member, IEEE,Z.He, Senior Member, IEEE, D.K.Wehe, Senior Member, IEEE,

More information

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system

More information

1 of :32

1 of :32 Home Page Products Price List Links & PDFs DISCONTINUED: SEE GAMMA-RAD Gamma Ray & X-Ray Spectroscopy System Hand-Held, High Efficiency NaI(Tl) Detector The GAMMA-8000 is a powerful, portable instrument

More information

Positron Annihilation Spectroscopy

Positron Annihilation Spectroscopy Positron Annihilation Spectroscopy (1) Angular Correlation θ N x, y = p x, y m C θ γ-ray (511keV ± E) 0 (2) Doppler Broadening Cp E = z 2 θ N p ~100µm 22 Na (e + Source) e - e + ~ 10-12 s Sample γ-ray

More information

TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY

TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY NUCLE A R FORENSIC S INTERN ATION A L TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY This document was designed and printed at Lawrence Livermore

More information

EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS

EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS NUCLEAR PHYSICS EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS MAGDALENA TOMA 1, OCTAVIAN SIMA 2, CARMEN CRISTACHE 1, FELICIA DRAGOLICI 1, LAURENÞIU

More information

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na Ellen Simmons 1 Contents Introduction Review of the Types of Radiation Charged Particle Radiation Detection Review of Semiconductor

More information

Efficiency and Attenuation in CdTe Detectors

Efficiency and Attenuation in CdTe Detectors Efficiency and Attenuation in CdTe Detectors Amptek Inc. Bob Redus, May 5, 00 Amptek s XR-00T-CdTe is a high performance x-ray and gamma ray detector system. Like Amptek s other XR00 products, a detector

More information

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING W. Hennig 1, H. Tan 1, A. Fallu-Labruyere 1, W. K. Warburton 1, J. I. McIntyre 2, A. Gleyzer 3 XIA, LLC 1, Pacific Northwest National Laboratory

More information

The Versatile Array of Neutron Detectors at Low Energy

The Versatile Array of Neutron Detectors at Low Energy The Versatile Array of Neutron Detectors at Low Energy S. V. Paulauskas, R. Grzywacz, M. Madurga, D. Miller, S. Padgett, W. A. Peters, and the VANDLE Collaboration Why use the time-of-flight technique?

More information

SCIENCE CHINA Physics, Mechanics & Astronomy

SCIENCE CHINA Physics, Mechanics & Astronomy SCIENCE CHINA Physics, Mechanics & Astronomy Article September 2013 Vol.56 No.9: 1735 1739 doi: 10.1007/s11433-013-5185-3 Application of wavelet transform in -ray spectra analysis YU GuoLiang 1,2*, GU

More information

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor World Journal of Nuclear Science and Technology, 2014, 4, 96-102 Published Online April 2014 in SciRes. http://www.scirp.org/journal/wjnst http://dx.doi.org/10.4236/wjnst.2014.42015 Characteristics of

More information

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC A. Specification Whole body counting method is used to detect the gamma rays emitted by radio nuclides,

More information

EXPERIMENTAL DETERMINATION OF THE URANIUM ENRICHMENT RATIO

EXPERIMENTAL DETERMINATION OF THE URANIUM ENRICHMENT RATIO NUCLER PHYSICS EXPERIMENTL DETERMINTION OF THE URNIUM ENRICHMENT RTIO. LUC Horia Hulubei National Institute for Physics and Nuclear Engineering (IFIN-HH), 407 tomistilor Street, PO Box MG-6, Magurele,

More information

ON THE OMNIPRESENT BACKGROUND GAMMA RADIATION OF THE CONTINUOUS SPECTRUM

ON THE OMNIPRESENT BACKGROUND GAMMA RADIATION OF THE CONTINUOUS SPECTRUM ON THE OMNIPRESENT BACKGROUND GAMMA RADIATION OF THE CONTINUOUS SPECTRUM R, Banjanac, D. Maletić, D. Joković, N. Veselinović, A. Dragić, V. Udovičić, I. Aničin Institute of Physics, University of Belgrade,

More information

arxiv: v1 [physics.ins-det] 3 Feb 2011

arxiv: v1 [physics.ins-det] 3 Feb 2011 Nuclear Instruments and Methods in Physics Research A 00 (2018) 1 5 Alogo.pdf Nuclear Instruments and Methods in Physics Research A Scintillation decay time and pulse shape discrimination in oxygenated

More information

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must

More information

Random Coincidence between two Independent Pulses

Random Coincidence between two Independent Pulses Random Coincidence between two Independent Pulses Sean O Brien June 16, 2006 1 Random Coincidence Rates In nuclear experimentation there are genuine coincident events, those that are detected by two or

More information

Gamma Spectroscopy. Calligaris Luca Massironi Andrea Presotto Luca. Academic Year 2006/2007

Gamma Spectroscopy. Calligaris Luca Massironi Andrea Presotto Luca. Academic Year 2006/2007 Gamma Spectroscopy Calligaris Luca Massironi Andrea Presotto Luca Academic Year 2006/2007 Abstract Here we propose the results of a number of experiments with gamma rays. In the first part we concentrated

More information

The Mössbauer Effect

The Mössbauer Effect Experimental Physics V85.0112/G85.2075 The Mössbauer Effect Spring, 2005 Tycho Sleator, David Windt, and Burton Budick Goals The main goal of this experiment is to exploit the Mössbauer effect to measure

More information

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS ANTOANETA ENE 1, I. V. POPESCU 2, T. BÃDICÃ 3, C. BEªLIU 4 1 Department of Physics, Faculty

More information

Pulse-shape shape analysis with a Broad-energy. Ge-detector. Marik Schönert. MPI für f r Kernphysik Heidelberg

Pulse-shape shape analysis with a Broad-energy. Ge-detector. Marik Schönert. MPI für f r Kernphysik Heidelberg Pulse-shape shape analysis with a Broad-energy Ge-detector Marik Barnabé é Heider Dušan Budjáš Oleg Chkvorets Stefan Schönert MPI für f r Kernphysik Heidelberg Outline 1. Motivation and goals 2. BEGe detector

More information

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production Experiment N2: Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production References: 1. Experiments in Nuclear

More information

Gamma Ray Spectrometry

Gamma Ray Spectrometry Gamma Ray Spectrometry Practical Information Document compiled by: Nafaa Reguigui September 2006. 1 Gamma Ray Spectrometry 1. OBJECTIVES 2. INTRODUCTION 3. INSTRUMENTATION BASICS 3.1. Nuclear instrumentation

More information

THE MÖSSBAUER EFFECT

THE MÖSSBAUER EFFECT THE MÖSSBAUER EFFECT Resonant gamma ray fluorescence is a useful tool in determining a variety of nuclear and solid state properties. The discovery of the Mössbauer effect greatly increased the accuracy

More information

Beta Spectroscopy. Glenn F. Knoll Radiation Detection and Measurements, John Wiley & Sons, Inc. 2000

Beta Spectroscopy. Glenn F. Knoll Radiation Detection and Measurements, John Wiley & Sons, Inc. 2000 Advanced Laboratory Experiments Universität Siegen Prof. Dr. I. Fleck Beta Spectroscopy Abstract The experiment on beta spectroscopy introduces the student into the field of special relativity and weak

More information

Deuteron activation cross section measurements at the NPI cyclotron

Deuteron activation cross section measurements at the NPI cyclotron Nuclear Physics Institute Řež EAF 2011 Deuteron activation cross section measurements at the NPI cyclotron E. Šimečková, P. Bém, M. Honusek, J. Mrázek, M. Štefánik, L. Závorka Nuclear Physics Institute

More information

Turkey s first photonuclear reactions performed at Akdeniz University: Matter, Anti-matter, Pure Energy and Alchemy

Turkey s first photonuclear reactions performed at Akdeniz University: Matter, Anti-matter, Pure Energy and Alchemy (Ülkemizin ilk fotonükleer reaksiyonu Akdeniz Üniversitesinde gerçekleştirildi: Madde, Anti-Madde, Saf Enerji ve Simya) Turkey s first photonuclear reactions performed at Akdeniz University: Matter, Anti-matter,

More information

DUNPL Preliminary Energy Calibration for Proton Detection

DUNPL Preliminary Energy Calibration for Proton Detection DUNPL Preliminary Energy Calibration for Proton Detection DUNPL Professor Alexander Komives By: Josh Wyant and Andrew Bever Report Number: TR-PD01 April 28, 2005 Throughout the semester we have been working

More information

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide Nuclear Medicine Imaging Systems: The Scintillation Camera List of Nuclear Medicine Radionuclides Tc99m 140.5 kev 6.03 hours I-131 364, 637 kev 8.06 days I-123 159 kev 13.0 hours I-125 35 kev 60.2 days

More information

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT A Monte Carlo simulation has been developed using the Geant4 software

More information

COMPTON SCATTERING OF GAMMA RAYS

COMPTON SCATTERING OF GAMMA RAYS COMPTON SCATTERING OF GAMMA RAYS v2.7 Last revised: R. A. Schumacher, January 2017 I. INTRODUCTION Compton scattering is the name given to the scattering of high-energy gamma rays from electrons. The gamma

More information

Warsaw University of Technology, Faculty of Physics. Laboratory of Nuclear Physics & Technology. Compton effect

Warsaw University of Technology, Faculty of Physics. Laboratory of Nuclear Physics & Technology. Compton effect Warsaw University of Technology, Faculty of Physics Laboratory of Nuclear Physics & Technology Compton effect Author: MSc. Eng. Dariusz Aksamit, Dariusz.Aksamit@pw.edu.pl, Faculty of Physics on the basis

More information