Fundamentals of Particle Counting
|
|
- Hilda Stanley
- 5 years ago
- Views:
Transcription
1 Fundamentals of Particle Counting 1
2 Particle Counting: Remains the most significant technique for determining the cleanliness level of a fluid Useful as a tool for qualification and monitoring cleanroom operations The preferred means of classification testing for cleanrooms (ISO 14644) Required for continuous monitoring in GMP for aseptic operations (US, EU and International GMPs)
3 Semiconductor Flat Panel Aerospace Pharmaceutical High Technology Industries Needing Contamination Control Data Storage
4 Automotive Food/Beverage Defense/Homeland Security Medical High Technology Industries Needing Contamination Control Air Quality
5 Optical Solar Coatings Hospital High Technology Industries Needing Contamination Control Electronics
6 What is a Particle Counter? A Device Used to Count and Size Discrete Particles in a Fluid. A Fluid is: A substance, as a liquid or gas, that is capable of flowing and that changes its shape at a steady rate when acted upon by a force tending to change its shape. Air Liquid Gas Utilizing Either: Light Obscuration (Light Blocking) Light Scattering
7 Components of a Particle Counter Sample Input User Interface: Touch Screen Display Printer
8 Components of a Particle Counter Sample Input HEPA Filter Exhaust
9 Components of a Particle Counter Sample Input The Particle Detector Light Stop Photo Detector Counter Electronics Laser Diode Vacuum Pump Flow Control
10 Particle Counter Terms Particle: Solid Material Ranging in Size from 5nm to 100µm Particle Counter: Device Used to Count and Size Individual Particles Sample Time: Duration of Sample Sample Flow Rate: Volume of Fluid Passing Through the Particle Counter per Unit Time Cubic foot / minute Milliliter / minute View Volume (Percentage): Percentage of Sample Measured Sample Volume: Sample Flow Rate X View Volume X Sample Time Counting Efficiency: Comparison of Particle Counter Under test to More Sensitive Instrument Refractive Index: The ratio of the velocity of light in vacuum to the velocity of light in a medium
11 Particle Counter Terms Sensitivity: Minimum Detectable Particle Size Zero Count Level (False Count): Number of False Counts When Inlet Fluid Stream is Particle Free Concentration Limits: Maximum Particle Concentration Permitted for Particle Counter Operation allowing no more then a 10% Coincidence Error Coincidence Error: Counting Error from More Then One Particle Entering View Volume at the Same Time Channels: Particle Sizing Bins Cumulative: Mode of Particle Counter Operation Indicating Particle Counts for the Particle Channel Size of Interest and all Particles Larger (for classification purposes, this is the mode to use) Differential: Mode of Particle Counter Operation Indicating Particle Counts for the Particle Channel Size of Interest up Until The Next Larger Particle Size Channel
12 Important Terminology Flow Rate Amount of air that is measured over a period of time in a particle counter. Expressed in volume / time. (Cubic feet / minute, Liters / minute) 0.1 CFM or 1/10 th of a Cubic Foot of Air / Minute (2.83 Liters / Minute) 1.0 CFM or 1 Cubic Foot of Air / Minute (28.3 Liters / Minute) 50 LPM or 1.77 CFM (20 Minute Sample time for 1 Cubic Meter) 2.0 CFM or 56.6 LPM (17.6 Minute Sample time for 1 Cubic Meter) 75 LPM or CFM (13.3 Minute Sample time for 1 Cubic Meter) 100 LPM or 3.53 CFM (10 Minute Sample time for 1 Cubic Meter)
13 Important Terminology Cubic 1 ft Cubic Meter Foot 1 Meter 3 = 35.3 Feet 3 1 M 1 ft 1 ft 1 M 35 Minute Sample
14 Important Terminology Class Maximum Allowable Concentration of Particles Specified in a maximum number of particles per volume of air (either particles / cubic meter or cubic foot). Grade Air Cleanliness Levels specified for the manufacture of sterile medicinal products There are 4 grades A The local zone for high risk operations B For aseptic preparation and filling, this is the background environment for the grade A Zone C & D Clean areas for carrying out less critical stages in the manufacture of sterile products
15 Particle Size Comparison Chart
16 Relative Size Micron Scale Andenovirus Micron Herpes Virus 0.1 Micron Human Blood Cell 7 Micron E-Coli 0.5 Micron Human Hair Micron 35 Micron Visible Particles
17 Relative Size Gas Molecules (AMC) 2-50 Å ( µm) 0.3 µm particle 0.1 µm particle
18 The Problem: Contamination is all Around Us. 20µm Common Dust Mite
19 Particle Counting Basics Two Particle Detection Methods 1) Light Extinction (Light Blocking) Attenuation of light signal. Measurement of particles >1.0 micron 2) Light Scattering Redirected light energy (Scattering) Measurement of particles >0.05 micron
20 Light Interaction with a Particle Reflection Diffraction Absorption Refraction Internal Scattering When a particle is illuminated by a light beam, it is redirected or absorbed Reflection redirects light through surface interaction Refraction redirects light from the light entering the particle
21 Particle Counter Operation Laser Source Photo Detector
22 Types of Light Scattering: Mie Examples of Mie Scattering Clouds Fog I d2 λ 4 I = intensity of light scattered d = diameter of the Particle λ = wavelength of light Direction of Incident Light
23 Types of Light Scattering: Rayleigh Example of Raleigh blue sky I d6 λ 4 I = intensity of light scattered d = diameter of the Particle λ = wavelength of light Direction of Incident Light
24 Sizing Light scattered by a single particle in a specific direction in relation to the original direction has a unique signature which relates to the size, shape and material of the particle. As size is dominant the light intensity collected by the receiving optics of an instrument can be calibrated to the diameter of a sphere of a reference material, typically Polystyrene Latex. There is a dependency of the scattered signal on particle shape and material. However, it is a common practice to report particle size information in equivalent Polystyrene Latex sphere diameters.
25 Sizing Assumptions in Particle Counting: 1. All Particles are Spherical in Nature 2. All Particles have the Same Refractive Index of Polystyrene Latex
26 MCA Graph C O U N T S Peak to Valley Signal / Noise Millivolts Ideally Signal to Noise Ratio should be > 2:1
27 Counting Efficiency Test Question: If you have 100 particles coming through your particle counter and the instruments counting efficiency is 50% how many particles would you expect to count?
28 50% Counting Efficiency Example of Counting Efficiency at 0.5 Micron Particles C o u n t s Nominal Size 50% 70% 30% 50% point of in a monodisperse distribution Smaller than nominal size Larger than nominal size Particle size Definition: In a monodisperse distribution, the sensor detects all particles at the nominal size and larger. ISO requires that the counting efficiency of the smallest detectable size of a sensor be between 30 and 70%.
29 Noise Cumulative 1.0um Cumulative 0.7um Cumulative 0.5um Diff 0.3um Cumulative 0.3um Diff 0.5um Diff 0.7um Diff 1.0um 0.3um 0.5um 0.7um 1.0um 50% counting efficiency
30 Counting Efficiency Measurement Test Set-up Mono-Dispersed Aerosol Reference Particle Counter With 100% Counting Efficiency (for Particle Size of Interest) FLOW CONTROL P.S.L. Particles DISPERSION CHAMBER Tubing Lengths Precisely the Same Length Flow Rates on Both Counter Must be Equal AEROSOL GENERATOR Mono-Dispersed Aerosol Particle Counter Under Test FLOW CONTROL Components Needed: Particle Source (Polystyrene Latex Spheres) Aerosol Generator Dispersion Chamber Airborne Particle Counter (Test Unit) Airborne Particle Counter (Reference Instrument) External vacuum source when required Vacuum Pump If required for units without Built in Pump
31 Scope of ISO Instruments that conform to this part of ISO are used for the classification of air cleanliness in cleanrooms and associated controlled environments in accordance with ISO , as well as the measurement of number and size distribution of particles in various environments. The following are within the scope of this part of ISO 21501: 1. Size calibration 2. Verification of size setting 3. Counting efficiency 4. Size resolution 5. False count rate 6. Maximum particle number concentration 7. Sampling flow rate 8. Sampling time 9. Sampling volume 10. Calibration interval 11. Test report
32 Coincidence Error / Concentration Limits If particle counter is used in environments where the concentrations of particles are too high, more than one particle at a time can enter the view volume. This results in coincidence errors. Sensor View Volume = Normal This example shows one 0.5µm particle in the view volume and its output pulse. Sensor View Volume = High Concentration Exceeding Concentration Limit This example shows two 0.5µm particles in the view volume at the same time. The counter would report a size larger than 0.5µ and only one count. Per ISO 21501: Concentration limits are represented with no more then a 10% coincidence error.
33 Sample Delivery A vacuum pump is used to pull the sample air through the sensor. The flow is measured in cubic feet per minute (cfm). The use of some type of flow control is required to maintain the correct flow. 1.0 cfm pump 28.3 LPM 0.1 cfm pump 2.83 LPM
34 Critical Orifice In remote counters a critical orifice is used to regulate the flow through the sensor. A critical orifice is a precision opening that will only allow a certain amount of air flow once the critical vacuum pressure is applied. In the case of Lighthouse counters the critical pressure is 18 Hg. Regulated flow (1.0cfm or 0.1cfm Depending on the sensor) Critical orifice 18 Hg vacuum pressure
35 Counting Electronics Threshold Circuit This circuit is duplicated from one to 8 Channels depending on the model of particle counter. Analog Signal in +V +V CH3 CH2 Digital signals to counting circuitry +V CH. 3 CH. 2 CH. 1 CH1
36 Counting Electronics cont d Threshold Circuit This circuit is duplicated from one to 8 Channels depending on the model of particle counter. Analog Signal in +V +V CH3 CH2 Digital signals to counting circuitry +V CH. 3 CH. 2 CH. 1 CH1
37 Counting Modes (Cumulative & Differential) Cumulative Differential
38 Spec Sheet
39 Laminar Air Flow And Isokinetic Sampling LAMINAR AIR FLOW: UNIDIRECTIONAL, NON-TURBULENT. AIR FLOW. STREAM LINES DO NOT CROSS. A KEY ELEMENT TO FUNCTION OF CLASS 100 (ISO 5) OR BETTER ISOKINETIC SAMPLING: INLET VELOCITY AND DIRECTION TO SAMPLER ARE THE SAME AS THE FREE STREAM VELOCITY AND DIRECTION SUB-ISOKINETIC ERRORS SUPER-ISOKINETIC ERRORS WHAT ABOUT TURBULENT AIR FLOW SAMPLING?
40 Isokinetic Sampling HEPA Filter Laminar Flow Stream Ft/Min ( M/Sec) Air Stream Lines ISOKINETIC Probes ISOKINETIC SAMPLING: Inlet velocity and direction to the sample probe are the same as the air stream velocity and direction
41 Sub-Isokinetic Sampling HEPA Filter Laminar Flow Stream Ft/Min ( M/Sec) Air Stream Lines SUB-ISOKINETIC SAMPLING ERROR ISOKINETIC Probes SUB-ISOKINETIC SAMPLING ERROR: Underestimate the population of small particles because the probes flow rate is too low.
42 Super-Isokinetic Sampling HEPA Filter Laminar Flow Stream Ft/Min ( M/Sec) Air Stream Lines SUPER-ISOKINETIC SAMPLING ERROR ISOKINETIC Probes SUPER-ISOKINETIC SAMPLING ERROR: Overestimate the population of small particles because the probes flow rate is too high.
43 Turbulent Flow Cleanrooms HEPA Filter HEPA Filter ISOKINETIC Probes Non Laminar (Uni-Directional) Cleanrooms are TURBULENT flow Cleanrooms. Isokinetic Sampling is NOT possible in this type of cleanroom. Recommended proceedures for particle counting remain as if sampling in a Laminar (Uni-Directional Cleanroom)
44 Summary Counting Efficiency: Comparison of Particle Counter Under test to More Sensitive Instrument Sensitivity: Minimum Detectable Particle Size Zero Count Level (False Count): Number of False Counts When Inlet Fluid Stream is Particle Free Concentration Limits: Maximum Particle Concentration Permitted for Particle Counter Operation allowing no more then a 10% Coincidence Error Cumulative: Mode of Particle Counter Operation Indicating Particle Counts for the Particle Channel Size of Interest and all Particles Larger (for classification purposes, this is the mode to use) Differential: Mode of Particle Counter Operation Indicating Particle Counts for the Particle Channel Size of Interest up Until The Next Larger Particle Size Channel
45 Summary ISO Calibration Standard Yearly Calibration Annex 1: Sampling a Cubic Meter of Air
46 Thank You 46
Determination of particle size distribution Single particle light interaction methods. Part 4:
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 21501-4 Second edition 2018-05 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 21501-2 First edition 2007-05-15 Determination of particle size distribution Single particle light interaction methods Part 2: Light scattering liquid-borne particle counter
More informationTechnical Paper 3 Verifying Particle Counter Calibration: Size Verification
Technical Paper 3 Verifying Particle Counter Calibration: Size Verification By D. John Hunt, Pacific Scientific Instruments. Originally printed in Pollution Equipment News Publication, June 1996. Volume
More informationApproval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date
ATS-SOI-3660 Page: 1 of 6 Approval Block Prepared by: Signature Date Evan Parnell 08 NOV 2013 Reviewed by: Signature Date Brian Flynn 08 NOV 2013 Approved by: Signature Date Kristal Jewell 08 NOV 2013
More informationGRIMM Aerosol Spectrometer and Dust Monitors. Measuring principle
Grimm Aerosol-Technik GRIMM Aerosol Spectrometer and Dust Monitors Measuring principle Eng. Wolfgang Brunnhuber 1 Agenda Part A physical background general principles of optical particle detection Part
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 21501-4 First edition 2007-05-15 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne particle counter for
More informationCALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS
CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS APPLICATION NOTE CC-119 (A4) This application note works through the As Found Calibration Certificate of the AeroTrak Counters
More informationTSI AEROTRAK REMOTE 4-20 ma SENSORS LIFE SCIENCE
TSI AEROTRAK REMOTE 4-20 ma SENSORS LIFE SCIENCE APPLICATION NOTE CC-118 (A4) Introduction There is a need for end users to integrate AeroTrak Remote Particle Counters into third party building management
More informationPARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY
WHITEPAPER ENGLISH PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY PARTICLE MEASUREMENT Particle measurement in cleanrooms. WP1508006-0100-EN, V1R0, 2015-08 PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY
More informationOPTICAL PARTICLE SIZER MASS CALIBRATION METHOD
OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method
More informationSPECIFICATIONS PARTICLE SENSOR KS-28B Higashimotomachi, Kokubunji, Tokyo , Japan
SPECIFICATIONS PARTICLE SENSOR KS-28B 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 99020-4E 07-04 Printed in Japan Outline The KS-28B is a particle sensor designed to measure the number
More informationLatest Technology and Standardization Trends for Liquid-borne Particle Counters
Latest Technology and Standardization Trends for Liquid-borne Particle Counters Kaoru Kondo Rion Co., Ltd. (Original paper is K.Kondo, Bunseki,Vol.9, p499 (2012)) Abstract Fine particles contaminating
More informationSL CONTENT Page No 1 Devices Used 3
1 SL CONTENT Page No 1 Devices Used 3 1 Aerocet-531 Mass Particle Counter/Dust Monitor 3 2 Isokinetic Probe 3 3 Temperature and Humidity Sensing device 4 4 Hygrometer 4 5 Ball Flow Meter 4 2 Flow Rate
More informationTraceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping
Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping Division of Nano metrology and Materials Measurement National
More informationParticle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator.
Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator Hiromu Sakurai hiromu.sakurai@aist.go.jp Particle Measurement Research Group National
More informationApplication of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada
Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids October 26 Ottawa, Canada Summary The introduction of a growing number of targeted protein-based drug formulations
More informationQuality Test Center PERiMETER: Applications
Quality Test Center PERiMETER: Applications Surface characterization (coarseness) Cleanliness conditions fulfilling packaging Ultra precision cleaning Porosity and damage analysis CT-based coordinate measuring
More informationISO STANDARD INTERNATIONAL. Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness
INTERNATIONAL ISO STANDARD 4644- First edition 999-05-0 Cleanrooms and associated controlled environments Part : Classification of air cleanliness Salles propres et environnements maîtrisés apparentés
More informationMeasurement and Control of Liquid-borne Particles in the Semiconductor Manufacturing Process
Measurement and Control of Liquid-borne Particles in the Semiconductor Manufacturing Process Kaoru Kondo RION Co., Ltd. (Original paper is K.Kondo: Cleaning Technology, Vol.22, No.12, p.37 (Japanese) )
More informationCD stage. ISO/CD Hydraulic fluid power Calibration of automatic particle counters for liquids ISO TC 131/SC 6 ISO TC 131/SC 6/WG 1 N313
ISO TC 131/SC 6 ISO TC 131/SC 6/WG 1 N313 Secretariat: BSI ISO/CD 11171 Hydraulic fluid power Calibration of automatic particle counters for liquids CD stage Warning This document is not an ISO International
More information429 LIGHT DIFFRACTION MEASUREMENT OF PARTICLE SIZE
Search USP29 429 LIGHT DIFFRACTION MEASUREMENT OF PARTICLE SIZE Light diffraction is one of the most widely used techniques for measuring the size of a wide range of particles from very fine to very coarse.
More informationPARTICLE SIZE ANALYTICAL RANGES AND APPLICABILITY. m mm (10-6 m) nm (10-9 m)
P A R T I C L E S I Z E A N A L Y S I S PARTICLE SIZE Accurately determining particle size has become essential in many industries, as it is a fundamental physical characteristic that must be selected,
More informationKnow thy Enemy : Understand and Control Particle Contamination. March 10th, 2008 Kirk Buecher
Know thy Enemy : Understand and Control Particle Contamination March 10th, 2008 Kirk Buecher Agenda Background Hach Ultra Analytics Particles Particle Sources, types and definitions Size, behavior, movement
More informationPharmaceutical Particulate Matter
Pharmaceutical Particulate Matter Analysis and Control Thomas A. Barber Baxter Healthcare Illustrations by John G. Williams Photomicrography Section by Damian S. Neuberger Interpharm Press Buffalo Grove,
More informationThe new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch. Symposium Stockholm Mai
The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch Symposium Stockholm Mai 2014 1 Outline Optical Particle Detection Method 11-R Main features of the GRIMM 11-R Measurement principle
More informationControlled Environment Testing Report
Client: Controlled Environment Testing Report Services Conducted: Product Testing Project Contact: Stefan Hocom Project Manager: Stefan E. Hocom 323 Spreckels Drive, Ste. B Aptos, CA 95003 The final approval
More informationSTANDARD OPERATING PROCEDURE (SOP) FOR THE FIELD OPERATION OF THE AERODYNE AEROSOL MASS SPECTROMETER (AMS)
1 of 9 STANDARD OPERATING PROCEDURE (SOP) FOR THE FIELD OPERATION OF THE AERODYNE AEROSOL MASS SPECTROMETER (AMS) Prepared by : Frank Drewnick, John Jayne Reviewed by: Volker A. Mohnen (PMTACS-NY QA Officer)
More informationMaintianing Validation and Calibration. Version 01. Rev 01.
T E C H N I C A L P A P E R Maintianing Validation and Calibration Version 01. Rev 01. Validation, Calibration Assure Investigation Accuracy In 1984, I purchased a 77 charcoal gray, Datsun 280Z from a
More informationTesting for particles in injectable products
Testing for particles in injectable products 20-Nov-2018 MONITORING PHARMACEUTICALS Liquid sample technology that is compliant with USP must be part of any testing strategy into winning the war on
More informationBasic Instrumentation. Learning Objectives:
Proteomics Basic Instrumentation Basic Instrumentation Handling instruments plays an important role in laboratory working on a daily basis. The use of most instruments is almost inevitable and hence a
More informationRecent Advances and Challenges in Nanoparticle Monitoring for the Semiconductor Industry. December 12, 2013
Recent Advances and Challenges in Nanoparticle Monitoring for the Semiconductor Industry December 12, 2013 Agenda Introduction Wafer Environment Nano-Contamination Requirements State-of-the-Art Monitoring
More informationLaser Safety for LIGO
Laser Safety for LIGO Peter King 8 th Annual DOE Laser Safety Officer Workshop September 12 th, 2012 LIGO G1200825 v1 An Overview of LIGO LIGO is the Laser Interferometer Gravitational Wave Observatory
More informationPTG-NIR Powder Characterisation System (not yet released for selling to end users)
Powder Characterisation System - PTG-NIR PTG-NIR Powder Characterisation System (not yet released for selling to end users) The PTG-NIR automatic powder and granule inspection system with integrated NIR
More informationThe Removal of Nanoparticles from Nanotrenches Using Megasonics
NSF Center for Micro and Nanoscale Contamination Control The Removal of Nanoparticles from Nanotrenches Using Megasonics Pegah Karimi 1, Tae Hoon Kim 1, Ahmed A. Busnaina 1 and Jin Goo Park 2 1 NSF Center
More informationBackground Statement. This document describes a test method for particle removal performance of nm-rated liquid filters using GNP and OPC.
Background Statement for SEMI Draft Document 5297 New Standard: Test Method for Particle Removal Performance of Liquid Filter Rated 20-50 nm with Liquid-Borne Particle Counter Notice: This background statement
More informationFTIR Spectrometer. Basic Theory of Infrared Spectrometer. FTIR Spectrometer. FTIR Accessories
FTIR Spectrometer Basic Theory of Infrared Spectrometer FTIR Spectrometer FTIR Accessories What is Infrared? Infrared radiation lies between the visible and microwave portions of the electromagnetic spectrum.
More informationTechnical Note. How to make Concentration Measurements using NanoSight LM Series Instruments
Technical Note How to make Concentration Measurements using NanoSight LM Series Instruments Last updated 17/06/09 Introduction NanoSight LM series allow the user to estimate the concentration of the nanoparticles
More informationCleanrooms and associated controlled environments. Part 1: Classification of air cleanliness by particle concentration
INTERNATIONAL STANDARD ISO 14644-1 Second edition 2015-12-15 Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness by particle concentration Salles propres et environnements
More informationThe molecules that will be studied with this device will have an overall charge of
The Basics of the Rotation of Polarized Light The molecules that will be studied with this device will have an overall charge of zero but will have localized polarities that can be used to orient the molecule.
More informationDirect reading aerosol monitors
Health and Safety Executive Direct reading aerosol monitors 27 th June 2018 Andrew Thorpe Introduction Direct reading Aerosol Monitors (DRAMs) are often used as a complementary approach to conventional
More informationImproved Process Measurement & Control. OD 600 vs. AU Measurements in Fermentation Cell Density
Finesse, LLC 3350 Scott Boulevard #1 Santa Clara, CA 95054 9351 Irvine Boulevard Irvine, CA 92618 800-598-9515 www.finesse-inc.com P01 OD 600 vs. AU Measurements in Fermentation Cell Density This technical
More informationAbsorption and scattering
Absorption and scattering When a beam of radiation goes through the atmosphere, it encounters gas molecules, aerosols, cloud droplets, and ice crystals. These objects perturb the radiation field. Part
More informationAS 101: Day Lab #2 Summer Spectroscopy
Spectroscopy Goals To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are related To see spectral lines from different elements in emission and
More information9/12/2011. Training Course Remote Sensing - Basic Theory & Image Processing Methods September 2011
Training Course Remote Sensing - Basic Theory & Image Processing Methods 19 23 September 2011 Introduction to Remote Sensing Michiel Damen (September 2011) damen@itc.nl 1 Overview Electro Magnetic (EM)
More informationReview: ISO Colloidal systems Methods for zeta potential determination
Review: ISO 13099 Colloidal systems Methods for zeta potential determination Mark Bumiller mark.bumiller@horiba.com www.horiba.com/particle New ISO Standards www.iso.org Outline ISO standards Zeta potential
More informationHOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY
HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY Accurate temperature measurement is key for controlling the stability and repeatability of many temperature-critical processes.
More informationPhysics 30: Chapter 5 Exam Wave Nature of Light
Physics 30: Chapter 5 Exam Wave Nature of Light Name: Date: Mark: /33 Numeric Response. Place your answers to the numeric response questions, with units, in the blanks at the side of the page. (1 mark
More informationProduct data sheet Palas U-SMPS 1050 / 1100 / 1200
Product data sheet Palas U-SMPS 1050 / 1100 / 1200 Applications Filter test Aerosol research Environmental and climatic studies Inhalation experiments Interior and workplace measurements Benefits Particle
More informationParticle size analysis -Chapter 3
Particle size analysis -Chapter 3 Importance of PSA Size and hence surface area of particles affect: The rate of drug dissolution and release from dosage forms Flow properties of granules and powders.
More informationSolar Flat Plate Thermal Collector
Solar Flat Plate Thermal Collector INTRODUCTION: Solar heater is one of the simplest and basic technologies in the solar energy field. Collector is the heart of any solar heating system. It absorbs and
More informationDifferential Mobility Particle Sizer (Aerosol measurements)
Institute for Atmospheric and Climate Science - IACETH Atmospheric Physics Lab Work Differential Mobility Particle Sizer (Aerosol measurements) Abstract A differential mobility particle sizer (DMPS) is
More informationLecture 26. Regional radiative effects due to anthropogenic aerosols. Part 2. Haze and visibility.
Lecture 26. Regional radiative effects due to anthropogenic aerosols. Part 2. Haze and visibility. Objectives: 1. Attenuation of atmospheric radiation by particulates. 2. Haze and Visibility. Readings:
More informationUniversity of Pécs Institute of Pharmaceutical Technology and Biopharmacy
University of Pécs Institute of Pharmaceutical Technology and Biopharmacy Particle Definition In a continuous phase the particle is an (mostly in gaseous or liquid material) existing, dispersed, interface
More informationVisualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source
3rd International EUVL Symposium NOVEMBER 1-4, 2004 Miyazaki, Japan Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source H. Tanaka, A. Matsumoto, K. Akinaga, A. Takahashi
More informationFRITSCH "analysette 22" NanoTec The Result of 30 Years of Experience in Particle Size Measurement. 1 The Inventor of "Reverse Fourier Optics"
FRITSCH "analysette 22" NanoTec The Result of 30 Years of Experience in Particle Size Measurement Wolfgang Mutter FRITSCH GmbH Manufacturers of Laboratory Instruments Industriestrasse 8 55743 Idar-Oberstein
More informationTi-Pure TITANIUM DIOXIDE. DETERMINATION OF PARTICLE SIZE DISTRIBUTION OF RPS ( Horiba (LA-900) Procedure) METHOD: T WP
Ti-Pure TITANIUM DIOXIDE DETERMINATION OF PARTICLE SIZE DISTRIBUTION OF RPS ( Horiba (LA-900) Procedure) METHOD: T4400.274.03WP 1 TITANIUM DIOXIDE (Ti-Pure ) Determination of Particle Size Distribution
More informationPerformance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3
Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3 1 CANBERRA Industries, Meriden, CT 06450 2 Los Alamos National Laboratory,
More informationCompact Hydrogen Peroxide Sensor for Sterilization Cycle Monitoring
Physical Sciences Inc. VG15-012 Compact Hydrogen Peroxide Sensor for Sterilization Cycle Monitoring January 26, 2015 Krishnan R. Parameswaran, Clinton J. Smith, Kristin L. Galbally-Kinney, William J. Kessler
More informationMULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.
Exam Name MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) The current definition of the standard meter of length is based on 1) A) the length of
More informationThe Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)
The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of
More informationand Environmental Science Centre
1. Purpose The purpose of this document is to familiarize the user with the mode of function of the FT-Raman available at the facility, and to describe the sampling procedure. 2. Introduction Raman spectroscopy
More informationAIR-O-CELL CSI. Bioaerosol Sampling Cassette For the collection of viable and non-viable airborne particles. Laboratory & User Manual
AIR-O-CELL CSI Bioaerosol Sampling Cassette For the collection of viable and non-viable airborne particles Laboratory & User Manual Air-O-Cell CSI Sampling Cassette The Air-O-Cell Collector for SEM Identification
More informationPlastic Coated Silica/Silica (Low OH) FIBER CROSS SECTION Polyimide and Acrylate Coated. Nylon and Tefzel Coated
DESCRIPTION When looking for a high quality fiber with superior transmission and a numerical aperture (N.A.) of 0.22 for efficient light coupling, the is the fiber of choice. The Anhydroguide fiber is
More informationFEDERAL STANDARD 209E. Airborne Particulate Cleanliness Classes in Cleanrooms and Clean Zones
FEDERAL STANDARD 209E Airborne Particulate Cleanliness Classes in Cleanrooms and Clean Zones Revised 1992 by the Institute of Environmental Sciences 940 E. Northwest Highway Mount Prospect, Illinois 60056
More informationAPPLICATION INFORMATION
A-1995A APPLICATION INFORMATION Particle Characterization THE SIZING OF NON-SPHERICAL,SUB-MICRON PARTICLES USING POLARIZATION INTENSITY DIFFERENTIAL SCATTERING (PIDS ) Introduction Nearly all laser-based
More informationApplied Fluid Mechanics
Applied Fluid Mechanics 1. The Nature of Fluid and the Study of Fluid Mechanics 2. Viscosity of Fluid 3. Pressure Measurement 4. Forces Due to Static Fluid 5. Buoyancy and Stability 6. Flow of Fluid and
More informationLecture # 04 January 27, 2010, Wednesday Energy & Radiation
Lecture # 04 January 27, 2010, Wednesday Energy & Radiation Kinds of energy Energy transfer mechanisms Radiation: electromagnetic spectrum, properties & principles Solar constant Atmospheric influence
More informationIs NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports?
Is NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports? A. Geach 1 1 CINRG Systems Inc., Burlington, Ontario, Canada Abstract In 2015 NIST released
More information4) If a woman weighs 125 lb, her mass expressed in kilograms is x kg, where x is. A) less than 15. B) greater than 15. Answer: A
Topic 1 Units, Conversions, and Estimation 1.1 Conceptual Questions 1) The current definition of the standard meter of length is based on A) the distance between the earthʹs equator and north pole. B)
More informationZeta Potential Analysis using Z-NTA
Zeta Potential Analysis using Z-NTA Summary Zeta Potential Nanoparticle Tracking Analysis (Z-NTA) adds measurements of electrostatic potential to simultaneous reporting of nanoparticle size, light scattering
More informationControl of Particulate Matter Contamination in Healthcare Manufacturing
Control of Particulate Matter Contamination in Healthcare Manufacturing Thomas A. Barber Interpharm Press Denver, Colorado IMHBIHHW PRESS Contents iii CONTENTS PREFACE Acknowledgments xiii xv 1. INTRODUCTION
More informationStandardization of Optical Particle Counters
Standardization of Optical Particle Counters Dean Ripple Bioprocess Measurements Group NIST, Gaithersburg WCBP, January 25, 2012 Protein Particulates in Biotherapeutics Proteins in solution partially denature
More informationATM 10. Severe and Unusual Weather. Prof. Richard Grotjahn.
ATM 10 Severe and Unusual Weather Prof. Richard Grotjahn http://atm.ucdavis.edu/~grotjahn/course/atm10/index.html Lecture topics: Optics: Scattering Sky Colors and Rays Optics: refraction Mirages and Refraction
More informationPoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo
ICEPP, the University of Tokyo E-mail: sawada@icepp.s.u-tokyo.ac.jp The MEG experiment yielded the most stringent upper limit on the branching ratio of the flavorviolating muon decay µ + e + γ. A major
More informationTHE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS
THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS INTRODUCTION In recent years, researchers involved in many unrelated scientific disciplines have acquired an interest in accurately
More informationElectroMagnetic Radiation (EMR) Lecture 2-3 August 29 and 31, 2005
ElectroMagnetic Radiation (EMR) Lecture 2-3 August 29 and 31, 2005 Jensen, Jensen, Ways of of Energy Transfer Energy is is the the ability to to do do work. In In the the process of of doing work, energy
More informationDAY LABORATORY EXERCISE: SPECTROSCOPY
AS101 - Day Laboratory: Spectroscopy Page 1 DAY LABORATORY EXERCISE: SPECTROSCOPY Goals: To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are
More informationENVR 116 Aerosol Technology Laboratory Session Fall 2005
ENVR 116 Aerosol Technology Laboratory Session Fall 2005 INSTRUMENT CALIBRATION USING PSL SPHERES The objective of this lab is to become familiar with the operation, characteristics, and limitations of
More informationAirborne particle deposition in cleanrooms: Calculation of product contamination and required cleanroom class
Airborne particle deposition in cleanrooms: Calculation of product contamination and required cleanroom class W Whyte 1, K Agricola 2 and M Derks 3 1 School of Engineering, University of Glasgow, UK; 2
More informationThe Elzone II Particle Count and Size Analyzer
The Elzone II 5390 Particle Count and Size Analyzer Elzone II Overview of Basic Theory Operates using the Electrical Sensing Zone (ESZ) Principle, also known as the Coulter Principle. First commercially
More informationA new versatile condensation particle counter for research and environmental monitoring
A new versatile condensation particle counter for research and environmental monitoring, Maximilian Weiss Palas GmbH, Greschbachstr. 3b, 76229 Karlsruhe, Germany, http://www.palas.de, spielvogel@palas.de
More informationSensitive Detection and Identification of Isovanillin Aerosol Particles at the pg/cm 3 Mass Concentration Level Using Raman Spectroscopy*
Sensitive Detection and Identification of Isovanillin Aerosol Particles at the pg/cm 3 Mass Concentration Level Using Raman Spectroscopy* R. L. Aggarwal 1, S. Di Cecca, L. W. Farrar, Shabshelowitz, A.,
More informationChapter 24 Photonics Question 1 Question 2 Question 3 Question 4 Question 5
Chapter 24 Photonics Data throughout this chapter: e = 1.6 10 19 C; h = 6.63 10 34 Js (or 4.14 10 15 ev s); m e = 9.1 10 31 kg; c = 3.0 10 8 m s 1 Question 1 Visible light has a range of photons with wavelengths
More informationOcean Optics XIV Conference, Kona, Hawaii 1998
INTRODUCTION Ocean Optics XIV Conference, Kona, Hawaii 1998 COASTAL AEROSOL PHASE FUNCTION MEASUREMENTS WITH A CUSTOM POLAR NEPHELOMETER By John N. Porter, Tom F. Cooney, Craig Motell University of Hawaii
More informationElectromagnetic Waves
ELECTROMAGNETIC RADIATION AND THE ELECTROMAGNETIC SPECTRUM Electromagnetic Radiation (EMR) THE ELECTROMAGNETIC SPECTRUM Electromagnetic Waves A wave is characterized by: Wavelength (λ - lambda) is the
More informationITS. One of the highest resolutions for particle sizing and counting. Multisizer 3 COULTER COUNTER SCIENCE
ITS SCIENCE ITS SCIENCE Sp. z o.o. Sp. K. Poświętne 25B, 26-670 Pionki +48 48 612 71 65 www.itsscience.pl One of the highest resolutions for particle sizing and counting. Multisizer 3 COULTER COUNTER Blood
More informationby Ray Garvey, Mong Ching Lin, and John Mountain Computational Systems Inc. 835 Innovation Drive, Knoxville, TN (423)
THE MISAPPLICATION OF "COMPOSITE CORRELATION OF CLEANLINESS LEVELS" by Ray Garvey, Mong Ching Lin, and John Mountain Computational Systems Inc. 835 Innovation Drive, Knoxville, TN 37932 (423) 675-2120
More informationMaking Accurate Field Spectral Reflectance Measurements By Dr. Alexander F. H. Goetz, Co-founder ASD Inc., Boulder, Colorado, 80301, USA October 2012
Making Accurate Field Spectral Reflectance Measurements By Dr. Alexander F. H. Goetz, Co-founder ASD Inc., Boulder, Colorado, 80301, USA October 2012 Introduction Accurate field spectral reflectance measurements
More informationThe solution for all of your
The solution for all of your nanoparticle sizing and zeta potential needs. DelsaNano Series Blood Banking Capillary Electrophoresis Cell Analysis Centrifugation Genomics Lab Automation Lab Tools Particle
More informationJRE Group of Institutions ASSIGNMENT # 1 Special Theory of Relativity
ASSIGNMENT # 1 Special Theory of Relativity 1. What was the objective of conducting the Michelson-Morley experiment? Describe the experiment. How is the negative result of the experiment interpreted? 2.
More informationApplied Fluid Mechanics
Applied Fluid Mechanics 1. The Nature of Fluid and the Study of Fluid Mechanics 2. Viscosity of Fluid 3. Pressure Measurement 4. Forces Due to Static Fluid 5. Buoyancy and Stability 6. Flow of Fluid and
More informationNTUA. A. Georgakopoulou. A. Papayannis1, A. Aravantinos2 NATIONAL TECHNICAL UNIVERSITY OF ATHENS TECHNOLOGICAL EDUCATIONAL INSTIDUTION OF ATHENS SIENA
High Spectral Resolution LIDAR Receivers, to Measure Aerosol to Molecular Scattering Ratio in Bistatic Mode, for use in Atmospheric Monitoring for EAS Detectors E. Fokitis1, P. Fetfatzis1, 1, S. Maltezos1
More informationANALYSIS OF LOW DENSITY PARTICLES USING DIFFERENTIAL CENTRIFUGAL SEDIMENTATION
ANALYSIS OF LOW DENSITY PARTICLES USING DIFFERENTIAL CENTRIFUGAL SEDIMENTATION Conventional Centrifugal Methods Centrifugal sedimentation of particles suspended in a fluid is a well known method (1, 2)
More informationLaser Beam Interactions with Solids In absorbing materials photons deposit energy hc λ. h λ. p =
Laser Beam Interactions with Solids In absorbing materials photons deposit energy E = hv = hc λ where h = Plank's constant = 6.63 x 10-34 J s c = speed of light Also photons also transfer momentum p p
More informationCHEM 254 EXPERIMENT 7. Phase Diagrams - Liquid Vapour Equilibrium for two component solutions
pressure CHEM 254 EXPERIMENT 7 Phase Diagrams - Liquid Vapour Equilibrium for two component solutions The partial pressures of the components of an ideal solution of two volatile liquids are related to
More informationNanophysics: Main trends
Nano-opto-electronics Nanophysics: Main trends Nanomechanics Main issues Light interaction with small structures Molecules Nanoparticles (semiconductor and metallic) Microparticles Photonic crystals Nanoplasmonics
More informationLIQUID LIGTH GUIDE. Una Pale
LIQUID LIGTH GUIDE Una Pale XV gymnasium, Faculty of electrical engineering and computing, University of Zagreb, Croatia 1. Introduction Light guide is a fiber-optic cable which is used to transfer signal
More informationSECTION 8 Part I Typical Questions
SECTION 8 Part I Typical Questions 1. For a narrow beam of photons, the relaxation length is that thickness of absorber that will result in a reduction of in the initial beam intensity. 1. 1/10. 2. 1/2.
More informationUV-Vis optical fiber assisted spectroscopy in thin films and solutions
UV-Vis optical fiber assisted spectroscopy in thin films and solutions Description UV-Visible absorption and transmission spectra provide fundamental information for all experiments related to the attenuation
More informationParticles, drops, and bubbles. Lecture 3
Particles, drops, and bubbles Lecture 3 Brownian Motion is diffusion The Einstein relation between particle size and its diffusion coefficient is: D = kt 6πηa However gravitational sedimentation tends
More information