1 Radiation Monitoring Systems & Waste Characterization Lecture 4 - Neutron Detectors M.Taiuti MASTER UNIVERSITARIO DI II LIVELLO IN SCIENZE E TECNOLOGIE DEGLI IMPIANTI NUCLEARI
2 Neutron Detectors What does it mean to detect a neutron? need to produce some sort of measurable quantitative i (countable) electrical l signal can t directly detect slow neutrons Need to use nuclear reactions to convert neutrons into charged particles Then we can use one of the many types of charged particle detectors gas proportional counters and ionization chambers scintillation detectors semiconductor detectors 2
3 Nuclear Reactions for Neutron Detectors n+ 3 He 3 H+ 1 H MeV n + 6 Li 4 He + 3 H MeV n + 10 B 7 Li* + 4 He 7 Li + 4 He MeV γ MeV (93%) 7 Li + 4 He +2.8 MeV (7%) n+cd Cd* γ-ray spectrum n Gd Gd* γ-ray spectrum n Gd Gd* γ-ray spectrum n U fission fragments + ~160 MeV n Pu fission fragments + ~160 MeV 197 Au(4.906 ev), 115 In( 1.46 ev), 181 Ta(4.28 ev), 238 U(6.67, ev); energy-selective detectors, narrow resonances, prompt capture gamma rays 3
4 Converter Materials 3 He is inevitably it in the gaseous state, t usually at several atmospheres pressure 6 Li is usually in chemically combined form, e.g., LiF, Li 2 O, Li 6 Gd(BO 3 ) 3, or in chemical compounds combined heterogeneously or homogeneously with intrinsic scintillators B is usually in chemically combined form, e.g., BF 3 (gas) or H 3 BO 3 in solution, sometimes as a thin coating in elemental form Gd as a converter is usually as the oxide or silicate U is usually thinly coated in oxide form, Ta in thin foil Lecture 4 - Neutron Detectors 4
5 Cross Sections n + 3 He 3 H + 1 H n + 6 Li n + 10 B 10 6 n Cd
6 Cross Sections n Gd n Gd n U 10 6 n Pu
7 An Exercise What is the wavelength of a non-relativistic i i neutron with kinetic energy T? Lecture 4 - Neutron Detectors 7
8 Semiconductor Detectors n + 6 Li 4 He+ 3 H MeV Lecture 4 - Neutron Detectors 8
9 Semiconductor Detectors cont d ~1,500, holes and electrons produced d per neutron (~ Coulomb) this can be detected directly without further amplification but... standard d device semiconductors do not contain enough neutron-absorbing nuclei to give reasonable neutron detection efficiency put neutron absorber on surface of semiconductor? develop boron phosphide p semiconductor devices? Lecture 4 - Neutron Detectors 9
10 Coating with Neutron Absorber Layer must be thin (a few microns) for charged particles il to reach hdetector detection efficiency is low Most of the deposited energy doesn t reach detector poor pulse height discrimination 10
11 Gas Detectors n + 3 He 3 H+ 1 H MeV ~25,000 ions and electrons produced per neutron (~ Coulomb) Lecture 4 - Neutron Detectors 11
12 Gas Detectors Ionization Mode Neutron electrons drift to anode, producing a charge pulse with no gas multiplication typically employed in low-efficiency i beam-monitor detectors t Proportional Mode Heavy particle (M 1 ) range * Ionization tracks Neutron capture event Light particle (M 2 ) range if voltage is high enough, electron collisions ionize gas atoms producing even more electrons gas amplification increases the collected charge proportional to the initial charge produced. gas gains of up to a few thousand are possible, above which proportionality is lost. 12
13 Fission Chamber Fission chambers use neutron-induced fission to detect neutrons. The chamber is usually similar in construction to that of an ionization chamber, except that the coating material is highly enriched 235 U. The neutrons interact with the 235 U, causing fission. One of the two fission fragments enters the chamber, while the other fission fragment embeds itself in the chamber wall. One advantage of using 235 U coating rather than boron is that the fission fragment has a much higher energy level than the α particle from a boron reaction. Neutron-induced d fission i fragments produce many more ionizations i i in the chamber per interaction than do the neutron- induced alpha particles. This allows the fission chambers to operate in higher gamma fields than an uncompensated ion chamber with boron lining. φ14mm 200 mm Cathode Fissile material Ionizing gas(ar + 5%N2) Electric insulator Anode Housing Double coaxial MI Electric insulator Cable φ6 mm Lecture 4 - Neutron Detectors 13
14 Gas Detectors Proportional counters (PCs) come in a variety of different forms. Simple detector (shown previously) Linear position-sensitivesensitive detector (LPSD): the anode is resistive, read out from both ends the charge distributes between the ends according to the position of the neutron capture event in the tube. usually cylindrical. 2-d position-sensitive detector (MWPC) many parallel resistive wires extend across a large thick area of fill gas. Each wire operates either as in LPSD or without position information as in a simple PC OR two mutually perpendicular arrays of anode wires. Each is read separately as an LPSD to give two coordinates for the neutron capture event MWPCs usually have a planar configuration 14
15 Reuter-Stokes LPSD 15
16 Pulse Height Discrimination Lecture 4 - Neutron Detectors 16
17 Pulse Height Discrimination cont d Can set discriminator levels to reject undesired events (fast neutrons, gammas, electronic noise) Pulse-height discrimination can make a large improvement in background Discrimination capabilities are an important criterion in the choice of detectors ( 3 He gas detectors are very good) Lecture 4 - Neutron Detectors 17
18 2-d Detector MAPS Detector Bank Lecture 4 - Neutron Detectors 18
19 Multi-Wire Proportional Counter Array of discrete detectors Remove walls to get multi-wire counter Lecture 4 - Neutron Detectors 19
20 MWPC cont d Segment the cathode to get x-y position ii Lecture 4 - Neutron Detectors 20
21 Resistive Encoding of a Multi-wire Detector Instead of reading every cathode strip individually, the strips can be resistively coupled (cheaper & slower) Position of the event can be determined from the fraction of the charge reaching each end of the resistive network (charge-division encoding) 21
22 Resistive Encoding of a Multi-wire Detector cont d Position i of the event can also be determined from the relative time of arrival of the pulse at the two ends of fthe resistive network (rise-time encoding) There is a pressurized gas mixture around the electrodes Lecture 4 - Neutron Detectors 22
23 Micro-Strip Gas Counter Drift Cathode Electrostatic Field Lines High Field (Avalanche) Region Cathode Strips Insulator Collector Anode Anode Strips Electrodes printed lithographically small features high spacial resolution, high field gradients charge localization and fast recovery
24 Sizes of Proportional Counters PCs and LPSDs come in many sizes. Diameters from ~ 5. mm to 50 mm. Fill gas pressures are highest h for small diameters, up to 40 atm, and lowest for large diameters 2.~ 3. atm Lengths vary from cm to meters; the longer detectors, up to about 3. m long, are typically those of larger diameter MWPCs are usually flat and square, but sometimes rectangular, even curved, or bananashaped Typical dimension 0.5 ~ 1.0 m Lecture 4 - Neutron Detectors 24
25 Detection Efficiency Detectors t rarely register all the incident id neutrons. The ratio of the number registered to the number incident is the efficiency Full expression: Approximate expression for low efficiency: Where: σ = absorption cross-sectionsection N = number density of absorber t = thickness N = cm -3 atm -1 for a gas For 1 cm thick 3 He at 1 atm and 1.8 Å neutrons: 25
26 Spatial Resolution Spatial resolution (how well the detector tells the location of an event) is always limited by the charged-particle il range and dby the range of neutrons in the fill gas, which depend on the pressure and composition of the fill gas And by the geometry: simple PCs: δ z ~ diameter; 6 mm - 50 mm LPSDs: δ z ~ diameter, δ y ~ diameter ; 6 mm - 50 mm MWPC: δ z and δ y ~ wire spacing; 1 mm - 10 mm Lecture 4 - Neutron Detectors 26
27 Scintillation Detectors Light output t from liquid id organic scintillators t shows different pulse shape for γ and fast neutrons: γ interacts by photoelectric, Compton and pair production neutron interacts by scattering on proton, transferring part of the kinetic energy Lecture 4 - Neutron Detectors 27
28 Scintillation Detectors n + 6 Li 4 He+ 3 H MeV 28
29 Some Common Scintillators for Neutron Detectors ZnS(Ag) is the brightest t scintillator t known, an intrinsic i i scintillator that is mixed heterogeneously with converter material, usually Li 6 F in the Stedman recipe, to form scintillating composites. These are only semitransparent. But it is somewhat slow, decaying with ~10 μsec halftime GS-20 (glass,ce 3+ ) is mixed with a high concentration of Li 2 Oin the melt to form a material transparent to light Li 6 Gd(BO 3 ) 3 (Ce 3+ ) (including 158 Gd and 160 Gd, 6 Li,and 11 B), and 6 LiF(Eu) are intrinsic i i scintillators t that t contain high h proportions of converter material and are typically transparent An efficient gamma ray detector with little sensitivity to neutrons, used in conjunction with neutron capture gamma-ray converters, is YAP (yttrium aluminum perovskite, YAl 2 O 3 (Ce 3+ )) Lecture 4 - Neutron Detectors 29
30 Some Common Scintillators for Neutron Detectors Material Density of 6 Li atoms (cm -3) Scintillation efficiency Photon wavelength (nm) Photons per neutron Li glass (Ce) % 395 ~7,000 LiI (Eu) % 470 ~51,000 ZnS (Ag) - LiF % ~160, Li 6 Gd(BO 3 ) 3 (Ce), 3.3x10 22 ~40,000 ~ 400 YAP NA 350 ~18,000 per MeV γ Lecture 4 - Neutron Detectors 30
31 Anger Camera Detector Photomultiplier outputs are resistively encoded to give x and y coordinates Entire assembly is in a light-tight box 31
32 Anger Camera Detector Prototype t scintillator-based t area-position-sensitive neutron detector Designed to allow easy expansion into a 7x7 photomultiplier array with a 15x15 cm 2 active scintillator area. Resolution is expected to be ~1.5x mm 2 Lecture 4 - Neutron Detectors 32
33 Crossed-Fiber Position-Sensitive Scintillation Detector Outputs to multi-anode photomultiplier tube 1-mm Square Wavelength-shifting fibers Outputs to coincidence single-anode photomultiplier tube Scintillator screen
34 Crossed-Fiber Position-Sensitive Scintillation Detector The scintillator t screen for this 2-d ddetector t consists of a mixture of 6 LiF and silver-activated ZnS powder in an epoxy binder Neutrons incident on the screen react with the 6 Li to produce a t and an α particle. Collisions with these charged particles cause the ZnS(Ag) to scintillate at a wavelength of approximately 450 nm. The photons are absorbed in the wavelength-shifting fibers where they converted to 520 nm photons emitted in modes that propagate out the ends of the fibers The optimum mass ratio of 6 LiF:ZnS(Ag) was determined to be 13 1:3. A mixture containing i 40 mg/cm 2 of 6 LiF and 120 mg/cm 2 of ZnS(Ag) is used in this screen design. This mixture has a measured neutron conversion efficiency of over 90% Lecture 4 - Neutron Detectors 34
35 Crossed-Fiber Position-Sensitive Scintillation Detector Design Parameters (ORNL I&C) size: 25 cm x 25 cm thickness: 2 mm number of fibers: 48 for each axis multi-anode photomultiplier li tube: Phillips XP1704 coincidence tube: Hamamatsu 1924 resolution: < 5 mm shaping time: 300 nsec count rate capability: ~ 1 MHz time-of-flight Resolution: 1 μsec 35
36 Crossed-Fiber Position-Sensitive Scintillation Detector Lecture 4 - Neutron Detectors 36
EEE4106Z Radiation Interactions & Detection
EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation
More informationNuclear Physics and Astrophysics
Nuclear Physics and Astrophysics PHY-30 Dr. E. Rizvi Lecture 4 - Detectors Binding Energy Nuclear mass MN less than sum of nucleon masses Shows nucleus is a bound (lower energy) state for this configuration
More informationChapter 11: Neutrons detectors
Chapter 11: Neutrons detectors 1 Contents Principles of neutrons detection Slow neutron detection methods Fast neutron detection methods 2 Introduction Neutrons are uncharged particles cannot be directly
More informationIntroduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose
Introduction to Radiological Sciences Neutron Detectors Neutron counting Theory of operation Slow neutrons Fast neutrons Types of detectors Source calibration Survey for Dose 2 Neutrons, what are they?
More informationPHYS 3446 Lecture #12
PHYS 3446 Lecture #12 Wednesday, Oct. 18, 2006 Dr. 1. Particle Detection Ionization Detectors MWPC Scintillation Counters Time of Flight 1 Announcements Next LPCC Workshop Preparation work Each group to
More informationDetekce a spektrometrie neutronů. neutron detection and spectroscopy
Detekce a spektrometrie neutronů neutron detection and spectroscopy 1. Slow neutrons 2. Fast neutrons 1 1. Slow neutrons neutron kinetic energy E a) charged particles are produced, protons, α particle,
More informationPhoton Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011
Photon Instrumentation First Mexican Particle Accelerator School Guanajuato Oct 6, 2011 Outline The Electromagnetic Spectrum Photon Detection Interaction of Photons with Matter Photoelectric Effect Compton
More informationDETECTORS. I. Charged Particle Detectors
DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State
More informationGLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY
GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad
More informationParticle Energy Loss in Matter
Particle Energy Loss in Matter Charged particles, except electrons, loose energy when passing through material via atomic excitation and ionization These are protons, pions, muons, The energy loss can
More informationNeutron Detection. n interactions with matter n detection High/low energy n detectors
Neutron Detection Example of n detection: Well logging Reservoir/Formation Evaluation Brief introduction to neutron generation Continuous sources Large accelerators Pulsed neutron generators n interactions
More informationParticle Energy Loss in Matter
Particle Energy Loss in Matter Charged particles loose energy when passing through material via atomic excitation and ionization These are protons, pions, muons, The energy loss can be described for moderately
More informationContents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung
Contents Marcel MiGLiERiNi Nuclear Medicine, Radiology and Their Metrological Aspects. Radiation in Medicine. Dosimetry 4. Diagnostics & Therapy 5. Accelerators in Medicine 6. Therapy Planning 7. Nuclear
More informationPHY492: Nuclear & Particle Physics. Lecture 25. Particle Detectors
PHY492: Nuclear & Particle Physics Lecture 25 Particle Detectors http://pdg.lbl.gov/2006/reviews/contents_sports.html S(T ) = dt dx nz = ρa 0 Units for energy loss Minimum ionization in thin solids Z/A
More informationChapter 4 Scintillation Detectors
Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,
More informationRadioactivity. Lecture 6 Detectors and Instrumentation
Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or
More informationApplied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra
22.101 Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra References: W. E. Meyerhof, Elements of Nuclear Physics (McGraw-Hill, New York, 1967),
More informationGas-filled Detectors
Gas-filled Detectors Radiation Gas-filled Detectors In a gas-filled detector, the io9nization provides electrons and positive ions. The acceleration of these charged particles obeys the simple equation
More informationOak Ridge National Laboratory* P. 0.Box 2008 Oak Ridge, TN (615) Presentation at. Les Houches, France.
NEUTRON SCINTILLATORS USING WAVELENGTH SHIFTING FIBERS D. P. Hutchinson, V. C. Miller and J. A. Ramsey Oak Ridge National Laboratory* P. 0.Box 2008 Oak Ridge, TN 37831-6006 (615)574-4730 Presentation at
More informationV. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber
CYRIC Annual Report 2001 V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber Kawata N. Baba M. Kato M.*, Miura T.**, and Yamadera A.***, Cyclotron and Radioisotope Center, Tohoku
More informationRadiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons
Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na Ellen Simmons 1 Contents Introduction Review of the Types of Radiation Charged Particle Radiation Detection Review of Semiconductor
More informationPerformance of high pressure Xe/TMA in GEMs for neutron and X-ray detection
Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection R. Kreuger, C. W. E. van Eijk, Member, IEEE, F. A. F. Fraga, M. M. Fraga, S. T. G. Fetal, R. W. Hollander, Member, IEEE, L. M.
More informationIntroduction to neutron rem meters
LA-UR-15-28285 Introduction to neutron rem meters Tom McLean, LANL CSU neutron class Fort Collins, CO Oct. 27-29 2015 Introduction: talk outline Discussion (brief) of neutron remmeters Gas proportional
More informationRadiation Detection. 15 th Annual OSC Readiness Training Program.
Radiation Detection 15 th Annual OSC Readiness Training Program www.oscreadiness.org GM Detectors 15 th Annual OSC Readiness Training Program www.oscreadiness.org 1 A closer look 15 th Annual OSC Readiness
More informationhν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?
Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important
More informationLast Lecture 1) Silicon tracking detectors 2) Reconstructing track momenta
Last Lecture 1) Silicon tracking detectors 2) Reconstructing track momenta Today s Lecture: 1) Electromagnetic and hadronic showers 2) Calorimeter design Absorber Incident particle Detector Reconstructing
More informationRadionuclide Imaging MII Detection of Nuclear Emission
Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor
More informationMEASUREMENT AND DETECTION OF RADIATION
MEASUREMENT AND DETECTION OF RADIATION Second Edition Nicholas Tsoulfanidis University of Missouri-Rolla Ж Taylor &Francis * Publishers since I79H CONTENTS Preface to the First Edition Preface to the Second
More informationAppendix A2. Particle Accelerators and Detectors The Large Hadron Collider (LHC) in Geneva, Switzerland on the Border of France.
Appendix A. Particle Accelerators and Detectors The Large Hadron Collider (LHC) in Geneva, Switzerland on the Border of France. Prepared by: Arash Akbari-Sharbaf Why Build Accelerators? Probe deeper From
More informationRadioactivity and Ionizing Radiation
Radioactivity and Ionizing Radiation QuarkNet summer workshop June 24-28, 2013 1 Recent History Most natural phenomena can be explained by a small number of simple rules. You can determine what these rules
More informationWeek 10: Chap. 14 Slow Neutron Detection
Week 10: Chap. 14 Slow Neutron Detection (Gamma ray) Backgrounds Slow Neutron Detection -- nuclear reactions --- spectra properties -- Proportional Counters --- fill gas --- lined detectors Fast neutron
More informationSCINTILLATION DETECTORS AND PM TUBES
SCINTILLATION DETECTORS AND PM TUBES General Characteristics Introduction Luminescence Light emission without heat generation Scintillation Luminescence by radiation Scintillation detector Radiation detector
More informationThe interaction of radiation with matter
Basic Detection Techniques 2009-2010 http://www.astro.rug.nl/~peletier/detectiontechniques.html Detection of energetic particles and gamma rays The interaction of radiation with matter Peter Dendooven
More informationExperimental Particle Physics
Experimental Particle Physics Particle Interactions and Detectors Lecture 2 2nd May 2014 Fergus Wilson, RAL 1/31 How do we detect particles? Particle Types Charged (e - /K - /π - ) Photons (γ) Electromagnetic
More informationGeneral Overview of Gas Filled Detectors
GAS-FILLED DETECTOR General Overview of Gas Filled Detectors Gas-Filled Detectors Ion chamber Proportional counter G-M (Geiger-Miller) counter Diagram of a Generic Gas-Filled Detector A Anode High-voltage
More informationChemical Engineering 412
Chemical Engineering 412 Introductory Nuclear Engineering Lecture 26 Radiation Detection & Measurement II Spiritual Thought 2 I would not hold the position in the Church I hold today had I not followed
More informationExperimental Particle Physics
Experimental Particle Physics Particle Interactions and Detectors Lecture 2 17th February 2010 Fergus Wilson, RAL 1/31 How do we detect particles? Particle Types Charged (e - /K - /π - ) Photons (γ) Electromagnetic
More informationA new detector for neutron beam monitoring
A new detector for neutron beam monitoring European Organization for Nuclear Research (CERN), Geneva, Switzerland in collaboration with Commissariat à l Energie Atomique (CEA), Saclay, France, Instituto
More informationChapter Four (Interaction of Radiation with Matter)
Al-Mustansiriyah University College of Science Physics Department Fourth Grade Nuclear Physics Dr. Ali A. Ridha Chapter Four (Interaction of Radiation with Matter) Different types of radiation interact
More informationEnergetic particles and their detection in situ (particle detectors) Part II. George Gloeckler
Energetic particles and their detection in situ (particle detectors) Part II George Gloeckler University of Michigan, Ann Arbor, MI University of Maryland, College Park, MD Simple particle detectors Gas-filled
More informationExperimental Particle Physics
Experimental Particle Physics Particle Interactions and Detectors 20th February 2007 Fergus Wilson, RAL 1 How do we detect Particles? Particle Types Charged (e - /K - /π - ) Photons (γ) Electromagnetic
More informationRadiation (Particle) Detection and Measurement
Radiation (Particle) Detection and Measurement Radiation detection implies that the radiation interacts (e.g. leaves at least part of its energy) in the material. A specific material is chosen, because
More informationThe outline. 1) Detector parameters: efficiency, geometrical acceptance, dead-time, resolution, linearity. 2) gaseous ionization chambers
The outline 1) Detector parameters: efficiency, geometrical acceptance, dead-time, resolution, linearity 2) gaseous ionization chambers 3) proportional counters- ionization measurement 4) silicon detectors
More informationLarge area scintillation detector for dosimetric stand with improved light collection
NUKLEONIKA 2011;56(4):317 321 ORIGINAL PAPER Large area scintillation detector for dosimetric stand with improved light collection Bronisław Machaj, Jan Mirowicz, Ewa Kowalska Abstract. In order to improve
More informationCopyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy
Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must
More informationIonization Detectors. Mostly Gaseous Detectors
Ionization Detectors Mostly Gaseous Detectors Introduction Ionization detectors were the first electrical devices developed for radiation detection During the first half of the century: 3 basic types of
More informationGamma and X-Ray Detection
Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular
More informationParticle Detectors Tools of High Energy and Nuclear Physics Detection of Individual Elementary Particles
Particle Detectors Tools of High Energy and Nuclear Physics Detection of Individual Elementary Particles Howard Fenker Jefferson Lab May 31, 2006 Outline of Talk Interactions of Particles with Matter Atomic
More informationPHYS 3650L - Modern Physics Laboratory
PHYS 3650L - Modern Physics Laboratory Laboratory Advanced Sheet Photon Attenuation 1. Objectives. The objectives of this laboratory exercise are: a. To measure the mass attenuation coefficient at a gamma
More informationThe Ones That Got Away: Neutron Leakage Spectra
The Ones That Got Away: Neutron Leakage Spectra Some neutrons are able to exit the surface before being captured The energies of these neutrons depend on their scattering history Fast neutrons that scatter
More informationENEN/Reactor Theory/ Laboratory Session 1 DETERMINATION OF BASIC STATIC REACTOR PARAMETERS IN THE GRAPHITE PILE AT THE VENUS FACILITY
p1 Summary DETERMINATION OF BASIC STATIC REACTOR PARAMETERS IN THE GRAPHITE PILE AT THE VENUS FACILITY P. Baeten (pbaeten@sccen.be) The purpose of this laboratory session is the determination of the basic
More informationPARTICLES REVELATION THROUGH SCINTILLATION COUNTER
14-25 JUNE 2004 SUMMER STAGE PARTICLES REVELATION THROUGH SCINTILLATION COUNTER by Flavio Cavalli and Marcello De Vitis Liceo Scientifico Statale Farnesina Tutor: Marco Mirazita 1) COSMIC RAYS - The Muons
More informationPoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo
ICEPP, the University of Tokyo E-mail: sawada@icepp.s.u-tokyo.ac.jp The MEG experiment yielded the most stringent upper limit on the branching ratio of the flavorviolating muon decay µ + e + γ. A major
More informationIntroduction. Principle of Operation
Introduction Ionizing radiation that is associated with radioactivity cannot be directly detected by our senses. Ionization is the process whereby the radiation has sufficient energy to strip electrons
More informationA novel design of the MeV gamma-ray imaging detector with Micro-TPC
Elsevier Science 1 Journal logo A novel design of the MeV gamma-ray imaging detector with Micro-TPC R.Orito *,H.Kubo,K.Miuchi,T.Nagayoshi,A.Takada,T.Tanimori,M.Ueno Department of Physics,Graduate School
More informationChem 481 Lecture Material 3/20/09
Chem 481 Lecture Material 3/20/09 Radiation Detection and Measurement Semiconductor Detectors The electrons in a sample of silicon are each bound to specific silicon atoms (occupy the valence band). If
More informationDESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE
DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE Hiroo Sato 1 and Yoichi Sakuma 2 1 International University of Health and Welfare, Kitakanemaru 2600-1, Ohtawara 324-8501
More informationChemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy
Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental
More informationChapter Seven (Nuclear Detectors)
Al-Mustansiriyah University College of Science Physics Department Fourth Grade Nuclear Physics Dr. Ali A. Ridha Chapter Seven (Nuclear Detectors) Ionizing radiation is rarely detected directly. Instead,
More informationNuclear and Particle Physics 4b Physics of the Quark Gluon Plasma
Nuclear and Particle Physics 4b Physics of the Quark Gluon Plasma Goethe University Frankfurt GSI Helmholtzzentrum für Schwerionenforschung Lectures and Exercise Summer Semester 2016 1 Organization Language:
More informationRADIATION DETECTION AND MEASUREMENT
RADIATION DETECTION AND MEASUREMENT SECOND EDITION GLENN F. KNOLL Professor of Nuclear Engineering The University of Michigan Ann Arbor, Michigan WILEY JOHN WILEY & SONS New York Chichester Brisbane Toronto
More information08 - Miscellaneous and historical detectors
08 - Miscellaneous and historical detectors Jaroslav Adam Czech Technical University in Prague Version 2 Jaroslav Adam (CTU, Prague) DPD_08, Miscellaneous and historical detectors Version 2 1 / 25 Streamer
More informationMCRT L8: Neutron Transport
MCRT L8: Neutron Transport Recap fission, absorption, scattering, cross sections Fission products and secondary neutrons Slow and fast neutrons Energy spectrum of fission neutrons Nuclear reactor safety
More informationLecture # 3. Muhammad Irfan Asghar National Centre for Physics. First School on LHC physics
Lecture # 3 Muhammad Irfan Asghar National Centre for Physics Introduction Gaseous detectors Greater mobility of electrons Obvious medium Charged particles detection Particle information easily transformed
More informationAbsorption and Backscattering ofβrays
Experiment #54 Absorption and Backscattering ofβrays References 1. B. Brown, Experimental Nucleonics 2. I. Kaplan, Nuclear Physics 3. E. Segre, Experimental Nuclear Physics 4. R.D. Evans, The Atomic Nucleus
More informationDevelopment of a new MeV gamma-ray camera
Development of a new MeV gamma-ray camera ICEPP Symposium February 16, 2004 Hakuba, Nagano, Japan Kyoto University Atsushi Takeda H. Kubo, K. Miuchi, T. Nagayoshi, Y. Okada, R. Orito, A. Takada, T. Tanimori,
More informationAngular Distribution of Neutrons from. by Frank Genevese Physical Review Vol.76, # 9 (Nov 1, 1949) A paper on experiment Paper Club, 22, Feb, 2011
Angular Distribution of Neutrons from the Photo-Disintegration of Deuteron by Frank Genevese Physical Review Vol.76, # 9 (Nov 1, 1949) A paper on experiment Paper Club, 22, Feb, 2011 The Outline of the
More informationR&D on Astroparticles Detectors. (Activity on CSN )
R&D on Astroparticles Detectors (Activity on CSN5 2000-2003) Introduction Results obtained with the R&D activity (2000-2003) with some drift chambers prototypes are reported. With different photocathode
More informationAnalysis of γ spectrum
IFM The Department of Physics, Chemistry and Biology LAB 26 Analysis of γ spectrum NAME PERSONAL NUMBER DATE APPROVED I. OBJECTIVES - To understand features of gamma spectrum and recall basic knowledge
More informationCHARGED PARTICLE INTERACTIONS
CHARGED PARTICLE INTERACTIONS Background Charged Particles Heavy charged particles Charged particles with Mass > m e α, proton, deuteron, heavy ion (e.g., C +, Fe + ), fission fragment, muon, etc. α is
More informationPerformance of the Gamma-ray Imaging Detector with Micro-TPC
Performance of the Gamma-ray Imaging Detector with Micro-TPC Cosmic-Ray Group Dept. of Physics, Graduate School of Science, Kyoto University R.Orito Orito,H.Kubo,K.,H.Kubo,K.Miuchi,T.Nagayoshi, A.Takada,A.Takeda,T.Tanimori
More informationelectrons out of, or ionize, material in their paths as they pass. Such radiation is known as
Detecting radiation It is always possible to detect charged particles moving through matter because they rip electrons out of, or ionize, material in their paths as they pass. Such radiation is known as
More informationRDCH 702 Lecture 8: Accelerators and Isotope Production
RDCH 702 Lecture 8: Accelerators and Isotope Production Particle generation Accelerator Direct Voltage Linear Cyclotrons Synchrotrons Photons * XAFS * Photonuclear Heavy Ions Neutrons sources Fission products
More informationAbsorption and Backscattering of β-rays
Experiment #54 Absorption and Backscattering of β-rays References 1. B. Brown, Experimental Nucleonics 2. I. Kaplan, Nuclear Physics 3. E. Segre, Experimental Nuclear Physics 4. R.D. Evans, The Atomic
More informationSCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION
SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system
More informationRadiation Detection and Measurement
Radiation Detection and Measurement Third Edition Glenn F. Knoll Professor of Nuclear Engineering and Radiological Sciences University of Michigan Ann Arbor, Michigan John Wiley & Sons, Inc. New York/Chichester/Weinheim/Brisbane/Toronto/Singapore
More informationToday, I will present the first of two lectures on neutron interactions.
Today, I will present the first of two lectures on neutron interactions. I first need to acknowledge that these two lectures were based on lectures presented previously in Med Phys I by Dr Howell. 1 Before
More informationExperiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado
Experiment 6 1 Introduction The Compton Effect Physics 2150 Experiment No. 6 University of Colorado In some situations, electromagnetic waves can act like particles, carrying energy and momentum, which
More informationTotal probability for reaction Yield
Total probability for reaction Yield If target has thickness d, and target material has # nuclei/volume: n 0 [part./cm 3 ] Y=σ n 0 d The yield gives the intensity of the characteristic signal from the
More informationNeutron Interactions Part I. Rebecca M. Howell, Ph.D. Radiation Physics Y2.5321
Neutron Interactions Part I Rebecca M. Howell, Ph.D. Radiation Physics rhowell@mdanderson.org Y2.5321 Why do we as Medical Physicists care about neutrons? Neutrons in Radiation Therapy Neutron Therapy
More informationPHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. II - Particle Detectors - Tsipenyuk Yu.M.
PARTICLE DETECTORS Tsipenyuk Yu.M. Russian Academy of Sciences, Moscow, Russia Keywords: gaseous counters, ionization chamber, proportional counter, Geiger Müller counter, semiconductor detector, neutron
More informationEmphasis on what happens to emitted particle (if no nuclear reaction and MEDIUM (i.e., atomic effects)
LECTURE 5: INTERACTION OF RADIATION WITH MATTER All radiation is detected through its interaction with matter! INTRODUCTION: What happens when radiation passes through matter? Emphasis on what happens
More informationRange of Alpha Particles in Air
Range of Alpha Particles in Air Oliver Kelly December 6, 2000 Abstract A fixed scintillation counter was used to detect and hence allow measurement of the range in air at S.T.P. of 4.88MeV α particles
More informationABSORPTION OF BETA AND GAMMA RADIATION
ABSORPTION OF BETA AND GAMMA RADIATION The purpose of this experiment is to understand the interaction of radiation and matter, and the application to radiation detection and shielding Apparatus: 137 Cs
More informationCompton Camera. Compton Camera
Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract
More informationProportional Counters
Proportional Counters 3 1 Introduction 3 2 Before we can look at individual radiation processes, we need to understand how the radiation is detected: Non-imaging detectors Detectors capable of detecting
More informationBeam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.
Beam Loss Monitors When energetic beam particles penetrates matter, secondary particles are emitted: this can be e, γ, protons, neutrons, excited nuclei, fragmented nuclei... Spontaneous radiation and
More informationDetector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples
Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,
More information05 - Scintillation detectors
05 - Scintillation detectors Jaroslav Adam Czech Technical University in Prague Version 2 Jaroslav Adam (CTU, Prague) DPD_05, Scintillation detectors Version 2 1 / 39 Scintillation detector principles
More informationEE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors
EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification
More informationTechnical University of Denmark
Technical University of Denmark Page 1 of 10 pages Written test, 12 December 2012 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: None. Pocket calculator not allowed "Weighting":
More informationGEM: A new concept for electron amplification in gas detectors
GEM: A new concept for electron amplification in gas detectors F. Sauli, Nucl. Instr. & Methods in Physics Research A 386 (1997) 531-534 Contents 1. Introduction 2. Two-step amplification: MWPC combined
More informationATOMIC WORLD P.1. ejected photoelectrons. current amplifier. photomultiplier tube (PMT)
ATOMIC WORLD P. HKAL PAPER I 0 8 The metal Caesium has a work function of.08 ev. Given: Planck constant h = 6.63 0 34 J s, charge of an electron e =.60 0 9 C (a) (i) Calculate the longest wavelength of
More informationPHY492: Nuclear & Particle Physics. Lecture 24. Exam 2 Particle Detectors
PHY492: Nuclear & Particle Physics Lecture 24 Exam 2 Particle Detectors Exam 2 April 16, 2007 Carl Bromberg - Prof. of Physics 2 Exam 2 2. Short Answer [4 pts each] a) To describe the QCD color quantum
More informationRadiation Detection and Measurement
Radiation Detection and Measurement Fourth Edition Glenn F. Knoll Professor Emeritus of Nuclear Engineering and Radiological Sciences University of Michigan Ann Arbor, Michigan WILEY John Wiley & Sons,
More informationChapter VIII: Nuclear fission
Chapter VIII: Nuclear fission 1 Summary 1. General remarks 2. Spontaneous and induced fissions 3. Nucleus deformation 4. Mass distribution of fragments 5. Number of emitted electrons 6. Radioactive decay
More informationPhotoelectron spectroscopy Instrumentation. Nanomaterials characterization 2
Photoelectron spectroscopy Instrumentation Nanomaterials characterization 2 RNDr. Věra V Vodičkov ková,, PhD. Photoelectron Spectroscopy general scheme Impact of X-ray emitted from source to the sample
More informationCherenkov Detector. Cosmic Rays Cherenkov Detector. Lodovico Lappetito. CherenkovDetector_ENG - 28/04/2016 Pag. 1
Cherenkov Detector Cosmic Rays Cherenkov Detector Lodovico Lappetito CherenkovDetector_ENG - 28/04/2016 Pag. 1 Table of Contents Introduction on Cherenkov Effect... 4 Super - Kamiokande... 6 Construction
More informationCosmic Rays Detector. Use of Coincidence Detector for Measures of Cosmic Rays. Lodovico Lappetito. RivelatoreRaggiCosmici_ENG - 6/22/2015 Page 1
Cosmic Rays Detector Use of Coincidence Detector for Measures of Cosmic Rays Lodovico Lappetito RivelatoreRaggiCosmici_ENG - 6/22/2015 Page 1 Table of Contents Design and Components... 3 Detector Design...
More informationSample Examination Questions
Sample Examination Questions Contents NB. Material covered by the AS papers may also appear in A2 papers. Question Question type Question focus number (section A or B) 1 A Ideal transformer 2 A Induced
More information