Performance of a New Commercial Electrical Mobility Spectrometer

Size: px
Start display at page:

Download "Performance of a New Commercial Electrical Mobility Spectrometer"

Transcription

1 Aerosol Science and Technology ISSN: (Print) (Online) Journal homepage: Performance of a New Commercial Electrical Mobility Spectrometer M. Heim, G. Kasper, G. P. Reischl & C. Gerhart To cite this article: M. Heim, G. Kasper, G. P. Reischl & C. Gerhart (2004) Performance of a New Commercial Electrical Mobility Spectrometer, Aerosol Science and Technology, 38:S2, 3-14, DOI: / To link to this article: Published online: 17 Aug Submit your article to this journal Article views: 499 View related articles Citing articles: 35 View citing articles Full Terms & Conditions of access and use can be found at

2 Aerosol Science and Technology, 38(S2):3 14, 2004 Copyright c American Association for Aerosol Research ISSN: print / online DOI: / Measurement Methods Performance of a New Commercial Electrical Mobility Spectrometer M. Heim, 1 G. Kasper, 1 G. P. Reischl, 2 and C. Gerhart 3 1 Institut für Mechanische Verfahrenstechnik und Mechanik, Universität Karlsruhe (TH), Karlsruhe, Germany 2 Institut für Experimentalphysik, Universität Wien, Wien, Austria 3 Grimm Aerosol Technik, Ainring, Germany A commercial electrical mobility spectrometer (Grimm DMA and Grimm CPC) was examined with respect to size, concentration and time-dependent responses of the CPC, the transfer function of the DMA, as well as the performance of the data inversion algorithm. For comparison with previously published results, the responses were also measured for two other commercial CPCs (TSI Model 3010 and TSI Model 3022). All measurements were performed using sodium chloride aerosol produced by an atomizer, except for size-dependent counting efficiency measurements, where an evaporation and condensation method was used to produce particles down to a size of 2 nm. The concentration-dependent CPC counting efficiency was measured against a reference instrument working in the single particle count mode following dilution. The CPC time response was evaluated for monodisperse aerosol by measuring the step response using a solenoid valve technique. The transfer function of the Grimm DMA was determined between 10 and 100 nm with two verified identical DMAs in series following the approach of Fissan et al. Measured tandem-dma responses were compared to simulated responses based on the diffusion-broadened transfer function of Stolzenburg to define the factors influencing resolution. Finally the data inversion algorithm was tested by comparison with other devices. INTRODUCTION Electrical mobility spectrometers (EMS) such as the scanning mobility particle sizer (SMPS) or the differential mobility particle sizer (DMPS) are among the most commonly used in- Received 27 June 2003; accepted 29 July Address correspondence to Michael Heim, Institut für Mechanische Verfahrenstechnik und Mechanik, Universität Karlsruhe (TH), Strasse am Forum 8, D Karlsruhe, Germany. michael.heim@ mvm.uni-karlsruhe.de struments for measuring particle size distributions in the submicron particle size range. A typical EMS consists of a differential mobility analyzer (DMA) to classify particles due to their size and charge into electrical mobility fractions and a counter such as a condensation particle counter (CPC) or a Faraday cup electrometer (FCE). The performance of such a system depends on the performance of these two elements. The main criteria for a CPC are size-dependent counting efficiency, concentrationdependent counting accuracy, and its time response function to a sharp concentration change. The quality of a DMA is usually determined by measuring its transfer function, which gives information about the sharpness of the classification and the particle losses that occur in the instrument. Much work was done in the past to test commercial CPCs and DMAs. Wiedensohler et al. (1997) determined the sizedependent counting efficiency of 26 CPCs by comparing them with two TSI 3025 CPCs calibrated previously against a FCE. A good review of measurement data published on the characteristics of the TSI CPCs is also given by Sem (2002). The measurement of the particle size-dependent counting efficiency generally requires monodisperse particles, usually generated by an evaporation/condensation method and subsequent DMA classification, and a sensitive FCE as concentration reference. This is the approach taken here. Data on concentration-dependent counting efficiencies of CPCs were published by Quant et al. (1992) and by Ankilov et al. (2002a). Quant et al. used a previously calibrated TSI 3022 CPC as the reference instrument, whereas Ankilov et al. worked with the mean concentration of all participating test instruments. The present study uses a TSI 3010 CPC as reference but in the single-particle count mode, which is considered absolute because inlet losses in that particle size range are negligible (Wen and Kasper 1986; Niida et al. 1988). 3

3 4 M. HEIM ET AL. Quant et al. (1992) published time response measurements using a solenoid valve to switch between polydisperse ambient air and filtered air; Wang et al. (2002) determined the time response from a concentration pulse of spark-source aerosol. In either case it is important to assure that the upper concentration level remains below the saturation range of the single particle count mode; otherwise the CPC response slows down. It is not entirely clear from either article whether that saturation limit was reached or not, because the latter data show a slower response than Quant et al., although the same optical detector was used (also see our data later on). The setup was therefore altered for the present purposes by switching between well-defined concentration levels generated with monodisperse aerosol. The DMA transfer function can be obtained either by using highly monodisperse particles of variable and well-known size as input aerosols, which is quite difficult, aside from certain molecules (Fernandez de la Mora et al. 1998; Tanaka and Takeuchi 2003). The more practical option is to use tandem DMA techniques, which require a mathematical deconvolution of the data, e.g., by the method of Fissan et al. (1996). This latter approach is predicated upon the use of two identical DMAs, a condition that is rarely met with sufficient accuracy, even for nominally identical devices (Karlsson and Martinsson 2003), unless specifically proven. This route was taken in the present article. ESSENTIAL DEVICE CHARACTERISTICS Figure 1 shows the geometry of the Grimm DMA Model The outer electrode has an i.d. of 40 mm, the center electrode an o.d. of 20 mm; the distance between the aerosol inlet and outlet is 441 mm. The aerosol enters the classifier through a tangential inlet similar to the Vienna DMA (Winkelmayr 1991) and is then distributed uniformly via a narrow annular slit before merging with the particle-free sheath air flow. Sheath air distribution, excess air, and the monodisperse outlet design are comparable to the that of Knutson and Whitby (1975). Typical flow rates of the classifier are Q a = Q s = 0.3 l/min and Q sh = Q ex = 3 l/min, which are controlled by a closed-loop system using rotary vane pumps and a flow controller included in the housing of the CPC. The classifier housing also includes a single-stage impactor and a high voltage module, as well as the possibility to connect an additional 241 Am charger (Grimm Neutralizer 5.422) between the impactor and the aerosol inlet of the classifier. The Grimm neutralizer was not available for the examination of the system; therefore, an available 85 Kr charger was used for the present measurements. The Grimm CPC Model is a continuous flow device like several others based on the concept of Bricard et al. (1976), using butanol as working fluid. Condenser and saturator temperatures are 10 C and 35 C, respectively, resulting in a temperature difference of 25 C. Instead of a butanol reservoir in the saturator, a sensor-controlled micropump is used to wet the felt, thereby preventing accidental flooding of the optics by the working fluid. A drainage system with a peristaltic pump is used to drain the Figure 1. Schematic drawing of the Grimm DMA. All dimensions are given in millimeters. condensate from the system to prevent contamination of the working fluid by condensing water from the condenser walls. The optical detection unit consists of a combination of laser and a fast photodiode, which detects scattered light at an angle of 90.Anintegrated pulse height analyzer detects droplet sizes at two trigger levels, depending on particle concentration level. Concentrations below 500 cm 3 are detected at the higher trigger level, which is less sensitive to noise, and higher concentrations up to the upper level of the single particle mode of 6000 cm 3 are measured at the lower level. Concentrations above cm 3 are determined by a photometric signal from the optical detector. Between these two ranges a mix of both signals (photometric and single-particle counting) is used to obtain a smooth transition between both measurement principles. The coincidence correction in this range is included by using the two trigger signals from the pulse height analyzer. The transition region is also used to control the possibility of misadjustments between the photometric mode and the single-particle count mode resulting from a contaminated optics or a diluted working fluid. The tested Grimm EMS system uses a data reduction algorithm based on a mixture of the scanning mode (Wang and Flagan 1990) and an advanced method of the classical stepping

4 NEW COMMERCIAL ELECTRICAL MOBILITY SPECTROMETER 5 mode (Knutson 1976; Winklmayr 1987). The method was developed by the Vienna Group and is an improved version of the algorithm described by Reischl et al. (1991). By stepping the voltage and using each concentration detected by the CPC with a frequency of 1 Hz, this method derives a faster measurement than a classical stepping technique. The detected number size distribution function of the classified sample aerosol F C (d P ) from the DMA at voltage V is given by the original number size distribution function F(d P ), the ideal transfer function Tr(Z(d P, i)), the system detection efficiency E(d P ), and the charge distribution of the particles α(d P ) (Fuchs 1963): F C (d P ) = dn C d ln d P = α(d P, i) Tr(Z(d P, i)) F(d P ) E(d P ) i=1 This equation can only be solved analytically to give a unique value of the original number particle size distribution under the following restrictions (Winklmayr 1987): knowledge on the DMA Transfer function Tr(Z(d P, i)), knowledge on the charge distribution α(d P ), knowledge on the system detection efficiency E(d P ), sharply limited upper particle size, and fixed voltage steps/mobility channels. The calculation includes a correction for the time response characteristic of the CPC in terms of flow and voltage stepping, the aerosol residence time in the DMA, and particle losses in the DMA. The measurement is done within a downscan, beginning [1] with the largest particle size. The channel width is adjusted to the breadth of the ideal transfer function (Reischl 1991). The normal operational mode uses 44 voltage channels, with an integration time of 2 s. The fast mode only uses 22 voltage channels; however, the information from the counts detected in the time between the voltage step and steady-state conditions being reached are utilized to calculate an additional series of 22 pseudovoltage channels. This results in a total number of 44 channels. SIZE-DEPENDENT CPC RESPONSE The experimental setup for the determination of the counting efficiency is shown in Figure 2. Sodium chloride aerosol in the size range from 2 to 20 nm was produced by evaporation/condensation in a tube furnace (Kasper and Berner 1978). Particles larger than 20 nm were generated by atomizing a salt solution to obtain droplets and dried in a diffusion dryer (not shown in Figure 2). Dry pressurized air was used for the generation of the particles as well as for DMA operation to avoid hygroscopic growth effects. The aerosol was diluted in a bypass diluter to ensure that all measurements were made in the more accurate single-particle count mode (below cm 3 ). The neutralized aerosol ( 85 Kr; 20 mci) was classified using a short Vienna-type DMA (R 2 = 33 mm, R 1 = 25 mm, L = 114 mm) operated under symmetric flow conditions Q sh = Q ex = 25 l/min. A FCE (University of Karlsruhe design; 0.2 fa detection limit) operated with the same flow rate as the CPC was used as a concentration reference for two CPCs, a Grimm Model and a TSI Model The efficiency curve for the TSI 3010 counter is shown in Figure 3, together with other published data for the same type of instrument. A three-parameter sigmoidal fit Figure 2. Experimental setup used for the determination of the counting efficiency.

5 6 M. HEIM ET AL. Figure 3. Measured counting efficiency of the TSI 3010 CPC (d 50 = 10.9 nm) together with previously published data for the same instrument. (Mertes et al. 1995) was applied to the data to calculate the 50% counting efficiency diameter d 50 for all measurements shown in Table 1. Figure 4 shows a comparison between measured counting efficiency curves of the Grimm device and other commercial CPCs. The data include measurements taken in two different years using the same Grimm device, giving an indication of the reproducibility. The resulting cutoff diameter for both measurements is 7.5 nm. The TSI 3010 counter used by Russell et al. (1996) was modified to operate with a higher temperature difference (25 K) between saturator and condenser, comparable to the temperature difference used by the Grimm counter. The lower cutoff diameter of 5.7 nm obtained by the latter authors could be a result of a different condenser design, consisting of eight tubular channels merging at the exit of the condenser (Keady 1988), and hence a better supersaturation profile. Table 1 Cutoff diameters for all instruments used Model d 50 Material Source TSI nm NaCl Kesten et al. (1991) TSI 3010 a 5.7 nm NaCl Russell et al. (1996) Grimm nm NaCl This work TSI nm NaCl Ankilov (2002b) TSI nm NaCl This work TSI nm Ag Mertes et al. (1995) TSI nm Ag Buzorius (2001) TSI nm Ag Wiedensohler et al. (1997) a ( T = 25 K). Figure 4. Counting efficiency for NaCl particles for different commercial CPCs. The two measurements for the Grimm CPC were taken two years apart. The TSI 3010 CPC used by Russell et al. (1996) was operated with a higher temperature difference of 25 K between saturator and condenser. CONCENTRATION-DEPENDENT CPC RESPONSE Beside the particle size dependence, the single-particle count mode has a concentration-dependent efficiency caused by optical coincidence effects (Raasch and Umhauer 1984). The calibration of the photometric mode can also result in erroneous concentration values, e.g., if the final droplet size reached in the CPC is not identical to the droplet size used during factory calibration due to a dilution of the butanol by condensing water, or due to contaminated detection optics. The experimental setup shown in Figure 5 was used to measure the concentration-dependent counting efficiency for the Grimm 5.403, the TSI 3022, and the TSI 3010 CPC with NaCl aerosol having a median diameter (d 50 )of52nm, a standard deviation (σ g )of1.44, and total concentrations between 10 and cm 3.Astatic mixer was installed to assure a uniform concentration profile after the dilution unit. The aerosol was then divided into four separate streams, one for each CPC. An optional second diluter (Palas, Injection-Type VKL) could be installed between CPC 4 and the sampling port for the measurement of higher aerosol concentrations. Initial measurements were performed without the second diluter, in order to compare all CPC models in the single-particle count mode. Figure 6 shows the ratio of measured to average particle number concentrations (c meas /c avg )versus average particle number concentration c avg.inthe examined concentration range, all CPC models are within 10% of the average value. Quant et al. (1992) also found a good agreement for the TSI 3022, TSI 3025, and the TSI 3010 for concentrations below 10 4 cm 3, where all instruments operate in the single-particle count mode. To obtain an appropriate reference concentration above 10 4 cm 3, the injector diluter was used upstream of CPC 4 (identified

6 NEW COMMERCIAL ELECTRICAL MOBILITY SPECTROMETER 7 Figure 5. Experimental setup used for the determination of the concentration-dependent CPC response. as TSI 3010 Ref. in Figure 6). The dilution ratio was determined experimentally by comparing the diluted concentration to the average concentration of the first three CPCs in the range below 10 4 cm 3.Aleast-squares fit was then used to calculate the dilution ratio. The obtained dilution ratio allowed a reference concentration with a relative accuracy of ±20% to be calculated for concentrations of up to cm 3. The measured concentrations for all three counters examined are shown in Figure 7 as a function of the calculated reference concentration. Figure 6. Results of the concentration comparisons in the single-particle count mode. All concentrations below 10 4 cm 3. Figure 7. Results of the concentration comparisons against a reference concentration calculated from the concentration measured by a TSI 3010 CPC in the single-particle count mode behind an injector diluter.

7 8 M. HEIM ET AL. For the TSI 3010, a deviation from the reference concentration is observed commencing at approximately cm 3, and reaching a saturation value at cm 3.Asthis instrument is designed for single-particle counting only, this behavior is most likely caused by coincidence effects. This is consistent with the data observed by Quant et al. (1992); however, the saturation value determined here is approximately four times higher. The Grimm and the TSI 3022 CPC both use a factorycalibrated photometric mode for concentrations above 10 4 cm 3. According to Figure 7, the measured concentration of the Grimm CPC stays almost completely within the validity range (5% deviation), while the TSI model 3022 tends to slightly overestimate the validity range of the reference concentration (20% deviation) between cm 3 and cm 3. The last factory calibration of the TSI 3022 was one year ago, and because of this it is possible that the photometric mode was slightly misadjusted. Also, according to the TSI manual the calibration of the TSI 3022 is performed with monodisperse 50 nm NaCl particles, whereas the Grimm is factory calibrated with a polydisperse NaCl aerosol with a mean diameter of about 50 nm. A polydisperse aerosol was also used for the present measurements. CPC TIME RESPONSE In order to measure transient aerosol concentrations or to perform fast particle size distribution scans (Collins et al. 2002), the time response of the particle counter must be known. This characteristic is usually measured in terms of the response of the CPC to a particle concentration step function. The experimental setup for this part of the study (Figure 8) is based on monodisperse NaCl particles with a mean diameter of 100 nm generated by narrow-band DMA classification. This size was chosen to avoid the influence of time-dependent diffusional losses (Okuyama et al. 1986), which can effect the time response. The necessary concentration step signal was generated by switching a solenoid three-way valve from the inlet of the CPC to the DMA or the high efficiency particle (HEPA) filter. The start time t 0 for the measurement was defined by a trigger signal from the solenoid valve with an uncertainty of 50 ms, due to the switching time of the valve. The time-resolved response was measured by using the digital pulse output from the optics with a frequency of 60 Hz. (This required minor electrical modifications on the Grimm CPC). Due to coincidence effects, a CPC s time response may appear to depend on aerosol concentration (Heim 2003). This is also the case for the time response of the Grimm Model shown in Figure 9 for three different starting concentrations N 0 (the initial dead time is not shown). Time response measurements in this study were therefore made at concentrations below 5000 cm 3. According to Russell et al. (1995), the ideal step response of a CPC can be described by a series connection of residence time Figure 8. Experimental setup used for the determination of the CPC time response.

8 NEW COMMERCIAL ELECTRICAL MOBILITY SPECTROMETER 9 Figure 9. Time responses of the Grimm CPC (the initial dead time is not shown) for a concentration step for three different starting concentrations N 0,neglecting the coincidence effect. distributions (RTD) in a plug flow reactor (PFR) and a continuously stirred tank reactor (CSTR; see, for example, Levenspiel 1972). Given its inner geometry and a Reynolds number Re 25 for the flow in the main volume (saturator), it is unlikely that the Grimm CPC will behave as an ideally mixed volume (CSTR). A more physical description is based on a laminar flow reactor (LFR) in series with a PFR, the latter to describe inlet phenomena resulting in a deviation of an ideal laminar flow. The response function of a CPC is then described by the following expression: 1 t < F(t) = ( 1 4 τlfr 2 ) (t τ PFR ) 2 t ( τlfr ) 2 + τ PFR ( ) τlfr 2 + τ PFR. The response function of the Grimm is shown in Figure 10, giving characteristic time constants of τ LFR = 3.37 s for the LFR and τ PFR = 0.63 s for the PFR. Agreement with the model is very good up to about 10 s. A small deviation for longer times may be due to nonideal effects such as the surface roughness of the saturator felt, which disturbs the boundary layer of the laminar flow and leads to a decay that is slower than predicted. Step response measurements for two other CPCs (TSI Model 3022 and TSI Model 3010) are shown in Figure 11. The response of the Model 3010 is significantly faster due to its three-fold higher volume flow rate, and a different condenser and saturator geometry (Keady 1988). The responses of the TSI Model 3022 and the Grimm Model are comparable, due to similar internal flow geometries and the same operating flow rate of 0.3 l/min. However, the LFR PFR model according to Equation (2) does not fit well to the Model 3022 reponse. Therefore, a [2] Figure 10. Time response of the Grimm CPC for a concentration step. The response can be approximated by a combination of the response for a PFR and an LFR, and it is described in Equation (2). more pragmatic approach was taken that models the response function by a series connection of a PFR and a parallel connection of two CSTRs (Equation (3)). This approach gives a good numerical representation of the response functions of all three instruments: F(t) = 1 t <τ PFR ( ) ( ) t τpfr t τpfr a exp + (1 a) exp t τ PFR. τ CSTR1 τ CSTR2 Figure 11. Comparison of the step response of the Grimm 5.403, the TSI 3022 and the TSI 3010 CPCs. The solid lines represent fits according to Equation (3) based on the parameters given in Table 2. [3]

9 10 M. HEIM ET AL. Table 2 Parameters to describe the time response of three CPCs for a particle size of 100 nm according to Equation (3) Model a τ PFR τ CSTR1 τ CSTR2 Grimm s 0.97 s 5.72 s TSI s 0.71 s 7.32 s TSI s 0.76 s 2.61 s Herein a denotes a factor that can vary between 0 and 1, τ PFR the systems dead time, and τ CSTR1 and τ CSTR2 the characteristic times of both CSTRs. The corresponding values for all three counters are presented in Table 2. DMA TRANSFER FUNCTION In order to obtain information concerning the transfer function of the Grimm DMA , tandem measurements were made using two identical instruments connected in series. The transfer function of the DMA was derived from the resulting tandem signal using a deconvolution technique similar to the one described by Stratmann et al. (1997). The instrument setup is shown in Figure 12. The NaCl particle concentration was controlled so that all measurements could be performed in the more accurate singleparticle count mode of the Grimm CPC, where distortions of the transfer function due to space charge effects (Alonso et al. 2001) are also considered to be negligible. The air flow rates of both classifiers were controlled by means of critical orifices to assure a constant flow. The fixed voltage for the first classifier (Generator DMA) was produced either by an HP 6674A (0 60 Volt) or an HV supply based on the Brandenburg HV module ( V). For the second classifier (Analyzer DMA) operated in a stepping mode, we used the internal HV supply. The mobility distribution of the Generator DMA was measured in 22 steps, with an interval length of 50 s, resulting in a total measuring time of 18 min. Before and after each Figure 12. Instrumental setup for the tandem measurements.

10 NEW COMMERCIAL ELECTRICAL MOBILITY SPECTROMETER 11 measurement, the particle number concentration downstream of the Generator DMA was measured to ensure that the concentration had not changed during measurement. To avoid any influence from the shape of the NaCl aerosol size distribution, the salt solution in the atomizer was adjusted so that the particle size selected by the Generator DMA was at the peak of the particle size distribution, where concentration changes within the range of the expected width of the transfer function could be neglected. Diffusional losses in the interconnecting tubes between the classifiers and the CPC were measured and also taken into account according to the method of Fissan et al. (1996). Karlsson et al. (2003) showed that manufacturing imperfections of DMAs with identical design can result in substantially different transfer functions. Therefore, the equality of both instruments was determined by using a third, short Vienna-type DMA (R 2 = 33 mm, R 1 = 25 mm, L = 114 mm) as the Generator DMA and the two Grimm s as Analyzer DMAs. Figure 13 shows that the tandem signals measured with the two Grimm DMAs are very close, so that the assumption of identical transfer functions is justified. For the deconvolution process we used a triangular transfer function (Martinsson et al. 2001) with broadening parameter (µ) and loss parameter (λ) according to Equation (4): [ ( )] f (λ, µ, Z 0, Z) = λ µ 1 + µ Qsh Z 1 ; Q a Z 0 1 Qa µ Qsh Z Z0 1, [ )] f (λ, µ, Z 0, Z) = λ µ 1 + µ Qsh (1 ZZ0 ; Q a 1 Z Z0 1 + Qa µ Qsh, [4] Z f (λ, µ, Z 0, Z) = 0; < 1 Q a or Z > 1 Z 0 µ Q sh Z 0 + Q a. µ Q sh Here Q a and Q sh denote the aerosol and sheath air flow rates, Z the mobility, and Z 0 the center mobility of the classifier. Transfer functions are usually described in terms of three parameters, height (h), area (A), and full width at half maximum (FWHM), which become dimensionless in the normalized mobility space. The parameters can be calculated from λ and µ (Equation (5)) and are included in Figures 14 16: h = λ µ, FWHM = Q a, [5] µ Q sh A = λ Qa Q sh. With decreasing particle size the height and the area of the transfer function decrease, most likely due to diffusional losses inside the classifier. The area does not reach the ideal value of 0.1 (Knutson and Whitby 1975) even for the largest particle sizes used in this study, where diffusional losses are expected to be negligible. The reason for this can be explained by particle losses due to electrophoretic migration in the exit region of the classifier (Zhang and Flagan 1996) induced by the local electric field near the insulator, where the aerosol passes from a region of high voltage at the outlet of the center electrode to the grounded bottom section of the classifier. For the largest particle size examined, the FWHM reaches the ideal value of 0.1, which can be interpreted as a sign that Figure 13. Tandem signals for both Grimm DMAs examined, downstream of a short Vienna DMA at a selected particle size of 52 nm. Figure 14. Height of the transfer function of the Grimm DMA as function of the particle diameter. The solid line represents a fit to the measurement data.

11 12 M. HEIM ET AL. Figure 15. Area of the transfer function of the Grimm DMA as function of the particle diameter. The solid line represents a fit to the measurement data. only a very minor part of the broadening is caused by geometric imperfections or flow disturbances inside the classifiers. For smaller particle sizes, the FWHM increases due to diffusional broadening (Kousaka et al. 1986). In order to prove this, simulations of the tandem measurements were performed based on the diffusion-broadened transfer function described by Stolzenburg (1988). The simulations account for the diffusional broadening inside the classifying zone of the DMA, but not for losses in the aerosol inlet and outlet of the classifier. In order to compare the model and the experiment, a loss correction function was applied to the measured data. The function was derived from the area ratio of the experimental and ideal transfer functions as shown in Equation (6). The particle diameter term (d P )inequation (6) was derived from the voltage of the first Figure 16. Full width at half maximum of the Grimm DMA as function of the particle diameter. The solid line represents a fit to the measurement data. Figure 17. Comparison between loss-corrected tandem measurements (loss correction: Equation (6)) for the Grimm DMA and simulated tandem signals based on Stolzenburg s diffusion-broadened transfer function for three different particle sizes and a flow rate of 0.3 l/min to 3.0 l/min. DMA. Figure 17 shows a comparison between the simulations and the tandem measurements for three different particle sizes. All three particle sizes show good agreement between the measurements and the simulations, indicating that the resolution of the classifier at the given flow rate was only limited by particle diffusion: η DMA = Area exp Area ideal = (1 exp ( d P )). [6] DATA INVERSION The data inversion algorithm included in the software of the Grimm EMS System (Version 1.3b) was tested by comparison with a TSI SMPS model 3934, consisting of a 3071 classifier and a 3010 CPC using the same flow ratio of 0.3 l/min to 3.0 l/min for both systems. Therefore an additional particle-free air flow was added between the outlet of the TSI DMA and the TSI 3010 CPC. The scan time for the TSI instrument was 3 min (60 s downscan and 120 s upscan); for the Grimm EMS system, which was operated in the fast mode, it was 198 seconds. The particle size distributions measured by the two instruments are shown in Figure 18. It has to be mentioned that the Grimm software displays the size distribution data referring to the natural logarithm instead of the decimal logarithm used by TSI. Therefore, the data points of the Grimm EMS System are corrected by a factor of for the comparison. The two particle size distributions in Figure 18 look very similar, with median diameters of 67.5 nm (Grimm) and 68.7 nm (TSI); the total concentration shows a difference of 16%, a value of the order that is apparently typical for identical EMS Systems (Dahmann et al. 2002).

12 NEW COMMERCIAL ELECTRICAL MOBILITY SPECTROMETER 13 Figure 18. Size distribution of a NaCl Aerosol measured with the Grimm EMS system and the TSI SMPS model SUMMARY The performance of a new CPC (Grimm Model 5.403) and a new DMA (Grimm Model ) were thoroughly examined using NaCl aerosol and compared to other commercial instruments. The size-dependent 50% counting efficiency diameter of the Grimm and the TSI 3010 CPC were determined to 7.5 nm and 10.9 nm, respectively. The counting accuracy for aerosol concentrations between 10 and cm 3 was measured for three different CPCs by comparing with a reference concentration derived from a diluted concentration measured by a TSI 3010 CPC in the single-particle count mode. In the photometric mode, both TSI Model 3022 and Grimm Model determined the concentrations almost correctly with a maximum relative error of 20%. The TSI Model 3010, which uses only the single-particle count mode, saturates above about cm 3. The responses to a step function input signal for the Grimm 5.403, the TSI 3022, and the TSI 3010 CPCs each exhibit a longterm behavior that can be very well represented numerically by a delay time followed by two superimposed exponential decay functions. The transfer function of the Grimm DMA was determined in the size range between 10 and 100 nm for a flow ratio of 0.3/3.0 (l/min) by using two virtually identical instruments in series. It was shown by comparison with simulations based on Stolzenburg s diffusion-broadened transfer function that particle diffusion was the only broadening factor. For a polydisperse test aerosol, the Grimm EMS and TSI SMPS systems both measured the same median diameter within 1 nm, with a difference in total concentration of 16%. REFERENCES Alonso, M., Alguacil, F. J., Watanabe, Y., Nomura, T., and Kousaka, Y. (2001). Experimental Evidence of DMA Voltage Shift Due to Space-Charge, Aerosol Sci. Technol. 35: Ankilov et al. (2002a). Intercomparison of Number Concentration Measurements by Various Aerosol Particle Counters, Atmos. Res. 62: Ankilov et al. (2002b). Particle Size Dependent Response of Aerosol Counters, Atmos. Res. 62: Bricard, J., Delattre, P., and Pourprix, M. (1976). Detection of Ultra-Fine Particles by Means of a Continuous Flux Condensation Particle Counter. In Fine Particles: Aerosol Generation, Measurement, Sampling, and Analysis, edited by B. Y. H. Liu. Academic Press, New York, pp Buzorius, G. (2001). Cut-Off Sizes and Time Constants of the TSI 3010 Operating at 1 3 lpm Flow Rates, Aerosol Sci. Technol. 35: Collins, D. R., Flagan, R. C., and Seinfeld, J. H. (2002). Improved Inversion of Scanning DMA Data, Aerosol Sci. Technol. 36:1 9. Dahmann, D., et al. (2002). Intercomparison of Mobility Particle Sizers (MPS), Gefahrstoffe Reinhaltung Luft 61: Fernandez de la Mora, J., de Juan, L., Eichler, T., and Rosell, J. (1998). Differential Mobility Analysis of Molecular Ions and Nanometer Particles, Trends Anal. Chem. 17: Fissan, H., Hummes, D., Stratmann, F., Büscher, P., Neuman, S., Pui, D. Y. H., and Chen, D. (1996). Experimental Comparison of Four Differential Mobility Analyzers for Nanometer Aerosol Measurements, Aerosol Sci. Technol. 24:1 13. Fuchs, N. A. (1963). On the Stationary Charge Distribution on Aerosol Particles in a Bipolar Ionic Atmosphere, Geof. Pura Appl. 56: Heim, M., Kasper, G., Weber, A. P., and Reischl, G. P. (2003). Response Function of Submicron Particle Counters, J. Aerosol Sci. 34:S1, S583 S584. Karlsson, M. N. A., and Martinsson, B. G. (2003). Methods to Predict the Transfer Function Size Dependence of Individual DMAs, J. Aerosol Sci. 34: Kasper, G., and Berner, A. (1978). A Generator for the Production of Extremely Monodisperse Sodium Chloride Aerosols (in German), Staub 38: Keady, P. B. (1988). Condensation Nucleus Counter, U.S. Patent, Number 4,790,650. Kesten, J., Reineking, A., and Porstendörfer, J. (1991). Calibration of a TSI 3025 Ultrafine Condensation Particle Counter, Aerosol Sci. Technol. 15: Knutson, E. O. (1976). Extended Electric Mobility Method for Measuring Aerosol Particle Size and Concentration. In Fine Particles: Aerosol Generation, Measurement, Sampling and Analysis, edited by B. Y. H. Liu, Academic Press Inc., New York, pp Knutson, E. O., and Whitby, K. T. (1975). Aerosol Classification by Electrical Mobility: Apparatus, Theory and Applications, J. Aerosol Sci. 6: Kousaka, Y., Okuyama, K., Adachi, M., and Mimura, T. (1986). Effect of Brownian Diffusion on Electrical Classification of Ultrafine Aerosol Particles in Differential Mobility Analyzer, J. Chem. Eng. Japan 19: Levenspiel, O. (1972). Chemical Reaction Engineering, 2nd ed., John Wiley & Sons, New York. Martinsson, B. G., Karlsson, M. N. A., and Frank, G. (2001). Methodology to Estimate the Transfer Function of Individual Differential Mobility Analyzers, Aerosol Sci. Technol. 35: Mertes, S., Schröder, F., and Wiedensohler, A. (1995). The Particle Detection Efficiency of the TSI-3010 CPC as Function of the Temperature Difference Between Saturator and Condenser, Aerosol Sci. Technol. 23: Niida, T., Wen, H. Y., Udischas, R., and Kasper, G. (1988). Counting Efficiency of Condensation Nuclei Counters in N 2,Ar,CO 2 and He, J. Aerosol Sci. 19: Okuyama, K., Kousaka, Y., Yamamoto, S., and Hosokawa, T. (1986). Particle Loss of Aerosols with Particle Diameters Between 6 and 2000 nm in Stirred Tank, J. Colloid Interface Sci. 110: Quant F. R., Caldow, R., Sem, G. J., and Addison, T. J. (1992). Performance of Condensation Particle Counters with Three Continuous-Flow Designs, J. Aerosol Sci., 23:S1, Raasch, J., and Umhauer, H. (1984). Coincidence Errors in Particle Light- Scattering Sizing Analysis (in German), VDI Fortschrittberichte, Reihe 3, Nr. 95.

13 14 M. HEIM ET AL. Reischl, G. P. (1991). Measurement of Ambient Aerosols by the Differential Mobility Analyzer Method: Concepts and Realization Criteria for the Size Range Between 2 and 500 nm, Aerosol Sci. Technol. 14:5 24. Russell, L. M., Flagan, R. C., and Seinfeld, J. H. (1995). Asymetric Instrument Response Resulting from Mixing Effects in Accelerated DMA-CPC Measurements, Aerosol Sci. Technol. 23: Russell, L. M., Zhang, S., Flagan, R. C., and Seinfeld, J. H. (1996). Radially Classified Aerosol Detector for Aircraft-Based Submicron Aerosol Measurments, J. Atmos. Oceanic Technol. 13: Sem, G. J. (2002). Design and Performance Characteristics of Three Continuous- Flow Condensation Particle Counter: A Summary, Atmos. Res. 62: Stolzenburg, M. R. (1988). An Ultrafine Aerosol Size Distribution Measuring System, Ph.D. thesis, University of Minnesota, Mineapolis, MN. Stratmann, F., Kauffeldt, T., Hummes, D., and Fissan, H. (1997). Differential Electrical Mobility Analysis: A Theoretical Study, Aerosol Sci. Technol. 26: Tanaka, H., and Takeuchi, K. (2003). Experimental Transfer Function for a Low-Pressure Differential Mobility Analyzer by Use of a Monodisperse C 60 Monomer, J. Aerosol Sci. 34: Wang, J., McNeill, V. F., Collins, D. R., and Flagan, R. C. (2002). Fast Mixing Condensation Nucleus Counter: Application to Rapid Scanning Differential Mobility Analyzer Measurements, Aerosol Sci. Technol. 36: Wang, S. C., and Flagan, R. C. (1990). Scanning Electrical Mobility Spectrometer, Aerosol Sci. Technol. 13: Wen, H. Y., and Kasper, G. (1986). Counting Efficiency of Six Commercial Particle Counters, J. Aerosol Sci. 17: Wiedensohler, A., et al. (1997). Intercomparison Study of the Size-dependent Counting Efficiency of 26 Condensation Particle Counters, Aerosol Sci. Technol. 27: Winklmayr, W. (1987). Untersuchungen Des Ultrafeinen Aerosols in Der Urbanen Atmosphäre von Wien, Ph.D. thesis, Naturwissenschaftliche Fakultät der Universität Wien, Vienna. Winklmayr, W., Reischl, G. P., Linder, A. O., and Berner, A. (1991) A New Electromobility Spectrometer for the Measurement of Aerosol Size Distributions in the Size Range from 1 to 1000 nm, J. Aerosol Sci. 22: Zhang, S., and Flagan, R. C. (1996). Resolution of the Radial Differential Mobility Analyzer for Ultrafine Particles, J. Aerosol Sci. 27:

Rapid Measurements of Aerosol Size Distributions Using a Fast Integrated Mobility Spectrometer (FIMS)

Rapid Measurements of Aerosol Size Distributions Using a Fast Integrated Mobility Spectrometer (FIMS) Rapid Measurements of Aerosol Size Distributions Using a Fast Integrated Mobility Spectrometer (FIMS) Jason Olfert, Brookhaven National Laboratory Jian Wang, Brookhaven National Laboratory Measurement

More information

The Scanning DMA Transfer Function

The Scanning DMA Transfer Function Aerosol Science and Technology, 38:833 850, 2004 Copyright c American Association for Aerosol Research ISSN: 0278-6826 print / 1521-7388 online DOI: 10.1080/027868290503082 The Scanning DMA Transfer Function

More information

Mobility-Based Particle Size Classification and Detection down to 2nm Theory to Practice

Mobility-Based Particle Size Classification and Detection down to 2nm Theory to Practice Mobility-Based Particle Size Classification and Detection down to 2nm Theory to Practice Robert C. Anderson Presented at SEMATECH November 12, 2012 Current and Future Defectively Issues from Components

More information

Particle counting efficiencies of new TSI condensation particle counters

Particle counting efficiencies of new TSI condensation particle counters Aerosol Science 38 (27) 674 682 www.elsevier.com/locate/jaerosci Technical note Particle counting efficiencies of new TSI condensation particle counters M. Hermann a,, B. Wehner a, O. Bischof b, H.-S.

More information

Product data sheet Palas U-SMPS 1050 / 1100 / 1200

Product data sheet Palas U-SMPS 1050 / 1100 / 1200 Product data sheet Palas U-SMPS 1050 / 1100 / 1200 Applications Filter test Aerosol research Environmental and climatic studies Inhalation experiments Interior and workplace measurements Benefits Particle

More information

Differential Mobility Particle Sizer (Aerosol measurements)

Differential Mobility Particle Sizer (Aerosol measurements) Institute for Atmospheric and Climate Science - IACETH Atmospheric Physics Lab Work Differential Mobility Particle Sizer (Aerosol measurements) Abstract A differential mobility particle sizer (DMPS) is

More information

Product data sheet Palas Nanoparticle measurement system DEMC 2000

Product data sheet Palas Nanoparticle measurement system DEMC 2000 Product data sheet Palas Nanoparticle measurement system DEMC 2000 Applications Aerosol science Calibration of condensation particle counters (CPC) Monodisperse particle source System component of an SMPS

More information

ELECTROSTATIC CLASSIFIER MODEL 3082

ELECTROSTATIC CLASSIFIER MODEL 3082 ELECTROSTATIC CLASSIFIER MODEL 3082 THESE INSTRUMENTS HAVE BEEN USED IN A BROAD VARIETY OF RESEARCH AND HAVE EARNED A WELL-DESERVED REPUTATION FOR BEING HIGHLY RELIABLE AND EXTREMELY VERSATILE. UNDERSTANDING,

More information

Causes of Concentration Differences Between a Scanning Mobility Particle Sizer and a Condensation Particle Counter

Causes of Concentration Differences Between a Scanning Mobility Particle Sizer and a Condensation Particle Counter Aerosol Science and Technology, 37:916 923, 2003 Copyright c American Association for Aerosol Research ISSN: 0278-6826 print / 1521-7388 online DOI: 10.1080/02786820390229057 Causes of Concentration Differences

More information

Jianfei Peng et al. Correspondence to: Jianfei Peng Min Hu and Renyi Zhang

Jianfei Peng et al. Correspondence to: Jianfei Peng Min Hu and Renyi Zhang Supplement of Atmos. Chem. Phys., 17, 10333 10348, 2017 https://doi.org/10.5194/acp-17-10333-2017-supplement Author(s) 2017. This work is distributed under the Creative Commons Attribution 3.0 License.

More information

The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch. Symposium Stockholm Mai

The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch. Symposium Stockholm Mai The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch Symposium Stockholm Mai 2014 1 Outline Optical Particle Detection Method 11-R Main features of the GRIMM 11-R Measurement principle

More information

A new versatile condensation particle counter for research and environmental monitoring

A new versatile condensation particle counter for research and environmental monitoring A new versatile condensation particle counter for research and environmental monitoring, Maximilian Weiss Palas GmbH, Greschbachstr. 3b, 76229 Karlsruhe, Germany, http://www.palas.de, spielvogel@palas.de

More information

PLEASE SCROLL DOWN FOR ARTICLE. Full terms and conditions of use:

PLEASE SCROLL DOWN FOR ARTICLE. Full terms and conditions of use: This article was downloaded by: [California Institute of Technology] On: 28 February 2011 Access details: Access Details: [subscription number 917270853] Publisher Taylor & Francis Informa Ltd Registered

More information

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for: A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for: generating monodisperse aerosol by mobility classification with automatic concentration detection

More information

Particle Size Distribution Measurements with the Novel 1 nm-smps

Particle Size Distribution Measurements with the Novel 1 nm-smps Particle Size Distribution Measurements with the Novel 1 nm-smps Fast Scanning at few Nanometers Torsten Tritscher 1 F. Dahlkötter 1, C. Kykal 1, J.H.T. Scheckman 1, J. Spielvogel 1, T. Krinke 1, E. Filimundi

More information

MEASURING NANOPARTICLE SIZE DISTRIBUTIONS IN REAL-TIME: KEY FACTORS FOR ACCURACY

MEASURING NANOPARTICLE SIZE DISTRIBUTIONS IN REAL-TIME: KEY FACTORS FOR ACCURACY MEASURING NANOPARTICLE SIZE DISTRIBUTIONS IN REAL-TIME: KEY FACTORS FOR ACCURACY APPLICATION NOTE SMPS-003 Introduction The benefits of sizing aerosolized submicrometer particles using an electrical mobility

More information

DMA Size-Selection and Electrostatic Deposition of Particle Size Standards Down to 10nm

DMA Size-Selection and Electrostatic Deposition of Particle Size Standards Down to 10nm DMA Size-Selection and Electrostatic Deposition of Particle Size Standards Down to 10nm Ben Hunt, William Dick, Zeeshan Syedain MSP Corporation 1 AAAR 2013 3IM.4 Presentation Outline Particle Deposition

More information

Measuring aerosol size distributions with the fast integrated mobility spectrometer

Measuring aerosol size distributions with the fast integrated mobility spectrometer Aerosol Science 39 (28) 94 956 www.elsevier.com/locate/jaerosci Measuring aerosol size distributions with the fast integrated mobility spectrometer Jason S. Olfert a, Pramod Kulkarni b, Jian Wang a, a

More information

ISO INTERNATIONAL STANDARD. Determination of particle size distribution Differential electrical mobility analysis for aerosol particles

ISO INTERNATIONAL STANDARD. Determination of particle size distribution Differential electrical mobility analysis for aerosol particles INTERNATIONAL STANDARD ISO 15900 First edition 2009-05-15 Determination of particle size distribution Differential electrical mobility analysis for aerosol particles Détermination de la distribution granulométrique

More information

Introduction and Initial Field Data of a Novel, Portable Nanoparticle Sizing Instrument

Introduction and Initial Field Data of a Novel, Portable Nanoparticle Sizing Instrument Introduction and Initial Field Data of a Novel, Portable Nanoparticle Sizing Instrument Torsten Tritscher 1, Thomas J. Krinke 1, Axel F. Zerrath 2, Erik Willis 2 and Oliver F. Bischof 1 1 TSI GmbH, Particle

More information

Chapter 2: Radial Differential Mobility Analyzer for One

Chapter 2: Radial Differential Mobility Analyzer for One 14 Chapter 2: Radial Differential Mobility Analyzer for One Nanometer Particle Classification * 2.1. Introduction Nanometer-sized aerosol particles are encountered extensively in research ranging from

More information

NANOPARTICLE MEASUREMENT SYSTEMS

NANOPARTICLE MEASUREMENT SYSTEMS NANOPARTICLE MEASUREMENT SYSTEMS Reliable determination of particle size and number in the nanoscale range from 4 nm to 1,400 nm 1 APPLICATIONS Research and development, environmental measurements Quality

More information

Particle Number Counter Calibration/Validation Procedures. Report to the Department for Transport

Particle Number Counter Calibration/Validation Procedures. Report to the Department for Transport Particle Number Counter Calibration/Validation Procedures Report to the Department for Transport Restricted Commercial ED47382004/PNC Issue Number Draft Revision 4 August 2007 Restricted Commercial PNC

More information

Bipolar Neutralization using Radioactive, X-ray, and AC Corona Methods

Bipolar Neutralization using Radioactive, X-ray, and AC Corona Methods Bipolar Neutralization using Radioactive, X-ray, and AC Corona Methods Jacob Swanson, Jean de La Verpillière, and Adam Boies University of Cambridge js2011@cam.ac.uk 2013 Cambridge Particle Meeting 1 Outline

More information

Differential Mobility Analysis of Aerosols: A Tutorial

Differential Mobility Analysis of Aerosols: A Tutorial Differential Mobility Analysis of Aerosols: A Tutorial Richard C. Flagan Division of Chemistry and Chemical Engineering 210-41 California Institute of Technology 1 Abstract Differential mobility analysis

More information

SOP SMPS BaltimorePM supersite Version /29/01 Page 1 of 13 SMPS

SOP SMPS BaltimorePM supersite Version /29/01 Page 1 of 13 SMPS Page 1 of 13 SOP DRAFT - MEASUREMENT AND ANALYSIS OF PARTICLE NUMBER SIZE DISTRIBUTIONS IN AMBIENT AIR USING THE Identification code: SOP APPROVALS OP Working SOP pages Issue Date: /. Local PI: / / Revision

More information

Mobility particle size spectrometers: Calibration procedures and measurement uncertainties

Mobility particle size spectrometers: Calibration procedures and measurement uncertainties Aerosol Science and Technology ISSN: 0278-6826 (Print) 1521-7388 (Online) Journal homepage: http://www.tandfonline.com/loi/uast20 Mobility particle size spectrometers: Calibration procedures and measurement

More information

A Differential Mobility Analyzer (DMA) for Size Selection of Nanoparticles at High Flow Rates

A Differential Mobility Analyzer (DMA) for Size Selection of Nanoparticles at High Flow Rates Aerosol Science and Technology ISSN: 0278-6826 (Print) 1521-7388 (Online) Journal homepage: http://www.tandfonline.com/loi/uast20 A Differential Mobility Analyzer (DMA) for Size Selection of Nanoparticles

More information

Corona Discharge Characteristics and Particle Losses in a Unipolar Corona-needle Charger Obtained through Numerical and Experimental Studies

Corona Discharge Characteristics and Particle Losses in a Unipolar Corona-needle Charger Obtained through Numerical and Experimental Studies J Electr Eng Technol.2017; 12(?): 1921-718 http://doi.org/10.???/jeet.2017.12.3.1921 ISSN(Print) 1975-0102 ISSN(Online) 2093-7423 Corona Discharge Characteristics and Particle Losses in a Unipolar Corona-needle

More information

Brian Sargent, Assistant Director, Designee 3/21/18. See Thread for Approval

Brian Sargent, Assistant Director, Designee 3/21/18. See  Thread for Approval Cali Jones Cali Jones, Coordinator 3/21/18 Brian Sargent Brian Sargent, Assistant Director, Designee 3/21/18 See Email Thread for Approval 3/21/18 @ 3:30 p.m. 3/26/18 @ 3:30 p.m. 1818 Cali Jones Cali Jones

More information

A New Volatility Tandem Differential Mobility Analyzer to Measure the Volatile Sulfuric Acid Aerosol Fraction

A New Volatility Tandem Differential Mobility Analyzer to Measure the Volatile Sulfuric Acid Aerosol Fraction 760 JOURNAL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY VOLUME 16 A New Volatility Tandem Differential Mobility Analyzer to Measure the Volatile Sulfuric Acid Aerosol Fraction D. A. ORSINI, A.WIEDENSOHLER, AND

More information

Recent progress and future prospects for traceable submicron aerosol measurements

Recent progress and future prospects for traceable submicron aerosol measurements Recent progress and future prospects for traceable submicron aerosol measurements 9th June 2010 Richard Gilham richard.gilham@npl.co.uk Summary Requirements for traceability Application to submicron aerosol

More information

Comprehensive Measurement of Atmospheric Aerosols with a Wide Range Aerosol Spectrometer

Comprehensive Measurement of Atmospheric Aerosols with a Wide Range Aerosol Spectrometer Comprehensive Measurement of Atmospheric Aerosols with a Wide Range Aerosol Spectrometer Juergen Spielvogel, Lothar Keck, Xiaoai Guo, Markus Pesch Email: jsp@grimm-aerosol.com GRIMM Aerosol Technik GmbH

More information

New Instruments from GRIMM

New Instruments from GRIMM New Instruments from GRIMM Markus Pesch Symposium at Stockholm, 01.11.2016 1 Outline Motivation for new Aerosol measurement devices Objectives for measurements of particle number and sizes GRIMM EDM 665

More information

Generation and Evaluation of Monodisperse Sodium Chloride and Oleic Acid Nanoparticles

Generation and Evaluation of Monodisperse Sodium Chloride and Oleic Acid Nanoparticles Generation and Evaluation of Monodisperse Sodium Chloride and Oleic Acid Nanoparticles Tzu Ming Chen, Hung Min Chein Energy & Environment Research Laboratories, Industrial Technology Research Institute,

More information

NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER

NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER APPLICATION NOTE NANOSCAN-002 (A4) Scanning Mobility Particle Sizer (SMPS ) Spectrometer TSI s SMPS Spectrometer is widely used as the standard

More information

Calibration checks of particle counter using primary and other techniques at the laboratory level

Calibration checks of particle counter using primary and other techniques at the laboratory level Calibration checks of particle counter using primary and other techniques at the laboratory level Shankar G. Aggarwal, Ph.D. CSIR-National Physical Laboratory, New Delhi-110 012, India @ Particulate Workshop

More information

High-Efficiency Unipolar Charger for Sub-10 nm Aerosol Particles Using Surface-Discharge Microplasma with a Voltage of Sinc Function

High-Efficiency Unipolar Charger for Sub-10 nm Aerosol Particles Using Surface-Discharge Microplasma with a Voltage of Sinc Function Aerosol Science and Technology ISSN: 0278-6826 (Print) 1521-7388 (Online) Journal homepage: http://www.tandfonline.com/loi/uast20 High-Efficiency Unipolar Charger for Sub-10 nm Aerosol Particles Using

More information

A method of testing size resolution of a particle sizing instrument Toward traceable particle measurements

A method of testing size resolution of a particle sizing instrument Toward traceable particle measurements A method of testing size resolution of a article sizing instrument Toward traceable article measurements K. Ehara, Y. Sato, H. Sakurai, and K. Takahata National Metrology Institute of Jaan National Institute

More information

Intercomparison of Mobility Particle Size Spectrometers

Intercomparison of Mobility Particle Size Spectrometers Intercomparison of Mobility Particle Size Spectrometers Project No.: MPSS- 01-3- Basic information: Location of the quality assurance:, lab: 11 Delivery date: April 1, 01 Setup in the laboratory: April

More information

System for In Situ Characterization of Nanoparticles Synthesized in a Thermal Plasma Process

System for In Situ Characterization of Nanoparticles Synthesized in a Thermal Plasma Process Plasma Chemistry and Plasma Processing, Vol. 25, No. 5, October 2005 ( 2005) DOI: 10.1007/s11090-005-4991-4 System for In Situ Characterization of Nanoparticles Synthesized in a Thermal Plasma Process

More information

Intercomparison of Mobility Particle Size Spectrometers

Intercomparison of Mobility Particle Size Spectrometers Intercomparison of Mobility Particle Size Spectrometers Project No.: MPSS-2015-2-1 Basic information: Location of the quality assurance: TROPOS, lab: 118 Delivery date: - Setup in the laboratory: December

More information

Generation of monodisperse aerosols through condensation nuclei control

Generation of monodisperse aerosols through condensation nuclei control Air Pollution XV 505 Generation of monodisperse aerosols through condensation nuclei control H. M. Kadlimatti 1, S. Gangamma 2 & S. K. Varghese 3 1 Department of Mechanical Engineering, Basaveshwar Engineering

More information

A Wide-Range Particle Spectrometer for Aerosol Measurement from

A Wide-Range Particle Spectrometer for Aerosol Measurement from Aerosol and Air Quality Research, 10: 125 139, 2010 Copyright Taiwan Association for Aerosol Research ISSN: 1680-8584 print / 2071-1409 online doi: 10.4209/aaqr.2009.10.0062 A Wide-Range Particle Spectrometer

More information

Nirma University Institute of Technology Chemical Engineering Department, Handouts -RRP- CRE-II. Handouts

Nirma University Institute of Technology Chemical Engineering Department, Handouts -RRP- CRE-II. Handouts Handouts Handout 1: Practical reactor performance deviates from that of ideal reactor s : Packed bed reactor Channeling CSTR & Batch Dead Zones, Bypass PFR deviation from plug flow dispersion Deviation

More information

Helsinki, Finland. 1 Aerodyne Research Inc., Billerica, MA, USA. 2 Chemistry Department, Boston College, Chestnut Hill, MA, USA

Helsinki, Finland. 1 Aerodyne Research Inc., Billerica, MA, USA. 2 Chemistry Department, Boston College, Chestnut Hill, MA, USA Supplementary material to article: Relationship between aerosol oxidation level and hygroscopic properties of laboratory generated secondary organic aerosol (SOA) particles (paper #0GL0 to Geophysical

More information

Particle Number Counter Calibration Procedure. Report to the Department for Transport

Particle Number Counter Calibration Procedure. Report to the Department for Transport Particle Number Counter Calibration Procedure Report to the Department for Transport Restricted Commercial ED47382004/PNC Issue Number Draft Revision 3 July 2007 Restricted Commercial PNC Calibration

More information

MEASURING NANOPARTICLE EXPOSURE

MEASURING NANOPARTICLE EXPOSURE MEASURING NANOPARTICLE EXPOSURE APPLICATION NOTE NSAM-001 The Nanoparticle Surface Area Monitor reports the surface area of inhaled particles deposited in the lung Nanoparticle Exposure There is increasing

More information

Intercomparison of Mobility Particle Size Spectrometers

Intercomparison of Mobility Particle Size Spectrometers Intercomparison of Mobility Particle Size Spectrometers Project No.: MPSS-2015-1-1 Basic information: Location of the quality assurance: TROPOS, lab: 118 Delivery date: October 05, 2015 Setup in the laboratory:

More information

Aerosol Charging Using Pen-Type UV Lamps

Aerosol Charging Using Pen-Type UV Lamps Aerosol and Air Quality Research, 11: 791 81, 11 Copyright Taiwan Association for Aerosol Research ISSN: 168-8584 print / 71-149 online doi: 1.49/aaqr.11.7.13 Aerosol Charging Using Pen-Type UV Lamps Lin

More information

USING THE ELECTRET FILTER TO REMOVE THE SUBMICRON AEROSOLS

USING THE ELECTRET FILTER TO REMOVE THE SUBMICRON AEROSOLS USING THE ELECTRET FILTER TO REMOVE THE SUBMICRON AEROSOLS SH Yang 1,*, GWM Lee 1, CH Luo 2, CC Wu 1, KP Yu 1, CL Lou 1 1 Graduate Institute of Environmental Engineering, National Taiwan University, 71,

More information

Particle Collection and Concentration for Cyclone Concentrators

Particle Collection and Concentration for Cyclone Concentrators Aerosol Science and Technology, 39:113 123, 2005 Copyright c American Association for Aerosol Research ISSN: 0278-6826 print / 1521-7388 online DOI: 10.1080/027868290907138 Particle Collection and Concentration

More information

Highly Biased Hygroscopicity Derived from Size-Resolved Cloud Condensation Nuclei Activation Ratios without Data Inversion

Highly Biased Hygroscopicity Derived from Size-Resolved Cloud Condensation Nuclei Activation Ratios without Data Inversion ATMOSPHERIC AND OCEANIC SCIENCE LETTERS, 2014, VOL. 7, NO. 3, 254 259 Highly Biased Hygroscopicity Derived from Size-Resolved Cloud Condensation Nuclei Activation Ratios without Data Inversion DENG Zhao-Ze

More information

Intercomparison Study of the Size-Dependent Counting Efficiency of 26 Condensation Particle

Intercomparison Study of the Size-Dependent Counting Efficiency of 26 Condensation Particle Aerosol Science and Technology SSN: 0278-6826 (Print) 1521-7388 (Online) Journal homepage: https://www.tandfonline.com/loi/uast20 ntercomparison Study of the Size-Dependent Counting Efficiency of 26 Condensation

More information

Particuology 9 (2011) Contents lists available at SciVerse ScienceDirect. Particuology. j o ur nal homep ag e:

Particuology 9 (2011) Contents lists available at SciVerse ScienceDirect. Particuology. j o ur nal homep ag e: Particuology 9 (2011) 606 610 Contents lists available at SciVerse ScienceDirect Particuology j o ur nal homep ag e: www.elsevier.com/locate/partic Evaluation of particle growth systems for sampling and

More information

Calibration of PMP Condensation Particle Number Counters

Calibration of PMP Condensation Particle Number Counters Calibration of PMP Condensation Particle Number Counters Effect of material on linearity and counting efficiency B. Giechaskiel, S. Alessandrini, F. Forni, M. Carriero, A. Krasenbrink European Commission,

More information

Characterisation of a new Fast CPC and its application for atmospheric particle measurements

Characterisation of a new Fast CPC and its application for atmospheric particle measurements Atmos. Meas. Tech., 4, 823 833, 2011 doi:10.5194/amt-4-823-2011 Author(s) 2011. CC Attribution 3.0 License. Atmospheric Measurement Techniques Characterisation of a new Fast CPC and its application for

More information

Supplementary information

Supplementary information 1 Supplementary information Instrument and operation 3 6 7 8 9 10 11 1 13 1 1 16 HTDMA: Briefly, the custom-built HTDMA consists of two long DMAs (3081L, TSI Inc.), a humidifier (PD-0T-1MSS, Perma Pure

More information

Intercomparison of Mobility Particle Size Spectrometers

Intercomparison of Mobility Particle Size Spectrometers Intercomparison of Mobility Particle Size Spectrometers Project No.: MPSS- 201- - 1 Basic information: Principal Investigator: Home Institution: Participant: Instrument No.1: Eija Asmi Finnish Meteorological

More information

Evaluation of the NanoAerosol Generator by Kanomax FMT Inc. in Aerosolization of Size Standard Nanoparticle and Protein

Evaluation of the NanoAerosol Generator by Kanomax FMT Inc. in Aerosolization of Size Standard Nanoparticle and Protein Aerosol and Air Quality Research, 18: 549 554, 2018 Copyright Taiwan Association for Aerosol Research ISSN: 1680-8584 print / 2071-1409 online doi: 10.4209/aaqr.2017.09.0352 Technical Note Evaluation of

More information

Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping

Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping Division of Nano metrology and Materials Measurement National

More information

Nanoparticle Trajectories In An Electrostatic Precipitator: Numerical Simulation And Experimental Validation

Nanoparticle Trajectories In An Electrostatic Precipitator: Numerical Simulation And Experimental Validation Extended Abstract Nanoparticle Trajectories In An Electrostatic Precipitator: Numerical Simulation And Experimental Validation Motivation 16 th ETH-Conference on Combustion Generated Nanoparticles June

More information

Chemical Reaction Engineering

Chemical Reaction Engineering Lecture 32! Chemical Reaction Engineering (CRE) is the field that studies the rates and mechanisms of chemical reactions and the design of the reactors in which they take place.!! 1! Lecture 32 Thursday

More information

Improving the Nanoparticle Resolution of the ELPI

Improving the Nanoparticle Resolution of the ELPI Aerosol and Air Quality Research, 1: 36 366, 21 Copyright Taiwan Association for Aerosol Research ISSN: 168-8584 print / 271-149 online doi: 1.429/aaqr.29.1.6 Improving the Nanoparticle Resolution of the

More information

COUNTS SURFACE VOLUME MASS COLLECTION GENERATION

COUNTS SURFACE VOLUME MASS COLLECTION GENERATION COUNTS SURFACE VOLUME MASS COLLECTION GENERATION GRIMM and Nanoparticles Dear customer, COUNTING SIZING Jürgen Spielvogel Nano Division Manager jsp@grimm-aerosol.com for almost 0 years GRIMM provides you

More information

Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator.

Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator. Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator Hiromu Sakurai hiromu.sakurai@aist.go.jp Particle Measurement Research Group National

More information

An Ultrafine Aerosol Condensation Nucleus Counter

An Ultrafine Aerosol Condensation Nucleus Counter Aerosol Science and Technology ISSN: 0278-6826 (Print) 1521-7388 (Online) Journal homepage: http://www.tandfonline.com/loi/uast20 An Ultrafine Aerosol Condensation Nucleus Counter Mark R. Stolzenburg &

More information

Defined calibration of the particle measuring system according to PMP

Defined calibration of the particle measuring system according to PMP Defined calibration of the particle measuring system according to PMP Dipl.-Ing. Martin Schmidt Dr.-Ing. Maximilian Weiß, Dipl.-Ing. Leander Mölter, Managing Partner Palas GmbH, Karlsruhe, Germany Contents

More information

High Temperature Condensation Particle Counter (HT- CPC)

High Temperature Condensation Particle Counter (HT- CPC) High Temperature Condensation Particle Counter (HT- CPC) Jeng K Rongchai and Nick Collings Outline Motivation The High Temperature CPC (HT-CPC) Desirable characteristics and working fluids Modelling Experiments

More information

Chapter 1. Introduction

Chapter 1. Introduction Chapter 1. Introduction 1-1 1.1. Overview In the past twenty years, charged droplets and strong electric fields have quietly revolutionized chemistry. In combination with an atmospheric-sampling mass spectrometer,

More information

Analysis of Data from the 2009 SOOT Experiment

Analysis of Data from the 2009 SOOT Experiment Analysis of Data from the 2009 SOOT Experiment Renyi Zhang Department of Atmospheric Sciences and Department of Chemistry Center for Atmospheric Chemistry and the Environment Texas A&M University College

More information

Intercomparison of Mobility Particle Size Spectrometers

Intercomparison of Mobility Particle Size Spectrometers Intercomparison of Mobility Particle Size Spectrometers Project No.: Principal Investigator: Home Institution: Participant: Candidate: Made by: Counter (SN): Software: MPSS-2016-5-1 Christine Braban NERC

More information

Site audit report Birkenes, Norway

Site audit report Birkenes, Norway Site audit report Birkenes, Norway Dr. Thomas Tuch, WCCAP World Calibration Centre for Aerosol Physics (WCCAP), Leibniz Institute for Tropospheric Research, Leipzig, Germany Summary: Measurements of physical

More information

7. Timescale for hygroscopic conversion of calcite mineral particles through heterogeneous reaction with nitric acid

7. Timescale for hygroscopic conversion of calcite mineral particles through heterogeneous reaction with nitric acid Supplementary Information for: 7. Timescale for hygroscopic conversion of calcite mineral particles through heterogeneous reaction with nitric acid Ryan C. Sullivan, 1, Meagan J. K. Moore, 1 Markus D.

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Chapter 5: Nanoparticle Production from Cathode Sputtering. in High-Pressure Microhollow Cathode and Arc Discharges

Chapter 5: Nanoparticle Production from Cathode Sputtering. in High-Pressure Microhollow Cathode and Arc Discharges 96 Chapter 5: Nanoparticle Production from Cathode Sputtering in High-Pressure Microhollow Cathode and Arc Discharges 5.1. Introduction Sputtering is a fundamental aspect of plasma operation and has been

More information

A. Wiedensohler, W. Birmili, A. Nowak, A. Sonntag, K. Weinhold, M. Merkel, B. Wehner, T. Tuch, S. Pfeifer, M. Fiebig, et al.

A. Wiedensohler, W. Birmili, A. Nowak, A. Sonntag, K. Weinhold, M. Merkel, B. Wehner, T. Tuch, S. Pfeifer, M. Fiebig, et al. Mobility particle size spectrometers: harmonization of technical standards and data structure to facilitate high quality long-term observations of atmospheric particle number size distributions A. Wiedensohler,

More information

DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER

DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER Metrol. Meas. Syst., Vol. XVIII (011), No. 1, pp. 19-136 METROLOGY AND MEASUREMENT SYSTEMS Index 330930, ISSN 0860-89 www.metrology.pg.gda.pl DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER

More information

EXPERIMENT No.1 FLOW MEASUREMENT BY ORIFICEMETER

EXPERIMENT No.1 FLOW MEASUREMENT BY ORIFICEMETER EXPERIMENT No.1 FLOW MEASUREMENT BY ORIFICEMETER 1.1 AIM: To determine the co-efficient of discharge of the orifice meter 1.2 EQUIPMENTS REQUIRED: Orifice meter test rig, Stopwatch 1.3 PREPARATION 1.3.1

More information

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT 62 CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT 5.1 INTRODUCTION The primary objective of this work is to investigate the convective heat transfer characteristics of silver/water nanofluid. In order

More information

Study of the Mechanism of Nucleation in the Polluted Atmospheric Boundary Layer

Study of the Mechanism of Nucleation in the Polluted Atmospheric Boundary Layer Study of the Mechanism of Nucleation in the Polluted Atmospheric Boundary Layer A DISSERTATION SUBMITTED TO THE FACULTY OF UNIVERSITY OF MINNESOTA BY Modi Chen IN PARTIAL FULFILLMENT OF THE REQUIREMENTS

More information

Effect of aging on cloud nucleating properties of atmospheric aerosols

Effect of aging on cloud nucleating properties of atmospheric aerosols Effect of aging on cloud nucleating properties of atmospheric aerosols Yan Ma Nanjing University of Information Science & Technology The 1st Regional GEOS-Chem Asia Meeting 2015 年 5 月 8 日 Indirect effect

More information

Particle Impaction Patterns from a Circular Jet

Particle Impaction Patterns from a Circular Jet Aerosol Science and Technology ISSN: 0278-6826 (Print) 1521-7388 (Online) Journal homepage: http://www.tandfonline.com/loi/uast20 Particle Impaction Patterns from a Circular Jet Virendra Sethi & Walter

More information

FACULTY OF CHEMICAL & ENERGY ENGINEERING FLUID MECHANICS LABORATORY TITLE OF EXPERIMENT: MINOR LOSSES IN PIPE (E4)

FACULTY OF CHEMICAL & ENERGY ENGINEERING FLUID MECHANICS LABORATORY TITLE OF EXPERIMENT: MINOR LOSSES IN PIPE (E4) FACULTY OF CHEMICAL & ENERGY ENGINEERING FLUID MECHANICS LABORATORY TITLE OF EXPERIMENT: MINOR LOSSES IN PIPE (E4) 1 1.0 Objectives The objective of this experiment is to calculate loss coefficient (K

More information

Experimental investigation on up-flow boiling of R1234yf in aluminum multi-port extruded tubes

Experimental investigation on up-flow boiling of R1234yf in aluminum multi-port extruded tubes Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 2016 Experimental investigation on up-flow boiling of R1234yf in aluminum multi-port

More information

GAW - WCCAP recommendation for aerosol inlets and sampling tubes

GAW - WCCAP recommendation for aerosol inlets and sampling tubes GAW - WCCAP recommendation for aerosol inlets and sampling tubes Alfred Wiedensohler, Wolfram Birmili, Leibniz Institute for Tropospheric Research, Leipzig, Germany John Ogren, NOAA ESRL GMD, Boulder,

More information

COOLING TOWER MODEL IN MNR

COOLING TOWER MODEL IN MNR COOLING TOWER MODEL IN MNR 1 COOLING TOWER MODEL The cooling tower model used in Liu s study [2] was adapted from the model developed by Eric Weber (1988). One modification is the use of a counterflow

More information

Fundamentals of Particle Counting

Fundamentals of Particle Counting Fundamentals of Particle Counting 1 Particle Counting: Remains the most significant technique for determining the cleanliness level of a fluid Useful as a tool for qualification and monitoring cleanroom

More information

Validation of a new flow-reactor for the study of secondary organic aerosol (SOA) formation

Validation of a new flow-reactor for the study of secondary organic aerosol (SOA) formation Validation of a new flow-reactor for the study of secondary organic aerosol (SOA) formation M. Duncianu*(1,2), V. Riffault (1,2), A. Tomas (1,2), P. Coddeville (1,2) (1) Université Lille Nord de France,

More information

Supersaturation in the Wyoming CCN Instrument

Supersaturation in the Wyoming CCN Instrument OCTOBER 2006 S N I D E R E T A L. 1323 Supersaturation in the Wyoming CCN Instrument JEFFERSON R. SNIDER Department of Atmospheric Science, University of Wyoming, Laramie, Wyoming MARKUS D. PETTERS Department

More information

EFFECT OF BAFFLES GEOMETRY ON HEAT TRANSFER ENHANCEMENT INSIDE CORRUGATED DUCT

EFFECT OF BAFFLES GEOMETRY ON HEAT TRANSFER ENHANCEMENT INSIDE CORRUGATED DUCT International Journal of Mechanical Engineering and Technology (IJMET) Volume 10, Issue 03, March 2019, pp. 555-566. Article ID: IJMET_10_03_057 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=10&itype=3

More information

Supplement of Total sulfate vs. sulfuric acid monomer concenterations in nucleation studies

Supplement of Total sulfate vs. sulfuric acid monomer concenterations in nucleation studies Supplement of Atmos. Chem. Phys., 15, 3429 3443, 2015 http://www.atmos-chem-phys.net/15/3429/2015/ doi:10.5194/acp-15-3429-2015-supplement Author(s) 2015. CC Attribution 3.0 License. Supplement of Total

More information

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH Submitted By: ABDULLAH IBN ABDULRAHMAN ID: 13456789 GROUP A EXPERIMENT PERFORMED

More information

Using a Fast-Scanning Electrical Nanoparticle Sizer to Characterize Nanoparticles from Laser Ablation

Using a Fast-Scanning Electrical Nanoparticle Sizer to Characterize Nanoparticles from Laser Ablation Using a Fast-Scanning Electrical Nanoparticle Sizer to Characterize Nanoparticles from Laser Ablation Chaolong Qi 1, Da-Ren Chen 1*, Meng-Dawn Cheng 2 1 Environmental Engineering Science Program Washington

More information

Model 2300XP PSL & Process-Particle Wafer Deposition System

Model 2300XP PSL & Process-Particle Wafer Deposition System Model 2300XP PSL & Process-Particle Wafer Deposition System Deposit PSL spheres on wafers to create NISTtraceable PSL size standards for - calibrating wafer inspection systems - providing fab-wide and

More information

Aerosol size distribution estimation and associated uncertainty for measurement with a Scanning Mobility Particle Sizer (SMPS)

Aerosol size distribution estimation and associated uncertainty for measurement with a Scanning Mobility Particle Sizer (SMPS) Journal of Physics: Conference Series Aerosol size distribution estimation and associated uncertainty for measurement with a Scanning Mobility Particle Sizer (SMPS) To cite this article: L Coquelin et

More information

Experimental Study on the Non-reacting Flowfield of a Low Swirl Burner

Experimental Study on the Non-reacting Flowfield of a Low Swirl Burner Experimental Study on the Non-reacting Flowfield of a Low Swirl Burner Hang Yin & Ren Dai School of Energy and Powering Engineering, University of Shanghai for Science and Technology Box 25, 516# Jungong

More information

Related U.S. Application Data 60 Provisional application No. 60/005,098, Oct. 12, (51) Int. Cl... G01N 15/02

Related U.S. Application Data 60 Provisional application No. 60/005,098, Oct. 12, (51) Int. Cl... G01N 15/02 United States Patent Russell et al. 19 USOO5922976A 11 Patent Number: 5,922,976 (45) Date of Patent: Jul. 13, 1999 54 METHOD OF MEASURING AEROSOL PARTICLES USING AUTOMATED MOBILITY-CLASSIFIED AEROSOL DETECTOR

More information

Meetprincipes voor meting van nanodeeltjes in de lucht

Meetprincipes voor meting van nanodeeltjes in de lucht Meetprincipes voor meting van nanodeeltjes in de lucht Andreas Schmidt-Ott Nanostructured Materials, Faculty of Applied Sciences, TU Delft Qauntum dots van verschillende afmetingen geven licht in verschillende

More information

Experimental Investigation of the Velocity Distribution near the Swirl Generator of a Uniflow Cyclone for Performance Data Prediction

Experimental Investigation of the Velocity Distribution near the Swirl Generator of a Uniflow Cyclone for Performance Data Prediction Experimental Investigation of the Velocity Distribution near the Swirl Generator of a Uniflow Cyclone for Performance Data Prediction M. Pillei 1,2,*, R. Goller 1, T. Kofler 1, A. Wierschem 2, M. Kraxner

More information