Position sensitive detection of thermal neutrons with solid state detectors (Gd Si planar detectors)

Size: px
Start display at page:

Download "Position sensitive detection of thermal neutrons with solid state detectors (Gd Si planar detectors)"

Transcription

1 Nuclear Instruments and Methods in Physics Research A 424 (1999) Position sensitive detection of thermal neutrons with solid state detectors (Gd Si planar detectors) G. Bruckner*, A. Czermak, H. Rauch, P. Weilhammer Institut Laue Langevin, B.P. 156, F Grenoble, France Atominstitut der O$ sterreichischen Universita( ten, Stadionallee 2, A-1020 Wien, Austria Institute of Nuclear Physics, PL Krakow, Poland CERN, CH-1211 Geneva 23, Switzerland Abstract Recent advances in the technology of position sensitive silicon detectors and the corresponding electronics allow the construction of fast time response thermal neutron detectors. These detectors also exhibit excellent position resolution by combination of silicon detectors with thin Gd converter foils. We constructed several one- and two-dimensional prototype detectors, using DC and AC coupled silicon strip detectors, pad detectors and different VLSI readout electronics. The position resolution and the detector efficiency for different converters at wavelengths from 1.1 to 3.3 A were determined at the TRIGA reactor in Vienna and at the ILL in Grenoble. Spatial resolutions of less than 100 μm and efficiencies up to 40% have been achieved. The results of these measurements are compared with a Monte Carlo simulation of the detector operation. These detectors can also be used for phase topography experiments using perfect crystal neutron interferometers. In certain cases an increase of the sensitivity in the order of 100 can be anticipated compared to beam attenuation radiography Elsevier Science B.V. All rights reserved. Keywords: Position sensitive neutron detector; Silicon; Gadolinium; Image 1. Introduction The central concept of these new detectors is to combine position sensitive Si planar sensors with external Gd converter foils [1 3]. The thermal neutrons are absorbed in the Gd foil and the resulting conversion electrons are detected in the silicon devices. Fig. 1 shows a schematic drawing of * Corresponding author: Tel: # ; fax: # ; bruckner@ill.fr. a Gd Si neutron detector consisting of one Gd converter foil and two Si detectors. A position sensitive Si planar detector consists of an array of many pn-diodes on one single substrate. Each pn-junction works as a detector for ionizing particles. The position of the detected particle is given by the individual address of the diode that collects the charge. Each diode is equipped with an individual amplifier chain to ensure real time readout. Typical wafers have a diameter of 4 6 in. and a thickness of about 300 μm. The single structures can be as small as 10 μm and are limited in their /99/$ see front matter 1999 Elsevier Science B.V. All rights reserved. PII: S ( 9 8 )

2 184 G. Bruckner et al./nuclear Instruments and Methods in Physics Research A 424 (1999) Fig. 2. Spectra of the reaction products of the neutron capture in Gd for different neutron wavelengths, as seen in a Siplanar detector in backward direction. Fig. 1. Schematic of a 4π Gd Si-neutron detector consisting of one converter foil and two Si-detectors with corresponding electronics. extent only by the size of the wafer. A position resolution of less than 10 μm has been achieved with these devices [4], as well as the construction of large detectors with areas up to m, for several high-energy physics experiments [5 7]. 2. External converter Considering only one converter layer (no sandwich detectors), the (n, γ) reaction in Gd, which produces also a reasonable number of conversion electrons, is the only suitable one for high detector efficiency. The proton, α-particle and triton from the Li and B reactions have a very short range. Therefore, these converters can only be used in situations, which require a low detection efficiency, e.g. beam monitors, or for the detection of cold or ultra cold neutrons. One can also think of doping the Si detectors with Li, but the usual doping concentrations of about 10/cm are much too low for efficient absorption of thermal neutrons. The (n, γ) reaction in Gd produces a quite complex conversion electron spectra (Figs. 2 and 3) Fig. 3. Spectra of the reaction products of the neutron capture in natural Gd for different neutron wavelengths, as seen in a Si-planar detector in backward direction. with significant energy lines between 29 and about 200 kev. Due to the high absorption cross section of Gd (σ " b for natural Gd, b for Gd) the converter can be made very thin, so that the range of the conversion electrons is sufficient to reach the Si-detector. The band-gap in Si of 1.11 ev results in an average energy of about 3.6 kev for the electron hole creation. Therefore, the device must be capable to measure a charge of less than 5000 e to detect the low-energy conversion electrons,

3 G. Bruckner et al./nuclear Instrument and Methods in Physics Research A 424 (1999) which have already lost part of their energy when escaping the converter foil. 3. Planar diode detectors and VLSI readout electronics The detectors are made of high resistivity silicon in planar technology and operated in fully depleted mode [8,9]. For a typical detector less than 100 V applied to the pn-junction ensures, that the whole detector works as one large depletion region. This mode decreases the detector noise, because it reduces the capacity between strip and backplane, and the thermally generated leakage currents. In addition, it guarantees that the entrance window on the rear side is as thin as the one on the front side and that there are no dead sections between the strips. While CCDs are made of only partly sensitive shift registers and contain inefficient channel stops to separate the colums, Si planar detectors are continuously efficient devices. Another important distinction between these detectors and CCDs concerns their readout. CCDs are integrating devices with charge transfer, that means serial readout. Therefore, a high-resolution time information is very difficult to obtain. Contrary Si detectors are constructed to measure single events with parallel readout electronics. Provided the lower level discriminator is set sufficiently high, the energy distribution of the conversion electrons does not effect the detector response, as in the case when the spectra would be detected by an integrating device such as film or CCD. The theoretical time resolution of the Gd Si detector is given by the charge collection time in Si, the transit time of the thermal neutrons through the converter foil and the lifetime of the excited Gd atoms after the neutron capture. None of these intervals is longer than some nanoseconds, which allows the construction of a very fast detector. For parallel readout it is necessary to equip each detector diode with the full amplification and readout electronics (more sophisticated readout techniques can reduce the number of channels). For many channels and/or a high position resolution this can only be done by the use of very large-scale integrated electronics (VLSI) [10]. Although the technique used for the detector fabrication is from a general point of view similar to the one used for electronic circuits, it differs in some essential aspects which prevent so far the construction of detector and electronics on the same wafer. 4. Efficiency The probability P (x) that the neutron will be absorbed in the converter between x and x#dx after surviving the distance x is given by the properties of the converter material namely the absorption cross section σ and number of converter atoms per unit volume N P (x)"(1!e ) Nσ dx. (1) For every stopped neutron the probability, that an electron will be emitted is given by the conversion factors of the corresponding (n, γ) reaction. As the decay pattern of Gd is quite complex some simplifications have to be made for the calculations of the detector efficiency. Taking the values of the dominant isotopes Gd and Gd from the nuclear data sheets, we used only the lowest energy levels which have the highest conversion factors, E (kev) "88.97, , , for Gd and E (kev)"79.51, , , for Gd [11]. The energy distribution of the conversion electrons was calculated, by subtracting the binding energies of the K, L and M shells. Although each conversion electron is accompanied by several γ-rays as well as X-rays and Auger electrons, which follow the ionization of the atom, only the conversion electrons, and the Auger electrons with an energy of more than 20 kev have been taken into account. To each electron created between x and x#dx, an energy E and a creation probability can be assigned according to the energy distribution, together with an isotropically distributed emission angle. From these values the probability for the electrons to reach the detector volume has been calculated by evaluating the escape probability successively for every layer the electron has to penetrate (Gd-converter, Al-contacts, p-doping) using the following semiempirical equations [12]. The range R in (μm) is a function of the initial

4 186 G. Bruckner et al./nuclear Instruments and Methods in Physics Research A 424 (1999) energy E (kev), the atomic weight A (g/mol), the atomic number Z, and the density ρ (g/cm) R (μm)" ) A ) E (kev). (2) Z ) ρ (g/cm) The residual energy E after penetrating a distance s in a given material can be expressed in terms of the reduced range y"s/r E "(1!y) ) E. (3) The electron transmission in forward direction is also a function of the reduced range and a parameter γ, which accounts for multiple scattering and diffusion losses. I "exp γ ) y I! γ"0.187 ) Z. (4) 1!y For all electrons, which enter the Si detector with a remaining energy of at least 15 kev, the efficiency of the Si detector can be assumed to be 100%. Our calculations show, that the detector efficiency depends critically on the cross section of the element and is therefore a function of the neutron energy (Fig. 5). The extremely high cross section of Gd results in different characteristic of the efficiency in forward and backward direction, as can be seen in Fig. 4. This has to be kept in mind when comparing Gd with other converters as for example Li. The efficiency, defined as the number of registered neutrons divided by the number of incident neutrons, was measured in backward direction for six different wavelengths in comparison to a calibrated He counter. The results for natural Gd and an enriched (90.5%) Gd converter are shown in Fig. 5. Figs. 2 and 3 show the spectra of the reaction products of the neutron capture in Gd and nat. Gd as seen by the Si-planar detector for the measurements made at the ILL. For thermal neutrons we can conclude an efficiency of more than 35% for the enriched Gd converter and about half of this value for natural Gd. One can clearly see the increase towards higher wavelengths. The measured values are constantly higher than the calculated ones. The dashed line in Fig. 5 gives the calculated values, but with a constant offset. From the spectra in Figs. 2 and 3 one can assume, Fig. 4. Calculated detection efficiency for neutrons with 1.8 A as a function of the converter thickness for different converters. Fig. 5. Detector efficiencies in backward direction for different wavelengths, measured at the Atominstitut in Vienna (ATI) and the ILL in Grenoble.

5 G. Bruckner et al./nuclear Instrument and Methods in Physics Research A 424 (1999) Fig. 6. Thermal neutron images and histograms for simply formed absorbers as seen by the pad detector. Original size 1616 pads with μm each, respectively, mm. that this is caused by X-rays and γ-rays following the neutron capture, which are partly detected. The X-ray lines of Gd can be seen in both spectra, although, it is more dominant for natural Gd. Conversion electrons, γ-rays and X-rays are emitted simultaneously compared to the time resolution of the amplifier of about 1 μs, therefore, one captured neutron can cause only one signal in the detector. Nevertheless, the electromagnetic radiation can increase the overall efficiency in the case when the electron is not detected. Otherwise, the detection of γ-rays and X-rays causes only a spread of the energy spectra. 5. Position sensitive detectors With our first measurements about three years ago, we could demonstrate that a one-dimensional neutron detector can be constructed from a single-sided strip detector [2]. The position resolution achieved with this early system is compared with the results of a pad detector and the latest measurements with a double-sided strip detector in Fig. 7. Fig. 7. Comparison of the position resolution, defined by an Lorentzian edge spread function, for three different detectors. A pad detector offers the advantage of a twodimensional position information on a single-sided detector design [13]. The sensor used for neutron detection has 256 pads with pitch of 300 μm, corresponding to an area of mm. For the readout we used two VA2 chips, each consisting of 128 channels with preamplifier, shaping amplifier and output multiplexer. The external trigger for this electronic system was provided by processing the backplane signal of the detector with a shorter shaping time (500 ns) than that of the VA2 amplifiers (1.2 μs) [13,14]. Fig. 6 shows the images and histograms of some simply shaped neutron absorbers taken with the pad detector, in comparison with a negative film pattern and a scanner image, respectively. The almost triangular shaped hole was VLSI chip produced by Integrated Detector and Electronics AS (IDE AS), Veritasveien 9, N-1322 Hovik, Norway.

6 188 G. Bruckner et al./nuclear Instruments and Methods in Physics Research A 424 (1999) Fig. 8. Transmission image of a micro switch next to a nut as seen by the double-sided strip detector. Original size mm. prepared by cutting an absorber into two pieces and making a notch in one side. The cut does not shield the neutrons completely, as can be seen in the histogram. As for the previous setup the position resolution was determined by measuring the detector response to an edge. An Lorentzian edge spread function was fitted to the data near the edge, which gives a position resolution of FWHM "210 μm (see Fig. 7). Figs. 8 and 9 show some images obtained with our latest system with neutrons of 2.1 A. It is based on a double-sided strip detector with a pitch of 50 μm strips were read out by five plus four 128 channel CMOS VLSI chips developed by IDE AS, which contain a low-noise charge sensitive preamplifier, a shaping amplifier and a discriminator for each channel. The self-triggering capability of this circuit is essential for neutron experiments, where the trigger cannot be provided by an external source. The system was setup in a way, so that only single hits were accepted with a shaping time of 1.2 μs and a threshold of 20 kev. The detector area of mm was only partly covered by the natural Gd converter, resulting in an effective area of mm. The measurements confirmed, that the distance between Gd converter foil and detector is the critical parameter for the position resolution. To take advantage of the extremely high resolution of the strip detector, the Gd converter has to be placed onto the detector surface or deposited directly on the Si wafer. The position resolution for neutrons. FWMH "95 μm has been obtained by shielding about half of the detector with a Li containing absorber and evaluating the data near to the edge. This value includes not only the variances produced in the detector but also the divergence of the neutron beam, the imperfection of the edge, the scattering of the neutrons at the absorber and geometric misalignments. We therefore expect a further improvement of this value for the near future.

7 G. Bruckner et al./nuclear Instrument and Methods in Physics Research A 424 (1999) References [1] B. Feigl, H. Rauch, Nucl. Instr. and Meth. 61 (3) (1968) 349. [2] G. Bruckner, H. Rauch, J. Neutron Res. 4 (1996) 141. [3] C. Petrillo, F. Sacchetti, O. Toker, N.J. Rhodes, Nucl. Instr. and Meth. A 378 (1996) 54. [4] J. Straver et al., Nucl. Instr. and Meth A 348 (1994) 485. [5] K. Borer et al., Nucl. Instr. and Meth A 253 (1987) 548. [6] Proc. 5th Int. Workshop on Vertex Detectors, Nucl. Instr. and Meth A 386 (1) (1997). [7] Proc. 2nd Int. Symp. on Development and Application of Semiconductor Tracking Detectors, Nucl. Instr. and Meth. A 383 (1) (1996). [8] Proc. 7th Symp. on Semiconductor Detectors, Nucl. Instr. and Meth. A 377 (1994). [9] G. Lutz, Nucl. Instr. and Meth 367 (1995) 21. [10] E. Nyga rd et al., Nucl. Instr. and Meth. A 301(1991) 506.P.Aspell et al. Nucl. Instr. and Meth. A 315 (1992) 425. [11] Nuclear Data Sheets 49 (2) (1986) and Nuclear Data Sheets 56 (2) (1989). [12] K. Kanaya, S. Okoyama, J.Phys. D 5 (1972). [13] P. Weilhammer et al., Nucl. Instr. and Meth. A 383 (1996) [14] G. Bruckner, H. Rauch, A. Rudge, P. Weilhammer, W. Dulinsky, SPIE 2867 (1996) 554. Fig. 9. Transmission image of a relay, mm, taken at a wavelength of 2.1 A with the double-sided strip detector. Acknowledgements This work has partly been supported by the EURATOM fusion program.

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH Semiconductor X-Ray Detectors Tobias Eggert Ketek GmbH Semiconductor X-Ray Detectors Part A Principles of Semiconductor Detectors 1. Basic Principles 2. Typical Applications 3. Planar Technology 4. Read-out

More information

CCD readout of GEM-based neutron detectors

CCD readout of GEM-based neutron detectors Nuclear Instruments and Methods in Physics Research A 478 (2002) 357 361 CCD readout of GEM-based neutron detectors F.A.F. Fraga a, *, L.M.S. Margato a, S.T.G. Fetal a, M.M.F.R. Fraga a, R. Ferreira Marques

More information

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection R. Kreuger, C. W. E. van Eijk, Member, IEEE, F. A. F. Fraga, M. M. Fraga, S. T. G. Fetal, R. W. Hollander, Member, IEEE, L. M.

More information

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland Available on CMS information server CMS NOTE 199/11 The Compact Muon Solenoid Experiment CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland 11 February 199 Temperature dependence of the

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

Characterization of 3D thermal neutron semiconductor detectors

Characterization of 3D thermal neutron semiconductor detectors Characterization of 3D thermal neutron semiconductor detectors J.Uher 1, C.Fröjdh 2, J.Jakůbek 1, C.Kenney 3, Z.Kohout 4, V.Linhart 1, S.Parker 5, S.Petersson 2, S.Pospíšil 1, G.Thungström 2 (1),, Czech

More information

Semiconductor Detectors

Semiconductor Detectors Semiconductor Detectors Summary of Last Lecture Band structure in Solids: Conduction band Conduction band thermal conductivity: E g > 5 ev Valence band Insulator Charge carrier in conductor: e - Charge

More information

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers Nuclear Instruments and Methods in Physics Research A 441 (2000) 459}467 Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers Z. He *, W.Li, G.F. Knoll, D.K. Wehe,

More information

Light ion recoil detector

Light ion recoil detector Light ion recoil detector Overall design The detector for light (target-like) particles is a substantial part of the R3B setup. It allows registration of recoils in coincidence with the heavy fragments,

More information

Method of active correlations in the experiment 249 Cf+ 48 Ca n

Method of active correlations in the experiment 249 Cf+ 48 Ca n Method of active correlations in the experiment 249 Cf+ 48 Ca 297 118 +3n Yu.S.Tsyganov, A.M.Sukhov, A.N.Polyakov Abstract Two decay chains originated from the even-even isotope 294 118 produced in the

More information

Lecture 18. New gas detectors Solid state trackers

Lecture 18. New gas detectors Solid state trackers Lecture 18 New gas detectors Solid state trackers Time projection Chamber Full 3-D track reconstruction x-y from wires and segmented cathode of MWPC z from drift time de/dx information (extra) Drift over

More information

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples)

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples) Detecting high energy photons Interactions of photons with matter Properties of detectors (with examples) Interactions of high energy photons with matter Cross section/attenution length/optical depth Photoelectric

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

Energy calibration of the threshold of Medipix for ATLAS

Energy calibration of the threshold of Medipix for ATLAS Energy calibration of the threshold of Medipix for ATLAS Céline Lebel Université de Montréal lebel@lps.umontreal.ca presenting for the Institut of Experimental and Applied Physics of the Czech Technical

More information

Peter Fischer, ziti, Universität Heidelberg. Silicon Detectors & Readout Electronics

Peter Fischer, ziti, Universität Heidelberg. Silicon Detectors & Readout Electronics Silicon Detectors and Readout Electronics Peter Fischer, ziti, Universität Heidelberg 1 Content of the Lecture (sorted by subject) Introduction: Applications of silicon detectors Requirements, measured

More information

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Position Sensitive Germanium Detectors for the Advanced Compton Telescope Position Sensitive Germanium Detectors for the Advanced Compton Telescope R.A. Kroeger 1, W.N. Johnson 1, J.D. Kurfess 1, B.F. Phlips, P.N. Luke 3, M. Momayezi 4, W.K. Warburton 4 1 Naval Research Laboratory,

More information

D. Meier. representing the RD42 Collaboration. Bristol University, CERN, CPP Marseille, Lawrence Livermore National Lab, LEPSI

D. Meier. representing the RD42 Collaboration. Bristol University, CERN, CPP Marseille, Lawrence Livermore National Lab, LEPSI Diamond as a Particle Detector D. Meier representing the RD42 Collaboration Bristol University, CERN, CPP Marseille, Lawrence Livermore National Lab, LEPSI Strasbourg, Los Alamos National Lab, MPIK Heidelberg,

More information

III. Energy Deposition in the Detector and Spectrum Formation

III. Energy Deposition in the Detector and Spectrum Formation 1 III. Energy Deposition in the Detector and Spectrum Formation a) charged particles Bethe-Bloch formula de 4πq 4 z2 e 2m v = NZ ( ) dx m v ln ln 1 0 2 β β I 0 2 2 2 z, v: atomic number and velocity of

More information

Advances in the Micro-Hole & Strip Plate gaseous detector

Advances in the Micro-Hole & Strip Plate gaseous detector Nuclear Instruments and Methods in Physics Research A 504 (2003) 364 368 Advances in the Micro-Hole & Strip Plate gaseous detector J.M. Maia a,b,c, *, J.F.C.A. Veloso a, J.M.F. dos Santos a, A. Breskin

More information

Nuclear cross-section measurements at the Manuel Lujan Jr. Neutron Scattering Center. Michal Mocko

Nuclear cross-section measurements at the Manuel Lujan Jr. Neutron Scattering Center. Michal Mocko Nuclear cross-section measurements at the Manuel Lujan Jr. Neutron Scattering Center Michal Mocko G. Muhrer, F. Tovesson, J. Ullmann International Topical Meeting on Nuclear Research Applications and Utilization

More information

CHARACTERISTICS OF LIGHT CHARGED PARTICLE EMISSION IN THE TERNARY FISSION OF 250 CF AND 252 CF AT DIFFERENT EXCITATION ENERGIES

CHARACTERISTICS OF LIGHT CHARGED PARTICLE EMISSION IN THE TERNARY FISSION OF 250 CF AND 252 CF AT DIFFERENT EXCITATION ENERGIES CHARACTERISTICS OF LIGHT CHARGED PARTICLE EMISSION IN THE TERNARY FISSION OF 25 CF AND 252 CF AT DIFFERENT EXCITATION ENERGIES S. VERMOTE AND C. WAGEMANS Department of Physics and Astronomy, University

More information

Chem 481 Lecture Material 3/20/09

Chem 481 Lecture Material 3/20/09 Chem 481 Lecture Material 3/20/09 Radiation Detection and Measurement Semiconductor Detectors The electrons in a sample of silicon are each bound to specific silicon atoms (occupy the valence band). If

More information

Alpha-particle Stopping Powers in Air and Argon

Alpha-particle Stopping Powers in Air and Argon Alpha-particle Stopping Powers in Air and Argon Mohammad QH 1* and Maghdid HA 2 1 Department of Physics, College of Science, Raparin University, Sulaimanyah, Iraqi Kurdistan, Iraq 2 Department of Physics,

More information

GEM-based photon detector for RICH applications

GEM-based photon detector for RICH applications Nuclear Instruments and Methods in Physics Research A 535 (2004) 324 329 www.elsevier.com/locate/nima GEM-based photon detector for RICH applications Thomas Meinschad, Leszek Ropelewski, Fabio Sauli CERN,

More information

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na Ellen Simmons 1 Contents Introduction Review of the Types of Radiation Charged Particle Radiation Detection Review of Semiconductor

More information

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER V.A. MOROZOV 1, N.V. MOROZOVA 1, T. BĂDICĂ 2, D. DELEANU 2,3, D. GHIŢĂ 2, S. PASCU 2,3 1 Joint

More information

EXPERIMENTS CHARACTERIZING THE X-RAY EMISSION FROM A SOLID-STATE CATHODE USING A HIGH-CURRENT GLOW DISCHARGE

EXPERIMENTS CHARACTERIZING THE X-RAY EMISSION FROM A SOLID-STATE CATHODE USING A HIGH-CURRENT GLOW DISCHARGE EXPERIMENTS CHARACTERIZING THE X-RAY EMISSION FROM A SOLID-STATE CATHODE USING A HIGH-CURRENT GLOW DISCHARGE A.B. KARABUT AND S.A. KOLOMEYCHENKO FSUE SIA LUCH 24 Zheleznodorozhnaja Street, Podolsk, Moscow

More information

Detection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics

Detection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics Detection of X-Rays Solid state detectors Proportional counters Microcalorimeters Detector characteristics Solid State X-ray Detectors X-ray interacts in material to produce photoelectrons which are collected

More information

OPTIMIZATION OF A NOVEL SOLID-STATE SELF POWERED NEUTRON DETECTOR

OPTIMIZATION OF A NOVEL SOLID-STATE SELF POWERED NEUTRON DETECTOR International Conference on Mathematics, Computational Methods & Reactor Physics (M&C 009) Saratoga Springs, New York, May 3-7, 009, on CD-ROM, American Nuclear Society, LaGrange Park, IL (009) OPTIMIZATION

More information

Detectors in Nuclear and High Energy Physics. RHIG summer student meeting June 2014

Detectors in Nuclear and High Energy Physics. RHIG summer student meeting June 2014 Detectors in Nuclear and High Energy Physics RHIG summer student meeting June 2014 Physics or Knowledge of Nature Experimental Data Analysis Theory ( application) Experimental Data Initial Conditions /

More information

Efficiency and Attenuation in CdTe Detectors

Efficiency and Attenuation in CdTe Detectors Efficiency and Attenuation in CdTe Detectors Amptek Inc. Bob Redus, May 5, 00 Amptek s XR-00T-CdTe is a high performance x-ray and gamma ray detector system. Like Amptek s other XR00 products, a detector

More information

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy Thomas Niedermayr, I. D. Hau, S. Terracol, T. Miyazaki, S. E. Labov and S. Friedrich Former colleagues: M. F. Cunningham, J. N.

More information

Development of a Pixel Detector for Ultra-Cold Neutrons

Development of a Pixel Detector for Ultra-Cold Neutrons Development of a Pixel Detector for Ultra-Cold Neutrons S. Kawasaki,a, G. Ichikawa a, M. Hino d, Y. Kamiya b, M. Kitaguchi d, S. Komamiya a,b, T. Sanuki c, S. Sonoda a a Graduate School of Science, University

More information

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler Energetic particles and their detection in situ (particle detectors) Part II George Gloeckler University of Michigan, Ann Arbor, MI University of Maryland, College Park, MD Simple particle detectors Gas-filled

More information

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy Taka Tanaka (SLAC/KIPAC) 9/19/2007 SLAC Advanced Instrumentation Seminar Outline Introduction CdTe Diode

More information

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 2001-22, p.37-42 EVALUATION OF ABSORPTION EFFICIENCY FOR NIS TUNNEL JUNCTION DETECTOR WITH SEGMENTED ABSORBERS R. Nouchi, I. Yamada,

More information

Characteristics of the Large-Area Stacked Microstructured Semiconductor Neutron Detector

Characteristics of the Large-Area Stacked Microstructured Semiconductor Neutron Detector Invited Paper Characteristics of the Large-Area Stacked Microstructured Semiconductor Neutron Detector S.L. Bellinger *a, R.G. Fronk a, T.J. Sobering b, D.S. McGregor a a S.M.A.R.T. Laboratory, Department

More information

Correction of the electric resistivity distribution of Si wafers using selective neutron transmutation doping (SNTD) in MARIA nuclear research reactor

Correction of the electric resistivity distribution of Si wafers using selective neutron transmutation doping (SNTD) in MARIA nuclear research reactor NUKLEONIKA 2012;57(3):363 367 ORIGINAL PAPER Correction of the electric resistivity distribution of Si wafers using selective neutron transmutation doping (SNTD) in MARIA nuclear research reactor Mikołaj

More information

A fast triple GEM detector for high-rate charged-particle triggering

A fast triple GEM detector for high-rate charged-particle triggering Nuclear Instruments and Methods in Physics Research A 478 (2002) 245 249 A fast triple GEM detector for high-rate charged-particle triggering G. Bencivenni a, W. Bonivento b,1, C. Bosio c, A. Cardini b,

More information

PHOTOELECTRON COLLECTION EFFICIENCY AT HIGH PRESSURE FOR A GAMMA DETECTOR ENVISAGING MEDICAL IMAGING

PHOTOELECTRON COLLECTION EFFICIENCY AT HIGH PRESSURE FOR A GAMMA DETECTOR ENVISAGING MEDICAL IMAGING 822 PHOTOELECTRON COLLECTION EFFICIENCY AT HIGH PRESSURE FOR A GAMMA DETECTOR ENVISAGING MEDICAL IMAGING C.D.R. Azevedo 1, C.A.B. Oliveira 1, J.M.F. dos Santos 2, J.F.C.A. Veloso 1 1.University of Aveiro,

More information

GaN for use in harsh radiation environments

GaN for use in harsh radiation environments 4 th RD50 - Workshop on radiation hard semiconductor devices for very high luminosity colliders GaN for use in harsh radiation environments a (W Cunningham a, J Grant a, M Rahman a, E Gaubas b, J Vaitkus

More information

Lessons learned from Bright Pixels and the Internal Background of the EPIC pn-ccd Camera

Lessons learned from Bright Pixels and the Internal Background of the EPIC pn-ccd Camera Lessons learned from Bright Pixels and the Internal Background of the EPIC pn-ccd Camera Elmar Pfeffermann, Norbert Meidinger, Lothar Strüder, Heinrich Bräuninger, Gisela Hartner Max-Planck-Institut für

More information

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification

More information

Mass Determination of Rn and Hg isotopes using MASHA

Mass Determination of Rn and Hg isotopes using MASHA Mass Determination of Rn and Hg isotopes using MASHA Alfred M. Sehone Lumkile Msebi Oleg Lishyk Stanislav Stanishevski Yuliya Brechko Supervisor Krupa Lubosh Flerov Laboratory of Nuclear reactions, JINR,

More information

& SUN ZhiJia 1;2. Abstract

& SUN ZhiJia 1;2. Abstract Study on the imaging ability of the 2D neutron detector based on MWPC TIAN LiChao 1;2;3*, CHEN YuanBo 1;2, TANG Bin 1;2,ZHOU JianRong 1;2,QI HuiRong 1;2, LIU RongGuang 1;2, ZHANG Jian 1;2, YANG GuiAn 1;2,

More information

A Complete Simulation of a Triple-GEM Detector

A Complete Simulation of a Triple-GEM Detector 1638 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 4, AUGUST 2002 A Complete Simulation of a Triple-GEM Detector W. Bonivento, A. Cardini, G. Bencivenni, F. Murtas, and D. Pinci Abstract Since some

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Measurement of the Ionizing Energy Depositions after Fast Neutron Interactions in Silicon

Measurement of the Ionizing Energy Depositions after Fast Neutron Interactions in Silicon Measurement of the Ionizing Energy Depositions after Fast Neutron Interactions in Silicon B. Bergmann a), I. Caicedo a), E. Fröjdh c), J. Kirstead b), S. Pospisil a), H. Takai b), D. Turecek a) a) Institute

More information

Physics sources of noise in ring imaging Cherenkov detectors

Physics sources of noise in ring imaging Cherenkov detectors Nuclear Instruments and Methods in Physics Research A 433 (1999) 235}239 Physics sources of noise in ring imaging Cherenkov detectors For the ALICE HMPID Group Andreas Morsch EP Division, CERN, CH-1211

More information

Solid State Device Fundamentals

Solid State Device Fundamentals Solid State Device Fundamentals ENS 345 Lecture Course by Alexander M. Zaitsev alexander.zaitsev@csi.cuny.edu Tel: 718 982 2812 Office 4N101b 1 Outline - Goals of the course. What is electronic device?

More information

Thin Calorimetry for Cosmic-Ray Studies Outside the Earth s Atmosphere. 1 Introduction

Thin Calorimetry for Cosmic-Ray Studies Outside the Earth s Atmosphere. 1 Introduction Thin Calorimetry for Cosmic-Ray Studies Outside the Earth s Atmosphere Richard WIGMANS Department of Physics, Texas Tech University, Lubbock TX 79409-1051, USA (wigmans@ttu.edu) Abstract Cosmic ray experiments

More information

Advantages / Disadvantages of semiconductor detectors

Advantages / Disadvantages of semiconductor detectors Advantages / Disadvantages of semiconductor detectors Semiconductor detectors have a high density (compared to gas detector) large energy loss in a short distance diffusion effect is smaller than in gas

More information

A Comparison between Channel Selections in Heavy Ion Reactions

A Comparison between Channel Selections in Heavy Ion Reactions Brazilian Journal of Physics, vol. 39, no. 1, March, 2009 55 A Comparison between Channel Selections in Heavy Ion Reactions S. Mohammadi Physics Department, Payame Noor University, Mashad 91735, IRAN (Received

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Simulating the Charge Dispersion Phenomena in Micro Pattern Gas Detectors with a Resistive Anode

Simulating the Charge Dispersion Phenomena in Micro Pattern Gas Detectors with a Resistive Anode Simulating the Charge Dispersion Phenomena in Micro Pattern Gas Detectors with a Resistive Anode M. S. Dixit a b and A. Rankin a a Department of Physics Carleton University 1125 Colonel By Drive Ottawa

More information

Determination of the shielding power of different materials against gamma radiation

Determination of the shielding power of different materials against gamma radiation Determination of the shielding power of different materials against gamma radiation Chow Wing Yan, Yeung Chun Lap S.K.H Tsang Shiu Tim Secondary School Wong Ka Wing Baptist Lui Ming Choi Secondary School

More information

Charge sharing in silicon pixel detectors

Charge sharing in silicon pixel detectors Nuclear Instruments and Methods in Physics Research A 487 (2002) 113 122 Charge sharing in silicon pixel detectors K. Mathieson a, *, M.S. Passmore b, P. Seller c, M.L. Prydderch c, V. O Shea a, R.L. Bates

More information

This paper should be understood as an extended version of a talk given at the

This paper should be understood as an extended version of a talk given at the This paper should be understood as an extended version of a talk given at the Abstract: 1 st JINA workshop at Gull Lake, 2002. Recent experimental developments at LANL (Los Alamos, NM, USA) and CERN (Geneva,

More information

GEM at CERN. Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM

GEM at CERN. Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM GEM at CERN Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM MicroStrip Gas Chamber Semiconductor industry technology: Photolithography Etching Coating Doping A. Oed Nucl. Instr. and Meth. A263 (1988) 351.

More information

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector C. Fiorini, member, IEEE, A. Gola, M. Zanchi, A. Longoni, P. Lechner, H. Soltau, L. Strüder Abstract In this work

More information

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland Available on CMS information server CMS CR -2018/225 The Compact Muon Solenoid Experiment Conference Report Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland 27 September 2018 (v2, 19 November

More information

Application Note. The Continuous Air Monitoring (CAM) PIPS Detector Properties and Applications

Application Note. The Continuous Air Monitoring (CAM) PIPS Detector Properties and Applications Application Note The Continuous Air Monitoring (CAM) PIPS Detector Properties and Applications Introduction The increasing demand for safety in nuclear installations calls for continuous survey of airborne

More information

arxiv: v2 [physics.ins-det] 8 Feb 2013

arxiv: v2 [physics.ins-det] 8 Feb 2013 Preprint typeset in JINST style - HYPER VERSION arxiv:1302.0278v2 [physics.ins-det] 8 Feb 2013 Investigation of gamma ray detection performance of thin LFS scintillator with MAPD readout E.Guliyev a, F.Ahmadov

More information

Impact of high photon densities on AGIPD requirements

Impact of high photon densities on AGIPD requirements Impact of high photon densities on AGIPD requirements Julian Becker University of Hamburg Detector Laboratory new data 1. Heating estimations 2. Confined breakdown 3. Range switching in adjacent pixels

More information

Radiation (Particle) Detection and Measurement

Radiation (Particle) Detection and Measurement Radiation (Particle) Detection and Measurement Radiation detection implies that the radiation interacts (e.g. leaves at least part of its energy) in the material. A specific material is chosen, because

More information

Bolometry. H. Kroegler Assciazione Euratom-ENEA sulla Fusione, Frascati (Italy)

Bolometry. H. Kroegler Assciazione Euratom-ENEA sulla Fusione, Frascati (Italy) Bolometry H. Kroegler Assciazione Euratom-ENEA sulla Fusione, Frascati (Italy) Revised May 28, 2002 1. Radiated power Time and space resolved measurements of the total plasma radiation can be done by means

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

Radiation Detector 2016/17 (SPA6309)

Radiation Detector 2016/17 (SPA6309) Radiation Detector 2016/17 (SPA6309) Semiconductor detectors (Leo, Chapter 10) 2017 Teppei Katori Semiconductor detectors are used in many situations, mostly for some kind of high precision measurement.

More information

Towards Proton Computed Tomography

Towards Proton Computed Tomography SCIPP Towards Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz, CA 95064 V.

More information

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P ) Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P2522015) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik Experiment:

More information

Alpha-Energies of different sources with Multi Channel Analyzer

Alpha-Energies of different sources with Multi Channel Analyzer Physical Structure of Matter Radioactivity Alpha-Energies of different sources with Multi Channel Analyzer What you can learn about Decay series Radioactive equilibrium Isotopic properties Decay energy

More information

M. De Napoli, F. Giacoppo, G. Raciti, E. Rapisarda, C. Sfienti. Laboratori Nazionali del Sud (LNS) INFN University of Catania. IPRD Oct.

M. De Napoli, F. Giacoppo, G. Raciti, E. Rapisarda, C. Sfienti. Laboratori Nazionali del Sud (LNS) INFN University of Catania. IPRD Oct. M. De Napoli, F. Giacoppo, G. Raciti, E. Rapisarda, C. Sfienti Laboratori Nazionali del Sud (LNS) INFN University of Catania IPRD08 1-4 Oct. Siena Silicon carbide (SiC) is expected to be applied to high-power

More information

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector A. H. D. Rasolonjatovo 1, T. Shiomi 1, T. Nakamura 1 Y. Tsudaka 2, H. Fujiwara 2, H. Araki 2, K. Matsuo 2, H. Nishizawa 2 1 Cyclotron

More information

Measurement of the transverse diffusion coefficient of charge in liquid xenon

Measurement of the transverse diffusion coefficient of charge in liquid xenon Measurement of the transverse diffusion coefficient of charge in liquid xenon W.-T. Chen a, H. Carduner b, J.-P. Cussonneau c, J. Donnard d, S. Duval e, A.-F. Mohamad-Hadi f, J. Lamblin g, O. Lemaire h,

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Lecture 8. Detectors for Ionizing Particles

Lecture 8. Detectors for Ionizing Particles Lecture 8 Detectors for Ionizing Particles Content Introduction Overview of detector systems Sources of radiation Radioactive decay Cosmic Radiation Accelerators Interaction of Radiation with Matter General

More information

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo ICEPP, the University of Tokyo E-mail: sawada@icepp.s.u-tokyo.ac.jp The MEG experiment yielded the most stringent upper limit on the branching ratio of the flavorviolating muon decay µ + e + γ. A major

More information

ORTEC. Review of the Physics of Semiconductor Detectors. Interaction of Ionizing Radiation with Semiconductor Detectors. Heavy Charged Particles

ORTEC. Review of the Physics of Semiconductor Detectors. Interaction of Ionizing Radiation with Semiconductor Detectors. Heavy Charged Particles ORTEC Review of the Physics of Historically, semiconductor detectors were conceived as solid-state ionization chambers. To obtain a high-electric-field, low-current, solid-state device for detection and

More information

Particle Energy Loss in Matter

Particle Energy Loss in Matter Particle Energy Loss in Matter Charged particles, except electrons, loose energy when passing through material via atomic excitation and ionization These are protons, pions, muons, The energy loss can

More information

Semiconductor-Detectors

Semiconductor-Detectors Semiconductor-Detectors 1 Motivation ~ 195: Discovery that pn-- junctions can be used to detect particles. Semiconductor detectors used for energy measurements ( Germanium) Since ~ 3 years: Semiconductor

More information

New irradiation zones at the CERN-PS

New irradiation zones at the CERN-PS Nuclear Instruments and Methods in Physics Research A 426 (1999) 72 77 New irradiation zones at the CERN-PS M. Glaser, L. Durieu, F. Lemeilleur *, M. Tavlet, C. Leroy, P. Roy ROSE/RD48 Collaboration CERN,

More information

Count. Count. 200mm. PMT scintillation. PMT flourescence. PMT fluorescence. PMT fluorescence. PMT fluorescence. 50mm. 42mm

Count. Count. 200mm. PMT scintillation. PMT flourescence. PMT fluorescence. PMT fluorescence. PMT fluorescence. 50mm. 42mm Proceedings of the Second International Workshop on EGS, 8.-12. August, Tsukuba, Japan KEK Proceedings 200-20, pp.316-323 Simulation of Ionization and Scintillation Signals in a Liquid Ionization Drift

More information

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2 Photons Produced in PbF 2 R. Dolenec a, S. Korpar b,a, P. Križan c,a, R. Pestotnik a, A. Stanovnik d,a a, Ljubljana, Slovenia b Faculty of Chemistry and Chemical Engineering, University of Maribor, Slovenia

More information

pp physics, RWTH, WS 2003/04, T.Hebbeker

pp physics, RWTH, WS 2003/04, T.Hebbeker 3. PP TH 03/04 Accelerators and Detectors 1 pp physics, RWTH, WS 2003/04, T.Hebbeker 2003-12-16 1.2.4. (Inner) tracking and vertexing As we will see, mainly three types of tracking detectors are used:

More information

3.1 Introduction to Semiconductors. Y. Baghzouz ECE Department UNLV

3.1 Introduction to Semiconductors. Y. Baghzouz ECE Department UNLV 3.1 Introduction to Semiconductors Y. Baghzouz ECE Department UNLV Introduction In this lecture, we will cover the basic aspects of semiconductor materials, and the physical mechanisms which are at the

More information

arxiv: v1 [physics.ins-det] 29 Jun 2011

arxiv: v1 [physics.ins-det] 29 Jun 2011 Investigation of Large LGB Detectors for Antineutrino Detection P. Nelson a,, N. S. Bowden b, a Department of Physics, Naval Postgraduate School, Monterey, CA 99, USA b Lawrence Livermore National Laboratory,

More information

Measurement of the n_tof beam profile in the second experimental area (EAR2) using a silicon detector

Measurement of the n_tof beam profile in the second experimental area (EAR2) using a silicon detector Measurement of the n_tof beam profile in the second experimental area (EAR) using a silicon detector Fidan Suljik Supervisors: Dr. Massimo Barbagallo & Dr. Federica Mingrone September 8, 7 Abstract A new

More information

Silicon Detectors. Particle Physics

Silicon Detectors. Particle Physics Mitglied der Helmholtz-Gemeinschaft Silicon Detectors for Particle Physics 9. August 2012 Ralf Schleichert, Institut für Kernphysik Outline Different Cameras Silicon Detectors Taking Pictures in Particle

More information

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors Lecture 2 Introduction to semiconductors Structures and characteristics in semiconductors Semiconductor p-n junction Metal Oxide Silicon structure Semiconductor contact Literature Glen F. Knoll, Radiation

More information

Impact of high photon densities on AGIPD requirements

Impact of high photon densities on AGIPD requirements Impact of high photon densities on AGIPD requirements Julian Becker University of Hamburg Detector Laboratory 1. Si-type influence on charge collection time 2. Measurements on charge collection time 3.

More information

Preliminary Results From The First Flight of Atic: The Silicon Matrix

Preliminary Results From The First Flight of Atic: The Silicon Matrix Preliminary Results From The First Flight of Atic: The Silicon Matrix J. H. Adams, Jr. (1) for the ATIC Collaboration (1) NASA Marshall Space Flight Center, Huntsville, AL 35812, USA james.h.adams@msfc.nasa.gov,

More information

Digital Hadron Calorimetry for the Linear Collider using GEM Technology

Digital Hadron Calorimetry for the Linear Collider using GEM Technology Digital Hadron Calorimetry for the Linear Collider using GEM Technology University of Texas at Arlington Andrew Brandt, Kaushik De, Shahnoor Habib, Venkat Kaushik, Jia Li, Mark Sosebee, Andy White* 1,

More information

Components of a generic collider detector

Components of a generic collider detector Lecture 24 Components of a generic collider detector electrons - ionization + bremsstrahlung photons - pair production in high Z material charged hadrons - ionization + shower of secondary interactions

More information

ATLAS New Small Wheel Phase I Upgrade: Detector and Electronics Performance Analysis

ATLAS New Small Wheel Phase I Upgrade: Detector and Electronics Performance Analysis ATLAS New Small Wheel Phase I Upgrade: Detector and Electronics Performance Analysis Dominique Trischuk, Alain Bellerive and George Iakovidis IPP CERN Summer Student Supervisor August 216 Abstract The

More information

A new detector for neutron beam monitoring

A new detector for neutron beam monitoring A new detector for neutron beam monitoring European Organization for Nuclear Research (CERN), Geneva, Switzerland in collaboration with Commissariat à l Energie Atomique (CEA), Saclay, France, Instituto

More information

Simulation for LHC Radiation Background

Simulation for LHC Radiation Background Simulation for LHC Radiation Background Optimisation of monitoring detectors and experimental validation M. Glaser1, S. Guatelli2, B. Mascialino2, M. Moll1, M.G. Pia2, F. Ravotti1 1 CERN, Geneva, Switzerland

More information

Geant4 simulations of the lead fluoride calorimeter

Geant4 simulations of the lead fluoride calorimeter Geant4 simulations of the lead fluoride calorimeter A.A. Savchenko a, *, A.A. Tishchenko a, S.B. Dabagov a,b, A. Anastasi b,c, G. Venanzoni b, M.N. Strikhanov a (et al.) a National Research Nuclear University

More information

Single Photon detectors

Single Photon detectors Single Photon detectors Outline Motivation for single photon detection Semiconductor; general knowledge and important background Photon detectors: internal and external photoeffect Properties of semiconductor

More information

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440 Jazan University College of Science Physics Department جاهعة جازان كلية العل وم قسن الفيزياء Lab Manual Nuclear Physics (2) 462 Phys 8 th Level Academic Year: 1439/1440 1 Contents No. Name of the Experiment

More information

Experimental Particle Physics

Experimental Particle Physics Experimental Particle Physics Particle Interactions and Detectors Lecture 2 2nd May 2014 Fergus Wilson, RAL 1/31 How do we detect particles? Particle Types Charged (e - /K - /π - ) Photons (γ) Electromagnetic

More information