Corona Veo the Next Generation Charged Aerosol Detector

Size: px
Start display at page:

Download "Corona Veo the Next Generation Charged Aerosol Detector"

Transcription

1 Corona Veo the Next Generation Charged Aerosol Detector Presenters Name Title Date 1 The world leader in serving science

2 Contents Introduction to Charged Aerosol Detection How Charged Aerosol Technology Works Key Features/Product Variations of Corona Veo Detectors Detection of Analytes in Different Matrices Inverse Gradient Solution for Uniform Response New Approach to a Generic Analytical Method Wrap-up Questions & Discussion 2

3 Introduction to Charged Aerosol Detection Used to quantitate any non-volatile and most semi-volatile analytes with LC Provides consistent analyte response independent of chemical structure Neither a chromophore, nor the ability to ionize, is required for detection Dynamic range of over four orders of magnitude from a single injection (sub-ng b to µg quantities on column) Mass sensitive detection provides relative quantification without the need for reference standards Compatible with gradient conditions for HPLC, UHPLC, and Micro LC Comparison of Charged Aerosol Detection to UV and MS 3

4 Charged Aerosol Detection How It Works The liquid eluent from the LC column enters the detector (1) where it undergoes nebulization by combining with a concentric stream of nitrogen gas or air (2). The fine droplets are carried by bulk gas flow to the evaporation sector (3) where desolvation occurs to form particles, while any larger droplets are drained to waste (4). Signal is directly proportional to the analyte quantity The dry particles exit from evaporation (5) and are combined with another gas stream that first passes over a high voltage Corona charger (6). The charged gas then mixes Any high h mobility species are removed by an ion trap (8) while with the dry particles, where the remaining charged particles pass to a collector, where the excess charge transfers to the charges from the passing particles are measured with a very particle s surface(7). sensitive electrometer (9). The resulting signal is conveyed to a chromatographic data software for quantitation (10). 4

5 Particle Charging for Charged Aerosol Detection Analyte particles with charged surface Charges remain on particles surface. Particles remain intact t and do not ionize. The more surface area, the more charge is carried by the particle Dry analyte particles Charged nitrogen gas Mixing Chamber 5

6 CoronaVeo Charged Aerosol Detector A novel, near universal, mass sensitive detector for routine LC determinations of any non-volatile and many semi-volatile analytes Uses the fastest growing g and most rapidly adopted universal LC detection technology since diode array detection (PDA or DAD) Provides unbiased detection for a wide variety of analyte classes New design incorporates unique, proprietary technologies only available from Thermo Fisher Scientific Thermo Scientific Dionex Corona Veo Charged Aerosol detector 6

7 Corona Veo Detector What's New? This entirely new detector incorporates many design and performance improvements: Radically new Thermo Scientific Dionex FocusJet concentric nebulization system improves sensitivity and precision All new aerosol conditioning and evaporation schemes widen the scope of applications to include capillary and micro LC, as well as UHPLC Usability and serviceability are enhanced by countless improvements, many of which came from our customers 7

8 FocusJet Concentric Nebulizer Technology The advanced concentric nebulizer design ensures uniform aerosol atomization pattern for superior droplet size selection Exchangeable design gives users ready access to improve serviceability and increase uptime Inlet accepts conventional male fittings, including Viper fingertight fittings for virtually dead-volume-free connections FocusJet Concentric Nebulizer Thermo Scientific Dionex Viper Fitting 8

9 FocusJet Nebulizer and Spray Chamber The inlet liquid and nebulization gas streams combine coaxially into an aerosol spray of uniform droplets at the nebulizer tip The fine droplets are transported into the heated evaporation sector Any large droplets fall and are expelled by a precision micro-pump To Evaporation Sector Impactor Coaxial N 2 gas flow Aerosol Concentric Nebulizer Fused silica capillary Spray chamber FocusJet Concentric Nebulizer Tip Inlet Gas Cross-flow In (N 2 ) Nebulizer Active drain pump 9

10 Standard Features Corona Veo Detector Exchangeable nebulizer the FocusJet nebulizer/spray chamber system improves performance Inlet flow capability accepts flows from 0.20 to 2.0 ml/min Front access to all fluidic connections improves serviceability and maximizes uptime Touch screen user interface provides clear operating commands and displays practical diagnostic messaging Revised fluid waste management a non-pressurized waste collection scheme simplifies operation Evaporation control two selectable evaporation temperatures permit stable operation with the typical range of eluents Active detector thermostatting to maintain stable performance and improve reproducibility Expanded response linearity through integration of a selectable Power function 10

11 Advanced Features Corona Veo RS Detector Includes all the features of the standard Corona Veo detector with the addition of: Programmable evaporation temperature a flexible tool for further optimizing analyte response over an expanded range of eluents (ambient +5 to +100 o C) Extended inlet flow capability accepts inlet flow rates for HPLC, UHPLC, capillary, and micro LC (as low as 10 µl/min.) Electronic gas pressure regulator (EGR) ensures optimized performance over the entire flow rate range Integrated stream switching (SSV) for automated flow diversion to alternate devices or for peak elimination and collection Advanced digital signal processing for fast data acquisition speeds necessary with UHPLC (up to 200 Hz with Thermo Scientific Dionex Chromeleon 7.2 Chromatography Data System) 11

12 Expanding the Range of Non-Linear Response Non-linear response with charged aerosol detection Linear response utilizing the power function Use a Power Function Value (PFV) to expand the linear calibration range for a non-linear aerosol detector Enhanced calibration using the Power Function 12

13 Applying Power Function to a Corona Veo Detector Improves visualization of the linear calibration range Acts to improve accuracy of linear interpolation at each end Has no effect on the limit of detection (LoD) Extend the linear calibration range using the Power Function 13

14 Product Variations of Corona Veo Detectors Model/Feature Corona Veo Detector Corona Veo RS Detector Flow Rate Range ml/min ml/min Evaporation Temp 35 C or 50 C Ambient +5 C to +100 C Integrated Stream Switching Valve (SSV) Not offered 6-port, 2-position micro valve (TTL controlled) Internal Gas Regulation User Adjustable Electronic control Data Acquisition Rate Up to 100 Hz Up to 200 Hz (with CM 7.2) Inlet Gas Pressure (reqd) Standalone Interface Liquid Waste Management Enhanced Linearity Analog Signal Output (0 1 VDC) bar (70 80 psig; kpa) Integrated color LCD touch screen (rationalized GUI) Non-pressurized (uses general LC waste stream) Via Power Function (User applied) Installable Option 14

15 Electronic Gas Regulation Corona Veo RS Detector Internal Electronic Gas Regulator Provides ultra-precise gas pressure to ensure efficient nebulization and stable particle flow through the entire flow path Ensures optimum detector performance over the entire inlet flow range of the Corona Veo RS detector (10 µl 2.0 ml per min.) Both analytical and micro operation modes are selectable via remote control from a chromatography data system 15

16 Integrated Stream Switching Corona Veo RS Detector Stream Switching Valve (SSV) Simplify system operation by enabling access to alternate devices without the need to unplumb/replumb modules (UV, MS, etc.) Diverts flow to waste in event of gas or pump flow errors, to protect against damage Improve chromatography results for complex samples by allowing elimination of harmful salts or collection of unretained peaks 16

17 Detection of a Broad Range of Analytes in Different Matrices 17

18 Glycan Analysis for Bovine Fetuin Separation of Oligosaccharide Alditols Column: Thermo Scientific GlycanPac AXH μm, mm Mobile phase A: 80% Acetonitrile Mobile phase B: 80 mm Ammonium formate, ph 4.4 Gradient: 2.5 % B to 25% B from 1 to 40 min Flow rate: 0.4 ml/min Inj. volume: 5 μl Col. temp: 30 o C Evap. temp: 50 o C Native Glycans 18

19 Determination of Adjuvants Analysis of Plant Saponins Column: Thermo Scientific Hypersil GOLD PFP 1.9 um, mm Mobile phase A: 0.1% Formic acid in water Mobile phase B: 0.1% Formic acid in 10:90 acetonitrile: reagent alcohol Gradient: 35% B to 83%B in 6 min to 90% B in 10 min Flow rate: 0.46 ml/min Inj. volume: 2 μl Col. temp: 45 o C Evap. temp: 50 o C 210 nm 19

20 Determination of Carbohydrates in Juice Analysis of Simple Sugars Column: Amino, 3 μm, mm Mobile phase: Acetonitrile:water (92:8) Flow rate: 0.8 ml/min Inj. volume: 2 μl Col. temp; 60 o C Post-column temp: 25 o C Evap. temp: 75 o C Sample preparation: Add 20 ml of 85% ACN to 1 gram juice Simplified sample preparation Dilute-and-shoot method 20

21 Extracts from Natural Products Analysis of Hoodia Plant Extracts Column: Mobile phase A: Mobile phase B: Gradient: Flow rate: Inj. volume: Col. temp: Evap. temp: Thermo Scientific Accucore C μm, mm Water Acetonitrile 35%B to 90%B in 10 min 0.5 ml/min 2 μl 40 o C 35 o C Hoodigosides are oxypregnane steroidal glycosides abundant in Hoodia gordonii 21

22 Separation of Herbal Extracts Analysis of Milk Thistle Extracts Column: Thermo Scientific Acclaim 120 C µm, mm Mobile phase A: 80:20 (v/v) 3mM Ammonium formate, 0.3% formic Acid:methanol Mobile phase B: 20:80 (v/v) 3mM Ammonium formate, 0.3% formic Acid:methanol Gradient: 15% B to 45% B in 4.4 min Flow rate: 0.5 ml/min Inj. volume: 2 μl Col. temp: 40 C Evap. temp: 35 C 22

23 Characterization of Algae-Based Biofuels Analysis of Algal Oils Column: Accucore C18, 2.6 μm, mm Mobile phase A: Methanol/water/acetic acid (600:400:4) Mobile phase B: Tetrahydrofuran/acetonitrile (50:950) Mobile phase C: Acetone/acetonitrile (900:100) Gradient: Time FlowRate %A %B %C (min) (ml/min) Flow rate: 1.0 ml/min Inj. volume: 2 μl Col. temp: 40 o C Evap. temp: 40 o C 23

24 Analysis of Industrial Descaling Agents Measurement of Polyacrylic Acid In Boiler Water Column: Acclaim SEC-300, 5 μm, mm Mobile phase A: Acetonitrile Mobile phase B: Water Isocratic: 10:90 Flow rate: 0.35 ml/min Inj. volume: 100 μl Col. Temp: 30 o C Evap. temp: 55 o C 24

25 Active Ingredient Composition Analysis of Gentamicin standard (200 μg/ml) Column: Acclaim RSLC PolarAdvantage II, 2.2 μm, mm Mobile phase A: 0.025:95:5 HFBA:water:acetonitrile Mobile phase B: 0.3:95:5 TFA:water:acetonitrile Gradient: - 0to15min1to 1.5min,1 10%B to 7min,10 to 100%B - 7 to 10min,100%B - 4 min. pre-injection equilibration Flow rate: 0.45 ml/min Inj. volume: 1 μl Col. temp: 15 o C Evap. temp: 80 o C 25

26 Impurity Testing at Low Flow Rate Analysis of Amikacin Sulfate USP (1.0 mg/ml) Column: C18,.2μm, mm Mobile phase A: 0.2% Pentafluoropropionic acid in water Mobile phase B: Methanol Gradient: Time %A %B (min) Flow rate: 50 µl/min Inj. volume: 0.5 μl Col. Temp: 55 o C Evap. temp: 35 o C 26

27 Determination of Drug Discovery Mass Balance Charged Aerosol UV Column: Acclaim 300 C18, 3 μm, mm Mobile phase A: 20 mm Ammonium acetate, ph 4.5 Mobile phase B: Acetonitrile Gradient: 2% B to 98% B in 30 min, Inverse Gradient Flow rate: 0.8 ml/min Inj. volume: 2 μl Col. temp: 30 o C Evap. temp: 35 o C 27

28 Formulation Testing Analysis of Diclofenac-Sodium Salt (1 mg/ml) Column: Mobile phase: Flow rate: Inj. volume: Col. temp: Evap. temp: Acclaim Trinity P1, 3μm, mm 75% Acetonitrile:25% 200mM Ammonium acetate ph ml/min 5 μl 30 o C 60 o C Measurement of Chloride Impurity 28

29 Inverse Gradient Solution for Uniform Response with Charged Aerosol Detection 29

30 Inverse Gradient Compensation for Uniform Response Conventional Gradient Elution Inverse Gradient Compensation 30

31 The Effects of Gradient and Mass on Calibration 7 Standard Gradient (Single Pump) Inverse Gradient (Dual Pump) Peak Area (CAD) Sulfanilamide Famotidine Perphanzine R² = R² = R² = R² = R² = R² = 1 Sulfanilamide Famotidine Perphanzine Mass on Column (ng) Mass on Column (ng) Inverse gradient extends the consistency of response 31

32 Solution for Uniform Response with Gradients A dual-gradient pump is the heart of this exclusive solution Inverse gradient kits with Viper connections provide easy installation Kit contains a media CD with eworkflows for Chromeleon CDS software Thermo Scientific Dionex UltiMate 3000 X2 Dual Gradient Pump Analytical Flow path Make-up Flow path Inverse Gradient Configuration A unique solution provided only by Thermo Scientific 32

33 A Radically New Approach to a Generic Method HILIC Excellent for polar molecules Reversed Phased Excellent for non- polar molecules Universal Charged Aerosol Detection 33

34 Combination of Composition and Flow Gradients Programmed gradients on both pumps mau RP+HILIC/UV UV method misses peaks 5 8 Resulting gradient on HILIC column min RP+HILIC/Charged Aerosol Additional peaks found, more Constant flow at 1.5 ml/min pa consistent response for others min 34

35 Next Steps Toward a True Generic Method The high organic mobile phase used in HILIC separations offer increased run speeds under UHPLC conditions. If volatile substances with chromophores are present, a UV-vis or DAD could be coupled before the charged aerosol detector to give complementary data. Combining this technique with LC-MS as an orthogonal detector would enable simultaneous confirmation of the identity of degradation products. RP-HILIC-UV-Charged Aerosol has great potential for generic screening 35

36 Wrap-Up A Corona Veo charged aerosol detector is the only universal LC detector that you need in your lab! Can provide consistent inter-analyte responses across a wide range analytes An ideal replacement for older, less capable HPLC detectors (RI, ELSD, etc.) An outstanding complementary detector to UV, DAD, and MS Inverse gradient techniques offer a unique solution for even better mass balance determinations Opens up exciting new possibilities for development of truly generic analytical methods 36

37 Thank You for Your Attention For a Cleaner, Healthier, Safer World 37

Pharmaceutical Analysis of API and Counter Ions in Complex Formulations in a Single Injection

Pharmaceutical Analysis of API and Counter Ions in Complex Formulations in a Single Injection Pharmaceutical Analysis of API and Counter Ions in Complex Formulations in a Single Injection Matthew Neely Pittcon Conference & Expo 2014 March 2-6, 2014 1 The world leader in serving science Outline

More information

Charged Aerosol Detection 101

Charged Aerosol Detection 101 Charged Aerosol Detection 101 Dr. Alexander Schwahn European Sales Support Expert for Biopharma Industry Thermo Fisher Scientific, Reinach, Switzerland The world leader in serving science Outline Introduction

More information

Discover what you re missing

Discover what you re missing Discover what you re missing Thermo Scientific Corona Veo Charged Aerosol Detector Hidden peaks revealed The analyte detection challenge No single liquid chromatography (LC) detector delivers ideal results.

More information

Charged Aerosol Detection See what other universal detectors are missing. Technical Guide

Charged Aerosol Detection See what other universal detectors are missing. Technical Guide Charged Aerosol Detection See what other universal detectors are missing Technical Guide Charged Aerosol Detection No single liquid chromatography (LC) detector delivers ideal results. Often, one analyte

More information

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners APPLICATION NOTE 21734 Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners Authors Aaron Lamb, Joanne Jones, Thermo Fisher Scientific, Runcorn, UK Keywords Environment, food

More information

No Chromophore? No Problem! Universal HPLC Detection. The world leader in serving science

No Chromophore? No Problem! Universal HPLC Detection. The world leader in serving science No Chromophore? No Problem! Universal HPLC Detection The world leader in serving science Topics Key concepts behind Thermo Scientific Charged Aerosol Detection (CAD) Universal Detection and Uniform Response

More information

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany Generic Method Approach for Pharmaceutical Drug Discovery and Development using Reversed-Phase Hydrophilic Interaction Liquid Chromatography with Universal Charged Aerosol Detection Frank Steiner, Michael

More information

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of

More information

Agilent 385-ELSD Evaporative Light Scattering Detector

Agilent 385-ELSD Evaporative Light Scattering Detector Agilent 8-ELSD Evaporative Light Scattering Detector For Low Molecular Weight Compounds Data Sheet Introduction Evaporative light scattering detectors are ideal for detecting analytes with no UV chromophore

More information

Agilent 385-ELSD Evaporative Light Scattering Detector

Agilent 385-ELSD Evaporative Light Scattering Detector Agilent 8-ELSD Evaporative Light Scattering Detector For Low Molecular Weight Compounds Data Sheet Introduction Evaporative light scattering detectors are ideal for detecting analytes with no UV chromophore

More information

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Mark Tracy, Xiaodong Liu, and Ian Acworth, Thermo Fisher Scientific, Sunnyvale, CA,

More information

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford,

More information

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection Ian Acworth, Bruce Bailey, Paul Gamache, and John Waraska Thermo Fisher Scientific, Chelmsford, MA, USA Introduction

More information

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview

More information

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution Marc Plante, Bruce Bailey, Christopher Crafts, Ian N. Acworth Thermo Fisher Scientific, Chelmsford,

More information

Enhancement of Linearity and Response in Charged Aerosol Detection

Enhancement of Linearity and Response in Charged Aerosol Detection Enhancement of Linearity and Response in Charged Aerosol Detection Christopher Crafts, Marc Plante, Bruce Bailey, Ian Acworth, Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: This work

More information

Varian 385-LC. ELSD for Low Molecular Weight Compounds

Varian 385-LC. ELSD for Low Molecular Weight Compounds D ATA S H E E T Varian 8-LC EVAPORATIVE LIGHT SCATTERING DETECTOR ELSD for Low Molecular Weight Compounds Evaporative light scattering detectors are ideal for detecting analytes with no UV chromophore

More information

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: Examples

More information

penta-hilic UHPLC COLUMNS

penta-hilic UHPLC COLUMNS penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and

More information

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation Technical Overview The Agilent InfinityLab D-LC Solution with Active Solvent Modulation Achieving Improved Resolution and Sensitivity for Challenging Combinations of Separation Conditions Author Sonja

More information

Varian 380-LC. ELSD for Routine HPLC and GPC Analysis. Key Benefits

Varian 380-LC. ELSD for Routine HPLC and GPC Analysis. Key Benefits D ATA S H E E T EVAPORATIVE LIGHT SCATTERING DETECTOR ELSD for Routine HPLC and GPC Analysis Evaporative light scattering detectors are ideal for detecting analytes with no UV chromophore as they do not

More information

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals Analytical Methods to Qualify and Quantify and ylated Biopharmaceuticals Christopher Crafts, Bruce Bailey, Marc Plante, Ian Acworth Thermo Fisher Scientific, Chelmsford, MA Application Note 1015 Key Words

More information

Agilent 1260 Infinity Evaporative Light Scattering Detector (ELSD)

Agilent 1260 Infinity Evaporative Light Scattering Detector (ELSD) Agilent 16 Infinity Evaporative Light Scattering Detector (ELSD) Data Sheet Introduction Evaporative light scattering detectors (ELSDs) are ideal for detecting analytes with no UV chromophore as they do

More information

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance David Thomas, Bruce Bailey, Marc Plante and Ian Acworth Thermo Fisher Scientific,

More information

Fast methods for the determination of ibuprofen in drug products

Fast methods for the determination of ibuprofen in drug products APPLICATION NOTE 779 Fast s for the determination of ibuprofen in drug products Authors Sylvia Grosse, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Germering, Germany Keywords Pharmaceutical,

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Marco Karsten, Bas, Dolman, Giovanni Maio, Frank Steiner, Holger Franz, Frank Arnold and Remco Swart LC Packings,

More information

for Acclaim Mixed-Mode HILIC-1 Column

for Acclaim Mixed-Mode HILIC-1 Column for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific Thermo Scientific Acclaim Trinity P2 Column Product Manual P/N: October 2013 Part of Thermo Fisher Scientific Product Manual for Acclaim Trinity P2 Columns Acclaim Trinity P2, 3µm, Analytical Column, 3.0

More information

penta-hilic UHPLC COLUMNS

penta-hilic UHPLC COLUMNS penta-hilic UHPLC COLUMNS Highly retentive, proprietary penta-hydroxy-ligand penta-hilic Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford,

More information

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview

More information

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Application Note Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns Impact of Instrument Dispersion on Performance of HPLC Capillary Columns Wendy Roe, Richard A. Henry, and Hillel Brandes Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA T413131 sigma-aldrich.com/analytical

More information

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water APPLICATION NOTE 70923 Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water Authors Chen Jing, Dai Zhenyu, Xu Qun, and Liang Lina, Thermo Fisher Scientific, Shanghai, People

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

Acclaim Mixed-Mode WCX-1

Acclaim Mixed-Mode WCX-1 Acclaim Mixed-Mode WCX-1 Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28 PRODUCT MANUAL for the Acclaim Mixed-Mode WCX-1 Columns 4.6 x 150 mm, P/N (068353) 4.6 x 250 mm, P/N (068352)

More information

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis Automated Switching Between D-LC and Comprehensive D-LC Analysis The Agilent 90 Infinity D-LC Solution Technical Overview Author Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Abstract This

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection and Select Impurities Analysis Using a ydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection Bruce Bailey, Ph. D. Thermo Fisher Scientific, Chelmsford, MA, USA Application

More information

UPLC Intact MASS Analysis Application Kit

UPLC Intact MASS Analysis Application Kit (Part Number: 176001519) CONTENTS I. INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMN III. CONNECTING THE COLUMN TO LCT PREMIER

More information

ChromTech HPLC System

ChromTech HPLC System ChromTech HPLC System High-prssure Gradient System (A) High-pressure pump: 2 pcs UV/VIS detector: 1 piece Gradient control module: 1 unit High-pressure gradient mixer: 1 piece Single channel chromatography

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Cation Exchange HPLC Columns

Cation Exchange HPLC Columns Cation Exchange HPLC Columns Hamilton offers seven polymeric packing materials for cation exchange separations. Type Recommended Application(s) PRP-X00 PRP-X00 PRP-X800 HC-0 HC-7 Ca + HC-7 H + HC-7 Pb

More information

Selecting Detectors for Compounds with No Optical Absorbance

Selecting Detectors for Compounds with No Optical Absorbance C146-E114 Selecting Detectors for Compounds with No Optical Absorbance Technical Report vol.15 1. Features of Absorbance Detectors In HPLC, the detector is selected in accordance with the analyte. Many

More information

Dionex IonPac AS28-Fast-4µm column

Dionex IonPac AS28-Fast-4µm column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column

More information

for Acclaim Trinity TM P1

for Acclaim Trinity TM P1 for Acclaim Trinity TM P Product Manual for Acclaim TRINITY P Columns Page of 4 Product Manual for Acclaim Trinity TM P Columns 07556 Acclaim Trinity P, 3µm, 4.6 x 00mm 07556 Acclaim Trinity P, 3µm, 4.6

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts

More information

Expanded capabilities in ammonium and amine detection

Expanded capabilities in ammonium and amine detection PRODUCT SPECIFICATIONS Thermo Scientifi c Dionex SC-CERS 500 Salt Converter-Cation Electrolytically Regenerated Suppressor Expanded capabilities in ammonium and amine detection Benefits Broadened ammonium

More information

HPLC Column Selection: Solve the Separation Mystery. The world leader in serving science

HPLC Column Selection: Solve the Separation Mystery. The world leader in serving science HPLC Column Selection: Solve the Separation Mystery The world leader in serving science verview Introduction Historical verview in Stationary Phases Basics in Column Selection Resolution, Efficiency and

More information

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone Mia Summers and Kenneth J. Fountain Waters Corporation, 34 Maple St.,

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs Value of Eclipse Plus C18 and ph Selectivity to Analyze Ae-Containing Drugs Application Pharmaceuticals Authors John W. Henderson Jr., William J. Long, and Cliff Woodward Agilent Technologies 285 Centerville

More information

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS Richard Jack, Xiaodong Liu, Leo Wang, and Chris Pohl OT70806_E 08/13S 1 The world leader in serving science

More information

Nexera UC Unified Chromatography

Nexera UC Unified Chromatography Nexera UC Unified Chromatography The latest addition to the chromatography toolbox Dr. Gesa J. Schad Shimadzu Europa GmbH A brief history of SFC ϒ Late 1800 s: it was found that heavy, non-volatile organic

More information

for Acclaim Mixed-Mode WCX-1

for Acclaim Mixed-Mode WCX-1 for Acclaim Mixed-Mode WCX- Product Manual for the Acclaim Mixed-Mode WCX- Column Page of 8 Product Manual for Acclaim Mixed-Mode WCX- Columns Acclaim Mixed-Mode WCX-, 3µm, Analytical column, 3.0 x 50mm

More information

Agilent LC & LC-MS Cannabis Testing

Agilent LC & LC-MS Cannabis Testing Agilent LC & LC-MS Cannabis Testing Potency Analysis for Cannabinoids Mike Adams 1, A. Roth 1, Joan Stevens 2, Karen Kaikaris 1, & Sue D Antonio 2 1 CWC LABS, 2 Agilent Technologies, UV detection with

More information

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof

More information

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

HPLC Winter Webinars Part 2: Sample Preparation for HPLC HPLC Winter Webinars Part 2: Sample Preparation for HPLC Jon Bardsley, Application Chemist Thermo Fisher Scientific, Runcorn/UK The world leader in serving science What am I Going to Talk About? What do

More information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors

More information

Acclaim HILIC-10 Columns

Acclaim HILIC-10 Columns Thermo Scientific Acclaim HILIC- Columns Product Manual P/N: January 3 Part of Thermo Fisher Scientific Product Manual for Acclaim HILIC- Analytical Column Acclaim HILIC-, 3 µm, Analytical, 4.6 x 5 mm

More information

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent Infinity LC System with Diode Array Detector Application Note Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany [mau]

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core

More information

HPLC Analysis of Non-volatile Analytes Using

HPLC Analysis of Non-volatile Analytes Using HPLC Analysis of Non-volatile Analytes Using Charged Aerosol Detection Paul H. Gamache, Ryan S. McCarthy, Scott M. Freeto, Darwin J. Asa, Matthew J. Woodcock, Katherine Laws and Roderic O. Cole ESA Inc.,

More information

Too Polar for Reversed Phase What Do You Do?

Too Polar for Reversed Phase What Do You Do? Too Polar for Reversed Phase What Do You Do? June 20, 2013 Mark Powell Columns and Consumables Technical Support Page 1 C8 or C18 Doesn t Always Do the Job Typical reversed phase conditions involve water/buffer

More information

How proteins separate on reverse-phase HPLC

How proteins separate on reverse-phase HPLC 1 Reverse Phase How proteins separate on reverse-phase HPLC RP chromatography separates proteins through the interaction of the hydrophobic foot of the protein with a nonpolar surface of the particle RP

More information

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Food Author Michael Woodman Agilent Technologies, Inc. 285 Centerville Road Wilgton, DE 1988-161 USA Abstract Soft drinks,

More information

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for Clinical Toxicology Page of 0 Materials and Methods Biomass Component Extraction: The uneaten cooked plant specimen was prepared for chemical analysis as follows. The sample was frozen, diced, pulverized

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

ION CHROMATOGRAPHY SYSTEM S 150

ION CHROMATOGRAPHY SYSTEM S 150 ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved

More information

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry The end of mass-speculation Pre-cleaned and certified vials for mass spectrometry Bonded, ultra-pure closures LC/MS and GC/MS Manufacturing the Ultimate Vial for Mass Spectrometry More than a quarter century

More information

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system Application Note Food Testing & Agriculture Authors Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Keiko

More information

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR Tanya Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION ACQUITY UPLC PDA LINEARITY There are two characteristics that

More information

HPLC analysis of VFAs, Furfural and HMF

HPLC analysis of VFAs, Furfural and HMF HPLC analysis of VFAs, Furfural and HMF Introduction The following chromatographic analysis is developed in order to analyse volatile fatty acids and carbohydrate degradation products; furfural, hydroxymethylfurfural

More information

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection Application Note Small Molecules The Agilent 26 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection Method Development Using Method Scouting Wizard Authors Stefan Bieber Analytical

More information

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20544 Key Words Hydrophilic, HILIC,

More information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Application Note: 56 Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Yang Shi, Catherine Lafontaine, Matthew Berube, John Fink, François Espourteille Thermo

More information

LC and LC/MS Column Selection Flow Chart

LC and LC/MS Column Selection Flow Chart LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times The Secrets of Rapid HPLC Method Development Choosing Columns for Rapid Method Development and Short Analysis Times Rapid Analysis Is More Than Run Time It is developing a method to meet a goal and developing

More information

Acclaim AmG C18 Column

Acclaim AmG C18 Column CHROMATOGRAPHY Acclaim AmG C8 Column Rugged Reversed-Phase Column Designed for Aminoglycoside Antibiotics Analysis Product Specifications The Thermo Scientific Acclaim AmG C8 column is a high performance

More information

Packings for HPLC. Packings for HPLC

Packings for HPLC. Packings for HPLC Summary of packings for HPLC In analytical HPLC, packings with particle sizes of 3 to 10 µm are preferred. For preparative separation tasks, also particles with diameters larger than 10 µm are applied.

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Analysis of Pesticides by Supercritical Fluid Chromatography/ Mass Spectrometry Optimizing the Agilent 126 Infi nity Analytical SFC System in Combination with the Agilent 646 Triple Quadrupole LC/MS Application

More information

APPLICATION NOTE Goal. Introduction. Keywords

APPLICATION NOTE Goal. Introduction. Keywords APPLICATION NOTE 72398 A highly sensitive high-performance liquid chromatographycharged aerosol detection method for the quantitative analysis of polysorbate 8 in protein solution Authors Zhen Long, Guobin

More information

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep

More information

Evaporative Light Scattering Detector (ELSD)

Evaporative Light Scattering Detector (ELSD) BÜCHI Labortechnik AG Evaporative Light Scattering Detector (ELSD) Martin Verstegen ELSD 3300 page 1 Overview of Common HPLC Detectors 1. UV/Vis 2. RI 3. Fluorescence 4. Mass Spectrometry 5. ELSD page

More information

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,

More information

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank

More information

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Prep 150 LC System: Considerations for Analytical to Preparative Scaling Andrew Aubin and Jo-Ann Jablonski Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The Prep 150 LC System is an affordable, highly reliable system for preparative chromatography and is suitable

More information

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun

More information

Determination of trace anions in concentrated hydrofluoric acid

Determination of trace anions in concentrated hydrofluoric acid APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion

More information

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Application Note Inorganic Ions, Water Testing, Minerals, Metals, Basic Chemicals Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Authors Anne Mack, Adam Bivens

More information

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis

More information

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Thermo Scientific Accucore XL HPLC Columns. Technical Manual Thermo Scientific Accucore XL HPLC Columns Technical Manual Thermo Scientific Accucore XL HPLC Columns Based on Core Enhanced Technology using µm solid core particles, Accucore XL HPLC columns allow users

More information