PHOENIX MARS LANDER 2007 THERMAL AND EVOLVED GAS ANALYZER CALIBRATION REPORT. Version /22/2009

Size: px
Start display at page:

Download "PHOENIX MARS LANDER 2007 THERMAL AND EVOLVED GAS ANALYZER CALIBRATION REPORT. Version /22/2009"

Transcription

1 PHOENIX MARS LANDER 2007 Prepared By: TEGA TEAM - -

2 Table of Contents Background TEGA Engineering Subsystem AD590 Temperature Sensors PRT Temperature Sensors Engineering Voltages and Currents Status Channels TA_MANIFOLD_PRES (0), TA_OUTLET_PRES (2) Oven and Shield Error Signals TA_T_HEATER_TEMP (23) Oven and Shield Voltage and Current Pulse Width Monitors TA_FULL_DETECT (42), TA_FULL_DETECT_RAW (43) Calculating the Calibration Curves for the Oven Pt Sense Winding TA0: TA: TA3: TA5: TA6: Further Discussion Converting TEGA DSC telemetry to true power Pressure Transducer Calibration EGA Mass to Voltage Conversions Theory Practice Calibration (coefficient derivation) In-Situ Calibration Ground Data Reduction References

3 Background Phoenix TEGA is part of the instrument package onboard the 2007 Phoenix Mars Lander. The Phoenix mission, the first chosen for NASA's Scout program, began exploration of the Northern plains of Mars during TEGA stands for Thermal and Evolved Gas Analyzer and is directed toward sampling and understanding the chemistry and mineralogical components of the soil through thermal (heat capacity) and gas analysis of soil samples delivered by the Robotic Arm from various sampling locations beneath and beside the Lander. An additional capability is using the Evolved Gas Analyzer to measure trace components of the Martian Atmosphere. A detailed description is contained in the TEGA Instrument Paper (Boynton et al 2008) but a brief description is presented below. TEGA has three maor components: The Control Electronics (CE); the Thermal Analyzer (TA); and the Evolved Gas Analyzer (EGA). The control electronics (CE) is enclosed in a payload package beneath the Lander deck and contains a microcontroller and power conditioning and control system as well as a housekeeping component. The housekeeping component measures currents, voltages, pressures, and temperatures of the various sensors scattered over the instrument. The Thermal Analyzer sitting above on the deck receives the sample from the Robotic Arm seals it into an oven and runs a calorimetry scan under control of the CE. The TA is also supplied with two gas supplies and valves to route the gas from the supply tanks and eight ovens to the EGA. The Evolved Gas Analyzer is a 4 channel magnetic sector mass analyzer that measures the concentrations of gasses either from the ovens or the atmosphere. The primary purpose of this document is to provide the critical information required to convert the raw data from the instrument into standard units corrected to the degree possible for the understood effects of environmental and instrument behavior. Some of these conversions were calculated in-situ in real-time with the maority taking place as a post process to the receipt of the data on the ground

4 2 TEGA Engineering Subsystem The TEGA Engineering Subsystem is based on a radiation hardened Analog Devices AD7872 A/D converter. This is a signed, 4-bit converter with an input range of -3V to +3V. A summer at the input to the ADC is used for level shifting, allowing the system to process input signals of 0 to +6V, -3 to +3V, and 0 to -6V (Figure ). Channel Selector + - G= ADC +3V 0V -3V Polarity Selector Figure : Analog Signal Processing Figure 2 shows the transfer function of the ADC. The TEGA flight software converts the ADC output to a signed, 6-bit integer using the rules in Table. Figure 2: ADC Transfer Function Input Range Conversion 0 to +6V Invert bit 3-3 to +3V Sign Extend by copying bit 3 into bits 4 and 5-4 -

5 0 to -6V Invert bit 3 then subtract 0x3FFF Table : Converting ADC output to signed, 6-bit integer 2. AD590 Temperature Sensors Most of the temperatures in TEGA were measured with Analog Devices AD590 temperature sensors. These sensors pass a current that is proportional to the temperature, with a nominal slope of µa/ºk and absolute accuracy of ±.7ºK from 28 to 423ºK (-55 to +50ºC). All of the flight sensors were tested for linearity down to -00ºC. The excitation voltage for the sensors was +2V, and the output current was read out with a 5.4K, 0.% resistor, resulting in a nominal voltage to the ADC of 5.4mV/ºK (42.08dN/ºK). Each readout channel was characterized before launch with a precision current source, and the results are shown in Table 2. Name (Channel #) Location A B TA_MANIFOLD_TEMP (3) Valve manifold in the TA TA_PLUMBING TEMP (6) Plumbing between cal tank and valve manifold TA_PLUMBING_2_TEMP (7) Plumbing between valve manifold and transfer tube unction TA_CAL_TANK_TEMP (9) Cal gas tank near heater TA_CAL_TANK_COLD_TEMP (46) Cal gas tank at coldest spot TA_TRANS_TUBE_TEMP (24) Transfer tube between TA and EGA TA_EGA_MAN_TEMP (8) Valve Manifold in the EGA TA_EGA_PLUMB_TEMP (4) EGA plumbing between transfer tube and ion source TA_CPU_TEMP (0) FPGA-A, the hottest component on the CPU board TA_PWR_SPLY TEMP () Thermal plate near shield power converters TA_PWR_SPLY_2_TEMP (2) Filter board near linear optocouplers TA_PWR_CNTL TEMP (3) Center of thermal plate TA_PWR_CNTL_2_TEMP (4) Power control board near oven/shield power monitor circuit TA_A2D_TEMP (5) On the A/D converter on the analog board TA_EGA_ELECT_BOX_TEMP (22) EGA electronics frame EGA_PROC_TEMP (87) EGA CPU Chip lid Table 2: AD590 conversions. Temperature (C) = raw / A + B. This assumes a perfect, µa/ºk sensor. 2.2 PRT Temperature Sensors For locations where the temperatures to be measured were likely to fall outside of the -55 to +50ºC range of an AD590 sensor, PRT sensors were used. These sensors change their - 5 -

6 resistance proportionally to the temperature, with a slight nonlinearity. The conversion for these sensors is done in two parts: ) Calculate the resistance of the sensor from the raw ADC data R = raw * M + N 2) Use a polynomial to convert the resistance to temperature T = A + B * (R/R0) + C * (R/R0)^2 + D * (R/R0)^3 + E * (R/R0)^4 Table 3 gives the coefficients for the 6 PRT sensors, and Table 4 gives their locations in the instrument. Channel R0 M N A B C D E TA_OVEN_TEMP (20) NOTE E E NOTE 2 TA_SHLD_TEMP (2) NOTE E E NOTE 2 TA_EGA_BAKEOUT_TEMP (5) E E TA_EGA_GEC_TEMP (25) E E TA_COVER TEMP (6) E E TA_COVER_2_TEMP (47) E E TA_INPUT_FUNNEL LO_TEMP (7) E E TA_INPUT_FUNNEL_2_LO_TEMP (9) E E EGA_MAGNET_TEMP_ (85) 00 Note Note EGA_MAGNET_TEMP_2 (86) 00 Note Note Table 3: PRT conversion coefficients Channel TA_OVEN_TEMP (20) TA_SHLD_TEMP (2) TA_EGA_BAKEOUT_TEMP (5) TA_EGA_GEC_TEMP (25) TA_COVER TEMP (6) TA_COVER_2_TEMP (47) TA_INPUT_FUNNEL LO_TEMP (7) TA_INPUT_FUNNEL_2_LO_TEMP (9) EGA_MAGNET_TEMP_ (85) EGA_MAGNET_TEMP_2 (86) Table 4: PRT Sensor Locations Location TEGA Sample Oven Heated Shield around the Sample Oven EGA Ion Source Housing Inside of the EGA GEC Getter mounting post On the body of the North side cover actuator On the body of the South side cover actuator Backside of the TA0 funnel Backside of the TA5 funnel Main EGA magnet, Ion Source end Main EGA magnet, Ion exit end Note : The two EGA magnet PRT sensors use a significantly different readout circuit. For these V R + F V R two sensors, RG R = R S, V R + F V R RG where V=raw/3276.6, R S =3240, V R =5, R F =0000, and R G =

7 Combining and simplifying, step becomes R = raw 8023 NOTE 2: The sense windings in the ovens and shields are 99.99% pure Platinum wire. The datasheet lists the α for this wire at , making it distinct from the standard Pt curves ( , , and ). NOTE 3: Each oven and shield has its own R0. These values are given in Table 5. Cell # Oven R 0 Shield R Table 5: Oven and Shield R 0 Values 2.3 Engineering Voltages and Currents There are many voltages and currents that are monitored primarily to assess instrument health and safety. All of these monitors use a linear, result = A + raw * B conversion. A few of the monitors require some temperature compensation. Channel A B Unit Description Note TA_BUS_A_VOLT (26) E E- Spacecraft bus voltage, measured AFTER the input filters, V 03 i.e. as close to the actuators as possible TA_BUS_A_CUR (34).3978E E- 04 A Bus A (RPC06) current draw NOTE 4 TA_BUS_B_CUR (35) -.090E E- 04 A Bus B (RPC03 + RPC09) current draw NOTE 5 TA_EGA_CUR (36) -.236E E- 04 A EGA PECM current draw NOTE 6 TA_CPU_PLUS_5_VOLT (29) E E- 04 V TEGA CPU board +5V rail voltage monitor TA_CPU_PLUS_5_CUR (37) E E- 05 A TEGA CPU board +5V rail current monitor TA_ANLG_PLUS_2_VOLT (30) E-4 V TEGA Analog board +2V rail voltage monitor TA_ANLG_PLUS_2_CUR (38) E-5 A TEGA Analog board +2V rail current monitor TA_ANLG_MINUS_2_VOLT (3) E-4 V TEGA Analog board 2V rail voltage monitor TA_ANLG_MINUS_2_CUR (39) E-5 A TEGA Analog board 2V rail current monitor TA_PLUS_5_VREF () E-4 V Direct readout of a precision +5V reference TA_PRES_SENSE_FD_BK (8) E-4 V +0V reference to pressure sensors TA_AGD_0_3 (27) E-4 V Unused MUX input, tied to analog ground TA_AGD_3_ (28) E-4 V Unused MUX input, tied to analog ground EGA_ION_PUMP_VOLT (8) E- V EGA Ion Pump High Voltage monitor EGA_ION_PUMP_CUR (82) E-2 µa EGA Ion Pump Current monitor EGA_SWEEP_VOLTAGE (83) E- V EGA Sweep Voltage monitor - 7 -

8 EGA_MULTIPLIER_VOLT (80) E- V EGA Channel Electron Multiplier high voltage monitor EGA_EMISSION_CUR (72) E- µa EGA Ion Source Emission Current monitor NOTE 7 EGA_TRAP_CUR (7) E- µa EGA Ion Source Trap Current monitor NOTE 7 EGA_FILAMENT_CUR_ (78) E-4 A EGA Filament # drive current monitor EGA_FILAMENT_CUR_2 (79) E-4 A EGA Filament #2 drive current monitor EGA_PLUS_5_VOLT (75) E-4 V EGA CPU board +5V rail monitor EGA_PLUS_2_VOLT (76) E-3 V EGA Analog +2V rail monitor EGA_MINUS_2_VOLT (77) 0-8.9E-4 V EGA Analog -2V rail monitor EGA_GEC_CUR (84) E-4 A EGA Getter current monitor Table 6: Voltage and Current Monitors NOTE 4: This sensor has a significant sensitivity to both voltage and temperature. The initial coefficients get us to +57/-99mA. Adding temperature compensation gets us to +75/-67mA: result = result +.08E E-03 * TA_PWR_SPLY_2_TEMP (temperature in ºC) Adding voltage compensation improves this to ±0mA: result = result E E-04 * RAW_TA_BUS_A_VOLT E-08 * RAW_TA_BUS_A_VOLT^ E-3 * RAW_TA_BUS_A_VOLT^3 (note that this uses the raw ADC output for the bus A voltage) NOTE 5: This sensor is sensitive to temperature. The initial coefficients get us to ±90mA. Applying temperature compensation gets us to ±20mA: result = result E E-03 * TA_PWR_SPLY_2_TEMP (temperature in ºC) NOTE 6: Initial coefficients get us to ±25mA. Temperature compensation improves this to ±0mA. result = result E E-04 * TA_PWR_SPLY_2_TEMP (temperature in ºC) NOTE 7: Not valid after the filament short on SOL Status Channels There are several channels that are used to report digital status. EGA_STATUS_BITS (70) This is a 6 bit status word that shows the internal status of the EGA. Bit Function 0 Emission ON Filament selected 2 Filament 2 selected 3 High Voltage Enabled 4 CEM High Voltage Enabled 5 Sweep High Voltage Enabled 6 Currently Sweeping 7 Emission Current Setting (0 = low, = high) 00-23V (0V) 9,8 Emission Energy (NOTE 8) 0-27V (-3V) 0-37V (-3V) -90V (-66V) 0 Emission Fault Magnet Sensor Fault 2 Magnet Sensor 2 Fault 3-5 Unused Table 7: EGA Status Bits - 8 -

9 NOTE 8: The filament short on sol 4 altered the emission energy control circuit. After sol 4 the emission energy was 24V less. The post-short values are in parentheses. EGA_FILAMENT SEL (73), EGA_FILAMENT_2_SEL (74) Prior to the filament short on Sol 4, these two channels provided positive feedback that the intended filament was truly selected. The selected filament should read V, and the other filament should read.6666v. The filament short altered the circuit configuration such that these readings are not meaningful after Sol 4. EGA_AVG_IDLE_CALLS (88), EGA_MIN_IDLE_CALLS (89) These two channels track the load on the EGA CPU. The Idle function is called repeatedly whenever the CPU is not busy. This function simply accumulates statistics on how often it is called. The collected statistics are read out, and the accumulations reset every time an EGA engineering packet is created. EGA_AVG_IDLE_CALLS is the average number of times the idle function was called per second since the last engineering packet. EGA_MIN_IDLE_CALLS is the minimum number of idle calls in any second since the last engineering packet. Together, these channels give an indication of both the average and the peak load on the CPU. 2.4 TA_MANIFOLD_PRES (0), TA_OUTLET_PRES (2) There are two pressure sensors in the TEGA plumbing. The Manifold Pressure Sensor is located in the plenum between the carrier and calibration valves and the TA inlet valves. This sensor is used to regulate the gas flow through the ovens. The Outlet Pressure Sensor monitors the pressure at the TA inlet to the EGA. Both sensors exhibit significant nonlinearity at very low pressures, but become quite linear as the pressure increases (see section 5). As a result, there are two equations for converting the raw value to pressure for each sensor. Channel Threshold A B C Constraint TA_MANIFOLD_PRES raw < E-08 raw E-05 0 NOTE 9 raw < E-07 TA_OUTLET_PRES NOTE 0 raw E-05 0 Table 8: Pressure Calibration Parameters Pressure = A + B*raw_pressure + C*raw_pressure^2 NOTE 9: TA Manifold at +65. There were occasions where the manifold was not run at +65C. See section 5 for temperature corrections. NOTE 0: EGA Manifold at Oven and Shield Error Signals TA_OVEN_ERR (44), MEM_OVEN_ERR (57), TA_SHLD_ERR (45), and MEM_SHLD_ERR (58) The oven and shield control systems use a high-gain difference amplifier to subtract the commanded temperature from the measured temperature. This difference is read out every 3.3mS and used by the software to calculate the proper width for the next pulse

10 The exact conversion from the error signal to degrees is different for each TA and also varies with the temperature, but a reasonable estimate is 63 dn / degree for the oven error and 763 dn / degree for the shield error. This measurement is reported two ways. The TA_OVEN_ERR and TA_SHLD_ERR channels take an extra reading of these channels, while the MEM_OVEN_ERR and MEM_SHLD_ERR report the last scheduled (3.3mS) reading. 2.6 TA_T_HEATER_TEMP (23) The T heater is a heater / temperature sensor assembly on a plumbing unction between each oven and the manifold. There is one T heater for each oven. The assembly consists of a coil of heater wire wound on the plumbing unction wired in parallel with a 00Ω PRT. The PRT is mounted on the metal support for the unction. This assembly is driven by a PWM controller that measures the parallel resistance in between the pulses to determine the temperature. As usual, there is a twist: The current source used to measure the resistance of the assembly has a significant temperature coefficient. Calculating the temperature requires several steps: ) Calculate the sense current (I S ) based on the smoothed TA_PWR_CNTL_2_TEMP engineering channel TA_PWR_CNTL_2_TEMP I S = 000 2) Calculate the resistance of the assembly, R T. 4 2 raw RT = 9.27 I S 3) Use R T and the known resistance of the heater winding (R H ) to calculate the resistance of the PRT, R S. RH RT RS = RH RT 4) Calculate the temperature of the PRT (T S ) from R S using our standard PRT equation R S RS RS RS T S = A + B C D E R0 R0 R0 R0 Where R 0 = 00, A = , B = 236.9, C = , D = , and E = ) And finally, calculate the temperature of the actual plumbing (T T ) from T S. T T T = S

11 Table 9 gives the R H value for each of the cells. Cell # R H Table 9: T Heater Winding Resistances 2.7 Oven and Shield Voltage and Current MEM_OVEN_VOLT (52), TA_OVEN_PLUS_5_VOLT (32) MEM_OVEN_CUR (53), TA_OVEN_PLUS_5_CUR (40) MEM_SHLD_VOLT (54), TA_SHIELD_PLUS_30_VOLT (33) MEM_SHLD_CUR (55), TA_SHIELD_PLUS_30_CUR (4) In order to accurately calculate the power going to the oven and shield, we need to measure the voltage and current during the power pulse. Since the A/D converter takes roughly 6µS to do a conversion, it would not be practical to measure all four of these during a single pulse. Instead, the readings are taken in a round-robin fashion with one reading taken on each pulse. The reading is taken approximately 5µS after the start of the pulse to allow the pulse to reach full voltage and the readout circuits to stabilize. These readings are not accurate until the pulse width is greater than about 8µS (pulse_width > 60). The MEM_ channels report the most recent of these timed readings. The TA_ channels are read out at some random time with respect to the power pulse. They should not be used. The conversions for these channels are linear, with a linear temperature correction for the TA_PWR_CNTL_2_TEMP sensor, which is located in the middle of the circuit with the temperature sensitivity. Table 0 gives the conversion coefficients for these 4 channels. Channel A B C Accuracy MEM_OVEN_VOLT E E E-04 ± mv MEM_OVEN_CUR E E E-04 ± 4mA MEM_SHLD_VOLT E E E+00 ± 0mV MEM_SHLD_CUR E E E-04 ± 7mA Result = A + B * raw + C * TA_PWR_CNTL_2_TEMP (TA_PWR_CNTL_2_TEMP in ºC) Table 0: Conversion coefficients for Oven and Shield Voltage and Current 2.8 Pulse Width Monitors MEM_OVEN_WIDTH (60) MEM_SHLD_WIDTH (6) MEM_T_WIDTH (59) - -

12 These channels report the width of the most recent pulse from the respective PWM controller. The oven and shield controllers are 6 bits with a resolution of 50nS, and the T heater controller is 0 bits with µs resolution. While calorimetric calculations are possible using these channels, the preferred method is to use the data in the DSC science data packets. 2.9 TA_FULL_DETECT (42), TA_FULL_DETECT_RAW (43) The flow of soil into the oven is monitored by an optical sensor. As soil particles fall into the oven, they modulate the light falling on a photodetector. The TA_FULL_DETECT_RAW channel is used immediately before a soil loading operation to get light and dark readings that are used to set the thresholds for soil detection. There is no calibration per se for this channel, but lower values indicate brighter light. The TA_FULL_DETECT channel is the output of an analog integrator that sums the light to dark transitions on the detector. During a soil loading operation, high time resolution readings are collected on these channels. These readings are sent in the LED science data packets

13 3 Calculating the Calibration Curves for the Oven Pt Sense Winding The coefficients used to convert the resistance of the oven and shield sense windings to temperature were determined for TEGA- and carried over to TEGA-2 without any further validation. Additionally, the room temperature resistance of the windings were measured twice during TEGA-2 assembly, and those measurements came out slightly different. The Callendar van Dusen method is applied to the TEGA-2 flight data to check the validity of the coefficients and to select which set of resistance measurements lead to believable temperatures. The Callendar van Dusen method uses four calibration points to fully define the R vs T curve for a given alloy of Pt. The calibration points are: R 0 at T=0ºC R 00 at T=00ºC R H at some T H >> 00ºC R L at some T L << 0ºC The Callendar van Dusen coefficients are R00 R0 α = (eq. ) 00 R0 RH R0 TH R0 α δ = (eq. 2) TH TH R R T T L T + δ 00 β (eq. 3) TL TL L 0 L L 0 00 R α = 3 Given only α, one can get a first order approximation of R for a given T with = R0 ( + α T ) (eq. 4). Given α and δ, one can calculate R T very accurately for R T T T temperatures 0ºC using R T = R0 + R0α T δ (eq. 5). To get an accurate R T for temperatures below 0ºC we also need the β coefficient, and the equation 3 T T T T is R T = R0 + R0α T δ β (eq. 6)

14 0 fort 0 α δ α δ By substituting A = α +, B =, and C = 2 α β (eq. 7, 8, and 9), fort < R T = R + AT + BT + CT T 00. [ ] we can rewrite eq. 6 in the more familiar form ( ) 0 These four data points are not available in the flight data. The data points that can be teased from the flight data are: R R at T = room temperature, measured during assembly for each oven T R, a good measurement of room temperature R H at T=359ºC for each oven. This point is from the mystery peak in the high temp day 2 data, which we believe is the Curie transition in the Ni oven body. Since the Curie point for Ni is sometimes given in the literature as 357ºC and sometimes as 36ºC, the average of the two is used for these calculations. α, the normalized slope of R vs T from 0-00ºC, is given as in the California Fine Wire datasheet. This is the same α as in eq. above. Since room temperature is not far from 0ºC, eq. can be used to calculate R 0 from R R RR and T R. Solving eq. for R 0 gives R0 =. For oven 4, this yields α TR + = R = Ω The value used for the TA4 R 0 was 33.68Ω..5x0-3 Sol 25 "Mystery Peak" Oven Duty Cycle Oven Sense Resistance Figure 3: Sol 25 Mystery Peak for TA4 90 Figure 3 shows the mystery peak in the high temp day 2 run for TA4 on sol 25. After some discussion, the science team concluded that the Curie point is at the peak of the transition. This peak is at R = Ω. Plugging this into eq. 2 yields δ = = There are no good candidates for R L and T L, so we can not calculate β. However, since there are no transitions of interest below 0ºC, this should not be an issue

15 Plugging these values for α and δ into equations 7 and 8 gives: A = = and B = = The numbers used during surface operations were A=3.9080x0-3 and B= x0-7. Using the new A and B values to recalculate the temperatures for the TA4 run shows that we were 0.4 degrees hotter at the top of the ramp (Figure 4). 000 old_ta_oven_temp new_ta_oven_temp 800 Oven Temperature TEGA Time Figure 4: Comparison of old and new conversions for TA4 0x0 3 Applying the same procedure to the rest of the surface runs yields the following: 3. TA0: R R = and T R = = R = Ω x0-3 Sol 70 "Mystery Peak" Oven Duty Cycle Oven Sense Resistance Figure 5: Sol 70 TA0 Mystery Peak. Peak at ohms - 5 -

16 From these numbers, we get δ = = TA: R R = and T R = = R = Ω (was ).5.0 x Figure 6: Sol 34 TA Mystery Peak. Peak at ohms δ = = TA3: R R = and T R = 24.8 = R = Ω (was ) x Figure 7: Sol 5 TA3 Mystery Peak. Peak at ohms - 6 -

17 δ = = TA5: R R = and T R = = R = Ω (was ) x Figure 8: Sol 77 TA5 Mystery Peak. Peak at ohms δ = = TA6: R R = and T R = = R = Ω (was ) x Figure 9: Sol 47 TA6 Mystery Peak. Peak at ohms - 7 -

18 δ = = The δ values are summarized in Table, and shown graphically in Figure 0, along with the average value

19 TA # δ No data No data Table : Summary of δ Values δ Mean δ =.77 δ TA # Figure 0: δ Values and Averages 6 7 This spread is also far too wide for these values to be useful, especially considering that δ/00 2 is the T 2 coefficient. Also, since δ describes a fundamental property of the Pt wire, this is enough of a spread to warrant further examination. As a quick check, the δ value for each TA was used to calculate a new temperature scale, then each mystery peak was plotted against this new temperature scale. Oven Duty Cycle 2.0x Average δ Res_ov_duty_0 Res_ov_duty_ Res_ov_duty_3 Res_ov_duty_4 Res_ov_duty_5 Res_ov_duty_ Calculated Oven Temperature Figure : Mystery peaks plotted each with its own temperature scale - 9 -

20 Figure shows these results. The markers show the location chosen as the peak. Since all of the markers lie exactly at 359 C, we can conclude that the δ values were calculated correctly for the given input data. To see ust how harmful this spread really is, the temperature scale for each TA was recalculated using the average δ of.7703 for all of the TAs. The results are shown in Figure 2. The markers now lie at ± 4.45 C. This isn t great, but it isn t horrible, either. Oven Duty Cycle 2.0x Average δ Res_ov_duty_0 Res_ov_duty_ Res_ov_duty_3 Res_ov_duty_4 Res_ov_duty_5 Res_ov_duty_ Calculated Oven Temperature Figure 2: Mystery Peaks Using a Common δ of.77 and the earlier R 0 measurements But how bad is this spread at 000 C? To get a rough idea, the TA4 max resistance was converted to temperature using δ =.346 and δ = (the two end members). This yields max temperatures of C and C, or 06.7 ± 60.2 degrees. This is really pretty horrible, but will have to do for this release of the RDR data. 3.6 Further Discussion As it turns out, there were two measurements of the R R and T R taken for each oven: One right after they finished cooling from the annealing bake, and a later one after the ovens were integrated into the TA assemblies. The earlier measurement was probably more accurate since it was done under very controlled conditions with a 6-digit, 4-wire ohmmeter. Unfortunately, the derived R 0 values do not agree between the two measurements, and consequently, neither do the derived δ values. All of the above calculations are based on the earlier measurement. TA # First measurement Second measurement δ based on first measurement δ based on second measurement

21 Table 2: Comparison of first and second oven measurements Table 2 shows both sets of R 0 and δ values. Using the later set of R 0 measurements, the spread in the δ values is somewhat larger than the spread using the later set (.0 vs 0.85), and TA0 becomes a serious outlier. Leaving out TA0 and repeating the analysis of the spread in temperatures using the mean δ value (in this case.65), shows the mystery peaks at ± 2.5 degrees (Figure 3). Calculating the spread at the max temperature using the TA3 data and δ =.369 and δ =.858 (the two end members ignoring TA0) yields temperatures of C and C, or ± 3.4 degrees. This is better than the previous spread, but it leaves out poor TA0. Oven Duty Cycle 2.0x Average δ Res_ov_duty_0 Res_ov_duty_ Res_ov_duty_3 Res_ov_duty_4 Res_ov_duty_5 Res_ov_duty_ Calculated Oven Temperature Figure 3: Mystery Peaks Using a Common δ of This situation begs further analysis. Since δ is a fundamental property of the Pt wire, it should be constant for all ovens. This analysis shows that there is some second-order effect that we do not understand. One possibility is that Pt wire changes its resistance in response to strain as well as temperature. Since each oven is built by hand, there could be considerable variation in the tightness of the wind and the subsequent strain placed on the wire by thermal expansion of the oven. Further analyses that could be done include putting an oven and a bare loop of the Pt wire in a controlled environment to map out the curve, but considering the variability we have seen it is not clear if one more data point would help or hurt. If time and money permit, we will revisit this problem and release updated RDR data

22 4 Converting TEGA DSC telemetry to true power The TEGA PWM controller sends one power pulse to the oven and shield every mS. For each pulse, TEGA reads out the voltage (V), current (I), and pulse width W (W). The power in the pulse is V I. These powers are then summed over some number of pulses (N) for downlink to bring the data rate down to a reasonable value. TEGA reads the voltage and current as 4-bit integers in arbitrary units. It does not convert the voltage and current readings into true volts and amps prior to doing the multiplication in order to save time in the DSC interrupt routine. It is the responsibility of the ground data system to do this conversion after the fact. The true voltage is V B D, and the true current is I B 2 D2, where B, B 2, D, and D 2 are determined during instrument characterization. In order to calculate the true power in N pulses, we need the sum N W ( V B D ) ( I B D ) 2 2 = power = (Equation ) N. Since TEGA has no knowledge of B, B 2, D, or D 2, we need to develop a method where those numbers can be applied on the ground. We can write Equation as: B B N 2 = W V I B D N 2 = W V B + D D power = (Equation 2) N With this form, TEGA can calculate the 4 individual sums for downlink, and the gains and offsets can be applied on the ground. These 4 sums are in the DSC telemetry packet as pulse ( ). N W = ), power ( W V I N = 2 D N = W ), current ( W I These are assembled from the EDR data as follows: Pulse = N W = Power = W V I = Current = W I N N = = OVEN_PULSE I N = = OVEN_PWR_HI * OVEN_PWR_LO = OVEN_CUR_HI * OVEN_CUR_LO N 2 = W ), and voltage ( W V N =

23 N Voltage = W V = OVEN_VOLT_HI * OVEN_VOLT_LO = N = SUM_COUNT From the instrument characterization we know that D and D 2 are a dependent on the temperature of the readout circuit, so we substitute D = A C TA_PWR_CNTL_2_TEMP and D = A C TA_PWR_CNT L_2_TEMP Table 3 lists the A N, B N, and C N coefficients for the flight unit. Note that these are the same coefficients used to convert the MEM_OVEN_VOLT, MEM_OVEN_CUR, MEM_SHLD_VOLT, and MEM_SHLD_CUR engineering channels. (See Table 0 in TEGA Engineering Explained) A B C A 2 B 2 C 2 Oven E E E E E E-04 Shield E E E E E-04 Table 3: Oven and Shield Power Coefficients

24 5 Pressure Transducer Calibration Two pressure transducers are used to control and monitor the pressure and gas flow in the TEGA instrument. One Sensor is mounted on the inlet side of the Thermal Analyzer(TA) manifold and is used to provide feedback for the pressure regulation valves from the two gas supply tanks. The second sensor is mounted in the Evolved Gas Analyzer(EGA) manifold and is used to monitor the gas pressure adacent to the Mass Analyzer port and the pressure down-flow of the ovens. The transducer signals are amplified and conditioned by electronics located in the Lower Payload Electronics bay to give nominal signals in the pressure range from a few mb up to Bar(4.7 PSI absolute). The TA Manifold sensor is a 9C030A and the EGA sensor is a 9C05A transducer both originally from the Honeywell(Sensym) and repackaged to provide a light weight interface to the manifolds. Both transducers read pressure in the absolute sense in that they can read pressures down to a vacuum. Deficiencies recognized after the initial calibration required a recalibration done insitu on Mars to update the parameters used to transform raw counts to pressure and flow. The Sol 4 Checkout along with EGA runs through Sol 66 are used to provide the calibration constants used in the final processed data. Given an estimate for the absolute pressure at the landing site, the Mass Spectrometer was used to locate several additional points at higher pressures by using the initial Net Counts/mBar to extrapolate this measurement to higher count rates at higher pressures. A fit for of the pressure as given by the count rate in the Mass Analyzer on Sol 66 was then transferred to the Manifold pressure transducer using data derived from the Sol 4 Checkout. The Sol 4 Checkout consisted of several measurements made while both the manifolds were still capable of being held at a static and equalized pressure by a diaphragm allowing the transfer of the calibration from the TA_OUTLET_PRES to the TA_MANIFOLD_PRES. Later on Sol 4 a puncture mechanism breeched this diaphragm allowing flow to the Mass Analyzer within the EGA and exhausting to Mars. Cross checks for the converted pressure values was correlated against other information related to the volume and flow characteristics of the TEGA plumbing system. These checks demonstrated good agreement between calculated pressures and expected values based on measured values of the Flight Model and equivalent runs in the Engineering Qualification model. The calibration values for the TA_OUTLET_PRES and TA_MANIFOLD_PRES are two part fits shown in table that convert the raw signal to a final pressure in Bar. A second order polynomial is used below the indicated threshold and a linear fit above. The equation for Manifold Pressure is valid for times when the TA_MANIFOLD_TEMP is at or near +65 C and the Outlet Pressure conversion is valid when the EGA_MANIFOLD_TEMP is at or near +35 C. An additional temperature compensation equation is also shown in table 2 for the TA_MANIFOLD_PRES but is not applied in the exported data. Although the devices are temperature compensated it was found that the output signal is not correct when the device is changing temperature rapidly or has a persistent thermal gradient across the transducer. The compensation provided can be

25 applied only when the temperatures are stable and cannot be universally applied on a point by point basis during or within a few minutes after abrupt temperature changes. No temperature Compensation is available at the present time for the OUTLET Pressure Sensor. The non-temperature compensated conversions are relevant to the maority of the data taken during the mission. Channel Threshold A B C Constraint TA_MANIFOLD_PRES DN < E-08 Note TA_MANIFOLD_PRES DN E-05 Note TA_OUTLET_PRES DN < E-07 Note 2 TA_OUTLET_PRES DN E-05 Note 2 Pressure=A+B*raw_pressure+C*raw_pressure^2, C is zero if not shown Table. Pressure Calibration Parameters Note : TA Manifold at +65 Note 2: EGA Manifold at +35 Channel A B Constraint TA_MANIFOLD_PRES E-05 Note 3,4 TA_MANIFOLD_PRES(T arb )= TA_MANIFOLD_PRES+A+B*T Table 2. Temperature Compensation for Manifold Sensor Note 3: Manifold temperature stable and absolute manifold pressure less than 00mB Note 4: Temperature in Celsius

26 0.20 Sol 4 Checkout TA Manifold Pressure High Temp Estimated Manifold Pressure ( Bar) Coefficient values ± one standard deviation K0 =0.628 ± K = ±.4e-005 K2 =2.8325e-008 ± 2.3e-009 Above 3760 Coefficient values ± one standard deviation a = ± b =8.3527e-005 ± 4.4e-007 TA_OUTLET_PRES_SAM fit_ta_outlet_pres_sam_ht fit_ta_outlet_pres_sam_lin_ht TA Manifold (RAW DN) Figure. Example Conversion for the Manifold Pressure The Pressure Transducers provided as part of the TEGA package are primarily to control and monitor the gas flow through the system and as such are required to have an absolute accuracy no better than 0% and a resolution on the order of a tenth of a mb. The measurements made on Sol 66 used as the calibration basis have an estimate of 7.5mB for the ambient Mars atmosphere compared to a value of 7.85 mb from MET reported post mission (about 4.5% error) under steady state conditions

27 6 EGA Mass to Voltage Conversions 6. Theory The Evolved gas Analyzer (EGA) is a magnetic sector mass spectrometer. Ions generated in the ion source are accelerated by a sweep voltage, pass though a magnetic field generated by a permanent magnet where they are deflected and arrive at one of the four detectors. When a given mass is commanded to be scanned the instrument automatically determines the correct sweep voltage to apply. This section outlines the mass-to-voltage equations used in the instrument, the voltage-to-mass equations used in the ground data reduction as well as the in-situ fine tuning of the calibration parameters. The relationship between the mass-to-charge ratio, magnetic field strength, radius of 2 2 curvature to the detector, and accelerating voltage is expressed in m / z = B r / 2V (Equation 3). 2 2 m / z = B r / 2V (Equation 3 ) where: r = radius of arc of ions deflected in the magnetic field V = acceleration voltage applied to ions leaving the ion source (sweep voltage) B = magnetic field strength m/ z = mass to charge ratio of ion For this instrument all ions are assumed to have a single charge ( z = ), making the mass to charge ratio equal to mass ( m /) = m. In order to get a specific mass to reach a detector the sweep voltage, V, is selected via a 6 bit DAC over a range of 2000 to 0 volts. ( ) V = dn (Equation 4) where: V = Voltage dn = 6 bit DAC code Since the magnetic field strength B varies as a function of temperature it becomes necessary to adust the sweep voltage to compensate for the magnet temperature. During surface operations the main magnet experienced temperatures from -50 to 0 C. Over this range the function of the field strength versus temperature is mostly linear

28 6.2 Practice There are two modes in which the EGA can sample masses, sweep mode and hop mode. In sweep mode the voltage is changed in steps in order to cover an entire mass range with even sized steps in mass space. In hop mode only the center of peaks are sampled with either 5 or 7 voltage settings evenly separated in mass space. (Figure 4) Figure 4: Sweep modes vs. Hop modes Sweep mode and hop mode have different requirements of the mass-to-voltage calculations with regards to precision. Small errors in the conversion only affect the start and end voltage for a sweep mode but can make a hop mode fall on the side of a peak or completely miss the peak. 6.3 Calibration (coefficient derivation) In order for hop modes to be precise it was necessary to understand the effect of the magnet temperature on the location of peaks. To determine the exact coefficients for the mass-to-voltage equations several masses on all four channels were continuously scanned and tracked in voltage space while the EGA was slowly ramped in the Mars simulation chamber from -50 to 20 C. Figure 5 shows an example of this data for mass 44 on channel

29 Figure 5: Sweep voltage vs. magnet temperature for mass 44 on channel 4 For each channel the mass, voltage, and magnet temperature data was fit to an equation with the form show in Equation 3 (see Figure 6). This equation is, in part, obtained from equation by setting the magnetic field strength B to a linear form ( B = gx + h), combining constants, and solving for V. The linear part of equation 3 ( mx + n) was added to correctly fit the data although the m term was typically fixed to zero. 2 ax + bx + c f ( M, x) = V = + mx + n (Equation 3) M where: V = Voltage in DAC steps (see equation 2) x = Magnet temperature (average of the two raw magnet temperature readings, see TEGA engineering explained) M = Mass in AMU a,b,c,m,n = Coefficients to be fit Table shows the actual results from these fits which were loaded into the EGA flight software. Channel Channel 3 a a b b c c m 0 m 0 n n Channel 2 Channel 4 a a b b c c m 0 m

30 N n Table : EGA flight software default mass-to-voltage coefficients. When the filaments in the ion source are run at high emission it is necessary to add 20 to the n term from their nominal low emission values. This result was determined experimentally by running the instrument at high and low emission and quantifying the shift in peak location. Subsequently, if the instrument is well calibrated at high emission is it necessary to remove 20 from the n term when switching to low emission. Figure 6: Results from fitting channel 4 data to equation 3. Mass and magnet temperature are displayed on the independent x and y axis, sweep voltage is on the dependent z axis. 6.4 In-Situ Calibration The mass-to-voltage equations described above provides an accurate (~ ± 0. AMU) mass-to-voltage conversion over a large temperature range. There are however many factors, such as filament warm up, that can effect the calibration in subtle ways. These errors are significant enough that tightly clustered mass hop points would not be centered on a peak simply using the default coefficients. To correct for this an in-situ calibration was performed several times during each run

31 To fine tune the calibration, for each channel, two well known masses (M, M 2 ) with large peaks are swept over with high resolution (Figure 7). The EGA flight software then determines the center of those peaks in voltage space (V, V 2 ) and updates the internal coefficients. Figure 7: Actual calibration sweeps for channel 3 using masses 8 and 44. The dotted black line shows where the instrument determined the peak centers to be. Once the voltage at the center of two peaks has been found the instrument solves equation 4 which is a simplified version of equation 3 where the magnet temperature dependent terms have been replaced with A and B. Once solved the c and n coefficients are update for each channel. V A = B (Equation 4) M n + n where: (From Equation 3) - 3 -

32 A = ax 2 B = mx + n M = Mass + bx + c V = Voltage x = Magnet temperature during the calibration sweeps Using M, M 2, V, V 2 from the calibration scan solve equation 4 as a system of equations using substitution. B = A = ( M V M 2 V2 ) ( M M 2 ) ( M V ) ( M B) (Equation 5) where: M =Mass of peak M 2 =Mass of peak 2 V =Voltage at center of peak V 2 = Voltage at center of peak 2 Finally equation 6 describes how the c and n coefficients are updated after a successful calibration. c = n = A 2 ( ax + bx) ( B mx) (Equation 6) Figure 8 shows the effect of a calibration scan on a tightly spaced hop sample on mass 44, channel 3. The black line shows where the peak center was calculated using equation 3 both before and after calibration. The green dots show where the hop points would have been sampled

33 Figure 8: The effect of an in-situ calibration on a tightly spaced hop mode. The left side shows where hop points (green) would have been placed using the default coefficients. The right side is after the coefficients were updated with a calibration scan. Before each calibration scan the instrument was typically commanded to reload the default coefficient values. If a calibration scan fails for some reason (not enough counts, could not find peak center) the default coefficients remained in place. 6.5 Ground Data Reduction When processing data returned from the EGA it is necessary to invert the processes described above. The EGA science data is in terms of counts at a certain voltage setting, so converting from voltage to mass is required. Starting with equation 3 and solving for M results in equation 7. M 2 ( ax + bx + c) = V ( mx + n) (Equation 7) The calibration sweeps are also returned so adustments to the coefficients over the course of a run can be replicated on the ground. 6.6 References Herbert, Christopher G., and Robert A.W. Johnstone. Mass Spectrometry Basics. Florida: CRC Press, 2003 Wright, Wilfred, and Malcolm McCaig. Permanent Magnets. Great Brittan: Oxford University Press,

Design Engineering MEng EXAMINATIONS 2016

Design Engineering MEng EXAMINATIONS 2016 IMPERIAL COLLEGE LONDON Design Engineering MEng EXAMINATIONS 2016 For Internal Students of the Imperial College of Science, Technology and Medicine This paper is also taken for the relevant examination

More information

Legacy Calibration of the Automatic Weather Station Model 2 of the United States Antarctic Program

Legacy Calibration of the Automatic Weather Station Model 2 of the United States Antarctic Program Legacy Calibration of the Automatic Weather Station Model 2 of the United States Antarctic Program G. A. Weidner 2, J. E. Thom 1, and M. A. Lazzara 1 1 Antarctic Meteorological Research Center Space Science

More information

Remote Display Unit. Installers Handbook Copyright 2001 AirSense Technology Ltd. LM Remote Display Unit Installers Handbook Issue 1.

Remote Display Unit. Installers Handbook Copyright 2001 AirSense Technology Ltd. LM Remote Display Unit Installers Handbook Issue 1. Remote Display Unit Installers Handbook Copyright 2001 AirSense Technology Ltd AirSense, ClassiFire, FastLearn PipeCAD, SenseNET, Stratos-HSSD and Stratos-Quadra are trademarks. HSSD is a Registered Trademark.

More information

Temperature Sensors & Measurement

Temperature Sensors & Measurement Temperature Sensors & Measurement E80 Spring 2014 Contents Why measure temperature? Characteristics of interest Types of temperature sensors 1. Thermistor 2. RTD Sensor 3. Thermocouple 4. Integrated Silicon

More information

An Ultra Low Resistance Continuity Checker

An Ultra Low Resistance Continuity Checker An Ultra Low Resistance Continuity Checker By R. G. Sparber Copyleft protects this document. 1 Some understanding of electronics is assumed. Although the title claims this is a continuity checker, its

More information

Speed of Light in Air

Speed of Light in Air Speed of Light in Air Electromagnetic waves represent energy in the form of oscillating electric and magnetic fields which propagate through vacuum with a speed c = 2.9979246x10 8 m/s. Electromagnetic

More information

ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS

ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS 1 Introduction... 3 2 Objective... 3 3 Supplies... 3 4 Theory...

More information

Sensors Conditioning Module

Sensors Conditioning Module Sensors Conditioning Module Version 1.00 Microstar Laboratories, Inc. This manual is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced, or translated

More information

Designing a Thermostat Worksheet

Designing a Thermostat Worksheet Designing a Thermostat Worksheet Most of us have a thermostat in our homes to control heating and cooling systems of our home. These important devices help us save energy by automatically turning off energy

More information

ICPMS Doherty Lecture 1

ICPMS Doherty Lecture 1 ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms

More information

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3 AE 3051, Lab #16 Investigation of the Ideal Gas State Equation By: George P. Burdell Group E3 Summer Semester 000 Abstract The validity of the ideal gas equation of state was experimentally tested for

More information

The Phoenix Mars Mission is the first of the NASA

The Phoenix Mars Mission is the first of the NASA FOCUS: HARSH ENVIRONMENT MASS SPECTROMETRY Phoenix Mars Mission The Thermal Evolved Gas Analyzer John H. Hoffman, Roy C. Chaney, and Hilton Hammack University of Texas at Dallas, Richardson, Texas, USA

More information

Resistivity and Temperature Coefficients (at 20 C)

Resistivity and Temperature Coefficients (at 20 C) Homework # 4 Resistivity and Temperature Coefficients (at 0 C) Substance Resistivity, Temperature ( m) Coefficient, (C ) - Conductors Silver.59 x 0-0.006 Copper.6 x 0-0.006 Aluminum.65 x 0-0.0049 Tungsten

More information

PHY3901 PHY3905. Hall effect and semiconductors Laboratory protocol

PHY3901 PHY3905. Hall effect and semiconductors Laboratory protocol PHY3901 PHY3905 Hall effect and semiconductors Laboratory protocol PHY3901 PHY3905 Hall effect and semiconductors Laboratory protocol Objectives Observe and qualitatively understand the phenomenon of transverse

More information

1. Mark the correct statement(s)

1. Mark the correct statement(s) 1. Mark the correct statement(s) Figure to the right shows a mass measurement scale using a spring. 1.1 The span of the scale is a) 16 kg b) 21 kg c) 11 kg d) 5-16 kg 1.2 The range of the scale is a) 16

More information

Differential Pressure Sensor

Differential Pressure Sensor Differential Pressure Sensor MDP200 Series Features Pressure range up to ±500Pa with high accuracy of ±3.0% m.v. Pressure based on thermal micro-flow measurement Outstanding hysteresis and repeatability

More information

Thermal and structural design constraints for radiometers operating in geostationary orbits. G. E. Zurmehly, R. A. Hookman

Thermal and structural design constraints for radiometers operating in geostationary orbits. G. E. Zurmehly, R. A. Hookman Thermal and structural design constraints for radiometers operating in geostationary orbits G. E. Zurmehly, R. A. Hookman ITT Aerospace/Communications Division Fort Wayne, Indiana ABSTRACT For a radiometer

More information

Thermal Coatings for In-vacuum Radiation Cooling LIGO-T C R. Abbott, S. Waldman, Caltech 12 March, 2007

Thermal Coatings for In-vacuum Radiation Cooling LIGO-T C R. Abbott, S. Waldman, Caltech 12 March, 2007 Thermal Coatings for In-vacuum Radiation Cooling LIGO-T070054-00-C R. Abbott, S. Waldman, Caltech 12 March, 2007 1. Overview and Background 1.1. There are instances in LIGO where the use of electronics

More information

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit. Current Transducer LDSR 0.3-TP/SP1 I P R N = 300 ma For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit. Features Closed

More information

1 Written and composed by: Prof. Muhammad Ali Malik (M. Phil. Physics), Govt. Degree College, Naushera

1 Written and composed by: Prof. Muhammad Ali Malik (M. Phil. Physics), Govt. Degree College, Naushera ELECTROMAGNETISM Q # 1. Describe the properties of magnetic field due to current in a long straight conductor. Ans. When the heavy current is passed through a straight conductor: i. A magnetic field is

More information

Using a Mercury itc with thermocouples

Using a Mercury itc with thermocouples Technical Note Mercury Support Using a Mercury itc with thermocouples Abstract and content description This technical note describes how to make accurate and reliable temperature measurements using an

More information

LM34 - Precision Fahrenheit Temperature Sensor

LM34 - Precision Fahrenheit Temperature Sensor - Precision Fahrenheit Temperature Sensor Features Typical Application Calibrated directly in degrees Fahrenheit Linear +10.0 mv/ F scale factor 1.0 F accuracy guaranteed (at +77 F) Parametric Table Supply

More information

MAGNETIC DEFLECTION. OBJECTIVE: To observe the effect of a magnetic field on an electron beam. To measure the Earth s magnetic field.

MAGNETIC DEFLECTION. OBJECTIVE: To observe the effect of a magnetic field on an electron beam. To measure the Earth s magnetic field. MAGNETIC DEFLECTION OBJECTIVE: To observe the effect of a magnetic field on an electron beam. To measure the Earth s magnetic field. THEORY: Moving charges exert forces on one another that are not observed

More information

(Refer Slide Time: 01:16)

(Refer Slide Time: 01:16) Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras Module - 4 Lecture - 50 Case Studies This will be lecture 50 in

More information

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors Slide 1 Electronic Sensors Electronic sensors can be designed to detect a variety of quantitative aspects of a given physical system. Such quantities include: Temperatures Light (Optoelectronics) Magnetic

More information

Superconductors A class of materials and compounds whose resistances fall to virtually zero below a certain temperature, T C T C is called the critical temperature The graph is the same as a normal metal

More information

Last Revision: August,

Last Revision: August, A3-1 HALL EFFECT Last Revision: August, 21 2007 QUESTION TO BE INVESTIGATED How to individual charge carriers behave in an external magnetic field that is perpendicular to their motion? INTRODUCTION The

More information

Keywords: sensor compensation, integrated circuit, ic, compensation algorithm, piezoresistive, prt, pressure sensors, temperature

Keywords: sensor compensation, integrated circuit, ic, compensation algorithm, piezoresistive, prt, pressure sensors, temperature Maxim > Design Support > Technical Documents > Tutorials > ASICs > APP 2024 Maxim > Design Support > Technical Documents > Tutorials > Sensors > APP 2024 Keywords: sensor compensation, integrated circuit,

More information

Lab E3: The Wheatstone Bridge

Lab E3: The Wheatstone Bridge E3.1 Lab E3: The Wheatstone Bridge Introduction The Wheatstone bridge is a circuit used to compare an unknown resistance with a known resistance. The bridge is commonly used in control circuits. For instance,

More information

PhysicsAndMathsTutor.com

PhysicsAndMathsTutor.com Electricity May 02 1. The graphs show the variation with potential difference V of the current I for three circuit elements. PhysicsAndMathsTutor.com When the four lamps are connected as shown in diagram

More information

AP Physics Electricity and Magnetism #3 Capacitors, Resistors, Ohm s Law, Electric Power

AP Physics Electricity and Magnetism #3 Capacitors, Resistors, Ohm s Law, Electric Power Name Period AP Physics Electricity and Magnetism #3 Capacitors, Resistors, Ohm s Law, Electric Power Dr. Campbell 1. The two plates of a capacitor hold +2500 µc and -2500 µc of charge, respectively, when

More information

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano Preparatory Course (task NA 3.6) Basics of experimental testing and theoretical background Module I Module I: traditional test instrumentation and acquisition systems Prof. Ramat, Stefano Transducers A

More information

TEMPERATURE SENSOR. Model TH-T. Roctest Limited, All rights reserved.

TEMPERATURE SENSOR. Model TH-T. Roctest Limited, All rights reserved. INSTRUCTION MANUAL TEMPERATURE SENSOR Model TH-T Roctest Limited, 2005. All rights reserved. This product should be installed and operated only by qualified personnel. Its misuse is potentially dangerous.

More information

PROCESS CONTROL BASIS FOR A COST-EFFECTIVE SELECTIVE SOLDERING PROCESS

PROCESS CONTROL BASIS FOR A COST-EFFECTIVE SELECTIVE SOLDERING PROCESS PROCESS CONTROL BASIS FOR A COST-EFFECTIVE SELECTIVE SOLDERING PROCESS Christian Ott Kreuzwertheim, Germany christian.ott@seho.de Heike Schlessmann heike.schlessmann@seho.de Reiner Zoch reiner.zoch@seho.de

More information

LAB I WHAT IS IN IT AND WHY?

LAB I WHAT IS IN IT AND WHY? LAB I WHAT IS IN IT AND WHY? Study Questions: 1. Make a list of 3 examples for each of the following material properties: 1. Mechanical properties 2. Electrical properties 3. Optical properties 4. Magnetic

More information

LAB I WHAT IS IN IT AND WHY?

LAB I WHAT IS IN IT AND WHY? LAB I WHAT IS IN IT AND WHY? Study (Pre-Lab) Questions (Complete these BEFORE coming to lab, there will be Study/Pre-Lab Questions for each lab this quarter): 1. Make a list of 3 examples for each of the

More information

EXPERIMENT 9 Superconductivity & Ohm s Law

EXPERIMENT 9 Superconductivity & Ohm s Law Name: Date: Course number: MAKE SURE YOUR TA OR TI STAMPS EVERY PAGE BEFORE YOU START! Lab section: Partner's name(s): Grade: EXPERIMENT 9 Superconductivity & Ohm s Law 0. Pre-Laboratory Work [2 pts] 1.

More information

4/3/2019. Advanced Measurement Systems and Sensors. Dr. Ibrahim Al-Naimi. Chapter one. Introduction to Measurement Systems

4/3/2019. Advanced Measurement Systems and Sensors. Dr. Ibrahim Al-Naimi. Chapter one. Introduction to Measurement Systems Advanced Measurement Systems and Sensors Dr. Ibrahim Al-Naimi Chapter one Introduction to Measurement Systems 1 Outlines Control and measurement systems Transducer/sensor definition and classifications

More information

DATA SHEET. Thermocouple module with digital I²C-Interface - THMOD-I²C. Characteristic features. Areas of application. Features.

DATA SHEET. Thermocouple module with digital I²C-Interface - THMOD-I²C. Characteristic features. Areas of application. Features. Description Characteristic features Industrial temperature measuring method Wide measuring range, -270...+1360 C Digital I²C-interface Simple integration to micro-controller Scope of supply with thermocouple,

More information

Catalysis CAPABILITIES

Catalysis   CAPABILITIES Catalysis www.extrel.com CAPABILITIES Contents Extrel instruments have been recognized for their exceptional performance by the world s leading researchers for more than 50 years. Reliability and flexibility

More information

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY /6.071 Introduction to Electronics, Signals and Measurement Spring 2006

MASSACHUSETTS INSTITUTE OF TECHNOLOGY /6.071 Introduction to Electronics, Signals and Measurement Spring 2006 MASSACHUSETTS INSTITUTE OF TECHNOLOGY 22.07/6.07 Introduction to Electronics, Signals and Measurement Spring 2006 Lab6. Resistor Networks 2. Experiment. Power rating of resistors. Power rating is a very

More information

Ref: HLSR 16-PW; HLSR 32-PW; HLSR 40-PW-000; HLSR 50-PW-000,

Ref: HLSR 16-PW; HLSR 32-PW; HLSR 40-PW-000; HLSR 50-PW-000, Digital Current Transducer HLSR-PW series I P N = 16... 50 A Ref: HLSR 16-PW; HLSR 32-PW; HLSR 40-PW-000; HLSR 50-PW-000, Bitstream output from on onboard Sigma Delta modulator. For the electronic measurement

More information

Safety Precautions WARNING If critical situations that could lead to user s death or serious injury is assumed by mishandling of the product.

Safety Precautions WARNING If critical situations that could lead to user s death or serious injury is assumed by mishandling of the product. Safety Precautions Observe the following notices to ensure personal safety or to prevent accidents. To ensure that you use this product correctly, read this User s Manual thoroughly before use. Make sure

More information

Test of the Qualification Model of UG electronics for the AMS TRD Gas System

Test of the Qualification Model of UG electronics for the AMS TRD Gas System Test of the Qualification Model of UG electronics for the AMS TRD Gas System Bruno Borgia, Alessandro Bartoloni, Francesca R. Spada April, 9 AMS Roma / rev. Contents Introduction. The UG-crate................................

More information

STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING

STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING 1 YEDITEPE UNIVERSITY ENGINEERING FACULTY MECHANICAL ENGINEERING LABORATORY 1. Objective: Strain Gauges Know how the change in resistance

More information

THE FRANCK-HERTZ EXPERIMENT

THE FRANCK-HERTZ EXPERIMENT Rice University Physics 332 THE FRANCK-HERTZ EXPERIMENT I. INTRODUCTION... 2 II. THEORETICAL CONSIDERATIONS... 3 III. MEASUREMENT AND ANALYSIS... 5 Revised July 1991 I. Introduction By the early part of

More information

Grounding and Shielding

Grounding and Shielding Grounding and Shielding Veljko Radeka, BNL Outline Primary guidelines The Liquid Argon Calorimeter system Prevention of EMI and isolation Safety grounds Interface with other subsystems Status of implementation

More information

Cryogenic Instrumentation I Thermometry OUTLINE Thermometry Pt (pure metal) Temperature Ranges of Thermometer Application Typical Resistive Thermal

Cryogenic Instrumentation I Thermometry OUTLINE Thermometry Pt (pure metal) Temperature Ranges of Thermometer Application Typical Resistive Thermal Cryogenic Instrumentation I 1. Thermometry 2. anges of Application 3. Constant Volume 4. Thermocouples 5. Time esponse Data 6. 4 Terminal esistance Measurement OUTLINE 8. Pt (pure metal) 9. Typical esistive

More information

QNET Experiment #05: HVAC System Identification. Heating, Ventilation, and Air Conditioning Trainer (HVACT) Student Manual

QNET Experiment #05: HVAC System Identification. Heating, Ventilation, and Air Conditioning Trainer (HVACT) Student Manual Quanser NI-ELVIS Trainer (QNET) Series: QNET Experiment #05: HVAC System Identification Heating, Ventilation, and Air Conditioning Trainer (HVACT) Student Manual Table of Contents 1. Laboratory Objectives...1

More information

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES (RPB-01) Approved by: Date: SECTION PAGE # 1.0 Scope 2.0 Definitions 3.0 References 4.0 Inspection

More information

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO 81 Name Date Partners Lab 5 - ELECTRON CHARGE-TO-MASS RATIO OBJECTIVES To understand how electric and magnetic fields impact an electron beam To experimentally determine the electron charge-to-mass ratio

More information

Sensors and Actuators Sensors Physics

Sensors and Actuators Sensors Physics Sensors and Actuators Sensors Physics Sander Stuijk (s.stuijk@tue.nl) Department of Electrical Engineering Electronic Systems HEMOESISIVE SENSOS (Chapter 16.3) 3 emperature sensors placement excitation

More information

Testing and Final Construction of the Superconducting Magnet for the Alpha Magnetic Spectrometer

Testing and Final Construction of the Superconducting Magnet for the Alpha Magnetic Spectrometer Testing and Final Construction of the Superconducting Magnet for the Alpha Magnetic Spectrometer The MIT Faculty has made this article openly available. Please share how this access benefits you. Your

More information

Modern Physics Laboratory MP2 Blackbody Radiation

Modern Physics Laboratory MP2 Blackbody Radiation Purpose MP2 Blackbody Radiation In this experiment, you will investigate the spectrum of the blackbody radiation and its dependence on the temperature of the body. Equipment and components Tungsten light

More information

Smartec Pressure Sensor (bridge output)

Smartec Pressure Sensor (bridge output) DATASHEET Pressure sensor family 1/5 Smartec Pressure Sensor (bridge output) Features * Commercial grade * DIP version for high volume production * Gauge or absolute pressure * Resistive bridge technology

More information

HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY

HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY Accurate temperature measurement is key for controlling the stability and repeatability of many temperature-critical processes.

More information

CHV Series Vector Control Inverter Options. Operating Instructions for Tension Control Card

CHV Series Vector Control Inverter Options. Operating Instructions for Tension Control Card CHV Series Vector Control Inverter Options Operating Instructions for Control Card 1. Model and Specifications 1.1 Model The model of tension card is CHV00ZL. CHV series inverter can conduct constant

More information

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces

More information

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed

More information

A F/4 B F/8 C 2F D 4F E 8F. Answer: Because F M A. /r 2 or eight times what it was 8F. Answer:

A F/4 B F/8 C 2F D 4F E 8F. Answer: Because F M A. /r 2 or eight times what it was 8F. Answer: Test 7 Section A 2 Core short answer questions: 50 marks Section B 2 Detailed studies short answer questions: 120 marks Suggested time: 90 2100 minutes Section A: Core short answer questions Specific instructions

More information

Circuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore

Circuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore Circuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore Lecture No. # 08 Temperature Indicator Design Using Op-amp Today,

More information

Calibration of the Modular Neutron Array (MoNA)

Calibration of the Modular Neutron Array (MoNA) Calibration of the Modular Neutron Array (MoNA) Robert Pepin Gonzaga University ~1~ Calibration of the Modular Neutron Array (MoNA): Figure 1 - A rendering of the Modular Neutron Array In recent years

More information

Downloaded from

Downloaded from CHAPTER 12 ELECTRICITY Electricity is a general term that encompasses a variety of phenomena resulting from the presence and flow of electric charge. These include many easily recognizable phenomena such

More information

Sensing, Computing, Actuating

Sensing, Computing, Actuating Sensing, Computing, Actuating Sander Stuijk (s.stuijk@tue.nl) Department of Electrical Engineering Electronic Systems HEMOESISIVE SENSOS AND LINEAIZAION (Chapter.9, 5.11) 3 Applications discharge air temperature

More information

High performance DC link capacitor/bus sourcing dual Infineon HybridPACK TM Drive inverters for EV applications

High performance DC link capacitor/bus sourcing dual Infineon HybridPACK TM Drive inverters for EV applications High performance DC link capacitor/bus sourcing dual Infineon HybridPACK TM Drive inverters for EV applications Michael A. Brubaker, SBE Inc., USA, MichaelB@sbelectronics.com Terry A. Hosking, SBE Inc.,

More information

Cosmic Ray Telescope for the Effects of Radiation to Spacecraft Thermal Interface Control Document

Cosmic Ray Telescope for the Effects of Radiation to Spacecraft Thermal Interface Control Document Effective Date: 01/11/2007 Expiration Date: 01/11/2012 Lunar Reconnaissance Orbiter Project Cosmic Ray Telescope for the Effects of Radiation to Spacecraft Thermal Interface Control Document December 19,

More information

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION Columbia Physics: Lab -6 (ver. 10) 1 EXPERMENT -6 e/m OF THE ELECTRON GENERAL DSCUSSON The "discovery" of the electron by J. J. Thomson in 1897 refers to the experiment in which it was shown that "cathode

More information

Electromotive Force. The electromotive force (emf), ε, of a battery is the maximum possible voltage that the battery can provide between its terminals

Electromotive Force. The electromotive force (emf), ε, of a battery is the maximum possible voltage that the battery can provide between its terminals Direct Current When the current in a circuit has a constant magnitude and direction, the current is called direct current Because the potential difference between the terminals of a battery is constant,

More information

NPSAT1 Solar Cell Measurement System

NPSAT1 Solar Cell Measurement System NPSAT1 Solar Cell Measurement System Presented by Captain John Salmon, USMC Space Systems Academic Group 777 Dyer Rd., Bldg. 233 Code (SP/Sd), Rm. 125 Monterey, CA 93943 (831) 656-7521 Topics NPSAT1 Overview

More information

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO 79 Name Date Partners OBJECTIVES OVERVIEW Lab 5 - ELECTRON CHARGE-TO-MASS RATIO To understand how electric and magnetic fields impact an electron beam To experimentally determine the electron charge-to-mass

More information

Glossary Innovative Measurement Solutions

Glossary Innovative Measurement Solutions Glossary GLOSSARY OF TERMS FOR TRANSDUCERS, LOAD CELLS AND WEIGH MODULES This purpose of this document is to provide a comprehensive, alphabetical list of terms and definitions commonly employed in the

More information

National 5 Physics. Electricity and Energy. Notes

National 5 Physics. Electricity and Energy. Notes National 5 Physics Electricity and Energy Notes Name. 1 P a g e Key Area Notes, Examples and Questions Page 3 Conservation of energy Page 10 Electrical charge carriers and electric fields and potential

More information

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER UNIT G485 Module 1 5.1.2 Magnetic Fields 11 Thus, in order for the particle to suffer NO DEFLECTION and so exit the device at Y : From which : MAGNETIC FORCE UP = ELECTRIC FORCE DOWN BQv = EQ THE MASS

More information

Developing a Compact Doppler/Magnetograph for Solar Diagnostics

Developing a Compact Doppler/Magnetograph for Solar Diagnostics Developing a Compact Doppler/Magnetograph for Solar Diagnostics Ivan Bernal (1), Dr. Neil Murphy (2) Victor Valley College (1), Jet Propulsion Laboratory (2) Abstract: Line-of-sight velocity and magnetic

More information

Thermal Modeling of Power Devices in a Surface Mount Configuration

Thermal Modeling of Power Devices in a Surface Mount Configuration AN5 Thermal Modeling of Power Devices in a Surface Mount Configuration Thermal Modeling of Power Devices in a Surface Mount Configuration AN5 AN5 Introduction This Application Note demonstrates how to

More information

Laboratory 14: Ratio of Charge to Mass for the Electron

Laboratory 14: Ratio of Charge to Mass for the Electron Laboratory 14: Ratio of Charge to Mass for the Electron Introduction The discovery of the electron as a discrete particle of electricity is generally credited to the British physicist Sir J. J. Thomson

More information

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5.

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5. HMS-5000 Manual Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K - Manual version: ver 5.01- www.ecopia21.co.kr - Table of contents - 1. Hardware Installation

More information

Community College of Allegheny County Unit 9 Page #1. Thermocouples R1 = 1K

Community College of Allegheny County Unit 9 Page #1. Thermocouples R1 = 1K 10K Community College of Allegheny County Unit 9 Page #1 Thermocouples +12V Thermocouple Junction Vin Copper Wire Constantan Wire + - 3 2 741 7 4 1 5-12V 6 V Vout R1 = 1K Rf = 100K Engineers are not expected

More information

Digital Current Transducer HO-SW series I P N = A. Ref: HO 100-SW; HO 150-SW; HO 200-SW; HO 250-SW

Digital Current Transducer HO-SW series I P N = A. Ref: HO 100-SW; HO 150-SW; HO 200-SW; HO 250-SW Digital Current Transducer HO-SW series I P N = 100... 250 A Ref: HO 100-SW; HO 150-SW; HO 200-SW; HO 250-SW Bitstream output from on onboard Sigma Delta modulator. For the electronic measurement of current:

More information

Harnessing the Power of Arduino for the Advanced Lab

Harnessing the Power of Arduino for the Advanced Lab P P Herbert Jaeger + Harnessing the Power of Arduino for the Advanced Lab (Final Version) ALPhA Immersion Workshop July 27 29, 2017 Department of Physics Indiana University Purdue University Ft. Wayne,

More information

costech instruments ECS 4010 Elemental Combustion System CHNS-O

costech instruments ECS 4010 Elemental Combustion System CHNS-O instruments ECS 4010 Elemental Combustion System CHNS-O costech Costech Analytical Tecnologies Inc. 26074 Avenue Hall, Suite 14 Valencia, CA 91355 - USA Phone: (661) 297-2395 Fax: (661) 297-5492 e-mail:

More information

EE 42/100 Lecture 3: Circuit Elements, Resistive Circuits. Rev D 1/22/2012 (4:19PM) Prof. Ali M. Niknejad

EE 42/100 Lecture 3: Circuit Elements, Resistive Circuits. Rev D 1/22/2012 (4:19PM) Prof. Ali M. Niknejad A. M. Niknejad University of California, Berkeley EE 100 / 42 Lecture 3 p. 1/22 EE 42/100 Lecture 3: Circuit Elements, Resistive Circuits ELECTRONICS Rev D 1/22/2012 (4:19PM) Prof. Ali M. Niknejad University

More information

Circuit Analysis and Ohm s Law

Circuit Analysis and Ohm s Law Study Unit Circuit Analysis and Ohm s Law By Robert Cecci Circuit analysis is one of the fundamental jobs of an electrician or electronics technician With the knowledge of how voltage, current, and resistance

More information

NEW YORK CENTRAL SYSTEM HOT BOX DETECTORS JUNE 1, 1959

NEW YORK CENTRAL SYSTEM HOT BOX DETECTORS JUNE 1, 1959 NEW YORK CENTRAL SYSTEM HOT BOX DETECTORS JUNE 1, 1959 N E W Y O R K C E N T R A L S Y S T E M HOT BOX DETECTORS INTRODUCTION The purpose of the information contained herein is to give all concerned a

More information

MET 301 EXPERIMENT # 2 APPLICATION OF BONDED STRAIN GAGES

MET 301 EXPERIMENT # 2 APPLICATION OF BONDED STRAIN GAGES MET 301 EPERIMENT # 2 APPLICATION OF BONDED STRAIN GAGES 1. Objective To understand the working principle of bonded strain gauge and to study the stress and strain in a hollow cylindrical shaft under bending,

More information

SC125MS. Data Sheet and Instruction Manual. ! Warning! Salem Controls Inc. Stepper Motor Driver. Last Updated 12/14/2004

SC125MS. Data Sheet and Instruction Manual. ! Warning! Salem Controls Inc. Stepper Motor Driver.   Last Updated 12/14/2004 SC125MS Stepper Motor Driver Salem Controls Inc. Last Updated 12/14/2004! Warning! Stepper motors and drivers use high current and voltages capable of causing severe injury. Do not operate this product

More information

INTERVALOMETER POWER CONSUMPTION

INTERVALOMETER POWER CONSUMPTION INTERVALOMETER POWER CONSUMPTION One of the key things I wanted from the design was to keep the power consumption low. By putting the PIC to sleep one can get ridiculously low power consumption but it

More information

Single Event Radiation Test Report. Microsemi LX Analog Input RAD Tolerant Telemetry Controller

Single Event Radiation Test Report. Microsemi LX Analog Input RAD Tolerant Telemetry Controller Single Event Radiation Test Report Microsemi LX7730 64 Analog Input RAD Tolerant Telemetry Controller Doc. N :TDRS-0020-4 Page 1/33 TABLE OF CONTENT 1. PURPOSE AND SCOPE... 3 2. APPLICABLE DOCUMENT...

More information

TE 75R RESEARCH RUBBER FRICTION TEST MACHINE

TE 75R RESEARCH RUBBER FRICTION TEST MACHINE TE 75R RESEARCH RUBBER FRICTION TEST MACHINE Background: The Research Rubber Friction Test Machine offers the ability to investigate fully the frictional behaviour of rubbery materials both in dry and

More information

Introduction to Blackbody Sources

Introduction to Blackbody Sources Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with

More information

Energy loss of alpha particles - Prelab questions

Energy loss of alpha particles - Prelab questions Energy loss of alpha particles - Prelab questions 1. Write down the decay path from 226 Ra to 206 Pb. Show the intermediate nuclides and the nuclear reactions which cause each transformation (α/β ± decay).

More information

BMXART0814 analog input module M340-8 inputs - temperature

BMXART0814 analog input module M340-8 inputs - temperature Characteristics analog input module M340-8 inputs - temperature Complementary Analog/Digital conversion Analogue input resolution Input impedance Permitted overload on inputs Common mode rejection Differential

More information

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision ARGUS VI Static Vacuum Mass Spectrometer Static Vacuum ARGUS VI Multicollection Low Volume Precision Based on more than 20 years of experience in noble gas mass spectrometry instrumentation, we have developed

More information

Cosmic Ray Telescope for the Effects of Radiation to Spacecraft Thermal Interface Control Document

Cosmic Ray Telescope for the Effects of Radiation to Spacecraft Thermal Interface Control Document Effective Date: February 14, 2006 Expiration Date: February 14, 2011 Lunar Reconnaissance Orbiter Project Cosmic Ray Telescope for the Effects of Radiation to Spacecraft Thermal Interface Control Document

More information

Modified Transient Plane Source (MTPS): Theory of Operation

Modified Transient Plane Source (MTPS): Theory of Operation Pg of 0 Modified Transient Plane Source (MTPS): Theory of Operation The C-Therm TCi Thermal Conductivity Analyzer is based on the Modified Transient Plane Source (MTPS) technique. It employs a one-sided,

More information

University of Connecticut Department of Electrical and Computer Engineering. ECE 4901: Fall 2017 Spring 2018

University of Connecticut Department of Electrical and Computer Engineering. ECE 4901: Fall 2017 Spring 2018 University of Connecticut Department of Electrical and Computer Engineering ECE 4901: Fall 2017 Spring 2018 Team 1817 (Hubbell): Electrical plug, connector, and receptacle temperature sensor Jim Lin (EE)

More information

LEP 2 Energy Calibration and the Spectrometer

LEP 2 Energy Calibration and the Spectrometer LEP 2 Energy Calibration and the Spectrometer Goals of LEP 2 m W measurement LEP 2 Energy Model the Magnetic Extrapolation Cross-checking with the Flux Loop (no details!) Cross-checking with the Synchrotron

More information

NICMOS Status and Plans

NICMOS Status and Plans 1997 HST Calibration Workshop Space Telescope Science Institute, 1997 S. Casertano, et al., eds. NICMOS Status and Plans Rodger I. Thompson Steward Observatory, University of Arizona, Tucson, AZ 85721

More information

THERMOCOUPLE CHARACTERISTICS TRAINER

THERMOCOUPLE CHARACTERISTICS TRAINER THERMOCOUPLE CHARACTERISTICS TRAINER (Model No : ) User Manual Version 2.0 Technical Clarification /Suggestion : / Technical Support Division, Vi Microsystems Pvt. Ltd., Plot No :75,Electronics Estate,

More information