Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation

Size: px
Start display at page:

Download "Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation"

Transcription

1 Mass Spectrometry Fundamental LC-MS Electrospray Ionisation Instrumentation i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual.

2 Aims and Objectives Aims and Objectives Aims Explain the function of the major components of an electrospray Interface Investigate methods of optimising signals using electrospray Ionisation Objectives At the end of this Section you should be able to: List and describe the most important components of an electrospray Ionisation interface Demonstrate an understanding of the principles of optimising instrument response when using electrospray ionisation

3 Content Electrospray source design 3 Introduction 3 Electrospray capillary design 3 Sprayer Sampling Plate Configuration 6 Cluster Ions 9 Prevention of Cluster Ion Sampling 10 Source Cleaning 12 Ion Optics 13 Ring Electrode 13 Ion Bridges 14 Collision Induced Dissociation 14 Ion Declustering 16 References 16 Crawford Scientific 2

4 Electrospray source design Introduction Electrospray is the dispersion of a liquid into electrically charged droplets, combining the two processes of droplet formation and droplet charging. The process of droplet charging is affected by three main variables: Eluent flow rate Liquid surface tension Electrolyte concentration If these parameters are not maintained at an optimised minimum level, the electrospray process will become unstable. If any of the variables increases significantly it maybe difficult for the electric field to produce the desired charged aerosol necessary for ion production in the API interface. Any effects observed due to a significant increase in any of the variables may be countered to a certain degree by increasing the capillary voltage (and hence the effective field strength at the capillary tip), but electrical discharge may occur, resulting in a decrease in instrument response and an unstable electrospray. Electrospray capillary design Standard electrospray capillaries are constructed from stainless steel or a coaxial arrangement of fused silica and stainless steel. If the capillary is of fused silica design, electrical contact is usually made by clamping the capillary in a metal union, however, capillaries with silver or gold deposits upstream from the tip have also been employed for improved electrical contact. Conventional Electrospray Capillary (practical upper flow rate limited to 10-20μL/min) Crawford Scientific 3

5 The metal composition used for stainless steel capillaries is of great importance in the oxidative and reductive processes occurring during droplet charging and as such continuity of composition for the metal capillaries should be ensured to give long term robustness to analytical determinations. Drying gas The practical upper limit to eluent flow in pure electrospray is μl/min depending upon the solvent composition. Capillary design may be modified to increase the tolerance of the electrospray process to increases in eluent flow rate, liquid surface tension or electrolyte concentration. One successful approach used in order to increase electrospray flow rate is the introduction of a nebulising gas via a concentric tube around the capillary (pneumatically assisted ESI). [1] High-flow electrospray sources (>5-10 L/min.) are normally combined with a supply of heat within the API source housing to assist the evaporation of solvents. The evaporation of large amounts of solvent is important to ensure that ion evaporation occurs in the optimum position within the API source to ensure maximum transmission and reduction of ion solvent clusters in the nozzle-skimmer region of the source. Source heating needs to be optimized for each analytical determination to ensure the production of the maximum amount of analyte ions in the source. Practically the use of heated nitrogen blown into the source housing is normally employed to aid desolvation in Crawford Scientific 4

6 high flow electrospray source designs. Perkin Elmer (Sciex) (Applied Biosystems, Foster City, CA), introduced one of the earliest instruments to incorporate pneumatically assisted electrospray in their IonSpray design. This design used a silica capillary within a stainless steel needle that was housed within a concentric PTFE tube. The design of the PTFE tube is such that gas flow rates at the capillary tip are around 200 m/s. When physically connecting the t-piece or HPLC column to the API interface housing, PEEK (polyetheretherketone) (0.1mm i.d.) tubing is preferred to fused silica tubing due to the possible adsorption of analyte species to residual silanol species on the inner surface of the silica capillary. Most modern instruments employ stainless steel or platinum capillaries for electrospray to avoid similar problems with fused silica that can adversely affect the quantitative response of the instrument. For low flow rate applications microelectrospray needles are available which consist of either coated or uncoated silica capillaries with a drawn tip to allow micro-droplet formation. [18] Filters in the needle assembly help to prevent blocking of the fine capillary tip (<0.5 m), and the coating of the needle helps electrical contact, negating the need for coaxial sheath liquids. Potentials of between 0.5 and 1kV are normally sufficient to produce efficient electrospray at eluent flow rates of less than 1 l/min. Nanoflow capillary instruments are also available which use fused silica capillaries with 20 m i.d. that can be used in the flow rate range nl/min. [19] The sample solution is fed to these capillaries via nitrogen-pressurised nanovials and can provide stable electrospray for numbers of hours with sample volumes of 1ml and less. Application areas of nanoflow electrospray systems include the study of proteins and other solutions where sample size is extremely limited. [2.3] Crawford Scientific 5

7 Sprayer Sampling Plate Configuration The position of the sprayer (capillary) relative to the ion sampling orifice within the sampling plate is of significant importance. By avoiding directly spraying at the sampling plate orifice the maintenance interval required for API interface types may be extended and the instance of charged droplet being sampled (as opposed to gas phase ions) is lowered hence increasing instrument response. [3] The core of the electrosprayed aerosol contains larger diameter droplets than in the extremities of the spray. It is expected that the efficiency of ion production (via ion evaporation), will be higher at the perimeter of the aerosol. The off-axis response is also more stable than the one observed when using on-axis. [3,4] In most modern instruments the sprayer position may be adjusted in one, two or three axes by means of micrometer screws that will allow optimisation relative to the sampling orifice. On-axis spray Off-axis spray Diagonal Crawford Scientific 6

8 The distance and potential difference between the tip of the capillary and the sampling plate determine the electric field that creates the electrospray and will influence the performance of the spray. Optimisation of the sprayer position and capillary voltage are interrelated and should be optimised empirically together. Pepperpot (FISONS-Micromass) Cross Flow (Waters) Crawford Scientific 7

9 Perhaps the most important practical advance in sprayer positioning has come with the introduction of orthogonal source design, where the sprayer is positioned orthogonally to the sampling orifice. This design has several advantages including, reduced down-time of the source with decreased coating of the source elements, sampling of fewer charged droplets relative to ions and the ability to tolerate higher flow rates. In orthogonal spray, neutrals and non-volatile materials collide with a plate perpendicularly located to the spray axis. Orthogonal design can be combined with a second orthogonal extraction (LCZ or Z-Spray from Waters) or a with an off-axis extraction, such as in the aqa (Thermo-Finnigan) disposition. [5] Orthogonal designs allow the use of eluent flow rates up to 1ml/min with electrospray and gives the advantage that eluent systems containing non-volatile buffers may be used for extended numbers of samples before source cleaning is required. Orthogonal sources are available from several other manufacturers. Crawford Scientific 8

10 Cluster Ions Polar molecules in the gas phase (water, solvent and eluent molecules), tend to form clusters with ions. [23] Cluster ions will appear in the mass spectrum and they are often so large that they are far outside the detectable mass range of a typical quadrupole ( Da): X n H O X H O) 2 ( 2 n Cluster ions may collide with the source elements in the early stages of the spectrometer and give rise to ion bursts which can result in noisy baselines in the Total or Selected Ion Chromatogram (TIC or SIC). One commonly applied solution to this problem allows ions to pass into the sampling orifice but excludes water vapour and other neutral species from the entrance to the vacuum system. This is achieved by forcing ions and neutrals in the opposite direction by the action of an electric field and/or a flow of dried gas. i Cluster formation Crawford Scientific 9

11 Cluster ions passing into the vacuum region of the source may be de-clustered by collision with rapidly moving background gas molecules, imparting enough energy to break the hydrogen bonds between the ion and the solvating cluster molecules. Increasing the temperature and the potential difference in the first vacuum region of the spectrometer in order to accelerate ions will both assist with ion declustering. Prevention of Cluster Ion Sampling A Sciex (Applied Biosystems, Foster City, CA), API source is shown. The region between the interface plate and the sampling orifice plate is continuously flushed with dry nitrogen that flows into the API source as well as into the vacuum region of the spectrometer. The gas flow into the source helps to repel water, neutrals and other potential contaminants, such as dirt and buffer salts, away from the sampling orifice, increasing the intervals between source maintenance. Ions in the region of the sampling orifice are driven into the vacuum region by the gas flow combined with a 600V potential difference between the interface plate and the sampling plate. The nitrogen employed in this sense is often referred to as a Curtain Gas. [6,7] i Gas curtain As the nitrogen curtain gas and ions undergo expansion into the nozzle-skimmer region of the source, significant cooling occurs. If the source and curtain gas are not heated, there is a significant possibility of cluster ion formation occurring in this region. Therefore most modern sources use both heated drying and curtain gas. A method preventing cooling effects experienced through expansion and the re-formation of cluster ions uses a heated transfer tube. [25] This device will preheat the mixture of ions and neutrals prior to expansion using a heated tube of approximately 20 cm long ( o C). A further advantage of this type of device is the ability to generate ions (via ion evaporation) from droplets that may be sampled into the nozzle-skimmer region. An Agilent Technologies (Palo Alto, CA, USA), source is shown which employs an ion transfer tube with metallised ends (also known as a 'dielectric capillary'), that allows the application of an accelerating voltage that can be used to promote in-source dissociation of cluster ions. Crawford Scientific 10

12 Orthogonal source with dielectric capillary (heated transfer tube) Crawford Scientific 11

13 Source Cleaning Contamination of the sampling orifice or tube can prove to be detrimental to the performance of the instrument, in some cases leading to very frequent source cleaning when dealing with samples in dirty matrices or when using non-volatile solvents and buffers. The layer of contamination inside an orifice or tube will attract a build up of charge that can effectively stop the passage of ions, while the flow of neutrals is not affected, thus significantly decreasing the instrument response. Cleaning regimes will differ for instruments from different manufacturers but may include some physical abrasion (aluminium powder is a popular choice) and / or wipe cleaning of the sampling cone and other source components followed by sonication in a range of solvents matched to the polarity of the contaminants (hexane, acetone and methanol are all popular choices for source cleaning). i Dirty layer formation The use of curtain gas, off axis spray and/or orthogonal spray can significantly increase the interval between source cleaning, by directing the deposition of contaminant species away from the sampling plate, orifice or ion transfer tube. Crawford Scientific 12

14 Ion Optics Ring Electrode In the nozzle-skimmer region of the mass spectrometer ions are transported in the free-jet expansion of gas, some of which will pass through the skimmer into the higher vacuum region of the mass spectrometer. Focusing of the ions into a narrow beam is not possible due to the effects of free jet expansion, which tends to direct the ions and entrained neutrals and gases into a barrel type shockwave, away from the axis of the spectrometer. However, ions may be forced to remain closer to the axis of the beam if a tube or ring lens is placed between the nozzle (sampling plate) and skimmer. [8] A voltage applied to the ring will reduce the spread of ions (but not water vapour or neutrals), away from the axis with a subsequent increase in instrument sensitivity (mainly due to a reduction in the noise). i Ring electrode operation Crawford Scientific 13

15 Ion Bridges In the high vacuum region of the spectrometer after the skimmer lenses, the pressure is still not low enough to focus ions using traditional ion lenses. In most cases the ions are focused using a radio frequency (RF) only hexapole or octapole ion bridge, that guides the ion beam into the analyzing quadrupole system. Ion bridges work on a similar principle to a quadrupole mass analyser but operate with radio frequency potential only. The quadrupole DC component is removed, and the application of a radio frequency potential allows ions of all m/z values to pass from one region of the spectrometer to another. [9] The multipole ion bridge has the advantage of slightly focusing the ion beam, therefore increasing transmission as well as allowing the removal of a significant amount of neutral species which are not held within the multipole ion bridge, so increasing signal to noise ratio. i Ion bridges operation Collision Induced Dissociation Classical API spectra tend to show very little fragmentation due to the 'soft' nature of the ionisation processes. That is, during the formation of ions, the analyte molecules do not receive enough energy to break the intra-molecular bonds. However, fragmentation may easily be induced in one of the higher-pressure regions and structural information can be gathered. Acceleration of ions between the sampling orifice and the skimmer, or between the skimmer and the RF-only multipole results in collisions of ions with the background gas. This process is known by various different names depending upon the instrument manufacturer and includes 'in-source collision induced dissociation (CID)', 'nozzleskimmer fragmentation', 'cone-voltage fragmentation', etc. By increasing the potential difference between the skimmer and the quadrupole V(S-Q) or between the nozzle and skimmer V(N-S)), the energy imparted to the analyte molecule through increasing frequency and energy of collisions can be enough to cause intra-molecular bonds to be broken and for fragmentation to occur. [10,11] In the case of larger molecules, such as peptides and proteins, the excess energy can often be absorbed in several vibrational modes and high potential differences are required to fragment these kind of molecules. Crawford Scientific 14

16 Fragmentation i CID working principle The major advantage of this technique for the production of spectra containing a greater amount of structural information is its simplicity. Only one voltage needs to be adjusted and there is no need for switching or adjustment of collision gases and no retuning of ion optics. It is possible to test the degree of CID using probes. [11] In negative ion mode the degree of CID can be estimated using the drug naproxen that normally shows the [M-H] - ion as the only at m/z 229. With small nozzle-skimmer potential differences the naproxen molecule readily loses CO 2 giving rise to a base peak at m/z 185. In the positive ion mode the [M+H] + ion of dibutyl phthalate at m/z 279 can be used as a test ion. Fragmentation to m/z 149 takes place readily even under mild CID conditions, making declustering of ions impossible with these labile sample ions. i Mass spectrum of naproxen (used to estimate the degree of CID in negative ion mode) Crawford Scientific 15

17 Ion Declustering Collision Induced Dissociation can also be used to improve the baseline noise and increase the signal-to-noise ratio in LC-MS experiments. [9,12] When ions pass through the sampling orifice into the vacuum region the background density of neutral gas ions falls rapidly. If ions moving in a low-density gas are accelerated by the nozzle-skimmer region mild-cid may be affected, which will be enough to cleave the hydrogen bonds inside the ion-water or ion-neutral gas clusters. Further, heating of the ion clusters, which occurs during collisions, will also aid desolvation of the cluster ions. Moderate acceleration of clusters is effective and widely used to decluster ions, leading to a reduction in baseline noise and the numbers to cluster ions detected. A disadvantage of this type of approach is the moderate scattering of the ion beam that is associated with ion-neutral gas collisions that may lead to a small reduction in the numbers of ions passing through the skimmer element and into the mass analyser (i.e. reduced spectrometer transmission). Declustering i Ion declustering process by using CID References 1. A. P. Bruins, T. R. Covey and J. D. Henion, Anal. Chem. 59, (1987), M. R. Emmett and R. M. Caprioli. J. Am. Soc. Mass Spectrom. 5, (1994), M. S. Wilm and M. Mann. Anal. Chem. 68, (1996), A. P. Bruins, T. R. Covey and J. D. Henions. Anal. Chem. 59, (1987), K. Hiraoka. Rapid Commun. Mass Spectrom. 6, (1992), J. Abian. The Coupling of Gas and Liquid Chromatography with Mass Spectrometry. J. Mass Spectrom. 34, (1999), C. K. Meng and J. B. Fenn. Org. Mass. Spectrom. 26, (1991), C. M. Whitehouse, R. N. Dreyer, M. Yashamita, J. B. Fenn. Anal. Chem. 57, (1985), I. C. Mylchreest, M. E. Hail and J. R. Herron. United States Patent 5, 157, 260, October 20, M. E. Hail and I. C. Mylchreest. Presented at the 41 st ASTM Conference on Mass Spectrometry and Allied Topics. May 31 Jun 1, 1993, San Francisco, CA, R. D. Smith, J. A. Loo, C. J. Barinaga, C. G. Edmonds and H. R. Hudspeth. J. Am. Soc. Mass Spectrom. 1, (1990), R. D. Smith and C. J. Barinaga. Rapid Commun. Mass Spectrom. 4, (1990), A. P. Bruins in Electrospray Ionisation Mass Spectrometry R. B. Cole [ed.] John Wiley and Sons Int. 1997, Crawford Scientific 16

18

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

M M e M M H M M H. Ion Sources

M M e M M H M M H. Ion Sources Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could

More information

Quattro Micro - How does it work?

Quattro Micro - How does it work? Quattro Micro - How does it work? 1 Introduction This document is designed to familiarise you with the principles behind how the Quattro Micro works. The level of this document is designed as Level One

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Chapter 1. Introduction

Chapter 1. Introduction Chapter 1. Introduction 1-1 1.1. Overview In the past twenty years, charged droplets and strong electric fields have quietly revolutionized chemistry. In combination with an atmospheric-sampling mass spectrometer,

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

Lecture 15: Introduction to mass spectrometry-i

Lecture 15: Introduction to mass spectrometry-i Lecture 15: Introduction to mass spectrometry-i Mass spectrometry (MS) is an analytical technique that measures the mass/charge ratio of charged particles in vacuum. Mass spectrometry can determine masse/charge

More information

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Chemistry Instrumental Analysis Lecture 34. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry

More information

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Selecting an LC/MS Interface Becky Wittrig, Ph.D. Selecting an LC/MS Interface Becky Wittrig, Ph.D. RESTEK CORPORATION LC/MS Interfaces I. Background of LC/MS I. Historical Perspective II. Reasons for use II. Interfaces I. Transport devices II. Particle

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

API Optimization, Source Design, and Other Considerations of LC/MS

API Optimization, Source Design, and Other Considerations of LC/MS API Optimization, Source Design, and Other Considerations of LC/MS Instrument Design Considerations: The Atmosphere-Vacuum Interface and Pumping System The hows and whys of the vacuum? We ve got to get

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n 20.2 Ion Sources electron ionization produces an M + ion and extensive fragmentation chemical ionization produces an M +, MH +, M +, or M - ion with minimal fragmentation MALDI uses laser ablation to transfer

More information

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer 38 Current Trends in Mass Spectrometry November 6 Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer This article demonstrates the improved

More information

HPLC solvents and mobile phase additives

HPLC solvents and mobile phase additives HPLC solvents and mobile phase additives The following is intended to be a guide for the choice of solvent and mobile phase additives. Of course the choice of solvents for LC will be dictated primarily

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

The Theory of HPLC. Ion Pair Chromatography

The Theory of HPLC. Ion Pair Chromatography The Theory of HPLC Ion Pair Chromatography i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual.

More information

Mass Spectrometry. Fundamental LC-MS. Atmospheric Pressure Chemical Ionisation (APCI)

Mass Spectrometry. Fundamental LC-MS. Atmospheric Pressure Chemical Ionisation (APCI) Mass Spectrometry Fundamental LC-MS Atmospheric Pressure Chemical Ionisation (APCI) i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance

More information

Mass spectrometry.

Mass spectrometry. Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged

More information

Interface (backside) & Extraction Lens

Interface (backside) & Extraction Lens Plasma Interface Interface (backside) & Extraction Lens Extraction Lens (-2000 volts) ION OPTICS Tip of the sampler cone is positioned to be in the region of maximum ionization Ions no longer under control

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Choosing the metabolomics platform

Choosing the metabolomics platform GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation

More information

Extrel Application Note

Extrel Application Note Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

BIOINF 4399B Computational Proteomics and Metabolomics

BIOINF 4399B Computational Proteomics and Metabolomics BIOINF 4399B Computational Proteomics and Metabolomics Sven Nahnsen WS 13/14 3. Chromatography and mass spectrometry Overview Recall last lecture Basics of liquid chromatography Algorithms to predict and

More information

MS-based proteomics to investigate proteins and their modifications

MS-based proteomics to investigate proteins and their modifications MS-based proteomics to investigate proteins and their modifications Francis Impens VIB Proteomics Core October th 217 Overview Mass spectrometry-based proteomics: general workflow Identification of protein

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information

Technical Note. Introduction

Technical Note. Introduction Technical Note Analysis and Characterization of Psilocybin and Psilocin Using Liquid Chromatography - Electrospray Ionization Mass Spectrometry (LC-ESI-MS) with Collision-Induced-Dissociation (CID) and

More information

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) By Arianne Bazilio & Jacob Weinrich December 2012 Contents Introduction... 2 Sample Introduction... 3 Torch... 4 Interface... 6

More information

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray Edward W. Sheehan 1, Ross C. Willoughby 1, Robert Classon 2, and Dan Dodgen 2 1 Chem-Space Associates, Inc., Pittsburgh,

More information

Electrospray ionization (ESI) has enabled the study

Electrospray ionization (ESI) has enabled the study Ion Funnels for the Masses: Experiments and Simulations with a Simplified Ion Funnel Ryan R. Julian, Sarah R. Mabbett, and Martin F. Jarrold Department of Chemistry, Indiana University, Bloomington, Indiana,

More information

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS

More information

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun

More information

Supporting Information. Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency

Supporting Information. Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency Supporting Information Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency Ta-Hsuan Ong, Ted Mendum, Geoff Geurtsen, Jude Kelley, Alla Ostrinskaya, Roderick Kunz Chemical,

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

CHROMATOGRAPHY AND MASS SPECTROMETER

CHROMATOGRAPHY AND MASS SPECTROMETER 22 CHROMATOGRAPHY AND MASS SPECTROMETER 22.1 INTRODUCTION We know that the biochemistry or biological chemistry deals with the study of molecules present in organisms. These molecules are called as biomolecules

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

Dry Deposition of Biogenic Terpenes via Cationic. Oligomerization on Environmental Aqueous Surfaces

Dry Deposition of Biogenic Terpenes via Cationic. Oligomerization on Environmental Aqueous Surfaces Supporting Information for Dry Deposition of Biogenic Terpenes via Cationic Oligomerization on Environmental Aqueous Surfaces Shinichi Enami* The Hakubi Center for Advanced Research, Kyoto University,

More information

CHAPTER-2 Formation of Adenine from CH 3 COONH 4 /NH 4 HCO 3 the Probable Prebiotic Route for Adenine

CHAPTER-2 Formation of Adenine from CH 3 COONH 4 /NH 4 HCO 3 the Probable Prebiotic Route for Adenine CHAPTER-2 Formation of Adenine from CH 3 COOH 4 /H 4 HCO 3 the Probable Prebiotic Route for Adenine ITRODUCTIO Mimicking the probable primitive earth conditions, a number of experiments for the generation

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

Liquid chromatography^tandem mass spectrometric analysis of surface and waste. (LC)^atmospheric pressure chemical ionisation

Liquid chromatography^tandem mass spectrometric analysis of surface and waste. (LC)^atmospheric pressure chemical ionisation trends in analytical chemistry, vol. 19, no. 4, 2000 249 Liquid chromatography^tandem mass spectrometric analysis of surface and waste water with atmospheric pressure chemical ionisation I: instrumentation

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Vacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular

Vacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular Vacuum Technology Vacuum Pumps Two general classes exist: Gas transfer physical removal of matter Mechanical, diffusion, turbomolecular Adsorption entrapment of matter Cryo, sublimation, ion Mechanical

More information

Introduction to LC-MS

Introduction to LC-MS Wednesday April 5, 2017 10am Introduction to LC-MS Amy Patton, MS Laboratory Manager, Pinpoint Testing, LLC Little Rock, AR DESCRIPTION: Amy Patton, laboratory manager for Pinpoint Testing, will begin

More information

LC MS analysis of metabolites. Basis of Chromatography

LC MS analysis of metabolites. Basis of Chromatography GBS 724 LC MS analysis of metabolites Stephen Barnes, PhD 4 7117; sbarnes@uab.edu Basis of Chromatography A moving (mobile) phase passes over an inert, stationary phase The compounds differentially interact

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

Potential sources of contamination on Mass Spectrometers and suggested cleaning procedures

Potential sources of contamination on Mass Spectrometers and suggested cleaning procedures Potential sources of contamination on Mass Spectrometers and suggested cleaning procedures Purpose: This document provides information on potential sources of background contamination commonly encountered

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision

More information

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Mass Spectrometry. Fundamental LC-MS. Mass Analysers Mass Spectrometry Fundamental LC-MS Mass Analysers i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Ionization Techniques Part IV

Ionization Techniques Part IV Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of

More information

Chapter 8. Online field-induced droplet ionization. mass spectrometry probes heterogeneous reaction

Chapter 8. Online field-induced droplet ionization. mass spectrometry probes heterogeneous reaction Chapter 8. nline field-induced droplet ionization 8-1 mass spectrometry probes heterogeneous reaction chemistry at the air droplet interface: soft-sampling the products of the oleic acid ozone reaction

More information

New Approaches in the Analysis of Amadori Compounds

New Approaches in the Analysis of Amadori Compounds 7 TH Wartburg Symposium on Flavor Chemistry and Biology Eisenach, April 21 st - April 23 rd, 2004 New Approaches in the Analysis of Amadori Compounds T. Davidek, I. Blank, K. Kraehenhuehl, J. Hau and S.

More information

The Importance of Field Strength in the Low Field Portion of a Differential Ion Mobility Spectrometry Waveform

The Importance of Field Strength in the Low Field Portion of a Differential Ion Mobility Spectrometry Waveform The Importance of Field Strength in the Low Field Portion of a Differential Ion Mobility Spectrometry Waveform Jesus I. Martinez-Alvarado, Brandon G. Santiago, Gary L. Glish University of North Carolina,

More information

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC CHROMATOGRAPHY Laboratory technique for the Separation of mixtures Chroma -"color" and graphein

More information

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Application Note Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

Characterization of typical chemical background interferences in atmospheric pressure ionization liquid chromatography-mass spectrometry y

Characterization of typical chemical background interferences in atmospheric pressure ionization liquid chromatography-mass spectrometry y RAPID CMMUNICATINS IN MASS SPECTRMETRY Rapid Commun. Mass Spectrom. 2006; 20: 3145 3150 Published online in Wiley InterScience (www.interscience.wiley.com) DI: 10.1002/rcm.2715 Characterization of typical

More information

Technical know-how in thermal analysis measurement

Technical know-how in thermal analysis measurement Technical articles Technical know-how in thermal analysis measurement Evolved gas analysis by thermogravimetry-differential thermal analysis-mass spectrometry (TG-DTA-MS) technique Kazuko Motomura* and

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson

More information

Mass Spectrometry (MS)

Mass Spectrometry (MS) Mass Spectrometry (MS) Alternative names: Mass spectrometric (selective) detector (MSD) Spectrometry - methods based on interaction of matter and radiation Mass spectrometry - method based on formation

More information

Fast Atom Bombardment (FAB)-MS

Fast Atom Bombardment (FAB)-MS Fast Atom Bombardment (FAB)-MS FAB-MS was developed by Barber and coworkers in the early 1980s and soon became a very powerful technique for desorption ionization of thermally labile molecules. In FAB

More information

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Particle Position Relative Mass Relative Charge Proton Nucleus 1 +1 Neutron Nucleus 1 0 Electron Orbitals 1/ Atomic Symbol

Particle Position Relative Mass Relative Charge Proton Nucleus 1 +1 Neutron Nucleus 1 0 Electron Orbitals 1/ Atomic Symbol Atomic Structure Details of the three Sub-atomic (fundamental) Particles Particle Position Relative Mass Relative Charge Proton Nucleus 1 +1 Neutron Nucleus 1 0 Electron Orbitals 1/1840-1 Behaviour of

More information

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS. ESS The RGA freenotes Theory page 1 of 14 RGA Theory Notes Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

More information

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD Chapter 4 DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD 4.1 INTRODUCTION Sputter deposition process is another old technique being used in modern semiconductor industries. Sputtering

More information

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading

More information

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP Application Note MBMS for Preformed Ions, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster Presented at 45th ASMS Conference on Mass Spectrometry, June 1-5, 1997) In order to accurately characterize a plasma

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Chapter 6. Neutral droplets in high electric fields as a. source of ions. Introduction to field-induced droplet

Chapter 6. Neutral droplets in high electric fields as a. source of ions. Introduction to field-induced droplet Chapter 6. Neutral droplets in high electric fields as a source of ions. Introduction to field-induced droplet ionization mass spectrometry 6-1 Adapted from Grimm, R. L. and Beauchamp, J. L. J. Phys. Chem.

More information

Supporting Information

Supporting Information Supporting Information Highly Efficient Ionization of Phosphopeptides at Low ph by Desorption Electrospray Ionization Mass Spectrometry Ning Pan, a, b Pengyuan Liu, b Weidong Cui, c Bo Tang, a Jingmin

More information

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap Pro LC/MS/MS Accurate, High-Throughput Protein Identification

More information

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Jonathan Williams Waters Corporation, Milford, MA, USA A P P L I C AT

More information

Biochemistry. Biochemical Techniques HPLC

Biochemistry. Biochemical Techniques HPLC Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know

More information

Characteristics and origins of common chemical noise ions in negative ESI LC MS

Characteristics and origins of common chemical noise ions in negative ESI LC MS Research Article Received: 9 November 2010 Accepted: 17 April 2011 Published online in Wiley Online Library: 2011 (wileyonlinelibrary.com) DOI 10.1002/jms.1924 Characteristics and origins of common chemical

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 6 Spectroscopic Techniques Lecture - 6 Atomic Spectroscopy

More information

Application Note. Edgar Naegele. Abstract

Application Note. Edgar Naegele. Abstract Agilent MassHunter Fast computer aided analysis of LC/ESI-TF data from complex natural product extracts Part 1: Analysis of Agilent 6210 TF data with the Molecular Feature Extractor in MassHunter Workstation

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

UPLC Intact MASS Analysis Application Kit

UPLC Intact MASS Analysis Application Kit (Part Number: 176001519) CONTENTS I. INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMN III. CONNECTING THE COLUMN TO LCT PREMIER

More information

1) In what pressure range are mass spectrometers normally operated?

1) In what pressure range are mass spectrometers normally operated? Exercises Ionization 1) In what pressure range are mass spectrometers normally operated? Mass spectrometers are usually operated in the high vacuum regime to ensure mean free paths significantly longer

More information

11 Patent Number: 5,818,041 Mordehai et al. (45) Date of Patent: *Oct. 6, MASS SPECTROMETER SYSTEMAND 56) References Cited

11 Patent Number: 5,818,041 Mordehai et al. (45) Date of Patent: *Oct. 6, MASS SPECTROMETER SYSTEMAND 56) References Cited United States Patent 19 USOO5818041A 11 Patent umber: Mordehai et al. (45) Date of Patent: *Oct. 6, 1998 54 MASS SPECTROMETER SYSTEMAD 56) References Cited METHOD FORTRASPORTIG AD AALYZIG OS U.S. PATET

More information

Agilent ESI and APCI sources: for polar to non-polar compounds

Agilent ESI and APCI sources: for polar to non-polar compounds 1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent ESI and APCI sources: for polar to non-polar compounds Nebulizer Pressure Corona current Nebulizer Pressure Vaporizer Vcap Vcap Drying Gas

More information