1 Bureau International des Poids et Mesures Comparison of the standards for absorbed dose to water of the ARPANSA and the BIPM for 60 Co gamma radiation P J Allisy-Roberts and D T Burns BIPM J F Boas, R B Huntley and K N Wise ARPANSA Otober 2000 Pavillon de Breteuil, F SEVRES edex

2 Comparison of the standards for absorbed dose to water of the ARPANSA and the BIPM for 60 Co gamma radiation by P J Allisy-Roberts and D T Burns Bureau International des Poids et Mesures, Sèvres J F Boas, R B Huntley and K N Wise Australian Radiation Protetion and Nulear Safety Ageny, Lower Plenty Road, Yallambie, Vitoria 3085, Australia Abstrat A omparison of the standards for absorbed dose to water of the Australian Radiation Protetion and Nulear Safety Ageny and of the Bureau International des Poids et Mesures (BIPM) has been arried out in 60 Co gamma radiation. The Australian standard is based on a graphite alorimeter and the subsequent onversion from absorbed dose to graphite to absorbed dose to water using the photon fluene saling theorem. The BIPM standard is ionometri using a graphite-walled avity ionization hamber. The omparison result is (standard unertainty ). 1. Introdution An indiret omparison of the standards of absorbed dose to water of the Australian Radiation Protetion and Nulear Safety Ageny (ARPANSA), Vitoria, Australia and of the Bureau International des Poids et Mesures (BIPM) was arried out in 60 Co gamma radiation. The Australian primary standard for absorbed dose is a graphite alorimeter as desribed in [1] and [2], the absorbed dose to water for 60 Co radiation being derived from the absorbed dose to graphite using the photon fluene saling theorem. The BIPM primary standard is a graphite avity ionization hamber of panake geometry as desribed in [3]. This absorbed dose to water omparison is the first suh omparison made between the two laboratories. The omparison was undertaken using two ionization hambers belonging to the ARPANSA as transfer standards. The hambers were alibrated at the ARPANSA before and after the measurements made at the BIPM in April The result of the omparison is derived from the ratios of the alibration fators of the transfer hambers determined at the two laboratories. 1

3 2. Determination of absorbed dose to water 2.1 The BIPM standard At the BIPM, the absorbed dose rate to water is determined from where I/m W e s,a = ( I m)( W / e) s Ψ ( µ ρ ) ( 1+ ε ) Πk, (1) Dw, BIPM /,a w, en w, w, i is the mass ionization urrent measured by the standard, is the mean energy expended in dry air per ion pair formed, is the eletroni harge is the ratio of the mean mass stopping powers of graphite and air, Ψ w, is the ratio of the photon energy fluene in water and graphite at the referene point in the water phantom, ( µ en ρ) w, is the ratio of the mean mass energy-absorption oeffiient for water to that in graphite averaged over the unperturbed and perturbed photon spetrum at the referene point, ( 1+ ε ) is the ratio of the absorbed dose to the ollision omponent of kerma, at the w, referene point in water to the same ratio at the referene point in graphite, and Π k i is the produt of the other orretion fators to be applied to the standard. The ionometri method is desribed fully in [3], whih also gives the values of the physial onstants and the orretion fators k i for the BIPM standard together with their unertainties. The ombined relative standard unertainty in the absorbed dose is , the detailed unertainty budget being given in Table 1 of this report. Absorbed dose is determined at the BIPM under the referene onditions defined by the Consultative Committee for Ionizing Radiation (CCRI, previously the CCEMRI) [4]: the distane from the soure to the referene plane (the entre of the detetor) is 1 m; the field size in air at the referene plane is 10 m 10 m, the photon fluene rate at the entre of eah side of the square being 50 % of the photon fluene rate at the entre of the square; the referene depth in water is 5 g m 2. 2

4 Table 1. Physial onstants, orretion fators and relative standard unertainties for the BIPM ionometri standard for absorbed dose to water Quantity BIPM value BIPM relative standard unertainty (1) 100 u i ; (v i ) 100 u j Dry air density (2) / (kg m 3 ) W/e /(J C 1 ) s,a k av (air avity) ( en ρ) w, µ Ψ w, (photon fluene ratio) (1+ε) w, (dose to kerma ratio) k ps (PMMA envelope) < k pf (phantom window) k rn (radial non-uniformity) < k s (reombination losses) < k h (humidity) Volume (4) /m I (ionization urrent) 0.01 ; (7) 0.02 Quadrati summation Combined relative standard unertainty of D w,bipm (3) (1) In Tables 1 to 3 and 8, u i represents the Type A relative standard unertainty u A(x i ) / x i estimated by statistial means; v i represents the number of degrees of freedom; u j represents the Type B relative standard unertainty u B(x i ) / x i estimated by other means. (2) At 0 C and kpa. (3) Combined unertainty for the produt of ( W / e) s, a (4) Using standard hamber serial number CH4-1. 3

5 2.2 The ARPANSA standard Method of measurement At the ARPANSA the absorbed dose to water is derived from the alorimetri determination of the absorbed dose to graphite by the relation where D w Ψ D ( µ ρ) w w w w, ARPANSA en i = D Ψ β Πk, (2) is the absorbed dose to graphite at the referene point in graphite, is the ratio of the photon energy fluene at the referene points in water and graphite, 1 ( µ ) w en ρ is the ratio of the mean mass energy-absorption oeffiients for water and graphite for the photon energy spetra at the orresponding referene points, 1 w β = β w β where β is the ratio of the absorbed dose to the ollision omponent of kerma, at the referene point 1 in water (w) or in graphite (), and Πk i is the produt of the orretion fators to be applied to the standard. The fators ( ) w w w µ en ρ and β are derived by alulation [5], and the photon fluene ratio Ψ is determined using the "dose ratio" method [6, 7] as desribed in setion Absorbed dose to water is determined at the ARPANSA under the following referene onditions whih are slightly different from those at the BIPM: the distane from the soure to the surfae of the water phantom is 1 m; the field size at the referene plane (nominally 1.05 m from the soure) is 10.5 m 10.5 m; the referene depth in water is 5 g m Determination of absorbed dose to graphite at the ARPANSA The ARPANSA standard of absorbed dose to graphite is a Domen-type alorimeter [8] onstruted by the Österreihishes Forshungszentrum Seibersdorf (ÖFZS) as desribed by Witzani et al. [9]. The absorbed dose rate to graphite, D, at the referene point in graphite is given by ( P m) kgapkzkrn kan t D = / k. (3) 1 The different notation ompared with that of equation (1) reflets the different experimental arrangements at the BIPM and at the ARPANSA. 4

6 The physial quantities and orretion fators in equation (3) are as desribed below and are listed with their relative standard unertainties in Table 2. Table 2. Physial quantities, orretion fators and relative standard unertainties for the determination of absorbed dose to graphite at the ARPANSA Quantity ARPANSA ARPANSA relative standard unertainty value 100 s i ; (ν i ) 100 u i P (power alulation) 0.08 ; (152) 0.04 Repeatability 0.05 ;(13) m (ore mass) /g k gap (alorimeter gaps) < k z (graphite depth) ; (4) 0.03 k rn (radial non-uniformity) ; (80) 0.04 k an (axial non-uniformity) <0.01 ; (5) 0.05 k t (soure deay) 0.01 Quadrati summation Combined relative standard unertainty of D 0.13 The radiation power absorbed in the graphite ore, P This is alulated [1, 2] from voltage and resistane measurements. The alorimeter is normally operated in the quasi-isothermal mode [9] in whih the eletrial power input to the alorimeter ore in the absene of radiation is mathed as losely as possible to the antiipated radiation power. The eletri heating is swithed off at the same time as the radiation soure is swithed on, so that the rate of heating of the ore remains approximately onstant. The radiation power input an thus be determined readily against the ARPANSA working standards of resistane and voltage. The mass of the alorimeter ore, m This was measured at the ÖFZS and orreted for impurities and buoyany. Corretion fator for the alorimeter gaps, k gap The differene between the absorbed dose rate at the entre of the alorimeter ore and that at the same position in a solid graphite phantom is alulated using Monte-Carlo odes [5] and the orretion fator, k gap, is derived from these results. The ARPANSA gap orretion is alulated for a 35 mm radius field (ross-setion approximately equivalent to a 65 mm 65 mm field) inident on the alorimeter with the entre of the ore at a depth in graphite of 30 mm (5.37 g m 2 ) and at a distane of 650 mm from the soure. 5

7 Corretion fator for the referene depth in graphite, k z The desired referene depth is not ahieved exatly with the available ombination of graphite build-up plates. Thus a orretion is applied whih is obtained by interpolation from attenuation measurements. Corretion fator for radial non-uniformity of the 60 Co beam over the alorimeter ore, k rn This was obtained experimentally by measuring the radial profile of the beam using an NE 2561 thimble ionization hamber. Corretion for the axial non-uniformity of the 60 Co beam over the alorimeter ore, k an This was obtained from the departure from linearity of the measured depth-dose distribution over the alorimeter ore. Normalization fator for the referene date and time, k t The alorimeter measurements were orreted to the referene date and time of at 12:00 Australian Eastern Daylight Time 2. The half life of 60 Co was taken as d, σ = 0.5 d [10] Conversion of absorbed dose to graphite into absorbed dose to water by the "dose ratio" method at the ARPANSA w In the "dose ratio" method, the photon fluene saling theorem [11] is used to determine Ψ for a point soure of radiation by saling the phantom dimensions, measurement depths and distanes from the soure in the inverse ratio ( ) of the eletron densities of water and the graphite (using the tabulated value of water density at 20 C, g m 3, the measured graphite bulk density, g m 3, and the Z/A ratios). Under these onditions and assuming that Compton sattering is the only interation mehanism, the ratio of primary to sattered radiation energy fluenes will be the same at orresponding points, i.e. the energy spetra will have the same shape. Furthermore, the ratio of the primary photon energy fluenes at these mapped referene points will be in the inverse ratio of the square of their distanes from the soure: w 2 ( r rw ) kpgkn Ψ =, (4) where r and r w are the distanes from the soure to the referene points in graphite and water respetively, and k pg k n are orretions needed for failure to ompletely satisfy the requirements of the photon fluene saling theorem as explained in the following paragraphs. 2 AEDT = UCT+ 11 h, where AEDT is Australian Eastern Daylight Time and UCT is Universal Coordinated Time 6

8 Distanes from the soure to the referene points in graphite and water, r and r w The position of the soure entre was approximated by appliation of the inverse square law to ionization hamber urrent measurements in air along the beam axis, orreted for air attenuation, as desribed by Wise [5]. The distane from the soure to the referene point in water is fixed by the referene onditions at m, and from the saling theorem, the distane to the referene point in graphite is mm. The unertainty in the graphite density of g m 3 leads to an unertainty in the saled graphite referene distane of 1.82 mm. The saled referene depth in graphite is (31.03 ± 0.09) mm. The omponents of the unertainty in (r /r w ) 2 are onsidered in [1] and inlude ontributions from the water phantom whih also enter into the unertainty in the water-depth orretion for the transfer hamber. Corretion fators for the limitations of the photon fluene saling theorem as applied to the graphite and water phantoms, k pg and k n The fator k pg orrets for the failure to sale the graphite phantom geometry to that of the water phantom. The fator k n orrets for the small number of non-compton interations whih are not proportional to the eletron density. These fators have been evaluated by Wise [5] using Monte-Carlo simulations with orrelated sampling. Thus D w an be alulated from D provided that r, r w, k pg, k n and the physial quantities in (2) and (3) are known. In pratie, two additional orretions are applied and the absorbed dose to water at the referene point is given by D w w ( µ en ρ) β kwin air w w, ARPANSA DΨ k =. (5) The physial quantities and the orretion fators k win and k air are desribed below and are listed in Table 3 together with their relative standard unertainties. Ratios of mass energy-absorption oeffiients ( µ ) w en ρ and of the quotients of absorbed w dose to the ollision omponent of kerma, β These were obtained using published data for the oeffiients [12] and Monte-Carlo simulations of the energy spetra at the referene points in the water and graphite phantoms as desribed by Wise [5]. Corretion fator for the differene in attenuation of the front window of the water phantom and that of the same thikness of water, k win For the ARPANSA water phantom used for measurements with 60 Co radiation, the beam passes through a PMMA 3 window of thikness 2 mm and the orretion fator was evaluated theoretially [5]. Corretion fator for air attenuation over the distane between the graphite alorimeter and the water phantom, k air This was derived from the simulated energy spetrum of the radiation beam and the attenuation oeffiients of air [12] at the appropriate energies. 3 Polymethylmetharylate 7

9 Table 3. Physial quantities, orretion fators and relative standard unertainties for onversion of absorbed dose to graphite to absorbed dose to water by the "dose ratio" method Quantity ARPANSA value ARPANSA relative standard unertainty 100 u i ; (ν i ) 100 u i D (see Table 2) (r /r w ) k n (non-compton interations) < k pg (phantom geometry) < ( µ ) w en ρ β w, (dose to kerma ratio) <0.01 <0.01 k win (water equivalene) k air (attenuation) Quadrati summation Combined relative standard unertainty of D w,arpansa Proedure for the omparison 3.1 The use of transfer hambers The omparison of the ARPANSA and BIPM standards was made indiretly by means of the alibration fators N for the ARPANSA transfer hambers given by Dw = D, (6) N D w, lab w, lab I lab where D w,lab is the absorbed dose rate to water and I lab is the ionization urrent of a transfer hamber normalized to the same referene date, eah measured at the ARPANSA or the BIPM. Current measurements are orreted for the effets and influenes desribed in this setion. The D w, BIPM value is the mean of measurements performed at the BIPM under the referene onditions over a period of three months before and after this omparison. By onvention it is given at the referene date of , 0:00 UCT, as is the value of I BIPM (The value for the half life of 60 Co reommended by the IAEA [10] is used at both laboratories). The relative standard unertainty of the mean of these eight measurements is The D w, ARPANSA value is derived from the mean of 14 alorimeter measurements made sine the installation of a new 60 Co soure at the ARPANSA in Marh 1995 and orreted to a 8

10 referene date of at 12:00 AEDT. The relative standard unertainty of the distribution of these 14 measurements is The ionization hamber urrents are the mean of measurements made before and after the measurements at the BIPM, normalized to the same referene date and time. The two laboratories determine absorbed dose by methods that are quite different and the only signifiant orrelation is that due to the use of ( µ en ρ) w, or ( µ ) w en ρ. Note that these ratios are slightly different but are assumed to be fully orrelated. The unertainty of the result of the omparison is obtained by summing in quadrature the remaining unorrelated unertainties of D w,bipm, D w, ARPANSA and the ontributions arising from the use of transfer standards. The ARPANSA transfer standards used for this omparison are graphite avity ionization hambers manufatured by Nulear Enterprises (Type NE 2561, serial numbers 070 and 328). Their main harateristis are listed in Table 4. Table 4. Charateristis of the transfer hambers type NE 2561 Charateristi Serial numbers 070 and 328 Dimensions Inner diameter / mm Wall thikness / mm Cavity length / mm Cavity entre from tip / mm Eletrode Diameter / mm 1.00 Volume Air avity / m Wall Build-up ap Material Density / (g m 3 ) Material Thikness / mm graphite < 0.01% impurity 1.80 Delrin 3.87 Applied voltage Negative polarity / V 210 The alibration proedures are desribed briefly below and are disussed in more detail in [1, 2] and [13] for the ARPANSA and the BIPM, respetively. 3.2 Measurement onditions and orretions The ARPANSA water phantom is a ube of side length 300 mm, with a wall thikness of 12 mm exept for the beam entrane window whih is a 60 mm diameter irle of thikness 2 mm. The BIPM phantom is a ube of similar size with a square window of side length 150 mm and 4 mm thik. Positioning of transfer hambers in the water phantoms At the ARPANSA, the depth of the entre of the transfer hamber was adjusted to 50 mm (standard unertainty 0.05 mm) using a depth gauge and a omputerized motion ontrol system; this distane inludes the front window of the phantom. The ionization urrent is orreted to a depth of 5 g m 2 using tabulated depth dose data [14] whih give a relative 9

11 gradient of mm 1. At the BIPM, the entre of the transfer hamber is plaed at 5 g m 2 aounting for the window (its thikness and density, its distortion due to the water pressure and its non-equivalene to water in terms of interation oeffiients) and for the water density at its measured temperature, and so no orretion is required for depth. In eah laboratory, any small differene (not more than 1 mm) between the referene distane (1 m and 1.05 m for the BIPM and ARPANSA respetively) and the distane at whih the measurement is made, is orreted by the inverse square law. Waterproof sleeve Eah hamber was supplied with a thin-walled (0.5 mm) PMMA waterproof sleeve manufatured at the ARPANSA. At the ARPANSA, a orretion fator of is applied to the urrent measured using the transfer hamber in the water phantom to aount for the inreased attenuation of the PMMA sleeve over that of the same thikness of water. For onsisteny the same orretion was applied to the measurements of urrent at the BIPM. Humidity, temperature and pressure During alibration, the relative humidity at the BIPM was between 48 % and 52 %; the air temperature was around 21 C and was stable to better than 0.01 C during a series of measurements. At the ARPANSA the relative humidity an vary between 30 % and 90 %; the temperature was around 23 C and was stable to within 0.02 C during a series of measurements. At both laboratories, the temperature of an air avity in the water phantom is measured with a alibrated thermistor. The measured ionization urrent is normalized to a temperature of K and a pressure of kpa, as part of eah laboratory's measurement system. At the ARPANSA, orretions of between 0.02 % and 0.09 % (standard unertainty 0.01 %) were made to orret the results to 50 % relative humidity. These orretions were alulated from an empirial fit, as desribed in [1, 2]. No suh orretion is required at the BIPM as humidity is losely ontrolled. Colleting voltage A olleting voltage of 210 V (negative polarity) was supplied at eah laboratory. The olleting voltage was applied at the BIPM for at least 30 minutes before measurements were made. At the ARPANSA, measurement results showing an initial drift were exluded. Measurement of harge The harge Q olleted by the hambers was measured using the loal measurement system at the BIPM and at the ARPANSA. The hambers were irradiated for at least 30 minutes before measurements ommened at the BIPM. The measured urrent was orreted at the BIPM for the leakage urrent of about 0.02 %. No orretion was made for this at the ARPANSA, as the ombined effet of bakground, leakage and bias urrent was typially less than 0.01 %. The unertainty due to this is inluded in the unertainty for ionization hamber measurements. To detet any gross hanges in the transfer hambers during transport, a hek soure of 90 Sr belonging to the ARPANSA (NE2562, serial number 024) was used to irradiate the hambers in a onstant geometry. The onsequent ionization urrents measured at both the ARPANSA and the BIPM were normalized for temperature and pressure, to 50 % relative humidity as desribed above, and for radioative deay to a ommon referene date of The half life of 90 Sr was taken as 28.8 years. The mean result for the two hambers, expressed as a urrent ratio R I = I ARPANSA / I BIPM, was [15]. As disussed in [1, 2], variations within a range of about 0.3 % have been noted for 90 Sr measurements with the NE2561 hambers at 10

12 the ARPANSA. At this level of unertainty, no hanges due to transport were observed and no differene between the measurement systems of the two laboratories ould be identified. Reombination and polarity The value of the reombination orretion fator for the NE 2561 hambers measured and applied at the ARPANSA is [1, 2]. For onsisteny, as initial reombination predominates and is independent of the dose rate, the same value was applied at the BIPM, the measurement onditions being similar. The transfer hambers were used with negative polarity at both laboratories and no orretion fator was applied for the polarity effet. Radial non-uniformity No orretion to the measurements was made at either the ARPANSA or the BIPM for the radial non-uniformity of the beam over the ross-setion of the sensitive volume of the transfer hambers. For NE 2561 hambers this effet is less than 0.02 % at eah laboratory. The transfer hamber orretion fators and their assoiated unertainties are summarized in Table 5. Table 5. Corretions and ombined relative standard unertainties for transfer hamber measurements at the ARPANSA and at the BIPM in their respetive water phantoms Measurement Value applied at the ARPANSA and the BIPM Relative standard unertainty 100 u /x Combined unertainty (1) of the ratio ARPANSA BIPM I BIPM /I ARPANSA Referene distane Depth position and orretion (2) Waterproof sleeve Temperature, pressure and humidity orretions Charge measurement Reombination Radial non-uniformity Deay orretion Combined relative standard unertainty (1) Unertainties of the orrelated values have been removed. (2) At the ARPANSA, the orretion shown is to the referene depth of 5 g m 2, the ombined unertainty arises from the atual depth set using the gauge (0.02 %), the front window thikness (0.02 %) and the ambient water temperature, 17 C to 23 C (0.01 %). At the BIPM the hamber is plaed at the referene depth so no orretion is needed; the ombined unertainty arises from the phantom window, the water density orretion and the depth position measurement unertainty. 11

13 4. Results of the omparison Reproduibility of measurements The short-term relative standard deviation of the mean ionization urrent, measured by eah transfer hamber, was at the ARPANSA (three series eah of 50 measurements of 400 pc for eah hamber) and at the BIPM (four series eah of 30 measurements of 1000 pc for eah hamber). The differenes in the urrents using the 90 Sr hek soure measured by the ARPANSA before and after alibration at the BIPM are ompatible with these short-term variations as shown in Table 6. Table 6. Stability of the ARPANSA transfer hambers and hek soure measurements I Sr, ARPANSA, /pa Ratio Date Chamber 070 Chamber /328 Marh May/June Relative differene % % % Table 7 gives the relevant values for the alulation of omparison,, expressed in the form R Dw Dw Dw,ARPANSA N D, w using (6), and the results of the R = N N. (7) Dw,BIPM In the stated unertainty u of R Dw, the orrelated unertainties arising from the use of transfer hambers and from ( µ en ρ) w, have been removed. Table 8 summarizes the unertainty omponents. The omparison result, , is taken from the mean value for both transfer hambers, with a ombined standard unertainty of The differene ( ) between the results RDw for the two hambers is ompatible with the statistial unertainty ( ) of the harge measurements. Table 7. Calibration fators and the results of the omparison Transfer hamber D w,arpansa / (mgy s 1 ) I ARPANSA / pa N Dw,ARPANSA/ (Gy µc 1 ) D w,bipm / (mgy s 1 ) I BIPM / pa N Dw,BIPM/ (Gy µc 1 ) RDw u NE NE mean values

14 Table 8. Estimated relative standard unertainties of the alibration fator, of the transfer hambers and of Quantity R D w Relative standard unertainty ARPANSA BIPM N Dw, Combined unertainty of R D w 100 u i 100 u j 100 u i 100 u j 100 u i 100 u j Absorbed dose rate to water (Tables 1 and 3) Transfer hamber measurements (Table 5) Relative standard unertainties of Relative standard unertainty of N Dw, lab R 0.29 Dw The two transfer hambers were also used in a onurrent omparison of air kerma standards [15], from whih the result was R K = Consequently it is possible to obtain the ratio N D w N K, as shown in Table 9, to hek the onsisteny of the hambers. The values obtained at eah laboratory, orreted for the fators desribed in this report, reflet the onsisteny of the hambers ( ), and the differene between the ARPANSA and the BIPM values ( ) reflets the differene in the omparison results for absorbed dose to water (1.0024) and air kerma (1.0028). Table 9. Comparison of N D w N K ratios at the ARPANSA and the BIPM Laboratory Transfer hamber N Dw / (Gy µc 1 ) N K / (Gy µc 1 ) N D w N K ARPANSA BIPM

15 5. Conlusion The ratio of the ARPANSA and BIPM primary standard determinations of absorbed dose to water is , with a standard unertainty of The result will be used as the basis for an entry to the BIPM key omparison database and the determination of degrees of equivalene between the ten national metrology institutes (NMIs) whih have finalized suh omparisons. The standard unertainty of the distribution of the results of the BIPM omparisons for these ten NMIs, shown in Figure 1 is Figure 1. International omparison of absorbed dose to water standards in 60 Co γ radiation Dw, NMI / Dw, BIPM ARP BEV BNM ENEA LSDG NIST NMi NPL NRC PTB National Metrology Institute Aknowledgements We would like to thank L H Kotler (ARPANSA) for modelling the ARPANSA 60 Co soure and for the initial Monte-Carlo simulation of the spetrum, and M. Boutillon and J. Miles for their helpful improvements to the text. Referenes [1] Huntley R B, Boas J F and Van der Gaast, H (1998) The 1997 determination of the Australian standards of exposure and absorbed dose at 60 Co, ARL/TR 126, ISSN ARL, Vitoria, Australia, 60 pages. [2] Huntley R B, Wise K N and Boas J F (2000) The Australian Standard of Absorbed Dose, Proeedings of the NPL Calorimetry Workshop, 8-10 Deember 1999, in press. [3] Boutillon M and Perrohe A-M (1993) Ionometri determination of absorbed dose to water for obalt-60 gamma rays, Phys. Med. Biol

16 [4] BIPM (1979) Comparaisons d'étalons de dose absorbée, BIPM Comité Consultatif pour les Etalons de Mesure des Rayonnements Ionisants, Setion (I) vol 4 (Paris/ Offilib) p RI(6). [5] Wise K N (2000) Monte Carlo methods used to develop the Australian absorbed dose standard, ARPANSA Tehnial Report ARPANSA/TR (in preparation). [6] Burns J E and Dale J W G (1990) Conversion of absorbed dose alibration from graphite into water, NPL Report RSA (EXT) 7 April 1990 Teddington U.K., 188 pages. [7] Burns J E (1994) Absorbed dose alibrations in high energy photon beams at the National Physial Laboratory: onversion proedure, Phys. Med. Biol [8] Domen S R and Lamperti P J (1974) A heat-loss ompensated alorimeter: theory, design and performane, J. Res. NBS 78A [9] Witzani J, Duftshmid K E, Strahotinsky Ch and Leitner, A (1984) A graphite absorbed dose alorimeter in the quasi-isothermal mode of operation, Metrologia [10] IAEA (1991), X- and gamma-ray standards for detetor alibration, IAEA-TEC-DOC-619, 157 pages. [11] Pruitt J S and Loevinger R (1982) The photon-fluene saling theorem for Compton sattered radiation, Med. Phys [12] Higgins P D, Attix F H, Hubbell J H, Seltzer S M, Berger M J and Sibata C H (1992) Mass energy-transfer and mass energy-absorption oeffiients, inluding in-flight positron annihilation for photon energies 1 kev to 100 MeV, NIST Report NISTIR-4812, Gaithersburg MD, 69 pages. [13] Boutillon M (1996) Measurement onditions used for the alibration of ionization hambers at the BIPM, Rapport BIPM-96/1, 19 pages. [14] BJR/IPEMB (1996) Central axis depth dose data for use in radiotherapy: Brit. J. Radiol. Suppl. 25, 188 pages. [15] Allisy-Roberts P J, Boutillon M, Boas J F and Huntley R B (1998) Comparison of the air kerma standards of the ARPANSA and the BIPM for 60 Co γ rays, Rapport BIPM-98/4, 10 pages. (Otober 2000) 15

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Rapport BIPM 02/01 Bureau International des Poids et Mesures Comparison of the standards of air kerma of the Yugoslavia and the BIPM for 60 Co γ rays by P.J. Allisy-Roberts, D.T. Burns, C Kessler BIPM

More information

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, S. Vörös 2 and B. Hofstetter-Boillat 2 1

More information

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays Final Report 2010-01-12 Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays C. Kessler 1, P.J. Allisy-Roberts 1, D.T. Burns 1, A.S. Guerra 2,

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays D.T. Burns, W. Tiefenböck* and J. Witzani* Bureau International des Poids et Mesures, Pavillon

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays D T Burns 1, C Kessler 1, E Mainegra-Hing 2, H Shen 2 and M R McEwen 2 1 Bureau International

More information

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation ey comparison BIPM.RI(I)-1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation C. essler 1, D. Burns 1,. Wang 2,Y. Fan 2,S. Jin 2, X. Yang 2 1 Bureau International des Poids

More information

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays Rapport BIPM-9712 Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays by P.J. Allisy-Roberts and M. Boutillon Bureau International des Poids et Mesures, F-92312 Sevres Cedex,

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de

More information

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays Rapport BIPM-96/6 Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays by P.J. Allisy-Roberts and M. Boutillon Bureau International des Poids et Mesures, F-92312 Sevres

More information

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, F. Delaunay 2, M. Donois 2 1 Bureau International des Poids

More information

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays D T Burns Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312

More information

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1) Rapport BIPM92/1 Determination of absorbed dose to water for 60Co by the scaling theorem M. Boutillon and A.M. Perroche (1) Bureau International des Poids et Mesures, F92312 Sevres Cedex Abstract A simple

More information

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays C. Kessler 1, P.J. Allisy-Roberts 1 D.T. Burns 1, P. Roger 1, L.A. de Prez 2, J.A. de Pooter 2 P.M.G.

More information

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays C Kessler 1, D T Burns 1, P Roger 1, V Sochor 2 1 Bureau International des Poids et

More information

SHIELDING MATERIALS FOR HIGH-ENERGY NEUTRONS

SHIELDING MATERIALS FOR HIGH-ENERGY NEUTRONS SHELDNG MATERALS FOR HGH-ENERGY NEUTRONS Hsiao-Hua Hsu Health Physis Measurements Group Los Alamos National Laboratory Los Alamos, New Mexio, 87545 USA Abstrat We used the Monte Carlo transport ode Los

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

"Research Note" ANALYSIS AND OPTIMIZATION OF A FISSION CHAMBER DETECTOR USING MCNP4C AND SRIM MONTE CARLO CODES *

Research Note ANALYSIS AND OPTIMIZATION OF A FISSION CHAMBER DETECTOR USING MCNP4C AND SRIM MONTE CARLO CODES * Iranian Journal of Siene & Tehnology, Transation A, Vol. 33, o. A3 Printed in the Islami Republi of Iran, 9 Shiraz University "Researh ote" AALYSIS AD OPTIMIZATIO OF A FISSIO CHAMBER DETECTOR USIG MCP4C

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties

More information

Atomic and Nuclear Physics

Atomic and Nuclear Physics Atomi and Nulear Physis X-ray physis Compton effet and X-ray physis LD Physis Leaflets P6.3.7. Compton effet: Measuring the energy of the sattered photons as a funtion of the sattering angle Objets of

More information

Atomic and Nuclear Physics

Atomic and Nuclear Physics Atomi and Nulear Physis X-ray physis Compton effet and X-ray physis LD Physis Leaflets P6.3.7. Compton effet: Measuring the energy of the sattered photons as a funtion of the sattering angle Objets of

More information

A. Shirani*and M. H. Alamatsaz

A. Shirani*and M. H. Alamatsaz IJST (013) A1: 9-34 Iranian Journal of Siene & Tehnology http://www.shirazu.a.ir/en Calulion of exposure buildup fators for point isotropi gamma ray soures in strified spherial shields of wer surrounded

More information

Monte Carlo Simulation of Electron and Radiative Emission from Silicon Diodes

Monte Carlo Simulation of Electron and Radiative Emission from Silicon Diodes SIMULATION OF SEMICONDUCTOR DEVICES AND PROCESSES Vol. 4 Edited by W. Fihtner, D. Aemmer - Zurih (Switzerland) September 12-14,1991 - Hartung-Gorre 521 Monte Carlo Simulation of Eletron and Radiative Emission

More information

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017 COMPARISON OF ABSORBED DOSE TO WATER AT HIGH PHOTON ENERGIES AT A REFERENCE FACILITY BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017 Bureau International des Poids et Mesures Pavillon

More information

Abstract An indirect comparison of the standards for reference air kerma rate for

Abstract An indirect comparison of the standards for reference air kerma rate for Comparison BIPM.RI(I)-K8 of high dose-rate Ir-19 brachyerapy standards for reference air kerma rate of e VSL and e BIPM J.T. Alvarez 1, J. A. de Pooter, C. Andersen 3, A.H.L. Aalbers, P.J. Allisy-Roberts

More information

Heat exchangers: Heat exchanger types:

Heat exchangers: Heat exchanger types: Heat exhangers: he proess of heat exhange between two fluids that are at different temperatures and separated by a solid wall ours in many engineering appliations. he devie used to implement this exhange

More information

THEORETICAL PROBLEM No. 3 WHY ARE STARS SO LARGE?

THEORETICAL PROBLEM No. 3 WHY ARE STARS SO LARGE? THEORETICAL PROBLEM No. 3 WHY ARE STARS SO LARGE? The stars are spheres of hot gas. Most of them shine beause they are fusing hydrogen into helium in their entral parts. In this problem we use onepts of

More information

Non-Markovian study of the relativistic magnetic-dipole spontaneous emission process of hydrogen-like atoms

Non-Markovian study of the relativistic magnetic-dipole spontaneous emission process of hydrogen-like atoms NSTTUTE OF PHYSCS PUBLSHNG JOURNAL OF PHYSCS B: ATOMC, MOLECULAR AND OPTCAL PHYSCS J. Phys. B: At. Mol. Opt. Phys. 39 ) 7 85 doi:.88/953-75/39/8/ Non-Markovian study of the relativisti magneti-dipole spontaneous

More information

Part G-4: Sample Exams

Part G-4: Sample Exams Part G-4: Sample Exams 1 Cairo University M.S.: Eletronis Cooling Faulty of Engineering Final Exam (Sample 1) Mehanial Power Engineering Dept. Time allowed 2 Hours Solve as muh as you an. 1. A heat sink

More information

DIGITAL DISTANCE RELAYING SCHEME FOR PARALLEL TRANSMISSION LINES DURING INTER-CIRCUIT FAULTS

DIGITAL DISTANCE RELAYING SCHEME FOR PARALLEL TRANSMISSION LINES DURING INTER-CIRCUIT FAULTS CHAPTER 4 DIGITAL DISTANCE RELAYING SCHEME FOR PARALLEL TRANSMISSION LINES DURING INTER-CIRCUIT FAULTS 4.1 INTRODUCTION Around the world, environmental and ost onsiousness are foring utilities to install

More information

Measuring & Inducing Neural Activity Using Extracellular Fields I: Inverse systems approach

Measuring & Inducing Neural Activity Using Extracellular Fields I: Inverse systems approach Measuring & Induing Neural Ativity Using Extraellular Fields I: Inverse systems approah Keith Dillon Department of Eletrial and Computer Engineering University of California San Diego 9500 Gilman Dr. La

More information

Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation

Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation P. J. Allisy-Roberts, C. Kessler, D.T. Burns Bureau International des Poids et Mesures, F-92312 Sèvres

More information

Class XII - Physics Electromagnetic Waves Chapter-wise Problems

Class XII - Physics Electromagnetic Waves Chapter-wise Problems Class XII - Physis Eletromagneti Waves Chapter-wise Problems Multiple Choie Question :- 8 One requires ev of energy to dissoiate a arbon monoxide moleule into arbon and oxygen atoms The minimum frequeny

More information

E γ. Electromagnetic Radiation -- Photons. 2. Mechanisms. a. Photoelectric Effect: photon disappears. b. Compton Scattering: photon scatters

E γ. Electromagnetic Radiation -- Photons. 2. Mechanisms. a. Photoelectric Effect: photon disappears. b. Compton Scattering: photon scatters III. letromagneti Radiation -- Photons. Mehanisms a. Photoeletri ffet: γ photon disappears b. Compton Sattering: γ photon satters. Pair Prodution: γ e ± pair produed C. Photoeletri ffet e Sine photon is

More information

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation

More information

The Determination of Surface Albedo of Thailand Using Satellite Data

The Determination of Surface Albedo of Thailand Using Satellite Data The Determination of Surfae Albedo of Thailand Using Satellite Data Serm Janjai 1,*, Wathara Wanvong 2 and Jarungsaeng Laksanaboonsong 1 1 Laboratory of Tropial Atmospheri Physis, Department of Physis,

More information

A model for measurement of the states in a coupled-dot qubit

A model for measurement of the states in a coupled-dot qubit A model for measurement of the states in a oupled-dot qubit H B Sun and H M Wiseman Centre for Quantum Computer Tehnology Centre for Quantum Dynamis Griffith University Brisbane 4 QLD Australia E-mail:

More information

Calibration of Piping Assessment Models in the Netherlands

Calibration of Piping Assessment Models in the Netherlands ISGSR 2011 - Vogt, Shuppener, Straub & Bräu (eds) - 2011 Bundesanstalt für Wasserbau ISBN 978-3-939230-01-4 Calibration of Piping Assessment Models in the Netherlands J. Lopez de la Cruz & E.O.F. Calle

More information

Millennium Relativity Acceleration Composition. The Relativistic Relationship between Acceleration and Uniform Motion

Millennium Relativity Acceleration Composition. The Relativistic Relationship between Acceleration and Uniform Motion Millennium Relativity Aeleration Composition he Relativisti Relationship between Aeleration and niform Motion Copyright 003 Joseph A. Rybzyk Abstrat he relativisti priniples developed throughout the six

More information

Answers to Coursebook questions Chapter J2

Answers to Coursebook questions Chapter J2 Answers to Courseook questions Chapter J 1 a Partiles are produed in ollisions one example out of many is: a ollision of an eletron with a positron in a synhrotron. If we produe a pair of a partile and

More information

Theory. Coupled Rooms

Theory. Coupled Rooms Theory of Coupled Rooms For: nternal only Report No.: R/50/TCR Prepared by:. N. taey B.., MO Otober 00 .00 Objet.. The objet of this doument is present the theory alulations to estimate the reverberant

More information

Combined Electric and Magnetic Dipoles for Mesoband Radiation, Part 2

Combined Electric and Magnetic Dipoles for Mesoband Radiation, Part 2 Sensor and Simulation Notes Note 53 3 May 8 Combined Eletri and Magneti Dipoles for Mesoband Radiation, Part Carl E. Baum University of New Mexio Department of Eletrial and Computer Engineering Albuquerque

More information

Supplementary Information. Infrared Transparent Visible Opaque Fabrics (ITVOF) for Personal Cooling

Supplementary Information. Infrared Transparent Visible Opaque Fabrics (ITVOF) for Personal Cooling Supplementary Information Infrared Transparent Visible Opaque Fabris (ITVOF) for Personal Cooling Jonathan K. Tong 1,Ɨ, Xiaopeng Huang 1,Ɨ, Svetlana V. Boriskina 1, James Loomis 1, Yanfei Xu 1, and Gang

More information

Wave Propagation through Random Media

Wave Propagation through Random Media Chapter 3. Wave Propagation through Random Media 3. Charateristis of Wave Behavior Sound propagation through random media is the entral part of this investigation. This hapter presents a frame of referene

More information

The Concept of Mass as Interfering Photons, and the Originating Mechanism of Gravitation D.T. Froedge

The Concept of Mass as Interfering Photons, and the Originating Mechanism of Gravitation D.T. Froedge The Conept of Mass as Interfering Photons, and the Originating Mehanism of Gravitation D.T. Froedge V04 Formerly Auburn University Phys-dtfroedge@glasgow-ky.om Abstrat For most purposes in physis the onept

More information

RADIATION DOSIMETRY AND MEDICAL PHYSICS CALCULATIONS USING MCNP 5

RADIATION DOSIMETRY AND MEDICAL PHYSICS CALCULATIONS USING MCNP 5 RADIATION DOSIMETRY AND MEDICAL PHYSICS CALCULATIONS USING MCNP 5 A Thesis by RANDALL ALEX REDD Submitted to the Offie of Graduate Studies of Texas A&M University in partial fulfillment of the requirements

More information

New Methods for Measuring the hermal Emissivity of Semi-transparent and Opaque Materials By D. Demange, M. Bejet, and B. Dufour ONERA - DMSC - Fort de Palaiseau, Chemin de la Hunière 91761 Palaiseau, Frane

More information

QCLAS Sensor for Purity Monitoring in Medical Gas Supply Lines

QCLAS Sensor for Purity Monitoring in Medical Gas Supply Lines DOI.56/sensoren6/P3. QLAS Sensor for Purity Monitoring in Medial Gas Supply Lines Henrik Zimmermann, Mathias Wiese, Alessandro Ragnoni neoplas ontrol GmbH, Walther-Rathenau-Str. 49a, 7489 Greifswald, Germany

More information

+Ze. n = N/V = 6.02 x x (Z Z c ) m /A, (1.1) Avogadro s number

+Ze. n = N/V = 6.02 x x (Z Z c ) m /A, (1.1) Avogadro s number In 1897, J. J. Thomson disovered eletrons. In 1905, Einstein interpreted the photoeletri effet In 1911 - Rutherford proved that atoms are omposed of a point-like positively harged, massive nuleus surrounded

More information

Failure Assessment Diagram Analysis of Creep Crack Initiation in 316H Stainless Steel

Failure Assessment Diagram Analysis of Creep Crack Initiation in 316H Stainless Steel Failure Assessment Diagram Analysis of Creep Crak Initiation in 316H Stainless Steel C. M. Davies *, N. P. O Dowd, D. W. Dean, K. M. Nikbin, R. A. Ainsworth Department of Mehanial Engineering, Imperial

More information

Mean Activity Coefficients of Peroxodisulfates in Saturated Solutions of the Conversion System 2NH 4. H 2 O at 20 C and 30 C

Mean Activity Coefficients of Peroxodisulfates in Saturated Solutions of the Conversion System 2NH 4. H 2 O at 20 C and 30 C Mean Ativity Coeffiients of Peroxodisulfates in Saturated Solutions of the Conversion System NH 4 Na S O 8 H O at 0 C and 0 C Jan Balej Heřmanova 5, 170 00 Prague 7, Czeh Republi balejan@seznam.z Abstrat:

More information

UTC. Engineering 329. Proportional Controller Design. Speed System. John Beverly. Green Team. John Beverly Keith Skiles John Barker.

UTC. Engineering 329. Proportional Controller Design. Speed System. John Beverly. Green Team. John Beverly Keith Skiles John Barker. UTC Engineering 329 Proportional Controller Design for Speed System By John Beverly Green Team John Beverly Keith Skiles John Barker 24 Mar 2006 Introdution This experiment is intended test the variable

More information

Calculation of Desorption Parameters for Mg/Si(111) System

Calculation of Desorption Parameters for Mg/Si(111) System e-journal of Surfae Siene and Nanotehnology 29 August 2009 e-j. Surf. Si. Nanoteh. Vol. 7 (2009) 816-820 Conferene - JSSS-8 - Calulation of Desorption Parameters for Mg/Si(111) System S. A. Dotsenko, N.

More information

Panel Session on Data for Modeling System Transients Insulated Cables

Panel Session on Data for Modeling System Transients Insulated Cables Panel Session on Data for Modeling System Transients Insulated Cables Bjørn Gustavsen SINTEF Energy Researh N-7465 Trondheim, Norway bjorn.gustavsen@energy.sintef.no Abstrat: The available EMTP-type programs

More information

The universal model of error of active power measuring channel

The universal model of error of active power measuring channel 7 th Symposium EKO TC 4 3 rd Symposium EKO TC 9 and 5 th WADC Workshop nstrumentation for the CT Era Sept. 8-2 Kosie Slovakia The universal model of error of ative power measuring hannel Boris Stogny Evgeny

More information

Acoustic Attenuation Performance of Helicoidal Resonator Due to Distance Change from Different Cross-sectional Elements of Cylindrical Ducts

Acoustic Attenuation Performance of Helicoidal Resonator Due to Distance Change from Different Cross-sectional Elements of Cylindrical Ducts Exerpt rom the Proeedings o the COMSOL Conerene 1 Paris Aousti Attenuation Perormane o Helioidal Resonator Due to Distane Change rom Dierent Cross-setional Elements o Cylindrial Duts Wojieh ŁAPKA* Division

More information

A Longitudinal Aerodynamic Data Repeatability Study for a Commercial Transport Model Test in the National Transonic Facility

A Longitudinal Aerodynamic Data Repeatability Study for a Commercial Transport Model Test in the National Transonic Facility NASA Tehnial Paper 3522 A Longitudinal Aerodynami Data Repeatability Study for a Commerial Transport Model Test in the National Transoni Faility R. A. Wahls and J. B. Adok Langley Researh Center Hampton,

More information

Temperature-Gradient-Driven Tearing Modes

Temperature-Gradient-Driven Tearing Modes 1 TH/S Temperature-Gradient-Driven Tearing Modes A. Botrugno 1), P. Buratti 1), B. Coppi ) 1) EURATOM-ENEA Fusion Assoiation, Frasati (RM), Italy ) Massahussets Institute of Tehnology, Cambridge (MA),

More information

Comparisons between IRD and BIPM exposure and air-kerma standards. for cobalt-60 gamma rays. by C.E. de Almeida* and M.-T. Niatel

Comparisons between IRD and BIPM exposure and air-kerma standards. for cobalt-60 gamma rays. by C.E. de Almeida* and M.-T. Niatel Rapport BIPM-86/12 Comparisons between IRD and BIPM exposure and air-kerma standards for cobalt-60 gamma rays by C.E. de Almeida* and M.-T. Niatel Bureau International des Poids et Mesures, F-92310 Sevres

More information

Application of the Dyson-type boson mapping for low-lying electron excited states in molecules

Application of the Dyson-type boson mapping for low-lying electron excited states in molecules Prog. Theor. Exp. Phys. 05, 063I0 ( pages DOI: 0.093/ptep/ptv068 Appliation of the Dyson-type boson mapping for low-lying eletron exited states in moleules adao Ohkido, and Makoto Takahashi Teaher-training

More information

Chapter 9. The excitation process

Chapter 9. The excitation process Chapter 9 The exitation proess qualitative explanation of the formation of negative ion states Ne and He in He-Ne ollisions an be given by using a state orrelation diagram. state orrelation diagram is

More information

Supporting Information for

Supporting Information for Eletroni Supplementary Material (ESI) for Nanosale This journal is The Royal Soiety of Chemistry 013 Supporting Information for Exitation polarization modulation in loalization mirosopy allows to resolve

More information

COMPUTER METHODS FOR THE DETERMINATION OF THE CRITICAL PARAMETERS OF POLLUTED INSULATORS

COMPUTER METHODS FOR THE DETERMINATION OF THE CRITICAL PARAMETERS OF POLLUTED INSULATORS COMPUTER METHODS FOR THE DETERMINATION OF THE CRITICAL PARAMETERS OF POLLUTED INSULATORS I. F. GONOS S. A. SUFLIS F. V. TOPALIS I.A. STATHOPULOS NATIONAL TECHNICAL UNIVERSITY OF ATHENS DEPARTMENT OF ELECTRICAL

More information

EYAL'C- \TION OF TELEMETRY RECORDS RECEIVED

EYAL'C- \TION OF TELEMETRY RECORDS RECEIVED j 1 AS -SR-El 85- -...?- EYAL'C- \TON OF TELEMETRY RECORDS RECEED FROM EXPLORER \- 1ND POXEER Y THE UNERSTY OF CHCAGO LABORATORES FOR APPLED SCENCES LAS-SR-El85 Otober 196 EALUATON OF TELEMETRY RECORDS

More information

Vector Field Theory (E&M)

Vector Field Theory (E&M) Physis 4 Leture 2 Vetor Field Theory (E&M) Leture 2 Physis 4 Classial Mehanis II Otober 22nd, 2007 We now move from first-order salar field Lagrange densities to the equivalent form for a vetor field.

More information

RESEARCH CONCERNING THE LOCAL LOSS OF STABILITY UNDER EXTERNAL PRESSURE AND TENSION OF OIL INDUSTRY TUBULARS

RESEARCH CONCERNING THE LOCAL LOSS OF STABILITY UNDER EXTERNAL PRESSURE AND TENSION OF OIL INDUSTRY TUBULARS ГОДИШНИК НА МИННО-ГЕОЛОЖКИЯ УНИВЕРСИТЕТ СВ. ИВАН РИЛСКИ, Том 48, Св. I, Геология и геофизика, 25 ANNUAL OF THE UNIVERSITY OF MINING AND GEOLOGY ST. IVAN RILSKI, Vol. 48, Part I, Geology and Geophysis,

More information

Comparison of Alternative Equivalent Circuits of Induction Motor with Real Machine Data

Comparison of Alternative Equivalent Circuits of Induction Motor with Real Machine Data Comparison of Alternative Equivalent Ciruits of Indution Motor with Real Mahine Data J. radna, J. auer, S. Fligl and V. Hlinovsky Abstrat The algorithms based on separated ontrol of the motor flux and

More information

Supplementary information for: All-optical signal processing using dynamic Brillouin gratings

Supplementary information for: All-optical signal processing using dynamic Brillouin gratings Supplementary information for: All-optial signal proessing using dynami Brillouin gratings Maro Santagiustina, Sanghoon Chin 2, Niolay Primerov 2, Leonora Ursini, Lu Thévena 2 Department of Information

More information

Duct Acoustics. Chap.4 Duct Acoustics. Plane wave

Duct Acoustics. Chap.4 Duct Acoustics. Plane wave Chap.4 Dut Aoustis Dut Aoustis Plane wave A sound propagation in pipes with different ross-setional area f the wavelength of sound is large in omparison with the diameter of the pipe the sound propagates

More information

New Potential of the. Positron-Emission Tomography

New Potential of the. Positron-Emission Tomography International Journal of Modern Physis and Appliation 6; 3(: 39- http://www.aasit.org/journal/ijmpa ISSN: 375-387 New Potential of the Positron-Emission Tomography Andrey N. olobuev, Eugene S. Petrov,

More information

ELECTROMAGNETIC WAVES WITH NONLINEAR DISPERSION LAW. P. М. Меdnis

ELECTROMAGNETIC WAVES WITH NONLINEAR DISPERSION LAW. P. М. Меdnis ELECTROMAGNETIC WAVES WITH NONLINEAR DISPERSION LAW P. М. Меdnis Novosibirs State Pedagogial University, Chair of the General and Theoretial Physis, Russia, 636, Novosibirs,Viljujsy, 8 e-mail: pmednis@inbox.ru

More information

Physics 486. Classical Newton s laws Motion of bodies described in terms of initial conditions by specifying x(t), v(t).

Physics 486. Classical Newton s laws Motion of bodies described in terms of initial conditions by specifying x(t), v(t). Physis 486 Tony M. Liss Leture 1 Why quantum mehanis? Quantum vs. lassial mehanis: Classial Newton s laws Motion of bodies desribed in terms of initial onditions by speifying x(t), v(t). Hugely suessful

More information

Nuclear Shell Structure Evolution Theory

Nuclear Shell Structure Evolution Theory Nulear Shell Struture Evolution Theory Zhengda Wang (1) Xiaobin Wang () Xiaodong Zhang () Xiaohun Wang () (1) Institute of Modern physis Chinese Aademy of SienesLan Zhou P. R. China 70000 () Seagate Tehnology

More information

Heat propagation and stability in a small high T superconductor. coil

Heat propagation and stability in a small high T superconductor. coil Ž. Physia C 310 1998 372 376 Heat propagation and stability in a small high T superondutor oil T. Kiss a,), V.S. Vysotsky a, H. Yuge a, H. Saho a, Yu.A. Ilyin a, M. Takeo a, K. Watanabe b, F. Irie a Graduate

More information

Developing Excel Macros for Solving Heat Diffusion Problems

Developing Excel Macros for Solving Heat Diffusion Problems Session 50 Developing Exel Maros for Solving Heat Diffusion Problems N. N. Sarker and M. A. Ketkar Department of Engineering Tehnology Prairie View A&M University Prairie View, TX 77446 Abstrat This paper

More information

An Adaptive Optimization Approach to Active Cancellation of Repeated Transient Vibration Disturbances

An Adaptive Optimization Approach to Active Cancellation of Repeated Transient Vibration Disturbances An aptive Optimization Approah to Ative Canellation of Repeated Transient Vibration Disturbanes David L. Bowen RH Lyon Corp / Aenteh, 33 Moulton St., Cambridge, MA 138, U.S.A., owen@lyonorp.om J. Gregory

More information

Improvements in the Modeling of the Self-ignition of Tetrafluoroethylene

Improvements in the Modeling of the Self-ignition of Tetrafluoroethylene Exerpt from the Proeedings of the OMSOL onferene 010 Paris Improvements in the Modeling of the Self-ignition of Tetrafluoroethylene M. Bekmann-Kluge 1 *,. errero 1, V. Shröder 1, A. Aikalin and J. Steinbah

More information

Universities of Leeds, Sheffield and York

Universities of Leeds, Sheffield and York promoting aess to White Rose researh papers Universities of Leeds, Sheffield and York http://eprints.whiterose.a.uk/ This is an author produed version of a paper published in Journal of Composites for

More information

Directional Coupler. 4-port Network

Directional Coupler. 4-port Network Diretional Coupler 4-port Network 3 4 A diretional oupler is a 4-port network exhibiting: All ports mathed on the referene load (i.e. S =S =S 33 =S 44 =0) Two pair of ports unoupled (i.e. the orresponding

More information

Canimals. borrowed, with thanks, from Malaspina University College/Kwantlen University College

Canimals. borrowed, with thanks, from Malaspina University College/Kwantlen University College Canimals borrowed, with thanks, from Malaspina University College/Kwantlen University College http://ommons.wikimedia.org/wiki/file:ursus_maritimus_steve_amstrup.jpg Purpose Investigate the rate of heat

More information

23.1 Tuning controllers, in the large view Quoting from Section 16.7:

23.1 Tuning controllers, in the large view Quoting from Section 16.7: Lesson 23. Tuning a real ontroller - modeling, proess identifiation, fine tuning 23.0 Context We have learned to view proesses as dynami systems, taking are to identify their input, intermediate, and output

More information

DETERMINATION OF CARRIER TEMPERATURE FROM JUNCTION I(V) MEASUREMENTS

DETERMINATION OF CARRIER TEMPERATURE FROM JUNCTION I(V) MEASUREMENTS Journal of Eletron Devies, Vol. 15, 2012, pp. 1269-1273 JED [ISS: 1682-3427 ] DETERMIATIO OF CARRIER TEMPERATURE FROM JUCTIO I(V) MEASUREMETS 1 Mohamed H. Boukhatem, 2 Mario El Tahhi, 3 Pierre Mialhe 1

More information

22.54 Neutron Interactions and Applications (Spring 2004) Chapter 6 (2/24/04) Energy Transfer Kernel F(E E')

22.54 Neutron Interactions and Applications (Spring 2004) Chapter 6 (2/24/04) Energy Transfer Kernel F(E E') 22.54 Neutron Interations and Appliations (Spring 2004) Chapter 6 (2/24/04) Energy Transfer Kernel F(E E') Referenes -- J. R. Lamarsh, Introdution to Nulear Reator Theory (Addison-Wesley, Reading, 1966),

More information

Case I: 2 users In case of 2 users, the probability of error for user 1 was earlier derived to be 2 A1

Case I: 2 users In case of 2 users, the probability of error for user 1 was earlier derived to be 2 A1 MUTLIUSER DETECTION (Letures 9 and 0) 6:33:546 Wireless Communiations Tehnologies Instrutor: Dr. Narayan Mandayam Summary By Shweta Shrivastava (shwetash@winlab.rutgers.edu) bstrat This artile ontinues

More information

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV

More information

Four-dimensional equation of motion for viscous compressible substance with regard to the acceleration field, pressure field and dissipation field

Four-dimensional equation of motion for viscous compressible substance with regard to the acceleration field, pressure field and dissipation field Four-dimensional equation of motion for visous ompressible substane with regard to the aeleration field, pressure field and dissipation field Sergey G. Fedosin PO box 6488, Sviazeva str. -79, Perm, Russia

More information

Model-based mixture discriminant analysis an experimental study

Model-based mixture discriminant analysis an experimental study Model-based mixture disriminant analysis an experimental study Zohar Halbe and Mayer Aladjem Department of Eletrial and Computer Engineering, Ben-Gurion University of the Negev P.O.Box 653, Beer-Sheva,

More information

A Heuristic Approach for Design and Calculation of Pressure Distribution over Naca 4 Digit Airfoil

A Heuristic Approach for Design and Calculation of Pressure Distribution over Naca 4 Digit Airfoil IOSR Journal of Engineering (IOSRJEN) ISSN (e): 2250-3021, ISSN (p): 2278-8719 PP 11-15 www.iosrjen.org A Heuristi Approah for Design and Calulation of Pressure Distribution over Naa 4 Digit Airfoil G.

More information

Metal: a free electron gas model. Drude theory: simplest model for metals Sommerfeld theory: classical mechanics quantum mechanics

Metal: a free electron gas model. Drude theory: simplest model for metals Sommerfeld theory: classical mechanics quantum mechanics Metal: a free eletron gas model Drude theory: simplest model for metals Sommerfeld theory: lassial mehanis quantum mehanis Drude model in a nutshell Simplest model for metal Consider kinetis for eletrons

More information

22.01 Fall 2015, Problem Set 6 (Normal Version Solutions)

22.01 Fall 2015, Problem Set 6 (Normal Version Solutions) .0 Fall 05, Problem Set 6 (Normal Version Solutions) Due: November, :59PM on Stellar November 4, 05 Complete all the assigned problems, and do make sure to show your intermediate work. Please upload your

More information

Determination of the Aerodynamic Characteristics of Flying Vehicles Using Method Large Eddy Simulation with Software ANSYS

Determination of the Aerodynamic Characteristics of Flying Vehicles Using Method Large Eddy Simulation with Software ANSYS Automation, Control and Intelligent Systems 15; 3(6): 118-13 Published online Deember, 15 (http://www.sienepublishinggroup.om//ais) doi: 1.11648/.ais.1536.14 ISSN: 38-5583 (Print); ISSN: 38-5591 (Online)

More information

Stabilization of the Precision Positioning Stage Working in the Vacuum Environment by Using the Disturbance Observer

Stabilization of the Precision Positioning Stage Working in the Vacuum Environment by Using the Disturbance Observer Proeedings of the 4th IIAE International Conferene on Industrial Appliation Engineering 216 Stabilization of the Preision Positioning Stage Working in the Vauum Environment by Using the Disturbane Observer

More information

Semiconductor light sources Outline

Semiconductor light sources Outline Light soures Semiondutor light soures Outline Thermal (blakbody) radiation Light / matter interations & LEDs Lasers Robert R. MLeod, University of Colorado Pedrotti 3, Chapter 6 3 Blakbody light Blakbody

More information

Cavity flow with surface tension past a flat plate

Cavity flow with surface tension past a flat plate Proeedings of the 7 th International Symposium on Cavitation CAV9 Paper No. ## August 7-, 9, Ann Arbor, Mihigan, USA Cavity flow with surfae tension past a flat plate Yuriy Savhenko Institute of Hydromehanis

More information

Analysis of discretization in the direct simulation Monte Carlo

Analysis of discretization in the direct simulation Monte Carlo PHYSICS OF FLUIDS VOLUME 1, UMBER 1 OCTOBER Analysis of disretization in the diret simulation Monte Carlo iolas G. Hadjionstantinou a) Department of Mehanial Engineering, Massahusetts Institute of Tehnology,

More information

The simulation analysis of the bridge rectifier continuous operation in AC circuit

The simulation analysis of the bridge rectifier continuous operation in AC circuit Computer Appliations in Eletrial Engineering Vol. 4 6 DOI 8/j.8-448.6. The simulation analysis of the bridge retifier ontinuous operation in AC iruit Mirosław Wiślik, Paweł Strząbała Kiele University of

More information

RESEARCH ON RANDOM FOURIER WAVE-NUMBER SPECTRUM OF FLUCTUATING WIND SPEED

RESEARCH ON RANDOM FOURIER WAVE-NUMBER SPECTRUM OF FLUCTUATING WIND SPEED The Seventh Asia-Paifi Conferene on Wind Engineering, November 8-1, 9, Taipei, Taiwan RESEARCH ON RANDOM FORIER WAVE-NMBER SPECTRM OF FLCTATING WIND SPEED Qi Yan 1, Jie Li 1 Ph D. andidate, Department

More information

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM Rapport BIPM-09/02 PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM D T Burns, C Kessler, P Roger BIPM A R El-Sersy NIS (Egypt) July 2009 BUREAU INTERNATIONAL DES POIDS ET

More information

The Reason of Photons Angular Distribution at Electron-Positron Annihilation in a Positron-Emission Tomograph

The Reason of Photons Angular Distribution at Electron-Positron Annihilation in a Positron-Emission Tomograph Advanes in Natural Siene ol 7, No,, pp -5 DOI: 3968/66 ISSN 75-786 [PRINT] ISSN 75-787 [ONLINE] wwwsanadanet wwwsanadaorg The Reason of Photons Angular Distribution at Eletron-Positron Annihilation in

More information