Measurement Uncertainty in the DTA Temperature Calibration

Size: px
Start display at page:

Download "Measurement Uncertainty in the DTA Temperature Calibration"

Transcription

1 International Journal of Pure and Applied Physics ISSN Volume 6, Number 4 (2010), pp Research India Publications Measurement Uncertainty in the DTA Temperature Calibration Yasser A. Abdelaziz, Essam M. Ibrahim and Mostafa M. Mekawy National Institute for Standards, Giza, Egypt Abstract In order to make good use of the Differential Thermal Analysis (DTA) and to obtain reliable results with good reproducibility and accuracy, it is necessary to carefully calibrate temperature measuring system of the apparatuses using standard reference materials with (99.999%) purity. The uncertainties in temperature calibration of the DTA were estimated according to the Guide to the Expression of Uncertainty in Measurement, GUM. Repeatability test was performed to access the accuracy of the instrument experimentally. The results demonstrate that the DTA temperature calibration uncertainty between 0.2 and 1 K by calibration at fixed melting points (zinc, aluminum and gold). The measurements traceability of DTA is also described in this paper. Keywords: DTA; Uncertainty; Reference Materials; Traceability. Introduction The determination of thermal properties of materials is a very important factor for a many applications at different temperatures. DTA has become one of today's routine techniques for the characterization of materials. The accuracy of the DTA measurements is affected by its temperature measurements within the DTA measurement cell to obtain reliable results with good reproducibility and accuracy, it is necessary to carefully calibrate the apparatuses using standard reference materials as described before in ASTM Standards, E793-85, standard methods [1]. DTA temperature calibration is necessary; to give a confidence in the accuracy of its data, to establish the traceability of the measured values to SI units; and to assess the uncertainty of measurement. Ultra pure reference material is the mostly wide method for the temperature calibration of DTA. The International Temperature Scale of 1990 (ITS-90) reported defined melting points of many substances [2], which can be used as reference material for DTA temperature calibration. Some papers were published, dealing with temperature calibration of DTA [3-5],

2 430 Yasser A. Abdelaziz et al but not deeply described the uncertainty calculations. In this work we use ISO-GUM [6] for determination of the uncertainty due to the systematic errors, when calibration of temperature measurements of the DTA using standard reference materials. Also we present a brief description of the DTA calibration and traceability chain. In the mathematical analysis of DTA temperature calibration, there are several factors to be considered. Those factors include measurement uncertainty, measurement resolution and purity of reference materials. These calculations can be helpful for the calibration laboratories, as one of the most important requirement for laboratory accreditation according to ISO/ IEC [7]. Experimental Sample preparation; High purity metal samples of at least % purity were obtained from Johnson Matthy Co, UK and certified to confirm their values with ITS-90 [2]. The sample was prepared by weighing 15 mg metal into Platinum crucible. The tube was connected to a GAS system. The metal sample was heated to red heat and the system evacuated to 1 mm Hg. The fused silica tube was sealed as close as possible to the sample, and the fused silica rod severed as close as possible to the bottom of the metal sample. The severed side of the sample was ground to achieve a smooth surface. Apparatus DTA (SHIMADZU DTA-50) involves heating or cooling a test sample and a reference under identical conditions, while recording any temperature difference between the sample and reference. This differential temperature is then plotted against time, or against temperature. Changes in the sample which lead to the absorption or evolution of heat can be detected relative to the reference. Differential temperatures can also arise between two samples when their response to the applied heat treatment is not identical. DTA can therefore be used to study thermal properties and phase changes which do not lead to a change in enthalpy. The baseline of the DTA curve should then exhibit discontinuities at the transition temperatures and the slope of the curve at any point will depend on the microstructural constitution at that temperature. The area under the DTA peak is due to the enthalpy change and is not affected by the heat capacity of the sample. The design of the differential thermal analysis apparatus is shown in (Fig. 1) consists of the sample holder comprising thermocouples, sample containers, ceramic or metallic block and the heating furnace. This system is also provided with a temperature programmer and a recording system. The essential requirements of the furnace are that, it should provide a stable and sufficiently large hot zone and must be able to respond rapidly to commands from the temperature programmer.

3 Measurement Uncertainty in the DTA Temperature Calibration 431 Figure 1: Schematic diagram of the DTA system. The sample holder assembly consists of a thermocouple each for the sample and reference, surrounded by a block to ensure a uniform heat distribution. The sample is contained in a small platinum crucible designed with an indentation on the base to ensure a snug fit over the thermocouple bead. The thermocouples should not be placed in direct contact with the sample to avoid contamination and degradation, although sensitivity may be compromised. The sample assembly is isolated against electrical interference from the furnace wiring with an earthed sheath, often made of platinum coated ceramic material. The sheath can also be used to contain the sample region within a controlled atmosphere or a vacuum. Measurements The temperatures of the sample and reference substance, when heated at a constant rate, rise as indicated by the two broken lines in (Fig.2). The thermally inert reference substance keeps rising in temperature at a constant rate. When endothermic change (such as fusion) takes place in the sample, on the other hand, heat energy supplied from the exterior is consumed by the change so that the temperature stops rising. Therefore, when the change is complete, substantial difference will result between the sample and the reference temperatures. This causes a large amount of heat energy to be transferred to the sample, resulting in a sharp rise in the sample temperature signal. Consequently, the sample temperature rises as high as it would be if the temperature had not stopped rising by the endothermic change. The above temperature difference ΔT (called the DTA signal) is detected and amplified to be recorded as a peak which is shown by a solid line in (Fig. 2). Under test sample temperature is measured by thermocouple. The output from the thermocouple for the reference substance is amplified and A/D converted before being sent to the CPU circuit which compares the signal from the thermocouple with the temperature program. Voltage applied to the

4 432 Yasser A. Abdelaziz et al heater of the furnace is controlled according to the comparison result. The sample temperature is displayed in digital form on the front panel and simultaneously output to the recorder or data processor. Thus, to obtain a stable baseline, the DTA-50 uses the temperature signal from the reference substance for controlling temperature also the signal from sample is shown on the display and the recorder. To detect temperature, the DTA-50 uses a unique Dumbbell-shaped detector which uses a thermocouple of Pt and Pt-10% rhodium alloy. This detector provides high sensitivity and good response Figure 2: Typical DTA Curves. Temperature Calibration There are a certified number of materials which be used as temperature calibration. These standards used under the dynamic conditions of DTA experiment, will enabled us to calibrate our instrument. We can compare our temperature data with those obtained by other instruments. Calibration of the peak areas (A) : Peak area (A) = ± K. H. m Where; K is the calibration constant and m is the sample mass. To determine K, a standard of high purity and known enthalpy of fusion ( H) is required. The appearance of the fusion peak will be affected by sample configuration [4]. The area under the peak is then measured, so K is determined and we can determine the enthalpy values for other materials by reference to our calibration standard. Therefore in DTA measurements, we should calibrate peak areas using a standard which provides a reference peak in the same temperature range as the test sample. Factors affecting DTA curves The effect of sample factor: Amount of sample:- The peak areas are proportional to the mass of the sample involved in the thermal

5 Measurement Uncertainty in the DTA Temperature Calibration 433 transition. So, we prefer to use a few milligrams of powdered solid sample. The particle size:- Phase transitions tend to be less affected by particle size vibrations, fine powders are preferable. Sample packing:- When the sample interacts with the atmosphere surrounding it, some effects on thermal records may expected. The ease of escape of volatiles may be influenced by packing density, so a reproducible method of packing the sample is desirable. There are other sample characteristics, These include its heat capacity and its thermal conductivity, which will vary with physical state or chemical constitution. Diluents:- There are materials will not react with the sample. These are used for modifying some properties of the sample. For example: We may want to bulk up a small sample in order to fill the container to an acceptable degree. The effect of the variation of instrumental factors on DTA: The heating rate: An increase in heating rate increases the procedural peak temperature and increases peak areas to a small degree. Sample holder materials: Glass, ceramic or metal containers are used in DTA. Metal containers, e.g. Shallow aluminum pans are common in DSC. Thermocouple: The sample and reference containers are in direct contact with the thermocouples. Atmosphere around sample: A flowing gas is preferable to a static atmosphere which is liable to change as soon as sample degradation or decomposition occurs. Estimate and evaluating uncertainty components The Guide to the Expression of Uncertainty in Measurement, more known as ISO- GUM [6], was elaborated by the International Organization for Standardization in order to establish a harmonized methodology for uncertainty estimation. Type A evaluation of uncertainty "Repeatability of measurements" Repeatability tests were performed by making several measurements for each metal

6 434 Yasser A. Abdelaziz et al reference material- and determining the onset temperature of melting, the results obtained are given in Table 1. Table 1: measurement repeatability of DTA temperature calibration at zinc, aluminum and gold. Metal Zinc Aluminum Gold Reference value according to ITS- 90 o C Sample No Melting temperature Using DTA C Each metal was placed in the DTA cell and cycled at least five times at a heating rate of 1 o C min - 1, and the mean melting point temperature was calculated. Although the repeatability for a specific sample has a low associated standard deviation, of between 0.02 C and 0.44 C, there was up to a 0.3 C difference in melting point value between experiments and nominal values. This difference could be attributed to the differences in the placement of the bead in the sample crucible. All of the tests reported in Table 1 were carried out with the bead in a platinum crucible. The above data set in the DTA system are used as the correction coefficients during this correction. Q = (a+bt+ct 2 ) X Qm Where; T: Temperature o C Qm: Signal before correction Q: Signal after correction The ratios of the heat quantity between the reference values and measured values were found and plotted against the temperature. Coefficients a, b and c were then

7 Measurement Uncertainty in the DTA Temperature Calibration 435 found through quadratic regression of these plots with respect to temperature. Type B evaluation of uncertainty All the following uncertainty calculation carried out experimentally at temperature calibration C (melting point of gold) using the factors affecting DTA curves, which described in this paper item 2.4. Some of these factors are not effecting in the temperature calibration at C (melting point of gold). Uncertainty in Temperature Measurement Resolution Platinum-/Platinum-Rhodium (10%) thermocouples were employed and the recorder consisted of a digital voltmeter on special software. The analog to digital conversion process which transforms probe resistance into digital values resolves to 0.01 C. Based on a rectangular distribution of the halfinterval, the uncertainty component of temperature resolution is then u(tres) = 0.01 * 0.5/ 3 (= ) Uncertainty of the Temperature Reference material The reference material has an uncertainty of ± C according to their certificates. Assuming rectangular distribution of the half interval, the uncertainty of the temperature reference standard, is then u(tref) = / 3 (= C) Heating rate In most DTA apparatus the rate of heating determines directly the linear dimensions of the temperature scale. A change, therefore, in the rate of heating would alter the temperature scale. Maintenance of a characteristic rate of heating will insure usually a constant temperature scale. Occasionally, however, due to variation in line voltage, the rate of heating alters somewhat, and consequently the scale also alters, is then u(trat) = 0.01/ 3 (= C) The existence of temperature differences within the sample due to position with respect to the thermocouples For example, it was found that by placing the sample at a distance of about 1 to 2 mm from the thermocouple junction increased the required amount of indicator from 2 to 4 mgm, which can lead to 0.1 C, is then u(tpos) = 0.1/ 3 (= C) Uncertainty in sample weight The uncertainty in the balance used in weight of the sample can be calculated from the calibration certificate (±0.40 mg). This uncertainty value experimentally equivalent to about ± C, is then u(twegt) = 0.010/1 (= C) Combined Standard Uncertainty of DTA the Temperature Calibration The standard uncertainty components and the resulting combined standard uncertainty

8 436 Yasser A. Abdelaziz et al of the DTA Temperature Calibration temperature, uc(tc), are listed in the following table. The combined uncertainty was computed as the square root of the sum of the variances with the equation, uc(tc) = [u 2 (Trel) + u 2 (Tref) + u 2 (Trat) + u 2 (Tpos) + u 2 (Twegt)] Expanded Uncertainty Utilizing a coverage factor k=2, the expanded uncertainty, U, is listed in the following table at various temperatures using the following formula. U = k * uc(tc) Table 2: Expanded Uncertainty of DTA Temperature Calibration with Coverage Factor k=2. Source of uncertainty Value ± Probability distribution Divisor c i Value ± Reference material Normal Temperature Rectangular Measurement Resolution Heating rate Rectangular sample position Rectangular Sample weight Repeatability Normal Combined uncertainty Normal Expanded uncertainty Normal (k = 2) Then the expanded uncertainty is; ±0.89 C at the melting point of gold. The traceability concept In the DTA measurements the SI units are the joule, the mole, the kilogram and the Kelvin. The definition of measurements traceability according to the International Vocabulary of Basic and General Terms in Metrology (VIM) [8]; "property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty" Traceability to the SI in chemical measurements of any other quantity requires that the measurements be made using a primary method of measurement, which is correctly applied and stated with an evaluated uncertainty. Also it possible to use indirect methods, include combinations of non-primary methods with associated uncertainties where the evaluation requires a study of the links to national or international measurement standards of each SI unit and includes an estimate of the uncertainty of the method, or comparison with reference materials,

9 Measurement Uncertainty in the DTA Temperature Calibration 437 which themselves are linked to the SI through a chain of other comparisons, culminating in a measurement using a primary method. [9] The heat and temperature measurements of the DTA systems were achieved through the SI units of joule, the mole, the kilogram and the Kelvin. All the temperature measurements are traceable to the ITS-90 fixed points [2]. Also the intensive units, J/mol, J/kg and J/(mol/K) are traceable to the measurements of Joule effect probe using a certified reference materials. Conclusion DTA measurements with an uncertainty in terms of SI units of 0.2 and 1 K are a realistic goal when the rules outlined above and the advices on calibration are followed at three melting points of reference materials. It is important to determine the different calibration factors for temperature calibration. All obtained results are very close to reference value of reference materials which gives us confidence in our results. The traceability of DTA calibration is to ITS-90 fixed points. This paper show how the uncertainty evaluations in DTA calibration, in case, the previous papers were explained only the calibration process. Finally, we can say that our DTA is an excellent instrument that, if operated carefully and accordingly to its theory of operation, can produce very accurate results. References [1] Annual Book of ASTM Standards, vol , E793-85, p. 437 and E794-85, p. 440, American Society for Testing and Materials, Philadelphia (1985). [2] Preston-Thomas, H., The International Temperature Scale of 1990, Metrologia 27, 3-10 (1990). [3] E. Gmelin, S.M. Sarge, Thermochim. Acta 347 (2000) 9. [4] A.N. Sembira t, J.G. Dunn, "High temperature calibration of DTA and DSC apparatus using encapsulated samples", Thermochimica Acta 274 (1996) [5] E. Gmelina, S.M. Sarge, "Temperature, heat and heat tow rate calibration of differential scanning calorimeters", Thermochimica Acta 347 (2000) [6] Guide to the Expression of Uncertainty in Measurement, ISO, Switzerland, [7] R. Kaarls, T.J Quinn, The Comite Consulatif pour la Quantite' de Matie're: a brief review of its origin present activities Metrologia 34 (1997) 1. [8] The International Vocabulary of Basic and General Terms in Metrology, BIPM, France (VIM). [9] C.A. Nieto de Castro*, M.J.V. Lourenc,o, M.O. Sampaio, "Calibration of a DSC: its importance for the traceability and uncertainty of thermal measurements", Thermochimica Acta 347 (2000)

10 438 Yasser A. Abdelaziz et al

Thermal Methods of Analysis

Thermal Methods of Analysis Thermal Methods of Analysis Calorie-something we know What is calorie? Can you see or touch a calorie? How is it measured? Working out in gym Change in weight Loss of calories-burning of fat? (10 km=500calories/9cal

More information

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors 1. Scope... 2 2. Introduction... 2 3. Selection of thermocouples

More information

5.5. Calibration and Test Procedures

5.5. Calibration and Test Procedures 5.5. Calibration and Test Procedures - List of Calibration Procedures 1. C-18-010-2000, Calibration procedure of melting point measuring apparatus 2. C-18-011-2000, Calibration procedure of calorimeter

More information

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed Thermal Methods of Analysis Theory, General Techniques and Applications Prof. Tarek A. Fayed 1- General introduction and theory: Thermal analysis (TA) is a group of physical techniques in which the chemical

More information

Automated volume measurement for weihts using acoustic volumeter

Automated volume measurement for weihts using acoustic volumeter IMEKO 20th TC3, 3rd TC16 and 1st TC22 International Conference Cultivating metrological knowledge 27th to 30th November, 2007. Merida, Mexico. Automated volume measurement for weihts using acoustic volumeter

More information

Home-built Apparatus for Measuring Thermal Conductivity of Glass and Polymer Materials

Home-built Apparatus for Measuring Thermal Conductivity of Glass and Polymer Materials Mater. Res. Soc. Symp. Proc. Vol. 1657 2014 Materials Research Society DOI: 10.1557/opl.2014. 375 Home-built Apparatus for Measuring Thermal Conductivity of Glass and Polymer Materials William R. Heffner

More information

Joint Committee for Traceability in Laboratory Medicine Terminology. R. Wielgosz and S. Maniguet

Joint Committee for Traceability in Laboratory Medicine Terminology. R. Wielgosz and S. Maniguet Joint Committee for Traceability in Laboratory Medicine Terminology R. Wielgosz and S. Maniguet Terminology 1. Understanding the words and phrases we are using 2. Terminology in ISO standards 3. Vocabulary

More information

Thermal Analysis measurements

Thermal Analysis measurements Thermal Analysis measurements R W McCallum Ames Laboratory And Materials Science and Engineering Phase vs Phase Field phase set of states of a macroscopic physical system that have relatively uniform chemical

More information

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating

More information

High Pressure DSC Differential Scanning Calorimeter

High Pressure DSC Differential Scanning Calorimeter High Pressure DSC Differential Scanning Calorimeter Introduction The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis technique to measure endothermic and exothermic transitions

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

Vocabulary of Metrology

Vocabulary of Metrology Vocabulary of Metrology For Understating Uncertainty and Traceability Prepared by: Kim, Sang Ho Engineering Team Leader Modal Shop, Inc A PCB Group Company SI National Metrology Institute Secondary Calibration

More information

ISO INTERNATIONAL STANDARD. Plastics Differential scanning calorimetry (DSC) Part 4: Determination of specific heat capacity

ISO INTERNATIONAL STANDARD. Plastics Differential scanning calorimetry (DSC) Part 4: Determination of specific heat capacity INTERNATIONAL STANDARD ISO 11357-4 First edition 2005-09-15 Plastics Differential scanning calorimetry (DSC) Part 4: Determination of specific heat capacity Plastiques Analyse calorimétrique différentielle

More information

Analysis of calorimetric measurement bismuth and tin system

Analysis of calorimetric measurement bismuth and tin system International Journal of Research in Engineering and Innovation Vol-1, Issue-1 (May-2017), 29-33 International Journal of Research in Engineering and Innovation (IJREI) journal home page: http://www.ijrei.com

More information

PCM specific heat capacity c p (T) measurements

PCM specific heat capacity c p (T) measurements PCM specific heat capacity c p (T) measurements DSC on PCM Workshop AIT, Vienna, 04. - 05. April 2016 Daniel Lager Engineer Energy Department - Sustainable Thermal Energy Systems AIT Austrian Institute

More information

Calorimetry: differential scanning calorimetry (DSC), isothermal titration calorimetry (ITC)

Calorimetry: differential scanning calorimetry (DSC), isothermal titration calorimetry (ITC) Calorimetry: differential scanning calorimetry (DSC), isothermal titration calorimetry (ITC) Dr. Yin Li Department of Biophysics, Medical School University of Pecs Thermal Analysis IUPAC definition - a

More information

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5

More information

Thermochimica Acta 531 (2012) Contents lists available at SciVerse ScienceDirect. Thermochimica Acta

Thermochimica Acta 531 (2012) Contents lists available at SciVerse ScienceDirect. Thermochimica Acta hermochimica Acta 531 (2012 6 11 Contents lists available at SciVerse ScienceDirect hermochimica Acta journa l h o me page: www.elsevier.com/locate/tca hermodynamic properties of potassium nitrate magnesium

More information

Study on a thermal diffusivity standard for the laser flash method measurements 1

Study on a thermal diffusivity standard for the laser flash method measurements 1 Study on a thermal diffusivity standard for the laser flash method measurements 1 M. Akoshima 2,3 and T. Baba 2 1 Paper presented at the Seventeenth European conference on Thermophysical Properties, September

More information

Uncertainty Analysis for Heat-flux DSC Measurements

Uncertainty Analysis for Heat-flux DSC Measurements Uncertainty Analysis for Heat-flux DSC Measurements metrology-focused coordinated R&D integration of national research programmes supported by the European Commission collaboration between National Measurement

More information

4. Thermometry. Temperature and Heat Flow Temperature Scales Thermometers

4. Thermometry. Temperature and Heat Flow Temperature Scales Thermometers 4. Thermometry Measuring temperature by sensation is very imprecise. That is why we need a temperature scale and a thermometer to measure temperature more accurately. Temperature and Heat Flow Temperature

More information

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Experimental Design 8.2.2013 Doc. Martti Heinonen etunimi.sukunimi@mikes.fi Outline 1.

More information

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017 MICROMATTER TECHNOLOGIES INC. #1-8333 130 th Street, Surrey, B.C. V3W 7X4, Canada Phone: +1.604.594.9720 or 1.866.966.2286 Fax: +1.604.594.9726 info@micromatter.com www.micromatter.com Technical Note 2015-01A

More information

Apparatus to measure high-temperature thermal conductivity and thermoelectric power of small specimens

Apparatus to measure high-temperature thermal conductivity and thermoelectric power of small specimens Apparatus to measure high-temperature thermal conductivity and thermoelectric power of small specimens T. Dasgupta and A. M. Umarji a Materials Research Centre, Indian Institute of Science, Bangalore-560012,

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. ESTIMATION OF UNCERTAINTY OF MEASUREMENT POLICY...

More information

A basic introduction to reference materials. POPs Strategy

A basic introduction to reference materials. POPs Strategy A basic introduction to reference materials POPs Strategy 2009-2010+ A tutorial 16 September 2009 Angelique Botha R&D metrologist Contents Why do we need reference materials? comparability of results metrological

More information

Introduction to metrology

Introduction to metrology Introduction to metrology Emma Woolliams 4 April 2017 http://www.emceoc.org About NPL Founded in 1900 World leading National Metrology Institute ~750 staff; 550+ specialists in Measurement Science plus

More information

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition INTERNATIONAL STANDARD ISO 6974-1 Second edition 2012-05-15 Natural gas Determination of composition and associated uncertainty by gas chromatography Part 1: General guidelines and calculation of composition

More information

Measurement Uncertainty, March 2009, F. Cordeiro 1

Measurement Uncertainty, March 2009, F. Cordeiro 1 Measurement Uncertainty, March 2009, F. Cordeiro 1 Uncertainty Estimation on the Quantification of Major Proteins in Milk by Liquid Chromatography Contribution to a Chemical Reference Measurement System

More information

Introduction to Blackbody Sources

Introduction to Blackbody Sources Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with

More information

Chapter 31. Thermal Methods

Chapter 31. Thermal Methods Chapter 31. Thermal Methods Thermal analysis: Physical property of a substance or its reaction products is measured as a function of temperature. * TGA: Thermogravimetric Analysis ( 熱重分析法 ) * DTA: Differential

More information

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM 83.8058 K TO 1234.93 K: METHODS AND UNCERTAINTIES Gregory F. Strouse National Institute of Standards and Technology, Gaithersburg, Maryland, USA ABSTRACT

More information

CHEM*3440. Thermal Methods. Thermogravimetry. Instrumental Components. Chemical Instrumentation. Thermal Analysis. Topic 14

CHEM*3440. Thermal Methods. Thermogravimetry. Instrumental Components. Chemical Instrumentation. Thermal Analysis. Topic 14 Thermal Methods We will examine three thermal analytical techniques: Thermogravimetric Analysis (TGA) CHEM*3440 Chemical Instrumentation Topic 14 Thermal Analysis Differential Thermal Analysis (DTA) Differential

More information

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014) SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Palenque 1 A, Bello Horizonte,. (Hereinafter called

More information

Thermal Analysis Premium

Thermal Analysis Premium Thermal Analysis Premium HP DSC 2+ STAR e System Innovative Technology Versatile Modularity Swiss Quality DSC Measurements under Pressure for Accelerated Materials Testing Double Safety System The Right

More information

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB) JAB NOTE4 ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) nd edition: January 5 01 1 st edition: March 5 00 Japan Accreditation Board (JAB) Initial edition: 00-0-5 1/ nd

More information

Measurement & Uncertainty - Basic concept and its application

Measurement & Uncertainty - Basic concept and its application Measurement & Uncertainty - Basic concept and its application Feb. 7. 2012 JongOh CHOI (choijongoh@kriss.re.kr) Knowledge and measurement When you can measure what you are speaking about, and express it

More information

Essentials of expressing measurement uncertainty

Essentials of expressing measurement uncertainty Essentials of expressing measurement uncertainty This is a brief summary of the method of evaluating and expressing uncertainty in measurement adopted widely by U.S. industry, companies in other countries,

More information

Universal Standard Protocols for Temperature and Material Characterization Calibration with Pharmaceuticals by Thermal Analysis

Universal Standard Protocols for Temperature and Material Characterization Calibration with Pharmaceuticals by Thermal Analysis Universal Standard Protocols for Temperature and Material Characterization Calibration with Pharmaceuticals by Thermal Analysis Abstract: Manik Pavan.Maheswaram, Dhruthiman.Mantheni, Shravan Thakur Singh,

More information

Doubt-Free Uncertainty In Measurement

Doubt-Free Uncertainty In Measurement Doubt-Free Uncertainty In Measurement Colin Ratcliffe Bridget Ratcliffe Doubt-Free Uncertainty In Measurement An Introduction for Engineers and Students Colin Ratcliffe United States Naval Academy Annapolis

More information

CH3511 EXPERIMENT: Determination of Thermal Properties Using a Differential Scanning Calorimeter

CH3511 EXPERIMENT: Determination of Thermal Properties Using a Differential Scanning Calorimeter CH3511 EXPERIMENT: Determination of Thermal Properties Using a Differential Scanning Calorimeter INTRODUCTION A Mettler Toledo 823E Differential Scanning Calorimeter (DSC) will be used to determine the

More information

DSC PT 10. Applications

DSC PT 10. Applications DSC PT 10 DSC PT 10 The differential scanning calorimetry method is widely used to examine and characterize substances, mixtures, and materials. This technique is internationally standardized under DIN

More information

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 NPL REPORT CMAM 73 Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 Peter Hanes October 2001 The National Physical Laboratory is operated on behalf

More information

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE Roulette, T., J. Roulette, and S. Pons. Results of ICARUS 9 Experiments Run at IMRA Europe. in Sixth International Conference on Cold Fusion, Progress in New Hydrogen Energy. 1996. Lake Toya, Hokkaido,

More information

ISO/IEC requirements. General matters of quality system, requirements for documentation and the list of the documents required for a

ISO/IEC requirements. General matters of quality system, requirements for documentation and the list of the documents required for a ISO/IEC 17025-2006 requirements. General matters of quality system, requirements for documentation and the list of the documents required for a quality system package: quality policy, quality manual, quality

More information

Temperature Measurements

Temperature Measurements Engineering 80 Spring 2015 Temperature Measurements SOURCE: http://www.eng.hmc.edu/newe80/pdfs/vishaythermdatasheet.pdf SOURCE: http://elcodis.com/photos/19/51/195143/to-92-3_standardbody to-226_straightlead.jpg

More information

ENTHALPY CHANGE CHAPTER 4

ENTHALPY CHANGE CHAPTER 4 ENTHALPY CHANGE CHAPTER 4 ENTHALPY Is the total energy of a system. E k = Kinetic energy. Vibrational Rotational Translational E due to motion H = E k + E p E P = Potential energy Attractive force b/w

More information

ME 365 EXPERIMENT 5 FIRST ORDER SYSTEM IDENTIFICATION APPLIED TO TEMPERATURE MEASUREMENT SYSTEMS

ME 365 EXPERIMENT 5 FIRST ORDER SYSTEM IDENTIFICATION APPLIED TO TEMPERATURE MEASUREMENT SYSTEMS ME 365 EXPERIMENT 5 FIRST ORDER SYSTEM IDENTIFICATION APPLIED TO TEMPERATURE MEASUREMENT SYSTEMS Objectives: In this two week experiment, we will gain familiarity with first order systems by using two

More information

A NEW MEASUREMENT AND EVALUATION METHOD FOR DSC OF PCM SAMPLES

A NEW MEASUREMENT AND EVALUATION METHOD FOR DSC OF PCM SAMPLES A NEW MEASUREMENT AND EVALUATION METHOD FOR DSC OF PCM SAMPLES H Mehling, E Günther, S Hiebler, Bavarian Center for Applied Energy Research (ZAE Bayern), Walther-Meißner-Str. 6, D-85748 Garching, Germany.

More information

Quantities and Measurement Units

Quantities and Measurement Units RCPA-AACB Chemical Pathology Course 2018 Quantities and Measurement Units Graham White Chemical Pathology SA Pathology at Flinders Medical Centre Quantities Quantity: property of a phenomenon, body, or

More information

What is measurement uncertainty?

What is measurement uncertainty? What is measurement uncertainty? What is measurement uncertainty? Introduction Whenever a measurement is made, the result obtained is only an estimate of the true value of the property being measured.

More information

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY 15th EFLM Continuing Postgraduate Course in Clinical Chemistry and Laboratory Medicine How to assess the quality of your method? Zagreb, 24-25 October 2015 HOW TO ASSESS THE MEASUREMENT UNCERTAINTY Centre

More information

11B, 11E Temperature and heat are related but not identical.

11B, 11E Temperature and heat are related but not identical. Thermochemistry Key Terms thermochemistry heat thermochemical equation calorimeter specific heat molar enthalpy of formation temperature enthalpy change enthalpy of combustion joule enthalpy of reaction

More information

15. Compare the result with the value you have taken above Compare the calculated pressure value with the actual pressure value that you have

15. Compare the result with the value you have taken above Compare the calculated pressure value with the actual pressure value that you have 105) to convert it to. 15. Compare the result with the value you have taken above. 17. 16. Compare the calculated pressure value with the actual pressure value that you have taken from the, is it the same?

More information

C q T q C T. Heat is absorbed by the system H > 0 endothermic Heat is released by the system H < 0 exothermic

C q T q C T. Heat is absorbed by the system H > 0 endothermic Heat is released by the system H < 0 exothermic PLEASE REORD ALL DATA DIRETLY INTO YOUR LAB NOTEBOOKS Introduction Heating a substance is one of the simplest processes carried out in the chemical laboratory, and is usually accompanied by a rise in the

More information

R SANAS Page 1 of 7

R SANAS Page 1 of 7 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Approved By: Chief Executive Officer:

More information

Analyzing & Testing Business Unit. Tau-R Mode for Advanced DSC Analysis Applications Newsletter 1/ /09, Dr. Stefan Schmölzer

Analyzing & Testing Business Unit. Tau-R Mode for Advanced DSC Analysis Applications Newsletter 1/ /09, Dr. Stefan Schmölzer Analyzing & Testing Business Unit Tau-R Mode for Advanced DSC Analysis Applications Newsletter 1/2009 7/09, Dr. Stefan Schmölzer Tau-R Mode for Advanced DSC Analysis Why is a correction of DSC measurements

More information

C80. Calvet Calorimeter From ambient to 300 C. A trademark of KEP Technologies group

C80. Calvet Calorimeter From ambient to 300 C. A trademark of KEP Technologies group C80 Calvet Calorimeter From ambient to 300 C A trademark of KEP Technologies group C80 SENSOR The C80 Calorimeter is one of the most powerful, yet flexible calorimeters available. The high precision Calvet

More information

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING

More information

Thermokinetic investigation of cellulose pyrolysis impact of initial and final mass on kinetic results

Thermokinetic investigation of cellulose pyrolysis impact of initial and final mass on kinetic results Journal of Analytical and Applied Pyrolysis 62 (2002) 165 177 www.elsevier.com/locate/jaap Thermokinetic investigation of cellulose pyrolysis impact of initial and final mass on kinetic results S. Völker

More information

Every physical or chemical change in matter involves a change in energy.

Every physical or chemical change in matter involves a change in energy. Sec. 2.1 Energy Objectives: 1. Explain that physical and chemical changes in matter involve transfers of energy 2. Apply the law of conservation of energy to analyze changes in matter 3. Distinguish between

More information

Experiment 3. Electrical Energy. Calculate the electrical power dissipated in a resistor.

Experiment 3. Electrical Energy. Calculate the electrical power dissipated in a resistor. Experiment 3 Electrical Energy 3.1 Objectives Calculate the electrical power dissipated in a resistor. Determine the heat added to the water by an immersed heater. Determine if the energy dissipated by

More information

Understanding Hot-Wire Anemometry

Understanding Hot-Wire Anemometry Thermal Minutes Understanding Hot-Wire Anemometry Introduction Hot-wire anemometry is a technique for measuring the velocity of fluids, and can be used in many different fields. A hot-wire anemometer consists

More information

Differential Scanning Calorimeter Analysis of Hydrino-Producing Solid Fuel

Differential Scanning Calorimeter Analysis of Hydrino-Producing Solid Fuel Differential Scanning Calorimeter Analysis of Hydrino-Producing Solid Fuel Dr. Gilbert L. Crouse, Jr. 1 Introduction A new theory of electron behavior in the atom has been developed in the last two decades

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale Temperature Scales The Celsius, Fahrenheit, and Kelvin Temperature Scales: Temperature, and Temperature Dependent on Physical Properties Physics Enhancement Programme Dr. M.H. CHAN, HKBU 9 T F T 5 T T

More information

Carbonized Electrospun Nanofiber Sheets for Thermophones

Carbonized Electrospun Nanofiber Sheets for Thermophones Supporting Information Carbonized Electrospun Nanofiber Sheets for Thermophones Ali E. Aliev 1 *, Sahila Perananthan 2, John P. Ferraris 1,2 1 A. G. MacDiarmid NanoTech Institute, University of Texas at

More information

Measurement uncertainty and legal limits in analytical measurements

Measurement uncertainty and legal limits in analytical measurements UNCERTAINTY Measurement uncertainty and legal limits in analytical measurements MIRELLA BUZOIANU, Reference Materials Laboratory, National Institute of Metrology (INM), Bucharest, Romania 1 Introduction

More information

Lecture 36: Temperatue Measurements

Lecture 36: Temperatue Measurements Lecture 36: Temperatue Measurements Contents Principle of thermocouples Materials for themocouples Cold junction compensation Compensating wires Selection of thermocouples Illustration of gas temperature

More information

novaa 800 D Atomic Absorption Spectrometer

novaa 800 D Atomic Absorption Spectrometer Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis INTERNATIONAL STANDARD IS0 14111 First edition 1997-03-15 Natural gas - Guidelines to traceability in analysis Gaz na turel - Lignes directrices pour la tracabilit6 en analyse Reference number IS0 14111

More information

Material Testing Overview (THERMOPLASTICS)

Material Testing Overview (THERMOPLASTICS) Material Testing Overview (THERMOPLASTICS) Table of Contents Thermal Conductivity... 3 Specific Heat... 4 Transition Temperature and Ejection Temperature... 5 Shear Viscosity... 7 Pressure-Volume-Temperature

More information

Calibration of Thermocouples

Calibration of Thermocouples European Association of National Metrology Institutes Calibration of Thermocouples EURAMET/cg-08/v.01 Previously EA-10/08 July 2007 Calibration Guide EURAMET/cg-08/v.01 CALIBRATION OF THERMOCOUPLES July

More information

Phase Change Diagram. Rank Solids, liquids and gases from weakest attractive forces to strongest:

Phase Change Diagram. Rank Solids, liquids and gases from weakest attractive forces to strongest: Unit 11 Kinetic molecular theory packet Page 1 of 13 Chemistry Unit 11 Kinetic Theory Unit Quiz: Test Objectives Be able to define pressure and memorize the basic pressure units. Be able to convert to/from:

More information

Section 7. Temperature Measurement

Section 7. Temperature Measurement Section 7 Temperature Measurement 7/25/2017 Engineering Measurements 7 1 Working Definition Temperature is a measure of the average kinetic energy of the molecules that make of a substance. After time,

More information

Laboratory 12: Three Thermodynamics Experiments

Laboratory 12: Three Thermodynamics Experiments Laboratory 12: Three Thermodynamics Experiments Experiment 1: Coefficient of Linear Expansion of Metals The fact that most objects expand when heated is common knowledge. The change in the linear dimensions

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3 AE 3051, Lab #16 Investigation of the Ideal Gas State Equation By: George P. Burdell Group E3 Summer Semester 000 Abstract The validity of the ideal gas equation of state was experimentally tested for

More information

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,

More information

How do you really know what the temperature is? Michael de Podesta

How do you really know what the temperature is? Michael de Podesta How do you really know what the temperature is? Michael de Podesta TECO: Madrid September 2016 GOLDEN RULE OF TALKS One talk: one thing This Talk: TWO THINGS! Thing 1: The definitions of the SI units of

More information

HFM 100 Series. Thermal Conductivity Meter for measurement of insulation and construction materials.

HFM 100 Series. Thermal Conductivity Meter for measurement of insulation and construction materials. HFM 100 Series Conforms to International Standards ASTM C518, ISO 8301, and EN 12667 Thermal Conductivity Meter for measurement of insulation and construction materials. Hot Disk TPS -160 to 1000 C HFM

More information

Recent Advances on TG-DSC Accurate Measurements 1

Recent Advances on TG-DSC Accurate Measurements 1 Recent Advances on TG-DSC Accurate Measurements 1 C. M. Santos 2, M. M. S. Ribeiro 3,4, M. J. V. Lourenço 3, F. J. V. Santos 3 and C. A. Nieto de Castro 3,5 2 FISIPE, Fibras Sintéticas de Portugal, SA,

More information

Final Report August 2010

Final Report August 2010 Bilateral Comparison of 100 pf Capacitance Standards (ongoing BIPM key comparison BIPM.EM-K14.b) between the CMI, Czech Republic and the BIPM, January-July 2009 J. Streit** and N. Fletcher* *Bureau International

More information

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY Proceedings of the 6th International Conference on Mechanics and Materials in Design, Editors: J.F. Silva Gomes & S.A. Meguid, P.Delgada/Azores, 26-30 July 2015 PAPER REF: 5376 STATISTICAL PROCESS CONTROL

More information

THERMOCOUPLE CHARACTERISTICS TRAINER

THERMOCOUPLE CHARACTERISTICS TRAINER THERMOCOUPLE CHARACTERISTICS TRAINER (Model No : ) User Manual Version 2.0 Technical Clarification /Suggestion : / Technical Support Division, Vi Microsystems Pvt. Ltd., Plot No :75,Electronics Estate,

More information

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium? 4.1 4 UNDERSTANDING THERMAL EQUILIBRIUM What is thermal equilibrium? 1. ( Heat, Temperature ) is a form of energy that flows from a hot body to a cold body. 2. The SI unit for ( heat, temperature) is Joule,

More information

Traceable Mass Determination and Uncertainty Calculation

Traceable Mass Determination and Uncertainty Calculation Traceable Mass Determination and Uncertainty Calculation Falko Hilbrunner, Thomas Fehling, Siego Mühlich, Mario Schreiber Sartorius Lab Instruments GmbH Weender Landstrasse 94-108 Goettingen, Germany ABSTRACT

More information

Calibration Traceability Guidelines

Calibration Traceability Guidelines Calibration Traceability Guidelines Check the website of accrediting bodies for the accredited laboratories individual scope of accreditation www.a2la.org ts.nist.gov/standards/accreditation/index.cfm

More information

Calorimetry. Chapter 2. Differential Scanning heat flux calorimetry

Calorimetry. Chapter 2. Differential Scanning heat flux calorimetry Chapter 2 Calorimetry In this Chapter, the technique of differential scanning heat flux calorimetry is explained. We used a salt, of which the heat capacity is well-known, NaF, to test the equipment. After

More information

International Temperature Scale of 1990 From Wikipedia, the free encyclopedia

International Temperature Scale of 1990 From Wikipedia, the free encyclopedia International Temperature Scale of 1990 From Wikipedia, the free encyclopedia The International Temperature Scale of 1990 (ITS-90) published by the Consultative Committee for Thermometry (CCT) of the International

More information

Differential Scanning Calorimetry

Differential Scanning Calorimetry Differential Scanning Calorimetry Springer-Verlag Berlin Heidelberg GmbH G. W. H. H6hne. W. F. Hemminger H.-J. Flammersheim Differential Scanning Ca lori metry 2nd revised and enlarged edition With 130

More information

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam Cryostat for Fixed-Point Calibration of Capsule-Type SPRTs I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam International Journal of Thermophysics Journal of Thermophysical Properties and Thermophysics and Its

More information

Ch 100: Fundamentals for Chemistry

Ch 100: Fundamentals for Chemistry Ch 100: Fundamentals for Chemistry Chapter 4: Properties of Matter Lecture Notes Physical & Chemical Properties Physical Properties are the characteristics of matter that can be changed without changing

More information

APPLICATION NOTE. Characterization and Classification of Recycled Polyamides by Means of Identify. Dr. Ekkehard Füglein

APPLICATION NOTE. Characterization and Classification of Recycled Polyamides by Means of Identify. Dr. Ekkehard Füglein APPLICATION NOTE Characterization and Classification of Recycled Polyamides by Dr. Ekkehard Füglein Introduction In the field of thermal analysis up to now, researchers had to compare own data with printed

More information

Name Partner. Thermal Physics. Part I: Heat of Vaporization of Nitrogen. Introduction:

Name Partner. Thermal Physics. Part I: Heat of Vaporization of Nitrogen. Introduction: Name Partner Thermal Physics Part I: Heat of Vaporization of Nitrogen Introduction: The heat of vaporization of a liquid, L v, is the energy required to vaporize (boil) a unit mass of substance. Thus if

More information

Temperature Measurement

Temperature Measurement MECE 3320 Measurements & Instrumentation Temperature Measurement Dr. Isaac Choutapalli Department of Mechanical Engineering University of Texas Pan American Introduction Temperature is one of the most

More information

Sensing, Computing, Actuating

Sensing, Computing, Actuating Sensing, Computing, ctuating Sander Stuijk (s.stuijk@tue.nl) Department of Electrical Engineering Electronic Systems 2 THERMOELECTRIC EFFECT (Chapter 5.11) 3 Thermocouple cylinder head temperature (thermocouple)

More information