Chapter 2. Time-Resolved Raman Spectroscopy. Technical aspects and instrumentation

Size: px
Start display at page:

Download "Chapter 2. Time-Resolved Raman Spectroscopy. Technical aspects and instrumentation"

Transcription

1 Chapter 2. Time-Resolved Raman Spectroscopy Technical aspects and instrumentation 20

2 2.1 Photon dispersion in scattering media When measuring from a small focal volume in a transparent sample or a surface the Raman signals arrive at the detector collectively and practically instantaneously; specifically, in the time it takes photons to travel through air from the sample to the detector. These photons will arrive within a very narrow distribution in time as they will have traveled a practically identical pathlength. When studying scattering samples, the Raman signal is considerably broadened in time in its arrival at the detector however, due to the random, multiple scattering events within the translucent medium creating a greater variation in pathlength. 1,2 In the case of the scattering sample, a pulsed excitation source that is well characterized in duration and energy is required to distinguish from where within the sample the Raman signal is created. Additionally, when measuring this broader distribution of photons emanating from within scattering samples, a fast detector is also required capable of gated measurements that temporally overlap well with the broadened Raman profiles. A detector gate that is too narrow in time will be too limiting in the number of photons collected and intensity will be significantly reduced, while a gate that is too broad will reduce temporal/depth resolution within a sample Ti:sapphire pulsed laser excitation A critical aspect of the TRRS system is the use of a pulsed laser. The laser pulse length affects the temporal and spectral resolution of measurements; A laser pulse that is too long (~ns) will not provide sufficient depth discrimination in a scattering sample, while a laser pulse that is too short (~fs) will not have the well-defined photon energy needed for Raman measurements. A sufficiently high repetition rate coupled with a low peak power is also required to obtain enough measureable signal, while minimizing photodegradation. The TRRS system used in this thesis work employs a Ti:sapphire laser (Coherent, Mira 900P) pumped by a CW 532 nm Verdi laser. The Ti:sapphire produces 3 ps laser pulses at a repetition rate of 76 MHz, or a pulse every 13.2 ns. The high laser pulse repetition rate allows for sufficient signal accumulated over a second, despite the low peak power of each individual pulse. The spectral broadening associated with excitation via such short laser pulses can be approximated by the Heisenberg uncertainty principle, which describes the relationship between the uncertainty in time and energy. If the 3 ps laser pulse length is considered as full width at half maximum (FWHM) to be 2σ, and where h is Planck s constant: 21

3 σ (t) σ (E)= h/4π; FWHM(t)/2 FWHM(E)/2 = h/4π; ( s)/2 FWHM(E)/2=( m 2 kg/s) / 4(3.14); FWHM(E)/2= J, or 1.76 cm -1 ; FWHM(E) = 3.5 cm -1 This expected value corresponds well with practical measurements on this system by Efremov et al., 2007, 4 in which the measured width of an Argon calibration lamp line (2.4 cm -1 ) was compared to the laser line width (5.1 cm -1 ) at 405 nm observed with the same spectrograph settings. The extra contribution to the bandwidth from the laser can be approximated as a propagation of errors; (5.1) 2 (2.4) 2 + (FWHM laser) 2 ; FWHM laser 4.5 cm -1 An additional advantage of this laser system is that its fundamental wavelength is tunable over a broad range of red to NIR wavelengths: ca. 690 nm 980 nm. Through the use of a frequency doubler/tripler (TP-2000B, u-oplaz) the fundamental wavelength can be doubled or tripled in frequency to create additional blue and UV ranges of excitation, specifically including nm and nm. This flexibility in excitation wavelength is advantageous for studying a wide variety of samples under resonant or near resonant conditions. For measurements described in this and subsequent thesis chapters, various excitation wavelengths were employed including fundamental 720 nm, and frequency doubled 412 nm and 460 nm. Different filters and other optics also are needed for different wavelength ranges. We employed long-pass filters from Semrock: type 735 AELP, 420 AELP and 470 AELP, respectively. The exact cut-off wavelength can be optimized by tilting the edge filter. The Raman signal is backscattered to a triple-grating, 50-cm spectrograph (Acton) which has two CCD cameras mounted to different exit ports; a CCD camera (Andor, DV420-OE) for CW Raman measurements, and an ICCD (La Vision, Picostar) for gated detection. The detector spectral sensitivity plays an important role in determining which excitation wavelength to implement for a particular measurement, as the quantum efficiency of the ICCD camera is greatest for UV and visible wavelengths, and drops off dramatically at wavelengths beyond ~800 nm. The detector performance also plays a crucial role in TRRS through the creation of a detector gate, and will therefore be discussed in detail in the next section. 22

4 2.3 Gated detection via an intensified charge-coupled device (ICCD) camera For TRRS measurements via an intensified charge coupled device (ICCD) camera, the detector gate size is regulated by voltage parameters within the intensifier of the ICCD. Specifically, two potentials, average voltage (V av ) and clamping voltage (V cl ), are applied between the photocathode screen and the multichannel plate (MCP), as shown in Figure 2.1. A positive average voltage keeps the camera closed in between gate pulses, while the clamping voltage relative to this determines the gate size. 4,5 Optimum voltage settings of the intensifier must be chosen to create a gate that provides high signal intensity, while maintaining good temporal resolution. The gate should also achieve fluorescence rejection by being relatively narrow in time and having sharp or steep opening and closing slopes. 4,5 The narrow gate width is important for rejecting fluorescence of longer lifetimes, while the steep closing slope is most critical for rejecting fluorescence of shorter lifetimes. 3,4 Efremov et al., 2007, 4 provides a detailed overview of the quantification of fluorescence rejection, as well as the spectral resolutions possible with the TRRS system used in this work, under varying parameters. The specific optimal voltage values of the intensifier vary with excitation wavelength, and a characterization of the gate size provided by various settings of the intensifier, performed at 412 nm excitation, can be found in Figure 2.2. At this wavelength, a combination of V av and V cl of 19.6 and 10.3, respectively, provides a gate of approximately 250 ps, and is the best compromise between intensity and resolution. Table 2.1 summarizes the gate widths created by this and various other voltage combinations for which the delay time versus intensity profiles are shown in Figure 2.2. These profiles of Raman signal intensity versus detector delay time correlate with detector gate size, and they are a convolution of laser pulse width and temporal broadening from the sample. The FWHM (Full Width at Half Maximum) of these profiles is approximately equal to the detector gate width when measuring non-retarded Raman photons from a small detection volume of a transparent sample. In that case the contributions of the laser pulse width and pathlength differences are negligible. A thin (1 mm) quartz cuvette of cyclohexane was used to measure the profiles in Figure 2.2. Since the optimal intensifier settings depend on the kinetic energy of the electrons released from the photocathode, different settings were needed for measurements in the NIR. For λ ex = 720 nm we used V av = 21.0 V and V cl = 9.1 V. 23

5 OFF ON Figure 2.1. Intensifier of the ICCD camera. Photons are converted to electrons by the photocathode. The intensifier is OFF when the average and clamping voltage settings do not facilitate the flow of electrons through the multichannel plate (MCP). The intensifier is ON when the average and clamping voltage settings allow electrons to reach the multichannel plate (MCP) where they are multiplied. These electrons subsequently knock photons off of a phosphor screen to be detected by the CCD camera. Resulting Gate Width (ps) V Av (Volts) V Cl (Volts) (FWHM) Table 2.1. The approximate gate widths as FWHM provided by various combinations of V Av and V Cl settings of the ICCD camera for an excitation wavelength of 412 nm. I. E. Iping Petterson, unpublished work. 24

6 Net Peak Intensity ps 500 ps 175 ps 150 ps 200 ps Net Delay Time (ps) Figure 2.2. ICCD gate profiles; net peak intensity of a representative Raman peak of cyclohexane plotted versus detector delay time for various settings of the average voltage and clamping voltage (see Table 2.1), respectively, of the ICCD camera. The FWHM of these profiles is practically identical with the detector gate width. Excitation 412 nm. I. E. Iping Petterson, unpublished work. 2.4 Depth profiling measurements The short detector gate of 250 ps can be delayed in time after the laser pulse, as shown in Figure 2.3A. When a very short delay time is used, the detector will collect Raman photons that are backscattered primarily from the surface of the sample. However, when a longer delay time is used, a greater percentage of photons emanating from deeper within a scattering sample will be detected. These photons have traveled deeper into the sample, and thereby have a much longer total pathlength. Their travel time within the sample and to scatter back towards the detector will also be longer and with a broader temporal distribution than those photons from the surface, as shown in Figure 2.3B. By first using a short detector delay to obtain a spectrum from primarily the sample surface, and then obtaining subsequent spectra while increasing the delay in small increments, a depth profile 25

7 of a sample can be formed. This principle is put into practice and discussed further in Chapter 3 of this thesis. A 1 st layer 2 nd layer B Figure (A) Raman photons created on the surface and within the second layer of a scattering medium have different pathlengths and arrive at the detector at different times. (B) Diagram of the profiles in time of Raman photons from first and second layers of a 2-layer sample when measured with a pulsed laser system and ICCD camera with a 250 ps gate. Raman photons from the second layer have a less intense, broadened time profile. Implementing a relatively short delay time for the detector gate also allows for the avoidance of considerable amount of fluorescence background. 26

8 2.5 Effect of focusing In conventional Raman measurements of the surface of a sample, the focal distance of the excitation and collection optics plays a critical role in the efficiency of Raman signal detection. In performing TRRS measurements of a two-layer sample, this is true for the intensity of the signal from the surface as well. However, a different relationship is found between Raman signal intensity from the second layer and accurate focus on the sample surface. Figure 2.4 shows Raman intensity versus detector delay time profiles, recorded at different sample distances from the objective (0 mm- 4 mm) for a 2-layer sample, of which the first layer was a 3 mm thick layer of Teflon, with a second layer of a 2 mm thick block of Arnite. While the distance from the sample surface does have a significant effect on the intensity of signal from the first layer material, the intensity of Raman signal from the second layer, Arnite, appears to remain more or less constant for all distances from the objective we measured. This indicates that for depth measurements, there is an advantage towards improved ratio of signal from deeper layers to surface signal by not focusing on the surface layer. Teflon 1st Layer 3.5 mm distance Intensity Arnite 2nd Layer Delay time (ps) Figure 2.4. Effect of focusing on Raman intensity in depth profile measurements with TRRS. The first layer intensity changes significantly with sample distance, having a maximum corresponding with the collection lens focus, in this case 3.5 mm. The sample distance is not critical for the intensity of second layer Raman signal (for distances 0-4 mm). I. E. Iping Petterson, unpublished results. 27

9 2.6 Advanced Raman techniques combined with TRRS Spatially Offset Raman Spectroscopy and TR-SORS As Time-Resolved Raman Spectroscopy is a depth Raman technique that makes the distinction between surface and deeper Raman photons emanating from a sample on the basis of their temporal difference by detecting with a delayed time gate, Spatially-Offset Raman Spectroscopy (SORS) 6 makes this discrimination spatially. The SORS approach was first reported by Matousek, et al., In SORS, a lateral offset is created between the laser excitation point and the Raman photon collection point; effectively exciting the sample at a different location from where the Raman signal is collected. Due to the random-walk propagation of photons in a scattering material, the further from the excitation spot, the higher the probability that collected photons have emanated from deeper within the sample. This spatial offset is typically most effective on the scale of a few mm, as the total number of collected photons significantly decreases with an offset, thus with too large a spatial offset, too few photons will make it back to the detector. The optimal offset distance is dependent on the thickness and scattering properties of the first-layer material, however even a small offset will significantly reduce the contribution of Raman signal from the surface, as can be seen in Figure 2.5. This will provide an advantage for signal from deeper layers in a 2 layer sample, similar to the effect noted for such a sample when adjusting the sample distance. Although SORS is usually carried out with CW lasers and detectors, with our TRRS setup, combining Spatially-Offset excitation geometry with Time-Resolved detection is also possible, and this approach may be advantageous in particular sample scenarios. 7 Figure 2.5 shows a comparison of signal intensity from first and second layers in a 2-layer scattering sample when measured with conventional Raman spectroscopy and SORS (2.5A), and TRRS and TR-SORS (2.5B). This provides an overview of the effect of these different techniques on the signal intensity from first and second layers. Chapters 3 and 6 of this thesis discuss these effects and the combination of Time-Resolution with Spatially-Offset Raman spectroscopy in greater detail. 28

10 Intensity A 1st Layer 2nd Layer Intensity B 1st Layer 2nd Layer RS SORS 0 TRRS TR-SORS Figure 2.5. Intensity differences in measurements of a 2-layer sample of Teflon; 2 mm (first layer), and Arnite; 2 mm (second layer), when measured by (A) Continuous wave Raman spectroscopy (RS) versus SORS with a 5 mm offset, and by (B) TRRS with a 300 ps detector delay time versus TR-SORS with a 300 ps delay time plus 5 mm spatial offset. It should be noted that while intensity can be compared among bars within Figure A or B, intensity differences between frames A and B are arbitrary. I.E. Iping Petterson, unpublished results based on the spectra of Chapter 3. 29

11 Surface-Enhanced (Resonance) Raman Spectroscopy and TR-SERS Raman signals can be selectively enhanced by using a technique called Surface-Enhanced Raman Spectroscopy (SERS). This selective enhancement is provided primarily by roughened metal surfaces or nanoparticles, most frequently Au or Ag In general, SERS is understood to provide Raman scattering enhancement by various mechanisms: physical, chemical and resonance effects. These may include electronic associations between the plasmon of the charged metal nanoparticle and particular molecular vibrations of the analyte in resonance with the excitation wavelength. The electronic associations between plasmons and the analyte have been observed even if the molecule is only in the vicinity of the metal nanoparticle, occurring by plasmon charge transfer, however the strongest enhancement is generally provided for adsorbed molecules. 9,10 In this latter scenario SERS provides an additional advantage in that binding to metal nanoparticles by the analyte of interest, or by an impurity, may also quench their fluorescence. Additionally, when a molecule is excited with an incident excitation beam of an energy that coincides with or approaches that of a molecular transition to an excited electronic state (for example S 0 to S 1 ), the vibrations (= geometry changes) that are associated with this electronic transition will be in resonance with the excitation energy and can be selectively enhanced. This phenomenon is also possible without the presence of SERS nanoparticles, in which case it is called Resonance Raman Scattering (RRS), a mode of Raman spectroscopy introduced in Section 1.2. Improvements in enhancement can be obtained by creating a resonance between the absorption or molecular transition of the molecules and the resonance of the metal plasmon, through exciting the sample with approximately this wavelength. The strongest enhancement is generally obtained when the plasmon resonance of the metal and molecular absorption both overlap with the excitation wavelength. In this scenario absorption by the molecule and nanoparticle is also then at its maximum, which can have a detrimental effect on the Raman scattering intensity when concentrations of molecules and analytes are too high. Often in these resonant scenarios not all molecular vibrations are equally enhanced, however this loss of information may be advantageous for simplicity in studying samples in which a particular structure of interest is found within a complex matrix. 30

12 The type of nanoparticle used for SERS enhancement of an analyte is also an important consideration as it must be compatible chemically, as well as in its spectroscopic properties. Ag and Au nanoparticles are commonly used, and their specific properties are dependent on the particle sizes and shapes that have been synthesized. Adjusting these physical parameters of the nanoparticles allows them to be tuned to provide greatest enhancement at certain wavelengths. Also generally speaking Au can be more chemically inert that Ag, which may also be advantageous for certain applications. SERS is a technique that can be combined with numerous geometries and detection techniques, including TR, SORS, and TR-SORS. Chapter 5 of this work will discuss these combinations with SERS in greater detail. 31

13 2.7 References 1 Matousek, P., Towrie, M., Stanley, A., and Parker, A.W. Efficient rejection of fluorescence from Raman spectra using picosecond Kerr gating. Appl. Spectrosc., 1999, 53(12): Everall, N., Hahn, T., Matousek, P., Parker, A.W., and Towrie, M. Photon migration in Raman spectroscopy. Appl. Spectrosc., 2004, 58(5): Efremov, E.V., Ariese, F., and Gooijer, C. Achievements in resonance Raman spectroscopyreview of a technique with a distinct analytical chemistry potential. Analytica Chimica Acta, 2008, 606(2): Efremov, E.V., Buijs, J.B., Gooijer, C., and Ariese, F. Fluorescence rejection in resonance Raman spectroscopy using a picosecond-gated intensified CCD camera. Appl. Spectrosc., 61, 2007, Kentech, Didcot UK. High Rate Imager tutorial. tutorials. 23/11/2011 updated. 6 Matousek, P., Clark, I. P., Draper, E.R.C., Morris, M.D., Goodship, A. E., Everall, N., Towrie, M., Finney, W.F., and Parker, A.W. Subsurface Probing in Diffusely Scattering Media Using Spatially Offset Raman Spectroscopy. Appl. Spectrosc., 2005, 59(4): Iping Petterson, I. E., Dvorak, P., Buijs J.B., Gooijer, C., and Ariese, F. Time-resolved spatially offset Raman spectroscopy for depth analysis of diffusely scattering layers, Analyst, 2010, 135: Lee, P. and Meisel, D. Adsorption and surface-enhanced Raman of dyes on silver and gold sols. J. Phys. Chem., 1982, 86: Le Ru, E.C. and Etchgoin, P.G. A quick overview of surface-enhanced Raman spectroscopy. In: Principles of Surface-Enhanced Raman Spectroscopy and related plasmonic effects. Amsterdam, the Netherlands: Elsevier, pgs McNay, G., Eustace, D., Smith, W. E., Faulds, K., and Graham, D. Surface-Enhanced Raman Scattering (SERS) and Surface Enhanced Resonance Raman Scattering (SERRS): A Review of Applications. Appl. Spectrosc., 2011,65(8):

Chapter 1. Introduction. Concepts of Raman spectroscopy

Chapter 1. Introduction. Concepts of Raman spectroscopy Chapter 1. Introduction Concepts of Raman spectroscopy 11 1.1 Light scattering and the Raman effect Types of electromagnetic scattering When photons from a monochromatic light source such as a laser illuminate

More information

Chapter 3. Time resolved Raman spectroscopy for depth analysis of multi-layered mineral samples

Chapter 3. Time resolved Raman spectroscopy for depth analysis of multi-layered mineral samples Chapter 3 Time resolved Raman spectroscopy for depth analysis of multi-layered mineral samples 3 Jan-Hein Hooijschuur, Ingeborg E. Iping Petterson, Gareth R. Davies, Cees Gooijer and Freek Ariese Journal

More information

Richard Miles and Arthur Dogariu. Mechanical and Aerospace Engineering Princeton University, Princeton, NJ 08540, USA

Richard Miles and Arthur Dogariu. Mechanical and Aerospace Engineering Princeton University, Princeton, NJ 08540, USA Richard Miles and Arthur Dogariu Mechanical and Aerospace Engineering Princeton University, Princeton, NJ 08540, USA Workshop on Oxygen Plasma Kinetics Sept 20, 2016 Financial support: ONR and MetroLaser

More information

HYPER-RAYLEIGH SCATTERING AND SURFACE-ENHANCED RAMAN SCATTERING STUDIES OF PLATINUM NANOPARTICLE SUSPENSIONS

HYPER-RAYLEIGH SCATTERING AND SURFACE-ENHANCED RAMAN SCATTERING STUDIES OF PLATINUM NANOPARTICLE SUSPENSIONS www.arpapress.com/volumes/vol19issue1/ijrras_19_1_06.pdf HYPER-RAYLEIGH SCATTERING AND SURFACE-ENHANCED RAMAN SCATTERING STUDIES OF PLATINUM NANOPARTICLE SUSPENSIONS M. Eslamifar Physics Department, BehbahanKhatamAl-Anbia

More information

Time resolved optical spectroscopy methods for organic photovoltaics. Enrico Da Como. Department of Physics, University of Bath

Time resolved optical spectroscopy methods for organic photovoltaics. Enrico Da Como. Department of Physics, University of Bath Time resolved optical spectroscopy methods for organic photovoltaics Enrico Da Como Department of Physics, University of Bath Outline Introduction Why do we need time resolved spectroscopy in OPV? Short

More information

Highly Efficient and Anomalous Charge Transfer in van der Waals Trilayer Semiconductors

Highly Efficient and Anomalous Charge Transfer in van der Waals Trilayer Semiconductors Highly Efficient and Anomalous Charge Transfer in van der Waals Trilayer Semiconductors Frank Ceballos 1, Ming-Gang Ju 2 Samuel D. Lane 1, Xiao Cheng Zeng 2 & Hui Zhao 1 1 Department of Physics and Astronomy,

More information

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source 3rd International EUVL Symposium NOVEMBER 1-4, 2004 Miyazaki, Japan Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source H. Tanaka, A. Matsumoto, K. Akinaga, A. Takahashi

More information

Supplementary Information for. Vibrational Spectroscopy at Electrolyte Electrode Interfaces with Graphene Gratings

Supplementary Information for. Vibrational Spectroscopy at Electrolyte Electrode Interfaces with Graphene Gratings Supplementary Information for Vibrational Spectroscopy at Electrolyte Electrode Interfaces with Graphene Gratings Supplementary Figure 1. Simulated from pristine graphene gratings at different Fermi energy

More information

Fast and Slow Ligand Exchange at the Surface of Colloidal Gold Nanoparticles

Fast and Slow Ligand Exchange at the Surface of Colloidal Gold Nanoparticles Fast and Slow Ligand Exchange at the Surface of Colloidal Gold Nanoparticles Rebecca Dinkel 1, Björn Braunschweig 1,2 * and Wolfgang Peukert 1,2 1 Institute of Particle Technology (LFG), Friedrich-Alexander

More information

MEFT / Quantum Optics and Lasers. Suggested problems Set 4 Gonçalo Figueira, spring 2015

MEFT / Quantum Optics and Lasers. Suggested problems Set 4 Gonçalo Figueira, spring 2015 MEFT / Quantum Optics and Lasers Suggested problems Set 4 Gonçalo Figueira, spring 05 Note: some problems are taken or adapted from Fundamentals of Photonics, in which case the corresponding number is

More information

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or

More information

CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK

CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK 161 CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK 7.1 SUMMARY OF THE PRESENT WORK Nonlinear optical materials are required in a wide range of important applications, such as optical

More information

SWOrRD. For direct detection of specific materials in a complex environment

SWOrRD. For direct detection of specific materials in a complex environment SWOrRD For direct detection of specific materials in a complex environment SWOrRD Swept Wavelength Optical resonant Raman Detector RAMAN EFFECT Raman scattering or the Raman effect ( /rɑːmən/) is the inelastic

More information

New Developments in Raman Spectroscopic Analysis

New Developments in Raman Spectroscopic Analysis New Developments in Raman Spectroscopic Analysis Mike Kayat B&W Tek, Inc 19 Shea Way Newark, DE 19713 United States of America +1 302 368 7824 mikek@bwtek.com 1 Overview Raman spectroscopy is now an established

More information

1. Transition dipole moment

1. Transition dipole moment 1. Transition dipole moment You have measured absorption spectra of aqueous (n=1.33) solutions of two different chromophores (A and B). The concentrations of the solutions were the same. The absorption

More information

Supporting Information: Optical Spectroscopy

Supporting Information: Optical Spectroscopy Supporting Information: Optical Spectroscopy Aminofluorination of Cyclopropanes: A Multifold Approach through a Common, Catalytically Generated Intermediate Cody Ross Pitts, Bill Ling, Joshua A. Snyder,

More information

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 1. Examples of advantages and disadvantages with laser-based combustion diagnostic techniques: + Nonintrusive + High

More information

Implementation and evaluation of data analysis strategies for time-resolved optical spectroscopy

Implementation and evaluation of data analysis strategies for time-resolved optical spectroscopy Supporting information Implementation and evaluation of data analysis strategies for time-resolved optical spectroscopy Chavdar Slavov, Helvi Hartmann, Josef Wachtveitl Institute of Physical and Theoretical

More information

Chemical Analysis on complex biological systems by Raman spectroscopy. Janina Kneipp

Chemical Analysis on complex biological systems by Raman spectroscopy. Janina Kneipp Chemical Analysis on complex biological systems by Raman spectroscopy Janina Kneipp Analytical Strategies J. Kneipp Nov 2014 1 Raman scattering and resonant Raman scattering S 1 resonant Raman scattering

More information

Laser heating of noble gas droplet sprays: EUV source efficiency considerations

Laser heating of noble gas droplet sprays: EUV source efficiency considerations Laser heating of noble gas droplet sprays: EUV source efficiency considerations S.J. McNaught, J. Fan, E. Parra and H.M. Milchberg Institute for Physical Science and Technology University of Maryland College

More information

High quantum efficiency S-20 photocathodes for photon counting applications

High quantum efficiency S-20 photocathodes for photon counting applications High quantum efficiency S-20 photocathodes for photon counting applications D. A. Orlov a,*, J. DeFazio b, S. Duarte Pinto a, R. Glazenborg a and E. Kernen a a PHOTONIS Netherlands BV, Dwazziewegen 2,

More information

ECE280: Nano-Plasmonics and Its Applications. Week8

ECE280: Nano-Plasmonics and Its Applications. Week8 ECE280: Nano-Plasmonics and Its Applications Week8 Surface Enhanced Raman Scattering (SERS) and Surface Plasmon Amplification by Stimulated Emission of Radiation (SPASER) Raman Scattering Chandrasekhara

More information

Resonance Raman measurements utilizing a deep UV source

Resonance Raman measurements utilizing a deep UV source Resonance Raman measurements utilizing a deep UV source Adam Willitsford a, C. Todd Chadwick b, Hans Hallen b, and C. Russell Philbrick a a The Pennsylvania State University, Department of Electrical Engineering,

More information

Vibrational Spectra of Chloroform, Freon-11 and Selected Isotopomers in the THz Frequency Region

Vibrational Spectra of Chloroform, Freon-11 and Selected Isotopomers in the THz Frequency Region Vibrational Spectra of Chloroform, Freon-11 and Selected Isotopomers in the THz Frequency Region Christa Haase, Jinjun Liu, Frédéric Merkt, Laboratorium für physikalische Chemie, ETH Zürich current address:

More information

Application of IR Raman Spectroscopy

Application of IR Raman Spectroscopy Application of IR Raman Spectroscopy 3 IR regions Structure and Functional Group Absorption IR Reflection IR Photoacoustic IR IR Emission Micro 10-1 Mid-IR Mid-IR absorption Samples Placed in cell (salt)

More information

3. Excitation and Detection of Fluorescence

3. Excitation and Detection of Fluorescence 3. Excitation and Detection of Fluorescence In this chapter, we examine key experimental components and methods to observe weakly fluorescing objects. We consider in turn the excitation source, the detectors,

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION doi:1.138/nature1878 I. Experimental setup OPA, DFG Ti:Sa Oscillator, Amplifier PD U DC U Analyzer HV Energy analyzer MCP PS CCD Polarizer UHV Figure S1: Experimental setup used in mid infrared photoemission

More information

I. Proteomics by Mass Spectrometry 1. What is an internal standard and what does it accomplish analytically?

I. Proteomics by Mass Spectrometry 1. What is an internal standard and what does it accomplish analytically? Name I. Proteomics by Mass Spectrometry 1. What is an internal standard and what does it accomplish analytically? Internal standards are standards added intentionally to all samples, standards and blanks.

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Supplementary Information Speckle-free laser imaging using random laser illumination Brandon Redding 1*, Michael A. Choma 2,3*, Hui Cao 1,4* 1 Department of Applied Physics, Yale University, New Haven,

More information

No. 9 Experimental study on the chirped structure of the construct the early time spectra. [14;15] The prevailing account of the chirped struct

No. 9 Experimental study on the chirped structure of the construct the early time spectra. [14;15] The prevailing account of the chirped struct Vol 12 No 9, September 2003 cfl 2003 Chin. Phys. Soc. 1009-1963/2003/12(09)/0986-06 Chinese Physics and IOP Publishing Ltd Experimental study on the chirped structure of the white-light continuum generation

More information

Ultraviolet-Visible and Infrared Spectrophotometry

Ultraviolet-Visible and Infrared Spectrophotometry Ultraviolet-Visible and Infrared Spectrophotometry Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

Supplementary Materials

Supplementary Materials Supplementary Materials Sample characterization The presence of Si-QDs is established by Transmission Electron Microscopy (TEM), by which the average QD diameter of d QD 2.2 ± 0.5 nm has been determined

More information

Photothermal Spectroscopy Lecture 2 - Applications

Photothermal Spectroscopy Lecture 2 - Applications Photothermal Spectroscopy Lecture 2 - Applications Aristides Marcano Olaizola (PhD) Research Professor Department of Physics and Engineering Delaware State University, US 1 Outlook 1. Optical characterization

More information

Chem 524 Lecture Notes Raman (Section 17) 2013

Chem 524 Lecture Notes Raman (Section 17) 2013 Chem 524 Lecture Notes Raman (Section 17) 2013 For HTML of 2005 notes, click here XIII. Molecular Light Scattering and Raman Spectroscopy (Read Ch. 16) A. Elastic Scattering o = s - basis for Dynamic Light

More information

Molecular spectroscopy

Molecular spectroscopy Molecular spectroscopy Origin of spectral lines = absorption, emission and scattering of a photon when the energy of a molecule changes: rad( ) M M * rad( ' ) ' v' 0 0 absorption( ) emission ( ) scattering

More information

requency generation spectroscopy Rahul N

requency generation spectroscopy Rahul N requency generation spectroscopy Rahul N 2-11-2013 Sum frequency generation spectroscopy Sum frequency generation spectroscopy (SFG) is a technique used to analyze surfaces and interfaces. SFG was first

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2014 Electronic Supplementary Information Unique ultrafast energy transfer in a series of phenylenebridged

More information

Stimulated Raman scattering of XeCl 70 ns laser pulses in silica fibres

Stimulated Raman scattering of XeCl 70 ns laser pulses in silica fibres J. Opt. A: Pure Appl. Opt. 1 (1999) 725 729. Printed in the UK PII: S1464-4258(99)00367-0 Stimulated Raman scattering of XeCl 70 ns laser pulses in silica fibres Nikolai Minkovski, Ivan Divliansky, Ivan

More information

FEMTOSECOND MID-INFRARED SPECTROSCOPY OF HYDROGEN-BONDED LIQUIDS

FEMTOSECOND MID-INFRARED SPECTROSCOPY OF HYDROGEN-BONDED LIQUIDS Laser Chem., 1999, Vol. 19, pp. 83-90 Reprints available directly from the publisher Photocopying permitted by license only (C) 1999 OPA (Overseas Publishers Association) N.V. Published by license under

More information

Laser Dissociation of Protonated PAHs

Laser Dissociation of Protonated PAHs 100 Chapter 5 Laser Dissociation of Protonated PAHs 5.1 Experiments The photodissociation experiments were performed with protonated PAHs using different laser sources. The calculations from Chapter 3

More information

Intrinsic beam emittance of laser-accelerated electrons measured by x-ray spectroscopic imaging

Intrinsic beam emittance of laser-accelerated electrons measured by x-ray spectroscopic imaging Intrinsic beam emittance of laser-accelerated electrons measured by x-ray spectroscopic imaging G. Golovin 1, S. Banerjee 1, C. Liu 1, S. Chen 1, J. Zhang 1, B. Zhao 1, P. Zhang 1, M. Veale 2, M. Wilson

More information

Prediction and Optimization of Surface-Enhanced Raman Scattering Geometries using COMSOL Multiphysics

Prediction and Optimization of Surface-Enhanced Raman Scattering Geometries using COMSOL Multiphysics Excerpt from the Proceedings of the COMSOL Conference 2008 Hannover Prediction and Optimization of Surface-Enhanced Raman Scattering Geometries using COMSOL Multiphysics I. Knorr 1, K. Christou,2, J. Meinertz

More information

Development of optimized Raman Spectroscopy setup for species detection in flames

Development of optimized Raman Spectroscopy setup for species detection in flames Development of optimized Raman Spectroscopy setup for species detection in flames Henrik Johansson Thesis submitted for the degree of Bachelor of Science Project duration: 2 months Supervised by Christian

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

Surface Plasmon Amplification by Stimulated Emission of Radiation. By: Jonathan Massey-Allard Graham Zell Justin Lau

Surface Plasmon Amplification by Stimulated Emission of Radiation. By: Jonathan Massey-Allard Graham Zell Justin Lau Surface Plasmon Amplification by Stimulated Emission of Radiation By: Jonathan Massey-Allard Graham Zell Justin Lau Surface Plasmons (SPs) Quanta of electron oscillations in a plasma. o Electron gas in

More information

Survey on Laser Spectroscopic Techniques for Condensed Matter

Survey on Laser Spectroscopic Techniques for Condensed Matter Survey on Laser Spectroscopic Techniques for Condensed Matter Coherent Radiation Sources for Small Laboratories CW: Tunability: IR Visible Linewidth: 1 Hz Power: μw 10W Pulsed: Tunabality: THz Soft X-ray

More information

Diagnósticos em Plasmas

Diagnósticos em Plasmas Tecnologia a Plasma para o Processamento de Materiais Diagnósticos em Plasmas Diagnósticos Ópticos João Santos Sousa, nº50901 Semestre Inverno 2004/2005 21 de Janeiro de 2005, 9h-10h, sala F8 Contents

More information

CHEM Outline (Part 15) - Luminescence 2013

CHEM Outline (Part 15) - Luminescence 2013 CHEM 524 -- Outline (Part 15) - Luminescence 2013 XI. Molecular Luminescence Spectra (Chapter 15) Kinetic process, competing pathways fluorescence, phosphorescence, non-radiative decay Jablonski diagram

More information

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. CHEM 322 Name Exam 3 Spring 2013 Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. Warm-up (3 points each). 1. In Raman Spectroscopy, molecules are promoted

More information

LABORATORY OF ELEMENTARY BIOPHYSICS

LABORATORY OF ELEMENTARY BIOPHYSICS LABORATORY OF ELEMENTARY BIOPHYSICS Experimental exercises for III year of the First cycle studies Field: Applications of physics in biology and medicine Specialization: Molecular Biophysics Fluorescence

More information

Introduction to laser-based combustion diagnostics

Introduction to laser-based combustion diagnostics Introduction to laser-based combustion diagnostics (Lecture 1b) Lecture prepared for course in laser-based combustion diagnostics by Per-Erik Bengtsson and Joakim Bood Division of Combustion Physics at

More information

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept. Spectrum of Electromagnetic Radiation Electromagnetic radiation is light. Different energy light interacts with different motions in molecules. CHEM*344 Chemical Instrumentation Topic 7 Spectrometry Radiofrequency

More information

Precision VUV spectroscopy of Ar I at 105 nm

Precision VUV spectroscopy of Ar I at 105 nm J. Phys. B: At. Mol. Opt. Phys. 32 (999) L5 L56. Printed in the UK PII: S0953-4075(99)05625-4 LETTER TO THE EDITOR Precision VUV spectroscopy of Ar I at 05 nm I Velchev, W Hogervorst and W Ubachs Vrije

More information

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a). 7.1. Low-Coherence Interferometry (LCI) Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a). The light is split by the beam splitter (BS) and

More information

χ (3) Microscopic Techniques

χ (3) Microscopic Techniques χ (3) Microscopic Techniques Quan Wang Optical Science and Engineering University of New Mexico Albuquerque, NM 87131 Microscopic techniques that utilize the third order non-linearality (χ (3) ) of the

More information

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Section I Q1. Answer (i) (b) (ii) (d) (iii) (c) (iv) (c) (v) (a) (vi) (b) (vii) (b) (viii) (a) (ix)

More information

Temperature-dependent spectroscopic analysis of F 2 + ** and F 2 + **-like color centers in LiF

Temperature-dependent spectroscopic analysis of F 2 + ** and F 2 + **-like color centers in LiF Journal of Luminescence 91 (2000) 147 153 Temperature-dependent spectroscopic analysis of F 2 + ** and F 2 + **-like color centers in LiF Neil W. Jenkins a, *, Sergey B. Mirov a, Vladimir V. Fedorov b

More information

Abstract... I. Acknowledgements... III. Table of Content... V. List of Tables... VIII. List of Figures... IX

Abstract... I. Acknowledgements... III. Table of Content... V. List of Tables... VIII. List of Figures... IX Abstract... I Acknowledgements... III Table of Content... V List of Tables... VIII List of Figures... IX Chapter One IR-VUV Photoionization Spectroscopy 1.1 Introduction... 1 1.2 Vacuum-Ultraviolet-Ionization

More information

Raman and stimulated Raman spectroscopy of chlorinated hydrocarbons

Raman and stimulated Raman spectroscopy of chlorinated hydrocarbons Department of Chemistry Physical Chemistry Göteborg University KEN140 Spektroskopi Raman and stimulated Raman spectroscopy of chlorinated hydrocarbons WARNING! The laser gives a pulsed very energetic and

More information

Understanding Nanoplasmonics. Greg Sun University of Massachusetts Boston

Understanding Nanoplasmonics. Greg Sun University of Massachusetts Boston Understanding Nanoplasmonics Greg Sun University of Massachusetts Boston Nanoplasmonics Space 100pm 1nm 10nm 100nm 1μm 10μm 100μm 1ns 100ps 10ps Photonics 1ps 100fs 10fs 1fs Time Surface Plasmons Surface

More information

Initial Hydrogen-Bonding Dynamics of. Photoexcited Coumarin in Solution with. Femtosecond Stimulated Raman Spectroscopy

Initial Hydrogen-Bonding Dynamics of. Photoexcited Coumarin in Solution with. Femtosecond Stimulated Raman Spectroscopy Electronic Supplementary Material (ESI) for Journal of Materials Chemistry C. This journal is The Royal Society of Chemistry 2015 Electronic Supplementary Information (ESI) for: Initial Hydrogen-Bonding

More information

A few Experimental methods for optical spectroscopy Classical methods Modern methods. Remember class #1 Generating fast LASER pulses

A few Experimental methods for optical spectroscopy Classical methods Modern methods. Remember class #1 Generating fast LASER pulses A few Experimental methods for optical spectroscopy Classical methods Modern methods Shorter class Remember class #1 Generating fast LASER pulses, 2017 Uwe Burghaus, Fargo, ND, USA W. Demtröder, Laser

More information

College of Chemistry and Chemical Engineering, Shenzhen University, Shenzheng, Guangdong, P. R. China. 2

College of Chemistry and Chemical Engineering, Shenzhen University, Shenzheng, Guangdong, P. R. China. 2 Electronic Supplementary Material (ESI) for Physical Chemistry Chemical Physics. This journal is the Owner Societies 5 Supplementary Information Remarkable Effects of Solvent and Substitution on Photo-dynamics

More information

Supplementary Materials for

Supplementary Materials for wwwsciencemagorg/cgi/content/full/scienceaaa3035/dc1 Supplementary Materials for Spatially structured photons that travel in free space slower than the speed of light Daniel Giovannini, Jacquiline Romero,

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Chem Homework Set Answers

Chem Homework Set Answers Chem 310 th 4 Homework Set Answers 1. Cyclohexanone has a strong infrared absorption peak at a wavelength of 5.86 µm. (a) Convert the wavelength to wavenumber.!6!1 8* = 1/8 = (1/5.86 µm)(1 µm/10 m)(1 m/100

More information

Surface-enhanced raman scattering from a layer of gold nanoparticles

Surface-enhanced raman scattering from a layer of gold nanoparticles VNU Journal of Science, Mathematics - Physics 26 (2010) 187-192 Surface-enhanced raman scattering from a layer of gold nanoparticles Nguyen The Binh *, Nguyen Thanh Dinh, Nguyen Quang Dong, Vu Thi Khanh

More information

plasma optics Amplification of light pulses: non-ionised media

plasma optics Amplification of light pulses: non-ionised media Amplification of light pulses: non-ionised media since invention of laser: constant push towards increasing focused intensity of the light pulses Chirped pulse amplification D. Strickland, G. Mourou, Optics

More information

Supplementary Material for In situ frequency gating and beam splitting of vacuum- and extreme-ultraviolet pulses

Supplementary Material for In situ frequency gating and beam splitting of vacuum- and extreme-ultraviolet pulses Supplementary Material for In situ frequency gating and beam splitting of vacuum- and extreme-ultraviolet pulses Rajendran Rajeev, Johannes Hellwagner, Anne Schumacher, Inga Jordan, Martin Huppert, Andres

More information

Lecture 5. X-ray Photoemission Spectroscopy (XPS)

Lecture 5. X-ray Photoemission Spectroscopy (XPS) Lecture 5 X-ray Photoemission Spectroscopy (XPS) 5. Photoemission Spectroscopy (XPS) 5. Principles 5.2 Interpretation 5.3 Instrumentation 5.4 XPS vs UV Photoelectron Spectroscopy (UPS) 5.5 Auger Electron

More information

(002)(110) (004)(220) (222) (112) (211) (202) (200) * * 2θ (degree)

(002)(110) (004)(220) (222) (112) (211) (202) (200) * * 2θ (degree) Supplementary Figures. (002)(110) Tetragonal I4/mcm Intensity (a.u) (004)(220) 10 (112) (211) (202) 20 Supplementary Figure 1. X-ray diffraction (XRD) pattern of the sample. The XRD characterization indicates

More information

1 Mathematical description of ultrashort laser pulses

1 Mathematical description of ultrashort laser pulses 1 Mathematical description of ultrashort laser pulses 1.1 We first perform the Fourier transform directly on the Gaussian electric field: E(ω) = F[E(t)] = A 0 e 4 ln ( t T FWHM ) e i(ω 0t+ϕ CE ) e iωt

More information

Investigation of fundamental mechanisms related to ambient gas heating and hydrodynamics of laser-induced plasmas

Investigation of fundamental mechanisms related to ambient gas heating and hydrodynamics of laser-induced plasmas Investigation of fundamental mechanisms related to ambient gas heating and hydrodynamics of laser-induced plasmas P. J. Skrodzki Acknowledgements This work is supported by the DOE/NNSA Office of Nonproliferation

More information

Fundamental investigation on CO 2 laser-produced Sn plasma for an EUVL source

Fundamental investigation on CO 2 laser-produced Sn plasma for an EUVL source Fundamental investigation on CO 2 laser-produced Sn plasma for an EUVL source Yezheng Tao*, Mark Tillack, Kevin Sequoia, Russel Burdt, Sam Yuspeh, and Farrokh Najmabadi University of California, San Diego

More information

Multidimensional femtosecond coherence spectroscopy for study of the carrier dynamics in photonics materials

Multidimensional femtosecond coherence spectroscopy for study of the carrier dynamics in photonics materials International Workshop on Photonics and Applications. Hanoi, Vietnam. April 5-8,24 Multidimensional femtosecond coherence spectroscopy for study of the carrier dynamics in photonics materials Lap Van Dao,

More information

Supplementary Figure 1 Schematics of an optical pulse in a nonlinear medium. A Gaussian optical pulse propagates along z-axis in a nonlinear medium

Supplementary Figure 1 Schematics of an optical pulse in a nonlinear medium. A Gaussian optical pulse propagates along z-axis in a nonlinear medium Supplementary Figure 1 Schematics of an optical pulse in a nonlinear medium. A Gaussian optical pulse propagates along z-axis in a nonlinear medium with thickness L. Supplementary Figure Measurement of

More information

Diffuse reflection BBSFG optical layout

Diffuse reflection BBSFG optical layout Diffuse reflection BBSFG optical layout Figure 1 shows the optical layout of the broad bandwidth sum frequency generation (BBSFG) system. A Nd:YVO 4 laser (a, Spectra-Physics MillenniaVs) pumps the Ti:Sapphire

More information

Sum-Frequency Generation Spectra of Thin Organic Films on Silver Enhanced Due to Surface Plasmon Excitation

Sum-Frequency Generation Spectra of Thin Organic Films on Silver Enhanced Due to Surface Plasmon Excitation E. V. Alieva et al.: Sum-Frequency Generation Spectra of Thin Organic Films on Ag 109 phys. stat. sol. (a) 175, 109 (1999) Subject classification: 78.30.Jw; 73.20.Mf; 78.66.Qn; S12 Sum-Frequency Generation

More information

Chemistry Instrumental Analysis Lecture 5. Chem 4631

Chemistry Instrumental Analysis Lecture 5. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 5 Light Amplification by Stimulated Emission of Radiation High Intensities Narrow Bandwidths Coherent Outputs Applications CD/DVD Readers Fiber Optics Spectroscopy

More information

"Molecular Photochemistry - how to study mechanisms of photochemical reactions?"

Molecular Photochemistry - how to study mechanisms of photochemical reactions? "Molecular Photochemistry - how to study mechanisms of photochemical reactions?" Bronislaw Marciniak Faculty of Chemistry, Adam Mickiewicz University, Poznan, Poland 2014/2015 - lecture 4 Contents 1. Introduction

More information

Vibrational Spectroscopies. C-874 University of Delaware

Vibrational Spectroscopies. C-874 University of Delaware Vibrational Spectroscopies C-874 University of Delaware Vibrational Spectroscopies..everything that living things do can be understood in terms of the jigglings and wigglings of atoms.. R. P. Feymann Vibrational

More information

Photoelectron Spectroscopy using High Order Harmonic Generation

Photoelectron Spectroscopy using High Order Harmonic Generation Photoelectron Spectroscopy using High Order Harmonic Generation Alana Ogata Yamanouchi Lab, University of Tokyo ABSTRACT The analysis of photochemical processes has been previously limited by the short

More information

Measurements of plasma temperature and electron density in laser-induced copper plasma by time-resolved spectroscopy of neutral atom and ion emissions

Measurements of plasma temperature and electron density in laser-induced copper plasma by time-resolved spectroscopy of neutral atom and ion emissions PRAMANA c Indian Academy of Sciences Vol. 74, No. 6 journal of June 2010 physics pp. 983 993 Measurements of plasma temperature and electron density in laser-induced copper plasma by time-resolved spectroscopy

More information

Time-resolved spectroscopy

Time-resolved spectroscopy Time-resolved spectroscopy Chih-Wei Luo ( 羅志偉 ) Department of Electrophysics, National Chiao Tung University, Taiwan Ultrafast Dynamics Lab Outline 1. Introduction of pulses 2. Spectroscopic methods for

More information

Nanocomposite photonic crystal devices

Nanocomposite photonic crystal devices Nanocomposite photonic crystal devices Xiaoyong Hu, Cuicui Lu, Yulan Fu, Yu Zhu, Yingbo Zhang, Hong Yang, Qihuang Gong Department of Physics, Peking University, Beijing, P. R. China Contents Motivation

More information

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at 2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various

More information

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Electromagnetic Spectrum - Molecular transitions Widely used in chemistry. Perhaps the most widely used in Biological Chemistry.

More information

Application of Raman Spectroscopy for Noninvasive Detection of Target Compounds. Kyung-Min Lee

Application of Raman Spectroscopy for Noninvasive Detection of Target Compounds. Kyung-Min Lee Application of Raman Spectroscopy for Noninvasive Detection of Target Compounds Kyung-Min Lee Office of the Texas State Chemist, Texas AgriLife Research January 24, 2012 OTSC Seminar OFFICE OF THE TEXAS

More information

Application of time resolved area normalized emission spectroscopy to multicomponent systems

Application of time resolved area normalized emission spectroscopy to multicomponent systems JOURNAL OF CHEMICAL PHYSICS VOLUME 115, NUMBER 15 15 OCTOBER 2001 Application of time resolved area normalized emission spectroscopy to multicomponent systems A. S. R. Koti and N. Periasamy a) Department

More information

LC-4: Photoelectric Effect

LC-4: Photoelectric Effect LC-4: Photoelectric Effect Lab Worksheet Name In this lab you investigate the photoelectric effect, one of the experiments whose explanation by Einstein forced scientists into accepting the ideas of quantum

More information

25 Instruments for Optical Spectrometry

25 Instruments for Optical Spectrometry 25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.

More information

Lecture 0. NC State University

Lecture 0. NC State University Chemistry 736 Lecture 0 Overview NC State University Overview of Spectroscopy Electronic states and energies Transitions between states Absorption and emission Electronic spectroscopy Instrumentation Concepts

More information

Chapter 24 Photonics Question 1 Question 2 Question 3 Question 4 Question 5

Chapter 24 Photonics Question 1 Question 2 Question 3 Question 4 Question 5 Chapter 24 Photonics Data throughout this chapter: e = 1.6 10 19 C; h = 6.63 10 34 Js (or 4.14 10 15 ev s); m e = 9.1 10 31 kg; c = 3.0 10 8 m s 1 Question 1 Visible light has a range of photons with wavelengths

More information

Hiromitsu TOMIZAWA XFEL Division /SPring-8

Hiromitsu TOMIZAWA XFEL Division /SPring-8 TUPLB10 (Poster: TUPB080) Non-destructive Real-time Monitor to measure 3D- Bunch Charge Distribution with Arrival Timing to maximize 3D-overlapping for HHG-seeded EUV-FEL Hiromitsu TOMIZAWA XFEL Division

More information

Laser Physics OXFORD UNIVERSITY PRESS SIMON HOOKER COLIN WEBB. and. Department of Physics, University of Oxford

Laser Physics OXFORD UNIVERSITY PRESS SIMON HOOKER COLIN WEBB. and. Department of Physics, University of Oxford Laser Physics SIMON HOOKER and COLIN WEBB Department of Physics, University of Oxford OXFORD UNIVERSITY PRESS Contents 1 Introduction 1.1 The laser 1.2 Electromagnetic radiation in a closed cavity 1.2.1

More information

Sfb 658 Colloquium 11 May Part II. Introduction to Two-Photon-Photoemission (2PPE) Spectroscopy. Martin Wolf

Sfb 658 Colloquium 11 May Part II. Introduction to Two-Photon-Photoemission (2PPE) Spectroscopy. Martin Wolf Sfb 658 Colloquium 11 May 2006 Part II Introduction to Two-Photon-Photoemission (2PPE) Spectroscopy Martin Wolf Motivation: Electron transfer across interfaces key step for interfacial and surface dynamics

More information

CHAPTER 3 RESULTS AND DISCUSSION

CHAPTER 3 RESULTS AND DISCUSSION CHAPTER 3 RESULTS AND DISCUSSION 3.1 CHAPTER OUTLINE This chapter presents the data obtained from the investigation of each of the following possible explanations: (1) Experimental artifacts. (2) Direct

More information

Performance Limits of Delay Lines Based on "Slow" Light. Robert W. Boyd

Performance Limits of Delay Lines Based on Slow Light. Robert W. Boyd Performance Limits of Delay Lines Based on "Slow" Light Robert W. Boyd Institute of Optics and Department of Physics and Astronomy University of Rochester Representing the DARPA Slow-Light-in-Fibers Team:

More information

Spectroscopy tools for PAT applications in the Pharmaceutical Industry

Spectroscopy tools for PAT applications in the Pharmaceutical Industry Spectroscopy tools for PAT applications in the Pharmaceutical Industry Claude Didierjean Sr. Technology and Applications Consultant Real Time Analytics Mettler Toledo AutoChem, Inc. claude.didierjean@mt.com

More information