Liquid chromatography^tandem mass spectrometric analysis of surface and waste. (LC)^atmospheric pressure chemical ionisation

Size: px
Start display at page:

Download "Liquid chromatography^tandem mass spectrometric analysis of surface and waste. (LC)^atmospheric pressure chemical ionisation"

Transcription

1 trends in analytical chemistry, vol. 19, no. 4, Liquid chromatography^tandem mass spectrometric analysis of surface and waste water with atmospheric pressure chemical ionisation I: instrumentation Paul G.M. Kienhuis*, Rene B. Geerdink RIZA, P.O. Box 17, 8200 AA Lelystad, The Netherlands Liquid chromatography (LC) coupled to tandem mass spectrometry (MS /MS), i.e. LC^MS /MS, is a technique for the analysis of medium polarity to ionic compounds. This and the following paper highlight the potential and limitations of atmospheric pressure chemical ionisation (APCI) combined with MS /MS in the analysis of water. Part I describes the APCI interface and the scan modes of a triple quadrupole mass spectrometer. It gives a general introduction to the instrumentation, from a practical point of view. Part II gives more detailed information, in which the instrumentation is optimised for and used in the analysis of surface water and waste water. Attention is paid to the use of methanol versus acetonitrile as organic modi er in combination with APCI, the application of the RF-only daughter scan mode, and the quantitation of known compounds and the identi cation of unknown compounds found in surface and waste water. z2000 Elsevier Science B.V. All rights reserved. Keywords: Liquid chromatography^tandem mass spectrometry; Thermospray; Tandem mass spectrometry; RFD scan mode; Instrumentation; Atmospheric pressure chemical ionization; Daughter scan mode; Collisionally induced dissociation 1. Introduction *Corresponding author In order to analyse surface and waste water at the Institute for Inland Water Management and Waste Water Treatment (RIZA) in The Netherlands, several methods based on mass spectrometric detection are used to cover the range from volatile to medium polar compounds. The volatile ones (with a target list of about 60 compounds) are analysed by on-line purge and trap concentration followed by gas chromatography (GC)^mass spectrometry (MS) detection. The less volatile compounds (about 400) are analysed by off-line extraction on XAD-4, or liquid^liquid extraction with dichloromethane, followed by GC^MS detection. The thermolabile or medium polar compounds (about 70) are analysed by on-line SPE extraction or large loop injections, followed by liquid chromatography (LC)^atmospheric pressure chemical ionisation (APCI)-tandem MS (MS /MS) detection. The aim of each method is to concentrate or extract as many compounds as possible, to quantify the target compounds and to identify the unknown peaks. This article highlights the LC^APCI-MS /MS analysis. We started in 1990 with a combination of LC^ UV^Thermospray-MS /MS (on a triple-stage quadrupole (TSQ)-70 instrument, Finnigan MAT) and subsequently purchased an atmospheric pressure ionisation (API) interface in 1996 in order to improve the ionisation of the more polar and ionic compounds. API has largely replaced the thermospray interface, because of its ease of operation, the possibility of ionising large molecules, the improved ability to handle higher ow-rates, the stability of the ionisation process, the possibility of generating fragment ions on a single quadrupole mass spectrometer, and the introduction of relative inexpensive `desktop' instruments. The API interface can be used in two different ionisation modes: APCI which, just like thermospray, is suitable for the neutral compounds, and electrospray ionisation /00/$ ^ see front matter ß 2000 Elsevier Science B.V. All rights reserved. PII: S (99)

2 250 trends in analytical chemistry, vol. 19, no. 4, 2000 (ESI), which is suitable for the ionic compounds [ 1^4 ]. Both thermospray and API are soft ionisation techniques. In the mass spectra, ions related to the intact molecule of a compound are observed mainly, ionised by detachment [M3H] 3 or attachment of a proton [M+H ], ammonia ([M+NH 4 ], ammoniated molecule) or other cations or anions. However, in order to identify a compound and to improve the reliability of an analysis, more speci c ions are necessary [ 5 ]. Fragmentation by means of collisionally induced dissociation (CID) is, at the moment, the most common way to get additional ions. CID can be performed either internally in an API source ( by a process called source CID [ 6 ] or cone voltage fragmentation [ 7 ]) or in a tandem mass spectrometer. This article is divided into two parts, and describes the design, optimisation and application of the LC^APCI-MS /MS instrumentation. Part I gives a general introduction and rules-of-thumb to optimise the APCI interface and the scan modes and settings of a triple quadrupole mass spectrometer. Detailed discussions on the instrumentation, the theoretical principles, and mechanisms can be found elsewhere [ 7^10 ]. Part II gives more speci c information concerning the application of the instrumentation to the analysis of surface and waste water. As a reference, the method used in combination with thermospray is discussed [ 11], followed by the problems (such as the choice of the organic modi er) encountered while changing from thermospray to APCI. Attention is paid to the use of several scan modes, the analysis and identi- cation of target compounds, of compounds often found in surface water, and the so-called CI^CID spectrum library. 2. Ionisation In GC^MS, a low ow of an inert gas (helium) carries the compounds into the ion source. With LC^MS, a relatively large amount of solvent enters the ionisation interface and strongly affects the ion- Fig. 1. In the Finnigan MAT APCI interface, the mobile phase (L) is passed into the atmospheric pressure region (AP) of the interface through a fused silica capillary (150 Wm ID) as sample inlet tube. A sheath-gas (Sh) ow (N 2 ), and eventually an auxiliary (A) ow (N 2 ), is used to nebulise the solvent and to keep the interface dry. The resultant spray is heated by a heater block (V, vaporiser) and sprayed against a spray shield. The nebulised solvent and analyte molecules are partly ionised by a corona discharge (D). The extent of solvent-vapour mixture in the interface is directed to waste, through two tubes (W) into a hood. The droplets formed in the tubes are combined in a ask. To prevent back-pressure, a large-bore tube and ask (which has been used previously in our laboratory as a cold nger in the thermospray exit line) is applied. Some other users apply a small suction pump to the drain connection. The heated capillary (C) samples the atmospheric-pressure region of the interface continuously. Ions and neutrals in the gas or liquid phase are drawn into the capillary and transported to the reduced pressure (RP) region by a reducing pressure gradient. The capillary is heated to reduce the amount of solvent. A tube-lens (T) directs the ions towards the skimmer (Sk) ori ce, which is off-axis of the capillary in order to reduce the amount of aerosol particles entering the high-vacuum (HV) region of the mass spectrometer. Between the skimmer ori ce and the rst mass lter, an octapole (O) is used to focus and transport the ions.

3 trends in analytical chemistry, vol. 19, no. 4, Table 1 Overview and general instrumental settings of the more important parameters that will in uence the sensitivity of the Finnigan MAT APCI interface Parameters Instrumental settings Mobile phase parameters % Organic modi er Type of modi er Other eluent constituents such as salt ( molarity), acidity Instrumental parameters Flow-rate Up to 2 ml / min; generally 0.1^1 ml / min Corona current 4^6 WA Sheath-gas ow-rate 10^100 psi Auxiliary gas ow-rate Normally not used Vaporiser temperature 350^425³C Heated capillary temperature 170^200³C Capillary voltage Dependent on most of the other parameters Tube-lens voltage Strongly dependent on most of the other parameters isation. The way in which the eluent composition affects the ionisation process is different for thermospray and APCI. Thermospray ionisation is performed in the highvacuum envelope of the mass spectrometer. To prevent freezing at the ori ce, the eluent is heated in a vaporiser before entering the source. To obtain the best response, the extent of heat must be such that only part of the solvent has been evaporated before leaving the vaporiser as a spray of droplets. Evaporation of the droplets continues in the high vacuum of the source whereas ion^molecule reactions are taking place during evaporation and in the gas-phase. The ion residence time in the ion source is short, i.e. of the order of Ws [9,10]. APCI is performed at atmospheric pressure. The heat, necessary to evaporate the eluent, is now admitted after nebulisation of the eluent. Ions are formed by means of a corona discharge. The ion residence time in the source is much longer, of the order of ms. Ion^molecule reactions in the gas-phase dominate the ionisation process [ 9,10,12,13 ]. These differences in ionisation and reaction time result in different in uences of the analyte and solvent properties on the ions formed during ionisation. The Finnigan APCI interface has a heated capillary to guide the ions into the high-vacuum envelope of the mass spectrometer (Fig. 1). The settings discussed below are speci c for this interface: other types of API interfaces may act in different ways. The rst choice to be made is in the selection of the eluent composition (Table 1). It is important to keep in mind that ionisation via APCI is chemistry in the gas-phase. The process starts with the generation of N 2 and O 2 3 ions by means of a continuous corona discharge. These ions react with the solvent and analyte molecules. Owing to the relatively long residence time in the APCI source, the ions measured are the result of an equilibrium in the gasphase. The eluent composition affects this equilibrium; for example, more compounds can be ionised in the positive-ion mode when methanol is used as organic modi er instead of acetonitrile. An explanation can be found in the proton af nity of the compounds involved (Table 2). The proton af nity of methanol is lower than for acetonitrile, so more analytes can be ionised by proton attachment in the presence of methanol. The proton af nity, as such, does not fully explain the ionisation process, but it is important to realise that solvents and compounds in the eluent can affect the ionisation process. Therefore, in the analysis of positive ions, it is important that the components of the applied solvent composition have a lower proton af nity than the analyte(s) to be measured. On the other hand, Table 2 Proton af nity and gas-phase acidity of LC^MS-related compounds [ 14 ] Compound Proton af nity in kj / mol Ammonia Acetic acid Acetonitrile Methanol Formic acid Water Gas-phase acidity in kj / mol

4 252 trends in analytical chemistry, vol. 19, no. 4, 2000 Table 3 In uence of several instrumental parameters on the ion current of m/z215 [M+H ] of metribuzin continuously infused at a concentration of 200 ng / ml Flow-rate (ml/min) Heated capillary (³C) Sheath-gas pressure (PSI) Capillary voltage (V) Tube-lens voltage (V) m/z215, countsu After each change of the ow-rate, temperature of the heated capillary, or the sheath-gas ow, the intensity of m/z215 is shown before and after the capillary and tube-lens voltages have been optimised. Mobile phase methanol /water (50% v /v, 1 ml acetic acid /l). Vaporiser 450³C, No auxiliary gas ow. when a target compound has a high proton af nity (in general, nitrogen-containing compounds [ 12,13 ]), it might be useful to consider acetonitrile in order to reduce the ionisation of interfering compounds. The choice of a solvent also affects the amount of internal energy an ion gains during ionisation, and thus the degree of CID fragmentation later on [ 15 ]. In the negative-ion mode, the gas-phase acidity affects the ionisation process. An increased tendency of a compound to ionise by detachment of a proton is related to a lower value of the gas-phase acidity. Therefore, when a lower ph of the high performance liquid chromatography (HPLC) eluent is necessary for separation of acidic compounds, the S /N ratio is better if acetic acid is used to acidify the HPLC mobile phase instead of the stronger formic acid, although the difference in gas-phase acidity is small (Table 2). Examples of these equilibrium phenomena will be given in Part II. The Finnigan APCI interface can handle a maximum ow-rate of about 2 ml / min. Flow-rates between 0.1 and 1 ml / min are generally used. No large differences in sensitivity are observed at different ow-rates after optimisation of all parameters at each ow-rate. It is important to know that almost all the instrumental parameters indicated in Table 1 are dependent on the solvent ow-rate, except for the corona discharge current. The auxiliary gas ow (N 2 ) is generally not necessary. For many compounds it has a large effect, with `no auxiliary gas' as the best option. It is also useful to optimise the auxiliary gas ow in the LC mode (where the compound is only in the ion source for a short time) and not in the FIA mode (where the compound is continuously available). The sheath-gas (N 2 ) pressure is more important, and should be increased at increasing ow-rates. The whole pressure-gauge range of 0^100 psi has to be used to optimise the signal intensity and stability. The optimum values for the capillary and tubelens voltage are affected by most of the parameters, but especially by the vaporiser and capillary temperature. The optimal range of the tube-lens voltage is rather small, and must be checked after each change of a setting. An example of the in uence of some parameters on the signal intensity of m/z215 of metribuzin is shown in Table 3. On the left side of the table, the parameters: ow-rate, capillary temperature and sheath-gas pressure, are shown. In the middle are the applied capillary and tube-lens voltages, and to the right the resulting signal intensity of m/z215. At the chosen parameter settings of the rst line, a signal intensity of 75U10 6 is reached after optimisation of the voltages. On the second line, the ow-rate has been changed to 0.5 ml / min. A signal intensity of 37U10 6 counts is reached at the same voltages as in the rst line, but the intensity increases to 40U10 6 in the third line after optimisation of the capillary voltage and the tube-lens voltage. In the same way, the sheath-gas ow is changed at lines 4^5, and the capillary temperature at lines 6^7. Every time, the signal intensity before and after the optimisation of the capillary and the tube-lens voltages are shown. The optimum values for the vaporiser and capillary temperature (after optimisation of the capillary- and tube-lens voltage) are rather at, and less ow-rate dependent than the sheath-gas ow. A vaporiser temperature of 350^425³C and a heated capillary temperature of 170^200³C are common

5 trends in analytical chemistry, vol. 19, no. 4, Fig. 2. Scan modes of a triple quadrupole mass spectrometer. settings, but for speci c compounds, a higher capillary temperature may be favourable [ 16 ]. Garcia et al. [ 17 ] reported the optimisation of 10 variables of the Finnigan APCI interface in the positive-ion mode for ibuprofen, 2-(4-isobutylphenyl)propionic acid, as the target compound, using fractional factorial- and response-surface modelling. Their results differ from the settings shown in Table 1. They found a ow-rate of 0.1 ml / min or lower, in combination with a high water content, a high auxiliary gas ow, a low tube-lens voltage of 55 V, and a heated capillary temperature of 225³C, to be optimum. As indicated by the authors, these optimised parameters might be speci c for ibuprofen. Schaefer and Dixon [ 18 ] investigated the quenching effect of several common buffers on the ionisation of ibuprofen and other model compounds, in the negative-ion mode with APCI on a Sciex triple quadrupole mass spectrometer. The

6 254 trends in analytical chemistry, vol. 19, no. 4, 2000 results show the quenching effect of the more strongly acidic buffer components on the weakly acidic compounds. They advise the use of the base, N-methylmorpholine, as buffer component in the analysis of weakly acidic compounds. Spliid and Koppen [ 19 ] described the optimisation of the Finnigan APCI interface for a group of pesticides at a ow-rate of 0.2 ml / min and with methanol as organic modi er. The optimisation of the inlet parameters such as capillary temperature, vaporiser temperature, corona current and sheathgas pressure was discussed and the results are comparable with our own ndings. In our laboratory, the mass spectrometer, including the analyser part, is serviced twice a year. After each service, the whole instrument has to be set properly. In order to tune and calibrate the mass spectrometer in combination with the APCI interface in the desired mass range (10^700 m/z), a mixture of analytes at a concentration of 200 Wg/l is used both for the positive and negative-ionisation mode (Table 4). The mixture is continuously introduced at a ow-rate of 0.5 ml / min at the optimal APCI settings speci c for the mixture. The resulting general tune- le is used as starting point to optimise the source parameters for the different applications for which the mass spectrometer is used. In this way, for each application, a tune- le is created which is speci c for each application at the same ow-rate, solvent composition, and temperature settings as applied in the analysis. After tuning the APCI with the analyte mixture, no contamination of the inlet system was seen after cleaning of the spray shield and ushing of the pump, tubing and heated capillary with methanol: water (50/50, v /v). Tuning and calibration of the whole instrument are normally only performed after cleaning of the whole analyser part of the TSQ, and proved to be very stable. The source (vaporiser, corona needle, heated capillary, skimmer and rst lenses) is cleaned more often. After cleaning, the source parameters are checked. Generally, no large differences in settings are found after cleaning. 3. Fragmentation With electron-impact ionisation, both ionisation and fragmentation of the molecules occur at the same time. The electron beam `hits' the molecules. An electron is lost, but the total process is so energetic that most of the ionised molecules fragment. These fragment ions are essential to con rm the identity of a compound [ 5,20 ]. Soft ionisation processes are less energetic and produce mainly ions containing the intact molecule. Additional processes are required to produce fragment ions. CID is the most common way to do so. The ions are accelerated to increase their kinetic energy and pass through a chamber with a collision gas. In an API interface, CID can be performed in the API source [ 6,7 ]. In a triple-stage quadrupole mass spectrometer, a quadrupole or octapole acts as the collision chamber, with an inert gas as collision gas. Parent ions enter the collision chamber and the fragment ions generated are called daughter ions Source CID By the application of an additional voltage in the region between the heated capillary and the skimmer, or in the region between the skimmer and the octapole, the ions are accelerated (Fig. 1). Nitrogen from the sheath-gas and solvent molecules acts as collision gas. Source CID offers the possibility of fragmenting ions in combination with a single quadrupole mass spectrometer, but only one `mode' is available. All ions are involved in the CID; they cannot be selected as in a tandem mass spectrometer. Table 4 Tune masses of the mixture used to tune and calibrate the mass spectrometer in the positive- and negative-ion modes Compounds Positive ions m/z Negative ions m/z Ethylenethiourea [M+H ] (ETU) DNOC [M3H] 3 Metribuzin [M+H ] Warfarin [M3H] 3 Tris(2-butoxyethyl) [M+H ] phosphate Fluazinam [M3H] 3 Reserpine [M+H ] [M3H] 3 Ions of the solvent: Water [H 2 O] Acetic acid [M3H] 3 A mixture of seven compounds is used at a concentration of 200 ng /l each, dissolved in methanol /water (50:50, v /v) with 1 ml acetic acid /l.

7 trends in analytical chemistry, vol. 19, no. 4, Fig. 3. Quadrupole offset voltages in the MS /MS modes. The offset voltage of a scanning quadrupole is about 3^5 V. Therefore, after leaving the source, the ions will gain an additional kinetic energy of 3^5 ev before passing the quadrupole. In the MS /MS modes, especially, the offset voltage of Q3 is important. Graph A shows a situation for positive ions without collision gas. The ions are accelerated to an additional kinetic energy of 20 ev before passing the collision cell, and reduced back to 5 ev by the offset voltage of Q3. In Graph B, collision gas at a low pressure of about 0.5 mtorr argon is present in the collision cell. The kinetic energy of the ions will be reduced by one or some collisions resulting in a divergent and insuf cient remaining energy to pass Q3 properly. To compensate for this effect, the offset voltage of Q3 must be increased ( Graph C). For the different MS /MS scan modes, Eq. 1 DOFF ms=ms ˆ DOFF ms COFF 13 M d =M p MSMSC=100 Š 1 is used by the instrument to calculate the offset voltage of Q3, where DOFF ms=ms represents the offset voltage of Q3 in the MS /MS scan modes; DOFF ms represents the offset voltage of Q3 in the Q3MS scan mode; COFF the applied collision-offset voltage; M d the m/zof the daughter ion; M p the m/zof the parent ion, and MSMSC the MSMS correction factor. TheMSMSCrangesfrom0to100anddependsonthepressureandcompound[ 26 ]: acommonvalueatalowgaspressureisabout70. At higher collision gas pressures, the ions will have multiple collisions and therefore their remaining kinetic energy is much lower. In this situation (Graph D), the offset voltage of Q3 is added to the actual collision offset voltage using an MSMSC=0, resulting in Eq. 2: DOFF ms=ms ˆ DOFF ms COFF 2 In the RFD scan mode, an MSMSC = 0 must be used because the cut-off mass value is used by the TSQ as M p, resulting in unrealistic offset voltages. As an example, in the positive-ion mode: DOFF ms = 35V,COFF=320 V, M p = cut-off mass = 70, M d = 200 m/zand MSMSC = 70. The resulting DOFF ms=ms will be +15 V and the positively charged daughter ions cannot pass Q3. The voltage difference can be changed from scan to scan. This can be used to cover a voltage range to generate a whole range of fragment ions. Josephs [ 21] uses this option to acquire spectra of drugs at CID offset voltages of 0, 320, 330 and 340 V on alternate scans. The averaged spectrum of a peak is used to build a spectral library, which is then used for automated

8 256 trends in analytical chemistry, vol. 19, no. 4, 2000 Fig. 4. In uence of the collision-gas (argon) pressure on the peak performance of atrazine ( m/z216/218) at 0, 1, 2 and 5 mtorr. Collision offset 320 V, MSMSC = 0.

9 trends in analytical chemistry, vol. 19, no. 4, library searching of peaks detected in unknown formulations TSQ scan modes The TSQ is a tandem mass spectrometer with three quadrupoles (Fig. 2) in-line [ 22 ]. A quadrupole can act as a mass analyser by a combination of AC (RF range, about 1 MHz) and DC voltages [ 23 ]. If only an RF voltage is used, all ions above a certain m/zratio (the so-called `cut-off mass') are not only allowed to pass the quadrupole, but are also focused. This focusing property is used in the central quadrupole (Q2), which is pressurised with a collision gas and used as the collision chamber. The extent of fragmentation depends on the internal energy of the ions gained during ionisation, the voltage difference between the ion source and the collision chamber (collision offset), and the pressure and the properties of the collision gas. Typically, argon is used as collision gas. Recent versions of the TSQ use an octapole instead of a quadrupole to improve the focusing effect. If next to the RF an additional DC voltage is applied, the quadrupole is able to act as a mass analyser or mass lter. It actually works as a bandpass lter: only ions of one particular m/zratio are allowed to pass. All other ions will collide on one of the rods. At a resolution of one m/z(nominalmass resolution, peak-width at half height is 0.65^0.7 mass units), about 10% of the ions of the selected m/zvalue will pass the quadrupole The rst and third quadrupoles can arbitrarily be used as an RF-only quadrupole or as the mass lter, resulting in several single MS and MS /MS scan modes. In Fig. 2, most of the possible combinations are shown. In the single MS scan modes, either the rst (Q1MS) or the third quadrupole (Q3MS) acts as a mass lter, and no collision gas is admitted to the collision cell. The result resembles a single quadrupole system. The Q3MS gives a better sensitivity compared to Q1MS, because of the focusing effect of Q1 and Q2 [ 24 ]. In the RF-only daughter (RFD) scan mode, Q1 is operated as an RF-only device, the collision cell is pressurised with collision gas, and Q3 acts as a mass lter. All ions above a pre-set cutoff m/z are allowed to enter the collision cell and are allowed to fragment. Unfragmented parent ions are scanned together with the generated fragment (daughter) ions. This scan mode (SRM) can be compared to a single quadrupole mass spectrometer with a separated ionisation and fragmentation chamber. In the RF-only parent scan mode (not shown in Fig. 2), Q1 is used as mass lter, Q2 is pressurised and Q3 acts as an RF-only quadrupole. The ion current measured for each m/z selected by Q1 is acquired for that speci c mass. If no ions are lost in the collision cell, the resulting mass spectrum should be identical to a spectrum acquired in the Q1MS scan mode. No useful application for this scan mode is reported in the literature. In the MS /MS scan modes, both Q1 and Q3 are used as mass lters. The most common and speci c combination is the daughter-ion scan mode (DAU). Here, Q1 is set on a xed m/z value and the selected ions are passed into the collision cell. ThedaughterionsarescannedwithQ3atanappropriate scan range. In the parent-ion scan mode (PAR), Q1 is scanning and Q3 is set on a xed m/ z value. The result is a spectrum of all ions producing the speci c daughter ion. As an example, m/z 149 is used to search for phthalates. In the acquired spectrum, all parent ions producing m/z 149 are shown. In the neutral-loss scan mode (NEU), both Q1 and Q3 are scanning, but with a constant m/z difference. The result now is a spectrum of all parent ions losing a speci c neutral part in the collision cell. The selected reaction monitoring mode (SRM) can be compared with the selected ion monitoring mode of a mass spectrometer with a single mass analyser. By analysing only a selected number of ions, instead of a mass range, the analysis time for each ion is increased, improving the S /N ratio. This mode is very useful for analysing a limited number of target compounds. The Q1 selects the parent ions one by one, and the speci c daughter ions of each parent ion are acquired in Q3. It is also possible to adjust the collision offset for each single reaction to achieve the best response. The speci city of this mode is often used to reduce, or even skip, the chromatographic part of the analysis [25] Collision-gas pressure and collision-offset voltage In order to achieve an effective CID process, three parameters must be set: õ A collision gas must be present in Q2 at the right pressure.

10 258 trends in analytical chemistry, vol. 19, no. 4, 2000 Fig. 5. In uence of argon pressure on the signal intensity of atrazine at a low and a medium collision-offset voltage ( m/z 216, [M+H] at 37 V,andm/z174, fragment ion at 320 V). õ õ The ions must have enough kinetic and internal energy to get collisions resulting in fragment ions. The ions entering the third quadrupole must have such a kinetic energy that they can be analysed properly. For the collision-gas pressure, two options are available. Until about 1990^1994, it was quite common to apply a low gas pressure (about 0.5 mtorr), resulting in one or a few collisions per ion [ 26 ]. To produce fragment ions, a difference of 10^50 V is needed in the offset voltage between the source and the collision cell to accelerate the ions between the source and the collision cell. Owing to the low collision-gas pressure, only a few collisions can take place and therefore, most of the fragmented and unfragmented ions leave Q2 with a rather high velocity. In the Q3MS scan mode, the offset voltage of Q3 is set to 3^5 V. Using the same offset voltage in the MS /MS scan modes, the kinetic energy of the ions will be reduced before entering Q3. For example, with a collision offset of 320 V, positive ions will gain an additional kinetic energy of 20 ev between the source and the collision cell. With an offset of 35 V for Q3, the kinetic energy will be reduced by 15 ev, resulting in a remaining energy of 5 ev. Without collision gas in the collision cell, this works very well (Fig. 3A). However, when the cell is pressurised, the ions will lose some of their kinetic energy at each collision and therefore cannot enter Q3 properly (Fig. 3B). To compensate for this effect, a correction factor (MSMSC) is used (Fig. 3C). The actual value of the correction factor, however, depends strongly on the actual collision-gas pressure and the properties of the ions involved, resulting in a detection system which is dif cult to optimise, and not robust [ 27 ]. Nowadays, a higher collision-gas pressure (2^4 mtorr argon) is common. In this pressure range, the offset voltage of Q3 is calculated by adding the offset voltage normally applied in the Q3MS mode (3^5 V) to the actual collision-offset voltage (MSMSC = 0): thus, at a collision offset of 320 V, the offset voltage of Q3 is 323 to 325 V. Now, the collision gas must reduce the kinetic energy so that the ions will pass Q3 with the proper kinetic energy (Fig. 3D). The in uence of the collision-gas pressure on the resolution is shown in Fig. 4. The resolution of m/z216 and 218 of atrazine is shown at various collision-gas pressures, while a collisionoffset voltage of 320 V is applied. In the TSQ-700, a collision-gas pressure of about 1.5 mtorr is suf cient to establish the nominal mass resolution again. At higher pressures, up to 5 mtorr, the reso-

11 trends in analytical chemistry, vol. 19, no. 4, lution does not increase further. The collision gas is also needed to fragment ions. A low collision-offset voltage can be used to analyse the intact parent ions (RFD scan mode) or the higher mass daughters of very unstable molecules (RFD and MS /MS scan modes). Higher collision-offset voltages are needed to fragment the parent ions. The collisiongas pressure affects the signal intensity of the ions at the different collision-offset voltages. At a low collision-offset voltage, a high collisiongas pressure reduces the signal intensity of the ions. In Fig. 5, the signal intensity of m/z216 of atrazine is shown. At the applied collision-offset voltage of 37 V, the ions do not fragment. The signal intensity reduces at increasing collision-gas pressure, because not all ions are able to pass the collision cell at higher gas pressures. The optimum gas pressure is around 1.5 mtorr argon. The minimum gas pressure needed to give a normal resolution is also 1.5 mtorr, so a pressure between 1.5 and 2 mtorr is optimal. At a medium collision-offset voltage, it is better to use a higher collision-gas pressure. In Fig. 5, the signal intensity of m/z 174 of atrazine is shown. At the applied offset voltage of 320 V, m/z 216 fragments to m/z 174. The fragmentation process becomes more ef cient at higher collision-gas pressures, until an optimum is reached at 3^5 mtorr. When both low and higher collision-offset voltages have to be applied in the same run, a collision-gas pressure must be chosen to meet both requirements. A compromise is found between 2 and 3 mtorr. 4. Applications With the instrumentation and general rules described here, a variety of types of applications can be performed. APCI or ESI together with MS / MS are often used, and are very successful in the analysis of a limited number of target compounds in combination with short LC-columns and short analysis times. For these methods, the solvents and systemparameterscanbechosentobeoptimalforthe compounds involved. In the analysis of surface water and waste water, however, choices must be made in order to analyse as many compounds as possible in a limited numberofruns.thewaysinwhichwedealwiththis goal, together with the results of optimisation and analyses, are described in Part II. Acknowledgements The authors thank Wilfried Niessen of Hyphen MassSpec Consultancy, Leiden, The Netherlands, for critically reading the manuscript and for his helpful suggestions. References [ 1] I. Ferrer, D. Barceloè, Analysis 26 (6) (1998) M118. [ 2 ] R.D. Voyksner, Environ. Sci. Technol. 28 (1994) 118A. [ 3 ] A.P. Bruins, Trends Anal. Chem. 13 (1994) 37. [ 4 ] A.P. Bruins, Trends Anal. Chem. 13 (1994) 81. [ 5 ] L.Y.T. Li, D.A. Campbell, P.K. Bennet, J. Henion, Anal. Chem. 68 (1996) [ 6 ] A. Cary, Technical Report No. 620, 7000 Series Finnigan MAT, [ 7 ] B.L.M. van Baar, in: D. Barceloè (editor), Applications of LC^MS in Environmental Chemistry, J. Chromatogr. Lib., Vol. 59, Elsevier, Amsterdam, 1996, p. 71. [ 8 ] P.J. Arpino, Mass Spectrom. Rev. 9 (1990) 631. [ 9 ] W.M.A. Niessen and J. van der Greef, Liquid Chromatography-Mass Spectrometry, Chromatographic Science Series, Vol. 58, Marcel Dekker, New York, [ 10 ] W.M.A. Niessen, in: D. Barceloè (editor), Applications of LC^MS in Environmental Chemistry. J. Chromatogr. Lib., Vol. 59, Elsevier, Amsterdam, 1996, p. 3. [ 11] P.G.M. Kienhuis, R.B. Geerdink, A. Sijpersma, Anal. Methods Instrum. 2 (1995) 236. [ 12 ] J. Sunner, G. Nicol, P. Kerbarle, Anal. Chem. 60 (1988) [ 13 ] J. Sunner, G. Nicol, P. Kerbarle, Anal. Chem. 60 (1988) [ 14 ] S.G. Lias, J.E. Bartmess, J.F. Liebman, J.L. Holmes and R.D. Levin, J. Phys. Chem. Ref. Data 17, Suppl. 1 (1988). [ 15 ] R. Wolf, 13th LC /MS Symposium on Liquid Chromatography, Montreux, [ 16 ] A. Rosell-Meleè, J.F. Carter, J.R. Maxwell, J. Am. Soc. Mass Spectrom. 7 (1996) 965. [ 17 ] D.M. Garcia, S.K. Huang, W.F. Stansbury, J. Am. Soc. Mass Spectrom. 7 (1996) 59. [ 18 ] W.H. Schaefer, F. Dixon Jr., J. Am. Soc. Mass Spectrom. 7 (1996) [ 19 ] N.H.Spliid,B.Koppen,J.Chromatogr.A736 (1996)105. [ 20 ] P.G.M. Kienhuis, J. Chromatogr. 647 (1993) 39. [ 21] J.L. Josephs, Application Report No. 244, Finnigan MAT, [ 22 ] E. de Hoffmann, J. Mass Spectrom. 31 (1996) 129. [ 23 ] P.E. Miller, M. Bonner Denton, J. Chem. Educ. 63 (1986) 617. [ 24 ] P.H. Dawson, J.B. French, J.A. Buckley, D.J. Douglas, D. Simmons, Org. Mass Spectrom. 17 (1982) 205. [ 25 ] R.B. Geerdink, in: D. Barceloè (editor), Applications of LC^MS in Environmental Chemistry, J. Chromatogr. Lib., Vol. 59, Elsevier, Amsterdam, 1996, p [ 26 ] S.A.McLuckey,J.Am.Soc.MassSpectrom.3 (1992)599. [ 27 ] J. Abian, G. Durand, D. Barcelo, J. Agric. Food Chem. 41 (8) (1993) 1264.

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Selecting an LC/MS Interface Becky Wittrig, Ph.D. Selecting an LC/MS Interface Becky Wittrig, Ph.D. RESTEK CORPORATION LC/MS Interfaces I. Background of LC/MS I. Historical Perspective II. Reasons for use II. Interfaces I. Transport devices II. Particle

More information

M M e M M H M M H. Ion Sources

M M e M M H M M H. Ion Sources Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer 38 Current Trends in Mass Spectrometry November 6 Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer This article demonstrates the improved

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Quattro Micro - How does it work?

Quattro Micro - How does it work? Quattro Micro - How does it work? 1 Introduction This document is designed to familiarise you with the principles behind how the Quattro Micro works. The level of this document is designed as Level One

More information

Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation

Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation Mass Spectrometry Fundamental LC-MS Electrospray Ionisation Instrumentation i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be

More information

This document is a preview generated by EVS

This document is a preview generated by EVS TECHNICAL REPORT RAPPORT TECHNIQUE TECHNISCHER BERICHT CEN/TR 15641 August 2007 ICS 67.050 English Version Food analysis - Determination of pesticide residues by LC- MS/MS - Tandem mass spectrometric parameters

More information

Technical Note. Introduction

Technical Note. Introduction Technical Note Analysis and Characterization of Psilocybin and Psilocin Using Liquid Chromatography - Electrospray Ionization Mass Spectrometry (LC-ESI-MS) with Collision-Induced-Dissociation (CID) and

More information

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid

More information

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Lecture 15: Introduction to mass spectrometry-i

Lecture 15: Introduction to mass spectrometry-i Lecture 15: Introduction to mass spectrometry-i Mass spectrometry (MS) is an analytical technique that measures the mass/charge ratio of charged particles in vacuum. Mass spectrometry can determine masse/charge

More information

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS) Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions

More information

API Optimization, Source Design, and Other Considerations of LC/MS

API Optimization, Source Design, and Other Considerations of LC/MS API Optimization, Source Design, and Other Considerations of LC/MS Instrument Design Considerations: The Atmosphere-Vacuum Interface and Pumping System The hows and whys of the vacuum? We ve got to get

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

Mass spectrometry.

Mass spectrometry. Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

Introduction to LC-MS

Introduction to LC-MS Wednesday April 5, 2017 10am Introduction to LC-MS Amy Patton, MS Laboratory Manager, Pinpoint Testing, LLC Little Rock, AR DESCRIPTION: Amy Patton, laboratory manager for Pinpoint Testing, will begin

More information

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap Pro LC/MS/MS Accurate, High-Throughput Protein Identification

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

New Approaches in the Analysis of Amadori Compounds

New Approaches in the Analysis of Amadori Compounds 7 TH Wartburg Symposium on Flavor Chemistry and Biology Eisenach, April 21 st - April 23 rd, 2004 New Approaches in the Analysis of Amadori Compounds T. Davidek, I. Blank, K. Kraehenhuehl, J. Hau and S.

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

HPLC solvents and mobile phase additives

HPLC solvents and mobile phase additives HPLC solvents and mobile phase additives The following is intended to be a guide for the choice of solvent and mobile phase additives. Of course the choice of solvents for LC will be dictated primarily

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision

More information

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 129 Infi nity LC, and 646 Triple Quadrupole LC/MS Application Note Clinical Research Authors Linda Côté and Christophe

More information

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson

More information

Mass Spectrometry. Fundamental LC-MS. Atmospheric Pressure Chemical Ionisation (APCI)

Mass Spectrometry. Fundamental LC-MS. Atmospheric Pressure Chemical Ionisation (APCI) Mass Spectrometry Fundamental LC-MS Atmospheric Pressure Chemical Ionisation (APCI) i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot

More information

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,

More information

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS Application Note Authors E. Michael Thurman and Imma Ferrer Center for Environmental

More information

Introduction. Chapter 1. Learning Objectives

Introduction. Chapter 1. Learning Objectives Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Selecting Detectors for Compounds with No Optical Absorbance

Selecting Detectors for Compounds with No Optical Absorbance C146-E114 Selecting Detectors for Compounds with No Optical Absorbance Technical Report vol.15 1. Features of Absorbance Detectors In HPLC, the detector is selected in accordance with the analyte. Many

More information

Presentation Basic Introduction to Instrumentation Matrix Effects Challenges

Presentation Basic Introduction to Instrumentation Matrix Effects Challenges Keys to Implementing Mass Spectrometry in the Clinical Laboratory Paul J. Taylor Chromatography Mass Spectrometry Satellite Meeting, AACB 20th September, 2013 Dept of Clinical Pharmacology, Princess Alexandra

More information

Construction and Use of API Single- Quadrupole MS Spectral Libraries.

Construction and Use of API Single- Quadrupole MS Spectral Libraries. Paul Zavitsanos Product Marketing Manager September 11, 2003 Construction and Use of API Single- Quadrupole MS Spectral Libraries. Time: 11:00 a.m. EDT Telephone Number: 888-840-7687 International Number:

More information

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance

More information

CHROMATOGRAPHY AND MASS SPECTROMETER

CHROMATOGRAPHY AND MASS SPECTROMETER 22 CHROMATOGRAPHY AND MASS SPECTROMETER 22.1 INTRODUCTION We know that the biochemistry or biological chemistry deals with the study of molecules present in organisms. These molecules are called as biomolecules

More information

Supporting Information

Supporting Information 1 Supporting Information 2 3 Discovery and implications of C 2 and C 3 perfluoroalkyl sulfonates in aqueous film forming foams (AFFF) and groundwater 4 Krista A. Barzen-Hanson a and Jennifer A. Field b*

More information

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) 1.0 Equipment: Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) Thermo Exactive Orbitrap Mass Spectrometer Direct Analysis

More information

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n 20.2 Ion Sources electron ionization produces an M + ion and extensive fragmentation chemical ionization produces an M +, MH +, M +, or M - ion with minimal fragmentation MALDI uses laser ablation to transfer

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

Mass Spectrometry (MS)

Mass Spectrometry (MS) Mass Spectrometry (MS) Alternative names: Mass spectrometric (selective) detector (MSD) Spectrometry - methods based on interaction of matter and radiation Mass spectrometry - method based on formation

More information

Mass Spectrometry. Electron Ionization and Chemical Ionization

Mass Spectrometry. Electron Ionization and Chemical Ionization Mass Spectrometry Electron Ionization and Chemical Ionization Mass Spectrometer All Instruments Have: 1. Sample Inlet 2. Ion Source 3. Mass Analyzer 4. Detector 5. Data System http://www.asms.org Ionization

More information

Choosing the metabolomics platform

Choosing the metabolomics platform GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation

More information

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4.

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4. Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4. Field Desorption 5. MS MS techniques Matrix assisted

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Training Courses. Chromatography training from the Crawford Scientific experts

Training Courses. Chromatography training from the Crawford Scientific experts Training Courses Chromatography training from the Crawford Scientific experts HPLC Fundamentals This one-day course introduces the fundamentally important concepts associated with HPLC analysis including

More information

INTRODUCTION OVERVIEW

INTRODUCTION OVERVIEW COMPARISON OF CCS(N 2 ) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION S WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION Kevin Giles, Martin Palmer, Keith Richardson, Nick Tomczyk Waters MS Technologies

More information

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Improved Screening for 2 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Mary lackburn, Jia Wang, Jonathan eck, Charles Yang, Dipankar Ghosh, Thermo Fisher Scientific, San Jose, C,

More information

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Xiang He and Marta Kozak ThermoFisher Scientific, San Jose, CA,

More information

Analytical Technologies and Compound Identification. Daniel L. Norwood, MSPH, PhD SCĪO Analytical Consulting, LLC.

Analytical Technologies and Compound Identification. Daniel L. Norwood, MSPH, PhD SCĪO Analytical Consulting, LLC. Analytical Technologies and Compound Identification Daniel L. Norwood, MSPH, PhD SCĪ Analytical Consulting, LLC. Extractables/Leachables Characterization (USP) Characterization is the discovery,

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

LC MS analysis of metabolites. Basis of Chromatography

LC MS analysis of metabolites. Basis of Chromatography GBS 724 LC MS analysis of metabolites Stephen Barnes, PhD 4 7117; sbarnes@uab.edu Basis of Chromatography A moving (mobile) phase passes over an inert, stationary phase The compounds differentially interact

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

Liquid Chromatography in conjunction with Mass Spectrometry (LC-MS)

Liquid Chromatography in conjunction with Mass Spectrometry (LC-MS) Review Article Liquid Chromatography in conjunction with Mass Spectrometry (LC-MS) Shelke PG*, Tripathi AS, Dewani AP, Bakal RL, Mohale DS and Chandewar AV P. Wadhwani College of Pharmacy, Yavatmal, Maharashtra,

More information

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS

More information

Chapter 1. Introduction

Chapter 1. Introduction Chapter 1. Introduction 1-1 1.1. Overview In the past twenty years, charged droplets and strong electric fields have quietly revolutionized chemistry. In combination with an atmospheric-sampling mass spectrometer,

More information

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories APPLICATION NOTE 65200 Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories Authors Neloni Wijeratne, Claudia Martins, Mary

More information

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) By Arianne Bazilio & Jacob Weinrich December 2012 Contents Introduction... 2 Sample Introduction... 3 Torch... 4 Interface... 6

More information

Journal of Chemical and Pharmaceutical Research, 2012, 4(1): Research Article

Journal of Chemical and Pharmaceutical Research, 2012, 4(1): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2012, 4(1):231-238 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Mass spectrometric study of naproxen dimer anions

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Jonathan Williams Waters Corporation, Milford, MA, USA A P P L I C AT

More information

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry PO-CON18E Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) ASMS 21 TP 3 Jing Dong, Satoshi Yamaki, Yuki Hashi, Naoki Hamada Shimadzu China,

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent

More information

Ionization Techniques Part IV

Ionization Techniques Part IV Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector

More information

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction Determination of Pesticide Residues in Oats by Automated QuEChERS and LC/QQQ Application Note Abstract The QuEChERS (Quick-Easy-Cheap-Effective-Rugged-Safe) sample extraction method was developed for the

More information

FAIMS Technology at a Glance

FAIMS Technology at a Glance FAIMS Technology at a Glance Field asymmetric ion mobility spectrometry (FAIMS), also known as differential mobility spectrometry (DMS), is a gas detection technology that separates and identifies chemical

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Application Note: 56 Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Yang Shi, Catherine Lafontaine, Matthew Berube, John Fink, François Espourteille Thermo

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

MS-based proteomics to investigate proteins and their modifications

MS-based proteomics to investigate proteins and their modifications MS-based proteomics to investigate proteins and their modifications Francis Impens VIB Proteomics Core October th 217 Overview Mass spectrometry-based proteomics: general workflow Identification of protein

More information

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS PO-CON1745E Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS ASMS 2017 MP 189 Shailendra Rane 1, Ashutosh Shelar 1, Shailesh Damale 1, Rashi Kochhar 1, Purshottam Sutar 1,

More information

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information